m550 operating manual

m550 operating manual
Model 550 OPERATING MANUAL
Product Overview
The Model 550 is a linear amplifier designed to drive a number of different Conoptics ElectroOptic Modulators. The amplifier is AC coupled and is designed to accept bipolar input signal
formats.
ConOptics, Inc. 19 Eagle Road, Danbury, CT 06810 Phone 800-748-3349 Fax 203-790-6145
Email: [email protected], or visit www.conoptics.com
CAUTION: BEFORE APPLYING POWER
Please review this manual and become familiar with all safety markings and instructions.
Insure that the equipment mains (line) voltage as specified on the rear panel is in accordance
with desired AC source. Safe operation of this equipment may be jeopardized by use other
than specified by Conoptics.
INTENDED USE
This equipment is designed for general laboratory use. It is intended to be used with a number
of Conoptics Electro-Optic Modulators only. The Model 550 amplifier is designed to drive a 50
ohm, Single Ended Line. Any use other than in direct connection with supplied optical
modulators can result in severe damage to the amplifier. This amplifier must only be used in
accordance with the instructions provided in the manual.
WARNING: proper connections
The Model 550 amplifier is designed to work into a 50 ohm load only. Running the amplifier
open circuit or into high VSWR loads will damage the amplifier
SAFTEY GROUND
ConOptics, Inc. 19 Eagle Road, Danbury, CT 06810 Phone 800-748-3349 Fax 203-790-6145 Email:
[email protected], or visit www.conoptics.com
2
This equipment is equipped with a protective earth terminal. This terminal is provided by a
common or ground wire in the power cord. The main power source must supply an
uninterrupted return to an earth ground. The equipment must not be used if this earth return is
not supplied.
SERVICE CAUTION
Only qualified personnel must perform adjustment, maintenance, or repair of the equipment.
Hazardous voltages are present within the instrument and contact may result in personal injury.
Any attempt to trouble shoot or adjust any controls within the system should be discussed with
Conoptics first at [email protected]
Model 550 SYSTEM SPECIFICATIONS
Cabinet
Input Impedance
Driver and power supply in single cabinet
50 ohms
ConOptics, Inc. 19 Eagle Road, Danbury, CT 06810 Phone 800-748-3349 Fax 203-790-6145 Email:
[email protected], or visit www.conoptics.com
3
Gain
Max. Output Drive Level
Detected Rise/Fall Time
Input Signal Format
Input Power (AC)
Dimensions
Cooling
Operating Environment
Weight
45 db
125 v, p-p into 50 ohms
Determined by modulator
Ac coupled bipolar signal, +10dbm max input level
Typical, 360w
19” Rack Mountable,
Forced air
Designed for laboratory use (indoor only)
Temperature range +5deg C to +50deg C ambient
Humidity 20%-80% RH up to 32deg C, Altitude <3000M
20lbs (9.07kg)
ConOptics, Inc. 19 Eagle Road, Danbury, CT 06810 Phone 800-748-3349 Fax 203-790-6145 Email:
[email protected], or visit www.conoptics.com
4
Connection Diagram
ConOptics, Inc. 19 Eagle Road, Danbury, CT 06810 Phone 800-748-3349 Fax 203-790-6145 Email:
[email protected], or visit www.conoptics.com
5
Model 550 Electronics
ConOptics, Inc. 19 Eagle Road, Danbury, CT 06810 Phone 800-748-3349 Fax 203-790-6145 Email:
[email protected], or visit www.conoptics.com