JUMO TDA-300 and JUMO TDA-3000 Handheld Thermometer with

JUMO TDA-300 and JUMO TDA-3000 Handheld Thermometer with
JUMO GmbH & Co. KG
Delivery address: Mackenrodtstraße 14
36039 Fulda, Germany
Postal address: 36035 Fulda, Germany
Phone:
+49 661 6003-0
Fax:
+49 661 6003-607
E-mail:
[email protected]
Internet:
www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow, Essex CM20 2DY, UK
Phone: +44 1279 635533
Fax:
+44 1279 635262
E-mail: [email protected]
Internet: www.jumo.co.uk
JUMO Process Control, Inc.
6733 Myers Road
East Syracuse, NY 13057, USA
Phone: 315-437-5866
1-800-554-5866
Fax:
315-437-5860
E-mail: [email protected]
Internet: www.jumousa.com
Data Sheet 702540
Page 1/6
JUMO TDA-300 and JUMO TDA-3000
Handheld Thermometer
with data logger
Brief description
TDA-300 and TDA-3000 are handheld digital temperature indicators. They are used
together with precision thermocouples or RTDs to measure temperatures on surfaces, in
liquids, in melts, and in soft materials.
Interchangeable temperature probes allow rapid adaptation to different applications.
Both instruments feature a data logging function. On Type TDA-3000, the recorded
measurement data can be read out via an USB interface. No special PC program is
necessary for visualization, since the measurement data are available as ASCII files in
CSV format (evaluation via spreadsheet programs).
The plastic housing is proof against shock and breakage and resistant to corrosive
substances. Thermometers, temperature probes and accessories are all available for
delivery from stock.
Displays
Light is on during active
data recording
Number of recorded
measurement data
Probe break or
short-circuit
Measurement /
measurement point/
user number
Shows the range
within which the
measurement lies
(over/underrange)
Battery charge status
Measured
temperature
Message window
(date, time, ...)
TDA-300 (Type 702540/...)
TDA-3000 (Type 702541/...)
Comparison of models
TDA-300
TDA-3000
Data logger
99 measurements
9999 measurements
TAG numbers
(meas. point designation)
5 (11 characters each)
99 (11 characters each)
User
1
99 (11 characters each)
Interfaces
-
USB, type Mini-B
Enclosure protection
IP67
IP54
Key features
k Measurement input for Pt100,
NiCr-Ni K, Fe-Con J
and Cu-Con T
k DKD calibration certificate
k Data logger for 9999 measurements
k Simple readout via USB and
evaluation of data from a PC
(TDA-3000 only)
k Limit monitoring
k Peak/bottom value acquisition
k Battery AA, Mignon LR6,
with a long service life
2013-12-03/00447944
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Data Sheet 702540
Page 2/6
Technical data
Input
Measurement input
Range limits
- resolution 1°C
- resolution 0.1°C
RTD
Pt100
to EN 60751
Thermocouple
NiCr-Ni K
to EN 60584
Thermocouple
Fe-Con J
to EN 60584
Thermocouple
Cu-Con T
to EN 60584
-200 to +850°C
-199.9 to +850.0°C
-200 to +1372°C
-199.9 to +999.9°C
-200 to +1200°C
-199.9 to +999.9°C
-50 to +400°C
-50.0 to +400.0°C
Measurement offset
-99.9 to +99.9°C
Sampling rate
Input filter
0.5 seconds
1st order digital filter; filter constant adjustable from 0 — 100 seconds
Unit
°C or °F
Accuracy
Measurement accuracy of display
at 23°C ambient temperature
Accuracy of cold junction
(thermocouple only)
Accuracy class of
temperature probes
±(0.1% + 1 digit) or ±0.3°C; the larger value applies
-
±0.5°C for 5 to 40°C
±1°C for -20 to +5°C
and for 40 to 50°C
±0.5°C for 5 to 40°C
±1°C for -20 to +5°C
and for 40 to 50°C
±0.5°C for 5 to 40°C
±1°C for -20 to +5°C
and for 40 to 50°C
Class A
Class 2
Class 2
Class 2
Measuring circuit monitoring
Symbol in display
Probe short-circuit,
probe/cable break,
wrong connection
Electrical data
Supply
Battery service life
1 alkaline battery, type LR6 AA
400 hrs continuous operation at 23°C ambient temperature
Power consumption
10mW (average value)
Environmental influences
Device type
TDA-300
Operating temperature range
Temperature error
Climatic conditions
± 0.01% of measuring span for 5 to 40°C ambient temperature
± 0.02% of measuring span for -20 to +5°C and 40 to 50°C ambient temperature
rel. humidity 95% annual average, no condensation
EN 61326-1
Class B
to general requirements
Electromagnetic compatibility (EMC)
- interference emission
- immunity to interference
IP enclosure protection
TDA-3000
-20 to +50°C
IP67
IP54
Housing
Dimensions (W x H x D)
Weight
57 x 152 x 46mm
approx. 150g including battery
Display screen
Screen type
FSTN LCD
Display of measured temperature
through 4 digits
Display of memory number
through 4 digits
Display of total number of
measured temperatures
through 4 digits
Messages (date, time, ...)
11 characters (68 x 8 pixels)
Data logging info
Probe break/short-circuit
2013-12-03/00447944
through symbol; light is on during data recording,
flashes when ready for automatic data recording
through symbol;
light is on when the sensor is not connected or
in the event of a break or short-circuit.
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Data Sheet 702540
Battery charge status
Page 3/6
3-stage display
Temperature unit
°C or °F (12 x 8 pixels)
Data logging function
Device type
TDA-300
TDA-3000
Recording type
Recording interval
Contents of data record
manual or automatic
any (manually), 1 — 3600 seconds (automatically)
temperature, TAG No. (meas. point designation), user,
limits, and date/time
Recording capacity
99 measurements
9999 measurements
Data storage
in SRAM
(volatile memory)
in EEPROM
(non-volatile memory)
Loss of data when battery is discharged
or has to be replaced
Data are retained for about 10 years,
memory can be rewritten about 100,000
times.
Limit monitoring
Limit monitoring
upper and lower limit can be set for each measurement point
Temperature within the limits
Temperature above or below
the limit
Measurement points
Device type
Number of measurement points
TAG number
(meas. point designation)
TDA-300
TDA-3000
5
99
consisting of up to 11 characters (digits, letters, and symbols)
User
Device type
TDA-300
TDA-3000
Number of users
1
99
User names
-
consisting of up to 11 characters
(digits, letters, and symbols)
USB interface (Universal Serial Bus) - only for TDA-3000
Speed
USB 2.0 (theoretically 12 Mbps max.)
Connection
Mini-B connector
Connecting cable
included in delivery
Supply
PC operating system
via PC
®
for Windows 2000, XP, Vista, 7 (32-bit and 64-bit)
Additional functions
Functions
2013-12-03/00447944
Peak/bottom value storage,
real-time clock (date and time),
function locking and
self-diagnostics
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Data Sheet 702540
Page 4/6
Temperature probes and adapters
RTD Pt100 with handle and attached connecting cable
Type
Construction
Description
Immersion probe
The immersion probe with handle is
particularly
suitable
for
measuring
temperatures in liquids.
The temperature sensor is located in the
probe tip, embedded in a heat transfer
compound. The handle, which is provided
with an anti-kink spring, is made from
temperature-resistant plastic.
Max. measurement temperature: 250°C
Max. temperature of handle:
100°C
Max. temperature of cable:
180°C
702546/01-100
Insertion probe
Ø4
702546/02-100
100
Thanks to its measurement tip, this probe is
particularly suitable for measuring the core
temperature in food and other soft materials.
The silicone handle is covered by a protective
sleeve and is resistant to corrosive media
such as oil and fatty acids.
Max. measurement temperature: 250°C
Max. temperature of handle:
180°C
Max. temperature of cable:
180°C
Thermocouples NiCr-Ni K with handle and attached compensating cable
Construction
Ø 1,5
Type
Flexible immersion probe
(mineral-insulated
thermocouple)
Description
200 (100)
The immersion probe tip is suitable for
measuring temperatures in liquids.
Max. measurement temperature: 1150°C
Max. temperature of handle:
100°C
Max. temperature of cable:
180°C
100 (150)
Thanks to its measurement tip, this probe is
particularly suitable for measuring the core
temperature in food and other soft materials.
The silicone handle is covered by a protective
sleeve and is resistant to corrosive media
such as oil and fatty acids.
Max. measurement temperature: 250°C
Max. temperature of handle:
180°C
Max. temperature of cable:
180°C
702545/01-...
702545/02-...
Surface probe
702545/03-004
Surface probe
702545/03-015
Ø4
Insertion probe
The surface probe is particularly suitable for
measurement on very small and uneven
objects
with
poor
heat
conduction,
for instance electronic components, glass,
ceramic.
The thermocouple is mounted on a spring
plate so that the probe can also be applied
obliquely to the surface.
Max. measurement temperature: 400°C
Max. temperature of handle:
100°C
Max. temperature of cable:
180°C
This surface probe enables highly accurate
and easily reproducible measurements on flat
surfaces.
Since the spring bands of the probe are
transversely linked, the measurements are
largely independent of the applied pressure
and the contact angle.
Max. measurement temperature: 500°C
Max. temperature of handle:
110°C
Max. temperature of cable:
180°C
The cable length for all probes is about 1500mm. The insertion probes are protected to IP67.
Probes with type J and T thermocouples on request.
2013-12-03/00447944
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Data Sheet 702540
Page 5/6
Adapters for existing probes
Type
Adapter for
RTD
Pt100
Diagram
Description
1700
Existing RTDs can be connected up via a
connector-coupling combination
(type Mini-Flach) in copper.
702546/04-000
Adapter for
thermocouple
type K
702545/04-000
Dimensions
Carrying case
2013-12-03/00447944
The adapter is 1700mm long and can be used
in ambient temperatures up to 100°C.
1700
The adapter is 1700mm long and can be used
in ambient temperatures up to 100°C.
Existing thermocouples can be connected up
via a thermal connector-coupling
(type Mini-Flach).
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Data Sheet 702540
Order details: Handheld thermometer with data logger
(1) Basic version
702540/
TDA-300, Type 702540/88-000
702541/
TDA-3000, Type 702541/88-000
(1)
Order code
Order example
702541
/
(2)
88
-
(3)
000
/
88
-
000
Order details: Temperature probe and adapter
(1) Basic version
702545/
Thermocouple K
702546/
Pt100 RTD
x x
x x
x
x x
01
02
03
04
x x
x
x
x x
x
x
000
004
015
100
150
200
(2) Probe type
immersion probe
insertion probe
surface probe
adapter with 1500mm cable, for existing probes
(3) probe diameter/length
with probe type 04 only
4 mm dia. (probe type 03 only)
15 mm dia. (probe type 03 only)
length: 100 mm (probe types 01 and 02 only)
length: 150 mm (probe type 02 only)
length: 200 mm (probe type 01 only)
(1)
Order code
Order example
(2)
/
702545
/
01
Standard accessories
- 1 Operating Manual
- 1 battery
- 1 carrying hook
- 1 USB connecting cable (TDA-3000 only)
Accessories
- Silicone heat transfer compound (20 ml tube, for
temperature measurements up to 200°C,
part no. 94091460
- Carrying case for thermometer, two probes,
heat transfer compound and accessories,
part no. 00453912
2013-12-03/00447944
(3)
-
200
Page 6/6
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project