User Manual

Waterproof Weighing Scale

ELW Plus / ESW Plus

ELW Plus Wipower

© Excell Precision Limited 2016. All rights reserved Worldwide.

The information contained herein is the property of Excell Precision Limited and is supplied without liability for errors or omissions. No part may be reproduced or used except as authorized by contract or other written permission. The copyright and the foregoing restriction on reproduction and use extend to all media in which the information may be embodied.

EXCELL PRECISION CO., LTD.

Table of Content

Instruction for Use ............................................................................................................................. 2

Preparing To Use the Scale ............................................................................................................. 2

Chapter 1 Introduction ...................................................................................................................... 4

1-1 Features and Specifications.................................................................................................. 4

1-2 Power Supply .......................................................................................................................... 5

1-3 Panel and Keyboard Introduction ........................................................................................ 6

1-4 Operating the Scale ............................................................................................................... 6

1-5 Self-Test Mode ........................................................................................................................ 8

1-6 Error Messages ...................................................................................................................... 9

1-7 Weight Unit .............................................................................................................................. 9

Chapter 2 Advanced Functions .....................................................................................................10

2-1 Advanced Function Setting Table.......................................................................................10

2-2 Advanced Function Setting Workflow................................................................................ 11

2-3 General Function Setting .......................................................................12

2-3-1 Automatic Backlight Function Setting ............................................13

2-3-2 Automatic Power-off Timer Setting .................................................14

2-3-3 Hi/Lo/OK Function Setting ................................................................15

2-3-4 Restore to the Default Setting .........................................................16

2-3-5 Noise Filter Setting ...........................................................................17

2-3-6 Hold Function Setting ......................................................................18

2-3-7 Auto Unit Weight Averaging Setting .............................................19

2-4 Weight Calibration ........................................................................................20

Appendix 1 7-Segment Display C haracters ................................................................................21

02026100 1 ZSME300000351

EXCELL PRECISION CO., LTD.

Thank you for purchasing EXCELL waterproof weighing scale. In order to use the scale properly, please read this instruction carefully before use. If you have a problem concerning the scale, please contact your supplier.

Instruction for Use

1. Please keep the scale in a cool dry place. Do not store it at high temperature.

2. Avoid objects impacting with the scale. Do not drop loads onto the scale or subject the weigh pan to any strong shock loads.

3. The load placed on the weigh pan must not exceed the maximum weighing capacity of the scale.

4. If the scale is not going to be used for some time, please clean it and store it in a plastic bag in dry conditions. A desiccant sachet may be included to prevent any moisture build up.

5. Please operate or charge the scale in an open area. Do not squeeze the power cord to avoid wire on fire.

Preparing To Use the Scale

1. Adjust the four levelling feet (if fitted) to set the scale pan level using the spirit level bubble located at the front of the scale.

2. Avoid operating the scale in direct sunlight or drafts of any kind.

3. If possible avoid connecting the scale to ac power outlet sockets which are adjacent to other appliances to minimise the possibility of interference affecting the performance of the scale.

4. Remove any weight that might be on the weigh pan before the scale is switched on and avoid leaving weight on the pan for long periods of time

5. All goods weighed should be placed in the centre of the weigh pan for accurate weighing. The overall dimensions of the goods being weighed should not exceed the dimension of the weigh pan.

6. We suggest to warm up the scale for 15~20 minutes before operation to ensure best accuracy.

7. When symbol keep flashing on the display, please charge the batteries soon.

8. Introduction of Storage Battery

02026100 2 ZSME300000351

EXCELL PRECISION CO., LTD.

Due to the storage battery adopt the advanced free -maintaining technique, customers need not to replenish electrolyte.

The scale should be recharged every 3 months to prevent failure of the internal rechargeable battery.

1. The battery should be charged for 8~10 hours.

2. The temperature of battery should below 45℃.

Maintaining

1. Please do not discharge with over-current when using the battery. Please charge the battery after discharging current.

2. Please take down the battery when the scale is not used for a long time or break the connection of cathode.

3. Do not short the battery terminals to check whether there is current. Please check whether the connection point is firm to guarantee good connection.

4. The battery should be replaced by specialized person. No reverse-battery or the

product will be damaged. a) Anode of battery should be connected with Anode of product battery(usually red cable) b) Cathode of battery should be connected with Cathode of product battery (usually brown cable or black cable) c) See the picture following:

Brown cable(or black cable) connected with Cathode of battery

Red cable connected with

Anode of battery

Safety warnings

1. The electrolyte of battery is caustic which causes metal, cotton, etc to corrode.

2. The hydrogen will be resolved when using or charging the battery and it will cause explosion when approaches fire.

No burning

Caution Corrosion

Warning explosion

Children faraway

02026100 3 ZSME300000351

EXCELL PRECISION CO., LTD.

Chapter 1 Introduction

1-1 Features and Specifications

Features:

 Sealed waterproof silica gel strip blocks water from infiltrating into the scale.

 Surrounded by waterproof grade sheeting to ensure the water free.

 1/3,000~1/6,000 display resolution available.

 ESW Plus adopts stainless steel housing while ELW Plus and ELW Plus Wipower plastic ABS housing.

 High speed of 24bits AD fast reacts and shortens the weighing operation duration.

 Selectable units: Kilogram (kg), gram (g), ounce (oz), and pound (lb) weighing units available.

 Built-in rechargeable battery can be easily replaced.

 Well-designed protection point for transportation.

 Low power indication and auto power off.

Specifications:

Model Capacity Division Resolution

ELW Plus-3

ELW Plus Wipower-3

ESW Plus-3

ELW Plus-6

ELW Plus Wipower-6

ESW Plus-6

3kg

6kg

0.5g

1g

1g

2g

1/6,000

1/3,000

1/6,000

1/3,000

ELW Plus-15

ELW Plus Wipower-15

ESW Plus-15

15kg

2g

5g

ELW Plus-30

ELW Plus Wipower-30

ESW Plus-30

30Kg

5g

10g

Operating Temperature -10°C ~ 40°C (14°F ~ 104°F)

Dimensions : ELW Plus:270 x 124 x 310mm (W x H x D)

ELW Plus Wipower:270 x 124 x 310mm (W x H x D)

ESW Plus: 240 x 120 x 280mm (W x H x D)

Weight of the scale: ELW Plus: 3.3 kg approximately;

ELW Plus Wipower: 3.3 kg approximately;

ESW Plus: 3.5 kg approximately;

1/7,500

1/3,000

1/6,000

1/3,000

02026100 4 ZSME300000351

EXCELL PRECISION CO., LTD.

1-2 Power Supply

Power Supply and Power Consumption

 DC

6V / 4Ah rechargeable battery

 AC 100V~240V (±10

%) adapter

 29 mA (system no backlight) about 153 hours

80 mA (system backlight) about 50 hours

 Wireless Charging (ELW Plus Wipower only)

Power Supply adapter 100V~240V(50~60Hz)

Output:>5W (meets

WPC1.12_Qi Wireless Power Supply’s) a. While the scale is charging, the spacing between charging plate and scale shall be

within 2mm. Metal objects are prohibited between the spacing; otherwise, It might cause overheating or burning. b. Charging without battery will cause charging indicator to flash red and green. Please avoid that. c. How to use Wireless charging

Raise the scale and aim the point

Plug in wireless adapter

Position hole is located the bottom of the scale near the center. Slide the charging plate and attach to the position hole

This side up

Try to move around the charging plate and ens ure it is unmovable

Red indicates it is charging

Green indicates it is fully charged

Low Battery Warning

When symbol keep flashing on the display, please recharge the batteries soon.

02026100 5 ZSME300000351

EXCELL PRECISION CO., LTD.

1-3 Panel and Keyboard Introduction

Icon Introduction

STABLE

: The weight is stable.

GROSS

: The scale is in the gross mode. The display shows the goods and any container weight. This Gross status indication is on when the TARE function is used.

Pcs

: The unit weight is not sufficient. When the icon is on, the counting function is operational but the count may contain errors.

.

.

.

.

.

.

.

.

.

.

: The sampling size is not sufficient When the icon is on, the counting function is operational but the count may contain errors.

Keyboard Function

ON/ZERO KEY

This key possesses two functions: Power On and Zero function.

OFF KEY

When the scale is switched on, press the OFF key, the scale will switch off.

TARE KEY

The tare function will not operate during the following conditions:

1. When the scale powers on if the weight is negative and after a container is placed on the weigh pan if the weight is still below zero.

2. The tare value is over the full scale capacity.

UNITS KEY

Press the UNITS key to switch weight units; the icons will indicate the active units.

NET/GROSS KEY

In the Tare mode, the screen displays the “TARE” icon; press the NET/GROSS key to switch between the “Net value” and the “Gross value”.

1-4 Operating the Scale

POWER ON

When the scale is off, press the ON/ZERO key, the scale will switch on.

POWER OFF

When the scale is on, press the OFF key, the scale will switch off.

02026100 6 ZSME300000351

EXCELL PRECISION CO., LTD.

ZERO

When the weigh pan is empty (free of load) and the display is not showing zero, press the ON/ZERO key to ze ro the scale. At zero, the “0” indication is on.

 When the weight value is within the zero range, the zero function operates to zero the scale or cancel the tare function.

 Zero range: The OIML & NTEP models have a zero range of

2% of Full Scale. The

Sri Lanka model has a zero range of

4% of Full Scale.

SWITCHING UNITS

Press the UNITS key to switch weight units, the icons or arrows will indicate the active units as appropriate. The units available are dependent on the exact scale model.

 After power off, the scale will memorize the active units. When the scale is powered on again, it displays the previously active units.

TARE FUNCTION

(1) Put a container on the weigh pan and after the weight is stable, press the TARE key to zero the weig ht of the container. The screen displays the “Net” icon.

(2) Put the goods in the container, the screen displays the net weight value of the goods.

(3) Remove the full container; the screen displays the negative weight value of the container. At this time pressing the TARE key again will cancel the tare and the scale reverts back to zero. The “Net” icon is switched off.

 The tare function can be operated continually to the full weighing capacity of the scale.

 Continual tare operation is adding or removing tare objects on weigh pan and pressing the TARE key each time.

NET/GROSS FUNCTION

In the Tare mode, the screen displays the “Net” icon, press the NET/GROSS key to switch between the “Net value” and the “Gross value”.

 When the GROSS icon is on, the weight value on the display is the total amount of the tare value and net value.

 At the Gross status, only OFF and NET/GROSS keys are functional.

 NET/GROSS key is only used in Tare mode.

SIMPLE COUNTING FUNCTION

(1) Use the UNITS key to enter into the “PCS” mode

(2) Press the NET/GROSS key to select the counting sample size

(S = 10, S = 20, S = 50, S = 100, S=200).

The LCD shows , , , , in order

.

(3) Put the samples on the weigh pan and press the UNITS key, the screen displays

“------”. After the sampling process is complete, put the goods on the weigh pan and the screen shows the quantity of the items.

 The sample weight should be heavier then the minimum capacity of the sca le (20d), If

.

.

.

.

.

.

.

.

.

 The weight of a sample should be heavier than the 0.2d (d=division), or the arrow pointing to the icon will be on.

Pcs

.

.

.

.

.

.

.

.

.

contain errors.

02026100 7 ZSME300000351

EXCELL PRECISION CO., LTD.

 To power off in this mode, the scale will memorize the “Pcs” unit. When the scale is powered on again, it directly e nters the simple counting mode.

 While the

“Auto unit weight average” function is available in the Advanced Function, the goods on the weigh pan are 5pcs greater than the sample size and less than double the sample size, the scale will automatically re-sample the unit weight.

1-5 Self-Test Mode

Set the switch SWA1 on the bottom of machine to the LOCK position.When power is off, hold NET/GROSS , and press ON/ZERO key, Wait till display shows to enter “Self-Test Mode”.

Press

ON/ZERO

key

Press

ON/ZERO

key

Press

ON/ZERO

key

Press

ON/ZERO

key

Press

UNITS

key

Press

UNITS

key

Press

UNITS

key

Press

UNITS

key

INTERNAL VALUE MODE

(must hook up Load Cell to test)

 Press

TARE

to enter, and the display shows internal value

 Please check whether the internal value has changed obviously with weight changing.

 Please check the backlight.

 Press

ON/ZERO

key to back to the last screen , the display shows

KEYPAD TEST MODE

 Press

TARE to enter, display shows

Keypad

’s internal code:

TARE key= 07 UNITS key= 06 NET/GROSS key=05

 Press

ON/ZERO key to back to the last screen , the display shows

FIRMWARE VERSION DISPLAY MODE

 Press

TARE

to enter , display shows the firmware version

 Press

TARE

key again, the display shows maintenance number for 2 secs

 Press

ON/ZERO key to back to the last screen, display shows

BACK TO THE LAST SCREEN

Press

TARE

key to exit self-test mode, the scale will restart automatically.

02026100 8 ZSME300000351

EXCELL PRECISION CO., LTD.

1-6 Error Messages

Initial zero is higher than the zero range when switching the indicator on.

(Approval models)

Initial zero is lower than the zero range when switching the indicator on.

(Approval models)

Internal value is below Zero.

The weight of the object is over 9 divisions of the maximum capacity.

A/D IC value is over the maximum range.

1-7 Weight Unit

kg 1 g = 0.001 kg g 1 g = 1 g lb 1 g = 0.002204623 lb oz 1 g = 0.03527396 oz

02026100 9 ZSME300000351

EXCELL PRECISION CO., LTD.

Chapter 2 Advanced Functions

2-1 Advanced Function Setting Table

Below is an overview of the advanced functions. For detailed settings refer to the following sections:

DISPLAY LEVEL 2 FUNCTIONS DISPLAY LEVEL 1 FUNCTION

Exit the ADVANCED

FUNCTION setting mode

--- ---

Return to theADVANCED

FUNCTION setting menu

Automatic backlight function setting

Automatic power-off timer setting

General Function setting mode

Hi/Lo/OK function setting

Restore the default settings

Noise filter setting

Hold function setting

Auto unit weight averaging setting

External Weight

Calibration

--- ---

02026100 10 ZSME300000351

EXCELL PRECISION CO., LTD.

2-2 Advanced Function Setting Workflow

In the weighing mode, press the

NET/GROSS

and

ON/ZERO

keys at the same time to enter the Advanced Function setting mode. The LCD shows .

Overall workflow of the Advanced Function setting mode:

Press UNITS key

Press ON/ZERO key

Press UNITS key

Press ON/ZERO key

Press ON/ZERO key

Press UNITS key

General Function setting mode

External Weight Calibration

Exit the Advanced Function setting mode

Refer to the following sections for the detailed operation procedures of each function setting.

02026100 11 ZSME300000351

EXCELL PRECISION CO., LTD.

2-3 General Function Setting

There are 7 functions in the general function setting mode from

to .

Workflow of the General Function setting:

ACTIONS

DISPLAY NOTE

Press

TARE

key

ON/ZERO key

Upward key (from 0 to 9)

UNITS key

Downward key (from 9 to 0)

TARE key

Move the cursor to right

NET/GROSS key

Move the cursor to left

Use the keys to select the function

Key in

“00” and press

TARE key to return to the Advanced function setting menu

Return to the Advanced Function Setting Mode Menu

Automatic Backlight Function Settings

Automatic Power-off Timer Settings

Hi/Lo/OK Settings

Restore the Default Settings

Noise Filter Settings

Hold Function Settings

Auto Unit Weight Averaging Setting

Refer to the following sections for detailed operation procedures of each setting.

02026100 12 ZSME300000351

EXCELL PRECISION CO., LTD.

2-3-1 Automatic Backlight Function Setting

Select in the General Function setting mode to change the backlight function setting.

ACTIONS DISPLAY

NOTE

Press

TARE

key

Use

ON/ZERO or

UNITS

key to select function

“on” or “off

Press

TARE

key

Key in the parameter “00”

Press

TARE

key to complete the general setting procedures

Automatic backlight function

LCD displays the last status

When the weight is over 10d, the display backlight will be on. After the weight is stable for 10 seconds or when the scale returns to zero, the display backlight switches off.

02026100 13 ZSME300000351

EXCELL PRECISION CO., LTD.

2-3-2 Automatic Power-off Timer Setting

Select in the General Function setting mode to change the automatic power-off timer setting.

ACTIONS

DISPLAY NOTE

Press

TARE

key

Use ON/ZERO or UNITS key to key in parameter

Automatic power-off function

LCD displays the last status

Press

TARE

key

Key in the parameter

“00”

Press TARE key to complete the general setting procedures

Automatic power-off timer setting

Use

ON/ZERO

or

UNITS

key to key in parameter

0

No auto power-off

1

2

When the scale is idle for 1 minute, the scale automatically switches off

When the scale is idle for 2 minutes, the scale automatically switches off

9

When the scale is idle for 9 minutes, the scale automatically switches off

ON/ZERO key

Upward key (from 0 to 9)

UNITS key

Downward key (from 9 to 0)

TARE key

Move the cursor to right

NET/GROSS k ey

Move the cursor to left.

When the weight on weigh pan is less than 10d or keeps idle for the set time, the scale will automatically switch off.

02026100 14 ZSME300000351

EXCELL PRECISION CO., LTD.

2-3-3 Hi/Lo/OK Function Setting

Select in the General Function setting mode to set the

Hi/Lo/OK function. This function is available in all unit modes. In one specific unit mode, enter to set the Hi/Lo/OK values.

When the high limit and low limit are both set as “0”, the Hi/Lo/OK function is disabled.

ACTIONS

DISPLAY

NOTE

LCD displays the last status

Press TARE key

Key in the desired high limit value

Press TARE key

Key in the desired low limit value

Press TARE key

Key in the parameter setting for alarm

Press TARE key

Key in the parameter

“00”

Press TARE key to complete the general setting procedures

.

.

.

.

lb lb lb lb

LCD displays the last status

LCD displays the last status

Alarm setting

(a) (b) (c )

The definition of a, b, c positions:

(a)

1 = beep sound on

0 = beep sound off

(b)

1 = beep sound on when stable

0 = beep sound on when unstable

(c)

1 = beep sound on when the weight is between high and low limits

0 = beep sound on when the weight is not between high and low limits and higher than 10d

ON/ZERO key

Upward key (from 0 to 9)

UNITS key

Downward key (from 9 to 0)

TARE key

Move the cursor to right

NET/GROSS key

Move the cursor to left.

02026100 15 ZSME300000351

EXCELL PRECISION CO., LTD.

2-3-4 Restore to the Default Setting

Select in the General Function setting mode to restore to the default setting.

ACTIONS

DISPLAY NOTE

Press TARE key restoration)

Return (Cancel the

Use ON/ZERO or UNITS keys to select

“return” or “format”

Press TARE key setting

Restore to the default

Key in the parameter

“00”

Press TARE key to complete the general setting procedures

 The default setting includes the following:

ON/ZERO key

Upward key (from 0 to 9)

UNITS key

Downward key (from 9 to 0)

TARE key

Move the cursor to right

NET/GROSS k ey

Move the cursor to left.

1) External weight calibration

2) HI/LO/OK setting value

3) Noise filter setting (External)

4) Sampling setting for the counting function

 Setting is only available for non-approval models.

02026100 16 ZSME300000351

EXCELL PRECISION CO., LTD.

2-3-5 Noise Filter Setting

Select in the General Function setting mode to set the noise filter setting.

ACTIONS

DISPLAY

NOTE

Returning to zero point setting

LCD displays the last status

Press TARE key

Returning to the zero point setting

Use

ON/ZERO

or

UNITS

key to key in the

Use ON/ZERO or UNITS key to key in the parameters

Press TARE key parameters or zero point

Default setting = 0

0

No skip 5

skip 5d

1

skip 1d 6

skip 6d

2

skip 2d 7

skip 7d

3

skip 3d 8

skip 8d

4

skip 4d 9

skip 9d

 When the weight on the scale is over

1/3 full capacity, the function is on.

Use ON/ZERO or UNITS key to key in the parameters

Digital switch & Stabilization range

setting

LCD displays the last parameter setting

Press TARE key

Digital switch & Stabilization range

setting

Use

ON/ZERO

or

UNITS

keys to key in the parameters. Default setting = 0

Use ON/ZERO or UNITS key to key in the parameters

Parameter 0 ~ 9, the larger the number the more stable the weight.

Filter parameter setting

LCD displays the last parameter setting

Press TARE key

Key in the parameter

“00”

Press TARE key to complete

Filter parameter setting

Use ON/ZERO or UNITS keys to key in the parameters. Default setting = 5

Parameter 0 ~ 9, the larger the number, the faster the filter response. Fast response can lead to weight instability. general setting procedures

Setting is only available for non-approval models.

02026100 17 ZSME300000351

EXCELL PRECISION CO., LTD.

2-3-6 Hold Function Setting

Select in the General Function setting mode to set the hold function.

ACTIONS

DISPLAY

NOTE

Press TARE key

Hold function setting

LCD displays the last parameter setting

Use ON/ZERO or UNITS key

to key in the parameters

Hold function setting

Use

ON/ZERO

or

UNITS

keys to key in the

Press TARE key parameters

Default setting = 0

0

Hold function disabled

1

“Peak hold” mode

2

“Stable hold 1” mode

3

“Stable hold 2” mode

Key in the parameter

“00”

Press TARE key to complete the general setting procedures

= Hold function disabled

= “Peak hold” mode

ON/ZERO key

Upward key (from 0 to 9)

UNITS key

Downward key (from 9 to 0)

TARE key

Move the cursor to right

NET/GROSS key

Move the cursor to left.

Keep displaying the maximum weight when the weight is continually changing

To exit this mode, press any key.

=

“Stable hold 1” mode

When the weight is stable, the LCD shows the current weight value. To exit this mode, press any key.

= “Stable hold 2” mode

When the weight is stable, the LCD shows the current weight value. When the weight returns to zero (<10d), the hold mode is cancelled automatically.

 Setting is unavailable for Brazil models.

02026100 18 ZSME300000351

EXCELL PRECISION CO., LTD.

2-3-7 Auto Unit Weight Averaging Setting

Select

Averaging.

ACTIONS

in the General Function setting mode to set Auto Unit Weight

DISPLAY

NOTE

Press TARE key

Use ON/ZERO or UNITS key

to key in the parameters

Auto unit weight averaging setting

LCD displays the last parameter setting

Press TARE key

Auto unit weight averaging setting

Use

ON/ZERO

or

UNITS

keys to key in the parameters

Default setting = 0

0

Auto unit weight averaging

1

Non auto unit weight averaging

Key in the parameter

“00”

Press TARE key to complete the general setting procedures

ON/ZERO key

Upward key (from 0 to 9)

UNITS key

Downward key (from 9 to 0)

TARE key

Move the cursor to right

NET/GROSS key

Move the cursor to left.

 Setting is unavailable for Brazil models.

02026100 19 ZSME300000351

EXCELL PRECISION CO., LTD.

2-4 Weight Calibration

In the weighing mode, press the NET/GROSS and ON/ZERO keys at the same time to enter the Advanced Function setting mode. The LCD shows and use the NET/GROSS or UNITS key to select to enter the weight calibration mode.

ACTIONS

DISPLAY

NOTE

Press TARE key

.

When the digit on the right flashes, ensure that the weigh pan is empty. Then

Press the TARE key.

After stabilization

Place the displayed weight on the weigh pan. Press the

TARE key. The scale calculates the full capacity internal value and after stabilization, the buzzer beeps 3 times. Remove the

.

weight from the weigh pan and the procedure is completed.

kg kg kg

The LCD displays the calibration weight value (the maximum capacity of 1 st

unit) and the unit. The digit on the right side keeps flashing.

The scale is calculating the internal value of the zero point. After stabilization, the LCD shows the calibration weight value.

ON/ZERO key

Upward key (from 0 to 9)

UNITS key

Downward key (from 9 to 0)

TARE key

Move the cursor to right

NET/GROSS key

Move the cursor to left.

 Unavailable for approval models

 Weight calibration conditions:

The calibration weight value placed on the weight pan must be over 100e, and must be within the 90% and 110% of the full weight

02026100 20 ZSME300000351

EXCELL PRECISION CO., LTD.

Appendix 1 7-Segment Display Characters

Digit 7 segments letter Alphabet 7 segments letter Alphabet 7 segments letter

0 A N

1 B O

2 C P

3 D Q

4 E R

5 F S

6 G T

7 H U

8 I V

9 J W

K X

L Y

M Z

02026100 21 ZSME300000351