- No category
advertisement
Agilent Technologies
11752D/E Option K01 Connector Gage Kit
User’s Guide
Manufacturing Part Number: 11752-90008
Print Date: September 2008
© Copyright Agilent Technologies, Inc. 2000, 2008 i
Warranty Statement
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT
TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE
MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL
WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL
OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR
PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN.
SHOULD AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH
WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT
WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL
CONTROL.
DFARS/Restricted Rights Notice
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of
Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than
Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or
DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
ii
Safety Notes
The following safety notes are used throughout this document. Familiarize yourself with each of these notes and its meaning before performing any of the procedures in this document.
WARNING Warning denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and met.
CAUTION Caution denotes a hazard. It calls attention to a procedure that, if not correctly performed or adhered to, could result in damage to or destruction of the instrument. Do not proceed beyond a caution sign until the indicated conditions are fully understood and met.
Definitions
• Specifications describe the performance of parameters covered by the product warranty
(temperature –0 to 55
°C, unless otherwise noted.)
• Typical describes additional product performance information that is not covered by the product warranty. It is performance beyond specification that 80% of the units exhibit with a 95% confidence level over the temperature range 20 to 30
°C. Typical performance does not include measurement uncertainty.
• Nominal values indicate expected performance or describe product performance that is useful in the application of the product, but is not covered by the product warranty.
• Characteristic Performance describes performance parameter that the product is expected to meet before it leaves the factory, but is not verified in the field and is not covered by the product warranty. A characteristic includes the same guard bands as a specification.
iii
iv
Contents
Contents-1
Contents
Contents-2
11752D/E Option K01
User’s Guide 1
11752D/E Option K01 Introduction
Introduction
This document contains information about initial inspection, performance tests, adjustments and operation of the Agilent 11752DK01 and 11752EK01 Connector Gages.
Description
The Agilent 11752D/EK01 connector gages are used to measure connector pin depth of
2.4 mm or 3.5 mm connectors. The connector gage kit contains the following.
• male connector gage
• female connector gage
• male master gage
• female master gage
• protective caps
• storage box (foam sets)
Verifying the Shipment
After the test set has been unpacked, inspect the test set and all accessories for any signs of damage that may have occurred during shipment. If your test set or any accessories appear to be damaged or missing, refer to
“Contacting Agilent” on page 11 . Use the
original or comparable packaging materials to transport the instrument.
If the gage kit is being returned to Agilent for calibration or service, attach a tag indicating the type of service required, return address, model number, and serial number. Also, mark the container FRAGILE to assure careful handling. In any correspondence, refer to the
instrument by model number and serial number. Refer to “Shipping Your Analyzer to
Agilent for Service or Repair” on page 11
.
2 User’s Guide
11752D/E Option K01 General Specifications
General Specifications
The specifications listed in note below are the performance standards or limits against which the connector gage kit is tested.
NOTE Gage resolution depends on your ability to discern the position of the gage needle between divisions. The specification for this characteristic is based on a statistically significant number of people making the same measurement and independently indicating their readings.
Characteristics and
Conditions
Measurement Range
Typical Limits Comments
Gage Resolution
Gage Calibration Accuracy
Gage Repeatability
Master Accuracy
Total Uncertainty:
Worst Case
RSS
–0.005 in (–0.127 mm) to
+0.005 in (+0.127 mm
0.00002 in (0.00051 mm)
0.00004 in (0.00102 mm)
0.000025 in (0.00064 mm)
0.00004 in (0.00102 mm)
0.000125 in (0.00318 mm)
0.000062 in (0.00157 mm)
In 0.0001 inch (0.00254 mm) increments.
Fifth of an increment
Quarter of an increment
Sum of resolution, repeatability, gage and master accuracy limits.
Root sum of the squares.
Operating Environment:
Temperature Range
Humidity (26 °C max dry bulb) Maximum Altitude
Storage and Shipping
Environment:
Temperature
Relative Humidity
Altitude
Net Weight
20 to 26 °C (+68 to +79 °F)
0 to 80%
4.5 km (15,000 ft)
–55 to +75 °C less than 95% at 40 °C less than 15.3 km (50,000 ft)
1 kg (2 lb 3 oz)
No temperature fluctuations greater than 1 °C.
User’s Guide 3
11752D/E Option K01 General Specifications
Operating Precautions
Your fingers are a heat source; handle the devices as little as possible during calibration.
Before using the connector gages, they must be inspected, cleaned, and zeroed.
CAUTION To protect the sensitive microcircuits that may be connected to the connectors you are calibrating, always wear an electrostatic discharge-grounding strap connected to a conductive bench mat when working near sensitive equipment.
CAUTION Do not use hydrocarbons such as acetone, trichlorethylene, carbon
CAUTION Do not spray any liquid solvent directly onto connector or gage surfaces.
Connector wear eventually degrades performance. To prolong the life of your gages:
• Clean all surfaces that come in contact during calibration.
• Do not measure a damaged connector; doing so could damage the gage. Replace damaged connectors.
• Keep the protector caps on the gages and masters when not in use to prevent contamination.
• Turn the connector nut (not the gage itself) when making connection to avoid damage to the connector and gage.
Inspecting and Cleaning the Gages
Before using the connector gages, they must be inspected, cleaned and zeroed.
Using an illuminated magnifying glass, carefully inspect the connector gage and the gage master as well as the connector you are about to measure. Foreign material on the gage, gage master or connector will render the measurement inaccurate.
To clean connector surfaces we recommend using compressed air, directing the air where it is needed through a plastic (do not use metal) nozzle. If the compressed air does not remove the foreign material, use a cotton swab (or if the swab is too large a round wooden toothpick wrapped in a single layer of lint-free cleaning cloth soaked in isopropyl alcohol.
Dry the connector and gage surfaces with a brief blast of the compressed air.
4 User’s Guide
11752D/E Option K01 General Specifications
Zeroing the Gage
Connect the gage to the appropriate gage master and tighten only the connecting nut (do not turn gage or master) until finger tight. Then the connecting nut to 90 N-cm (8 lb-in).
Turn the zero set knob until the indicator is lined up with the zero indicator on the dial.
Gages should be checked often, to make sure that the zero setting has not changed.
Generally, when the gage pointer, on a gage that has been recently zeroed, does not line up exactly with the zero mark, the gage or calibration block needs cleaning.
Figure 1 Connector Gage
User’s Guide 5
11752D/E Option K01 General Specifications
Measuring Connectors
Measuring the recession of the center conductor, behind the outer conductor mating plane in a connector, is done in exactly the same way as zeroing the gage, except of course that the zero set knob is not re-set when the measurement is made.
Hold the gage by the barrel only and tighten only the connecting nut until finger tight. Do not turn the gage or the device. Recession or protrusion will show as a reading either counterclockwise or clockwise from zero. Refer to
NOTE Protrusion of center conductor shoulder (male pin and end of the female center pin) is not allowable on any connector.
For maximum accuracy, measure the connector several times and take an average of the readings.
Service
Do not attempt to repair the gage or gage master. Any attempt to do so will void your warranty.
It is recommended that the gages be calibrated annually at an Agilent Technologies service office. Refer to
“Agilent Support, Services, and Assistance” on page 11 .
6 User’s Guide
11752D/E Option K01
Replaceable Parts
Figure 2 11752D-K01 Connector Gage and Accessories
Replaceable Parts
User’s Guide 7
11752D/E Option K01
Figure 3 11752E-K01 Connector Gage and Accessories
Replaceable Parts
11752-60100
8 User’s Guide
11752D/E Option K01 Safety and Regulatory Information
Safety and Regulatory Information
Declaration of Conformity
For a copy of the manufacturer’s Declaration of Conformity for this apparatus, contact your
local Agilent Technologies office or sales representative. Refer to “Contacting Agilent” on page 11
.
Statement of Compliance
This product has been designed and tested in accordance with the standards listed on the
Manufacturer’s Declaration of Conformity, and has been supplied in a safe condition. The documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
General Safety Considerations
Before Applying Power
Verify that the product is configured to match the available main power source. If this product is to be powered by autotransformer, make sure the common terminal is connected to the neutral (grounded) side of the ac power supply.
User’s Guide 9
11752D/E Option K01 Safety and Regulatory Information
Regulatory Information
This section contains information that is required by various government regulatory agencies.
Instrument Markings
The instruction documentation symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the documentation.
This symbol indicates that the instrument requires alternating current (ac) input.
This symbol indicates separate collection for electrical and electronic equipment, mandated under EU law as of August 13, 2005. All electric and electronic equipment are required to be separated from normal waste for disposal (Reference WEEE
Directive, 2002/96/EC).
This symbol indicates that the power line switch is ON.
This symbol indicates that the power line switch is in the STANDBY position.
This symbol indicates that the power line switch is in the OFF position.
This symbol is used to identify a terminal which is internally connected to the product frame or chassis.
ICES/NMB-001
The CE mark is a registered trademark of the European Community. (If accompanied by a year, it is when the design was proven.)
The CSA mark is a registered trademark of the Canadian Standards Association.
This instrument complies with Canada: CSA 22.2 No. 000000061010-1, Second
Edition.
This is a symbol of an Industrial Scientific and Medical Group 1 Class A product.
This is a marking to indicate product compliance with the Canadian
Interference-Causing Equipment Standard (ICES-001).
Direct Current.
This is a required mark signifying compliance with an EMC requirement. The C-Tick mark is a registered trademark of the Australian Spectrum Management Agency.
China RoHS regulations include requirements related to packaging, and require compliance to China standard GB18455-2001.
This symbol indicates compliance with the China RoHS regulations for paper/fiberboard packaging.
10 User’s Guide
11752D/E Option K01 Agilent Support, Services, and Assistance
Agilent Support, Services, and Assistance
Service and Support Options
The analyzer’s standard warranty is a one-year return to Agilent Technologies service warranty.
NOTE There are many other repair and calibration options available from the
Agilent Technologies support organization. These options cover a range of service agreements with varying response times. Contact Agilent for additional information on available service agreements for this product.
Refer to “Contacting Agilent” on page 11
.
Contacting Agilent
Assistance with test and measurements needs and information or finding a local Agilent office are available on the Web at: http://www.agilent.com/find/assist
If you do not have access to the Internet, please contact your Agilent field engineer.
NOTE In any correspondence or telephone conversation, refer to the Agilent product by its model number and full serial number. With this information, the
Agilent representative can determine the warranty status of your unit.
Shipping Your Analyzer to Agilent for Service or Repair
IMPORTANT Agilent Technologies reserves the right to reformat or replace the internal hard disk drive in your analyzer as part of its repair. This will erase all user information stored on the hard disk. It is imperative, therefore, that you make a backup copy of your critical test data located on the analyzer’s hard disk before shipping it to Agilent for repair.
If you wish to send your network analyzer to Agilent Technologies for service or repair:
• Include a complete description of the service requested or of the failure and a description of any failed test and any error message.
• Ship the analyzer using the original or comparable antistatic packaging materials.
• Contact Agilent for instructions on where to ship your analyzer.
User’s Guide 11
11752D/E Option K01 Agilent Support, Services, and Assistance
12 User’s Guide
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 8 Introduction
- 8 Description
- 8 Verifying the Shipment
- 9 General Specifications
- 10 Operating Precautions
- 10 Inspecting and Cleaning the Gages
- 11 Zeroing the Gage
- 12 Measuring Connectors
- 12 Service
- 13 Replaceable Parts
- 15 Safety and Regulatory Information
- 15 Declaration of Conformity
- 15 Statement of Compliance
- 15 General Safety Considerations
- 16 Regulatory Information
- 17 Agilent Support, Services, and Assistance
- 17 Service and Support Options
- 17 Contacting Agilent
- 17 Shipping Your Analyzer to Agilent for Service or Repair