- Industrial & lab equipment
- Measuring, testing & control
- Laser levels
- Gammex
- PROBE+ GLD-250
- User's Guide
- 40 Pages
Gammex PROBE+ GLD-250, GLD-450 PROBE+ Diode Laser User's Guide
Below you will find brief information for PROBE+ GLD-250, PROBE+ GLD-450. The PROBE+ Diode Lasers are intended to be used for aligning patients to medical diagnostic imaging and radiation therapy equipment which utilizes ionizing radiation. These lasers are meant to be used in medical clinic or hospital settings by trained medical staff. The lasers are not designed to be used around flammable anesthetics, which present a risk of explosion or fire. The lasers are not intended to be used in a sterile or corrosive environment.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
PROBE+ Diode Laser GLD-250 GLD-450 User’s Guide Table of Contents Introduction and Intended Use. . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 General Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Storage and Transportation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Enviromental Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Location of Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Operator Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Installation Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Simplified Room Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Pre-Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Laser Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1 Movement of Crosshair and Sagittal Lines . . . . . . . . . . . . . . . . . 1 4 Coarse Carriage Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 5 Fine Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 5 Additional Coarse Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 6 Line Adjustment Components . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 7 Fine Line Positioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 8 Curvature Adjustments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 9 Laser Focus Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 9 Single Line Adjustments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 0 Line Rotation Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1 Centering Optics over Laser . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1 Second Side Laser Installation . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1 Installation on Angled Walls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 4 Ceiling Laser Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 6 Sagittal Laser Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 8 Calibration and Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 9 Removing the Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 9 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 0 Ceiling and Sagittal Lasers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 0 Side Lasers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 0 Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 0 Troubleshooting, Repair and Maintenance . . . . . . . . . . . . . . . . . 3 1 Replacing Circuit Boad and Laser Diode Assembly . . . . . . . . . . 3 2 Dismantling/Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 3 Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 4 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 5 Product Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 6 Product Ordering Infomation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 7 NOTE: Installation Template was shipped with your patient positioning unit. For replacement template, please use Gammex part number: 007657 Installation and User Guide Dear Valued Customer Congratulations on your recent purchase of a Gammex Laser Alignment System! This diode laser product is the result of many years of continuous quality improvement of laser technologies. If, after carefully reading this manual, you have any questions, please contact our Sales and Service Department at 1-800-Gammex 1. Introduction This Installation and User’s Guide includes the information you will need to safely and efficiently install, operate and maintain your medical alignment laser. This guide is only intended for use with Gammex PROBE+ Diode Laser models GLD–450 (Crosshair) and GLD–250 (Sagittal). Intended Use These lasers are intended to be used for aligning patients to medical diagnostic imaging and radiation therapy equipment which utilizes ionizing radiation. They are intended to be used in medical clinic or hospital settings by trained medical staff. The lasers are not designed to be used around flammable anesthetics, which present a risk of explosion or fire. The lasers are not intended to be used in a sterile or corrosive environment. No other uses intended or implied. 1 Introduction PROBE+ Diode Laser Definitions Symbols Indicates user to consult accompanying document (safety) Class II equipment This symbol indicates that when the end-user wishes to discard this product, it must be sent to a separate collection facilities for recovery and recycling. General Safety Signal words are used according to international standards. The meanings of these signal words are: DANGER - Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury. This signal word is to be limited to the most extreme situations. WARNING - Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. CAUTION - Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices. Permitted for property damage only accidents. Laser Apertures Please Do Not Look Into Laser Introduction 2 Installation and User Guide LASER SAFETY WARNING - Risk of incorrect diagnosis or treatment. Isocenter of alignment lasers must coincide with isocenter of diagnostic or treatment equipment. Test alignment of lasers before use daily for first month of operation and weekly after first month. Newly installed equipment will shift as building walls and floor settle and as building materials dry out. Realign lasers to isocenter before use, if they have moved. CAUTION - Laser Light, risk of eye damage. • To prevent eye damage, do not stare into beam and do not allow patients to stare into beam. The normal blink reflex will prevent eye damage. If a patient is unable to blink normally the laser operator must prevent laser from shining in patient's eye. Laser is a class II laser. • Replace laser diode only with assembly provided by Gammex to ensure that laser power levels do not exceed class II limits. • Do not adjust sealed potentiometers as power levels could exceed class II limits and damage the laser diode. GENERAL • Sagittal laser - Single vertical line projecting laser • Crosshair laser - Crossed line projected from laser • Iso-center - Center point where the X, Y, and Z planes coincide. Storage and Transportation • Do not drop • Non-corrosive environment • Storage Temperature: -20 - +75° C (-4 -167° F) • Storage Humidity: 0 - 85% Relative Humidity (no dew condensation) Environmental Limits: • Temperature: 0-60° C (32 -140° F) • Humidity: 0 - 85 % relative humidity (no dew or condensation) • No flammable anesthetics • No corrosives 3 Introduction PROBE+ Diode Laser Location of Labels Safety labels are located in two places on the laser: 1. inside on the laser baseplate. 2. top outside of housing Location of Labels 4 Installation and User Guide Operator Controls A - Cover B - Red Mylar Window A B C - Power-On Indicator D - ON/OFF Switch (Push on/push off) • To turn the laser on: push the ON/OFF button (D). The Power-On indicator (C) will be lighted. • To turn the laser off: push the On/Off switch (D). The Power-On indicator goes dark. • Some facilities like to connect all the lasers to one wall switch so all lasers can be turned on and off from a C D convenient location. CAUTION - Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous raditation exposure. 002741 DECAL/CAUTION CLASS II DIODE LASER CAUTION LASER LIGHT DO NOT STARE INTO BEAM DIODE LASER 1.0 MILAWAT MAXIMUM OUTPUT 630-670nm WAVE LENGTH CLASS II LASER PRODUCT 5 Operator Controls PROBE+ Diode Laser Introduction GENERAL All Gammex PROBE+ Laser Alignment Systems provide an optical aid for rapid and accurate patient positioning and repositioning for therapy and simulator machines. TECHNICAL DESCRIPTION Various combinations of the GLD-450 crosshair PROBE+ and the GLD-250 sagittal (straight line) PROBE+ are used with different systems. The GLD-450 lasers, mounted on each side, beam crosshairs along the X-axis. The crosshair GLD-450 laser, mounted on the ceiling, beams a crosshair vertically along the Z-axis. The Sagittal GLD-250 laser beams a straight line in the Z-Y plane. Installation Requirements ELECTRICAL Each laser is provided with a universal power supply (transformer) with an input of 100 to 240 volts at 50 to 60 Hz. The universal power supply plugs into a wall outlet with interchangeable “snap-in” plugs to fit most outlets. A 1.2 meter (4 foot) power cord supplies 12 volt dc power to the laser with a screwon power plug TOOLS NEEDED In addition to some basic tools, such as screwdrivers, hammer and ladder; you may need an electric hammer drill, plumb bob, level (water or laser level), masking tape and 15M (50 feet) of string. (The mounting hardware, and 4 allen (hex) wrenches (.050”, 1/8”, 3/32”, & 5/64”) are included with the lasers.) Typical Linear Accelerator room Layout shown in Figure 1. CAUTION: The location of the lasers is determined in relation to the therapy machine’s crosshair and isocenter. Before installing lasers, ask the physicist to verify that the field light aligns with the machine’s isocenter. Pre-Installation 6 Installation and User Guide ° Figure 1 7 Example Room Layout PROBE+ Diode Laser Figure 2 Use a level to line up the laser exit reference point at the height of isocenter. Pre-Installation 8 Installation and User Guide Pre-Installation Use a level, see Fig. 2, to line up the laser exit reference point at the height of isocenter. Use a level to verify: PREPARING THE ROOM • Gantry is level each time it is rotated 1. Unpack each laser, add the cord and plug into a wall outlet. • Template is placed accurately on wall 3. Set gantry and collimator to 0° and level the collimator. • Crosshairs on wall are at 90° angles 2. Turn the linear accelerator lamp on. 4. Lower the table as low as possible. COLLIMATOR CROSSHAIR ACCURACY The collimator-projected crosshair is used as a reference when determining the laser locations. The light field should coincide with isocenter, the gantry should be level and, if collimator crosshair displacement exists, it should be compensated for. 1. Project a crosshair onto a piece of paper taped to the table and mark the center with a dot. 2. Rotate through 360° and mark the center again. 3. If there is a difference between the two dots, find the center of the distance between the two dots. Mark this point and use this point as the crosshair center reference point. Use this point to line up the laser. USING THE LEVEL Use a level, see Fig. 2, to determine the height of the laser exit reference point when installing lasers in a simulator room or in a room with angled walls. In standard treatment rooms, the level should also be used to verify the laser location. 1. Raise the table to the front pointer (isocenter). 2. Line up with the table edge and with the wall. Figure 2 3. The line on the wall indicates the horizontal axis of the laser exit reference point. 9 Pre-Installation PROBE+ Diode Laser Figure 3 Rotate the gantry 90° and project a crosshair onto the wall. Figure 4 Use a string to verify the location of the side lasers. Side Laser Installation 10 Installation and User Guide Installing Lasers SIDE LASER INSTALLATION For all GLD-450 installations. For angled walls, refer to section Installation on Angled Walls. Use a level to verify: • Gantry is level each time it is rotated • Template is placed accurately on wall • Crosshairs on wall are at 90° angles 1. Rotate the gantry 90° to the 270° position and project a crosshair onto the wall. Figure 3 2. Use a pencil and straight rule to trace the crosshair onto the wall. 3. Line up the mounting template from back of this manual, so that the projected crosshair lines up with the laser exit point on the mounting template. Use a level to line up the mounting template. 4. Mark the outer edge of the mounting template and also mark the mounting holes. 5. Rotate the gantry 180° to the 90° position and repeat steps 24 on the opposite wall. 6. Return the gantry to 0°. 7. Tape a piece of paper to the table. 8. Project a crosshair onto the paper, keep the field as wide as possible. 9. Hold the ends of a plumb bob on each wall until string shadow is in line with projected crosshair. Figure 4 10. The point at which the string hits the wall is the center of the laser exit point. Mark the center on the wall. Use the mounting template to verify the location of the mounting holes. 11. Drill the mounting holes. 12. Remove the PROBE+ Laser from its carton. Loosen the screw at the top of the cover. Slide the cover up, to disengage it, and lift the cover off. Remove any packing material from the optics. 13. Attach washers and screws with mounting spacers as shown in Figure 5. Mount the laser to the wall Figure 5. Laser must be firmly mounted. Oversized holes provide flexibility to line up the exit reference, Use suitable anchors. Wall anchors for masonry are provided in the mounting kit. Use a 3/8” diameter drill for the provided anchors. 11 Side Laser Installation PROBE+ Diode Laser B A E F Figure 5 Mount the base plate to the wall with the supplied mounting hardware. C D Use the template to indicate where on the wall to mount the laser. A: B: C: D: E: F: Side Laser Installation 12 Exit Reference Base Plate Laser Assembly End Plate Mounting Screws Washers Installation and User Guide Installing Lasers SIDE LASER INSTALLATION (continued) 14. Do not replace the cover at this time. 15. Plug in the power supply. Attach the threaded power plug to the bottom of the laser. Turn on the laser. The push switch will toggle on and off. The LED near the power switch should light and the laser lines project. 16. Rotate the gantry 90° (so that the crosshair is projected onto the laser) and tape a piece of paper to the gantry head, covering the projected light field. 17. Adjust the laser so that the laser crosshair lines up with the collimator cross hair. Refer to the following pages for the methods for making laser adjustments. 18. Place a piece of paper directly in front of the laser. 19. Verify that the center of the projected collimator crosshair is centered on the top laser lens. 20. Adjust for straight lines if necessary. Refer to the following pages. 21. Turn off the laser. 22. Place the cover on the laser. Slide it downward until seated, and then finger-tighten the screw at the top of the cover. Take care not to jolt or bang the laser. 23. Check the alignment of the laser several times. Use the fine adjustments for minor adjustments. 24. Return the gantry to 0°. The PROBE+ laser has both coarse and fine adjustments on the horizontal and vertical position of the crosshair. The fine adjustments have limited range and are set in the middle of range during factory calibration. If the limit of range is reached, turn the fine adjustment several turns in the other direction. Then use the coarse adjustment to move close to the desired position. Finally, use the fine adjustment to finalize the correct position. At installation the coarse adjustments should be used to position the crosshair close to the right position and then the fine adjustment can be used to accurately locate the crosshair. After positioning the crosshair close to the proper location with the coarse adjustment, use the fine adjustment to finely position the crosshair horizontally. Turn the adjustment to rotate the optics, moving the crosshair left or right. Focus, angularity and line adjustments at installation: Adjustments are available if required, and these are explained on the following pages. Brightness of the laser beam is factory set at the maximum allowable. Increased brightness will damage the laser. Consult the factory for reducing brightness. 13 Side Laser Installation PROBE+ Diode Laser Movement of Crosshair and Sagittal Lines Figure 6 Laser Positioning Adjustments 14 Installation and User Guide Coarse Carriage Adjustments The carriage on which the laser is seated can be moved both horizontally and vertically using the tools included in calibration kit. Course Vertical Adjustment C B B D D A Course Horizontal Adjustment Coarse vertical adjustment of the laser carriage. Loosen the large allen (hex) screws (B) with a 1/8" allen (hex) wrench; do not loosen too much. Now use the wrench to turn the setscrew (A), which moves the carriage vertically. When vertical adjustment is satisfactory, tighten the large allen (hex) screws (B) firmly. Coarse horizontal adjustment of the laser carriage. Loosen the large allen (hex) screws (D) with a 1/8" hex wrench; do not loosen too much. Now use the wrench to turn the setscrew (C), which moves the carriage horizontally. When horizontal adjustment is satisfactory, tighten the large allen (hex) screws (D) firmly. 15 Laser Positioning Adjustments PROBE+ Diode Laser Additional Coarse Adjustment The PROBE+ vertical line optics can be coarsely adjusted by rotating the turret. To make this adjustment, loosen the setscrew shown below using a .050” allen (hex) wrench. The entire optics assembly can be rotated by hand, left or right, to move the crosshair horizontally. This is a very coarse adjustment. The setscrew must then be re-tightened. A fine-tuning adjustment can then be used to accurately position the crosshair. Laser Positioning Adjustments 16 Installation and User Guide Line Adjustment Components The PROBE+ has both coarse and fine line positional adjustments available. The adjustment components are identified in Figure 7 below. A F G E B I H F H C D Figure 7 A. Mirror adjustment screw B. Beam splitter adjustment: C. Locking setscrew: Coarse rotational adjustment D. Horizontal Adjustment Screw F. Lens adjustment screws G. Lens angle adjustment screws H. Line focus adjustment lever I. Focus lock screw E. Vertical Adjustment Screw 17 Laser Positioning Adjustments PROBE+ Diode Laser Fine Line Positioning Both of the following adjustments can be made with the cover either on or off. These adjustments may need to be repeated multiple times during the installation process and are available for customer use. FINE HORIZONTAL POSITIONING After positioning the crosshair close to the proper location with the coarse adjustment, use the fine adjustment screw (Figure 7 D) to finely position the crosshair horizontally as shown below. FINE VERTICAL POSITIONING After positioning the crosshair close to the proper location with the coarse carriage adjustment, Use the fine adjustment screw (Figure 7 E) to finely position the crosshair vertically as shown below. Laser Positioning Adjustments 18 Installation and User Guide Line Clarity and Additional Positional Adjustments Line curvature, focus, and angularity are factory set and should not require adjustments at installation. However if required, they are explained on the following pages. These adjustments can only be made with the cover removed. CURVATURE ADJUSTMENTS The line curvature may be adjusted by using the two adjustment screws (Figure 7 F) on the right hand side of the optics. A 3/32" allen (hex) wrench is required. LASER FOCUS ADJUSTMENT The focus of the PROBE+ can be adjusted as follows: Remove the cover, if necessary. The focus adjustment is a silver colored lever (Figure 7 H) on the operators right hand side of the optics. On the opposite (left hand) side of the optics is a set screw (Figure 7 I) in a deep angled hole. Loosen the locking set screw using a 0.050” allen (hex) wrench. Focus the crosshair by moving the lever to concentrate the crosshair into fine lines. Re-tighten the setscrew (Figure 7 I). Replace the cover and check the crosshair focus and alignment. 19 Laser Positioning Adjustments PROBE+ Diode Laser Additional Positional Adjustments Figure 8 SINGLE LINE ADJUSTMENTS The two adjustments on the left hand side of the optics move either line of the crosshair up or down with respect to the other line. Use a 3/32" allen (hex) wrench. MIRROR ADJUSTMENT SCREW This upper screw moves the horizontal line up and down. Figure 8 Figure 9 BEAM SPLITTER ADJUSTMENT SCREW This lower screw moves the vertical line up and down. Figure 9 Laser Positioning Adjustments 20 PROBE+ Diode Laser Installation and User Guide Line Rotational Adjustments Figure 10 ROTATE HORIZONTAL LINE The horizontal line of the crosshair can be rotated as follows. Loosen the locking set screw, as shown, on the right side of the upper lens holder. Use a .050" allen (hex) wrench. Rotate the lens holder by hand. Use care not to touch the lens or upset the factory calibration of the lens or the mirror. A B Figure 10 A ROTATE VERTICAL LINE The vertical line of the crosshair can be rotated as follows. Loosen the locking set screw, as shown, on the right side of the lower lens holder. Use a .050" allen (hex) wrench. rotate the lens holder by hand. Use care not to touch the lens or upset the factory calibration of the lens or the mirror. Figure 10 B CENTERING OPTICS OVER LASER The PROBE+ optics are factory set to be centered on the laser diode. Adjustment must be done by factory. Second Side Laser Installation Repeat the process, on the opposite wall, used to mount the first laser. 21 Laser Positioning Adjustments PROBE+ Diode Laser Figure 11 Use a sheet of typing paper to verify the placement of the side lasers. . Figure 12 Use a front pointer to verify that the side lasers hit isocenter. Additional Installations 22 PROBE+ Diode Laser Installation and User Guide Figure 13 Use a string to verify the location of the side lasers 23 Additional Installations PROBE+ Diode Laser Installation on Angled Walls 1. When mounting on Angled Walls, cut out, fold and tape to the mounting template the strip of paper provided with the template. This strip approximates the distance to the center of rotation of the top lens of the laser optics. Use a level to verify: • Gantry is level each time it is rotated • Template is placed accurately on wall • Crosshairs on wall are at 90° angles 2. Use a level to determine the height of the laser exit reference point. (Refer to Pages 9 & 10). 3. Set gantry at 0°, tape a piece of paper to the table. 4. Project a crosshair onto the paper, keeping the field as wide as possible. 5. Hold the ends of a string on each wall until the string shadow is in line with the projected crosshair. Several people may be needed. See Figure 13. 6. On each wall use the template with its extension. Move the template until the string touches the X on the extension. That is the center of the laser exit reference point. Level the template and mark the mounting holes. 7. Drill the mounting holes. 8. Mount the laser. Adjust the turret of the laser by loosening the setscrew and rotating the optics until the laser crosshair lines up with the gantry pointer. Tighten the setscrew. Use the fine adjustments to make any additional corrections horizontally. QUICK laser lines overlap front pointer lines up with the laser lines lines are straight up and down lines are level on the horizontal axis lines are sharp and focused at the isocenter 9. Repeat the process on the other side laser. 10. Laser should form a continuous plane. 11. Place a piece of paper in the path (Figure 11 & 12) and verify and correct as needed: NOTE: If there is no front pointer, raise the table to isocenter and align the lasers with the table edge. Additional Installations 24 PROBE+ Diode Laser Installation and User Guide X Figure 14 Use a plumb bob to determine the placement of the ceiling laser. 25 Additional Installations PROBE+ Diode Laser Ceiling Laser Installation 1. Level gantry at 0° and place the table at isocenter height. 2. Tape a piece of paper to the table. Use a level to verify: 3. Project a crosshair and mark the center point with a pencil. • Gantry is level each time it is rotated 4. Move collimator out of the way. • Template is placed accurately on wall • Crosshairs on wall are at 90° angles 5. Hold a plumb bob from the ceiling so that the bottom weight hits the dot on the table. The point at which the string hits the ceiling indicates the laser exit reference point. Mark this with a pencil. Figure 14 6. Use the template to mark the location of the mounting holes. 7. Drill the mounting holes. 8. Unplug the laser power supply and remove the cover. 9. Attach washers and screws and mount the laser to the ceiling. 10. Plug in power supply and turn on laser. 11. The center of the laser crosshair should correspond with the dot on the table. 12. Raise and lower the table. The crosshair should align for the entire course. Refer to Laser Positioning Adjustments on pages 14 through 21. NOTE: If mounted in a recess, cut a hole in the ceiling tile large enough to allow the entire cross to be projected. Additional Installations 26 PROBE+ Diode Laser Installation and User Guide Figure 15 Use a string and the ceiling laser crosshair to determine the placement of the sagittal laser. 27 Additional Installations PROBE+ Diode Laser Sagittal Laser Installation The Sagittal Laser is mounted high on the wall (room layout). Minimum height: 7', the higher the better. 1. Set the table to isocenter height. Use a level to verify: • Gantry is level each time it is rotated • Template is placed accurately on wall • Crosshairs on wall are at 90° angles 2. Use the crosshair from the ceiling laser as a reference and tape a plumb bob string to the center of the table so that the string shadow aligns with the crosshair. (Point A in Figure 15) If your system does not contain a ceiling laser, use the collimator-projected crosshair. Be sure to check for collimator accuracy. (Refer to page 9). 3. Move the table up and down to verify tracking of the string shadow. 4. The point at which the string hits the wall indicates the center of the vertical laser exit reference point. (Point B in Figure 15) 5. Use the template to determine the location of the mounting holes. 6. Drill the mounting holes. 7. Unplug the laser power supply and remove the cover. QUICK 8. Attach washers and screws and mount the laser to the wall. 9. Plug in power supply turn on laser. laser lines overlap front pointer lines up with the laser lines lines are straight up and down lines are sharp and focused 10. Verify that the laser line tracks the crosshair projected onto the table. Additional Installations 28 PROBE+ Diode Laser Installation and User Guide Calibration and Adjustments – PROBE+ Warning - Risk of incorrect diagnosis of treatment. • Isocenter of alignment lasers must coincide with isocenter of diagnostic of treatment equipment. Test alignment of laser daily, before use, for first month of operation and weekly after first month. Newly installed equipment will shift as building walls and floor settle and as building materials dry out. Re-align lasers to isocenter before use, if they have moved. • The laser must be aligned with the diagnostic or therapy equipment to be safely useful. Only a GAMMEX service technician, Medical Physicist, Biotech Engineer or Service Technician should perform alignment. Caution - Laser Light Risk of eye damage. Do not stare into beam. • Alignment procedures need to be performed with laser on. • Look only at laser light reflected from a diffuse surface (like paper or a wall). See Simplified Room Layout for general concept of laser alignment. Removing the Cover Some adjustments and repairs require the removal of the cover. Remove this cover, as follows: First, turn off the laser. Unplug the power cord from the AC receptical and from the PROBE+ laser. Loosen the screw at the top of the cover using your fingers. Slide the cover up, to disengage it, and lift the cover off and away. Plug in the power cord and on laser for adjustments. To replace the cover after making adjustments, reverse the process. Turn off the laser. Unplug the power cord. Tilt the top of the cover slightly away from the wall. Slide the bottom tab on the cover under the bottom end plate. Tilt the top of the cover toward the wall until the screw is aligned with the hole in the top end plate. Secure the cover by finger tightening the screw. Take care not to jar the laser. Plug in the power cord. 29 Laser Calibration and Adjustments PROBE+ Diode Laser Calibration and Adjustments – PROBE+ GENERAL Test the alignment of the lasers daily, before use, for the first month after installation. After the first month test the system at least weekly. Isocenter of the alignment lasers must coincide with the isocenter of the diagnostic or treatment equipment. CEILING AND SAGITTAL LASERS Tape a piece of typing paper to the table under the gantry (Leveled and Set at 0°). Project a crosshair on the paper from the gantry. Mark the center point. Move the gantry out of the way. Project the Ceiling laser crosshair on the paper. Raise and lower the table. The center of the crosshair should align with the dot on the paper. In the same way the Sagittal line should pass through the dot on the paper. SIDE LASERS Turn on both side lasers. Use a piece of typing paper held perpendicular to the laser light. The two-projected laser crosshairs should line up on the paper (Figure 11). With the gantry at 0° (Leveled and Set at 0°) use a front pointer (Figure 12) to verify that both side lasers hit isocenter. If there is not a front pointer, raise the table to isocenter and project a crosshair on the table to verify that the side lasers hit isocenter. ADJUSTMENTS In a new installation some building movements can be expected. In most cases the fine adjustments made by turning the knobs as shown on pages 17 should be sufficient to make any adjustments. The fine adjustments have limited range. If the fine adjustment is at the end of the range, turn the knob several turns in the opposite direction. Use a coarse adjustment to come close to the desired position and use the fine adjustment to zero in. Pages 14 through 21 outline various adjustments available on the PROBE+. In most cases only the fine horizontal and vertical adjustments will be needed. Laser Calibration and Adjustments 30 PROBE+ Diode Laser Installation and User Guide Troubleshooting PROBLEM NO LASER LIGHT POSSIBLE CAUSE POSSIBLE REMEDY Wall switch not on. Check switch by turning laser off and on. Wall switch not in full on position. Verify switch lever is in the full on position. Plug/connector not plugged in. Verify plug is in the fully Plug partially plugged in. plugged in position. Power supply not functioning. INTERMITTENT LASER Use voltmeter to measure the voltage output of the power supply wall plug. If voltage less than 11 or greater than 13 VDC, the power supply will need to be replaced. Call Gammex Sales and Service for assistance. Use voltmeter and check power supply. See above. Repair Warning - Use only repair parts specified by Gammex. Some parts may be similar, but may result in unit drifting out of alignment, early failure or being out of compliance with laser regulations. Maintenance & Cleaning Once installed, Gammex Lasers require no maintenance other than cleaning of mirrors, beam splitters, and lenses according to standard optical practices. Occasional outside dusting of the red plastic window in the cover may be necessary. The window is very delicate. Use only a camera lens brush or other soft brush for this purpose. Clean other outside surfaces of the laser, receiver and transmitter with a soft cloth slightly moistened with water or a mild detergent solution, then wipe the surface with a dry cloth. Do not use any type of abrasive pad, scouring powder or chemical solvent such as thinner, benzene, alcohol or disposable wipes as these may damage the finish. Should a laser fail to function properly, contact Gammex Sales and Service. 31 Troubleshooting PROBE+ Diode Laser Repair REPLACING THE CIRCUIT BOARD AND LASER DIODE ASSEMBLY CAUTION: Laser Light. Risk of eye damage. If the Laser Light fails it can be easily replaced with a new laser diode. -Lasers must be less than 1 mw to meet Class II power levels. The laser diode and circuit board must be replaced as a set since they are adjusted at the Gammex factory to meet Class II power limits. -Do not dismantle the laser diode assembly as this could cause exposure to laser radiation exceeding Class II limits. No user repairable parts are inside this assembly. -If the laser diode or circuit board fails, they can be easily replaced with a new assembly. TOOLS REQUIRED: 5/64" .050” & 1/16” allen (hex) wrenches. 1. Turn the power off. Remove the 12 vdc power plug. Remove the power plug cord. Remove the cover by unscrewing the latch on the top and rotating it down and out to clear the switch and power jack. 2. In performing this replacement, handle the material carefully and take care not to lose the hardware. 3. Loosen the swivel screw using a 0.050” allen (hex) wrench and rotate the optics module 90 degrees to either side. 4. Remove the flat head screw using a 5/64” allen (hex) wrench. It is easier to remove these screws, and slide out the focus module, if the tension on the optics has been released. To release the tension on the optics module, pinch the top of carriage, squeezing it between your thumb and fore finger. It will rotate against the spring tension. 5. Rotate the optics module to the 90 degree position on the other side (Rotate 180 degrees). 6. Remove the second flat head screw using a 5/64” allen (hex) wrench. This will free the focus module (with the small circuit board). Slide this module out to the side where the cable comes out. Take care not to disrupt the optics module. Repair and Parts Information 32 PROBE+ Diode Laser Installation and User Guide REPLACING THE LASER DIODE 7. Remove the two button head screws using a 1/16” hex wrench, that hold the laser diode/circuit board assembly to the focus module. Remove the laser/circuit board assembly. 8. Remove the two # 6 button head screws, using a Phillips screwdriver, that hold the large circuit board to its bracket. The complete laser/electronics assembly has now been removed. 9. Replace the two circuit boards and laser assembly. 10.Reassemble the unit reversing the dis-assembly process. Do not put the cover on, at this time. Plug in the power cord and turn the laser on. 11.Focus the laser and check the alignment and make adjustments (See the Calibration and Adjustments Section.). 12.Lasers are shipped with brigthness at the maximum allowed. Do not turn up the brightness over this level. It will cause the laser to burn out. The brightness is controlled on the large circuit board by the silver screw on a potentiometer. This screw has a single slot and points perpendicular to the circuit board. 13.Turn off the PROBE+. Unplug the power cord. Replace the cover. Reattach the power cord and turn the PROBE+ on. check the alignment again. Return this assembly to Gammex. Call Gammex Sales and Service at 1-800-GAMMEX-1 DISMANTLING The PROBE+ can be removed from the wall by following the reverse of the installation instructions. DISPOSAL The PROBE+ laser contains steel, aluminum, lead, lead alloys copper, glass, brass, lead-free paint and plastic. Recycle according to your applicable laws. 33 Repair and Parts Information PROBE+ Diode Laser Parts List PART NO. DESCRIPTION 601955 PCB, Diode, PROBE + Sub Assembly 007622 POWER SUPPLY 12VDC Universal Medical Grade POWER SUPPLY PLUGS Set of 4 006223 WINDOW: RED LENS Repair and Parts Information 34 PROBE+ Diode Laser Installation and User Guide Specifications Laser Beam Output Maximum Power....................................Less than 0.5 mW (each beam)Laser Class II Range ..............................................................................................Up to 6.0 m(20 ft.) Line Width........................................................................0.5 mm(0.02in) @ 2 m (7 ft.) Beam (Line) Divergence .....................................................................................15-17° Drift .................................................................................................No measurable drift WaveLength .......................................................................6350Å (635 nm) visible red Visibility ...............................................................Clearly visible in strong ambient light Intensity .....................................................................User adjustable to desired levels Pulse Duration and Repetition Rate..............................Continuous (No Pulse) Output Laser Beam Adjustment Horizontal Range of Vertical Projection...............330.2 cm (±130 in.) @ 3.05 m (10ft.) Vertical Range of Horizontal Projection.......Plus 100 cm /minus 300cm @ 3 m (10 ft.) Line angle adjustment range ...................................................................±180 degrees Mounting Turret................................................................................ adjustable to ±45° Laser Dimensions Length................................................................................................17.3 cm (6.8 in.) Width..................................................................................................14.0 cm (5.5 in.) Depth. ..................................................................................................8.0 cm (3.1 in.) Weight.................................................................................................1.8 kg (4.0 lbs.) Universal Power Supply................includes adapters for U.S. and international use. Power Requirements........................................................................12 VDC, 300mA Environmental Conditions Operating Temperature:................................................................................0 - +60oC Storage Temperature:................................................................................-20 - +75oC Operating Humidity: ....................30 - 85% Relative Humidity (no dew condensation) Storage Humidity:........................30 - 85% Relative Humidity (no dew condensation) Certification Complies with Center for Devices and Radiological Health regulations for Class II Lasers. Compliance Information Complies with 21 CFR 1040.10 and.11:April, 2008 for Class II visible lasers. Complies with IEC 60825-1:2007 for Class II visible lasers. Complies with IEC 61140:2001 for Class II equipment. CE marked. Power supply UL, cUL, GS and CE marked. 35 Specifications PROBE+ Diode Laser Technical Support Technical support is available from Gammex Service Representatives who are available to work with facility planners, architects and hospital personnel throughout the planning and installation process. Contact the Gammex Service department at 1-800-232-9699. Product Warranty Gammex Inc. warrants its patient positioning equipment to be free from defects in materials and workmanship under normal use and service for 1 year from the date of shipment. The sole obligation of Gammex Inc. under this warranty is to repair or replace without charge or to refund the purchase price, at the option of Gammex Inc., of any parts which its examination shall have disclosed to be defective, provided that Buyer shall have given to Gammex Inc., a written notice of the claimed defect no later than 7 days after the end of the warranty period (one (1) year of the date of shipment of such equipment to Buyer). At the request of Gammex Inc., Buyer, at its expense, shall return the claimed defective part to Gammex Inc. THIS CLASS II LASER PRODUCT COMPLIES WITH DHHS REQUIREMENTS PURSUANT TO 21 CFR, SUB-CHAPTER J. DISCLAIMER OF OTHER WARRANTIES The aforesaid warranty rights are buyer's exclusive remedies and are in lieu of any other remedies, obligations, or rights, including, without limitation, any other warranties, expressed or implied (e.g., implied warranties of merchantability or fitness for a particular purpose). Under no circumstances shall Gammex Inc. be liable for any incidental, indirect, special or consequential damages or for any other loss, damage, penalty or expense of any kind, including, without limitation, loss of profits or overhead, reimbursement, personal injury or property damage. The aforesaid warranty obligation of Gammex Inc. constitutes its sole liability, and under no circumstances, shall the maximum liability of Gammex Inc., under any legal theory(e.g., contract, warranty, negligence, promissory estoppel, strict liability, misrepresentation, tort) and for any reason whatsoever(e.g., defect, delay or otherwise) exceed the purchase price of the defective part regardless whether the claim is asserted by buyer or any other person or entity. The liabilities of Gammex Inc. as above set forth, shall not be extended because of advice given by it in connection with the design, installation or use of the equipment or parts therefore. Your warranty may be registered at http://www.gammex.com/warranty.asp Product Warranty 36 Product Ordering Information and Customer Service Gammex is committed to satisfying our customers’ needs. If you have any questions, comments or suggestions regarding our products and service, please call or fax us. Please contact a GAMMEX dealer or representative for a quotation or for a detailed description of our ordering policies, warranties, delivery policy, conditions of sale, damaged goods policy, and returned goods policy. In the United States, Sales Department hours are: 7:30 a.m. through 5:00 p.m. (Central Time). SALES/CUSTOMER SERVICE: 608-828-7000 TOLL FREE (U.S. and Canada only): 800-426-6391 FAX: 608-828-7500 Sales e-mail: [email protected] In the United States, Service Department hours are: 7:30 a.m. through 5:00 p.m. (Central Time). SERVICE DEPARTMENT: 608-828-7000 TOLL FREE (U.S. and Canada only): 800-232-9699 FAX: 608-828-7500 Service e-mail: [email protected] GAMMEX, Inc. 7600 Discovery Dr. P.O. Box 620327 Middleton, WI 53562-0327 USA On the internet at: www.gammex.com VISA, MasterCard and American Express accepted. All products and specifications are subject to change without notice. © 1997 GAMMEX. All rights reserved. Printed in U.S.A. 006991-00-03 GAMMEX 7600 Discovery Dr. P.O. Box 620327 Middleton, WI 53562-0327 USA 1 800 GAMMEX 1 (426-6391) 1 608 828 7000 Fax: 1 608 828 7500 www.gammex.com e-mail: [email protected] GAMMEX-RMI LTD Broadway Business Centre 32A Stoney Street Nottingham NG1 1LL United Kingdom +44 0115 92474188 Fax: +44 0115 9247189 e-mail: [email protected] GAMMEX-RMI Gmbh Bergstrasse 2 35398 Giessen Germany +49 6403 774 2293 Fax: +49 6403 774 2526 e-mail: [email protected]
advertisement
Key Features
- Class II Laser safety
- Visible red light
- Adjustable intensity
- Easy installation
- Precise alignment
- Durable construction
- User-friendly controls