Dräger X-am 5000 Multi-Gas Monitor Technical Manual

Dräger X-am 5000 Multi-Gas Monitor Technical Manual

Below you will find brief information for Multi-Gas Monitor X-am 5000. This manual will guide you through the basic functions and features of the device. You can learn about switching on and off, entering the workplace, and using the quick menu. The manual covers calibration, maintenance, and troubleshooting procedures. It also includes information on replacing batteries and sensors and offers valuable insights into safety instructions.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

X-am 5000 Multi-Gas Monitor Technical Manual | Manualzz
Dräger X-am® 5000
approved as type MQG 0010
Multi-Gas Monitor
04723999_01.eps
Technical Manual
Content
Content
For Your Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Tests and Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Marking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Intended operating area and operating conditions . . . . . . . . . . . . . . . . . . . . . . . . . . .
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5
5
6
6
What is What . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Front panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Rear panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Special symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8
8
8
8
9
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Standard gas configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Standard device configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Activating the Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Switching on the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Switching off the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Before entering the workplace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
During operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Calling the Info Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Calling the Info-Off Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Calling the Quick Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Possible functions of the quick menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Quick menu "Delete peak values" . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Calling the Calibration Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Calibration menu functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
12
12
12
12
14
15
16
16
16
16
17
18
Identifying Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Concentration pre-alarm A1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Concentration main alarm A2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
STEL / TWA exposure alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Battery pre-alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Battery main alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Device alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
18
18
18
19
19
19
19
Operation with pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
With Dräger Pump X-am 1/2/5000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
With manual pump adapter and rubber ball pump . . . . . . . . . . . . . . . . . . . . . . . . . .
Observe the following during measuring mode with pump . . . . . . . . . . . . . . . . . . . .
20
20
20
20
Configuring the Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Device settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
2
Content
Read Database and Display Graphically . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Faults, Cause and Remedy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Warning messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Fault messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Maintenance intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Carry out manual bump test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Manual implementation without the documentation of results in the device
memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Menu implementation with the documentation of results in the device memory . 34
Automatic implementation with the Bump Test Station . . . . . . . . . . . . . . . . . . . . 35
Calibrating the Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Carrying out the fresh air calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Carrying out 1-button calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Calibrating the sensitivity for an individual measuring channel . . . . . . . . . . . . . . 42
Span calibration for CatEx . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Replacing the batteries / rechargeable batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Charging the rechargeable batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Charging with the multiple charging station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Charging with charging module and plug-in power pack or vehicle charging
adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Replacing the Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Sensor warm-up acceleration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
X-am 5000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Sensor data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Order List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
3
For Your Safety
For Your Safety
General safety statements
 Before using this product, carefully read the associated Instructions for Use. This
document does not replace the Instructions for Use.
Definitions of alert icons
The following alert icons are used in this document to provide and highlight areas of
the associated text that require a greater awareness by the user. A definition of the
meaning of each icon is as follows:
DANGER
Indicates an imminently hazardous situation which, if not avoided, will result in death
or serious injury.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death
or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, could result in physical injury, or damage to the product or environment. It may also be used to alert
against unsafe practices.
NOTICE
Indicates additional information on how to use the product.
Intended Use
Portable gas detection instrument for the continuous monitoring of the concentration of
several gases in the ambient air within the working area and in explosion-hazard areas.
X-am 5000, depending on the device type and configuration of DrägerSensors:
independent measurement of one up to five gases.
4
Tests and Approvals
Tests and Approvals
Marking
Supply unit 83 22 237;
approved as Type ABT 0100
Temperature class T4
-20 °C ≤ Ta ≤ +50 °C
when used with alkaline batteries
Duracell Procell MN 1500*)
0736
-XX
PO Ex ia I X
0 Ex ia IIC T3 X
PB Ex d ia l X
1 Ex d ia IlC T4/T3 X
Dräger Safety
Type: MQG 0010
Temperature class T3
-20 °C ≤ Ta ≤ +40 °C
when used with NiMH batteries
GP 180AAHC*) (1800 mAh)
23560 Lübeck, Germany
C
US
C22.2 No.152
0158
I M1 / II 1G
for comb. sensor
I M2 / II 2G
Um=4.6V Im=1.3A
Ex ia I/IIC T3 Ma/Ga Ex d ia I/IIC T4/T3 Mb/Gb
BVS 10 ATEX E 080X
IECEx BVS 10.0053X
ANZEx 11.2003X
PFG 10 G 001X
Intrinsically safe Ex ia, CSA 11 1800517
CSA: Class I, Div. 1, Gr.. A,B,C,D TC T4/T3
Class I, Zone 0, A/Ex ia IIC T3 /Ga
Class I, Zone 1, A/Ex d ia IIC T4/T3 /Gb
-20°C ืTa ื+50/+40°C: see Battery Pack!
For TC T4/T3: see Battery Pack!
Warning: Read manual for safety precautions.
Avertissement: Lire le manuel avant utilisation.
Do not change or charge batteries in haz loc.
or with alkaline batteries
Varta Type 4006*)
Varta Type 4106*)
Panasonic LR6 Powerline
Supply unit 83 18,704;
approved as HBT 0000
Temperature class T4
-20 °C ≤ Ta ≤ +50 °C
Supply unit 83 22,244;
approved as HBT 0100
Temperature class T4
-20 °C ≤ Ta ≤ +50 °C
*) Not subject to the Metrological Performance Test BVS10ATEXE080X and PFG10G001X.
Serial no.1) on separate sticker
WARNING
Read the safety measures in the Instructions for Use.
Do not replace or charge batteries in potentially explosive areas. Danger of explosion!
The BVS 10 ATEX E 080 X technical suitability test is based on the calibration with the
target gas.
1)
The year of manufacture is shown by the third letter of the serial no.: D = 2012, E = 2013, F = 2014, H
= 2015, J = 2016, K = 2017, L = 2018, M = 2019, N = 2020 etc.
Example: Serial no. AREH-0054: the third letter is E, which means year of manufacture 2013.
5
Tests and Approvals
Intended operating area and operating conditions
Hazardous areas classified by zones
The device is intended for use in explosion-hazard areas or mines, in which firedamp
classified by zone 0, zone 1 or zone 2 may occur. It is determined for use within a temperature range of –20 °C to +50 °C, and for areas in which gases of explosion groups
IIA, IIB or IIC and temperature class T3 or T4 (depending on the batteries and
rechargeable battery) may be present. For zone 0, the temperature class is limited to
T3.
If used in mines, the device is only to be used in areas known to have a low risk of
mechanical impact.
Hazardous areas classified by divisions
This device is intended to be used in hazardous areas or mines susceptible to firedamp
classified as Class I & II, Div. 1 or Div. 2 within a temperature range of -20 °C to +50 °C
and where gases or dusts of groups A, B, C, D and temperature class T3 or T4
(depending on battery pack and batteries) may be present.
For applications in accordance with CSA (Canadian Standards Association), the
following should be observed:
Only the performance of the detector part of this instrument for flammable gases has
been tested. The instrument has not been approved for use in mines by the CSA.
WARNING:
The sensitivity must be tested on a daily basis before first use with a known
concentration of the gas to be measured in accordance with 25 to 50 % of the
concentration limit value. The accuracy must be 0 to +20 % of the actual value. The
accuracy can be corrected via calibration.
Safety instructions
WARNING
To reduce the danger of explosion, do not mix new batteries with old batteries and do
not mix batteries made by different manufacturers.
WARNING
Always disconnect the device from the power pack before carrying out any maintenance operations.
WARNING
Substitution of components may impair intrinsic safety.
CAUTION
Not tested in an oxygen-enriched atmosphere (>21 % O2).
6
Tests and Approvals
WARNING
High off-scale readings may indicate an explosive concentration.
CAUTION
Use only supply units ABT 0100 (83 22 237), HBT 0000 (83 18 704) or HBT 0100
(83 22 244). Check the supply unit for approved batteries and applicable temperature
class.
Note the following for CSA (Canadian Standards Association) applications:
For the CSA approval only the functions of the device component that is used to measure flammable gases are tested. The device is not approved by CSA for use in mining.
WARNING:
Before daily use, test the sensitivity with a known concentration of the applicable gas
corresponding to 25 to 50% of the maximum concentration. The accuracy must be
within a range of 0 to +20% of the actual value. Perform a calibration to correct the
accuracy if necessary.
7
What is What
What is What
Front panel
1
Gas entry
Alarm LED
Buzzer
OK key
key
Display
Tool for replacing
the sensor
2
2
3
7
X-am 5000
6
5
00223999_02.eps
1
2
3
4
5
6
7
4
0
Rear panel
IR interface
Fastening clip
Type plate
Charging contacts
Power pack
Serial no.
1
2
6
for 5 measuring channels
only:
1 Measured gas
display
with unit
2 Measured value
display
3 Special symbols
other:
4 Measured gas
display
5 Measured gas
display with unit
6 Special symbols
4
5
Display
CH4
%LEL
O2
Vol%
CO
ppm
H 2S
ppm
NH3
ppm
1
CH4
%LEL
O2
Vol%
CO
ppm
2
3
4
The following only shows the device version with 5 measuring channels.
8
00323999_02.eps
3
5
6
00423999_02_en.eps
1
2
3
4
5
6
What is What
Special symbols
Fault message, refer to page 15
Warning message, refer to page 15
The peak value display for all measuring gases,
refer to page 15
The exposure evaluation display (TWA) for measuring gases, e.g., H2S and
CO,
refer to page 15
The exposure evaluation display (STEL) for measuring gases, e.g,. H2S and
CO,
refer to page 15
The device is set to function test with gas (bump test),
refer to page 33
The device is set to the fresh air calibration function,
refer to page 39
The device is set to the 1-button calibration function,
refer to page 40
The device is set to the single gas calibration function,
refer to page 42
The function for password entry is active,
refer to page 17
Battery / rechargeable battery 100 % full
Battery / rechargeable battery 2/3 full
Battery / rechargeable battery 1/3 full
Battery / rechargeable battery empty
9
Configuration
Configuration
Standard gas configuration
CatEx 125 PR
[%LEL]
CatEx 125 Mining PR
[%LEL]
XXS O2 [vol. %]
XXS O2 100 [vol. %]
XXS CO [ppm]
XXS CO LC [ppm]
XXS CO HC [ppm]
XXS CO H2-CP [ppm]
XXS H2 [ppm]
XXS H2S [ppm]
XXS H2S LC [ppm]
XXS H2S HC [ppm]
XXS H2S/CO [ppm]
XXS NO [ppm]
XXS NO2 [ppm]
XXS SO2 [ppm]
XXS PH3 [ppm]
XXS PH3 HC [ppm]
XXS HCN [ppm]
XXS HCN PC [ppm]
XXS NH3 [ppm]
XXS CO2 [vol. %]
XXS Cl2 [ppm]
XXS H2 HC [vol. %]
XXS OV [ppm]
XXS OV A [ppm]
XXS Odorant [ppm]
XXS Amine [ppm]
XXS COCI2 [ppm]
XXS O3 [ppm]
XXS NO2 LC [ppm]
10
self-latching
can be
acknowledged
setpoint
Alarm A2 1)
self-latching
can be
acknowledged
Alarm A1 1)
Measuring
range 1)
setpoint
DrägerSensor
0 to 100
20
Yes
No
40
No
Yes
0 to 100
20
Yes
No
40
No
Yes
0 to 25
0 to 100
0 to 2,000
0 to 2.000
0 to 10,000
0 to 2.000
0 to 2.000
0 to 200
0 to 100
0 to 1,000
0 to 200 H2S
0 to 2,000 CO
0 to 200
0 to 50
0 to 100
0 to 20
0 to 2,000
0 to 50
0 to 50
0 to 300
0 to 5
0 to 20
0 to 4
0 to 200
0 to 200
0 to 40
0 to 100
0 to 10
0 to 10
0 to 50
19 2)
18 2)
30
30
600
30
200
10
5
100
10 H2S
30 CO
25
5
1
0.1
5
10
5
50
0.5
0.5
0.8
10
10
10
10
0,1
0,1
0,5
No
No
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No
23
24
60
60
1,200
60
400
20
10
200
20 H2S
60 CO
50
10
2
0.2
10
20
10
100
1
1
1.6
20
20
20
20
0,2
0,2
1,0
No
No
No
No
No
No
No
Yes
No
No
No
No
No
No
No
No
No
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
No
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
No
No
Configuration
1)
2)
Different settings can be selected to meet customer requirements on delivery. The current setting can
be checked and changed with the Dräger CC Vision software. A version of Dräger CC-Vision suitable
for the Dräger X-am 5000/2000 can be downloaded on the product page of the X-am 5000: www.draeger.com.
In the case of O2 A1 is the lower alarm setpoint: an alarm is issued if the value is too low.
Standard device configuration
NOTICE
Only trained personnel are permitted to make changes to the device configuration.
Dräger X-am® 50001)
Bump test mode 2)
Fresh-gas adjustment2)
Operating signal 2) 3)
Switch off 2)
LEL factor 2)
(ch4)
STEL 2) 4) 5)
(short-term average)
TWA 2) 5) 6)
(shift average)
Alarm A1 7)
Alarm A1 at O2 sensor 7)
Alarm A2 7)
1)
2)
3)
4)
5)
6)
7)
Extended bump test
ON
ON
allowed
4.4 (vol. %)
(4.4 vol. % corresponds to 100 %LEL)
STEL function - disabled
Average value duration = 15 minutes
TWA function - disabled
Average value duration = 8 hours
can be acknowledged, non-latching, pre-alarm,
rising flank
cannot be acknowledged, latching, like main alarm,
falling flank
cannot be acknowledged, latching, main alarm,
rising flank
X-am® is a registered trademark of Dräger.
Different settings can be selected to meet customer requirements on delivery. The current setting can
be checked and changed with the Dräger CC Vision software.
A periodic short signal indicates the operating capacity of the instrument. If there is no operating signal,
correct operation cannot be guaranteed.
STEL: average value of an exposure over a short period, generally 15 minutes.
Interpretation only if the sensor is designed for this.
TWA: shift averages are workplace limit values for generally eight hours per day of exposure for five
days a week during a working life.
Latching and acknowledgement of alarms A1 and A2 can be configured with the Dräger CC Vision PC
software.
Changing the standard configuration: See “Configuring the Device” on page 21.
WARNING
After a basic initialization has been carried out with the PC software Dräger CC
Vision, individual alarm settings may have been changed.
11
Activating the Device
Activating the Device
Before using the device for the first time, insert the supplied batteries or a charged
NiMH power pack T4 (order no. 83 18 704), refer to Replacing the Batteries, page 47.
Charge the rechargeable batteries if necessary, page 48.
The Dräger X-am 5000 is ready for operation.
Operation
Switching on the device









Press and hold the OK key for approx. 3 seconds until the countdown » 3 . 2 . 1 «
shown in the display has elapsed.
All the display segments, including the visual, audible and vibration alarms, are activated for a short time.
The software version is displayed.
The device performs a self test.
The next sensor which is next due for calibration is displayed with the days remaining until the next calibration, e.g., » Ex %LEL CAL 20 «.
The time until the bump test interval elapses is displayed in days, e.g., » bt 123 «.
« (STEL)1) for all toxic
All A1 and A2 alarm thresholds and »
« (TWA)1) and »
gases (e. g. H2S or CO) are displayed consecutively.
During the warm-up period of the sensors, the respective display of the measured
value flashes and the special symbol »
« (for warning) is displayed. No alarms
are issued during the running-in period of the sensors.
Press the OK key to cancel the display of the activation sequence.
Switching off the device


Press and hold the OK key and
key at the same time until the countdown
» 3 . 2 . 1 « shown in the display has elapsed.
Before the device is switched off, the visual, audible and vibration alarms are activated for a short time.
Before entering the workplace
WARNING
Before any measurements relevant to safety are made, check the adjustment with a
bump test, adjust if necessary and check all alarm elements. If national regulations
apply, a bump test must be performed according to the national regulations. Faulty
adjustment may result in incorrect measuring results, with possible serious consequences.

Switch on the device. The current measured values are shown in the display.
1) Only when activated in the instrument configuration. Delivery condition: not activated.
12
Operation



Observe any warning »
« or fault messages »
«.
The device can be operated normally. If the warning message does not go
out automatically during operation, the device must be maintained after
the end of use.
The device is not ready to measure and requires maintenance.
If one of these special symbols is displayed, appropriate measures, refer to page 25
to page 28, must be taken.
Check that the gas inlet opening on the device is not covered.
WARNING
Explosion hazard! To reduce the risk of flammable or explosive atmospheres igniting,
it is essential that the warning notices below are observed:
 Fractions of catalytic poisons in the measuring gas (e.g. volatile silicon, sulphur,
heavy metal compounds or halogenated hydrocarbon) can damage the Cat Ex
sensor. If the CatEx sensor can no longer be calibrated to the target concentration,
the sensor must be replaced.
 In case of measurements in oxygen-deficient atmosphere (<8 vol. % O2) the CatEx
sensor may show incorrect displays; in this case, a reliable measurement with a
CatEx sensor is not possible.
 In an oxygen enriched atmosphere (>21 vol. % O2), the explosion protection cannot be guaranteed; remove instrument from the Ex area.
 High values outside the display area indicate an explosive concentration where applicable.
13
Operation
During operation
During operation, the measured values for every measured gas are displayed.
If a measuring range is exceeded or a negative drift occurs, the following displays are
shown instead of the measured value display:
»
« (concentration too high) or
»
« (measuring range not reached) or
»
« (blocking alarm).


Excess concentrations of flammable materials can lead to a lack of oxygen.
At O2 concentrations below 8 Vol.-%, on the Ex channel instead of the measured
value a fault is displayed with »
« if the measured value is below the pre-alarm
threshold and the device is equipped with an O2 sensor (not with setting CH4 with
measuring range >100 %LEL).
In the event of an alarm, the corresponding displays, including the visual, audible and
vibration alarms, are activated – see “Identifying Alarms” on page 18.
NOTICE
Special states in which there is no measuring operation (quick menu, calibration
menu, warm-up of sensors, password input) are indicated by a visual signal (slow
flashing of the alarm LED
).
If the measuring range is exceeded significantly on the CatEx channel (very high
concentration of flammable materials), a blocking alarm is triggered. This CatEx
blocking alarm can be acknowledged manually by switching the device off and back on
again in fresh air.
In configuration setting CH4 with measuring range 100 vol. %, no blocking alarm is
triggered as the heat conductance measurement principle is used.
14
Operation
CAUTION
The measuring range 0 to 100 vol. % CH4 is not suitable for monitoring explosive
mixtures in the measuring range from 0 to 100%LEL.
WARNING
When using a CatEx sensor in the Dräger X-am 5000, a calibration of zero point and
sensitivity must be carried out after an impact load that results in a fresh air display
not equal to zero.
After the measuring range of the TOX measuring channels has been exceeded temporarily (up to one hour), checking the measuring channels is not necessary.
Calling the Info Mode

In measuring mode, press the
OK
key for approx. 3 seconds.
If any warning or fault messages exist, the corresponding information or error codes
are displayed (page 25 to page 31).
Press the OK key successively for the next display.
The peak values and the exposition values TWA1) and STEL1) are displayed.
Warning messages are displayed. Numerical codes of warning messages:
see page 25.
OK key
Fault messages are displayed. Numerical codes of fault messages: see
page 28.
OK key
The peak values = the maximum measured values in the case of, e.g., CO,
H2S, ... or the minimum measured values in the case of O2 within the storage
interval are displayed
OK key
The average values of the exposures based on a shift of, e.g., 8 hours (TWA)
of all the active sensors for the exposure evaluation are displayed
OK key
The short-term values (STEL) = average values of the concentrations over
the average value duration of all the active sensors for the exposure evaluation are displayed
OK key
The device is in measuring mode again

If no key is pressed for 10 seconds, the device returns automatically to measuring
mode.
1)
Only when activated in the device configuration. Delivery status: not activated.
15
Operation
Calling the Info-Off Mode
When the device is in a deactivated state, press the
key.
The name of the gas, measuring unit and measuring range limit value are displayed for
all channels.
Pressing the
OK
key again exits the Info Off mode (or via timeout).
Calling the Quick Menu

Only the fresh air calibration is activated in the quick menu on delivery. The PC software Dräger CC Vision can be used to activate the bump test for the quick menu
and/or the function for displaying and deleting peak values.

In measuring mode, press the
key three times.
If no functions have been activated in the quick menu, the device remains in measuring mode.

You can select the activated functions of the quick menu by pressing the

Press the
OK
key.
key to call the selected function.
Possible functions of the quick menu
Bump test, refer to page 33
Fresh-gas adjustment, refer to page 39
Delete peak values, refer to page 16

Press the

If no key is pressed for 60 seconds, the device returns automatically to measuring
mode.
key to cancel the active function and to switch to measuring mode.
Quick menu "Delete peak values"
ch4
%LEL
O2
Vol%
CO
ppm
H 2S
ppm
NH3
ppm
04823999_01_en.eps
After the function has been selected, the
current peak values are displayed; the
peak values special symbol appears in
the display at the same time.
16
Operation

The peak values can be deleted by
pressing the OK key for 5 sec.
The adjacent display appears.
Press the
OK
key to end the function.
ch4
%LEL
O2
Vol%
CO
ppm
H 2S
ppm
NH3
ppm
04923999_01_en.eps

Calling the Calibration Menu

The calibration menu can only be accessed by entering a password.
Password on delivery: » 001 «
The default password on delivery can be changed using the PC software
Dräger CC Vision.



In measuring mode, press the
key for at least 4 seconds.
The function for entering the password is selected.
The »
« special symbol (for the "enter password" function) is displayed.

The display shows » 000 «, with the
first digit flashing.
Use the
key to set the flashing digit.





Press the OK key, the second digit
starts flashing.
Use the
key to set the flashing digit.
Press the OK key, the third digit starts
flashing.
Use the
key to set the flashing digit.

Press the OK key to confirm the password once it has been set completely.

The calibration menu functions can
now be selected by pressing the
key.

Press the
function.
OK
02323999_01.eps

key to call the selected
17
Identifying Alarms
Calibration menu functions
Fresh air calibration, refer to page 39
1-button calibration, refer to page 40
Single gas calibration, refer to page 42

Press the

If no key is pressed for 10 minutes, the device automatically returns to measuring
mode.
key to cancel the active function.
Identifying Alarms
An alarm is displayed visually, audibly and through vibration in a specific pattern.
Concentration pre-alarm A1
The alarm is indicated by an intermittent alarm message:
Display » A1 « and measured value alternating: not for O2!

The pre-alarm A1 is not self-latching and stops when the concentration has dropped
below the alarm setpoint A1.
 In the case of A1 a single tone is audible and the alarm LED flashes.
Acknowledging the pre-alarm:
 Press the OK key. Only the audible alarm and the vibration alarm are switched off.
Concentration main alarm A2
The alarm is indicated by an intermittent alarm message:
Display » A2 « and measured value alternating:
In the case of A2 a double tone is audible and the alarm LED flashes
twice.
For O2:» A1 « and measured value alternating = oxygen deficiency
» A2 « and measured value alternating = oxygen surplus
DANGER
Leave the area immediately. Danger to life! A main alarm is self-latching and cannot
be acknowledged or cancelled.
After leaving the area, if the concentration is less than the alarm setpoint A2:
 Press the OK key. The alarm messages are switched off.
If the measuring range is exceeded significantly on the CatEx channel (very high
concentration of flammable materials), a blocking alarm is triggered. This CatEx
blocking alarm can be acknowledged manually by switching the device off and back on
again in fresh air.
18
Identifying Alarms
In configuration setting CH4 with measuring range 100 vol. %, no blocking alarm is
triggered as the heat conductance measurement principle is used.
CAUTION
The measuring range 0 to 100 vol. % CH4 is not suitable for monitoring explosive
mixtures in the measuring range from 0 to 100 %LEL.
STEL / TWA exposure alarm
The alarm is indicated by an intermittent alarm message:
Display » A2 « and »
« (TWA) or »
« (STEL) and measured
value alternating:
CAUTION
Leave the area immediately. After this alarm, the deployment of personnel is subject to the relevant national regulations.


STEL and TWA alarms cannot be acknowledged or canceled.
Switch off the device. The values for the exposure evaluation are deleted after the
device is switched on again.
Battery pre-alarm
The alarm is indicated by an intermittent alarm message:
Flashing special symbol »
« on the right side of the display:
Acknowledging the pre-alarm:
 Press the OK key. Only the audible alarm and the vibration alarm are switched off.
 The battery still lasts approx. 20 minutes after the first battery pre-alarm.
Battery main alarm
The alarm is indicated by an intermittent alarm message:
Flashing special symbol »
« on the right side of the display:
The battery main alarm cannot be acknowledged or canceled:
 The device is automatically switched off again after 10 seconds.
 Before the device is switched off, the visual, audible and vibration alarms are activated for a short time.
Device alarm
The alarm is indicated by an intermittent alarm message:
Special symbol »
« on the right side of the display:



The device or one or several sensor channels are not ready for operation.
For remedies, refer to page 25 to page 31.
If necessary, commission the Dräger Service Center to eliminate the error.
19
Operation with pump
Operation with pump
With Dräger Pump X-am 1/2/5000
Accessories:
Dräger Pump X-am 1/2/5000, sampling hose and probes, refer to Order List, see
“Accessories” on page 60..
Commissioning and performing the measurement:
 Refer to the Instructions for Use of the Dräger Pump X-am 1/2/5000.
With manual pump adapter and rubber ball pump
Accessories:
For manual pump adapter, rubber ball pump, sampling hose and probes, refer to Order
List, see “Accessories” on page 60..
Commissioning and performing the measurement:
 Refer to the Instructions for Use of the accessories used.
Observe the following during measuring mode with pump





20
The required waiting time when flushing the hose probe:
Before every measurement, flush the Dräger sampling hose or the Dräger probes
with the air sample to be measured.
It is absolutely necessary to flush the extension hose to eliminate or minimize the
effects which may interfere with measurements when using a sampling hose or a
probe, e.g., memory effects, dead volume.
The flushing phase depends on various factors, e.g., type and concentration of the
gas or vapor to be measured, material, length, diameter, and age of the sampling
hose or probe. Generally, when using a sampling hose (new, dry, clean), a typical
flushing time of approx. 3 seconds is required for each meter. This flushing time applies in addition to the sensor response time (see the Instructions for Use of the gas
measuring device used).
Example:
In the case of a sampling hose with a length of 10 m, the flushing time is approx.
30 seconds and the sensor response time is in addition approx. 60 seconds. Therefore, the total time before reading the gas measuring device is approx. 90 seconds.
The flow rate alarm is delayed by 10 to 30 seconds depending on the length of the
hose.
Configuring the Device
Configuring the Device
To individually configure a device with standard configuration, the device must be connected with a PC.
IR
USB 2.0
0
Dräger
CC-Vision
X-am 5000
USB DIRA with USB cable
(order no. 83 17 409)
USB 2.0
0
Dräger
CC-Vision
Calibration cradle (order no. 83 18 752)
with inserted
USB DIRA with USB cable (order no. 83 17 409)
0
Dräger
CC-Vision
E-Cal
01423999_01_en.eps
USB 2.0 / COM
E-Cal module Dräger X-am 125
(order no. 83 18 754)
The installed PC software Dräger CC Vision is used for configuration.
 Observe the documentation and online help of the software.

A version of Dräger CC-Vision suitable for the Dräger X-am 5000 can be downloaded
on the product page of the X-am 5000: www.draeger.com.
21
Configuring the Device
Device settings
NOTICE
Only trained personnel are permitted to make changes to the device configuration.
The following changes can be made to the device parameters for a device:
Designation
Password
Operating signal LED 1)
Operating signal horn1)
Switch-off mode
Shift length (TWA) 2) (in minutes)
Short-term exposure limit (STEL) 3) 4) (in
minutes)
User ID(12 characters)
Switch database on or off
Overwrite database
Database mode
Database interval
Field
Numeric field (3-figure)
Yes/No
Yes/No
“Switch off permitted” or
“Switch off prohibited” or
“Switch off prohibited at A2”
60 - 14400
(setting for exposure alarm)
0 - 15
(setting for exposure alarm)
Alphanumeric field
On/Off
Yes/No
Peak/Average
1 s / 10 s / 30 s / 1 min / 2 min / 5 min /
10 min / 30 min
Date
(date on the PC)
Time
(time on the PC)
Warning after expiry of calibration interval Yes/No
Error after expiry of calibration interval
Yes/No
Delay until error after expiry of calibration 0 - 10
interval (days)
Automatic detection of Bump Test Station Yes/No
Activate sensitivity calibration following
Yes/No (relates only to a device connegative bump test
nected to the Dräger Bump Test Station)
Bump test mode
“extended bump test” or “quick bump test”
or “bump test deactivated”
Warning after expiry of bump test interval Yes/No
Error after expiry of bump test interval (if Yes/No
warning activated)
Bump test interval (days)
1 - 732
Delay until error after expiry of cal. interval 0 - 10
(days)
Activate user service life
Yes/No
22
Configuring the Device
User service life (days)
(if activated)
Running in
LEL category
1)
2)
3)
4)
0 - 999
Yes/No
“---” or “PTB” or “IEC” or “NIOSH” (if this is
changed, the LEL factor will be altered to
match)
At least one of the two operating signals must be switched on.
Corresponds to the averaging time and is used to calculate the exposure value TWA.
Only evaluated if the sensor is provided for the purpose.
Corresponds to the averaging time and is used to calculate the exposure value STEL.
Sensor settings
The following changes can be made to the sensor parameters for the sensors:
Designation
Field
Alarm threshold A1 (in measurement unit) 0 - A2
Alarm threshold A2 (in measurement unit) A1 – Measuring range limit value
Type of evaluation1)
Inactive, TWA, STEL, TWA+STEL
Alarm threshold STEL
0 – Measuring range limit value
(in measurement unit)1)
Alarm threshold TWA
0 – Measuring range limit value
(in measurement unit)1)
Calibration interval (days)
0 - 180 (sensor-dependent)
Unit (sensor-dependent)
Vol%, %UEG, %LEL, %LIE, ppm, mbar,
ppb, mg/m3
Gas name: “Ex” (CatEx sensor only)
Yes/No
1)
Only evaluated if the sensor is provided for the purpose.
Testing the parameters
In order to ensure that the values have been correctly transferred to the gas measuring
device:
 Press the touch button Data from X-am 1/2/5x00
 Check parameters.
23
Read Database and Display Graphically
Read Database and Display Graphically
To read the database of the device and display it graphically, the device must be connected with a PC.
IR
USB 2.0
0
Dräger
GasVision
X-am 5000
USB DIRA with USB cable
(order no. 83 17 409)
USB 2.0
0
Dräger
GasVision
Calibration cradle (order no. 83 18 752)
with inserted
USB DIRA with USB cable (order no. 83 17 409)
0
Dräger
GasVision
E-Cal
E-Cal module Dräger X-am 125
(order no. 83 18 754)
The installed PC software Dräger GasVision is used for reading and displaying the
database.
 Observe the documentation and online help of the software.
24
01823999_01_en.eps
USB 2.0 / COM
Faults, Cause and Remedy
Faults, Cause and Remedy
Fault
Cause
Remedy
Not possible to switch on
the device
Discharge the power pack
Charge the power pack,
page 48.
Discharge the alkaline batter- Insert new alkaline batteries,
ies
page 47.
Not possible to switch off
the device
Display » – – «
The device is not set to meas- Select measuring mode.
uring mode
The device is configured to
"Disable prohibited"
Configure the device to "Disable
allowed" with Dräger CC Vision.
Measuring range calibrated
incorrectly
Recalibrate the measuring
range, page 38.
Electronics or sensors
defective
Must be repaired by
DrägerService.
To display the numerical codes of the warning and fault messages in the info mode,
page 15.
Warning messages
Special symbol »
«
Cause
and displayed numerical
code:
Remedy
152
Customer's service life counter about to elapse
Reset the service life counter
using Dräger CC Vision.
153
Database 90 % full
Read the database soon and
clear memory afterwards.
154
Database full
Read the database and clear
memory.
155
Interval for the function test
with gas (bump test) elapsed
Carry out the function test,
page 33.
159
Calibration not possible. The
menu function cannot be car- Determine the message code
ried out because of a mesvia the info menu and switch it
sage which is preventing the off, if necessary.
function (e.g., sensors in
warm-up phase).
251
DrägerSensor CatEx 125 PR Wait until warm-up time is
in warm-up phase
complete.
252
DrägerSensor CatEx 125 PR Wait until warm-up time is
in warm-up phase
complete.
25
Faults, Cause and Remedy
Special symbol »
«
Cause
and displayed numerical
code:
Remedy
253
Ex concentration has drifted
into the negative range
Carry out fresh air calibration,
page 39.
254
The temperature is too high
Operate the device within the
allowed temperature range.
255
The temperature is too low
Operate the device within the
allowed temperature range.
256
The calibration interval for
Carry out span calibration for
DrägerSensor CatEx 125 PR DrägerSensor CatEx 125 PR,
has elapsed
page 42.
257
Alarm setpoint A2 setting is
greater than 60 %LEL
271
The calibration interval for
Carry out span calibration for
thermal conduction for
DrägerSensor CatEx 125 PR,
DrägerSensor CatEx 125 PR page 42.
has elapsed
272
Sensor is switched off due to
excess gas
Restart the device
351
DrägerSensor XXS EC1 in
the warm-up phase
Wait until warm-up time is
complete.
352
DrägerSensor XXS EC1 in
the warm-up phase
Wait until warm-up time is
complete.
353
EC1 concentration has drifted Carry out fresh air calibration,
into the negative range
page 39.
354
The temperature is too high
Operate the device within the
allowed temperature range.
355
The temperature is too low
Operate the device within the
allowed temperature range.
356
The calibration interval for
DrägerSensor XXS EC1 has
elapsed
Carry out span calibration for
DrägerSensor XXS EC1,
page 42.
357
Alarm setpoint A2 setting is
greater than 60 %LEL
Set alarm setpoint to less than
60 %LEL.
451
DrägerSensor XXS EC2 in
the warm-up phase
Wait until warm-up time is
complete.
452
DrägerSensor XXS EC2 in
the warm-up phase
Wait until warm-up time is
complete.
453
EC2 concentration has drifted Carry out fresh air calibration,
into the negative range
page 39.
454
The temperature is too high
Operate the device within the
allowed temperature range.
455
The temperature is too low
Operate the device within the
allowed temperature range.
26
Set alarm setpoint to less than
60 % LEL.
Faults, Cause and Remedy
Special symbol »
«
Cause
and displayed numerical
code:
Remedy
456
The calibration interval for
DrägerSensor XXS EC2 has
elapsed
Carry out span calibration for
DrägerSensor XXS EC 3,
page 42.
457
Alarm setpoint A2 setting is
greater than 60 %LEL
Set alarm setpoint to less than
60 %LEL.
551
DrägerSensor XXS EC3 in
the warm-up phase
Wait until warm-up time is complete.
552
DrägerSensor XXS EC3 in
the warm-up phase
Wait until warm-up time is
complete.
553
EC3 concentration has drifted Carry out fresh air calibration,
into the negative range
page 39.
554
The temperature is too high
Operate the device within the
allowed temperature range.
555
The temperature is too low
Operate the device within the
allowed temperature range.
556
The calibration interval for
DrägerSensor XXS EC3 has
elapsed
Carry out span calibration for
DrägerSensor XXS EC 3,
page 42.
557
Alarm setpoint A2 setting is
greater than 60 %LEL
Set alarm setpoint to less than
60 %LEL.
575
Calibration interval for the
compensation channel has
elapsed
Adjust the sensitivity of the compensation channel.
576
Calibration required because
of overgassing.
Adjust the sensitivity of the compensation channel.
651
DrägerSensor XXS EC 4 in
the warm-up phase
Wait until warm-up time is
complete.
652
DrägerSensor XXS EC 4 in
the warm-up phase
Wait until warm-up time is
complete.
653
EC 4 concentration has
Carry out fresh air calibration,
drifted into the negative range page 39.
654
The temperature is too high
Operate the device within the
allowed temperature range.
655
The temperature is too low
Operate the device within the
allowed temperature range.
656
The calibration interval for
Carry out span calibration for
DrägerSensor XXS EC 4 has DrägerSensor XXS EC 4,
elapsed
page 42.
657
Alarm setpoint A2 setting is
greater than 60 %LEL
Set alarm setpoint to less than
60 %LEL.
27
Faults, Cause and Remedy
Fault messages
Special symbol »
«
Cause
and displayed numerical
code:
Remedy
102
The customer's service life
counter has elapsed
Reset the service life counter
using Dräger CC Vision.
103
The device is defective
The device must be repaired by
DrägerService.
104
Check sum error program
code
The device must be repaired by
DrägerService.
105
The bump test interval has
elapsed
Carry out bump test, page 35.
106
The calibration interval has
Carry out span calibration,
elapsed (at least 1 calibration page 40 or page 42.
interval has elapsed)
107
Bump test error (at least
1 channel has a bump test
error)
Carry out bump test, page 35 or
carry out span calibration,
page 40 or page 42.
108
The device is defective
The device must be repaired by
DrägerService.
109
The menu function cannot be Determine the error code via the
carried out because of an
info menu and switch it off, if
error.
necessary.
111
Failed alarm element test:
alarm light.
Repeat alarm element test with
Dräger X-dock.
112
Failed alarm element test:
alarm horn.
Repeat alarm element test with
X-dock.
113
Failed alarm element test:
Vibration motor.
Repeat alarm element test with
X-dock.
114
Defective parameter check
Correct parameters and repeat
test using X-dock
115
Device is disabled by X-dock. Activate device with X-dock.
116
Failed software update.
The device must be repaired by
DrägerService.
117
User parameters not feasible
Check configuration of user
parameters and adjust
201
The zero point calibration of Carry out fresh air calibration,
DrägerSensor CatEx 125 PR page 39.
is not valid
202
The span calibration of
Carry out span calibration,
DrägerSensor CatEx 125 PR page 40 or page 42.
is not valid
28
Faults, Cause and Remedy
Special symbol »
«
Cause
and displayed numerical
code:
Remedy
203
The measurement value of
Carry out fresh air calibration,
DrägerSensor CatEx 125 PR page 39.
is in the negative range
204
DrägerSensor CatEx 125 PR Check DrägerSensor
is not inserted or defective
CatEx 125 PR, page 51.
205
Error during the function test Repeat the function test. Caliwith gas (bump test) of
brate or replace DrägerSensor
DrägerSensor CatEx 125 PR CatEx 125 PR, if necessary
page 51.
207
Failed rise time test.
Repeat rise time test with Xdock.
208
User parameters not feasible
Check configuration of user
parameters and adjust
218
Blocking alarm not plausible.
Adjust the sensor.
221
DrägerSensor CatEx 125 PR Use in the sensor in an environcannot be operated due to
ment containing at least 8 vol. %
oxygen deficiency
O 2.
222
No valid zero point calibration Carry out fresh air calibration,
of DrägerSensor CatEx
page 39.
125 PR for thermal conduction
223
No valid span calibration of
Carry out span calibration for
DrägerSensor CatEx 125 PR thermal conduction, page 40 or
for thermal conduction
page 42.
224
Device incorrectly configured Change sensor for applicable
by Dräger CC-Vision.
channel with Dräger CC-Vision.
301
The zero point calibration of Carry out fresh air calibration,
DrägerSensor XXS EC1 is not page 39.
valid
302
The span calibration of
Carry out span calibration. Carry
DrägerSensor XXS EC1 is not out page 42 or fresh air calibravalid
tion, page 39.
303
The measured value of
Carry out fresh air calibration,
DrägerSensor XXS EC 1 is in page 39.
the negative range
304
DrägerSensor XXS EC1 is not Check DrägerSensor XXS EC1,
inserted or defective
page 51.
305
Error during the function test
with gas (bump test) of
DrägerSensor XXS EC1
Repeat function test. Calibrate
or replace DrägerSensor XXS
EC1, if necessary page 51.
306
Failed filter test.
Repeat filter test with X-dock.
29
Faults, Cause and Remedy
Special symbol »
«
Cause
and displayed numerical
code:
Remedy
307
Failed rise time test.
Repeat rise time test with Xdock.
308
User parameters not feasible
Check configuration of user
parameters and adjust
324
Device incorrectly configured Change sensor for applicable
by Dräger CC-Vision.
channel with Dräger CC-Vision.
326
Error during warm-up acceler- Disconnect and reconnect
ation Dräger Sensor XXS
power pack or replace the senEC1
sor. Sensor must not be loaded
with gas within the first 5 minutes.
401
The zero point calibration of Carry out fresh air calibration,
DrägerSensor XXS EC2 is not page 39.
valid
402
The span calibration of
Carry out span calibration,
DrägerSensor XXS EC2 is not page 42.
valid
403
The measured value of
Carry out fresh air calibration,
DrägerSensor XXS EC 2 is in page 39.
the negative range
404
DrägerSensor XXS EC2 is not Check DrägerSensor XXS EC2,
inserted or defective
page 51.
405
Error during the function test
with gas (bump test) of
DrägerSensor XXS EC2
Repeat function test. Calibrate
or replace DrägerSensor XXS
EC2, if necessary page 51.
406
Failed filter test.
Repeat filter test with X-dock.
407
Failed rise time test.
Repeat rise time test with Xdock.
408
User parameters not feasible
Check configuration of user
parameters and adjust
424
Device incorrectly configured Change sensor for applicable
by Dräger CC-Vision.
channel with Dräger CC-Vision.
426
Error during warm-up acceler- Disconnect and reconnect
ation Dräger Sensor XXS
power pack or replace the senEC2
sor. Sensor must not be loaded
with gas within the first 5 minutes.
501
The zero point calibration of Carry out fresh air calibration,
DrägerSensor XXS EC3 is not page 39.
valid
30
Faults, Cause and Remedy
Special symbol »
«
Cause
and displayed numerical
code:
Remedy
502
The span calibration of
Carry out span calibration,
DrägerSensor XXS EC3 is not page 42.
valid
503
The measured value of
DrägerSensor XXS EC3 is in
the negative range
504
DrägerSensor XXS EC3 is not Check DrägerSensor XXS EC3,
inserted or defective
page 51.
505
Error during the function test
with gas (bump test) of
DrägerSensor XXS EC3
Repeat function test. Calibrate
or replace DrägerSensor XXS
EC3, if necessary page 51.
506
Failed filter test.
Repeat filter test with X-dock.
507
Failed rise time test.
Repeat rise time test with Xdock.
508
User parameters not feasible
Check configuration of user
parameters and adjust
524
Device incorrectly configured Change sensor for applicable
by Dräger CC-Vision.
channel with Dräger CC-Vision.
525
The span calibration for the
compensation channel is not
valid
526
Error during warm-up acceler- Disconnect and reconnect
ation Dräger Sensor XXS
power pack or replace the senEC3
sor. Sensor must not be loaded
with gas within the first 5 minutes.
601
The zero point calibration of Carry out fresh air calibration,
DrägerSensor XXS EC4 is not page 39.
valid
602
The span calibration of
Carry out span calibration,
DrägerSensor XXS EC4 is not page 42.
valid
603
The measured value of
DrägerSensor XXS EC4 is in
the negative range
604
DrägerSensor XXS EC4 is not Check DrägerSensor XXS EC4,
inserted or defective
page 51.
605
Error during the function test
with gas (bump test) of
DrägerSensor XXS EC4
Repeat function test. Calibrate
or replace DrägerSensor XXS
EC 4, if necessary page 51.
606
Failed filter test.
Repeat filter test with X-dock.
Carry out fresh air calibration,
page 39.
Carry out span calibration for
compensation electrode.
Carry out fresh air calibration,
page 39.
31
Maintenance
607
Failed rise time test.
Repeat rise time test with Xdock.
608
User parameters not feasible
Check configuration of user
parameters and adjust
624
Device incorrectly configured Change sensor for applicable
by Dräger CC-Vision.
channel with Dräger CC-Vision.
626
Error during warm-up acceler- Disconnect and reconnect
ation Dräger Sensor XXS
power pack or replace the senEC4
sor. Sensor must not be loaded
with gas within the first 5 minutes.
Maintenance
Maintenance intervals
The device should be inspected and maintained by suitably qualified persons annually.
Comparisons:
 EN 60079-29-2 – Gas detectors - Selection, installation, use and maintenance of detectors for flammable gases and oxygen
 EN 45544-4 – Electrical apparatus used for the direct detection and direct concentration measurement of toxic gases and vapours - Part 4: Guide for selection, installation, use and maintenance
 National regulations
Recommended calibration interval for measuring channels Ex, O2, H2S, SO2, NO2 and
CO: 6 months.
Calibration interval of other gases: refer to the Instructions for Use of the respective
DrägerSensors.

Depending on device configuration:
Replace the alkaline batteries or charge the battery – refer to page 47 to page 48 –
after each use, at the latest after the battery alarm has been triggered or after 2
weeks.


Calibrating the device – page 38.
In regular intervals, according to the sensors used and the operating conditions. For
sensor-specific calibration data, refer to the Instructions for Use/data sheets of the
sensors used1).
Before you carry out safety-related relevant measurements, the zero point and sensitivity of the devices should be tested in accordance with national regulations.



Inspection by suitably qualified persons – every year.
The inspection intervals must be established in each individual case and shortened
if necessary, depending on technical safety considerations, engineering conditions
1)
Instructions for Use/data sheets for DrägerSensors can be downloaded on the product page of the X-am
5000 at the following Internet address: www.draeger.com. See also the enclosed Instructions for Use and
data sheets for the sensors used.
32
Maintenance


and the technical requirements of the equipment.
We recommend that a service agreement be concluded with Dräger and that repairs
also be carried out by them.
Replace the sensors, page 51 – if necessary, when it is not possible to calibrate the
sensors anymore.
Carry out manual bump test
Manual implementation without the documentation of results in the device
memory
 Prepare a test gas cylinder, the vol0.5 L/min
ume flow must be 0.5 L/min and the
gas concentration must be higher
than the alarm setpoint concentration
to be tested.
Example test gas cylinder 68 11 130
= mixed gas with 50 ppm CO, 15 ppm
H2S, 2.5 vol. % CH4, 18 vol. % O2
 Connect the test gas cylinder with the
calibration cradle (83 18 752).
 Vent the test gas into a fume cupboard or into the open air (with a hose
connected to the second connector of
the calibration cradle).
0
00523999_01_en.eps
CAUTION
Do not inhale the test gas. Risk to
health! Observe the hazard warnings of
the relevant Safety Data Sheets.
 Switch on the device and insert it into
the calibration cradle – press downwards until it engages.
 Open the test gas cylinder valve to let test gas flow over the sensors.
 Recommendation: Wait until the device displays the test gas concentration with sufficient tolerance –
Ex: ±20% of the test gas concentration 1)
O2: ±0.6 vol. % 1)
TOX: ±20% of the test gas concentration 1).
Wait al least until the alarm setpoint A1 or A2 is exceeded.
 If the alarm setpoints are exceeded, the device displays the gas concentration in alternation with » A1 « or » A2 « depending on the test gas concentration.
 Close the test gas cylinder valve and remove the device from the calibration cradle
 If the concentration has now fallen under the A1 alarm setpoint:
 Acknowledge the alarm.
 If the displays are outside of the above-mentioned ranges:
 Calibrating the device, refer to page 38.
1)
Upon application of the Dräger mixed gas (order no. 68 11 130) the displays should be within this
range.
33
Maintenance
Menu implementation with the documentation of results in the device memory
The "Quick bump test" or the "Extended bump test" is selected using the Dräger CC
Vision PC software. The "Quick bump test" checks whether the gas concentration has
exceeded the Alarm 1 threshold (with oxygen, the check is whether the concentration
has fallen below the Alarm 1 threshold). In the case of the “Extended bump test”, a
check is made as to whether the gas concentration has reached the set bump test concentration within a tolerance window.
Setting on delivery: Extended bump test.

Prepare a test gas cylinder, the volume flow must be 0.5 L/min and the gas concentration must be higher than the alarm setpoint concentration to be tested.
Example test gas cylinder 68 11 130 = mixed gas with 50 ppm CO, 15 ppm H2S, 2.5
vol. % CH4, 18 vol. % O2

Connect the test gas cylinder with the
calibration cradle (83 18 752).
Vent the test gas into a fume cupboard or into the open air (with a hose
connected to the second connector of
the calibration cradle).

0.5 L/min
CAUTION




34
The current gas concentration values
and the special symbol »
« (for
bump test) flash.
Press the OK key to start the function
test with gas.
Open the test gas cylinder valve to let
test gas flow over the sensor.
If gas concentration exceeds the
alarm thresholds A 1 or A 2 the corresponding alarm will occur.
Exit the function test with gas:
00523999_01_en.eps
0
ch4
%LEL
O2
Vol%
CO
ppm
H 2S
ppm
02423999_01_en.eps
Do not inhale the test gas. Risk to
health!
Observe the hazard warnings of the relevant Safety Data Sheets.
 Switch on the device and insert it into
the calibration cradle – press downwards until it engages.
 Call the quick menu and select the
function test with gas (bump test),
page 16.
Maintenance
After the preset bump test concentration
is reached or a gas alarm is triggered
(with the "Quick bump test"):
 The display containing the current
gas concentration changes with the
display » OK «.
 The bump test that was carried out is
documented with the result and date
in the device memory.




Close the test gas cylinder valve and
remove the device from the calibration cradle.
If the concentration values have now
fallen under the A1 alarm setpoints,
the device returns to the measuring
mode.
02523999_01_en.eps

ch4
%LEL
O2
Vol%
CO
ppm
H 2S
ppm
If the set bump test concentration is not reached within the specified time, the alarm
mode is activated to indicate failure.
The fault message »
« appears
and »
« is displayed instead of
ch4
the measured value on the faulty
%LEL
measuring channel.
O2
In this case, repeat the function test
Vol%
with gas or calibrate the device,
CO
ppm
page 38.
The function test with gas can also be
carried out automatically.
The "Bump Test Station" is required for
this function, refer to page 35.
02623999_01_en.eps
H 2S
ppm
Automatic implementation with the Bump Test Station
Prerequisite:
The device must first be configured for the automatic function test with gas (bump test)
using the PC software Dräger CC Vision.
 Activating the device for the automatic function test.
 Composition of test gas (mixed gas) – standard on delivery: 50 ppm CO, 15 ppm
H2S, 2.5 vol. % CH4, 18 vol. % O2
 Define which measuring channels should participate in the automatic function test.
All measuring channels participate in the function test by default.

Prepare the Bump Test Station according to the instructions.

Switch on the device and insert it into the receptacle of the Bump Test Station until
it engages.
35
Maintenance
Ex
%UEG
O2
Vol%
CO
ppm
H2 S
ppm
03823999_01.eps
0
The function test with gas is started
automatically. The special symbol
»
« (for bump test) flashes.

If a gas alarm (Quick bump test) is triggered and the preset bump test concentration (Extended bump test) is
reached within the specified time, the
display shows the current gas concentration, alternating with » OK «.
ch4
%LEL
O2
Vol%
CO
ppm
H 2S
ppm


Remove the device from the Bump Test Station.
If the concentration values have now fallen under the A1 alarm setpoints, the device
returns to the measuring mode.

An error will be triggered if the preset bump test concentration is not reached within
the specified time.
36
02523999_01_en.eps

Maintenance

ch4
%LEL
O2
Vol%
CO
ppm
H 2S
ppm
02623999_01_en.eps

The fault message »
« appears
and »
« is displayed instead of
the measured value on the faulty
measuring channel.
In this case, repeat the function test
with gas or calibrate the device,
page 38.
The function test with gas can also be carried out manually, refer to page 33 and
page 34.
The PC program Dräger CC Vision can be used to enable the "Automatic calibration
after incorrect bump test" option.
37
Maintenance
Calibrating the Device
Calibration may not be possible due to device and channel errors.
Allow the sensors to warm up before the calibration!
Warm-up time: see Instructions for Use/data sheets of the installed DrägerSensors (at
www.draeger.com).
Calibration interval:
 Observe the relevant specifications in the Instructions for Use/data sheets of the
DrägerSensors installed.
 For critical applications, observe the recommendations in EN 60079-29-21) or EN
45544-42) and national regulations. We recommend that you calibrate the channels
after 6 months.
CAUTION
Do not inhale the test gas. Risk to health!
Observe the hazard warnings of the relevant Safety Data Sheets.



1)
2)
38
Improve the zero point accuracy – carry out the fresh air calibration, page 39.
Set the sensitivity of all sensors to the value of the test gas – carry out the 1-button
calibration, page 40.
Set the sensitivity of a sensor to the value of the test gas – calibrate the sensitivity,
page 42.
EN 60079-29-2 – Gas measuring device - Selection, installation, use and maintenance of apparatus
for the measurement of combustible gases or oxygen
EN 45544-4 – Electrical devices for the direct detection and direct concentration measurement of toxic
gases and vapors – Part 4: Guidelines for selection, installation, use and maintenance.
Maintenance
Carrying out the fresh air calibration
To improve the zero point accuracy, you can carry out a fresh air calibration.
NOTICE
If none of the sensors fitted permits calibration with fresh air (e.g. only O3, only IRCO2), fresh air calibration is not offered as a menu function.


Calibrate the device to fresh air, free of measured gases or other interfering gases.
Not all sensors are included in the fresh air calibration1). Sensors which have not
warmed up or which are faulty prevent a calibration.
In the case of sensors which are in the warm-up phase, the message » 159 « is displayed with the special symbol »
« (for warning message).
In the case of a sensor or device error, the message » 109 « is displayed with the
special symbol »
« (for a fault message).
The message is cleared after 5 seconds and the function is available again in the
menu
During the fresh air calibration, the zero point of all sensors (with the exception of
DrägerSensor XXS O2) is set to 0.
In the case of DrägerSensor XXS O2, the display is set to 20.9 vol. %.
Switch on the device.
Depending on device configuration:
 Call the quick menu and select the
Fresh Air Calibration function,
page 16.
or
 Call the calibration menu and select
the Fresh Air Calibration function,
page 17.


1)
The current gas concentration values
flash.
When the measured values have stabilized:
Press the OK key to perform the fresh
air calibration.
ch4
%LEL
O2
Vol%
CO
ppm
H 2S
ppm
NH3
ppm
03123999_01_en.eps


Fresh air calibration / zero point adjustment is not supported by the DrägerSensor XXS O3. A zero
point calibration / adjustment of these sensors can be conducted using the Dräger CC-Vision PC software. To do so, a suitable zero gas that is free of ozone (e.g. N2) should be used.
39
Maintenance

ch4
%LEL
O2
Vol%
CO
ppm
H 2S
ppm
NH3
ppm
03223999_01_en.eps

The display containing the current
gas concentration changes with the
display » OK «.
Press the OK key to confirm the calibration or wait for approx. 5 seconds.


ch4
%LEL
O2
Vol%
CO
ppm
H 2S
ppm
NH3
ppm
03323999_01_en.eps

If a fault occurred during the fresh air
calibration.
The fault message »
« appears
and »
« is displayed for the respective sensor instead of the measured value.
In this case, repeat the fresh air calibration.
If necessary, replace the sensor,
page 51.
Carrying out 1-button calibration
NOTICE
If no sensors are enabled for 1-button calibration by the Dräger CC Vision PC program, the 1-button calibration menu function will not be offered.


40
All sensors that are enabled by the Dräger CC Vision PC program take part in the 1button calibration.
In the case of the 1-button calibration, the sensitivity of all sensors is set to the value
of the test gas.
When using the test gas cylinder 68 11 130 = mixed gas with 50 ppm CO, 15 ppm
H2S, 2.5 vol. % CH4,
18 vol. % O2.
Maintenance

CAUTION
Do not inhale the test gas. Risk to
health!
Observe the hazard warnings of the relevant Safety Data Sheets.








0.5 L/min
CH4
%LEL
O2
Vol%
CO
ppm
H 2S
ppm
0
02723999_01en.eps

If a mixed gas with another composition is used, the specified concentration values in the device must be
changed to the target values of the
mixed gas used using the PC software Dräger CC Vision.
Connect the test gas cylinder with the
calibration cradle.
Vent the test gas into a fume cupboard or into the open air (with a hose
connected to the second connector of
the calibration cradle).
Switch on the device and insert it into
the calibration cradle until it engages.
ch4
Call the calibration menu, enter the
%LEL
password and select the 1-button
O2
calibration function, page 17.
Vol%
OK
Press the
key to start the 1-button
CO
calibration.
ppm
Open the test gas cylinder valve to let
H 2S
ppm
test gas flow over the sensor.
The currently displayed measured
values start to flash.
The flashing stops after a static
measured value has been reached.
The calibration is now carried out automatically.
The displayed measured values
change to the values according to the gas supplied.
The automatic stability monitoring can be overridden by pressing the OK key. A calibration then takes place immediately. If it is detected that no test gas has been applied, the 1-button calibration will be aborted. The channels will then indicate » n/
a «. If only one sensor is taking part in the 1-button calibration, a calibration will be
performed in each case when the OK key is pressed.
02823999_01_en.eps

41
Maintenance





02923999_01_en.eps

ch4
%LEL
O2
Vol%
CO
ppm
H 2S
ppm
Close the test gas cylinder valve and
remove the device from the calibration cradle.
If a fault occurred during the 1-button
calibration.
The fault message »
« appears
and »
« is displayed for the respective sensor instead of the measured value.
In this case, repeat the 1-button calibration or carry out a single gas calibration, refer to page 42.
If necessary, replace the sensor,
page 51.
ch4
%LEL
O2
Vol%
CO
ppm
H 2S
ppm
03023999_01_en.eps

When the calibration is completed
and the displayed measured values
have stabilized:
The display containing the current
gas concentration changes with the
display » OK «.
Press the OK key or wait for 5 seconds to quit the calibration.
The device changes to the measuring
mode
Calibrating the sensitivity for an individual measuring channel
 The span calibration can be carried out specifically for individual sensors.
 In the case of the span calibration, the sensitivity of the selected sensor is set to the
value of the test gas used.
42
Maintenance

Connect the test gas cylinder with the
calibration cradle.
Vent the test gas into a fume cupboard or into the open air (with a hose
connected to the second connector of
the calibration cradle).

0.5 L/min
X-am 5000
CH4
%LEL
O2
Vol%
CO
ppm
H2S
ppm
0
CAUTION
Do not inhale the test gas. Risk to
health!
Observe the hazard warnings of the relevant Safety Data Sheets.

Switch on the device and insert it into the calibration cradle.

Press and hold the [ + ] key for 5 seconds to call the calibration menu, enter the
password and select the single gas calibration function, page 17.

Press the OK key to start the channel
selection.
The display flashes the gas of the first
measuring channel, e.g., » Ex - %LEL
«.
ch4
%LEL
CO2
ppm
Press the OK key to carry out the calibration of this measuring channel, or
Use the
key to select another
measuring channel (O2 - vol. %, H2S
- ppm or CO - ppm).
O2
Vol%
CO
ppm
H 2S
ppm
NH3
ppm



02723999_01_en.eps
Use a standard test gas.
Allowed test gas concentration:
Ex: 40 to 100 %LEL
O2 10 to 25 vol. %
CO: 20 to 999 ppm
H2S: 5 to 99 ppm
Test gas concentration of other gases:
refer to the Instructions for Use of the
respective DrägerSensors.
03523999_01_en.eps

43
Maintenance
O2
Vol%
CO
ppm
H 2S
ppm
NH3
ppm

ch4
%LEL
03623999_01_en.eps
If the displayed measurement value is
stable:
 Press the OK key to perform the
calibration.
 The display containing the current
gas concentration changes with the
display » OK «.
 Press the OK key or wait for approx. 5
seconds to end the calibration of this
measuring channel.
05823999_en.eps
Span calibration for CatEx
 If the measuring range end value < = 100%LEL, the adjustment for the heat tinting
is offered.

Display in the case of channel selection:
 Open the test gas cylinder valve to let
ch4
test gas flow over the sensor.
%LEL
 Press the OK key to start the calibraCO2
tion for the heat tinting or press the
ppm
-key to select the next sensor.

The next measuring channel is
offered for calibration.
After the calibration of the last
measuring channel, the device
changes to measuring mode.

Close the test gas cylinder valve and remove the device from the calibration cradle.
44
Maintenance
If the gas CH4 (measurement range
up to 100 vol. %) is selected on the
CatEx channel, the heat conductivity
is calibrated, display with
channelselection:

Press the OK key to perform the
calibration for heat conductivity.

Press the
sensor.

Press the OK key to perform the
calibration of the selected measuring
channel.
The test gas concentration is
displayed.
Press the OK key to confirm the test
gas concentration or use the [ + ] key
to change the test gas concentration
and complete the process by
pressing the OK key.
The measurement value flashes.






03723999_01_en.eps
O2
Vol%
CO
ppm
H 2S
ppm
NH3
ppm
CH4
VOL%
CO2
ppm
O2
Vol%
CO
ppm
H 2S
ppm
NH3
ppm
key to select the next
O2
VOL%
O2
Vol%
CO
ppm
H 2S
ppm
NH3
ppm
05223999_01_de.eps

ch4
%LEL
CO2
ppm
05123999_01_en.eps
If a fault occurred during the span
calibration.
 The fault message »
« appears
and »
« is displayed for the
sensor instead of the measured
value.
 In this case, repeat the calibration.
If necessary, replace the sensor,
page 51.
Open the test gas cylinder valve to let
test gas flow over the sensor.
The displayed, flashing measurement value changes to the value according to the
supplied test gas.
Press the
key to select the next sensor.
45
Maintenance
Note on adjusting the Ex channel to nonane as the measured gas:
 When calibrating the Ex channel, propane can be used as a substitute for the
calibration gas.
 When using propane to adjust the Ex channel to nonane, the display must be set to
2x the test gas concentration used.
Note on the use in underground mining:
 When calibrating the Ex channel to methane as the measured gas, the indication on
the instrument must be set to a value 5 % (relative) higher than the test gas concentration used.
46
Maintenance
Replacing the batteries / rechargeable batteries
WARNING
Explosion hazard! To reduce the risk of flammable or explosive atmospheres igniting,
it is essential that the warning notices below are observed:
Do not throw used batteries into fire or try to open them by force.
Do not replace or charge batteries in areas at risk of an explosion hazard.
Do not mix new batteries with used batteries, and do not mix batteries from different
manufacturers or of different types.
Remove batteries before maintenance work.
Batteries / rechargeable batteries are part of the Ex approval.
Only the following types may be used:
– Alkaline batteries – T3 – (non rechargeable!)
Panasonic LR6 Powerline,
Varta Type 41061) (power one) or
Varta Type 40061) (industrial)
– Alkaline batteries – T4 – (non rechargeable!)
Duracell Procell MN15001)
– NiMH rechargeable batteries – T3 – (rechargeable)
GP 180AAHC1) (1800) max. 40 °C ambient temperature.
1) Not subject to the Metrological Performance Test BVS10 ATEX E 080X and PFG 10 G 001X.

Insert the power pack into the device
and tighten the screw, the device
switches on automatically.
2
3
00623999_01.eps
Completely replace the power pack
T4 (with sealed rechargeable batteries, order no. 83 18 704).
–
+
3
+
–
Switching off the device:
 Press and hold the OK key and the
key at the same time.
1 Loosen the screw (2.0 mm hexagon
socket) on the power pack and remove the power pack.
2 Replace the alkaline batteries with
new ones or the rechargeable NiMH
batteries with charged ones – ensure
correct polarity.
1
After replacing the power pack T4, it is recommended that a complete charging is carried out.
After the batteries have been replaced:
 The settings and data are stored when the battery is replaced. The sensors warm
up again.
47
Maintenance
Charging the rechargeable batteries
WARNING
Explosion hazard! To reduce the risk of flammable or explosive atmospheres igniting,
it is essential that the warning notices below are observed:
Do not charge underground or in explosion-hazard areas! Danger of explosion!
The chargers are not designed in accordance with the regulations for firedamp and
explosion protection.
Charge the NiMH power pack T4 (HBT 0000 type) or T4 HC (HBT 0100 type) using
the corresponding Dräger charger. Ambient temperature during the charging process:
0 to +40 °C.
Even if the device is not used, we recommend that you store the device in the charger
(Charging module X-am 1/2/5000, order no. 83 18 639)!
To maintain the lifetime of the batteries, charging is temperature controlled and only
performed in a temperature range of 5 to 35 °C.
When this temperature range is left, the charging process is automatically interrupted
and automatically continued after the temperature range has been reached again.
The charging time is typically 4 hours.
A new NiMH power pack reaches its full capacity after three complete charging/discharging cycles. Never store the device for extended periods without being connected
to a power source (maximum of 2 months) because the internal buffer battery will drain.
Charging with the multiple charging station
 A maximum of 20 devices can be charged at the same time on the power pack (order
no. 83 18 805) of the multiple charging station.
 When attaching the charging modules, disconnect the power pack from the mains
supply!
Attaching charging modules
1 Turn the slots of the interlock into a horizontal position by using a screwdriver or
coin.
2 Insert the projecting tongue of the charging module (at the same time, current entry)
until it engages.
1 Close the interlock with a quarter turn (slot is positioned vertically).
48
Maintenance
1
2
X-am 5000
1
2
X-am 5000
1
2
X-am 5000
Ex
%LEL
CO2
ppm
Ex
%LEL
CO2
ppm
Ex
%LEL
CO2
ppm
O2
Vol%
CO
ppm
H2S
ppm
NH3
ppm
O2
Vol%
CO
ppm
H2S
ppm
NH3
ppm
O2
Vol%
CO
ppm
H2S
ppm
NH3
ppm
0
0
00723999_01_en.eps
0
1


Attach additional charging modules in the same way.
Always connect or disconnect the charging modules individually and not in groups
in order to prevent the charging station from becoming damaged. During transportation, the power pack and the charging modules should also always be handled individually and without inserted devices.

Position the device on an even and level surface.

Connecting the power pack to the
mains.
The green "Mains" LED lights.
Insert the device into the charging
module.
Display LED on the charger:
Charge
Fault
Full
2
If a fault occurs:
Remove the device from the charging
module and insert it again.
If the fault still occurs, have the charging
module repaired.
X-am 5000
Ex
%LEL
CO2
ppm
O2
Vol%
CO
ppm
H2S
ppm
NH3
ppm
0
1 3
2
00823999_01_en.eps
1

It takes approx. 4 hours to fully charge
an empty rechargeable battery.
A short circuit of the charging contacts in the charging modules, e.g., by metallic
objects that have fallen in, does not result in damage to the charging station. It should,
however, be avoided due to possible heating hazards and incorrect displays on the
charging module.
In the event of a short circuit or if the power pack is overloaded:
3 The red "Overload" LED lights, and an audible alarm sounds.
 After the fault has been corrected, the alarm is switched off automatically and the
charging process is restarted.
49
Maintenance

In the event of a power failure, the devices already charged will be protected from
discharging.
Charging with charging module and plug-in power pack or vehicle charging
adapter
 When using the power pack (order no. 83 16 994), up to 5 devices can be charged
at the same time, with power pack (order no. 83 15 635) up to 2 devices.
 The power pack contained in the rechargeable battery and charging set (order no.
83 18 785) is suitable for charging a device.
 When using the vehicle charging adapter (order no. 45 30 057) it is recommended
that you supply every charging module separately.
2
X-am 5000
1
Ex
%LEL
CO2
ppm
O2
Vol%
CO
ppm
H2S
ppm
NH3
ppm
83 16 994 (100 ... 240 V)
or
83 15 635 (100 ... 240 V)
83 17 754
The charging process is carried out analog to charging with the multiple charging station.
50
00923999_01_en.eps
0
Replacing the Sensors
Replacing the Sensors
CAUTION
Damage to components!
There are components in the instrument that are sensitive to electric charge. Before
opening the instrument to replace the sensor, ensure that the person performing the
work is earthed to avoid damage to the device. Earthing can be safely ensured, e. g.
via an ESD workstation (electrostatic discharge).
To replace the sensors of the device, connect the device with a PC.
Replace the sensors using the PC program Dräger CC Vision.
USB 2.0
IR
0
Dräger
CC-Vision
X-am 5000
USB DIRA with USB cable
(order no. 83 17 409)
USB 2.0
0
Dräger
CC-Vision
Calibration cradle (order no. 83 18 752)
with inserted
USB DIRA with USB cable (order no. 83 17 409)
USB 2.0 / COM
0
Dräger
CC-Vision
E-Cal
01423999_01_en.eps


E-Cal module Dräger X-am 125
(order no. 83 18 754)
51
Replacing the Sensors
Next:
 Carry out a fresh air calibration, page 39.
and then:
 Calibrate the sensitivity:
either
carry out 1-button calibration, page 40
or
carry out span calibration, page 42.
52
Replacing the Sensors
Electrochemical sensors
WARNING
Do not throw into fire,
Do not force open. Danger! Acid burn risk!
Sensors of type XXS O3 and XXS NO2 LC contain small quantities of nanomaterials.
Like batteries, only dispose of as special waste,
in accordance with local waste disposal regulations. Further information can be
obtained from the relevant local authority and from appropriate waste disposal
companies.
The DrägerSensor CatEx 125 PR should be disposed of as electronic waste.
53
Sensor warm-up acceleration
Sensor warm-up acceleration
There is a function available for accelerating the warm-up procedure for selected EC
sensors. The function shortens the time taken until the unit is ready to make measurements, i.e. the time taken to save the display and alarm evaluation of the measurements. The time to activate the calibration is not changed.


In order to use the sensor warm-up acceleration function in the Dräger X-am 5000,
at least one sensor suitable for the purpose must be fitted and registered in the unit
using the Dräger CC-Vision PC software.
After that, activate the function by ticking the "Accelerated warm-up" checkbox ("Device" menu) and updating the device data with Dräger CC-Vision.
NOTICE
A selection menu for activating the process will only be displayed after the device is
restarted by plugging in a power pack.
54
5:00
120:00
001
05623999_01.eps
Restart the device by plugging in a
power pack.
 All the display segments, including the
visual, audible and vibration alarms,
are activated for a short time.
 The software version is displayed.
 The instrument performs a self test.
 The sensor warm-up acceleration
menu is displayed.
The steps in the selection menu are
limited by a 25-second time-out, which
then changes to the standard setting
of unaccelerated warm-up and proceeds with the device start-up procedure.
 It is necessary to select between two
times for the sensor warm-up acceleration,
— the shortened time achievable as a result of the warm-up acceleration,
— the unshortened time required by the
installed sensors to save the measurement display (measurement does not
flash any more).
 Select the shortened warm-up time
with the
button and confirm it with
the OK button.
 The function for entering the password
is called up.
 Enter the digits with the
button and
confirm with the OK button.
05523999_01.eps

Sensor warm-up acceleration


Start of the warm-up acceleration.
The remaining time is displayed and
decremented.
WARNING
04:08
05723999_01.eps
This function requires that the device is
in a gas-free environment. A warning
screen is displayed and the remaining
time during which no gas may be
applied is shown It is essential to
ensure that the sensor is in the fresh air
during this time (for approx. 5 minutes)!
Any gas applied during this period could
later result in faulty alarms or suppressed alarms!
If the procedure is unsuccessful, e.g. as a result of a defective sensor, this will be indicated by error code » X07 « on the channel for the affected sensor (X). In the event of
a fault, the device will also switch to the unaccelerated mode.
Dräger recommends that the device not be switched off and on again for a period of
two hours when the accelerated warm-up is in use.
55
Cleaning
Cleaning
The device does not need any special care.
 Dirt and deposits can be removed from the device by washing it with cold water. A
sponge can be used for wiping if necessary.
NOTICE
Abrasive cleaning implements (brushes etc.), cleaning agents and cleaning solvents
can destroy the dust and water filters.

Carefully dab dry the device using a cloth.

Dräger recommends storing the instrument in the charger module (order no.
83 18 639).
Dräger recommends checking the charge of the power supply at least every
three weeks if the instrument is not stored in the charger module.
Storage

Disposal
This product must not be disposed of as municipal waste. This is indicated by
the adjacent icon.
You can return this product to Dräger free of charge. For information please
contact the national sales organisations and Dräger.
Batteries and rechargeable batteries must not be disposed of as municipal
waste. This is indicated by the adjacent icon. Collect and dispose of batteries
and rechargeable batteries at battery collection centres, in accordance with
applicable regulations.
56
Technical Data
Technical Data
X-am 5000
Ambient conditions:
during operation and
storage
–20 to +50 °C for NiMH power packs type:
HBT 0000 and HBT 0100,
with alkali single cells type: Duracell Procell MN 15001)
-20 to +40 °C for NiMH single cells type:
GP 180AAHC1) and for alkali single cells type:
Panasonic LR6 Powerline
0 to +40 °C for alkali single cells type:
Varta 40061), Varta 41061),
700 to 1300 hPa
10 to 90% (to 95% briefly) relative humidity
Device data
Protection class
IP 67 for devices with sensors
Alarm volume
Typically 90 dB (A) in 30 cm distance
Operation time
– Alkaline battery
Typically 12 hours under normal conditions
– NiMH
power pack:
T4 (HBT 0000)
T4 HC (HBT 0100)
Typically 12 hours under normal conditions
Typically 13 hours under normal conditions
Dimensions
approx. 130 mm x 48 mm x 44 mm (H x W x D)
Weight
approx. 220 g to 250 g
Storage time
X-am 5000
Sensors
1 year
1 year
Position of use
any
Refresh interval for dis- 1 s
play and signals
CE markings
Electromagnetic compatibility (Directive 2004/108/EC)
Explosion protection (Directive 94/9/EC)
Approvals2):
see “Tests and Approvals” on page 5.
The BVS 10 ATEX E 080 X technical suitability test is based
on the adjustment with the target gas.
1) Not subject to the Metrological Performance Test BVS10ATEXE080X and PFG10G001X.
2) The technical approvals are valid for the X-am 5000 gas detection instrument and the calibration cradle. The explosion-protection
approvals are only valid for the X-am 5000 gas detection instrument; the calibration cradle must not be used in the Ex zone.
57
58
Ex
EN 60079-29-1 7)
EN 50271
Possible poisoning
5 %LEL
35 seconds
1 %LEL/
8 hours
7 seconds for methane
40 seconds for nonane2)
0 to 100 %LEL3)
0 to 100 vol. % for methane
–––
–––
17 seconds for methane
25 seconds for propane
Catalytic oxidation
EN 50104 8)
(measurement of
oxygen deficiency
and oxygen
surplus)
EN 50271
0.3 vol. %
5 minutes
–––
–––
–––
0 to 25 vol. %
2 %
of the
measurement
value
EN 45544-1/-2 9)
EN 50271
0 to 200 ppm
H2S 4)
2 ppm
1 % of the
measurement
value/month
5 minutes
–––
6 seconds
6 seconds
XXS H2S
Electrochemical
15 seconds
XXS O2
Electrochemical
10 seconds
XXS H2S LC
2 %
of the
measurement
value
EN 45544-1/-2 9)
EN 50271
0 to 100 ppm
H2S 5)
0.4 ppm
1 % of the
measurement
value/month
5 minutes
–––
6 seconds
Electrochemical
18 seconds
XXS CO
3 %
of the
measurement
value
EN 45544-1/-2 10)
EN 50271
0 to 2000 ppm
CO 6)
6 ppm
1 % of the
measurement
value/month
5 minutes
–––
6 seconds
Electrochemical
25 seconds
1) The Technical Manual and Instructions for Use/data sheets for the sensors used and the PC software CC-Vision for Dräger X-am 5000 can be downloaded on the product page for the X-am 5000 at the following
website: www.draeger.com See also the enclosed Instructions for Use and data sheets for the sensors used.
2) For decreasing concentrations, the adjustment time for nonane is about 50 seconds.
3) For alkanes from methane to nonane, LEL-values in accordance with EN 60079-20-1 For flow speeds of 0 to 6 m/s the deviation of the display can be 5 - 10 % of the measurement value. When calibrating for
propane, the deviation of the display in the range of 80 to 120 kPa can be up to 6 %LEL.
4) Certified for 1 to 100 ppm
5) Certified for 0.4 to 100 ppm
6) Certified for 3 to 500 ppm
7) The device responds to most combustible gases and vapours. The sensitivities differ depending on the type of gas. We recommend a calibration using the target gas to be measured. For the range of alkanes, the
sensitivity decreases from methane to nonane.
8) The measuring signals can be affected negatively by ethane, ethene, ethyne, carbon dioxide and hydrogen. No O2 measurement in helium.
9) The measuring signals can be affected additively by sulphur dioxide, nitrogen dioxide and hydrogen, and negatively by chlorine.
10) The measuring signals can be affected additively by acetylene, hydrogen and nitrogen monoxide.
Standards
(Measuring function for explosion protection and
measurement of oxygen deficiency and surplus as
well as toxic gases, DEKRA EXAM, Essen, Germany:
BVS 10 ATEX E 080 X3), PFG 10 G 001 X)
Linearity error
Warm-up time
Effect of sensor poisons
Hydrogen sulphide H2S, 10 ppm
Hydrogen sulphides, heavy metals, gases containing
silicone, sulphur or polymerizable substances
Zero point deviations (DIN EN 45544)
Device drift
Measuring range
Measurement value configuration time t0...50
Measuring principle
Measurement value configuration time t0...90
Extract: For details, see the Instructions for Use/data sheets for the sensors used1)
Sensor data
Order List
Order List
Name and Description
Order No.
Dräger X-am 5000
83 20 000
Unlimited 1 to 5 multi gas monitor with exchangeable sensors.
With selectable special calibration.
Standard calibration of the Ex sensor: Methane.
Inclusive to country-specific adjustable standard alarm setpoints.
Power supply units:
NiMH supply unit T4
83 18 704
NiMH supply unit HBT 0100 T4 HC
1)
83 22 244
Battery holder ABT 0100 (without alkaline batteries)
83 22 237
Alkaline batteries T3 (2 pcs)1)
83 22 239
1)
Alkaline batteries T4 (2 pcs)
83 22 240
Battery and charger kit (contains NiMH supply unit T4, charger
module for Dräger X-am 1/2/5000 and power supply unit)
83 18 785
HC battery and charger kit
(contains NiMH supply unit HBT 0100 T4 HC, charger module
for Dräger X-am 1/2/5000 and power supply unit)
83 22 785
Chargers:
Charging module for Dräger X-am 1/2/5000
83 18 639
Power pack with connecting cord (worldwide)
for a maximum of 20 charging modules Dräger X-am 1/2/5000
83 15 805
Plug-in power pack (worldwide)
for a maximum of 5 charging modules Dräger X-am 1/2/5000
83 16 994
Plug-in power pack (worldwide)
for a maximum of 2 charging modules Dräger X-am 1/2/5000
83 15 635
Vehicle connecting line 12V/24V
for Dräger X-am 1/2/5000 charging module
45 30 057
Vehicle installation set
for 1 Dräger X-am 1/2/5000 charging module
83 18 779
59
Order List
Name and Description
Order No.
Accessories
The accessories are not subject to BVS10 ATEX E 080X and
PFG 10 G 001X.
Pump accessories:
Dräger Pump X-am 1/2/5000
83 19 400
Case for Dräger Pump X-am 1/2/5000
83 19 385
Rubber ball pump
68 01 933
Manual pump adapter
83 19 195
Dust and water filter
83 13 648
Extension hoses and probes:
Measuring probe 0.5 m
64 08 238
Measuring probe 1.5m
64 08 239
Plug-in telescopic probe
68 01 954
Telescopic probe 100 with accessories
83 16 530
Telescopic probe 150 stainless steel
83 16 533
Tester 90
83 16 532
Float probe with accessories
83 18 371
Viton hose
12 03 150
Hose
11 80 681
Accessories for measured value acquisition and
configuration:
Dräger GasVision (time limited test version available from
www.draeger.com/software)
Licence code for Dräger GasVision
(full version)
83 25 646
Dräger CC Vision (full version available from www.draeger.com/
software)
PC communication set 1 Dräger X-am 1/2/5000
with USB connection and Dräger CC Vision
83 18 761
PC communication set 2 Dräger X-am 1/2/5000
with USB connection, Dräger CC Vision and barcode reader
83 18 762
USB DIRA with USB cable (USB infrared adaptor for communication Dräger X-am 1/2/5000 – PC)
83 17 409
60
Order List
Name and Description
Order No.
Calibration accessories:
Dräger X-dock, e. g. X-dock 5300 X-am 125
83 21 880
Bump Test Station, including mixed gas cylinder
83 19 130
Calibration cradle Dräger X-am 1/2/5000
83 18 752
Mixed gas cylinder
2.5 vol. % CH4, 18 vol. % O2, 15 ppm H2S, 50 ppm CO
68 11 130
Test gas cylinder propane, 0.9 vol. % C3H8 in air
68 11 118
On demand controller
83 16 556
Standard controller
68 10 397
Other accessories:
Rubberboot Dräger X-am 1/2/5X00
83 21 506
Carrying case
83 18 755
Spare parts
DrägerSensor CatEx 125 PR, 0 to 100 %LEL
(0 to 100 vol. % methane)
68 11 050
DrägerSensor CatEx 125 PR, 0 to 100 %LEL
68 13 080
DrägerSensor XXS O2, 0 to 25 vol. %
2)
DrägerSensor XXS O2 100, 0 to 100 vol. %
68 10 881
68 12 385
1)
68 10 882
DrägerSensor XXS H2S, 0 to 200 ppm1)
68 10 883
DrägerSensor XXS CO, 0 to 2000 ppm
more DrägerSensors
1)
2)
3)
on request 3)
Not subject to the Metrological Performance Test BVS10ATEXE080X and PFG10G001X.
Expected service life of the sensors: O2, CO and H2S>5 years, CatEx > 3 years.
Data sheets for all sensors that can be used for the device can be downloaded on the product page of
the X-am 5000 at the following Internet address: www.draeger.com.
61
Declaration of Conformity
Declaration of Conformity
62
Declaration of Conformity
63
Dräger Safety AG & Co. KGaA
Revalstraße 1
D-23560 Lübeck
Germany
Phone +49 451 8 82- 0
Fax +49 451 8 82- 20 80
www.draeger.com
90 23 999 - TM 4638.210 en
© Dräger Safety AG & Co. KGaA
Edition 13 - December 2014
(Edition 01 - April 2007)
Subject to alteration

advertisement

Key Features

  • Continuous monitoring of multiple gases
  • Visual and audible alarms
  • Customizable settings
  • Data logging and graphical display
  • Bump test and calibration functions

Frequently Answers and Questions

What is the purpose of the bump test?
The bump test helps to ensure the device is functioning properly and detecting the target gas within the correct range. It involves exposing the sensor to a known concentration of the gas to check for a proper response.
How do I replace the batteries?
The battery replacement is outlined in the manual, and its recommended to replace batteries in safe environments.
What are the different types of alarms available?
The device has a pre-alarm (A1) and a main alarm (A2). The pre-alarm warns about a potential hazardous concentration, while the main alarm indicates a dangerous level that requires immediate action.

Related manuals

Download PDF

advertisement