BK PRECISION 9120 A, 9121 A, 9124 Single output programmable DC power supply Instruction Manual
Below you will find brief information for Single output programmable DC power supply 9120A, Single output programmable DC power supply 9121A, Single output programmable DC power supply 9124. These models provide you with clean and reliable power with high resolution and accuracy combined with fast transient response times and excellent temperature stability. They feature a convenient interface for adjusting voltage and current values, storing and recalling operating states or enabling/disabling the output. They are suitable for bench or rack mounted operation. The highly compact 912x series is a laboratory grade power supply well suited for applications in design, production or use in university labs.
Advertisement
Advertisement
INSTRUCTION
MANUAL
Models 9120A, 9121A, 9124
SINGLE OUTPUT PROGRAMMABLE
DC POWER SUPPLY SERIES
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
Safety information
Please review the following safety precautions before operating our equipment.
General information
The following safety precautions should be observed before using this product and any associated instrumentations.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the product. Refer to this manual for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product may be impaired.
Before performing any maintenance, disconnect the line cord and all test cables.
Protection from electric shock
Operators of this instrument must be protected from electric shock at all times. The responsible body must ensure that operators are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human contact.
Under these circumstances personnel must be trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 volts, no conductive part of the circuit may be exposed.
Definition of users
Responsible body is the individual or group responsible for the use and maintenance of equipment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
This product should only be used as intended. Users must be trained in electrical safety procedures and proper use of the instrument. Users must be protected from electric shock and contact with hazardous live circuits.
Service is only to be performed by qualified service personnel.
Safety symbols and terms
C
onnect to safety earth ground using the wire recommended in the user manual.
This symbol on an instrument indicates that the user should refer to the operating instructions located in the manual.
Certification
We certify that this product met its published specifications at time of shipment from the factory.
2
TABLE OF CONTENTS
1. Introduction.............................................................................................................. 4
Description ............................................................................................................................................... 4
Features .................................................................................................................................................... 4
2. Quick Reference ...................................................................................................... 5
2.1 The Front Panel .......................................................................................................................................... 5
2.2 The Rear Panel ........................................................................................................................................... 6
2.3 Preliminary Checkout................................................................................................................................. 6
1. Check the list of supplied items............................................................................................................ 6
2. Connect the power cord and turn on the power supply ........................................................................ 7
3. Checkout Procedure ............................................................................................................................. 7
1.4 Output Verification ..................................................................................................................................... 8
Current Output Check .............................................................................................................................. 8
1.5 If the power supply does not turn On ......................................................................................................... 8
1.6 To Adjust the Carrying Handle................................................................................................................... 9
1.7 To Rack Mount the Instrument................................................................................................................... 9
3. Front-panel Operation............................................................................................ 11
3.1 Front Panel Keys ...................................................................................................................................... 11
Numerical Keys/Secondary Shift Functions........................................................................................... 11
Shift Functions ....................................................................................................................................... 11
Primary Function Keys........................................................................................................................... 12
Up/Down and Entry key......................................................................................................................... 12
3.2 Front-panel Operation Overview.............................................................................................................. 12
3.3 Setting the Voltage.................................................................................................................................... 12
3.4 Setting the Current.................................................................................................................................... 13
3.5 Save and Recall Operation ....................................................................................................................... 13
3.6 Menu Operation........................................................................................................................................ 13
3.6.1 Menu Description .......................................................................................................................... 13
3.6.2 Menu Function .............................................................................................................................. 15
3.7 Output Operation...................................................................................................................................... 20
3.8 Remote Sense and digital port functions .................................................................................................. 20
3.9 Digital Volt Meter (DVM)........................................................................................................................ 21
3.10 Milliohm Meter ...................................................................................................................................... 22
4. Remote Operation .................................................................................................. 23
4.1 Communication cable............................................................................................................................... 23
4.2 Communication between Power Supply and PC...................................................................................... 24
4.3 SCPI Command Overview ....................................................................................................................... 24
Common IEEE488.2 Commands ........................................................................................................... 24
Essential SCPI Commands ..................................................................................................................... 25
Port Configuration Command ................................................................................................................ 27
Trigger Commands ................................................................................................................................. 27
SCPI Condition Register ........................................................................................................................ 27
4.4 SCPI Command Description .................................................................................................................... 29
Common IEEE488.2 Commands ........................................................................................................... 29
Essential SCPI Commands ..................................................................................................................... 32
Output Commands.................................................................................................................................. 35
List File Commands ............................................................................................................................... 36
Measurement commands ........................................................................................................................ 38
Interface Configuration Commands ....................................................................................................... 40
Trigger commands.................................................................................................................................. 41
Calibration commands............................................................................................................................ 41
5 Specifications.......................................................................................................... 43
5.1 Specifications ........................................................................................................................................... 43
5.2 Supplemental Characteristics ................................................................................................................... 44
3
1. Introduction
Description
Models 9120A, 9121A and 9124 are fully programmable, linear DC Power Supplies that provide you with clean and reliable power, high resolution and accuracy combined with fast transient response times and excellent temperature stability. The front panel keys and the control knob provide a convenient interface for adjusting
Voltage and Current, storing and recalling operating states or enabling/disabling the output. This power supply is suitable for either bench or rack mounted operation. The 912x is a compact, laboratory grade power supply well suited for applications in design, production or use in university labs.
Features
Very high accuracy and resolution: 0.1mV, 0.1mA
Low ripple and low noise
Fast settling time ﹤20ms
5
1
2
digit digital voltage meter and m
Ω
meter
Convenient data entry via knob or numerical key pad
Over Temperature (OTP) protection
Compact foot print ( 2U x 1/2)
Bright and easy to read display (VFD technology)
Excellent temperature stability
Output on/off control
SCPI compatible command set, communicate via USB or RS232 port
Application Software for front panel emulation and simple test sequence generation
Rack mount kit available
Closed case calibration
Remote Sense Function
Discrete Fault Indicator/Remote Inhibit (DFI/RI). Can be used to turn off power supplies simultaneously.
List Mode: Generate, store and execute test sequences without the need for an external computer
4
2. Quick Reference
2.1 The Front Panel
1 2
3 4 5 6 7
1
VFD display
2
Rotary knob
3
4
Power switch
Numeric keys, auxiliary. functions
5
6
7
Function keys
Up/Down keys and “Enter” key
Output terminals
5
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
2.2 The Rear Panel
1 2
3 4 5 6
1
Digital Voltmeter terminals
2
AC power inlet and fuse compartment
3
4
5
Ventilation holes
Quick connect terminal for Remote sensing and digital port functions (digital I/O, DFI/RI and ext. trigger)
Interface connector for remote control
6
2.3 Preliminary Checkout
AC Power selection switch (110V / 220V)
The following steps help you verify that the power supply is ready for use.
1. Check the list of supplied items
Verify that you have received the following items with your power supply. If anything is missing, contact your authorized B+K Precision distributor.
6
2. Connect the power cord and turn on the power supply
When you turn on the power supply, the front-panel display will light up briefly while the power supply performs its power-on self-test. All the VFD annunciators will turn on at once. Check for any missing display segments.
Refer to section 1.5 in this chapter, if the power supply does not turn on.
3. Checkout Procedure
At power up, the instrument will automatically perform a self test routine. During this time, the following should be displayed:
System Test, Please wait! followed by
0.000V 0.00000A
0.000V 0.000V
The first row displays the actual output voltage value and current and the state of power supply. The second row displays the voltage measured by the DVM (on left) and the Set Value for the voltage of the power supply.
To obtain additional information about the instrument, press and hold the SHIFT button during Power Up. On the display you will see the following:
First row: V/A rating and DVM voltage range second row: Firmware version and serial number
Sourc: XXV XA Meas: XXV
Ver: 1.67 SN:5975002002
Press “Esc” to exit, VFD displays as follows:
In case the self test routine is not successful, you may see one of the following:
If the EEPROM was damaged or the latest operation data in lost, the VFD will display:
ERR EEPROM
If the calibration data stored in the EEPROM is lost, the VFD will display
ERROR CAL
If the latest operating state of the power supply in EEPROM is lost, the VFD will display:
Error Config Data
7
Warning: The power supply is shipped from the factory with a power-line cord that has a plug appropriate for your location. Your power supply is equipped with a 3-wire grounding type power cord; the third conductor being the ground. The power supply is grounded only when the power-line cord is plugged into an appropriate receptacle. Do not operate your power supply without adequate cabinet ground connection.
1.4 Output Verification
The following procedures verify that the power supply outputs the correct voltage and current levels and properly responds to entries from the front panel.
Voltage Output Check
The following steps verify basic voltage functions without load.
1) Turn on the power supply.
2) Enable the outputs
3
)Set the voltage value
Set a different voltage values. Make sure that the set value and output value are the same. Also check if the output current value is zero or close to zero A.
4) Ensure that the voltage can be adjusted from zero to the maximum rated value.
Current Output Check
The following steps check the basic current functionality by shorting the power supply’s output.
1) Turn on the power supply.
3) Connect a short across the (+) and (-) output terminals with an insulated test lead. Use a wire size sufficient to handle the maximum current.
4) Enable the output.
5) Adjust the voltage value to 1.0 volt. Ensure that the that he CC annunciator is lit (power supply is in CC operation mode)
6) Adjust the current value. Set a different Current value and check if the actual Current value is the same as the set Current value. Also verify that the output voltage value is nearly zero.
7) Ensure that the current can be adjusted from zero to the full rated value.
8) Turn off the power supply and remove the short wire from the output terminals.
1.5 If the power supply does not turn On
Use the following steps to help resolve problems you might encounter when turning on the instrument.
1. Verify that there is AC power applied to the power supply.
8
To rack mount a single instrument, order rack mount kit IT-E151
To rack mount two instruments side-by-side, order rack mount kit IT-E151, In this case you don’t need to use the front cover panel.
10
Verify that the power cord is firmly plugged into the power receptacle on the rear panel of the power supply.
Make sure the power outlet you are using is working properly and verify that the power supply is turned on.
2. Verify the power-line voltage setting.
Make sure the voltage selector switch is set according to the present line voltage (110VAC or 220VAC).
Change the voltage setting if it’s not correct.
3. Verify that the correct power-line fuse is installed.
Model
Fuse Description(110VAC
)
Fuse Description
(220VAC)
9120A
9121A
T2.5A 250V
T2.5A 250V
9124 T2.5A 250V
Replace blown fuses according to the table above.
1.6 Adjusting the Carrying Handle
T1.25A 250V
T1.25A 250V
T1.25A 250V
To adjust the position, grasp the handle by the sides and pull outward. Then rotate the handle to the desired position.
Bench-top viewing positions Carrying position
1.7 Rack Mounting the Instrument
You can mount the power supply in a standard 19-inch rack cabinet using the IT-E151 rack mount kit.
Note: Remove the carrying handle and the two plastic ears before rack-mounting the instrument. To remove the handle, grasp the handle on the side, pull outwards and rotate it to a special position where the arrow on the handle and the arrow on the plastic ears are in opposite directions. Now you can pull the handle outwards. After removing the handle, you can remove the two plastic ears with a screw driver.
9
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
3. Front-panel Operation
So far we have covered the quick start chapter which briefly introduced the front panel operation and how to check basic voltage and current functionality. This chapter describes in detail how to operate the instrument manually via the front-panel keys.
This chapter is divided into the following sections: z Front-Panel Operation Overview z Setting the Voltage z Setting the Current z Save/Recall Operation z Menu Operation z On/Off Operation z Remote Sense and digital port functions z m
Ω
Meter z Digital Voltage Meter
3.1 Front Panel Keys
V-set I-set
Menu
Trigger
Save Recall Enter
V/m
Ω
0.1W
1W
8 9
Local
10W
··
Shift On/Off
Numerical keys
Secondary Shift Functions
Numerical Keys/Secondary Shift Functions
Primary Function keys
Shift Functions
V/mΩ: Toggle between DVM and m
Ω
Meter mode
0.1W: Set the range of the m
Ω
Meter to 0.1W
1W: Set the range of the m
10W: Set the range of the m
Ω
Meter to 1W
Ω
Meter to 10W
Menu Set the parameters of the power Supply
List Generate programs in List Mode
Trigger Generate a single trigger pulse (when configured for Immediate mode)
Local Enable front panel operation when in remote mode
0 – 9 Numerical keys for direct entry of values
11
Primary Function Keys
V-set
Set the voltage value
I-set
Save
Recall
Set the current value
Save the current operating data to internal memory
Recall operating data from internal memory
On/Off
Set the output state of the power supply
Shift
Use to access secondary functions
Up/Down and Entry key
▲
:Up key
▼
:Down key
Enter: Press to confirm numerical entries
3.2 Front-panel Operation Overview
1) The power supply at shipment is preconfigured for front-panel operation. At power-on, the power supply will automatically start up in front panel operation mode.
2) When the power supply is in remote operation mode, the front-panel is disabled. You can revert to Local mode, by pressing the Local button or by sending the appropriate SCPI command. Toggling between front-panel and remote operation modes will not result in any change of the output parameters.
3) The output of the power supply can be enabled or disabled from the front panel by pressing
4) The VFD annunciators display the present operating status of the power supply. At power up, the following is displayed:. The top row shows the actual output voltage and output current value and the state of the power supply. The second line shows the voltage value as measured by the DVM and the Set value of the
Voltage. The bottom right field is also used to display the Current Set Value, the Menu parameters and the
Ohm meter range.
3.3 Setting the Voltage
The Voltage can be adjusted from 0V to the maximum rated voltage of each model. There are 2 ways to set the constant voltage value.
Solution1
:
Press the
▲
and ▼ keys or the knob to change the value of the least significant digit
12
Solution2:
1. Press
0
Enter
3.4 Setting the Current
The Current output is adjustable from 0A to the maximum current value of each model.
2. Enter a numerical value or use the
▲ and
▼ keys to change the current value
3.5 Save and Recall Operation
You can store up to 50 different operating states in memory locations 1 through 50. Each operating state includes a constant voltage value, constant current value, maximum output voltage value and voltage step value.
To save a setting:
Set the desired Voltage and Current value
Press the save
Save
key. Use the knob to scroll to one of the memory locations 1 – 50. Press to assign and store the current settings to the selected memory location
To recall a setting:
ENTER
Press the Recall key. Use the knob to scroll to the memory location where the settings you want to recall are stored. Press
ENTER
to recall and activate those settings
You can also use the SCPI command:*SAV
、*RCL to save and recall.
3.6 Menu Operation
3.6.1 Menu Description
Press Shift :Menu to enter menu mode. The menu parameters will be displayed in the bottom right field of the display. Use the
▲
and ▼ keys to scroll through the menu list and press
ENTER
to select a menu and view the parameters. Press
ESC
MENU
to return to the higher level menu and to return to the main operating mode.
Config
Config Init.
Out Recall
On
Off<Default>
PWR-ON Recall
On
Off<Default>
Key Sound Set
Return to the factory default setup value.
Set the Power ON/OFF state after power up.
“Remembers” and restores the Power ON/OFF state of the power supply before power was turned off.
Disable this function.
Recall operating parameters of power supply after power up
“Remembers” and restores the operating parameters of the power supply (voltage, current settings..) before power was turned off.
Disable this function.
Keypad sound setting.
13
Off
Knob Lock Set
On
Off< Default >
Remote Sense
Disable key sound
Enable/disable the rotary knob.
Lock the rotary knob.
Unlock the rotary knob.
Setup voltage measurement Mode.
On
Off< Default >
ShortCut Recall
The power supply will measure the input voltage at the remote sense connector.
The power supply will measure the input voltage at the front panel connector.
Shortcut of the recall function
Off<Default>
Meter Rate Set
Disable this function
Set the update speed of the power supply meter
Low< Default >
Baudrate set
Low speed
Set the baud rate
Baudrate 4800 < Default
>
Baudrate 9600
Baudrate 19200
Baudrate 38400
Comm. Parity Configure the parity bit.
None< Default >
Even
Odd
Address Set Set the communication address (range from 0 to 30)
Address=
**
Port Mode
Trigger< Def >
RI/DFI
DIGITAL I/O
Trig Source
Immediat<Def>
External
Bus
RI Mode
Off< Default >
Select mode of digital port
Setting the trigger mode
Pressing trigger pulse
Shift
+
Trigger
keys will generate a
Ext. Trigger signal is applied to the digital port in the rear panel.
Remote command trigger mode.
Configure the Remote Inhibit (RI) mode
Disable this function
Latching
Live
DFI Source
Off< Default >
QUES
OPER
ESB
RQS
Configure the Discrete Fault Indicator (DFI) mode
Question Bit
Operation Bit
Event State Bit
Require Bit
Key Lock Set Setting keypad password.
Press
Enter
directly to disable the key lock function.
Password= ﹡﹡﹡﹡
Exit
14
Exit
System Set
Max Volt. set
Max= ﹡﹡﹡﹡
Step Volt Set
Set the Maximum Voltage.
Set the voltage step
Exit
List Set
Step=﹡﹡﹡﹡
Configure list files
Call ListFile
Edit ListFile
Recall list operation file.
Edit list operation file.
Continuous
Once
Repeat
Step
Once
Repeat
Save Mode Set
8 X 25 Steps
Users can allocate 4 types of memory space to save the list file.
4 X 50 Steps
2 X 100 Steps
1 X200 Steps
Exit
Out On Timer Output timer. If Timer State is set to ON, power supply output will turn off after the timer elapsed.
Timer State Setting POWER ON timer state
On Enable Output Timer. Output will turn off after value set in menu “Timer Set” counted down to zero.
Off< Default >
Timer Set Setting time of POWER ON timer.
Timer= ﹡﹡S
Exit
3.6.2 Menu Function
ShortCut Recall
This parameter enables a Shortcut version of the Recall functionality available by pressing the Recall button ( see
3.5)
To enable this function, do the following:
1
) Press Shift
+
MENU
When Config is displayed, press
2
) Press ▲、▼ to scroll to ShortCut Recall then press
ENTER
ENTER
3
) Press ▲、▼to turn this feature On,press
4) Press Esc twice to exit from the menu
ENTER
to confirm.
Now you can conveniently recall up to 9 settings by simply pressing one of the corresponding number keys 1 – 9.
The setting previously stored at that location, (using the Save function), will be retrieved and activated. If the selected location is empty, EEPROM ERROR will be displayed.
15
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
Setting the Baud Rate (>BAUDRATE)
This parameter configures the baud rate for serial communication. Possible values are 4800,9600,19200 or
38400. When operating the power supply in remote mode, make sure that you configure identical baud rate settings for the power supply and the computer. The default setting is 4800.
Setting Address (>ADDRESS)
With this parameter, it is possible to address each instrument. The address range is 0 to 30. The default address is zero.
Port Mode
A 4-pin connector in the rear panel is provided for digital input and output signals. This digital port can be configured to provide Fault/Inhibit, External Trigger or Digital I/O functions. The signal level is TTL.
TRIGGER: Pins 1 and 2 can be used to apply external trigger sources to the power supply. These pins can also be used to control the list operation
RI/DFI
:
The Inhibit Input pin can be used to control the output state of the power supply (RI function).
The Fault Output pin (DFI function) can be used to indicate internal faults of the power supply.
DIGITAL I/O
: Read and control output and input state of the 2 available pins.
Mode
Pin
1 Trigger in Inhibit Input Digital Input
2 GND GND GND
3 No Use Fault Indicator Output Digital Output
4 No GND
Trigger Operation
The power supply supports 3 different trigger modes. Immediate, External and Bus. Configure one of the trigger sources before performing trigger operating.
• Trigger Key:
When this function is enabled, you can generate an immediate trigger pulse by pressing
Trigger.
Shift
• External trigger signal (TTL):
When this function is enabled, the power supply can be triggered with a TTL pulse applied to pin 1 of the terminal connector in the rear. The TTL on pulse width should be at least 5 ms.
• Bus
When this function is enabled, you can trigger the power supply by sending a *TRG or TRIgger command to the power supply
Remote Inhibit (RI) - Input
Used to turn off the output of the power supply. Can be used to turn off several power supplies simultaneously.
The RI input has 3 modes: LATCHING
、L IVE and OFF
16
LATCHING When the TTL signal at the RI port transitions from TTL High to Low, the output of the power supply will turn off.
LIVE The output state of the power supply changes according to the signal level applied to the RI port.
If the level is TTL high, the power supply output is on; if the level at the RI port is TTL Low, the output of the power supply is off.
OFF The signal applied to the RI port does not affect the output state of the power supply.
Discrete Fault Interupt (DFI) - Output
Used to indicate that a fault has occurred in power supply.
The DFI function can be activated by state changes of the QUES, OPER, ESB, RQS bits
QUES
:
The output level of DFI reflects the state of the QUES bit. When the QUES bit is 1, the DFI output goes to a low level. When QUES bit is 0, DFI the output goes High
OPER
:
The output level of DFI reflects the state of the OPER bit.
ESB
:
The output level of DFI reflects the state of the ESB bit.
RQS
:
OFF
:
The output level of DFI reflects the state of the RQS bit.
The output level of the DFI port remains high.
KEY LOCK
It is possible to set a password to lock the function keys. After setting the password, all the function keys on the front panel will be locked except the OUT On/Off key. You must enter the correct password to enable the keys again. If you don’t want to lock the function keys and you selected the >KEY LOCK menu, simply pres the Enter key, do NOT enter any numbers
When shipped from the factory, no password is set and the function keys are unlocked. The start bit of your desired password shouldn’t be 0.
List Set
This mode allows you to create a sequence of steps, store it into the power supplies non volatile memory and execute it. The input parameters for generating a list include the name of the list file, the input steps (no more than 200 steps), the step time (the minimum is 1mS) and the value of each step. The list file can be stored in
ROM with a capacity of 4K from where it can be recalled. This memory area is divided into 4 areas A, B, C, D.
Each area can store a certain amount of groups, and each group has an assigned maximum capacity according to the table below:
Memory area Number of groups/ area Capacity/group
Example: Group C con store 4 groups, each group has a maximum capacity of 1 KByte.
17
The power supply executes a list in CONTINIOUS or Step mode. In CONTINIOUS mode, the power supply will start executing the list once, as soon as a trigger signal is received, then it will wait until the next trigger signal is received.
Before you edit the list file, set the trigger source in the menu to “immediate”.
Example:
1) Press
Shift
+
2) Press up or down key to select “Config”, press
Enter
3) Press up or down key to select “Trig Source”, press
to confirm.
Enter
to confirm.
4) Press up or down key to select “Immediat”, press
Memory area
Enter
to confirm.
Number of groups/ area Capacity/group
Example: Group C can store 4 groups, each group has a maximum capacity of 1 KByte.
The power supply executes a list in CONTINIOUS or Step mode. In CONTINIOUS mode, the power supply will start executing the list once, as soon as a trigger signal is received, then it will wait until the next trigger signal is received.
Operation
:
Generate file: 2 cycles, continuous mode
1
)
Press
Shift
+
Menu into menu operation.
2
)
VFD displays Config
,press▼to select List Set,press Enter to confirm.
3
)
VFD displays Call ListFile
,press▼to select Edit ListFile, press
Enter
to confirm.
4
)
VFD displays Continuous
,press Enter
to confirm.
5
)
VFD displays Repeat
,press
Enter
to confirm
6
)
VFD displays Second, Select ms then press
Enter
to confirm
7
)
VFD displays List Count= _
,press numeric key or move the rotary knob, set number of cycles ( in this example, count is 2)
,press
Enter
to confirm.
8
)
VFD displays1th=
*.****V, press numeric key or move the rotary knob, set the maximum voltage, press
Enter
to confirm.
9
)
VFD displays 1th=
*.****A, press numeric key or move the rotary knob, set the maximum current, press
Enter to confirm.
10
) VFD displays 1th=*mS, press numeric key or move the rotary knob, set delay time, press Enter to confirm.
11
) VFD displays 2th=*.****V, press numeric key or move the rotary knob, set the maximum voltage, press
Enter to confirm.
12
) VFD displays 2th=*.****A, press numeric key or move the rotary knob, set the maximum current, press
Enter
to confirm.
13
) VFD displays 2th=*mS,press numeric key or move the rotary knob, set delay time, press Enter
18
to confirm.
14
) VFD display Store File_,press numeric key or move the rotary knob, set the register number(1 to 8), press
Enter
to confirm.
15
) Press
Esc
two times to escape menu operation.
Execute file
16
) Press
Shift
+
List
to set the list file, then press
Shift
+
Trigger
to run the list file. Press
Shift
+
List to stop.
If you have created several list files, you can call the list file that you need by “Call ListFile” function in the menu. Press
Shift
+
Esc
two times to exit menu operation. And then press
Trigger
to run the file that you called. Press
Shift
+
List
Shift
+
List to stop running.
to set it, press
Continuous mode: Once the trigger signal is received, the programmed list will be executed once. Once finished, the instrument will pause until the next trigger signal is received.
Trigger Trigger
Step mode: The power supply will advance to the next step only after it receives a trigger signal.
Trigger
Trigger
Trigger Trigger
Trigger
Trigger
Trigger Trigger
Digital I/O
When the digital port of the power supply mode is in DIGITAL I/O mode and the power supply is configured for remote operation, you can send SCPI command (DIGital:INPut[:STATe?] and DIGital:OUTPut[:STATe?])to
19
read and set the state of the output and input port.
3.7 Output Operation
For front panel operation, press
On/Off
to enable/disable the output. If the power supply is in remote control mode, you can send SCPI command (OUTPut ON|OFF) to change the state of output.
3.8 Remote Sense and digital port functions
8 pin connector in rear panel
+
-
+S
-
S INH GND FLT GND
+S,-S Remote sense pins
Remote voltage sensing is used to maintain good regulation at the load and reduce the degradation of regulation that would occur due to the voltage drop in the leads between the power supply and the load. By connecting the supply for remote voltage sensing, voltage is sensed at the load rather than at the supply's output terminals. This will allow the supply to automatically compensate for the voltage drop in the load leads and improve regulation.
You must set remote sense mode before you start remote test function.
+,- Output
These pins are identical to the out put terminals in the front
INH Inhibit Input pin
This pin has multiple functions: a) When “Port Mode” in the menu is set to “Trigger”, INH is configured as trigger Input. b) When “Port Mode” in the menu is set to “RI/DFI”, INH can be used to turn the power supply output on/off.
INH can be configured in 3 ways:
LATCHING
:When the level of INH port changes from high to low, the output of power supply turns off.
LIVE
:
The output state of power supply changes according to the level of the INH port. If the level
OFF
: of INH is TTL High, the output is on. If the level at pin INH is TTL Low, the output of the power supply is off.
The level state of INH does not affect the output state of the power supply. c) When “Port Mode” in the menu is set to “DIGITAL I/O”, INH port functions as a digital input pin. The status of the pin can be read via SCPI command
FLT: Fault Output pin
This pin hat multiple functions: a) When “Port Mode” in the menu is set to “Trigger”, FLT has no function. b) When “Port Mode” in the menu is set to “RI/DFI”, the FLT pin can be used to determine the reason of a power supply fault.
The status of FLT can be derived from bits QUES, OPER, ESB, RQS or it can be disabled
QUES
: The output level of FLT reflects the state of the QUES bit. When the QUES bit is 1, FLT output a
20
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
Low signal, otherwise, if QUES bit is 0, FLT goes High.
OPER
: The output level of FLT reflects the of the OPER bit.
ESB
: The output level of FLT reflects the state of the ESB bit.
RQS
: The output level of FLT reflects the state of the RQS bit.
OFF
: The output level of FLT remains high. c) When “Port Mode” in menu is set to “DIGITAL I/O”, FLT functions as an output pin. The status of the port can be programmed via SCPI command.
3.9 Digital Volt Meter (DVM)
The power supply provides a built-in Digital Volt meter which can measure DC volts in a range from 0 - 40V.
The connectors are in the rear panel. The voltage value is displayed on the bottom left field of the display.
To measure voltages, connect the leads as shown here:
+
DVM
Load
By default, the power supply is in DVM mode. Press and m
Ω
Meter mode.
Shift
followed by V/m
Ω
to toggle between Voltmeter
21
3.10 Milliohm Meter
The instrument is also equipped with a Milliohm Meter which can accurately measure resistance up to 10Ω. To protect the resistor, make sure to select an appropriate power range before connecting it to the power supply. To measure resistance, connect the resistor as shown below:
+
+
DV M
OUTPUT
Load
To measure resistance:
1) Enable Ohmmeter mode: Press
0.1W
)
2) Press
Shift
Shift
followed by V/m
Ω
(Display will show ---.—m
Ω
Range
followed by 0.1W /1W /10W to select an appropriate range for the resistor you are measuring.
Measurement range
Test voltage (Output)
0.1W
1V
1W
3.3V
10W
10V
22
4. Remote Operation
The DB9 interface connector on the rear panel of the power supply can be connected to a RS-232 or USB interface via the included adapter cable. This chapter describes how to use a computer to control the output of the power supply.
4.1 Communication cable
RS232 Communication cable IT-E131
The DB9 interface connector on the rear panel of the power supply provides a TTL voltage level interface. Use the communication cable (IT-E131) to connect the DB9 interface connector of the power supply to the RS-232 interface connector of the computer
IT-E131 communication cable
Power
Note: It is not possible to connect the DB9 connector of the power supply via a standard
RS232/USB cable to a PC’s RS232 port or USB port.
USB Communication cable IT-E132
The DB9 interface connector on the rear panel of the power supply provides a TTL voltage level interface. Use the communication cable (IT-E132) to connect the DB9 interface connector of the power supply to the USB interface connector of the computer.
Power
Note: Before you can use the USB communication cable, you must install the USB driver which can be found on the included installation disk.
23
4.2 Communication between Power Supply and PC
Before putting the instrument into remote operation, make sure that the baud rate, parity bit and communication address settings on the power supply and computer side are identical, otherwise communication will not be possible.
1. Address Range is 0 to 30. Default setting is 0.
2. Baud rate: 4800,9600,19200 and 38400 are selectable, default setting is 9600
3. Parity and Data bits: None/8 bits (default setting)
Even/8 bits
Odd/8 bits
4. Stop bits:1 (fixed)
5. Start Bits: 1 (fixed)
Data Frame Format
Parity=None
Start
Bit
8 Data Bits
Stop
Bit
Parity=Odd,
Even
Start
Bit
8 Data Bits
Parity
Bit
End of String character is ’\n’ (0x0a)
DB9 Interface Details
Stop
Bit
The DB9 connector in the rear panel of the power supply provides a TTL level signal .It can be connected to a standard PC interface via the IT-E131 or IT-E132 isolated communication cable.
Note: Setting of the address is optional. It is not required to communicate with the instrument. The address can be set from the front panel and is stored in non volatile memory. This feature is useful when connecting several instruments, e.g. via USB hub. In this scenario, Windows assigns a virtual COM port to each device which is unknown prior to establishing communications with the instrument (could be different each time). In this case, the address will allow the user to correlate each virtual COM port with the address assigned to each instrument.
4.3 SCPI Command Overview
Common IEEE488.2 Commands
*CLS
*ESE
24
*ESE?
*ESR?
*IDN?
*OPC
*OPC?
*PSC
*PSC
?
*RST
*SRE
*SRE?
*STB?
*TRG
*SAV
*RCL
Essential SCPI Commands
SYSTem
:ERRor[:NEXT]?
:VERSion?,
:ADDRess?
:REMote
:LOCal
:RWLock
STATus
:QUEStionable
[:EVENt]?
:CONDition?
:ENABle <VALUE>
:ENABle?
:OPERation
:[EVENt]?
: CONDition?
:ENABle <VALUE>
:ENABle?
Calibration Command
CALibration
:SECure
[:STATe] {<ON|OFF>,<quoted code>}
[:STATe]?
:VOLTage
:LEVel {<level> }
[:DATA] {<numeric value>}
:CURRent
:LEVel {<level> }
[:DATA] {<numeric value>}
:DVM
:SAVe
:INITital
25
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
Output Commands
OUTPut
[:STATe] {<bool>}
[:STATe]?
:TIMer
[:STATe] {<bool>}
[:STATe]?
:DATA {<timer>}
:DATA?
[SOURce:]
MODE {<FIXed|LIST|DRM>}
MODE?
VOLTage
[:LEVel] {<n>}
[:LEVel]?
:PROTection
CURRent
[:LEVel] {<n>}
[:LEVel]?
LIST
:STATe?
[:LEVel] {<n>}
[:LEVel]?
:MODE {<mode>}
:MODE?
:STEP {<step>}
:STEP?
:COUNt {<n>}
:COUNt?
:CURRent
[:LEVel] {<n>,<n>}
[:LEVel]? {<n>}
:VOLTage
[:LEVel] {<n>,<n>}
[:LEVel]? {<n>}
:WIDth {<n>,<n>}
:WIDth? {<n>}
:NAME {<string code>}
:NAME?
:AREA {1|2|4|8}
:AREA?
:SAVe {1|2|3|4|5|6|7||8}
:RCL {1|2|3|4|5|6|7|8}
Input Measure Command
MEASure
[:SCALar]
:VOLTage[:DC]?
:CURRent[:DC]?
:POWer[:DC]?
:DVM[:DC]
?
:RESistance[DC]?
26
Port Configuration Commands
[SOURce:]
SYSTem
: SENSe [:STATe] {<bool>}
[:STATe]?
PORT
RI
:MODE?
DFI
:SOURce?
DIGital
:INPut[:STATe]?
SENSe
:RESistance:
:RANGe {LOW | MIDdle | HIGH>}
:RANGe?
Trigger Command
TRIGger
[:IMMediate]
:SOURce {<source>}
SCPI Condition Register
You can obtain the state of the power supply and read parameters from the operation register. The different states of the power supply can be read from 7 condition registers. These registers are status byte register, standard event register, quest condition register and operation status register. The status byte register stores the information of 3 other registers. The following table provides the details on each register’s meaning.
4
5
7
0
2
3
3
BIT Signal Meaning
4
0
0
1
2
3
1
2
CAL
WTG
CV
CC
RI
OV
OT
UNR
Operation status register
The power supply is calculating new calibration parameter.
The power supply is waiting for a trigger signal.
The power supply is in constant voltage condition.
The power supply is in constant current condition.
Show the input level condition of RI
Quest condition register
Over voltage
Over temperature
The output of the power supply is unregulated.
OPC
QYE
DDE
EXE
CME
PON
QUES
Standard event status register
Operation of power supply is completed.
Query error. Data of output array is missing.
Device-dependent error. Data stored in register is missing or error occurs in preliminary checkout.
Execution error. Command parameter overflows or the condition is not right.
Command error. Syntax or semantic error occurs when receiving information.
Power on. It is 1when power supply is reset.
Status byte register
If a quest enable condition changes, QUES is 1.
27
5
6
7
ESB
MSS
RQS
OPER
If a standard event status enable register changes, ESB is 1.
If a operation event enable register changes, OPER is 1.
Structure of condition register:
Quest condition register
condition event enable
OV
OT
Unr n.u. n.u. n.u
0
1
2
3
4
5
0
1
2
3
4
5
0
1
2
3
4
5
L
O
G
I
C
O
R n.u n.u
6
7
6
7
6
7
Operation event register Status byte register
condition event enable event enable
CV n.u
I
C
O
R
3
4
5
6
7
0
1
2
3
4
5
6
7
0
1
2
Standard event status register
event enable
OPC
0 0 n.u
QYE
1
2
1
2
DDE
EXE
3
4
3
4
CME
n.u
5
6
5
6
PON
7 7
C
O
R
L
O
G
I
C
O
R
L
O
G
I
28
4.4 SCPI Command Description
Common IEEE488.2 Commands
*CLS
This command clears the following registers:
Standard event status register
Quest condition register
Operation event register
Status byte register
Error code
Command syntax
:*CLS
Parameter
:None
*ESE
This command sets the parameter of the standard event enable register. The value determines which bit value of the standard event register is set to 1. The ESB bit in the of status byte register will change to 1 to reflect the changes.
Command syntax
:*ESE <NRf>
Parameter
:0~255
Reset value
:Consult *PSC command
Example
:*ESE 128
Query syntax
:*ESE?
Return parameter
:<NR1>
Reference command
:*ESR? *PSC *STB?
Bit map of standard event status enable register
Bit position
Bit Name
7 6 5
32
4
16
3
8
2
4
1
DDE Device-dependent error
QYE Query error
OPC Operation complete
0
PON Not used CME EXE DDE QYE Not used OPC
Bit Weight
PON Power-on
128
CME Command error
EXE Execution error
*ESR?
This command queries the standard event status register. After executing this command, the standard event status register is reset. The bit definition of the standard event status register is identical to the standard event status enable register
Query syntax
:*ESR?
Parameter
:None
Return parameter
:<NR1>
Reference command
: *CLS *ESE *ESE? *OPC
29
*IDN?
This command reads information about the power supply. The return value contains 4 segments divided by a comma.
Query syntax
:*IDN?
Parameter
:None
Return parameter
:<AARD> segment description
BK Precision manufacturer
XXXX product mode
XXXXXX product serial number
VX
.XX software version number
For example
:BK PRECISION,9120A,000004,V1.01
*OPC
When all commands sent to the instrument prior to this command have been executed, OPC of the standard event status register will be set to 1.
Command syntax
:*OPC
Parameter
:None
Query syntax
:*OPC?
Return parameter
:<NR1>
*PSC
This command controls whether or not the power supply sends when it is reset.
1 OR ON: When power supply is reset, operation event enable register, query event enable register and standard event status register are all reset.
0 OR OFF: The data of status byte register, operation event enable register, quest event enable register and standard event status enable register is stored in nonvolatile register, and is recalled when power supply is reset.
Command syntax
:*PSC <bool>
Parameter
: 0|1|ON|OFF
Query syntax
:*PSC?
Return parameter
:0|1
Reference command
:*ESE *SRE STAT:OPER:ENAB STAT:QUES:ENAB
*RST
This command resets the power supply to its default setting.
CAL:SEC:STAT OFF
OUTP OFF, CURR MAX
VOLT:PROT:MAX
VOLT: MIN
TRIG:SOUR BUS
SYST:SENS OFF
PORT:MODE TRIG
RI:MODE OFF
DFI:SOUR OFF
VOLT:PROT:STAT OFF
30
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
Command syntax
:*RST>
Parameter
:None
*SRE
This command can set the parameter of the status byte enable register. The value of this parameter determines which bit value of the status byte register is 1 and the byte will enable RQS of status byte register is 1. The bit definition of the status byte enable register is the same as the status byte register.
Command syntax
:*SRE <NRf>
Parameter
:0~255
Reset value
:Consult *PSC command
Example
:*SRE 128
Query syntax: *SRE
?
Return parameter
:<NR1>
Reference Command
:*ESE *ESR? *PSC *STB?
*STB?
This command reads data from the status byte register. After executing this command, the status byte register is reset.
Query syntax
:*STB?
Parameter
:None
Return parameter
:<NR1>
Reference command
: *CLS *ESE *ESR
Bit map of the standard event status enable register
Bit Name OPER RQS ESB no use QUES no use no use no use
*TRG
When the power supply’s trigger source is set to Bus command, this command will generate a trigger signal. Its function is identical to the [SYSTem:]TRIGger command.
Command syntax
:*TRG
Parameter
:None
Reference command
:TRIG TRIG:SORU
*SAV
This command saves the operating parameters of the power supply to non volatile memory. The parameters include constant current, constant voltage, maximum voltage value and step voltage values.
Command syntax
:*SAV<NRf>
Parameter
:1~50
Example
:*SAV 3
Reference command
:*RCL
31
*RCL
This command recalls the parameter saved with the *SAV command.
Command syntax
:*RCL<NRf>
Parameter
:1~50
Example
:*RCL 3
Reference command
:*SAV
Essential SCPI Commands
SYSTem:ERRor[:NEXT]?
This command queries the error code and error information of the power supply.
(0) No error
(1) Too many numeric suffices in Command Spec
(10) No Input Command to parse
(14) Numeric suffix is invalid value
(16) Invalid value in numeric or channel list, e.g. out of range
(17) Invalid number of dimensions in a channel list
(20) Parameter of type Numeric Value overflowed its storage
(30) Wrong units for parameter
(40) Wrong type of parameter(s)
(50) Wrong number of parameters
(60) Unmatched quotation mark (single/double) in parameters
(65) Unmatched bracket
(70) Command keywords were not recognized
(80) No entry in list to retrieve (number list or channel list)
(90) Too many dimensions in entry to be returned in parameters
(101) Command Execution error
(100) Too many command
(110) Rxd error Parity
1. Error EEPROM
2.
3.
4.
Config data error
Error Calibration data
Factory Data error
Command syntax
:SYST:ERR?
Parameter
:None
Return parameter
:〈NR1〉,〈SRD〉
SYSTem:VERSion?
This command queries the software vers i on.
Command syntax
:SYST:VERS?
Parameter
:None
Return parameter
:<NR2>
SYSTem:ADDRess?
This command queries the address of the power supply
Command syntax
:SYST:ADDR?
32
Parameter
:None
Return parameter
:<NR2>
SYSTem:REMote
This command puts the power supply in remote control mode.
Command syntax
:SYST:REM
Parameter
:None
Query syntax
:None
SYSTem:LOCal
This command configures the instrument for front panel operation
Command syntax
:SYST:LOC
Parameter
:None
Query syntax
:None
SYSTem:RWLock[:STATe]
This command also sets the instrument to remote control mode. When using this command, it is not possible to press LOCAL key on the front panel to revert back to manual mode.
Command syntax
:SYST:RWL
STATus:QUEStionable[:EVENt]?
This command queries the parameters of the quest event register. After execution, the quest event register is reset.
Query syntax
:STATus:QUEStionable[:EVENt]?
Parameter
:None
Return parameter
:<NR1>
Reference command
: STATus:QUEStionable:ENABle
Bit map of standard event status enable register
Bit name no use no use no use no use no use unr OT OV
STATus:QUEStionable:CONDition?
This command queries the parameters of the quest condition register. When a bit of the quest condition changes, the corresponding bit value in the quest event register will be set to 1.
Query syntax
:STATus:QUEStionable: CONDition?
Parameter
:None
Return parameter
:<NR1>
STATus:QUEStionable:ENABle
This command sets the parameter of the quest event enable register. This parameter determines which bit of the quest event register is set to 1. If a QUES condition changes, the QUES bit of status byte register will be set to
33
1.
Command syntax
:STATus:QUEStionable:ENABle <NRf>
Parameter
:0~255
Reset value
:Consult *PSC command
Example
:STATus:QUEStionable:ENABle 128
Query syntax
:STATus:QUEStionable:ENABle?
Return parameter
:<NR1>
Reference command
:*PSC
STATus:OPERation[:EVENt]?
This command queries the parameters of the operation event register. After executing this command the operation event register is reset.
Query syntax
:STATus: OPERation [:EVENt]?
Parameter
:None
Return parameter
:<NR1>
Reference command
: STATus: OPERation:ENABle
Bit map of standard event status enable register
Bit Name no use no use no use RI CC CV WTG CAL
STATus:OPERation:CONDition?
This command queries the parameters of the operation condition. When a parameter of the operation condition register changes, the corresponding bit in the operation event register will be set to 1.
Query syntax
:STATus: OPERation: CONDition?
Parameter
:None
Return parameter
:<NR1>
STATus:OPERation:ENABle
This command sets the parameter of the operation event enable register. The parameter determines which bit value of quest event register is set to 1. If a OPER condition changes, the OPER bit of the status byte register will be set to 1.
Command syntax
:STATus: OPERation:ENABle <NRf>
Parameter
:0~255
Reset value
:Consult *PSC command
Example
:STATus: OPERation:ENABle 128
Query syntax
:STATus: OPERation:ENABle?
Return parameter
:<NR1>
Reference command
:*PSC
34
Output Commands
OUTPut[:STATe]
This command sets the power supply output on or off.
Command syntax
:OUTPut[:STATe] <bool>
Parameter
:0|1|ON|OFF
*RST value
:OFF
Query syntax
:OUTPut:STATe?
Return parameter
:0|1
OUTPut:TIMer[:STATe]
This command sets the output timer state of the power supply.
Command syntax
:OUTPut:TIMer[:STATe] <bool>
Parameter
:0|1|ON|OFF
*RST value
: OFF
Query syntax
:OUTPut:TIMer[:STATe]?
Return parameter
:0|1
OUTPut:TIMer:DATA
This command sets the time of the ou t put timer.
Command syntax
:ONPut:TIMer:DATA <NR1>
Parameter
:<NR2>
*RST value
:1
Query syntax
:OUPut:TIMer:DATA?
Return parameter
:<NR2>
[SOURce:]MODE
This command configures the power supply for command fixed mode, list mode or DVM mode.
FIXed Command fixed mode
LIST List mode
DRM Digital milliohm meter
Command syntax
:[SOURce:]MODE <mode>
Parameter
:FIXed|LIST|DRM
*RST value: FIXed
Example
: MODE FIX
Query syntax
:[SOURce:] MODE?
Return parameter
:<CRD>
[SOURce:]CURRent [:LEVel]
This command sets the current value of the power supply.
Command syntax
:[SOURce:]CURRent [:LEVel] <NRf>
Parameter
:MIN TO MAX|MIN|MAX
Unit
:A mA
35
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
*RST value: MIN
Example
: CURR 3A, CURR 30mA, CURR MAX, CURR MIN
Query syntax
:[SOURce:]CURRent [:LEVel]?
Parameter
:[MIN|MAX]
Example
:CURR?, CURR? MAX, CURR?MIN
Return parameter
:<NR2>
[SOURce:]VOLTage[:LEVel]
This command sets the voltage value of the power supply.
Command syntax
:[SOURce:]VOLTage[:LEVel] <NRf>
Parameter
:MIN TO MAX|MIN|MAX
Unit
:V mV kV
*RST value
:MAX
Query syntax
:[SOURce:]VOLTage[:LEVel]?
Parameter
:[MIN|MAX]
Return parameter
:<NR2>
List File Commands
[SOURce:]LIST:MODE
This command sets the trigger condition for executing the list file
CONTinious List operation is continuous mode.
STEP List operation is step mode.
Command syntax
:[SOURce:]LIST:MODE <CRD>
Parameter
: CONTinious|STEP
Query syntax
:[SOURce:]LIST:MODE?
Return parameter
:<CRD>
[SOURce:]LIST:STEP
This command sets the operation mode of the list file.
ONCE List operate once
REPeat Repeat list operation
Command syntax
:[SOURce:]LIST:STEP <SRD>
Parameter: ONCE|REPeat
Query syntax
:[SOURce:]LIST:STEP?
Return parameter
:<CRD>
[SOURce:]LIST:COUNt
This command sets the number of steps for the list operation.
Command syntax
:[SOURce:]LIST:COUNt <NRf>
Parameter
:2~400
36
Query syntax
:[[SOURce:]LIST:COUNt?
Parameter
:None
Return parameter
:<NR1>
[SOURce:]LIST :CURRent[:LEVel]
This command sets the current step.
Command syntax
:[SOURce:]LIST:CURRent[:LEVel]
Parameter
:0~30A
Unit
:A mA
Example
:LIST:CURR 1,3A;
Query syntax
:[SOURce:]LIST:CURRent:[LEVel]?
Parameter
:None
Example
:LIST:CURR? 1;
Return parameter
:<NR2>
[SOURce:]LIST :VOLTage[:LEVel]
This command sets the voltage step.
Command syntax
:[SOURce:]LIST:VOLTage [:LEVel] <NRf>
Parameter
:0~360V
Unit
:V mV
Example
:LIST:VOLT 1,3V;
Query syntax
:[SOURce:]TRANsition: VOLTage:TLEVel?
Parameter
:None
Example
:LIST:VOLT? 1;
Return parameter
:<NR2>
[SOURce:]LIST:WIDth
This command sets the minimum step time.
Command syntax
:[SOURce:]LIST:WIDth <NRf>
Parameter
:MIN TO MAX|MIN|MAX
Unit
:S mS
Example
:LIST:WID 1, 100mS;
Query syntax
:[SOURce:]LIST:WIDth?
Parameter
:None
Example
:LIST:WID? 1;
Return parameter
:<NR2>
[SOURce:]LIST:NAME
This command sets the name for the list file. Make sure the file name does not exceed 8 characters.
Command syntax
:[SOURce:]LIST:NAME <name>
37
Parameter
:<SRD>
Example
:LIST:NAME ‘TEST’;
Query syntax
:[SOURce:]LIST:NAME?
Return parameter
:<SRD>
[SOURce:]LIST:AREA
This command divides up the storage area for the list file in one of the 4 ways listed below.
1
.1 group per store area, 400 steps
2
.2 groups per storage area, each group contains 200 steps.
4
.4 groups per storage area, each group has 100 steps.
8
.8 groups of storage area, each group has 50 steps.
Command syntax
:[SOURce:]LIST:AREA <NR1>
Parameter
:1|2|4|8
Example
:LIST:AREA 1
Query syntax
:[SOURce:]LIST:AREA?
Return parameter
:<NR1>
[SOURce:]LIST:SAVe
This command saves the list file to a register (non volatile m e mory).
Command syntx
:[SOURce:]LIST:SAVe <NR1>
Parameter
:1~8
Example
:LIST:SAV 1
[SOURce:]LIST:RCL
This command can recall the list file saved before from the register.
Command syntax
:[SOURce:]LIST:SAV <NR1>
Parameter
:1~8
Example
:LIST:SAV 1
Measurement commands
MEASure[:SCALar]:VOLTage[:DC]?
This command queries the input voltage of the power supply.
Command syntax
:MEASure[:SCALar]:VOLTage[:DC]?
Parameter
:None
Return parameter
:〈NR2〉
Return parameter unit
:V
Example
:MEAS:VOLT?
MEASure[:SCALar]:CURRent[:DC]?
This command queries the input current of the power supply.
Command syntax
:MEASure[:SCALar]:CURRent[:DC]?
38
Parameter
:None
Return parameter
:〈NR2〉
Return parameter unit
:A
Example
:MEAS:CURR?
MEASure[:SCALar]:POWer[:DC]?
This command queries the input power of the power supply.
Command syntax
:MEASure[:SCALar]:POWer?
Parameter
:None
Return parameter
:〈NR2〉
Return parameter unit
:W
Example
:MEAS:POW?
MEASure[:SCALar]:DVM[:DC]?
This command queries the voltage reading of the digital volt meter.
Command syntax
:MEASure[:SCALar]:DVM?
Parameter
:None
Return parameter
:〈NR2〉
Return parameter unit
:V
Example
:MEAS:DVM?
SENSe:RESistance:RANGe
This command sets the range of the milliohm meter.
LOW: 0.01W resistance range
MIDDLE: 0.1W resistance range
HIGH: 1W resistance range
Command syntax
:SENSe:RESistance:RANGe
Parameter
:LOW | MIDdle | HIGH
Example
:RES:RANG LOW
Query syntax
:SENSe:RESistance:RANGe?
Return parameter
:<SRD>
MEASure[:SCALar]:RESistance[:DC]?
This command queries the resistance reading of the milliohm meter.
Command syntax
:MEASure[:SCALar]: RESistance?
Parameter
:None
Return parameter
:〈NR2〉
Return parameter unit
:R
Example
:MEAS:RES?
39
Interface Configuration Commands
[SOURce:]SYSTem:SENSe [:STATe]{<bool>}
This command enables/disables the power supply’s remote sense function.
Command syntax
:SYSTem: SENSe [:STATe] <bool>
Parameter
:0|1|ON|OFF
Query syntax
:SYSTem:SENSe [:STATe]?
*RST value
:0
[SOURce:]PORT:FUNCtion
This command sets the mode of the port in the rear panel.
TRIGGER function: Pin1, pin2 are configured as external trigger source for the power supply and to control the list operation.
RI/DFI function
:
Inhibit Input controls the output state of the power supply. The Fault Output can indicate the reason for internal failure.
。
DIGITAL I/O function
:Read and control the state of the digital port
Command syntax
:SOURce:PORT:FUNCtion
Parameter
:TRIGger|RIDFi|DIGital
Query syntax
:SOURce:PORT:FUNCtion?
*RST value
:TRIGger
[SOURce:]RI:MODE
This command sets the input mode of the RI input pin
Command syntax
:SOURce:RI:MODE
Parameter
:OFF|LATChing|LIVE
Query syntax
:SOURce:RI:MODE?
*RST value
:OFF
[SOURce:]DFI:SOURce
This command sets output source of the DFI output pin.
Command syntax
:SOURce:DFI:SOURce
Parameter
:OFF|QUES|OPER|ESB|RQS
Query syntax
:SOURce:DFI:SOURce?
*RST value
:OFF
[SOURce:]DIGital:OUTPut[:STATe]
This command sets the output state of the digital port. This command can be used when the mode of the port is set to DIGITAL.
Command syntax
:SOURce:OUTPut[:STATe]
Paremeter
:OFF|ON|0|1
40
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
[SOURce:]DIGital:INPut[:STATe]?
This command sets the input state of port. This command can be used when the mode of the port is set to
DIGITAL.
Command syntax
:SOURce:DIG:INPut[:STATe] ?
Trigger commands
TRIGger[:IMMediate]
When trigger source is command mode, this command will give a trigger signal. Its function is as the same as
*TRG command.
Order syntax
:* TRIGger[:IMMediate]
Parameter
:None
Reference order
:TRIG TRIG:SORU
TRIGger:SOURce
This command sets the trigger mode of the power supply.
• IMMediate:
When this function is enabled, you can generate an immediate trigger pulse by pressing
Trigger.
• EXTernal trigger signal (TTL):
Shift
plus
When this function is enabled, the power supply can be triggered with a TTL pulse applied to pin 1 of the terminal connector in the rear. The TTL on pulse width should be at least 5 ms.
• BUS Command
When this function is enabled, you can trigger the power supply by sending a *TRG or TRIgger command to the power supply
Command syntax
:TRIGger:SOURce <mode>
Parameter
:IMMediate|EXTernal|BUS
*RST value: KEY
Calibration commands
CALibration:SECure:[STATe]
Set protection mode enable or disable when calibrating the power supply.
Command syntax
:CALibration:SECure:[STATe ]{ON|OFF>,[<password>]}
Parameter
:0|1|ON|OFF, ‘5811
Example
:CAL:SEC 1, ‘5811; CAL:SEC OFF
Query syntax
:CALibration:SECure:STATe?
Parameter
:None
CALibration:VOLTage:LEVel
This command sets the voltage calibration point.
.
Command syntax
:CALibration:VOLTage:LEVel <point>
41
Parameter
:P1|P2
CALibration:VOLTage [:DATA]
{<numeric value>}
Returns the actual output voltage value of calibration point.
Command syntax
:CALibration:VOLTage [:DATA] <NRf>
Parameter
:<NRf>
Example
:CAL:VOLT 30.0002V
CALibration:CURRent:LEVel
This command sets the current calibration points. P1,P2,P3,P4.
Command syntax
:
CALibration:CURRent:LEVel <point>
Parameter
:P1|P2
CALibration:CURRent [:DATA]
{<numeric value>}
Return actual output current value to calibration point.
Command syntax
:
CALibration:CURRent [:DATA] <NRf>
Parameter
:<NRf>
Example
:CAL:VOLT 3.0002A
CALibration:DVM:LEVel
This command can set current calibration point. P1,P2,P3,P4.
Command syntax
:CALibration:DVM:LEVel <point>
Parameter
:P1|P2|P3|P4
CALibration:DVM [:DATA]
{<numeric value>}
Return actual output current value to calibration point.
Command syntax
:CALibration:DVM [:DATA] <NRf>
Parameter
:<NRf>
Example
:CAL:VOLT 3.0002A
CALibration:SAVe
This command saves the calibration coefficient into nonvolatile register.
Command syntax
:CALibration:SAVe
Parameter
:None
CALibration: INITial
This command resets the current calibration coefficients to thei r default values.
Command syntax: CALibration: INITial
Parameter: None
42
5 Specifications
5.1 Specifications
Parameter
±(
±(
Output Ratings
( 0 °C~40 °C)
Load Regulation
%of output+offset)
Line Regulation
%of output+offset)
Programming resolution
Readback/Meter resolution
9120A 9121A 9124
Voltage 0 ~32V 0 ~20V
Current 0~3A 0~5A
LVP*
Voltage
0 ~33V 0 ~21V
0~72V
0~1.2A
0 ~73V
<0.01
%+2mV <0.01%+2mV
Current
<0.05
%+1mA <0.05%+0.3mA
Voltage
Current
<0.01
%+1mV <0.01%+1mV
<0.05
%+0.1mA
≤0.05
%+0.05mA
Voltage 0.1mV
Current 0.1mA
Voltage 0.1mV 0.1mV
Current 0.01mA 0.05mA
Voltage 0.5mV
0.1mV
0.05mA
0.5mV
0.01mA
2mV
Front Panel setting resolution
Current 0.1mA 0.1mA
Programming accuracy,
12 months (25 °C ± 5 °C)
±(
%of output+offset)
Voltage <0.03%+3mV
Current <0.05%+2mA
≤0.03
≤0.05
%+6mV
%+1mA
Readback / Meter accuracy
12months (25 °C ± 5 °C)
±(
%of output+offset)
Ripple & Noise
(20Hz ~20MHz)
Voltage <0.02%+3mV
≤0.02
%+5mV
Current <0.05%+2mA
≤0.05
%+1mA
Voltage
Current
≤ 4mV p-p
≤ 3mV p-p
≤ 3mArms ≤ 3mArms
≤ 5mV p-p
≤ 3mArms
Voltage <0.02%+3mV
≤0.02%+5mV
Temperature coefficient,
(0 °C~40 °C)
±(
%of output+offset)
Readback temperature, coefficient,
±(
%of output+offset)
Current
<0.05
%+2mA <0.05%+0.5mA
Voltage <0.02%+3mV
Current
<0.05
%+2mA
≤0.02%+5mV
≤0.05
%+0.5mA
DVM Accuracy
DVM Resolution
Voltage
Voltage
0~12V range: 0.02%+2mV
0~40V range: 0.02%+3mV
0~12V range: 0.1mV
0~40V range: 1mV
Ohm meter
Weight mΩ net
Accuracy: 0.1%, for Voltage and Current >= 10% of FS
(full scale)
Accuracy: 0.3% for Voltage and Current >= 3% of FS
19.8 lbs, (9 kg)
Dimensions WxHxD
214.5mm × 88.2mm × 354.6mm
8.45(W) x 3.47(H) x 13.9(D) inch
*) LVP: Limit Voltage Protection. Limits the voltage than can be set either via the front panel or remote control command.
NOTE: Specifications and information are subject to change without notice. Please visit www.bkprecision.com
for the most current product information.
43
5.2 Supplemental Characteristics
State Storage Memory
50 user-configurable stored states
Recommended Calibration Interval
1 year
AC Input Ratings (selectable via switch on the rear panel)
220AV±10%
,47~63Hz
110AV±10%
,47~63Hz
Maximum output power
9120A 9121A 9124
96W 100W 86W
Cooling
Fan cooled
Operating Temperature
0 to 40 °C
Storage Temperature
-20 to 70 °C.
Environmental Conditions
Designed for indoor use in an installation category II, pollution degree 2 environments. Designed to operate at maximum relative humidity of 95% and at altitudes of up to 2000 meters.
Dimensions:
214.5mm (W) x 88.2 mm (H) x 354.6mm (D) all units in mm
44
Visit us at www.TestEquipmentDepot.com
Advertisement
Key features
- Very high accuracy & resolution
- Low ripple & low noise
- Fast settling time
- 5 1/2 digit DVM & mΩ meter
- Convenient data entry
- Over Temp. protection
- Compact footprint
- Bright & easy to read display
- Excellent temp. stability
- SCPI compatible command set