Brooks GF100 / GF120 / GF125 Mass Flow Controllers

Brooks GF100 / GF120 / GF125 Mass Flow Controllers
DS-TMF-GF100-Series-MFC-eng
December, 2016
Data Sheet
GF Series
GF100/GF120/GF125
Thermal Mass Flow
GF100 Series
High Purity/Ultra-High Purity
Digital Thermal Mass Flow Devices
Overview
Designed for semiconductor, MOCVD, and other gas flow control applications that require
a high purity all-metal flow path, the Brooks® GF100 Series mass flow controllers and
meters deliver outstanding performance, reliability, and flexibility. Highlights of the GF
Series industry-leading features include: ultra fast 300 millisecond settling time,
MultiFlo™ gas and range programmability, optional pressure transient insensitivity (PTI),
local display, extremely low wetted surface area, and corrosion resistant Hastelloy® sensor
tube and valve seat. The GF100 Series has been marathon tested to over three times the
semiconductor industry standard for reliability, ensuring repeatable low-drift performance over time. An independent diagnostic/service port permits users to troubleshoot or
change flow conditions without removing the mass flow controller from service.
The flagship GF125 is a second generation multi-variable, pressure transient insensitive
mass flow controller. This product builds upon Brooks' leadership position in pressure
transient insensitive (PTI) mass flow controller technology, minimizing process gas flow
variation due to pressure and temperature fluctuations. The GF125 enables customers to
simplify and reduce the size and cost of gas panels by eliminating the need for point of
use pressure regulators, pressure transducers, and associated hardware.
MultiFlo™ gas and range programmability, a patented technology developed and refined
by Brooks over the last 10 years, has changed the mass flow controller industry by
offering customers the ability to select new gas calibrations and full scale ranges without
the trouble and cost of removing the mass flow controller from the gas line. The GF Series
fourth generation MultiFlo technology continues to lead the market with the most
accurate and broadest range performance through extensive refinement and physical
validation on critical process gases.
Pr
oduct Description
Product
The GF100 Series is a highly configurable platform based on a novel modular
architecture. Already widely adopted by semiconductor, vacuum thin film, solar, and
related customers, the GF100 Series feature set was carefully selected to enable drop-in
replacement and upgrade of most brands of metal-seal mass flow controllers, including
the former Celerity, UNIT, Tylan, and Mykrolis brands. With the wide range of options and
features available, the GF100 Series provides users with a path to simplification and
standardization, greatly reducing spares inventory and support costs.
1
Featur
es and Benefits
Features
User In
terface
Interface
• Easy installation, start-up, and operation
Con
venien
ort
Convenien
venientt Service PPort
• Easy diagnostics and troubleshooting
• Independent of tool communication
Magnetically Coupled Valve
• Corrosion resistant all metal valve
• Diaphragm free design, unaffected by
differential pressure
• High purge rate capability
• Tested for over seven million cycles
with no failures
Pr
essur
Pressur
essuree Sensor
• Stable flow control
regardless of upsets or
fluctuations in delivery
pressure
Second Gener
ation T-Rise Sensor
Generation
• Excellent long-term stability
(<+0.5% F.S/yr.)
• Improved signal to noise ratio
• High purge rate capability
• Lower temperature operation for gases
prone to thermal decomposition
Temper
atur
emperatur
aturee Sensor
• Accurate flow and temperature measurement over
full temperature range
• Real-time compensation for ambient temperature
flucuations enable precise gas chemistry control
2
Featur
es
Features
Benefits
MultiFlo Gas and Range Configurability
Ability to reconfigure the mass flow controller for new gas calibrations and full scale ranges without
the time and costs of removing the device from the gas line.
User Accessible Service Port with Advanced
Diagnostics with User-Friendly Interface
Convenient interface to diagnostics for maximum uptime. Ensures device is operating within user
specified limits for high yield and maximum uptime.
Corrosion Resistant Hastelloy T-Rise Sensor
Provides unmatched long-term sensor stability ensuring maximum yield and throughput.
Pressure Transient Insensitivity (PTI), and
Safe Delivery System (SDS) Options
Improves yield. Reduces overall gas panel costs.
Pr
oduct Description
Product
Brooks MultiFlo technology offers unparalleled flexibility; a
single device can be programmed for thousands of different
gas and flow range configurations.
Re-programming is simple and fast; a new gas and range can
be programmed in under 30 seconds. Brooks provides a full
gas database to ensure the true value of MultiFlo is realized:
• Dramatically reduces inventory costs
• Mass flow controller full scale flow range can re-scaled down
typically by a factor of 3:1 with no impact on accuracy,
turndown or leak by specifications, for optimum process and
inventory flexibility
• Up to 40% fewer configurations required to support typical
etch and CVD processes verses our closest competitor
• Widest process gas coverage through extensive gas library
• Mass flow controllers can be replaced in only a few minutes
• Off-the shelf spares programmability enables rapid process
recovery
• Maximum flexibility for research applications
80
60
40
20
0
0
1
2
3
4
5
Time in Seconds
-20
Set-Point (% FS)
Brooks GF125
Competitor A
Competitor B
MFC
MFC
MFC
MFC
1.6L Xe 2.6L H2
MFC
MFC
MFC
MFC
2.6L CO 2.1L HBr 2.6L N2 1.6L C12
MFC
=
3.6L He 3.6L Ar
MFC
GF Series
MFC
MFC
MFC
Comparison of GF vs Competitors response time
to ascending & decending setpoints
MFC
MFC
1.8L H2Se 2.0L NH3 1.9L PH3 1.7L COS
MFC
MultiFlo
MultiFlo™ Gas and Range Configurability
A major advancement over traditional single point gas
conversion factors, Brooks MultiFlo technology delivers up to a
three-times improvement in process gas accuracy. This is
achieved through advanced gas modeling optimized through
actual gas testing providing compensation for non-linear gases.
MultiFlo also allows the device to be quickly and easily
configured for another gas and/or flow range without
sacrificing accuracy or rangability. Selecting a new gas
automatically creates a new calibration curve, establishes
optimized PID settings for dynamic control, automatically
compensates for gas density effects, and ensures smooth,
overshoot-free transitions between flow rates with excellent
steady-state stability.
100
Set Point (% of Full Scale)
Ultra Fast Response
By combining Brooks' patented flow sensor technology with a
high speed ARM processor and fast acting diaphragm free valve
assembly, the GF100 Series delivers up to 3 times faster
response and settling time compared to other mass flow
controllers, enabling:
• Improved wafer throughput by reducing nonproductive flow
settling steps
• Critical Etch processes requiring ultrafast 1-2 second etch steps
• Reduced diverted gas consumption and associated
abatement costs
• Time-sensitive gas delivery steps in Atomic Layer Deposition
• For processes requiring a slow ramped gas turn-on or time
critical transitions between flow rates. A user programmable
ramp function is provided
1.5L SiH4 1.3L NF3 1.5L COF2
...
MultiFloTM technology allows one GF Series to be programmed
for thousands of different gases and flow ranges
# of
Platforms
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
GF1xx Series
Range
3 - 10
11 - 30
31 - 92
93 - 280
281 - 860
861 - 2,600
2,601 - 7,200
7,201 - 15,000
15,001 - 30,000
30,001 - 40,000
40,001 - 55,000
Competitor A
2 Models
Range
10
17.5
30
55
100
175
300
550
1,000
1,750
3,000
5,500
10,000
22,000
30,000
50,000
Competitor B
4 Models
Range
1-5
6 - 14
15 - 27
28 - 38
39 - 71
72 - 103
104 - 192
193 - 279
280 - 754
755 - 2,037
2,038 - 5,500
5,501 - 11,000
11,001 - 30,000
30,0001 - 50,000
The Brooks Advantage! Less platforms means more process
flexibility and lower cost of spares.
3
Pr
oduct Description
Product
MultiFloTM Configurator Accessory Kits:
778Z010ZZZ
Basic MultiFlo Configurator Kit
*Software, MultiFlo Configurator
A331710003
Cable Assembly 2.5mm
214F027AAA
USB-RS485 Converter with DB-9 female
778Z011ZZZ
A331710003
214F027AAA
A332295001
A332297002
A332297001
Basic MultiFlo Configurator Kit
w/Power Supply and Adapter Cables
*Software, MultiFlo Configurator
Cable Assembly 2.5mm
USB-RS485 Converter with DB-9 female
Power Supply MFC
Cable, Power, 9-Pin
Cable, Power, DeviceNet
* MultiFlo Configurator Software is available on the Brooks
Instrument website at: www.BrooksInstrument.com/MultiFlo
Pressure Flucuations in Non-Pressure Transient Insensitivty MFC
Pressure Transient Insensitivity (PTI) (GF125 only)
Cost and space constraints are driving gas panel designers to
remove point of use pressure regulators and pressure
monitoring components, placing more burden on the mass flow
controller to control accurately under dynamic pressure
conditions. Conventional mass flow controllers react strongly to
small inlet pressure fluctuations resulting in unstable
performance and unpredictable accuracy (see Non-Pressure
Insensitive MFC). This drove Brooks to develop Pressure
Transient Insensitive mass flow controller technology (PTI-MFC).
The GF125 PTI-MFC is a second generation PTI-MFC utilizing a
patented control algorithm that inverts the pressure signal,
compares it to the pre-fluctuation signal and drives real-time
valve position compensation to maintain stable flow. Enhanced
pressure transient insensitivity is achieved through faster
sensing, faster processing, and a reduction in internal deadvolume between the sensors and valve orifice.
4
Stable Pressure with Pressure Transient Insensitivty GF Device
GF125 Reliability - Zero Drift - 40oC
0.60
Zero Output (% F.S.)
Advanced Thermal Flow Measurement Sensor
Brooks' proprietary sensor technology combines:
• Improved signal to noise performance for improved accuracy
at low setpoints
• Improved reproducibility at elevated temperatures through
new isothermal packaging, onboard conditioning
electronics with ambient temperature sensing and
compensation
• Improved long-term stability through enhanced sensor
manufacturing and burn in process
• Highly corrosion resistant Hastelloy C-22 sensor tube
• Optimized temperature profile for gases prone to thermal
decomposition
• Unique orthogonal sensor mounting orientation
-Eliminates sensor drift caused by valve heating effects
-Eliminates thermal siphoning effects for the most common
mounting orientations
0.40
0.20
0.00
-0.20
-0.40
-0.60
0K
50K
100K
250K
500K
750K
Number of Cycles
GF125-1
GF125-2
Upper Limit (+0.5% F.S.)
Lower Limit (-0.5% F.S.)
GF125-3
1M
Pr
oduct Description
Product
High Purity Flow Path
All metal, corrosion resistant flow path with reduced surface area
and un-swept volumes for faster dry-down during purge steps:
• SEMI F-20 compliant wetted flow path
• 4 μ inch Ra max surface finish standard (10 μ inch Ra on GF100)
• Highly corrosion resistant Hastelloy C-22 valve seat and jet orifice
Extensive Mechanical Configuration Support
GF100 Series supports all metal seal / UHP industry gas
connection interface standards for full OEM and process coverage
• Downport 80mm and 92mm C-seal and W-Seal, on 1.125" and
1.5" bodies
• Downport 80mm CS seal on 1.5" body
• 124 mm 1/4” VCR Male on 1.5" body
High Purity Flow Path
Enhanced Diagnostics
The mass flow controller remains the most complex and critical
component in gas delivery systems. When dealing with UHP gas
distribution or highly toxic or corrosive gases, removing the mass
flow controller to determine if it is faulty should be the last resort.
In response to this, Brooks pioneered smarter mass flow controllers
with embedded self test routines and introduced an independent
diagnostic/service port to provide the user with a simple interface,
for troubleshooting without disturbing flow controller operation.
MFC on Gas Box
User Interface
The user interface has a high visibility LCD display that provides
a local indication of Flow (%), Temperature (°C), Pressure (PSIA/
KPa) and Network Address, selectable through the Display button.
A Zero button provides a simple means to re-zero the mass flow
controller as part of scheduled maintenance.
Communication Interface
The GF100 Series supports analog 0-5 Vdc, RS485, and
DeviceNet™ communication protocols. A range of low profile
adapter cables facilitate replacing older mass flow controllers
with the GF100 Series eliminating the need to carry mass flow
controllers of same gas/range but different electrical connectors.
Communication Interface
5
Pr
oduct Specifications (Standar
d GF Series)
Product
(Standard
Performance
GF100
GF120
Full Scale Flow Range (N 2 Eq.)
Flow Accuracy
+1% S.P. > 35-100%,
Repeatability & Reproducibility
+ 0.5% F.S. (included in accuracy)
< 1 sec
300ms (3-860 sccm N2 Eq.)
400ms (861-7200 sccm N2 Eq.)
500ms (7201-30000 sccm N2 Eq.)
<700ms (30001-55000 sccm N2 Eq.)
700ms
Normally Open Valve
<1.5 sec
Pressure Insensitivity
Control Range
+0.35% F.S. 2-35%
< + 0.15% S.P.
Linearity
Response Time (Settling Time)
Normally Closed Valve
GF125
3 sccm to 55 slm
Not Applicable
< 5% SP up to 5 psi/sec upstream press. spike
2-100% (Normally Closed Valve)
3-100% (Normally Open Valve)
MultiFlo
standard
#of Bins
11 bins
Valve Shut Down (N.C. Valve)
Valve Shut Down (N.O. Valve)
< 1% of F.S.
2% of F.S.
Zero Stability
< + 0.5% F.S. per year
Temperature Coefficient
Span: 0.05% S.P. per oC, Zero: 0.005% F.S. per oC
Ratings
Operating Temperature Range
Differential Pressure Range*
10-50 oC
3-860 sccm = 7-45 psid, 861- 7200 sccm = 10-45 psid, 7201-55000 sccm = 15-45 psid
*Argon gas applications require an additional 10 psid differential pressure.
Low vapor pressure gases require an inlet pressure of > 100 Torr, with vacuum on outlet
(example SiCl4). Contact Brooks Technical Support for more information.
Maximum Operating Pressure
500 psia max
Leak Integrity (external)
100 psia max
1x10 -10 atm. cc/sec He
Mechanical
Valve Type
Wetted Materials
Surface Finish
Normally Closed
Normally Open
Meter (no valve)
GF100: SEMI F20 HP Compliant, 316L VIM/VAR, Hastelloy C-22, 316L Stainless Steel, 304 Stainless Steel, KM-45
GF120/GF125: SEMI F20 UHP Compliant, 316L VIM/VAR, Hastelloy C-22, 316L Stainless Steel, 304 Stainless Steel, KM-45
4μ inch Ra (0.1 μm Ra)
10μ inch Ra
Diagnostics & Display
Status Lights
MFC Health, Network Status
Alarms
Display Type
Viewing Angle / Viewing Distance
Units Displayed / Resolution
Control Valve Output, Network Interruption
Top Mount Integrated LCD
Fixed / 10 feet
Flow (%), Temp. (oC), Pressure (psia, kPa) / 0.1 (unit)
Electrical
Electrical Connection
RS485/Analog via 9-Pin “D” connector, DeviceNetTM via 5-Pin “M12” connector
Digital Communication
RS485+ (model specific), DeviceNet (model specific), RS485 Diagnostic Port (all models)
Diagnostic /Service Port
RS485 via 2.5mm jack
Power Supply/Consumption
DeviceNet: 545mA max. @ +11-25 Vdc., 250mA max. @ 24Vdc
RS485/Analog: 6 Watts max @ +15Vdc. (+10%) or +24 Vdc (±10%)
Compliance
EMC
Environmental Compliance
EC Directive 2004/108/EC CE: EN61326: 2006 (FCC Part 15 & Canada IC-subset of CE testing)
RoHS Directive (2011/65/EU)
REACH Directive EC 1907/2006
NOTE: See the following Safe Delivery System (SDS) section for optional detailed specifications
6
®
Pr
oduct Description GF120 Safe Delivery System (SDS ® ) Option
Product
The GF120 Safe Delivery System (SDS ®) is Brooks’ state-of-the-art low
pressure drop mass flow controller for the delivery of sub atmospheric
safe delivery system (SDS) gases used in Implant and Etch processes. The
Brooks GF120 (SDS) models are available in full scale flow ranges 4-25
sccm (option GF120XSL) or >25 sccm to 1 slpm (option GF120XSD).
These expensive, hazardous gases are adsorbed onto a solid medium
within the gas cylinder, remaining below atmospheric pressure despite
containing up to 15 times more dopant than conventional pressurized
sources.
The amount of gas that can be extracted from the SDS controlled cylinder
is highly dependent upon the final cylinder pressure. This is illustrated in
SDS desorption species information in the SDS Gas Use Rate Efficiency
Curve. Most of the gas is released at pressures below 100 Torr. The
minimum cylinder pressure that can be reached is limited by the
conductance of the mass flow controller regulating the flow. Most mass
flow controllers require a 50 Torr differential pressure at flow rates of 5
sccm. At this 50 Torr limit, only ~65% of the dopant can be extracted
from the adsorbent medium at normal operating temperatures. The
GF120 (SDS) low pressure operation enables a further 30% of the
dopant to be extracted, driving significant cost savings in SDS gases
and equipment OEE.
Pr
oduct Specifications (GF120XSD and GF120XSL) Options
Product
Performance
Full Scale Flow Range (N 2 Eq.)
Gases Supported
MultiFlo Programmable
Flow Accuracy
Repeatability & Reproducibility
Zero Stability
GF120XSL
GF120XSD
4 - 25 sccm
>25 to 1 slpm
AsH 3, PH3, BF3, SiF4, Ar, Xe, N2O, N2, GeF4, AsF5, PF3,
H2Se, HMDSO, HMDSN, H2O***
Not Configurable
+/-1% S.P. >35% F.S. +/-0.35% F.S. <35% F.S.
<+/- 0.15% S.P.
<=0.6% F.S. per year
Settling Time (to within + 2% F.S.)
< 3 sec
Warm Up Time
Leak Integrity
minimum of 30 minutes
1X10-11 atm. cc/sec He
Valve Shut Down (Leaky by)
Operating Conditions
Minimum Operating Inlet Pressure*
Maximum Pressure
Pressure Insensitivity
DifferentialPressure**
Valve Configuration
Ambient Temperature Range
Zero Temperature Coefficient
<1% F.S.
GF120XSD
GF120XSL
4 to 20 sccm < 10 Torr
>20 to 50 sccm < 20 Torr
>50 sccm to 1 slpm < 50 Torr
500 psia max
Not Available
10 Torr-30 psid typical (1.33-207 kPa typical)
Normally Closed
10 OC-50OC
Span: 0.05% SP per OC, Zero: 0.005% F.S. per OC
*Performance at minimum inlet pressure will be gas and flow range dependent. Consult Technical Support for details.
**Typical pressure drop. Actual pressure drop will be gas and flow dependent. Consult Technical Support for details.
***Consult factory for other gases.
7
Electrical
Pr
oduct Specifications
In
terface Options
Product
Interface
Base I/O Options
I/O Options Using Base Model and Adapter Cable
PDC Or
dering Code G1
Ordering
Description: Industry standard
Analog / RS485 interface
123
PDC Or
dering Code GX
Ordering
Description: OEM specific
Analog / RS485 interface.
Display and top plate
re-oriented 180o
A range of low profile adapter cables have been developed to support
replacing older generation MFC's with different pinout configurations.
The base MFC will be either a G1, TX or SX configuration, depending on
the product being replaced.
PDC Or
dering Code UX
Ordering
Description: SX base I/O with
7003550 adapter for
compatibility with Unit UDU15
PDC Or
dering Code: T1
Ordering
Description: TX base I/O with
7003551 adapter for compatibility
with IFlow DB15 & TN 15 pin
PDC Or
dering Code: FX / JX
Ordering
Description: SX base I/O with
7003069 (FX)/7001814 (JX)
adapter for compatibility with
Unit UDF9/UDJ9
PDC Or
dering Code: EX
Ordering
Description: GX base I/O with
7003083 adapter for compatibility
with Unit “E”, IN “L”, “R”
PDC Or
dering Code: KX
Ordering
Description: G1 base I/O with
7003298 adapter for compatibility
with Unit UDK15
PDC Or
dering Code: B
Ordering
BXX
Description: G1 base I/O with
7003590 adapter for
compatibility with Brooks 15-Pin D
123
PDC Or
dering Code D
Ordering
DXX
Description: Industry standard
ODVA compliant DeviceNet
interface
PDC Or
dering Code TX
Ordering
Description: Industry standard
Analog only interface
PDC Or
dering Code SX
Ordering
Description: Industry standard
Analog 9-Pin Sub D connector
and dual RJ11 RS485 ports
PDC Or
dering Code BB
Ordering
Description: Industry standard
ODVA compliant DeviceNet
interface, Plus a separate
Analog 0-5 Vdc Connector
All Base I/O options include:
Diagnostic port communication RS485 via 2.5mm jac
jackk
Other adapter options ar
tact Br
ooks
aree available for the GF Series. Please con
contact
Brooks
Customer Service for mor
formation.
moree in
information.
8
Pr
oduct Dimensions
Product
Downport Configurations
Electrical Connector
G1/GX/TX Specific Dimensions
Electrical Connector
SX Specific Dimensions
Electrical Connector
DX/BB Specific Dimensions
E
3.29in
84mm
3.67in
93mm
.98in
25mm
5.09in
129mm
G
4.45in
113mm
3.75in
95mm
F
A
Fitting Option Code
D
C
B
Dim A
Dim B
Dim C
Dim D
Dim E
Dim F
Dim G
CX
C-SEAL
92mm [3.62in]
105mm [4.13in]
22mm [0.86in]
28mm [1.12in]
83mm [3.28in]
25mm [1.00in]
127mm [5.00in]
EX
W-SEAL
79.8mm[3.14in]
93mm [3.66in]
30mm [1.18in]
39mm [1.53in]
72mm [2.82in]
25mm [1.00in]
127mm [5.00in]
92mm [3.62in]
105mm [4.13in]
127mm [5.00in]
WX
Seal Type
28mm [1.12in]
83mm [3.28in]
25mm [1.00in]
DX
C-SEAL
79.8mm[3.14in]
93mm [3.66in]
22mm [0.86in]
28mm [1.12in]
72mm [2.82in]
25mm [1.00in]
127mm [5.00in]
YX
W-SEAL
W-SEAL
79.8mm[3.14in]
93mm [3.66in]
22mm [0.86in]
22mm [0.86in]
28mm [1.12in]
72mm [2.82in]
25mm [1.00in]
127mm [5.00in]
AX
C-SEAL
92mm [3.62in]
105mm [4.13in]
30mm [1.18in]
39mm [1.53in]
83mm [3.28in]
25mm [1.00in]
127mm [5.00in]
BX
W-SEAL
92mm [3.62in]
105mm [4.13in]
30mm [1.18in]
39mm [1.53in]
83mm [3.28in]
25mm [1.00in]
127mm [5.00in]
LX
C-SEAL
92mm [3.62in]
105mm [4.13in]
22mm [0.86in]
28mm [1.12in]
83mm [3.28in]
25mm [1.00in]
127mm [5.00in]
VCR Configurations
9
Model Code
Code Description
I.
Base Model Code
Code Option
GF
100
120
125
II.
Package / Finish Specifications
III.
Configurability
C
X
MultiFlo capable. Standard bins or specific gas/range may be selected.
Not MultiFlo capable. Specific gas/range required. (must select w/ SD, SL or HA special application)
IV.
Special Application
XX
SL
SD
Standard
Safe Delivery System (GF120 Only) Full scale flow range; 4 to 25 sccm, Nitrogen Equivalent
Safe Delivery System (GF120 Only) Full scale flow range; >25 sccm to 1 slpm, Nitrogen Equivalent
V.
Valve Configuration
O
C
M
Normally Open valve (not available with SD, SL or HA options)
Normally Closed valve (must select with SD, SL or HA special application)
Meter (No Valve)
VI.
Gas or SH MultiFlo Bin
XXXX XXXX
SH40 010C
SH41 030C
SH42 092C
SH43 280C
SH44 860C
SH45 2.6L
SH46 7.2L
SH47 015L
SH48 030L
SH49 040L
SH50 055L
VII. Fitting
10
Option Description
High Purity/Ultra High Purity Digital Mass Flow Controllers
VX
VS
CX
DX
EX
WX
YX
AX
BX
LX
AS
Flow range 3 sccm -55 slpm N2 Eq.; + 1.0% SP Accuracy; 1 sec Response; 10 Ra
Flow range 3 sccm -55 slpm N2 Eq.; + 1.0% SP Accuracy; 700 msec Response; 4 Ra
Pressure Transient Insensitive (PTI) Flow range 3 sccm -55 slpm N2 Eq.; + 1.0% SP Accuracy;
300-700 msec Response; 4 Ra
Specific Gas Code & Range, i.e. “0004” = Argon and “010L” = 10 slpm (must select w/ SD, SL or
HA special application).
Standard Configuration #40, 3-10 sccm Nitrogen Equivalent (0° C Reference)
Standard Configuration #41, 11-30 sccm Nitrogen Equivalent (0° C Reference)
Standard Configuration #42, 31-92 sccm Nitrogen Equivalent (0° C Reference)
Standard Configuration #43,93-280 sccm Nitrogen Equivalent (0° C Reference)
Standard Configuration #44, 281-860 sccm Nitrogen Equivalent (0° C Reference)
Standard Configuration #45, 861-2600 sccm Nitrogen Equivalent (0° C Reference)
Standard Configuration #46, 2601-7200 sccm Nitrogen Equivalent (0° C Reference)
Standard Configuration #47, 7201-15000 sccm Nitrogen Equivalent (0° C Reference)
Standard Configuration #48, 15001-30000 sccm Nitrogen Equivalent (0° C Reference)
Standard Configuration #49, 30001-40000 sccm Nitrogen Equivalent (0° C Reference)
Standard Configuration #50, 40001-55000 sccm Nitrogen Equivalent (0° C Reference)
1-1/2” body width, 124mm 1/4” VCR male
1-1/8” body width, 124mm 1/4” VCR male
1-1/8” body width, 92mm C Seal
1-1/8” body width, 79.8mm C Seal
1-1/2” body width, 79.8mm W Seal
1-1/8” body width, 92mm W Seal
1-1/8” body width, 79.8mm W Seal
1-1/2” body width, 92mm C Seal
1-1/2” body width, 92mm W Seal
1-1/8” body width, 92mm C Seal w/Poke Yoke
1-1/2” body width, 92mm 0.440” large bore C Seal (only for bins SH45-SH50)
VIII. Downstream Condition
A
V
Atmosphere
Vacuum; Default for SD, SL and HA special application
IX.
O
Default Sensor Orientation
Sensor
Model Code continued on next page.
Model Code (con
tinued)
(continued)
Code Description
X. Connector
Code Option
BX
EX
FX
GX
G1
JX
KX
SX
TX
T1
UX
BB
Option Description
Cable adapter to 15 pin D Brooks (Unit “B”,”N”) adapts G1 base
Cable adapter to Card Edge (w/out VTP), RS485 through RJ11 jacks (Unit”E”; IN “L”, “R”) adapts GX base
(Not Available on 79.8mm fitting DX, YX, EX)
Cable adapter with 9 pin STEC pin-out & jack screws (w/VTP) (Unit”F”,”O”) adapts SX base
9-Pin D with RS485 (Unit”G”); display and overlay 180o orientation (Not Available on 79.8mm fitting DX, YX, EX)
9-Pin D with RS485 (Unit”G”) (Not Available on 79.8mm fitting DX, YX, EX)
Cable adapter with 9 pin STEC pin-out & jack screws (w/VTP) (Unit”J”,”W”) adapts SX base
Cable adapter to MKS 15-Pin D (Unit “K”) adapts G1 base
9 pin D with STEC pin-out (w/VTP) (Unit”S”,”Q”)
9 pin D with UDT9 pin-out (UDT9)
Cable adapter to 15 pin D (IFlow DB15 & TN 15 pin) adapts TX base (Not Available on 79.8mm fitting DX, YX, EX)
Cable adapter to 15 pin D (w/VTP) (Unit & TN “U”) adapts SX base
DeviceNetTM Analog (Not Available on 79.8mm fitting DX, YX, EX)
DeviceNet Standard Configuration Parameters
Poll IO
Poll IO
Poll IO External
Power On Full Scale Full Scale Full Scale Instance Instance
State
Baud
Connector
State
Setting
Setting
Setting Producer Consumer Transition
Rate
I/O
D0
D1
D2
D3
D4
D5
D6
D7
D8
D9
DA
DB
DC
DD
DE
DX
Code Description
XI. Customer Special Request
DeviceNet
DeviceNet
DeviceNet
DeviceNet
DeviceNet
DeviceNet
DeviceNet
DeviceNet
DeviceNet
DeviceNet
DeviceNet
DeviceNet
DeviceNet
DeviceNet
DeviceNet
DeviceNet
Code Option
XXXX
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
Idle
Count
Idle
Count
Idle
SCCM
Idle
Count
Executing Count
Idle
Count
Idle
Count
Idle
Count
Idle
Count
Executing Count
Idle
Count
Idle
Count
Idle
Count
Executing Count
Executing Sccm
To be defined by CSR
Integer
Integer
Float
Integer
Integer
Integer
Integer
Integer
Integer
Integer
Integer
Integer
Integer
Integer
Float
6000h
6000h
7FFFh
6000h
6000h
6000h
7FFFh
7FFFh
6000h
6000h
7FFFh
6000h
7FFFh
7FFFh
6000h
2
21
13
22
22
6
3
6
3
2
22
22
3
22
15
A
X
Auto Shut-Off (Included) Default for SD and SL special application
Auto Shut-Off (Not Included) (Must be selected for meter)
XIII. Auto Zero
X
Auto Zero (Not Included)
Sample Standard Model Code
I
II
III
GF
100
C
000
IV
XX
V
M
Executing
Executing
Executing
Executing
Executing
Executing
Executing
Executing
Executing
Executing
Executing
Executing
Idle
Executing
Executing
500KB
500KB
500KB
500KB
500KB
500KB
500KB
500KB
500KB
500KB
500KB
500KB
500KB
500KB
500KB
Option Description
Customer Special Request Number; required with “DX, BB” Conn. Option to define DNet settings
XII. Auto Shut-Off
XIV. Reference Temperature
7
7
19
7
8
8
7
8
7
7
7
8
7
8
19
0o C Reference Calibration (Standard) - Default Setting
VI
VII
- SH40010C - VX
Sample Safe Delivery System (SDS) Model Code
I
II
III
V
VI
IV
X
SD
C - XXXXXXXX GF
120
VII
EX
VIII
A
IX
0
X
GX
-
XI
XXXX
XII
A
XIII
X
XIV
- 000
VIII
V
IX
0
X
SX
-
XI
XXXX
XII
A
XIII
X -
XIV
000
11
Br
ooks Service and Support
Brooks
Brooks is committed to assuring all of our customers receive the ideal flow solution for their application, along with outstanding
service and support to back it up. We operate first class repair facilities located around the world to provide rapid response and
support. Each location utilizes primary standard calibration equipment to ensure accuracy and reliability for repairs and recalibration and is certified by our local Weights and Measures Authorities and traceable to the relevant International Standards.
Visit www.BrooksInstrument.com to locate the service location nearest to you.
TION
STAR
ARTT-UP SERVICE AND IN-SITU CALIBRA
CALIBRATION
ST
AR
Brooks Instrument can provide start-up service prior to operation when required. For some process applications, where ISO-9001
Quality Certification is important, it is mandatory to verify and/or (re)calibrate the products periodically. In many cases this service
can be provided under in-situ conditions, and the results will be traceable to the relevant international quality standards.
CUST
OMER SEMINARS AND TRAINING
CUSTOMER
Brooks Instrument can provide customer seminars and dedicated training to engineers, end users, and maintenance persons.
Please contact your nearest sales representative for more details.
Due to Brooks Instrument's commitment to continuous improvement of our products, all specifications are subject to change
without notice.
TRADEMARKS
Brooks is a trademark of Brooks Instrument, LLC
All other trademarks are the property of their respective owners.