Smoke Meter Manual
SMOKEMETER
Equipment Manual
TES1529J
Jan 2013
Mmmsauhbyx932 7Mar
Mar 2008
Smokemeter Equipment
TES1529 Iss J
Page 2
IMPORTANT
Every reasonable effort has been made to ensure that information within these Operating
Instructions is correct at the time of release, Crypton Ltd. can not accept responsibility for any
errors that may occur.
The information in these Operating Instructions is subject to change without notice, and does
not represent a commitment on the part of Crypton Ltd.
Service & Warranty
The reliability of this equipment is fully supported by our service agents, with repair workshops
and field service engineers to provide a full range of After-Sales Care, including installation,
contract maintenance, factory overhaul and emergency repairs on site. Please refer to the page
at the end of this manual for full details.
Note:
Your attention is drawn to our Terms & Conditions of Sale, particularly paragraph 2. If a service
engineer is called out under service warranty where, upon inspection and test the equipment is
found to be in full working order and no fault found, the user is liable to be charged the cost
incurred for this call out. Before calling out an engineer, ensure your equipment is faulty by
checking its operation, particularly mains supply and fault codes/self test if applicable.
Software Licence
Further to our general Terms & Conditions of Sale, the software described in these Operating
Instructions is furnished under the following Software Licence Agreement:
1. Licence: You may use the program on the machine provided. You may not copy, duplicate or
transmit the program in any form, in whole or in part, without the express prior written
permission of Crypton Ltd.
2. Term: The licence is effective until terminated. You may terminate the licence by destroying
the program together with all copies, or returning the program to Crypton Ltd.
3. Crypton Ltd. does not warrant that the program will meet your requirements or that its
operation will be uninterrupted or error free.
4. In no event will Crypton Ltd. be liable for any incidental or consequential damages, including
lost profits, business operation and the like, arising out of the use or inability to use this
program.
You acknowledge that you have read this Agreement, understand it, and agree to be bound by
its provisions.
WARNING:
Do not attempt to operate this equipment unless you have read and
understood these instructions.
WINDOWS® is a registered trademark of Microsoft® Corporation
Smokemeter Equipment
TES1529 Iss J
Page 3
CONTENTS
Page No.
HEALTH AND SAFETY
5
SMOKEHEAD
6
INSTALLATION
6
OPERATION
9
BATTERY CHARGING (Wireless only)
10
MAINTENANCE
12
OIL TEMPERATURE PROBE
16
ADAPTERS
16
CONNECTION
16
OPERATION
16
RPM TACHOMETERS
17
ACCOUSTIC – CDSS1
18
BATTERY – CDSS6/DX211-12
20
CONSUMABLES
21
TECHNICAL SPECIFICATIONS
23
AFTER SALES AND SUPPORT
25
Smokemeter Equipment
TES1529 Iss J
Page 4
HEALTH AND SAFETY
1.
The equipment is not weatherproof and should not be used outside in rain or snow.
2.
The internal circuits and components of the analyser should not be tampered with. No
internal parts are operator serviceable.
3.
Operators should exercise due caution with regard to the engine fan and associated
belts.
4.
Always provide ventilation by using an exhaust gas extraction system or by having an
adequate supply of fresh air.
Vehicle emissions are dangerous. In particular, Carbon Monoxide
(CO) from a vehicle exhaust is a highly poisonous gas. If breathed
in, it is absorbed by the blood and will result in greatly slowed
reactions and if absorbed in sufficient quantity can be fatal.
Published figures show that a concentration of 0.3% CO in the air
can be fatal if inhaled for 30 minutes.
5.
Before starting an engine, ensure that the gear selector is in neutral or park.
6.
Care should be taken with regard to scalding from the cooling system, burns from the
exhaust system and electric shocks from the ignition HT system.
7.
When disconnecting the fuel system of a hot engine beware of the fire hazard caused by
fuel spilling onto manifolds, ignition distributor, etc.
8.
Position the mains cable and sample pipe in such a manner that they do not present a
hazard to anyone.
9.
The earth lead of the mains supply cable MUST ALWAYS be connected to a good earth
point.
10.
Operators should not allow themselves to come into contact with water or other similar
conductors when using the equipment.
11.
Equipment should be sited in an area designated and clearly marked:
FOR USE BY AUTHORISED PERSONNEL ONLY
Smokemeter Equipment
TES1529 Iss J
Page 5
INSTALLING EQUIPMENT
A number of optional devices may be fitted for use with the Smoke Test. Please refer to the
operating instructions for details of usage.
DX260-1xx SMOKEHEAD (Cabled)
The DX260-1 Smokehead is plugged in to a connector positioned on its cradle base. The base is
then connected to the host equipment. The appropriate Sampling Pipe is then connected to the
Smokehead. A separate base unit (DX260-1IFP) is available.
The Smokehead should be returned
to base when not being used.
DX260-1 shown on the DX260-1IFP base.
DX260-1xxBT SMOKEHEAD (Wireless)
The DX260-1BT wireless Smokehead uses wireless technology to communicate directly with the
host computer. The Smokehead has rechargeable batteries built in to enable approximately 1.5
hours of continuous use between charges. A charger Base DX260-1CHP is available.
The Smokehead should be returned to its charger base between tests to keep the batteries
charged. The Smokehead will go into hibernation mode if it is not being communicated to by the
computer whilst it is off the charger base for longer than approximately 6 minutes.
DX260-1CHP Charger base
DX260-1xxCZ (Cabled) & DX260-1xxBZ (Wireless)
SMOKEHEADS For use on Zero Emissions Boxes.
Use correct Sampling Pipe
SP12 for RAM-1 (Cars)
SP13 for RAM-2 (HGV)
AS09188 Interface Box
Mains power input
Smokehead input
Signal output to PC
(not used on wireless systems)
Special versions of the DX260-1 Smokehead have been designed for use on a Zero Emissions
Box (ZEB). These Smokeheads have adaptor brackets and a different sample pipe to mount on
the ZEB. The Smokehead is plugged in to a specific interface box (AS09188) that is also
mounted on the ZEB. For a cabled system the Interface box is then connected to the host
equipment. For a wireless system the Interface acts only as a power supply for the smokehead
and should not be connected to the host PC.
Smokemeter Equipment
TES1529 Iss J
Page 6
SAMPLING PIPES
Connect the appropriate sampling pipe to the sampling head.
Sampling pipe DX210-4A
The sampling pipe DX210-4A can be used for all vehicles, but for convenience of use alternative
pipes are available. (In some countries a larger probe must be used for tailpipe diameters of
greater than 70mm i.e. trucks – use probe DX210-4A-L)
DX210-4A
DX.210-4A-EX is a 300mm longer version of the standard supplied probe and is available to be
purchased as an optional extra.
Sampling pipe XP12 & SP13
Sampling pipe SP12 and SP13 are used on a Zero Emissions Box. SP12 is used for testing cars
and SP13 for trucks.
SP12 (Cars)
SP13 (Trucks, HGV)
Sampling pipe DX210-6
Sampling pipe DX210-6 is used for vertical exhausts. Hang the sampling pipe and the sampling
head as shown in the illustration. For very high exhausts, an extension pole 1.5 m or 2 m length
may be fitted to the sampling head using an extension kit DX260-160L (1.5m) or DX260-160H
(2.0m) as illustrated. (In some countries a larger probe must be used for tailpipe diameters of
greater than 70mm i.e. trucks– use probe DX210-6-L)
DX210-6
Smokemeter Equipment
TES1529 Iss J
Page 7
Sampling pipe DX210-5C
The sampling pipe DX210-5C has remote control and is used to access mid-chassis exhausts on
large commercial vehicles. To use, hold the sampling pipe and the sampling head as shown in
the illustration with the arm rest under the forearm. The head of the pipe has a pivoting action
so that the 'aim' of the pipe can be changed by moving one arm forwards or backwards relative
to the other arm.
Use the hand lever to operate the jaws and clamp the pipe in position, then allow the remote
control and the sampling head to rest on the floor.
NOTE:
Before fitment of a sampling pipe, inspect its 'O' ring for damage and renew if
necessary (order part number DX210-5/60).
Periodically apply a smear of grease to the 'O' ring to assist fitment of the
sampling pipe to the sampling head.
CAUTION:
The results of free acceleration testing will be affected by changes in
sampling pipe diameter and/or length. Only genuine pipes must be used
and their design must not be altered in any way.
The smokehead body must always be positioned in clean air and away
from exhaust fumes when in use.
WARNING:
Sampling pipes can get very hot; take care when disconnecting from the
sampling head.
USING THE SMOKEHEAD OUTSIDE
If it is necessary or desirable to test a vehicle outside the workshop, the following points must
be borne in mind.
The sampling head is weather resistant and may be used in temperatures down to a minimum
of -15ºC. But in the case of severe weather and heavy rain – DO NOT USE OUTSIDE!
WARNING:
!
The power supply and PC are not weather protected and for the safety of both the
operator and the equipment, these parts MUST be kept under cover in inclement
weather conditions.
Smokemeter Equipment
TES1529 Iss J
Page 8
OPERATION
WARMING-UP
Switch on power to the base unit. The ‘Power On’ light will illuminate.
There is a warm-up time of 5 to 15 minutes dependent on the outside temperature. During this
time the Smokemeter will not function.
NOTE:
For details of software controlled procedures, refer to the appropriate software
manual.
WIRELESS SMOKEHEAD OPERATION (DX260-1BT)
For wireless systems the operation of the equipment is identical to the cabled systems but a
wireless Smokehead (DX260-1BT) is utilised.
Blue flashing LED indicates
battery charging operation
When the equipment is first
switched on the smokehead should
be firmly positioned onto its
charging base and the unit should
be allowed to charge for a minimum
of five to six hours (or preferably
left overnight) to fully charge the
batteries.
The equipment screen will prompt
the user to return the smokehead to
its charging base between tests.
This must be obeyed in order to
proceed further. Failure to return
the smokehead to its base will
result in a depletion of the battery
charge. The test will not continue
until the smokehead has been
returned.
Ensure the Smokehead
is firmly positioned on
its charger base
Smokemeter Equipment
TES1529 Iss J
Page 9
BATTERY CHARGING – Care, Modes & Endurance:
The DX260-1BT wireless Smokehead contains a Multi-Cell Rechargeable Battery designed to
give a maximum of 1.5 hours endurance from a full charge.
The Smokehead must always be replaced on the charging cradle in between tests.
To recharge, the Smokehead must be returned to and correctly located on its charging cradle.
This will maximise the battery endurance and is also necessary for test results to be printed.
On the handle of the Smokehead is a highly visible Blue Light emitting Diode (LED). This LED
gives indication of the status of the Smokehead whether or not it is on its charging Cradle.
The nominal endurance of the internal Battery pack is 1.5 hours continuous use from a full
charge and cold start. This will vary according to ambient temperature. For optimum use and
battery life, the smokehead should be removed from the charging cradle for the minimum
amount of time and returned to the charger immediately after testing or as prompted by the
smoke test Software.
The rate of charging is dependant on a combination of Battery Voltage and Battery
Temperature. If the Smokehead has seen reasonable use the charger will switch to ‘Fast
Charge’, taking approximately 1.5 hours to fully recharge. Once full terminal Voltage has been
reached, the battery charging rate is reduced to ‘Trickle Charge’. When charging for a protracted
period, for example overnight, a trickle charge of 15 hours will completely recharge the internal
battery. A blue LED on the Handle indicates the Charging Status.
It is strongly advised that the Charger Cradle be left permanently switched ON with the
Smokehead in situ overnight to ensure full operational usage at beginning of the working day.
It is also strongly recommended that the batteries be completely discharged on a
monthly basis by setting the program to Live Readings with the smokehead off the
charger. This could be done at the end of the working day.
Please note that rechargeable batteries have a finite lifespan, therefore the battery pack will
need to be replaced periodically if the smokehead is no longer sustaining sufficient charge for
correct operation. Please contact a Product Support for battery replacement information.
LED Indication
1. Smokehead removed from Charging Cradle:
LED ‘On’ continuously: Smokehead is running on internal Battery.
LED pulsing ‘ON’ for 2 seconds, ‘OFF’ for 0.5 seconds: Smokehead battery has reached a
Low Voltage condition, and has less than 10 minutes life left before Auto-shutdown.
LED ‘OFF’: Low Voltage condition. Smokehead will Auto Shut down. No more testing possible
until Smokehead battery has been recharged.
If the Smokehead has received no Communication from Host Computer for over 6 minutes it will
go into ‘Sleep’ Mode. Return the Smokehead to the Charging Unit to invoke a response.
If the Smokehead is taken outside its maximum operating range of about 100 metres, and
wireless communication is no longer possible, after 6 minutes the Smokehead will go into ‘Sleep’
mode.
To recover from ‘Sleep’ mode, the Smokehead should be put back in its Charging Cradle, and
the LED function observed.
Smokemeter Equipment
TES1529 Iss J
Page 10
2. Smokehead returned to Charging Cradle:
FAST CHARGE (2.5 flashes per second):
The Smokehead is in ‘Fast Charge’ Mode. The battery should fully charge within 1.5 hours.
TOP-UP CHARGE (1 flash per second):
The Fast Charge cycle has completed and the battery is being brought to full charge.
SLOW CHARGE (1 flash every 2 seconds):
The Smokehead battery is in ‘Slow Charge’ mode, the charging rate determined by either
Terminal Voltage or, Battery Cell temperature.
TRICKLE CHARGE (1 flash every 3 seconds):
LED Flashing ON/OFF shows the Smokehead battery is in ‘Trickle-Charge’ mode indicating that
the batteries are fully charged.
Fault Diagnosis:
FAST FLICKER (8 flashes per second or more):
The smokehead has detected poor charging contacts. Clean the contacts on both the cradle and
the smokehead feet and replace on charger.
LED NOT ILLUMINATED:
Return the Smokehead to Charging Cradle. LED should display according to Charging level
required.
LED NOT ILLUMINATED WHEN SMOKEHEAD IS ON CHARGING CRADLE:
Check that the Smokehead is correctly seated on the charging Cradle. The Smokehead has twin
connectors on each insulated foot, and may be connected either way round.
Check the Charging Cradle is connected to the 240 V AC Mains.
Check that the sprung connectors that transmit the Charging Current on the Charging Cradle
are clean and not stuck in a depressed condition.
Check that the Charging Connectors on the insulated mounts on underside of the Smokehead
are clean and free of dirt and obstructions.
To check the charging Voltage at the Charging Cradle: Connect a Digital Multimeter set to
measure up to 30V DC to the two charging Connectors on the charging Cradle, and depress the
Central Charging Switch. A Voltage of 28 V to 30 V DC should register on the Multimeter
Display.
The internal battery pack has a finite lifespan and will need to be replaced periodically. This
should be carried out by a qualified Service Engineer.
There are no user serviceable parts in the Smokehead.
Smokemeter Equipment
TES1529 Iss J
Page 11
MAINTENANCE OF THE SMOKEHEAD
Serial numbers
Serial number locations are as follows:


Smokehead internal - Underneath the lens housing (remove end cap for access).
Smokehead external – on underside of Smokehead.
Checking the Smokehead serial number
CLEANING SAMPLING PIPES
Under normal circumstances the sampling head does not require cleaning. If a sampling pipe
becomes restricted with deposits, the test results can be affected.
To clean a sampling pipe, remove the sample probe from the smokehead. Blow through the
sample pipe only, using compressed air. DO NOT BLOW THROUGH THE SMOKEHEAD.
WARNING:
Wear eye protection and provide a suitable container to
catch the soot that will be blown out from Sample Pipe.
DO NOT use compressed air on the Smokehead body.
Smokemeter Equipment
TES1529 Iss J
Page 12
LENS CLEANING
Two lenses are fitted in the smokehead, one at either end.
The Smokemeter checks the cleanliness of the lenses at the start of each test, and the PC will
provide a warning when cleaning is required.
Note orientation of lens housing
(arrow up)
Accessing a lens
The cleaning procedure is as follows.
1.
If fitted, remove the sampling pipe.
2.
Unscrew the knurled knob at the sampling pipe end of the sampling head and remove
the end cap.
3.
Carefully pull out the lens housing and wipe the lens with a soft cloth (if required, a little
methylated spirit will assist cleaning).
4.
Refit the lens housing with the arrow pointing upwards (a little silicon grease on the 'O'
ring will assist fitment – but do not get on lens!).
5.
Refit the end cap and tighten the knurled knob.
NOTE: Ensure the end cap locates fully against the body of the sampling head before
tightening the knob. If not correctly located, the lens housing is probably not
fully home.
6.
Clean the lens at the other end of the smokehead in the same way.
NOTE: The end cap for the sampling pipe must be fitted at the end away from the
sampling head cable.
Smokemeter Equipment
TES1529 Iss J
Page 13
SMOKEHEAD CABLE AND FAN RENEWAL
Should the cable become damaged and require renewal, the procedure is as follows.
1.
Unscrew the knurled knob (1) and remove the end cap.
2.
Remove the four countersunk screws (2) which secure the internal fan and grille to the
smokehead end plate.
3.
Remove the four panhead screws (3) securing the end plate and carefully ease the plate
clear of the housing.
4.
Disconnect the 4-pin multiplug (4a). If carrying out a fan replacement disconnect the 2pin plug (4b) and replace the fan. Carefully fit the fan and fan grille to the end plate (4
countersunk screws).
5.
Remove the 4 wires from the 4-way plug. Undo the metal cable clamp from the end
plate. Loosen the cable gland strain relief and remove the cable from the end plate.
6.
Re-fit the new cable through the cable gland strain relief and secure to the end plate with
the metal clamp (Important: secure the clamp on the earth braid of the cable). Re-fit the
4-way plug. Ensure the cable gland strain relief is tightened.
7.
Connect the 4-pin multiplug (4a) and the 2-pin plug (4b).
8.
Locate the end plate to the smokehead (cable on left hand side) and loosely secure with
the four panhead screws.
9.
Check that the lens housing is fully in, then fit the end cap and secure with the knurled
knob. Fully tighten the knurled knob then slacken half a turn.
10.
Push the end plate upwards as far as it will go, and then fully tighten the 4 panhead
screws.
11.
Fully tighten the knurled knob on the end cap.
Smokemeter Equipment
TES1529 Iss J
Page 14
SMOKEHEAD BATTERY REPLACEMENT (DX260-1BT ONLY)
Should the battery pack require renewal, the procedure is as follows.
1.
Unscrew the knurled knob (1) and open up both end caps.
2.
Remove the four panhead screws (2) & (3) securing the end plates and carefully ease the
plates clear of the housing.
3.
At the fan end, disconnect the 2-pin plug (4a) and also the 2-pin plug (4b) connected to
the fan.
4.
At the probe end, disconnect the 3-pin plug (5).
5.
Depending on which version - loosen the two screws (6a) or the plastic spacer (6b)
securing the battery pack and remove old battery pack, do not remove by pulling the
cables. Ensure the connectors do not foul any components during removal/replacement
of the board.
6.
Replace with new battery pack (note orientation). Discard the two securing screws
(earlier versions) or the plastic spacer (later versions)– no longer required.
7.
Re-connect the 2-pin plug (4a), the 2-pin plug (4b). and the 3-pin plug (5).
8.
Relocate the end plates to the smokehead and loosely secure with 4 panhead screws (7).
9.
Check that the lens housing is fully in, then fit the end caps and secure with the knurled
knob. Fully tighten the knurled knob then slacken half a turn (8).
10.
Push the end plate upwards as far as it will go, and then fully tighten the 4 panhead
screws.
11.
Fully tighten the knurled knob on the end caps.
NOTE: Ensure the battery is disposed of in accordance with current rules and regulations
applicable and enforced in your area. When the equipment is first switched on the Smokehead
should be firmly positioned onto its charging base and the unit should be allowed to charge for
a minimum of five to six hours (or preferably left overnight) to fully charge the batteries.
Smokemeter Equipment
TES1529 Iss J
Page 15
OIL TEMPERATURE PROBES
ADAPTERS
The DX211-11 and DX211-11BT Temperature Probe Adapters are used in conjunction with the
DX210-25A or DX210-26 oil probes for engine temperature measurements. The DX211-11 is
hard-wired to the smokemeter; the DX211-11BT communicates via wireless technology.
DX211-11
DX211-11BT
Cabled Temperature Probe Adapter
Wireless Temperature Probe Adapter
Connection to the Smokemeter
The DX211-11 utilises an RS232 communications cable for connection to the Smokemeter. The
unit connects to the smokemeter via a serial port. When not in use the flexible probe should be
stored in a protective holster
The DX211-11BT is a wireless device that will communicate with a wireless enabled
Smokemeter. A charging holster is supplied which should be used to store and charge the unit
when not in use. The DX211-11BT batteries will be charging at all times whilst stored in the
holster, indicated by the status indication light on the probe adapter glowing red.
Charging option for the DX.211-11BT
In addition to the trolley mounted storage/charger base, the DX.211-11CH
product option offers remote wall mounted storage/charging (standard UK
13A power outlet is required)
Temperature probes
There are two flexible temperature probes, either of which can be plugged into either adapter.
The DX210-26 is a standard 0.75M probe, the DX210-25A is a 2.5M long-reach probe.
Connect the flexible oil probe to the socket of the DX211-11 or DX211-11BT. This connector
includes a snap-lock feature that ensures a reliable connection between the two devices.
Adjusting oil temperature probe length
Before testing, the probe must be adjusted to the correct length as follows:
Lay it alongside the vehicle dipstick (see illustration) and adjust the position of the depth stop
until the effective length of the probe is approximately 10mm shorter than the dipstick.
DX210-26 / DX210-25A
WARNING: ENSURE THAT THE OIL TEMPERATURE PROBE IS SET TO THE CORRECT
DIPSTICK LENGTH AND THAT IT IS CLEAN FROM DEBRIS BEFORE INSERTING INTO
THE ENGINE. FAILURE TO DO SO MAY RESULT IN SERIOUS DAMAGE TO THE PROBE
AND/OR ENGINE.
Smokemeter Equipment
TES1529 Iss J
Page 16
ENGINE TACHOMETERS
RPM TACHO - Acoustic CDSS1
The CDSS1 RPM Adapter is connected to the Tacho input port of the AS09185 Tacho Adapter.
The CDSS1 can derive its power via AS09133 data lead.
CDSS1
AS09133
AS09185
CDSS1 Tacho
input
CDSS1 12v
Adapter input
Data output to PC
RPM TACHO - Battery CDSS6 (DX211-12)
The CDSS6 RPM Tacho provides a signal via an RS232 comms lead DX230-261. The comms lead
is connected directly to the host equipment. The CDSS6 derives its power from the vehicle
battery. There is no oil temperature probe facility with this device.
DX230-261
(CDSS6 only)
CDSS6/CDSS6BT
(DX211-12/DX211-12BT)
RPM TACHO – Battery Wireless
CDSS6BT (DX211-12BT)
The CDSS6BT RPM Tacho is identical to the CDSS6 except that it provides a signal to the host
PC via a wireless connection. It will only work with wireless enabled equipment and does not
use the DX230-261 comms lead. The CDSS6BT derives its power from the vehicle battery. There
is no oil temperature probe facility with this device.
Smokemeter Equipment
TES1529 Iss J
Page 17
USING THE ENGINE TACHOMETERS
Ensure the vehicle engine is switched off before connecting engine speed adapters to the
vehicle. The following types of engine speed adapter are available:
ACOUSTIC TACHOMETER CDSS1
Acoustic sensor type for Petrol and Diesel engines
This adapter measures the noise and vibration of an engine, and converts this information to
engine speed.
Note: The CDSS1 connects to the host equipment via a signal interface cable (AS09133). This
cable can also carry the power supply for the CDSS1 from the AS09185 Tacho Adapter (the
CDSS1 12v external power supply can be connected to either the AS09185 Tacho Adapter or
CDSS1).
The engine speed signals produced are equivalent to a conventional ignition system (i.e. one
pulse per engine cycle). When defining the ignition system on the host analyser, It is therefore
imperative that a conventional system is selected, irrespective of that fitted to the vehicle.
Selecting DIS (Distributor-less Ignition System) for example will result in the readings being
divided by 2.
Connections
Status LEDs.
Red/Green
Multicolour Status LED.
Red/yellow/green.
Power and signal
connection
Acoustic/Vibration Sensor
Input
Serial Output
(for Service and test
use only)
Connection Bar for
an Inductive Clamp
Microphone for detecting
acoustic noise
Magnetic base with Accelerometer for
detecting vibration noise
External Power Supply
+12V dc (not used)
Smokemeter Equipment
TES1529 Iss J
Page 18
Suggested mounting Positions for the sensor (CDSS1)
The sensor has a powerful magnetic base and should therefore be attached to an iron or steel
part of the engine, such as a bolt or lifting bracket. Always ensure that the Sensor is attached to
the Engine itself and not to a part of the vehicle where the vibrations may be cushioned by the
engine mountings e.g. wing or bulkhead.
Examples of good mounting positions are: Lifting brackets, Cylinder head bolts (where they are
accessible), Oil drain plug etc.
CDSS1 Operation
Prior to starting the vehicle attach the magnetic sensor to a part of the engine. The LED on the
sensor should be RED
1. Switch engine OFF.
Red
Attach Sensor to a suitable point as described in ‘Suggested mounting positions’.
The Status LED should be red.
2. Start Engine and allow to idle.
Yellow
After approx 3 – 10 seconds the status LED will change to Yellow.
3. Rev Engine to high speed (2000 – 6000 rpm) and hold.
Note: In the case of a Diesel this may be 70 – 90% of Governor.
Green
Hold Engine speed until yellow status LED changes to Green. This should take a maximum of 5
seconds. If the status LED does not change, try another mounting position.
Smokemeter Equipment
TES1529 Iss J
Page 19
BATTERY TACHO CDSS6, CDSS6BT (DX211-12, DX211-12BT)
Battery sensing type for Petrol and Diesel Engines
This adapter measures the frequency modulation ripple of the alternator across the battery
terminals, and converts this information into engine speed signal.CDSS6BT and DX211-12BT
are wireless variants
Connections
STATUS LED
ON:
Power/Initialising
FLASH: Measurement in progress
SIGNAL OUTPUT
via comms lead DX230-261
(Not CDSS6BT)
BATTERY CLIPS
Red to battery +VE
Black to battery -VE
CDSS6/CDSS6BT RPM Operation
Ensure the CDSS6 is connected to the host PC via the DX230-261 Comms lead (Not required on
CDSS6BT).
Before starting the vehicle, attach the crocodile clips on CDSS6 power/sensing cable to battery
terminals, ensuring that the polarity is correct.
Start engine, switch on vehicle lights and allow to idle.
The unit will search for a steady speed each time it is powered up, if the engine speed falls or
the engine type is changed. This is the initialisation period and may last up to 17 seconds.
During this time, the status LED will flash slowly and the rpm will read zero.
Once the speed has been detected, readings will appear on the host analyser. The CDSS6
updates 10 readings per second.
Note:
Please note that on a small number of vehicles the readings may be unstable, inaccurate or
unreadable. To minimise the risk of this:Load the battery as much as possible by doing the following:
Switch on the vehicle headlights/Main beam/Fog/Spots,
Switch on the heated rear window,
However - Switch off the cabin fan as this can introduce interference.
Keep the engine speed stable during the initialisation time <17Sec.
If, after initialisation the speed is stable but reading incorrectly, change the cylinder setting on the
‘Tacho Set-up’ screen until the correct value appears.
Smokemeter Equipment
TES1529 Iss J
Page 20
CONSUMABLES
DX210-4A
Sampling pipe (Standard)
DX210-4A-EX Sampling pipe (Std + 300mm, optional extra)
DX210-4A-L
Sampling pipe (Large)
For large aperture exhausts (>70mm)
Battery Tacho Power/Sensing cable
CABL3016
CDSS6 to vehicle battery
Tacho Comms Cable
AS09133
DX230-261
5m Standard (for CDSS1)
2m Standard (for CDSS6)
Battery Tacho Power/Sensing Cable
CABL3023
CDSS6 to vehicle accessory socket
Sampling Pipe (Special)
(For use with Zero Emissions Box only)
SP12
SP13
– For RAM-1 (Cars)
– For RAM-2 (HGV)
Cable – Smokehead Cable Assembly
DAS00497
DAS00556
DAS00559
10m Standard
20m (Optional)
0.6m (Use with AS09188 Interface Module)
Cable – Serial comms cable from PC to VOSA Smartcard
reader (UK only).
AS08804 (Note: May already be fitted if host PC has EGA)
DX260-1BT/01 Battery Pack
(For use with DX260-1BT Head only)
DX230-130/16 Fan Kit
(Replacement fan for DX260-1 & DX260-1BT)
Oil Temperature Probes
DX210-25A
DX210-26
2.5m (Use with DX211-11 or DX211-11BT)
0.75m (Use with DX211-11 or DX211-11BT)
Smokemeter Equipment
TES1529 Iss J
Page 21
Temperature Probe Adapters (use with DX210-25A
and DX210-26 probes)
DX211-11
– 5m cabled Adapter
DX211-11BT – Wireless Probe Adapter
Smokemeter Equipment
TES1529 Iss J
Page 22
TECHNICAL SPECIFICATION
A.
SMOKE SAMPLING HEAD
Sampling type
Partial flow.
Light source
Green LED, wave length 560 mm.
Light intensity
8-step autorange for optimum light level
Effective optical path length
250 mm.
Physical response time
Less than 0.4 seconds (varies with gas velocity).
Range
0 - 100% opacity, 0 - 10.00 m-1.
Accuracy
1.0% full scale (static).
Measuring cell temperature
70ºC min.
Warm-up time
10 minutes approx. at - 15ºC ambient
Operating conditions
-15º to +40ºC. 10-90% relative humidity (non condensing).
Wireless Range (DX260-1BT only) Class 1 wireless Module provides up to 100 metres range
(maximum range outdoors – line of sight).
Smoke temperature
Checked at the inlet to the measuring cell.
Smoke pressure
Controlled at ambient ± 3.75 mbar.
Sampling pipes
Flexible stainless steel, internal diameter 10 mm, lengths
875 mm (standard) 1175 (std EX option)
1,255 mm (under-chassis and vertical exhausts).
Specials available for use with Zero Emissions Boxes.
Lens cleaning
Wipe with soft cloth every 250 vehicle tests approx.
Calibration
Automatic electronic calibration (patented).
Manual calibration (using neutral density filters).
Security
Sampling head serial number and software version held in
permanent memory.
Fault diagnosis
Condition monitoring for - Dirty lenses, calibration failure,
incorrect measuring cell temp.- Excessive smoke temp. etc.
Power
24V DC, 60W (DX260-1 only) - via Interface Box
Internal battery (DX260-1BT only) - requires 30V 240W
Charging Base
Housing
Aluminium extrusion with aluminium die cast end caps.
Protection
IP22.
Dimensions
450 x 200 x 80 mm.
Weight
3.2 Kg. (DX260-1)
3.9 Kg. (DX260-1BT)
Smokemeter Equipment
B.
TES1529 Iss J
Page 23
CONNECTING CABLE - SAMPLING HEAD TO POWER SUPPLY (DX260-1 only)
Construction
4-core screened.
Sheath
Polyurethane (for low temperature operation).
Length
10m (standard), 20m max. (optional).
C.
POWER SUPPLY - Interface Box (for DX260-1)/Charger Base (for DX260-1BT)
Power
100–250V AC, 50-60Hz, 3.2A (Interface Box - DX260-1)
100-260v AC, 50-60Hz, 8.0A . (Charger base - DX260-1BT)
Operating conditions
0 to + 40ºC. 10-90% relative humidity (non condensing).
Input (Interface Box - DX260-1)
Smoke sampling head.
Output (Interface Box - DX260-1) RS232, 9-way 'D' (opacity, smoke temp.) 9,600 BAUD
Security
Interface unit serial number and software version held in
permanent memory.
Protection
IP20.
D.
OIL TEMPERATURE PROBE ADAPTER
Measurement range
7°C – 116°C
Accuracy
5°C below 45°C
1°C above 45°C
Response
<1 Second
Handle Dimensions
22mm Ø x 170mm Length (DX211-11 - cabled)
35mm Ø x 130mm Length (DX211-11BT - wireless)
Cable Length (cabled only)
5 Metres
Wireless Range (Wireless only)
Class 1 wireless Module provides up to 100 metres range
(maximum range outdoors – line of sight).
Smokemeter Equipment
E.
TES1529 Iss J
Page 24
TACHOMETERS
Battery Tachometers
Measurement range
400-9999 rpm
Accuracy
± 20 rpm below 2000 rpm
± 2% above 2000 rpm
Response
< 1 second
Initialisation
≤ 17 seconds
Power requirements
9-15V dc, < 0.5A (reverse protected)
Operating temperature
0 to +45°C
Humidity
< 90% non-condensing
Cable
5 metres comms lead.
(Cabled version only)
Wireless range (Wireless only)
Class 1 wireless Module provides up to 100 metres range
(max range – line of sight).
Dimensions
230mm x 100mm x 40mm
Weight
790g
Vibration Tachometer
Measurement range
400-6000 rpm
Accuracy
± 10 rpm within range 400-6000 rpm
Response
< 1 second
Initialisation
≤ 15 seconds
Power requirements
9-15V dc, < 0.35A
Operating temperature
0 to +50°C
Humidity
< 90% non-condensing
Cable
5 metres to sensor
Dimensions
227mm x 184mm x 48mm
Weight
1.18Kg
Smokemeter Equipment
TES1529 Iss J
Page 25
AFTER SALES SERVICE
Apart from the routine maintenance and adjustments stipulated in this manual the equipment must not be
tampered with in any way. All further servicing must be carried out only by an engineer from our Authorised
Agents. Failure to observe these conditions will invalidate the Guarantee.
UK Customers
If you require a Service Engineer to attend ON SITE within the UK, for any of the following reasons:




An equipment fault,
For machine calibration,
You need spare parts,
Equipment covered by this manual requires returning for factory overhaul.
Please contact our Product Support Helpline at the following number:
Tel:
0844 665 7610
Fax:
0844 665 7604
e-mail: [email protected]
Crypton can provide information and contracts covering the following: Car Data; Fault Code Information;
Diagnostic Information; Software Support Contracts; Software Updates & Accessories.
Overseas Customers
Service and spare parts cover outside the UK is provided by the agent from whom your equipment was
purchased.
Disposal of equipment

Do not dispose of this equipment as miscellaneous solid municipal waste but
arrange to have it collected separately.

The re-use or correct recycling of electronic equipment (EEE) is important in
order to protect the environment and the wellbeing of humans.

In accordance with European Directive WEEE 2002/96/EC, special collection
points are available for the delivery of waste electrical and electronic equipment.

The public administration and producers of electrical and electronic equipment are involved in
facilitating the processes of the re-use and recovery of waste electrical and electronic equipment
through the organisation of collection activities and the use of appropriate planning arrangements.

Unauthorised disposal of waste electrical and electronic equipment is punishable by law with
appropriate penalties.
Disposal of batteries

Batteries must be recycled or disposed of properly.

Do not throw batteries away as part of normal refuse disposal.

Do not throw batteries into open flame.