Equipment Interfaces User Guide

Equipment Interfaces User Guide
ExamWRITER Equipment
Integration
User’s Guide
February 2015
®
Limitation of Liability and Damages
The information contained herein is subject to change without notice and is provided "as-is" without warranty of any kind. Nothing
herein shall be construed as constituting an additional warranty. Eyefinity shall not be liable for technical or editorial errors or omissions contained herein.
Copyright © 2015 by Eyefinity, Inc. All rights reserved.
Eyefinity
10875 International Drive, Suite 200
Rancho Cordova, CA 95670-7371
Trademarks
OfficeMate, ExamWRITER, and Eyefinity are registered trademarks of Eyefinity, Inc. All other brands are trademarks or registered
trademarks of their respective owners.
®
Contents
Chapter 1
1
1
1
2
5
Using the CenterVue Interface
CenterVue Interface Overview
Setting Up the CenterVue DRS Interface
Capturing Images with CenterVue DRS
Reviewing CenterVue DRS Images
Chapter 2
7
7
7
8
Using the Eagle Eye 3000 Interface
Eagle Eye 3000 Overview
Setting Up the Eagle Eye 3000 Interface
Using the Eagle Eye 3000 Interface
Chapter 3
11
11
11
13
15
Using the EyeScape Interface
EyeScape Interface Overview
Setting Up the EyeScape Interface
Capturing EyeScape Images
Reviewing EyeScape Images
Chapter 4
17
17
17
18
20
Using the EyeTel Interface
EyeTel Interface Overview
Setting Up the EyeTel Interface
Capturing Images with EyeTel
Reviewing Images from EyeTel
Chapter 5
21
21
22
22
24
Using the Heidelberg Interface
Heidelberg Interface Overview
Setting Up the Heidelberg Interface
Capturing Heidelberg Images
Reviewing Heidelberg Images
®
ExamWRITER Equipment Integration User’s Guide
i
Contents
Chapter 6
27
27
27
30
40
47
Using the Humphrey® HFA II-i Interface
Humphrey Interface Overview
Setting Up the Humphrey Interface
Setting Up the Humphrey Equipment
Using the Humphrey Interface
Viewing Humphrey Analysis Reports
Chapter 7
49
49
49
50
51
53
Using the Humphrey® Matrix Interface
Humphrey Matrix Overview
Setting Up the Humphrey Matrix Equipment
Setting Up the Humphrey Matrix Interface
Using the Humphrey Matrix Interface
Viewing Humphrey Matrix Reports
Chapter 8
55
55
56
56
58
60
64
64
Using the Huvitz Interface
Huvitz Interface Overview
Setting Up the Huvitz Equipment
Setting Up the Huvitz Interface
Installing CDR-Mate
Setting Up CDR-Mate
Activating CDR-Mate to Receive Data From Huvitz
Using the Huvitz Interface
Chapter 9
67
67
67
68
70
Using the iMACAM Interface
iMACAM Interface Overview
Setting Up the iMACAM Interface
Using iMACAM Capture
Using iMACAM Review
Chapter 10
71
71
71
72
73
75
75
76
77
79
Using the Konan Interface
Konan Interface Overview
Setting Up the Konan Equipment
Setting Up the Non-CellChek Software
Setting Up the CellChek Software
Setting Up the Konan Interface
Setting Up the Konan Interface with the Non-CellChek Software
Setting Up the Konan Interface with the CellChek Software
Capturing Images with Konan
Reviewing Images with Konan
ii
ExamWRITER Equipment Integration User’s Guide
Contents
Chapter 11
81
81
81
82
83
85
Using the Kowa Interface
Kowa Interface Overview
Setting Up the Kowa Interface
Setting Up the Kowa Camera
Capturing Kowa Images
Reviewing Kowa Images
Chapter 12
87
87
88
88
89
91
Using the Marco Interface
Marco Interface Overview
Setting Up Marco RT & TRS Series Equipment
Setting Up Marco ARK, LM, RKT, & VL Series Equipment
Setting Up the Marco Interface
Using the Marco Interface
Chapter 13
95
95
96
97
100
101
102
Using the NIDEK NAVIS-EX Interface
NIDEK NAVIS-EX Interface Overview
Setting Up NIDEK NAVIS-EX
Configuring NAVIS-EX
Setting Up the NIDEK NAVIS-EX Interface
Using NIDEK NAVIS-EX Capture
Using NIDEK NAVIS-EX Review
Chapter 14
105
105
106
107
109
Using the Optos Interface
Optos Interface Overview
Setting Up the Optos Interface
Using Optos Capture
Using Optos Review
Chapter 15
111
111
112
112
116
118
119
121
124
Using the Reichert Equipment Interface
Reichert Interface Overview
Setting Up the ReichertCapture Software
Setting Up the Reichert Interface
Using the AL-200, AL-500, AL-700, AT-555, PT-100, Reichert 7, &
Reichert 7CR Devices
Using the RK-600 & RK-700 Devices
Using the iPac Pachymeter
Using the Auto Phoropter RS & Auto Phoropter Visutron900
Reviewing Reflex UBM & Foresee PHP Images
127
127
128
130
Using the ReSeeVit Interface
ReSeeVit Interface Overview
Setting Up the ReSeeVit Interface
Using ReSeeVit Capture
Chapter 16
ExamWRITER Equipment Integration User’s Guide
iii
Contents
Chapter 17
133
133
134
137
137
Using the Topcon IMAGEnet Interface
Topcon IMAGEnet Interface Overview
Setting Up the Topcon IMAGEnet Interface
Using the Topcon IMAGEnet Interface
Reviewing Topcon IMAGEnet Images
Chapter 18
141
141
142
142
145
147
Using the Topcon CL, KR, & RM Interface
Topcon Interface Overview
Setting Up the Topcon CL Equipment
Setting Up the Topcon KR & RM Equipment
Setting Up the Topcon Interface
Using the Topcon Interface
Chapter 19
149
149
149
150
151
Using the Topcon CT-80 Interface
Topcon CT-80 Interface Overview
Setting Up the Topcon CT-80 Equipment
Setting Up the Topcon CT-80 Interface
Using the Topcon CT-80 Interface
Chapter 20
153
153
153
154
155
Using the Topcon CV-5000, CV-5000S, CV-3000, & CV-2500
Interface
Topcon Interface Overview
Setting Up the Topcon Equipment
Setting Up the Topcon Interface
Using the Topcon Interface
Chapter 21
159
159
160
160
160
161
Using the Vmax Vision PSF Refractor Interface
Vmax Vision PSF Refractor Interface Overview
Setting Up the Vmax Vision PSF Refractor
Setting Up Shared Folders
Setting Up the Vmax Vision PSF Refractor Interface
Using the Vmax Vision PSF Refractor Interface
Chapter 22
165
165
166
168
172
173
Using the Woodlyn Interface
Woodlyn Interface Overview
Setting Up the Woodlyn Interface
Configuring the Woodlyn Interface
Using Woodlyn Capture
Using Woodlyn Review
iv
ExamWRITER Equipment Integration User’s Guide
Contents
Chapter 23
175
175
175
176
176
177
Using the Zeiss VISUPAC Interface
Zeiss VISUPAC Interface Overview
Setting Up the Zeiss VISUPAC Equipment
Setting Up the Zeiss VISUPAC Interface
Capturing Images with Zeiss VISUPAC
Reviewing Images with Zeiss VISUPAC
Chapter 24
179
179
180
180
181
Using the Zeiss i.Profiler & i.Scription Interface
Zeiss i.Profiers & i.Scription Interface Overview
Setting Up the Zeiss i.Profiler Equipment & i.Scription Software
Setting Up the Zeiss i.Profiler & i.Scription Interface
Using the Zeiss i.Profiler & i.Scription Interface
Chapter 25
187
187
187
188
190
Using the Zeiss FORUM Interface
Zeiss FORUM Interface Overview
Setting Up Zeiss FORUM
Setting Up the Zeiss FORUM Viewer Interface
Viewing Data and Images with Zeiss FORUM
Chapter 26
191
191
191
192
193
Using the OCULUS Interface
OCULUS Interface Overview
Setting Up the OCULUS Interface
Capturing OCULUS Visual Fields Data
Reviewing OCULUS Visual Fields Data
Appendix A
195
Setting Up Serial Cable Interfaces with Terminal Services
Index
199
ExamWRITER Equipment Integration User’s Guide
v
Contents
vi
ExamWRITER Equipment Integration User’s Guide
Using the CenterVue Interface
1
In this chapter:
CenterVue
Interface
Overview
•
CenterVue Interface Overview, 1
•
Setting Up the CenterVue DRS Interface, 1
•
Capturing Images with CenterVue DRS, 2
•
Reviewing CenterVue DRS Images, 5
ExamWRITER interfaces with the CenterVue DRS software before the capture
process when patient information is transferred from ExamWRITER into
CenterVue DRS to create or update retinal images and during the review process
when the newly created or existing CenterVue retinal images are available for
on-screen review.
You do not need to install any CenterVue software on the computer on which you
are using ExamWRITER. The CenterVue equipment interface saves equipment
data to the CenterVue database and ExamWRITER reads that database through
a wireless or wired network connection. The equipment data can be shared and
accessed at multiple computers, as long as the computers are networked
properly. The CenterVue interface uses a wireless router or wired Ethernet cable
to connect to the computer.
NOTES
• Back up your CenterVue DRS data. Contact CenterVue for
recommended backup procedures.
• If you began using CenterVue DRS before ExamWRITER, your
CenterVue Patient IDs will be different than your ExamWRITER
patient numbers; therefore, your CenterVue and ExamWRITER
patients will not be linked to each other. To automatically link your
CenterVue and ExamWRITER patients, run the
ExamWRITER/CenterVue Patient Synchronization Utility. To open
this utility and view instructions on how to use the utility, go to
www.officemate.net/omkb/Article.aspx?id=30730.
Setting Up the
CenterVue
DRS Interface
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
2. Select the Write captured equipment data to a file check box if you want
ExamWRITER to record data received from CenterVue in a file for
debugging purposes; otherwise, go to step 3.
®
ExamWRITER Equipment Integration User’s Guide
1
Chapter 1
Using the CenterVue Interface
Capturing Images with CenterVue DRS
3. Select the check box next to CenterVue Capture.
4. Type the network address in the IP Address text box.
NOTE
To find the device’s network address, go to the CenterVue
equipment, open the Settings panel and go to the Network tab.
5. Select the check box next to CenterVue Review.
6. Type the network address in the IP Address text box.
7. Click Save.
8. Click Exit.
9. Close and reopen ExamWRITER to activate your changes.
Capturing
Images with
CenterVue
DRS
2
To capture images from your CenterVue DRS equipment, perform the following
steps:
1. Create a new blank exam or open a saved exam in ExamWRITER.
2. Click the Interfaces icon on the ExamWRITER chart window.
ExamWRITER Equipment Integration User’s Guide
Chapter 1
Using the CenterVue Interface
Capturing Images with CenterVue DRS
3. Select CenterVue:Capture.
The CenterVue Capture window opens with the patient’s last name, first
name, middle initial, gender, date of birth, patient ID, preloaded from
ExamWRITER.
ExamWRITER Equipment Integration User’s Guide
3
Chapter 1
Using the CenterVue Interface
Capturing Images with CenterVue DRS
4. Depending on whether the patient is found, perform one of the following
steps.
If…
Then…
The patient is found on
the CenterVue
equipment
1. Go to the CenterVue equipment and
capture the necessary images.
If the patient is not
found, but you believe
the patient has a record
in the CenterVue
equipment
1. Note the OM Patient ID. Select and copy it if
needed.
2. Come back to ExamWRITER and click Exit
on the CenterVue Capture dialog box.
2. Click View Patients.
The DRS console opens.
3. Search for the patient and click the patient’s
name.
4. Click the Edit icon.
5. Type or past the OM Patient ID in the Code
text box.
6. Click Save.
7. Go to the CenterVue equipment and
capture the necessary images.
8. Come back to ExamWRITER and click Exit
on the CenterVue Capture dialog box.
If the patient is not found
1. Click Create New Patient.
The relevant demographic data is
transferred to the CenterVue DRS
equipment.
2. Go to the CenterVue equipment and
capture the necessary images.
3. Come back to ExamWRITER and click Exit
on the CenterVue Capture dialog box.
The CenterVue icon displays under the Images, Drawings and Graphics
category bar in ExamWRITER after you close and reopen ExamWRITER.
4
ExamWRITER Equipment Integration User’s Guide
Chapter 1
Using the CenterVue Interface
Reviewing CenterVue DRS Images
4. Follow the instructions in “Reviewing CenterVue DRS Images” on page 5 to
review the patient’s captured retinal images.
Reviewing
CenterVue
DRS Images
1. Open ExamWRITER.
2. Create a new blank exam or open a saved exam for patient for whom you
have previously captured retinal images.
ExamWRITER Equipment Integration User’s Guide
5
Chapter 1
Using the CenterVue Interface
Reviewing CenterVue DRS Images
3. Click the Interfaces icon on the ExamWRITER chart window and select
CenterVue:Review.
OR
Click the CenterVue icon under the ExamWRITER Images, Drawings and
Graphics category bar.
The DRS console opens and displays the patient’s history.
4. Follow the CenterVue DRS instructions to use the application.
5. Close the DRS console to return to ExamWRITER.
6
ExamWRITER Equipment Integration User’s Guide
Using the Eagle Eye 3000 Interface
2
In this chapter:
Eagle Eye
3000 Overview
•
Eagle Eye 3000 Overview, 7
•
Setting Up the Eagle Eye 3000 Interface, 7
•
Using the Eagle Eye 3000 Interface, 8
The Eagle Eye 3000 equipment interface saves equipment data to the
ExamWRITER database, which then can be accessed by all computers
networked to look at that database. The equipment data can be shared and
accessed at multiple computers, as long as the computers are networked
properly and the databases are properly mapped and shared. The Eagle Eye
3000 interface uses a serial RS-232 cable or null modem to connect to the
computer or Terminal Server.
Setting Up the
Eagle Eye
3000 Interface
NOTE
You must set up the Eagle Eye 3000 interface from the computer
connected to the Eagle Eye 3000 equipment.
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
2. Select the Write captured equipment data to a file check box if you want
ExamWRITER to record data received from the Eagle Eye 3000 in a file for
debugging purposes; otherwise, go to step 3.
3. Select the check boxes next to the equipment interfaces that you are setting
up.
®
ExamWRITER Equipment Integration User’s Guide
7
Chapter 2
Using the Eagle Eye 3000 Interface
Using the Eagle Eye 3000 Interface
4. Select a communication port number from the Comm Port drop-down
menu; otherwise, go to step 5.
The Comm Port default is 1.
5. Click Save.
6. Click and reopen ExamWRITER to activate your changes.
Using the
Eagle Eye
3000 Interface
1. Ensure that ExamWRITER is open.
2. Create a new blank exam or open a saved exam for an existing patient.
3. Click the Interfaces icon on the ExamWRITER chart window.
4. Select VisionMedInstrum:Eagle Eye 3000.
The Officemate and Examwriter Equipment Interface window opens.
5. Capture a measurement on the Eagle Eye 3000 pachymeter by following the
instructions in the Eagle Eye 3000 Owner’s Manual.
6. Press Print on the Eagle Eye 3000 to transmit the average measurement of
the four measurements displayed on the pachymeter to ExamWRITER.
8
ExamWRITER Equipment Integration User’s Guide
Chapter 2
Using the Eagle Eye 3000 Interface
Using the Eagle Eye 3000 Interface
7. Click OD or OS to accept the measurement and record it for the appropriate
eye.
NOTES
• Click Redo to re-capture the measurement.
• Click Cancel to cancel the test.
8. Click Test Other Eye to instruct ExamWRITER to wait for additional
measurement data from the Eagle Eye 3000.
The Officemate and Examwriter Equipment Interface window opens.
9. Capture a measurement on the Eagle Eye 3000 pachymeter by following the
instructions in the Eagle Eye 3000 Owner’s Manual.
10. Press Print on the Eagle Eye 3000 to transmit the average measurement of
the four measurements displayed on the pachymeter to ExamWRITER.
11. Click OD or OS to accept the measurement and record it for the appropriate
eye.
NOTES
• Click Redo to re-capture the measurement.
• Click Cancel to cancel the test.
ExamWRITER Equipment Integration User’s Guide
9
Chapter 2
Using the Eagle Eye 3000 Interface
Using the Eagle Eye 3000 Interface
12. Click Save/Exit to save the OD and OS measurements in the patient’s
ExamWRITER database.
The average pachymetry measurements will be displayed under the
Examination / Special Eye Health Tests category bar in ExamWRITER.
10
ExamWRITER Equipment Integration User’s Guide
Using the EyeScape Interface
3
In this chapter:
EyeScape
Interface
Overview
•
EyeScape Interface Overview, 11
•
Setting Up the EyeScape Interface, 11
•
Capturing EyeScape Images, 13
•
Reviewing EyeScape Images, 15
ExamWRITER interfaces with Synemed EyeScape Digital Imaging Systems
before the capture process when patient information is transferred from
ExamWRITER into EyeScape to create or update retinal images and during the
review process when the newly created or existing EyeScape retinal images are
available for on-screen review.
The EyeScape equipment interface saves equipment data to the EyeScape
database and ExamWRITER reads that database through a mapped drive. The
equipment data can be shared and accessed at multiple computers, as long as
the computers are networked properly and the databases are properly mapped
and shared. The EyeScape interface uses a serial RS-232 cable or null modem to
connect to the computer. The EyeScape interface can only use one camera per
Terminal Server.
NOTES
• You must install ExamWRITER and the EyeScape software on the
computer to which the EyeScape equipment is connected; then
you can install the EyeScape software on the server on which
ExamWRITER is installed. Contact Synemed to determine if the
EyeScape software will run on an OS server and if any additional
licensing needs to be purchased when using it with Terminal
Server.
• Back up your EyeScape data! Contact Synemed for
recommended backup procedures.
Setting Up the
EyeScape
Interface
NOTE
You must set up the EyeScape interface from the computer
connected to the camera equipment.
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
®
ExamWRITER Equipment Integration User’s Guide
11
Chapter 3
Using the EyeScape Interface
Setting Up the EyeScape Interface
2. Select the Write captured equipment data to a file check box if you want
ExamWRITER to record data received from EyeScape in a file for debugging
purposes; otherwise, go to step 3.
3. Select the check boxes next to the equipment interfaces that you are setting
up.
4. Select the drive and shared folder where the EyeScape database is stored.
NOTE
Typically, the EyeScape database is stored in the Program
Files\Synemed\EyeScapeNE directory on the computer acting
as the Synemed “server.” To locate the exact database path,
open Eyescape, click Program, and then select Service
Settings. The Service Setup window opens and displays the
database path that you must select in the Setup ExamWRITER
Equipment Interfaces window.
5. Click Save.
12
ExamWRITER Equipment Integration User’s Guide
Chapter 3
Using the EyeScape Interface
Capturing EyeScape Images
6. Click Exit.
7. Close and reopen ExamWRITER to activate your changes.
Capturing
EyeScape
Images
1. Ensure that ExamWRITER is open.
2. Create a new blank exam or open a saved exam for an existing patient.
3. Click the Interfaces icon on the ExamWRITER chart window.
4. Select Synemed:Capture.
The EyeScape window opens with the patient’s name and date of birth
pre-loaded from ExamWRITER.
5. Follow the EyeScape capture instructions to capture the patient’s retinal
images.
ExamWRITER Equipment Integration User’s Guide
13
Chapter 3
Using the EyeScape Interface
Capturing EyeScape Images
6. After you have completed capturing the images, click Exit Program.
The EyeScape NE icon will be displayed under the Images, Drawings and
Graphics category bar in ExamWRITER after you close and reopen
ExamWRITER.
7. Follow the instructions in “Reviewing EyeScape Images” on page 15 to
review the patient’s captured retinal images.
14
ExamWRITER Equipment Integration User’s Guide
Chapter 3
Using the EyeScape Interface
Reviewing EyeScape Images
Reviewing
EyeScape
Images
NOTES
• The EyeScape review software can only be used from within
ExamWRITER when a patient’s retinal images have previously
been captured using ExamWRITER and the EyeScape capture
interface.
• Both the EyeScape review software and ExamWRITER must be
installed and set up on all workstations on which you want to
review images. Also, a path to the location where the EyeScape
database is located must be set up. For information on setting up
the path to the EyeScape database in the ExamWRITER
equipment interface, go to “Setting Up the EyeScape Interface”
on page 11. 1. Ensure that ExamWRITER is open.
2. Create a new blank exam or open a saved exam for an existing patient for
whom you have previously captured retinal images.
3. Click the Interfaces icon on the ExamWRITER chart window and select
Synemed:Review.
OR
Click the EyeScape NE icon under the ExamWRITER Images, Drawings
and Graphics category bar.
The EyeScape window opens and displays the patient’s history.
4. Follow the EyeScape review instructions to use the application.
5. Click Exit Program to close EyeScape to return to ExamWRITER.
ExamWRITER Equipment Integration User’s Guide
15
Chapter 3
Using the EyeScape Interface
Reviewing EyeScape Images
16
ExamWRITER Equipment Integration User’s Guide
Using the EyeTel Interface
4
In this chapter:
EyeTel Interface Overview
•
EyeTel Interface Overview, 17
•
Setting Up the EyeTel Interface, 17
•
Capturing Images with EyeTel, 18
•
Reviewing Images from EyeTel, 20
ExamWRITER interfaces with the EyeTel Imaging software before the capture
process when patient information is transferred from ExamWRITER into the EyeTel
DigiScope to create or update retinal images and during the review process when
the newly created or existing EyeTel DigiScope retinal images are available for
on-screen review.
The EyeTel equipment interface saves equipment data to the EyeTel database
and ExamWRITER reads that database through a mapped drive. The equipment
data can be shared and accessed at multiple computers, as long as the
computers are networked properly and the databases are properly mapped and
shared. The EyeTel equipment interface uses a serial RS-232 cable or null modem
to connect to the computer or Terminal Server. The EyeTel equipment interface
can use multiple serial ports per PC.
NOTES
• You must install the EyeTel Imaging software on the same
workstation as ExamWRITER.
• Before you begin using the EyeTel equipment interface, you must
set up the EyeTel equipment. Refer to the EyeTel DigiScope
Instrument User’s Guide for detailed information on setting up the
EyeTel equipment.
Setting Up the
EyeTel Interface
NOTES
• You must set up the EyeTel interface from the computer connected
to the EyeTel equipment.
• You must be using ExamWRITER version 7.3 or above to interface
with the EyeTel equipment.
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
®
ExamWRITER Equipment Integration User’s Guide
17
Chapter 4
Using the EyeTel Interface
Capturing Images with EyeTel
2. Select the Write captured equipment data to a file check box if you want
ExamWRITER to record data received from EyeTel in a file for debugging
purposes; otherwise, go to step 3.
3. Select the check boxes next to the equipment interfaces that you are setting
up.
4. Select the drive and folder where the equipment data is stored.
5. Click Save.
6. Click Exit.
7. Close and reopen ExamWRITER to activate your changes.
Capturing
Images with
EyeTel
1. Ensure that ExamWRITER is open.
18
ExamWRITER Equipment Integration User’s Guide
2. Create a new blank exam or open a saved exam for an existing patient.
3. Click the Interfaces icon on the ExamWRITER chart window.
Chapter 4
Using the EyeTel Interface
Capturing Images with EyeTel
4. Select EyeTel:Capture.
The EyeTel Imaging window opens after the patient’s first and last name,
social security number, patient ID, date of birth, and gender are pre-loaded
in EyeTel DigiScope from ExamWRITER.
5. Follow the EyeTel instructions to capture the patient’s retinal images on the
DigiScope.
The EyeTel Imaging icon will be displayed under the Images, Drawings and
Graphics category bar in ExamWRITER after you close and reopen
ExamWRITER.
6. Follow the instructions in “Reviewing Images from EyeTel” on page 20 to
review the patient’s captured retinal images.
ExamWRITER Equipment Integration User’s Guide
19
Chapter 4
Using the EyeTel Interface
Reviewing Images from EyeTel
Reviewing
Images from
EyeTel
NOTE
Both the EyeTel review software and ExamWRITER must be installed
and set up on all workstations on which you want to review images.
Also, a path to the location where the EyeTel database is located
must be set up. For information on setting up the path to the EyeTel
database in the ExamWRITER equipment interface, go to “Setting Up
the EyeTel Interface” on page 17. 1. Ensure that ExamWRITER is open.
2. Create a new blank exam or open a saved exam for an existing patient for
whom you have previously captured retinal images.
3. Click the Interfaces icon on the ExamWRITER chart window and select
EyeTel:Review.
OR
Click the EyeTel Imaging icon under the ExamWRITER Images, Drawings
and Graphics category bar.
The EyeTel Imaging window opens and then the DigiScope displays the
patient’s images and reports.
20
ExamWRITER Equipment Integration User’s Guide
Using the Heidelberg Interface
5
In this document:
Heidelberg
Interface
Overview
•
Heidelberg Interface Overview, 21
•
Setting Up the Heidelberg Interface, 22
•
Capturing Heidelberg Images, 22
•
Reviewing Heidelberg Images, 24
ExamWRITER interfaces with the Heidelberg HRT2/HRT3 and Spectralis OCT
before the capture process when patient information is transferred from
ExamWRITER into Eye Explorer to create or update retinal images and during the
review process when the newly created or existing Eye Explorer retinal images are
available for on-screen review.
The Heidelberg equipment interface saves equipment data to the Heidelberg
database and ExamWRITER reads that database through a mapped drive. The
equipment data can be shared and accessed at multiple computers, as long as
the computers are networked properly and the databases are properly mapped
and shared. The Heidelberg interface uses a serial RS-232 cable or null modem
to connect to the computer. The Heidelberg interface can only use one camera
per Terminal Server.
NOTES
• You must install ExamWRITER and the Eye Explorer software on
the computer to which the Heidelberg equipment is connected;
then you can install the Eye Explorer software on the server on
which ExamWRITER is installed. Contact Heidelberg Engineering
to determine if the Eye Explorer software will run on an OS server
and if any additional licensing needs to be purchased when using
it with Terminal Server.
• Back up your Heidelberg data! Contact Heidelberg Engineering
for recommended backup procedures.
®
ExamWRITER Equipment Integration User’s Guide
21
Chapter 5
Using the Heidelberg Interface
Setting Up the Heidelberg Interface
Setting Up the
Heidelberg
Interface
NOTES
• You must set up the Heidelberg interface from the computer
connected to the Heidelberg equipment.
• You must be using ExamWRITER version 7.3 or above to interface
with the Heidelberg equipment.
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
2. Select the Write captured equipment data to a file check box if you want
ExamWRITER to record data received from Heidelberg in a file for
debugging purposes; otherwise, go to step 3.
3. Select the check boxes next to the equipment interfaces that you are setting
up.
4. Select the drive and shared folder where the Heidelberg data is written.
NOTE
Typically, the shared folder where the Heidelberg HRT2/HRT3
data is written is HRT\heyex\Data\data on the computer acting
as the Heidelberg server. See the Heidelberg setup
documentation for more information on obtaining this path.
5. Click Save.
6. Click Exit.
7. Close and reopen ExamWRITER to activate your changes.
Capturing
Heidelberg
Images
1. Ensure that ExamWRITER is open.
22
ExamWRITER Equipment Integration User’s Guide
2. Create a new blank exam or open a saved exam for an existing patient.
3. Click the Interfaces icon on the ExamWRITER chart window.
Chapter 5
Using the Heidelberg Interface
Capturing Heidelberg Images
4. Select Heidelberg:Capture.
The Heidelberg Eye Explorer window opens with the patient’s name and
birth date pre-loaded from ExamWRITER.
5. Follow the Heidelberg instructions to capture the patient’s retinal images.
ExamWRITER Equipment Integration User’s Guide
23
Chapter 5
Using the Heidelberg Interface
Reviewing Heidelberg Images
6. Click Exit Program to return to ExamWRITER.
The Heidelberg HRT icon will be displayed under the Images, Drawing,
Graphics category bar in ExamWRITER after you close and reopen
ExamWRITER.
7. Follow the instructions in “Reviewing Heidelberg Images” on page 24 to
review the patient’s captured retinal images.
Reviewing
Heidelberg
Images
NOTES
• The Heidelberg review software can only be used from within
ExamWRITER when a patient’s retinal images have previously
been captured using ExamWRITER and the Heidelberg Eye
Explorer.
• Both the Heidelberg review software and ExamWRITER must be
installed and set up on all workstations on which you want to
review images. Also, a path to the location where the Heidelberg
database is located must be set up. For information on setting up
the path to the Heidelberg database in the ExamWRITER
equipment interface, go to “Setting Up the Heidelberg Interface”
on page 22.
1. Ensure that ExamWRITER is open.
2. Create a new blank exam or open a saved exam for an existing patient for
whom you have previously captured retinal images.
24
ExamWRITER Equipment Integration User’s Guide
Chapter 5
Using the Heidelberg Interface
Reviewing Heidelberg Images
3. Click the Interfaces icon on the ExamWRITER chart window and select
Heidelberg:Review.
OR
Click the Heidelberg HRT icon under the ExamWRITER Images, Drawing,
Graphics category bar.
The Heidelberg Eye Explorer window opens and displays the patient’s
images.
4. Follow the Heidelberg review instructions to review the images.
5. Click Exit Program to return to ExamWRITER.
ExamWRITER Equipment Integration User’s Guide
25
Chapter 5
Using the Heidelberg Interface
Reviewing Heidelberg Images
26
ExamWRITER Equipment Integration User’s Guide
Using the Humphrey® HFA II-i
Interface
6
In this chapter:
Humphrey
Interface
Overview
•
Humphrey Interface Overview, 27
•
Setting Up the Humphrey Interface, 27
•
Setting Up the Humphrey Equipment, 30
•
Using the Humphrey Interface, 40
•
Viewing Humphrey Analysis Reports, 47
ExamWRITER receives data reports from the Humphrey Field Analyzer HFA II-i
Series equipment and links the reports to the ExamWRITER exam record.
ExamWRITER does not send clinical data to the Humphrey equipment.
The Humphrey equipment interface saves data reports to the DATA folder on the
ExamWRITER server, which then can be accessed by all computers networked to
look at that folder. The data reports can be shared and accessed at multiple
computers, as long as the computers are networked properly and the folder is
properly mapped and shared. The Humphrey interface uses a mapped network
drive to connect to the computer or Terminal Server. Ensure that you have
purchased the networking license from Carl Zeiss Meditech, Inc., so that you can
properly network your Humphrey Field Analyzer and interface with ExamWRITER.
NOTE
Setting Up the
Humphrey
Interface
Even if you are familiar with using the Humphrey equipment, you will
still need to follow the instructions in “Using the Humphrey Interface”
on page 40 so that you correctly recall patient data and transfer the
tests.
Before you begin using the Humphrey equipment interface, set up the Humphrey
interface in ExamWRITER.
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
2. Select the Write captured equipment data to a file check box if you want
ExamWRITER to record data received from the Humphrey equipment in a
file for debugging purposes; otherwise, go to step 3.
®
ExamWRITER Equipment Integration User’s Guide
27
Chapter 6
Using the Humphrey® HFA II-i Interface
Setting Up the Humphrey Interface
3. Select the check boxes next to the equipment interfaces that you are setting
up.
4. Highlight the Humphrey Capture HFAIIi device.
5. Type the Humphrey equipment’s serial number in the Data File Start Chars
text box.
NOTES
• Your Humphrey equipment’s serial number is listed in the OA
Web License and on a sticker on the equipment.
• Omit the “I” in the serial number when you type it into the
Data File Start Chars text box. For example, if your serial
number is 123I-34432, type, 123-34432 in the Data File
Start Chars text box.
28
ExamWRITER Equipment Integration User’s Guide
Chapter 6
Using the Humphrey® HFA II-i Interface
Setting Up the Humphrey Interface
6. Select the drive and shared folder (EMR or EMR Export) where the
Humphrey equipment data is written.
7. Click Save.
8. Highlight the Humphrey TestConnection device and repeat steps 5–7.
9. Click Exit.
10. Close and reopen ExamWRITER to activate your changes.
ExamWRITER Equipment Integration User’s Guide
29
Chapter 6
Using the Humphrey® HFA II-i Interface
Setting Up the Humphrey Equipment
Setting Up the
Humphrey
Equipment
Before you begin using the Humphrey equipment interface, set up the Humphrey
Field Analyzer HFA II-i Series equipment.
NOTE
See the Humphrey HFA II-i Series User Manual for more information
on setting up the Humphrey equipment.
1. Ensure that the Humphrey Field Analyzer HFA II-i Series is connected to your
network with a network cable.
2. Ensure that ExamWRITER is open.
3. Press the System Setup icon on the Humphrey Main Menu window.
The System Setup window opens.
NOTE
30
In order to create and view the Overview, Change Analysis, and
Three-in-One reports, you must press Print Setup on the
System Setup window and then press Print Destination.
Ensure that Export Image Files is selected from the first
drop-down menu and EMR/PMS Host is selected from the
Export to drop-down menu.
ExamWRITER Equipment Integration User’s Guide
Chapter 6
Using the Humphrey® HFA II-i Interface
Setting Up the Humphrey Equipment
4. Press Communications Setup.
5. Press Data Export Setup.
ExamWRITER Equipment Integration User’s Guide
31
Chapter 6
Using the Humphrey® HFA II-i Interface
Setting Up the Humphrey Equipment
6. Select Shared Folder from the File Server Access Protocol drop-down
menu and follow the instructions below:
a. Press Shared Folder Setup.
a. Press User Name.
b. Type your user name and press Enter.
c. Press Password.
d. Type your password and press Enter.
7. Press Browse for Shared Folder.
The Data Export – Select Shared Folder window opens.
NOTE
If you need help browsing to shared folders, call Carl Zeiss
Meditec at 1.800.341.6968.
8. Navigate to and select the workgroup that you have shared.
9. Press Proceed.
10. Navigate to and select the computer that you have shared.
11. Press Proceed.
12. Navigate to and select the directory that you have shared (DATA or DATA
Export).
13. Press Select Folder.
32
ExamWRITER Equipment Integration User’s Guide
Chapter 6
Using the Humphrey® HFA II-i Interface
Setting Up the Humphrey Equipment
14. Deselect the Enable Patient Folder check box and press Save.
15. Press HFA Network Setup.
ExamWRITER Equipment Integration User’s Guide
33
Chapter 6
Using the Humphrey® HFA II-i Interface
Setting Up the Humphrey Equipment
16. If there is not an “X” in the Obtain HFA TCP/IP Settings Automatically
check box, press the empty check box to select it; otherwise, go to step 9.
17. Press Workgroup Name.
18. Type the workgroup or domain name of the computer with the shared drive
and press Enter.
19. Press Save.
20. Press EMR/PMS Export Setup.
34
ExamWRITER Equipment Integration User’s Guide
Chapter 6
Using the Humphrey® HFA II-i Interface
Setting Up the Humphrey Equipment
21. Select Shared Folder from the File Server Access Protocol drop-down
menu and follow the instructions below:
a. Press Shared Folder Setup.
a. Press User Name.
b. Type your user name and press Enter.
c. Press Password.
d. Type your password and press Enter.
22. Press Browse for Shared Folder.
The EMR/PMS Export – Select Shared Folder window opens.
NOTE
If you need help browsing to shared folders, call Carl Zeiss
Meditec at 1.800.341.6968.
23. Navigate to and select the workgroup that you have shared.
24. Press Proceed.
25. Navigate to and select the computer that you have shared.
26. Press Proceed.
27. Navigate to and select the directory that you have shared (EMR or EMR
Export).
28. Press Select Folder.
29. On the ExamWRITER main window, click Tools, and select Equipment
Integration Setup.
ExamWRITER Equipment Integration User’s Guide
35
Chapter 6
Using the Humphrey® HFA II-i Interface
Setting Up the Humphrey Equipment
30. Highlight the Humphrey Test Connection device and click
TestConnection.
A connection test occurs in ExamWRITER.
31. Press Proceed.
The “Connection test in progress, please wait...” message appears on the
Humphrey equipment. This test can take up to a few minutes to perform.
32. Press Save.
A second connection test may occur on the Humphrey equipment. This test
can take up to a few minutes. The System Setup window opens after the
connection test has completed.
36
ExamWRITER Equipment Integration User’s Guide
Chapter 6
Using the Humphrey® HFA II-i Interface
Setting Up the Humphrey Equipment
33. Press Work List Setup and follow the instructions below:
a. Select Retrieve Only from the Work List Mode drop-down menu.
b. Select Shared Folder from the File Server Access Protocol drop-down
menu.
34. Press Shared Folder Setup and follow the instructions below:
a. Press User Name.
b. Type your user name and press Enter.
c. Press Password.
d. Type your password and press Enter.
35. Press Browse for Shared Folder.
The Work List – Select Shared Folder window opens.
NOTE
If you need help browsing to shared folders, call Carl Zeiss
Meditec at 1.800.341.6968.
ExamWRITER Equipment Integration User’s Guide
37
Chapter 6
Using the Humphrey® HFA II-i Interface
Setting Up the Humphrey Equipment
36. Navigate to and select the workgroup that you have shared.
37. Press Proceed.
38. Navigate to and select the computer that you have shared.
39. Press Proceed.
40. Navigate to and select the directory that you have shared (EMR or EMR
Export).
41. Press Select Folder.
42. Press Save.
43. On the ExamWRITER main window, click Tools, and select Equipment
Integration Setup.
44. Highlight the Humphrey Test Connection device and click
TestConnection.
A connection test occurs on ExamWRITER.
45. Press Proceed on the Humphrey System Setup window.
The “Connection test in progress, please wait...” message appears on the
Humphrey equipment. This test can take up to a few minutes to perform.
46. Press Done.
38
ExamWRITER Equipment Integration User’s Guide
Chapter 6
Using the Humphrey® HFA II-i Interface
Setting Up the Humphrey Equipment
47. Press Save/Transmit Option and follow the instructions below:
a. Select EMR/PMS Host from the Transfer Destination drop-down menu.
b. Select Save and Transmit from the Save/Transmit Options drop-down
menu.
c. Select XML and Image Files from the Data Format drop-down menu.
48. On the ExamWRITER main window, click Tools, and select Equipment
Integration Setup.
ExamWRITER Equipment Integration User’s Guide
39
Chapter 6
Using the Humphrey® HFA II-i Interface
Using the Humphrey Interface
49. Highlight the Humphrey Test Connection device and click
TestConnection.
A connection test occurs on ExamWRITER.
50. Press Proceed on the Humphrey System Setup window.
The “Connection test in progress, please wait...” message appears on the
Humphrey equipment. This test can take up to a few minutes to perform.
Using the
Humphrey
Interface
1. Ensure that the Humphrey Field Analyzer HFA II-i Series is connected to your
network with a network cable.
2. Ensure that ExamWRITER is open.
3. Create a new blank exam or open a saved exam for an existing patient.
4. Click the Interfaces icon on the ExamWRITER chart window.
5. Select Humphrey: Capture HFAIIi.
The Officemate/Examwriter Equipment Interface window opens and
displays a “Waiting for Humphrey: Capture HFAIIi Exam Data…” message.
40
ExamWRITER Equipment Integration User’s Guide
Chapter 6
Using the Humphrey® HFA II-i Interface
Using the Humphrey Interface
ExamWRITER sends patient demographic information to the Humphrey
equipment.
NOTE
If you used your Humphrey equipment prior to integrating it with
ExamWRITER, you may need to change the patient’s number in
the Humphrey equipment to match the patient’s number in
ExamWRITER.
6. Insert a floppy disk into the floppy drive on the Humphrey equipment.
7. Press one of the ten test buttons on the Humphrey Main Menu window.
ExamWRITER Equipment Integration User’s Guide
41
Chapter 6
Using the Humphrey® HFA II-i Interface
Using the Humphrey Interface
8. Select Right or Left from the Select Eye window.
The Patient Data 1 window opens.
9. If the incorrect patient is displayed in the window, follow the instructions
below; otherwise, go to step 10:
a. Press Clear Patient Data.
b. Press Clear Data.
10. Press Recall Patient Data.
42
ExamWRITER Equipment Integration User’s Guide
Chapter 6
Using the Humphrey® HFA II-i Interface
Using the Humphrey Interface
11. Select Work List from the Source drop-down menu.
12. Press Proceed.
The Work List window opens.
13. Highlight the current patient.
14. Press Proceed.
The Patient Data 1 window opens.
15. Verify that the patient demographic data is correctly displayed on the
window.
ExamWRITER Equipment Integration User’s Guide
43
Chapter 6
Using the Humphrey® HFA II-i Interface
Using the Humphrey Interface
16. Press Proceed.
The Humphrey equipment checks for a floppy disk and ensures that you are
able to write to the disk. The test screening window opens.
17. Press Start on the test screening window.
The “Initialize Gaze Monitory. Instruct Patient to Look at Fixation Lights”
message appears.
44
ExamWRITER Equipment Integration User’s Guide
Chapter 6
Using the Humphrey® HFA II-i Interface
Using the Humphrey Interface
18. Press Start.
The “Initializing Gaze Successful” message appears.
19. Press Continue.
The test is performed and the “Do You Wish to Save This Test For: Patient?”
message appears.
20. Press Yes.
The “Saving Test In Progress. Please Wait” message appears and the
Humphrey data is transferred to ExamWRITER. ExamWRITER parses and
saves the data and receives the reports.
NOTE
If you want to test another eye, press Test Other Eye and repeat
steps 17–20; otherwise, go to step 21.
ExamWRITER Equipment Integration User’s Guide
45
Chapter 6
Using the Humphrey® HFA II-i Interface
Using the Humphrey Interface
21. Click Done after all of the reports have been received by ExamWRITER.
22. Press the Main Menu or Return icon on the Humphrey equipment.
The Main Menu window opens.
46
ExamWRITER Equipment Integration User’s Guide
Chapter 6
Using the Humphrey® HFA II-i Interface
Viewing Humphrey Analysis Reports
Viewing
Humphrey
Analysis
Reports
1. Open the Humphrey Analysis Reports window using one of the following
methods:

Click the Interfaces icon and select Humphrey:Review HFA.

Click on the Plan - Management tab and click the Zeiss HFA IIi icon.
2. Double-click on a report to open, view, and print it.
ExamWRITER Equipment Integration User’s Guide
47
Chapter 6
Using the Humphrey® HFA II-i Interface
Viewing Humphrey Analysis Reports
48
ExamWRITER Equipment Integration User’s Guide
Using the Humphrey® Matrix
Interface
7
In this chapter:
Humphrey
Matrix Overview
•
Humphrey Matrix Overview, 49
•
Setting Up the Humphrey Matrix Equipment, 49
•
Setting Up the Humphrey Matrix Interface, 50
•
Using the Humphrey Matrix Interface, 51
•
Viewing Humphrey Matrix Reports, 53
ExamWRITER receives data reports from the Humphrey Matrix equipment and
stores the reports in the ExamWRITER exam record. ExamWRITER does not send
clinical data to the Humphrey Matrix equipment.
The Humphrey Matrix equipment interface saves equipment data to the
ExamWRITER database, which then can be accessed by all computers
networked to look at that database. The Humphrey Matrix interface uses a RS-232
cable or null modem to connect to the computer. You must connect to the
computer using COM 1 port. The Humphrey Matrix interface cannot connect to a
Terminal Server.
Setting Up the
Humphrey
Matrix Equipment
Before you begin using the Humphrey Matrix equipment interface, you must set
up the Humphrey Matrix equipment. Refer to the Humphrey Matrix Visual Field
Instrument User’s Guide and the Humphrey Matrix Connectivity Cable
Instructions for detailed information on setting up the Humphrey Matrix
equipment.
You must be using the Humphrey Matrix 7.02 or above in order to interface with
ExamWRITER. If you are unsure of your Humphrey Matrix version number, press
the F6 key or click the F6 tab on the Matrix window to open the Matrix Help
window. The SBC Software Version in the System Information frame must be
07.02.01[0] or above and the MCU Software Version must be 02.03.01[0] or
above. If the software versions do not meet the requirements stated above,
contact your Humphrey Matrix vendor for support.
®
ExamWRITER Equipment Integration User’s Guide
49
Chapter 7
Using the Humphrey® Matrix Interface
Setting Up the Humphrey Matrix Interface
Setting Up the
Humphrey
Matrix Interface
Before you begin using the Humphrey equipment interface, set up the Humphrey
Matrix interface in ExamWRITER.
NOTES
• You must set up the Humphrey Matrix interface from the computer
connected to the Humphrey Matrix equipment.
• You must be using ExamWRITER version 7.3 or above to interface
with the Humphrey Matrix equipment.
• You must have the .NET Framework version 2.0.50727 or above
installed on the computer system running ExamWRITER and
connected to the Humphrey Matrix equipment. You can download
this .NET Framework version at
http://www.microsoft.com/downloads/details.aspx?FamilyID=085
6EACB-4362-4B0D-8EDD-AAB15C5E04F5&displaylang=en. The
.NET Framework requires Windows Installer version 3 or above.
You can download this Windows Installer version at
http://www.microsoft.com/downloads/details.aspx?familyid=8894
82fc-5f56-4a38-b838-de776fd4138c&displaylang=en.
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
2. Select the Write captured equipment data to a file check box if you want
ExamWRITER to record data received from the Humphrey Matrix equipment
in a file for debugging purposes; otherwise, go to step 3.
3. Deselect the check boxes next to the equipment interfaces that you are not
using.
4. Highlight the Humphrey Matrix device.
50
ExamWRITER Equipment Integration User’s Guide
Chapter 7
Using the Humphrey® Matrix Interface
Using the Humphrey Matrix Interface
5. Select 1 from the Comm Port drop-down menu.
NOTE
If you change the communication port after initially setting it up,
you must restart the computer so that the OM Matrix Gateway
is reset to the new communication port.
6. Click Save.
7. Click Exit.
8. Close and reopen ExamWRITER to activate your changes.
Using the
Humphrey
Matrix Interface
NOTE
Ensure that the OM Matrix Gateway installed and open on the
computer on which ExamWRITER is installed and that is connected
to the Humphrey Matrix equipment (go to
www.officemate.net/omkb/article.aspx?id=19629 if you do not have
the OM Matrix Gateway program and you need to download and
install it). If the OfficeMate logo is visible in the Windows task bar at
the bottom of the screen, then OM Matrix Gateway is open. If you
right-click on the OfficeMate logo in the Windows task bar, select
Shutdown, and select Yes when the prompt asks you to shutdown
the OM Matrix Gateway application, you must restart the computer
and the Humphrey Matrix equipment so that communication errors
do not occur.
1. Ensure that the Humphrey Matrix equipment is connected to the computer
that ExamWRITER is installed upon with a Humphrey Matrix Connectivity
Cable. For instructions on using the cable, read the Humphrey Matrix
Connectivity Cable Instructions.
2. Ensure that ExamWRITER is open.
ExamWRITER Equipment Integration User’s Guide
51
Chapter 7
Using the Humphrey® Matrix Interface
Using the Humphrey Matrix Interface
3. Create a new blank exam or open a saved exam for an existing patient.
4. Click the Interfaces icon on the ExamWRITER chart window.
5. Select Humphrey:Matrix Capture.
The Humphrey Matrix Integration window opens.
6. On the Humphrey Matrix computer system’s Main Menu window, click
Select Patient.
The View Patients window opens.
7. Select OMLocal from the Source drop-down menu.
8. Click Search.
9. Select the patient whose exam is open in ExamWRITER.
10. Perform the desired test on the Humphrey Matrix.
After the test is completed, the Humphrey Matrix icon will appear on the
ExamWRITER chart window under the Images, Drawings and Graphics
category bar on the Plan - Management tab.
52
ExamWRITER Equipment Integration User’s Guide
Chapter 7
Using the Humphrey® Matrix Interface
Viewing Humphrey Matrix Reports
Viewing
Humphrey
Matrix Reports
1. Open the Humphrey Matrix Reports window using one of the following
methods:

Click the Interfaces icon and select Humphrey:Review Matrix.

Click on the Plan - Management tab and click the Humphrey Matrix
icon.
ExamWRITER Equipment Integration User’s Guide
53
Chapter 7
Using the Humphrey® Matrix Interface
Viewing Humphrey Matrix Reports
2. Double-click on a report to open, view, and print it.
54
ExamWRITER Equipment Integration User’s Guide
Using the Huvitz Interface
8
In this chapter:
Huvitz Interface Overview
•
Huvitz Interface Overview, 55
•
Setting Up the Huvitz Equipment, 56
•
Setting Up the Huvitz Interface, 56
•
Installing CDR-Mate, 58
•
Setting Up CDR-Mate, 60
•
Activating CDR-Mate to Receive Data From Huvitz, 64
•
Using the Huvitz Interface, 64
ExamWRITER receives data from the Huvitz equipment and displays the data on
the ExamWRITER chart window. ExamWRITER does not send data to the Huvitz
equipment.
NOTE
The HRK7000 and HRK8000 Wavefront Autorefractor/Keratometer
only import autorefraction and keratometry data, and not wavefront
data, into ExamWRITER.
The Huvitz equipment interface saves equipment data to the ExamWRITER
database, which then can be accessed by all computers networked to look at that
database. The equipment data can be shared and accessed at multiple
computers, as long as the computers are networked properly and the databases
are properly mapped and shared. The Huvitz interface uses a serial RS-232 cable
®
ExamWRITER Equipment Integration User’s Guide
55
Chapter 8
Using the Huvitz Interface
Setting Up the Huvitz Equipment
or null modem to connect to the computer or Terminal Server. The Huvitz
equipment interface can use multiple serial ports per PC.
NOTES
• If you are using the Huvitz DRS3100 in a wireless environment,
you must hard-wire a computer in the exam room to the Huvitz
equipment by connecting the Huvitz DRS3100 PC translation box
to the serial port on the computer and sharing your export
directory. If you are using more than one Huvitz DRS3100 in a
wireless environment, the export directory on the network must be
accessible from the tablet PC or laptop. For more information on
selecting your export directory, go to “Setting Up CDR-Mate” on
page 60.
• If you are using multiple Huvitz pieces of equipment in multiple
exam rooms, go to
www.officemate.net/omkb/article.aspx?id=10868 for instructions
on how to properly set up your export directories.
Setting Up the
Huvitz Equipment
Set up your Huvitz equipment using the following settings. If you need help setting
up your Huvitz equipment, contact you equipment vendor.
Equipment
Settings
Huvitz MRK3100
Format: OLD
Baud Rate: 9600
Huvitz CLM3100
Format: OLD
Baud Rate: 9600
Huvitz HRK7000
BPS: 9600
RS232: V1
Mode: Topcon and Standard
Huvitz HRK8000
BPS: 9600
RS232: V1
Mode: Topcon and Standard
Setting Up the
Huvitz Interface
NOTE
You must set up the Huvitz interface from the computer connected to
the Huvitz equipment.
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
56
ExamWRITER Equipment Integration User’s Guide
Chapter 8
Using the Huvitz Interface
Setting Up the Huvitz Interface
2. Select the Write captured equipment data to a file check box if you want
ExamWRITER to record data received from Huvitz in a file for debugging
purposes; otherwise, go to step 3.
3. If you are using the Huvitz DRS3100, select the Verify Patient ID before
data is imported check box if you want ExamWRITER to verify the patient ID
for debugging purposes; otherwise, go to step 4.
4. Select the check boxes next to the equipment interfaces that you are setting
up.
NOTE
If you are using the Huvitz HRK7000 or HRK8000, select Huvitz
MRK3100.
5. If you are using the Huvitz CLM3100 or Huvitz MRK3100, select a
communication port number from the Comm Port drop-down menu;
otherwise, go to step 6.
The Comm Port default for the Huvitz CLM3100 is 1. The Comm Port default
for the Huvitz MRK3100 is 2.
ExamWRITER Equipment Integration User’s Guide
57
Chapter 8
Using the Huvitz Interface
Installing CDR-Mate
6. If you are using the Huvitz DRS3100, CDR3100, or HDR7000, complete the
following tasks:
a. Type the data file’s beginning characters in the Data File Start Chars
text box and select the data file’s ending characters (extension) from the
Data File End Char drop-down menu to specify the start and end
characters of data source files and identify specific files from which the
interface is importing data.
b. Select the drive and folder where the equipment data is stored;
otherwise, go to step 7.
7. Click Save.
8. Click Exit.
9. Close and reopen ExamWRITER to activate your changes.
Installing
CDR-Mate
NOTE
If you are using a Huvitz autorefractor or autolensometer, you do not
need to install CDR-Mate; instead, simply connect the autorefractor
or autolensometer to your computer’s serial port and go to “Using the
Huvitz Interface” on page 64.
1. Insert the CDR-Mate for Digital Refraction System CD-ROM.
2. Navigate to the Install folder on the CD-ROM.
3. Double-click setup.exe.
The InstallShield Wizard opens.
58
ExamWRITER Equipment Integration User’s Guide
Chapter 8
Using the Huvitz Interface
Installing CDR-Mate
4. Click Next.
5. Click Install.
The installation begins.
6. Click Finish when the installation completes.
ExamWRITER Equipment Integration User’s Guide
59
Chapter 8
Using the Huvitz Interface
Setting Up CDR-Mate
Setting Up
CDR-Mate
1. Double-click the CDRMate icon on your desktop.
2. Click Setup.
3. Select Options.
The Options window opens.
4. Click the Connection tab.
5. Select the Digital Refraction radio button in the Connection Target box.
6. Select 1.0 from the Protocol Version drop-down menu in the Connection
Target box.
7. Select the appropriate system ID from the System ID drop-down menu in
the Connection Target box. The system ID must be set to the same ID as the
Huvitz control panel system ID connected to the computer.
NOTE
To check the Huvitz control panel system ID, follow the
instructions below:
1. Press and hold down the Alt key on the Huvitz control
panel.
2. Press and release Menu.
3. Press and release Start.
4. Press and release ID.
The lower left corner of the Huvitz control panel screen
displays a box with the CDR System ID number.
5. Ensure that the ID number is the same as the system ID in
the Connection Target box.
6. Select COM1 from the Port Number drop-down menu in the Port
Connection box.
60
ExamWRITER Equipment Integration User’s Guide
Chapter 8
Using the Huvitz Interface
Setting Up CDR-Mate
7. Select 9600 from the Baud Rate drop-down menu in the Port Connection
box.
8. Select 1 from the Computer ID drop-down menu in the Port Connection box.
9. Select the Start CDRMate on Windows startup check box in the Startup &
Connect box.
10. Select the Connect to port when CDRMate starts up check box in the
Startup & Connect box.
11. Click the Data Receiving & Transfer tab.
12. Select the Automatic Export ON radio button.
ExamWRITER Equipment Integration User’s Guide
61
Chapter 8
Using the Huvitz Interface
Setting Up CDR-Mate
13. Click the ... (ellipses) located to the right of the Export Directory text box and
browse to and select OMATE32\DATA\HuvitzData or
OfficeMate\DATA\HuvitzData as the Export Directory.
NOTES
• If you do not have a HuvitzData folder in your
OMATE32\DATA or OfficeMate\DATA\HuvitzData folder,
create one now.
• If you are using the Huvitz DRS3100, the Export Directory
must be the same drive and folder that you selected when
you set up the Huvitz interface. For more information on
setting up the Huvitz interface, go to “Setting Up the Huvitz
Interface” on page 56.
14. Select Manually from the Deleting Record drop-down menu in the Data
Receiving Option box.
15. Do not change any of the options in the Data Transfer Option box.
62
ExamWRITER Equipment Integration User’s Guide
Chapter 8
Using the Huvitz Interface
Setting Up CDR-Mate
16. Click Advanced.
The Advanced Data Receiving Options window opens.
17. Select Text (CSV Format *.csv) from the Export Type drop-down menu.
18. Verify that the Create Sub Folder OFF radio button is selected.
19. Select By patient no. from the File Creation drop-down menu.
20. Do not change any of the options in the Manual Export box.
21. Click OK to close the Advanced Data Receiving Options window.
22. Click OK to close the Options window.
ExamWRITER Equipment Integration User’s Guide
63
Chapter 8
Using the Huvitz Interface
Activating CDR-Mate to Receive Data From Huvitz
Activating
CDR-Mate to
Receive Data
From Huvitz
1. Click File on the Huvitz CDR Mate window.
2. Select Connect if it is enabled.
NOTE
To view the connection status of your system, locate the
“Connected” or “Not Connected” message at the bottom left of
the Huvitz CDR Mate window.
3. Minimize the Huvitz CDR Mate window.
Using the
Huvitz Interface
1. Ensure that ExamWRITER is open.
64
ExamWRITER Equipment Integration User’s Guide
2. Create a new blank exam or open a saved exam for an existing patient.
3. Click the Interfaces icon on the ExamWRITER chart window.
Chapter 8
Using the Huvitz Interface
Using the Huvitz Interface
4. Select Huvitz:DRS3100, Huvitz:MRK3100, or Huvitz:CLM3100.
NOTE
If you are using the Huvitz HRK7000 or HRK8000, select
Huvitz:MRK3100.
The Officemate/Examwriter Equipment Interface window opens and
displays a “Waiting for Huvitz Exam Data…” message.
5. Press Print on the Huvitz CDR-3100 operation panel.
The operation panel displays all of the recorded data.
ExamWRITER Equipment Integration User’s Guide
65
Chapter 8
Using the Huvitz Interface
Using the Huvitz Interface
6. Press Print again on the Huvitz CDR-3100 operation panel.
The operation panel’s built-in printer prints the recorded data and displays a
“Printing result data” message. When the printing is complete, the operation
panel displays a “Transmitting result data” message. After all of the data has
been transmitted from the operation panel to the computer, a progress bar
appears on the Officemate/Examwriter Equipment Interface window.
After all of the received data is processed and saved to the OfficeMate
database, the ExamWRITER exam chart refreshes and all of the received
data from the operation panel is displayed on the Vision/Rx Spectacle and
Binocular tabs.
66
ExamWRITER Equipment Integration User’s Guide
Using the iMACAM Interface
9
In this chapter:
iMACAM
Interface
Overview
•
iMACAM Interface Overview, 67
•
Setting Up the iMACAM Interface, 67
•
Using iMACAM Capture, 68
•
Using iMACAM Review, 70
ExamWRITER interfaces with the iMACAM software before the capture process
when patient information is transferred from ExamWRITER into iMACAM Capture
to create or update retinal images and during the review process when the newly
created or existing iMACAM retinal images are available for on-screen review.
The iMACAM equipment data can be shared and accessed at multiple
computers, as long as the computers are networked properly and the databases
are properly mapped and shared. The iMACAM equipment interface uses an
iMACAM cable to connect to the computer.
NOTES
• iMACAM software must be installed on the same workstation as
ExamWRITER.
• If you are installing iMACAM software on a workstation running
Windows Vista, ensure that the User Account Control (UAC) check
box in the Control Panel is not selected.
• Back up your iMACAM data! Contact the manufacturer for
recommended backup procedures.
• When you upgrade your iMACAM software, you must select the
correct data destination. Contact the manufacturer for upgrade
instructions.
Setting Up the
iMACAM
Interface
NOTE
You must set up the iMACAM interface from the computer connected
to the iMACAM equipment.
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
®
ExamWRITER Equipment Integration User’s Guide
67
Chapter 9
Using the iMACAM Interface
Using iMACAM Capture
2. Select the Write captured equipment data to a file check box if you want
ExamWRITER to record data received from iMACAM in a file for debugging
purposes; otherwise, go to step 3.
3. Select the check boxes next to the equipment interfaces that you are setting
up.
4. Click Save.
5. Click Exit.
6. Close and reopen ExamWRITER to activate your changes.
Using
iMACAM
Capture
1. Ensure that ExamWRITER is open.
2. Create a new blank exam or open a saved exam for an existing patient.
NOTE
Ensure that you have recorded the patient’s date of birth in
ExamWRITER before opening the iMACAM Capture program.
3. Click the Interfaces icon on the ExamWRITER chart window.
68
ExamWRITER Equipment Integration User’s Guide
Chapter 9
Using the iMACAM Interface
Using iMACAM Capture
4. Select Imacam:Capture.
The Match/Confirm EMR Record window opens. If the patient’s information
is not already stored within the iMACAM Capture program, the patient’s
name, gender, and date of birth are loaded into the program from
ExamWRITER. If the patient’s information is already stored within the
iMACAM Capture program, the patient’s information in the program’s
database is displayed.
5. Click OK and follow the iMACAM Capture instructions to capture the
patient’s retinal images.
6. Close iMACAM Capture to return to ExamWRITER.
The iMACAM Advanced icon will be displayed under the Images, Drawings
and Graphics category bar in ExamWRITER after you close and reopen
ExamWRITER.
7. Follow the instructions in “Using iMACAM Review” on page 70 to review the
patient’s captured retinal images.
ExamWRITER Equipment Integration User’s Guide
69
Chapter 9
Using the iMACAM Interface
Using iMACAM Review
Using
iMACAM
Review
NOTES
• The iMACAM Review software can only be used from within
ExamWRITER when a patient’s retinal images have previously
been captured using ExamWRITER and the iMACAM Capture
interface.
• Both the iMACAM Review software and ExamWRITER must be
installed and set up on all workstations on which you want to
review images. 1. Ensure that ExamWRITER is open.
2. Create a new blank exam or open a saved exam for an existing patient for
whom you have previously captured retinal images.
NOTE
Ensure that you have recorded the patient’s date of birth in
ExamWRITER before opening the iMACAM Review program.
3. Click the Interfaces icon on the ExamWRITER chart window and select
Imacam:Review.
OR
Click the iMACAM Advanced icon under the ExamWRITER Images,
Drawings and Graphics category bar.
The iMACAM Review window opens and displays the patient’s images.
4. Follow the iMACAM Review instructions to use the application.
5. Close iMACAM Review to return to ExamWRITER.
70
ExamWRITER Equipment Integration User’s Guide
Using the Konan Interface
10
In this chapter:
Konan Interface Overview
•
Konan Interface Overview, 71
•
Setting Up the Konan Equipment, 71
•
Setting Up the Konan Interface, 75
•
Capturing Images with Konan, 77
•
Reviewing Images with Konan, 79
ExamWRITER interfaces with the Konan Specular Microscope software before the
capture process when patient information is transferred from ExamWRITER to
create or update retinal images and during the review process when the newly
created or existing Konan retinal images are available for on-screen review.
The Konan equipment interface saves equipment data to the Konan database
and ExamWRITER reads that database through a mapped drive. The equipment
data can be shared and accessed at multiple computers, as long as the
computers are networked properly and the databases are properly mapped and
shared. The Konan interface uses a serial RS-232 cable or null modem to connect
to the computer or Terminal Server. The Konan interface can only use one camera
per Terminal Server.
NOTES
• You must install ExamWRITER and the Konan software on the
computer to which the Konan equipment is connected; then you
can install the Konan software on the server on which
ExamWRITER is installed. Contact Konan Medical to determine if
the Konan software will run on an OS server and if any additional
licensing needs to be purchased when using it with Terminal
Server.
• Back up your Konan data! Contact Konan Medical for
recommended backup procedures.
Setting Up the
Konan Equipment
®
Before you begin using the Konan equipment interface, you must set up the
Konan Specular Microscope equipment software to properly interface with
ExamWRITER.
•
Setting Up the Non-CellChek Software, 72
•
Setting Up the CellChek Software, 73
ExamWRITER Equipment Integration User’s Guide
71
Chapter 10
Using the Konan Interface
Setting Up the Konan Equipment
Setting Up the Non-CellChek Software
If you are not using the CellChek software, follow the instructions below:
1. Open the Options window in the Konan software.
2. Ensure that the following options are set:

Replace Analysis Results – ON

Print with Analysis Results – OFF

Auto ID Input – ON

Show Overlay – ON

Search by ID – OFF

Print size – Half size

Printing format – Pattern 1

ID input method – Electronic chart

The data failed to record – Not save
3. Click Setting file for filing system.
The Setting File for Filing System window opens.
4. Ensure that the following options are set:
72

Timing of writing a File for Filing System – After analyzing

Auto delete function – OFF

Types of File for Filing System – Create only a JPEG file

Choice of the function – Create a File for Filing System
ExamWRITER Equipment Integration User’s Guide
Chapter 10
Using the Konan Interface
Setting Up the Konan Equipment
5. Type or select the drive and folder where the equipment data is stored in the
Destination to save the file text box. The equipment data is stored in the
OfficeMate\Konan or OMATE32\Konan folder.
6. Click OK.
Setting Up the CellChek Software
If you are using the CellChek software, follow the instructions below:
1. Double-click the Konan Analysis Software icon.
2. Click CellChek.
The CellChek window opens.
ExamWRITER Equipment Integration User’s Guide
73
Chapter 10
Using the Konan Interface
Setting Up the Konan Equipment
3. Click Set Up and select Electronic Medical Record.
The Electronic Medical Record window opens.
4. Set the Electronic Medical Record window options as shown here.
NOTE
74
The equipment data is stored in the OfficeMate\DATA\Konan or
OMATE32\DATA\Konan folder.
ExamWRITER Equipment Integration User’s Guide
Chapter 10
Using the Konan Interface
Setting Up the Konan Interface
Setting Up the
Konan Interface
NOTES
• You must set up the Konan interface from the computer connected
to the Konan equipment.
• You must be using ExamWRITER version 7.3 or above to interface
with the Konan equipment.
Follow one of these sets of instructions to set up the Konan interface, depending
on whether or not you are using the CellChek software:
•
Setting Up the Konan Interface with the Non-CellChek Software, 75
•
Setting Up the Konan Interface with the CellChek Software, 76
Setting Up the Konan Interface with the Non-CellChek Software
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
2. Select the Write captured equipment data to a file check box if you want
ExamWRITER to record data received from Konan in a file for debugging
purposes; otherwise, go to step 3.
3. Select the check boxes next to the equipment interfaces that you are setting
up.
4. Select the drive and folder where the equipment data is stored in the Path to
Equipment Data section. The capture and review equipment data is stored
in the OfficeMate\Konan or OMATE32\Konan folder. Ensure that the location
that you select is the same location that you selected in the Setting File for
Filing System window in the Konan equipment software. For information on
setting up the Konan equipment, go to “Setting Up the Non-CellChek
Software” on page 72.
ExamWRITER Equipment Integration User’s Guide
75
Chapter 10
Using the Konan Interface
Setting Up the Konan Interface
5. Click Save.
6. Click Exit.
7. Close and reopen ExamWRITER to activate your changes.
Setting Up the Konan Interface with the CellChek Software
1. On the ExamWRITER main window, click Tools, and select Equipment
Integration Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
2. Select the Write captured equipment data to a file check box if you want
ExamWRITER to record data received from Konan in a file for debugging
purposes; otherwise, go to step 3.
3. Select the check boxes next to the equipment interfaces that you are setting
up.
4. Select the drive and folder where the equipment data is stored in the Path to
Equipment Data section. The capture and review equipment data is stored
in the OfficeMate\DATA\Konan or OMATE32\DATA\Konan folder. Ensure that
the location you select is the same location that you selected in the
Electronic Medical Record window in the Konan equipment software. For
information on setting up the Konan equipment, go to “Setting Up the
CellChek Software” on page 73.
5. Click Save.
6. Click Exit.
7. Close and reopen ExamWRITER to activate your changes.
76
ExamWRITER Equipment Integration User’s Guide
Chapter 10
Using the Konan Interface
Capturing Images with Konan
Capturing
Images with
Konan
1. Ensure that ExamWRITER is open.
2. Create a new blank exam or open a saved exam for an existing patient.
3. Click the Interfaces icon on the ExamWRITER chart window.
4. Select Konan:Capture.
The Konan analysis software opens.
5. If you are using the new Konan CellChek software, follow the instructions
below; otherwise go to step 6.
a. Click CellChek on the main window.
The CellChek window opens.
ExamWRITER Equipment Integration User’s Guide
77
Chapter 10
Using the Konan Interface
Capturing Images with Konan
b. Click New Patient.
The Enter Patient Information window opens with the patient
demographic information that was transferred from ExamWRITER.
6. Follow the Konan software capture instructions to capture the patient’s
retinal images on the Specular Microscope.
7. Close the Konan capture software to return to ExamWRITER.
The Konan icon will be displayed under the Images, Drawings and Graphics
category bar in ExamWRITER after you close and reopen ExamWRITER.
8. Follow the instructions in “Reviewing Images with Konan” on page 79 to
review the patient’s captured retinal images.
78
ExamWRITER Equipment Integration User’s Guide
Chapter 10
Using the Konan Interface
Reviewing Images with Konan
Reviewing
Images with
Konan
NOTE
Both the Konan review software and ExamWRITER must be installed
and set up on all workstations on which you want to review images.
Also, a path to the location where the Konan database is located
must be set up. For information on setting up the path to the Konan
database in the Konan equipment software, go to “Setting Up the
Konan Equipment” on page 71. For information on setting up the
path to the Konan database in the ExamWRITER equipment
interface, go to “Setting Up the Konan Interface” on page 75. 1. Ensure that ExamWRITER is open.
2. Create a new blank exam or open a saved exam for an existing patient for
whom you have previously captured retinal images.
3. Click the Interfaces icon on the ExamWRITER chart window and select
Konan:Review.
OR
Click the Konan icon under the ExamWRITER Images, Drawings and
Graphics category bar.
The logo appears if the patient has
images. Click it to view them.
The Konan Specular Microscope Images window opens.
ExamWRITER Equipment Integration User’s Guide
79
Chapter 10
Using the Konan Interface
Reviewing Images with Konan
4. Double-click an image to view it.
5. Follow the Konan software review instructions to use the application.
6. Close the Konan review software to return to ExamWRITER.
80
ExamWRITER Equipment Integration User’s Guide
Using the Kowa Interface
11
In this chapter:
Kowa Interface Overview
•
Kowa Interface Overview, 81
•
Setting Up the Kowa Interface, 81
•
Setting Up the Kowa Camera, 82
•
Capturing Kowa Images, 83
•
Reviewing Kowa Images, 85
ExamWRITER interfaces with the Kowa Digital Fundus Camera before the capture
process when patient information is transferred from ExamWRITER into Kowa
Capture to create or update retinal images and during the review process when
the newly created or existing Kowa retinal images are available for on-screen
review.
The Kowa equipment interface saves equipment data to the Kowa database and
ExamWRITER reads that database through a mapped drive. The equipment data
can be shared and accessed at multiple computers, as long as the computers are
networked properly and the databases are properly mapped and shared. The
Kowa interface uses a serial RS-232 cable or null modem to connect to the
computer or Terminal Server. The Kowa interface can only use one camera per
Terminal Server.
NOTES
• You must install ExamWRITER and the Kowa software on the
computer to which the Kowa equipment is connected; then you
can install the Kowa software on the server on which
ExamWRITER is installed. Contact Kowa to determine if the Kowa
software will run on an OS server and if any additional licensing
needs to be purchased when using it with Terminal Server.
• Back up your Kowa data! Contact Kowa Optimed for
recommended backup procedures.
Setting Up the
Kowa Interface
NOTE
You must set up the Kowa interface from the computer connected to
the Kowa equipment.
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
®
ExamWRITER Equipment Integration User’s Guide
81
Chapter 11
Using the Kowa Interface
Setting Up the Kowa Camera
2. Select the Write captured equipment data to a file check box if you want
ExamWRITER to record data received from Kowa in a file for debugging
purposes; otherwise, go to step 3.
3. Select the check boxes next to the equipment interfaces that you are setting
up.
4. Select the drive and shared folder where the Kowa camera data is written.
5. Click Save.
6. Click Exit.
7. Close and reopen ExamWRITER to activate your changes.
Setting Up the
Kowa
Camera
1. Ensure that Kowa Capture is open.
82
ExamWRITER Equipment Integration User’s Guide
2. Click Setting and select Option Setting.
The SETTING (OPTION) window opens.
Chapter 11
Using the Kowa Interface
Capturing Kowa Images
3. Select the Link to other Dbase System check box.
4. Click OK.
Capturing
Kowa Images
1. Ensure that ExamWRITER is open.
2. Create a new blank exam or open a saved exam for an existing patient.
3. Click the Interfaces icon on the ExamWRITER chart window.
ExamWRITER Equipment Integration User’s Guide
83
Chapter 11
Using the Kowa Interface
Capturing Kowa Images
4. Select Kowa:Capture.
The Kowa Capture window opens with the patient’s last name, first name,
middle initials, gender, date of birth, and patient ID, pre-loaded from
ExamWRITER.
5. Follow the Kowa Capture instructions to capture the patient’s retinal images.
84
ExamWRITER Equipment Integration User’s Guide
Chapter 11
Using the Kowa Interface
Reviewing Kowa Images
6. Close Kowa Capture to return to ExamWRITER.
The Kowa icon will be displayed under the Images, Drawings and Graphics
category bar in ExamWRITER after you close and reopen ExamWRITER.
7. Follow the instructions in “Reviewing Kowa Images” on page 85 to review
the patient’s captured retinal images.
Reviewing
Kowa Images
NOTES
• The Kowa review software can only be used from within
ExamWRITER when a patient’s retinal images have previously
been captured using ExamWRITER and the Kowa Capture
Interface.
• Both the Kowa review software and ExamWRITER must be
installed and set up on all workstations on which you want to
review images. Also, a path to the location where the Kowa
database is located must be set up. For information on setting up
the path to the Kowa database in the ExamWRITER equipment
interface, go to “Setting Up the Kowa Interface” on page 81. 1. Ensure that ExamWRITER is open.
2. Create a new blank exam or open a saved exam for an existing patient.
ExamWRITER Equipment Integration User’s Guide
85
Chapter 11
Using the Kowa Interface
Reviewing Kowa Images
3. Click the Interfaces icon on the ExamWRITER chart window and select
Kowa:Review.
OR
Click the Kowa icon under the ExamWRITER Images, Drawings and
Graphics category bar.
The Kowa Capture window opens and displays the patient’s history.
4. Follow the Kowa review instructions to use the application.
5. Close Kowa to return to ExamWRITER.
86
ExamWRITER Equipment Integration User’s Guide
Using the Marco Interface
12
In this chapter:
Marco Interface Overview
•
Marco Interface Overview, 87
•
Setting Up Marco RT & TRS Series Equipment, 88
•
Setting Up Marco ARK, LM, RKT, & VL Series Equipment, 88
•
Setting Up the Marco Interface, 89
•
Using the Marco Interface, 91
ExamWRITER receives data from the Marco equipment and displays the data on
the ExamWRITER chart window. ExamWRITER does not send data to the Marco
equipment.
The Marco equipment interface saves equipment data to the ExamWRITER
database, which then can be accessed by all computers networked to look at that
database. The equipment data can be shared and accessed at multiple
computers, as long as the computers are networked properly and the databases
are properly mapped and shared. The Marco interface uses a serial RS-232 cable
or null modem to connect to the computer or Terminal Server. The Marco
equipment interface can use multiple serial ports per PC.
®
ExamWRITER Equipment Integration User’s Guide
87
Chapter 12
Using the Marco Interface
Setting Up Marco RT & TRS Series Equipment
Setting Up
Marco RT &
TRS Series
Equipment
The following digital refractor models interface with ExamWRITER and should be
configured according to Marco’s specifications:
•
OPD-Scan III
•
RT-900
•
RT-2100
•
RT-5100
•
RTS-1200
•
TRS-3100
•
TRS-5100
NOTES
• The Marco RT-900 (Evolution) requires that the control panel
interface be correctly set up so that k-readings from the
autorefractor are accepted by the Marco RT-900. If the control
panel interface is not set up correctly, the Marco RT-900 will not be
able to send the k-readings to ExamWRITER and they will not
appear on the chart window. Contact your Marco representative
for information on properly setting up the Marco RT-900.
• If you are using the Marco RT-5100 (EPIC 5100, TRS-5100), you
must upgrade your firmware with fix #RB-1.04.
• If you have questions about setting up Marco equipment, contact
your Marco representative.
Setting Up
Marco ARK,
LM, RKT, & VL
Series Equipment
The following models interface with ExamWRITER without a digital refraction
system and are covered in this section:
•
ARK-530A (referred to as ARK-A-Devices)
•
ARK-700A (referred to as ARK-A-Devices)
•
ARK-710A (referred to as ARK-A-Devices)
•
ARK-730A (referred to as ARK-A-Devices)
•
ARK-760A (referred to as ARK-A-Devices)
•
LM-600 (referred to as LM-Devices)
•
LM-1000 (referred to as LM-Devices)
•
LM-1200 (referred to as LM-Devices)
•
LM-1800 (referred to as LM-Devices)
•
RKT-7700
•
M3 TonoRef II
•
VL-3000
NOTE
88
Daisy-chaining, or connecting lensometers to autorefractors, is not
supported.
ExamWRITER Equipment Integration User’s Guide
Chapter 12
Using the Marco Interface
Setting Up the Marco Interface
You must set several parameters on your Marco equipment in order for the device
to transfer data to the ExamWRITER Equipment Interface.
1. Follow the instructions in your device manual to access the Parameter Set
screen and set each of the parameters.
2. When you are finished setting your parameters, exit Parameter Set mode,
turn off the device, and turn it back on.
The following table shows the parameter number and setting for each of the six
parameters on each device.
Device
KM Unit
ARK-530A
I/F Mode
I/F Format
#71: NCP10
#72: All
Baud Rate
Bit Length
CR Code
ARK-700A
#3: D
#31: NCP10
#32: All
#33: 9600
#34: 8
35: No
ARK-710A
#11: D
#51: NCP10
#52: All
#53: 9600
#54: 8
55: No
ARK-730A
#11: D
#51: NCP10
#52: All
#53: 9600
#54: 8
55: No
ARK-760A
#3: D
#51: NCP10
#52: All
#53: 9600
#54: 8
55: No
OFF
LM-600
Menu Page
5: #32
NCP10
#33: 9600
8
LM-1000
Menu Page
5: NCP10
Menu Page
5: 9600
Menu Page 5: Menu Page
8
5: No
LM-1200
Menu Page
4: NCP10
Menu Page
4: 9600
Menu Page 4: Menu Page
8
4: No
LM-1800
Menu Page
5: #32
NCP10
#33: 9600
8
OFF
RKT-7700
#49: NCP10
#50: All
M3 TonoRef II
#81: NCP10
#82: All
9600
8
ODD
1
VL-3000
VL-3000
PC
Setting Up the
Marco Interface
1. Follow the instructions provided by Marco to install the Marco Connect
software.
NOTE
To obtain the Marco Connect software, contact Marco at
800.874.5274 or go to http://www.marco.com.
2. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
3. Select the Write captured equipment data to a file check box if you want
ExamWRITER to record data received from Marco in a file for debugging
purposes; otherwise, go to step 3.
ExamWRITER Equipment Integration User’s Guide
89
Chapter 12
Using the Marco Interface
Setting Up the Marco Interface
4. Select the check box next to the equipment interface that you are using.
NOTES
• If you are using the Marco M3 TonoRef II, select the Marco
RKT-7700 equipment interface.
• If you are using the Marco TRS-3100, select the Marco
RT-5100 equipment interface.
5. Select a source from the Data Source drop-down menu.

Select Port if the equipment is connected directly to the workstation on
which ExamWRITER is running.

Select File if you are using Marco Connect to interface with your
equipment
6. If you selected Port in step 5, select a communication port number from the
Comm Port drop-down menu; otherwise, go to step 7.
The Comm Port default is 1.
7. If you selected File in step 5, type the first two characters of the device name
in the Data File Start Chars field; otherwise, go to step 10.
8. Select the file extension from the Data File End Chars drop-down menu. If
you are using Marco Connect, select .xml.
9. Select the location on the network where Marco Connect saves the data file:
a. Select a drive from the drop-down menu.
b. Select the folder in which Marco Connect saves the data file.
NOTE
The Marco Connect application allows you to specify where
you want to save the data files on the network.
10. Repeat steps 4–9 for each piece of Marco equipment you are using.
11. Click Save.
90
ExamWRITER Equipment Integration User’s Guide
Chapter 12
Using the Marco Interface
Using the Marco Interface
12. Click Exit.
13. Close ExamWRITER to activate your changes.
Using the
Marco Interface
1. Ensure that the Marco equipment is connected to the computer that
ExamWRITER is installed upon with an EP21-10-0809 serial cable or null
modem serial cable.
NOTE
Call Marco at 800.874.5274 to order an EP21-10-0809 serial
cable or null modem serial cable.
2. Ensure that ExamWRITER is open.
3. Create a new blank exam or open a saved exam for an existing patient.
4. Click the Interfaces icon on the ExamWRITER chart window.
5. Select the Marco device that corresponds to your Marco equipment.
The Officemate/Examwriter Equipment Interface window opens and
displays a “Waiting for Marco Exam Data…” message.
6. Press Print on the Marco equipment operation panel.
The operation panel displays all of the recorded data.
7. Press Print on the Marco equipment operation panel.
The operation panel’s built-in thermal printer prints the recorded data and
displays a “Printing result data” message. After all of the data has been
ExamWRITER Equipment Integration User’s Guide
91
Chapter 12
Using the Marco Interface
Using the Marco Interface
transmitted from the operation panel to the computer, a progress bar
appears on the Officemate/Examwriter Equipment Interface window.
After all of the received data is processed and saved to the OfficeMate
database, the ExamWRITER exam chart refreshes, and all of the received
92
ExamWRITER Equipment Integration User’s Guide
Chapter 12
Using the Marco Interface
Using the Marco Interface
data from the operation panel is displayed on the Vision/Rx Spectacle and
Binocular tabs.
NOTES
• The manifest prism data is displayed as binocular phoria
data.
• If the Phoria - Distance and Phoria - Near Test text boxes
display “Von Graefe” and you want to change the name of
the test, click in the Test text box and select a new test from
the Test drop-down menu.
• NT (tonometry) data is not saved on the IC card when you
use the Marco RT-5100 (EPIC 5100); therefore, if you are
using the IC card to transfer data from the
autorefractor/keratometer (ARK) equipment to the Marco
RT-5100 (EPIC 5100), the data will not appear in the
ExamWRITER IOP table.
• In order for the K radius values to appear in the K-Readings
boxes in ExamWRITER, both the Marco RT-2100 (TRS-2100
or EPIC 2100), RT-5100 (EPIC 5100), TRS-3100, or RT-900
(Evolution) and the autorefractor/keratometer (ARK)
equipment must be set for the K radius units to be either
diopters or diopters and millimeters. If the Marco equipment
is outputting the K radius values in millimeters only, then they
will not appear in ExamWRITER.
• If there is subjective data, the standard Marco VD of 13.75
will be displayed in the ExamWRITER Vertex Dist box.
• The aniseikonia measurement is for distance only. The
output of “00” is translated to “Absent”; otherwise, it is
displayed as “Present” in the Aniseikonia boxes.
• The pupillary distances are extracted from subjective (far
and near) measurements in Marco and recorded in the PD
box in ExamWRITER.
ExamWRITER Equipment Integration User’s Guide
93
Chapter 12
Using the Marco Interface
Using the Marco Interface
94
ExamWRITER Equipment Integration User’s Guide
Using the NIDEK NAVIS-EX
Interface
13
In this chapter:
NIDEK
NAVIS-EX
Interface
Overview
•
NIDEK NAVIS-EX Interface Overview, 95
•
Setting Up NIDEK NAVIS-EX, 96
•
Setting Up the NIDEK NAVIS-EX Interface, 100
•
Using NIDEK NAVIS-EX Capture, 101
•
Using NIDEK NAVIS-EX Review, 102
The NIDEK NAVIS-EX software can acquire images from these fundus
photography devices and display them in ExamWRITER:
•
AFC-230/AFC-210 Non-Mydriatic Autofocus Fundus Camera
•
AFC-330 Non-Mydriatic Auto Fundus Camera
•
F-10 Opthalmoscope
Contact your Marco representative to set up the camera equipment and contact
your NIDEK representative to set up the NAVIS-EX software.
ExamWRITER interfaces with the NIDEK NAVIS-EX software before the capture
process when patient information is transferred from ExamWRITER into NIDEK
NAVIS-EX Capture to create or update patient information and retinal images and
during the review process when the newly created or existing NIDEK NAVIS-EX
patient information and retinal images are available for review.
ExamWRITER and NAVIS-EX communicate with each other via a command line
interface located at C:\NIDEK\EMR\Client\MiddlewareCLI.exe.
The NIDEK NAVIS-EX equipment interface saves equipment data to the NIDEK
NAVIS-EX database and ExamWRITER reads that database through a mapped
drive. The equipment data can be shared and accessed at multiple computers,
as long as the computers are networked properly and the databases are properly
mapped and shared. The NIDEK NAVIS-EX interface uses a serial RS-232 cable
or null modem to connect to the computer. The NIDEK NAVIS-EX interface cannot
connect to a Terminal Server.
®
ExamWRITER Equipment Integration User’s Guide
95
Chapter 13
Using the NIDEK NAVIS-EX Interface
Setting Up NIDEK NAVIS-EX
Setting Up
NIDEK
NAVIS-EX
Before you begin using the NIDEK NAVIS-EX equipment interface, you must set
up the NAVIS-EX software to properly interface with ExamWRITER.
In order to use the NIDEK NAVIS-EX, you must:
•
Configure the NIDEK AFC-230, AFC-330, AFC-210, or F-10 camera. Refer to
the equipment manual for more information on configuring your camera.
•
Install ExamWRITER v9.5 or above.
•
Install NIDEK NAVIS-EX v1.2.1 or above. Refer to the NIDEK NAVIS-EX
Software Installation Manual for more information on installing NAVIS-EX.
•
Configure NAVIS-EX. For more information see “Configuring NAVIS-EX” on
page 97.
NOTES
• NIDEK NAVIS-EX software must be installed on the same
workstation as ExamWRITER.
• If you are installing NIDEK NAVIS-EX software on a workstation
running Windows Vista, ensure that the User Account Control
(UAC) check box in the Control Panel is not selected.
• Back up your NIDEK NAVIS-EX data! Contact the manufacturer for
recommended backup procedures.
• When you upgrade your NIDEK NAVIS-EX software, you must
select the correct data destination. Contact NIDEK Client Services
for upgrade instructions.
• Contact NIDEK or your certified networking technician if you have
questions about networking the NAVIS-EX software.
96
ExamWRITER Equipment Integration User’s Guide
Chapter 13
Using the NIDEK NAVIS-EX Interface
Setting Up NIDEK NAVIS-EX
Configuring NAVIS-EX
In order for NAVIS-EX and ExamWRITER to exchange data, NAVIS-EX must be
configured to accept data from ExamWRITER in an expected format and order.
NOTE
Refer to the NAVIS-EX Operator’s Manual for detailed information
about configuring the NAVIS-EX software.
1. Configure the NAVIS-EX Configuration window as follows (if you have not
already done so):
a. On the Patient List tab, set the sorting preference for the Patient List to
(from top to bottom): Patient ID, Name, Age, Sex, Comment, History.
ExamWRITER Equipment Integration User’s Guide
97
Chapter 13
Using the NIDEK NAVIS-EX Interface
Setting Up NIDEK NAVIS-EX
b. On the Patient Change tab, select the Use Patient Change check box.
2. Click Setup.
The Setting window opens.
98
ExamWRITER Equipment Integration User’s Guide
Chapter 13
Using the NIDEK NAVIS-EX Interface
Setting Up NIDEK NAVIS-EX
3. Configure the NAVIS-EX Setting window as follows (if you have not already
done so):
a. On the General tab, set the Sex Code for Male to 1 and Female to 2.
b. On the Data tab, set the Field List order (from top to bottom) to:
PatientID 1, Last Name 2, First Name 3, Middle Name 4, Sex 5,
BirthDay 6.
ExamWRITER Equipment Integration User’s Guide
99
Chapter 13
Using the NIDEK NAVIS-EX Interface
Setting Up the NIDEK NAVIS-EX Interface
Setting Up the
NIDEK
NAVIS-EX
Interface
NOTE
You must set up the NIDEK NAVIS-EX interface from the computer
connected to the camera equipment.
1. Read and follow the instructions in the knowledge base article at
www.officemate.net/omkb/article.aspx?id=31496.
2. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
3. Select the check boxes next to the equipment interfaces that you are setting
up.
4. Select the drive and folder where the NIDEK NAVIS-EX data files are stored:
C:\NIDEK\EMR\Client\.
5. Click Save.
6. Click Exit.
7. Close and reopen ExamWRITER to activate your changes.
100
ExamWRITER Equipment Integration User’s Guide
Chapter 13
Using the NIDEK NAVIS-EX Interface
Using NIDEK NAVIS-EX Capture
Using NIDEK
NAVIS-EX
Capture
NOTE
If you are using the AFC-330 Non-Mydriatic Auto Fundus Camera,
you not need to complete the instructions in this section. Instead,
simply open an exam in ExamWRITER for the patient whose image
you captured and then, from the camera, find the patient and
capture the image. If you have questions about capturing images
using the camera, contact your Marco representative.
1. Ensure that ExamWRITER is open.
2. Create a new blank exam or open a saved exam for an existing patient.
NOTE
Ensure that you have recorded the patient’s date of birth and
gender in ExamWRITER before opening the NIDEK NAVIS-EX
Capture program.
3. Click the Interfaces icon on the ExamWRITER chart window.
4. Select NIDEK NAVIS-EX:Capture.
NIDEK NAVIS-EX launches, captures the NIDEK camera image, and stores
it in NAVIS-EX. The NAVIS-EX Image Capture window opens with information
about the selected patient.
If the patient’s information is not already stored within the NIDEK NAVIS-EX
Capture program, the patient’s name, gender, and date of birth are loaded
into the program from ExamWRITER. If the patient’s information is already
stored within the NIDEK NAVIS-EX Capture program, the patient’s
information in the program’s database is displayed.
Data for current patient open in ExamWRITER is displayed.
If the patient’s demographic information is already stored in NAVIS-EX but
has since changed in ExamWRITER, the information will also be changed in
NAVIS-EX database.
5. If information changed, click OK in the confirmation box.
ExamWRITER Equipment Integration User’s Guide
101
Chapter 13
Using the NIDEK NAVIS-EX Interface
Using NIDEK NAVIS-EX Review
6. Click the Capture button on the toolbar to capture the patient’s retinal
images.
7. Close NIDEK NAVIS-EX Capture to return to ExamWRITER.
The NIDEK NAVIS-EX icon is displayed under the Images, Drawings and
Graphics category bar in ExamWRITER after you close and reopen
ExamWRITER.
Click the icon to open the
Review software for the
patient.
NOTE
If you click the NAVIS-EX icon, the NAVIS-EX Review software
opens with the current patient displayed.
8. Follow the instructions in “Using NIDEK NAVIS-EX Review” on page 102 to
review the patient’s captured retinal images.
Using NIDEK
NAVIS-EX
Review
NOTES
• The NIDEK NAVIS-EX Review software can only be used from
within ExamWRITER when a patient’s retinal images have
previously been captured using ExamWRITER and the NIDEK
NAVIS-EX Capture interface.
• Both the NIDEK NAVIS-EX Review software and ExamWRITER
must be installed and set up on all workstations (up to three) on
which you want to review images. 1. Ensure that ExamWRITER is open.
2. Create a new blank exam or open a saved exam for an existing patient for
whom you have previously captured retinal images.
NOTE
102
Ensure that you have recorded the patient’s date of birth and
sex in ExamWRITER before opening the NIDEK NAVIS-EX
Review program.
ExamWRITER Equipment Integration User’s Guide
Chapter 13
Using the NIDEK NAVIS-EX Interface
Using NIDEK NAVIS-EX Review
3. Click the Interfaces icon on the ExamWRITER chart window and select
NIDEK NAVIS-EX:Review.
OR
Click the NAVIS-EX icon under the ExamWRITER Images, Drawings and
Graphics category bar.
NOTES
• The NAVIS-EX icon is only displayed when images exist for
the selected patient.
• If you hover the mouse over the NAVIS-EX icon, the date in
the format YYYY/MM/DD and a description of the latest
image file is displayed. (The image description is not
entered into the ExamWRITER record.)
The NIDEK NAVIS-EX Review window opens and displays the selected
patient’s detail information and images.
If the patient’s information is not already stored within the NIDEK NAVIS-EX
Review program, the patient’s name, gender, and date of birth are loaded
into the program from ExamWRITER.
4. Follow the NIDEK NAVIS-EX Review instructions in the NAVIS-EX Operator’s
Manual to use the application.
5. Close NIDEK NAVIS-EX Review to return to ExamWRITER.
ExamWRITER Equipment Integration User’s Guide
103
Chapter 13
Using the NIDEK NAVIS-EX Interface
Using NIDEK NAVIS-EX Review
104
ExamWRITER Equipment Integration User’s Guide
Using the Optos Interface
14
In this chapter:
Optos Interface Overview
•
Optos Interface Overview, 105
•
Setting Up the Optos Interface, 106
•
Using Optos Capture, 107
•
Using Optos Review, 109
ExamWRITER interfaces with the Optos software before the capture process
when patient information is transferred from ExamWRITER into Optos Capture to
create or update retinal images and during the review process when the newly
created or existing Optos retinal images are available for on-screen review.
You must install ExamWRITER and the Optos Capture and Optos Review
applications on the computer to which the Optos equipment is connected; then
you can install the Optos Capture and Optos Review applications on the server on
which ExamWRITER is installed. Contact Optos at 800.854.3039 to determine if
the Optos Capture and Optos Review applications will run on an OS server and if
any additional licensing needs to be purchased when using it with Terminal
Server. Both applications can be used on the same workstation or on two
separate workstations.
The Optos equipment interface saves equipment data to the Optos database and
ExamWRITER reads that database through a mapped drive. The equipment data
can be shared and accessed at multiple computers, as long as the computers are
networked properly and the databases are properly mapped and shared. The
Optos equipment interface uses a serial RS-232 cable or null modem to connect
to the computer or Terminal Server.
®
ExamWRITER Equipment Integration User’s Guide
105
Chapter 14
Using the Optos Interface
Setting Up the Optos Interface
NOTES
• Back up your Optos data! Contact Optos at 800.854.3039 for
recommended backup procedures.
• If you began using Optos before ExamWRITER, your Optos
Patient IDs will be different than your ExamWRITER patient
numbers; therefore, your Optos and ExamWRITER patients will not
be linked to each other. To automatically link your Optos and
ExamWRITER patients, run the ExamWRITER/Optos Patient
Synchronization Utility. To open this utility and view instructions on
how to use the utility, go to
www.officemate.net/omkb/Article.aspx?id=10027.
• Ensure that the Optional Automatic ID (user can enter or
change ID) radio button is selected in the Admin - optos v2
application’s Application Usage Policy window. To open this
window, click Policy on the Admin - optos v2 main window toolbar,
select Manage Policies, and click the Patient ID tab.
• When you upgrade your Optos software, you must select the
correct data destination. Contact Optos at 800.854.3039 for
upgrade instructions.
• If you are using Optos Daytona, you must rename the
scheduler.exe file on your Optos server to capture.exe. Contact
Optos at 800.854.3039 if you have questions about renaming this
file.
Setting Up the
Optos Interface
NOTE
You must set up the Optos interface from the computer connected to
the Optos equipment.
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
2. Select the Write captured equipment data to a file check box if you want
ExamWRITER to record data received from Optos in a file for debugging
purposes; otherwise, go to step 3.
106
ExamWRITER Equipment Integration User’s Guide
Chapter 14
Using the Optos Interface
Using Optos Capture
3. Select the check boxes next to the equipment interfaces that you are setting
up.
4. Click Save.
5. Click Exit.
6. Close and reopen ExamWRITER to activate your changes.
Using Optos
Capture
1. Ensure that ExamWRITER is open.
2. Create a new blank exam or open a saved exam for an existing patient.
3. Click the Interfaces icon on the ExamWRITER chart window.
ExamWRITER Equipment Integration User’s Guide
107
Chapter 14
Using the Optos Interface
Using Optos Capture
4. Select Optos:Capture.
The optos v2 Capture window opens with the patient’s last name, first name,
middle initials, gender, date of birth, patient ID, iris category, practitioner,
notes, and exam status pre-loaded from ExamWRITER.
5. Follow the Optos Capture instructions to capture the patient’s retinal images.
108
ExamWRITER Equipment Integration User’s Guide
Chapter 14
Using the Optos Interface
Using Optos Review
6. Close Optos Capture to return to ExamWRITER.
The optomap Retinal Exam icon will be displayed under the Images,
Drawings and Graphics category bar in ExamWRITER after you close and
reopen ExamWRITER.
7. Follow the instructions in “Using Optos Review” on page 109 to review the
patient’s captured retinal images.
Using Optos
Review
NOTES
• The Optos Review software can only be used from within
ExamWRITER when a patient’s retinal images have previously
been captured using ExamWRITER and the Optos Capture
interface.
• Both the Optos Review software and ExamWRITER must be
installed and set up on all workstations on which you want to
review images. 1. Ensure that ExamWRITER is open.
2. Create a new blank exam or open a saved exam for an existing patient for
whom you have previously captured retinal images.
ExamWRITER Equipment Integration User’s Guide
109
Chapter 14
Using the Optos Interface
Using Optos Review
3. Click the Interfaces icon on the ExamWRITER chart window and select
Optos:Review.
OR
Click the optomap Retinal Exam icon under the ExamWRITER Images,
Drawings and Graphics category bar.
The Review – optos v2 window opens and displays the patient’s history.
4. Follow the Optos Review instructions to use the application.
5. Close Optos Review to return to ExamWRITER.
110
ExamWRITER Equipment Integration User’s Guide
Using the Reichert Equipment
Interface
15
In this chapter:
Reichert Interface Overview
•
Reichert Interface Overview, 111
•
Setting Up the ReichertCapture Software, 112
•
Setting Up the Reichert Interface, 112
•
Using the AL-200, AL-500, AL-700, AT-555, PT-100, Reichert 7, & Reichert
7CR Devices, 116
•
Using the RK-600 & RK-700 Devices, 118
•
Using the iPac Pachymeter, 119
•
Using the Auto Phoropter RS & Auto Phoropter Visutron900, 121
•
Reviewing Reflex UBM & Foresee PHP Images, 124
ExamWRITER interfaces with all Reichert equipment by reading data or images
obtained from the Reichert equipment and populating the patient’s exam with that
data or making the images available for on-screen review.
The Reichert Auto Phoropter RS and Auto Phoropter Visutron900 are the only
Reichert devices that send data to and receive data from ExamWRITER.
ExamWRITER first exports auto refraction or presenting spectacle data to the
Auto Phoropter RS or Auto Phoropter Visutron900. The Auto Phoropter RS or Auto
Phoropter Visutron900 then exports manifest Rx data to ExamWRITER.
The Reichert equipment interface saves equipment data to the ExamWRITER
database, which then can be accessed by all computers networked to look at that
database. The equipment data can be shared and accessed at multiple
computers, as long as the computers are networked properly and the databases
are properly mapped and shared. The Reichert interface uses a serial RS-232
cable or null modem to connect to the computer or Terminal Server. The Reichert
equipment interface can use multiple serial ports per PC.
®
ExamWRITER Equipment Integration User’s Guide
111
Chapter 15
Using the Reichert Equipment Interface
Setting Up the ReichertCapture Software
Setting Up the
ReichertCapture Software
Setting Up the
Reichert Interface
If you are setting up a Reichert AL-500, AL-700, PT-100, Reichert 7, Reichert 7CR,
RK-700, iPac, Auto Phoropter RS, or Auto Phoropter Visutron900, you need to
install the ReichertCapture software. For help on installing the ReichertCapture
software, see the instructions from Reichert that came with the software, or
contact a Reichert representative.
NOTE
If you are using a Reichert AL-200, AT-555, or RK-600 device with
ExamWRITER v10.0 or below, then you can only receive data from
these devices using a serial cable. If you are using these devices
with ExamWRITER v10.5 or above, you can choose to receive data
using a serial cable or the ReichertCapture software.
NOTES
• You must set up the Reichert interface from the computer
connected to the Reichert equipment.
• You must use ExamWRITER v7.3 or above to interface with the
Al-200 Lensometer, AT-555 Tonometer, or RK-600 Autorefractor.
You must use ExamWRITER v10.7 or above to interface with the
Reflex UBM and Foresee PHP. For all other Reichert equipment,
you must use ExamWRITER v10.5 or above.
• If you are using the AT-555 Tonometer, ensure that the device’s
firmware is version 2.4 or later. Refer to the documentation that
came with the device for information on checking the firmware
version.
• Ensure that you have the proper serial cable for your equipment.
112
For this device
Use this cable
AL-200 Lensometer
NULL Modem
RK-600 Autorefractor
Reichert custom cable
iPac Pachymeter
Wireless
All other Reichert devices
Standard serial cable
ExamWRITER Equipment Integration User’s Guide
Chapter 15
Using the Reichert Equipment Interface
Setting Up the Reichert Interface
1. Unless you are setting up the iPac, refer to the documentation that came
with your device to set the RS232 data bus parameters listed in the table
below. If you are setting up the iPac interface, go to step 3.
Model
Parameter
Setting
Notes
All
Character
8
Required
All
Parity
N
Required
All
Stop bits
1
Required
AL-200
PRSM
Base
Required to record the
presenting horizontal and
vertical prism data
RK-600
KERATO
Diopt
Required to record the
kerotometry readings in
ExamWRITER
RK-600
PRINT
Manual
Required
RK-600
REVIEW DATA
On
Required
2. Refer to the documentation that came with your device to set the baud rate:
Model
Baud Rate
AL-200, PT-100, RK-600,
Auto Phoropter RS, Auto
Phoropter Visutron900
9600
AL-500
38400
AL-700
38400 (fixed at 921600
when USB is selected)
AT-555, Reichert 7, Reichert
7CR
19200
RK-700
115200 (fixed at 921600
when USB is selected)
3. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
4. Select the Write captured equipment data to a file check box if you want
ExamWRITER to record data received from the Reichert equipment in a file
for debugging purposes; otherwise, go to step 5.
ExamWRITER Equipment Integration User’s Guide
113
Chapter 15
Using the Reichert Equipment Interface
Setting Up the Reichert Interface
5. Select the check boxes next to the equipment interfaces that you are setting
up.

If you are setting up an AL-200, AL-500, or AL-700 device, select the
Reichert:Lensometer check box.

If you are setting up an AT-555, PT-100, Reichert 7, or Reichert 7CR
device, select the Reichert:Tonometer check box.

If you are setting up an RK-600 or RK-700 device, select the
Reichert:Auto-Refractor check box.

If you are setting up the iPac Pachymeter, select the Reichert:iPac
check box.

If you are setting up an Auto Phoropter RS or Auto Phoropter
Visutron900, select the Reichert:Auto Phoropter Output and the
Reichert:Auto Phoropter Input check boxes.

If you are setting up the Reflex UBM, select the UBM Image Review
check box.

If you are setting up the Foresee PHP, select the PHP Image Review
check box.
6. If you are setting up a Reichert AL-200, AT-555, or RK-600 device, and are
connecting to the computer using a serial cable, follow the instructions
below; otherwise go to step 7.
a. Select Port from the Data Source drop-down menu.
b. Select a port number from the Comm Port drop-down menu. The default
port is 1.
c. Go to step 11.
7. If you are setting up a Reichert device that does not require a serial
connection, select File from the Data Source drop-down menu. If you
selected Auto Phoropter Input in step 5, you do not enter this information.
114
ExamWRITER Equipment Integration User’s Guide
Chapter 15
Using the Reichert Equipment Interface
Setting Up the Reichert Interface
8. Type the data file’s beginning characters in the Data File Start Chars text
box. If you selected Auto Phoropter Input, UBM Image Review, or PHP
Image Review in step 5, you do not enter this information.
Reichert Model
Data File Start Chars
AL-200
AL-200
AL-500
AL-500
AL-700
AL-700
AT-555
AT-555
iPac Pachymeter
iPac
Reichert7
Reichert7
Reichert7CR
Reichert7CR
RK-600
RK-600
RK-700
RK-700
Auto Phoropter RS
Auto
Auto Phoropter Visutron 900 Auto
9. If you selected UBM Image Review or PHP Image Review Select in step 5,
select .jpg from the Data File End Char drop-down menu. If you selected
any other equipment interfaces, except the Auto Phoropter Input, in step 5,
select .xml from the Data File End Char drop-down menu. If you selected
Auto Phoropter Input in step 5, you do not enter this information.
10. Select the drive and folder where the equipment data is stored in the Path to
Equipment Data section.
11. Click Save.
12. Click Exit.
ExamWRITER Equipment Integration User’s Guide
115
Chapter 15
Using the Reichert Equipment Interface
Using the AL-200, AL-500, AL-700, AT-555, PT-100, Reichert 7, & Reichert 7CR Devices
13. Close and reopen ExamWRITER to activate your changes.
Using the
AL-200,
AL-500,
AL-700,
AT-555,
PT-100,
Reichert 7, &
Reichert 7CR
Devices
1. Open ExamWRITER.
2. Create a new blank exam or open a saved exam for an existing patient.
3. Click the Interfaces icon on the ExamWRITER chart window.
4. Select the device that corresponds with your Reichert equipment.
The Officemate and Examwriter Equipment Interface window opens and
displays the following message: “Waiting for Reichert Data…”
5. Use the Reichert equipment to perform the examination or tests.
The data captured by the Reichert equipment is transmitted automatically at
the end of the test. A progress bar appears across the Officemate and
Examwriter Equipment Interface window indicating that the information is
being received.
NOTE
If the information does not transmit automatically at the end of
the examination or tests, press the Print button on the Reichert
equipment.
After all of the received data from the AL-200, AL-500, or AL-700 is
processed and saved to the ExamWRITER database, the patient’s
ExamWRITER record refreshes, and the data received from the Reichert
device is displayed on the Vision/Rx Spectacle tab.
After all of the received data from the AT-555, PT-100, Reichert 7, or Reichert
7CR is processed and saved to the ExamWRITER database, the patient’s
ExamWRITER record refreshes, and the data received from the Reichert
device is displayed in the IOP table on the Examination tab.
116
ExamWRITER Equipment Integration User’s Guide
Chapter 15
Using the Reichert Equipment Interface
Using the AL-200, AL-500, AL-700, AT-555, PT-100, Reichert 7, & Reichert 7CR Devices
6. Click the Vision/Rx Spectacle tab within the patient’s exam record to view
the Reichert AL-200, AL-500, or AL-700 data.
OR
Click the Exam - Special Tests tab within the patient’s exam record to view
the AT-555, PT-100, Reichert 7, or Reichert 7CR data.
ExamWRITER Equipment Integration User’s Guide
117
Chapter 15
Using the Reichert Equipment Interface
Using the RK-600 & RK-700 Devices
Using the
RK-600 &
RK-700
Devices
1. Open ExamWRITER.
2. Create a new blank exam or open a saved exam for an existing patient.
3. Use the Reichert equipment to perform the examination or tests.
NOTE
After three readings for both refraction and kerotometry, the
Reichert equipment displays the Review Data screen. Do not
press the Print icon before completing steps steps 4–5 below.
4. After the tests using the Reichert equipment are completed, click the
Interfaces icon on the ExamWRITER chart window.
5. Select Reichert:Auto-Refractor.
The Officemate and Examwriter Equipment Interface window opens and
displays the following message: “Waiting for Reichert Data…”
6. Press the Print button on the Review Data screen on the Reichert
equipment.
The Reichert device’s internal thermal printer prints the captured data, and
the Reichert equipment transmits the data to ExamWRITER. A progress bar
118
ExamWRITER Equipment Integration User’s Guide
Chapter 15
Using the Reichert Equipment Interface
Using the iPac Pachymeter
appears across the Officemate and Examwriter Equipment Interface window
indicating that the information is being received.
After all of the data is received and saved to the OfficeMate database, the
patient’s ExamWRITER record refreshes, and the data received from the
Reichert equipment is displayed on the Vision/Rx Spectacle tab.
7. Click the Vision/Rx Spectacle tab within the patient’s exam record to view
the Reichert data.
Using the iPac
Pachymeter
1. Open ExamWRITER.
2. Create a new blank exam or open a saved exam for an existing patient.
3. Click the Interfaces icon on the ExamWRITER chart window.
ExamWRITER Equipment Integration User’s Guide
119
Chapter 15
Using the Reichert Equipment Interface
Using the iPac Pachymeter
4. Select Reichert:iPac.
The Officemate and Examwriter Equipment Interface window opens and
displays the following message: “Waiting for Reichert:iPac Exam Data…”
5. Use the Reichert iPac to perform the examination or tests.
6. To send the data from the iPac to ExamWRITER, tilt the multi-function switch
on the iPac in the downward direction.
The iPac transmits the data to ExamWRITER. A progress bar appears
across the Officemate and Examwriter Equipment Interface window
indicating that the information is being received.
The data received from the Reichert iPac is displayed in the Pachymetry box
on the Exam - Special Tests tab.
120
ExamWRITER Equipment Integration User’s Guide
Chapter 15
Using the Reichert Equipment Interface
Using the Auto Phoropter RS & Auto Phoropter Visutron900
7. Click the Exam - Special Tests tab within the patient’s exam chart to view
the Reichert iPac data.
NOTE
Using the
Auto
Phoropter RS
& Auto
Phoropter
Visutron900
The iPac only provides data for the corneal thickness, and does
not calculate the adjusted IOP. ExamWRITER automatically
calculates the adjusted IOP after the corneal thickness and IOP
data fields are populated.
1. Open ExamWRITER.
2. Create a new blank exam or open a saved exam for an existing patient.
3. Click the Interfaces icon on the ExamWRITER chart window.
4. Select Reichert:Auto Phoropter Input.
The Reichert Rx window opens.
5. Select either the Auto-Refraction or Presenting Spectacle Rx radio button
to send the data you want to the Reichert equipment.
ExamWRITER Equipment Integration User’s Guide
121
Chapter 15
Using the Reichert Equipment Interface
Using the Auto Phoropter RS & Auto Phoropter Visutron900
6. Click Send.
The selected data is sent to the Reichert equipment and the Reichert
Launch file status dialog box informs you that the Reichert Input file was
successfully created.
7. Click OK to close the Reichert Launch file status dialog box.
8. Use the Reichert equipment to perform the examination or tests.
9. To send the data from the Reichert equipment to ExamWRITER, press the
COM button on the Reichert equipment.
10. Click the Interfaces icon on the ExamWRITER chart window.
122
ExamWRITER Equipment Integration User’s Guide
Chapter 15
Using the Reichert Equipment Interface
Using the Auto Phoropter RS & Auto Phoropter Visutron900
11. Select Reichert:Auto Phoropter Output.
The Officemate and Examwriter Equipment Interface window opens and
displays the following message: “Waiting for Reichert:Auto Phoropter Output
Exam Data…”
ExamWRITER opens the XML file created by the Reichert equipment. A
progress bar appears across the Officemate and Examwriter Equipment
Interface window indicating that the information is being received.
The data received from the Reichert equipment is displayed in the Manifest
box on the Vision/Rx tab.
ExamWRITER Equipment Integration User’s Guide
123
Chapter 15
Using the Reichert Equipment Interface
Reviewing Reflex UBM & Foresee PHP Images
12. Click the Vision/Rx tab within the patient’s exam chart to view the Reichert
equipment data.
Reviewing
Reflex UBM &
Foresee PHP
Images
NOTE
All Reflex UBM and Foresee PHP image file names must contain the
patient’s ExamWRITER ID number. Reflex UBM image file names
must use the Last,First YYMMDD-HHMMSS RecordNum-#.xxx
naming convention. Foresee PHP image file names must use the
Last_First_IDNum_EyeSide--V.4.0.1-YYYY.MM.DD-HH.MM.SS.jpg
naming convention.
1. Ensure that ExamWRITER is open.
2. Create a new blank exam or open a saved exam for an existing patient for
whom you have previously captured images.
124
ExamWRITER Equipment Integration User’s Guide
Chapter 15
Using the Reichert Equipment Interface
Reviewing Reflex UBM & Foresee PHP Images
3. Click the Interfaces icon on the ExamWRITER chart window and select
Reichert:UBM Image Review.
OR
Click the Interfaces icon on the ExamWRITER chart window and select
Reichert:PHP Image Review.
The UBM Images or PHP Images window opens and displays a list of
images.
4. Double-click on a row to display and review an image.
ExamWRITER Equipment Integration User’s Guide
125
Chapter 15
Using the Reichert Equipment Interface
Reviewing Reflex UBM & Foresee PHP Images
126
ExamWRITER Equipment Integration User’s Guide
Using the ReSeeVit Interface
16
In this chapter:
ReSeeVit
Interface
Overview
•
ReSeeVit Interface Overview, 127
•
Setting Up the ReSeeVit Interface, 128
•
Using ReSeeVit Capture, 130
ExamWRITER interfaces with the Veatch ReSeeVit software before the capture
process when patient information is transferred from ExamWRITER into the
ReSeeVit software to create or update retinal images and during the review
process when the newly created or existing ReSeeVit retinal images are available
for on-screen review.
The ReSeeVit equipment interface saves equipment data to the ReSeeVit
database and ExamWRITER reads that database through a mapped drive. The
equipment data can be shared and accessed at multiple computers, as long as
the computers are networked properly and the databases are properly mapped
and shared. The ReSeeVit interface uses a serial RS-232 cable or null modem to
connect to the computer or Terminal Server. The ReSeeVit interface can only use
one camera per Terminal Server.
NOTES
• ReSeeVit equipment cannot be interfaced with the multilocation
version of ExamWRITER.
• You must install ExamWRITER and the ReSeeVit software on the
computer to which the ReSeeVit equipment is connected; then
you can install the ReSeeVit software on the server on which
ExamWRITER is installed. Contact Veatch Ophthalmic
Instruments to determine if the ReSeeVit software will run on an
OS server and if any additional licensing needs to be purchased
when using it with Terminal Server.
• Back up your ReSeeVit data! Contact Veatch Ophthalmic
Instruments for recommended backup procedures or visit the
Veatch Web site at http://www.veatchinstruments.com.
®
ExamWRITER Equipment Integration User’s Guide
127
Chapter 16
Using the ReSeeVit Interface
Setting Up the ReSeeVit Interface
Setting Up the
ReSeeVit
Interface
To view your ReSeeVit images within ExamWRITER, you will need to adjust some
settings on each computer on which you want to view the images. This section
includes the following topics:
•
To set up the computer connected to the ReSeeVit equipment, 128
•
To set up other computers on the network, 129
 To set up the computer connected to the ReSeeVit equipment
NOTE
You must set up the ReSeeVit interface from the computer
connected to the ReSeeVit equipment.
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
2. Select the Write captured equipment data to a file check box if you want
ExamWRITER to record data received from ReSeeVit in a file for debugging
purposes; otherwise, go to step 3.
3. Deselect the check boxes next to any equipment you are not using as
needed.
4. Select the Veatch: Capture ReSeeVit check box.
5. Select one Veatch Review check box based on your setup:
If you are using…
Select this Veatch Review check box…
A single ReSeeVit camera The Veatch: Review ReSeeVit check box that
corresponds with your camera.
Multiple ReSeeVit
cameras
128
The Veatch Review ReSeeVit Multiple check
box.
ExamWRITER Equipment Integration User’s Guide
Chapter 16
Using the ReSeeVit Interface
Setting Up the ReSeeVit Interface
6. Select the drive and folder where the ReSeeVit software is installed.
NOTE
The ReSeeVit software is typically installed in the C:\Program
Files\Veatch Instruments\ReSeeVit 5.0 directory.
7. Click Save.
8. Click Exit.
9. Close ExamWRITER to save your changes.
10. Ensure that the CSODB.dll file is installed in the Program Files\Veatch
Instruments\ReSeeVit 5.0 folder.
NOTE
If the CSODB.dll is not installed, contact your Veatch
representative or Veatch support to obtain the CSODB.dll file.
 To set up other computers on the network
NOTE
These steps are only required if you need to be able to review
ReSeeVit images computers other than the computer connected to
the equipment. These steps must be performed on each computer
on which you will review ReSeeVit images.
1. On the ExamWRITER main window, click Tools, and select Equipment
Integration Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
2. Deselect the check boxes next to any equipment you are not using as
needed.
ExamWRITER Equipment Integration User’s Guide
129
Chapter 16
Using the ReSeeVit Interface
Using ReSeeVit Capture
3. Select one Veatch Review check box based on your setup:
If you are using…
Select this Veatch Review check box…
A single ReSeeVit camera The Veatch: Review ReSeeVit check box that
corresponds with your camera.
Multiple ReSeeVit
cameras
The Veatch Review ReSeeVit Multiple check
box.
4. Select the drive and folder where the ReSeeVit software is installed.
NOTE
The ReSeeVit software is typically installed in the Program
Files\Veatch Instruments\ReSeeVit 5.0 directory on the
computer connected to the equipment.
5. Click Save.
6. Click Exit.
7. Close and restart ExamWRITER to activate your changes.
Using
ReSeeVit
Capture
1. Ensure that ExamWRITER is open.
130
ExamWRITER Equipment Integration User’s Guide
2. Create a new blank exam or open a saved exam for an existing patient.
3. Click the Interfaces icon on the ExamWRITER chart window.
Chapter 16
Using the ReSeeVit Interface
Using ReSeeVit Capture
4. Select Veatch:Capture ReSeeVit.
The ReSeeVit 5.0 window opens with the patient’s last name, first name,
gender, date of birth, and prior exams ExamWRITER.
5. Follow the ReSeeVit capture instructions provided by Veatch to capture the
patient’s retinal images.
ExamWRITER Equipment Integration User’s Guide
131
Chapter 16
Using the ReSeeVit Interface
Using ReSeeVit Capture
6. Close the ReSeeVit software to return to ExamWRITER.
The ReSeeVit Exam icon will be displayed under the Images, Drawings and
Graphics category bar in ExamWRITER after you close and reopen the
patient’s exam.
7. Click the ReSeeVit Exam icon to review the patient’s captured image(s).
132
ExamWRITER Equipment Integration User’s Guide
Using the Topcon IMAGEnet
Interface
17
In this chapter:
•
Topcon IMAGEnet Interface Overview, 133
•
Setting Up the Topcon IMAGEnet Interface, 134
•
Using the Topcon IMAGEnet Interface, 137
•
Reviewing Topcon IMAGEnet Images, 137
These instructions cover the following Topcon products:
•
Topcon IMAGEnet 2000
•
Topcon IMAGEnet Lite
NOTE
Topcon
IMAGEnet
Interface
Overview
Topcon IMAGEnet Lite is not supported on Microsoft Windows 8.
ExamWRITER interfaces with the Topcon IMAGEnet 2000 and IMAGEnet Lite by
transferring information to the IMAGEnet capture software and by providing a
quick link from the patient’s exam in ExamWRITER to the captured images in
IMAGEnet’s capture software. ExamWRITER does not send data to the Topcon
IMAGEnet 2000 or IMAGEnet Lite equipment.
The Topcon IMAGEnet equipment interface saves equipment data to the
ExamWRITER database, which then can be accessed by all computers
networked to look at that database. The equipment data can be shared and
accessed at multiple computers, as long as the computers are networked
properly and the databases are properly mapped and shared. The Topcon
IMAGEnet interface uses a serial RS-232 cable or null modem to connect to the
computer. The Topcon IMAGEnet interface cannot connect to a Terminal Server.
®
ExamWRITER Equipment Integration User’s Guide
133
Chapter 17
Using the Topcon IMAGEnet Interface
Setting Up the Topcon IMAGEnet Interface
NOTES
• You must install ExamWRITER and the Topcon IMAGEnet software
on the computer to which the Topcon IMAGEnet equipment is
connected; then you can install the Topcon IMAGEnet software on
the server on which ExamWRITER is installed. Contact Topcon
Medical Systems to determine if the Topcon IMAGEnet software
will run on an OS server and if any additional licensing needs to be
purchased when using it with Terminal Server.
• Back up your Topcon IMAGEnet data! Contact Topcon Medical
Systems for recommended backup procedures.
Setting Up the
Topcon
IMAGEnet
Interface
NOTES
• You must set up the Topcon IMAGEnet 2000 or IMAGEnet Lite
interface from the computer with the Topcon IMAGEnet software
installed and connected to the Topcon IMAGEnet equipment.
• Even if you are using Terminal Services, you must install the
IMAGEnet software on each computer from which you plan to use
the software and you must set up a path on each computer to the
location where the Topcon IMAGEnet database is located.
• You must install the Topcon IMAGEnet interface on each computer
you want to view images captured with the Topcon IMAGEnet
software.
• Back up your IMAGEnet data. Contact Topcon for recommended
backup procedures.
• The CheckPatientInfo.dll file must be installed in the C:\Program
Files\TOPCON\IMAGEnet 2000 directory. If the
CheckPatientInfo.dll file is not there, contact Topcon support for
help.
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
2. Select the check boxes next to the equipment interfaces that you are setting
up.
134
ExamWRITER Equipment Integration User’s Guide
Chapter 17
Using the Topcon IMAGEnet Interface
Setting Up the Topcon IMAGEnet Interface
3. Select the drive and folder where the Topcon IMAGEnet software is installed
on each computer that you want to view images captured with the Topcon
IMAGEnet software.
NOTE
Typically, the path to the IMAGEnet software on the local
computer is C:\Program Files\TOPCON\IMAGEnet 2000. When
setting up the Topcon IMAGEnet 2000 or IMAGEnet Lite
interface on other computers on your network, select the drive
and folder of the computer that is connected to the Topcon
IMAGEnet equipment.
ExamWRITER Equipment Integration User’s Guide
135
Chapter 17
Using the Topcon IMAGEnet Interface
Setting Up the Topcon IMAGEnet Interface
NOTE
If you are using ExamWRITER through terminal services, you
will also need to type your computer’s network name in the
Computer Name field.
To locate your computer in Windows, open the System
Properties window. Click the Computer Name tab and locate
the name in the Full computer name field.
4. Click Save.
5. Click Exit.
6. Close and reopen ExamWRITER to activate your changes.
136
ExamWRITER Equipment Integration User’s Guide
Chapter 17
Using the Topcon IMAGEnet Interface
Using the Topcon IMAGEnet Interface
Using the
Topcon
IMAGEnet
Interface
1. Open ExamWRITER.
2. Create a new blank exam or open a saved exam for an existing patient.
3. Click the Interfaces icon on the ExamWRITER chart window.
4. Select your camera from the Interfaces menu.
The Topcon IMAGEnet application opens in capture mode with the patient’s
name and ID populated. The patient’s birth date and gender are also added
to the IMAGEnet patient record.
5. Capture images using the Topcon IMAGEnet application.
6. When you are finished capturing images, close the Topcon IMAGEnet
application and follow the instructions in “Reviewing Topcon IMAGEnet
Images” on page 137 to review the patient’s captured retinal images.
Reviewing
Topcon
IMAGEnet
Images
NOTE
Both the Topcon IMAGEnet review software and ExamWRITER must
be installed and set up on all workstations on which you want to
review images. The Topcon IMAGEnet review software must be
installed on all workstations, even if you are using Terminal Services.
Also, a path to the location where the Topcon IMAGEnet database is
located must be set up. For information on setting up the path to the
Topcon IMAGEnet database in the ExamWRITER equipment
interface, go to “Setting Up the Topcon IMAGEnet Interface” on page
134. ExamWRITER Equipment Integration User’s Guide
137
Chapter 17
Using the Topcon IMAGEnet Interface
Reviewing Topcon IMAGEnet Images
After you have captured images for a patient, the images are linked to the
patient’s ExamWRITER record. To review the patient’s captured images, perform
the following steps:
1. Open the patient’s exam for whom you have previously captured retinal
images in ExamWRITER.
The Topcon IMAGEnet logo appears on the exam under the Images,
Drawings and Graphics category bar. If you hold your cursor over the
IMAGEnet logo, the tooltip shows you the number of Topcon IMAGEnet
exams for that patient and the date of the last exam.
138
ExamWRITER Equipment Integration User’s Guide
Chapter 17
Using the Topcon IMAGEnet Interface
Reviewing Topcon IMAGEnet Images
2. Click the Topcon IMAGEnet icon.
The IMAGEnet Images windows opens, displaying a list of images for the
patient and the dates those images were captured.
3. Double-click the image you want to view.
The Topcon IMAGEnet application opens, displaying the patient’s images.
4. When your review is complete, close the Topcon IMAGEnet application.
ExamWRITER Equipment Integration User’s Guide
139
Chapter 17
Using the Topcon IMAGEnet Interface
Reviewing Topcon IMAGEnet Images
140
ExamWRITER Equipment Integration User’s Guide
Using the Topcon CL, KR, &
RM Interface
18
In this chapter:
Topcon Interface Overview
•
Topcon Interface Overview, 141
•
Setting Up the Topcon CL Equipment, 142
•
Setting Up the Topcon KR & RM Equipment, 142
•
Setting Up the Topcon Interface, 145
•
Using the Topcon Interface, 147
ExamWRITER receives data from the Topcon computerized lensmeter (CL
devices), autorefractor/keratometer (KR devices), and autorefractor (RM devices)
equipment and displays the data on the ExamWRITER chart window.
ExamWRITER does not send data to the Topcon equipment.
NOTE
The KR-8000, KR-8900, KR-8000PA, and KR-1
Autorefractor/Keratometer only import autorefraction and
keratometry data, and not topography images, into ExamWRITER.
The Topcon equipment interface saves equipment data to the ExamWRITER
database, which then can be accessed by all computers networked to look at that
database. The equipment data can be shared and accessed at multiple
computers, as long as the computers are networked properly and the databases
are properly mapped and shared. The Topcon interface uses a serial RS-232
cable or null modem to connect to the computer or Terminal Server. The Topcon
equipment interface can use multiple serial ports per PC.
NOTE
®
If you are using a digital refraction system with the KR-8000PA or
KR-7000P Autorefractor/Keratometer, it cannot be connected to the
computer that is running the ExamWRITER Topcon equipment
interface; instead, the KR-8000PA and KR-7000P
Autorefractor/Keratometer must be connected to a separate
computer so that the topographic data correctly transfers to the
digital refractor.
ExamWRITER Equipment Integration User’s Guide
141
Chapter 18
Using the Topcon CL, KR, & RM Interface
Setting Up the Topcon CL Equipment
Setting Up the
Topcon CL
Equipment
NOTE
If you have questions about the RS232C communication format or
setting up any of your Topcon equipment, contact your Topcon
dealer or Topcon Medical Systems using the address information on
the back cover of your Topcon equipment documentation. You can
also contact Topcon support at 1.866.922.6278.
Before you begin using the Topcon CL equipment interface, set up the Topcon CL
equipment. This section tells you how to set up your Topcon CL equipment,
including how
•
To select the output data for the CL devices, 142
•
To set the communication format for the CL devices, 142
 To select the output data for the CL devices
1. Press the MODE button.
2. Select MENU.
3. Choose RS-232C.
4. Chose STD1.
5. Press the Print switch to output the data.
RS232C DATA OUT is displayed on the Topcon CL equipment monitor.
When the transmission is complete, RS232C SUCCESS is displayed on the
monitor.
 To set the communication format for the CL devices
1. When you are in the INITIAL MENU screen, press the Measurement switch
and move the cursor to ON-LINE.
2. Choose ON-LINE.
In the ON-LINE screen you can modify the computer lensmeter data
receiving format, communication format, and communication speed.
3. Press the Measurement switch and move the cursor to DATA FORMAT.
4. Press the Print switch and choose STD1 (Topcon STD1 format).
5. Press the Measurement switch.
Setting Up the
Topcon KR &
RM Equipment
NOTE
If you have questions about the RS232C communication format, the
KR-8900 output data type, or setting up any of your Topcon
equipment, contact your Topcon dealer or Topcon Medical Systems
using the address information on the back cover of your Topcon
equipment documentation. You can also contact Topcon support at
1.866.922.6278.
In the ON-LINE screen you can modify the computer lensmeter data receiving
format, communication format, and communication speed. In the SETTINGS
142
ExamWRITER Equipment Integration User’s Guide
Chapter 18
Using the Topcon CL, KR, & RM Interface
Setting Up the Topcon KR & RM Equipment
screen you can modify the output data type. This section tells you how to set up
your Topcon KR and RM equipment, including how
•
To set the communication format for the KR & RM devices, 143
•
To set the communication format and speed and select the output data
information for the KR-1 only, 143
•
To set the RS232C communication speed, 144
•
To select the RS232C output data type for the KR-7000 only, 144
•
To select the output data type for the KR-8900 only, 144
•
To change the settings for the KR-800 only, 144
 To set the communication format for the KR & RM devices
NOTES
• For information about setting up the KR-7000, go to “To select the
RS232C output data type for the KR-7000 only” on page 144.
• For information about setting up the KR-1, go to “To set the
communication format and speed and select the output data
information for the KR-1 only” on page 143.
1. When you are in the INITIAL MENU screen, press the Measurement switch
and move the cursor to ON-LINE.
2. Choose ON-LINE.
3. Press the Measurement switch and move the cursor to DATA FORMAT.
4. Press the Print switch and choose STD1 (Topcon STD1 format).
5. Press the Measurement switch.
 To set the communication format and speed and select
the output data information for the KR-1 only
1. Tap Settings on the touch-screen monitor.
2. Tap Index.
3. Tap Comm.
4. Set the Format to STD1.
5. Set the Baud Rate to 2400.
6. Set the Output Data to ALL.
7. Set the Output Port to ON.
ExamWRITER Equipment Integration User’s Guide
143
Chapter 18
Using the Topcon CL, KR, & RM Interface
Setting Up the Topcon KR & RM Equipment
 To set the RS232C communication speed
NOTE
If you are using the Topcon color mapping software and topographer
and the KR-8000PA autorefractor/keratometer, you must set the
RS232C communication speed baud rate to the appropriate rate to
import your topography images into the color mapping software and
then reset it to 2400, as instructed below, to import measurements
from the KR-8000PA into ExamWRITER. You cannot import
topography images into ExamWRITER.
1. When you are in the INITIAL MENU screen, press the Measurement switch
and move the cursor to ON-LINE.
2. Choose ON-LINE.
3. Press the Measurement switch and move the cursor to DATA FORMAT.
4. Press the Print switch and choose 2400 (baud rate 2400).
5. Press the Measurement switch.
 To select the RS232C output data type for the KR-7000 only
1. When you are in the INITIAL MENU screen, press the Measurement switch
and move the cursor to ON-LINE.
2. Choose ON-LINE.
3. Set the RS232C format to OLD FORMAT.
4. Press the Print switch and choose 9600 (baud rate 9600).
5. Press the Measurement Switch.
 To select the output data type for the KR-8900 only
1. Press the Menu Switch.
1. When you are in the SETTINGS MENU screen, press the Measurement
Switch.
2. Set the Output Data Type to Yes.
 To change the settings for the KR-800 only
1. Press the button with the wrench symbol.
2. Press the Comm button.
3. Set the Output to ON.
4. Set the Baud Rate to 2400.
5. Set the Serial Connection to STD1.
144
ExamWRITER Equipment Integration User’s Guide
Chapter 18
Using the Topcon CL, KR, & RM Interface
Setting Up the Topcon Interface
Setting Up the
Topcon Interface
NOTES
• You must set up the Topcon CL, KR, and RM interface from the
computer connected to the Topcon equipment.
• You must be using ExamWRITER version 7.3 or above to interface
with the Topcon CL equipment.
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
2. Select the Write captured equipment data to a file check box if you want
ExamWRITER to record data received from Topcon in a file for debugging
purposes; otherwise, go to step 3.
3. Select the check boxes next to the equipment interfaces that you are setting
up.

If you are setting up a Topcon CL equipment interface, select the
Topcon CL Devices check box.

If you are setting up a Topcon KR equipment interface, select the
Topcon ARK Devices check box.
NOTE
At this time, you must use a Topcon digital refraction system to
interface a Topcon RM device with ExamWRITER.
ExamWRITER Equipment Integration User’s Guide
145
Chapter 18
Using the Topcon CL, KR, & RM Interface
Setting Up the Topcon Interface
4. Select one of the following options from the Data Source drop-down menu:

If you are connecting this computer to the Topcon equipment with a
serial cable, select Port and then select a communication port number
from the Comm Port drop-down menu. The Comm Port default is 1.

If you are not connecting this computer to the Topcon equipment with a
serial cable, select File and then type the data file’s beginning
characters in the Data File Start Chars text box and select the data file’s
ending characters (extension) from the Data File End Char drop-down
menu to specify the start and end characters of data source files and
identify specific files from which the interface is importing data.
5. Click Save.
6. Click Exit.
7. Close and reopen ExamWRITER to activate your changes.
146
ExamWRITER Equipment Integration User’s Guide
Chapter 18
Using the Topcon CL, KR, & RM Interface
Using the Topcon Interface
Using the
Topcon Interface
1. Ensure that the Topcon CL, KR, or RM device is connected to the computer
that ExamWRITER is installed upon with a null modem serial cable.
2. Ensure that ExamWRITER is open.
3. Create a new blank exam or open a saved exam for an existing patient.
4. Click the Interfaces icon on the ExamWRITER chart window.
5. Select the Topcon device that corresponds to your Topcon equipment.
The Officemate/Examwriter Equipment Interface window opens and
displays a “Waiting for Topcon Exam Data…” message.
ExamWRITER Equipment Integration User’s Guide
147
Chapter 18
Using the Topcon CL, KR, & RM Interface
Using the Topcon Interface
6. Use the Topcon device to perform the exam tests
The test data is automatically transmitted to the computer at the end of the
test. You can also press Print on the Topcon device’s operation panel to
transmit the data to the computer.
The device’s built-in thermal printer prints the recorded data and displays a
“Printing result data” message. After all of the data has been transmitted
from the operation panel to the computer, a progress bar appears on the
Officemate/Examwriter Equipment Interface window.
After all of the received data is processed and saved to the OfficeMate
database, the ExamWRITER exam chart refreshes and all of the received
data from the Topcon device is displayed on the Vision/Rx Spectacle tab.
148
ExamWRITER Equipment Integration User’s Guide
Using the Topcon CT-80
Interface
19
In this chapter:
Topcon CT-80
Interface
Overview
•
Topcon CT-80 Interface Overview, 149
•
Setting Up the Topcon CT-80 Equipment, 149
•
Setting Up the Topcon CT-80 Interface, 150
•
Using the Topcon CT-80 Interface, 151
ExamWRITER receives data from the Topcon CT-80 equipment and displays the
data on the ExamWRITER chart window. ExamWRITER does not send data to the
Topcon CT-80 equipment.
The Topcon CT-80 equipment interface saves equipment data to the
ExamWRITER database, which then can be accessed by all computers
networked to look at that database. The equipment data can be shared and
accessed at multiple computers, as long as the computers are networked
properly and the databases are properly mapped and shared. The Topcon CT-80
interface uses a serial RS-232 cable or null modem to connect to the computer or
Terminal Server. The Topcon CT-80 equipment interface can use multiple serial
ports per PC.
Setting Up the
Topcon CT-80
Equipment
Before you begin using the Topcon equipment interface, set up the Topcon CT-80
equipment.
1. Press Menu on the Topcon CT-80 monitor.
2. Highlight RS-232C Mode.
3. Press the button on the joystick.
4. Press the measurement switch.
5. Ensure that the FORMAT is set to MODE 2; if it is not set to MODE 2, use the
cursor buttons to highlight FORMAT, press the measurement switch to select
FORMAT, and press the measurement switch to select MODE 2.
Selecting MODE 2 will ensure that data is transmitted to ExamWRITER after
each measurement is made.
®
ExamWRITER Equipment Integration User’s Guide
149
Chapter 19
Using the Topcon CT-80 Interface
Setting Up the Topcon CT-80 Interface
6. Ensure that SPEED [BPS] is set to 2400; if it is not set to 2400, use the
cursor buttons to highlight SPEED [BPS], press the measurement switch to
select SPEED [BPS], and press the measurement switch to select 2400.
Setting Up the
Topcon CT-80
Interface
NOTE
You must set up the Topcon CT-80 interface from the computer
connected to the Topcon CT-80 equipment.
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
2. Select the Write captured equipment data to a file check box if you want
ExamWRITER to record data received from the Topcon CT-80 in a file for
debugging purposes; otherwise, go to step 3.
3. Select the check boxes next to the equipment interfaces that you are setting
up.
4. Select 2400 from the Baud Rate drop-down menu.
5. Select a communication port number from the Comm Port drop-down
menu.
The Comm Port default is 1.
6. Click Save.
7. Click Exit.
8. Close and reopen ExamWRITER to activate your changes.
150
ExamWRITER Equipment Integration User’s Guide
Chapter 19
Using the Topcon CT-80 Interface
Using the Topcon CT-80 Interface
Using the
Topcon CT-80
Interface
1. Ensure that the Topcon CT-80 is connected to the computer that
ExamWRITER is installed upon with a serial cable or serial-to-USB adapter
cable.
NOTE
You can purchase serial cables and serial-to-USB adapter
cables from Topcon at 1.800.223.1130.
2. Ensure that ExamWRITER is open.
3. Create a new blank exam or open a saved exam for an existing patient.
4. Click the Interfaces icon on the ExamWRITER chart window.
5. Select Topcon:CT-80.
The Officemate/Examwriter Equipment Interface window opens and
displays a “Waiting for Topcon:CT-80 Exam Data…” message.
6. Perform the measurements on the Topcon CT-80. For information on
performing these measurements, see the Topcon CT-80 Owner’s Manual.
ExamWRITER Equipment Integration User’s Guide
151
Chapter 19
Using the Topcon CT-80 Interface
Using the Topcon CT-80 Interface
7. Press Print on the Topcon CT-80 operation panel.
The average of the measurements that you preformed is transmitted to
ExamWRITER. After all of the data has been transmitted from the operation
panel to the computer, a progress bar appears on the
Officemate/Examwriter Equipment Interface window.
After all of the received data is processed and saved to the OfficeMate
database, the ExamWRITER exam chart refreshes and all of the received
data from the operation panel is displayed in the IOP table on the
Examination tab.
152
ExamWRITER Equipment Integration User’s Guide
Using the Topcon CV-5000,
CV-5000S, CV-3000, &
CV-2500 Interface
20
In this chapter:
Topcon Interface Overview
•
Topcon Interface Overview, 153
•
Setting Up the Topcon Equipment, 153
•
Setting Up the Topcon Interface, 154
•
Using the Topcon Interface, 155
ExamWRITER receives data from the Topcon KB-50/CV-5000 (EXAM-5000),
CV-5000S, KB-IDS/CV-3000, and CV-2500 equipment and displays the data on
the ExamWRITER chart window. ExamWRITER does not send data to the Topcon
equipment.
The Topcon equipment interface saves equipment data to the ExamWRITER
database, which then can be accessed by all computers networked to look at that
database. The equipment data can be shared and accessed at multiple
computers, as long as the computers are networked properly and the databases
are properly mapped and shared. The Topcon interface uses a serial RS-232
cable or null modem to connect to the computer or Terminal Server. The Topcon
equipment interface can use multiple serial ports per PC.
Setting Up the
Topcon
Equipment
Before you begin using the Topcon equipment interface, you must set up the
Topcon equipment. This section tells you how to set up the equipment, including
how
•
To set up the Topcon KB-50/CV-5000 (EXAM-5000) or CV-5000S equipment,
153
•
To set up the Topcon KB-IDS/CV-3000 or CV-2500 equipment, 154
 To set up the Topcon KB-50/CV-5000 (EXAM-5000) or
CV-5000S equipment
Eyefinity requires that you set DIP switch 2 for the KB-50/CV-5000 and CV-5000S
equipment to connect to a computer to No. 1=ON and No. 2=OFF.
®
ExamWRITER Equipment Integration User’s Guide
153
Chapter 20
Using the Topcon CV-5000, CV-5000S, CV-3000, & CV-2500 Interface
Setting Up the Topcon Interface
In addition to setting the DIP switches, follow the instructions below to set up the
KB-50/CV-5000 or CV-5000S controller:
1. Hold down the Shift key and press the Menu key on the KB-50/CV-5000 or
CV-5000S controller.
The Settings window opens.
2. Press Other.
The Other window opens.
3. In the KB section, set the Print Switch item to Print and Export.
See the Topcon Instruction Manual CV 1DIAL Controller KB-50 for more
information on setting up the Topcon KB-50/CV-5000 and CV-5000S equipment.
 To set up the Topcon KB-IDS/CV-3000 or CV-2500 equipment
Eyefinity requires that you set DIP switch 2 for the KB-IDS/CV-3000 and CV-2500
equipment to connect to a computer to No. 1=ON and No. 2=OFF.
See the Topcon Instruction Manual CV 1DIAL Controller KB-1DS for more
information on setting up the Topcon KB-IDS/CV-3000 and CV-2500 equipment.
Setting Up the
Topcon Interface
NOTE
You must set up the Topcon interface from the computer connected
to the Topcon equipment.
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
2. Select the Write captured equipment data to a file check box if you want
ExamWRITER to record data received from Topcon in a file for debugging
purposes; otherwise, go to step 3.
3. Select the Topcon:KB-1DS/CV3000 check box.
154
ExamWRITER Equipment Integration User’s Guide
Chapter 20
Using the Topcon CV-5000, CV-5000S, CV-3000, & CV-2500 Interface
Using the Topcon Interface
4. Select a communication port number from the Comm Port drop-down
menu.
The Comm Port default is 1.
5. Click Save.
6. Click Exit.
7. Close and reopen ExamWRITER to activate your changes.
Using the
Topcon Interface
1. Ensure that the Topcon KB-50/CV-5000 (EXAM-5000), CV-5000S,
KB-IDS/CV-3000, or CV-2500 is connected to the computer that
ExamWRITER is installed upon with a serial cable or serial-to-USB adapter
cable.
NOTE
You can purchase serial cables and serial-to-USB adapter
cables from Topcon at 1.800.223.1130.
2. Ensure that ExamWRITER is open.
3. Create a new blank exam or open a saved exam for an existing patient.
4. Click the Interfaces icon on the ExamWRITER chart window.
ExamWRITER Equipment Integration User’s Guide
155
Chapter 20
Using the Topcon CV-5000, CV-5000S, CV-3000, & CV-2500 Interface
Using the Topcon Interface
5. Select Topcon:KB-1DS/CV3000.
The Officemate/Examwriter Equipment Interface window opens and
displays a “Waiting for Topcon:KB-1DS/CV3000 Exam Data…” message.
156
ExamWRITER Equipment Integration User’s Guide
Chapter 20
Using the Topcon CV-5000, CV-5000S, CV-3000, & CV-2500 Interface
Using the Topcon Interface
6. Press Print on the Topcon KB-50/CV-5000 (EXAM-5000), CV-5000S,
KB-IDS/CV-3000, or CV-2500 operation panel.
The operation panel’s built-in thermal printer prints the recorded data and
displays an “RS-232C” message and then a “Printing” message. After all of
the data has been transmitted from the operation panel to the computer, a
progress bar appears on the Officemate/Examwriter Equipment Interface
window.
After all of the received data is processed and saved to the OfficeMate
database, the ExamWRITER exam chart refreshes and all of the received
data from the operation panel is displayed on the Vision/Rx Spectacle and
Binocular tabs.
NOTE
In order to obtain a reading that stereopsis is present when
using the Topcon KB-50/CV-5000 (EXAM-5000), CV-5000S,
KB-IDS/CV-3000, or CV-2500, the patient must respond to the
first orientation (OD 45º/OS 135º) with [F1] (Diffreup) and the
second orientation with [F2] (DiffreDepth). This combination of
responses is the only combination that will transfer “PRES” into
the ExamWRITER Stereopsis Dist box. Any other combinations
of patient test responses will transfer “ABS” into the
ExamWRITER Stereopsis Dist box.
ExamWRITER Equipment Integration User’s Guide
157
Chapter 20
Using the Topcon CV-5000, CV-5000S, CV-3000, & CV-2500 Interface
Using the Topcon Interface
158
ExamWRITER Equipment Integration User’s Guide
Using the Vmax Vision PSF
Refractor Interface
21
In this chapter:
Vmax Vision
PSF Refractor
Interface
Overview
®
•
Vmax Vision PSF Refractor Interface Overview, 159
•
Setting Up the Vmax Vision PSF Refractor, 160
•
Setting Up Shared Folders, 160
•
Setting Up the Vmax Vision PSF Refractor Interface, 160
•
Using the Vmax Vision PSF Refractor Interface, 161
ExamWRITER exports auto refraction, presenting spectacle Rx, or manifest Rx
data to the Vmax Vision PSF Refractor. The PSF Refractor then exports final Rx
data to ExamWRITER.
The Vmax equipment data can be shared and accessed at multiple computers,
as long as the computers are networked properly and the databases are properly
mapped and shared. The Zeiss VISUPAC interface uses a mapped network drive
to connect to the computer or Terminal Server.
ExamWRITER Equipment Integration User’s Guide
159
Chapter 21
Using the Vmax Vision PSF Refractor Interface
Setting Up the Vmax Vision PSF Refractor
Setting Up the
Vmax Vision
PSF Refractor
Set up the Vmax Vision PSF Refractor on the same network on which
ExamWRITER is installed. The PSF Refractor does not need to be set up on the
same computer as ExamWRITER. For help with installing the PSF Refractor, see
the instructions from Vmax Vision that came with the device.
Setting Up
Shared
Folders
You must have specific shared files set up on your computer so that the Vmax
Vision PSF Refractor and OfficeMate can share data.
1. Check your computer to see if the following folders exist:

C:\Officemate\VMaxShare\Data\Vmax Capture

C:\Officemate\VMaxShare\Data\Vmax Launch
2. If the folders listed in step 1 do not exist, create the folders.
3. Right-click the VMaxShare folder.
4. Select Properties.
5. Click the Sharing tab.
6. Click Permissions.
7. Ensure that the Allow Full Control check box is selected.
8. Click OK.
Setting Up the
Vmax Vision
PSF Refractor
Interface
NOTE
You must be using ExamWRITER v10.5 or above to interface with the
Vmax Vision PSF Refractor.
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
2. Select the Write captured equipment data to a file check box if you want
ExamWRITER to record data received from the PSF Refractor in a file for
debugging purposes; otherwise, go to step 3.
3. Select the VMax Launch check box.
4. Type the name of the computer on which the PSF Refractor resides in the
Computer Name text box.
NOTES
• Press Enter to type multiple computer names.
• To find the name of your computer on Windows, open the
System Properties window. Click the Computer Name tab
and locate the name in the Full computer name field.
5. Select C:\Officemate\VMaxShare\Data\Vmax Launch as the location
where equipment data is stored in the Path to Equipment Data section.
6. Select the VMax Capture check box.
160
ExamWRITER Equipment Integration User’s Guide
Chapter 21
Using the Vmax Vision PSF Refractor Interface
Using the Vmax Vision PSF Refractor Interface
7. Select C:\Officemate\VMaxShare\Data\Vmax Capture as the location
where equipment data is stored in the Path to Equipment Data section.
8. Click Save.
9. Click Exit.
10. Close ExamWRITER to activate your changes.
Using the
Vmax Vision
PSF Refractor
Interface
1. Open ExamWRITER.
2. Create a new blank exam or open a saved exam for an existing patient.
3. Click the Interfaces icon on the ExamWRITER chart window.
4. Select VMax:VMax Launch.
The VMax Rx window opens.
5. Select the Auto Refraction, Presenting Spectacle Rx, or Manifest Rx
radio button and click Send.
The selected data, the patient name, and patient ID are sent to the Vmax
Vision PSF Refractor and the VMax Launch file status dialog box informs you
that the patient file was successfully created.
ExamWRITER Equipment Integration User’s Guide
161
Chapter 21
Using the Vmax Vision PSF Refractor Interface
Using the Vmax Vision PSF Refractor Interface
6. Click OK to close the VMax Launch file status dialog box.
7. Use the PSF Refractor to perform the examination or tests.
8. Click the Store button on the interface of the PSF Refractor.
The PSF Refractor exports the data to an XML file.
9. Click the Interfaces icon on the ExamWRITER chart window.
162
ExamWRITER Equipment Integration User’s Guide
Chapter 21
Using the Vmax Vision PSF Refractor Interface
Using the Vmax Vision PSF Refractor Interface
10. Select VMax:VMax Capture.
The Officemate and Examwriter Equipment Interface window opens and
displays the following message: “Waiting for VMax:VMax Capture Exam
Data…”
ExamWRITER opens the XML file created by the PSF Refractor. A progress
bar appears across the Officemate and Examwriter Equipment Interface
window indicating that the information is being received.
The data received from the PSF Refractor is displayed in the Final Spectacle
Rx box on the Vision/Rx tab. The same data is displayed in the XML file
format in the Notes box on the Vision/Rx tab.
NOTE
To view the full XML data, double-click the Notes box.
ExamWRITER Equipment Integration User’s Guide
163
Chapter 21
Using the Vmax Vision PSF Refractor Interface
Using the Vmax Vision PSF Refractor Interface
11. Click the Vision/Rx tab within the patient’s exam record to view the data
from the PSF Refractor.
164
ExamWRITER Equipment Integration User’s Guide
Using the Woodlyn Interface
22
In this chapter:
Woodlyn
Interface
Overview
•
Woodlyn Interface Overview, 165
•
Setting Up the Woodlyn Interface, 166
•
Configuring the Woodlyn Interface, 168
•
Using Woodlyn Capture, 172
•
Using Woodlyn Review, 173
ExamWRITER interfaces with the Woodlyn Medmont software before the capture
process when patient information is transferred from ExamWRITER into Woodlyn
Capture to create or update retinal images and during the review process when
the newly created or existing Woodlyn retinal images are available for on-screen
review.
You must install ExamWRITER and the Woodlyn Capture and Woodlyn Review
applications on the computer to which the Woodlyn equipment is connected; then
you can install the Woodlyn Capture and Woodlyn Review applications on the
server on which ExamWRITER is installed. Contact Woodlyn to determine if the
Woodlyn Capture and Woodlyn Review applications will run on an OS server and
if any additional licensing needs to be purchased when using it with Terminal
Server. Both applications can be used on the same workstation or on two
separate workstations.
The Woodlyn equipment interface saves equipment data to the EyeTel database
and ExamWRITER reads that database through a mapped drive. The equipment
data can be shared and accessed at multiple computers, as long as the
computers are networked properly and the databases are properly mapped and
shared. The Woodlyn equipment interface uses a serial RS-232 cable or null
modem to connect to the computer or Terminal Server. The Woodlyn interface can
only use one camera per Terminal Server.
NOTES
• Back up your Woodlyn data! Contact Woodlyn for recommended
backup procedures.
• You may need to use a dongle, provided by Woodlyn, to use the
Woodlyn Medmont software.
®
ExamWRITER Equipment Integration User’s Guide
165
Chapter 22
Using the Woodlyn Interface
Setting Up the Woodlyn Interface
Setting Up the
Woodlyn
Interface
NOTE
Follow the instructions below only on the OfficeMate server. You can
then enable the Woodlyn Capture and Woodlyn Review options in
ExamWRITER on the Setup ExamWRITER Equipment Interfaces
window on the computer where the Woodlyn equipment is
connected.
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
2. Select the Write captured equipment data to a file check box if you want
ExamWRITER to record data received from Woodlyn in a file for debugging
purposes; otherwise, go to step 3.
3. Select the check boxes next to the equipment interfaces that you are setting
up.
4. Select the drive and folder where the equipment data is stored.
5. Click Update DB Settings to update your database settings.
The SetupOM_EquipIF window opens.
6. Click OK.
7. Click Save.
8. Click Exit.
166
ExamWRITER Equipment Integration User’s Guide
Chapter 22
Using the Woodlyn Interface
Setting Up the Woodlyn Interface
9. Close and reopen ExamWRITER to activate your changes.
ExamWRITER Equipment Integration User’s Guide
167
Chapter 22
Using the Woodlyn Interface
Configuring the Woodlyn Interface
Configuring
the Woodlyn
Interface
Configuring the Woodlyn interface allows you to create patients in the Woodlyn
Medmont software and transfer the new patient’s information to ExamWRITER.
NOTE
You must set up the Woodlyn interface before you configure the
Woodlyn Medmont software! For information about setting up the
Woodlyn interface, go to “Setting Up the Woodlyn Interface” on page
166.
1. Open Windows Explorer.
2. Navigate to the Program Files\Medmont\Medmont Studio 4 directory
(most likely on your C drive).
3. Double-click MedmontConfigure.exe.
The Select Language window opens.
4. Select the language in which you want the Medmont Studio to run and click
OK.
The Medmont Studio Configuration Tool window opens.
168
ExamWRITER Equipment Integration User’s Guide
Chapter 22
Using the Woodlyn Interface
Configuring the Woodlyn Interface
5. Select the appropriate configuration for the computer and click Next.
6. Select appropriate legacy data information and click Next.
7. Specify the location where Medmont Studio will store its database and
related files and click Next.
ExamWRITER Equipment Integration User’s Guide
169
Chapter 22
Using the Woodlyn Interface
Configuring the Woodlyn Interface
8. Select OfficeMateSQL (if you are using a Microsoft SQL database) or
OfficeMateAccess (if you are using a Microsoft Access database) from the
Practice Management Integration drop-down menu and click Next.
The Confirm change of Integration window opens.
9. Click Yes to confirm the change of integration.
10. Select the location of the ExamWRITER data files and click Next.
11. If an existing Medmont Studio database has been detected, select to add
new or reset the system settings and click Next.
170
ExamWRITER Equipment Integration User’s Guide
Chapter 22
Using the Woodlyn Interface
Configuring the Woodlyn Interface
12. Select the Medmont Studio components that you want to make available
and click Next.
13. Select the software drivers, if necessary, for any new hardware devices that
you intend to install on this computer and click Next.
14. Click Finish.
ExamWRITER Equipment Integration User’s Guide
171
Chapter 22
Using the Woodlyn Interface
Using Woodlyn Capture
Using
Woodlyn
Capture
1. Ensure that ExamWRITER is open.
2. Create a new blank exam or open a saved exam for an existing patient.
3. Click the Interfaces icon on the ExamWRITER chart window.
4. Select Woodlyn:Capture Woodlyn.
The Medmont Studio window opens with the patient’s last name, first name,
middle initial, gender, date of birth, and patient ID pre-loaded from
ExamWRITER.
5. Follow the Woodlyn Capture instructions to capture the patient’s retinal
images.
6. Close Woodlyn Capture to return to ExamWRITER.
The Woodlyn icon will be displayed under the Images, Drawings and
Graphics category bar in ExamWRITER after you close and reopen
ExamWRITER.
7. Follow the instructions in “Using Woodlyn Review” on page 173 to review the
patient’s captured retinal images.
172
ExamWRITER Equipment Integration User’s Guide
Chapter 22
Using the Woodlyn Interface
Using Woodlyn Review
Using
Woodlyn
Review
NOTES
• The Woodlyn Review software can only be used from within
ExamWRITER when a patient’s retinal images have previously
been captured using ExamWRITER and the Woodlyn Capture
interface.
• Both the Woodlyn Review software and ExamWRITER must be
installed and set up on all workstations on which you want to
review images. 1. Ensure that ExamWRITER is open.
2. Create a new blank exam or open a saved exam for an existing patient for
whom you have previously captured retinal images.
3. Click the Interfaces icon on the ExamWRITER chart window and select
Woodlyn:Review DV2000, Woodlyn:ReviewE300, or Woodlyn:Review
M500.
OR
Click the Woodlyn icon under the ExamWRITER Images, Drawings and
Graphics category bar.
The Medmont Studio window opens and displays the patient’s history.
4. Follow the Woodlyn Review instructions to use the application.
5. Close Woodlyn Review to return to ExamWRITER.
ExamWRITER Equipment Integration User’s Guide
173
Chapter 22
Using the Woodlyn Interface
Using Woodlyn Review
174
ExamWRITER Equipment Integration User’s Guide
Using the Zeiss VISUPAC
Interface
23
In this chapter:
Zeiss VISUPAC
Interface
Overview
•
Zeiss VISUPAC Interface Overview, 175
•
Setting Up the Zeiss VISUPAC Equipment, 175
•
Setting Up the Zeiss VISUPAC Interface, 176
•
Capturing Images with Zeiss VISUPAC, 176
•
Reviewing Images with Zeiss VISUPAC, 177
ExamWRITER interfaces with the Zeiss VISUPAC software before the capture
process when patient information is transferred from ExamWRITER to create or
update images and during the review process when the newly created or existing
images are available for on-screen review.
ExamWRITER interfaces with the Zeiss VISUPAC capture application, but uses
the ExamWRITER File Viewer window to review images. The Zeiss VISUPAC
software must be installed on the same computers as ExamWRITER.
The Zeiss VISUPAC equipment interface saves equipment data to the Zeiss
VISUPAC database and ExamWRITER reads that database through a mapped
drive. The equipment data can be shared and accessed at multiple computers,
as long as the computers are networked properly and the databases are properly
mapped and shared. The Zeiss VISUPAC interface uses a mapped network drive
to connect to the computer or Terminal Server.
Setting Up the
Zeiss VISUPAC
Equipment
®
Before you begin using the Zeiss VISUPAC equipment interface, you must set up
the Zeiss VISUPAC equipment and software to properly interface with
ExamWRITER. Contact your Zeiss representative to set up the Zeiss VISUPAC
equipment and software.
ExamWRITER Equipment Integration User’s Guide
175
Chapter 23
Using the Zeiss VISUPAC Interface
Setting Up the Zeiss VISUPAC Interface
Setting Up the
Zeiss VISUPAC
Interface
NOTE
You must be using ExamWRITER version 8.0 or above to interface
with the Zeiss VISUPAC.
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
2. Select the check boxes next to the equipment interfaces that you are setting
up.
3. Select the drive and folder where the equipment data is stored. Ensure that
the location you select is the same location that you selected in the Zeiss
VISUPAC software settings. For information on setting up the Zeiss VISUPAC
equipment, contact your Zeiss representative.
NOTES
• The file path to the Zeiss Capture VisuPac equipment data is
typically the EIS/HIS folder on your network drive.
• The file path to the Zeiss Review VisuPac equipment data is
typically the OMATE32/DATA/HIS or OfficeMate/DATA/HIS
folder on your network drive.
4. Click Save.
5. Click Exit.
6. Close and reopen ExamWRITER to activate your changes.
Capturing
Images with
Zeiss VISUPAC
1. Ensure that ExamWRITER is open.
176
ExamWRITER Equipment Integration User’s Guide
2. Create a new blank exam or open a saved exam for an existing patient.
3. Click the Interfaces icon on the ExamWRITER chart window.
Chapter 23
Using the Zeiss VISUPAC Interface
Reviewing Images with Zeiss VISUPAC
4. Select Zeiss:Capture VisuPac.
5. Follow the Zeiss VISUPAC software capture instructions to capture the
patient’s images.
6. Close the Zeiss VISUPAC capture software to return to ExamWRITER.
The Zeiss icon will be displayed under the Images, Drawings and Graphics
category bar in ExamWRITER after you close and reopen ExamWRITER.
7. Click Process Images.
8. Follow the instructions in “Reviewing Images with Zeiss VISUPAC” on page
177 to review the patient’s captured images.
Reviewing
Images with
Zeiss VISUPAC
1. Ensure that ExamWRITER is open.
2. Create a new blank exam or open a saved exam for an existing patient for
whom you have previously captured images.
3. Click the Interfaces icon on the ExamWRITER chart window and select
Zeiss:Review VisuPac.
OR
Click the Zeiss icon under the ExamWRITER Images, Drawings and
Graphics category bar.
The VisuPac Images window opens
ExamWRITER Equipment Integration User’s Guide
177
Chapter 23
Using the Zeiss VISUPAC Interface
Reviewing Images with Zeiss VISUPAC
4. Double-click the image you want to review.
The image opens in the File Viewer window.
5. When you are finished reviewing the images, close the File Viewer window
to return to ExamWRITER.
178
ExamWRITER Equipment Integration User’s Guide
Using the Zeiss i.Profiler &
i.Scription Interface
24
In this chapter:
Zeiss i.Profiers
& i.Scription
Interface
Overview
•
Zeiss i.Profiers & i.Scription Interface Overview, 179
•
Setting Up the Zeiss i.Profiler Equipment & i.Scription Software, 180
•
Setting Up the Zeiss i.Profiler & i.Scription Interface, 180
•
Using the Zeiss i.Profiler & i.Scription Interface, 181
The Zeiss i.Profiler and i.Scription interface process is described below:
1. ExamWRITER exports patient demographic data to the Zeiss i.Profiler
equipment.
2. The Zeiss i.Profiler equipment exports auto refraction data to ExamWRITER;
the data is displayed in the Auto Refraction box on the Vision/Rx tab on the
ExamWRITER chart window.
3. ExamWRITER exports the auto refraction Rx and manifest Rx, if recorded, to
the Zeiss i.Scription software.
4. The Zeiss i.Scription software calculates the Rx to the one-hundredth
decimal point and exports the final Rx or the manifest Rx to ExamWRITER;
the data is displayed in the Final Spectacle Rx box on the Vision/Rx tab on
the ExamWRITER chart window.
®
ExamWRITER Equipment Integration User’s Guide
179
Chapter 24
Using the Zeiss i.Profiler & i.Scription Interface
Setting Up the Zeiss i.Profiler Equipment & i.Scription Software
Setting Up the
Zeiss i.Profiler
Equipment &
i.Scription
Software
 Set up the Zeiss i.Profiler equipment on the same network on which
ExamWRITER is installed. The equipment does not need to be set up on the
same computer as ExamWRITER. For help on installing the equipment, see
the instructions from Zeiss that came with the equipment.
 Install the Zeiss i.Scription software on the same network on which
ExamWRITER is installed. The software does not need to be installed on the
same computer as ExamWRITER. For help on installing the software, see the
instructions from Zeiss that came with the software.
NOTE
Setting Up the
Zeiss i.Profiler
& i.Scription
Interface
Although you can use the Zeiss i.Profiler equipment without the
i.Scription software, you cannot use the i.Scription software
without the i.Profiler equipment.
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
2. Select the Write captured equipment data to a file check box if you want
ExamWRITER to record data received from Zeiss in a file for debugging
purposes; otherwise, go to step 3.
3. Select the check box (Zeiss iScription Export, Zeiss iScription Import, or
Zeiss iProfilerPlus) next to the equipment interface that you are using.
4. Select the location on the network where Zeiss saves the data file:
a. Select a drive from the drop-down menu.
b. Select the folder in which Zeiss saves the data file.
5. If you selected the Zeiss iScription Export or Zeiss iProfilerPlus check
boxes in step 3, type the name of the i.Profiler Plus in the Computer Name
180
ExamWRITER Equipment Integration User’s Guide
Chapter 24
Using the Zeiss i.Profiler & i.Scription Interface
Using the Zeiss i.Profiler & i.Scription Interface
text box. For example, type IPROFPXXXXXX, where XXXXXX is the i.Profiler
Plus serial number.
6. Repeat steps 3–5 for each piece of Zeiss equipment or software (Zeiss
iScription Export, Zeiss iScription Import, and Zeiss iProfilerPlus) that you
are using.
7. Click Save.
8. Click Exit.
9. Close ExamWRITER to activate your changes.
Using the Zeiss
i.Profiler &
i.Scription
Interface
1. Ensure that the Zeiss IP2.MatheTray program is running.
NOTE
The IP2.MatheTray.exe is located in the Program Files\CZV
folder. This program is running if you see a Zeiss icon in your
Windows taskbar.
2. Ensure that ExamWRITER is open.
3. Create a new blank exam or open a saved exam for an existing patient.
4. Click the Interfaces icon on the ExamWRITER chart window.
5. Select Zeiss:iProfilerPlus.
ExamWRITER exports patient demographic data to the Zeiss i.Profiler
equipment, imports auto refraction data from the Zeiss i.Profiler equipment,
ExamWRITER Equipment Integration User’s Guide
181
Chapter 24
Using the Zeiss i.Profiler & i.Scription Interface
Using the Zeiss i.Profiler & i.Scription Interface
and displays the data in the Auto Refraction box on the Vision/Rx tab on the
ExamWRITER chart window.
6. If you want to export the manifest Rx to the Zeiss i.Scription software,
complete the instructions below; otherwise, go to step 7.
a. Click your cursor in the Auto Refraction box on the Vision/Rx tab on the
ExamWRITER chart window.
b. Click Copy to Manifest to copy the auto refraction Rx to the manifest Rx;
otherwise, skip to step 7.
c. If you want to create a prismatic Rx in the Zeiss i.Scription software,
ensure that the H Prism and V Prism values are recorded for the
patient.
d. Click Save/Exit.
7. Click the Interfaces icon on the ExamWRITER chart window.
8. Select Zeiss:iScription Export.
The auto refraction Rx and manifest Rx, if recorded, is exported from
ExamWRITER and imported into the Zeiss i.Scription software.
9. Open the i.Scription software from the task bar.
The i.Scription Software - Review window opens.
182
ExamWRITER Equipment Integration User’s Guide
Chapter 24
Using the Zeiss i.Profiler & i.Scription Interface
Using the Zeiss i.Profiler & i.Scription Interface
10. Click i.Scription.
The i.Scription Software - iScription window opens.
11. If you recorded H prism and V prism values in the patient’s manifest Rx in
ExamWRITER and you want to create a prismatic Rx for the patient, click
Prismatic prescription.
12. Click Calculate i.Scription.
The Rx is calculated to the one-hundredth decimal point and displayed in
the i.Scription table.
13. Click one of the following buttons:

i.Scription to Final Rx to use the calculated i.Scription Rx as the
patient’s final Rx.

Manifest to Final Rx to use the calculated i.Scription Rx as the patient’s
manifest Rx.
ExamWRITER Equipment Integration User’s Guide
183
Chapter 24
Using the Zeiss i.Profiler & i.Scription Interface
Using the Zeiss i.Profiler & i.Scription Interface
14. Click Export.
The final Rx or manifest Rx is exported to ExamWRITER and the Pontus
Information window opens.
15. Click OK.
16. Close the i.Scription software.
17. Click the Interfaces icon on the ExamWRITER chart window.
184
ExamWRITER Equipment Integration User’s Guide
Chapter 24
Using the Zeiss i.Profiler & i.Scription Interface
Using the Zeiss i.Profiler & i.Scription Interface
18. Select Zeiss:iScription Import.
The Rx exported from the i.Scription software is imported into the Final
Spectacle Rx box on the Vision/Rx tab on the ExamWRITER chart window.
ExamWRITER Equipment Integration User’s Guide
185
Chapter 24
Using the Zeiss i.Profiler & i.Scription Interface
Using the Zeiss i.Profiler & i.Scription Interface
186
ExamWRITER Equipment Integration User’s Guide
Using the Zeiss FORUM
Interface
25
In this chapter:
Zeiss FORUM
Interface
Overview
•
Zeiss FORUM Interface Overview, 187
•
Setting Up Zeiss FORUM, 187
•
Setting Up the Zeiss FORUM Viewer Interface, 188
•
Viewing Data and Images with Zeiss FORUM, 190
The Zeiss FORUM Viewer is a Web-based application used to remotely archive,
view, and manage diagnostic data and images from any computer within the local
area network (LAN) for all Carl Zeiss Meditec diagnostic instruments.
ExamWRITER interfaces with the Zeiss FORUM Viewer to review patient exam
data and images. When ExamWRITER patient information is linked with Zeiss
data and images, exam data and images are available for review. The Zeiss
FORUM Viewer interface uses a mapped network drive to connect to the
computer or Terminal Server.
NOTE
Setting Up
Zeiss FORUM
Data and images can be viewed on any Windows workstation at the
main office or at other practice locations, at home, or with any
Internet connection. Refer to the Zeiss FORUM documentation for
more information on Internet connection setup requirements.
Before you begin using the Zeiss FORUM equipment interface, you must set up
the Zeiss FORUM equipment and software to properly interface with
ExamWRITER.
Contact your Zeiss representative to set up the Zeiss FORUM equipment and
software.
®
ExamWRITER Equipment Integration User’s Guide
187
Chapter 25
Using the Zeiss FORUM Interface
Setting Up the Zeiss FORUM Viewer Interface
In order to use the Zeiss FORUM Viewer, you must:
•
Install ExamWRITER v9.5.3 or above.
•
If you are using ExamWRITER v10.6, v10.7, or v10.8, download and install
the latest Zeiss.dll file at
http://www.officemate.net/omkb/Article.aspx?id=32403.
•
Install the Zeiss FORUM Viewer 3.1 on your workstations. Refer to the Zeiss
documentation or Carl Zeiss Meditec Customer Care for more information
and help on installing Zeiss FORUM.
NOTE
•
•
The Zeiss FORUM Viewer 2.6 and below is no longer
supported.
Configure the following parts of the Zeiss FORUM in the Zeiss Configuration
Administration:

IDs in the Patient ID Configuration window. IDs in the Zeiss Configuration
Administration must match the IDs in ExamWRITER.

The HIS folder. This folder must exist on the Zeiss FORUM server in the
root directory (C:\HIS).

The DICOM import folder.

The latest license files. Ensure that these files are in the correct FORUM
folder (C:\Program Files\CZM\Forum).
Set up a mapped drive on your workstations to the HIS folder on the Zeiss
FORUM server.
Setting Up the
Zeiss FORUM
Viewer Interface
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
188
ExamWRITER Equipment Integration User’s Guide
The Setup ExamWRITER Equipment Interfaces window opens.
Chapter 25
Using the Zeiss FORUM Interface
Setting Up the Zeiss FORUM Viewer Interface
2. Select the check box next to Zeiss Forum Viewer.
3. Type the name of the Zeiss FORUM server where the Zeiss FORUM software
is installed in the Computer Name box.
NOTE
• If you are running ExamWRITER locally on multiple
workstations and you want to set up the Zeiss FORUM
Viewer interface on each of those computers, type the name
of the Zeiss FORUM server where the Zeiss FORUM software
is installed.
• To find the name of the Zeiss FORUM server in Windows,
open the System Properties window. Click the Computer
Name tab and locate the name in the Full computer name
field.
4. Select the drive and folder where the Viewer data files are stored. Ensure
that the location that you select is the same location that you selected in the
Zeiss FORUM software settings. For information on setting up the Zeiss
FORUM equipment, contact your Zeiss representative.
NOTE
The file path to the Zeiss FORUM data is in the HIS folder on
your network drive.
5. Click Save.
6. Click Exit.
7. Close and reopen ExamWRITER to activate your changes.
ExamWRITER Equipment Integration User’s Guide
189
Chapter 25
Using the Zeiss FORUM Interface
Viewing Data and Images with Zeiss FORUM
Viewing Data
and Images
with Zeiss
FORUM
1. Ensure that ExamWRITER is open.
2. Create a new blank exam or open a saved exam for an existing patient for
whom you have previously captured images.
3. Click the Interfaces icon on the ExamWRITER chart window and select
Zeiss:Forum Viewer.
The Zeiss FORUM Viewer prompts you for the password.
4. Type the Zeiss FORUM password.
The Carl Zeiss Forum Patient Directory window opens displaying patient
names, IDs, and birth dates.
5. Double-click the name of the patient whose data you want to review.
The patient data and any images open in the Carl Zeiss Forum Patient
Directory window. Images open in a preview gallery at the bottom of the
window.
NOTE
If not all of the equipment is present, image file names are
displayed instead of the images.
6. When you are finished reviewing the data and images, close the Carl Zeiss
Forum Patient Directory window to return to ExamWRITER.
190
ExamWRITER Equipment Integration User’s Guide
Using the OCULUS Interface
26
In this chapter:
OCULUS Interface Overview
•
OCULUS Interface Overview, 191
•
Setting Up the OCULUS Interface, 191
•
Capturing OCULUS Visual Fields Data, 192
•
Reviewing OCULUS Visual Fields Data, 193
ExamWRITER interfaces with the OCULUS Easyfield C Perimeter and
Binoptometer 4P application before the capture process when patient information
(name and date of birth) is transferred from ExamWRITER into OCULUS and
during the review process when the OCULUS visual fields data is available for
on-screen review.
The OCULUS equipment interface saves equipment data to the OCULUS
database and ExamWRITER reads that database through a mapped drive. The
equipment data can be shared and accessed at multiple computers, as long as
the computers are networked properly and the databases are properly mapped
and shared. The OCULUS equipment interface uses a USB cable to connect to
the computer.
NOTES
• You must install ExamWRITER, the OCULUS application, and the
OCULUS driver on the computer to which the OCULUS
equipment is connected.
• Back up your OCULUS data! Contact OCULUS for recommended
backup procedures.
Setting Up the
OCULUS Interface
NOTE
You must set up the OCULUS interface from the computer
connected to the OCULUS equipment.
1. On the ExamWRITER main window, click Tools, select Equipment
Integration Setup, and select Equipment Interface Setup.
The Setup ExamWRITER Equipment Interfaces window opens.
2. Select the Debugging mode/Write to file check box if you want
ExamWRITER to record data received from OCULUS in a file for debugging
purposes; otherwise, go to step 3.
3. Select the check boxes next to the equipment interfaces that you are setting
up.
®
ExamWRITER Equipment Integration User’s Guide
191
Chapter 26
Using the OCULUS Interface
Capturing OCULUS Visual Fields Data
4. Select the drive and folder where the equipment data is stored (for example,
C:\Easy or C:\Bino4).
5. Click Save.
6. Click Exit.
7. Close and reopen ExamWRITER to activate your changes.
Capturing
OCULUS
Visual Fields
Data
1. Ensure that ExamWRITER is open.
2. Create a new blank exam or open a saved exam for an existing patient.
3. Click the Interfaces icon on the ExamWRITER chart window.
4. Select Oculus: Capture Easyfield or Oculus: Capture Binoptometer.
The OCULUS application opens with the patient’s name and date of birth
pre-loaded from ExamWRITER.
192
ExamWRITER Equipment Integration User’s Guide
Chapter 26
Using the OCULUS Interface
Reviewing OCULUS Visual Fields Data
5. Follow the OCULUS instructions to capture the patient’s visual fields data.
6. Close the OCULUS application to return to ExamWRITER.
The OCULUS icon will be displayed under the Images, Drawings and
Graphics category bar in ExamWRITER after you close and reopen
ExamWRITER.
7. Follow the instructions in “Reviewing OCULUS Visual Fields Data” on page
193 to review the patient’s captured visual fields data.
Reviewing
OCULUS
Visual Fields
Data
1. Ensure that ExamWRITER is open.
2. Create a new blank exam or open a saved exam for an existing patient for
whom you have previously captured OCULUS visual fields data.
ExamWRITER Equipment Integration User’s Guide
193
Chapter 26
Using the OCULUS Interface
Reviewing OCULUS Visual Fields Data
3. Click the Interfaces icon on the ExamWRITER chart window and select
Oculus: Review Easyfield or Oculus: Review Binoptometer.
OR
Click the OCULUS icon under the ExamWRITER Images, Drawings and
Graphics category bar.
The OCULUS application opens with the patient’s visual fields data.
4. Follow the OCULUS instructions to use the application.
5. Close the OCULUS application to return to ExamWRITER.
194
ExamWRITER Equipment Integration User’s Guide
Setting Up Serial Cable Interfaces
with Terminal Services
A
If you are using ExamWRITER with Terminal Services, you can connect your serial
cable-based equipment to your local computer even though your ExamWRITER
software is running on a remote computer. In order to use serial cable-based
ExamWRITER equipment interfaces with Terminal Services, you must allow your
local serial port to be available during a terminal server session.
NOTES
• Depending on the policies of your network, local serial port
mapping might be disabled for some or all remote connections.
• You might see a security warning message asking if you want to
connect your local disk drives, local ports, or smart card to the
remote computer. To enable one or more of these connections,
select the appropriate check boxes. If you choose not to enable
these connections, a remote connection is still established, but
these resources are not available to the remote computer.
• The following instructions require that you have Microsoft Remote
Desktop Connection 6 (or later) installed on the local computer.
Information about Remote Desktop Connection can be found on
Microsoft’s Web site at
http://support.microsoft.com/default.aspx/kb/925876.
 To allow your local serial port to be available during a
terminal server session
1. Click Start.
2. Select Programs or All Programs.
3. Select Accessories.
4. Select Remote Desktop Connection.
The Remote Desktop Connection window opens.
®
ExamWRITER Equipment Integration User’s Guide
195
Appendix A
Setting Up Serial Cable Interfaces with Terminal Services
5. Click Options.
The Remote Desktop Connection window expands to show you more
options.
6. Click the Local Resources tab.
7. Click More in the Local devices and resources box.
8. Select the Serial ports check box.
196
ExamWRITER Equipment Integration User’s Guide
Appendix A
Setting Up Serial Cable Interfaces with Terminal Services
9. Click OK.
10. Click the General tab.
11. Click Save in the Connection settings box.
NOTE
The serial port information that you set up becomes the default
setting for all Remote Desktop connections on this computer
and is saved in a file called Default.rdp.
ExamWRITER Equipment Integration User’s Guide
197
Appendix A
Setting Up Serial Cable Interfaces with Terminal Services
12. Follow the instructions in the appropriate chapter in the ExamWRITER
Equipment Integration User’s Guide to set up and use your ExamWRITER
equipment interface.
198
ExamWRITER Equipment Integration User’s Guide
Index
EyeTel interface, 17
capturing images, 18
overview, 17
reviewing images, 20
setting up, 17
Symbols
.NET Framework, 50
B
Binoptometer 4P, 191
F
C
FORUM interface
overview, 187
reviewing data and images, 190
setting up, 187, 188
CDR-Mate
activating, 64
installing, 58
setting up, 60
CellChek software, 73, 77
CenterVue
capture, 2
H
Heidelberg camera
setting up, 22
Heidelberg interface, 21
capturing images, 22
Eye Explorer, 21
reviewing images, 24
setting up, 22
CenterVue interface, 1
computer name
entering, 136
locating, 136
HRT2, 21
D
HRT3, 21
daisy-chaining, 88
E
Eagle Eye 3000 interface, 7
setting up, 7
using, 8
Easyfield C Perimeter, 191
ExamWRITER/Optos Patient Synchronization
Utility, 1, 106
Eye Explorer, 21
EyeScape interface, 11
capturing images, 13
reviewing images, 15
setting up, 11
®
Humphrey analysis reports, 47
Change Analysis, 30
Overview, 30
Three-in-One, 30
Humphrey equipment
EMR/RMS export setup, 34
HFA network setup, 33
save/transmit option, 39
setting up, 30
test connection, 35, 38, 39
user manual, 30
work list setup, 37
Humphrey interface, 27
recalling patient data, 42
setting up, 27
using, 40
ExamWRITER Equipment Integration User’s Guide
199
Index
Humphrey Matrix
OM Matrix Gateway, 51
reports, 53
Humphrey Matrix equipment
setting up, 49
Humphrey Matrix interface, 49
setting up, 50
using, 51
Huvitz interface, 55
setting up, 56
using, 64
I
Konan interface, 71
capturing images, 77
overview, 71
reviewing images, 79
setting up, 75
setting up equipment, 71
Kowa camera
setting up, 82
Kowa interface, 81
capturing images, 83
reviewing images, 85
setting up, 81
L
i.Profiler equipment
setting up, 180
license agreement, ii
i.Profiler interface, 179
setting up, 180
using, 181
M
Marco ARK, LM, RKT, & VL series equipment
setting up, 88
i.Scription interface, 179
export, 182
final Rx, 183
import, 185
manifest Rx, 183
setting up, 180
using, 181
Marco Connect, 90
Marco interface, 87
daisy-chaining, 88
firmware fix, 88
using, 91
Marco RT series equipment
setting up, 88
i.Scription software
setting up, 180
Marco RT-900, 88
iMACAM Capture, 68
Marco TRS series equipment
setting up, 88
iMACAM interface, 67
overview, 67
setting up, 67
MCU Software Version, 49
iMACAM Review, 70
N
IMAGEnet 2000 interface
using, 137
IMAGEnet 2000 interface. See Topcon IMAGEnet
2000 interface
installing CDR-Mate, 58
IP2.MatheTray, 181
NIDEK NAVIS-EX Capture, 101
NIDEK NAVIS-EX interface, 95
overview, 95
setting up, 100
setting up NIDEK software, 96
NIDEK NAVIS-EX Review, 102
K
Konan
CellChek software, 73, 77
200
ExamWRITER Equipment Integration User’s Guide
Index
O
OCULUS interface, 191
Binoptometer 4P, 191
capturing visual fields data, 192
Easyfield C Perimeter, 191
overview, 191
reviewing visual fields data, 193
setting up, 191
OM Matrix Gateway, 51
ReSeeVit interface, 127
overview, 127
setting up, 128
S
SBC Software Version, 49
serial cable interfaces
setting up to use with Terminal Services, 195
serial port, 195
Optos Capture, 107
Spectralis OCT, 21
Optos interface, 105
Daytona, 106
overview, 1, 105
setting up, 1, 106
T
Terminal Server
using Topcon IMAGEnet with, 136
Optos Review, 5, 109
Optos/ExamWRITER Patient Synchronization
Utility, 1, 106
Terminal Services
serial cable interfaces, 195
serial port, 195
P
Topcon CL equipment
communication format, 142
output data, 142
setting up, 142
prismatic Rx, 183
PSF Refractor. See Vmax Vision PSF Refractor
Topcon CL, KR, and RM interface, 141
setting up, 145
using, 147
R
Reichert
serial cables, 112
Reichert interface, 111
RS232 data bus baud rate, 113
RS232 data bus parameters, 113
setting up, 112
using the AL-200, 116
using the AL-500, 116
using the AL-700, 116
using the AT-555, 116
using the Auto Phoropter RS, 121
using the Auto Phoropter Visutron900, 121
using the Foresee PHP, 124
using the PT-100, 116
using the Reflex UBM, 124
using the Reichert 7, 116
using the Reichert 7CR, 116
using the RK-600, 118
using the RK-700, 118
ReichertCapture software, 112
Remote Desktop connection, 195
ReSeeVit Capture, 130
Topcon CT-80 equipment
setting up, 149
Topcon CT-80 interface, 149
setting up, 150
using, 151
Topcon CV2500 equipment
DIP switches, 154
setting up, 154
Topcon CV-5000S equipment
controller, 154
DIP switches, 153
setting up, 153
Topcon IMAGEnet 2000 interface, 133
reviewing images, 137
setting up, 134
using, 137
Topcon IMAGEnet Lite interface, 133
Topcon interface, 153
setting up, 154
using, 155
ExamWRITER Equipment Integration User’s Guide
201
Index
Topcon KB-50/CV-5000 equipment
controller, 154
DIP switches, 153
setting up, 153
Topcon KB-IDS/CV3000 equipment
DIP switches, 154
setting up, 154
Topcon KR equipment
communication format, 143
RS232C communication speed, 144
setting up, 142
Topcon KR-1 equipment
communication format, 143
communication speed, 143
output type, 143
Topcon KR-7000 equipment
RS232C output data, 144
Topcon KR-800 equipment
settings, 144
Woodlyn interface, 165
configuring, 168
dongle, 165
overview, 165
setting up, 166
Woodlyn Review, 173
Z
Zeiss FORUM interface
overview, 187
reviewing data and images, 190
setting up, 188
setting up Viewer interface, 187
Zeiss VISUPAC interface
capturing images, 176
overview, 175
reviewing images, 177
setting up, 176
setting up equipment, 175
Topcon KR-8800PA equipment
RS232C communication speed, 144
Topcon KR-8900 equipment
output data type, 144
Topcon RM equipment
communication format, 143
RS232C communication speed, 144
setting up, 142
V
VISUPAC interface
capturing images, 176
overview, 175
reviewing images, 177
setting up, 176
setting up equipment, 175
Vmax Vision PSF Refractor
setting up, 160
Vmax Vision PSF Refractor interface
capture, 163
launch, 161
overview, 159
setting up, 160
using, 161
W
Windows Installer, 50
Woodlyn Capture, 172
202
ExamWRITER Equipment Integration User’s Guide
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project