Model 8011

Model 8011
Model 8011 High Current Test Socket Kit
Connection instructions
Keithley Instruments, Inc.
28775 Aurora Road
Cleveland, Ohio 44139
1-888-KEITHLEY
http://www.keithley.com
Model 8011 High Current Test Socket Kit connection instructions
WARNING
Before proceeding, please read, understand, and follow all safety precautions provided at the
end of this document (page 14).
The Model 8011 High Current Test Socket Kit is designed for testing high power devices with 2600A Series
SourceMeter instruments and the Model 2651A High Power SourceMeter Instrument at currents up at 50 A pulsed
and voltages up to 40 V.
The Model 8011 includes two test boards, the 8011-DTB and the 8011-CTB, and cables and connectors for
simplifying connections between the SourceMeter instruments and the device. The 8011-DTB Device Test Board is
designed for testing type TO-247 packaged parts and axial components. The 8011-CTB Customizable Test Board
provides a blank space for user customization — the user can use the socket type of his or her choice.
Figure 1. Model 8011 High Current Test Socket Kit*
*Not shown: PA-1032 (these instructions); full set of cables; 8010-317 insulator plugs; and 2600-KIT.
PA-1032 Rev. A / April 2011
1
Model 8011 Connection Instructions
Power and environmental specifications
For indoor use only.
Specification
Value
Voltage limit
40 V DC maximum
Current limit
15 A DC maximum
Unpack and inspect the High Current Test Socket Kit
To unpack and inspect the High Current Test Socket Kit:
1. Inspect the box for damage.
2. Open the top of the box.
3. Remove the bags that contain the boards and accessories.
4. Carefully lift the High Current Test Socket Kit out of the box.
5. Inspect the materials for any obvious signs of physical damage. Report any damage to the shipping agent
immediately.
You should have received:
•
TO-247 Device Test Board (8010-120; to re-order, use Keithley Model number 8010-DTB)
•
Customizable Test Board (8010-130; to re-order, use Keithley Model number 8010-CTB)
•
PA-1032 Model 8011 High Current Test Socket Kit Connection Instructions (this document)
•
Four nylon hex standoffs with inserts on each end (ST-272-1)
•
Four Phillips flat-head screws (#6-32 x 3/8)
•
Two 8 in. (203.2 mm) stackable banana cables — one black (CA-560-0), one red (CA-560-2)
•
One mounting plate (8011-301)
•
Two 8-pin female plug connectors with screw terminal connections (CS-1629-8)
•
Four 9 in. (228.6 mm) black cables with one stackable banana plug and one unterminated end (CA-567-0)
•
Four 9 in. (228.6 mm) red cables with one stackable banana plug and one unterminated end (CA-567-2)
•
One 120 in. (304.6 cm) green/yellow ground cable with lugs (CA-568-120)
•
One #6-32 Keps® washer nut
•
Two insulator plugs (quantity includes a spare plug) (8010-317)
•
Two screw terminal connector kits (2600-KIT)
2
PA-1032 Rev. A / April 2011
Model 8011 Connection Instructions
Assembly instructions
NOTE
If you are customizing the 8010-130 Customizable Test Board, make customizations before assembling
the Model 8011.
See Figure 2 below.
Tools needed
•
Medium Phillips head screwdriver
•
5/16 in. nut driver or socket wrench
Assembly instructions
1. Select the TO-247 Device Test Board (8010-120) or Customizable Test Board (8010-130), as appropriate for
your application.
2. Connect one end of the ground cable to the ground screw at the center of the mounting plate.
3. Attach the ground cable with the #6-32 Keps® washer nut.
4. Thread the four nylon hex standoffs onto the four threaded posts on the mounting plate.
5. Place the 8010-120 or 8010-130 board on the standoffs.
6. Thread four of the Phillips flat-head screws through the 8010-120 or 8010-130 board.
7. Tighten the screws.
8. Attach the other end of the ground cable to protective earth (ground).
WARNING
The ground wire must be attached to a known safety earth (ground). Failure to attach the
ground wire to a known safety earth (ground) may result in electric shock.
Specifications are subject to change without notice.
PA-1032 Rev. A / April 2011
3
Model 8011 Connection Instructions
Figure 2. Model 8011 assembly
Four #6-32 x 3/8
Phillips flat-head screws
8010-130 Customizable Test Board
or
8010-120 TO-247 Device Test Board
Four ST-272-1 hex nylon standoffs
One #6-32 Keps® washer nut
One CA-568-120 green/yellow
ground cable with lugs
One 8011-301 mounting plate
Connect ground cable lug to protective earth (ground)
4
PA-1032 Rev. A / April 2011
Model 8011 Connection Instructions
Mounting instructions
The 8011-301 mounting plate should be used to secure your test board (8010-120 or 8010-130). It can also be
used to secure the assembled Model 8011 to a work surface using the mounting holes shown in Figure 3.
Figure 3. 8011-301 mounting plate dimensions
.156 in.
(3.962 mm)
5.250 in.
(133.35 mm)
Mounting holes
2.625 in.
(66.675 mm)
.750 in
(19.05 mm)
4.500 in
(114.3 mm)
6.000 in
(152.4 mm)
Specifications are subject to change without notice.
PA-1032 Rev. A / April 2011
5
Model 8011 Connection Instructions
NOTE
You can attach your own test board or testing fixture to the 8010-130 Customizable Test Board using the
#6-32 screw mounts provided.
Figure 4. 8010-130 Customizable Test Board dimensions
3.000 in.
(76.2 mm)
1.000 in.
(25.4 mm)
.156 in.
(3.962 mm)
3.750 in.
(95.25 mm)
1.750 in.
(44.45 mm)
4.250 in.
(107.95 mm)
#6-32
.750 in.
(19.05 mm)
3.500 in.
(88.9 mm)
.250 in.
(6.35 mm)
5.000 in.
(5 mm)
6
PA-1032 Rev. A / April 2011
Model 8011 Connection Instructions
General guidelines
The High Current Test Socket Kit provides several types of connections to allow you to make connections to meet
the needs of your application.
In general, when making connections:
•
Remove power from the SourceMeter instrument before making connections.
•
Make connections to the SourceMeter instrument before making connections to the Model 8011 boards.
•
Connect the force leads of the SourceMeter instrument to terminals 4, 5, and 6 of the Model 8011 boards.
Connect the sense leads to terminals 1, 2, and 3.
•
Install the device under test (DUT) after making the SourceMeter instrument connections.
Use-cases are shown below that demonstrate some typical connections.
Use-cases
Two-terminal DUT using a Model 2651A set to local sense
NOTE
These instructions provide general information about connections to the Model 2651A. For detail on the
connections, see the Model 2651A user documentation.
Tools needed
•
Small flat-head screwdriver
Instructions
1. Remove power from the Model 2651A.
2. Make connections as shown in Table 1 and Figure 5.
3. Install the 8010-317 insulator plug into the test socket.
4. Install the 2-terminal DUT into the axial posts labeled 4, 1, 3 and 6 on the 8010-120 board.
5. Return power to the Model 2651A.
6. Press the CONFIG key.
7. Press the SRC or MEAS key (the sense mode can be set from either the V-SOURCE or the V-MEAS menu).
8. If you pressed the SRC key, select V-SOURCE > SENSE-MODE.
9. If you pressed the MEAS key, select V-MEAS > SENSE-MODE.
10. Select 2-WIRE to enable local sense.
11. Test the device.
Specifications are subject to change without notice.
PA-1032 Rev. A / April 2011
7
Model 8011 Connection Instructions
Table 1. Connections for two-terminal device test (local sense)
Cable end
Cable
Cable end
Model 2651A rear panel
CA-552-1 (supplied with the Model 2651A)
Open
LO of CA-552-1
Black CA-560-0
8010-120 blue terminal 4
HI of CA-552-1
Red CA-560-2
8010-120 yellow terminal 6
Figure 5. Connections for two-terminal device test (local sense)
TO-247 Device Test
Board (8010-120)
2-Terminal DUT
From
Model 2651A
Plug
CA-560-0
(black)
8011
HIGH CURRENT TEST SOCKET KIT
RATINGS:
40V MAX.
15A DC MAX.
50A PULSE MAX.
HI
CA-552-1
(supplied with
Model 2651A)
CA-560-2
(red)
8010-317 insulator plug
(insert into 3-pin socket)
Two-terminal DUT using a Model 2651A set to remote sense
Tools needed
•
Small flat-head screwdriver
Instructions
1. Remove power from the Model 2651A.
2. Make connections as shown in Table 2 and Figure 6.
3. Install the 8010-317 insulator plug into the test socket.
4. Install the 2-terminal DUT into the axial posts labeled 4, 1, 3 and 6 on the 8010-120 board.
5. Return power to the Model 2651A.
8
PA-1032 Rev. A / April 2011
Model 8011 Connection Instructions
6. Press the CONFIG key.
7. Press the SRC or MEAS key (the sense mode can be set from either the V-SOURCE or the V-MEAS menu).
8. If you pressed the SRC key, select V-SOURCE > SENSE-MODE.
9. If you pressed the MEAS key, select V-MEAS > SENSE-MODE.
10. Select 4-WIRE to enable remote sense.
11. Test the device.
Table 2. Connections for two-terminal device test (remote sense)
Cable end
Cable
Cable end
Model 2651A rear panel
CA-552-1 (supplied with the Model 2651A)
Open
LO of CA-552-1
Black CA-560-0
8010-120 blue terminal 4
HI of CA-552-1
Red CA-560-2
8010-120 yellow terminal 6
Model 2651A rear panel
CA-557-1 (supplied with the Model 2651A)
CS-1629-8
Sense LO of CS-1629-8
Black CA-567-0
8010-120 blue terminal 1
Sense HI of CS-1629-8
Red CA-567-2
8010-120 yellow terminal 3
Figure 6. Connections for two-terminal device test (remote sense)
CA-567-0
(black)
S LO
G
From
Model 2651A
G
G
G
S HI
CA-557-1
(supplied with
Model 2651A)
CA-567-2
(red)
CS-1629-8
2-Terminal DUT
Plug
CA-560-0
(black)
From
Model 2651A
8011
HIGH CURRENT TEST SOCKET KIT
RATINGS:
40V MAX.
15A DC MAX.
50A PULSE MAX.
HI
CA-552-1
(supplied with
Model 2651A)
Specifications are subject to change without notice.
PA-1032 Rev. A / April 2011
CA-560-2
(red)
8010-120 TO-247
Device Test Board
8010-317 insulator plug
(insert into 3-pin socket)
9
Model 8011 Connection Instructions
Three-terminal DUT using a Model 2651A and a single or dual channel
SourceMeter instrument
NOTE
If you are using a Model 2600A, do not set the output above 40 V. The Model 8011 is rated to 40 V.
You can use the Model 8011 with the 8010-120 board to test a three-terminal device using a Model 2651A High
Power System SourceMeter® Instrument and another SourceMeter instrument. This use-case is appropriate for the
following SourceMeter instruments:
•
Model 2601A Single-Channel System SourceMeter Instrument
•
Model 2611A Single-Channel System SourceMeter Instrument (200 V)
•
Model 2602A Dual-Channel System SourceMeter Instrument
•
Model 2612A Dual-Channel System SourceMeter Instrument (200 V)
Tools and equipment needed
•
Small flat-head screwdriver
•
Cable created using the 2600-KIT screw terminal connector kits (included with the Model 8011); the cable must
use wire that can support 3 A and 200 V
Instructions
1. Remove power from the SourceMeter instruments.
2. Make connections as shown in Table 3 and Figure 7.
3. Install the three-terminal DUT into the socket on the 8010-120 board.
4. Return power to the SourceMeter instruments.
5. On both SourceMeter Instruments:
a. Press the CONFIG key.
b. Press the SRC or MEAS key (the sense mode can be set from either the V-SOURCE or the V-MEAS
menu).
c.
If you pressed the SRC key, select V-SOURCE > SENSE-MODE.
d. If you pressed the MEAS key, select V-MEAS > SENSE-MODE.
e. Select 4-WIRE to enable remote sense.
6. Test the device.
10
PA-1032 Rev. A / April 2011
Model 8011 Connection Instructions
Table 3. Connections for three-terminal device test
Cable end
Cable
Cable end
Model 2651A rear panel
CA-552-1 (supplied with the Model 2651A)
Open
Model 2651A rear panel
CA-557-1 (supplied with the Model 2651A)
CS-1629-8
Series 2600A rear panel
CA-557-x (recommended; separately purchased)
CS-1629-8
LO of CA-552-1
Black CA-560-0
8010-120 yellow terminal 6
HI of CA-552-1
Red CA-560-2
8010-120 white terminal 5
Sense LO of CS-1629-8
Black CA-567-0
8010-120 yellow terminal 3
Sense HI of CS-1629-8
Red CA-567-2
8010-120 white terminal 2
Sense LO of CS-1629-8
Black CA-567-0
8010-120 yellow terminal 3
LO of CS-1629-8
Black CA-567-0
8010-120 yellow terminal 3
Sense HI of CS-1629-8
Red CA-567-2
8010-120 blue terminal 1
HI of CS-1629-8
Red CA-567-2
8010-120 blue terminal 4
Figure 7. Connections for three-terminal device test
CS-1629-8
LO
S LO
G
HI
G
G
G
S HI
Model 2601A, 2611A,
2602A, or 2612A
CA-567-0
(black)
2600-KIT
2600-KIT
CA-567-2
(red)
Customer-supplied cable
CS-1629-8
S LO
G
Model 2651A
CA-557-1
(supplied with
Model 2651A)
CA-567-0
(black)
G
D
S
3-Terminal
DUT
G
G
G
S HI
CA-567-2
(red)
CA-560-0
(black)
8011
HIGH CURRENT TEST SOCKET KIT
RATINGS:
40V MAX.
15A DC MAX.
50A PULSE MAX.
Model 2651A
HI
CA-552-1
(supplied with
Model 2651A)
Specifications are subject to change without notice.
PA-1032 Rev. A / April 2011
CA-560-2
(red)
8010-120 TO-247
Device Test Board
Detail of stacked
banana plugs
at terminal 3
11
Model 8011 Connection Instructions
Three-terminal DUT using a Model 2651A and a low-current SourceMeter
instrument
NOTE
Do not set the Model 2635A or Model 2636A above 40 V. The Model 8011 is rated to 40 V.
You can use the Model 8011 with the 8010-120 board to test a three-terminal device using a Model 2651A High
Power System SourceMeter® Instrument and a low-current SourceMeter instrument. This use case is appropriate
for the following SourceMeter instruments:
•
Model 2635A Single-Channel System SourceMeter Instrument (1 fA, 10 A Pulse)
•
Model 2636A Dual-Channel System SourceMeter Instrument (1 fA, 10 A Pulse)
Tools needed
•
Small flat-head screwdriver
Instructions
NOTE
This use case calls for three 2600-ALG-2 cables. Normally, the Model 2635A ships with one of these
cables and the Model 2636A ships with two. For additional cables, order Keithley Instruments catalog
number 2600-ALG-2. This cable is listed as part number 237-308-2 on Model 2635A and 2636A parts lists.
To set up a test of a three-terminal device:
1. Remove power from the SourceMeter instruments.
2. Make connections as shown in Table 4 and Figure 8.
3. Install the three-terminal DUT into the socket on the 8010-120 board.
4. Return power to the SourceMeter instruments.
5. On both SourceMeter Instruments:
a. Press the CONFIG key.
b. Press the SRC or MEAS key (the sense mode can be set from either the V-SOURCE or the V-MEAS
menu).
c.
If you pressed the SRC key, select V-SOURCE > SENSE-MODE.
d. If you pressed the MEAS key, select V-MEAS > SENSE-MODE.
e. Select 4-WIRE to enable remote sense.
6. Test the device.
12
PA-1032 Rev. A / April 2011
Model 8011 Connection Instructions
Table 4. Connections for three-terminal device test
Cable end
Cable
Cable end
Model 2651A rear panel
CA-552-1 (supplied with the Model 2651A)
Open
Model 2651A rear panel
CA-557-1 (supplied with the Model 2651A)
CS-1629-8
Model 2635A or
Model 2636A rear panel
Sense HI
2600-ALG-2
Red alligator clip to 8010-120 blue
terminal 1
Model 2635A or
Model 2636A rear panel HI
2600-ALG-2
Red alligator clip to 8010-120 blue
terminal 4
Model 2635A or
Model 2636A rear panel
Sense LO
2600-ALG-2
Red and black alligator clips to
8010-120 yellow terminal 3
LO of CA-552-1
Black CA-560-0
8010-120 yellow terminal 6
HI of CA-552-1
Red CA-560-2
8010-120 white terminal 5
Sense LO of CS-1629-8
Black CA-567-0
8010-120 yellow terminal 3
Sense HI of CS-1629-8
Red CA-567-2
8010-120 white terminal 2
Figure 8. Connections for three-terminal device test
GND (not connected)
Model 2635A or Model 2636A
(green)
Sense HI
2600-ALG-2
(237-308-2)
GUARD
GND
(black)
(not connected)
(not connected)
(green)
Hi
Model 2635A or Model 2636A
2600-ALG-2
(237-308-2)
GUARD
(black)
I
eH
d)
Se
(re
2600-ALG-2
(237-308-2)
ns
I
d)
Sense LO
H
Model 2635A or Model 2636A
(not connected)
(green)
(re
GND
(not connected)
Model 2651A
CA-557-1
(supplied with
Model 2651A)
G
D
S
ack
(bl
CA-567-2
(red)
3-Terminal
DUT
G
G
G
S HI
LO
S LO
G
)
CA-567-0
(black)
CS-1629-8
S
s
en
e
LO
(re
d)
CA-560-0
(black)
8011
HIGH CURRENT TEST SOCKET KIT
RATINGS:
40V MAX.
15A DC MAX.
50A PULSE MAX.
Model 2651A
CA-552-1
(supplied with
Model 2651A)
HI
Specifications are subject to change without notice.
PA-1032 Rev. A / April 2011
CA-560-2
(red)
8010-120 TO-247
Device Test Board
13
Model 8011 Connection Instructions
NOTE
Alligator clips are connected the screw posts on the underside of the board, as shown here.
Se
ns
e
HI
(re
d)
HI
se
Sen
LO
)
(red
LO
(red)
(black)
Contact information
If you have any questions after reviewing this information, please contact your local Keithley Instruments
representative or call Keithley Instruments corporate headquarters (toll-free inside the U.S. and Canada only) at
1-888-KEITHLEY (1-888-534-8453), or from outside the U.S. at +1-440-248-0400. For worldwide contact numbers,
visit the Keithley Instruments website (http://www.keithley.com).
Safety precautions
WARNING
The 8011 High Current Test Socket Kit is designed to work exclusively with the 2651A High
Power System SourceMeter® Instrument and should never be connected to voltage sources
that may exceed 40 V DC. The socket should be used within a comprehensive test system (for
example, a protective enclosure or safety barrier) that provides adequate safety measures to
prevent hazards such as accessible voltage, fire and explosion from causing harm.
The following safety precautions should be observed before using this product and any associated instrumentation. Although
some instruments and accessories would normally be used with nonhazardous voltages, there are situations where hazardous
conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions
required to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using
the product. Refer to the user documentation for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product warranty may be impaired.
The types of product users are:
Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring that the
equipment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety procedures and proper use of the
instrument. They must be protected from electric shock and contact with hazardous live circuits.
14
PA-1032 Rev. A / April 2011
Model 8011 Connection Instructions
Maintenance personnel perform routine procedures on the product to keep it operating properly, for example, setting the line
voltage or replacing consumable materials. Maintenance procedures are described in the user documentation. The procedures
explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, perform safe installations, and repair products. Only properly trained
service personnel may perform installation and service procedures.
Keithley Instruments products are designed for use with electrical signals that are rated Measurement Category I and
Measurement Category II, as described in the International Electrotechnical Commission (IEC) Standard IEC 60664. Most
measurement, control, and data I/O signals are Measurement Category I and must not be directly connected to mains voltage or
to voltage sources with high transient over-voltages. Measurement Category II connections require protection for high transient
over-voltages often associated with local AC mains connections. Assume all measurement, control, and data I/O connections
are for connection to Category I sources unless otherwise marked or described in the user documentation.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test
fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than
30 V RMS, 42.4 V peak, or 60 VDC are present. A good safety practice is to expect that hazardous voltage is present in any
unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators
are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential
human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock.
If the circuit is capable of operating at or above 1000V, no conductive part of the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance-limited
sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective
devices to limit fault current and voltage to the card.
Before operating an instrument, ensure that the line cord is connected to a properly-grounded power receptacle. Inspect the
connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input
power disconnect device must be provided in close proximity to the equipment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under
test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting
cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth)
ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the
voltage being measured.
The instrument and accessories must be used in accordance with its specifications and operating instructions, or the safety of
the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating
information, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with the same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use
of a lid interlock.
If a
screw is present, connect it to safety earth ground using the wire recommended in the user documentation.
The ! symbol on an instrument means caution, risk of danger. The user should refer to the operating instructions located in
the user documentation in all cases where the symbol is marked on the instrument.
The
symbol on an instrument means caution, risk of electric shock. Use standard safety precautions to avoid personal
contact with these voltages.
The
The
symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns.
symbol indicates a connection terminal to the equipment frame.
Specifications are subject to change without notice.
PA-1032 Rev. A / April 2011
15
Model 8011 Connection Instructions
If this Hg symbol is on a product, it indicates that mercury is present in the display lamp. Please note that the lamp must be
properly disposed of according to federal, state, and local laws.
The WARNING heading in the user documentation explains dangers that might result in personal injury or death. Always read
the associated information very carefully before performing the indicated procedure.
The CAUTION heading in the user documentation explains hazards that could damage the instrument. Such damage may
invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protection from electric shock and fire, replacement components in mains circuits — including the power
transformer, test leads, and input jacks — must be purchased from Keithley Instruments. Standard fuses with applicable national
safety approvals may be used if the rating and type are the same. Other components that are not safety-related may be
purchased from other suppliers as long as they are equivalent to the original component (note that selected parts should be
purchased only through Keithley Instruments to maintain accuracy and functionality of the product). If you are unsure about the
applicability of a replacement component, call a Keithley Instruments office for information.
To clean an instrument, use a damp cloth or mild, water-based cleaner. Clean the exterior of the instrument only. Do not apply
cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with
no case or chassis (e.g., a data acquisition board for installation into a computer) should never require cleaning if handled
according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the
factory for proper cleaning/servicing.
16
PA-1032 Rev. A / April 2011
Was this manual useful for you? yes no