Leica confocal microscope Manual
The Leica Confocal Microscope is a powerful tool for imaging biological samples. It uses a laser to excite fluorescent molecules in the sample, and a pinhole to reject out-of-focus light. This results in high-resolution images with excellent optical sectioning. The Leica Confocal Microscope is used for a wide variety of applications, including cell biology, developmental biology, neuroscience, and materials science.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
Leica Confocal Manual
Neuroscience Imaging Core
Rightmire Hall
Ohio State University
Director: Tony Brown
060 Rightmire Hall
292-1205
[email protected]
Facility Manager: Paula Monsma
062 Rightmire Hall
292-3025
[email protected]
This manual prepared by Paula Monsma.
Leica Confocal Manual .................................................... 1
Getting Started ................................................................. 4
Start the Leica Confocal Software ................................... 4
Left Monitor ...................................................................... 5
Information Window ......................................................... 6
Beam Path Setting Window ............................................. 8
Format Button ................................................................ 10
Mode Button .................................................................. 10
Obj button ...................................................................... 11
MicCtrl Button ................................................................ 11
Pinh Button .................................................................... 12
Speed Button ................................................................. 12
Right Monitor ................................................................. 13
The Viewer Window ................................................... 13
Viewer Toolbar ............................................................... 14
Collecting an image ....................................................... 15
The Beam Path Settings Window ............................... 15
Take the first image .................................................... 16
Create an overlay of the two channels ....................... 17
Optimize the gain and offset of the two channels ....... 17
Averaging ....................................................................... 19
Averaging taking a single image ................................ 19
Crosstalk ........................................................................ 20
Check for Crosstalk .................................................... 20
Adjust beam path settings for crosstalk ...................... 22
Change the detection slot ........................................... 23
Set up a sequential scan ................................................ 25
Optimize the settings for each channel ...................... 25
Save the sequential setting ........................................ 26
Load the sequential scan settings .............................. 27
Acquire an image using the sequential scan settings . 28
Zoom Functions ............................................................. 30
Zooming on a specific area ........................................ 30
Centered Zoom .......................................................... 31
DIC................................................................................. 32
Take a DIC image ...................................................... 32
2 CMN Imaging Core
1/16/2013
Saving the Data ............................................................. 33
Saving the experiment ................................................ 33
Saving the overlay image ........................................... 35
Experiments ................................................................... 36
Create a new experiment ........................................... 36
Open a saved experiment .......................................... 37
Renaming images ...................................................... 38
Browsing all the images in an experiment .................. 38
Adding a scale bar ......................................................... 39
Taking a Z series ........................................................... 41
Set the Start and End Points ...................................... 41
Set the number of sections ......................................... 42
Collect the series ........................................................ 43
Viewing the Z-series ................................................... 44
Create a 3D Projection and Animation ........................... 45
Select the Projection Type ......................................... 46
Preview the Projection ................................................ 46
Export an Animation ...................................................... 48
Archiving Data - Create a data disk ............................... 49
The Nero Express Window ......................................... 50
Logging off of the computer ........................................... 53
Shutting Down ............................................................... 54
3 CMN Imaging Core
1/16/2013
Getting Started
1.
Turn on the nitrogen tank valve
2. Turn on the mercury bulb
(top)
3. Turn on the microscope
(bottom)
4.Turn on the fans for the lasers
(red switches)
5. Turn on the three lasers (keys)
7. Log in as a user
Start the Leica Confocal Software
Double click on the Leica Confocal Software icon.
4
6. Turn on the computer
CMN Imaging Core
1/16/2013
Left Monitor
Pull down menus and buttons-
Many of the functions here can be done somewhere else in the software. The most important button here is the
“SAVE” button.
Experiments Window –
Images are sent here as files.
Images can be saved, renamed, deleted and exported from this window.
Blank area for opened windows to be placed so that they don’t cover up other information.
Information Window – All of the settings for the current image are available here.
5 CMN Imaging Core
1/16/2013
Information Window
6 CMN Imaging Core
1/16/2013
Menu buttons – Each opens a new set of tools. Generally, these are used in a left to right order.
Z Series controls
Tools for acquiring images.
7 CMN Imaging Core
1/16/2013
Beam Path Setting Window
Laser Line Settings-
Indicates which laser lines are active and at what power.
Window-Saved beam path settings. Save your own settings here.
Transmitted light
PMT – make active to take DIC images.
Dichroic
(Beam Splitter)
8 CMN Imaging Core
1/16/2013
Select the detection slots for each PMT
Seq –Opens Sequential Scan Settings
Save – Saves the current Laser line and detection slot settings.
Close – Closes the Beam Path Settings window
Many different options for formatting the image.
9 CMN Imaging Core
1/16/2013
Format Button
Choose the resolution of the image to be collected. The default is
512 x 512.
Mode Button
Determines the type of scan that will occur. Default is xyz. Modes with “t” are timed.
10 CMN Imaging Core
1/16/2013
Obj button
Displays the objective currently in use. Also, can change the objective in use by clicking on the desired objective.
MicCtrl Button
To easily change the microscope settings between scanning and looking at your sample thru the eyepieces, select the “MicCtrl” button.
Choose “Scan” if you want acquire images with the confocal software.
Choose “Visual” if you want to look through the eyepieces of the microscope. If you get an error message when you try to acquire an image in the software check this se tting to make sure that “Scan” is selected.
11 CMN Imaging Core
1/16/2013
Pinh Button
Displays the current pinhole setting. The default for each objective is
1 Airy Disk. Increasing the pinhole will increase the signal, but will also reduce the confocality of the image.
Speed Button
Displays the current scan speed, which is equivalent to exposure.
Default is 400 Hz. Increasing the speed will decrease photobleaching and brightness of image, as well as the time it takes to acquire the image.
12 CMN Imaging Core
1/16/2013
Right Monitor
The Viewer Window
13
Viewer Toolbar- Buttons that will facilitate taking and displaying an image.
Viewer Window – The acquired images will be displayed here.
CMN Imaging Core
1/16/2013
Viewer Toolbar
Single, Tiled, Gall. and Ovl –Ways of determining the type of view that you would like in the viewer window.
Ch 1 thru Ch 4 –
Selecting/Deselecting will determine which channels are displayed in the viewer window.
Q LUT and Lut – Buttons for optimizing the gain and offset, as well as adjusting the display of the image.
Buttons for acquiring images. Many buttons are duplications of what is under the Acquire menu.
14 CMN Imaging Core
1/16/2013
Collecting an image
Select the MicCtrl button. Set it to scan.
The Beam Path Settings Window
Select the “Beam” button. The Beam Path Settings window will open.
To select excitation and detection settings, double click on one of the options under the Leica directory in the Beam Path Settings window
(blue arrow). The settings for FITC-TRITC are selected here.
15 CMN Imaging Core
1/16/2013
Take the first image
Select “Single Scan” to take one scan of your image.
The first scan on the right monitor. This image will be flipped on a horizontal axis as compared to what is seen through the eyepieces of the microscope.
16 CMN Imaging Core
1/16/2013
Create an overlay of the two channels
Choose the “Ovl” button to create a merged image of the two channels.
Optimize the gain and offset of the two channels
Select the “Q LUT” button. This will create an image where pixels of maximum value are displayed as blue and pixels with a value of zero are displayed as green. Pixels with values in the mid ranges are displayed as reds and whites. Adjusting this will change the way the data is collected by the detectors. This is not simply changing the way the image looks on the screen.
17 CMN Imaging Core
1/16/2013
Select “Continuous” to start a continuous scan. Select each channel, and using the control panel, use the smart gain knob and adjust the gain to show maximum values and use the smart offset to adjust the offset to show minimum values. Select the other channel and adjust the gain and offset for that channel.
This image has been adjusted, so that there are some maximum pixels and some minimum pixels, but nothing is saturated. To change the display back to colored image, simply click on Q LUT twice more.
The gain and offset will need to be checked and adjusted each time a setting is changed.
18 CMN Imaging Core
1/16/2013
Averaging
Averaging taking a single image
Select “Aver” Select single scan. Choose the number of frames you would like to average. Averaging two to four frames is most common.
Select “Single Scan”. An image will be created in the Experiments window.
The single scan is saved as an image in the Experiments window.
The small red check on the file icon indicates that this is the image that is currently being shown on the right monitor.
19 CMN Imaging Core
1/16/2013
Crosstalk
Check for Crosstalk
Reduce the 488 laser line to zero. Take a scan. Check the gain and offset.
Because there is no signal in the FITC channel, we would say that there isn’t crosstalk in this direction.
20 CMN Imaging Core
1/16/2013
Check for crosstalk from the other laser line
Reduce the 543 laser line to zero. Set the 488 laser line to its previous setting. Take a single scan.
Because we see a signal in the TRITC channel, this shows crosstalk.
Since we see signal in this channel it can mean that the 488 laser line is also exciting TRITC (excitation crosstalk) or that the emission curve for FITC is so large and the signal so strong that the detection slot for
TRITC is actually detecting FITC.
21 CMN Imaging Core
1/16/2013
Alexa 546
Spectra
488 FITC 543 TRITC
Looking at the spectra for both channels, you can see that the
488 laser line will excite both fluorophores, FITC more than
Alexa 546, so there is possibility of excitation crosstalk. When we look at the
488 FITC 543 TRITC detection slots (shaded areas) for both channels, we see that the tail of the emission spectra for FITC is clearly detected in
Alexa
488
Spectra the TRITC channel. So, if we excite with just the 488 laser at a power low enough that it doesn’t excite the TRITC channel, we will still see data in the TRITC channel, because the FITC emission curve is being detected in the TRITC channel.
Adjust beam path settings for crosstalk
Reduce the 488 laser line to 7% to see if that helps. The 488 laser line is very powerful and is generally used at a low level. Take a scan, adjust the gain and offset.
22 CMN Imaging Core
1/16/2013
There is still a large amount of crosstalk.
Change the detection slot
Using the slider, change the detection slot for the TRITC channel, so that the tail of the FITC emission is no longer being detected. Take a scan.
23 CMN Imaging Core
1/16/2013
The signal in the TRITC channel is reduced, but there is still some crosstalk.
Because the emission tail of the FITC is so large and there is so much overlap into the TRITC channel, this would be a good situation to use sequential scan.
24 CMN Imaging Core
1/16/2013
Set up a sequential scan
A sequential scan will take a scan with the settings for one channel and then the settings for another channel. It’s a good way to reduce crosstalk between channels.
Optimize the settings for each channel
Here, we are optimizing the settings for the FITC channel. We widened the detection slot and set the laser setting at reasonable level.
Take a scan. Optimize the gain and offset.
25 CMN Imaging Core
1/16/2013
Save the sequential setting
Select “Save” (blue arrow) in the Beam Path Settings window. The
“Name of Parameter Setting” (yellow arrow) dialog will open. Name the setting (here Seq Green Convallaria) and click “OK”. The setting will be saved under “User” (red arrow) in the Beam Path Settings window.
Here we are optimizing the settings for the TRITC channel.
26 CMN Imaging Core
1/16/2013
The resulting image.
Save this setting as for the other channel.
Load the sequential scan settings
Select “Seq” (blue arrow) in the Beam Path Settings window. The
Sequential Scan Settings window will open. Select one of the channel settings that you saved (red arrow and highlighted) and click
“Add” (yellow arrow) in the Sequential Scan Settings window. Then select the settings for the other channel and click “Add”.
27 CMN Imaging Core
1/16/2013
Here the settings for both channels have been added. Now choose a mode.
Acquire an image using the sequential scan settings
With the “Sequential Scan Settings” window open, select “Single
Scan” or “Continuous”. The resulting image will be collected using the sequential scan settings.
28 CMN Imaging Core
1/16/2013
The resulting image. Check and adjust the gain and offset.
29 CMN Imaging Core
1/16/2013
Zoom Functions
Zooming on a specific area
To choose an area to zoom in on, select “Z. In”. Go to your image and draw a box around the desired area. Collect an image using either Single Scan or Continuous scan. This is not simply zooming in on the already acquired image, it is scanning smaller area with the same resolution.
Resulting zoomed image.
30 CMN Imaging Core
1/16/2013
The amount of zoom is indicated in the information box.
Centered Zoom
Select the “Zoom” button. It will tell you what the current zoom is, as well as allow you to choose a specific zoom. This will zoom in on the center of the image. This is handy if you have several areas of interest in one larger image because you can use it to reduce the zoom to 1.00, so that you can see your original image again.
31 CMN Imaging Core
1/16/2013
DIC
Take a DIC image
To take a DIC image, make sure that the the objective prism and the condenser prism are correct for the objective you are using, the transmitted light detector selection knob is set to scan, and that the analyzer is not in the light path. Then check the “Active” box for the transmitted light PMT.
DIC Image Overlay Image
Take an image, adjust the gain and offset.
32 CMN Imaging Core
1/16/2013
Saving the Data
Saving the experiment
The program places all the images that you acquire in temporary memory until you save them. The images that are automatically sent to the experiment are the individual channels. If you have created an overlay of both channels, you will need to specifically save that. If the program crashes, upon restart it will notify you that there are temporary files and it will ask if you want to save them. If you exit out of the program purposely and choose not to save your images, they are not recoverable.
To save the experiment
Select “Save” at the top of the screen.
Shortcut Icon on the desktop
Each lab will have a user folder on the D drive in the “users” folder, as well as a shortcut to that folder on the desktop. Save images only on the D drive. Here, McKinney lab is the user folder. To save, select
“McKinney lab’s Documents”.
33 CMN Imaging Core
1/16/2013
Rename the experiment (red arrow) and click “Save” (blue arrow).
The experiment will be saved as a folder. The folder will contain all of the images acquired in the experiment.
The software creates a file that tells what the names of the files are in the experiment folder. If you wish to change the name of your files and still be able to open those files in the Leica software, you must
rename them from within the Leica software. If you go into
Windows and rename the individual files there, when you try to reopen these images in the Leica software you will not be able to see them. A good way to eliminate this problem is to create a permanent archival copy of your raw data on CD or DVD. When you want to use the images, make a copy of the image and change that.
34 CMN Imaging Core
1/16/2013
Saving the overlay image
In order to save an image that looks like the Viewer window, right click on the images in the right monitor. Choose “Send to”, then
“Experiment”, and then “All(snapshot)”. This will send an image that when you open it in Photoshop, will have the individual channels as well as the overlay in one composite image.
The icon for the snapshot (arrow) in the experiment window.
35 CMN Imaging Core
1/16/2013
Experiments
Create a new experiment
This would be useful if you have several very different samples you want to image during the same session. You can rename the experiment so that it’s clear which set of images came from which experiment without having to rename each image that is taken.
Select “New” to create a new experiment. The new experiment will be created in the Experiment window.
36 CMN Imaging Core
1/16/2013
Open a saved experiment
Se lect “Open”. An open window will open. Browse until you find your experiment folder. Select it and open it. Here “4-4-05
MolProbesSlide” was selected.
When the experiment folder is open, only the .lei file is visible.
Double click it to open. This will open all the images in this folder in the Leica software.
37 CMN Imaging Core
1/16/2013
Renaming images
To rename an image, right click on the name of the file and select rename.
Browsing all the images in an experiment
Choose “Browse”. This will display all the images in the experiment in a thumbnail format.
38 CMN Imaging Core
1/16/2013
Adding a scale bar
Check “Scale” in the Display menu, under Settings.
A scale bar will be added to your images. To format the scale bar to a specific length, choose the “Scale” button (arrow). A Scale Bar dialog will open.
39 CMN Imaging Core
1/16/2013
Type in the length of the scale bar and choose apply.
The resulting scale bar. In order to save the scale bar, you must right click on the image and send it to the experiment as a snapshot.
40 CMN Imaging Core
1/16/2013
Taking a Z series
A Z series is a set of optical slices through the specimen, each at a different Z plane. It provides a more three-dimensional picture of the specimen.
Set the Start and End Points
Select the “Series” (red arrow) button. This will bring up the “Setting
Z/YPosition for Spatial Image Series” dialog. Start a Continuous scan. Using the control panel knob, change the Z position until you are at the top of the specimen. Select the “Begin” button (green arrow) to set this as the beginning of the Z series. Change the Z position in the opposite direction until you are at the bottom of the specimen and choose the “End” button (blue arrow) to set this as the end of the Z series.
41 CMN Imaging Core
1/16/2013
Set the number of sections
Here the beginning and end points have already been chosen. Select the “Sect” button to choose the number of sections or slices that will be taken. The software automatically chooses an “optimized” number of sections. To change the number of sections, select “Others” and the Z-configuration dialog will open.
Select the number of sections. A good rule of thumb is to double the optimized step size.
42 CMN Imaging Core
1/16/2013
Here, the step size has been changed to 0.24. This decreases the number of sections to 194.
Collect the series
Select the “Series” button to start the collection of the Z-series. The
“Setting Z/Y-Position for Spatial Image Series” will show the progress of the Z-series. The green line in the box is the begin point. The red line in the box is the end point. The yellow line in the box is the current position.
43 CMN Imaging Core
1/16/2013
Viewing the Z-series
These are all the images in the Z-series. The select this view, choose
“Gall” for the gallery view, “Single” to only have one image, both “Ch
1” and “Ch 2” since both channels are being captured and “Ovl” to have the overlay image.
The Z series file is named “Series” with a number, in this case
Series027.
44 CMN Imaging Core
1/16/2013
Create a 3D Projection and Animation
Select the “Process” menu. Choose “3D Visualization and
Animation”.
In 3D Visualization and Animation, choose “Projections with
Animation”.
45 CMN Imaging Core
1/16/2013
Select the Projection Type
There are six different choices of how the projection is created. Each of the windows displays a preview of what the projection will look like with that method. In this case, “Average Proj.” was chosen.
Preview the Projection
Choose “Preview” (blue arrow) to create a preview in the Result window. Choose “Apply” (red arrow) to create a file in the
Experiments window that can be saved (red circle).
46 CMN Imaging Core
1/16/2013
A finished 3D Animation.
The projection will also be displayed in the Viewer window.
47 CMN Imaging Core
1/16/2013
Export an Animation
Right click on the 3D Animation file in the Experiments window. Here it is named 3DAnimAver_00. Choose “Export”.
A “Save As” window will open. Name the animation and choose from the “Save as type” menu. Here “Export-Animation-Avi (*.avi) has been chosen. Choose “Save”. The animation will be saved as an
AVI file.
48 CMN Imaging Core
1/16/2013
Archiving Data - Create a data disk
The computer has a CD/DVD writer as well as USB ports. Loading a
CD or DVD into the CD/DVD drive will cause burning software to open automatically. If it does not open automatically, click on Nero
SmartStart icon on the desktop.
Nero SmartStart icon
Choose “Data”. Choose “Make a Data CD” or “Make a Data DVD”, depending on the type of media you are using.
49 CMN Imaging Core
1/16/2013
The Nero Express Window
Open up your user folder. Select the items in the folder that you would like to save to the CD and drag them over to the CD project.
50 CMN Imaging Core
1/16/2013
The blue line indicates how much space the selected items consume.
Select “Next” to go to the next screen.
Select “Burn” to save the selected data to the CD.
51 CMN Imaging Core
1/16/2013
When the burn process is finished, a dialog will open. Click “OK”.
Then this window will open. Choose “Exit” to close the burning program.
52 CMN Imaging Core
1/16/2013
Logging off of the computer
After exiting out of the confocal software, log off of the computer.
Click on the “Start” button.
Choose “Log Off”.
53 CMN Imaging Core
1/16/2013
Shutting Down
7. Turn off the
nitrogen tank
4. Turn off the microscope
(bottom)
5. Turn off the mercury bulb
(top)
3.Turn off the lasers
(keys), but not the red buttons
6. At least 15 minutes
after #3! Turn off the fans for the lasers (red switches)
1. Shut down the
Leica software
2. Log out as a user
Alternative Shutdown Procedure:
1. Turn off the lasers (keys only) and the mercury bulb.
Wait 15 minutes before covering the microscope or turning off the fans for the laser (step 5).
2. Finish working with the software (save your data, burn a disk, etc). Shut down software. Log out as user.
3. Clean up microscope (clean off oil, leave 10x objective in place with the turret down).
4. Turn off the microscope.
5. Turn off the fans for the lasers (at least 15 minutes after
step 1!!).
6. Cover microscope.
7. Turn off the nitrogen tank. Shut off lights in the room.
54 CMN Imaging Core
1/16/2013
Troubleshooting
This section is intended to offer help in solving various problems. If y ou are still having problems with the confocal system, please don’t hesitate to contact Paula.
Hardware Legend Can’t Get All Values
If there is a problem with the way the microscope has initialized (or the microscope wasn’t on before the software was started), you will get this message – “Hardware legend can’t get all values.” (You will see this error message in other situations as well.) Click “OK”. The next screen will look like this.
DM IRE2 (SDK) is the microscope.
Make sure that the microscope is on and click “Initialize”.
You will see this message when the microscope is initialized and usable. Click “OK” and the software is ready to use.
55 CMN Imaging Core
1/16/2013
Device SCSI Controller V2 cannot find a Leica
Scanner on SCSI bus
If you see “Device SCSI Controller V2 cannot find a Leica Scanner on
SCSI bus”, the scan head has not completely booted up.
Click OK. Then you will see this message - “Hardware legend can’t get all the values”.
Click OK again. Then the software will load, but it will look like this and you won’t be able to take images.
56 CMN Imaging Core
1/16/2013
What to do:
1. Shut down the Leica software and wait a few minutes before restarting the software. Frequently, all the system needs is more time.
2. If step 1 doesn’t work, then the scan head might not have started up completely. In this case, turn off the lasers and the scan button (lower red switch), wait a few seconds and turn the scan button and the lasers back on.
Wait a few minutes and restart the software.
3. If neither step
1 or step 2 works, turn off the lasers and the scan button, as well as the microscope. Wait a few minutes and restart the lasers, microscope and scan button. Wait a few more minutes and restart the software.
4. If nothing works, do a complete shutdown of the system and then restart the whole system.
57 CMN Imaging Core
1/16/2013
advertisement
Key Features
- High resolution imaging
- Excellent optical sectioning
- Wide range of applications
- User-friendly software