Internal Preamplifier Installation Note

Internal Preamplifier Installation Note
Keysight X-Series Signal Analyzers
Internal Preamplifier
Notice: This document contains references to Agilent.
Please note that Agilent’s Test and Measurement
business has become Keysight Technologies. For
more information, go to www.keysight.com.
Installation Note
Part Number: N9010-90056
Published in USA
January 2015
Notice.
The information contained in this document is subject to change without notice.
Keysight Technologies makes no warranty of any kind with regard to this material, including but
not limited to, the implied warranties of merchantability and fitness for a particular purpose.
Keysight Technologies shall not be liable for errors contained herein or for incidental or
consequential damages in connection with the furnishing, performance, or use of this material.
© Copyright 2012 - 2015 Keysight Technologies, Inc.
Internal Preamplifier Installation Kit
Internal Preamplifier Installation Kit
Products Affected:
CXA N9000A
EXA N9010A
MXA N9020A
PXA N9030A
UXA N9040B
Serial Numbers:
All
Options
CXA
P03 - Preamplifier, 3.0 GHz
P07 - Preamplifier, 7.5 GHz
P13 - Preamplifier, 13.6 GHz
P26 - Preamplifier, 26.5 GHz
EXA
P03 - Preamplifier, 3.6 GHz
P07 - Preamplifier, 7.0 GHz
P13 - Preamplifier, 13.6 GHz
P26 - Preamplifier, 26.5 GHz
P32 - Preamplifier, 32 GHz
P44 - Preamplifier, 44 GHz
MXA
P03 - Preamplifier, 3.6 GHz
P08 - Preamplifier, 8.4 GHz
P13 - Preamplifier, 13.6 GHz
P26 - Preamplifier, 26.5 GHz
PXA
P03 - Preamplifier, 3.6 GHz
P08 - Preamplifier, 8.4 GHz
P13 - Preamplifier, 13.6 GHz
P26 - Preamplifier, 26.5 GHz
P43 - Preamplifier, 43 GHz
P44 - Preamplifier, 44 GHz
P50 - Preamplifier, 50 GHz
UXA
P08 - Preamplifier, 8.4 GHz
P13 - Preamplifier, 13.6 GHz
P26 - Preamplifier, 26.5 GHz
To Be Performed By:
(X) Keysight Service Center
(X) Personnel Qualified by Keysight
(X) Customer
Estimated Installation Time:
Estimated Adjustment and Verification Time:
Installation Note N9010-90056
0.5 Hours
3.5 Hours (see Note 2)
3
Internal Preamplifier Installation Kit
Introduction
This kit contains all of the instructions required to install any of the internal preamplifier options into
an X-Series signal analyzer.
NOTE
1. At the time of manufacture the hardware related to this option was fully
adjusted and the option performance was verified to be within its warranted
specifications. Within one year of the initial calibration date of the instrument
this option is fully calibrated with no further adjustment or verification
testing.
To determine the initial calibration date, locate the original calibration
certificate that was shipped with the instrument at the time of purchase. The
Date of Calibration is printed on the original calibration certificate.
To ensure that this newly installed option has been installed properly, the
procedure that follows includes a functional check.
No factory test data is available for this option. If you require test data for this
option then a full calibration is required after installation of this option.
Arrangements regarding the level of calibration must be made with the
calibration provider.
2. If this instrument has been adjusted as part of a repair or calibration during
its first year, or if the instrument is more than one year old, additional
adjustments and performance verification tests are required to ensure that
this newly installed option is functioning properly. However, the completion
of these tests does not guarantee that the instrument meets all warranted
specifications.
If the option being installed is N9010A-P07, and the analyzer’s serial number
prefix is less than MY/SG/US5051, additional adjustments and performance
verification tests are required.
If the option being installed is N9010A-P13 or N9010A-P26, and the
analyzer’s serial prefix is less than MY/SG/US5322, additional adjustments
and performance tests are required.
If the option being installed is N9010A-P07, the analyzer’s instrument
software must be A.07.00 or later. Earlier software versions do not recognize
N9010A-P07 as a valid option.
If the option being installed is N9010A-P13 or N9010A-P26, the analyzer’s
instrument software must be A.12.09 or later. Earlier software versions do not
recognize N9010A-P13 or N9010A-P26 as valid options.
3.If the instrument is N9030A with Options 543, 544, or 550 and the preamp
option to be installed is P03, P08, P13, or P26, instrument software must be
A.10.52 or later. If the software is earlier, the preamp softkey will not appear.
4
Installation Note N9010-90056
Internal Preamplifier Installation Kit
4. The latest revision of the X-Series signal analyzer software may be
downloaded from:
http://www.keysight.com/find/Xseries_software.
5. This option is licensed for one instrument model/serial number
combination. The license key will only install on the designated instrument.
Contents
Quantity
Description
Part Number
1
Installation Note
This note
1
Option Upgrade Entitlement Certificate
5964-5178
Tools Required
— Keysight Calibration and Adjustment Software, N7814A
— (revision E.09.00 or later required for PXA with frequency range above 26.5 GHz)
— (revision E.10.00 or later required for EXA with frequency range above 26.5 GHz)
— (revision E.11.00 or later required for EXA being upgraded with N9010A-P13 or
N9010A-P26)
— Personal computer with internet access and USB port
— USB storage device with > 2 GB free memory
Installation Procedure over USB
1. Locate the Option Upgrade Entitlement Certificate (5964-5178) from the kit.
2. Redeem the Option Upgrade Entitlement Certificate by following the instructions on the
Certificate.
3. After redeeming your Option Upgrade Entitlement Certificate you will receive an email with an
attached License File.
4. Locate a USB storage device. Perform a virus scan on this device before use.
5. Save the License File to the root directory of the USB Storage Device.
6. Connect the USB Storage Device to the signal analyzer USB port. Windows will detect the new
hardware and may display the configuration menu shown in Figure 1. This menu may be
configured according to your preferences.
Installation Note N9010-90056
5
Internal Preamplifier Installation Kit
Figure 1
USB Storage Device Configuration Menu
7. The signal analyzer will automatically consume the License File. (This may take a few minutes)
When the License File is consumed the Keysight License Manager will display a “Successful
License Installation” message as shown in Figure 2.
Figure 2
Successful License Installation
Alternate Installation Procedure
The License File can be manually installed over USB or LAN by placing the license file in the
following folder on the signal analyzer
C:\Program Files\Agilent\licensing
Verify the Installation
1. Cycle the power on the signal analyzer.
2. Press System, Show, System to display a list of installed options.
3. Verify that the newly installed option appears on the list.
6
Installation Note N9010-90056
Internal Preamplifier Installation Kit
Functional Check:
Within one year of the initial calibration date of the instrument this option is fully calibrated with no
further adjustment or verification testing. To ensure that this newly installed option has been
enabled correctly the following functional check is recommended.
1. Press System, Show, Errors, Clear Error Queue. This will remove out-of-date errors.
2. Press System, Alignments, Align Now, All.
3. Watch for any errors during the instrument alignment.
4. Press System, Show, Errors to view the instrument error log.
5. Verify that there were no errors during the alignment.
Optional Functional Check
If Option P03 or P07 (N9000AK-P07) was installed:
Press Mode, Spectrum Analyzer.
Press Input/Output, RF Calibrator, 50 MHz
Press Frequency, enter 50 MHz to display the 50 MHz calibrator.
Press Span, 1 MHz.
Press Amplitude, More, Internal Preamp, select Low Band to turn preamp on.
The displayed signal amplitude should not change more than 0.2 dB.
If Option P07 (N9010AK-P07), P08, P13, P26, P32, P43, P44, or P50 was installed:
Press Mode, Spectrum Analyzer.
Press Input/Output, RF Calibrator, 4.8 GHz
Press Frequency, enter 4.8 GHz to display the 4.8 GHz calibrator.
Press Span, 1 MHz.
Press Amplitude, More, Internal Preamp, select Full Range to turn preamp on.
The displayed signal amplitude should not change more than 0.2 dB.
Installation Note N9010-90056
7
Internal Preamplifier Installation Kit
Utilities, Adjustments, and Performance Verification Tests
Calibration Software and specified test equipment is required to perform the adjustments, and can
be used to automate the performance verification testing.
Obtain Keysight X-Series Signal Analyzer Calibration Application SW, N7814A. Revision E.09.xx or
later required for PXA with frequency range above 26.5 GHz. Information on how to obtain this
software can be found at:
http://www.keysight.com/find/calibrationsoftware
Utilities Required
None
Adjustments Required
If the instrument is more than one year old, or if this instrument has been adjusted as part of
calibration during its first year, the following adjustments are the minimum set required to ensure
that this newly installed option is functioning properly.
NOTE
If the option being installed is N9010A-P07 and the analyzer’s serial number
prefix is less than MY/SG/US5051, the adjustments listed below for the EXA
must be performed.
If the option being installed is N9010A-P13 or N9010A-P26 and the analyzer’s
serial number prefix is less than MY/SG/US5322, the adjustments listed
below for the EXA must be performed.
CXA (N9000A) Ad justments
Freq Response Below 3.0 GHz, Preamp On
Freq Response Above 3.0 GHz, Preamp On (Option P07, P13, or P26)
EXA (N9010A) Ad justments
Freq Response Below 3.6 GHz, Preamp On
Freq Response Above 3.6 GHz, Preamp On (Option P07, P13, P26, P32, or P44)
MXA (N9020A) & PXA (N9030A) Ad justments
Freq Response Below 3.6 GHz, Preamp On
Freq Response Above 3.6 GHz, Preamp On (Option P08, P13, P26, P43, P44, or P50)
Effective DANL (Option NFE) Adjustment (N9030A only)
8
Installation Note N9010-90056
Internal Preamplifier Installation Kit
Performance Testing Required
If the instrument is more than one year old, or if the instrument has been adjusted as part of a
calibration during its first year, the following performance verification tests are the minimum set
required to ensure that this newly installed option is functioning properly. Performing only these
tests does not guarantee the instrument meets all specifications.
NOTE
If the option being installed is N9010A-P07, and the analyzer’s serial number
prefix is less than MY/SG/US5051, the performance verification tests listed
below for the EXA must be performed.
If the option being installed is N9010A-P13 or N9010A-P26 and the analyzer’s
serial number prefix is less than MY/SG/US5322, the performance tests listed
below for the EXA must be performed.
CXA (N9000A) Performance Verification Tests
Freq Response Below 3.0 GHz, Preamp On
Freq Response Above 3.0 GHz, Preamp On (Option P07, P13, or P26)
Displayed Average Noise Level
EXA (N9010A) Performance Verification Tests
Freq Response Below 3.6 GHz, Preamp On
Freq Response Above 3.6 GHz, Preamp On (Option P07, P13, P26, P32, or P44)
Displayed Average Noise Level
MXA (N9020A) & PXA (N9030A) Performance Verification Tests
Freq Response Below 3.6 GHz, Preamp On
Freq Response Above 3.6 GHz, Preamp On (Option P08, P13, P26, P43, P44, or P50)
Displayed Average Noise Level
Effective DANL (Option NFE) (N9030A only)
A full calibration is required to assure the instrument meets all specifications
The end user must ultimately determine whether they want a full calibration to be performed after
the installation of this upgrade or not. If a full calibration is required, arrangements regarding the
level of calibration must be made between the end user and the calibration provider.
For assistance, contact your nearest Keysight Technologies Sales and Service Office. To find your
local Keysight office access the following URL, or if in the United States, call the following
telephone number:
Installation Note N9010-90056
9
Internal Preamplifier Installation Kit
http://www.keysight.com/find/assist
1-800-829-4444 (8 am - 8 pm ET, Monday - Friday)
10
Installation Note N9010-90056