SMX2064/2060, SM2064/2060 Series

SMX2064/2060, SM2064/2060 Series
SMX2064/2060, SM2064/2060 Series
7-1/2 Digit Digital Multimeters
Introduction
SMX2064
The SMX2064/2060 and SM2064/2060 series are 7-1/2 digit high-speed digital multimeters in PXI and PCI form factors,
which maintain high accuracy at high measurement rates. The measurement speed is up to 20,000 readings/second
for the SMX2064 and SM2064. The SMX206x and SM206x series provides a comprehensive set of DMM capabilities,
including 2-wire, 4-wire, and 6-wire guarded resistance measurements; inductance and capacitance; leakage and temperature;
frequency and timing; sourcing of voltage and current; and much more. The SMX2064/2060 and SM2064/2060 series
digital multimeters are easy to setup and use, have sophisticated analog and digital circuitry to provide repeatable
measurements, and are protected to handle any unexpected situation your measurement environment may encounter.
With high performance and variable applications, the SMX2064/2060 and SM2064/2060 series are suitable for automated
production testing, laboratory automation, and portable/field testing.
Specifications
SM2060
Specifications subject to change without notice.
For the most current and complete specifications, please refer to the user manual.
DC Functions
Features
■
■
■
■
■
■
■
■
■
■
■
■
■
Flexible, full-featured auto-ranging DMM
7-1/2 digit resolution
20,000 readings/second (SMX2064, SM2064)
1,400 readings/second (SMX2060, SM2060)
DC & AC voltage & current, 2-wire, 4-wire,
6-wire (SMX2064, SM2064) ohms
Frequency and time measurements
Capacitance & inductance measurements
(SMX2064, SM2064)
Voltage & current sourcing (SMX2064, SM2064)
Precise trigger capabilities
330 V isolation barrier
Self-calibrating
Plug-and-Play software with graphical user interface
Fully programmable with Visual Basic, C++, LabVIEW™,etc.
■ Operating Systems
• Windows 98/NT/2000/XP/2003
DC Voltage
240 mV
Accuracy ± (% of reading + Volts) [1]
Full scale
7-1/2 Digits
Resolution
Input
24 hours
90 Days
One Year
Resistance
23°C±1°C
23°C±5°C
23°C±5°C
0.005 + 2 µV
240 mV
240.00000 mV
10 nV
>10 GΩ
0.003 + 1 µV
0.004 + 1.5 µV
2.4 V
2.4000000 V
100 nV
>10 GΩ
0.002 + 3 µV
0.0025 + 5 µV
0.003 + 5 µV
24 V
24.000000 V
1 µV
10 MΩ
0.004 + 120 µV
0.005 + 130 µV
0.006 + 150 µV
240 V
240.00000 V
10 µV
10 MΩ
0.003 + 250 µV
0.004 + 300 µV
0.005 + 0.5 mV
330 V
330.00000 V
10 µV
10 MΩ
0.005 + 550 µV
0.01 + 700 µV
0.015 + 0.8 mV
[1] With Aperture set to ≥ 0.5 sec, and within one hour from Self Calibration (S-Cal)
Accuracy ±(% of reading + Amps) [1]
DC current
240 mV
Full scale
7-1/2 Digits
Resolution
Max Burden
24 hours
90 Days
One Year
Voltage
23°C±1°C
23°C±5°C
23°C±5°C
240 nA[2]
240.0000 nA
2.4 µA[2]
2.400000 nA
0.1 pA
100 µV
0.07 + 40 pA
0.1 + 45 pA
0.17 + 60 pA
1 pA
100 µV
0.05 + 70 pA
0.08 + 90 pA
0.21 + 150 pA
0.13 + 0.8 nA
24 µA[2]
24.00000 µA
10 pA
100 µV
0.05 + 400 pA
0.08 + 600 pA
240 µA[2]
240.000 µA
10 nA
2.5 mV
0.052 + 200 nA
0.07 + 300 nA
0.1 + 400 nA
2.4 mA
2.40000 mA
10 nA
25 mV
0.05 + 300 nA
0.06 + 400 nA
0.07 + 550 nA
24 mA
24.0000 mA
1 µA
250 mV
0.05 + 350 nA
0.065 + 450 nA
0.08 + 550 nA
240 m
240.000 mA
100 nA
55 mV
0.05 + 50 µA
0.055 + 60 µA
0.065 + 80 µA
2.4 A
2.40000 A
10 µA
520 mV
0.3 + 60 µA
0.4 + 70 µA
0.45 + 90 µA
[1] With Aperture set to ≥0.96 sec, and within one hour from zero (Relative control).
[2] Available only with the SMX2064 and SM2064.
2-Wire Resistance
■ Recommended Software
• VB/VC++/BCB/Delphi
• DAQBench
Accuracy ± (% of reading + Ω) [1]
Range [3]
Full scale
7-1/2 Digits
Resolution
Source
24 hours
90 Days
One Year
Current
23°C±1°C
23°C±5°C
23°C±5°C
0.008 + 2 mΩ
24 Ω[2]
24.000000 Ω
1 µΩ
10 mA
0.0038 + 1.4 mΩ
0.005 + 1.6 mΩ
240 Ω
240.00000 Ω
10 µΩ
1 mA
0.0037 + 4.5 mΩ
0.0046 + 5 mΩ
0.007 + 6 mΩ
2.4 kΩ
2.4000000 kΩ
100 µΩ
1 mA
0.0023 + 28 mΩ
0.004 + 32 mΩ
0.006 + 33 mΩ
24 kΩ
24.000000 kΩ
1 mΩ
100 µA
0.0025 + 300 mΩ
0.004 + 330 mΩ
0.006 + 350 mΩ
240 kΩ
240.00000 kΩ
10 mΩ
10 µA
0.0055 + 3.2 Ω
0.006 + 4 Ω
0.007 + 5 Ω
2.4 MΩ
2.4000000 MΩ
100 mΩ
1 µA
0.018 + 40 Ω
0.03 + 50 Ω
0.04 + 70 Ω
24 MΩ
24.0000 MΩ
100 Ω
100 nA
0.12 + 400 Ω
0.13 + 500 Ω
0.2 + 600 Ω
240 MΩ[2]
240.000 MΩ
1 kΩ
10 nA
0.8 + 20 kΩ
1.0 + 30 kΩ
1.3 + 50 kΩ
[1] With Aperture set to ≥0.5 Sec, and within one hour from Self Calibration (S-Cal).
[2] Ranges are only with the SMX2064, SM2064.
[3] Test voltages are 2.4 V max with the exception of the 24 Ω and 240 Ω ranges at 240 mV.
4-Wire Resistance
Accuracy ± (% of reading + Ω) [1]
Range [3]
Full scale
7-1/2 Digits
Resolution
Source
24 hours
90 Days
One Year
Current
23°C±1°C
23°C±5°C
23°C±5°C
0.008 + 1 mΩ
24 Ω[2]
24.000000 Ω
1 µΩ
10 mA
0.0038 + 0.7 mΩ
0.005 + 0.8 mΩ
240 Ω
240.00000 Ω
10 µΩ
1 mA
0.0037 + 3 mΩ
0.0046 + 4 mΩ
0.007 + 5 mΩ
2.4 kΩ
2.4000000 kΩ
100 µΩ
1 mA
0.0023 + 28 mΩ
0.004 + 32 mΩ
0.006 + 33 mΩ
24 kΩ
24.000000 kΩ
1 mΩ
100 µA
0.0025 + 300 mΩ
0.004 + 330 mΩ
0.006 + 350 mΩ
240 kΩ
240.00000 kΩ
10 mΩ
10 µA
0.0055 + 3.2 Ω
0.007 + 4 Ω
0.007 + 5 Ω
2.4 MΩ
2.4000000 MΩ
100 mΩ
1 µA
0.018 + 40 Ω
0.03 + 50 Ω
0.04 + 70 Ω
24 MΩ
24.0000 MΩ
100 Ω
100 nA
0.12 + 400 Ω
0.13 + 500 Ω
0.2 + 600 Ω
[1] With Aperture set to ≥0.5 Sec, and within one hour from Self Calibration (S-Cal).
[2] Ranges are only with the SMX2064, SM2064.
[3] Test voltages are 2.4 V max with the exception of the 24 Ω and 240 Ω ranges at 240 mV.
3-9
www.adlinktech.com
DMM
1
DAQ
2
PXI
Inductance
Diode Characterization
Accuracy ± (% of reading + henrys)
Available DC current
Uncertainty
100 nA, 1 µA, 10 µA, 100 µA and 0.02%
1 mA (SMX2064 and SM2064
10 mA & 12.5 mA)
4V
Typical Current
Value
Typical Voltage Value
Uncertainty
1%
0.02%
AC Functions
Full scale
7-1/2 Digits
Resolution
10 Hz - 20 Hz
20 Hz - 47 Hz
47 Hz - 10 kHz
10 kHz - 50 kHz
50 kHz-100 kHz
240 mV
240.00000 mV
100 nV
3.2 + 430 µV
0.4 + 200 µV
0.15 + 120 µV
0.27 + 230 µV
2.0 + 400 µV
2.4 V
2.4000000 V
1 µV
3.2 + 2.5 mV
0.4 + 1.7 mV
0.065 + 1.2 mV
0.35 + 1.5 mV
2.1 + 2 mV
24 V
24.000000 V
10 µV
3.3 + 20 mV
0.4 + 16 mV
0.073 + 13 mV
0.22 + 25 mV
1.5 + 40 mV
240 V
240.00000 V
100 µV
3.3 + 200 mV
0.4 + 150 mV
0.06 + 130 mV
0.30 + 200 mV
1.6 + 300 mV
330 V
330.00000 V
100 µV
3.3 + 200 mV
0.45 + 250 mV
0.09 + 230 mV
0.32 + 300 mV
1.6 + 400 mV
One Year Accuracy ± (% of reading + Amps), 23˚C±10˚C
AC Current (true RMS)
Full scale
6-1/2 Digits
Resolution
Max Burden 10 Hz - 20 Hz[1] 20 Hz - 47 Hz[1] 47 Hz - 1 kHz[1] 1 kHz -10 kHz[1]
2.4 mA
2.400000 mA
1 nA
25 mV
2.9 + 4 µA
1.0 + 4 µA
0.12 + 4 µA
24 mA
24.00000 mA
10 nA
250 mV
2.8 + 30 µA
1.0 + 30 µA
0.16 + 30 µA
0.4 + 40 µA
240 mA
240.0000 mA
100 nA
55 mV
2.8 + 400 µA
1.0 + 400 µA
0.2 + 220 µA
0.4 + 400 µA
2.4 A
2.400000 A
1 µA
520 mV
2.7 + 5 mA
0.9 + 6 mA
0.35 + 4 mA
0.5 + 5 mA
0.22 + 4 µA
[1] All AC Current ranges have typical measurement capability to 20 kHz.
Time Functions
Frequency and Period
ACV Mode
Input Impedance
Frequency Range
Period Range
Resolution
Uncertainty
1 MΩ with < 300 pF
2 Hz - 300 kHz
0.5 s - 3.33 µs
5 1/2 digits
± 0.002% of reading
Frequency Range
Period Range
Resolution
Uncertainty
2 Hz - 500 kHz
0.5 s - 2.0 µs
5 1/2 digits
± 0.01% of reading
Voltage range
24 mV - 250 V
ACI Mode
Input Signal
Input Impedance
range
10 Ω(3 mA & 30 mA)
0.33 mA - 2.5 A
Pulse Width
Frequency Range
Resolution
Width Range
Typical Uncertainty
2 Hz to 100 kHz
1 µs
2 µs to 1 s
0.01% of reading ±4 µs
Selected VAC
Threshold range
Threshold DAC
Highest allowed
Typical one year
Range
(DC level)
resolution
input Vp-p
setting uncertainty
240 mV
-1.0 V to +1.0 V
0.5 mV
1.900 V
0.2% + 4 mV
2.4 V
-10.0 V to +10.0 V
5.0 mV
19.00 V
0.2% + 40 mV
24 V
-100.0 V to +100.0 V
50 mV
190.0 V
0.2% + 0.4 V
240 V
-400 V to 400 V
500 mV
850.0 V
0.2% + 4 V
Threshold DAC
Totalizer
Maximum Count
Allowed rate
Condition
10^9
1 to 30,000 events per second
Uses Threshold DAC
Capacitance and Inductance Specifications (SMX2064 and SM2064 only)
Capacitance
Accuracy ±(% of reading + farads) [1]
Full scale
Reading
Resolution
One Year 23°C±5°C
Measurement
Time
Measurement
Rate (rps)
1,200 pF
1,199.9 pF
0.1 pF
1.5 ± 0.25 pF
52.3 ms
19.1
12 nF
11.999 nF
1 pF
1.2 ± 5 pF
118 ms
8.5
120 nF
119.99 nF
10 pF
1.0
127 ms
7.9
1.2 µF
1.1999 µF
100 pF
1.0
175 ms
5.7
12 µF
11.999 µF
1 nF
1.0
480 ms
2.1
119.99 µF
10 nF
1.0
50.3 ms
19.9
1.2 mF
1.1999 mF
100 nF
1.2
151.5 ms
6.6
12 mF
50.000 mF
1 µF
2
170 ms
5.9
[1] Within one hour of zero, using relative control. Accuracy is specified for values higher than 5% of the selected range with the exception of
the 1,200 pF range.
3.0% + 500 nH
240 µH
50 kHz
330.00 µH
10 nH
2.0% + 3 µH
2.4 mH
4 kHz
3.3000 mH
100 nH
1.5% + 25 µH
1.5% + 200 µH
23°C±5°C [1]
24 mH
1.5 kHz
33.000 mH
1 µH
240 mH
1 kHz
330.00 mH
10 µH
2.5 + 3 mH
2.4 H
100 Hz
3.3000 H
100 µH
3.0 + 35 mH
[1] Accuracy is specified for values greater than 5% of the selected range.
Other measurement functions of SMX2064 and SM2064 6-wire guarded resistance,
extended resistance, AC peak-to-peak voltage, AC crest factor, AC median value, leakage
current, RTD temperature, thermocouple temperature
Source Function (SMX2064 and SM2064 only)
■ DC Voltage Source
· Output range: -10.000 V to +10.000 V
· DAC resolution: 18 bits (closed loop), 12 bits (open loop)
■ AC Voltage Source
· Output range: 900 mV to 8 V
· DAC resolution: 12 bits
· Frequency range/resolution: 1 Hz to 200 kHz/2 mHz
■ DC Current Source
· Output range: 1.25 µA to 12.5 mA
5
6
Trigger Functions
■ External Hardware Trigger (at DIN-7 connector)
· Trigger input voltage level range: +3 V to +15 V
· Minimum trigger input current: 1 mA
■ Analog Threshold Trigger
· Trigger point: selectable positive or negative transition of set threshold
· Captures up to 120 post-trigger readings for apertures > 625 µs
· Captures up to 80 post-trigger readings for apertures > 625 µs
■ Delayed Hardware Trigger
· Up to 65 m Sec with 1 µs resolution
· Up to 1 s with 2 µs resolution
General Specifications
■ Overload Protection (voltage inputs): 330 VDC, 250 VAC
■ Isolation: 330 VDC, 250 VAC from earth ground
■ Maximum Input (Volt x Hertz):
· 8x106 volts x Hz normal mode input
· 1x106 volts x Hz common mode input
■ Calibration: Calibrations are performed by Signametrics in a computer at 23˚C
internal temperature rise. All calibration constants are stored in a text file.
■ Operating Temperature: -10˚C to 65˚C
■ Storage Temperature: -40˚C to 85˚C
■ Power requirements: +5 V, 300 mA maximum
■ Dimensions (not including connectors):
· SMX2064/2060: 160 mm x 100 mm
· SM2064/2060: 208 mm x 112 mm
■ Safety: Designed to IEC 1010-1, Installation Category II
Ordering Information
■ SMX2064
7-1/2 digits PXI Digital Multimeter with LCR Meter
7
8
9
10
11
12
■ SMX2060
7-1/2 digits PXI Digital Multimeter
■ SM2064
7-1/2 digits PCI Digital Multimeter with LCR Meter
■ SM2060
7-1/2 digits PCI Digital Multimeter
4
Accessories
120 µF
1 nH
cPCI & Industrial
Systems
Range
33.000 µH
Software &
Utilities
Active edge polarity
Positive or negative transition
75k Hz
Vision
Polarity
Positive or negative pulse widths
24 µH
3
Serial Comm
0.1 Ω (330 mA & 2.5 A)
One Year Accuracy
Distributed I/O
Input RMS
Resolution
Motion
Range
Full Scale 4 1/2 Digits
PAC
Range [3]
Test Frequency
GPIB & Bus
Expansion
One Year Accuracy ± (% of reading + Volts), 23˚C±5˚C
AC Voltage (true RMS)
Range
Modular
Instruments
Maximum Diode
Voltage Compliance
www.adlinktech.com 3-10