Extech Instruments MN62-K True RMS Multimeter User manual

Add to my manuals
19 Pages

advertisement

Extech Instruments MN62-K True RMS Multimeter User manual | Manualzz

USER GUIDE 

 

 

True RMS Digital Multimeter 

 

With Non‐Contact Voltage Detector and Low Pass Filter 

Model MN62 

Introduction 

 

 

 

Thank you for selecting the Model MN62 True RMS Auto‐ranging Multimeter. This meter measures 

AC/DC  Voltage,  AC/DC  Current,  Variable  Frequency  Voltage  (VFC),  Resistance,  Capacitance, 

Frequency,  Duty  Cycle  (AC  only),  Diode  Test,  and  Continuity;  a  Non‐Contact  Voltage  Detector  is  also built‐in. This device is shipped fully tested and calibrated and, with proper use, will provide  years of reliable service. Please visit our website (www.extech.com) to check for the latest version  of this User Guide, Product Updates, Product Registration, and Customer Support. 

Features 

True RMS AC Measurements 

Non‐contact Voltage Detector 

 4000 Count Multifunction LCD Display 

 Automatic and Manual Ranging 

 MAX‐MIN Reading Memory 

Relative Mode 

Data Hold 

 Variable Frequency Voltage mode (VFC) offers low pass filter 

 Automatic Power OFF (APO) 

 Rugged, Battery Operated, and Portable 

2

 MN62‐en‐GB_V1.1   12/15 

Safety 

The MN62 is designed in compliance with standards such as IEC/EN61010‐1, pollution grade II,  overvoltage CAT III 600V and double insulation standards. Please comply with the instructions  provided in this guide otherwise the protection provided by the instrument may be affected. 

This symbol adjacent to another symbol, terminal or operating device indicates  that the operator must refer to an explanation in the Operating Instructions to  avoid personal injury or damage to the meter. 

WARNING

This WARNING symbol indicates a potentially hazardous situation, which if not  avoided, could result in death or serious injury. 

CAUTION

MAX

600V

This CAUTION symbol indicates a potentially hazardous situation, which if not  avoided, may result damage to the product. 

This symbol advises the user that the terminal(s) so marked must not be  connected to a circuit point at which the voltage with respect to earth ground  exceeds (in this case) 600V AC rms or DC

This symbol adjacent to one or more terminals identifies them as being  associated with ranges that may, in normal use, be subjected to particularly  hazardous voltages. For maximum safety, the meter and its test leads should  not be handled when these terminals are energized. 

This symbol indicates that a device is protected throughout by double  insulation or reinforced insulation. 

 

PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY 

OVERVOLTAGE CATEGORY I 

Equipment  of  OVERVOLTAGE  CATEGORY  I  is  equipment  for  connection  to  circuits  in  which  measures are taken to limit the transient overvoltages to an appropriate low level. 

  Note – Examples include protected electronic circuits. 

OVERVOLTAGE CATEGORY II 

Equipment  of  OVERVOLTAGE  CATEGORY  II  is  energy‐consuming  equipment  to  be  supplied  from  the fixed installation. 

  Note – Examples include household, office, and laboratory appliances. 

OVERVOLTAGE CATEGORY III 

Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations. 

  Note – Examples include switches in the fixed installation and some equipment for industrial use  with permanent connection to the fixed installation. 

OVERVOLTAGE CATEGORY IV 

Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation. 

Note – Examples include electricity meters and primary over‐current protection equipment 

3

 MN62‐en‐GB_V1.1   12/15 

 

 

CAUTIONS 

 Improper use of this meter can cause damage, shock, injury or death. Read and understand this user  manual before operating the meter. 

 Always remove the test leads before replacing the battery or fuses. 

 Inspect the condition of the test leads and the meter itself for any damage before operating the meter.  

 Use great care when making measurements if the voltages are greater than 30VAC rms or 60VDC. These  voltages are considered a shock hazard. 

 Always discharge capacitors and remove power from the device under test before performing Diode, 

Resistance or Continuity tests. 

 Voltage checks on electrical outlets can be difficult and misleading because of the uncertainty of  connection to the recessed electrical contacts. Other means should be used to ensure that the terminals  are not "live". 

 If the equipment is used in a manner not specified by the manufacturer, the protection provided by the  equipment may be impaired. 

 This device is not a toy and must not reach children’s hands. It contains hazardous objects as well as small  parts that the children could swallow. 

 In case the device is going to be unused for an extended period of time, remove the batteries and store  them separately. 

 Expired or damaged batteries can cause cauterization on contact with the skin. Use protective gloves  when handling. 

 Do not short‐circuit the batteries. Do not throw batteries into a fire. 

This meter has been designed for safe use, but must be operated with caution. The rules listed below must  be carefully followed for safe operation. 

1. NEVER apply voltage or current to the meter that exceeds the specified maximum: 

Function 

V DC or V AC 

µA/mA AC/DC 

A AC/DC 

Input Protection Limits

Maximum Input

600VDC/AC rms

500mA 600V fast acting fuse

10A 600V fast acting fuse

Frequency, Resistance, Capacitance, Duty Cycle, 

Diode Test, Continuity 

600VDC/AC rms

2. USE EXTREME CAUTION when working with high voltages. 

3. DO NOT measure voltage if the voltage on the "COM" input jack exceeds 600V above earth ground. 

4. NEVER connect the meter leads across a voltage source while the function switch is in the current,  resistance, or diode mode. Doing so can damage the meter. 

5. ALWAYS discharge filter capacitors in power supplies and disconnect the power when making resistance  or diode tests. 

6. ALWAYS turn off the power and disconnect the test leads before opening the covers to replace the fuse  or batteries. 

7. NEVER operate the meter unless the back cover and the battery and fuse covers are in place and fastened  securely. 

8. NEVER change the measurement function by turning the function dial while testing. 

9. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the  equipment may be impaired. 

4

 MN62‐en‐GB_V1.1   12/15 

Meter Description 

1. Non‐Contact Voltage Detector LED lamp 

2. LCD Display 

3. MAX‐MIN 

4. RANGE button 

1

2

5. HOLD‐Backlight   button 

6. Function switch 

7. ‘A’ probe input jack 

8. ‘mAµA’ probe input jack 

9. COM (‐) input jack 

3

4

5

 

10. Positive (+) input jack  6

11. SELECT‐Variable Frequency Control button 

12. HZ‐% button 

7

8

13. REL (Relative) button 

 

Note: Tilt stand, test lead holders, and battery/fuse compartment are on rear of unit. 

Button Descriptions

 

13

12

11

10

9

RANGE button 

The default mode is Auto Range. To manually range, press RANGE to step through the ranges.  

Press and hold RANGE to return to Auto Range mode. Auto Range is displayed in Auto mode. 

MAX‐MIN button 

Press to activate the MAX/MIN recording mode (MAX will appear). The meter will display and  hold the maximum reading. Press again (MIN will appear). The meter will display and hold the  minimum reading. To exit, press and hold MAX‐MIN for 2 seconds. 

REL (Relative) button 

The displayed value, at the time of the button press, becomes the Reference value with which  subsequent readings are compared. See dedicated Relative mode section. 

HOLD and Backlight   button   

Short press to freeze displayed reading (H icon is visible). Press again to release reading. 

Press and hold to switch backlight ON or OFF. 

Hz‐% button 

Frequency (Hz) and Duty Cycle (%) modes. Available only in the AC Voltage/Current and 

Frequency modes. 

SELECT‐VFC button 

Short presses select AC or DC for voltage and current measurements. 

In AC Voltage VFC mode, press and hold to access VFC (Variable Frequency Control) where a  low pass filter is employed for measuring the voltage of a variable frequency/speed drive. 

5

 MN62‐en‐GB_V1.1   12/15 

Display Description 

1

2

3

4

5

6

7

89 10 11

22

21

19

18

17

20

VFC

12

13

16

1. Relative mode 

2. See units prefix list below 

15

14

3. Ohms (Resistance) 

4. Hertz (Frequency) 

5. See units prefix list below 

6. Voltage 

7. See units prefix list below 

8. Amperes (Amps) Current 

9. Farads (Capacitance) 

10. Percent (%) duty cycle (for AC only) 

11. Battery status icon 

12. APO (Auto Power OFF) active  

13. Measurement reading digits 

14. Non‐contact voltage detector (displays EF in reading area) 

15. Auto Range is shown in Auto Range mode 

16. MAX‐MIN memory active 

17. Data Hold active 

18. Continuity test (Resistance) 

19. Diode test 

20. Minus (negative) sign 

21. AC (alternating current) or DC (direct current) mode 

22. VFC (low pass filter) Variable Frequency Control 

Measurement unit prefixes 

n  nano (10

‐9

) (capacitance) 

µ  micro (10

‐6

) (amps, cap)  m  milli (10

‐3

) (volts, amps)    k  kilo (10

3

) (ohms)   

M  mega (10

6

) (ohms) 

6

 

 MN62‐en‐GB_V1.1   12/15 

Operating Instructions 

WARNING: Risk of electrocution. High‐voltage circuits, both AC and DC, are very dangerous and should be  measured with great care. 

1. ALWAYS turn the function switch to the OFF position when the meter is not in use. 

2. If OL appears in the display during a measurement, the value exceeds the selected range. 

Change to a higher range.  

3. If the meter does not switch on when the function dial is turned to any position other than 

OFF, please check the batteries. Refer to the Battery Installation section of this guide. 

AUTO POWER OFF (APO) 

The APO feature turns the meter off after 15 minutes of inactivity. To disable APO, hold the 

SELECT button while turning the meter on. The meter will emit 5 beeps indicating that APO has  been disabled (the APO clock icon will not be visible on the display). Turn the meter off and then  on again to re‐enable APO (the APO clock icon will be visible on the display when APO is active).  

AUTOMATIC and MANUAL Range 

The meter defaults to AUTO Range mode when it is switched ON. In this mode the meter  automatically selects the best range for the measurements being made and is generally the best  mode for most measurements. For measurement situations requiring that a range be manually  selected, perform the following: 

1. Press the RANGE button. The “AUTO RANGE” display indicator will turn off. 

2. Press the RANGE button to step through the available ranges. 

3. To exit the Manual Ranging mode and return to Auto ranging, press and hold the RANGE  button for 2 seconds. 

MAX/MIN Memory 

1. Press the MAX/MIN button to activate the MAX/MIN recording mode. The display icon MAX  will appear. The meter will display and hold the maximum reading and will update only when  a new maximum reading occurs. 

2. Press the MAX/MIN button again and the display icon MIN will appear. The meter will  display and hold the minimum reading and will update only when a new minimum reading  occurs. 

3. To exit MAX/MIN mode press and hold the MAX/MIN button for 2 seconds. 

Relative mode 

The relative measurement feature allows the user to make measurements relative to a stored  reference value. A reference voltage, current, etc. can be stored where subsequent  measurements can be made in comparison to that stored reference value. The displayed value is  the difference between the reference value and the measured value.   

1. Perform a measurement as described in the operating instructions. 

2. Press the REL button to store the reading; the Relative indicator will appear on the display. 

3. The display will now indicate the difference between the stored and measured value. 

4. Press the REL button to exit the relative mode. The Relative indicator will switch OFF. 

7

 MN62‐en‐GB_V1.1   12/15 

Display Backlight 

Press and hold the    button to turn the backlight ON/OFF. The backlight will automatically turn  off after 30 seconds. 

Data Hold 

The Data Hold function freezes the reading in the display. Press the HOLD button to switch Data 

Hold ON/OFF. When the Data Hold mode is active the display will show the H icon. 

DC VOLTAGE MEASUREMENTS  

CAUTION:  Do not measure DC voltages if a motor on the circuit is being switched ON or OFF. Large voltage  surges may occur that can damage the meter. 

1. Set the function switch to the position.  

2. Press the SELECT button to switch to DC if necessary. 

3. Insert the black test lead banana plug into the negative COM jack. 

Insert the red test lead banana plug into the positive V jack. 

4. Touch the black test probe tip to the negative side of the circuit.  

Touch the red test probe tip to the positive side of the circuit. 

5. Use the RANGE button to manually range the reading. Press and  hold the RANGE button to return to Auto Range Mode. Read the  voltage in the display.  

 

AC VOLTAGE MEASUREMENTS 

WARNING:  Risk of Electrocution. The probe tips may not be long enough to contact the live parts inside  some 240V outlets for appliances because the contacts are recessed deep in the outlets. As a result, the  reading may show 0 volts when the outlet actually has voltage on it. Make sure the probe tips are touching  the metal contacts inside the outlet before assuming that no voltage is present. 

 

CAUTION:  Do not measure AC voltages if a motor on the circuit is being switched ON or OFF. Large voltage  surges may occur that can damage the meter. 

1. Set the function switch to the V position. 

2. Press the SELECT button to switch to AC if necessary. 

3. Insert the black test lead banana plug into the negative COM jack. 

Insert red test lead banana plug into the positive V jack. 

4. Touch the black test probe tip to the neutral side of the circuit. 

Touch the red test probe tip to the “hot” side of the circuit. 

5. Read the voltage in the display.  

6. Use the RANGE button to manually range the reading. Press and  hold the RANGE button to return to Auto Range Mode. 

8

 MN62‐en‐GB_V1.1   12/15 

AC VOLTAGE MEASUREMENTS with VFC, Hz, and % Duty Cycle 

WARNING:  Risk of Electrocution. The probe tips may not be long enough to contact the live parts inside  some 240V outlets for appliances because the contacts are recessed deep in the outlets. As a result, the  reading may show 0 volts when the outlet actually has voltage on it. Make sure the probe tips are touching  the metal contacts inside the outlet before assuming that no voltage is present. 

 

CAUTION:  Do not measure AC voltages if a motor on the circuit is being switched ON or OFF. Large voltage  surges may occur that can damage the meter. 

1. Set the function switch to the VFC position.  

2. Insert the black test lead banana plug into the negative COM jack. 

Insert red test lead banana plug into the positive V jack. 

3. Press and hold the VFC button until the meter beeps and VFC is  briefly displayed. The meter is now applying a low pass filter (Note: 

VFC is Variable Frequency Control). Press and hold the VFC button  again to exit the VFC mode (END will briefly display). 

4. Touch the black test probe tip to the neutral side of the circuit. 

Touch the red test probe tip to the “hot” side of the circuit. 

5. Read the voltage in the display.  

6. Press the Hz/% button to indicate Hz

7. Read the frequency in the display. 

8. Press the Hz/% button again to indicate %

9. Read the % duty cycle in the display. 

%  Duty  Cycle  Note:  Duty  Cycle  measurements  are  AC  only  for  this  meter.  Measurement  signals must be ‘zero crossing’ and must reach ‐0.7V for the meter to properly display Duty 

Cycle %. 

9

 MN62‐en‐GB_V1.1   12/15 

DC/AC MILLIVOLT MEASUREMENTS  

CAUTION:  Do not measure DC/AC voltages if a motor on the circuit is being switched ON or 

OFF. Large voltage surges may occur that can damage the meter. 

1. Set the function switch to the mV position.  

2. Insert the black test lead banana plug into the negative COM jack. 

Insert the red test lead banana plug into the positive V jack. 

3. Press the SELECT button to select DC or AC millivolts. 

4. Touch the black test probe tip to the negative side of the circuit.  

Touch the red test probe tip to the positive side of the circuit. 

(Figure at right shows DC measurement only). 

5. Read the voltage in the display.  

6. For AC, press the Hz/% button to indicate Hz

7. Read the frequency in the display. 

8. For AC, press the Hz/% button again to indicate %

9. Read the % duty cycle in the display. 

DC CURRENT MEASUREMENTS 

CAUTION:  Do not make 5A current measurements for longer than 30 seconds. Allow 15  minutes between tests. Exceeding 30 seconds may cause damage to the meter and/or the test  leads. 

1. Insert black test lead banana plug into the negative COM  jack. 

2. For current measurements up to 6000µA DC, set the  function switch to the µA position and insert the red test  lead banana plug into the µA/mA jack. 

3. For current measurements up to 600mA DC, set the function  switch to the mA position and insert the red test lead  banana plug into the µA/mA jack. 

4. For current measurements up to 10A DC, set the function  switch to the A position and insert the red test lead banana  plug into the 10A jack. 

5. Press the SELECT button to indicate DC on the display. 

6. Remove power from the circuit under test, then open up the  circuit at the point where you wish to measure current. 

7. Touch the black test probe tip to the negative side of the circuit. Touch the red test probe  tip to the positive side of the circuit. 

8. Apply power to the circuit. 

9. Read the current in the display.  

10

 MN62‐en‐GB_V1.1   12/15 

 

 

AC CURRENT (FREQUENCY, DUTY CYCLE) MEASUREMENTS 

CAUTIONS: Do not make 5A current measurements for longer than 10 seconds. Allow 15  minutes between tests. Exceeding 10 seconds may cause damage to the meter and/or the test  leads.  

Before connecting the meter in a series circuit turn off the power supply to the series loop. 

Start with the highest range possible if unsure of expected current measurement. 

Fuses are provided in the 10A and µA/mA input jacks. Do not connect test leads in parallel to  any circuit when test leads are connected to these jacks. 

Pressing the HZ‐% button, while taking AC current measurements, is allowed. 

1. Insert the black test lead banana plug into the negative COM jack. 

2. For current measurements up to 6000µA AC, set the  function switch to the µA position and insert the red  test lead banana plug into the µA/mA jack. 

3. For current measurements up to 600mA AC, set the  function switch to the mA position and insert the red  test lead banana plug into the µA/mA jack. 

4. For current measurements up to 10A AC, set the  function switch to the A position and insert the red  test lead banana plug into the 10A jack. 

5. Press the SELECT button to indicate “AC” on the  display. 

6. Remove power from the circuit under test, then open  up the circuit at the point where you wish to measure  current. 

7. Touch the black test probe tip to the neutral side of  the circuit.  

Touch the red test probe tip to the “hot” side of the circuit. 

8. Apply power to the circuit. 

9. Read the current in the display. 

10. Press the Hz/% button to indicate “Hz”. 

11. Read the frequency in the display. 

12. Momentarily press the Hz/% button again to indicate “%”. 

13. Read the % duty cycle in the display. 

14. Press the Hz/% button to return to the current‐only measurement. 

%  Duty  Cycle  Note:  Duty  Cycle  measurements  are  AC  only  for  this  meter.  Measurement  signals must be ‘zero crossing’ and must reach ‐0.7V for the meter to properly display Duty 

Cycle %. 

11

 MN62‐en‐GB_V1.1   12/15 

RESISTANCE MEASUREMENTS 

WARNING: To avoid electric shock, disconnect power to the unit under test and discharge all  capacitors before taking resistance measurements. Remove batteries and unplug line cords. 

1. Set the function switch to the Ω    position. 

2. Insert the black test lead banana plug into the negative COM jack. 

Insert the red test lead banana plug into the positive Ω jack. 

3. Press the SELECT button to indicate Ω on the display. 

4. Touch the test probe tips across the circuit or part under test. It is  best to disconnect one side of the part under test so the rest of the  circuit will not interfere with the resistance reading. 

5. Read the resistance in the display. 

Resistance Measurement Notes 

The display shows OL when the measured resistance open‐circuit or  resistance value exceeds maximum range. 

When measuring low resistance, a test‐lead measurement error of 0.1 to 

0.2

  will exist. Short the test leads and use the Relative mode to null the test‐lead error. 

Check the test leads for loosening or other breakage and verify that the lead resistance is 0.5

   or less by shorting the leads and viewing the displayed reading. Replace test leads if necessary. 

Several seconds are required for stability when taking high resistance measurements. 

Do not input signals > 60VDC or 30VAC. 

 

CONTINUITY CHECK 

WARNING: To avoid electric shock, never measure continuity on circuits or wires that have  voltage on them. 

1. Set the function switch to the Ω          position. 

2. Insert the black lead banana plug into the negative COM jack. 

Insert the red test lead banana plug into the positive Ω jack. 

3. Press the SELECT button to indicate Ω on the display 

4. Touch the test probe tips to the circuit or wire you wish to check.  

5. If the resistance is > 150Ω there will be a break in the circuit but no  beeper will sound. If the resistance is  ≤  10Ω, the beeper signal will  sound. If the circuit is open, the display will indicate “OL”. 

Continuity Measurement Notes 

 Switch OFF power to any devices on the measurement circuit 

 Do not input signals > 60VDC or 30VAC 

12

 MN62‐en‐GB_V1.1   12/15 

 

 

DIODE TEST 

1. Set the function switch to the Ω          position. 

2. Insert the black test lead banana plug into the negative COM  jack and the red test lead banana plug into the positive V jack. 

3. Press the SELECT button to indicate   and V on the display. 

4. Touch the test probes to the diode under test. Forward voltage  will typically indicate 0.400 to 0.700V. Reverse voltage will  indicate “OL”. Shorted devices will indicate near 0V and an open  device will indicate “OL” in both polarities. 

 

Diode Measurement Notes 

Switch OFF power to any devices on the measurement circuit 

Do not input signals > 60VDC or 30VAC 

CAPACITANCE MEASUREMENTS 

WARNING: To avoid electric shock, disconnect power to the unit under test and discharge all  capacitors before taking any capacitance measurements. Remove the batteries and unplug the  line cords. 

1. Set the rotary function switch to the Capacitance symbol position. 

2. Insert the black test lead banana plug into the negative COM jack. 

Insert the red test lead banana plug into the positive V jack. 

3. Touch the test leads to the capacitor to be tested. 

4. Read the capacitance value in the display 

 

Capacitance Measurement Notes 

The meter will display a fixed reading with no input connected, this is  the internal fixed capacitance; please use the Relative mode to null the  effect of the internal fixed capacitance (especially for low capacitance  measurements). 

The display will show OL when the capacitor is short‐circuited or if the  capacitance value exceeds the maximum range of the meter. 

High capacity capacitors require a longer time period to stabilize than  lower capacity devices. 

To prevent damage to the instrument and personal injury, release all  residual charges to devices under test before taking measurements.  

  

13

 MN62‐en‐GB_V1.1   12/15 

 

 

 

FREQUENCY, DUTY CYCLE MEASUREMENTS (ELECTRONIC) 

1. Set the rotary function switch to the Hz/% position. 

2. Insert the black lead banana plug into the negative COM jack and  the red test lead banana plug into the positive Hz jack. 

3. Touch the test probe tips to the circuit under test. 

4. Read the frequency on the display. 

5. Press the Hz/% button to indicate “%”. 

6. Read the % duty cycle in the display. 

 

 

  meter. Measurement signals must be ‘zero crossing’ and must reach 

 

 

 

 

 

 

Frequency (Electronic) Measurement Note: Do not input signals > 60VDC or 30VAC 

NON‐CONTACT VOLTAGE DETECTOR 

WARNING: To avoid electric shock, do not use this meter in the Non‐Contact Voltage Detector  mode if the meter does not power up when the function dial is set to the NCV position. 

1. Set the rotary function switch to the NCV position. 

2. Hold the meter top near to a source of AC voltage or  electromagnetic field. The test leads are not necessary in  the NCV mode. 

3. The front LED lamp will light and the beeper will sound  when an AC voltage or electromagnetic field is sensed. 

The beeper sound changes with changing intensity of the  detected energy source. 

4. The display shows EF (Electromotive Force) in this mode. 

14

 MN62‐en‐GB_V1.1   12/15 

Maintenance 

 

WARNING:  To avoid electric shock, disconnect the test leads from any source of voltage before  removing the back cover or the battery or fuse covers. 

WARNING: To avoid electric shock, do not operate your meter until the battery and fuse covers  are in place and fastened securely. 

This meter is designed to provide years of dependable service, if the following care instructions are  performed: 

1. KEEP THE METER DRY. If it gets wet, wipe it off. 

2. USE AND STORE THE METER IN NORMAL TEMPERATURES. Temperature extremes can  shorten the life of the electronic parts and distort or melt plastic parts. 

3. HANDLE THE METER GENTLY AND CAREFULLY. Dropping it can damage the electronic parts  or the case. 

4. KEEP THE METER CLEAN. Wipe the case occasionally with a damp cloth. DO NOT use  chemicals, cleaning solvents, or detergents. 

5. USE ONLY FRESH BATTERIES OF THE RECOMMENDED SIZE AND TYPE. Remove old or weak  batteries so they do not leak and damage the unit. 

6. IF THE METER IS TO BE STORED FOR A LONG PERIOD OF TIME, the batteries should be  removed to prevent damage to the unit. 

BATTERY INSTALLATION 

WARNING: To avoid electric shock, disconnect the test leads from any source of voltage before removing the battery cover.

1. Turn power off and disconnect the test leads from the meter. 

2. Open the rear battery cover (behind the tilt stand) by removing the screw using a Phillips  head screwdriver. 

3. Replace the two (2) 1.5V AA batteries in the battery holder, observing correct polarity. 

4. Put the battery cover back in place. Secure with the screw. 

 

Never dispose of used batteries or rechargeable batteries in household waste. 

As consumers, users are legally required to take used batteries to appropriate collection  sites, the retail store where the batteries were purchased, or wherever batteries are sold. 

Disposal: Do not dispose of this instrument in household waste. The user is obligated to take  end‐of‐life  devices  to  a  designated  collection  point  for  the  disposal  of  electrical  and  electronic equipment.  

 

WARNING: To avoid electric shock, do not operate the meter until the battery cover is in place  and fastened securely. 

 

NOTE: If the meter does not work properly, check the fuses and batteries to make sure that they  are still good and that they are properly inserted.  

15

 MN62‐en‐GB_V1.1   12/15 

FUSE REPLACMENT 

WARNING: To avoid electric shock, disconnect the test leads from any source of voltage before  removing the meter cover. 

1. Disconnect the test leads from the meter. 

2. Remove the meter’s protective rubber holster (jacket). 

3. Loosen the center Phillips head screw on the back of the meter while pulling the battery  compartment cover away from the meter. When the screw is loosened sufficiently, the  battery compartment cover can be removed. Refer to diagram below. 

4. Remove the batteries. 

5. Fuse F1 will be visible (directly below the battery compartment); replace fuse F1: FF0.5A H 

600V (CE) (ϕ6x32mm). 

6. Remove the remaining two Phillips head screws located on the back of the meter near the  bottom. 

7. Open the meter housing by pulling the two halves of the meter assembly apart. 

8. Fuse F2 will be now be visible (directly below F1); replace fuse F2: F10A H 600V (CE) 

(ϕ6x25mm). 

9. When replacing fuses, gently remove the old fuse and install the new fuse into the holder. 

Always use a fuse of the proper size and value. 

10. Re‐assemble and secure the meter housing, batteries, and battery cover before attempting  to use the meter. 

WARNING: To avoid electric shock, do not operate the meter until the meter housing and  battery compartment cover are fastened securely. 

16

 MN62‐en‐GB_V1.1   12/15 

Specifications 

Specifications stated for ambient conditions 23 o C ±5 o C (73.4

o F ±9 o F); Relative Humidity < 75% 

Electrical Specifications 

 

Function 

DC Voltage 

 

AC Voltage 

DC Current 

AC Current 

  

Range 

40.00mV 

400.0mV 

4.000V 

Resolution 

10V 

0.1mV

1mV

40.00V 

400.0V 

10mV

0.1V

600V 

DCV Notes:

40.00mV 

400.0mV 

1V

10M Input Impedance 

Max input voltage ±600V 

10V 

0.1mV

4.000V 

40.00V 

1mV

10mV

Accuracy

(0.7% of reading + 5 digits) 

(0.5% of reading + 2 digits) 

(0.7% of reading + 3 digits) 

(1.0% reading + 3 digits) 

(0.8% reading + 3 digits) 

400.0V 

600V 

VFC 200.0~600V

0.1V

1V

0.1/1V

(1.0% reading + 3 digits) 

(4.0% reading + 3 digits) 

ACV Notes:

10M Input Impedance 

Max input voltage ±600Vrms 

Frequency response: 45~1kHz (for VFC: 45~400Hz) 

True RMS 

All AC voltage ranges are specified from 5% of range to 100% of range 

Crest Factor: ≤3 at full scale up to 400V, decreasing linearly to ≤1.5 at 600V 

400.0A 

4000A 

40.00mA 

0.1A 

1A 

(0.7% reading + 2 digits) 

400.0mA 

4.000A 

10.00A 

DCA notes: 

10A 

0.1mA

1mA

10mA

(1.0% reading + 3 digits) 

Overload protection for A/mA range: F1 fuse (ϕ6x32)mm; FF0.5A H 600V (CE)

Overload protection for 10A range: F2 fuse (ϕ6x25)mm F 10A H 600V (CE) 

400.0A 

4000A 

0.1A 

1A 

(1.0% reading + 3 digits) 

40.00mA  10A 

400.0mA 

4.000A 

0.1mA

1mA

(1.2% reading + 3 digits) 

10.00A 

ACA notes: 

10mA

True RMS 

Frequency response: 45~1kHz 

All AC voltage ranges are specified from 5% of range to 100% of range 

Overload protection same as DC Current overload protection (see above) 

17

 MN62‐en‐GB_V1.1   12/15 

Function 

Resistance 

Capacitance 

Range 

400.0 

Resolution 

0.1 

Accuracy

(1.0% reading + 2 digits) 

4.000k 

40.00k 

1 

10 

(0.8% reading + 2 digits) 

400.0k  100 

4.000M  1k 

40.00M 

Resistance notes:

10k 

Open circuit voltage 1V approx. 

Overload protection: 600V 

9.999nF

1pF

(1.2% reading + 3 digits) 

(1.5% reading + 5 digits) 

99.99nF~999.9µF

9.999mF~99.99mF

10pF~0.1µF

1µF~10µF

(Relative) (4.0% reading + 10 digits) 

(4.0% reading + 5 digits) 

(10% reading)  

Capacitance notes:

Overload protection: 600V 

For capacitor measurements ≤ 1µF use Relative measurements to ensure measurement  accuracy

9.999Hz~9.999MHz

0.001Hz~0.001MHz 

(0.1% reading + 4 digits) 

 

Frequency 

(electronic) 

Duty Cycle 

Continuity 

Diode 

1%~99% 0.1%

Frequency and Duty Cycle notes:

Overload protection 600V 

Input range sensitivity:  

≤ 100kHz: 100mVrms 

Not defined

>100kHz: 200mVrms 

>1MHz: 500mVrms 

5MHz~10MHz: 900mVrms 

Duty Cycle %  is only applicable for zero crossing AC measurements ≤ 100kHz 

Beeper sounds when resistance is ≤ 10. Display resolution: 0.1; 

Overload protection: 600V

Open circuit voltage is approx. 3.2V. Silicon PN junction bias ranges 0.5 to 0.8V

Resolution: 1mV; Overload protection: 600V 

 

Note: Accuracy specifications consist of two elements: 

 (% reading) Accuracy of the measurement circuit 

 (+ digits) Accuracy of the analog to digital converter 

18

 MN62‐en‐GB_V1.1   12/15 

General Specifications 

Display  4,000 count backlit liquid crystal display with bargraph 

Over‐range indication  OL is displayed 

Auto Power Off  15 minutes (approximately) with disable feature 

Polarity 

Measurement Rate 

Automatic (no indication for positive); Minus (‐) sign for negative  

2 to 3 times per second, nominal 

Battery Status Indication Battery icon   provides battery operating voltage status 

Battery Power 

Fuse Protection 

2 x 1.5V AA batteries 

A/mA range: F1 fuse (ϕ6x32mm); FF0.5A H 600V (CE)

  10A range: F2 fuse (ϕ6x25mm) F 10A H 600V (CE)

Operating Temperature  0 o

C to 40 o

C (32 o

F to 104 o

F) 

 

Storage Temperature  ‐10 o

C to 50 o

C (14 o

F to 122 o

F) 

Operating Humidity  ≤ 75% RH for 0~30 o

C (32~86 o

F); ≤ 50% RH for 30~40 o

C (86~104 o

F)

 

 

Operating Altitude 

Weight  

Dimensions  

Safety 

Approvals 

UL Listed 

2000m (7000’) maximum 

350g (12.3oz.) includes holster and batteries 

175 x 80 x 48.5mm (6.9 x 3.1 x 1.9”) includes holster 

EN61010‐1: CATIII 600V; Pollution Degree 2  

CE/UL 

The UL mark does not indicate that this product has been evaluated for  the accuracy of its readings. 

 

Copyright © 2015 FLIR Systems, Inc. 

All rights reserved including the right of reproduction in whole or in part in any form 

www.extech.com    

19

 MN62‐en‐GB_V1.1   12/15 

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement