Practical tools for field calibration Intrinsically safe MC2-IS

Practical tools for field calibration Intrinsically safe MC2-IS

Practical tools for field calibration

Intrinsically safe

MC2-IS

Beamex MC2: Practicality for calibration

Practicality in calibration. The Beamex MC2 series includes three different hand-held calibrators for field use: the MC2 pressure/electrical calibrator, the

MC2 temperature/electrical calibrator and the MC2 multifunction calibrator. The MC2 is a compact-sized and easy-to-use hand-held calibrator. It has a large graphical display, a menu-based interface and full numerical keyboard. Being a Beamex calibrator,

MC2 represents the high, uncompromised quality standards evident in other Beamex calibration equipment.

Beamex MC2-IS: Practical tool for calibration in hazardous environments

The Beamex MC2-IS intrinsically safe multifunction calibrator is an ATEX and IECEx certified calibrator designed for use in potentially explosive environments such as offshore platforms, oil refineries, chemical and petrochemical plants where inflammable gases may be present. It connects to almost 20 available Beamex intrinsically safe external pressure modules. The calibrator has a compact size and design.

Beamex MC2 and Beamex MC2-IS – a series of 4 different practical calibrators.

MC2 Main features

Available in three versions:

MC2 Pressure/Electrical Calibrator

MC2 Temp./Electrical Calibrator

MC2 Multifunction Calibrator

Internal / external pressure modules

Compact size and design

User-friendly

MC2-IS Main features

ATEX and IECEx certified Intrinsically Safe Multifunction Process

Calibrator

Connects to almost 20 available Beamex intrinsically safe external pressure modules

Compact size and design

User-friendly

MC2 Feature comparison

Features

Internal pressure module

Connection for external pressure modules

Current measurement

(with internal and external supply)

Voltage measurement

Frequency measurement

Pulse counting

Switch sensing

Internal HART

®

compatible

24 VDC loop supply

Current generation

(with internal and external supply)

Voltage generation

Frequency generation

Pulse generation mV measurement / simulation

Resistance measurement / simulation

RTD measurement / simulation

TC measurement / simulation

1) 20 VDC loop supply

2) With external supply

MC2-PE

Pressure

/ Electrical

-

-

-

-

-

-

-

-

MC2-TE

Temperature

/ Electrical

-

MC2-MF

Multifunction

MC2-IS

Intrinsically

Safe

(barometric only)

(1

(2

3

Features of the MC2 and MC2-IS

Accuracy guaranteed

The Beamex MC2 and Beamex MC2-IS calibrators are extremely accurate process calibrators. As proof of this, each

MC2 and MC2-IS calibrator is delivered with a traceable, accredited calibration certificate.

Compact and user-friendly

The MC2 and MC2-IS calibrators are compact-sized, lightweight portable calibrators with a large graphical display and multilingual interface. They have a full numerical keyboard.

Using the MC2 and MC2-IS is quick and easy.

Robust field calibrator

The MC2 and MC2-IS are robust and made for tough use. The impact protectors and membrane keyboard make them field compatible and weatherproof.

Wide range of configuration possibilities

The MC2 and MC2-IS provide a number of configuration possibilities, such as both internal and external pressure modules. For instance, the MC2-IS connects to almost 20 available Beamex intrinsically safe external pressure modules.

Make it safe with the MC2-IS

The MC2-IS is an intrinsically safe, ATEX and IECEx certified, multifunction calibrator. It is designed for use in potentially explosive environments, such as offshore platforms, oil refineries, chemical and petrochemical plants where inflammable gases may be present.

4

BEAMEX - WORLD-CLASS CALIBRATION SOLUTIONS

MC2 - General specifications for all models

General Specifications

General MC2

Display

Weight

Dimensions

Keyboard

Battery type

Charging time

Charger supply

60 mm x 60 mm (2.36” x 2.36”), 160 x 160 pixels LCD, backlit

720 ... 830 g (1.59 ... 1.83 lbs)

215 mm (8.5”) x 102 mm (4”) x 49 mm (1.9”) (d/w/h)

Membrane keyboard

Rechargeable NiMH, 4000 mAh, 3.6V DC

5 hours

100...240 VAC, 50-60 Hz

Battery operation

13 ... 24 hours in measurement mode, back light off.

8 ... 12 hours when sourcing an average of 12 mA to loop, with back light on.

Operating temperature -10...50°C (14...122°F)

Operating temperature when charging batteries

Storage temperature

Humidity

Warmup time

Max. input voltage

Safety

EMC

0 … 35°C (32...95°F)

-20 to 60°C (-4 to 140°F)

0 to 80% R.H. non condensing

Specifications valid after a 5 minute warmup period.

30 V AC, 60 V DC

Directive 73/23/EEC, EN 61010-1

Directive 89/336/EEC, EN 61326

Ex

Warranty

-

Standard: 2 years for MC2; 1 year for battery pack.

1)

MC2-IS

60 mm x 60 mm (2.36” x 2.36”), 160 x 160 pixels LCD, backlit

1070 g (2.3 lbs)

215 mm (8.5”) x 102 mm (4”) x 49 mm (1.9”) (d/w/h)

Membrane keyboard

Rechargeable NiMH, 1700 mAh, 4.8V DC

12 hours

100...240 VAC, 50-60 Hz

12 hours in measurement mode, backlight off.

4 hours when sourcing a loop, backlight continuously on.

-10...50°C (14...122°F)

0 … 35°C (32...95°F)

(Must be charged at a non Ex area.)

-20 to 60°C (-4 to 140°F)

0 to 80% R.H. non condensing

Specifications valid after a 5 minute warmup period.

30 V DC

Directive 73/23/EEC, EN 61010-1

Directive 89/336/EEC, EN 61326

ATEX: II 1 G ia IIC T4 Ga (T

IECEx: Ex ia IIC T4 Ga (T a

= -10 ... +50 °C) a

= -10 ... +50 °C)

Standard: 2 years for MC2-IS; 1 year for battery pack.

1)

1) The warranty of the MC2 and MC2-IS will be extended up to 6 years if the product is calibrated on a yearly basis at Beamex’s Calibration Laboratory

Voltage measurement -1…60 V DC (MC2-IS -1...30 V DC)

Range MC2 Range MC2-IS Resolution

±0.25 V

±(0.25 ... 1 V)

1 ... 25 V

25 ... 60 V

±0.25 V

±(0.25 ... 1 V)

-

1 ... 30 V

0.001mV

0.01 mV

0.1 mV

1 mV

1 Year Uncertainty(±)

(2

0.02% RDG + 5 µV

0.02% RDG + 5 µV

0.02% RDG + 0.25 mV

0.02% RDG + 0.25 mV

Feature

Temperature coefficient

Input impedance

Supported units

Display update rate

Specification

< ±0.0015% RDG / °C outside of 18 ... 28°C

< ±0.0008% RDG / °F outside of 64.4 ... 82.4°F

>1 M W

V, mV, µV

3 / second

mA measurement ±100 mA

Range

±25mA

±(25 ... 100 mA)

Resolution

0.0001 mA

0.001 mA

1 Year Uncertainty(±)

(2

0.02% RDG + 1.5 µA

0.02% RDG + 1.5 µA

Feature

Temperature coefficient

Input impedance

Supported units

Display update rate

Specification

< ±0.0015% RDG / °C outside of 18 ... 28°C

< ±0.0008% RDG / °F outside of 64.4 ... 82.4°F

< 7.5

W mA, µA

3 / second

Loop supply

Feature

Maximum output current

Specification MC2

> 25 mA, short circuit protected

Output voltage 24 V ± 10%

Output impedance in HART ® compatible mode 300

W ± 20%

Specification MC2-IS

> 25 mA, short circuit protected

20 V ± 10% @ 0 mA, > 12 V @ 20 mA

250

W ± 20%

2) Uncertainty includes reference standard uncertainty, hysteresis, non-linearity, repeatability and typical long-term stability for the mentioned period. (k=2).

BEAMEX - WORLD-CLASS CALIBRATION SOLUTIONS

5

Electrical measurements

Frequency measurement 0.0027 … 50 000 Hz

Range

0.0027 ... 0.5 Hz

0.5 ... 5 Hz

5 ... 50 Hz

50 ... 500 Hz

500 ... 5000 Hz

5000 ... 50000 Hz

Resolution

0.000001 Hz

0.00001 Hz

0.0001 Hz

0.001 Hz

0.01 Hz

0.1 Hz

1 Year Uncertainty(±)

(1

0.01% RDG

0.01% RDG

0.01% RDG

0.01% RDG

0.01% RDG

0.01% RDG

Feature

Temperature coefficient

Input impedance

Trigger level

Minimum signal amplitude

Supported units

Gate period

Specification

Specification valid from -10 to 50°C (14 ... 122°F)

> 1 M

W

-1…14 V in 1 V steps and open collector inputs

2 Vpp (< 10 kHz), 3 Vpp (10...50 kHz)

Hz, kHz, cph, cpm, 1/Hz (s), 1/kHz (ms), 1/MHz (µs)

267 ms + 1 signal period

1) Uncertainty includes reference standard uncertainty, hysteresis, non-linearity, repeatability and typical long-term stability for the mentioned period. (k=2)

Pulse counting 0 ... 9 999 999 pulses

Feature Specification

Range

Input impedance

Trigger level

Minimum signal amplitude

0 to 9 999 999 pulses

> 1 M W

-1 … 14 V in 1 V steps and open collector inputs

2 Vpp (pulse length > 50 µs), 3 Vpp (pulse length 10 ... 50 µs)

Switch test

Feature

Potential free contacts

Voltage level detection

Specification

Test voltage (Trigger level)

Trigger level

Input impedance

MC2 MC2-IS

3 V, 0.13 mA (1 V) or 24 V, 35 mA (2 V) 3 V, 0.08 mA (1 V) or 20 V, 25 mA (2 V)

-1…14 V in 1 V steps

> 1 M W

-1…14 V in 1 V steps

> 1 M W

6

BEAMEX - WORLD-CLASS CALIBRATION SOLUTIONS

Pressure Measurements

Internal Pressure Modules (IPM)

Internal Module

MC2

(3

IPM200mC

IPM2C

IPM20C

IPM160

Barometric option

Feature

Temperature coefficient

Maximum overpressure

Pressure port

Media compatibility

Supported pressure units

Display update rate

-

-

-

-

Internal Module

MC2-IS

(4

Barometric option -IS

Unit Range

(2

Resolution 1 Year Uncertainty(±)

(1 kPa mbar iwc kPa bar psi kPa bar psi

MPa bar psi

±20

±200

±80

-100 to 200

-1 to 2

-14.5 to 30

-100 to 2000

-1 to 20

-14.5 to 300

0 ... 16

0 ... 160

0 ... 2400

0.001

0.01

0.01

0.01

0.0001

0.001

0.1

0.001

0.01

0.001

0.01

0.1

0.05% RDG + 0.05% FS

0.05% FS

0.05% FS

0.05% FS

Also enables absolute pressure measurement for the above pressure inputs.

When using the barometric option, add 0.1 kPa (0.0146 psi) uncertainty for absolute pressure measurement.

Specification

< ±0.001% RDG /°C outside 15 … 35°C.

< ±0.0006% RDG /°F outside 59 … 95°F

2 × Range

G 1/8” female with G 1/8” male (ISO 228/1) 60° internal cone adapter

IPM160: G 1/8” female

Wetted parts: AISI316 stainless steel, Nitrile rubber.

Pa, hPa, kPa, MPa, mbar, bar, lbf/ft2, psi, ozf/in2, gf/cm2, kgf/cm2, kgf/m2, kp/cm2, at, mmH2O, cmH2O, mH2O, iwc, ftH2O, mmHg, cmHg, mHg, inHg, mmHg(0°C), inHg(0°C), mmH2O(4°C; 60°F;

68°F/20°C), cmH2O(4°C; 60°F; 68°F/20°C), inH2O(4°C; 60°F; 68°F/20°C), ftH2O(4°C; 60°F;

68°F/20°C), torr, atm, + four (4) user-configurable units

2.5 / second

External Pressure Modules (EXT) Standard Accuracy

Non Intrinsically Safe Intrinsically Safe

EXT200mC-s

EXT2C-s

EXT200mC-s-IS

EXT2C-s-IS

Range

(2

±200 mbar

-1 ... 2 bar

±80 iwc

-14.5 ... 30 psi

Resolution

0.01 mbar

0.01 iwc

0.0001 bar

0.001 psi

EXT20C-s

EXT160-s

EXT20C-s-IS

EXT160-s-IS

-1 ... 20 bar -14.5 ... 300 psi

0 ... 160 bar 0 ... 2400 psi

0.001 bar

0.01 psi

0.01 bar

0.1 psi

1 Year Uncertainty(±)

(1

0.05% RDG + 0.05% FS

0.05% FS

0.05% FS

0.05% FS

External Pressure Modules (EXT) High Accuracy

Non Intrinsically Safe Intrinsically Safe

Barometric

EXT10mD

EXT100m

EXT400mC

EXT1C

EXT2C

EXT6C

EXT20C

EXT60

EXT100

EXT160

EXT250

EXT600

Barometric-IS

EXT10mD-IS

EXT100m-IS

EXT400mC-IS

EXT1C-IS

EXT2C-IS

EXT6C-IS

EXT20C-IS

EXT60-IS

EXT100-IS

EXT160-IS

EXT250-IS

EXT600-IS

EXT1000 EXT1000-IS

Range

(2

800 … 1200 mbar abs 23.6 … 35.4 inHg a

±10 mbar differential ±4 iwc differential

0 … 100 mbar gauge

±400 mbar

0 … 40 iwc

±160 iwc

±1 bar

-1 … 2 bar

-1 … 6 bar

-1 … 20 bar

-14.5 … 15 psi

-14.5 … 30 psi

-14.5 … 90 psi

-14.5 … 300 psi

0 … 60 bar

0 … 100 bar

0 … 160 bar

0 … 250 bar

0 … 600 bar

0 … 900 psi

0 … 1500 psi

0 … 2400 psi

0 … 3700 psi

0 … 9000 psi

0 … 1000 bar 0 … 15000 psi

1 Year Uncertainty(±)

(1

0.5 mbar (0.015 inHg)

0.1% RDG + 0.05% Span

0.025% RDG + 0.025% FS

0.025% RDG + 0.02% FS

0.025% RDG + 0.015% FS

0.025% RDG + 0.01% FS

0.025% RDG + 0.01% FS

0.025% RDG + 0.01% FS

0.025% RDG + 0.01% FS

0.025% RDG + 0.01% FS

0.025% RDG + 0.01% FS

0.025% RDG + 0.015% FS

0.025% RDG + 0.015% FS

0.025% RDG + 0.015% FS

1) Uncertainty includes reference standard uncertainty, hysteresis, non-linearity, repeatability and typical long-term stability for the mentioned period. (k=2)

2) The internal pressure module’s range may also be displayed in absolute pressure if a barometric module is used.

3) The MC2 Calibrator can hold one internal pressure module and the barometric option.

4) The MC2-IS does not have any internal pressure modules, but it does have a barometric option.

All external pressure modules (EXT) are also compatible with Beamex MC4, MC5, MC5-IS and MC5P Calibrators.

BEAMEX - WORLD-CLASS CALIBRATION SOLUTIONS

7

Electrical generation, measurement and simulation

mV measurement (T/C-terminals) -25 ... 150 mV

Range Resolution 1 Year Uncertainty(±)

(1

-25 ... 150 mV 0.001 mV 0.02% RDG + 4 µV

Feature

Temperature coefficient

Input impedance

Supported units

Display update rate

Specification

< ±0.0015% RDG / °C outside of 18 ... 28°C

< ±0.0008% RDG / °F outside of 64.4 ... 82.4°F

> 10 M

W

V, mV, µV

3 / second

mV generation (T/C-terminals) -25 ... 150 mV

Range Resolution

-25 ... 150 mV 0.001 mV

Feature

Temperature coefficient

Maximum load current

Load effect

Supported units

Specification

< ±0.0015% RDG / °C outside of 18 ... 28°C

< ±0.0008% RDG / °F outside of 64.4 ... 82.4°F

MC2: 5 mA

MC2-IS: 1 mA

< 5µV/mA

V, mV, µV

1 Year Uncertainty(±)

(1

0.02 % RDG + 4 µV

Voltage generation -3 ... 12 V

(MC2-IS: -3 ... 11 V)

Range MC2 Range MC2-IS Resolution

±0.25 V

-3 ... -0.25 V

0.25 ... 12 V

±0.25 V

-3 ... -0.25 V

0.25 ... 11 V

0.01 mV

0.1 mV

0.1 mV

Feature

Temperature coefficient

Maximum load current

Load effect

Supported units

Specification

< ±0.0015% RDG / °C outside of 18 ... 28°C

< ±0.0008% RDG / °F outside of 64.4 ... 82.4°F

MC2: 5 mA

MC2-IS: 1 mA

< 50 µV/mA

V, mV, µV

1 Year Uncertainty(±)

(1

0.02 % RDG + 0.1 mV

0.02 % RDG + 0.1 mV

0.02 % RDG + 0.1 mV

mA generation (source/sink) 0 ... 25 mA

(MC2-IS: only sink)

Range Resolution 1 Year Uncertainty(±)

(1

0 ... 25 mA 0.0001 mA 0.02 % RDG + 1.5 µA

Feature

Temperature coefficient

Max load impedance (source)

Max loop voltage (sink)

Supported units

Specification MC2

< ±0.0015% RDG / °C outside of 18 ... 28°C

< ±0.0008% RDG / °F outside of 64.4 ... 82.4°F

750

W (0 ... 20 mA), 600 W (20 ... 25 mA)

60 V mA, µA

Specification MC2-IS

< ±0.0015% RDG / °C outside of 18 ... 28°C

< ±0.0008% RDG / °F outside of 64.4 ... 82.4°F none

30 V mA, µA

1) Uncertainty includes reference standard uncertainty, hysteresis, non-linearity, repeatability and typical long-term stability for the mentioned period. (k=2).

8

BEAMEX - WORLD-CLASS CALIBRATION SOLUTIONS

Electrical generation, measurement and simulation

Resistance measurement 0 ... 4000 W

Range

0 ... 250 W

250 ... 2650

W

2650 ... 4000

W

Feature

Temperature coefficient

Measurement current

Supported units

Display update rate

Resolution

1 m W

10 m

W

100 m

W

Specification

< ±0.0015% RDG / °C outside of 18 ... 28°C

< ±0.0008% RDG / °F outside of 64.4 ... 82.4°F

Pulsed, bi-directional 1 mA (0..500

W), 0.2 mA (>500 W).

W, kW

3 / second

1 Year Uncertainty(±)

(1

4-wire connection:

0.02 % RDG + 3.5 m W

3-wire connection:

0.02% RDG + 13.5 m W

Resistance simulation 0 ... 4000 W

Range Resolution

0 ... 400 W

400 ... 4000 W

10 m W

100 m W

Feature

Temperature coefficient

Maximum Resistance excitation current

Settling time (pulsed currents)

Supported units

Specification MC2

< ±0.0015% RDG / °C outside of 18 ... 28°C

< ±0.0008% RDG / °F outside of 64.4 ... 82.4°F

5 mA (0 ... 650 W)

Iexc × Rsim < 3.25 V (650 ... 4000 W)

1 ms

W, kW

1 Year Uncertainty(±)

(1

0.04 % RDG or 30 m W (Whichever is greater)

0.04 % RDG or 30 m W (Whichever is greater)

Specification MC2-IS

< ±0.0015% RDG / °C outside of 18 ... 28°C

< ±0.0008% RDG / °F outside of 64.4 ... 82.4°F

4 mA (0 ... 812 W)

Iexc × Rsim < 3.25 V (812 ... 4000 W)

1 ms

W, kW

Frequency generation 0.0005 ... 10 000 Hz

Range

0.0005 ... 0.5 Hz

Resolution

0.000001 Hz

0.5 ... 5 Hz

5 ... 50 Hz

50 ... 500 Hz

500 ... 5000 Hz

5000 ... 10000 Hz

0.00001 Hz

0.0001 Hz

0.001 Hz

0.01 Hz

0.1 Hz

Feature

Temperature coefficient

Maximum load current

Output amplitude positive square wave

Output amplitude symmetric square wave

Duty cycle

Supported units

Jitter

Specification MC2

Specification valid from –10 to 50°C (14 ... 122°F)

5 mA

0 .. 12 Vpp ±(0.2 V+5%)

0 .. 6 Vpp ±(0.2 V+5%)

1 ... 99 % (0.0009 … 500 Hz), high / low time: min 25µs, max 1165 s

Hz, kHz, cph, cpm, 1/Hz (s), 1/kHz (ms), 1/MHz (µs)

< 0.28 µs

1 Year Uncertainty(±)

(1

0.01% RDG

0.01% RDG

0.01% RDG

0.01% RDG

0.01% RDG

0.01% RDG

Specification MC2-IS

Specification valid from –10 to 50°C (14 ... 122°F)

1 mA

0 .. 11 Vpp ±(0.2 V+5%)

0 .. 5.5 Vpp ±(0.2 V+5%)

1 ... 99 % (0.0009 … 500 Hz), high / low time: min 25µs, max 1165 s

Hz, kHz, cph, cpm, 1/Hz (s), 1/kHz (ms), 1/MHz (µs)

< 0.28 µs

Pulse generation 0 ... 9 999 999 pulses

Feature

Range

Resolution

Maximum load current

Specification MC2

0 to 9 999 999 pulses

1 pulse

5 mA

Output amplitude positive pulse 0 ... 12 Vpp ±(0.2 V+5%)

Output amplitude symmetric pulse 0 ... 6 Vpp ±(0.2 V+5%)

Pulse frequency

Duty cycle

0.0005 ... 10 000 Hz

1 ... 99 % (0.0009 … 500 Hz), high / low time: min 25µs, max 1165 s

Specification MC2-IS

0 to 9 999 999 pulses

1 pulse

1 mA

0 ... 11 Vpp ±(0.2 V+5%)

0 ... 5.5 Vpp ±(0.2 V+5%)

0.0005 ... 10 000 Hz

1 ... 99 % (0.0009 … 500 Hz), high / low time: min 25µs, max 1165 s

1) Uncertainty includes reference standard uncertainty, hysteresis, non-linearity, repeatability and typical long-term stability for the mentioned period. (k=2).

BEAMEX - WORLD-CLASS CALIBRATION SOLUTIONS

9

Temperature measurement and simulation

Thermocouple measurement and simulation

Thermocouple types available as standard

Type Range (°C) Range (°C)

B

(2

R

(2

S

(2

E

(2

J

(2

K

(2

N

(2

T

(2

0 ... 1820

-50 ... 1768

-50 ... 1768

-270 ... 1000

-210 ... 1200

-270 ... 1372

-270 ... 1300

-270 ... 400

0...200

200...400

400...1820

-50...0

0...100

100...1768

-50...0

0...1768

-270...-200

-200..1000

-210...1200

-270...-200

-200...1000

1000...1372

-270...-200

-200...1300

-270...-200

-200...-100

-100...400

U (4

L (4

C

(5

G

(6

D

(5

-200 ... 600

-200 ... 900

0 … 2315

0 … 2315

0 … 2315

-200...-100

-100...600

-200... 900

0 … 1000

1000 … 2000

2000 … 2315

0 … 100

100 … 2315

0 … 1000

1000 … 2000

2000 … 2315

Feature

Resolution

Temperature coefficient

Input impedance

Supported units

Display update rate

Maximum load current

Load effect -

-

Measurement

0.01 °C

< ±0.0015%

of thermovoltage / °C outside of 18...28°C

< ±0.0008%

of thermovoltage / °F outside of 64.4 ...82.4°F

>10 M

W

°C, °F, K

3 / second

0.3 °C

0.2 °C

0.25°C

0.4 °C

0.8 °C

1.2 °C

(3

1.0 °C

0.4 °C

0.8 °C

1.2 °C

1 Year Uncertainty (±)

(1

(3

2.0 °C

1.0 °C

1.0 °C

0.8 °C

0.6 °C

1.0 °C

0.7 °C

(3

0.25 °C

0.3 °C

(3

0.3 °C

0.4 °C

(3

0.4 °C

(3

0.3 °C

0.2 °C

Simulation

0.01 °C

< ±0.0015%

of thermovoltage / °C outside of 18...28°C

< ±0.0008%

of thermovoltage / °F outside of 64.4 ...82.4°F

-

°C, °F, K

-

MC2: 5 mA

MC2-IS: 1 mA

< 5 µV/mA

Internal Reference Junction

Range (°C) 1 Year Uncertainty

-10 … 50 °C ±0.25 °C

1) Uncertainty includes reference standard uncertainty, hysteresis, non-linearity, repeatability and typical long-term stability for the mentioned period. (k=2).

Uncertainty does not include reference junction uncertainty.

2) IEC 584, NIST MN 175, BS 4937, ANSI MC96.1

3) ±0.02 % of thermovoltage + 4 µV

4) DIN 43710

5) ASTM E 988 - 96

6) ASTM E 1751 - 95e1

10

BEAMEX - WORLD-CLASS CALIBRATION SOLUTIONS

Temperature measurement and simulation

RTD measurement and simulation

Sensor Type

Pt 50 ... 1000

Ni 100

Ni 120

Cu10

Range

-200 ... 200 °C

200 ... 600 °C

600 … 850 °C

-60 ... 180 °C

-80 ... 260 °C

-200 ... 260 °C

Resolution

0.01 °C

0.01 °C

0.01 °C

0.01 °C

0.01 °C

0.01 °C

Measurement 1 Year

Uncertainty (±)

(1

0.1 °C

0.2 °C

0.3 °C

0.1 °C

0.1 °C

0.2 °C

Simulation 1 Year

Uncertainty (±)

(1, 2

0.15 °C

0.25 °C

0.35 °C

0.15 °C

0.15 °C

0.8 °C

Feature

Temperature coefficient

Measurement current

Maximum Resistance excitation current

Measurement

< ±0.0015%

of resistance / °C outside of 18 ... 28°C

< ±0.0008%

of resistance / °F outside of 64.4 ... 82.4°F

Pulsed, 1 mA (0..500

W), 0.2 mA (>500 W).

-

°C, °F, K

3 / second

Simulation

< ±0.0015%

of resistance / °C outside of 18 ... 28°C

< ±0.0008%

of resistance / °F outside of 64.4 ... 82.4°F

-

MC2: 5 mA (0 ... 650 W)

Iexc × Rsim < 3.25 V (650 ... 4000

W)

MC2-IS: 4 mA (0 ... 812 W)

Iexc × Rsim < 3.25 V (812 ... 4000

W)

-

°C, °F, K Supported units

Display update rate

RTD types available as standard

Pt50 (385)

Pt100 (385)

Pt200 (385)

Pt400 (385)

Pt500 (385)

Pt1000 (385)

Pt100 (3926)

Pt100 (391)

Pt100 (375)

Pt100 (3923)

Ni100 (618)

Ni120 (672)

1) Uncertainty includes reference standard uncertainty, hysteresis, non-linearity, repeatability and typical long-term stability for the mentioned period. (k=2).

2) Specification valid with an excitation current >0.2 mA (0...400

W), >0.1 mA (400 ... 4000 W)

Cu10 (427)

Standard Accessories

• User guide

• Calibration certificate

• Internal rechargeable NiMH battery pack + battery charger

• Test leads and clips

• USB cable

• Adapter pressure connector - from G1/8” female to G 1/8” male with 60° internal cone (included in MC2-PE and MC2-MF models)

• Product safety information and EC declaration of conformity (included in MC2-IS)

Optional Accessories

• Pressure T-hose

• Soft carrying case

• Connection cable for external pressure modules

• Dry battery cartridge (not in MC2-IS)

• Calibration handpumps

All specifications are subject to change without prior notice.

© Beamex Oy Ab 2013. All rights reserved. Beamex is a trademark of Beamex Oy Ab.

HART

®

is a registered trademark of the HART communication Foundation.

BEAMEX - WORLD-CLASS CALIBRATION SOLUTIONS

11

w w w . b e a m e x . c o m

Portable calibrators

Workstations

Calibration software

Professional services

Industry solutions

For more information, please visit

www.beamex.com or contact [email protected]

Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertisement