model 8463 baby scale

Add to my manuals
12 Pages

advertisement

model 8463 baby scale | Manualzz

MODEL 8463

BABY SCALE

Operation Manual

8535-0232-0M Rev B 203 E. Daugherty, Webb City, MO 64870 USA

04/16

Ph: 417-673-4631

 Fax: 417-673-2153

www.detecto.com

Technical Support: Ph: 866-254-8261

[email protected]

2 8535-0232-0M Rev B

 8463 Operation

TABLE OF CONTENTS

INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 1

FEATURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 1

SETUP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 2

PRODUCT DESCRIPTION . . . . . . . . . . . . . . . . . . . . Page 4

OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 5

CARE AND MAINTENANCE . . . . . . . . . . . . . . . . . . Page 7

ERROR AND STATUS MESSAGES . . . . . . . . . . . .

Page 7

SALES SUPPORT AND ORDER ENTRY . . . . . . . .

Page 7

COPYRIGHT

All rights reserved. Reproduction or use, without expressed written permission, of editorial or pictorial content, in any manner, is prohibited. No patent liability is assumed with respect to the use of the information contained herein.

DISCLAIMER

While every precaution has been taken in the preparation of this manual, the Seller assumes no responsibility for errors or omissions. Neither is any liability assumed for damages resulting from use of the information contained herein. All instructions and diagrams have been checked for accuracy and ease of application; however, success and safety in working with tools depend to a great extent upon the individual accuracy, skill and caution. For this reason, the Seller is not able to guarantee the result of any procedure contained herein. Nor can they assume responsibility for any damage to property or injury to persons occasioned from the procedures. Persons engaging the procedures do so entirely at their own risk.

8535-0232-0M Rev B

 8463 Operation

3

INTRODUCTION

The Model 8463 Baby Scale is a 15kg capacity digital baby scale. The scale is equipped with a collapsible baby tray for easy storage.

This manual contains information on the operation of this scale. Please read it before attempting to use the scale and keep it handy for future reference.

FEATURES

Collapsible baby tray for easy storage

Weight Hold Display

Tare

Weight Percentile Calculation

Large power saving LCD display

Auto-Off during battery power

Low battery indicator

Capacity: 15kg

Graduations: 0kg to 7.5kg x 5g, 7.5kg to15kg x 10g

Accuracy:

±2.5g (0-2.5kg), ±

5g (2.5kg -7.5kg),

±10

g (7.5kg – 15kg)

Power: 9 volts DC supplied by six 1.5 volt AA size batteries or optional AC adapter.

Product dimension: 566mm x 350mm x 144mm

Product weight: 3.3 kg

8535-0232-0M Rev B

 8463 Operation

1

SETUP

Unpacking

Before beginning installation of your scale, make certain the instrument has been received in good condition. When removing the scale from its packing, inspect it for signs of damage, such as exterior dents and scratches. Keep the carton and packing material for return shipment if it should become necessary. It is the responsibility of the purchaser to file all claims for any damages or loss incurred during transit.

1. Remove the scale from the shipping carton and inspect it for any signs of damage.

2. Plug-in the optional AC adapter or install six AA size batteries as shown below.

Optional AC Adapter

To apply power to the scale using optional AC adapter (not included), insert the plug from the

AC adapter cable into the power adapter jack at the rear of the scale and then plug the AC adapter into the proper electrical outlet.

Power Adapter Jack

Optional Battery Power

The scale can use six AA size batteries (not included). If you wish to operate the scale from batteries, you must first obtain and install the batteries as shown in the battery holder at the bottom of the scale.

Battery Holder

6 AA Size

2 8535-0232-0M Rev B

 8463 Operation

SETUP, CONTINUED

1. Place scale on flat, stable surface.

2. Unfold tray until it snaps together.

3. Remove screws form each side of scale.

4. Unfold tray brackets. From rear of scale slide tray in track on each side of scale until screw holes in bracket align with holes on sides of scale.

5. Fasten tray to scale by re-installing the two screws.

8535-0232-0M Rev B

 8463 Operation

3

PRODUCT DESCRIPTION

4

Keys

+ / T

WP / Esc

- / Hold

Increase Value

Tare

Weight Percentile

Escape / Cancel

Decrease Value

Hold

Power

Enter

Press to increase the input value.

Press for Tare in weighing mode.

Press to enter weight percentile calculation mode.

Press to cancel input value.

Press to decrease the input value.

Press to hold / release displayed weight.

Press to turn scale on and off.

Press to enter or confirm input value.

Display

NET

Net weight, Tare function in use

HOLD

Hold function in use

MONTH Baby month

Male icon Female icon

<

Percentile value

-

Negative value

0

Weight at zero

~

Weight stable

8535-0232-0M Rev B

 8463 Operation

OPERATION

Weighing

1. Make sure that nothing is on the scale platform.

2. Press key.

3. The software version followed by full screen display will appear. Scale will then display

0.000 kg

along with

~

and

0

once a stable zero weight is established.

4. Place the baby on the tray and the weight will be displayed.

NEVER leave an infant or child unattended while on the scale. Failure to maintain control of infant or child at all times while they are on the scale can result in serious injury to infant or child.

Hold Weight Display

1. When the baby’s weight is displayed, press the

- / Hold

key once to hold the weight on the display. Remove the baby form the scale. The weight will continued to be displayed.

2. After recording the displayed weight, press the

- / Hold

key once to release the weight display and return to weighing mode.

Tare Function

Tare allows for the scale to weigh accurately with additional items on the scale such as blanket or towel.

1. In the weighing mode place the item (e.g. blanket) on the scale. The item weight will be displayed.

2. Press the

+ / T

key.

NET and

0.000 kg

weight will appear on the display.

3. Place the baby on the scale with the item (e.g. blanket) the weight of the baby will be displayed.

4. When the baby and the item are removed the scale will display a negative weight.

5. Press

+ / T to exit Tare function

8535-0232-0M Rev B

 8463 Operation

5

WEIGHT PERCENTILE

Weight Percentile reflects body weight relative to age and gender, and is influenced by recent changes in growth, health, or nutritional status. The percentile compares a child’s weight to other children and indicates what percentage of children of the same age and gender weigh less than the child being weighed.

For example, a 30 month old boy who is in the 30 th

weight percentile weighs more than 30% of other children his age, while 70% of children weigh more. This usually means he is smaller than average, but may not indicate any health problems. Concerns should be referred to a professional healthcare provider.

WEIGHT PERCENTILE CALCULATION

1. In the weighing mode press

WP / Esc.

2. Press

+ / T or

- / Hold to adjust age form 1 to 36 months.

3. Press to confirm or press

WP / Esc to return to weighing mode.

4. Press

+ / T or

- / Hold to adjust gender for male or female.

5. Press to confirm.

6. Place baby on scale. The weight will be shown, then the percentile. Weight and percentile display will cycle three times, then return to weighing mode.

6

NOTE:

< p3

will be displayed if the weight percentile is less than 3, and

P97<

will be displayed if the weight percentile is greater than 97.

8535-0232-0M Rev B

 8463 Operation

CARE AND MAINTENANCE

1. Do not disassemble the scale, other than replacing the battery it contains no user serviceable parts.

2. Remove batteries if the scales is not to be used for extended period of time.

3. Damage to the scale may occur as a result of improper handling. Avoid excessive impact or vibration to the scale such as dropping it.

4. Clean the scale with a lightly dampened cloth. Do not use solvents or immerse in water.

ERROR AND STATUS MESSAGES

Display Meaning

UStb Unstable Make sure scale is on flat, stable surface

Err0

Initial zero is too large. Remove all weight from scale.

Err2

Err3

EEP

NoCAL

Bat or

Overload Remove all weight immediately to prevent permanent damage

Scale has an initialization problem.

EEPROM Error

Calibration Error

Displayed when the battery voltage has dropped too low for accurate weighing. Remove the old battery and replace with a new battery before continuing operation.

SALES SUPPORT AND ORDER ENTRY

Mr. Andy Stevens, International Sales Manager (EMEA)

E-mail: [email protected]

Phone: +44 (0) 7787 834778

Factory UK Office

Ms. Allison Balaam, International Business and Sales Administrator (EMEA)

E-mail: [email protected]

Phone: +44 (0) 2078 594904

For Order Entry

E-mail: [email protected]

8535-0232-0M Rev B

 8463 Operation

7

8 8535-0232-0M Rev B

 8463 Operation

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement