- Industrial & lab equipment
- Measuring, testing & control
- Ahlborn
- ALMEMO 2450-1L
- Operating instructions
AHLBORN ALMEMO 2450-1L Universal measuring instrument Operating instructions 24 Pages
AHLBORN ALMEMO 2450-1L Universal measuring instrument Operating instructions
Add to My manuals24 Pages
Below you will find brief information for Universal measuring instrument ALMEMO 2450-1L. This instrument is a universal measuring device equipped with the patented ALMEMO connector system. This enables you to connect a wide variety of sensors, such as thermocouples, NTC probes, humidity sensors, and even your own existing sensors. It features a 2-row LCD display with a variety of functions, including peak value memory and relative measuring. It is powered by 3 AA batteries, has a maximum current consumption of 10 mA, and operates in temperatures ranging from -10 to +50 °C.
advertisement
____________________________
Operating instructions
Universal measuring instrument
ALMEMO
®
2450-1L
V2.3
06.11.2013
www.ahlborn.com
1. Operating controls
1. OPERATING CONTROLS
(1) Measuring inputs M0
M0 for a wide range of sensors
M10 to M30 3 additional channels
(2) LCD
Function field
(a) Function
(b) Measuring point, 2nd meas. value
(c) Units for 2nd measured value
Main field
(c) Units for 1st measured value
(e) 1st measured value
(f) Operating states :
LOBAT
REL
Battery voltage <3.8 V
Relative measuring
(3) Operating keys
ON OFF
Switch the device on
To switch device OFF, press and hold down
M▲
,
M▼
Meas. point selection
MAX , MIN
MEM
CLR
Max. / min. value cancel: press and hold down
Measured value memory
Relative measuring
Adjust sensor, cancel: press and hold down
To switch ON with press keys :
CLR
MEM
M▲
Reinitialization
Device configuration
Softwareversion
Rear of device
(4) Battery compartment
3 AA alkaline-manganese batteries
2 ALMEMO ® 2450-1L
Contents
2. CONTENTS
1. OPERATING CONTROLS..........................................................................2
3. GENERAL...................................................................................................5
3.1 Warranty...............................................................................................5
3.2 Scope of delivery.................................................................................6
3.3 Waste disposal....................................................................................6
4. SAFETY INSTRUCTIONS..........................................................................7
4.1 Special notes on use...........................................................................8
4.2 Handling batteries / rechargeable batteries correctly......................8
5. INTRODUCTION.........................................................................................9
5.1 Functions.............................................................................................9
5.1.1 Sensor programming....................................................................9
5.1.2 Measuring operations.................................................................10
6. INITIAL COMMISSIONING.......................................................................11
7. POWER SUPPLY.....................................................................................11
7.1 Battery operation and supply voltage monitoring..........................11
7.2 Sensor supply....................................................................................11
7.3 Switching ON / OFF, reinitialization.................................................12
7.4 Data buffering....................................................................................12
8. CONNECTING THE TRANSDUCERS......................................................12
8.1 Transducers.......................................................................................12
8.2 Measuring inputs and additional channels.....................................13
9. DISPLAY AND KEYPAD..........................................................................14
9.1 Display................................................................................................14
9.2 Keypad...............................................................................................15
10. MEASURING OPERATIONS....................................................................16
10.1 Measured value...............................................................................16
10.1.1 Selecting a measuring point.....................................................16
10.1.2 Measuring ranges.....................................................................17
10.1.3 Double display..........................................................................18
10.2 Peak value memory.........................................................................18
10.3 Measured value memory.................................................................18
10.4 Relative measuring.........................................................................19
11. DEVICE CONFIGURATION......................................................................19
11.1 Automatic switch OFF.....................................................................20
11.2 Device locking.................................................................................20
12. TROUBLE-SHOOTING............................................................................20
13. DECLARATION OF CONFORMITY.........................................................21
ALMEMO ® 2450-1L 3
2. Contents
14. APPENDIX................................................................................................22
14.1 Technical data.................................................................................22
14.2 Product overview................................................................................
.................................................................................................................. 22
14.3 Index.................................................................................................23
14.4 Your contact....................................................................................24
4 ALMEMO ® 2450-1L
General
3. GENERAL
Congratulations on your purchase of this new and innovative ALMEMO measuring instrument. Thanks to the patented ALMEMO
®
® connector the device configures itself automatically; its operation should be fairly straightforward. The device can, however, be used with such a wide range of sensors and peripherals and offers many different special functions. You are advised therefore to properly familiarize yourself with the way the sensors function and with the device's numerous possibilities and take the time to carefully read these operating instructions and the appropriate sections in the
ALMEMO ® Manual. This is absolutely necessary to avoid operating and measuring errors and to prevent damage to the device. To help you find the answers to your questions quickly and easily there is a comprehensive index at the end both of these instructions and of the Manual.
3.1 Warranty
Each and every device, before leaving our factory, undergoes numerous quality tests. We provide a guarantee, lasting two years from delivery date, that your device will function trouble-free. Before you send your device to us, please
observe the advisory notes in Chapter 12. Trouble-shooting In the unlikely
event that the device proves defective and you need to return it please wherever possible use the original packaging material for dispatch and enclose a clear and informative description of the fault and of the conditions in which it occurs.
This guarantee will not apply in the following cases :
The customer attempts any form of unauthorized tampering or alteration inside the device.
The device is used in environments and conditions for which it is not suited.
The device is used with unsuitable power supply equipment and / or peripherals.
The device is used for any purpose other than that for which it is intended.
The device is damaged by electrostatic discharge or lightning.
The user fails to observe and comply with the operating instructions.
The manufacturer reserves the right to change the product's characteristics in the light of technical progress or to benefit from the introduction of new components.
ALMEMO ® 2450-1L 5
3. General
3.2 Scope of delivery
When you unpack the device check carefully for any signs of transport damage and ensure that delivery is complete.
Measuring instrument ALMEMO
These operating instructions
ALMEMO ® Manual
® 2450-1L with 3 AA alkaline batteries
CD with the AMR-Control software and various useful accessories
In the event of transport damage please retain the packaging material and inform your supplier immediately.
3.3 Waste disposal
The pictogram showing a waste bin crossed through means that the product is subject to European Union regulations on segregated waste disposal. This applies both to the product itself and to any accessories marked with the same symbol. Disposal of any such item as unsorted domestic waste is strictly forbidden
•
•
•
•
•
Please dispose of all packaging materials according to the applicable national waste management regulations.
Please dispose of cardboard boxes, protective plastic packaging materials, and all preservative substances separately and in the proper manner.
The disposal of the device itself (also of device parts, accessories, and consumables) is subject to the applicable national and local waste management regulations and to the environmental protection legislation in force in the country of use.
Please dispose of all waste in the proper manner; this applies in particular to all parts and substances that constitute a hazard for the environment. This includes inter alia plastics, batteries, and rechargeable battery packs .
When disposing of goods, please wherever possible use the original packaging materials.
6 ALMEMO ® 2450-1L
Safety instructions
4. SAFETY INSTRUCTIONS
DANGER Danger to life and limb, risk of damage to equipment
Read the instructions carefully before starting to operate the device.
Please ensure that you comply with all general safety advice and the special safety instructions included in other chapters.
Such risks may occur in the following circumstances :
•
•
•
•
•
•
Failure to heed the operating instructions and all the safety notes these contain
Any form of unauthorized tampering or alteration inside the device
Use of the device in environments or conditions for which it is not suited
Use of the device with an unsuitable power supply and / or in conjunction with unsuitable peripheral equipment
Use of the device for any purpose other than that for which it is intended
Damage caused by electrostatic discharge or lightning.
DANGER Risk of fatal injury caused by dangerously high voltage
Such risks may occur in the following circumstances :
•
•
•
•
Use of the device with an unsuitable power supply and
/ or in conjunction with unsuitable peripheral equipment
Damage caused by electrostatic discharge or lightning
Do not run sensor lines in the vicinity of high-voltage power cables.
Before you touch any sensor lines, ensure that all static electricity has been discharged.
ALMEMO ® 2450-1L 7
4. Safety instructions
DANGER Warning - explosive atmospheres or substances
In the vicinity of various fuels or chemicals there is a risk of explosion.
Do not use the device in the close vicinity of blasting work or filling stations!
4.1 Special notes on use
•
•
If the device is brought into the work-room from a cold environment there is a risk that condensation might form on the electronics. In measuring operations involving thermocouples pronounced changes in temperature may cause substantial measuring errors. You are advised therefore to wait until the device has adjusted to the ambient temperature before starting to use it.
Before using the mains adapter make sure that the mains voltage is suitable.
•
•
Be sure to observe the maximum load capacity of the sensor power supply.
Sensors with their own integrated power supply are not electrically isolated from one another
4.2 Handling batteries / rechargeable batteries correctly
When inserting batteries / rechargeable batteries ensure that these are correctly polarized.
If the device will probably not be needed for a relatively long period of time or if the batteries are empty, remove the batteries; this will prevent battery acid leaking onto the device and damaging it.
Rechargeable batteries should be recharged as and when necessary.
You should never attempt to recharge an ordinary (nonrechargeable) battery; it may explode !
Batteries / rechargeable batteries must never be short-circuited or thrown onto the fire.
Batteries / rechargeable batteries are special waste and must not be discarded together with normal domestic waste.
8 ALMEMO ® 2450-1L
Introduction
5. INTRODUCTION
The ALMEMO
®
2450-1L is a new member in our family of unique measuring devices - all equipped with Ahlborn's patented ALMEMO
The intelligent ALMEMO
®
®
connector system.
connector offers decisive advantages when connecting sensors and peripherals because all parameters are stored in an EEPROM located on the connector itself; repeat programming is thus no longer necessary.
All sensors can be connected to all ALMEMO ing points apply to all devices in the ALMEMO described in detail in the ALMEMO each device.
®
measuring instruments in the same way. Programming and functioning are identical for all units. The follow-
®
measuring system; these are
®
Manual which is included in delivery with
Detailed explanation of the ALMEMO
®
system (Manual Ch 1)
Overview of the device functions and measuring ranges (Manual Ch 2)
Basic principles, operation, and technical data for all sensors (Manual Ch 3)
Options for connecting your own existing sensors (Manual Ch 4)
The operating instructions you are now reading cover only those features and controls that are specific to this device. Many sections therefore also refer to the more detailed description in the Manual; (see Manual, Section xxx).
5.1 Functions
The ALMEMO
®
2450-1L measuring instrument has just one measuring input suitable for most ALMEMO
®
sensors - with the single exception of resistance sensors. The measuring possibilities are numerous; there are 4 channels in the sensor connectors with over 40 measuring ranges. For operation purposes the device incorporates a large LCD display and a keypad.
5.1.1 Sensor programming
The measuring channels are programmed, completely and automatically, by the ALMEMO
®
connectors. The user can supplement or modify this programming; this applies to the other devices only. However, this device also behave according to all the programmed parameters.
Measuring ranges
Appropriate measuring ranges are available for all sensors with a non-linear characteristic, e.g. 7 thermocouple types, NTC probes, and flow transducers
(rotating vanes, thermoanemometers). For humidity sensors additional function channels are available for calculating humidity variables such as dew point, mixture ratio, vapor pressure, and enthalpy. Measured values from other sensors can also be acquired using the voltage and current ranges with individual scaling in the connector. Existing sensors can also be used - so long as the appropriate ALMEMO
®
connector is connected via its screw terminals. For digital input signals, frequencies, and pulses, adapter connectors are available with an integrated microcontroller. It is thus possible to connect numerous sensor
ALMEMO ® 2450-1L 9
5. Introduction types to any ALMEMO
®
measuring instrument and to change sensors without the need for any extra settings.
Units
The 2-character units display can be adapted for each measuring channel so that both the display and the printout always indicate the correct units, e.g.
when a transmitter is connected. Conversion between °C (Centigrade) and °F
(Fahrenheit) is performed automatically.
Measured value designation
Each sensor is identified by means of a 10-character alphanumeric name. This name is entered via the interface and will appear in the printout or on the computer display.
Correction of measured values
The measured value on each measuring channel can be corrected both in terms of zero-point and gain; this means that even sensors usually requiring initial adjustment (e.g. expansion, force) can be freely interchanged. Sensors with multi-point calibration can also be connected; (see Manual Section
6.3.13).
Scaling
The corrected measured value on each measuring channel can also be further scaled in terms of zero-point and gain - using the base value and factor. The decimal point position can be set by means of the exponent function.
Sensor locking
All sensor data stored in the connector EEPROM can be protected by means of a graduated locking function against undesired access.
5.1.2 Measuring operations
A total of up to 4 measuring channels are available for 1 transducer; i.e. it is also possible to evaluate double sensors, individually scaled sensors, and sensors with function channels. You can move forwards or backwards from one measuring channel to the next using the keypad. The selected measuring point is by default assigned preferred status and is scanned at half the measuring rate; all other active channels are also scanned but in the background (semicontinuous mode). The data is output on the display.
Measured values
The measured value for the selected measuring point is shown continuously with autozero and, as and when necessary, with measured value correction.
With most sensors, sensor breakage is detected automatically (except for connectors with shunt, dividers, or additional electronics).
Measuring functions
With some sensors, to achieve optimal measured value acquisition, certain special measuring functions are required. These include e.g. cold junction compensation for thermocouples and temperature and atmospheric pressure compensation for certain humidity variables.
10 ALMEMO ® 2450-1L
Functions
Maximum and minimum values
For each measuring operation the maximum value and minimum value are acquired and saved to memory. These values can then be displayed, output, or deleted from memory.
Measured value memory
A measured value in the display can be saved by simply pressing a key.
6. INITIAL COMMISSIONING
1. Connect sensor to socket M0 (1); see 8.
2. Ensure that the power supply is provided by 3 AA batteries;
3. To switch ON press key ON
4. Select measuring channels by pressing key M▲ (3), read out measured
5. Save the measured value by pressing key MEM
6. Relative measuring to a reference value by pressing key CLR (3)
Revert to normal measured value by pressing and holding down key CLR
7. Evaluating a measuring operation
Call up maximum / minimum values by pressing keys MAX / MIN (3)
To delete max. / min. value(s) press and hold down key MAX or MIN see
7. POWER SUPPLY
Power can be supplied to the measuring instrument with 3 AA alkaline batteries (included in delivery)
7.1 Battery operation and supply voltage monitoring
Power is supplied to the measuring instrument as delivered by 3 AA batteries.
At a current consumption of approx. 10 mA the operating time will be approx.
250 hours. The current operating voltage is displayed each time the the device is switched on; this gives you a basis for estimating the remaining operating time. When the remaining battery capacity drops to approx. 10 percent, the LO-
BAT arrow will appear in the display. If the batteries are completely discharged the device will switch off. To replace old batteries first unscrew the battery compartment cover (7) on the rear of the device.
7.2 Sensor supply
At the terminals + (plus) and – (minus) in the ALMEMO ® connector there is a 9volt sensor supply voltage available (maximum 150 mA) (self-healing fuse, 500 mA). Other voltages (12, 15, or 24 V or references for a potentiometer and strain gauge) can be obtained using special connectors; (see Manual 4.2.5 and
4.2.6).
ALMEMO ® 2450-1L 11
7. Power supply
7.3 Switching ON / OFF, reinitialization
To switch the device ON briefly press and release the key
ON OFF r
(3) in the middle of the keypad; to switch the device, OFF press and hold down the key
ON OFF r
. After the device is switched off all saved values and settings are re-
If interference (e.g. electrostatic) or a malfunction (e.g. battery failure) causes the device to behave abnormally, the device can be reinitialized. To activate
RESET press and hold down the key CLR r when switching on. This will restore all settings to the factory default status. Only the programming of the sensors in the ALMEMO
®
connectors remains unaffected.
7.4 Data buffering
The sensor's programming is stored in the EEPROM on the sensor connector and the device's calibration and programmed parameters are stored in the
EEPROM on the instrument itself, both on a fail-safe basis.
8. CONNECTING THE TRANSDUCERS
Only certain ALMEMO
®
sensors can be connected to ALMEMO
®
input socket
M0 (1), namely those programmed with measuring ranges as per Ch 10.1.2.
Attempting to connect any other sensor here will trigger an error message. To connect your own existing sensors you simply need the appropriate
ALMEMO
®
connector.
8.1 Transducers
The ALMEMO
®
Manual includes detailed descriptions of the ALMEMO
®
range of sensors (see Manual Ch 3) and instructions for connecting your own existing sensors to ALMEMO
®
instruments (see Manual Ch 4). All suitable sensors with an ALMEMO
®
connector (see above) have the measuring range and units already programmed and can thus be connected to the input socket without further adjustment. A mechanical coding system ensures that sensors and output modules can only be connected to the correct sockets. All ALMEMO
®
connectors incorporate two snap-lock levers; these snap into position as soon as the connector is inserted into the socket, thus preventing unintended disconnection if the cable is accidentally pulled. To withdraw the connector, both these levers must be pressed in at the sides.
For the ALMEMO
®
2450-1L version with the optional seal new specially designed sensors are available with spray-coated ALMEMO
®
connectors incorporating a double sealing lip to protect the socket unit against the effects of splashing water. For any unused sockets protective stoppers are available.
12 ALMEMO ® 2450-1L
Measuring inputs and additional channels
8.2 Measuring inputs and additional channels
The ALMEMO
®
2450-1L measuring instrument has one input socket (1) to which initially measuring channel M0 is assigned. ALMEMO
®
sensors can, however, if necessary, provide up to four channels. The additional channels can be used in particular for humidity sensors with four measuring variables
(temperature / humidity / dew point / mixture ratio) or for function channels.
Each sensor can if necessary be programmed with several measuring ranges or scaling settings; and two or three sensors, if pin assignment so permits, can be combined in a single connector (e.g. rH / NTC, mV / V, mA / V, etc.). The additional measuring channel numbers per connector go up in steps of 10 (e.g.
the first sensor has channels M0, M10, M20, M30).
Double connectors with 2 x differential voltage / differential current
(input D - B) are not possible.
On the measuring instrument this gives the following channel assignment : sensor channels
4. chan.
3. chan.
2. chan.
1. chan.
30
20
10
00
M0
P0
A2 A1 DC
Sensors combined within one connector and sensors with their own power supply are electrically interconnected and must therefore be operated in isolation. The voltage at the measuring inputs themselves must not exceed 5 volts
(between B, C, D, A and - ).
ALMEMO ® 2450-1L 13
9. Display and keypad
9. DISPLAY AND KEYPAD
9.1 Display
The display (2) on the ALMEMO ® 2450-1L measuring instrument is a 2-row
LCD arrangement; the main field comprises 5x 7-segment digits (e) plus 2x 16segment digits (d) for depicting the measured value;,the function field comprises 41/2x 7-segment digits (b) for depicting various measuring functions (a); there are also 2 arrows (f) for depicting the operating status.
Function field
Main field
Display of measuring functions in
Measuring point
Maximum value
Minimum value
Saved value
M
O
MAX
36
.
5
MIN
17
.
3
M
36
.
2
Temperature value from double sensors
Configuration of locking
Configuration of automatic OFF
26
.
5
°C
Loc
AOFF the
function field
14 ALMEMO ® 2450-1L
Display
Special operating states and faults
Display segment test :
Supply voltage runs automatically after switch ON
Display after segment test
Under 3.8 V : LOBAT arrow lights up
Relative measuring with respect to a reference value : REL arrow lights up
Checksum error in device calibration :
CALEr
Non-connected sensors, deactivated measuring points :
-----
Measuring range / function not permitted :
Multiplexer M5 (D-B) is not allowed:
Err
ErrI
Sensor breakage :
NiCr
flashes
Outside of measuring range, undershoots cold junction compensation or cold junction compensation breakage :
CJ
flashes
Overshoots values range (>65000) :
Overshoots measuring range :
Undershoots measuring range :
9.2 Keypad
To operate the device a keypad with 7 keys is provided :
65OOO
flashes
Maximum value flashes
Minimum value flashes
Function :
To switch ON the device : (see Section 7.3)
Key
ON OFF
To switch OFF the device : must be pressed and held down
Function :
ON OFF
Key
Selection of measuring points (see Section 10.1.1)
M
or
M
ALMEMO ® 2450-1L 15
9. Display and keypad
Displaying the maximum value : (see Section 10.2)
MAX
To delete press and hold down
Displaying the minimum value : (see Section 10.2)
MIN
To delete press and hold down
Zero-setting the measured value : (see Section 10.4)
CLR
To delete press and hold down
Saving the measured value : (see Section 10.3)
MEM
Displaying the battery voltage :
ON OFF
10. MEASURING OPERATIONS
On the ALMEMO
®
2450-1L all measuring channels, whenever available, are scanned semi-continuously at 2.5 mops; (see Manual 6.5).
Up to 4 measuring points can be displayed; (see Section 8.2)
10.1 Measured value
After switching ON first of all a segment test is performed; then the battery voltage appears and if the batteries are almost empty
(<3.8 V) the LOBAT arrow also appears.
The measured value is then displayed with the appropriate units in the main field and the measuring point is displayed in the function field. All special operating states possible for the measured value are explained in Section
.
10.1.1 Selecting a measuring point
By pressing key
M
you can select one after the other all active measuring points and have the current measured value displayed for each. By pressing key
M
you can return to the previous channel.
To increment the measuring channel press key :
M
To decrement the measuring channel press :
When switching between channels the abbreviation for the measuring range is
briefly displayed; (see 10.1.2).
16 ALMEMO ® 2450-1L
Measured value
10.1.2 Measuring ranges
With each channel switchover or sensor breakage the abbreviation for the measuring range appears in the display. For identification purposes the following table lists all possible measuring ranges supported by this device.
Transducer Sensor / conMeasuring range Units Abbre-
NiCr-Ni (K)
NiCroSil-NiSil (N)
Fe-CuNi (L)
Fe-CuNi (J)
Cu-CuNi (U)
Cu-CuNi (T)
PtRh10-Pt (S)
Ntc Typ N
Millivolt
Millivolt 1
Millivolt 2
Volt
Difference millivolt
Difference millivolt 1
Difference millivolt 2
Difference volt*
Sensor voltage
Milliampere
Percent (4-20mA)
Frequency
Digital input
nector
FT Axxx
ZA 9020-FSN
ZA 9000-FSL
ZA 9000-FSJ -200.0... +950.0
°C IrCo
ZA 9000-FSU -200.0... +600.0
°C CUCO
ZA 9000-FST
FS Axxx
-200.0... +400.0
°C CoCo
°C Pt10
FN Axxx
viation
-200.0...+1370.0
°C NiCr
-200.0...+1300.0
°C NiSi
-200.0... +900.0
°C FECO
ZA 9000-FS0
ZA 9000-FS1
ZA 9000-FS2
ZA 9000-FS3
0.0...+1760.0
-20.00...+100.00
°C Ntc
-10.000...+55.000
mV U 55
-26.000...+26.000
mV U 26
-260.00...+260.00
mV U260
-0.2600...+2.6000
V U2.60
ZA 9000-FS0D -10.000...+55.000
mV d 55
ZA 9000-FS1D -26.000...+26.000
mV d 26
ZA 9000-FS2D -260.00...+260.00
mV d260
ZA 9000-FS3D -0.2600...+2.6000
any 0.00...20.00
V d2.60
V UbAt
ZA 9601-FS1
ZA 9601-FS2
ZA 9909-AK1
ZA 9000-EK2
0.000...+26.000
0.00... 100.00
mA I032
% P420
0... 32000
Hz FrEq
0.0... 100.0
% Inp
Digitale interface
Snap-on head normal 20
Snap-on head normal 40
Snap-on head micro 20
Snap-on head micro 40
Macro
Water-Micro
ZA 9919-AKxx
FV A915-S120
-65000... +65000 diGi
0.30... 20.00
m/s S120
FV A915-S140
FV A915-S220
FV A915-S240
FV A915-MA1
FV A915-WM1
Relative air humidity, capacitive FH A646
Rel. air humidity, capacitive TC FH A646-C
Rel. air humidity, capacitive TC FH A646-R
Mixture ratio, capacitive
Dew-point temperature, cap.
Partial vapor pressure, cap.
Enthalpy, capacitive
FH A646
FH A646
FH A646
FH A646
0.40... 40.00
m/s S140
0.50... 20.00
m/s S220
0.60... 40.00
m/s S240
0.10... 20.00
m/s L420
0.00... 5.00
m/s L605
0.0... 100.0
%H °orH
0.0... 100.0
%H HcrH
0.0... 100.0
%H H rH
0.0 ... 500.0
g/k H AH
-25.0... 100.0
°C H dt
0.0 ...1050.0
mb H UP
0.0 ... 400.0
kJ H En
* Double connectors with differential voltage / differential current (D-B) cannot be used
ALMEMO ® 2450-1L 17
10. Measuring operations
Transducer Sensor / connector
Functions channels:
Maximum value of channel Mb1 any
Minimum value of channel Mb1 any
Alarm value of channel Mb1 any
Measured value of Mb1
Cold junction temperature any any
TC=Temperature compensation
Measuring range Units Abbreviation
Hi
Lo
Alrn
MESS
°C
CJ
10.1.3 Double display
On all double sensors incorporating a temperature sensor on the 1st channel the temperature value can at the same time be displayed in the function field.
Select 2 nd channel
Activate temperature display
Press and hold dow M▲
Return to channel display
Press and hold down M▲
10.2 Peak value memory
From the measured values acquired for each measuring point the highest and the lowest values are continuously recorded. To display these high/low peak values first the desired channel must be set (see Section 7.1) and then the
MAX
or
MIN
key must be pressed. As a check the display also includes the associated symbol.
To display the maximum value press key :
To display the minimum value press key :
To delete the maximum value press and hold down key :
To delete the minimum value press and hold down key :
MAX
MIN
MAX
MIN
.
.
.
.
To return to the measuring point display press key : .
As soon as you clear the memory, the current measured value will appear (because measuring is continuous).
10.3 Measured value memory
On the ALMEMO
®
2450-1L any measured value can be saved. To save the measured value press key :
MEM
.
The value most recently saved then appears in the function field preceded by the symbol
´M´. To return to the channel display press key :
M
.
18 ALMEMO ® 2450-1L
Relative measuring
10.4 Relative measuring
One very useful function is to zero the measured value at certain locations or at certain times as a reference value in order then to observe only the subsequent deviations. This function is independent of the locking status and does not modify the programming parameters in the connector.
To zero-set the measured value press key :
To display relative measuring press arrow :
CLR
.
REL
.
To return to normal measured value press and hold down key :
CLR
.
Setting to zero automatically deletes the maximum and minimum values for this channel. The MAX, MIN, and MEM functions are thus also available for relative measurement.
11. DEVICE CONFIGURATION
On the ALMEMO
®
2450 measuring instrument a number of parameters can be configured. To do so when switching ON press and hold down key
MEM
. The function field should then show an abbreviation for the parameter and the main field should show the value currently set.
To select from all possible parameters,
if any are available, press keys : or
M
.
Automatic switch OFF time in minutes: see 11.1
Loc
AOFF
To enter a value first press : starts flashing.
To modify the value press keys :
To delete parameters press :
Entry is completed by again pressing key :
ON
or
CLR
ON
To terminate configuration at any time press key:
MEM
and the value
M
ALMEMO ® 2450-1L 19
11. Device configuration
11.1 Automatic switch OFF
In menu item ´
AOFF´ an automatic device switch OFF time can be programmed in minutes; this will help save the batteries. This automatic device switch OFF will not take effect if the setting is ´- -´ or if a mains adapter or an interface cable is connected.
11.2 Device locking
The measured value in the main field of the display can be manipulated by pressing key
CLR
and setting it to zero. This function can be evaluated in different ways or even switched off in cases where there is a risk of accidentally activating relative measuring by zero-setting the measured value.
Loc parameter :
0 The offset is saved in RAM, base or zero-point - depending on locking
1 The offset is saved in RAM only.
2 Relative measuring is locked
12. TROUBLE-SHOOTING
The ALMEMO ® 2450-1L measuring instrument can be configured and programmed in many versatile ways. It is suitable for connecting a wide variety of different sensors. Given these numerous possibilities the device may in certain circumstances not behave quite as expected. The cause of such unexpected behavior is only very rarely a device defect; more usually it is incorrect operation by the user, an invalid setting, or unsuitable cabling. In such event try to pinpoint and clear the problem with the aid of the following tests.
Error:
No display, display malfunction, keys do not react
Remedy: Check the power supply; replace the batteries; switch off and then
on again; if necessary re-initialize (see 7.3).
Error:
Measured values are incorrect.
Remedy: Switch Device OFF / ON, press key and hold CLR . Check all the sensor programming very carefully, especially the base value and zero-point.
Error:
Fluctuating measured values or the system hangs in mid-operation.
Remedy: Check the cabling for any inadmissible electrical connections,
Unplug any suspicious sensors.
Connect hand-held sensors in air or phantoms (for thermocouples short-circuit AB) and check.
Connect the sensors again one at a time and check successively.
If a fault persists for any one connection, then check all wiring; if necessary, insulate the sensor and eliminate interference by using shielded or twisted wiring.
Error: ´
CALEr´is displayed when the device is switched on.
Remedy: The calibration of a measuring range may have become misad-
20 ALMEMO ® 2450-1L
Trouble-shooting justed. The device must be recalibrated at the factory.
If, after performing the above-listed checks and remedial steps, the device still fails to behave as described in the operating instructions, it must be returned to our factory in Holzkirchen with error description.
13. DECLARATION OF CONFORMITY
Ahlborn Mess- und Regelungstechnik GmbH declares herewith that measuring instrument ALMEMO ® 2590 carries the CE label and complies in full with the requirements of EU directives relating to low voltage and to electromagnetic compatibility (EMC) (89/336/EWG).
The following standards have been applied in evaluating the product.
Safety / security: EN 61010-1:2001
EMC: EN 61326: 2006
If a product is modified in any manner not agreed with us in advance, this declaration becomes void.
When using the sensor with an extension care must be taken to ensure that wiring is not laid alongside or close to high-voltage power cables and that it is, if necessary, properly shielded so as to prevent spurious interference being induced in the system.
The following advisory notes must be observed when operating the device :
Using the device in strong electromagnetic fields may aggravate measuring errors (<50 µV at 3 V/m and 1.5 meters thermocouple sensor). After exposure to such irradiation ceases, the device will again operate within its technical specifications
.
ALMEMO ® 2450-1L 21
14. Appendix
14. APPENDIX
14.1 Technical data
Measuring inputs :
Measuring channels:
A/D converter :
Measuring ranges:
(see Manual 2.3)
1 ALMEMO ® socket suitable for ALMEMO ® connectors maximum 3 additional channels for double sensors and function channels
Delta - sigma, 16-bit, 2.5 mops, adjustable 1 to 100 see Measuring range list Chap. 10.1.2 on page 20
Some measuring ranges differ from the standard
ALMEMO® ranges
9 volts, maximum 150 mA Sensor power supply :
Standard equipment :
LCD : Measured value : 5x 7-segment 15 mm,
2x 16-segment 9 mm
Function 4½ x 7-segment 9 mm, 9 symbols
7 silicone keys
1 Measured value on the RAM
Operation :
Memory :
Power supply :
Batteries :
Current consumption
Housing :
3 AA alkaline batteries approx. 10 mA (without input modules)
(LxWxH) 127 x 83 x 42 mm
ABS (acrylonitrile butadiene styrene), weight : approx. 260 g
Suitable conditions
Operating temperature
(storage temperature
Ambient relative humidity :
-10 to +50 °C
-20 to +60 °C)
10 to 90 % rH (non-condensing)
14.2 Product overview
Universal measuring instrument ALMEMO 2450-1L
1 measuring input , 2-row LCD, 7 keys, battery supply
Options
Measuring ranges for temperature display of 10 refrigerants
Top hat rail mounting
Order no.
MA 2450-1L
SB 0000-R
ZB 2450-HS
22 ALMEMO ® 2450-1L
14.3 Index
additional channels
AOFF
Automatic switch OFF
Battery operation
Connecting the transducers
Data buffering
Declaration of conformity
Device configuration
Device locking
Display
Double display faults
Function field
Functions
Functions channels
Housing
Initial commissioning
Introduction keypad
Keypad
Main field maximum value
Measured value
Measured value memory
Measuring inputs
Measuring inputs
Measuring operations
Measuring ranges minimum value
ON OFF operating states
Options
Order no.
Peak value memory
Power supply
Product overview reference value reinitialization
Relative measuring
Safety instructions
Scope of delivery
Selecting a measuring point
Index
16
17
18
15f.
15
22
22
18
11, 22
22
14
18
16
18
11
9
14
15
13
22
19
12
19
7
6
16
18
15
14
9
21
19
20
14
18
22
13
20
20
11
12
12
23
10
10.1.2
10.2
9.2
9.1
14.2
14.2
10.2
14.1
14.2
6
5
9
9.2
9.1
10.2
10.1
10.3
8.2
14.1
8.2
11.1
11.1
7.1
8
7.4
13
11
11.2
9
10.1.3
9.1
9.1
5.1
10.1.2
14.1
10.4
7.3
10.4
4
3.2
10.1.1
ALMEMO ® 2450-1L
14. Appendix
Sensor programming
Sensor supply
Standard equipment
Suitable conditions supply voltage monitoring
Switching ON / OFF
Technical data
Temperature compensation the measured value
To switch OFF
To switch ON
Transducers
Trouble-shooting
Warranty
Waste disposal
14.4 Your contact
AHLBORN Mess- und Regelungstechnik GmbH
Eichenfeldstraße 1
83607 Holzkirchen
Germany internet : http://www.ahlborn.com
e-mail : [email protected]
5.1.1
7.2
14.1
14.1
7.1
7.3
14.1
10.1.2
10.4
9.2
9.2
8.1
12
3.1
3.3
12
20
5
6
18
19
15
15
9
11
22
22
11
12
22
Even the greatest possible care cannot exclude the possibility of inaccuracies.
We reserve the right to make technical changes without advance notice.
24 ALMEMO ® 2450-1L
advertisement
Key Features
- Patented ALMEMO connector system
- 2-row LCD display
- Peak value memory
- Relative measuring
- Battery powered
- Operates in temperatures from -10 to +50 °C
Frequently Answers and Questions
What sensors can I connect to this instrument?
How do I switch the device on and off?
What are the maximum and minimum values?
What is relative measuring?
Related manuals
advertisement
Table of contents
- 2 1. OPERATING CONTROLS
- 5 3. GENERAL
- 5 3.1 Warranty
- 6 3.2 Scope of delivery
- 6 3.3 Waste disposal
- 7 4. SAFETY INSTRUCTIONS
- 8 4.1 Special notes on use
- 8 4.2 Handling batteries / rechargeable batteries correctly
- 9 5. INTRODUCTION
- 9 5.1 Functions
- 9 5.1.1 Sensor programming
- 10 5.1.2 Measuring operations
- 11 6. INITIAL COMMISSIONING
- 11 7. POWER SUPPLY
- 11 7.1 Battery operation and supply voltage monitoring
- 11 7.2 Sensor supply
- 12 7.3 Switching ON / OFF, reinitialization
- 12 7.4 Data buffering
- 12 8. CONNECTING THE TRANSDUCERS
- 12 8.1 Transducers
- 13 8.2 Measuring inputs and additional channels
- 14 9. DISPLAY AND KEYPAD
- 14 9.1 Display
- 15 9.2 Keypad
- 16 10. MEASURING OPERATIONS
- 16 10.1 Measured value
- 16 10.1.1 Selecting a measuring point
- 17 10.1.2 Measuring ranges
- 18 10.1.3 Double display
- 18 10.2 Peak value memory
- 18 10.3 Measured value memory
- 19 10.4 Relative measuring
- 19 11. DEVICE CONFIGURATION
- 20 11.1 Automatic switch OFF
- 20 11.2 Device locking
- 20 12. TROUBLE-SHOOTING
- 21 13. DECLARATION OF CONFORMITY