Order Reference Order No. Recommended accessories

Order Reference Order No. Recommended accessories
Order Reference
Order Reference
Order No.
SATURN GEO easy .........................................................................A 1855 06611
Scope of delivery
6 alcaline AA - type (LR6) batteries
2 measuring leads 1,5 m
1 connector cable (for RA 2-pole measurements)
2 alligator clips
1 carrying belt and hooks
1 operating instructions with CE-certificate
1 LEM NORMA test certificate
Recommended accessories:
Set for 3-pole measurements:............................................A 6045 10302
2 x 1 earth stake................................................................A 6045 10350
1 cable reel with 25m wire ..............................................A 6045 05102
1 cable reel with 50m wire ..............................................A 6045 05103
Set for 4-pole measurements:............................................A 6045 10301
4 x 1 earth stake...............................................................A 6045 10350
2 x 1 cable reel with 25m wire .......................................A 6045 05102
1 cable reel with 50m wire ..............................................A 6045 05103
Carrying case
for SATURN GEO easy and accessories.....................A 6030 00530
Unpacking: Check delivery for damage during transport.
Keep the packing material for later transport and check
scope of delivery.
-2-
Contents
Contents:
1
Safety Regulations..........................................................................................4
2
Description of instrument.............................................................................5
3
SETUP ................................................................................................................6
4
RA 2-pole, 3-pole measurements ...............................................................7
5
RA 4-pole measurements.............................................................................8
6
Soil resistivity...................................................................................................9
7
Troubleshooting ............................................................................................11
8
Technical Data...............................................................................................13
9
Maintenance, Service, Warranty...............................................................17
10 Other products ..............................................................................................19
-3-
Safety Regulations
1 Safety Regulations
This measuring device is only to be installed and operated by qualified
personnel and according to the technical data in compliance with the
safety precautions and regulations set forth below. Additionally, the use
of this equipment requires compliance with all legal and safety
regulations pertaining to each specific application. Similar regulations
apply to the use of accessories.
Operating electrical devices implies that parts of the device carry
dangerous voltages. Disregarding warning notices may lead to serious
physical injury and material damage.
It can be assumed that safe operation is no longer possible if the device
•
shows visible damage,
•
has been exposed to unfavourable conditions (e.g. storage beyond
the permissible climatic limits without adaptation to the ambient
climate, dewing etc.) or
•
has been exposed to major strain during transport (e.g. been
dropped from some height without visible external damage etc.).
No measurements must be performed on unprotected
measuring circuits.
While a measurement is in progress no earth stake must be touched!
Qualified Personnel
are persons familiar with the setting up, installation, starting off and
operation of the device and possess a formal qualification required for
such activities.
-4-
Description of Instrument
2 Description of instrument
This instrument has been manufactured complying with quality assurance
system DIN ISO 9001. The compliance with the actual EMC-regulations is
documented by the attached CE-sign.
Suited for all earth/ground measurements including those complying to
IEC 1024, ENV 61024, DIN VDE 0185, ÖVE-E 49 .
Applications:
•
earth resistance of various installations (high voltage pylons,
buildings, cell phone substations, RF transmitters etc.)
•
inspection and planning of lightning protection systems
•
soil resistivity (e.g. for geological examinations).
3
4
m ax 50 V
2
P2
S
1
P1
ES
5
6
7
easy
10
9
Œ
•
Ž
•
•
‘
’
“
”
•
8
Rotary switch for measuring function and ON/OFF
START button for starting the selected measuring function
Display unit, liquid crystal display (LCD)
Connection terminal “H” auxiliary earth electrode 4 mm ∅
Connection terminal “S” probe 4mm ∅
Connection terminal “ES” earth probe 4 mm ∅
Connection terminal “E” earth electrode 4 mm ∅
Battery compartment for 6 x Alcaline AA-Type (LR6) batteries or
NiCd (bottom side)
Technical data. Type plate with serial number (bottom side)
Carrying belt attachment (4x)
-5-
SETUP
3 SETUP
Please, read carefully the chapter ”safety regulations” before
powering up the instrument.
If you have problems see section ”Troubleshooting”.
Insertion of batteries
Œ
•
Ž
•
•
1
Switch off instrument
Disconnect all test leads
Open battery compartment
Insert batteries. Close battery compartment
Apply carrying belt as shown
3
2
4
5
+
-6-
RA 2-pole, 3-pole measurements
4 RA 2-pole, 3-pole measurements
For 2-pole measurements connect terminals H and S with the supplied
connector cable!
Œ
Select function RA 3-pole
Display is as shown on the right.
•
Connect test leads
Hint: Colour coded front plate and terminals provide appropriate connection.
Connect terminal E/C1 to earth electrode with short test lead (1,5 m).
Place 2 ground stakes. Minimum distance between earth electrode (E/C1),
probe (S/P2), and auxiliary earth (H/C2) is 20m!
Connect stakes with 25m and 50m wires to H/C2 and S/P2 as shown.
Ž
Press START
The ”active” symbol indicates that measurement is in progress.
Continuous measurement as long as START is pressed.
•
Symbol ” ” indicates completed measurement. The result is kept on the
display until a new measurement is started or the main switch is turned.
1
2
ma
x 50V
E
S
H
H/C2
E/C1 S/P2
eas y
>20 m
3
>2 0 m
4
active
E
-7-
ES
S
H
E
ES
S
H
RA 4-pole measurements
5 RA 4-pole measurements
Œ
Select function RA 4-pole
Display as shown on the right.
•
Connect test leads
Hint: Colour coded front plate and terminals provide appropriate connections.
Connect terminals E/C1 and ES/P1 to earth electrode with 2 short test
leads (1,5 m). Place 2 ground stakes. Minimum distance between earth
electrode (E/C1), probe (S/P2), and auxiliary earth (H/C2) is 20m! The ES
test lead eliminates the influence of the test leads.
Connect stakes with 25m and 50m wires to H/C2 and S/P2 as shown.
Ž
Press START
The ”active” symbol indicates that measurement is in progress. For
continuous measurement keep START pressed.
•
Symbol ” ” indicates completed measurement. The result is kept on
display until a new measurement is started or the rotary switch is turned.
2
1
ma
x 50V
P2
S
E
ES
S
H
ES/P1
E/C1
P1
ES
S/P2
H/C2
ea sy
>20 m
3
>20 m
4
acti ve
E
-8-
ES
S
H
E
ES
S
H
Soil resistivity
6 Soil resistivity
Œ
Select function
ρ
Display as shown on th e right.
•
Connect test leads
Hint: Colour coded front plate and terminals provide appropriate connections.
Place 4 ground stakes in a straight line with a distance of 4 m between
them. Connect test leads as shown.
Ž
Press START
The ”active” symbol indicates that measurement is in progress. For
continuous measurement keep START pressed.
•
Symbol ” ” indicates completed measurement. The result is kept on
display until a new measurement is started or the rotary switch is turned.
Note: The distance between the stakes must be 4 m. For other distances the
result must be calculated as:
ρ real = ρ reading * distance [m] / 4
-9-
Soil resistivity
1
activ e
m
E
ES
S
2
H
P2
S
4m
4m
4m
4m
4m
P1
4m
ES
E/C1 ES/P1 S/P2 H/C2
ea sy
4m
3
4m
4m
4
acti ve
m
E
- 10 -
ES
S
H
m
E
ES
S
H
Troubleshooting
7 Troubleshooting
Œ
External voltage (Uext) too high
If the external voltage applied to the instrument is too high, no
measurement can be started (see “Technical data” for Uext limit).
Hint: Reposition probe (S/P2) and restart measurement.
•
Auxiliary earth electrode resistance (R H) too high
If the auxiliary earth electrode resistance is too high it is not possible to
drive the current necessary for reliable measurements. The measurement
is blocked (see “Technical data” for Rh limit).
Hint: Check connection of test lead with terminal H/C2, check auxiliary earth
stake.
Ž
Probe resistance (Rs) too high
If the probe resistance is too high measurements are not reliable. The
measurement is blocked (see “Technical data” for Rs limit).
Hint: Check connection of test lead with terminal S/P2, check probe stake.
•
Weak batteries
If the batteries are weak the supply voltage may break down during
measurement. If there is enough energy to complete the measurement
”
” symbol is displayed - measuring results are valid. If not, a reset
occurs.
Hint: Replace batteries. Use 6 alkaline AA-type (LR6) batteries.
•
Is your R A measurement result reliable?
Probe S/P2 must be outside the potential gradient areas of E/C1 and H/C2
for accurate measurements. Normally a probe distance of more than 20m
is sufficient. However in some environmental conditions (depending mainly
on the soil resistivity) this may not suffice. To be sure reposition the probes
and take several measurements. If the readings are appr. the same your
measuring results are reliable, if not increase the probe distance.
- 11 -
Troubleshooting
1
3
U >Limit
ext
E
S
ma
x5
0V
S
H
>Li mit
E ES S
H
E/ C1
S/ P2
H/ C2
ea s y
4
2
>20 m
>20 m
Us
H
>Li mit
E
ES S
H
E
S
H
5
- 12 -
Ug
Technical Data
8 Technical Data
Display:
1999 digit LCD-display with special symbols, digit height 25 mm,
fluorescent backlight
User interface:
Instant measurement through TURN and START - one button concept
. The only operating elements are rotary switch and START button.
Robust &
waterproof:
Instrument is designed for tough environmental conditions
(rubber protective cover, IP56).
Temperature ranges:
Working temp.:
Operating temp.:
Storage temp.:
Reference temp.:
-10° C ... +50° C
0° C ... +35° C
-20° C ... +60° C
+23° C ±. 2° C
( +14° F ...+122° F )
( +32° F ... +95° F )
( -4° F ...+140° F )
( +73° F
± 4° F )
Temp. coefficient: ± 0,1 % of reading / K
Intrinsic error:
Refers to the reference temperature range and is guaranteed for 3
years.
Operating error:
Refers to the operating temperature range and is guaranteed for 3 years.
Climatic class:
C1 (IEC 654-1), -5° C...+45°C, 5%...95% RH
Protective type:
IP56 according to EN 60529
Safety:
Protection by
reinforced insulation
EMC (Emission):
IEC 61326-1:1997 Class B
CISPR16 (CISPR22), CISPR16-1
EMC (Immunity):
IEC61326-1:1997
IEC 61000-4-2 8 kVair / 4 kV contact (A)
IEC 61000-4-3 3 V/m (A)
IEC 61000-4-4 2 kV (A)
IEC 61000-4-5 2 kV (A)
IEC 61000-4-6 3V (A)
IEC 61000-4-8 30 A/m (A)
Quality system:
developed, designed and manufactured according to DIN ISO 9001
External voltage:
values
Uext,max = 24 V (DC, AC < 400Hz), measurement inhibited for higher
Uext rejection:
>120dB (162/3, 50, 60, 400Hz)
Measuring time:
typical 6 sec
Max. overload:
250 Vrms
Auxiliary power:
6 x 1,5 V mignon cells alkali-manganese (type AA LR6)
- 13 -
Technical Data
Battery life span:
typical > 3000 measurements
Dimensions:
240 x 180 x 110 mm
Weight:
1,1 kg (including batteries)
- 14 -
Technical Data
RA 3-pole ground resistance measurement
(IEC 1557-5)
Switch
position
Resolution
measuring range
Intrinsic error
Operating error
Ra 3-pole
0,001 … 10 Ω
0,001 Ω ... 19,99 kΩ
±(2% rdg + 3 d)
±(5% rdg + 3 d)
For 2-pole measurements connect terminals H and S with the supplied connector cable!
Measuring principle: Current and voltage measurement
Measuring voltage:
Short-circuit current:
Meas. frequency:
Probe resistance (Rs):
Auxiliary earth electrode resistance (Rh):
Additional error from Rh and Rs:
Monitoring of Rs and Rh with error indicator.
Automatic range selection.
Um = 48 Vac.
> 50 mA
128 Hz (125 Hz on request)
max 100 kΩ
max. 100 kΩ
Rh[kΩ] ž Rs[kΩ] / Ra[Ω] ž 0,2 %
RA 4-pole ground resistance measurement
(IEC 1557-5)
Switch
position
Resolution
measuring range
Intrinsic error
Operating error
Ra 4-pole
0,001 … 10 Ω
0,001 Ω ... 19,99 kΩ
±(2% rdg + 3 d)
±(5% rdg + 3 d)
Measuring principle: Current/voltage measurement
Measuring voltage:
Short-circuit current:
Measuring frequency:
Probe resistance (Rs+ Res):
Auxiliary earth electrode resistance (Rh):
Additional error from Rh and Rs:
Monitoring of Rs, and Rh with error indicator.
Automatic range selection.
Um = 48 Vac.
> 50 mA
128 Hz (125 Hz on request)
max. 100 kΩ
max. 100 kΩ
Rh[kΩ] ž Rs[kΩ] / Ra[Ω] ž 0,2 %
Soil resistivity (ρ) measurement
- 15 -
Technical Data
Switch
position
ρ
measuring range
Intrinsic error
Operating error
0,1 Ωm...99,9 kΩm ±(2% of reading + 3 digits) ±(5% of reading + 3 digits)
Measuring principle: Current and voltage measurement
Measuring voltage:
Short-circuit current:
Measuring frequency:
Probe resistance (Rs):
Auxiliary earth electrode resistance (Rh):
Monitoring of Rs and Rh with error indicator.
Automatic range selection.
Um = 48 Vac.
> 50 mA
128 Hz (125 Hz on request)
max. 100 kΩ
max. 100 kΩ
Note: ρ measurement specifications relate to stake distances of 4 m! Errors are
given as % of measured value (% of rdg) and digits (dig).
- 16 -
Maintenance, Service, Warranty
9 Maintenance, Service, Warranty
If the instrument is operated and handled properly it does not need specific
maintenance and service. Maintenance work must only be undertaken by
trained and qualified staff and during the warranty period only by authorised
service centres.
Calibration
As an additional service we offer a regular check up and calibration of your
instrument. Upon request and for a fee we supply company test certificates
or test certificates of the public calibration service.
Storage
In case the instrument is either not used or stored for a longer period,
remove batteries and keep them separately beforehand to avoid possible
damage by leaking battery electrolyte.
Service
Dear customer,
This instrument has been manufactured and tested according to the latest
technological innovations and in compliance with the
quality assurance system DIN ISO 9001.
If there should still be reason for complaint please refer to your nearest
service centre, giving a detailed description of the defect by using the
enclosed service card and including a copy of the invoice or the delivery
note. If the instrument is operated as agreed we guarantee fault free
function and calibration of the instrument for a period of TWO (2) years after
the date of the first end-user purchase.
This warranty does not apply to any damage caused by improper treatment
or operation or by impermissible overload.
Packing
Use for shipping the original packing only. The manufacturer doesn't
accept any granting security for damage that might have been formed by an
inappropriate packing.
Warranty
2
years
WARRANTY
2 years
2 years
warranty for the indicated accuracy of the device.
entitlement to repairs free of charge in case of
malfunction if instrument was used properly.
- 17 -
Maintenance, Service, Warranty
Note:
For reasons of clarity this manual does not contain
every detailed information about all models of this
product and therefore cannot comprise every
imaginable way of use, operation or maintenance
- 18 -
Other products
10 Other products
The LEM group offers a wide range of other products for testing the
efficiency of protective measures:
Earth/ground testers: HANDY GEO, SATURN GEO, UNILAP GEO (X)
Installation testers:
SATURN 100, UNILAP 100, UNILAP 100 (X)E
Insulation testers:
HANDY ISO, SATURN ISO, UNILAP ISO X, UNILAP
ISO 5kV
•
Multimeters,
•
Non-invasive transducers for analysis of voltage, current, power,
•
Power quality analysers,
•
Power measurement instruments.
Also available: Appropriate accessories, interfaces (RS232, IrDA), data
memory, PC-software for protocol generation,….
LEM provides complete measurement solutions – please, contact our
sales partners world-wide for additional information.
LEM policy is one of continuous product
improvement and the company reserves the
right to revise the above specifications without
notice.
- 19 -
Notes