advertisement
MARSIC300
Ship Emission Measuring Device
Installation and Initial Start-up
T E C H N I C A L I N F O R M A T I O N
Described product
MARSIC300
Manufacturer
SICK AG
Erwin-Sick-Str. 1
79183 Waldkirch, Germany
Germany
Legal information
This work is protected by copyright. Any rights derived from the copyright shall be reserved for SICK AG. Reproduction of this document or parts of this document is only permissible within the limits of the legal determination of Copyright Law. Any modifica‐ tion, expurgation or translation of this document is prohibited without the express writ‐ ten permission of SICK AG.
The trademarks stated in this document are the property of their respective owner.
© SICK AG. All rights reserved.
Original document
This document is an original document of SICK AG.
2 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
CONTENTS
Contents
8017585/V2-0/2015-09 | SICK
Subject to change without notice
1
About this document........................................................................
6
1.1
Function of this document.......................................................................
6
1.2
6
1.3
6
1.4
Symbols and document conventions......................................................
6
1.4.1
1.4.2
1.4.3
1.4.4
Warning symbols......................................................................
Warning levels / Signal words.................................................
Information symbols................................................................
Safety symbols on device........................................................
6
7
7
7
2
9
2.1
Installation information............................................................................
9
2.1.1
2.1.2
Notes on the gas supply..........................................................
Tube screw fitting.....................................................................
9
9
2.1.3
Information on power supply..................................................
10
2.2
10
2.3
10
2.4
11
2.5
Checklist for mechanical and electrical installation..............................
11
2.6
Installing the analyzer cabinet.................................................................
13
2.7
Laying the sample gas line.......................................................................
14
2.8
Laying the tube bundle cable...................................................................
15
2.9
Connecting the sample gas line to the analyzer.....................................
15
2.10 Connecting the tube bundle cable to the analyzer.................................
19
2.11 Gas connections on analyzer...................................................................
22
2.12 Electrical connections on the analyzer.................................................... 26
2.13 Installing the gas sampling system.........................................................
29
3
3.1
31
4
Menus (SOPAS ET)............................................................................. 33
4.1
33
4.1.1
Save parameters ..................................................................... 33
4.2
33
4.3
33
4.4
Menu tree (SOPAS ET)..............................................................................
34
4.5
35
4.5.1
Measuring screen 1 .. X...........................................................
35
4.6
37
4.6.1
4.6.2
4.6.3
4.6.4
Control values..........................................................................
37
Sensor values..........................................................................
39
Signals......................................................................................
39
Logbook....................................................................................
41
T E C H N I C A L I N F O R M A T I O N | MARSIC300 3
CONTENTS
4
4.6.5
4.6.6
4.6.7
System status..........................................................................
43
System info............................................................................... 43
Operating hours counter.........................................................
43
4.7
44
4.7.1
Measuring components..........................................................
44
4.7.2
4.7.3
Measuring screen....................................................................
I/O.............................................................................................
44
46
4.7.4
4.7.5
4.7.6
4.7.7
4.7.8
4.7.9
4.7.3.1
Hardware Plan......................................................... 46
4.7.3.2
Data.........................................................................
47
Sequence control program / Sampling point program.........
Temperature control................................................................
49
51
Pressure control.......................................................................
51
Flow rate...................................................................................
51
O2 sensor.................................................................................
Logbook....................................................................................
51
51
4.7.10
Instrument display...................................................................
52
4.7.11
System .....................................................................................
52
4.7.12
Emitter......................................................................................
53
4.7.13
Modbus..................................................................................... 53
4.8
54
4.8.1
Parameter................................................................................
54
4.8.1.1
Concentrations of span gases...............................
54
4.8.1.2
Adjustment factors.................................................
54
4.8.1.3
Start times..............................................................
54
Manual adjustment.................................................................
55 4.8.2
4.8.3
Automatic adjustment.............................................................
55
4.9
55
4.9.1
4.9.2
4.9.3
4.9.4
Tests.........................................................................................
Operating states......................................................................
System maintenance...............................................................
Service log................................................................................
55
56
57
58
4.9.5
4.9.6
System restart.......................................................................... 58
Confirm active messages........................................................
59
59
5
Maintenance...................................................................................... 61
5.1
Leak tightness check during initial startup............................................. 61
5.2
Leak tightness check with pressure test tool.........................................
62
6
6.1
Dimensional drawings..............................................................................
64
6.2
Measuring parameters.............................................................................
65
6.3
65
6.4
Sample gas conditions.............................................................................
66
6.5
6.6
Heated sample gas line............................................................................
66
T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
CONTENTS
6.7
68
6.8
Interfaces and protocols..........................................................................
68
6.9
69
69
6.11 Electrical connections in analyzer...........................................................
70
73
73
74
7
7.1
75
7.1.1
7.1.2
Device status (Discrete Inputs [1xxxx], Function Code 02)..
Measured variables (Input Register [3xxxx], Function Code
75
04)............................................................................................
75
7.1.3
7.1.4
Device-internal monitoring values (Input Register [3xxxx],
Function Code 04)...................................................................
76
Triggering the MARSIC 300 (Coils [0xxxx], Function Code
15, write multiple Coils)..........................................................
76
7.1.5
77
7.2
Error messages and possible causes......................................................
78
8
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 5
1 ABOUT THIS DOCUMENT
1
1.1
1.2
1.3
1.4
1.4.1
About this document
Function of this document
This document describes:
• Installation
• Initial start-up
• Operation via SOPAS ET
NOTICE
This technical documentation is only valid in combination with the MARSIC Operating
Instructions.
Target group
This document is addressed to technicians (persons with technical understanding) operating and maintaining the measuring system.
The technicians must have been trained on the device.
Requirements for the technician
• The technician must be familiar with the exhaust gas technology on the ship (over‐ pressure, toxic and hot flue gases) and able to avoid hazards when working on gas ducts
• The technician must be familiar with handling compressed gas cylinders (span gases)
• The technician must be able to avoid hazards caused by noxious span gases
• Only electricians are permitted to work on the electrical system or electrical subas‐ semblies
Further information
• SFU Gas Sampling System Operating Instructions
• Sample Gas Line Operating Instructions
• System documentation
• Optional: MPR (Meeting Point Router) Operating Instructions
• Optional: Instrument Air Conditioning Operating Instructions
• Optional: External Measuring Point Switchover Operating Instructions
• Optional: Pressure Test Tool Operating Instructions
Symbols and document conventions
Warning symbols
Symbol Significance
Hazard (general)
Hazard by voltage
Table 1: Warning symbols
6 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
1.4.2
1.4.3
1.4.4
ABOUT THIS DOCUMENT 1
Symbol Significance
Hazard in potentially explosive atmospheres
Hazard by explosive substances/mixtures
Hazard by toxic substances
Hazards by noxious substances
Hazard by high temperature
Hazard for the environment/nature/organic life
Table 1: Warning symbols
Warning levels / Signal words
DANGER
Risk or hazardous situation which will result in severe personal injury or death.
WARNING
Risk or hazardous situation which could result in severe personal injury or death.
CAUTION
Hazard or unsafe practice which could result in less severe or minor injuries.
Notice
Hazard which could result in property damage.
Note
Hints
Information symbols
Symbol Significance
Important technical information for this product
Important information on electric or electronic functions
Table 2: Information symbols
Safety symbols on device
The following safety symbols are on the device:
Symbol Significance
Warning for general hazard
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 7
1 ABOUT THIS DOCUMENT
Symbol Significance
Warning for hazard by electric voltage, possibly also by residual electric voltage
If you need to work on a subassembly marked with such a symbol: b Read the relevant Section in these Operating Instructions b Observe all the safety information in the relevant Section
8 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
2
2.1
2.1.1
2.1.2
INSTALLATION 2
Installation
Installation information
Notes on the gas supply
The operator is responsible for the correct laying of sample gas lines.
ATTENTION
Risk of contamination of the analyzer by contaminated instrument air.
b Only use instrument air according to the applicable specification (see Technical data) b If required, install a suitable instrument air conditioning
Tube screw fitting
Steel coupling b Push the tube into the tube screw fit‐ ting to the stop
Turn the cap nut finger-tight b Hold the fitting bolt steady and tighten the cap nut with 1 1/4 revolutions
Push-in fitting (pneumatic)
Pressure tool
8017585/V2-0/2015-09 | SICK
Subject to change without notice
1 Retaining ring b Connecting the tube: Push tube in b Disconnecting the tube: Press the retaining ring in and pull the tube out
It is easier to press the retaining ring in using the pressure tool provided.
T E C H N I C A L I N F O R M A T I O N | MARSIC300 9
2 INSTALLATION
2.1.3
2.2
2.3
Information on power supply
The operator is responsible for correct laying and connection of the electric lines.
WARNING
Danger to life by electric voltage b Only allow an authorized electrician to work on the electric system
ATTENTION
The analyzer power supply is configured for an individual power system
Check the configured power system against the system documentation provided.
• If the analyzer power system does not match the power system on board: Please contact SICK Customer Service
When selecting and laying the electric lines for power supply, observe the applicable local standards and guidelines.
Scope of delivery
Included in scope of delivery:
• SFU gas sampling system
• Analyzer, complete, pre-installed and configured
• Heated sample gas line, length as ordered
• Tube bundle cable non-terminated
• All screw fittings of the enclosure ducts
Please see the delivery documents for the scope of delivery.
Provision by operator
To be provided by operator, especially:
• Suitable flange on the exhaust duct (see "SFU Operating Instructions")
• Fixing accessories of enclosures (dowels, screws, etc.)
• Fixing accessories for heated sample gas line and tube bundle cable
• Mains line:
• External power disconnection unit
• Compressed air, instrument air as zero gas when necessary
Observe quality of operator's instrument air:
• Optional span gases
10 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
INSTALLATION 2
2.4
Installation overview
1
2
3 3
2
2.5
5 6 7
Figure 1: Installation - overview
1 Gas sampling system
Heated sample gas line (with 2 measuring points: 2 sample gas lines)
Tube bundle cable (with 2 measuring points: 2 tube bundle cables) with pneumatic and elec‐ tric lines
Analyzer
Energy supply
Interfaces
Instrument air inlet
Option: Instrument air conditioning
Sample gas outlet
Checklist for mechanical and electrical installation
Fitting system components
System component
Install SFU gas sampling system
Install analyzer cabinet
Optional: Install instrument air conditioning
Optional: Measuring point switchover (more than 2 measuring points)
Optional: Install MPR.
Reference
see "Installing the gas sampling system", page 29
see "Installing the analyzer cabinet", page 13
See enclosed “Instrument Air Conditioning Operating
Instructions”
See enclosed system documentation of measuring point switchover
See enclosed “MPR Operating Instructions”
Laying the sample gas line and tube bundle cable
Line
Lay the sample gas line
Lay the tube bundle cable
Laying and connecting the gas lines
Reference
see "Laying the sample gas line", page 14
see "Laying the tube bundle cable", page 15
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 11
2 INSTALLATION
Gas line
Heated sample gas line: Gas sampling system -> analyzer cabinet
Tube bundle cable: Gas sampling system -> analyzer cabinet
Sample gas outlet analyzer cabinet
Instrument air feed on pressure control unit
Optional auxiliary instrument air feed on pressure control unit
Optional: Span gas feed on span gas valve
Connection
Gas sampling system: see "Installing the gas sam‐ pling system", page 29
Analyzer cabinet: see "Gas connections on analyzer", page 22
Gas sampling system: see "Installing the gas sam‐ pling system", page 29
Analyzer cabinet: see "Gas connections on analyzer", page 22
Analyzer cabinet: see "Gas connections on analyzer", page 22
Analyzer cabinet: see "Gas connections on analyzer", page 22
Laying and connecting electric lines
Electric line
Power supply analyzer cabinet
Tube bundle cable for gas sampling system -> ana‐ lyzer cabinet
Ethernet
MPR
Connection
Analyzer cabinet: see "Electrical connections on the analyzer", page 26
Gas sampling system: see "Installing the gas sam‐ pling system", page 29
Analyzer cabinet: see "Electrical connections on the analyzer", page 26
Analyzer cabinet: see "Electrical connections on the analyzer", page 26
Analyzer cabinet: see "Electrical connections on the analyzer", page 26
12 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
2.6
Installing the analyzer cabinet
INSTALLATION 2
8017585/V2-0/2015-09 | SICK
Subject to change without notice
NOTICE
Observe clearances:
• Top: 30 cm
• Bottom: 20 cm
• Observe the clearance for the heated sample gas line (see Fig. above).
• Install the analyzer in a well ventilated area.
•
Observe the relevant ambient conditions: see "Ambient conditions", page 65 .
• Attach the enclosure with the mounting brackets provided on a suitable wall with sufficient load-bearing capacity
• Secure the enclosure with 4 x M10 screw fittings (2 at bottom, 2 at top) with prop‐ erty class 8.8 (or higher)
• Secure the screw fittings against loosening
• Drilling plan: See dimension drawings above
• Install the enclosure horizontal
T E C H N I C A L I N F O R M A T I O N | MARSIC300 13
2 INSTALLATION
2.7
Laying the sample gas line
1
Connection side without electric connec‐ tions on gas sampling system
Connection side with electric connec‐ tions on analyzer
Protective cap
2 x Pt100 connections (1 as reserve)
Power supply
Cable gland
Counter nut
WARNING
Risk of fire b Observe the laying instructions provided with the line b Minimum clearance to other lines (for example, tube bundle cable): 2 cm b Do not lay or roll up lines directly next to each other b Start laying at the analyzer
°
The end with the electric connection belongs on the analyzer
Important: The screw fitting for the housing duct must be located at the end
° of the electrical connection (analyzer side)
The end without electrical connection belongs on the gas sampling system
Roll-up excess length on the gas sampling system
Leave enough length for pulling the gas sampling system b Protect the line from damage (chafing through vibration, mechanical load) b Observe minimum bend radius of 300 mm
14 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
INSTALLATION 2
2.8
Laying the tube bundle cable
2.9
Figure 2: Tube bundle cable non-terminated
Tube bundle cable (system-specific example) with:
- Power supply
- Grounding conductor
- Signal lines
- Tube, gray: Zero gas/span gas
- 2 x tubes, black/blue: Main valve and backflush
Start laying the tube bundle cable at the analyzer and roll-up excess length at the gas sampling probe.
1.
Assembly on the analyzer:
see "Gas connections on analyzer", page 22 and
"Electrical connections on the analyzer", page 26
2.
Lay the tube bundle cable to the analyzer cabinet w Attach excess length to the gas sampling probe w Leave enough length for pulling the probe w Protect the line from damage (chafing through vibration, mechanical load) w Minimum bend radius: 300 mm
3.
Assembly on the gas sampling probe: see "Installing the gas sampling system", page 29
Connecting the sample gas line to the analyzer
WARNING
Danger to life by electric voltage b Only allow an authorized electrician to work on the electric system
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 15
2 INSTALLATION
The flexible wires of the heated sample gas line are numbered.
Line assignment:
see "Heated sample gas line", page 66
Figure 3: Heated sample gas line
1
Connection side without electric connec‐ tions on gas sampling system
Connection side with electric connec‐ tions on analyzer
Protective cap
2 x Pt100 connections (1 as reserve)
Power supply
Cable gland
Counter nut
1.
Unscrew counter nut from the cable gland and pull off the sample gas line.
2.
Lead sample gas line together with electrical connections from above through the housing opening on the analyzer.
3.
Push counter nut back over the sample gas line and electric connections.
16
1
Figure 4: Clamping ring screw connection
Sample gas line 1
Sample gas line 2
Cable gland
Counter nut
Clamping ring screw connection
(cell)
Cable duct
Figure 5: Foam insulation
4.
Screw counter nut tight on the cable gland
5.
Unscrew cell cover and remove.
CAUTION
Hot parts in cell.
b Let cell cool down before working on the cell.
6.
Remove protective cap from sample gas line
T E C H N I C A L I N F O R M A T I O N | MARSIC300
Figure 6: Finished assem‐ bly
8017585/V2-0/2015-09 | SICK
Subject to change without notice
INSTALLATION 2
7.
Insert sample gas line to stop in the clamping ring screw connection on the cell
With 2 sample gas lines: Observe inlets:
8.
°
°
Right inlet: Measuring point 1
Left inlet: Measuring point 2 see "Clamping ring screw connection"
9.
Attach red foam insulation on the clamping ring screw connection and bind together with a cable clip. No thermal bridges may remain.
10. Close cell again
11. Screw cable gland tight
12. Push electric lines downwards through the cable duct
13. Connect power supply
14. Connect Pt100
Electric terminal connection plan:
L
Heating
8017585/V2-0/2015-09 | SICK
Subject to change without notice
Plug
TUBE 1
Subassembly
Sample gas line 1
Pin
1
2
3
Assignment
L (L)
N (L)
PE
Tube bundle cable line number 1
T E C H N I C A L I N F O R M A T I O N | MARSIC300 17
2 INSTALLATION
Plug
FILTER1
Subassembly
Gas sampling system filter 1
(Lines from tube bundle cable)
Pin
1
2
3
PROBE1
TUBE2
FILTER2
PROBE2
TUBE3
Gas sampling system probe tube 1
(Lines from tube bundle cable)
Sample gas line 2
Gas sampling system filter 2
Gas sampling system probe tube 2
Sample gas line 3
4
5
6
1 ... 3
1 ... 3
4 ... 6
1 The connections must match the connections on the gas sampling system
Assignment
L (L)
N (L)
PE
L (L)
N (L)
PE (not connected)
As for TUBE1
As for FILTER1
As for PROBE1
Tube bundle cable line number 1
4x1.5
mm 2
1
2
1x4 mm 2
4x1.5
mm 2
GNYE
3
4
Plug
Pt100
DIGITAL
INPUTS
Subassembly
Sample gas line 1
Gas sampling system filter 1
Gas sampling system probe tube 1
Not connected
Sample gas line 2
Gas sampling system filter 2
Gas sampling system probe tube 2
Sample gas line 3
Digital input 1
Digital input 2
Digital input 3
Digital input 4
Pin Assignment
1
2
5
6
3
4
7
8
9, 10
Pt100 +
Pt100 –
Pt100 +
Pt100 –
Pt100 +
Pt100 –
As above
1
2
3
11, 12 As above
13, 14 As above
15
16
Pt100 +
Pt100 –
+ 24 V
+ Signal
- Signal
4
5 ... 8
GND
As above
9 ... 12 As above
13 ... 16 As above
Tube bundle cable line number 1
4x1.0
mm 2
3
4
1
2
4x1.0
mm 2
18 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
2.10
Plug Subassembly Pin Assignment
DIGITAL
OUTPUTS
Digital output 1
Digital output 2
Digital output 3
Digital output 4
Valves
1
2
3
4 ... 6
7 ... 9
NC
COM
NO
As above
As above
10 ... 12 As above
Internal VALVE
OUTPUTS
1 The connections must match the connections on the gas sampling probe
Connecting the tube bundle cable to the analyzer
INSTALLATION 2
Tube bundle cable line number 1
8017585/V2-0/2015-09 | SICK
Subject to change without notice
1
Tube bundle cable 1
Tube bundle cable 2 (optional)
Valve block
Electronics unit
T E C H N I C A L I N F O R M A T I O N | MARSIC300 19
2 INSTALLATION
Figure 7: Tube bundle cable non-terminated
Tube bundle cable (system-specific example) with:
- Power supply
- Grounding conductor
- Signal lines
- Tube, gray: Zero gas/span gas
- 2 x tubes, black/blue: Main valve and backflush
1
Power supply 4x1.5 mm2
Lines 1 and 2: Gas sampling system filter
Lines 3 and 4: Gas sampling system probe tube
Pt100 lines 4x1.0 mm2
Lines 1 and 2: Gas sampling system filter
Lines 3 and 4: Gas sampling system probe tube
Grounding conductor GNYE 4 mm2
PTFE tube DN4/6: Zero gas
PA tube black DN6/8; Imprint "1":
Main valve: Control air
Connections on analyzer and gas sampling probe must match
PA tube blue DN6/8; Imprint "2":
Backflush: Backflush air
Connections on analyzer and gas sampling probe must match
WARNING
Danger to life by electric voltage b Only allow an authorized electrician to work on the electric system
CAUTION
The tube bundle cable is unusable when it is cut off too short.
The tube bundle cable is unusable when a lead inside is damaged when removing the sheath.
b If the tube bundle cable is long enough: Leave a piece as "reserve" b Only remove the sheath from the tube bundle cable when you are capable of doing this
1.
Remove 1900 mm of the tube bundle cable sheath
2.
Lead the tube bundle cable through the screw fitting on the housing top
3.
Lead tubes and lines of the tube bundle cable downwards in the cable duct
4.
Connect the 3 gas lines of one tube bundle cable to the valve block (connections marked as "outlet" in the Figure below)
20 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
8017585/V2-0/2015-09 | SICK
Subject to change without notice
INSTALLATION 2
NOTICE
The gas connections on the valve block must match the gas connections of the gas sampling system (see "SFU Gas Sampling System Operating Instructions").
9
-
Figure 8: Valve block
1
Outlet: Zero gas measuring point 1
Outlet: Zero gas measuring point 2
(option)
Outlet: Control air measuring point 1
Outlet: Backflush air measuring point 1
Outlet: Control air measuring point 2
(option)
Outlet: Backflush air measuring point 2
(option)
Inlet: Zero gas
Inlet: Control/backflush air
Inlet: Auxiliary control air
Red plug = dummy plug
5.
Connect heating of gas sampling system to plug PROBE1 (also to plug PROBE2 for 2 measuring points):
NOTICE
The connections on MARSIC300 must match the connections of the gas sampling system.
Electric terminal connection plan:
1.
Connect Pt100 of the gas sampling system to plug PT100
NOTICE
The connections on MARSIC300 must match the connections of the gas sampling system.
T E C H N I C A L I N F O R M A T I O N | MARSIC300 21
2 INSTALLATION
2.11
22
Plug Subassembly Pin Assignment Tube bundle cable line number 1
Pt100 Sample gas line 1
Gas sampling system filter 1
Gas sampling system probe tube 1
Not connected
1
2
3
Pt100 +
Pt100 –
Pt100 +
6
7
4
5
8
9, 10
Pt100 –
Pt100 +
Pt100 –
As above
11, 12 As above
13, 14 As above
4x1.
0 mm 2
Sample gas line 2
Gas sampling system filter 2
Gas sampling system probe tube 2
DIGITAL
INPUTS
DIGITAL
OUTPUTS
Sample gas line 3
Digital input 1
Digital input 2
Digital input 3
Digital input 4
Digital output 1
1
2
15
16
3
4
5 ... 8
Pt100 +
Pt100 –
+ 24 V
+ Signal
- Signal
GND
As above
1
2
9 ... 12 As above
13 ... 16 As above
NC
COM
3
4 ... 6
7 ... 9
NO
As above
As above
Digital output 2
Digital output 3
Digital output 4 10 ... 12 As above
VALVE
OUTPUTS
Valves Internal
The connections must match the connections on the gas sampling probe
4x1.
0 mm 2
1
2
3
4
Gas connections on analyzer
WARNING
Hazard when pressure is too high
Hoses can burst when the pressure is too high b
Observe the maximum pressures of the gases provided by the operator: see "Sup‐ ply gases", page 73
T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
INSTALLATION 2
1
Sample gas outlet on cell
Fixing of line sample gas outlet on the cell
8017585/V2-0/2015-09 | SICK
Subject to change without notice
Figure 9: Gas connections MARSIC300
1
Heated sample gas line for measuring point 1
Heated sample gas line for measuring point 2 (option)
Tube bundle cable 1
Tube bundle cable 2 (option)
Instrument air
Instrument air as zero gas/span gas
(option)
Span gas (option)
Sample gas outlet
Connect instrument air
Connect instrument air to the pressure control unit.
ATTENTION
Risk of contamination of the analyzer by contaminated instrument air.
b Only use instrument air according to the applicable specification (see Technical data) b If required, install a suitable instrument air conditioning
T E C H N I C A L I N F O R M A T I O N | MARSIC300 23
2 INSTALLATION
Figure 10: Pressure reducer unit
1
Instrument air inlet with zero gas quality
Instrument air inlet for induc‐ tion air ejector only
Manual valve for instrument air selection
3 pressure reducers (adjusta‐ ble)
Manual valve
"closed"
Manual valve
"open"
The instrument air is used as both induction air for the ejector (cell) and zero/control air.
Instrument air can be connected in 2 ways: b One (1) common instrument air supply for ejector air and zero/control air (inlet 1) b Separate instrument air supply for:
Ejector air (inlet 2) and zero/control air (inlet 1)
■
°
°
Instrument air quality
The requirements for instrument air quality are lower when the air is only used as ejector air than when used for zero/control air (zero gas quality) ( see
""see "Supply gases", page 73 )
• When connected just as (1) instrument air supply with zero gas quality to be used as common air for both ejector air and zero/control air (on inlet 1):
°
Set manual valve to position "open"
• When connected as instrument air supply for the ejector (on inlet 2) and as instru‐ ment air supply with zero gas quality (on inlet 1):
°
Set manual valve to position "closed"
ATTENTION
Risk of contamination of the analyzer by contaminated instrument air.
b
Only use instrument air corresponding to the mandatory specification ( see "Supply gases", page 73
) b If required, install a suitable instrument air conditioning
Connect span gas (option)
WARNING
Hazard when pressure is too high
Hoses can burst when the pressure is too high b
Observe the maximum pressures of the gases provided by the operator: see "Sup‐ ply gases", page 73
24 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
Connect span gas to the span gas valve.
INSTALLATION 2
8017585/V2-0/2015-09 | SICK
Subject to change without notice
1
Span gas valve
Span gas inlet
Connect sample gas outlet
CAUTION
Acidic condensate in sample gas outlet - risk of blockage
Acidic condensate forms in the sample gas outlet b Always lay the sample gas outlet line running downwards so that no condensate can accumulate b Lay the line end in a suitable disposal device (drain or condensate collection con‐ tainer) b Do not kink the line and protect against frost
The analyzer is fitted with an outlet hose (DIN 8/10) about 30 cm long when delivered.
• Lay the sample gas outlet in a suitable exhaust duct
• The sample gas outlet must be open to the ambient pressure
Recommendation for lengthening the outlet hose:
To prevent the outlet line clogging, it is recommended to lay a (1) new line starting at the sample gas outlet of the cell to the disposal location.
To do this (see Fig. above):
1.
Open cell cover (loosen 4 screws on the side)
2.
Unscrew sample gas outlet on sample gas outlet (bracket piece)
3.
Loosen fixing of sample gas outlet line
4.
Remove the thermal insulation from the existing line and remove the line
5.
Lead a new line through the housing duct at the bottom on the housing floor and connect it to the sample gas outlet of the cell
6.
Refit the fixing
7.
Refit the thermal insulation
8.
Close cell again
To lengthen the existing hose piece: No cross-section narrowing may occur when con‐ necting the lengthening hose.
Sample gas outlet on housing:
T E C H N I C A L I N F O R M A T I O N | MARSIC300 25
2 INSTALLATION
2.12
1 Sample gas outlet at bottom rear of the housing
Further connections
NOTE
The following connections have already been described in the Manual but are repeated here for the sake of completeness.
Connect heated sample gas line to analyzer
Connect heated sample gas line to analyzer: see "Connect the sample gas line to the analyzer"
Connect tube bundle cable to analyzer
Connect tube bundle cable to analyzer:
see "Connecting the tube bundle cable to the analyzer", page 19
Electrical connections on the analyzer
Connect power supply
ATTENTION
The analyzer power supply is configured for an individual power system
Check the configured power system against the system documentation provided.
• If the analyzer power system does not match the power system on board: Please contact SICK Customer Service
The power supply is located on the left on the analyzer.
26 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
INSTALLATION 2
NOTICE
°
°
°
Install an external power disconnection unit which disconnects all connectors and fuses near the analyzer
The power disconnection unit must be marked clearly and be easily accessi‐ ble
Observe the maximum power input of the complete system: see "Power sup‐ ply", page 69
The onsite wiring system to the power source of the system must be installed and fused according to the relevant regulations
Always connect a protective ground to PE
° b Route the electric lines through the screw fittings of the enclosure b Connect the electric lines
Connect signal line (optional)
1 2 x signal line ducts
• Lead cable through the housing duct
Attach shield as shown in the corresponding Figure above.
Connect Ethernet (optional)
8017585/V2-0/2015-09 | SICK
Subject to change without notice
Plug
ETH0
ETH1
USB
SD card
CAN1
CAN2
RS422, RS488
RS232 (top plug)
O2 (bottom plug)
Connection for
Ethernet (e.g. SOPAS ET), MPR (remote maintenance)
Internal
Internal
SD card (on the right, next to USB)
Internal
Internal
Internal
Internal
O
2
sensor
T E C H N I C A L I N F O R M A T I O N | MARSIC300 27
2 INSTALLATION
Plug
DISP (top plug)
I/O-MOD (bottom plug)
Connection for
Display
Internal
1 Ethernet cable duct
• Lead cable through the housing duct
Attach shield as shown in the corresponding Figure above.
• Connect Ethernet to ETH0 (network or computer with SOPAS ET)
• Plug type: RJ 45
Connect MPR (remote maintenance) (option)
Plug
ETH0
ETH1
USB
SD card
CAN1
CAN2
RS422, RS488
RS232 (top plug)
O2 (bottom plug)
DISP (top plug)
I/O-MOD (bottom plug)
Connection for
Ethernet (e.g. SOPAS ET), MPR (remote maintenance)
Internal
Internal
SD card (on the right, next to USB)
Internal
Internal
Internal
Internal
O
2
sensor
Display
Internal
• Connect MPR (remote maintenance via SICK Meeting-Point Router) to ETH0
• Further information, see "MPR Operating Instructions”
Further connections
28 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
2.13
INSTALLATION 2
NOTE
The following connections have already been described in the Manual but are repeated here for the sake of completeness.
Connect heated sample gas line to analyzer
Connect heated sample gas line to analyzer: see "Connect the sample gas line to the analyzer"
Connect tube bundle cable to analyzer
Connect tube bundle cable to analyzer:
see "Connecting the tube bundle cable to the analyzer", page 19
Installing the gas sampling system
NOTICE b Observe the ambient conditions of the gas sampling system: See "SFU Gas Sam‐ pling System Operating Instructions" b Leave the sample gas line and tube bundle cable long enough to be able to pull the sample gas system b All connections must match the connections in the analyzer (
see "Connecting the tube bundle cable to the analyzer", page 19
)
Fitting the flange b Install the flange of the gas sampling system as described in the "SFU Gas Sam‐ pling System Operating Instructions"
°
Observe the 10° tilt of the probe tube during installation
Gas connections b Connect the following gas connections:
°
°
Heated sample gas line
Tube bundle cable
• Black line 1: Main valve
• Black line 2: Backflush
• PTFE tube Instrument air/span gas
Electrical connections b Connect the following electric lines of the tube bundle cable:
°
°
°
°
Filter power supply
Probe tube power supply
Filter Pt100 line
Probe tube Pt100 line
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 29
2 INSTALLATION
Tube bundle cable
1
Power supply 4x1.5 mm2
Lines 1 and 2: Gas sampling system fil‐ ter
Lines 3 and 4: Gas sampling system probe tube
Pt100 lines 4x1.0 mm2
Lines 1 and 2: Gas sampling system fil‐ ter
Lines 3 and 4: Gas sampling system probe tube
Grounding conductor GNYE 4 mm2
PTFE tube DN4/6: Zero gas
PA tube black DN6/8; Imprint "1": Main valve: Control air
Connections on analyzer and gas sam‐ pling probe must match
PA tube blue DN6/8; Imprint "2": Back‐ flush: Backflush air
Connections on analyzer and gas sam‐ pling probe must match
Fitting the gas sampling system on the flange
NOTICE
Risk of contamination of gas sampling system b First install the gas sampling system on the exhaust duct just before the analyzer is switched on b Install the gas sampling system: See "SFU Gas Sampling System Operating Instruc‐ tions"
30 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
3
3.1
INITIAL START-UP 3
Initial start-up
Initial start-up
Before switching on
1.
Check that all connections are connected properly as described in the "Installa‐ tion" Sections
2.
3.
Adjust the manual valve in the analyzer as described in
see "Gas connections on analyzer", page 22
4.
Open the external instrument air supply
5.
Set pressures: (
see "Gas connections on analyzer", page 22
)
6.
Check the power supply of the analyzer matches the setting on the ship network:
see "Electrical connections on the analyzer", page 26
7.
Check the voltage connection on the gas sampling system
8.
Install the gas sampling system on the flange of the exhaust duct: See "SFU Gas
Sampling System Operating Instructions"
Switching on
1.
Switch on the external power disconnection unit
2.
The green "Power" LED on the control panel goes on:
Energy supply is available
3.
The yellow and red LEDs go on sporadically
4.
BOOTING appears several times on the screen
5.
The Measuring screen appears
6.
The system heats up:
°
°
Only the green LED is on.
Display: INIT/HEATING UP
A downwards counter displays the maximum duration of the process.
7.
Display: CONDITIONING
8.
Only the green LED is on and the status bar shows MEASURING
The system is ready for operation
If a measured values blinks: The measured value is outside the calibration range
When the yellow or red LED is on: Press the Diag button and clear the error: Error
list see "Error messages and possible causes", page 78
Final tests and parameter settings
1.
Perform a leak tightness check:
see "Leak tightness check during initial startup", page 61
2.
Perform an internal adjustment: Display menu Adjustment/internal adjustment
3.
Connect computer with SOPAS ET on ETH0 (
see "Electrical connections in ana‐ lyzer", page 70
) (
)
If the customer network is already connected: Disconnect the customer network
4.
If required: Set the times for automatic adjustments w To change in SOPAS ET: Menu Adjustment/Parameter/Start time
5.
If required: Configure the measuring point switchover w In SOPAS ET: Menu Parameterization/Sequence Controls/Sampling point pro‐ gram
6.
Set the IP address of the customer network for the ETH0
° In the SOPAS ET user interface: Click the pen symbol in the device tile of the
MARSIC300
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 31
3 INITIAL START-UP
7.
Disconnect the computer from ETH0 and connect the customer network to ETH0
Initial start-up is completed.
32 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
4
4.1
4.1.1
4.2
4.3
MENUS (SOPAS ET) 4
Menus (SOPAS ET)
SOPAS ET
SOPAS ET can also be used to the configure MARSIC300 and to access the logbook.
SOPAS ET runs on an external PC connected to the MARSIC300 via the Ethernet inter‐ face.
You can download SOPAS ET free of charge from the SICK homepage (" DOWNLOADS ").
• Install SOPAS ET on a laptop.
•
Connection on MARSIC300: see "Electrical connections on the analyzer", page 26 .
• Open SOPAS ET.
• In SOPAS ET, search for device "MARSIC300” and connect (see SOPAS ET Help menus)
Save parameters
SOPAS ET can be used to backup and restore the MARSIC300 parameters: See the
Help menu in SOPAS ET.
Passwords
There are 2 passwords:
Password for
Device display access "Authorized
User"
SOPAS ET
Password
1234
HIDE
Change password
Changing the password:
see "Instrument display", page 52
Cannot be changed
Using the menus
Example: Menu: Parameterization/I/O/Data/External Data/Digital outputs
8017585/V2-0/2015-09 | SICK
Subject to change without notice
Entry field
1
Significance
Save current entries.
Create group b Click "Mark" and then click the desired lines.
b To cancel "Mark": Click "Mark" again and then click an empty line.
In editable menus: Call up the setting menu for lines.
b Click desired line. Then click "Edit".
Copy selected lines
Insert copied lines before a selected line
Call up "Next function group"
(The bottom lines in the example depend on the menu)
T E C H N I C A L I N F O R M A T I O N | MARSIC300 33
4 MENUS (SOPAS ET)
4.4
34
Menu tree (SOPAS ET)
The menu tree shown here shows the menus for user level "Authorized User".
Menu Reference
Measuring screen
Measuring screens
Diagnosis
Control values
Zero drift
Span gas drift
Internal adjustment drift
Reference energy
Intensity
Sensor values
Temperatures
Pressures
Flow rate
Cell
O2 sensor
Emitter
Motors
Hardware
Signals
External signals
Measuring signals (MVi)
Diagnosis internal (IDOi, ID
Boolean values (BVi, LIi)
Real values (RVi)
Modbus values (MBVi)
Modbus input values (MBIVi)
Modbus input flags (MBIFi)
Modbus reference flags (MBIRi)
Filtered values (FVi)
Integer values (IVi)
Real constant (RCi)
Logbook
System status
System info
Timemeter
Parameter
Measuring components
Global definitions
Measuring screen
Measuring Screen x
I/O
Hardware plan
Data
– External data
– OPC outputs (OPCOi)
– Modbus values (MBVi)
– Modbus input values (MBIVi)
– Modbus input flags (MBIFi)
– Modbus reference flags (MBIRFi)
Sequence control programs
Sampling point program (MPP)
Temperature control
Pressure control
Flow rate
O2 sensor
Logbook
T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
MENUS (SOPAS ET) 4
4.5
Instrument display
Device
Emitter
Modbus
Adjustment
Parameter
Concentrations
Adjustment factors
Start times
Manual adjustment
Automatic adjustment
Maintenance
Tests
Digital inputs
Digital outputs
Analog inputs
Analog outputs
Operating states
System maintenance
Service log
Restart system
Confirm active messages
Measuring screen
Menu: MEASURING SCREEN
Only the configured Measuring screens are shown.
b Double-click the desired Measuring screen
4.5.1
•
Configuring Measuring screens: see "Measuring screen", page 44
•
Scaling Measuring screens: see "Measuring screen 1 .. X", page 35
Measuring screen 1 .. X
Menu: MEASURING SCREEN/MEASURING SCREEN 1 .. X
The Measuring screen is shown as measured value box, bar graph or line writer depending on the configuration.
Measured value box
Here, the measured values are shown in numeric format.
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 35
4 MENUS (SOPAS ET)
1 Component
Unit
Measured value
Bar graph
Here, the measured values are shown as a bar graph.
1 Component
Unit
Measured value
Line writer and data storage
Here, the measured values are shown in numeric format as measured value box and in a time chart.
1 Component
Line writer - shows the activated measured val‐ ues (see under "Scaling
Measuring screens").
The line writer first starts the display when it is called up for the first time.
Changing the user level deletes the history of the line display.
36
Data storage functions b Start:
Start data storage.
Data that have been configured are stored (independent of the graphic represen‐ tation).
Maximum 65536 entries per curve are stored in a file and then a new file created automatically (with an incremental index).
w A dialog field appears above the line diagram to enter a target file name (.txt):
T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
MENUS (SOPAS ET) 4 b Stop:
Stop recording.
b View:
View recorded data.
A dialog field appears to select the file with the stored values.
The following then appears above the line diagram:
4.6
4.6.1
b Reset:
Delete all lines displayed (no effect on data storage).
Scaling Measuring screens
Tap on a display to call up a screen for scaling:
1 Font color (the line writer colors are predefined)
Measuring screen precision
Example:
-2: 123.45
-1: 1234.5
0 : 12345
1 : 123450
Scale start value
Scale end value
(for line writer: for y-axis)
Line display active / not active
(for line writer)
Diagnosis
Control values
Zero drift
Menu: DIAGNOSIS/CONTROL VALUES/ZERO DRIFT
This menu displays the drift "with zero gas" (since the last drift "reset") and can be reset.
This drift is recalculated as from the relevant adjustment.
8017585/V2-0/2015-09 | SICK
Subject to change without notice b Reset: Reset zero drift
Both "Reset" have the same effect
Span gas drift
Menu: DIAGNOSIS/CONTROL VALUES/DRIFT SPAN GAS
This menu displays the drift "with span gas" (since the last drift "reset") and can be reset.
T E C H N I C A L I N F O R M A T I O N | MARSIC300 37
4 MENUS (SOPAS ET)
This drift is recalculated as from the relevant adjustment.
b Reset: Reset zero drift
Internal adjustment drift
Menu: DIAGNOSIS/CONTROL VALUES/DRIFT INTERNAL ADJUSTMENT
This menu displays the drift "with internal standard" (this means without span gas)
(since the last drift "reset").
Reference energy
Menu: DIAGNOSIS/CONTROL VALUES/REFERENCE ENERGY
This menu displays the current reference energy (as percentage) and can be reset.
The energy is automatically monitored.
If a limit value is underflown (default setting: 60 %), the MARSIC300 switches to classi‐ fication " MAINTENANCE REQUEST “.
b Reset: Reset reference energy
Intensity
Menu: DIAGNOSIS/CONTROL VALUES/INTENSITY
This menu displays intensities (energies) and amplification levels of the measuring components.
Let SICK Customer Service evaluate this information.
38 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
4.6.2
4.6.3
MENUS (SOPAS ET) 4
Sensor values
Menu: DIAGNOSIS/SENSOR VALUES
This menu displays the diverse internal device sensor values:
• Temperatures
• Pressures
• Flow rate
• Cell
• O2 sensor
• Emitter
• Motors
• Hardware
These values serve as information.
A message is output when a value is outside its nominal range.
Signals
Menu: DIAGNOSIS/SIGNALS
NOTE
Configuration of signals listed below: see "I/O", page 46
and following.
External signals
Digital signals
Menu: DIAGNOSIS/SIGNALS/EXTERNAL SIGNALS/DIGITAL SIGNALS
This menu displays the current state of the digital signals (DIi, DOi, limit val‐ ues)
Digital values:
. = off (0)
I = on (1)
Analog input signals
Menu: DIAGNOSIS/SIGNALS /EXTERNAL SIGNALS /ANALOG INPUTS
This menu displays the current values of the analog input signals (AIi).
Analog output signals
Menu: DIAGNOSIS/SIGNALS /EXTERNAL SIGNALS /ANALOG OUTPUTS
This menu displays the current values of the analog out‐ put signals (AOi).
Measuring signals
Menu: DIAGNOSIS/SIGNALS/MEASURING SIGNALS
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 39
4 MENUS (SOPAS ET)
This menu displays the measuring sig‐ nals.
Let SICK Customer Service evaluate these signals.
Internal signals
Menu: DIAGNOSIS/SIGNALS/DIAGNOSIS INTERNAL
This menu displays the signals.
Let SICK Customer Service evaluate these signals.
Boolean values
Menu: DIAGNOSIS/SIGNALS/BOOLEAN VALUES
This menu displays the Boolean values
(BVi) and limit val‐ ues (LIi).
Real values
Menu: DIAGNOSIS/SIGNALS/REAL VALUES
This menu displays the current real val‐ ues (RVi).
40
Modbus values
This menu displays the current Modbus values and Modbus flags
T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
MENUS (SOPAS ET) 4
• Modbus values
• Modbus input values
• Modbus input flags
• Modbus reference flags
Further information on Modbus:
and
Filtered values
Menu: DIAGNOSIS/SIGNALS/FILTERED VALUES
This menu displays the current filtered values (FVi).
Integer values
Menu: DIAGNOSIS/SIGNALS/INTEGER VALUES
This menu displays the current integer values (IVi).
Real constants
Menu: DIAGNOSIS/SIGNALS/REAL CONSTANTS
This menu displays the current real constants (RCi).
4.6.4
Logbook
Menu: DIAGNOSIS/LOGBOOK
Maximum number of entries: 6000.
(Representation: uncompressed data storage)
8017585/V2-0/2015-09 | SICK
Subject to change without notice
Name Remark
Logbook fill level in %.
If the font is red: The logbook is full.
Warning mode: No further entries are made.
Ring buffer mode: The oldest entries are overwritten.
T E C H N I C A L I N F O R M A T I O N | MARSIC300 41
4 MENUS (SOPAS ET)
Name
1
Remark
Data storage:
Symbol not crossed out: Compressed.
Symbol crossed out: Uncompressed.
Significance and default setting:
Ring buffer mode warning mode, significance and default value: see "Logbook", page 51
Number of entries of selected filter.
Only filtered messages are shown.
• Failure (active)
• Failure (all)
• Maintenance request (active)
• Maintenance request (all)
• Uncertain (active)
• Uncertain (all)
• Other (active)
• Other (all)
• All active
• All
Attention: All logbook entries in MARSIC300 are deleted
All entries selected with the filter (see in the Table further above) are stored on the PC as log file in C:\Own files.
Format: CSV (comma-separated list). Viewable in, e.g., EXCEL.
To update the display: Click the display.
Scroll to older entries.
Scroll to current entries.
Consecutive number of the message.
Red LED: Message still pending.
Green LED: Message no longer pending.
9
42 T E C H N I C A L I N F O R M A T I O N | MARSIC300
Triggering unit: System, measured value name (sample gas component), subassem‐ bly, Evaluation module.
Number of times the error has occurred.
Significance and default setting:
Only with "compressed data storage".
Logbook message (error messages see MARSIC300 Operating Instructions).
F = Failure
M = Maintenance request
C = Check (function control / maintenance)
U = Uncertain
X = Extended
Format: yy-mm-dd
For "uncompressed": When message occurred.
For "compressed": Last time message occurred
Format: hh:mm:ss
For "uncompressed": When message occurred.
For "compressed": Last time message occurred
8017585/V2-0/2015-09 | SICK
Subject to change without notice
4.6.5
4.6.6
4.6.7
MENUS (SOPAS ET) 4
Name
Remark
Format: yy-mm-dd
For "uncompressed": When message was deleted
For "compressed": Last time message was deleted.
Format: hh:mm:ss
For "uncompressed":When message was deleted
For "compressed":Last time message was deleted.
System status
Menu: DIAGNOSIS/SYSTEM STATUS
1 LEDs in accordance with the operator panel
With measuring point switch‐ over: Current measuring point
With measuring point switch‐ over: Remaining time until measuring point switchover
Name of current sequence control program
Remaining time of current sequence control program
System info
Menu: DIAGNOSIS/SYSTEM INFO
This menu displays system status (see "MARSIC300 Operating Instructions") and sys‐ tem information.
• Serial number
• IP address
• Software version
Operating hours counter
Menu: DIAGNOSIS/TIMEMETER
This menu serves to view diverse operating hours counters.
1 Operating hours
This counter shows the total operating time ("Power on") of the Analyzer mod‐ ule.
Measuring duration
This counter shows the total time sam‐ ple gas was fed.
Operating time of the light source
Operating time of the filter of the gas sampling system
Operating time of the cell inlet filter
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 43
4 MENUS (SOPAS ET)
4.7
4.7.1
Parameterization
Measuring components
Global definitions
Menu: PARAMETERIZATION/MEASURING COMPONENT/GLOBAL DEFINITIONS
This menu displays the global device information (e.g.: number of filter wheels, number of activated components).
1 Active measuring components. Checkmark: Active
Internal use: Error threshold of reference energy as from which a message is generated.
Definition of measuring components
Menu: PARAMETERIZATION/MEASURING COMPONENTS/DEFINITION OF COMPONENTS
This menu displays the setting for the respective measuring component.
This menu only serves as information.
4.7.2
Measuring screen
Menu: PARAMETERIZATION/MEASURING SCREEN
Up to 8 Measuring screens can be configured and displayed:
Select the desired Measuring screen
Double-clicking a Measuring screen opens a screen in which the Measuring screen can be configured:
44 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
Measured value box
MENUS (SOPAS ET) 4
1 Selected Measuring screen
Do not display the Measuring screen
Line writer (see example below)
1 measured value box
4 measured value boxes
16 measured value boxes (see example below)
2 large, 8 small measured value boxes
6 bar graphs (see below)
9 3 bar graphs + 2 measured value boxes
3 bar graphs + 8 measured value boxes
(hidden)
Click "Save" to store
Example: Measuring screen 1 with layout III
(16 measured value boxes)
1 b Enter desired names (tags). MAR‐
SIC300 tags: see "Tags (Name)", page 59
Scaling the measured value box:
"Measuring screen 1 .. X", page 35
The Measuring screen then appears like this.
Component names and units can, for example, originate from the factory settings (e.g., RVi), from the source of a filter (e.g., FVi) or from the analyzer.
8017585/V2-0/2015-09 | SICK
Subject to change without notice
Bar graph
The Measuring screen then looks like this.
Example: Measuring screen 1 with bar graph layout I (6 bars)
1 b Enter desired names (tags). MAR‐
SIC300 tags: see "Tags (Name)", page 59
Scaling the measured value box:
"Measuring screen 1 .. X", page 35
T E C H N I C A L I N F O R M A T I O N | MARSIC300 45
4 MENUS (SOPAS ET)
Line writer
1 b Enter desired names (tags).
Scaling the measured value box:
"Measuring screen 1 .. X", page 35
Scaling the time axis [sec, min or hrs],
scaling the y-axis: see "Measuring screen 1 .. X", page 35
The Measuring screen then looks like this:
1 Measured value box
Line writer
4.7.3
4.7.3.1
I/O
Menu: PARAMETERIZATION/I/O
This menu displays the data interfaces.
Hardware Plan
Menu: Parameterization/I/O/Hardware plan
CAN bus address x
Displays the I/O modules present in the selected CAN bus gateway.
NOTE
The sequence of the specified modules must match the sequence of the plugged in modules (starting at the gateway).
46 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
MENUS (SOPAS ET) 4
1 Consecutive number of module.
Checkmark: Module is plugged in.
I/O module type.
4.7.3.2
Data
8017585/V2-0/2015-09 | SICK
Subject to change without notice
Menu: PARAMETERIZATION/I/O/DATA
Digital inputs (DIi)
Menu: PARAMETERIZATION/I/O/ DATA /EXTERNAL DATA/DIGITAL INPUTS
This menu displays the digital inputs.
Name
Index
Module
Name
Inverted
Remark
Consecutive number of the digital input (DI1, DI2, ....).
Topographic addressing (
). Generated automatically.
Set fixed.
Checkmark: Read in inverted.
Digital outputs (DOi)
Menu: PARAMETERIZATION/I/O/
DATA / EXTERNAL DATA/DIGITAL OUTPUTS
This menu displays the digital outputs.
Name
Index
Module
Source
Inverted
Remark
Consecutive number of the digital outputs (DO1, DO2, ....).
Topographic addressing (
). Generated automatically.
Tag.
Checkmark: Output inverted.
Analog inputs (AIi)
Menu: PARAMETERIZATION/I/O/ DATA / EXTERNAL DATA/ANALOG INPUTS
This menu displays the analog inputs.
Name
Index
Module
Name
Unit
Zero
Range start
Range end
Remark
Consecutive number of the analog inputs (AI1, AI2, ....).
Topographic addressing (
). Generated automatically.
Can be entered freely.
Unit of variable read in.
Select from input screen.
Enter scaling of connected analog signal.
Analog outputs (AOi)
Menu: PARAMETERIZATION/I/O/ DATA / EXTERNAL DATA/ANALOG OUTPUTS
This menu serves to configure the analog inputs.
Name
Index
Module
Source
Live Zero
Remark
Consecutive number of the analog output (AO1, AO2, ....).
Topographic addressing (
). Generated automatically.
Tag.
Select from input screen.
T E C H N I C A L I N F O R M A T I O N | MARSIC300 47
4 MENUS (SOPAS ET)
Name
Progr. perf.
Range 1/2 start/end
MR 1/2 active
Remark
Internal use.
Scaled analog output of measuring range 1 or 2.
Checkmark: Measuring range 1 or 2 active.
Both checkmarks set: Automatic measuring range switch-over.
Hysteresis: 10%.
OPC outputs
Menu: PARAMETERIZATION/I/O/DATA/OPC OUTPUTS
This menu assigns data from the MARSIC300 to the OPC output values.
Name
Index
Source
Remark
Consecutive number of the OPC output value.
Tag.
Modbus values (MBVi)
Menu: PARAMETERIZATION/I/O/DATA/MODBUS VALUES
The Modbus values determine which values are stored in the Modbus.
Edit window:
48
1 Index: Consecutive number of the Modbus value
Active Modbus value: Checkmark: Active
When the checkmark for "Auto Name/Unit" is set, the name and unit of the tag specified in the source are automatically used.
With Autostatus, the system status is used automatically as status of the value
When the checkmark for scaling is activated, this value is scaled. It is scaled in register range 4000 et seq. (Holding register) to 0 ... 10000.
Value name
Measuring range start
9 Measuring range end
Physical unit of value
Source: Tag for which the value is to be used.
Data type (real/Integer/Bool) (2 registers)
The status of the tag named here is used when Autostatus is not activated.
Checkmark: Use standard range register 4000 et seq. (Holding register)
Position: Offset in respective range
Checkmark: Use standard range register 1000 et seq.
Register type (Holding/Input/Coil/DI)
Checkmark: Use standard range register 2000 et seq.
Checkmark: Use standard range register 3000 et seq.
Modbus input values (MBIVi)
Menu: PARAMETERIZATION/I/O/DATA/MODBUS INPUT VALUES
T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
MENUS (SOPAS ET) 4
Serves to assign a name, unit and data type to the measured values stored by the Mas‐ ter in the Holding register, whereby i is the index. Two registers incrementing from the offset are used per measured value. A status is not considered.
Edit window:
1 Line (Index)
Input value name
Physical unit of input value
Data type (real/integer/Bool)
Modbus input flags (MBIFi)
Menu: PARAMETERIZATION/I/O/DATA/MODBUS INPUT FLAGS
Serves to assign a name to the Bool values (flags) stored by the Master in the Coils.
Edit window:
1 Line (Index)
Input flag name
Modbus reference flags (MBIRFi)
Menu: PARAMETERIZATION/I/O/DATA/MODBUS REFERENCE FLAGS
Serves to assign the reference sources (span gas / internal adjustment) to individual values to so that it can be seen what is currently active.
Edit window:
4.7.4
1 Index
Reference flag name
Source: Tag for which the value is to be used.
Sequence control program / Sampling point program
Sampling point program
Menu: PARAMETERIZATION/SEQUENCE CONTROL PROGRAM/SAMPLING POINT PROGRAM
This menu serves to configure "measuring at several sampling points".
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 49
4 MENUS (SOPAS ET)
CAUTION
The programs are automatically controlled.
The device no longer measures correctly when a program is interrupted with Start/
Pause/Stop/Continue.
b Only use these functions when you are sure you can assess the consequences
1 Display: Internal tag to enable the sampling point program
Display: Run-time for the current sampling point
Start the sampling point program
Interrupt the sampling point program (pause)
Display: Active sampling point
Display: Index of active sampling point
Cancel the sampling point program
Continue the sampling point program after "Pause"
9 Input: Number of sampling points
Minimum measuring time per sampling point
Checkmark: Fixed cycle time 285 s ("DeSOx time")
Checkmark: Fixed cycle time from "Cycle time"
Checkmark: Fixed sampling point time
Input: Cycle time
1 Sampling point index
Sampling point name
Sampling point sequence
Measuring time at the sampling point including run-in times
Run-in time
Checkmark active: Sampling point active
50
1 Display: Sampling point index
Checkmark: Sampling point active
Sampling point name
Overall duration at the sampling point
T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
4.7.5
MENUS (SOPAS ET) 4
Run-in time
Position in the sequence of the sampling point program
Temperature control
Menu: PARAMETERIZATION/TEMPERATURE CONTROL
This menu serves to set the unit of the temperature display.
All other displays serve as information.
1 Adjustable: Unit of temperature display
[°C, K, °F ]
4.7.6
4.7.7
Pressure control
Menu: PARAMETERIZATION/PRESSURE CONTROL
This menu serves to view the parameters for pressure control.
Flow rate
Menu: PARAMETERIZATION/FLOW RATE
This menu serves to configure the warning threshold for the gas flow rate.
1 Checkmark: Monitoring active
Input: Minimum flow rate error message
(default value: 100 l/h)
Input: Minimum flow rate warning threshold
(default value: 130 l/h)
Input: Maximum flow rate warning thresh‐ old (default value: 600 l/h)
Input: Maximum flow rate error message
(default value: 1000 l/h)
4.7.8
4.7.9
O2 sensor
Menu: PARAMETERIZATION/O2 SENSOR
This menu serves to view the configuration of the O
2
sensor.
Logbook
Menu: PARAMETERIZATION/LOGBOOK
This menu serves to configure the logbook.
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 51
4 MENUS (SOPAS ET)
Changing the setting deletes all entries.
If an error occurs:
1 An error counter is incremented.
The error message is stored.
When the logbook is full:
Message "Logbook error" is output and no fur‐ ther entries are stored.
The current entries overwrite the oldest entries.
A relevant message is not output.
4.7.10
Instrument display
Menu: PARAMETERIZATION/INSTRUMENT DISPLAY
This menu serves to configure how the instrument display shows information:
• The Measuring screen.
• The password for the operator panel on the display.
• If parameters have been changed and the Measuring screen is being shown on the instrument display: The display on the instrument display must be reinitialized so that the changes can be shown on the instrument display.
1 Press the menu-dependent button " MENU “ on the device.
2 Then press " MEAS “.
1 Password, comprising 4 digits. (Information: Is first used after a hardware reset (
Validity duration of the password-protected level. (Information: Is first used after a hardware reset (
Configuration (
see "Tags (Name)", page 59 ) of the Measuring screen
Start and end values of the display range of the bar and line charts. Start value must be lower than the end value (no plausibility check during input).
Counter format of the display: Number of decimal places.
4.7.11
System
Menu: PARAMETERIZATION/SYSTEM PARAMETERS
This menu serves to display and change system parameters.
52 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
MENUS (SOPAS ET) 4
1 System name
Checkmark: The IP configuration of the MARSIC300 can be changed in SOPAS ET.
Device language.
Accept with "Initialize“ ( see "Operating states", page 56
)
Use PC time on MARSIC300.
4.7.12
Emitter
Menu: PARAMETERIZATION/EMITTER
This menu serves to view the parameters for the emitter.
4.7.13
Modbus
MARSIC runs as "Slave".
The Modbus process communicates with the device process via shared memory.
A semaphore secures the access on both sides.
The Modbus process communicates with a connected device (e.g., evaluation com‐ puter) via TCP/IP.
Register assignment:
see "Modbus register", page 75
Modbus values and Modbus flags: see "Data", page 47
This menu serves to configure Modbus communication.
Menu: PARAMETERIZATION/MODBUS
8017585/V2-0/2015-09 | SICK
Subject to change without notice
1 Slave address of MARSIC300
Register swap
TCP port (standard: 502)
Simulation mode in which the values in registers 4200 et seq. are written scaled to registers
4000 et seq.
T E C H N I C A L I N F O R M A T I O N | MARSIC300 53
4 MENUS (SOPAS ET)
4.8
4.8.1
4.8.1.1
Adjustment
Lower measured value limit for Simulation mode
Upper measured value limit for Simulation mode
Modbus restart with acceptance of the settings
Parameter
Concentrations of span gases
Menu: ADJUSTMENT/PARAMETER/CONCENTRATIONS
This menu displays the concentrations of the span gases and internal adjustment filter.
The concentrations of the span gases can be set.
1 Enter the concen‐ tration of the test medium
Only internal signif‐ icance
4.8.1.2
Adjustment factors
Menu: ADJUSTMENT/PARAMETER/ADJUSTMENT FACTORS
This menu serves to set the correction factors (adjustment filter wheel optional).
• The measured value is computed with both adjustment factors.
• Factor "Internal adjust." is set automatically to "1.000" when factor "Span gas" is changed.
• Checkmark "Active": This measuring component is computed
4.8.1.3
1 Display: Current factor
Input: New factor
Confirm: Use new factor
Start times
Menu: ADJUSTMENT/PARAMETER/START TIMES
This menu displays the start times of the "Cyclic triggers (CT1 .. CT16)“.
1 Cyclic trigger name
Next start time
Deactivation via operator panel or menu: Parameter/
Variables and functions/
Cyclic trigger (CTi)
54 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
4.8.2
4.8.3
MENUS (SOPAS ET) 4
Manual adjustment
Menu: ADJUSTMENT/MANUAL ADJUSTMENT
Further information, see "MARSIC300 Operating Instructions”
1 Measured value of this component is set to zero
Measured value of this component is set to the nominal concentration of the span gas
Measured values of all components listed above are set to zero
Automatic adjustment
Menu: ADJUSTMENT/AUTOMATIC ADJUSTMENT
Further information, see "MARSIC300 Operating Instructions”
4.9
4.9.1
1 Start adjustment of program displayed
The available programs are system-specific: see System documentation.
Maintenance
Tests
Menu: MAINTENANCE/TESTS
This menu serves to test the digital and analog interfaces.
b Click the desired interface (mark) b Perform with „Test“.
b A menu to set the parameters appears b (Field „Save“ has no significance.)
Test digital inputs
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 55
4 MENUS (SOPAS ET)
Test digital outputs
Test analog inputs
1 Zero: 0 mA/4 mA
Measuring range start
Measuring range end
Test analog outputs
4.9.2
Operating states
Menu: MAINTENANCE/OPERATING STATES
This menu serves to switch the operating states of the MARSIC300 on.
1 0 mA / 4 mA
Topographic addressing
(
Generated automati‐ cally.
Measuring range start
56 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
4.9.3
System maintenance
Menu: MAINTENANCE/MAINTENANCE SYSTEM
This menu serves to start various maintenance procedures.
MENUS (SOPAS ET) 4
1 Switchover LED is on:
Status signal "Mainte‐ nance“ is switched on.
System Stop.
Switch to measuring operation.
Switch to measuring operation (after changes in menu:
Parameterization/Meas‐ uring components/...).
8017585/V2-0/2015-09 | SICK
Subject to change without notice
1 Standby
Switches the system to Standby to put it out of operation for some time.
Exit Standby
Switches the system back to regular measuring operation.
(After it was switched to Standby using item "1" of this menu).
Cancel system maintenance
Abort a program started in this menu.
T E C H N I C A L I N F O R M A T I O N | MARSIC300 57
4 MENUS (SOPAS ET)
4.9.4
Blow back probe
Triggers a backflush of the probe tube (as well as the inlet filter when fitted on the probe tube) with instrument air.
• Press "Start" to start the backflush.
• Operating state: "Maintenance".
• The measuring screen appears with a downwards counter to the end of the adjustment.
• The system switches back to operating state "Measure" when the adjustment has completed (when "Maintenance" was set manually beforehand: Back in
"Maintenance").
Leakage test
Start a leak tightness check.
Test pressure sensors
Checks the pressure sensors.
Perform this check when you have the impression that a pressure sensor is defec‐ tive.
• "Test OK" means: The pressure sensors are OK
• "Test failed" means: One of the pressure sensors is defective. Replace the pressure control module:
Adjust pressure sensors
After replacing the pressure control module: Perform this menu item.
NOTE
Further information, see "MARSIC300 Operating Instructions”
Service log
Menu: MAINTENANCE/SERVICE LOG
Completed maintenance work can be entered in this Table as clear text.
4.9.5
1 Date [dd.mm.yyyy]
Service engineer name
Completed maintenance work in clear text
System restart
Menu: MAINTENANCE/RESTART SYSTEM
This menu serves to restart the system.
1 Start "Restart system"
58 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
4.9.6
Confirm active messages
Menu: MAINTENANCE/CONFIRM ACTIVE MESSAGES
This menu resets all pending active messages.
LED " MAINTENANCE REQUEST“ goes off.
1 Reset active messages
MENUS (SOPAS ET) 4
4.10
Tags (Name)
Tags (name) signify states and variables.
The following Table lists tags relevant for the Measuring screens.
Tag
Operating state
S
Description
Operating state of MARSIC300
Direct value for current input 0..20 mA in mA
Triggering signal for digital output before an inversion that may be set
R/W
R/W
R
1
Input/output interfaces
AO i
(i=1..20) Physical value output (scaled)
Direct value for current output 0..20 mA in mA AO i
O
( i=1..20)
AO i
OR
( i=1..20)
Current active display range
Read in and converted physical value
R/W
1 = Initializing
2 = Heating
3 = Measuring
4 = Manual
5 = System stop
Meas. value
MV i
( i=1..6) Concentration (corrected with all factors) i at the measuring point currently active
Concentration (not corrected) i at the measuring point currently active MV i
CU
( i=1..6)
MV i
AU
( i=1..6)
MV i
AC
( i=1..6)
Absorption (not corrected) i at the measuring point currently active
Extinction (corrected after cross-sensitivity correction) i at the measuring point currently active
Variables
RV01..RV80
Floating point number
BV01..BV150
Boolean variable
FV01..FV20
LV01..LV20
Filter value
Limit value exceeded
R
R
R
R
R
R
R
R
0 = within limit value
1 = limit value exceeded
R
0 = measuring range 1
1 = measuring range 2
R/W AI i
( i=1..48)
AI i
I
( i=1..48)
DO i
( i=1..128)
R
R
R
B
R
B
R
R
R
R
R
R
B
R
R
B
I
I/R/B 2
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 59
4 MENUS (SOPAS ET)
Tag
DO i
O
( i=1..128)
Description
Direct relay state for switching signal output R
DI i
( i=1..64)
Heating control
Input signal after an inversion that may be set
HC i
( i=1..2)
HCPi
( i=1..2)
HCPiC
( i=1..2)
Actual value (temperature) of the internal heating control i
Actual value of internal PID heating control i in the configured unit
Actual value of internal PID heating control i in °C
1
2
R = read, W = write
I = integer value, R = real value, B = Boolean value
R
R
R
R/W 1
R
R/W
R
R
R
I/R/B 2
B
B
60 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
5
5.1
MAINTENANCE 5
Maintenance
Leak tightness check during initial startup
1.
Perform the leak tightness check when the device is running
WARNING
Risk of burns on hot cell
The cell is very hot (approx. 200 °C).
For the leak tightness check, the line on the sample gas outlet must be unscrewed when the cell is hot.
b Use heat-resistant gloves b Use heat-resistant tool
1.
Start program MAINTENANCE/MAINTENANCE SYS./AIRTIGHT TEST
2.
Wait until message "Close outlet - discon. purge" appears
Figure 11: Sample gas outlet (inside view)
1 Sample gas outlet
1 Sample gas outlet at bottom rear of the housing
8017585/V2-0/2015-09 | SICK
Subject to change without notice
1
Rear (thin) flush air line
Instrument air valve (shown
"open")
1 Sample gas outlet on cell
3.
On Analyzer module: Disconnect rear purge air line (press ring to do so)
4.
Close off sample gas outlet gas-tight:
°
Either at the end of the sample gas outlet line (the line does not end at the housing duct but runs through to the cell in the housing)
T E C H N I C A L I N F O R M A T I O N | MARSIC300 61
5 MAINTENANCE
5.2
°
The leak tightness check set has a suitable plug to close off the sample gas outlet line
Or on the cell at the sample gas outlet (10 mm clamping ring sealing plug, the plug is also in the leak tightness check set)
To do this, open the cell: Loosen 4 side screws and remove cover
5.
The pressure in the system slowly rises
Message "close air valve" appears when the pressure is ≥ 1200 hPa (after about
30 seconds) (the current pressure is shown in the "Measured value display")
6.
Close instrument air valve
°
The pressure no longer rises and measurement starts automatically after about 20 seconds: Measurement duration approx. 5 minutes
°
Pressure loss during this time must not exceed 20 hPa. A message is shown:
• "Test OK - open air valve": Test successful
• "Test failed - open air valve": Test unsuccessful: Analyzer switches to
"Maintenance request" state
7.
Open instrument air valve again
8.
Wait until message "Reopen outlet - connect purge" appears
9.
Reconnect rear purge air line
10. Restore sample gas inlet to its original state
Leak tightness check with pressure test tool
This Section describes the leak tightness check with the SICK "Leak test tool kit"
Cap
Probe tube
Sample gas inlets
Span gas inlet
Ejector block
Size
2 x clamping ring screw fittings 8/10
1 x clamping ring screw fitting 4/6
3 x clamping ring screw fitting 4/6
Procedure
1.
Switch analyzer to "Standby": see "System maintenance", page 57
2.
Flush system for 10 minutes in this state
Start removing the probe during the flush duration: See "SFU Gas Sampling Sys‐ tem Operating Instructions"
3.
Close off external instrument air supply
4.
Close probe tube at gas inlet
5.
Open cell enclosure
6.
Unscrew exhaust gas line 1 from sample gas outlet of ejector
62 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
MAINTENANCE 5
1 Ejector sample gas outlet
7.
Connect PTFE connecting hose of the test tool to sample gas output of ejector
8.
Unscrew all further lines marked in the Figure and close connections on the cell gas-tight
9.
Perform a leak tightness check: See "Pressure Test Tool Operating Instructions"
°
°
Test duration: 5 minutes
The "leak" pressure rise must be < 10 mbar (< 0.14 psi). If the pressure rise is higher, search for and clear the leak in the gas path.
10. Connect all lines again after successful leak tightness check
11. Refit the probe
12. Switch standby off again
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 63
6 TECHNICAL DATA
6 Technical data
NOTE
The Technical data depend to some extent on the individual equipment of your ana‐ lyzer.
b See the system documentation provided for the configuration of your analyzer.
6.1
Dimensional drawings
NOTICE
Observe clearances:
• Top: 30 cm
• Bottom: 20 cm
64 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
TECHNICAL DATA 6
6.2
Measuring parameters
Variants
DeSO x
DeNO x
Emission
Number of measured variables
Number of measured variables
Measuring method
Measuring method
Spectral range
Spectral range
Sample volume
Sample volume
NO
NO
2
CO
NH
3
CH
4
H
2
O
Component
SO
2
CO
2
O
2
Measuring point switchover
Measuring point switchover
Measured value characteristics
Measuring principle
Measuring precision
Detection limit
Sensitivity drift
Zero drift
Span drift
Setting time t
90
Components
SO
2
, CO
2
, H
2
O, optional O
2
NO, NO
2
, H
2
O, optional O
2
SO
2
, CO
2
, NO, NO
2
, CO, CH
4
, NH
3
, H
2
O, optional O
2
Max. 9
Hot extractive
2000 ... 11000 nm
200 ... 300 l/h
Measuring range
0 ... 30 ppm; 0 ... 2000 ppm
0 ... 25% by vol.
0 ... 21% by vol.
0 ... 300 ppm; 0 ... 2000 ppm
0 ... 200 ppm; 0 ... 500 ppm
0 ... 200 ppm; 0 ... 2000 ppm
0 ... 50 ppm; 0 ... 500 ppm
0 ... 500 ppm; 0 ... 10000 ppm
0 ... 40% by vol.
Max. 2 measuring points (optional 8 measuring points)
Photometric
< 2% of the respective full scale value
< 2% of the respective full scale value
< 2% of the respective full scale value per week
< 2% of the respective full scale value per week
< 2% of the respective full scale value per week
< 140 s, total measuring path as from probe extraction
6.3
Ambient conditions
Ambient conditions in operation
Installation location
Ambient temperature
Relative humidity
8017585/V2-0/2015-09 | SICK
Subject to change without notice
Below deck
+0 ... +45 °C
< 90% (without condensate)
T E C H N I C A L I N F O R M A T I O N | MARSIC300 65
6 TECHNICAL DATA
6.4
6.5
6.6
Ambient conditions in operation
Air pressure
Degree of protection
Ambient conditions in storage
Ambient temperature
Relative humidity
900 ... 1100 hPa
IP 44
Sample gas conditions
Sample gas at the measuring point Characteristic
Process temperature 10 ... 550 °C
Sample gas temperature subassem‐ bly:
Temperature:
•
• Sample gas probe
Sample gas line
• Cell
Process pressure
Dust load
• Approx. 200 °C
• Approx. 200 °C
• Approx. 200 °C
-20 ... +200 hPa relative
< 200 mg/m 3
Design
-20 ... +70 °C
< 90% (without condensate)
Design
Design
Material, general
Dimensions
Installation
Weight
Materials with media contact
Degree of protection
1 x wall housing
Steel plate, aluminium cast
see "Dimensional drawings", page 64
Wall fitting
Approx. 120 kg
• Stainless steel 1.4571
• PTFE
• Aluminium (coated)
IP 44
Heated sample gas line
Sample gas line
Length
Ambient temperature
Working temperature
Temperature control
Power supply
Power consumption
Degree of protection
Max. 35 m
–20 ... 80 °C
Max. 200 °C
1 x Pt100
1 x additional Pt100 as reserve
115 V or 230 V
90 VA/m
IP 54
66 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
L
Heating
TECHNICAL DATA 6
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 67
6 TECHNICAL DATA
6.7
Tube bundle cable
6.8
Tube bundle cable (system-specific example) with:
- Power supply
- Grounding conductor
- Signal lines
- Tube, gray: Zero gas/span gas
- 2 x tubes, black/blue: Main valve and backflush
1
Power supply 4x1.5 mm2
Lines 1 and 2: Gas sampling system fil‐ ter
Lines 3 and 4: Gas sampling system probe tube
Pt100 lines 4x1.0 mm2
Lines 1 and 2: Gas sampling system fil‐ ter
Lines 3 and 4: Gas sampling system probe tube
Grounding conductor GNYE 4 mm2
PTFE tube DN4/6: Zero gas
PA tube black DN6/8; Imprint "1": Main valve: Control air
Connections on analyzer and gas sam‐ pling probe must match
PA tube blue DN6/8; Imprint "2": Back‐ flush: Backflush air
Connections on analyzer and gas sam‐ pling probe must match
Interfaces and protocols
Operation and interfaces
Operation
Display and input
Analog outputs
Digital inputs/outputs
Data interface
Remote maintenance
Via LC-Display or SOPAS ET software, several operating levels, password-protected
Black-and-white foiled screen with function buttons
Status LEDs
• "Power"
• "Malfunction”
• "Maintenance request"
Optional
Optional
1 x Ethernet (Modbus TCP/IP)
SICK MPR (optional)
68 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
6.9
6.10
TECHNICAL DATA 6
Operation and interfaces
PC operation SOPAS ET via Ethernet
Power supply
Power supply
Supply voltage (preset)
Frequency
Power consumption
• Analyzer
• Heated sample gas line
• Gas sampling system
• Heated probe tube
IT network (without neutral conductor, not grounded)
• 3~230V, PE
• 3~208V, PE
TN(S) network (with neutral conductor, grounded)
• 3~230V, N, PE
• 3~115V, N, PE
Further variants optional (e.g.: 1~230V, 2~208V) b Refer to the system documentation provided for the default value.
50/60 Hz
Power consumption
• Approx. 1000 VA
• Approx. 95 VA/m
• Approx. 450 VA
• Approx. 450 VA
Line cross-sections (relative to leads with ferrules)
•
• CAN
• RS485
• Pt100 inputs
• 24 V DC valve outputs
• Digital inputs
• Relay outputs (potential-free)
• External heating circuits
Power supply
Line cross-section: 0.14 ... 1.5 mm 2
AWG28 ... AWG16
Line cross-section: 0.25 ... 2.5 mm 2
AWG30 ... AWG12
Line cross-section: 0.25 ... 4.0 mm 2
AWG30 ... AWG10
Line cross-section: 0.5 ... 6.0 mm 2
AWG20 ... AWG7
Interfaces (optional)
Digital outputs
Digital inputs
Analog outputs
Electrically isolated, relay changeover contact, 50 V, max. 4 A
Electrically isolated, 24 V, 0.3 A
2 outputs, 0/4 - 20 mA, electrically isolated, max. load 500
Ohm
Torques for screw fittings
Tighten all screw connections, for which no tightening torque or pretension force is specified in drawings or Assembly Instructions, according to VDI 2230.
Exceptions to this rule are all connections with screws that are not screw connections in the real sense. This includes hose clips, cable glands, screw fittings, gas connections, screws for circuit boards etc. Tighten these screw fittings as evenly as possible with a much lower torque (hose clips 1 Nm, other screw fittings according to manufacturer specifications).
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 69
6 TECHNICAL DATA
20
22
24
27
8
10
12
14
16
18
30
33
36
39
5
6
3.5
4
1.6
2
2.5
3
Select the next lowest torque valid for the screw for mixed materials and special screws such as relieved screws.
The friction coefficient serving as basis is (screw fitting without lubrication)
µk=µG=0.14. The calculated values are valid for room temperature (T=20°C).
Size M Pitch P Tightening torque Ma (Nm)
3.6
4.6
5.6
12.9
4
4
3.5
3.5
3
3
2.5
2.5
1.25
1.5
1.75
2
2
2.5
0.6
0.7
0.8
1
0.4
0.45
0.45
0.5
0.05
0.1
0.21
138
186
235
350
475
645
1080
1330
46
71
97
3.5
8.4
17
29
0.54
0.85
1.02
2
184
250
315
470
11
22
39
62
95
130
1.3
2
2.7
4.6
0.05
0.11
0.23
1
635
865
1440
1780
425
580
730
1100
1450
2000
2600
3400
85
135
210
300
5
10
25
49
8.8, A2 u.
A4-80
0.17
0.35
0.73
1.3
1.9
2.5
10.9
610
830
1050
1550
2100
2800
3700
4800
15
36
72
125
1.7
2.6
4.4
8.7
200
310
430
43
84
145
235
365
500
710
970
1220
1800
3.2
5.1
10
18
0.28
0.6
1.23
2
2450
3400
4300
5600
6.11
Electrical connections in analyzer
Power supply - connection / fuses
The power supply is located on the left on the analyzer.
Name
MAINS USV (3-pole)
MAINS (5-pole)
70 T E C H N I C A L I N F O R M A T I O N | MARSIC300
Supply
Power supply for electronics unit (internal)
External power supply
8017585/V2-0/2015-09 | SICK
Subject to change without notice
TECHNICAL DATA 6
Name
F1
F2
Electronics fuses
Supply
Internal
Internal
Name
ELECTRONIC
TUBE 1
FILTER/PROBE 1
TUBE 2
FILTER/PROBE 2
TUBE 3
CELL
DEVICE
Connections for heated components
Fuse for
Electronics
Sample gas line 1
Filter heater 1
Sample gas line 2
Filter heater 2
Sample gas line 3
Sample gas cell
Device
Plug
TUBE 1
Subassembly
Sample gas line 1
FILTER1 Gas sampling system filter 1
(Lines from tube bundle cable)
PROBE1
TUBE2
FILTER2
PROBE2
8017585/V2-0/2015-09 | SICK
Subject to change without notice
Gas sampling system probe tube 1
(Lines from tube bundle cable)
Sample gas line 2
Gas sampling system filter 2
Gas sampling system probe tube 2
Pin
1
1
2
2
3
3
4
5
6
1 ... 3
1 ... 3
4 ... 6
Assignment
L (L)
N (L)
PE
L (L)
N (L)
PE
L (L)
N (L)
PE (not connected)
As for TUBE1
As for FILTER1
As for PROBE1
Tube bundle cable line number
4x1.5
mm
1x4
2 mm 2
4x1.5
mm 2
T E C H N I C A L I N F O R M A T I O N | MARSIC300
1
2
1
GNYE
3
4
71
6 TECHNICAL DATA
Plug Subassembly
TUBE3 Sample gas line 3
1 The connections must match the connections on the gas sampling system
Connections for interfaces and SD card
Pin Assignment Tube bundle cable line number 1
Plug
ETH0
ETH1
USB
SD card
CAN1
CAN2
RS422, RS488
RS232 (top plug)
O2 (bottom plug)
DISP (top plug)
I/O-MOD (bottom plug)
Connection for
Ethernet (e.g. SOPAS ET), MPR (remote maintenance)
Internal
Internal
SD card (on the right, next to USB)
Internal
Internal
Internal
Internal
O
2
sensor
Display
Internal
Pt100 and signal connections
Plug Subassembly Pin Assignment Tube bundle cable line number 1
Pt100 Sample gas line 1
Gas sampling system filter 1
Gas sampling system probe tube 1
Not connected
Sample gas line 2
Gas sampling system filter 2
Gas sampling system probe tube 2
Sample gas line 3
1
2
3
Pt100 +
Pt100 –
Pt100 +
6
7
4
5
Pt100 –
Pt100 +
Pt100 –
8
9, 10 As above
11, 12 As above
13, 14 As above
15 Pt100 +
16 Pt100 –
4x1.0
mm 2
1
2
3
4
4x1.0
mm 2
72 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
6.12
TECHNICAL DATA 6
Plug Subassembly Pin Assignment Tube bundle cable line number 1
DIGITAL
INPUTS
DIGITAL
OUTPUTS
Digital input 1
Digital input 2
Digital input 3
Digital input 4
Digital output 1
Digital output 2
Digital output 3
Digital output 4
Valves
3
4
1
2
5 ... 8
+ 24 V
+ Signal
- Signal
GND
As above
1
2
9 ... 12 As above
13 ... 16 As above
NC
COM
3
4 ... 6
7 ... 9
NO
As above
As above
10 ... 12 As above
Internal VALVE
OUTPUTS
1 The connections must match the connections on the gas sampling probe
Circuit breakers
The circuit breakers are located at the bottom of the electronics unit.
The circuit breakers are labeled
When a circuit breaker has triggered: b Press the circuit breaker pin back in again.
If this does not work: w Wait for a few minutes (cooling down phase) and then press the pin back in again
If this does not work: Check the subassembly and replace when necessary
6.13
Supply gases
NOTICE
Risk of contamination of analyzer b Observe the specified quality of the instrument air b If required, provide for instrument air conditioning
Gas
Instrument air (zero gas quality)
Quality
Particle size max. 1 μm
Oil content max. 0.1 mg/m3
Pressure dew point max. –
40 °C
Purity class 2 (ISO 8573)
Inlet pressure
600 ... 700 kPa
(6.0 ... 7.0 bar)
Flow rate
Approx. 350 l/h
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 73
6 TECHNICAL DATA
6.14
Gas
Instrument air solely as induction air for ejector
External span gas
Quality
Particle size max. 5 μm
Oil content max. 1 mg/m3
Pressure dew point max.
+3 °C
Purity class 3 (ISO 8573)
Precision: ± 2 %
Concentration: 80% ... 100% of measuring range
The span gas must comply with the specifications of the standards to be applied (e.g.,
MARPOL Annex VI)
Inlet pressure
500 ... 700 kPa
(5.0 ... 7.0 bar)
Max. 400 kPa
(4.0 bar)
Tube connections
Connection
Sample gas inlet
Ejector induction air
Span gas inlet
Gas outlet
Flow rate
Approx. 1300 l/h
Approx. 350 l/h
Dimension
Clamping ring screw connection 6 mm
DN 6/8
Clamping ring screw connection 6 mm
DN 8/10
74 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
ANNEX 7
7
7.1
7.1.1
7.1.2
Annex
Modbus register
Device status (Discrete Inputs [1xxxx], Function Code 02)
Discrete Input No.
1000
1001
1002
1003
1004
Date/Time
Device group status malfunction
Device group status maintenance request
Device group status maintenance
Device group status outside specification
Measured data are live/current (0 = held, 1 = alive)
1008
1009
1010
1011
1012
1013
1014
1015
1016
1017
1018
1019
1024
1025
1026
Feed via measuring point_01 "Measured value from measuring point 1"
Feed via measuring point_02
Feed via measuring point_03
Feed via measuring point_04
Feed via measuring point_05
Feed via measuring point_06
Feed via measuring point_07
Feed via measuring point_08
Feed via measuring point_09
Feed via measuring point_10
Feed via measuring point_11
Feed via measuring point_12
Zeroize (Val./Adj.)
Span check (Val./Adj.)
Zeroize/span check (Val./Adj.), total signal
Measured variables (Input Register [3xxxx], Function Code 04)
If no measuring: Register = "0"
Input Register No
1000 [Float]
1002 [Float]
1004 [Float]
1006 [Float]
1008 [Float]
1010 [Float]
1012 [Float]
1014 [Float]
1016 [Float]
1018 [Float]
1020 [Float]
1022 [Float]
Measured variable
SO2
NH3
CO
CH4
O2
VOC
CO2
Ratio SO2/CO2
H2O
NO
NO2
NOx
1050 [DInt]
1052 [DInt]
Measured values are from measuring point "x" and alive.
"0" = maintenance or held measured value
Measured values are from measuring point "x" and alive.
"0" = maintenance
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 75
7 ANNEX
7.1.3
7.1.4
76
Input Register No
1054 [Float]
1056 [Float]
1058 [Float]
Measured variable
Optional: [pH washing water]
Optional: [PAH washing water (oil content)]
Optional: [Clouding washing water]
Device-internal monitoring values (Input Register [3xxxx], Function Code 04)
Input Register No Internal variable
2000 [Float] Sample gas flow
2002 [Float] T_Cell (heating measuring cell)
2004 [Float] T_Optics (heating optics)
2006 [Float] T_Ext1 (external heating 1)
2008 [Float] T_Ext2 (external heating 2)
... ...
2018 [Float] T_Ext7 (external heating 7)
2020 [Float] T (ambient)
2022 [Float] T (LPMS01)
2024 [Float] T (LPMS02)
2026 [Float] T (LPMS03)
2028 [Float] p(0)
2030 [Float] p(2)
2032 [Float] p (ambient)
2034 [Float] Δp
3000 [Float] SO2 low [0..500]
3002 [Float] SO2 high [0..2000]
Triggering the MARSIC 300 (Coils [0xxxx], Function Code 15, write multiple Coils)
Coil No.
Action
2000 Select measuring point_01
2001 Select measuring point_02
2002 Select measuring point_03
2003 Select measuring point_04
2004 Select measuring point_05
2005 Select measuring point_06
2006 Select measuring point_07
2007 Select measuring point_08
2008 Select measuring point_09
2009 Select measuring point_10
2010 Select measuring point_11
2011 Select measuring point_12
2012 Trigger Standby
2013 Trigger BlowBack
2014 Trigger Zeroize
2015 Trigger Span Set (internal)
2016 Trigger zero validation
2017 Trigger span validation (internal)
2018 Trigger adjustment O2
T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
7.1.5
ANNEX 7
Remarks:
• The external selection is activated as soon as release signal '1' is set. If this is not done, the measuring points are automatically triggered sequentially.
• Standby: 1 = Standby, 0 = Measuring operation
• Zero and span set: "1" The adjustment/validation process starts for 5 seconds; the trigger signal must then be disabled
VDI 4301 conform range (Holding Register [4xxxx], Function Code 03)
Holding Register No.
4000 [32 Bit Float, scaled according to VDI]
4002 [32 Bit DInt]
4004 [32 Bit Float, scaled according to VDI]
4006 [32 Bit DInt]
4008 [32 Bit Float, scaled according to VDI]
4010 [32 Bit DInt]
4012 [32 Bit Float, scaled according to VDI]
4014 [32 Bit DInt]
4016 [32 Bit Float, scaled according to VDI]
4018 [32 Bit DInt]
4020 [32 Bit Float, scaled according to VDI]
4022 [32 Bit DInt]
4024 [32 Bit Float, scaled according to VDI]
4026 [32 Bit DInt]
4028 [32 Bit Float, scaled according to VDI]
4030 [32 Bit DInt]
4032 [32 Bit Float, scaled according to VDI]
4034 [32 Bit DInt]
4036 [32 Bit Float, scaled according to VDI]
4038 [32 Bit DInt]
4040 [32 Bit Float, scaled according to VDI]
4042 [32 Bit DInt]
4044 [32 Bit Float, scaled according to VDI]
4046 [32 Bit DInt]
Measured variable
SO2
Device status
CO2
Device status
Ratio SO2/CO2
Device status
H2O
Device status
NO
Device status
NO2
Device status
NOx
Device status
NH3
Device status
CO
Device status
CH4
Device status
O2
Device status
VOC
Device status
4064 [32 Bit Float, scaled according to VDI]
4066 [32 Bit DInt]
4068 [32 Bit Float, scaled according to VDI]
4070 [32 Bit DInt]
4072 [32 Bit Float, scaled according to VDI]
4074 [32 Bit DInt]
4076 [32 Bit Float, scaled according to VDI]
4078 [32 Bit DInt]
4080 [32 Bit Float, scaled according to VDI]
4082 [32 Bit DInt]
Number of active measuring point
Device status
Number of triggered measuring point
Device status pH washing water
Device status
PAH washing water (oil content)
Device status
Clouding washing water
Device status
Remark: The device status is added for each measured value for conformity reasons.
Device status format. The individual status has the same significance as the device sta‐ tus of Discrete Inputs 1000 et seq.
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 77
7 ANNEX
1
2
Bit No.
0
3
4
Malfunction
Maintenance
Maintenance request
Outside the specification
Test operation according to VDI
7.2
Error messages and possible causes
Current pending messages are shown on the device display.
A combined list of messages is included in SOPAS ET (see "MARSIC300 Technical Data").
NOTE
The following Table only includes those messages with classification "X" that are important for information.
Messages not included in the following Table have no further significance for operation.
NOTE
Display of current device state data: Menu Diagnosis/System param.
Trigger: System
C = Classification:
"F" = Failure
"M": Maintenance request
"U": Uncertain
"E": Extended
78 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
ANNEX 7
Code
S001
Error text
Temperature too high
S002 Temperature too low
S004 Flow too low
C
F
Description
Measuring cell temperature too high
Optic head temperature too high
Temperature of heating for a subassembly too high
Possible clearance
When T < 356 °C: Replace electronics unit.
When T >= 356 °C: Check plug-in connector on electronics unit.
When plug OK: Replace cell.
When T < 356 °C:
When housing temperature >= 55 °C: Check housing fan.
When housing temperature < 55 °C: Replace electronics unit.
When T >= 356 °C: Check plug-in connector on electronics unit.
When plug OK: Replace Analyzer module.
Check device documentation to clarify which subassembly is affected.
F After x minutes
When T < 356 °C: Replace electronics unit.
When T >= 356 °C: Check subassembly plugin connector.
When plug OK: Replace subassembly
LPMS01 (1/2 control) temperature too high When housing temperature < 55°C: Electron‐ ics unit fan functioning?
Yes: Replace electronics unit.
No: Replace Analyzer module.
LPMS02 (power electronics) temperature too high
When housing temperature >= 55 °C: Check housing fan.
When housing temperature < 55 °C: Replace electronics unit.
When housing temperature >= 55 °C: Check housing fan.
Check system documentation to clarify which subassembly is affected (heating circuit 1 ..7).
F Flow too low
• Temperature displayed < -30 °C: Pt100 short circuit: Replace subassembly
For heated sample gas line: Connect reserve Pt100
For analyzer: Replace Analyzer module
For optic head: Replace Analyzer mod‐ ule
• Reset on circuit breaker under electron‐ ics unit possible: Check all cables
involved for damage ( see "Circuit break‐ ers", page 73
)
Check all plugs are plugged correctly.
• Reset not possible: Replace subassem‐ bly affected
Sample gas flow and instrument air flow too low: Replace cell
Sample gas flow too low and instrument air flow OK: Gas sampling system defective
Instrument air flow too low and sample gas flow OK: Check all hose connections.
When all hose connections OK: Replace valve block.
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 79
7 ANNEX
Code
S005
S006
S008
S009
S010
S011
S012
S013
S014
S015
S016
S024
S025
S026
S033
S034
S035
Error text
Pressure too high
Pressure too low
Chopper
Motor filter wheel 1
Motor filter wheel 2
Motor filter wheel 3
Emitter
5 Volt power
24 Volt power
Detector signal
Ref. energy too low
F Filter wheel motor does not detect reference position
F
F
F
F
F
C Description
F Pressure too high
F Pressure too low
F
No active component F When all "active" checkmarks of all compo‐ nents are inactive
Evaluation module failure F Evaluation module could not be started
Evaluation mod. file error F Files for Evaluation module not created
(espec, config, condition, measval)
Possible clearance
Only instrument air pressure too high:
• Check and set connected instrument air.
• Set correct pressure on pressure reducer unit.
Only sample gas pressure too high:
• Set sample gas pressure within device specification
Instrument air and sample gas pressure too high:
• Exhaust gas hose crimped/blocked
• Counter-pressure in exhaust duct too high
• Check all hose connections
If this does not work:
• Replace the pressure control module
• Otherwise: Replace Analyzer module
Replace pressure control module.
24 V power supply malfunction: Replace elec‐ tronics unit.
24 V power supply OK: Replace Analyzer mod‐ ule.
24 V power supply malfunction: Replace elec‐ tronics unit.
24 V power supply OK: Replace Analyzer mod‐ ule.
Emitter voltage incorrect: Replace electronics unit.
Emitter voltage OK and emitter power incor‐ rect: Replace emitter.
Replace electronics unit.
Replace electronics unit.
Replace Analyzer module.
If further error messages are pending: Clear corresponding error.
If no further error messages are pending:
Replace cell.
Check in SOPAS ET.
"Load backup": Menu Maintenance/Save parameters.
If error remains: "Load default setting" .
If error remains: Replace electronics unit.
Dev. zero point too high M Configured for measuring component Check zero gas for pressure and cleanness.
Perform maintenance on compressed air con‐ ditioning unit
Perform manual zero point adjustment (menu:
Adjustment/Zero point).
If deviation still too high: Replace Analyzer module
Check I/O module, check configuration: IO hardware plan
Configuration I/O module M CONF (I/O) configuration error, module found does not correspond to nominal configura‐ tion
Ref. energy too low M If further error messages are pending: Clear corresponding error.
If no further error messages are pending:
Replace cell.
80 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
ANNEX 7
S048
S049
S050
S057
S058
S072
S089
S090
S091
S092
S093
S094
S095
S096
S097
S098
S112
S113
Code
S036
S038
S039
S040
S041
S042
S043
S045
S046
S047
S114
S115
S116
Error text
O2 sensor failure
Channel 1 error
Channel 2 error
C Description
M Error bit O2 Error=1
M OV0 (I/O) signals that the desired current is
M not reached on Analog module connection
(node y, module z).
M Flow too high
Possible clearance
Please contact SICK Customer Service
Check I/O module, cable damage
Flow too high Test pressure sensors with program "Mainte‐ nance/Maint Sys./Test pressure sensors".
If this does not work: Replace pressure control module.
Please contact SICK Customer Service Flow too low
Controller IO or HC busy
Emitter weak dev. span gas adjust too high
M Flow too low
M BSY (I/O and HC3X) signals that the module microcontroller is still executing the previous command
M
M When F_Medium calculation is refused because outside tolerable range; Configured for measuring component
Replace emitter.
Perform adjustment with zero gas and span gas.
If error remains: Replace Analyzer module.
Perform "Internal reference" adjustment.
If error remains: Replace Analyzer module.
dev. Int. adjust too high dev. O2 adjust too high
M When F_Medium calculation is refused because outside tolerable range; Configured for measuring component
M When F_Medium calculation is refused because outside tolerable range; Configured for measuring component alarm O2 measure value M
FlashCard not recognized M FlashCard not recognized adjust factor is Zero M If one of factors F_Medium or F_Filter in range -0,000001 < x < 0,000001
Energy too high
Energy too low
U When at least one energy value > 5*Energie‐
MAX
U Energy too low
Module not found
Zero
AF
Communication problem
Adjustment canceled
Dark measurement
System start
Adjust. zero canceled
Backup done
Backup denied
AM failure IO-Node
Check sum error
Communication error
High/low voltage
Output without current
E
E
E
E
E
I/O (EXIST)
New zero recorded
F_Filter has been recalculated
Internal communication problem
Adjustment canceled
Check span gas, check entry for span gas con‐ centration.
Please contact SICK Customer Service
Check span gas, check entry for span gas con‐ centration
Please contact SICK Customer Service
If a further emitter error is pending: Replace emitter.
Otherwise replace Analyzer module
Please contact SICK Customer Service
Extinction value set to zero is displayed
New factor determined and measured value from adjustment displayed
Please contact SICK Customer Service
Restart adjustment. If this does not work:
Please contact SICK Customer Service
E
E
E
E
E
E
E
F BCK (I/O) shows the transfer made from
Master to Slave (controller) had an incorrect checksum and the Slave did not save the data.
F COM (I/O) communication error with an /I/O module.
F PF0 (I/O) signals the internal voltage monitor‐ ing of the 5 V and 24 V supply voltages exceed or underflow a range.
F TOO (HC3X)
Check I/O module, cable damage
Please contact SICK Customer Service
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 81
7 ANNEX
This Table contains solution proposals that can only be processed by specially trained personnel.
E039
E040
E097
E098
E099
E100
E101
E102
E103
E030
E031
E032
...
E034
E035
E036
E037
E038
E023
E024
E025
E026
E027
E028
E029
Code
E001
E002
E003
E004
E005
E006
E007
...
E009
E010
...
E012
E013
...
E021
E022
Trigger: Evaluation process
Error text
Operating system error
Temp. too low
Incorrect configuration
Incorrect configuration
Internal file error
Incorrect configuration
Internal file error
Incorrect configuration
Internal file error
Resolution too high/low
Numeric error
Incorrect configuration
Internal file error
Numeric error
Incorrect configuration
Incorrect configuration
Unknown error
Operating system error
Operating system error
Internal file error
Numeric error
Syntax error
Error during processing
Extinction too high
Internal file error
Internal file error
Evaluation uncertain
Medium temp. too high/low
Medium pressure too high/low
Medium flow too high/low
Measured value too high/low
Evaluation uncertain
Evaluation uncertain
C Possible clearance
U Please contact SICK Customer Service
82 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
ANNEX 7
Trigger: Sequence control programs
Code Triggered by subassem‐ bly
Internal heating
C Message
F Alarm from "device" M001
...
M009
M010
...
M029
M034
...
M045
M046
...
M057
External heating
System
Measuring point 1
...
Measuring point 12
F Alarm from "device"
X System xx disabled by user
M Flow alarm (measuring)
Code Triggered by subassem‐ bly
M058 System
C Message
F Flow alarm (measuring)
Clearance see above: S004
Next measuring point activation after message acknowl‐ edged
Clearance
Clearance see above: S004
Next measuring point activation after message acknowl‐ edged
No action necessary M060 Program
M062 Leak tightness check
M063
M064
M065
M066
M067
M Adjust zero not started
X Test passed
M Test failed
Pressure not reached
Air valve not closed
Leakage > Limit
X Deviation="xx"
Clearance
" device " = triggering subassembly
Clearance see above: S001 and S002
" device " = triggering subassembly
Clearance see above: S001 and S002
No action required
Code Triggered by subassem‐ bly
M069 Debug
M070 Light Source
M071 Filter Unit
M072 Valve driver module
C Message
X Internal message
M Lifetime exceeded
M Lifetime filter exceeded
F Temperature > Limit
M073 Power Supply X 115V
M074 Program
M075
M076 CELL
X Stop by internal failure
X Cancelled by user
M Lifetime filter exceeded
Code
M086 Pressure
M087
M088
Triggered by subassem‐ bly
C Message
X Sensors ok
X Sensors adjusted
M Sensors not OK
Repeat test and observe messages displayed;
Check seating of connections, disconnect heating hose from cell and close sample gas inlet off with dummy plug
(from leak tightness check set)
If leaky: Exchange pressure control module and cell other‐ wise exchange gas sampling system
Serves as information
"xx" = pressure loss [hPa] during measuring time.
Clearance
No action necessary
Replace emitter
Replace gas sampling system filter
Other temperature error pending? Then see S001 above;
Otherwise replace valve block
Serves as information
No action necessary
No action necessary
No action necessary
Replace cell inlet filter
Clearance
No action necessary
Sample gas outlet open to environment?
If no blockages: Exchange pressure control module
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 83
7 ANNEX
Code Triggered by subassem‐ bly
M089 Measuring point
C Message
M All disabled
M090
...
M101
M102
...
M113
System
Clearance
X Measuring sample point 1
...
Measuring sample point
12
X SP1 disabled by ext signal
...
SP12 disabled by ext sig‐ nal
See additional message: Clear pending error; check exter‐ nal signal
No action necessary
"SPx" = measuring point x
No action necessary
84 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
INDEX 8
8 Index
A
Adjustment factors............................................................................. 54
Automatic adjustment........................................................................ 55
B
C
Checklist for mechanical and electrical installation......................... 11
Concentrations of span gases........................................................... 54
Configuring the Measuring screens................................................... 44
Confirm active messages................................................................... 59
Connect Ethernet................................................................................ 27
Connect heated sample gas line ................................................ 26, 29
Connecting the tube bundle cable to the analyzer........................... 19
Connect instrument air ...................................................................... 23
Connect MPR ..................................................................................... 28
Connect power supply ....................................................................... 26
Connect sample gas outlet................................................................ 25
Connect signal line............................................................................. 27
Connect span gas............................................................................... 24
Connect tube bundle cable......................................................... 26, 29
Counters (in menus)........................................................................... 31
D
Dimensional drawings........................................................................ 64
downwards counter ............................................................................ 31
E
Electrical connections........................................................................ 70
Error messages................................................................................... 78
I
F
Fitting system components................................................................ 11
Installing the gas sampling system................................................... 29
Instrument display.............................................................................. 52
Internal adjustment drift.................................................................... 38
L
Laying the tube bundle cable............................................................. 15
Leak tightness check.......................................................................... 61
Leak tightness check with pressure test tool.................................... 62
Line writer ............................................................................. 36, 37, 46
M
Manual adjustment............................................................................ 55
8017585/V2-0/2015-09 | SICK
Subject to change without notice
Measured value box..................................................................... 35, 45
Measuring screen .............................................................................. 35
Menu tree (SOPAS ET)........................................................................ 34
Modbus register.................................................................................. 75
Modbus values.................................................................................... 48
N
O
of the tube bundle cable sheath........................................................ 20
Operating hours counter.................................................................... 43
Operating state ............................................................................ 43, 56
P
Parameter Measuring Screen..................................................... 44, 52
Power supply................................................................................. 69, 69
Power supply - connection.................................................................. 70
R
Reference energy................................................................................ 38
remote maintenance.......................................................................... 28
S
Safety symbols on device..................................................................... 7
Sampling point program..................................................................... 49
Save parameters ................................................................................ 33
Scaling Measuring screens................................................................ 37
Sequence control program................................................................. 49
Set the IP address............................................................................... 31
Software version................................................................................. 43
Start a leak tightness check.............................................................. 58
System name...................................................................................... 53
system parameters............................................................................. 52
System status............................................................................... 43, 43
T
Temperature control........................................................................... 51
T E C H N I C A L I N F O R M A T I O N | MARSIC300 85
8 INDEX
Tube bundle cable.............................................................................. 68
Z
zero gas quality............................................................................ 24, 73
86 T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/V2-0/2015-09 | SICK
Subject to change without notice
INDEX 8
8017585/V2-0/2015-09 | SICK
Subject to change without notice
T E C H N I C A L I N F O R M A T I O N | MARSIC300 87
Australia
Phone +61 3 9457 0600
1800 334 802 – tollfree
E-Mail [email protected]
Austria
Phone +43 (0)22 36 62 28 8-0
E-Mail [email protected]
Belgium/Luxembourg
Phone +32 (0)2 466 55 66
E-Mail [email protected]
Brazil
Phone +55 11 3215-4900
E-Mail [email protected]
Canada
Phone +1 905 771 14 44
E-Mail [email protected]
Czech Republic
Phone +420 2 57 91 18 50
E-Mail [email protected]
Chile
Phone +56 2 2274 7430
E-Mail [email protected]
China
Phone +86 4000 121 000
E-Mail [email protected]
Denmark
Phone +45 45 82 64 00
E-Mail [email protected]
Finland
Phone +358-9-2515 800
E-Mail [email protected]
France
Phone +33 1 64 62 35 00
E-Mail [email protected]
Gemany
Phone +49 211 5301-301
E-Mail [email protected]
Great Britain
Phone +44 (0)1727 831121
E-Mail [email protected]
Hong Kong
Phone +852 2153 6300
E-Mail [email protected]
Hungary
Phone +36 1 371 2680
E-Mail [email protected]
India
Phone +91–22–4033 8333
E-Mail [email protected]
Israel
Phone +972-4-6881000
E-Mail [email protected]
Italy
Phone +39 02 27 43 41
E-Mail [email protected]
Japan
Phone +81 (0)3 5309 2112
E-Mail [email protected]
Malaysia
Phone +603 808070425
E-Mail [email protected]
Netherlands
Phone +31 (0)30 229 25 44
E-Mail [email protected]
New Zealand
Phone +64 9 415 0459
0800 222 278 – tollfree
E-Mail [email protected]
Norway
Phone +47 67 81 50 00
E-Mail [email protected]
Poland
Phone +48 22 837 40 50
E-Mail [email protected]
Romania
Phone +40 356 171 120
E-Mail [email protected]
Russia
Phone +7-495-775-05-30
E-Mail [email protected]
Singapore
Phone +65 6744 3732
E-Mail [email protected]
Slovakia
Phone +421 482 901201
E-Mail [email protected]
Slovenia
Phone +386 (0)1-47 69 990
E-Mail [email protected]
South Africa
Phone +27 11 472 3733
E-Mail [email protected]
South Korea
Phone +82 2 786 6321
E-Mail [email protected]
Spain
Phone +34 93 480 31 00
E-Mail [email protected]
Sweden
Phone +46 10 110 10 00
E-Mail [email protected]
Switzerland
Phone +41 41 619 29 39
E-Mail [email protected]
Taiwan
Phone +886 2 2375-6288
E-Mail [email protected]
Thailand
Phone +66 2645 0009
E-Mail [email protected]
Turkey
Phone +90 (216) 528 50 00
E-Mail [email protected]
United Arab Emirates
Phone +971 (0) 4 88 65 878
E-Mail [email protected]
USA/Mexico
Phone +1(952) 941-6780
1 (800) 325-7425 – tollfree
E-Mail [email protected]
Vietnam
Phone +84 8 62920204
E-Mail [email protected]
More representatives and agencies at www.sick.com
SICK AG | Waldkirch | Germany | www.sick.com
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 43 System info
- 46 Hardware Plan
- 53 Modbus
- 58 System restart
- 61 Maintenance
- 61 Leak tightness check during initial startup
- 64 Technical data
- 66 Design