KILOTECH Weighing scale

KILOTECH Weighing scale
KILOTECH
KWS 2200
Weighing scale
Operation Manual
V5.13
Contents Subject to Change without Notice
──────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
Ce manuel est disponible en français à www.kilotech.com
ou auprès de notre service à la clientèle au 1-800-694-4445.
Ce manuel est disponible en Spanish à www.kilotech.com
ou auprès de notre service à la clientèle au 1-800-694-4445.
────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
1
CONTENT
1. Specifications ........................................................................................ 4 2. Faceplate ................................................................................................ 6 3. Summary of Key function ..................................................................... 7 4. Switch between operation mode ......................................................... 8 5. Normal Weighing mode ........................................................................ 9 6. Accumulation Mode: ........................................................................... 10 7. Check Weight ....................................................................................... 10 8. Count Weighing Mode ........................................................................ 10 9.Check Counts (counts compare) ........................................................ 11 10. Percent Weighing Mode .................................................................... 12 11.Check Percent (percentage compare) .............................................. 13 12. BMI Working Mode ............................................................................ 13 13. HOLD Function .................................................................................. 14 14. Data Compare Function .................................................................... 15 15. Accumulation ..................................................................................... 16 16. View or Set Time ................................................................................ 17 17. View or Set Date ................................................................................ 17 18. View Firmware Version ..................................................................... 17 19. Display Test ....................................................................................... 17 20. Keyboard and Buzzer Test ............................................................... 17 21. Enter Setup Mode: ............................................................................. 18 22. UPS world ship: ................................................................................. 18 23. Error codes and Troubleshooting ................................................... 18 24. Troubleshooting ................................................................................ 20 25. Display Character .............................................................................. 22 26. Packing List ....................................................................................... 23 ────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
2
KWS 2200 weighing scale Operation Manual
Thank you for purchasing the KWS 2200 weighting scale. Please read all operating instructions carefully
before use and keep the following points in mind:
* Avoid lengthy exposure to extreme heat or cold, your scale works best when operated at normal room
temperature. Always allow the scale to acclimate to a normal room temperature before use
* Allow sufficient warm up time. Turn the scale on and wait for a few minutes if possible, to give the internal
components a chance to stabilize before weighing.
* These electronic scales are precision instruments. Do not operate near an in-use cell phone, radio, computer
or other electronic device. These devices emit RF and can cause unstable scale readings. If your scale ever
performs poorly, try moving the scale to a different room or location.
* Avoid using in condition of heavy vibration and airflow.
* Read the weight reading in short time after loading. The output signature of load cell and electronic circuit
may be little influenced after weighing for a long time.
────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
3
1. Specifications
1.1 Housing and Outline Dimension:
A
B
C
D
E
F
355mm (14”)
82mm 3 (¼”)
305mm (12”)
44mm (1 ¾”)
89mm (3 ½”)
203mm (8”)
1.2 Power Source options:
•
AA Batteries (4x): The display segments will flash when the batteries are almost empty. When
“Lo.bAt” is displayed, you must replace the batteries immediately and the indicator will turn off
automatically after 10s.
• AC Adapter: 6-9VDC, ≥500mA, central positive:
• USB power supply
1.3 Real Clock:
built-in nonvolatile real time & date
1.4 Programmable settings:
pre-set tare weight
motion detection window
auto-power off time,
backlight working mode
hold function: (peak weight, dynamic weighing)
Check Weighing Mode;
Counting Mode;
Percentage Working Mode
1.4.7 Data Comparison and indicator of High, OK and Low is available in Weighing mode, Parts Counting
mode and Percentage Working mode.
1.4.8 Battery voltage is low and charging indicator
1.4.1
1.4.2
1.4.3
1.4.4
1.4.5
1.4.6
────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
4
1.4.9
1.4.10
1.4.11
Programmable what content will be output and when they are output on serial port.
BMI function
Available Measure Unit: kg, lb, lb:oz, PCS, %,
1.5 Display:
6-digit, 7-segment , 1"(25mm) LCDs with 16 annunciators and blue backlight
1.6 Keypad:
6 push buttons
1.7 Environment:
1.7.1 Working temperature: -10°C to 40°C
1.7.2 Storage temperature: -20°C to 70°C
1.7.3 Humidity: 10 to 80% RH without condensation
1.8 Communication:
1.8.1 Serial port0: Full-duplex RS232
1.8.2 Serial port1: USB (Virtual RS232)
1.8.3 Baud Rate: Selectable: 1200-2400-4800-9600-19200-38400 bps
1.8.4 Data Output Format: 8N1, 7O1, 7E1
1.8.5 Protocol: programmable
────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
5
2. Faceplate
Meaning of symbol on faceplate:
2.1
lights up when scale is stable
2.2
lights up when scale is at zero point and the gross weight is 0
2.3 NET
lights up when net weight is displayed, and the tare weight is not 0
2.4 TOTAL
lights up when display data is total times or total of weight, pieces or percentage
2.5
lb
lights up when measure unit is lb or lb:oz
2.6
oz
lights up when measure unit is oz or lb:oz
2.7
kg
lights up when measure unit is kg
2.8
%
lights up when in percentage weighing mode, measure unit is %
2.9
Pcs
lights up when in counting mode, measure unit is pieces
2.11 PEAK
lights up when PEAK-HOLD is active
2.12 HOLD
lights up when HOLD mode is active
2.13 HI
lights up when data compare is activated and current data (weight, pieces or percent) is over
the upper setting limitation.
2.14 OK
lights up when data compare is activated and current data (weight, pieces or percent) is
within the set limits
2.15 LO
lights up when data compare is activated and current data (weight, pieces or percent) is
below the set limit
2.16
lights up when battery is used or charged
────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
6
3. Summary of Key function
Key
Condition
Weighing/Counting/Percent,
press down > 3s
SETUP
Function
To enter setup mode
Input data mode, press down >3s
to input decimal point
Input data mode, press down <3s
Return to last sub-menu
Menu selection mode
Return to last sub-menu
Weighing/Counting/Percent,
press down <3s
PRINT
Weighing/Counting/Percent ,
FUNC
press down > 3s
Output data to serial communication port
select working mode: weighing ,counting or percent
Input data mode
The digit on flashed position add 1
Menu selection mode
To last item of current menu
ACCUMULATION function is enabled in configuration menu :
Weighing/Counting/Percent/BMI,
To add up current weight/pieces/percentage to memory,
press down <3s
display times and total of accumulation; in BMI working mode,
to input height
ACC
TOTAL
Weighing/Counting/Percent,
ACCUMULATION function is enabled in configuration menu,
press down >3s
To display times and total of accumulation
Input data mode
the digit on flashed position subtract 1
Menu selection mode
To Next item of current menu
Weighing mode, press down<3s
Change weighing units: kg->lb
Counting/Percent mode,
To enter getting piece weight or unit-percentage-weight mode
press down < 3s
(by way of sample or input directly).
UINT
Weighing /Counting/Percent mode, To input compare data (for weight , pieces or percentage )
DATA
press down> 3s
Display date or time mode,
press down >3s
Display voltage mode,
press down >3s
Display ADC code
high and low limitation
To set current date or time
To calibrate input voltage value
Select displaying code from no-filter, filter1, filter2
────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
7
of
Input data mode
Weighing/Counting/Percent ,
press down <3s
TARE
Weighing/Counting/Percent ,
PRESET press down > 3s
ON/OFF
Tare function
To input pre-set tare weight at selected weight unit
input data mode or
To confirm input data or current item selection, and go to next
Menu selection mode
item of current menu, or next operation
Display ADC code
Set or clear “tare” code
Power off mode
Power on
Weighing/Counting/Percent,
ZERO
Rotate the flashed position from left to right
press down <3s
Weighing/Counting/Percent,
press down > 3s
Zero function
Power off
Input data mode
ignore modification
Menu selection mode
Prepare to exit from current working mode
Note:
To access the secondary key function, you must press it for more than 3s.
4. Switch between operation mode:
Long press FUNC key, then use
to choose and confirm to switch into weighing mode,
counting mode or percentage mode
────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
8
5. Normal Weighing mode
5.1 Power on scale: when scale is off, short press ON/OFF key to turn on;
Power off scale: when scale is on, long press ON/OFF key to turn off the scale.
5.2 Switch units:
Short press UNIT key
5.3 ZERO:
The ZERO key is used to return the scale to zero. The scale will display and error message when the
scale is beyond zero range. When using a tare, the ZERO key is inactive.
5.4 TARE:
Place the item to tare on the platter and press the TARE key, the indicator will show zero, the NET
annunciator will be lighted. Please refer the ZERO and TARE limitations in 4.9 sections.
5.5 Preset tare weight: long press PRESET key, Pr.Tare will show, and the TARE annunciator flashes, it
means it is in preset TARE weight mode, use
to input tare weight.
There’s no limitation to preset tare weight, but must be greater than zero. Once you complete the setting
of the tare weight, the “NET” annunciator will be lighted.
Note: you can only store one tare weight; the new tare weight will automatically replace the old one
5.6 Clear tare weight: remove all weight from the scale and short press TARE key.
5.7 Limitation to ZERO and TARE operation under different conditions:
Weight
key function
Data in TARE
Standard
on
memory unit
Tare key
platform
no
No action
≤0
yes
Clear the tared weight
USA
No
>0
Tare
Yes
≤0
Canada
>0
no
No action
Yes
Clear the tared weight
No
Tare
Yes
No action
Zero key
Zero
Zero
NOTE: (1) ZERO only be active when scale is stable and weight is in SAZSM setting range.
(2) TARE only be active when scale is stable
(3) Clear TARE weight or ZERO scale, make indicator to enter displaying GROSS mode
(4) TARE weight, make indicator to enter displaying NET mode
────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
9
6. Accumulation Mode:
Press ACC key to add current weight to accumulation memory, the accumulation time’s counter will
increase by 1 and display total accumulative weight.
7. Check Weight
Check weight in weighing mode:
7.1 To use the check weight function, CONFIG-FUNC-COMPAR item should set to YES, and high and low
limitation of weight should be set correctly according to following steps:
7.2 In weighing mode, Press down DATA key more than 3s to input compare data of high and low.
7.3 After HIGH being shown, 000000 will be displayed, use PRINT, ACC, UNIT key to input high weight
number and press TARE key to confirm. Annunciator of Hi will be shown in this step or press ZERO key to
exit and back to weighing mode.
7.4 After Low being shown, 000000 will be displayed, use PRINT, ACC, UNIT key to input low weight number
and press TARE key to confirm. Annunciator of Lo will be shown in this step or press ZERO key to exit
and back to weighing mode.
NOTE: If High number is 0, equal or less than the low number, the comparison will be disabled, and the
input data has no limitation.
7.5 After the limitations have been set and check weight function is be active, one of annunciators HI, OK, LO
will light up, and the beeper will sound according to its setting in USER-BEEP.
8. Count Weighing Mode
8.1 In this mode, scale will weigh goods weight on scale, calculate and display its counts after the piece
weight of goods is obtained
8.2 To make counting function be available, CONFIG-FUNC-COUNT item should be set to YES in CONFIG
menu.
8.3 To enter counting working mode, in normal weighing or percent-weighing mode, long press FUNC key,
WEIGH/PERCEN will be shown, use ACC or PRINT key to select COUNT, then press TARE to confirm go
to parts counting mode. Before new piece weight is got, the last piece weight will be used.
8.4 In counting working mode, the function of ZERO, TARE, PRINT, SETUP, PRESET TARE, ACC, SETUP,
ON/OFF are available.
8.5 There’re two ways to obtain the piece weight: (1) input piece weight directly, refer operations of step 8.5.1
(2) weigh samples weight which quantity is known, refer operation of step 8.5.2
8.5.1
Input piece weight from keypad: in counting mode, press UNIT key, When InP.PWt is shown, press
TARE key to enter input piece weight mode.
8.5.1.1 When UNIT.KG is shown, use PRINT or ACC key to select the unit of piece weight, use TARE key to
────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
10
confirm and go to next, or press ZERO key to exit getting piece weight mode and back to counting
mode.
8.5.1.2 When last stored piece weight is shown, use PRINT, ACC, UNIT key to input new piece weight, press
SETUP key more than 3s to input decimal point. Press TARE key to confirm and save it, then go
back to counting mode. If the input piece weight is less than 0.5d, the indicator will display PWt.Er
and go back to counting mode.
8.5.2
Obtain piece weight by weighing samples weight which quantity: in counting mode, press UNIT key,
When InP.PWt is shown, use PRINT or ACC key to select SPL.PWT, press TARE key to weigh
samples (which quantity is known) weight, calculate piece weight. Press ZERO key to exit getting
piece weight mode and back to counting mode.
8.5.2.1 When SPL.Lo is shown, move away any sample on scale and press TARE key to confirm, before
scale is stable, SPL.Lo will be flashed. After it is stable, it will go to next step. Press ZERO key to exit
getting piece weight mode and back to counting mode.
8.5.2.2 When SPL.Hi is shown, put samples (its quantity is known) onto the scale, Press TARE key to
confirm reading weight. Before scale is stable, SPL.Hi will be flashed. After it is stable, it will go to
next step. Press ZERO key to exit getting piece weight mode and back to counting mode.
8.5.2.3 After INP.PCS being shown, 000000 will be displayed, use PRINT, ACC, UNIT key to input the
quantity of samples and Press TARE key to confirm. If the calculated piece weight is less than 0.5d,
the indicator will display PWt.Er and go back to counting mode. If successful, the scale will go back to
counting mode and you can start counting. The piece weight will be saved after the power off and can
be used next time.
9.Check Counts (counts compare) in Counting mode:
9.1 To make counts compare function be available, CONFIG-FUNC-COMPAR item should set to YES, and
high and low limitation of pieces should be set correctly according to following steps:
9.2 In counting working mode, Press down DATA key more than 3s to input compare data of high and low.
9.3 After HIGH being shown, 000000 will be displayed, use PRINT, ACC, UNIT key to input high quantity
number and press TARE key to confirm. Annunciator of Hi will be shown in this step. Press ZERO key to
exit getting piece weight mode and back to counting mode.
9.4 After Low being shown, 000000 will be displayed, use PRINT, ACC, UNIT key to input low quantity
number and press TARE key to confirm. Annunciator of Lo will be shown in this step. Press ZERO key to
exit getting piece weight mode and back to counting mode.
NOTE: If High number is 0 or is equal or less than low number, the comparison will be disabled.
9.5
After the limitations have been set and check weight function is be active, one of annunciators HI, OK,
LO will light up, and the beeper will sound according to its setting in USER-BEEP.
────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
11
10. Percent Weighing Mode
10.1
In this mode, scale will weigh goods weight on it, calculate and display its percentage after the
unit-percentage-weight of goods is obtained.
(NOTE: If % display format is set to 100%, 100.0% or 100.00% in CONFIG-FUNC-PERCEN menu
item, then, the unit-percentage-weight is the weight of 1%, 0.1% or 0.01%)
10.2
To make percent weighing function be available, CONFIG-FUNC-PERCEN menu item shouldn’t be
set to NONE.
10.3
To enter percent weighing mode, in normal weighing or counting mode, long press FUNC key,
WEIGH/COUNT will be shown, use ACC or PRINT key to select PERCEN, then press TARE to
confirm go to percent weighing mode. The last set unit-percentage-weight will be used.
10.4
In percent weighing mode, the function of ZERO, TARE, PRINT, SETUP, PRESET TARE, ACC,
SETUP, ON/OFF are available.
10.5
There’re two ways to obtain the unit-percentage-weight: (1) input weight and its percentage, then
scale calculates the unit-percentage-weight, refer operation of step 10.5.1
(2) weigh samples weight
which percentage is known, refer operation of step 10.5.2
10.5.1 Input weight and its percentage from keypad, and calculate unit-percentage-weight: in percent
weighing mode, press UNIT key, When InP.Pct is shown, press TARE key to enter this mode: Before
input weight, use PRINT or ACC key to select the percentage from 1%, 2%, 5%, 10%, 20%, 50% and
100%, this percentage is corresponding to the weight you will input in following steps
10.5.1.2
When UNIT.KG is shown, use PRINT or ACC key to select the unit of input weight, use TARE key
to confirm and go to next. Press ZERO key to exit and back to percent weighing mode.
10.5.1.3
When last stored unit-percentage-weight data is shown, use PRINT, ACC, UNIT key to input new
unit-percentage-weight, press SETUP key more than 3s to input decimal point. Press TARE key to
confirm
and
save
it,
then
go
back
to
percent
weighing
mode.
If
the
calculated
unit-percentage-weight is less than 0.5d, the indicator will display Pct.Er and go back to percent
weighing mode.
10.5.2 Obtain unit-percentage-weight by weighing samples weight which percentage is known: in percent
weighing mode, press UNIT key, When InP.Pct is shown, use PRINT or ACC key to select SPL.Pct,
press TARE key to weigh samples (which percentage is known) weight, calculate piece weight. Press
ZERO key to exit and back to percent weighing mode.
10.5.2.1 When SPL.Lo is shown, remove everything from the platter and press TARE key to confirm, before
scale is stable, SPL.Lo will be flashed. After it is stable, it will go to next step. Press ZERO key to
exit and back to percent weighing mode.
10.5.2.2 When SPL.Hi is shown, put samples (its percentage is known) onto the scale, Press TARE key to
confirm reading weight. Before scale is stable, SPL.Hi will be flashed. After it is stable, it will go to
next step. Press ZERO key to exit and back to percent weighing mode.
10.5.2.3 After
INP.PCT
being
shown,
000000(position
of
decimal
point
is
determined
by
────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
12
CONFIG-FUNC-PERCEN setting) will be displayed, use PRINT, ACC, UNIT key to input the
percentage of samples and Press TARE key to confirm. If the calculated unit-percentage-weight is
less than 0.5d, the indicator will display Pct.Er and go back to percent weighing mode, otherwise,
after the reasonable unit-percentage-weight being got, the scale will go back to percent weighing
mode. The got unit-percentage-weight can be saved after the power off and can be used next time.
11.Check Percent (percentage compare) in Percent weighing mode:
11.1
To make percentage compare function be available, CONFIG-FUNC-COMPAR menu item should set
to YES, and high and low limitation of percentage should be set correctly according to following steps:
11.2
In percent weighing mode, Press down DATA key more than 3s to input compare data of high and low.
11.3
After HIGH being shown, 000000 will be displayed, use PRINT, ACC, UNIT key to input high
percentage number and press TARE key to confirm. Annunciator of Hi will be shown in this step. Press
ZERO key to exit and back to percent weighing mode.
11.4
After Low being shown, 000000 will be displayed, use PRINT, ACC, UNIT key to input low percentage
number and press TARE key to confirm. Annunciator of Lo will be shown in this step. Press ZERO key
to exit and back to counting mode.
NOTE: If High number is 0 or is equal or less than low number, the comparison will be disabled.
11.5
After the limitations have been set and check weight function is be active, one of annunciators HI, OK,
LO will light up, and the beeper will sound according to its setting in USER-BEEP.
12. BMI Working Mode
12.1
To use the BMI mode. CONFIG-FUNC-BMI menu item should be set to YES.
12.2
To enter BMI Working mode:
8.2.1 Long press FUNC key, one of WEIGH/COUNT/PERCEN will be shown, use ACC or PRINT key to
select BMI, then press TARE to confirm go to BMI mode. When CONFIG-FUNC-ACCUMU is set to
no: press BMI (ACC) key, go to BMI mode.
12.3
Scale will show: “CM.xxx” (means: last input height is xxx cm) or “IN.xx.x” (means: last input height is
xx.x inch) will be displayed, and to wait for input height: (1)to change height unit to cm or inch by
pressing DATA key; (2) to change height number by using PRINT, ACC(BMI) keys (3)Press and hold
down PRINT or ACC(BMI) key will increase or decrease number fast, (4) Press TARE key to confirm
the input. Press ON/OFF key to exit input data mode and back to BMI working mode. The range of
height is 50-250cm(19.7-98.4inch)
12.4
and default is 170cm(66.9inch)
Press ACC key to select weight or BMI number to be displayed, when weight is displayed, the weight
unit can be selected by pressing UNIT key, and BMI and weight unit will be displayed at same time.
12.5
When current net weight is less than NLD.RNG, the indicator will go to display weight number if
CONFIG-FUNC-ACCUMU=
No;
or
the
indicator
will
back
to
original
working
mode
if
CONFIG-FUNC-ACCUMU= Yes.
────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
13
13. HOLD Function
13.1
HOLD function can be used to freeze display number. In this mode, scale can catch a dynamic number,
hold a stable number, or average a unstable number, then HOLD(freeze) this number temporary for
user to watch or record. This function can be used in normal weighing mode, counting mode and
percent weighing mode. After entering HOLD mode, the speed of A/D converter can be increased to
80Hz (if USER-HOLD-AD.H.SPD is set to YES) from original 10Hz for some dynamic weighing
applications. With the hold function, it is possible to weigh restless weighing samples such as live
animals, moving objects. The indicator provides special mode settings to accommodate sample’s
movements.
13.2
To make HOLD function be active, the CONFIG-FUNC-HOLD menu item must be set to YES; menu
items
of
USER-HOLD-HLD.MOD
/-AVG.TIM
/-HLD.TIM
/-DYN.RNG
/-STB.TIM,
USER-OTHER-NLD.RNG need be set to a value within limits.
To speedup sampling of weight, set USER-HOLD-AD.H.SPD menu item to YES.
To enter HOLD working mode, press down HOLD key when scale works in normal weighing mode,
counting mode or percent weighing mode.
13.3
There’re several HOLD mode to freeze display data:
(1) Positive Peak Number HOLD mode
(2) Negative Peak Number HOLD mode
(3) Toggle HOLD mode
(4) Average HOLD mode
(5) Auto HOLD mode
The following are details of these HOLD modes:
13.3.1 Positive Peak HOLD:
When USER-HOLD-HLD.MOD is set to PS.PEAK, the hold mode is positive peak hold mode. When
scale first enters this working mode, it will display the largest positive number that is from the time of
zero-pint set. After entering this working mode, scale will always catches and refresh positive larger
number and display it. To exit HOLD mode, press HOLD key again.
13.3.2 Negative Peak HOLD:
When USER-HOLD-HLD.MOD is set to NG.PEAK, the hold mode is negative peak hold mode. When
scale first enters this working mode, it will display the largest negative number that is from the time of
zero-point set. After entering this working mode, scale will always catches negative larger number and
display it. To exit HOLD mode, press HOLD key again.
13.3.3 Toggle HOLD:
When USER-HOLD-HLD.MOD is set to TOGGLE, the hold mode is toggle hold mode ---a manual Hold
function. After entering this working mode, scale will freeze and display number if scale is stable. Only
the weight that is over USER-OTHER-NLD.RNG (zero 'dead' band ) can be held. To exit HOLD mode,
press HOLD key again. If the time of scale being unstable is more than USER-HOLD-STB.TIM,
STB.ER will be shown, press TARE key to start averaging again, or press HOLD key to exit.
────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
14
13.3.4 Average HOLD:
When USER-HOLD-HLD.MOD is set to AVERAG, the hold mode is average hold mode. After entering
this working mode, scale will freeze and display number if scale is stable. If scale is not stable, but the
variation is less than USER-HOLD-DYN.RNG, scale will average data in USER-HOLD-AVG.TIM, then
freeze and display the number. Only the weight that is over USER-OTHER-NLD.RNG can be frozen.
Scale will exit HOLD mode according to the setting of USER-HOLD-HLD.TIM. If the time of scale
variation being over USER-OTHER-NLD.RNG is more than USER-HOLD-STB.TIM, STB.ER will be
shown, press TARE,UNIT,ACC or PRINT to start averaging again, or press HOLD key to exit.
13.3.5 Auto HOLD:
When USER-HOLD-HLD.MOD is set to AUTO, the hold mode is auto hold mode--- different subjects
can be weighed one after another without pressing any buttons. After entering this working mode, scale
will freeze and display number if scale is stable. If scale is not stable, but the variation is less than
USER-HOLD-DYN.RNG, scale will average data in USER-HOLD-AVG.TIM, then freeze and display
the number. Only the weight that is over USER-OTHER-NLD.RNG can be frozen. If the held weight is
moved away, and a new load put on the scale, scale will automatically hold new number of load. Scale
will exit HOLD mode according to the setting of USER-HOLD-HLD.TIM. If the time of scale variation
being over USER-OTHER-NLD.RNG is more than USER-HOLD-STB.TIM, STB.ER will be shown,
press TARE to start averaging again, or press HOLD key to exit.
13.4
In Positive or Negative Peak HOLD mode, the red HOLD (for LED version) or PEAK and HOLD (for
LCD version) annunciator will be lighted, in other HOLD mode, green HOLD (for LED version) or HOLD
(for LCD version) annunciator will be lighted. When HOLD annunciator flash, the displayed number is
live, When HOLD annunciator become steady, the displayed number is frozen.
14. Data Compare Function
14.1 Data compare function can be used in normal weighing mode, counting mode and percent weighing
mode, and call it as Check Weight, Check Counts and Check Percentage. When this function is enabled,
you can set a higher and a lower limitation of weight, counts or percentage independently, and these
limits can be saved permanently. Then, the current data of weight, counts or percentage will be compared
with the setting limitation, and corresponding annunciator will be lighted.
14.2 To make data compare function be available, CONFIG-FUNC-COMPAR menu item should set to YES,
and high and low limitation should be set correctly according to following steps:
14.3 In normal weighing mode, counting mode or percent weighing mode, Press down DATA key more than
3s to enter input compare data of high and low mode.
14.4 After HIGH being shown, last setting data of high will be displayed, use PRINT, ACC, UNIT key to input
new number of high and press TARE key to confirm. Annunciator of Hi will be shown in this step. Press
ZERO key to exit and back to original working mode.
14.5 After Low being shown, last setting of low will be displayed, use PRINT, ACC, UNIT key to input new
number of low and press TARE key to confirm. Annunciator of Lo will be shown in this step. Press ZERO
────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
15
key to exit and back to original working mode.
NOTE:
If High number is 0 or is equal or less than low number, the comparison will be disabled.
14.6 After a reasonable limitation is set and compare function is be active, one of annunciators HI, OK, LO will
be lighted, and the beeper will sound according to its setting in USER-BEEP.
14.7 For details, please refer to section of section5.13, 6.6 and 7.6
15. Accumulation
15.1 Accumulation function can be used in normal weighing mode, counting mode and percent weighing
mode, when this function is enabled, you can accumulate current net weight, piece, and percentage. Note,
only the load on scale is larger than USER-OTHER-NLD.RNG, the displayed positive number can be
added up. The accumulation times and total can be displayed or printed.
15.2 To make data accumulation function be available, CONFIG-FUNC-ACCUMU menu item should set to
MANUAL or AUTO, Following are details.
15.3 When CONFIG-FUNC-ACCUMU is set to MANUAL, the stable and positive displayed net weight (must
be larger than USER-OTHER-NLD.RNG), piece or percentage can be accumulated by long pressing
TOTAL key, and indicator will display accumulation times first, and then display total of number. To avoid
repeating accumulation for same load, one load only can be accumulated once. So, before a new load put
onto the scale, the original load should be removed and let load on scale be smaller than
USER-OTHER-NLD.RNG.
15.4 When CONFIG-FUNC-ACCUMU is set to AUTO, the stable and positive displayed net weight (must be
larger than USER-OTHER-NLD.RNG), piece or percentage can be accumulated automatically, and
indicator will display accumulation times first, and then display total of number. To avoid repeating
accumulation for same load, one load only can be accumulated once. So, before a new load put onto the
scale,
the
original
load
should
be
removed
and
let
load
on
scale
be
smaller
than
USER-OTHER-NLD.RNG.
15.5 To view total, when display number is zero, long pressing TOTAL key, and indicator will display
accumulation times first, and then display total of number.
NOTE:When HOLD function is enable, and scale is working in PEAK HOLD mode( CONFIG-HOLD=YES,
USER-HOLD-HLD.MOD=PS.PEAK/NG.PEAK), Accumulation function will be automatically disabled!!!
────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
16
16. View or Set Time
16.1 After entering SETUP mode (by press down SETUP key more than 3s), using PRINT or ACC key to
select MISC-TIME item, press TARE to display current time.
16.2 Time display Format is: xx:xx:xx(hh-mm-ss) , 24h format
16.3 Press down UNIT more than 3s to enter modification time mode. Use the:SETUP, PRINT, ACC, UNIT,
TARE keys to modify current time. The indicator will return to normal mode after 5 sec of inactivity.
16.4 Press SETUP key to return to last menu item, press ON/OFF key to prepare to exit this mode
17. View or Set Date
17.1 After entering SETUP mode (by press down SETUP key more than 3s), using PRINT or ACC key to
select MISC-DATE item, press TARE to display current time.
17.2 Date display Format is: xx.xx.xx(yy-mm-dd)
17.3 Press down UNIT more than 3s to enter modification date mode. Use the SETUP, PRINT, ACC, UNIT,
TARE keys to modify current date. If time of no operation is more than 5s, it will automatically exit
modification mode.
17.4 Press SETUP key to return to last menu item, press ON/OFF key to prepare to exit this mode
18. View Firmware Version
18.1 Press down SETUP until CONFIG is shown, using PRINT or ACC key to select MISC-VER item, press
TARE to display current Version.
18.2 Firmware Version display Format is:
Vxx.yy,
xx is hardware version, yy is software version
18.3 Press SETUP key to return to last menu item, press ON/OFF key to prepare to exit this mode
19. Display Test
19.1 Press down SETUP more than 3s to enter SETUP mode, using PRINT or ACC key to select
TEST-DISP.TST item, press TARE to enter test display mode and all segments will be lighted first.
19.2 In this mode, every pressing of ACC key will light next segment, every pressing of UNIT will light next
digit, press PRINT will automatically light all segments and all digits.
19.3 Press SETUP key to return to last menu item, press ON/OFF key to prepare to exit this mode
20. Keyboard and Buzzer Test
20.1 Press down SETUP more than 3s to enter SETUP mode, using PRINT or ACC key to select
TEST-key.tst item, press TARE to enter test keypad mode, and key. -- will be displayed.
20.2 In this mode, press a key, the value of this key will be displayed on -- position and buzzer will beep
whatever USER-BEEP-KEY item is set.
20.3 Press SETUP key to return to last menu item, press ON/OFF key to prepare to exit this mode
────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
17
21. Enter Setup Mode:
21.1 If need to set configuration parameters, set user parameters, calibrate the scale, set current date or time,
test some hardware…
It’s need to long press SETUP key to enter setup mode
21.2 After Entering Setup Mode, the main menu item CONFIG will be shown first.
21.3 In Setup mode, use
to select wanted submenu and wanted
menu item, select wanted choice, set wanted number, confirm and save data, or exit this mode. Please
refer the “4.Operation Menu Structure”.
22. UPS world ship:
22.1. To interface the scale with UPS world ship set the following parameters:
Enter USER menu, set following parameters, if RS232 is used, set these on COM1, if USB is used, set
these on COM2:
USER-COM1(or COM2)-BAUD.RT = 4800
USER-COM1(or COM2)-BYT.FMT = 7E1
USER-COM1(or COM2)-LAYOUT = SCP-12
22.2. Set scale port to NCI3835 in UPS WorldShip:
23. Error codes and Troubleshooting
25.1 Meaning of Symbols:
25.1.1
0﹉ ﹉ ﹉ ﹉ ------ Zero is over the setting range
25.1.2
0﹍ ﹍ ﹍ ﹍ ------ Zero point is below the setting range
25.1.3
Ad﹉ ﹉ ﹉ ------ Signal to ADC is over max. range)
25.1.4
Ad﹍ ﹍ ﹍ ------ Signal to ADC is below min. range
25.1.5
﹉﹉﹉﹉
------ Weight is over upper limitation, or display data is over limitation
25.1.6
﹍﹍﹍﹍
------ Weight is below lower limitation
25.1.7
EEP.E1
------ CONFIG or CAL parameters are not correctly set
25.1.8
EEP.E2
------ USER parameter is not correctly set
25.1.9
Lo.bAt
------ Battery voltage is lower than setting.
25.1.10 CAP.- - -
----
Next displaying content is Capacity
25.1.11 CAP.ER
-----
Parameters about Capacity is not correct
25.1.12 CAL.Px
-----
Calibration on point(x)
25.1.13 CAL.OFF ----25.1.14 CAL.ON
25.1.15 CAL.Er
---------
Calibration Seal Switch is on OFF position
Calibration Seal Switch is on ON position
Calibration error, maybe input data or loaded weight is incorrect, unstable, un-linear
25.1.16 CAL.End -----
Calibration is end
25.1.17 OFF
-----
Power OFF the indicator
25.1.18 STB.ER
------ Unstable time is lager than setting of USER-OTHER-NLD.RNG
25.1.19 AC.xxxx
----- Accumulation times is xxxx
────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
18
25.1.20 PR.TARE
----- To Preset TARE weight
25.1.21 COMP
---- To go to input COMPARE data mode
25.1.22 HIGH
----- To input HIGH limitation data of Comparison
25.1.23 LOW
----- To input LOW limitation data of Comparison
25.1.24 SPL.Lo ---------- Sample load weight of low point.
25.1.25 SPL.HI ----------- Sample load weight of high point.
25.1.26 SPL.PWT ------- Sample goods weight to calculate piece weight
25.1.27 INP.PCS --------- input pieces number of weighted goods
25.1.28 UNIT.KG --------- Unit kg is selected
25.1.29 UNIT.LB --------- Unit lb is selected
25.1.30 PWT.ER --------- Piece weight is error, it’s too small (<0.5d).
25.1.31 SPL.PCT -------- Sample goods weight to calculate
25.1.32 INP.PCT --------- input percentage of weighted goods
25.1.33 PCT.ER ---------- Unit-Percentage -Weight is too small (<0.5d).
────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
19
24. Troubleshooting
SYMPTOM
Ad﹉ ﹉ ﹉
Ad﹍ ﹍ ﹍
0﹉ ﹉ ﹉ ﹉
0﹍ ﹍ ﹍ ﹍
PROBABLE
Load
cell
wires
CAUSE
to
indicator
REMEDY
are
Make sure wires are ok and correctly
incorrectly connected, or shorted, or
connected. Replace load cell or ADC
opened; or ADC, load cell are damaged
chip, Service required.
Weight reading exceeds Power On Zero
limit.
Perform zero calibration.
Weight reading below Power On Zero
zero calibration.
or The weight value cannot be displayed
in the current unit of measure because it
exceeds 6 digits..
﹍﹍﹍﹍
EEP.E1
EEP.E2
Install platform on scale. Perform
limit.
Weight reading exceeds Overload limit,
﹉﹉﹉﹉
Make sure scale platform is empty.
Reduce load on scale until weight
value can be displayed. Use a more
appropriate unit of measure. Re-set
some parameters of CONFIG or
UAER.
Weight reading below Under load limit.
CONFIG or CAL parameters are not
correctly set
Install platform on scale.
Perform zero calibration
Re-set items in CONFIG, do
calibration
USER parameter is not correctly set
Re-set items in USER
Set PRIM.N/PRIM.d/SECND.n to
CAP.ER
Capacity parameters are not correct
correct number, make sure capacity
not more than 6 digit
CAL.Er
Calibration error, maybe input data or
Input correct data, load correct
loaded weight is too small, too big,
weight onto platform, Service
unstable, un-linear
required
Piece weight is error, it’s too small
PWT.ER
(<0.5d), The weight on the platform is
too small to define a valid reference
Use a greater weight for the sample.
weight.
Unit-Percentage -Weight is error, it’s too
small (the weight of 1%, 0.1%, or
PCT.ER
0.01%-determined by
Use a greater weight for the sample.
CONFIG-FUNC-PERCNT is less than
0.5d)
────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
20
USER-HOLD-STB.TIM
STB.ER
is
too
short,
USER-HOLD-HLD.RNG is too small,
other failure
or set USER-HOLD-HLD.RNG
bigger. Service required
Power cord not plugged in or properly
connected. Power outlet not supplying
Not turn on.
Set USER-HOLD-STB.TIM longer,
electricity. Battery discharged. Other
failure.
Check power cord connections.
Make sure power cord is plugged
into the power outlet. Check power
source. Replace batteries. Service
required.
Cannot zero the
Load on scale exceeds allowable limits.
Remove load on scale.
display or will not zero
Load on scale is not stable.
Wait for load to become stable.
when turned on.
Load cell damage.
Service required.
Cannot display weight
in desired weighing
unit.
Unit not set to enable, or d≥5oz,when
unit is lb:oz
Enable unit in CONFIG-UNITS
Battery symbol is
empty or Lo.bAt
is
Batteries are discharged.
Change batteries
shown
────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
21
25. Display Character
ASCII
LCD/LED Show
ASCII
LCD/LED Show
ASCII
0
A
N
1
B
O
2
C
P
3
D
Q
4
E
R
5
F
S
6
G
T
7
H
U
8
I
V
9
J
W
K
X
L
Y
M
Z
LCD/LED Show
────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
22
26. Packing List
No.
CONTENT
QTY
1
Indicator
1
2
User manual
1
3
Swivel bracket
1
4
AC Adapter
1
5
RS232 cable
1
6
USB interface
1
7
USB cable
1
9
1x CR2032 battery for RTC
1
10
AA Alkaline batteries
4
────────────────────────────────────────────────────────
KILOTECH INC.
www.kilotech.com
23
Was this manual useful for you? yes no
Thank you for your participation!