08-44.3 E Microcal 200_2000.cdr

08-44.3 E Microcal 200_2000.cdr
MicroCal
Accurate and Reliable
365 Days a Year
MULTIFUNCTION
CALIBRATORS
MicroCal 200/200+
MicroCal 2000+
Table IN - OUT Range
Sensor or
parameter
Tc J
Tc K
Tc T
Tc R
Tc S
Tc B
Tc C
Tc G
Tc D
Tc U
Tc L
Tc N
Tc E
Tc F
-210
-350
-270
454
-270
-454
-50
-60
-50
-60
50
140
0
32
0
32
0
32
-200
-330
-200
-330
-270
-450
-270
-454
0
32
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
1200°C
2200°F
1370°C
2500°F
400°C
760°F
1760°C
3200°F
1760°C
3200°F
1820°C
3310°F
2300°C
4172°F
2300°C
4172°F
2300°C
4172°F
400°C
760°F
760°C
1400°F
1300°C
2380°F
1000°C
1840°F
1400°C
2560°F
-190
-310
-160
-256
-130
-238
150
302
170
338
920
1688
0
32
190
374
0
32
-160
-256
-200
-328
0
32
-200
-328
0
32
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
1200°C
2192°F
1260°C
2300°F
400°C
752°F
1760°C
3200°F
1760°C
3200°F
1820°C
3308°F
2000°C
3632°F
2300°C
4172°F
2130°C
3866°F
400°C
752°F
760°C
1400°F
1300°C
2372°F
1000°C
1832°F
1400°C
2552°F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
0.01°C
0.1 °F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
MicroCal 200
Accuracy
(% of rdg)
±(0.02% + 0.1 °C)
±(0.02% + 0.18°F)
±(0.02% + 0.1 °C)
±(0.02% + 0.18°F)
±(0.02% + 0.1 °C)
±(0.02% + 0.18°F)
±(0.02% + 0.2 °C)
±(0.02% + 0.36°F)
±(0.02% + 0.2 °C)
±(0.02% + 0.36°F)
±(0.02% + 0.3 °C)
±(0.02% + 0.54°F)
±(0.02% + 0.2 °C)
±(0.02% + 0.36°F)
±(0.02% + 0.3 °C)
±(0.02% + 0.54°F)
±(0.02% + 0.3 °C)
±(0.02% + 0.54°F)
±(0.02% + 0.1 °C)
±(0.02% + 0.18°F)
±(0.02% + 0.1 °C)
±(0.02% + 0.18°F)
±(0.02% + 0.1 °C)
±(0.02% + 0.18°F)
±(0.02% + 0.1 °C)
±(0.02% + 0.18°F)
±(0.02% + 0.1 °C)
±(0.02% + 0.18°F)
-200
-330
-200
-330
-200
-330
-200
-330
-200
-330
-200
-330
...
...
...
...
...
...
...
...
...
...
...
...
850°C
1570°F
650°C
1210°F
600°C
1120°F
850°C
1570°F
850°C
1570°F
850°C
1570°F
-200
-328
-200
-328
-200
-328
-200
-328
-200
-328
-200
-328
...
...
...
...
...
...
...
...
...
...
...
...
850°C
1562°F
650°C
1210°F
600°C
1112°F
850°C
1562°F
530°C
986°F
850°C
1562°F
0.01°C
0.1 °F
0.01°C
0.1 °F
0.01°C
0.1 °F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
0.01°C
0.1 °F
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
+ 0.05°C)
+ 0.09°F)
+ 0.05°C)
+ 0.09°F)
+ 0.05°C)
+ 0.09°F)
+ 0.15°C)
+ 0.27°F)
+ 0.1 °C)
+ 0.18°F)
+ 0.1 °C)
+ 0.18°F)
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
+
+
+
+
+
+
+
+
+
+
+
+
0.05 °C)
0.09 °F)
0.05 °C)
0.09 °F)
0.05 °C)
0.09 °F)
0.15 °C)
0.27 °F)
0.1 °C)
0.18 °F)
0.1 °C)
0.18 °F)
0.01°C
0.1 °F
0.01°C
0.1 °F
0.01°C
0.1 °F
0.01°C
0.1 °F
0.01°C
0.1 °F
0.01°C
0.1 °F
±(0.005%
±(0.005%
±(0.005%
±(0.005%
±(0.005%
±(0.005%
±(0.005%
±(0.005%
±(0.005%
±(0.005%
±(0.005%
±(0.005%
+
+
+
+
+
+
+
+
+
+
+
+
0.05°C)
0.09°F)
0.05°C)
0.09°F)
0.05°C)
0.09°F)
0.15°C)
0.27°F)
0.1 °C)
0.18°F)
0.1 °C)
0.18°F)
-70
-100
-180
-300
-60
-80
0
32
...
...
...
...
...
...
...
...
150°C
310°F
150°C
310°F
180°C
360°F
150°C
310°F
-70
-94
-180
-292
-60
-76
0
32
...
...
...
...
...
...
...
...
150°C
302°F
150°C
302°F
180°C
356°F
150°C
302°F
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
+ 0.4 °C)
+ 0.72°F)
+ 0.05°C)
+ 0.09°F)
+ 0.05°C)
+ 0.09°F)
+ 0.05°C)
+ 0.09°F)
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
+
+
+
+
+
+
+
+
0.4 °C)
0.72 °F)
0.05 °C)
0.09 °F)
0.05 °C)
0.09 °F)
0.05 °C)
0.09 °F)
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
±(0.01%
±(0.01%
±(0.01%
±(0.01%
±(0.01%
±(0.01%
±(0.01%
±(0.01%
+
+
+
+
+
+
+
+
0.4 °C)
0.72°F)
0.05°C)
0.09°F)
0.05°C)
0.09°F)
0.05°C)
0.09°F)
-20
-0.2
-2
-5
0
0
0
0
0
1
1
1
0
0
0
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
Total range
Accuracy Range
Resolution
MicroCal 200+
Accuracy
(% of rdg)
± (0.01% + 0.1 °C)
± (0.01% + 0.18 °F)
± (0.01% + 0.1 °C)
± (0.01% + 0.18 °F)
± (0.01% + 0.1 °C)
± (0.01% + 0.18 °F)
± (0.01% + 0.2 °C)
± (0.01% + 0.36 °F)
± (0.01% + 0.2 °C)
± (0.01% + 0.36 °F)
± (0.01% + 0.3 °C)
± (0.01% + 0.54 °F)
± (0.01% + 0.2 °C)
± (0.01% + 0.36 °F)
± (0.01% + 0.3 °C)
± (0.01% + 0.54 °F)
± (0.01% + 0.3 °C)
± (0.01% + 0.54 °F)
± (0.01% + 0.1 °C)
± (0.01% + 0.18 °F)
± (0.01% + 0.1 °C)
± (0.01% + 0.18 °F)
± (0.01% + 0.1 °C)
± (0.01% + 0.18 °F)
± (0.01% + 0.1 °C)
± (0.01% + 0.18 °F)
± (0.01% + 0.1 °C)
± (0.01% + 0.18 °F)
MicroCal 2000+
Resolution
Accuracy
(% of rdg)
0.01°C
±(0.005% + 0.1 °C)
0.1 °F
±(0.005% + 0.18°F)
0.01°C
±(0.005% + 0.1 °C)
0.1 °F
±(0.005% + 0.18°F)
0.01°C
±(0.005% + 0.1 °C)
0.1 °F
±(0.005% + 0.18°F)
0.01°C
±(0.005% + 0.2 °C)
0.1 °F
±(0.005% + 0.36°F)
0.01°C
±(0.005% + 0.2 °C)
0.1 °F
±(0.005% + 0.36°F)
0.01°C
±(0.01% + 0.3 °C)
0.1 °F
±(0.01% + 0.54°F)
0.1 °C
±(0.01% + 0.2 °C)
0.1 °F
±(0.01% + 0.36 °F)
0.1 °C
±(0.01% + 0.3 °C)
0.1 °F
±(0.01% + 0.54 °F)
0.1 °C
±(0.01% + 0.3 °C)
0.1 °F
±(0.01% + 0.54 °F)
0.01°C
±(0.005% + 0.1 °C)
0.1 °F
±(0.005% + 0.18°F)
0.01°C
±(0.005% + 0.1 °C)
0.1 °F
±(0.005% + 0.18°F)
0.01°C
±(0.005% + 0.1 °C)
0.1 °F
±(0.005% + 0.18°F)
0.01°C
±(0.005% + 0.1 °C)
0.1 °F
±(0.005% + 0.18°F)
0.01°C
±(0.005% + 0.1 °C)
0.1 °F
±(0.005% + 0.18°F)
High Accuracy
Multifunctional Process Calibrators
MULTIFUNCTION
CALIBRATORS
Portable, Rugged and
Easy to Use
Measures and Sources 14 Type
of TCs and 10 RTDs
24 Vdc Loop Power Supply
In-Line Digit Setting
Pannel Mounting Kit
CU100
Ni100
Ni120
mV (L)
mV (H)
V
mA (In)
mA (Out)
Ω
IN
Ω
OUT
Frequency
Pulse counter
Pulse (Out)
±(0.02% + 2 µ
V)
±(0.02% + 10 µ
V)
±(0.02% + 0.08mV)
±(0.02% + 0.4 µ
A)
±(0.02% + 0.4 µ
A)
±(0.02% + 12 mΩ
)
±(0.02% +120 mΩ
)
±(0.02% + 20 mΩ
)
±(0.02% +200 mΩ
)
±(0.005% + 0.001 Hz)
±(0.005% + 0.001 Hz)
±(0.005% + 0.001 Hz)
infinite
1 pulse / min
1 pulse / min
± (0.01% + 2 µ
V)
± (0.01% +10 µ
V)
± (0.01% + 0.08mV)
± (0.01% + 0.4 µ
A)
± (0.01% + 0.4 µ
A)
± (0.01% +12 mΩ
)
± (0.01% +120 mΩ
)
± (0.01% + 20 mΩ
)
± (0.01% +200 mΩ
)
±(0.005% + 0.001 Hz)
±(0.005% + 0.001 Hz)
±(0.005% + 0.001 Hz)
infinite
1 pulse / min
1 pulse / min
• The MicroCal 2000+ relative accuracies shown above are stated for 360
days and the operative conditions are from +70°F to +77°F
• Typical 90 days relative accuracy can be estimated by dividing the "% of
reading" specifications by 1.6.
12
0.1 µ
V
1 µ
V
10 µ
V
0.1 µ
A
0.1 µ
A
1 mΩ
0.01 Ω
1 mΩ
0.01 Ω
0.001 Hz
0.01 Hz
0.1 Hz
1 count
1 pulse / min
1 pulse / h
±(0.0035% + 1 µ
V)
±(0.005% + 10 µ
V)
±(0.005% + 0.08mV)
±(0.005% + 0.4 µ
A)
±(0.005% + 0.4 µ
A)
±(0.005% + 12 mΩ
)
±(0.005% +120 mΩ
)
±(0.005% + 12 mΩ
)
±(0.005% +120 mΩ
)
±(0.005% + 0.001 Hz)
±(0.005% + 0.001 Hz)
±(0.005% + 0.001 Hz)
infinite
1 pulse / min
1 pulse / min
• Typical 2 years relative accuracy can be estimated by multiplying the “% of
reading" specifications by 1.4.
• All input ranges: additional error ±1 digit.
• Eurotron traceability chart and uncertainty can be supplied on request.
bt !
No dou Accurate
st
the Mo rdg)
f
o
e
n
O
5%
or
(0.003
alibrat
C
le
b
a
y!
Por t
le Toda
Availab
ron
ot
Instrum
e
Q
rti
Ce
Note:
• The MicroCal 200/200+ relative accuracies shown above are stated for 360
days and the operative conditions are from +64°F to +82°F
1 µ
V
10 µ
V
0.1 mV
0.1 µ
A
0.1 µ
A
1 mΩ
0.01 Ω
1 mΩ
0.01 Ω
0.001 Hz
0.01
0.1 Hz
1 count
1 pulse/min
1 pulse/h
°C
°F
°C
°F
°C
°F
°C
°F
fie
d
s
nt
+200mV
+2
V
+20 V
+50 mA
+50 mA
500 Ω
5000 Ω
500 Ω
5000 Ω
200 Hz
2000 Hz
20000 Hz
6
10 counts
6000 pulse/min
36000 pulse/h
°C
°F
°C
°F
°C
°F
°C
°F
ste
m
CU10
Eu
r
Pt1000 IEC
OIML
CalpMan 2007
Software for Automatic
Calibration and Documenting in
Compliance with ISO9001
Bulletin 08-44.3 E MicroCal 200_2000 - ed. 04-2007
Pt 500
MicroCal PM200
Optional Module
to Generate/Measure Pressure
Specifications may change without notice
Pt100 IEC
OIML, a 3926
Pt100
a 3902
Pt100 JIS
SAMA
Pt 200
Quality S
y
www.eurotron.com
All descriptions are related to a fully optioned instrument. See last page for the different configurations
MicroCal
To Keep your Test and Process Equipment
Accurately Calibrated for ISO 9000 Compliance
MULTIFUNCTION
CALIBRATORS
MicroCal
High Technology for Daily Use
in Field and Laboratory
General
SPECIFICATIONS
The MicroCal 200/200+ and MicroCal 2000+ are multifunctional calibrators
with insulated channels. The instruments are portable and developed to meet
the needs of instrumentation engineers and Quality managers, both in
laboratory and in field work.
This units are accurate, rugged, compact and easy to use. They are the ideal
solution to simulate and measure simultaneously: Voltage, current, resistance,
thermocouple, resistance thermometers, frequency and pulse.
Advanced flexibility and high performance has been achieved using a 32 bit
microprocessor and a fast A/D conversion technology. The calibrator’s internal
memory stores all data for normalized IEC, DIN and JIS thermoelectric sensors
for both IPTS68 and ITS90 International Temperature Scale
IN/OUT parameters:
Reference junction compensation:
Rj compensation drift:
Rj compensation error:
Common mode rejection:
Normal mode rejection:
Temperature stability:
=
Table top
The microprocessor performs automatic
linearization and cold junction
compensation to assure high accuracy. It is
possible to set the calibrator to execute
menu-driven calibration procedures for
your instruments in field work.
The case is equipped with 2 pivot feet to
change the vertical viewing angle when
using the instrument on the top of a table.
Quality System
Research, development, production,
inspection and certification activities are
defined by methods and procedures of the
Eurotron Quality System inspected for
compliance and certified ISO9001 by
GASTEC, a Dutch notified body.
Both Channel 1 (Out) and Channel 2 (In)
have the following operative mode
capabilities:
=
millivolts
=
volts
=
milliamperes (active and passive loop)
=
Ohms
=
temperature with thermocouples
=
temperature with resistance
thermometers
=
frequency
=
pulse
EMC Conformity
The instrument case, made in shockresistant injection molded ABS +
polycarbonate has an internal metal
coating to fulfill the prevision of the directive
89/336/CEE Electromagnetic
Compatibility.
Remote auxiliary inputs are available for:
and temperature
module
=
Two internal sensors and built-in hand
pump pressure module
=
Relative humidity
The ergonomic case design allows the user
the calibrator in three different ways:
=
Portable
Two different leather cases, with a cover
and shoulder strap, are available upon
request for instrument alone or
instrument, printer and accessories.
These are extremely useful for practical
use since they leave one hand free for
instrument tuning.
=
Panel mounting
It requires a panel cutout of 9.7” x 3.5”.
The instrument bezel flange butts against
the front of the mounting plate; two
lateral mounting brackets fit over the
instrument rear panel
Output impedance (emf output):
Input impedance (mV, V and Tc ranges):
Input impedance (mA ranges):
Source resistance effects:
Rtd and Ω
simulation excitation current:
Rtd and Ω
measurement excitation current:
Rtd connection:
Rtd cable compensation:
Rtd cable compensation error (Pt100):
Maximum load resistance:
Display:
Measurement sampling time:
Output noise (at 300 Hz):
Digital interface:
Channel 1_Channel 2 insulation:
Calculation functions:
Selection °C/°F/K:
In/Out data memory:
Logging mode:
Convert function:
Scale factor:
Square root:
Calibration:
Power supply:
Self contained operation:
Recharging time:
Line operation:
Line transformer insulation:
Firmware release identification:
Operating environment temperature range:
Storage temperature range:
Case:
Dimensions:
Weights:
6
MULTIFUNCTION
CALIBRATORS
mV, V, mA, Ω
, KΩ
, frequency, pulses
J, K, T, R, S, B, N, C, E, U, L, F, G, D
Pt100 IEC, OIML, USLAB, US, SAMA, JIS, Pt200, 500,
1000, 1000 OIML, Ni100, Ni120, Cu10, Cu100
internal automatic:
from -10 °C to +55 °C (14°F to 130°F)
external adjustable:
from -50 °C to +100 °C (-58°F to 210°F)
remote with external Pt100: from -10°C to +100 °C (14°F to 210°F)
± 0.015°C/°C (from -10 °C to +55 °C)
internal :
±0.15°C (±0.3°F)
remote :
±0.3°C (±0.6°F)
>140 dB at ac operation
>70 dB at 50 or 60 Hz
MicroCal 200/200+: for temperature exceeding the band +18°C to +28°C (from +64 °C to +84 °F)
full scale:
± 8 ppm/°C
zero:
± 0.2 mV /°C
MicroCal 2000+: for temperature exceeding the band +21°C to +25°C (from +70 °C to +77 °F)
full scale:
± 3 ppm/°C
zero:
± 0.2 mV /°C
< 0.5 Ω
with 0.5 mA maximum current
>10 MΩ
<140 Ω
@ 1 mA
±1 mV error for 1000 Ω
source resistance
MicroCal 200/200+:
from 0.01 to 5 mA
MicroCal 2000+:
from 0.01 to 2 mA
~ 0.4 mA @ 400Ω
~ 0.04 mA @ 4000Ω
2, 3, and 4 wires
up to 100 Ω
(each wire)
±0.005°C/ Ω
of total wire
1000 Ω
@ 20 mA
graphic LCD 240 x 64 dots display with LED backlight device
250 ms
<2 mVpp for ranges up to 200 mV f.s.,
<10 mVpp for ranges up to 2 V f.s.
<80 mVpp for ranges up to 20 V f.s.
full bidirectional TTL (a RS232 adapter normal or insulated, is available as an option)
50 Vdc
hold, max, min, offset, average
through the configuration procedure
20 data with manual or automatic recall
1500 input data items (optional memory card for memory extension)
displays the electrical equivalent of the engineering unit
5 different settings with zero and span programmable within -399999 and +999999
in combination with scale factor
self learning technique with automatic procedure
external charger and rechargeable Ni-Cd battery
6h on Tc and mV input/output (backlight Off)
3.5h with 20 mA simulation (backlight Off)
5h at 90% and 6h at 99% with instrument switched off.
The battery charging is active only with the instrument switched off.
100 - 120 - 230 Vac through the external battery charger
2500 Vac
release code on the display
from -10 °C to +55 °C (from 14 °C to 130 °F)
from -30 °C to +60 °C (from -22 °C to 140 °F)
Injection moulded ABS with internal metal coating
264 x 96 x 172 mm (10.4”x3.8”x6.8”) DIN size
nett: 8.8 lb. gross: 12.1 lb.
Signal type:
thermocouples type:
resistance thermometers:
11
MULTIFUNCTION
CALIBRATORS
MicroCal 200/200+
MicroCal 2000+
"Excellence in measurements"
Instrument Highlight
Comparative Table
Accuracy
MicroCal
200
MicroCal
200+
±0.02% rdg
± 0.01%rdg
MicroCal
2000+
± 0.0035% rdg
32-bit microprocessor and flash memory for
firmware upgrading through serial interface
All normalized IEC, DIN, JIS thermocouples
Pt, Ni, Cu resistance thermometers
mA, mV, V, W, frequency, pulse, counter
IPTS 68 and ITS 90 linearization
Current output mode directly
on active or passive loops
Bidirectional serial interface
Option
Communication bus for extension
to pressure and other optional modules
Optional dedicated external impact
type printerOptionOptionOption
Pannel mounting brackets
Option
Option
Traceable Report of Calibration
Option
Display & Keyboard
The large graphic display with back light
allows a simultaneous indication of the
measured and the simulated value (large
digit), together with all data related to the
signal as: engineering unit, type of sensor
or signal, temperature scale, cold
junction selection, battery level, etc.
A swap key allow to change the position
of the IN and OUT parameters on the
display.
The operative set-up mode is simplified
by a sequence of menu pages that only
require <Select> and <Enter>
instructions.
A full set of instruction pages are memory
stored for direct operator assistance.
The relevant instruction may be recalled
through the <Help> key.
RAM extension with PCMCIA memory card
Non-volatile memory with real-time clock
Option
Logging and direct real-time graph
with movable cursor to read the required
actual value
Option
Option
mV and V I/O display
6 digits
6 digits
Real Time Graphic Logging
MicroCal 200/200+
6 digit display
MicroCal 2000+
6 ½ digit display
7
6 ½ digits (0.1 µ
V)
MicroCal 200/200+
MicroCal 2000+
MULTIFUNCTION
CALIBRATORS
Firmware
The firmware is stored inside a flash
memory and allows a fast and easy
upgrade of the instrument using a
standard PC. Firmware includes the
following capabilities:
• multiple measurement and
generation mode
• signal processing: filter, average,
min, max
• PC downloadable test procedure
• DataLogger and graphic recorder
• switch test routine
• ramping and stepping for dynamic
testing
• user definable linearization
• user entry of probe specific
calibration coefficients.
Scale Factor - Square Root
All non-temperature ranges are fully
programmable to read both input and
output values in terms of engineering
units. Four programmable
alphanumeric characters are
available on the display to show the
symbol of the parameter being
measured (i.e. mbar, % RH, % CO, etc.)
The unit allows the mA reading and
output to be related to flow when using
a DP transmitter across a calibrated
flange.
compensation temperature (from -58
to +210°F) by keyboard.
or to calibrate totalizers and counters.
The instrument can be configured to
measure frequency and count pulse
Transmitter Simulation
(totalizer mode). Technical units in Hz,
The calibrator can be connected to pulse/h and pulse/m. The threshold is
your system inputs to simulate 4-20 mA adjustable from 0 to 20 V with 0.01 V
transmitters. It has an adequate power resolution.
to drive 20 mA into a load of 1000 Ω
in
source mode. For fast loop checks, Calculated Readings
preprogrammed steps can be The units allow measurements of
recalled to output 4, 8, 12, 16 and 20 unstable input signals through a
mA or % equivalent. Since the programmable averaging or min, max
instrument has two separate channels. identification. A hold feature is also
It is possible to read the error directly in included from a keyboard or external
% (In/Out). The operator can set and contact instructions.
change temperature values while
obtaining the equivalent output mA. Programmable Signal Converter
The mA mode may be connected The instrument can be used as a
directly either on passive or on active t e m p o r a r y s i g n a l c o n v e r t e r
replacement. Any input signal
current loops.
(including remote auxiliary inputs) can
be converted into any of the available
Frequency - Counts
The "Out" mode is designed to output signals. The galvanic insulation
generate zero based square pulse, between the input and output
with an adjustable amplitude, at a channels allow to use this feature in
frequency up to 20 KHz. A preset field work applications as well.
number of pulses may be
programmed and transmitted to test
Rj Compensation
Accurate and fast response
compensation, through a special low
thermal capacity design of binding
posts, incorporates a thin film, high
accuracy Pt100 as cold junction
reference. The internal reference
allows the maximum accuracy for the
14 to 130°F
temperature
range. An external
Pt100 sensor can
be connected for
s p e c i a l
application (from
14 to +210°C). It is
possible to
manually set the
8
MicroCal
To Document the Inspection
and Calibration Activities
"Excellence in measurements"
MULTIFUNCTION
CALIBRATORS
ORDERING CODE
=
direct numeric entry;
MicroCal 200
=
direct keyboard access to n. 20
3916 - A - B - C - D
=
programmable
memory stored
values.
MicroCal 200+
3918 - A - B - C - D
MicroCal 2000+
End
step
3928 - A - B - C - D
Each instrument is supplied with USB cable, PC
Software Package (Linman, Logman), leather
carrying case with shoulder strap, Report of Calibration
and instruction manual.
Start
Rise
Soack
Fall
Time
Report of Calibration
Each MicroCal unit is factor y
calibrated and certified against the
highest Eurotron Standards, which are
periodically certified by an
Internationally recognized laboratory
to ensure performance. The units are
also shipped with a Report of
Calibration stating the nominal and
actual values and the deviation errors.
4 Wire Resistance Thermometer
Although resistance and temperature
with resistance thermometer may be
measured with 2 or 3 wire standard
connections. The instrument is also
designed for 4-wire measurements for
higher accuracy and the ability to use
as a reference for your calibrations.
MicroCal PM200 Pressure Module
An optional pressure module is
available with one or two internal
sensors. The MicroCal PM200 is an
intelligent module and it is used to
extend the MicroCal capability to
pressure in/out. MicroCal PM200 can
incorporate pressure/vacuum pump,
volume adjuster and ventilation valve.
External sensors can be used to
measure pressure up to 700bar. See
MicroCal PM200 bulletin for more
specifications.
Remote Temperature Probe
A high accuracy probe is available on
request for general purpose
temperature measurement and/or
remote cold junction compensation.
Simulation Capability
Menu-driven set up to generate:
=
a continuous or step ramp output
where the total time, the starting
point, the final point and the size of
the steps are requested by the set-up
procedure to run the program;
=
a continuous programmable cycle
rises, soaks, falls; (see figure)
=
the signal value setting uses unique
in-line, single-digit setting mode or a
9
Table A
0
3
Options:
None
TTL/RS232 insulated converter
Table B
1
2
3
4
5
9
Line setting - Line cord plug
120 V 50/60Hz - USA plug
230 V 50/60Hz - Schuko plug
230 V 50/60Hz - UK plug
230 V 50/60Hz - European plug
100 V 50/60Hz - USA/Japan plug
special
Table C
0
6
Accessories
None
Brackets for panel mounting
Table D
1
Report of Calibration
Eurotron certificate
MicroCal PM200
Multifunction electrical signals
+ pressure calibrator
MicroCal
Easy to use
Easy to Order
MULTIFUNCTION
CALIBRATORS
CalPMan 2007 Software
OFF-LINE Procedure
¡
Automatic procedure run
¡
Built-in communication with
For over 20 years, Eurotron Instruments has
combined its own ISO 9001, electric,
thermoelectric and pressure
measurements, and development of
software application experiences.
CalpMan 2007 has been designed to be
used industries, where there are both
laboratory and maintenance on field
needs. The CalPMan 2007 is a WindowsTM
2000/XP software designed to plan,
manage and document all the
calibrations and the certifications of the
process instrumentation. The software can
manage the automatic procedures of
the Eurotron calibrators.
The CalPMan 2007 can also manage the
off-line procedures with all the Eurotron
documenting calibrators. The test data
can be transferred directly into the
instrument and the test can run by the
calibrator itself without any connection to
a computer. The final results will be stored
into the instrument, and they can be
transferred from it directly into the
CalPMan 2007 database. As transferred,
is it possible to print out the certificates
and manage the instrument database.
documenting calibrators
¡
Traceable and editable calibration
documents
¡
Database backup/Restore ability
The CalpMan 2007 Advanced has the
following advanced features:
¾
Up to 3 instruments of simultaneous
management;
¾
Advanced control of the test point;
¾
Multi-parameter procedures run;
¾
Simultaneous certification of more than
one temperature sensors;
¾
Password manager and advanced
data security ability.
Two software versions are available: the
CalpMan 2007 and the CalpMan 2007
Advanced.
The main features are:
¡
Easy to use;
¡
Compliant with ISO9001 requirements;
¡
Standards and instruments expired date
management;
FIVE STEPS FOR IN-FIELD DOCUMENTING CALIBRATION
the procedure
2 Transfer
into the calibrator
1
Prepare the procedure
using the CalPMan
Software
Print
the report
5
3
Select the instrument
(TAG) and run the procedure
4
Transfer the results
on a PC
The MicroCal 200/200+ and 2000+ Series are
documenting calibrators able to load test
strategies using the CalPMan 2007 software. It
can run them and store the results
automatically, calculate the errors, highlight the
out of range points, and download data on a
PC to store them and to generate different
reports. The calibrator internal memory can store
a whole week procedures.
Calibration
Procedure
IN
5.01
bar
InP
ç
ç
5.00
Run
Dis
Esc
bar
Tag Selection
Tag: TRX0GH-00
Pressure transmitter for PLC
PFK3JM-04 located on sector Q28
inside panel PNL8WC-23
Before: 25/07/97 17:07:00 (FAIL)
After : 25/07/97 17:27:00 (OK )
10
MicroCal
To Keep your Test and Process Equipment
Accurately Calibrated for ISO 9000 Compliance
MULTIFUNCTION
CALIBRATORS
MicroCal
High Technology for Daily Use
in Field and Laboratory
General
SPECIFICATIONS
The MicroCal 200/200+ and MicroCal 2000+ are multifunctional calibrators
with insulated channels. The instruments are portable and developed to meet
the needs of instrumentation engineers and Quality managers, both in
laboratory and in field work.
This units are accurate, rugged, compact and easy to use. They are the ideal
solution to simulate and measure simultaneously: Voltage, current, resistance,
thermocouple, resistance thermometers, frequency and pulse.
Advanced flexibility and high performance has been achieved using a 32 bit
microprocessor and a fast A/D conversion technology. The calibrator’s internal
memory stores all data for normalized IEC, DIN and JIS thermoelectric sensors
for both IPTS68 and ITS90 International Temperature Scale
IN/OUT parameters:
Reference junction compensation:
Rj compensation drift:
Rj compensation error:
Common mode rejection:
Normal mode rejection:
Temperature stability:
=
Table top
The microprocessor performs automatic
linearization and cold junction
compensation to assure high accuracy. It is
possible to set the calibrator to execute
menu-driven calibration procedures for
your instruments in field work.
The case is equipped with 2 pivot feet to
change the vertical viewing angle when
using the instrument on the top of a table.
Quality System
Research, development, production,
inspection and certification activities are
defined by methods and procedures of the
Eurotron Quality System inspected for
compliance and certified ISO9001 by
GASTEC, a Dutch notified body.
Both Channel 1 (Out) and Channel 2 (In)
have the following operative mode
capabilities:
=
millivolts
=
volts
=
milliamperes (active and passive loop)
=
Ohms
=
temperature with thermocouples
=
temperature with resistance
thermometers
=
frequency
=
pulse
EMC Conformity
The instrument case, made in shockresistant injection molded ABS +
polycarbonate has an internal metal
coating to fulfill the prevision of the directive
89/336/CEE Electromagnetic
Compatibility.
Remote auxiliary inputs are available for:
and temperature
module
=
Two internal sensors and built-in hand
pump pressure module
=
Relative humidity
The ergonomic case design allows the user
the calibrator in three different ways:
=
Portable
Two different leather cases, with a cover
and shoulder strap, are available upon
request for instrument alone or
instrument, printer and accessories.
These are extremely useful for practical
use since they leave one hand free for
instrument tuning.
=
Panel mounting
It requires a panel cutout of 9.7” x 3.5”.
The instrument bezel flange butts against
the front of the mounting plate; two
lateral mounting brackets fit over the
instrument rear panel
Output impedance (emf output):
Input impedance (mV, V and Tc ranges):
Input impedance (mA ranges):
Source resistance effects:
Rtd and Ω
simulation excitation current:
Rtd and Ω
measurement excitation current:
Rtd connection:
Rtd cable compensation:
Rtd cable compensation error (Pt100):
Maximum load resistance:
Display:
Measurement sampling time:
Output noise (at 300 Hz):
Digital interface:
Channel 1_Channel 2 insulation:
Calculation functions:
Selection °C/°F/K:
In/Out data memory:
Logging mode:
Convert function:
Scale factor:
Square root:
Calibration:
Power supply:
Self contained operation:
Recharging time:
Line operation:
Line transformer insulation:
Firmware release identification:
Operating environment temperature range:
Storage temperature range:
Case:
Dimensions:
Weights:
6
MULTIFUNCTION
CALIBRATORS
mV, V, mA, Ω
, KΩ
, frequency, pulses
J, K, T, R, S, B, N, C, E, U, L, F, G, D
Pt100 IEC, OIML, USLAB, US, SAMA, JIS, Pt200, 500,
1000, 1000 OIML, Ni100, Ni120, Cu10, Cu100
internal automatic:
from -10 °C to +55 °C (14°F to 130°F)
external adjustable:
from -50 °C to +100 °C (-58°F to 210°F)
remote with external Pt100: from -10°C to +100 °C (14°F to 210°F)
± 0.015°C/°C (from -10 °C to +55 °C)
internal :
±0.15°C (±0.3°F)
remote :
±0.3°C (±0.6°F)
>140 dB at ac operation
>70 dB at 50 or 60 Hz
MicroCal 200/200+: for temperature exceeding the band +18°C to +28°C (from +64 °C to +84 °F)
full scale:
± 8 ppm/°C
zero:
± 0.2 mV /°C
MicroCal 2000+: for temperature exceeding the band +21°C to +25°C (from +70 °C to +77 °F)
full scale:
± 3 ppm/°C
zero:
± 0.2 mV /°C
< 0.5 Ω
with 0.5 mA maximum current
>10 MΩ
<140 Ω
@ 1 mA
±1 mV error for 1000 Ω
source resistance
MicroCal 200/200+:
from 0.01 to 5 mA
MicroCal 2000+:
from 0.01 to 2 mA
~ 0.4 mA @ 400Ω
~ 0.04 mA @ 4000Ω
2, 3, and 4 wires
up to 100 Ω
(each wire)
±0.005°C/ Ω
of total wire
1000 Ω
@ 20 mA
graphic LCD 240 x 64 dots display with LED backlight device
250 ms
<2 mVpp for ranges up to 200 mV f.s.,
<10 mVpp for ranges up to 2 V f.s.
<80 mVpp for ranges up to 20 V f.s.
full bidirectional TTL (a RS232 adapter normal or insulated, is available as an option)
50 Vdc
hold, max, min, offset, average
through the configuration procedure
20 data with manual or automatic recall
1500 input data items (optional memory card for memory extension)
displays the electrical equivalent of the engineering unit
5 different settings with zero and span programmable within -399999 and +999999
in combination with scale factor
self learning technique with automatic procedure
external charger and rechargeable Ni-Cd battery
6h on Tc and mV input/output (backlight Off)
3.5h with 20 mA simulation (backlight Off)
5h at 90% and 6h at 99% with instrument switched off.
The battery charging is active only with the instrument switched off.
100 - 120 - 230 Vac through the external battery charger
2500 Vac
release code on the display
from -10 °C to +55 °C (from 14 °C to 130 °F)
from -30 °C to +60 °C (from -22 °C to 140 °F)
Injection moulded ABS with internal metal coating
264 x 96 x 172 mm (10.4”x3.8”x6.8”) DIN size
nett: 8.8 lb. gross: 12.1 lb.
Signal type:
thermocouples type:
resistance thermometers:
11
MicroCal
Accurate and Reliable
365 Days a Year
MULTIFUNCTION
CALIBRATORS
MicroCal 200/200+
MicroCal 2000+
Table IN - OUT Range
Sensor or
parameter
Tc J
Tc K
Tc T
Tc R
Tc S
Tc B
Tc C
Tc G
Tc D
Tc U
Tc L
Tc N
Tc E
Tc F
-210
-350
-270
454
-270
-454
-50
-60
-50
-60
50
140
0
32
0
32
0
32
-200
-330
-200
-330
-270
-450
-270
-454
0
32
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
1200°C
2200°F
1370°C
2500°F
400°C
760°F
1760°C
3200°F
1760°C
3200°F
1820°C
3310°F
2300°C
4172°F
2300°C
4172°F
2300°C
4172°F
400°C
760°F
760°C
1400°F
1300°C
2380°F
1000°C
1840°F
1400°C
2560°F
-190
-310
-160
-256
-130
-238
150
302
170
338
920
1688
0
32
190
374
0
32
-160
-256
-200
-328
0
32
-200
-328
0
32
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
1200°C
2192°F
1260°C
2300°F
400°C
752°F
1760°C
3200°F
1760°C
3200°F
1820°C
3308°F
2000°C
3632°F
2300°C
4172°F
2130°C
3866°F
400°C
752°F
760°C
1400°F
1300°C
2372°F
1000°C
1832°F
1400°C
2552°F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
0.01°C
0.1 °F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
MicroCal 200
Accuracy
(% of rdg)
±(0.02% + 0.1 °C)
±(0.02% + 0.18°F)
±(0.02% + 0.1 °C)
±(0.02% + 0.18°F)
±(0.02% + 0.1 °C)
±(0.02% + 0.18°F)
±(0.02% + 0.2 °C)
±(0.02% + 0.36°F)
±(0.02% + 0.2 °C)
±(0.02% + 0.36°F)
±(0.02% + 0.3 °C)
±(0.02% + 0.54°F)
±(0.02% + 0.2 °C)
±(0.02% + 0.36°F)
±(0.02% + 0.3 °C)
±(0.02% + 0.54°F)
±(0.02% + 0.3 °C)
±(0.02% + 0.54°F)
±(0.02% + 0.1 °C)
±(0.02% + 0.18°F)
±(0.02% + 0.1 °C)
±(0.02% + 0.18°F)
±(0.02% + 0.1 °C)
±(0.02% + 0.18°F)
±(0.02% + 0.1 °C)
±(0.02% + 0.18°F)
±(0.02% + 0.1 °C)
±(0.02% + 0.18°F)
-200
-330
-200
-330
-200
-330
-200
-330
-200
-330
-200
-330
...
...
...
...
...
...
...
...
...
...
...
...
850°C
1570°F
650°C
1210°F
600°C
1120°F
850°C
1570°F
850°C
1570°F
850°C
1570°F
-200
-328
-200
-328
-200
-328
-200
-328
-200
-328
-200
-328
...
...
...
...
...
...
...
...
...
...
...
...
850°C
1562°F
650°C
1210°F
600°C
1112°F
850°C
1562°F
530°C
986°F
850°C
1562°F
0.01°C
0.1 °F
0.01°C
0.1 °F
0.01°C
0.1 °F
0.1 °C
0.1 °F
0.1 °C
0.1 °F
0.01°C
0.1 °F
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
+ 0.05°C)
+ 0.09°F)
+ 0.05°C)
+ 0.09°F)
+ 0.05°C)
+ 0.09°F)
+ 0.15°C)
+ 0.27°F)
+ 0.1 °C)
+ 0.18°F)
+ 0.1 °C)
+ 0.18°F)
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
+
+
+
+
+
+
+
+
+
+
+
+
0.05 °C)
0.09 °F)
0.05 °C)
0.09 °F)
0.05 °C)
0.09 °F)
0.15 °C)
0.27 °F)
0.1 °C)
0.18 °F)
0.1 °C)
0.18 °F)
0.01°C
0.1 °F
0.01°C
0.1 °F
0.01°C
0.1 °F
0.01°C
0.1 °F
0.01°C
0.1 °F
0.01°C
0.1 °F
±(0.005%
±(0.005%
±(0.005%
±(0.005%
±(0.005%
±(0.005%
±(0.005%
±(0.005%
±(0.005%
±(0.005%
±(0.005%
±(0.005%
+
+
+
+
+
+
+
+
+
+
+
+
0.05°C)
0.09°F)
0.05°C)
0.09°F)
0.05°C)
0.09°F)
0.15°C)
0.27°F)
0.1 °C)
0.18°F)
0.1 °C)
0.18°F)
-70
-100
-180
-300
-60
-80
0
32
...
...
...
...
...
...
...
...
150°C
310°F
150°C
310°F
180°C
360°F
150°C
310°F
-70
-94
-180
-292
-60
-76
0
32
...
...
...
...
...
...
...
...
150°C
302°F
150°C
302°F
180°C
356°F
150°C
302°F
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
±(0.02%
+ 0.4 °C)
+ 0.72°F)
+ 0.05°C)
+ 0.09°F)
+ 0.05°C)
+ 0.09°F)
+ 0.05°C)
+ 0.09°F)
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
± (0.01%
+
+
+
+
+
+
+
+
0.4 °C)
0.72 °F)
0.05 °C)
0.09 °F)
0.05 °C)
0.09 °F)
0.05 °C)
0.09 °F)
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
±(0.01%
±(0.01%
±(0.01%
±(0.01%
±(0.01%
±(0.01%
±(0.01%
±(0.01%
+
+
+
+
+
+
+
+
0.4 °C)
0.72°F)
0.05°C)
0.09°F)
0.05°C)
0.09°F)
0.05°C)
0.09°F)
-20
-0.2
-2
-5
0
0
0
0
0
1
1
1
0
0
0
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
Total range
Accuracy Range
Resolution
MicroCal 200+
Accuracy
(% of rdg)
± (0.01% + 0.1 °C)
± (0.01% + 0.18 °F)
± (0.01% + 0.1 °C)
± (0.01% + 0.18 °F)
± (0.01% + 0.1 °C)
± (0.01% + 0.18 °F)
± (0.01% + 0.2 °C)
± (0.01% + 0.36 °F)
± (0.01% + 0.2 °C)
± (0.01% + 0.36 °F)
± (0.01% + 0.3 °C)
± (0.01% + 0.54 °F)
± (0.01% + 0.2 °C)
± (0.01% + 0.36 °F)
± (0.01% + 0.3 °C)
± (0.01% + 0.54 °F)
± (0.01% + 0.3 °C)
± (0.01% + 0.54 °F)
± (0.01% + 0.1 °C)
± (0.01% + 0.18 °F)
± (0.01% + 0.1 °C)
± (0.01% + 0.18 °F)
± (0.01% + 0.1 °C)
± (0.01% + 0.18 °F)
± (0.01% + 0.1 °C)
± (0.01% + 0.18 °F)
± (0.01% + 0.1 °C)
± (0.01% + 0.18 °F)
MicroCal 2000+
Resolution
Accuracy
(% of rdg)
0.01°C
±(0.005% + 0.1 °C)
0.1 °F
±(0.005% + 0.18°F)
0.01°C
±(0.005% + 0.1 °C)
0.1 °F
±(0.005% + 0.18°F)
0.01°C
±(0.005% + 0.1 °C)
0.1 °F
±(0.005% + 0.18°F)
0.01°C
±(0.005% + 0.2 °C)
0.1 °F
±(0.005% + 0.36°F)
0.01°C
±(0.005% + 0.2 °C)
0.1 °F
±(0.005% + 0.36°F)
0.01°C
±(0.01% + 0.3 °C)
0.1 °F
±(0.01% + 0.54°F)
0.1 °C
±(0.01% + 0.2 °C)
0.1 °F
±(0.01% + 0.36 °F)
0.1 °C
±(0.01% + 0.3 °C)
0.1 °F
±(0.01% + 0.54 °F)
0.1 °C
±(0.01% + 0.3 °C)
0.1 °F
±(0.01% + 0.54 °F)
0.01°C
±(0.005% + 0.1 °C)
0.1 °F
±(0.005% + 0.18°F)
0.01°C
±(0.005% + 0.1 °C)
0.1 °F
±(0.005% + 0.18°F)
0.01°C
±(0.005% + 0.1 °C)
0.1 °F
±(0.005% + 0.18°F)
0.01°C
±(0.005% + 0.1 °C)
0.1 °F
±(0.005% + 0.18°F)
0.01°C
±(0.005% + 0.1 °C)
0.1 °F
±(0.005% + 0.18°F)
High Accuracy
Multifunctional Process Calibrators
MULTIFUNCTION
CALIBRATORS
Portable, Rugged and
Easy to Use
Measures and Sources 14 Type
of TCs and 10 RTDs
24 Vdc Loop Power Supply
In-Line Digit Setting
Pannel Mounting Kit
CU100
Ni100
Ni120
mV (L)
mV (H)
V
mA (In)
mA (Out)
Ω
IN
Ω
OUT
Frequency
Pulse counter
Pulse (Out)
±(0.02% + 2 µ
V)
±(0.02% + 10 µ
V)
±(0.02% + 0.08mV)
±(0.02% + 0.4 µ
A)
±(0.02% + 0.4 µ
A)
±(0.02% + 12 mΩ
)
±(0.02% +120 mΩ
)
±(0.02% + 20 mΩ
)
±(0.02% +200 mΩ
)
±(0.005% + 0.001 Hz)
±(0.005% + 0.001 Hz)
±(0.005% + 0.001 Hz)
infinite
1 pulse / min
1 pulse / min
± (0.01% + 2 µ
V)
± (0.01% +10 µ
V)
± (0.01% + 0.08mV)
± (0.01% + 0.4 µ
A)
± (0.01% + 0.4 µ
A)
± (0.01% +12 mΩ
)
± (0.01% +120 mΩ
)
± (0.01% + 20 mΩ
)
± (0.01% +200 mΩ
)
±(0.005% + 0.001 Hz)
±(0.005% + 0.001 Hz)
±(0.005% + 0.001 Hz)
infinite
1 pulse / min
1 pulse / min
• The MicroCal 2000+ relative accuracies shown above are stated for 360
days and the operative conditions are from +70°F to +77°F
• Typical 90 days relative accuracy can be estimated by dividing the "% of
reading" specifications by 1.6.
12
0.1 µ
V
1 µ
V
10 µ
V
0.1 µ
A
0.1 µ
A
1 mΩ
0.01 Ω
1 mΩ
0.01 Ω
0.001 Hz
0.01 Hz
0.1 Hz
1 count
1 pulse / min
1 pulse / h
±(0.0035% + 1 µ
V)
±(0.005% + 10 µ
V)
±(0.005% + 0.08mV)
±(0.005% + 0.4 µ
A)
±(0.005% + 0.4 µ
A)
±(0.005% + 12 mΩ
)
±(0.005% +120 mΩ
)
±(0.005% + 12 mΩ
)
±(0.005% +120 mΩ
)
±(0.005% + 0.001 Hz)
±(0.005% + 0.001 Hz)
±(0.005% + 0.001 Hz)
infinite
1 pulse / min
1 pulse / min
• Typical 2 years relative accuracy can be estimated by multiplying the “% of
reading" specifications by 1.4.
• All input ranges: additional error ±1 digit.
• Eurotron traceability chart and uncertainty can be supplied on request.
bt !
No dou Accurate
st
the Mo rdg)
f
o
e
n
O
5%
or
(0.003
alibrat
C
le
b
a
y!
Por t
le Toda
Availab
ron
ot
Instrum
e
Q
rti
Ce
Note:
• The MicroCal 200/200+ relative accuracies shown above are stated for 360
days and the operative conditions are from +64°F to +82°F
1 µ
V
10 µ
V
0.1 mV
0.1 µ
A
0.1 µ
A
1 mΩ
0.01 Ω
1 mΩ
0.01 Ω
0.001 Hz
0.01
0.1 Hz
1 count
1 pulse/min
1 pulse/h
°C
°F
°C
°F
°C
°F
°C
°F
fie
d
s
nt
+200mV
+2
V
+20 V
+50 mA
+50 mA
500 Ω
5000 Ω
500 Ω
5000 Ω
200 Hz
2000 Hz
20000 Hz
6
10 counts
6000 pulse/min
36000 pulse/h
°C
°F
°C
°F
°C
°F
°C
°F
ste
m
CU10
Eu
r
Pt1000 IEC
OIML
CalpMan 2007
Software for Automatic
Calibration and Documenting in
Compliance with ISO9001
Bulletin 08-44.3 E MicroCal 200_2000 - ed. 04-2007
Pt 500
MicroCal PM200
Optional Module
to Generate/Measure Pressure
Specifications may change without notice
Pt100 IEC
OIML, a 3926
Pt100
a 3902
Pt100 JIS
SAMA
Pt 200
Quality S
y
www.eurotron.com
All descriptions are related to a fully optioned instrument. See last page for the different configurations