INSTRUCTION MANUAL
Digital Platform Scales
FG-150KAL — FG-150KAM — FG-150KBM
FG-60KAL ~~ FG-60KAM FG-60KBM
FG-30KAM - FG-30KBM
A&D Company, Limited
WM+PD4000966 - |
This manual and Marks
All safety messages are identified by the following, “WARNING” or “CAUTION”, of
‚| ANSI 7535.4 (American National Standard Institute: Product Safety Signs and
Labels). The meanings are as follows: |
A potentially hazardous situation which, if not avoided,
WARNING could result in death or serious injury.
| A potentially hazardous situation which, if not avoided,
/AN CAUTION may result in minor or moderate injury.
This is a hazard alert mark.
PS This mark informs you about the operation of the product.
ou This manual is subject to change without notice at any time to improve the
product. -
o Product specifications are subject to change without any obligation on the part
of the manufacturer. |
© 2004 A&D Company, Limited All rights reserved.
Contents
1. INTRODUCTION ............. ee... eenneveereeoneniere een eres arenero aerea reee errar nereeRecaa 2
2, UNPACKING .…........……ucrerirerersrastersensrrscereremeercenarenrenentensrencsnen en er SEVEN cena 2
3. NAMES AND FUNCTIONS .................e.eeerizcceccenererceveseucereecorareeoreoreecenererer recae. 3
4. SETTING UP ........ TEN 6
4-1. Attaching a display pod to the base (FG-KAL and FG-KAM)............——..... 6
4-2. Installing the scale ................eee een TÉR! Keen er 6
4-3. Power [email protected] rec aerena an ets terre nca ere cran encre ce 7
5. BASIC OPERATION .................e.mmnec ins eres nessa 8
5-1. Turning the power ON and OFF .............. нее. NT er ne00s 8
5-2, Selecting a weighing UNit..................... ==... ere re ee ee 8
5-3. Basic operation........ nemssnanenencereneerarersanams ana a ae mec ce see cncrreranren ane case c 00 nana rec 8
5-4, Weight display resolution............ EN EEE EEE 9
6. COUNTING MODE... ee, 10
7. COMPARATOR ................... vee enerevereens A 11
8. CALIBRATION ...............ecocrecconeececnncoreonccorerveneeenecenecee e REP 12
8-1. Calibration using a weight .....................e... RE EE 12
8-2. Gravity acceleration Correction ..............—— rem... .—. e... ervenonenos 13
9. FUNCTIONS..............ece.eecceccceneeocconcareccararerenneeneeteanee terreno неее vee 14
— 9-1. The procedure for setting parameters .............. ===... ea. 14
9-2. Function ISE ....................e..e.e.erereenenee rene sera centrer enverra rer enrra crc seance rence caen 15
10. OPTIONS................... A EG EEE nenes „ен 16
10-1. Installation of OP-23/0P-24.................. Lnrrooecrenreere ner ene PerrAeeaCUCNNa Lesssererees 16
10-2. OP-23 RS-232C serial interface ‚u.a 17
10-3. OP-24 RS-232C serial interface and Comparator relay output.............. 20
11. MAINTENANCE... rrr reser e seers see e eee ae aes 21
11-1. Notes on maintenance ......................... eee 21
11-2 Error Codes ............e.......e0merneeeeccrer ere eee. bs 21
12. SPECIFICATIONS .....................0iemeemn enn e PT 22
12-1. Specifications rcerononeonnonanenceanea ren ncanraererneeeOENreNENOFOFOrONE FONO ененененненененннне 22
— 12-2. Dimensions.............. e... reeves PO 23
GRAVITY ACCELERATION MAP ...................... TZ
1
This manual describes how this scale works and how to get the most out of it
in terms of performance.
FG series platform scales have the following features:
O The FG series has three kinds of weight display resolution, 1/3,000, 1/6,000
(1/7,500) and 1/12,000 (1/15,000).
QO There are 2 sizes of weighing pan. The FG-KAL has a larger pan and the FG-
KAM/KBM has a smaller pan. The FG-KAL/KAM has a display column and the FG-
KBM is a without column model. You can select a model that suits your own
application. |
О Аз power source, you can use an AC adapter or C size dry batteries.
O The counting function easily counts the number of articles of the same weight.
O The comparator function compares the display value with upper limit and lower limit.
The display shows the result and the optional FG-24 can output it as a relay signal.
Q The optional FG-23 and FG-24 has an RS-232C interface and can output the
weighing data to connect with a printer, computer and so on.
2. UNPACKING
When unpacking, check whether all of the following items are included:
“ FG30KAM/60KAM/150KAM
FG60KAL/150KAL
Column support foot
(FG-KAM only)
3
a ES
Protective cover is attached.
(All models)
Scale (Shape is different by model.)
Instruction manual
INSTRUCTION MANUAL
AC adapter
Ÿ
Please confirm that the AC adapter
type Is correct for your local ED
voltage and receptacie type. |
HI OK LO indicator DO
Ç a —
3. \
Display pod
Display
column
CAL switch
cover
AC adapter jack Cable cover
Weighing pan Column (FG-KAL/KAM)
| Support foot
(FG-KAM only)
of
pie
AE р
ri -
В а
аи :
y
и ase
и
я,
A Я
р | eveling foot
FG-KAL/KAM series
Metric models
= LN cs | |
LCD display e — =
STABLE indicator во
NET indicator || |=
kg | Weighing unit
ZERO indicator 17 ——% = |
В
+
_
ITY
[+4 )
a)
ON/OFF switch / |
Ne / | \ —
PRINT switch / RE-ZERO switch \
MODE switch /
3
U.S.A. models
7 D
|
MARA eng
ZERO
ON/OFF switch 4 7 7 | J
7 7 — И
PRINT switch/ — / ZERO switch] \ TARE switc
MODE switch
Indicators
STABLE O Indicates when the reading is stable.
МЕТ < Indicates when NET weight is displayed. (Tare function is used.)
ZERO | < Indicates when the scale zero is correct.
FOKELO Indicates when the scale zero is correct.
Weighing U units “kg” and “pcs” for metric models |
“ID”, “oz”, “kg” and “pcs” for U.S.A. models
Switches
_ON/OFF Switch
oD ON/OFF] Used to turn the power on or off. When turned on, the
scale will be automatically set to zero (power-on zero).
_PRINT Switch
Outputs the weight value to printer. In the setting mode,
this switch is used to increment the value of the selected
digit blinking.
_MODE Switch. |
MODE (ноге Switches the weighing unit. In the setting mode, this
switch is used to store a parameter and go to the next. |
| wd
_RE-ZERO Switch
Clears the display to zero. In the setting mode, this
switch is used to select a digit blinking to change its
value.
_ZERO Switch_ U.S.A. model
ZERO Zeroes the scale and turns the display zero. In the
> setting mode, this switch is used to select a digit blinking
to change its value. -
_TARE Switch U.S.A. model
Subtracts tare (container) weight on the weighing pan.
O The | RE-ZERO |, [ ZERO | and | TARE | switches work when the weight value is
stable.
O The | RE-ZERO | and | ZERO | switches will zero the scale if the weight is within
+2% of the weighing capacity (kg) around the power-on zero point. The ZERO
indicator 7 turns on.
0 If the weight exceeds +2% of the weighing capacity (kg), the | RE-ZERO | switch
Will tare the scale. In this case the ZERO and NET indicators turn on.
O The | TARE | switch will tare the scale when the weight is plus value. In this case
the ZERO and NET indicators turn on.
The ZERO operation clears tare operation previously done and NET indicator turns
off.
4-1. Attaching a display pod to the base (FG-KAL and FG-KAM)
1. First, remove the 4 screws from the bottom of the display column, _
a и” „ Screws
2. Set the display column to the base by pulling the cable | y e ro
1 e J
into the base.
[J Take care not pinch the cable between the
column and the base.
3. Tighten the 4 screws removed at step 1 to fix
the display column firmly.
4. Tilt the display pod forward by pressing in on
the two round side clamps, and slide off the
cable cover.
5. Pull the bundle of cable out from the top end of
display column, feeding all of slack of the
cable on the base into the display column.
{J Take care not scraich the cable.
pes : RRs
eu aL EE E
APA
Neither sla
Nor tensior
FEN
6. Put the bundle of cable back into the display column.
7. Make sure the cable is fitted to the 2 cable
guides and attach the cable cover.
Cable guide
8. Place the weighing pan on the base.
4-2. Installing the scale CC |
1. Select the place for installing the scale. Refer to “Cautions for installing the scale”
below.
2. Adjust the level of the base, using the spirit level and leveling feet. The FG-KAM has
an extra foot under the display column. Adjust this foot to reach floor after adjusting
the level of the base.
3. Tilt the display pod by pressing in on the two round side clamps.
Cautions for installing the scale
Consider the following conditions to get the most out of your scale.
J Install the scale where the temperature and relative humidity is stable, There is no
draft and a stable power source is available,
O Install the scale on a solid and level surface.
l Do not install the scale in direct sunlight.
0 Do not install the scale near heaters or air conditioners. В
QU Do not install the scale where there is flammable or corrosive gas present.
Q Do not install the scale near equipment which produces magnetic fields.
1 Do not install the scale where there is apt to be static electricity, in a place where
the relative humidity is lower than 45% RH. Plastic and isolators are apt to be
charged with static electricity. В | В
НВ Ро по! изе an unstable power source. ВЕ В О
НЫ When the scale is installed for the first time, or the scale has been moved, carry out
calibration as described in “8. CALIBRATION".
4-3. Power source
For the power source, the AC adapter or C size dry cells can be used.
When using the AC adapter
Use a stable power source. To use the AC adapter, insert the AC adapter plug into the
AC adapter jack on the rear side of display pod.
LI Confirm the AC adapter type is correct for your local voltage.
* When using the batteries
Prepare 4 x C size (R14P/LR14) dry batteries. The batteries are not included in the
product. The scale can be used continuously for about 150 hours using the alkaline
batteries.
2, Slide the battery cover off = В |
3. Push the battery case inside the display pod and take it out.
4. Insert four new dry cells into the battery case.
5. Push the battery case into the display pod as before.
6. Attach the battery cover. | |
1. Turn of the scale and disconnect the AC adapter if used.
Li Take great care of the polarity of baiteries. The polarity marks are
shown in the battery case.
LJ Replace used batteries with four new ones when “L 50” is displayed.
{1 Do not mix used and new batteries. It may cause damage to the
battery or product, if used.
LJ Do not mix the batiery type. It may cause damage to the battery or
product, | a |
Q The battery life depends on the ambient temperature,
(J Remove batteries from display pod when the scale is not to be used
for a long time. They may leak and cause damage. a
{J Damage due to battery leakage is not covered by the warranty.
>
6. Remove the object from the weighing pan.
O The switch -will zero the scale if the weight is within +2% of the
— weighing capacity (kg) around the power-on zero point. The ZERO indicator ? turns
on. When the weight exceeds +2% of the weighing capacity (kg), it will be
subtracted to zero as a tare weight. In this case the ZERO and NET indicators turn
on.
0 The | ZERO | switch will zero the scale if the weight is within +2% of the weighing
capacity (kg) around the power-on zero point. The ZERO indicator ? turns on.
When the weight exceeds +2% of the weighing capacity (kg), the switch does not
work.
0 The | TARE | switch will subtract the weight to zero as a tare weight when the
weight is a plus value. |
Precautions during operation
OU Make sure that the STABLE indicator is on whenever reading or storing a
value.
O Do not press switches with a sharp implement such € as a pencil.
НВ Ро not apply a shock load to the scale.
0 Do not place a load onto the pan that exceeds the capacity.
[1 Keep the scale free from foreign objects such as dust or liquid.
[1 Calibrate the scale periodically to keep weighing accuracy.
(See “8. CALIBRATION”.)
5-4. Weight display resolution
The FG series has three kinds of weight display resolution, NORMAL, HIGH and
HIGHER. The following is about “kg” and “Ib” display for reference. See the "2.
SPECIFICATIONS” in detail.
NORMAL: 1/3,000
HIGH: 1/6,000 ~ 1/7,500 (depending on capacity)
HIGHER: 1/12,000 ~ 1/15,000 (depending on capacity)
The factory setting is the HIGH resolution, but it is possible to change by Function
settings “F2”. Set this function according to the application.
O In some countries, the resolution will be limited NORMAL only, and the setting F2 is
not available.
In the COUNTING mode, the scale works as in the HIGHER resolution regardless
“of weight display resolution.
Determines a unit weight (the weight of one piece) from sample pieces, and calculates
how many pieces on the weighing pan using the unit weight.
1. Press the | MODE | switch to select “pes”.
(“pes” = pieces)
2. Press and hold the | MODE | switch to enter
the sample unit weight storing mode.
3. To select the number of samples, press the
PRINT | switch. It may be set to 5, 10, 20, 50
or 100.
. If necessary, place a container on the weighing
pan, and press the [ RE-ZERO | ( | TARE |)
switch. Confirm that the right side of the number
of samples shows zero.
, Place the correct number of samples on the pan
or in the container.
; Press the MODE | switch to calculate and
store the unit weight. Remove the sample. The
scale is set to count objects with this unit weight.
fo OX
© A
ZERO? Um)
и NM
© con
ZERO? + U msj
© = I -
ZERO CL nes
Confirm the display
>
9 3 À
“zero? EL U ms
|
4
© a
CL
The total weight of sample pieces should be more than below regardless
of number of sample pieces.
FG-30K: 259
FG-60K: 62.59
FG-150K: 1259
If not, the display shows “Lo ut” and returns to the previous display.
Increase the number of samples (go to step 3) and try again.
О if pressing the | MODE | switch without adding sample pieces or adding
enough weight to be acceptable as a unit weight, the display leaves
counting mode and switches to the next weight unit.
- 7. Place the objects to be counted on the pan.
BCS
0 Unit weigh is maintained even if the scale is powered off.
10
в
|
7. COMPARATOR
The results of the comparison are indicated by HI, OK or LO on the display.
The formula to compare is as follows:
LO < Lower limit value = OK = Upper limit value < HI
Operating conditions (see the “Function list F6”):
— F6-0: No comparison (comparator function disabled).
F6-1: To compare all data.
F6-2: To compare more than +4d or less than Ad.
F6-3: To compare all stable data.
F6-4: To compare stable data more than +4d or less than -4d.
F6-5: To compare stable data more than +4d.
d = minimum weight display (see “12-1 Specifications”)
In case of counting mode, “d” is equal to minimum weight display of kg mode.
J The upper limit and lower limit numerical values are common to each of the
weighing and counting mode.
0 Ignore the decimal point of setting value to apply it to each mode.
Example of FG30K / setting value is “001000”.
Display mode Limit value Display capacity
NORMAL resolution kg 10.00 kg 30.00 kg x 0.01 kg
HIGH resolution kg 1.000 kg 30.000 kg x 0.005 kg
HIGHER resolution kg 1.000 kg 30.000 kg x 0.002 kg
NORMAL resolution Ib 10.00 Ib - 60.00 kg x 0.02 lb
HIGH resolution Ib 10.00 Ib 60.00 kg x 0.01 Ib
HIGHER resolution Ib 1.000 Ib 60.000 kg x 0.005 Ib
Counting mode 1000 pcs
Entering the upper and lower limit values
1. Press the | MODE | switch to select weighing Г | nm A )
unit "kg", “Ih” or oz”
2, Press and hold the | MODE | switch to enter the
o \ |
upper limit setting mode. | Er "A in
3. Enter an upper limit value using the following switches. JER J
RE-ZERO | or | ZERO | To select the digit blinking to change.
PRINT | Increment the value of the selected digit.
The minus sign can be set at the next digit of the least | 7 N
significant digit. The | PRINT | switch alternates the я Г En | | ПО | |
minus sign on and off. The blinking “” shows minus E _
and no sign is plus.
4, After setting all of digits, press the | MODE | switch. Then, the upper limit is stored
and the display goes to the lower setting mode.
5. Set the lower limit in a similar way, and press the | MODE | switch to return to
weighing mode.
UI The upper and lower limits are > maintained even i the scale is powered off.
11
8-1. Calibration using a weight
This function adjusts the scale for accurate weighing.
Calibrate the scale in the following cases.
O When the scale is first used.
О When the scale has bee moved.
О When the ambient environment has
changed.
Q For regular calibration.
Connecto
Lock screv
Loose the lock screws on the rear side of Calibration (CAL) switch
the display pod, and remove the CAL
switch cover. Then, there is a calibration
switch on the board inside.
O For the FG-KB series (without column), you may once remove the weighing pan to
access lock screws easily.
CAL switch cover
Q Do not use a ballpoint pen and so on to press the calibration switch.
That may short-circuit and damage the scale.
1. Warm up the scale for at least half an hour with nothing on the weighing pan.
O Change Function setting "F1" or place something on the pan to disable the autc |
power-off function. |
5 Press and hold the calibration (CAL) switch until > pg \
| ro A -
L Li |
Ш
a
appears, and release the switch.
J The weighing unit must be “kg” or “Ib” to enter |
calibration mode. " uu
3. Make sure that there is nothing on the weighing pan,
and wait until the STABLE indicator turns on.
4. Press the MODE | switch. The scale calibrate the - U us
zero point, and the display shows, “Sfn |” and the Yann
weight value to calibrate (SPAN calibration). ) ELLE
Li The weight value is equal to the capacity. When
you enter with “kg” mode, then the value is “kg”.
Entering with “Ib”, then “Ib”. |
О If you do not need SPAN calibration, turn the power off to exit from the calibratio! -
procedure.
5. To calibrate with the different weight, change the displayed value using the followin:
switches. | | |
RE-ZERO | or | ZERO To select the digit blinking 10 change.
PRINT| Increment the value of the selected digit.
12
| <=
a
6. Place the calibration weight on the pan with the same (O J ; Lo An
value as displayed, and wait until the STABLE С |
indicator turns on. \
7. Press the | MODE | switch. The scale calibrate SPAN Fo
and | End | will appear. E Г С
Remove the weight from the pan, and turns the power
off.
LI Note | | | |
The value set in step 5 is cleared after the power is switched off.
If the scale will suppose to move to another location, set the gravity
acceleration value for the current location and calibrate the scale
according to the procedure above. See the next section to set the
value, ВЕ
8-2. Gravity acceleration correction
When the scale is first used or has been moved to different place, it should be
calibrated using a calibration weight. |
But if a calibration weight is not available, the gravity acceleration correction will
compensate the scale. Change the gravity acceleration value of the scale to the value
of area where the scale will be used. Refer to the gravity acceleration map appended
to the end of this manual. |
Li Note | |
It is not necessary to set the gravity acceleration correction when
calibrating the scale with a calibration weight at the place where it is
to be used. | - a
appears, and release the switch.
J The weighing unit must be “kg” or “Ib” to enter
calibration mode. |
2, Press the | PRINT | switch. | A
The display shows the gravity acceleration value
memorized in the scale.
3. To change the displayed value using the following switches.
RE-ZERO | or | ZERO | To select the digit blinking to change.
PRINT Increment the value of the selected digit.
4. Press the [MODE] switch. The display returns to [ £AL E | |
5. If necessary to calibrate the scale using a calibration weight, go to step 3 of "8-1.
Calibration using a weight”. To finish the setting, turn the power off.
1. Press and hold the calibration (CAL) switch until ©
FE A |
Lo
13
The scale has Function settings to expand your applications.
The parameters set in the Function settings are maintained even if the power switcher
off.
C
|
ho
9-1. The procedure for setting parameters
aire
1. Turn the power off.
à. Press and hold the | ZERO | switch and turn the
power on via the | ON/OFF | switch. Then, the first | E | -
function item and its parameter is displayed |
3, Set the parameter value using the | PRINT | switch. -
CI If you do not need to change the parameter, go to next
step without setting.
4. Press the | MODE | switch. Then the display goes to |
УЕ
- — +
Г
“=
|
МОР
——
TINT у >
J
I
|
|
МЕ
PAN
N—
the next function item.
{I In this stage, the new parameter is not stored in
fhe scale yet.
El To end changing the parameters, turn the power
off.
5. Repeat the steps 3 and 4 to the last item.
6. After setting the last item, press the | MODE switch, |
Then, will appear. ro
7. Press the | MODE | switch again. The parameters are
stored in the scale, and the scale will automatically boot
up.
CI If you turn the power off before this step is done,
none of the parameters will be changed,
14
9-2. Function list
ltem Description
Auto power-off Fu Auto power-off disabled Automatically
function e E | Auto power-off enabled for battery power off
use only
Eo Auto power-off enabled for battery
and AC adapter |
Display F 2-10 Normal (1/3,000 class) Legal for trade|
resolution + F 2 | High (1/6,000~1/7,500 class) will be Normal,
Fe-¢ Higher (1/12,000-1/15,000 class)
Weighing unit |* £ 9-0 Ib . | U.S.A. models
when powered | F 3- | oz | | only
on | F 3-2 |Кс
RS-232C e FY 0 2400 bps
Baud rate F 47 4800 bps.
F 4-2 19600 bps
RS-232C + F 5-10 Stream mode
Data output F 6- € Command mode
mode Е 5- © Print switch mode
Е 5- 3 Auto-print mode +/- data
F 5-4 Auto-print mode + data
Comparator + FE6-Ú Comparator disabled Conditions to
mode £ 6-1 Compares all data compare.
| Е 6- e Compares data > +4d or < -4d d = minimum
F 6-3 Compares all stable data display division
| F 6- 4 Compares stable data > +4d or < 4d
Filtering to F 1-0 Weak / fast response
weighing data | * F 1- | Normal / normal response
| Fe Strong / slow response
RS-232C e FB- 0 Reply is sent Reaction to the
Data format E B- 1 |Noreply except “Q” command command
+ Factory setting
15
The following options are available for the FG series:
0 OP-23 (FG-23) RS-232C serial interface
О ОР-24 (ЕС-24) RS-232C serial interface and Comparator relay output
[J OP-23 and OP-24 cannot be used together. |
40-1. installation of OP-23/0P-24
The OP-23/0P-24 has an interface board, an option panel and a DIN 8 pin connector.
The option panel and DIN connector are common to both options.
Before installation, prepare an interface cable using attached DIN connector. Or the
is a way to use the optional RS-232C cable (see “10-2. OP-23 RS-232C serie
interface”).
1. Disconnect the AC adapter from the scale. |
> Remove the four screws and the cover of option slot.
3. Thread the interface cable through the hole of option panel first, and connect tr
DIN connector to the interface board. | |
4. Connect the interface board to the connector in the display pod.
5. Attach and fix the option panel using the screws that removed in the step 2.
A
Option slot for OP-23/24 / Interface board
(Cover)
Option panel
Interface cable
16
10-2. OP-23 RS-232C serial interface
— This interface allows FG series to be connected with a multi-function printer or a
personal computer. |
[1 The RS-232C interface has the following four modes.
stream mode Outputs data continuously.
Command mode Controls the scale using commands from a computer.
Print switch mode Outputs data by pressing the switch..
Auto-print mode Outputs data which meets the conditions of auto-print.
LJ If necessary, setthe parameter of the data format and data output mode (F4, F5
and F8). |
LA DIN 8 pin connector (JA+TCP0586) is provided with the OP-23 for wiring.
re 0 There are optional cables to connect with a personal computer.
al AX-KO577A-200 FG to D-Sub 25 pin computer / RS-232C cable, 2m
AX-KO1786-200 FG to D-Sub 9 pin computer / RS-232C cable, 2m
AX-KO577A-200 AX-KO1 786-200
"a | DIN 7P D-Sub 25P DIN 7P D-Sub 9S
| 1 1 1 1
3 3 314
3 3 3 3
4 T4 4 4
5 - 5 5 5
в 8 SS
E 7 7 7 7
others | | 8
9
(DIN 7 pin plug P can connect with DIN 8 pin socket.)
Interface specifications
Transmission system EIA RS-232C
Transmission form Asynchronous, bi-directional, half-duplex
Data format Baud rate: 2400, 4800, 9600 bps
Data: 7 bits + parity 1bit (even) | | |
Start bit: 1 bit
Stop bit: 1 bit | | |
Code: ACH a |
Terminator: CrLr (Cg: ODh, Lr: OAR) |
| LSB MSB 1 (-15V~-5V)
| 101112 |3 | 4 | 5 |6 0 (5V-15V)
Ne —— A |
| | Stop bit
Parity bit
| Data bit
Start bit
17
Pin connections
View from outside ‚,
RXD
TXD
DSR
Mating connector:
DIN 8 pin (JA+TCP058)
Attached to FG-23.
NC
NC
NC
NC
: (Inside of OP-23)
— Data format
S|ITI, |+ ololOlOla|, TO|O| |kK|ag|Ca!Lr
Y / \ IN A
Header Data | Unit Terminator
Separator (“_” shows a space.) ;
O There are 4 types of headers:
| ST: Stable weighing data
QT : Stable counting data
US : Unstable weighing data (including counting data)
OL : Out of weighing range (Over)
O The data is normally 9 digits including decimal point and a sign.
0 There are 4 types of units:
_kg :Weighing data “gram”
_PC : Counting data “pes”
_ Ib :Weighing data “decimal pound”
_ oz Weighing data “decimal ounce”
O The terminator is always Calc.
O Example of output data: |
Weighing data “kg” st]. [+]oJo[1]2[3] . [4]5] |k]g Cal!
Counting data [e]T[.|+10[0[0[1[2]8]|4]5|_|P|C|Cail
Out of range “kg” (+) o], |+]Ja]9]a] aja]. Tala). IK|gíCall
Out of range “pcs” (-) ofc]. T-Telee]o]ofele]e] |P|C|CAll |
18
Data output mode
0 Stream mode — Function setting “F5-0”
The scale outputs the current display data. The data-update rate is approximately 10
times per second. This rate is the same as the display-update.
The scale does not output data while it is in setting mode.
Q Print switch mode Function setting “F5-2”
When the | PRINT | switch is pressed while the weighing data is stable (STABLE
indicator is on), the scale transmits the data.
D Auto-print mode +/- data Function setting “F5-3"
The scale transmits the weighing data when the display is stable (STABLE indicator
is on) and the data is more than +4d or less than -4d of weight data.
d = minimum weight display (see “12-1 Specifications”)
When in counting mode, “d” is equal to minimum weight display of kg mode.
The next output can be obtained after the display returns to between -4d and +4d.
— H Auto-print mode + data Function setting “F5-4"
The scale transmits the weighing data when the display is stable (STABLE indicator
is on) and the data is more than +4d of weight data.
d = minimum weight display (see “12-1 Specifications”)
When in counting mode, “d” is equal to minimum weight display of kg mode.
The next output can be obtained after the display returns to below +4d.
O Command mode — Function setting “F5-1
In the command mode, the scale is controlled by commands that come from the
personal computer and so on.
Command list
“Q” command Command to request the current weighing data.
Command
Reply [87| |+ |0|0|1|2|3]. 14 |5 |_ [к] а [ба]!
“Z” command Same operation as the | RE-ZERO ] or | ZERO | switch,
Command
Q This command works as | RE-ZERO | for the metric models
and as | ZERO | for U.S.A. models.
“T command Same operation as the | TARE | switch.
Command
This command works as | TARE | for U.S.A. models.
UI The metric models cannot accept this command.
19
Reply to the command
When the “F8-0" is selected, the scale reacts io the received command as follows.
Q For the “Q” command, the scale will send the data.
QO For the “7” and “T” commands, the scale will send the same code as a reply aft
executing the command.
Reply
Reply
When the command cannot execute because the scale is unstable, for examp!
“I” will be sent.
Reply
Q If the received command is not for the FG series, the scale will send *?”. The"
command for the metric model is included to this group.
Reply o
Q When “F8-1” is selected, there is no reply except the "Q” command.
10-3. OP-24 RS-232C serial interface and Comparator relay outpu
The OP-24 has an RS-232C series interface and relay output for the compara
function. It allows output of the HI, OK or LO signal results to an external device aí
solid state relay output. |
Ш The specification for the RS-232C interface is same as the OP-23 (FG-23). €
See "10-2. OP-23 RS-232C serial interface” for further information.
Interface specifications
Pin connections
View from outside |
RXD
TXD
DSR
Mating connector:
DIN 8 pin (JA+TCP058)
Attached to FG-24.
COM
OK |
HI Ш
LO |
: (Inside of OP-24)
Maximum rating for relay is as follows.
Q Maximum voltage: 50VDC
O Maximum current: 100mA DC
O Maximum ON resistance: 357 a | |
20
dy
e
2
= e TES 3%, = Fei
E В a el B H
He E TEE = *- FE
A EE CR 5
My = A >
= o В kd ae
11-1. Notes on maintenance
LJ Do not disassemble the scale. Contact your local A&D dealer if your scale
needs service or repair.
[J Please use the original packaging for transportation.
CO] Do not use organic solvenis to clean the scale. Use a warm lint free cloth
dampened with a mild detergent.
11-2 Error codes
Overload error | | |
| - ) Warning to indicate that an object beyond the scale
capacity has been placed on the pan. Remove the object
from the pan.
Range over notice | | В
| | — This will be shown if the weight sensor receives strong
=)
-f
force upward. Check if the weighing pan is touching
anything or if there is anything in the base. There is a
possibility that the weight sensor itself may have a failure.
The sample weight is too light to set the unit weight in the
counting mode. Increase the sample numbers.
Low battery
Warning to show that the batteries are exhausted.
VW VL
| | H | Replace them with new batteries.
Low power
| Warning to show that the voltage of main power source is
Г Г | too low.
parameters when the calibration, function setting, unit
weight registration, comparator limits setting and so on
have been done. Turn the power off once and try the
above procedure again. If this error happens again, there
is a defect in the memory device.
| г - | This may be shown that the scale fails to store
If you cannot cancel an error or other errors occur, request service from the store
where you purchased the product or to your A&D dealer.
21
12.
12-1. Specifications
SPECIFIC?
Display update
10 times per second
Operating temp.
10%C-40%C / 14°F~104°F, less than 85% R.H. (non-condensing)
*) Factory setting
22
FG-S0KAM | FG-60KAM | FG-150KAM
MODEL FG.30KBM | FG-60KBM_| FG-150KBM FG-60KAL |. FG-150K
a capacity 30 60 150 | 60 150
mr 0.01 0.02 0.05 0.02 0.05
ke) display 0.005 * 0.01 * 0.02 * 0.01 * 0.02 *
9 0.002 0.005 0.01 0.005 0.01
ый capacity 60 150 300 150 300
0.02 0.05 0.1 0.05 0.1
(e display 0.01 * 0.02 * 0.05 * 0.02% 0.05”
0.005 0.01 0.02 0.01 0.02
ho capacity 960 2400 4800 2400 © 4800
_ 0.5 1 2 1 2
(oz) display 0,2 * 0.5 * E 0,5 E
0.1 0.2 0.5 0.2 0.5
No. of samples 5 (can be changed to 10, 20, 50 or 100) pieces
Max. counts 120,000 pcs | 96,000 pcs 120,000 pcs | 96,000 pcs 120,000
Min. unit weight 0.25 9 0.625 д 1.25 9 0.625 9 1.25%
Repeatability
(Std. deviation) 0.005 kg 0.01 kg 0.02 kg 0.01 kg 0.02 |
Linearity error. +0.01 kg +0,02 kg +0.05 kg 0.02 kg +0.05
Sensitivity drift +20 ppm / °C (10°C-30°C / 50°F~86°F)
Display 7 segment LCD display (Character height 26 mm)
Power supply AC adapter or C size (R14P / LR14) x 4 batteries
Battery operating Approximately 150 hours with alkaline dry cell battery
Welgna pan 300 х 380 mm / 11.8 x 15.0 in. x on
FG-KAM: 300(W) x 624(D) x 781(H) mm
Dimension 11.8(W) x 24.6(D) x 30.7(H) in. 390(W) x 774 (D) x 781(F
EFG-KBM: 380(W) x 464(D) x 118(H) mm 15.4(W) x30.4(D) x 30.7
- 15.0(W) x 18.3(D) x 4.6(H) in. I
Weight FG-KAM: 11.2 kg 16.4 k
(approximately) FG-KBM: 9.7 kg A. —
Calibration weight | 30 kg 60 kg 150 kg 60 kg 150
(Factory setting) 60 Ib 150 Ib 300 Ib 150 Ib 300
veo
me
m
12-2, Dimensions
FG-60KAL
FG-150KAL
- FG-30KAM
FG-60KAM
FG-150KAM
FG-30KBM
FG-60KBM
FG-150KBM
78]
со
|
530 390
a N
| NL | =
289 | 440 300
380 300
ВЕ 2 y
ED que
82| 202 300 | 220
380 145 300
300 220
mm
23
Values of gravity at various locations
LATITUDE
90°
80°
70°
60°
50°
40°
30°
20°
10°
o”
Amsterdam 9.813 m/s* Manila 9.784 m/s”
Athens 9.807 m/s“ Melbourne 9.800 m/s“
Auckland NZ 9.799 m/s* Mexico City 9.779 m/s*
Bangkok 9.783 m/s* Milan 9.806 m/s“
Birmingham 9,813 m/s“ New York 9.802 m/s“
Brussels 9.811 m/s* Oslo 9.819 m/s”
Buenos Aires 9.797 m/s* Ottawa 9.806 m/s“
Calcutta 9,788 m/s“ Paris 9.809 m/s“
Cape Town 9.796 m/s” Rio de Janeiro 9.788 m/s”
Chicago 9.803 m/s Rome 9.803 m/s
Copenhagen 9.815 m/s“ San Francisco 9.800 m/s“
Cyprus 9.797 m/s? Singapore 9.781 m/s”
Djakarta 9.781 m/s Stockholm 9.818 m/s”
Frankfurt 9.810 m/s Sydney 9.797 m/s
Glasgow 9.816 m/s* Taichung 9.789 m/s?
Havana 9,788 m/s“ Taiwan 9,788 m/s“
Helsinki 9.81 9 me таре 9.790 m/s”
wa , 5
Lisbon 9.801 m/s? Vancouver, BC 9.609 ms?
London (Greenwich) 9.812 m/s* Washington DC 9.801 m/s”
Los Angeles 9.796 m/s? Wellington NZ 9.803 m/s?
Madrid 9.800 m/s* Zurich 9.807 m/s“
ALTITUDE Y
EE q
SSSS EEE
NEL ETRO
Sea Level
300m/1000 ft.
- - | 600m/2000 fi.
900m/3000 ft.
1200m/4000 ft
1500m/5000 ft.
1800m/6000 ft.
2100m/7000 ft.
= g(m/s’)
© eo © © en © [a] oh © © © с © ©
~ ACCELERATION DUE TO GRAVITY
24
LE LS AA NAN
A0
“|
IES
N AIN
ИАН
АА
ПВН
Gé
Au
NA
EE
Ne
a
=
World map
pri ИНН
Too
SE
>
«09 A prt
= €
OOOOOODCGC