product catalog
2005
ww
w.
hi
ok
i.c
o.
jp
1990
2003
2002
1978
0
Celebrating 70 Years
of Technological
Excellence
2001
About the Catalog
Searching for product pages and notes ...
Products in this catalog are grouped according to functions
so you can easily find the right instrument for your
application by referring to the list of product groups in the
table of contents on the first page, and moving directly to
the indicated section.
Dimensions and mass:
Exterior dimensions exclude protrusions, and are given in
order of width(W), height(H), and depth(D), in mm units.
Indicated weight represents an approximation of the mass
of the main unit only, not including case, accessories, etc.
About the marks ...
True RMS measuring capability for
accurate measurement of even distorted
waveforms.
New products in the 2004 Electrical
Measuring Instruments PRODUCT
CATALOG.
Models are available with interfaces compatible with LAN, GP-IB,
RS-232C, SCSI, FAX/modem and USB standards.
Measurement categories (Overvoltage categories)
To ensure safe operation of measurement products, IEC 61010 establishes safety standards for various electrical
environments, categorized as CAT I to CAT IV, and called measurement categories. These are defined as follows.
CAT I : Secondary electrical circuits connected to an AC electrical outlet through a transformer or similar device.
CAT II : Primary electrical circuits in equipment connected to an AC electrical outlet by a power cord (portable
tools, household appliances, etc.)
CAT III : Primary electrical circuits of heavy equipment (fixed installations) connected directly to the distribution
panel, and feeders from the distribution panel to outlets.
CAT IV : The circuit from the service drop to the service entrance, and to the power meter and primary overcurrent
protection device (distribution panel).
Higher-numbered categories correspond to electrical environments with greater momentary energy, so a
measurement product designed for CAT III environments can endure greater momentary energy than one designed
for CAT II. Using a measurement product
in an environment designated with a
higher-numbered category than that for
which the product is rated could result in
Distribution Panel
a severe accident, and must be carefully
Service Entrance
avoided.
Internal Wiring
Service Drop
Never use a CAT I measuring product in
CAT II, III, or IV environments.
T
The measurement categories comply
Outlet
with the Overvoltage Categories of the
Power Meter
IEC60664 Standards.
Fixed Installation
What is the CE Mark?
The CE mark certifies that a product
complies with electrical safety standards
established by European Community
directives (EC directives). These EC directives require
conformance of a product to EN/IEC standards for
electrical safety.
•HIOKI's products bearing the CE Mark are designed to
confirm to the Low Voltage and EMC directives based
on the EC directives.
•The Low Voltage directive is applicable to products
operating from 50 to 1000V AC and 75 to 1500V DC,
and require protection from electrical hazards such as
electric shock.
•The EMC directive requires suppression of emissions
of harmful electromagnetic radiation, and the ability
to withstand exposure to external electromagnetic
radiation without malfunction.
In some cases, power lines may carry
voltage spikes of several times the
normal supply voltage. For reasons of
safety, ordinary testers should not be
used to measure power lines carrying
more than 250V. When measuring such power lines,
always use a tester with built-in overcurrent protection
to guard against short circuits, such as Model 3008 and
CAT III marked products.
Note : An industrial power line refers to a high-capacity
supply circuit to equipment in factories or offices. A
high-capacity supply circuit refers generally to a line
carrying 20 A or more. This does not therefore include
supply lines protected by overcurrent protection (fuses) or
distribution breakers.
Notes on accuracy ...
The specifications in this catalog include figures for
"measurement accuracy" when referring to digital
measuring instruments, and for "measurement
tolerance" when referring to analog instruments.
The accuracy and tolerance figures in the product
specifications are defined in terms of full scale
(f.s.) value and displayed reading (rdg.) or digit
resolution (dgt.) as described below.
f.s. (maximum
full-scale)
display, or length of scale, ...
Signifies the maximum display (scale) value or the
length of the scale (in cases where the scale consists
of unequal increments or where the maximum value
cannot be defined). In general, this is the range
value (the value written on the range selector, or
equivalent) currently in use. However, be aware that
in cases where the maximum display value is 2000V
but the range value is only 600V, the maximum
display value (scale value) is still used as the f.s.
value.
1
rdg. (displayed or indicated value, ... reading value)
This signifies the value actually being measured, i.e.,
the value that is currently indicated or displayed by the
measuring instrument.
dgt. (digital resolution, ... digit)
Signifies the smallest display unit on a digital measuring
instrument, i.e., the value displayed when the last digit
on the digital display is "1". Essentially, this indicates
an error of 1 digit (based on decimal processing in
analog-to-digital conversion), but in actuality this is
the digit error combined with the f.s. error converted to
a fraction of a digit unit. The accuracy associated with
a particular measured value as shown in the product
specifications is derived from these values.
1. To avoid short circuits and electric
shock accidents when using a clamp-on
sensor, use only with power lines
carrying voltages within the rating limit
of the sensor.
2. To avoid short circuits and electric shock accidents
when the clamp-on sensor is open, do not use on bare
conductors.
Contents
About the Catalog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . P.1
Model Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . P.3
Recorders, Memory Recorders
P.4 to 15
Electronic Measuring Instruments P.16 to 26
Environmental Measuring Instruments P.27 to 31
Clamp Sensors
P.32 to 34
Power Measuring Instruments
P.35 to 39
Clamp Testers
P.40 to 45
Field Measuring Instruments
P.46 to 53
Options & Peripherals
P.54 to 60
Meter Relays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . P.54
Test Leads, Test Probes . . . . . . . . . . . . . . . . . . . . .P.55 to P.56
Recording Paper, Temperature Sensors . . . . . . . . . . . . . . P.57
All Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . .P.58 to P.60
Service, Traceability, Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . P.61
Company Guide & Internet Website . . . . . . . . . . . . . . . . . . . . . . . . . . . . P.62
2
Model Index
1000 or later ...Peripherals
2000 or later ...Meter relays
1195. . . . . . . p.58
1196. . . . . . . p.58
7000 or later ...Signal sources
7011. . . . . . .
7016. . . . . . .
7075. . . . . . .
7075-01 . . . . .
2103. . . . . . . p.54
2104. . . . . . . p.54
2300. . . . . . . p.28
p.26
p.26
p.26
p.26
3000 or later ...Multi meters, Current /Insulation /Earth /Rotation /Thermo /Illumination /Power meters
3008. . . . . . .
3030-10 . . . . .
3118-11, -12 . .
3120. . . . . . .
3126-01 . . . . .
3127-10 . . . . .
3128-10 . . . . .
3129. . . . . . .
3144-20 . . . . .
3151. . . . . . .
3153. . . . . . .
3154. . . . . . .
3156. . . . . . .
3157-01 . . . . .
3158. . . . . . .
3159. . . . . . .
3167. . . . . . .
3169-20,-21 . .
3173. . . . . . .
3186. . . . . . .
p.46
p.46
p.46
p.49
p.46
p.45
p.45
p.53
p.50
p.53
p.23
p.24
p.25
p.25
p.24
p.23
p.35
p.36
p.24
p.35
3187. . . . . . .
3193. . . . . . .
3194. . . . . . .
3196. . . . . . .
3237. . . . . . .
3238. . . . . . .
3239. . . . . . .
3244. . . . . . .
3245. . . . . . .
3246. . . . . . .
3255-50 . . . . .
3256-50,-51 . .
3257-50,-51 . .
3269. . . . . . .
3270. . . . . . .
3272. . . . . . .
3273-50 . . . . .
3274. . . . . . .
3275. . . . . . .
3276. . . . . . .
p.35
p.39
p.39
p.37
p.22
p.22
p.22
p.48
p.46
p.48
p.49
p.49
p.49
p.33
p.32
p.33
p.33
p.33
p.33
p.33
3280-10 . . . . .
3280-20 . . . . .
3281. . . . . . .
3282. . . . . . .
3283. . . . . . .
3284. . . . . . .
3285. . . . . . .
3286-20 . . . . .
3287. . . . . . .
3288. . . . . . .
3290. . . . . . .
3290-10 . . . . .
3331. . . . . . .
3332. . . . . . .
3403. . . . . . .
3404. . . . . . .
3412-50 . . . . .
3415-01 . . . . .
3416-01 . . . . .
3418. . . . . . .
3540-01,-02,-03 .
3541. . . . . . .
3550. . . . . . .
3551. . . . . . .
3555. . . . . . .
3560. . . . . . .
3632-20 . . . . .
3633-20 . . . . .
3634-20 . . . . .
3635-24,-25,-26 .
3636-20 . . . . .
3637-20 . . . . .
3638-20 . . . . .
3639-20 . . . . .
3640-20 . . . . .
3641-20 . . . . .
3645-20 . . . . .
3650. . . . . . .
3660. . . . . . .
3661-20 . . . . .
3423. . . . . . .
3441. . . . . . .
3442. . . . . . .
3443. . . . . . .
3444. . . . . . .
3445. . . . . . .
3446-01 . . . . .
3447-01 . . . . .
3451-11 to -15 . .
3452-11 to -13 . .
3453. . . . . . .
3454-10 . . . .
3454-11,-51 . .
3501. . . . . . .
3503/3504 . .
3511-50 . . . . .
3522-50 . . . . .
3532-50 . . . . .
3535. . . . . . .
3540. . . . . . .
p.31
p.30
p.30
p.30
p.30
p.30
p.27
p.27
p.52
p.52
p.51
p.52
p.52
p.16
p.16
p.18
p.20
p.20
p.19
p.17
8861. . . . . . . . p.5
8910. . . . . . . p.13
8936. . . . . . . . p.7
8937. . . . . . . . p.7
8938. . . . . . . . p.7
8939. . . . . . . . p.7
8940. . . . . . . . p.7
8946. . . . . . . . p.7
8947. . . . . . . .
8950. . . . . . . .
8951. . . . . . . .
8952. . . . . . . .
8953-10 . . . . . .
8954. . . . . . . .
8955. . . . . . . .
8956. . . . . . . .
p.7
p.9
p.9
p.9
p.9
p.9
p.9
p.5
8957. . . . . . . . p.5
8958. . . . . . . . p.5
8959. . . . . . . . p.5
8992. . . . . p.10,12
8993. . . . . . . p.12
8994. . . . . . . . p.9
8995. . . . . . . . p.5
9318. . . . . . . p.59
9319. . . . . . . p.59
9320. . . . . . . . p.8
9320-01 . . . . . . p.8
9321. . . . . . . . p.8
9321-01 . . . . . . p.9
9322. . . . . . . . p.8
9323. . . . . . . p.59
9324. . . . . . . . p.8
9325. . . . . . . . p.8
9326. . . . . . . p.56
9327. . . . . . . . p.9
9328. . . . . . . . p.9
9332. . . . . . . p.14
9333. . . . . . . p.14
9334. . . . . . . p.15
9335. . . . . . . p.15
9338-9340 . . p.59
9344-9349 . . p.59
9351. . . . . . . p.59
9355. . . . . . . p.59
9360. . . . . . . p.59
9363-9365 . . p.59
9371. . . . . . . p.59
9372. . . . . . . p.59
9375. . . . . . . p.59
9376. . . . . . . p.59
9377. . . . . . . p.59
9378. . . . . . . p.59
9380. . . . . . . p.59
9382. . . . . . . p.59
9384. . . . . . . p.59
9386. . . . . . . p.59
9388. . . . . . . p.59
9390. . . . . . . p.59
9391. . . . . . . p.59
9392. . . . . . . p.59
9393. . . . . . . p.59
9397. . . . . . . p.59
9398. . . . . . . p.59
9399. . . . . . . p.59
9418-10,-15 . . p.59
9420. . . . . . . p.59
9425. . . . . . . p.59
9433. . . . . . . p.59
9436. . . . . . . p.59
9438. . . . . . . p.56
9439. . . . . . . p.59
9440. . . . . . . p.59
9441. . . . . . . p.59
9442. . . . . . . p.59
9443-02 . . . . . p.59
9443-03 . . . . . p.59
9444. . . . . . . p.59
9445-02, -03 . . p.59
9446. . . . . . . p.59
9447. . . . . . . p.59
9451. . . . . . . p.59
9452. . . . . . . p.56
9453. . . . . . . p.56
9454. . . . . . . p.18
9455. . . . . . . p.56
9458. . . . . . . p.37
9459. . . . . . . p.37
9460. . . . . . . p.56
9461. . . . . . . p.56
9462. . . . . . . p.59
9463-9464 . . p.59
9465. . . . . . p. 56
9466. . . . . . p. 56
9467. . . . . . . p.56
9472-9479 . . p.57
9517. . . . . . . p.59
9518-01 . . . . . p.20
9540-01 . . . . . p.59
9549. . . . . . . . p.9
9555. . . . . . . p.34
9557. . . . . . . p.59
9558. . . . . . . p.59
9588. . . . . . . p.60
9593-01 . . . . . p.20
9598. . . . . . . p.60
9599. . . . . . . p.60
9600. . . . . . . p.39
9601. . . . . . . p.39
9602. . . . . . . p.39
9603. . . . . . . p.39
9604. . . . . . . p.39
9605. . . . . . . p.39
9612. . . . . . . p.10
9613. . . . . . . p.60
9614. . . . . . . p.60
9615. . . . . . . p.60
9617. . . . . . . p.60
9618. . . . . . . p.60
9624. . . . . . . p.37
9624-01 . . . . . p.37
9625. . . . . . . p.36
9626. . . . . . . p.60
9627. . . . . . . p.60
9631. . . . . . . p.60
9632. . . . . . . p.60
9633. . . . . . . p.60
9634. . . . . . . p.60
9635-9636 . . p.44
9637-9638 . . p.60
9639. . . . . . . p.60
9641. . . . . . . p.12
9642. . . . . . . p.12
9643. . . . . . . p.12
9645-01 . . . . . . p.9
9646. . . . . . . . p.9
9648. . . . . . . p.12
9649. . . . . . . p.12
9650. . . . . . . p.32
9651. . . . . . . p.32
9652. . . . . . . p.12
9653. . . . . . . p.12
9657. . . . . . . p.29
9657-10 . . . . . p.32
9658. . . . . . . p.29
9660-9661 . . p.32
9667. . . . . . . p.32
9668. . . . . . . p.32
9669. . . . . . . p.32
9663. . . . . . . . p.9
9670. . . . . . . p.37
9671. . . . . . . p.37
9675. . . . . . . p.32
p.42
p.42
p.43
p.43
p.43
p.43
p.43
p.44
p.42
p.42
p.45
p.45
p.35
p.39
p.31
p.31
p.27
p.27
p.27
p.27
p.17
p.17
p.21
p.21
p.21
p.18
p.29
p.29
p.29
p.29
p.29
p.29
p.29
p.29
p.29
p.29
p.29
p.27
p.50
p.50
3662-20 . . . . .
3663-20 . . . . .
3801-50 . . . . .
3802-50 . . . . .
3803. . . . . . .
3804. . . . . . .
3805. . . . . . .
3851-10 . . . . .
3852. . . . . . .
3853. . . . . . .
3854. . . . . . .
3909. . . . . . .
3911-20 . . . . .
3912-20 . . . . .
3930. . . . . . .
3931. . . . . . .
p.50
p.50
p.48
p.46
p.48
p.46
p.46
p.55
p.58
p.58
p.58
p.58
p.29
p.29
p.23
p.24
9677. . . . . . .
9678. . . . . . .
9680. . . . . . .
9686. . . . . . .
9694. . . . . . .
9699. . . . . . .
9700-10 . . . . .
9713. . . . . . .
9714. . . . . . .
9720. . . . . . .
9726. . . . . . .
9727. . . . . . .
9728. . . . . . .
ALLIGATOR CLIPs . .
CT-101A. . . .
CT-2MR . . . .
CT-5MRN . . .
HB-1 . . . . . .
HS-1 . . . . . .
HS-2 . . . . . .
220H . . . . . .
p.19
p.19
p.60
p.60
p.32
p.19
p.19
p.60
p.60
p.60
p.60
p.60
p.60
p.60
p.40
p.53
p.53
p.53
p.53
p.53
p.12
8000 or later ...Recorders and Peripherals
8205-10 . . . . . p.13
8206-10 . . . . . p.13
8420-51 . . . . . p.12
8421-51 . . . . . p.12
8422-51 . . . . . p.12
8715-01 . . . . . . p.4
8730-10 . . . . . p.11
8731-10 . . . . . p.11
8807-01 . . . . . p.10
8808-01 . . . . . p.10
8807-51 . . . . . p.10
8808-51 . . . . . p.10
8826. . . . . . . . p.6
8835-01 . . . . . . p.6
8855. . . . . . . . p.9
8860. . . . . . . . p.5
9000 or later ...Peripherals
9005-01 . . . . . p.32
9010. . . . . . . p.34
9010-10 . . . . . p.34
9014. . . . . . . p.55
9017. . . . . . . p.55
9018,-10 . . . . p.34
9021-01 . . . . . p.55
9032. . . . . . . p.31
9033. . . . . . . p.31
9050. . . . . . . p.53
9060. . . . . . . p.55
9067. . . . . . . p.55
9081. . . . . . . p.58
9084. . . . . . . p.58
9094. . . . . . . p.55
9132-10 . . . . . p.32
9140. . . . . . . p.55
9143. . . . . . . p.55
9148. . . . . . . p.58
9151-02 . . . . . p.58
9151-04 . . . . . p.58
9165. . . . . . . p.58
9168. . . . . . . p.55
9170. . . . . . . p.55
9177. . . . . . . p.55
9178. . . . . . . p.55
9180 to 83 . . p.57
9184. . . . . . . p.57
9185. . . . . . . p.55
9188. . . . . . . p.57
9190. . . . . . . p.55
9195. . . . . . . p.55
9196. . . . . . . p.55
9197. . . . . . . p.55
9198. . . . . . . p.55
9199. . . . . . . . p.8
9203. . . . . . . p.16
9207-10 . . . . . p.55
9208. . . . . . . p.55
9209. . . . . . . p.42
9211. . . . . . . p.31
9212. . . . . . . p.31
9213. . . . . . . p.31
3
9214. . . . . . .
9215. . . . . . .
9216. . . . . . .
9217. . . . . . .
9221. . . . . . .
9222. . . . . . .
9223. . . . . . .
9229. . . . . . .
9229-01 . . . . .
9231. . . . . . .
9232. . . . . . .
9233. . . . . . .
9234. . . . . . .
9235. . . . . . .
9236. . . . . . .
9237. . . . . . .
9257. . . . . . .
9261. . . . . . .
9262. . . . . . .
9263. . . . . . .
9267. . . . . . .
9268. . . . . . .
9269. . . . . . .
9270. . . . . . .
9271. . . . . . .
9272. . . . . . .
9274. . . . . . .
9276. . . . . . .
9277. . . . . . .
9278. . . . . . .
9279. . . . . . .
9287-10 . . . . .
9288. . . . . . .
9290-10 . . . . .
9292. . . . . . .
9293. . . . . . .
9294. . . . . . .
9296. . . . . . .
9297. . . . . . .
9299. . . . . . .
9303. . . . . . .
9305. . . . . . .
9315. . . . . . .
p.53
p.53
p.58
p.58
p.57
p.58
p.57
p.57
p.57
p.57
p.57
p.57
p.57
p.57
p.57
p.57
p.55
p.55
p.55
p.55
p.23
p.55
p.55
p.32
p.32
p.32
p.32
p.32
p.34
p.34
p.34
p.56
p.56
p.32
p.56
p.56
p.56
p.56
p.56
p.56
p.12
p.59
p.59
Recorders, Memory Recorders
Recorders, Memory Recorders Index
Handy units for ser vice and maintenance recording
Recorder (1ch)
100 sampling/sec.
No memory
. . . . . . . . . . . . . . p.13
8807-01
400 kS/sec.(2 ch)
256 kW (1 ch) memory
12 bits A/D resolution
Battery operation
. . . . . . . . . . . . . . p.10
8808-01
400 kS/sec.(4 ch)
256 kW (1 ch) memory
12 bits A/D resolution
Battery operation
. . . . . . . . . . . . . . p.10
8835-01
1 MS/sec.(8 ch)
4 MW (1 ch) memory
12 bits A/D resolution
. . . . . . . . . . . . . . . . p.6
8420-51/8421-51
Data logger (8 ch,16 ch)
100 ms to 1h interval
16 MW internal memory
16 bits A/D resolution
Battery operation
. . . . . . . . . . . . . . p.12
8422-51
Data logger (32 ch)
100 ms to 1h interval
16 MW internal memory
16 bits A/D resolution
Battery operation
. . . . . . . . . . . . . . p.12
WAVE
COMPARATOR
For simultaneous recording of multiple signals
8826
8860
8861
8730-10/8731-10
1 MS/sec. (32 ch)
4 MW (1 ch) memory
(expandable up to four)
12 bits A/D resolution
. . . . . . . . . . . . . . . . p.6
20 MS/sec. (12 bits 8 ch)
2 MS/sec. (16 bits 8 ch)
50 ms/all ch (16 bits 64 ch)
32 MW up to 1GW memory
. . . . . . . . . . . . . . . . p.5
20 MS/sec. (12 bits 16 ch)
2 MS/sec. (16 bits 16 ch)
50 ms/all ch (16 bits 128 ch)
64 MW up to 2GW memory
. . . . . . . . . . . . . . . . p.5
Geared for the
Production line
8730-10: 1ch input
8731-10: 2ch input
. . . . . . . . . . . . . . . p.11
For waveform capture
of high-speed signals
Recorders, Memory Recorders
8205-10
Options for MEMORY HiCORDER series
8855
8910
8 channels
20 MS/sec.(8 ch)
Max. 512 MW memory
12 bits A/D resolution
(16 bits A/D 1 MS/sec.)
. . . . . . . . . . . . . . . . p.9
CAN ADAPTER
..................................................
p.5,7,8,14,15
..............
p.13
For power line fault monitoring
8206-10
Recorder for power lines
100 sampling/sec.(2 ch)
AC voltage and current
No memory
. . . . . . . . . . . . . . p.13
8715-01
Recorder for power lines
400 kS/sec.(4 ch)
64 kW/ch memory
12 bits A/D resolution
Battery operation
8807-51
Harmonic analysis
400 kS/sec (2 ch)
256 kW (1 ch)Memory
12 bits A/D resolution
. . . . . . . . . . . . . . p.10
8808-51
Harmonic analysis
400 kS/sec (4 ch)
256 kW (1 ch)Memory
12 bits A/D resolution
. . . . . . . . . . . . . . p.10
4
8860 8861
MEMORY HiCORDER
HIOKI's Next Generation Recorder
High Performance Isolated Oscilloscope and Data Logger
All in One Complete Instrument
●High and Low Speed Dual Sampling - High speed at 20MS/s
(with 8956 input unit)
●Maximum 128 channels (8861) or 64 channels (8860) of data logging
●Large capacity memory, LAN/USB and other popular PC interfaces standard
●Intuitive operation using GUI/mouse/keyboard
STANDARD
STANDARD
Recorders, Memory Recorders
8861
8860
SPECIFICATIONS
Measurement ranges 5 mV to 20 V/division, 12 ranges (20 division f.s.),
using 8956 ANALOG UNIT /option resolution: 1/100 of range
DC to 10 MHz ±3 dB (using 8956 ANALOG UNIT /option)
Frequency band
Time axis
at memory function
Measurement
functions
Number of
input channels
Memory capacity
Data storage
5µs to 5 minutes/division, 26 settings;
external sampling (100 samples/division, desired setting)
Memory, Recorder,
Recorder & Memory(Version 2.00 or later),
FFT(Version 2.00 or later)
8860:Analog (up to 8 channels) + logic (16 channels standard)
or Logger Input (up to 64 channels)
8861:Analog (up to 16 channels) + logic (16 channels standard)
or Logger Input (up to 128 channels)
8860: 32 M words/9715×1 (Total 1 GW, 9715-03 (×1))
8861: 64 M words/9715×2 (Total 2 GW, 9715-03 (×2))
Floppy disk drive (optional, USB), Type II PC card slot, MO
drive (optional), Hard disk drive (optional)
USB1.1, LAN
OPTIONS
(The 8860 or 8861 cannot be used alone. Measurement requires optional input unit or similar peripheral.)
●Options (Factory fitted)
8995 A4 PRINTER UNIT
9715,-01,-02,-03 MEMORY BOARD(32MW to 1GW)*
9716 FD UNIT
9717 MO UNIT
9718 HD UNIT
9719 MEMORY BACK UP UNIT
●Options
9197 CONNECTION CORD (500V Max.)
9198 CONNECTION CORD (300V Max.)
9217 CONNECTION CORD (BNC-BNC)
9231 RECORDING PAPER (30m, 6rolls/1set)
9322 DIFFERENTIAL PROBE
9327 LOGIC PROBE
9320-01 LOGIC PROBE
9321-01 LOGIC PROBE
9723 CARRYING CASE (for 8860)
9724 CARRYING CASE (for 8861)
9642 LAN CABLE
9626 PC CARD 32MB
9627 PC CARD 64MB
9726 PC CARD 128MB
9727 PC CARD 256MB
9728 PC CARD 512MB
CLAMP ON SENSORs (refer to p.32-34)
Other common options (refer to p.8)
Interfaces
Recording and display 10.4-inch TFT color LCD, 216 mm × 30 m , thermal paper roll
Scaling, Vernier function, cursor measurement, comment
Other functions
insertion, other functions
100 to 240 V AC (50/60 Hz)
Power supply
Dimensions, mass
(base unit only)
Accessories
8860: 330 mmW×250 mmH×184.5 mmD, 8.0 kg
8861: 330 mmW×250 mmH×284.5 mmD, 10.5 kg
Power cord (1), input cord label (1), Application disk (1)
9716 FD UNIT
9715
MEMORY BOARD
9718
HD UNIT
*One MEMORY BOARD is required in Model 8860, and two MEMORY BOARDs
of the same capacity is required in Model 8861
8860
Input modules
8956
8957
8958
8959
8936
8937
8938
8939
8940
8946
8947
ANALOG UNIT
HIGH RESOLUTION UNIT
16ch SCANNER UNIT
DC/RMS UNIT
ANALOG UNIT
VOLTAGE/TEMP UNIT
FFT ANALOG UNIT
STRAIN UNIT
F/V UNIT
4ch ANALOG UNIT
CHARGE UNIT
8861
5
9717
MO UNIT
8835-01
MEMORY HiCORDER
High-visibility, Compact, Multi-channeled
Field measurement has never been easier
●Compact 4/8 ch recorder saves space with slim profile
●110mm-width recording paper and a large color display(6.4-inch)
●Maximum 8 analog channels and 16 logic
●Network Recording Instrument for LAN
SPECIFICATIONS
Measurement ranges
using 8936 ANALOG UNIT /option
Frequency band
Time axis at memory function
Functions
*Additional functions, using 9540-01
FUNCTION UP DISK /option
Number of input channels
using 8936, 8946 ANALOG UNIT /option
Memory capacity
Data storage
10mV to 50V/division, 12 ranges (10 division f.s.)
resolution; 1/160 of range
DC to 400kHz , ±3dB
100µs to 5 minutes/division, 20 ranges (1division =100samples)
Memory recorder, Recorder, RMS recorder (50/60Hz or
DC only), *Recorder and Memory, *FFT
OPTIONS
OPTION
OPTION
(The 8835-01 cannot be used alone. Measurement
requires optional input unit or similar peripheral.)
9221 RECORDING PAPER (30 m, 10 rolls /1 set)
9333 LAN COMMUNICATOR
9335 WAVE PROCESSOR
9388 CARRYING CASE
9439 DC POWER ADAPTER
(using between 10 to 28 V DC)
9540-01 FUNCTION UP DISK (to enhance the 8835-01)
9151-02/04 GP-IB CONNECTION CABLE (2m /4m)
CLAMP ON SENSORs (refer to p.32 - 34)
Other common options (refer to p.8)
OPTION
Recorders, Memory Recorders
4 analog channels plus 16 logic channels (using 8936)
8 analog channels plus 16 logic channels (using 8946)
12bits×4M words/channel (using 1 channel)
FDD×1, Type-III PC card×1;
9626/9627/9726/9727/9728 PC CARD
LAN, GP-IB, or RS-232C (option, using PC card)
Interfaces
Recording and display 110mm×30 m, roll type thermal paper,
Recording speed: 25mm/s, 6.4-inch color TFT LCD
Scaling, Waveform parameter calculations,
Other functions
Memory segmentation, Cursor readout, etc.
100 to 120V AC or 200 to 240V AC (50/60 Hz)
Power supply
* 10 to 28V DC, using the 9439 DC POWER ADAPTER
285W×220H×132D mm, 4.5kg
Dimensions, mass
Power cord(1), Recording paper(1 roll), Dust cover(1),
Accessories
PC card protector(1), Application disk(1)
*Note: An input cord is not supplied with the input unit. Requires the 9197 or 9198
8826
MEMORY HiCORDER
32-channel recorder with all isolated inputs
●Simultaneous sampling, display and recording of all 32 analog and 32
logic channels
●Large capacity memory of max. 16M-word
Memory expandable four times (option)
●High resolution of 12-bit, 1 M-sampling /second
●B4-size (paper width 264 mm) wide printer
●High-visibility waveforms displayed on a 10.4-inch color TFT liquid
crystal display
SPECIFICATIONS
Measurement ranges
5 mV to 20 V/division, 12 ranges
(normal f.s.; 20 division, wide f.s.; 24 division),
resolution: 1/80 of range
Frequency band
DC to 400 kHz, ±3dB
Time axis at memory function 100µs to 5 minutes/division, 20 ranges
(1 division =100 samples)
Functions
Memory recorder, Recorder (included X-Y),
RMS recorder, Recorder and Memory, FFT
Number of input channels 32 analog channels plus 32 logic channels
Memory capacity
(analog 12 bits)×4M words/channel (using 4ch)
* Expandable up to four, using 9599 MEMORY BOARD
Data storage
FDD×1, Type-III PC card×1; SRAM, flash ATA,
9626/9627/9726/9727/9728 PC CARD
Recording and display 264 mm×30 m, roll type thermal paper,
Recording speed: 25 mm/s, 10.4-inch color TFT LCD
Other functions
Scaling, Waveform judgment, Waveform processing
calculations, Waveform parameter calculations,
Memory segmentation, Logging print, Clock,
Cursor readout, Comment entry, etc.
Power supply
100 to 240 V AC, 50/60 Hz
Dimensions, mass
401W×235H×382D mm, 11 kg (excluding input units)
Accessories
Power cord(1), Recording paper(1 roll), Dust cover(1),
PC card protector(1), Application disk(1)
OPTIONS
(The 8826 cannot be used alone. Measurement requires optional input cord or similar
peripheral.)
9229 RECORDING PAPER (30 m, 6 rolls /1 set)
9229-01 RECORDING PAPER (perforated)
9598 MO UNIT...factory fitted only
9599 MEMORY BOARD (48 M words)
9151-02/04 GP-IB CONNECTION CABLE (2 m /4 m)
CLAMP ON SENSORs (refer to p.32 - 34)
Other common options (refer to p.8)
OPTION
STANDARD
OPTION
OPTION
*Note: An input cord is not supplied with the input unit. Requires the 9197 or 9198
6
INPUT Units
for 8835-01, 8841, 8842, 8860, 8861
9318
9318 CONVERSION CABLE
(to connect the clamp-on sensor to the 8940)
Dimensions and mass:
9319
Approx. 170W × 20H × 148D mm, approx. 290g
8936, 8938
8936 ANALOG UNIT
Input
Measurement range
Recorders, Memory Recorders
Maximum sampling rate
Accuracy
Zero-position
Frequency characteristics
Input resistance and capacitance
Input coupling
Max. allowable input
Max. grounding voltage
Accessories
Other functions
Accessories
(to connect the 3273 to the 8940)
Dimensions and mass:
Approx. 170W × 20H × 148D mm , approx. 300g
Number of channels: 2, Connector: Insulated BNC
* Input isolated from output, inter-channel isolation
5mV to 20V/ division, 12 ranges, full-scale (f.s.) = 20 divisions,
AC voltage for possible measurement/ display using the memory function
: 280V rms, Low-pass filter, 5/500/ 5k/ 100kHz, the measurement
resolution is 1/80 of range * When used with 8841, 8842
1 MS/s (simultaneous sampling of two channels)
DC amplitude: ±0.4%f.s. Zero-position: ±0.1%f.s.
-50% to 150%, 1 % step * With zero-adjustment function
DC to 400kHz ±3 dB, with AC coupling: 7Hz to 400kHz ±3dB
1MΩ, 30 pF approx. (at C 100kHz)
DC, GND, AC
400V DC (upper voltage which when applied to between input pins does not damage them)
370V AC, DC (upper voltage which when applied to input channel casing or
between input channels does not damage them)
None * Input cord optional
8938 FFT ANALOG UNIT
Anti-aliasing filter
9319 CONVERSION CABLE
Cutoff frequency 20, 40, 80, 200, 400, 800, 2k, 4k, 8k, 20k,
40kHz auto-select (linked to frequency range)
Same as the 8936 ANALOG UNIT
None *Input cord optional
8940
8940 F/V UNIT
Number of channels: 2*1, Voltage input: BNC terminal
*1 Input isolated from output, inter-channel isolation
Sensor connector terminal Number of channels: 2 (for current measurement)*2
*2 Models that allow unit insertion up to a total of 4 channels: 8841, 8842, 8720
Compatible current sensors 9270, 9271, 9272, 9277, 9278, 9279, 3273, 3273-50
Frequency: 0.05Hz to 5kHz/ division, 11 ranges, 5(r/min) to 500(r/min)/
Measurement range
division, 5 ranges, P50 Hz (40 to 60 Hz), P60 Hz (50 to 70 Hz)
Integration: 5 counts to 500 k counts/ division, Pulse duty ratio:
100 % f.s. Current: 5 mA to 100 A/ division, 10 ranges, linked to
use with type of the clamp-on sensor, Voltage: 0.5 mV to 2 V/
division, 12 ranges, Max. allowable input: 30 V rms or 60 V DC,
full-scale (f.s.) = 20 divisions, low-pass filter, 5/500/5k/ 100kHz
or OFF, the measurement resolution is 1/80*3 of range
*3 When used with 8841, 8842, and when used with 9279 CLAMP
ON SENSOR, the resolution is 1/64 of range
Max. sampling period 1µs (voltage, current)
Voltage input pull-up: ON (10kΩ)/ OFF
Other functions
Input coupling: DC, GND, AC (voltage, current), DC (others)
Max. grounding voltage 30V rms or 60V DC (upper voltage which when applied to input
channel casing or between input channels does not damage them)
None * Input cord and conversion cable optional
Accessories
Input
Dimensions and mass:
Approx. 170W × 20H × 148D mm , approx. 300g
8937
Dimensions and mass:
Approx. 170W × 20H × 148D mm , approx. 310g
8937 VOLTAGE/TEMPERATURE UNIT
8946
Number of channels: 2 each for voltage and temperature
* Input isolated from output, inter-channel isolation
Voltage input: isolated BNC, thermocouple input: plug-in terminal
Voltage measurement 500µV to 2V/ division; 12 settings, full-scale (f.s.) = 20 divisions,
low-pass filter: 5/500/5k/ 100kHz, the measurement resolution
range
is 1/80 of range * When used with 8841, 8842
10°C to 100°C/ division; 4 settings, full-scale (f.s.) = 20 divisions,
Temperature
low-pass filter: 5/ 500Hz, measurement resolution: 1/80 of range
measurement range
* When used with 8841, 8842
K: -200 to 1350°C, E: -200 to 800°C, J: -200 to 110°C,
Thermocouple range
T: -200 to 400°C, N: -200 to 1300°C, R: 0 to 1700°C,
S: 0 to 1700°C, B: 300 to 1800°C,
Reference junction compensation: internal/ external (switchable)
Voltage input: 1 MS/s, Temperature measurement: 4kS/s
Max. sampling rate
(2-channel simultaneous sampling)
Voltage input: DC amplitude ±0.4% of f.s. Zero-position ±0.15%
Accuracy
of f.s. Temperature measurement (K, E, J, T, N): ±0.1% of f.s.
±1°C, ±0.1% of f.s. ±2°C (-200 to 0°C), (R, S): ±0.1% of f.s.
±3°C, (B): ±0.1% of f.s. ±4°C (400 to 1800°C) Reference junction
compensation accuracy: ±0.1 % f.s. ±1.5°C (internal compensation)
Voltage input: -50% to 150%, 1% steps * With zero-adjust function
Zero position
Temperature measurement: -100% to 100%, 1% steps
Frequency characteristics Voltage input: DC to 400kHz + 1/-3dB
Temperature measurement: DC to 1 kHz + 1/-3dB
Voltage input: 1 MΩ, 50pF approx. (at C 100 kHz)
Input resistance and
Temperature measurement: 5.1MΩ
capacitance
DC, GND, AC
Input coupling
30V rms or 60V DC
Max. allowable input
(upper voltage which when applied to between input pins does not damage them)
Max. grounding voltage 30V rms or 60V DC (upper voltage which when applied to input
channel casing or between input channels does not damage them)
None * Input cord optional
Accessories
Inputs
Dimensions and mass:
Approx. 170W × 20H × 148D mm , approx. 250g
8939
8939 STRAIN UNIT
Inputs
Converter connector
Suitable converter
Measurement range
Maximum sampling rate
Accuracy (after auto-balancing)
Balancing
Zero position
Frequency characteristics
Max. allowable input
Max. grounding voltage
Accessories
7
Number of channels: 2, Connector: Adapter cable connector
* Input isolated from output, inter-channel isolation
Via adapter cable, TAJIMI PRC03-32A10-7M10.5
Strain gage converter, bridge impedance: 120Ω to 1kΩ,
gage factor 2.00, bridge voltage 2 ±0.05 V
20µe to 1000µe/ division; 6 settings, full-scale (f.s.) =
20 divisions, low-pass filter: 10 Hz, 30 Hz, 300 Hz, 3 kHz, OFF
the measurement resolution is 1/80 of range *Using 8841, 8842, 8720
1 MS/s (simultaneous sampling for 2 channels)
DC amplitude: ±(0.5 %f.s. + 2µe) Zero-position: ±0.5 %f.s.
Electronic auto-balancing, max. adjustment range ±10000µe
-50 % to 150 %; in 1% steps * With auto-balancing
DC to 20 kHz +1/-3 dB
10 V (DC + AC peak) (upper voltage which when applied to
between input pins does not damage them)
30 V rms or 60 V DC (upper voltage which when applied to input
channel casing or between input channels does not damage them)
Conversion cable (2), PRC 03-12A10-7M 10.5(2)
8946 4ch ANALOG UNIT
Inputs
Measurement range
Maximum sampling rate
Accuracy
Zero-position
Frequency characteristics
Input resistance and capacitance
Input coupling
Max. allowable input
Max. grounding voltage
Accessories
Number of channels: 4, Terminal: Metallic BNC
* Input isolated from output, inter-channel isolation
10 mV to 2 V/ division, 8 ranges, full-scale (f.s.) = 20 divisions,
low-pass filter, 5/ 500/ 5 k/ 50 kHz; the measurement resolution
is 1/80 of range * When used in 8841
1 MS/s (simultaneous sampling of four channels)
DC amplitude: ±0.5 % f.s. Zero-position: ±0.15 % f.s.
-50 % to 150 %, 1 % step * With zero-adjustment function
DC to 100 kHz ±3 dB
1 MΩ, 15 pF approx. (at C 100 kHz)
DC, GND
30 V rms or 60 V DC (upper voltage which when applied to
between input pins does not damage them)
30 V rms or 60 V DC (upper voltage which when applied to input
channel casing or between input channels does not damage them)
None * Input cord optional
Dimensions and mass:
Approx. 170W × 20H × 148D mm , approx. 310g
8947
8947 CHARGE UNIT
Number of channels: 2
Measurement objects can be selected individually for each channel.
Full isolation between inputs, and between inputs and recorder.
Common GND for voltage input and charge input channels.
Voltage and pre-amplifier internal inputs: BNC terminals
(With voltage input: input resistance, 1MΩ ; input capacitance, less than 200 pF)
Charge input: miniature connector (#10-32 UNF)
Charge input: piezoelectric charge output acceleration
Suitable converters
pickup sensors, Internal pre-amplifier input: acceleration
pickup sensors with built-in pre-amplifier
50m (m/s2)/DIV to 10 k (m/s2)/DIV, 12 ranges×6 types, the
Measurement ranges
measurement resolution is 1/80 to 1/32 of range (changes according
Charge input
to measurement sensitivity) Measurement sensitivity: 0.1 to 10pC
(miniature connector)
/ (m/s2), Pre-amplifier internal input measurement sensitivity:
Pre-amplifier internal
0.1 to 10mV/ (m/s2), Amplitude accuracy: ±2 % f.s., Frequency
input (BNC terminal)
characteristics: 1 to 50 kHz +1/ -3 dB, Low-pass filter: 500 / 5kHz,
Pre-amplifier driving power supply: 2mA ±20%, +15V ±5%,
Highest input charge : ±500pC (high sensitivity side 6 ranges),
±50000 pC (low sensitivity side 6 ranges)
500µV to 2V/DIV, 12 ranges, the measurement resolution is 1/80
Measurement ranges
to 1/32 of range (changes according to measurement sensitivity)
Voltage input
DC amplitude accuracy: ±0.4% f.s., Frequency characteristics:
(BNC terminal)
DC to 400kHz +1/ -3 dB, Low-pass filter: 5 / 500 / 5k/ 100kHz,
Input coupling: DC, AC, GND, Max. allowable input: 30V rms or 60V DC
Maximum sampling rate 1 MS/s (simultaneous sampling of two channels)
Max. grounding voltage 30V rms or 60V DC (upper voltage which when applied to input
channel casing or between input channels does not damage them)
None * Input cord optional
Accessories
Input
Common options for 8800 series
MEMORY HiCORDER
*Designated products are not CE-Mark compliant
Note: Product names appearing herein are trademarks or registered trademarks of the various companies.
Logic Signal Measurement
Case
Instrument
9321(9321-01)
LOGIC PROBE
9320(9320-01)
LOGIC PROBE
9397-01 CARRYING CASE
4-channel isolated, on/off
detection of AC/DC voltage
4-channels, on/off detection of
voltage/contact signal
9349 CARRYING CASE
9626 PC CARD
9627 PC CARD
9726 PC CARD
9727 PC CARD
9728 PC CARD
(for 8842 only)
with casters for convenient
transportation.
32 MB
64 MB
128 MB
256 MB
512 MB
9608 MEMORY BOARD
Expands instrument memory by 4×
Must specify when ordering: not user installable.
(8841, 42 only)
DC Power Supply
9433 DC POWER
ADAPTER
9231 RECORDING PAPER
30 m/ 98.43 feet, 6 rolls / 1 set
Supplies operating
power in the range 10
to 28 V DC.
(8841 and 8842)
9607 MO UNIT
Installs on the bottom of the instrument. Must specify when ordering:
not user installable.
(8841, 42 only)
PC Communication
Input Modules
9557 RS-232C CARD
9558 GP-IB CARD
(compliance with the PCMCIA
Standard)
(compliance with the PCMCIA
Standard) * With a GP-IB cable,
cord length : 2m (6.6 feet)
Recorders, Memory Recorders
Storage Media
(for 8841 and 8855)
High-voltage input
9324 POWER CORD
for logic terminal
9333 LAN
COMMUNICATOR
9322 DIFFERENTIAL
PROBE
software required to use
LAN connection with
Windows 95/ 98/ Me,
WindowsNT 4.0 / XP
for up to 2 kV DC, 1 kV AC
9325 POWER CORD
for 8940 sensor terminal
LAN CARD (HIOKI-tested)
(compliance with the PCMCIA Standard)
Manufactured by third parties.
Various input modules
Install by inserting into the instrument
Can be replaced by user
8936 ANALOG UNIT
8937 VOLTAGE/TEMPERATURE UNIT
8938 FFT ANALOG UNIT
8939 STRAIN UNIT
8940 F/V UNIT
8946 4ch ANALOG UNIT
8947 CHARGE UNIT
9335 WAVE PROCESSOR
software required to convert the
binary file to CSV text file, to remote
control with Windows 95/ 98/ NT
4.0 /Me/ 2000/ XP
9197 CONNECTION
CORD
9198 CONNECTION
CORD
for high voltage (up to 500V)
for low voltage (up to 300V)
Current Measurement, other options
9555
9279
9318
9272
Input from 20 to 1000 A
40 Hz to 1 kHz for 0.2 V AC output.
BNC terminal
9270
Enables observation of distorted AC current
waveforms. Input: up to 20 A, 5 to 50 kHz for
2 VAC out
*9271 CLAMP ON SENSOR
Banana-to-BNC, use to connect
to BNC terminal on Input
Module
For 100V/15A, convenient for
measuring 100 VAC line current
with clamp-on probe
9217 CONNECTION CORD
*9165 CONNECTION CORD
Insulation BNC-to-insulation
BNC, use to connect to
insulation-BNC terminal on
Input Module
Metal BNC-to-metal BNC,
use to connect to metal-BNC
terminal on Input Module
Insulation transformer, 400 V or
200 V AC input, 10 V AC output,
for AC power line measurement.
9271
*9270 CLAMP ON SENSOR
9199 CONVERSION ADAPTER *CT-101A LINE SPLITTER
*9303 PT
9278
9277
9018-10 CLAMP ON PROBE *9132-10 CLAMP ON PROBE
Input from 10 to 500 A
40 Hz to 3 kHz for 0.2 V AC
output. BNC terminal
9555
Enables observation of distorted AC current
waveforms. Input: up to 200 A, 5 to 50 kHz
for 2 VAC out
Paper width 70 (2.75") − 220
(8.66") mm
9277 UNIVERSAL CLAMP ON CT
Observe waveforms from DC to distorted AC.
Input up to 20 A, DC to 100 kHz for 2 VAC
out
3269/3272 POWER SUPPLY
9278 UNIVERSAL CLAMP ON CT
Observe waveforms from DC to distorted AC.
Input up to 200 A, DC to 100 kHz for 2 VAC
out
*9272 CLAMP ON SENSOR
*9279 UNIVERSAL CLAMP ON CT
Enables observation of distorted AC current
waveforms. Input selectable 20/200 A, 5 to 10
kHz for 2 VAC out
Note : Can only be used in combination with the
9555 SENSOR UNIT or 8940 F/V UNIT.
Observe waveforms from DC to distorted AC.
Input up to 500 A, DC to 20 kHz for 2 VAC
out
Note : Can only be used in combination with the
9555 SENSOR UNIT or 8940 F/V UNIT.
9318 CONVERSION CABLE
*9555 SENSOR UNIT
Connects 9270 to 9272, 9277 to 9279
clamp-on sensors to 8940 F/V UNIT.
220H
*220H PAPER WINDER
Used together with 9270 to 9272, 9277 to
9279 clamp-on sensors. Power supply unit.
9319 CONVERSION CABLE
3274
3273-50
Connects 3273, 3273-50 to
8940 F/V UNIT.
3273-50/3276 CLAMP ON PROBE
Wide (DC to 50 MHz/100 MHz) range, mA-level to 15 A
peak current. Requires power from 3272 or 3269 or 8940
F/V UNIT.
3274/3275 CLAMP ON PROBE
Wide (DC to 10 MHz) range, mA-level to 500 A rms
current.
Requires power from 3272 or 3269 only and requires
scaling operations
8
8855
MEMORY HiCORDER
8ch high-speed isolated inputs of 20MS/s, Max.
512 M words long storage memory
●Maximum 8 analog channels and 16 logic channels
●20MS/s, 8ch 12-bit high-speed isolated Input (8950/8951/8952)
●1MS/s, 8ch 16-bit HIGH RESOLUTION Input (8953-10)
●Max. 512 M words long storage memory
●Zoom and scroll functions providing enlarged and compressed
displays
●Standard LAN/SCSI interface
SPECIFICATIONS
Measurement ranges
using 8950 ANALOG UNIT /option
Frequency band
Time axis at memory function
Functions
Recorders, Memory Recorders
5 mV to 20 V/division, 12 ranges (20 division f.s.)
resolution: 1/100 of range
DC to 10 MHz ±3dB Sampling speed max.20MS/s
5µs to 5 minutes/division, 24 ranges (1 division =100 samples)
Memory, Recorder
(Rec & Memory, FFT/function in ver.2.00 or later)
Number of input channels 8 analog channels plus 16 logic channels
(analog inputs are isolated up to 370V)
Memory capacity
4M words/channel (Total 32MW)
Max.64M words/channel (Total 512MW)(Option)
Data storage
FDD×1, Type-II PC card×1; 9626-9728 PC CARD
(MO or HDD×1/option)
Interfaces
LAN, SCSI, GP-IB or RS-232C (Option, use PC card)
Display and recording 10.4 inch TFT color LCD
(option; 216mm×30m, roll type thermal paper)
Other functions
Scaling, Waveform judgment,
Waveform processing calculations,
Waveform parameter calculations,
Memory segmentation,Waveform zoom display,
Clock, Cursor readout, Comment entry, etc.
Power supply
100 to 240V AC 50/60 Hz Max. 180VA
Dimensions, mass
275W×285H×170D mm, 6.3kg
Accessories
Power cord(1), PC card protector(1)
Input cord label(1), Application disk(1)
OPTIONS
(The 8855 cannot be used alone. Measurement requires optional input unit or similar
peripheral.)
●Options (Factory fitted)
8994 PRINTER UNIT
9645 MEMORY BOARD (96MW)
9645-01 MEMORY BOARD (512MW)
9646 MO UNIT (1.3GB/640MB~128MB)
9663 HD UNIT (20GB)
●Options
8950 ANALOG UNIT
8951 VOLTAGE/CURRENT UNIT
8952 DC/RMS UNIT
8953-10 HIGH RESOLUTION UNIT
8954 VOLTAGE/TEMP UNIT
8955 F/V UNIT
9197 CONNECTION CORD (500V Max.)
9198 CONNECTION CORD (300V Max.)
9217 CONNECTION CORD (BNC-BNC)
9231 RECORDING PAPER (30m, 6 rolls
/1set)
9322 DIFFERENTIAL PROBE
9328 POWER CORD (for 8950/8953-9322)
9327 LOGIC PROBE
9321-01 LOGIC PROBE
9333 LAN COMMUNICATOR
9335 WAVE PROCESSOR
STANDARD
STANDARD
OPTION
9397-01 CARRYING CASE
9557 RS-232C CARD
9558 GP-IB CARD
9549 FUNCTION UP DISK
9642 LAN CABLE
9626 PC CARD 32 MB
9627 PC CARD 64 MB
9726 PC CARD 128 MB
9727 PC CARD 256 MB
9728 PC CARD 512 MB
CLAMP ON SENSORs (refer to p.32-34)
Other common options (refer to p.8)
OPTION
Input modules
Logic input
Install or replace simply by inserting
the module into the base unit
Note: Input cords are not provided.
Please purchase the appropriate
input cord for the probe type and
application separately.
9646 MO UNIT
Specify upon order,
cannot be installed by
user.
8950 ANALOG UNIT
8951 VOLTAGE/CURRENT UNIT
8952 DC/RMS UNIT
8953-10 HIGH RESOLUTION UNIT
8954 VOLTAGE/TEMP UNIT
8955 F/V UNIT
9327 LOGIC PROBE
4 channels, ON/OFF detection of
voltage/contact signals (high-speed type
for use with the 8855)
9549 FUNCTION UP DISK
Voltage, current, and power
waveforms on the secondary side
of an inverter
9645 MEMORY BOARD
(total 128 megawords)
Expands instrument memory by 4
times its original size. Specify upon
order; factory installation only.
9645-01 MEMORY BOARD
(total 512 megawords)
Expands instrument memory by 16
times its original size. Specify upon
order; factory installation only.
9
9663 HD UNIT
Specify upon order; factory
installation only. (20 GB)
Note: The MO and HDD
UNITs cannot be
mounted on the unit at
the same time.
By installing the power monitor
function in the 8855 MEMORY
HICORDER,you can monitor power
transient waveforms and view power
trend graphs. Use of this function
requires the optional 9549 FUNCTION
UP DISK,which is sold separately
9321-01 LOGIC PROBE
4 isolated channels, ON/OFF detection
of AC/DC voltage (small terminal-type
for use with the 8855, 8807-01, and
8808-01)
9323 ADAPTER CABLE
(Terminal conversion cable for
connecting the all-purpose 9321 LOGIC
PROBE and the 8855 when terminals do
not match.)
8807-01 8808-01
MEMORY HiCORDER
New concept incorporating detachable printer,
B5-sized handy recorder
●B5 book-sized, compact, and handy high-speed recorders
●2 analog channels (8807-01) 4 analog channels (8808-01) isolated
inputs (with 8 logic)
●PC card slot, 3-way power supply, and powerful trigger functions
●Fax/modem communication function (9332)
SPECIFICATIONS
Measurement ranges
OPTIONS
(The 8807-01 & 8808-01 cannot be used alone. Measurement requires optional INPUT
CORD or similar peripheral.)
8992
8808-01
Recorders, Memory Recorders
10 mV to 100 V*1/division, 13 ranges (10 division f.s.), resolution:
1/160 of range, *1 100 V/division excludes the RMS & memory
recorder functions
Frequency band
DC to 50 kHz, ±3 dB
Time axis at memory function 200µs to 5 minutes/division, 19 ranges (1 division =80 samples)
Functions
Memory recorder, Recorder, RMS & memory
recorder (50/60 Hz or DC only)
Number of input channels 8807-01: fixed input section, 2 analog*2 +8 logic
8808-01: fixed input section, 4 analog*2 +8 logic
*2 analog inputs are isolated up to 450V
Memory capacity
(analog 12 bits+ logic 4 bits)×256 k words*/channel * using CH1
Data storage
PCMCIA Type-II PC card ×1
9626-9728 PC CARD
Interfaces
RS-232C, Printer (8992 PRINTER UNIT can be connected)
Recording and display 112 mm×18 m, roll type thermal paper, Recording speed: 10 mm/s
(using AC adapter), 5 mm/s (using batteries), 5.7-inch STN color LCD
Power supply
9418-15 AC ADAPTER, LR6 (AA)×6 (Continuous use 1 hour,
LR6 batteries cannot be used with 8992 PRINTER UNIT),
9447 BATTERY PACK (Continuous use 3 hours)
Dimensions, mass
203W×170H×52D mm (printer detached)
280W×170H×52D mm (printer attached)
8807-01: 1.1 kg (printer detached), 1.5 kg (printer attached)
8808-01: 1.2 kg (printer detached), 1.6 kg (printer attached)
Accessories
LR6 (AA) Alkaline batteries(6), Alkaline battery
box(1), Shoulder belt(1), Application disk(1)
8992 PRINTER UNIT (print size 100 mm width)
9234 RECORDING PAPER (18m, 10 rolls /1 set)
9320-01 LOGIC PROBE (refer to p.8)
9321-01 LOGIC PROBE (refer to p.8)
9323 CONVERSION CABLE
9391 CARRYING CASE
9418-15 AC ADAPTER (universal 100 to 240VAC, 12VDC/2.5A output)
9447 BATTERY PACK (7.2V, 2400 mAh, recharging with the 9418-15)
9612 RS-232C CABLE (mini DIN 9-pin to Dsub 9-pin, 1.5 m)
CLAMP ON SENSORs (refer to p.32~34)
9332 WAVE COMMUNICATOR
9335 WAVE PROCESSOR
9643 CHARGE STAND
Other common options (refer to p.8)
*Note: An input cord is not supplied with the 8807-01 & 8808-01.
Requires the 9197 or 9198
8992+8807-01
STANDARD
OPTION
8807-51 8808-51
MEMORY HiCORDER
Instantaneous Analysis and Long-term Recording of Harmonic
Waves for Maintenance of Commercial Power Systems
Waveform
Bar graph
Numeric data
Power vector
10
8730-10 8731-10
WAVE COMPARATOR
A MEMORY HiCORDER geared for the
production line
●Easy installation into production lines for high speed measurement
and assessment
●Compare the signals of manufactured components and devices with
a memorized reference signal to conduct PASS/FAIL evaluations on a
waveform level
●8730-10: 1ch input/comparison 8731-10: 2ch input/2ch
simultaneous comparison
●Connect to a PC via LAN for networking capabilities
Recorders, Memory Recorders
OPTIONS
9333 LAN COMMUNICATOR
9335 WAVE PROCESSOR
9626 PC CARD 32M
9627 PC CARD 64M
STANDARD
9726 PC CARD 128M
9727 PC CARD 256M
9728 PC CARD 512M
STANDARD
Rack-mount the 8730-10 or 8731-10 on
existing facilities.
11
SPECIFICATIONS
Measurement ranges 100 mV to 5 V/division, 6 ranges (10 division f.s.)
resolution: 1/160 of range
Frequency characteristic DC to 400 kHz ±3 dB
8730-10: Analog 1 channel 8731-10: Analog 2 channels
Number of input
channels
12bits×50 k words/ch
Memory capacity
100µs to 5 minutes/division, 20 settings (1 division =100 samples)
Time axis
external sampling (up to 1 kHz, minimum sampling period 1 ms)
Measurement functions Memory recorder with Wave Comparator
PC card Type II slot
Data storage
RS-232C LAN (10BASE-T) External I/O
Interfaces
7.2-inch STN color LCD
Display
100 to 240 V AC (50/60 Hz)
Power supply
8730-10:288W×144H×190D mm, 3.6kg
Dimensions, mass
8731-10:288W×144H×190D mm, 3.7kg
Power cord(1), Application disk(1)
Accessories
Set up the instruments on the Control Box.
Bring the unit to the local testing site for
ultimate portability and convenience.
8420-51 21-51 22-51
MEMORY HiLOGGER
All analog inputs fully isolated!
Voltage/temperature loggers with PC networking support
●8420-51(8ch), 8421-51(16ch), 8422-51(32ch) All analog channels isolated.
●Temperature, voltage, humidity, cumulative pulses, rotations
●Max. 32-channel, 100ms high-speed multi-channel data collection
●Simultaneous numerical and graphical display
●Attachable Printer and I/O Module
●Up to 16 HiLOGGERs can be connected
and controlled via LAN.
Recorders, Memory Recorders
8421-51
8420-51/8421-51 SPECIFICATIONS
8420-51:Analog 8 channels isolated by Photo-MOS relays
Input Channels
8421-51:Analog 16 channels isolated by Photo-MOS relays
Pulse Inputs 4ch, Logic Inputs 16ch (using the 8993)
Measurement objects
Thermocouple : K,E,J,T,N,W(WRe5-26),R,S,and B
Platinum measurement resistance type : Pt-100 JPt-100
Analog voltage input : 100mV,1V,10V,100V and 1-5V f.s.
Recording interval
100ms to 1hour (5 sec. to 1 hour for humidity measurement)
A/D resolution
16 bits
Memory capacity
Internal : 16MW DRAM (32MB)
External : 9626-9728 PC CARD
Interfaces
RS-232C, LAN (10Base-T Ethernet connectors)
Printer (8992), DIGITAL I/O UNIT (8993)
Display and Recording 5.7-inch STN color LCD, 8992 Printer (Option)
112 mm×18 m, roll type thermal paper,
Recording speed: 2 mm/s
Function
Waveform compression and magnification, Event search,
Waveform scroll, cursor measurement, scaling,
automatic save, start condition retention, setting save,
comment entry, automatic set up, numerical calculation
Power supply
9418-15 AC ADAPTER, 9447 BATTERY PACK
Dimensions, mass
234W×170H×52D mm 1.4kg (instrument only)
310W×170H×52D mm 1.7kg (with printer )
302W×170H×52D mm 1.7kg (with Digital I/O Unit )
Accessories
9418-15 AC ADAPTER(1),
Terminal Cover(1), Screwdriver(1), Application disk(1)
STANDARD
STANDARD
8422-51 SPECIFICATIONS
Input Channels
8422-51:Analog 32 channels isolated by Photo-MOS relays
Pulse Inputs 4ch, Logic Inputs 16ch (using the 8993)
Measurement objects
Thermocouple : K,E,J,T,N,W(WRe5-26),R,S,and B
Analog voltage input : 100mV,1V,10V,100V and 1-5V f.s.
Recording interval
A/D resolution
Memory capacity
100ms to 1hour (16ch) (200 ms to 1 hour)
16 bits
Internal : 16MW DRAM (32MB)
External : 9626-9728 PC CARD
Interfaces
RS-232C, LAN (10Base-T Ethernet connectors)
Printer (8992), DIGITAL I/O UNIT (8993)
Display and Recording 5.7-inch STN color LCD, 8992 Printer (Option)
112 mm×18 m, roll type thermal paper,
Recording speed: 2 mm/s
Function
Waveform compression and magnification, Event search,
Waveform scroll, cursor measurement, scaling,
automatic save, start condition retention, setting save,
comment entry, automatic set up, numerical calculation
Power supply
9418-15 AC ADAPTER, 9447 BATTERY PACK
Dimensions, mass
234W×170H×52D mm 1.4kg (instrument only)
310W×170H×52D mm 1.7kg (with printer )
302W×170H×52D mm 1.7kg (with Digital I/O Unit )
Accessories
9418-15 AC ADAPTER(1),
Terminal Cover(1), Screwdriver(1), Application disk(1)
OPTIONS
8992 PRINTER UNIT (print size 100 mm width)
8993 DIGITAL I/O UNIT
9234 RECORDING PAPER (18m, 10 rolls /1 set)
9334 LOGGER COMMUNICATOR
9418-15 AC ADAPTER(100 to 240 V AC, 12 V DC/2.5 A output)
9447 BATTERY PACK(7.2 V, 2400 mAh, recharging with the 9418-15)
9612 RS-232C CABLE(mini DIN 9-pin to Dsub 9-pin, 1.5 m)
9721 RS-232C CABLE(mini DIN 9-pin to Dsub 9-pin, 1.5 m)
9641 CONNECTION CORD
9642 LAN CABLE
9643 CHARGE STAND
9648 CARRYING CASE
9649 PROTECTIVE CASE (Basic water-resistance)
9652-01 FIXED STAND
9681 HUMIDITY SENSOR
9329 TERMINAL UNIT
*Note: An input cord is not supplied with the 8420-51, 8421-51 & 8422-51
12
8910
●Select CAN-Bus information arbitrarily from recorder or data
logger and convert them into analog/logic signals
●Up to 12ch analog output + 24bit logic output
●Record both CAN adapter analog output and actual analog
data (i.e. sensor output) simultaneously
●Choose desired data from a PC or Memory HiCORDER
CAN ADAPTER
Record and Analyze CAN-Bus Signals
SPECIFICATIONS
Input
Number of output channels
Output resolution
Output voltage
Response speed
Interface
Functions
Functions
Recorders, Memory Recorders
CAN-Bus interface 2-channel (Receive only)
Up to 12 analog channels and 6 logic channels - 24bit
16bit
-5 to 5V (Analog), 0 to 5V (Logic)
Can follow up to a 1ms CAN-Bus refresh rate (1kS/sec max.)
RS-232C (For data selection settings only)
(1) Settings of CAN-Bus defined data (Various
Function
parameter settings to capture required data from
specifications
CAN-Bus)
(2) CAN-Bus signal input port settings
(3) Output channel settings (Settings to determine
output channels for captured data), etc.
Power supply
Dimension and Mass
Accessories
(1) From the 8910 Setting Software (standardly
equipped), Items (1), (2) and (3) above can be set.
(2) From Memory HiCORDERs (Models 8826,
8841 and 8842) and the 8910 Main Unit, Item
(3) above can be set via the RS-232C interface or
external recording media (FD, PC card and MO)
(1) AC adapter (100 to 240Vac universal, 12V/2.5A dc output)
(2) 10 to 30VDC (Can be supplied from a cigarette lighter socket in
an automobile.)
(3) Supplied from CAN-Bus signal input connector (10 to 30VDC)
180W×50H×100D mm, 940g
AC Adapter (1), RS-232C Cable (1), 9713-01 CAN Cable (1),
CD-R (including 8910 Setting Software)
Setting
Methods
Throttle position
Revolution
Speed
Water temp.
etc.
CAN
Network
8841/884
RS-232C Cable
Logic 6ch-24bit
Analog 12ch
8910
Other analog signals Available
number of channels on Memory HiCORDER
8826
8205-10 8206-10
MICRO HiCORDER
Easy data recording as convenient as a simple
tester, yet with broad functionality
●Record voltage and current variations simply with full line-up of
optional clamp on sensors of up to 1000A
●Input levels can be monitored on the LCD like an analog display
●Built-in thermal printer for printing data such as time and amplitude axis
8205-10
8205-10: SPECIFICATIONS
DC /AC Voltage: 0.1V to 500V f.s. 12 ranges
Measurement ranges
AC Current: 10A to 1000A f.s. 7 ranges (using CLAMP ON SENSOR / option)
10ms
Sampling period
20Hz to 30kHz, ±3 dB
Frequency band
20cm/minute to 2cm/hour, 5 ranges
Paper feed speed
Number of input channels Voltage or Current, 1 channel
Voltage: ±2% f.s. Current: ±4% f.s.
Accuracy
(using 9651 CLAMP ON SENSOR / option)
100 to 240V AC (50/60 Hz) or 9.5 to 14V DC, 2 way
Power supply
250W×122H×93.5D mm, 1.2 kg
Dimensions, mass
Power cord(1), Recording paper(1 roll), 9257 CONNECTION
Accessories
CORD(1), 9344 CARRYING CASE(1)
8206-10: SPECIFICATIONS
AC Voltage: 100/ 200/ 500 V extended scale, 3 ranges
Measurement ranges
AC Current: 10A to 1000A f.s. 7 ranges (using CLAMP ON SENSOR / option)
10 ms
Sampling period
30Hz to 30kHz, ±3dB
Frequency band
60 cm/hour to 2 cm/hour, 5 ranges
Paper feed speed
Number of input channels Voltage and Current, 2 channels alternate recording
Voltage: ±2% f.s. Current: ±4 % f.s.
Accuracy
(using 9651 CLAMP ON SENSOR / option)
100 to 240 V AC (50/60 Hz) or 9.5 to 14 V DC, 2 way
Power supply
250W×122H×93.5D mm, 1.2kg
Dimensions, mass
Power cord(1), Recording paper(1 roll), 9257 CONNECTION
Accessories
CORD(1), 9344 CARRYING CASE(1)
8206-10
OPTIONS
9650
13
9651
9668
9650 CLAMP ON SENSOR
(for 8205-10, 8206-10, 100A f.s.,40Hz~1kHz, 3m length)
9651 CLAMP ON SENSOR
(for 8205-10, 8206-10, 500A f.s.,40Hz~1kHz, 3m length)
9668 CLAMP ON SENSOR
(for 8205-10, 8206-10, 1000A f.s.,40Hz~1kHz, 3m length)
9235 RECORDING PAPER (15m, 10rolls /1 set)
9236-01 RECORDING PAPER (Climate-resistant, 15m, 10rolls/1set)
9326 CONNECTION CORD (for 8205-10 only)
220H PAPER WINDER (refer to p.8)
9332
WAVE COMMUNICATOR (for 8807-01, 8808-01, 8807-51 and 8808-51)
Remotely measure and transfer measurement
results through a telephone line and RS-232C
●RS-232C and telephone line connections
●Waveform data transfer and remote settings
●Transfer data files from remote locations
●Display waveforms on a PC
●Data conversion and importing into spreadsheet programs
SPECIFICATIONS
Compatible Recorders
Supplied Media
PC Operating
Environment
PC-Side Modem
Recorder Compatible
Modem
Communication
Methods
PC Usage
Limitations
Functions
Bottom left screen shows measured data displayed on a PC.
(For use with 8808-01 data)
Bottom right screen shows MEMORY HiCORDER connection
settings and data file acquisition, displaying connection conditions and
a list of measurement files.
9333
Recorders, Memory Recorders
8807-01, 8808-01,8807-51, 8808-51 MEMORY HiCORDERSs
(Vers. 2.00 or later)
One CD-ROM or one CD-R
IBM PC/AT Compatible
(with at least 800×600 display resolution)
Windows 95(SP1 or later), 98 or NT4.0(SP3 or later)
Fax modem supported by Windows 95, 98, or NT4.0, 2000, XP
Fax modem card for analog public telephone lines
(only specified models that have been pre -confirmed for operability)
Standard telephone line or RS-232C
(cannot be used at the same time)
While this application is running, other applications that
use the modem or RS-232C cannot be used
Communications Functions / Trigger Acquisition
Functions / Waveform Display Functions, Saving
Converted Data / Connection Destination Registration
/ Setting File Creation Functions
LAN COMMUNICATOR (for 8720, 8826, 8835-01, 8841, 8842 and 8855)
Connecting MEMORY HiCORDERs to LANs and
high-speed transfer of waveform data to PCs
●High-speed communication with LAN connections
●Gathering and managing waveform data on PCs, and the waveform
data collection function
●Remote and PC operations
●Data conversion and importing into spreadsheet programs
SPECIFICATIONS
Compatible Recorders
Recorder Operating
Environment
Supplied Media
PC Operating
environment
Communication System
Functions
8826 MEMORY HiCORDER (Vers. 2.30 or later)
8835-01 MEMORY HiCORDER*1 (Vers. 1.10 or later)
8841, 8842 MEMORY HiCORDERs (Vers. 2.30 or later)
8855 MEMORY HiCORDER
8720 VISUAL HiCORDER (Vers. 2.00 or later)
8730-10, 8731-10 WAVE COMPARATOR
(*1) 8835 not compatible with the 9333.
Compatible PC Card: 9578 10BASE-T LAN CARD
Connector: 10BASE-T
CD-ROM or CD-R (1 pc.)
IBM PC/AT Compatible
(1024×768 or higher screen resolution is recommended
when using the remote control functions)
Windows 95(OSR2 or later), 98, or NT4.0, 2000, XP
(network functions installed with a TCP/IP environment)
Ethernet, TCP/IP
Remote Control Applications / Waveform Data
Acquisition Applications / Waveform Viewer /
GP-IB Command Functions
Images on right show an 8841 screen during measurements on a PC in
real-time. The waveform screen is on the left and the control panel is on the
right. In addition to displaying measured waveforms in real time, you can
also use the keys displayed on the panel in the same way as the keys on the
MEMORY HiCORDER.
14
9334
LOGGER COMMUNICATOR
(for 8420-51, 8421-51 and 8422-51)
Program that enables data collection using
Ethernet and data analysis in Windows
Recorders, Memory Recorders
9335
WAVE PROCESSOR
Display, Convert, Calculate and Print Waveforms on a PC
*Note: The use of MO discs, PC cards, and floppy disks and the availability of a LAN
connection depend on the specifications of the recorder model in question.
15
(for MEMORY HiCORDER)
SPECIFICATIONS
For low resistance measurement
Batter y Testers
3540, -01, -02, -03
3541
3560
9203
3550, 3551
3555
Testing source DC
100 ms response
16 times/sec. sampling
Comparator (buzzer only)
Wide measurement range
0.1µΩ (20m Ω range) to
110MΩ High speed and
High precision
Prints out measurement
data at fixed intervals
for 3227, 3540, 3560,
3550, 3551
. . . . . . . . . . . . . . . p.17
. . . . . . . . . . . . . . . p.17
Testing source AC 1kHz
50/60 times/sec. sampling
Comparator output, full remote
control, RS-232C included
GP-IB or Printer interface option
. . . . . . . . . . . . . . . p.18
For medium to
high-capacity lead-acid
storage battery: UPS and
similar applications
Check battery deterioration
. . . . . . . . . . . . . . . p.21
For compact storage
batteries: portable
telephones and similar
applications
Check battery deterioration
. . . . . . . . . . . . . . . p.21
3511-50
3522-50
3532-50
3535
3503/3504
3501
7 measurement items
Testing source frequency
120Hz, 1kHz
Comparator output, RS-232C
included, GP-IB option
. . . . . . . . . . . . . . . p.18
14 measurement items
Testing source frequency
DC, 1mHz to 100kHz
Comparator output,
GP-IB or RS-232C option
. . . . . . . . . . . . . . . p.20
14 measurement items
Testing source frequency
42Hz to 5MHz
Comparator output,
GP-IB or RS-232C option
. . . . . . . . . . . . . . . p.20
14 measurement items
Testing source frequency
100kHz to 120MHz
Comparator output
GP-IB and RS-232C
. . . . . . . . . . . . . . . p.19
C, D Testing
Testing source frequency
120Hz, 1kHz
Comparator output, RS-232C
included, GP-IB option
C Testing
Testing source signal
4V DC/ one time
Indicator type
Signal Sources, Waveform Generators DMMs
7011
DC signal source
Voltage, Current,
Thermoelectric power
Measurement function
. . . . . . . . . . . . . . . p.26
7016
7075 (4ch)
DC signal source Voltage,
7075-01 (2ch)
Current, Measurement function Arbitrary waveform genarator
Function generator, Sweep
sequence function
. . . . . . . . . . . . . . . p.26 . . . . . . . . . . . . . . . p.26
PORTABLE WITHSTANDING HIGH VOLTAGE CONTACT
VOLTAGE HiTESTER
CHECKER
3237/3238/3239 3173
3931
High speed DMM
199999 count display
Monitor the voltage
on test terminals for
3153/3159/3158/3173
. . . . . . . . . . . . . . . p.22
Electronic Measuring Instruments
Inductance, Capacitance, or Impedance Meters
Basic AC withstand
voltage tester
3kV/30VA
Economical and easy to
operate
. . . . . . . . . . . . . . . p.24
. . . . . . . . . . . . . . . p.24
Safety Standards Measuring Instruments
Insulation Test
Equipment
Leakage current of Protective ground
Medical Equipment Test Equipment
AC Withstanding
Voltage Test
Equipment
Insulation/
Withstanding Test
Equipment
Automatic Insulation/
Withstanding
Test Equipment
3154
3156
3158
3159
3153
Testing voltage 25 /50
/100 /250 /500 /1000 V
Comparator output
Timer function
. . . . . . . . . . . . . . . p.24
Leakage current for use
in testing electric and
medical appliances
RS-232C
. . . . . . . . . . . . . . . p.25
3157-01
Testing source AC
Protective ground
tester indispensable for
standard certification
. . . . . . . . . . . . . . . p.25
5 kV/500VA
2000MΩ/5kV
AC withstanding voltage Insulation and
of electric appliances
withstanding voltage
tester
. . . . . . . . . . . . . . . p.24 . . . . . . . . . . . . . . . p.23
9999MΩ/5 kV AC/DC
Automatic insulation
and withstanding voltage
tester
. . . . . . . . . . . . . . . p.23
16
3541
RESISTANCE HiTESTER
Measure from very low (µΩ) to very high (MΩ)
resistances with a single instrument
●Wide Measurement Range
●0.1µΩ (20 mΩ range) to 110.000 MΩ
●High Speed & High Precision Measurements
●As fast as 0.6 ms with 70 ppm precision (in the 2 kΩ to 110 kΩ
range)
●Two Types of Temperature Correction
●Correction by Pt sensor or Infrared Thermometer
●Equipped with EXT I/O, GP-IB and RS-232C interfaces
●Easily integrates into automated production lines
SPECIFICATIONS
Measurement
Accuracy
Functions
Interface
Power supply
Dimensions
Mass
Accessories
Four-terminal resistance measurement
0.1µΩ (20 mΩ range) to 110.000 MΩ
Low power four-terminal resistance measurement
10µΩ (2Ω range) to 2.00000 kΩ
Temperature measurement (Pt) -10.0 to 99.9°C
Temperature measurement (analog) 0 to 2V
70ppm of rdg+15ppm of f.s. (2kΩ.20kΩ range slowΩ)
Temperature correction, temperature conversion, self
calibration, measurement fault detection, overflow detection,
offset voltage compensation, average, statistical calculation,
key lock, save/load, comparator, BIN measurement
GP-IB, RS-232C, EXT-I/O
100 to 240 VAC 50/60 Hz
Approx. 215W×80H×295D mm (excluding projections)
Approx. 2.6 kg
9287-10 CLIP TYPE LEAD, 9451 TEMPERATURE
PROBE, Power Cord, EXT I/O Male Connector
OPTIONS
Electronic Measuring Instruments
9452 CLIP TYPE LEAD
9453 FOUR TERMINAL LEAD
9454 ZERO ADJUSTMENT BOARD
9455 PIN TYPE LEAD (for ultra precision)
9461 PIN TYPE LEAD
9465 PIN TYPE LEAD
9467 LARGE CLIP TYPE LEAD
9300 CONNECTION CABLE (for multipolar connectors)
9637 RS-232C CABLE (9pin-9pin/cross/1.8m)
9638 RS-232C CABLE (9pin-25pin/cross/1.8m)
9151-02 GP-IB CONNECTOR CABLE (2m)
9151-04 GP-IB CONNECTOR CABLE (4m)
9670 PRINTER
9671 AC ADAPTER (for 9670)
9672 BATTERY PACK (for 9670)
9673 BATTERY CHARGER (for 9672)
9237 RECORDING PAPER (80mm × 25m, 4 rolls)
STANDARD
9287-10
CLIP TYPE LEAD
STANDARD
3540 3540-01 3540-02 3540-03
Offers selectable manual measurement or system
application
●4-terminal method milli-ohmmeter (Fast 100-ms Response)
●Comparator function memorizes up to seven tables
●Temperature compensation function measures temperature and
calculates value relative to copper at 20°C
SPECIFICATIONS
Measurement ranges
and Accuracy
Measurement current
Max. applied
measurement voltage
Sampling speed
Response time
Display
Measurement method
Open-circuit terminal voltage
Digital input/output
(-01, -02 and -03 Ver. only)
Comparator functions
Interface
Power supply
Dimensions, mass
Accessories
mΩ HiTESTER
30 mΩ to 30 kΩ, 7 ranges, 3500 full digits
±0.1 % rdg. ±6 dgt. (30 mΩ, 3Ω range)
±0.1 % rdg. ±4 dgt. (300 mΩ, 30Ω to 30 kΩ range)
100 mA (30 mΩ, 300 mΩ range) to 10µA (3 kΩ, 30kΩ range)
3.5 mV DC (30 mΩ range) to 350 mV DC (30 kΩ range)
16 times /second (fast mode), 4 times /second (slow mode)
100 ms (fast mode), 300 ms (slow mode)
3500 full digits, Liquid Crystal Display
Four-terminal measurement
4.0 V Max. (30 mΩ to 30 kΩ all ranges)
TTL output BCD, or other inputs /outputs for external control
Setting: Upper and lower limit, or reference value and % for
resistance, Up to 7 tables Output: 3 levels (Hi, In, Lo),
Open-collector, LED display, beep sound
External printer (-02 only), RS-232C (-03 only)
LR6 (AA) or R6P (AA)×6, or 9445-02, -03 AC ADAPTER (9V, 1A)
215W×61H×213D mm, 900 g, 1 kg (-01,-02,-03)
9287-10 CLIP-TYPE LEAD(1), 9451 TEMPERATURE PROBE(1),
Fuse(1), Ferrite Clamp(2), External Connector Socket (-01 only)
OPTIONS
Model 3540-03
The 3540 is the low-price version without external control interfaces, for
manual measurement. The 3540-01 adds BCD output and external control,
the 3540-02 includes a printer interface and the 3540-03 includes an RS-232C
interface.
17
9203 DIGITAL PRINTER
9233 RECORDING PAPER (10 m, 10 rolls /1set)
9425 CONNECTION CORD (for 9203-3540-02, 2m)
9445-02 AC ADAPTER (100 to 240 VAC, 9 V/1A output, for USA)
9445-03 AC ADAPTER (100 to 240 VAC, 9 V/1A output, for EU)
9452 CLIP-TYPE LEAD
9453 FOUR-TERMINAL LEAD
*9455 PIN-TYPE LEAD
9460 CLIP-TYPE LEAD WITH TEMPERATURE SENSOR
9461 PIN-TYPE LEAD
9467 LARGE CLIP-TYPE LEAD
9637 RS-232C CABLE (9pin-9pin)
9638 RS-232C CABLE (9pin-25pin)
*Note: The 9455 probe is a precision instrument. Exercise appropriate care when
handling it.
3560
AC mΩ HiTESTER
For measurement requirements from contact
resistance to internal resistance and voltage of
batteries
SPECIFICATIONS
Measurement ranges
Accuracy
Measurement current
Max. applied
measurement voltage
Sampling speed
●Fast response time approximately 84 ms (60 Hz)
●Low-power resistance measurement
●Battery measurement
●High-resolution (1µΩ in the 30mΩ. range)
30 mΩ to 3 kΩ, 6 ranges, ±0.5 % rdg. ±8 dgt. (all ranges)
In case of MEDIUM: Add 3 dgt. to the above dgt. error.
FAST: ±0.5 % rdg. ±8 dgt. (30 mΩ), ±0.5 % rdg. ±6 dgt.
(other ranges), However, in case of FAST, the display counter
decreases 4 digits in all ranges.
7.4 mA (30 mΩ range) to 1.5µA (3 kΩ range)
60 V DC (AC input is not possible)
50 times /s (FAST) to 1.56 times /s (SLOW) :at 50 Hz mode
60 times /s (FAST) to 1.88 times /s (SLOW) :at 60 Hz mode
Display
31000 full digits (resistance), 50000 full digits (voltage),
Fluorescent tube.
Measurement method 1 kHz AC four-terminal measurement
Open-circuit terminal voltage 20mV peak max. (30 mΩ to 3 kΩ all ranges)
Comparator functions Setting: Upper and lower limit, Up to 30 tables
Output: 3 levels (Hi, In, Lo) or (Pass, Fail), open-collector,
display, beep sound
Interface
RS-232C (standard), GP-IB or External printer (option)
Power supply
100 to 240 V AC, 50/60 Hz
Dimensions, mass
215W×80H×320D mm, 2.1 kg
Accessories
9287-10 CLIP-TYPE LEAD(1), Power cord(1)
OPTION
3511-50
STANDARD
Electronic Measuring Instruments
OPTIONS
9203 DIGITAL PRINTER
9233 RECORDING PAPER (10 m, 10 rolls /1set)
9425 CONNECTION CORD (for 9203-3560, 2m)
9452 CLIP-TYPE LEAD
9453 FOUR-TERMINAL LEAD
9454 ZERO ADJUSTMENT BOARD (when 9461 or 9465 is used)
*1 9455 PIN-TYPE LEAD
9461 PIN-TYPE LEAD
9465 PIN-TYPE LEAD
9466 REMOTE CONTROL SWITCH
9467 LARGE CLIP-TYPE LEAD
*2 9588 GP-IB INTERFACE
9589 PRINTER INTERFACE
9151-02/04 GP-IB CONNECTION CABLE (2 m /4 m)
*1Note: The 9455 probe is a precision instrument. Exercise appropriate care
when handling it.
*2Note: Non-CE mark product
LCR HiTESTER
Compact & powerful dedicated LCR measurement
in 5m second timeframes
●High speed measurement : 5ms (1 kHz) or 13ms (120 Hz)
●High precision accuracy : ±0.08 %
●Built-in comparator
SPECIFICATIONS
Measurement parameters ❘Z❘, θ, C, L, D, Q, R
Measurement method Source : open teminal voltage 50mV, 500mV, 1Vrms (AC)
sense: voltage, AC
120 Hz or 1 kHz
Source frequency
❘Z❘, R : 10 mΩ to 200.00 MΩ (depending on condition)
Measurement ranges
θ : -90.00 to +90.00°, C : 0.940 pF to 999.99 mF,
L : 1.600 µH to 200.00 kH, D : 0.0001 to 1.9900,
Q : 0.85 to 999.99
Basic accuracy
❘Z❘ : ±0.08% rdg. , θ: ±0.05°
Measurement times
Fast : 5 msec. to Slow : 300 msec. (at 1 kHz)
Fast : 13 msec. to Slow : 400 msec. (at 120 Hz)
Display
99999 full digits, LED
Comparator functions Setting : Upper and lower limit, absolute value,
Output : 3 levels (Hi, In, Lo), Open-collector, Isolated
External printer
9442 (use with the 9443-02 or -03/9444)
Power supply
100 to 240 V AC, 50/60Hz
Dimensions, mass
210W×100H×168D mm, 2.5 kg
Accessories
Power cord(1), Fuse(1)
OPTIONS
(The 3511-50 cannot be used alone. Measurement requires optional test fixture or
probe.)
OPTION
STANDARD
9140 FOUR-TERMINAL PROBE (DC to 100 kHz)
9143 PINCHER PROBE (DC to 5 MHz)
9261 TEST FIXTURE (cable connection type, DC to 5 MHz)
9262 TEST FIXTURE (direct connection type, DC to 5 MHz)
9263 SMD TEST FIXTURE (direct connection type, DC to 5 MHz)
9268 DC BIAS VOLTAGE UNIT (± 40 V DC max.)
9269 DC BIAS CURRENT UNIT (± 2 A DC max.)
9165 CONNECTION CORD (for 9268/9269; BNC to BNC, 1.5 m)
9166 CONNECTION CORD (for 9268/9269; BNC to clip, 1.5 m)
9151-02/04 GP-IB CONNECTION CABLE (2 m /4 m)
9518-01 GP-IB INTERFACE
9442 PRINTER
9443-02 AC ADAPTER (for the 9442, EU), 9443-03 (USA)
9444 CONNECTION CABLE (for the 3511-50/9442)
1196 RECORDING PAPER (25 m, 10 rolls/1 set, for the 9442)
18
3535
LCR HiTESTER
High-speed LCR meter with up to 120MHz
sampling
●Wide range from 100kHz to 120MHz
●High speed LCR testing (6ms/sample)
●Removable head amplifier
●"Load compensation function" for comparing standard component
and providing compensation
Electronic Measuring Instruments
The head amplifier
can be removed from
the main chassis.
3535
SPECIFICATIONS
Measurement parameters ❘Z❘, ❘Y❘, Q, Rp, Rs(ESR), G, X, B, θ, Ls, Lp, Cs, Cp, D(tanδ)
1kΩ range
10kΩ range
100kΩ range
100Ω to 2kΩ
1kΩ to 20kΩ
10kΩ to 300kΩ
Z.R
Measurement Range:
Reference Value
0.66pF to 15.9µF 0.066pF to 1.59nF
4.4fF to 159pF
C
0.133nH to 3.18mH 1.33µH to 31.8mH 13.3µH to 477mH
L
-180.00° to 180.00°
θ
Range
100kHz to 120MHz
Resolution setting
4digits (when using front panel to make setting)
100.0kHz to 1.000MHz
100Hz steps
1.000MHz to 10.000MHz 1kHz steps
Measurement
Frequency
10.00MHz to 100.0MHz
10kHz steps
100.0MHz to 120.0MHz
100kHz steps
When using GP-IB or RS-232C interfaces, resolution is 1Hz.
Accuray
±0.005% max. against set value
5mV to 1V, max. 20mA (up to 10.000MHz)
Open Terminal Voltage (V)
and Constant Voltage (CV) Modes 5mV to 500mV, max. 10mA (from 10.01MHz)
Resolution
1mV steps
Measurement
Accuracy
±(5%+5mV)×(2+log f) (f in terms of MHz)
Levels
200µA to 20mA, max. 1V (up tp 10.00MHz)
Constant Current (CC)
Mode
200µA to 10mA, max. 0.5V (from 10.01 MHz)
Resolution
10µA steps
Accuracy
±(10%+50µA)×(2+log f) (f in terms of MHz)
Basic accuracy
Z: ±0.5% rdg.; θ: ±0.3°
Output impedance
50Ω ±10Ω (at 100kHz)
Voltage
0.000V to 1.000V
Monitor
Current
0.000mA to 20.0mA
Current (when set at V or CV) 0.20mA to 20.00mA
Limit
Voltage (when set a CC)
0.005V to 1.000V
Average
OFF, 2, 4, 8, 16, 32, 64
Internal trigger, External trigger
Trigger
Trigger delay
0.01s to 9.99s; 0.01s resolution
Available for two measurement parameters; percentage, ∆%, or
absolute value settings (for ∆%, the offset of the measurement
Comparator
value from the standard value is displayed)
Panel Memory and Load Maximum 30 sets
Measurement value and judgment result using comparator
Zoom Display
Number of Lines Displayed Can set at 3, 4, or 5; may differ depending on parameter
Hard copy of measurement value or screen
Printer
(requires 9442, 9444)
GP-IB, RS-232C, EXT..I/O (All standard)
Interfaces
Operating Environment 10 to 40°C, 80%rh max., no condensation
Storage Environment -10 to 55°C, 80%rh max., no condensation
100V to 240V AC, 50/60Hz Approx.50VA
Power Supply
Approx. 360W×130H×360D mm ; 8.3kg
Dimensions, mass
OPTIONS
9678 CONNECTION
CABLE
HEAD AMP UNIT
(Model 3535 cannot be used alone. Measurement requires optional head amp unit
and test fixture or Probe. )
9700-10 HEAD AMP UNIT
9151-02/04 GP-IB CONNECTION CABLE (2m/4m)
*9677 SMD TEST FIXTURE
9442 PRINTER
9699 SMD TEST FIXTURE
9443-02 AC ADAPTER (for the 9442, EU), 9443-03 (USA)
9444 CONNECTION CABLE (for the 3535/9442)
9678 CONNECTION CABLE
1196 RECORDING PAPER (25m, 10rolls/1set, for the 9442)
*Note: Non-CE mark product
STANDARD
STANDARD
9700-10 HEAD AMP UNIT
9678 CONNECTION CABLE
(2m)
9677 SMD TEST FIXTURE
9699 SMD TEST FIXTURE
Operating Frequency : DC to 120 MHz
Measurable Object Size : 3.5 ±0.5 mm
Dimensions :
Approx.103W ✕ 37.1H ✕ 47.6D mm
Mass : Approx. 135 g
Operating Frequency : DC to 120 MHz
Measurable Object Size : W ; 1.0 to 4.0 mm
H ; 1.5 mm
Dimensions :
Approx.100.5W ✕ 28.6H ✕ 40.0D mm
Mass : Approx. 125 g
9700-10+3535
19
3522-50
LCR HiTESTER
Better functionality and performance at a low cost
●High speed measurement of 5 ms LCR meter
●Higher frequency range (DC or 1 mHz to 100 kHz)
●Fourteen parameters measured (High resolution and high accuracy)
●DC resistance measurement
SPECIFICATIONS
Measurement parameters ❘Z❘, ❘Y❘, θ, Rp(DCR), Rs(ESR, DCR), G, X, B, Cp, Cs,
Lp, Ls, D(tanδ), and Q
Measurement method Source: constant current 10µ to 100 mA(AC/DC), or
constant voltage 10 mV to 5 V (AC/DC)
open teminal voltage
DC, or 1mHz to 100kHz
Source frequency
❘Z❘, ❘R❘, X: 10.00 mΩ to 200.00 MΩ (depending on condition)
Measurement ranges
θ: -180.00 to +180.00°, C: 0.3200 pF to 1.0000 F,
L: 16.000 nH to 750.00 kH, D: 0.00001 to 9.99999,
Q: 0.01 to 999.99, ❘Y❘, G, B: 5.0000 nS to 99.999 S
❘Z❘: ±0.08 % rdg. , θ: ±0.05°
Basic accuracy
Fast: 5 msec. to Slow2: 828 msec.
Measurement times
typical values for displaying Z
99999 full digits, LCD with backlight display
Display
Comparator functions Setting: Upper and lower limit, percentage, or
absolute value, Output: 3 levels (Hi, In, Lo), Open-collector,
Isolated
9442 (use with the 9443-02 or -03/9446/9593-01)
External printer
100 to 240 V AC, 50/60Hz
Power supply
313W×125H×290D mm, 4.5 kg
Dimensions, mass
Power cord(1), Fuse(1)
Accessories
OPTIONS
OPTION
OPTION
3532-50
9140 FOUR-TERMINAL PROBE (DC to 100 kHz)
9143 PINCHER PROBE (DC to 5 MHz)
9261 TEST FIXTURE (cable connection type, DC to 5 MHz)
9262 TEST FIXTURE (direct connection type, DC to 5 MHz)
9263 SMD TEST FIXTURE (direct connection type, DC to 5 MHz)
9268 DC BIAS VOLTAGE UNIT (± 40 V DC max.)
9269 DC BIAS CURRENT UNIT (± 2 A DC max.)
9165 CONNECTION CORD (for 9268/9269; BNC to BNC, 1.5 m)
9166 CONNECTION CORD (for 9268/9269; BNC to clip, 1.5 m)
9151-02/04 GP-IB CONNECTION CABLE (2 m /4 m)
9518-01 GP-IB INTERFACE
9593-01 RS-232C INTERFACE
9442 PRINTER
9443-02 AC ADAPTER (for the 9442, EU), 9443-03 (USA)
9446 CONNECTION CABLE (for the 3522-50 /9442)
1196 RECORDING PAPER (25 m, 10 rolls /1 set, for the 9442)
Electronic Measuring Instruments
(The 3522-50 cannot be used alone. Measurement requires optional test fixture or
probe.)
LCR HiTESTER
Impedance meter with a wide test frequency range
●Higher frequency range (42 Hz to 5 MHz)
●High speed measurement of 5 ms LCR meter
●Interactive touch panel operation
●Wide setting range for measurement voltage and current
SPECIFICATIONS
Measurement parameters ❘Z❘, ❘Y❘, θ, Rp, Rs(ESR), G, X, B, Cp, Cs, Lp, Ls,
D(tanδ), and Q
Measurement method Source: constant current 10µ to 100 mA (42 Hz to 1 MHz),
50µ to 20 mA (1 MHz to 5 MHz), or constant voltage 10 mV
to 5 V (42 Hz to 1 MHz), 50 mV to 1 V (1 MHz to 5 MHz)
open teminal voltage
Source frequency
42 Hz to 5 MHz
Measurement ranges
❘Z❘, R, X: 10.00 mΩ to 200.00 MΩ(depending on condition)
θ : -180.00 to +180.00°, C: 0.3200 pF to 370.00 mF,
L : 16.000 nH to 750.00 kH, D : 0.00001 to 9.99999,
Q : 0.01 to 999.99, Y,G,B : 5.0000 nS to 99.999 S
Basic accuracy
❘Z❘: ±0.08% rdg. , θ: ±0.05°
Measurement times
Fast : 5 msec. to Slow2 : 140 msec.
typical values for displaying ❘Z❘
Display
99999 full digits, LCD with backlight display
Comparator functions Setting : Upper and lower limit, percentage, or
absolute value, Output : 3 levels (Hi, In, Lo), Opencollector, Isolated
External printer
9442 (use with the 9443-02 or -03/9446/9593-01)
Power supply
100 to 240 V AC, 50/60Hz
Dimensions, mass
352W×124H×323D mm, 6.5kg
Accessories
Power cord(1), Fuse(1)
OPTIONS
(The 3532-50 cannot be used alone. Measurement requires optional test fixture or
probe.)
OPTION
OPTION
9140 FOUR-TERMINAL PROBE (DC to 100 kHz)
9143 PINCHER PROBE (DC to 5 MHz)
9261 TEST FIXTURE (cable connection type, DC to 5 MHz)
9262 TEST FIXTURE (direct connection type, DC to 5 MHz)
Note: Measurement ranges are limited when using the 9140, 9143
9263 SMD TEST FIXTURE (direct connection type, DC to 5 MHz)
9268 DC BIAS VOLTAGE UNIT (± 40 V DC max.)
9269 DC BIAS CURRENT UNIT (± 2 A DC max.)
9165 CONNECTION CORD (for 9268/9269; BNC to BNC, 1.5 m)
9166 CONNECTION CORD (for 9268/9269; BNC to clip, 1.5 m)
9151-02/04 GP-IB CONNECTION CABLE (2 m /4 m)
9518-01 GP-IB INTERFACE
9593-01 RS-232C INTERFACE
9442 PRINTER
9443-02 AC ADAPTER (for the 9442, EU), 9443-03 (USA)
9446 CONNECTION CABLE (for the 3532-50 /9442)
1196 RECORDING PAPER (25 m, 10 rolls /1 set, for the 9442)
20
3550 3551 3555
Instantaneous determination of battery
deterioration
●Model 3551: support for high-capacity batteries
●Model 3555: for compact storage batteries (portable telephones)
●Model 3550: for medium-capacity lead-acid storage batteries
●Three-rank rating of battery state: Pass, Warning, or Fail
BATTERY HiTESTER
3550, 3551: SPECIFICATIONS
Resistance Measurement 3550: 30 mΩ to 3Ω, 3 ranges, 10µΩ resolution max.
3551: 3 mΩ to 300 mΩ, 3 ranges, 1µΩ resolution max.
Voltage Measurement 3 or 30 V DC, 2 ranges, 1 mV resolution max.
Temperature Measurement -10 to 60°C, 1 range, 0.1°C resolution (platinum sensor)
Sampling rate
0.83 times/second
Comparator functions
Setting: Upper and lower limit for resistance, and lower limit for
voltage, Output: LED, beep
Other functions
Data memory, Printer interface (9203 or Centronics)
Power supply
3550: LR6(AA)×6, 3551: LR6(AA)×6, or 9418-10
Dimensions, mass
3550: 196W×130H×50D mm, 710 g (including batteries)
3551: 196W×130H×65D mm, 860 g (including batteries)
Accessories
3550: 9460 CLIP-TYPE LEAD WITH TEMPERATURE
SENSOR(1), 9382 CARRYING CASE(1), Dust cover(1), LR6(6)
3551: 9465 PIN-TYPE LEAD(1), 9466 REMOTE CONTROL
SWITCH(1), 9377 CARRYING CASE(1), Dust cover(1), LR6(6)
Electronic Measuring Instruments
3555: SPECIFICATIONS
Resistance Measurement 300 mΩ to 30Ω, 3 ranges, 100µΩ resolution max.
Voltage Measurement 3 or 30V DC, 2 ranges, 1 mV resolution max.
Sampling rate
1.25 times/second
Comparator functions
Setting: Upper and lower limit, for resistance, and lower
limit for voltage, Output: LED, beep
Power supply
LR6(AA), 6 pieces (Continuous use of 18 hours)
Dimensions, mass
196W×130H×50D mm, 680g (including batteries)
Accessories
9461 PIN-TYPE LEAD (1), LR6 (6)
3550
OPTIONS
9203 DIGITAL PRINTER (for the 3550, 3551)
9425 CONNECTION CORD
(for connecting the 3550 /3551 to the 9203 / 2 m length)
*2 9467 LARGE CLIP-TYPE LEAD
9233 RECORDING PAPER (10 m, 10 rolls/1 set, for 9203)
9418-10 AC ADAPTER (for the 3551 only)
*1 9455 PIN-TYPE LEAD (for the 3550, 3555)
9287-10 CLIP-TYPE LEAD (for the 3555)
9382 CARRYING CASE (for the 3555)
*1Note: The 9455 probe is a precision instrument. Exercise appropriate care when
handling it.
*2Note: Non-CE mark product
3551
3551
Checking the battery in an
Uninterruptible
Power Supply (UPS) without
shutting down
3555
21
3555
On-the-spot testing
of compact storage
batteries
-- portable
telephones and
similar applications
3238 3239
DIGITAL HiTESTER
High-accuracy, multi-functional model (3238)
A new DMM with 4-terminal resistance
measurement! (3239)
●Samples at rates of up to 300 samples/sec. (3.3 ms/ sample)
●Comparator function provides high-speed pass/fail evaluation
●Equipped with external input and output for sequence control
●Useful Save/Load function helps work go faster
●Interface supports full remote operation
●AC/DC current and frequency functions
CAT II 300V
CAT I 500V
3238-01
3239-01
STANDARD
3239
3237
High-accuracy Type
3238 DIGITAL HiTESTER
3239 DIGITAL HiTESTER
3238-01 DIGITAL HiTESTER (with GP-IB) 3239-01 DIGITAL HiTESTER (with GP-IB)
OPTIONS
9010 CLAMP ON PROBE (10/20/50/100/200/500 A AC)
9018 CLAMP ON PROBE (10/20/50/100/200/500 A AC)
9132 CLAMP ON PROBE (20/50/100/200/500/1000 A AC)
9637 RS-232C CABLE (9pin-9pin, Reverse type/1.8m)
9638 RS-232C CABLE (9pin-25pin, Reverse type/1.8m)
9151-02/04 GP-IB CONNECTION CABLE (2 m /4 m)
9442 PRINTER
9443-02 AC ADAPTER (for the 9442, EU), 9443-03 (USA)
9444 CONNECTION CABLE (for 9442 printer)
1196 RECORDING PAPER (25 m, 10 rolls/1 set, for the 9442)
Electronic Measuring Instruments
3238
3238, 3239 SPECIFICATIONS
200m/2/20/200/1000V(±0.01% rdg. ±2dgt./2V)
DC voltage (DC V)
2/20/200/750V(±0.1% rdg. ±100dgt./45 to 10kHz) True RMS
AC voltage (AC V)
200m/2A(±0.1% rdg. ±6dgt./200mA)
DC current (DC A)
200m/2A(±0.3% rdg. ±100dgt./200mA, 45 to 3kHz) True RMS
AC current (AC A)
100/1k/10k/100k/300kHz
Frequency
(±0.015% rdg. ±2dgt./10 to 300kHz )
200/2k/20k/200k/2000k/20M/100MΩ
Resistance (Ω)
(±0.02% rdg. ±2dgt./2k to 200kΩ)
2k/20k/200k/2000kΩ
Resistance (LPΩ)
(±0.02% rdg. ±6dgt./2k to 200kΩ)
Open terminal voltage 6V DC max.(Ω, Diode check)
0.45V DC max.(LPΩ, Continuity check)
A built-in buzzer sounds when the resistance value is
Continuity check
less than 50.00Ω.
FAST approx. 300 samples/s ,
Sampling rate
MEDIUM approx.
8 to 9 samples/s , SLOW approx. 1 sample/s
LED max. 199999 (999999 for frequency)
Display
Comparator, Average, Zero Adjust, Trigger
Ancillary functions
and the Save/Load functions
External input/output,RS-232C, GP-IB(Option-01)
Interface
AC 100V/120V/220V/240 V, (50/60Hz)
Power supply
Dimensions and mass approx. 215W×80H×265D mm, 2.6 kg
9170 TEST LEAD(1)
Accessories
DIGITAL HiTESTER
High-speed DMM (3.3ms/sample)
Minimizing tact time with sequence control at a truly affordable price
●Samples at rates of up to 300 samples/sec. (3.3ms/ sample)
●Comparator function provides high-speed pass/fail evaluation
●Equipped with external input and output for sequence control
●Useful Save/Load function helps work go faster
●Interface supports full remote operation
3237 SPECIFICATIONS
DC voltage (DC V)
200m/2/20/200/1000V(±0.025% rdg. ±2dgt./2V)
AC voltage (AC V)
2/20/200/750V(±0.2% rdg. ±100dgt./45 to 3kHz) True RMS
Resistance (Ω)
200/2k/20k/200k/2000k/20M/100MΩ
(±0.05% rdg. ±2dgt./2k to 2MΩ)
Resistance (LPΩ)
2k/20k/200k/2000kΩ
(±0.05% rdg. ±6dgt./2k to 200kΩ)
Open terminal voltage 6V DC max.(Ω, Diode check)
0.45V DC max.(LPΩ, Continuity check)
Continuity check
A built-in buzzer sounds when the resistance value is
less than 50.00Ω.
Sampling rate
FAST approx. 300 samples/s , MEDIUM approx.
8 to 9 samples/s , SLOW approx. 1 sample/s
Display
LED max. 199999
Ancillary functions
Comparator, Average, Zero Adjust, Trigger
and the Save/Load functions
Interface
External input/output,RS-232C, GP-IB(Option-01)
Power supply
AC 100V/120V/220V/240 V, (50/60Hz)
Dimensions and mass approx. 215W×80H×265D mm, 2.6 kg
Accessories
9170 TEST LEAD(1)
3237
Economically Priced Type
3237
DIGITAL HiTESTER
3237-01 DIGITAL HiTESTER (with GP-IB)
OPTIONS
9010 CLAMP ON PROBE (10/20/50/100/200/500 A AC)
9018 CLAMP ON PROBE (10/20/50/100/200/500 A AC)
9132 CLAMP ON PROBE (20/50/100/200/500/1000 A AC)
9637 RS-232C CABLE (9pin-9pin, Reverse type/1.8m)
9638 RS-232C CABLE (9pin-25pin, Reverse type/1.8m)
9151-02/04 GP-IB CONNECTION CABLE (2 m /4 m)
9442 PRINTER
9443-02 AC ADAPTER (for the 9442, EU), 9443-03 (USA)
9444 CONNECTION CABLE (for 9442 printer)
1196 RECORDING PAPER (25 m, 10 rolls/1 set, for the 9442)
3237-01
STANDARD
CAT II 300V
CAT I 500V
22
3153
AUTOMATIC INSULATION / WITHSTANDING HiTESTER
Programmable testing, full remote control
Automatic Insulation Withstanding Tester
●Insulation resistance test (DC50V~1200V), Withstanding voltage test
(AC/DC5000V), full remote control in series
●Programable testing
(Testing Programs 32 files, Testing points 50 steps/file)
●Accurate testing voltage generation by PWM control method
●3930 HIGH VOLTAGE SCANNER (Option)
SPECIFICATIONS
Withstanding test
Testing voltage
AC 0.2 to 5.00 kV 500VA (maximum 30 minutes)
DC 0.2 to 5.00 kV 50VA (continuous)
Voltage setting method Digital setting
Waveform
Sin wave
Frequency
50/60Hz DC
Measurement range
Current: 0.01 to 100.0 mA, ±(2% rdg. +5dgt.)
10mA/100mA(AC)
AC (Average value rectified, RMS display)
Insulation test
Testing voltage
DC50 to 1200V
Measurement range
0.1 to 9999MΩ, 4 ranges
Judgment function
Contents: UPPER-FAIL / PASS / LOWER-FAIL
(Digital setting window comparator method)
Timer section
Setting range: 0.3 to 999 seconds
Interfaces
EXT I/O, EXT SW, RS-232C, GP-IB
Display
Fluorescent tube display (digital), Analog meter
Monitor function
Output voltage, detection current, Insulation resistance
Power supply
AC100-120V/AC200-240V, (50/60 Hz), 1000VA max.
Dimensions, mass
320W×155H×480D mm, 18 kg
Accessories
9615 H.V. TEST LEAD (high voltage side and return, 1 each),
Power cord (1), spare fuse (1)
Electronic Measuring Instruments
OPTIONS
9613 REMOTE CONTROL BOX (single)
9614 REMOTE CONTROL BOX (dual)
3930 HIGH VOLTAGE SCANNER
9267 SAFETY TEST DATA MANAGEMENT
SOFTWARE
9637 RS-232C CABLE
(Dsub 9pin-9pin, cross,1.8m)
9638 RS-232C CABLE
(Dsub 9pin-25pin,cross,1.8m)
3930
STANDARD
3159
STANDARD
INSULATION / WITHSTANDING HiTESTER
Perform insulation resistance and withstand
voltage testing in a single series
●Insulation resistance test (DC500V/1000V)
●Withstanding voltage test (AC5000V)
●Testing in series
(Insulation resistance test to Withstanding voltage test )
●Standard Interfaces (EXT I/O, EXT SW, RS-232C, STATUS OUT)
SPECIFICATIONS
Withstanding test
Testing voltage
0 to 2.5 kV / 0 to 5.0 kV AC, dual-range configuration
(Average value rectified, effective value display)
500 VA (maximum 30 minutes)
Voltage setting method Manual adjusted transformer
Waveform
Same as the power supply waveform
Frequency
Same as the power supply frequency
Measurement range
Current: 0.01 to 120 mA , ±(3% f.s.+20µA)
2mA/8mA/32mA/120mA
AC (Average value rectified, RMS display)
Insulation test
Testing voltage
DC500V/1000V
Measurement range
2MΩ to 2000MΩ(500V), 4MΩ to 2000MΩ(1000V)
Judgment function
Contents: UPPER-FAIL / PASS / LOWER-FAIL
(Digital setting window comparator method)
Timer section
Setting range: 0.5 to 999 seconds
Interfaces
EXT I/O, EXT SW, RS-232C
Display
Fluorescent tube display (digital), Analog meter
Monitor function
Output voltage, detection current, Insulation resistance
Power supply
120 V AC, 50/60 Hz (3159-01)
220 V AC, 50/60 Hz (3159-02)
230 V AC, 50/60 Hz (3159-03)
240 V AC, 50/60 Hz (3159-04)
Dimensions, mass
320W×155H×330D mm, 18 kg~21.5kg
Accessories
9615 H.V. TEST LEAD (high voltage side and return, 1 each),
Power cord (1), spare fuse (1)
OPTIONS
STANDARD
23
9267 SAFETY TEST DATA
MANAGEMENT SOFTWARE
9613 REMOTE CONTROL BOX (single)
9614 REMOTE CONTROL BOX (dual)
9637 RS-232C CABLE
(Dsub 9pin-9pin, cross,1.8m)
9638 RS-232C CABLE
(Dsub 9pin-25pin,cross,1.8m)
9267 SAFETY TEST DATA MANAGEMENT
SOFTWARE
3173
PORTABLE WITHSTANDING VOLTAGE HiTESTER
An Economical and
Simple way to Handle
Withstand Voltage Testing
●Measures between 0 to 3kV AC
●External Control (Standard)
3931
SPECIFICATIONS
Basic Specifications
[Voltage generator]
Output voltage range
0 to 3 kV AC (single range), 30 VA
Accuracy
±5% f.s.
Waveform
Power waveform, Synchronized to power
[Current detector]
Current cut-off
0.1 to 9.9 mA
Current cut-off accuracy Setting value ±(5%+0.05mA)
Evaluation method
Analog comparator PASS, FAIL a buzzer sound and external I/O
[Timer area]
Setting range
1 to 99 sec (1-second resolution)
Timer accuracy
Setting value ±(1%+50msec)
General Specifications
EXT I/O signal
START and STOP, PASS and FAIL, TEST
Power supply
120 V AC (3173-01), 220 V AC (3173-02)
230 V AC (3173-03), 240 V AC (3173-04) 50/60 Hz 50 VA
Dimensions, Mass
Approx. 149 (W) × 200 (H) × 215 (D) mm
Approx. 7.0 kg (for 120 to 240 V AC)
HIGH VOLTAGE CONTACT CHECKER
SPECIFICATIONS
■ Inputs
Max. allowable input voltage
Max. rated voltage to earth
Measurement range
Waveform
Threshold voltage adjustment
Voltmeter
Input impedance
5kV AC, 50/60 Hz
5kV AC, 50/60 Hz
200V to 5kV AC, 50/60 Hz
Commercial power waveform, sine waveform
Manual adjustment
Analog voltmeter, displays average rectified effective values
0 to 5kV AC, ± 5% f.s. (after warming up for at least 10 minutes)
High-voltage input, 150 MΩ ± 10%
(High; 100 MΩ, Low ; 50MΩ)
■ Detector
Threshold voltage setting range 200V to 5kV (rms)
Result output
3154
DIGITAL MΩ HiTESTER
For the laboratory to the production line, six test
voltages from 25 to 1000V
STANDARD
3158
SPECIFICATIONS
Measurement function Insulation resistance (Applied DC voltage method)
25, 50, 100, 250, 500, 1000 V DC
Testing voltage
Measurement range 25 to 50 V: 2MΩ to 200 MΩ, 3 ranges
100 to 250 V: 2MΩ to 2000 MΩ, 4 ranges
500 to 1000 V: 2MΩ to 4000 MΩ, 4 ranges
±2 % rdg. ±5 dgt. (at 25 to 100 V testing voltage, 0 to 20.00 MΩ)
Accuracy
(at 250 V testing voltage, 0 to 100.0 MΩ) (at 500 to 1000 V testing
voltage, 0 to 999 MΩ)
±5 % rdg. (at 25 to 50V testing voltage, 19.0 to 200.0MΩ)
(at 100V testing voltage, 19.0 to 2000MΩ) (at 250V testing voltage,
100.1 to 2000MΩ) (at 500 to 1000V testing voltage, 1000 to 4000MΩ)
Fast: less than 0.7 second, Slow:less than 1.5 seconds (at manual ranging)
Response times
Fast: 10 samples/s, Slow:1 sample/s
Sampling rates
Comparator functions: judgments PASS or FAIL Test time timer
Functions
functions: 0.5 to 99 second Delay time timer functions: 0.1 to 99 second
LED
Display
100 to 240 V AC (50/60 Hz)
Power supply
215W × 61H × 213D mm, 1.1 kg
Dimensions, mass
Power cord (1)
Accessories
AC WITHSTANDING VOLTAGE HiTESTER
Guarantees electrical and electronic equipment
safety
SPECIFICATIONS
Test function
Testing voltage
Voltage testing method
Transformer capacity
Voltage adjustment
Waveform
Frequency
Measurement items
Judgment function
STANDARD
Beeper, LED indication, EXT I/O output
Electronic Measuring Instruments
Monitor the voltage
on test terminals
Greatly increased
reliability for voltage
withstand tests
Timer section
Interfaces
Display
Monitor function
Power supply
Dimensions, mass
Accessories
Withstanding test
0 to 2.5kV / 0 to 5.0kV AC, dual-range configuration
Zero-toggle switch
500VA (maximum 30 minutes)
Manually adjusted transformer
Same as the power supply waveform
Same as the power supply frequency
Voltage: 0 to 5kV AC (Average value rectified, RMS display)
Current: 0.01 to 120mA AC (Average value rectified, RMS display)
Contents: UPPER-FAIL / PASS / LOWER-FAIL
(Digital setting window comparator method)
Setting range: 0.5 to 999 seconds
EXT I/O, EXT SW, RS-232C
Fluorescent tube display (digital), Analog meter
Output voltage, detection current
120V AC, 50/60Hz (3158-01) 220V AC, 50/60Hz (3158-03)
230V AC, 50/60Hz (3158-04)
320W × 155H × 263D mm, 16kg (3158-01) 18kg (3158-03, -04)
9615 H.V. TEST LEAD (high voltage side and return, 1 each), Power cord (1), spare fuse (1)
24
3156
LEAK CURRENT HiTESTER
Leakage Current Measurement Essential for
Electrical Safety
●Automatically compatible with Networks
stipulated by IEC/UL/JIS standards
●Automatic measurement function
(Measuring power supply polarity
switching as well as the normal state /
single failure state)
●Stores data for 100 units
●Power supply separation
Electronic Measuring Instruments
STANDARD
STANDARD
SPECIFICATIONS
Earth leakage current
Measurement mode
Leakage current between enclosure and earth /
Leakage current between enclosure and enclosure /
Leakage current between enclosure and line /
Patient leakage current I / Patient leakage current II /
Patient leakage current III / Patient auxiliary current
DC / AC / AC+DC (25mA max), AC peak (75mA max)
Target current
Measurement range DC / AC / AC+DC mode ; 50µA / 500µA / 5 mA / 25 mA
AC peak mode ; 500µA / 1 mA / 10 mA / 75 mA
Measurement system Indication of a current value calculated based the measured drop in
voltage caused by simulated resistance of the human body.
Measurement of true effective value.
The measurement section ; chassis-grounded and floating.
DC / AC / AC+DC mode ; ±(2.0 % rdg. + 6 dgt.)
Accuracy
AC peak mode ; ±(2.0 % rdg. + 2 dgt.)
DC mode ; ±(2.0 % rdg. + 3 dgt.)
1 MΩ ±1 % (Excluding voltmeter section, simulated resistance
Input resistance
of the human body)
For medical electrical equipment / For IEC 60990 /
Network (human
simulated resistance) For JIS / For UL / General-purpose 1 / General-purpose 2
110 % voltage application function / Wiring check function /
Functions
Automatic measurement function /
Application line selection function /
Ground fault prevention function /
3157-01
Functions
Interface
Power supply
Dimensions, mass
Accessories
Setting of single-fault condition /
Switching power supply polarity / Setting of measuring time /
Measurement delay / Maximum value hold /
Allowable value judgement / Data save / Clock / Data back up / etc.
EXT I/O , RS-232C , GP-IB
100, 120, 220, 240 V AC (default setting)
(50/60 Hz, Rated power ; 30VA)
Approx. 320W × 110H × 263Dmm, 4.0 kg
9170 TEST LEAD(2), 9195 ENCLOSURE PROBE(1),
9399 CARRYING CASE(1), Alligator clip(3)(2 red, 1 black)
AC Power cord(2), Spare fuse(2)
OPTIONS
9637 RS-232C CABLE (9-pin to 9-pin, crossing cable 1.8m)
9638 RS-232C CABLE (9-pin to 25-pin, crossing cable 1.8m)
9151-02 GP-IB CABLE (2.0m)
9151-04 GP-IB CABLE (4.0m)
9442 PRINTER
9443-02 AC ADAPTER (for printer, for use in EU)
9443-03 AC ADAPTER (for printer, for use in USA)
9444 CONNECTION CABLE (for printer)
1196 RECORDING PAPER (for printer)
9686 CARRYING CASE (with casters)
9267 SAFETY TEST DATA MEASUREMENT SOFTWARE
AC GROUNDING HiTESTER
Protective ground tester indispensable for
standard certification
SPECIFICATIONS
Low resistance, AC 4-terminal method
Measurement items
Current generator principle: PWM constant current control,
Generator section
Current setting range: 3.0A to 31.0A (0.1A resolution), into 0.1Ω load,
Maximum output power: 130VA (at output terminals)
Subject to derating according to ambient temperature (80% at 40°C)
Frequency: 50Hz or 60Hz sine wave
Soft start function: Apply current only after checking load connection
Resistance measurement: 0 to 1.800Ω (0.001Ω resolution), Accuracy:
Monitor section
±2% rdg. ±4 dgt. (after zero-adjust), Current monitoring range: 0 to
35.0A AC (0.1A resolution), Monitoring cycle: 2 times/second
Timer setting: Counts down time after start until preset time, or
Other functions
shows elapsed time after start, Setting range: 0.5 to 999 second,
Comparator: Pass/Fail evaluation using preset upper/lower limit, I/O
output, Memory function: max. 20 settings (with save/load)
Fluorescent tube digital display
Display
100 to 120V, 200 to 240V AC (auto-switching), 50/60Hz
Power supply
320W × 90H × 263D mm, 7kg
Dimensions, mass
Power cord(1), Spare fuse(1), Shorting bar(2)
Accessories
OPTION
25
OPTION
7016
SIGNAL SOURCE
SPECIFICATIONS
Signal Generator with DMM
●Constant voltage 0 to ± 1.5000V 0 to ± 15.000V
●Constant current 0 to ± 25.000mA
●Pulse generation and measurement
STANDARD
OPTIONS
3856-01 COMMUNICATION PACKAGE(RS-232C)
3856-02 COMMUNICATION PACKAGE(USB)
*9180 SHEATH TYPE TEMPERATURE PROBE
*9181 SURFACE TYPE TEMPERATURE PROBE
*9182 SHEATH TYPE TEMPERATURE PROBE
*9183 SHEATH TYPE TEMPERATURE PROBE(class1)
9472 SHEATH TYPE TEMPERATURE PROBE(class1)
9473 SHEATH TYPE TEMPERATURE PROBE(class1)
9474 SHEATH TYPE TEMPERATURE PROBE(class1)
9475 SHEATH TYPE TEMPERATURE PROBE(class1)
9476 SURFACE TYPE TEMPERATURE PROBE
3851-10 TEST LEAD (Lead length:1m/standard accessories)
*Non-CE mark products
DC SIGNAL SOURCE
All-in-one Signal source
SPECIFICATIONS
Constant voltage: 0 to ±2.5 V, 100µV resolution,
0 to ±25 V, 1 mV resolution, Sink /source: ±25 mA,
±0.03% of setting ±3 dgt.
Constant current: 0 to ±25 mA, 1µA resolution,
Sink /source; ±25 V, ±0.03 % of setting ±3 dgt.
Thermoelectric power TC (0°C)/TC(RJ)
K: -176.0 to 1372.0°C, E: -220.0 to 839.0°C, J: -208.0 to
1108.0°C, T: -169.0 to 400.0 °C, R: -50 to 100°C, 101 to
1769°C, S: -50 to 100°C, 101 to 1769°C, B: 300 to 600°C,
601 to 1820°C
±0.05% of setting ±0.5°C (at Thermoelectric power K)
Basic Accuracy
Measurement functions Voltage: 0 to ±2.5 V, 100µV resolution,
0 to ±25 V, 1 mV resolution, Input resistance: 1 MΩ,
and Accuracy
±0.03% rdg. ±2 dgt.
Current: 0 to ±25 mA, 1µA resolution,
Input resistance: 25Ω, ±0.03 % rdg. ±2 dgt.
-25.0 to 80.0°C, 0.1°C resolution (using the 9184)
Temperature
100Ω, ±0.2 % rdg.
Standard resistance
LR6(alkaline) 6 pieces, 9420 BATTERY PACK(Ni-Cd),
Power supply
or 9418-10 AC ADAPTER
104W×180H×58D mm, 590 g (excluding batteries)
Dimensions, mass
9168 INPUT CORD(1), 9170 TEST LEAD(1), Fuse(3)
Accessories
Generator functions
and Accuracy
●Check thermocouple temperature sensors
●Generate various signals for electronic circuit and equipment testing
●Calibrate industrial equipment all with a single unit
OPTIONS
9184 RJ SENSOR
(for reference contact compensation)
9380 CARRYING CASE
9418-10 AC ADAPTER
9420 BATTERY PACK (7.2V /700mAh)
7075 7075-01
Electronic Measuring Instruments
7011
Constant voltage: 0 to ±1.5V, 100µV resolution, 0 to ±15V,
1mV resolution,Sink/source: ±25mA, ±0.03%of setting ±3dgt.
Constant current: 0 to±25mA,1µ A resolution,
Sink/source:±12V, ±0.03% of setting ±5dgt.
Pulse signal genration:0.5to 4800Hz, 28 ranges, ±0.005% of
setting ±0.01Hz
DC
Voltage: 50mV to250V, 6 ranges, ±0.03% rdg. ±5dgt.
Measurement functions
AC Voltage: 50mV to250V, 6 ranges, ±0.7% rdg. ±20dgt.
and Basic accuracy
AC+DC Voltage: 50mV to250V, 6 ranges, ±0.8% rdg. ±25dgt.
DC Current: 50 to 500mA, 2 ranges, ±0.03% rdg. ±5dgt.
AC Current: -50to 500mA, 2 ranges, ±0.6% rdg. ±20dgt.
AC+DC Current: 50 to 500mA, 2 ranges, ±0.7% rdg. ±40dgt.
Measurement functions Resistance: 500Ω to50MΩ, 6 ranges, ±0.15% rdg. ±5dgt.
Diode&Continuity check: Possible
and Accuracy
Temperature :-40to1372°C, ±0.3%rgt. ±3°C
(K type thermocouple, °F display possible)
Frequency :100Hz to 200kHz, 5 ranges, ±0.02% rdg. ±3dgt.
LCD with backlignt 51,000 counts
Display
LR6(AA)8pieces, included Ni-MH battery pack, or
Power supply
included AC adapter
90W×192H×54D mm, 735g(instrument only)
Dimensions,mass
Carring case(1), AC adapter(1), Battery(8), 3851 TEST
Accessories
LEAD(1), Alligator clips(1), Yellow test lead(1), Test leads(1)
Generator functions
and Accuracy
WAVEFORM GENERATOR
Arbitrary waveform generator with four
independently controllable channels
●Even for complex signals, evaluation is made easy
●Easy touch panel operation
●Multiple channels, 4CH (7075), 2CH (7075-01)
●Large 128,000-Word/channel memory, sweep sequence functions
STANDARD
STANDARD
SPECIFICATIONS
Number of channels 7075: 4-channels, 7075-01: 2-channels
Function generator, Arbitrary waveform generator (for each channel)
Output functions
Max. output voltage 10 V range: 10 mV to 10 V o.c. (1 mV resolution)
1 V range: 1 mV to 1 V o.c. (0.1 mV resolution)
0.1 V range: 0.1 mV to 0.1 V o.c. (0.01 mV resolution)
(o.c.:open-circuit)
40 Ohm
Minimum load
impedance
50 Ohm±2 % (DC)
Output impedance
Waveform types: sine, square (fixed 50 % duty), triangle, ramp-up,
Function generator
ramp-down, pulse, noise, DC, Frequency range:sine (10 mHz to
mode
10 MHz), square (10 mHz to 10 MHz), triangle (10 mHz to
200 kHz), pulse (10 mHz to 200 kHz)
Arbitrary waveform Voltage axis resolution: 16 bits (64000 counts)
Waveform memory capacity: 128 kW/ch
generation mode
Filtering: 2-stage LPF, 50 Hz to 1 MHz, 14 steps,
Waveform input methods: FD/GP-IB (direct download from
MEMORY HiCORDER by FD or GP-IB), or RS-232C download
(at use of 7990), Arbitrary waveform clock: Max. 4ch,
Frequency range: 10 mHz to 10 MHz (10 mHz resolution)
5.7-inch LCD (with touch panel)
Display
FDD×1, MS-DOS format
Data storage
100/120/200/230 V AC/auto selects, (50/60 Hz)
Power supply
345W×130H×286Dmm, 7075: 7.8 kg, 7075-01:7.5 kg
Dimensions, mass
7990 WAVEFORM CREATION SOFTWARE (CD-R×1)
Accessories
OPTIONS
●Output cord
9165 CONNECTION CORD BNC to BNC, 1.5 m length
9166 CONNECTION CORD BNC to clip, 1.5 m length
●PC communication
9151-02 GP-IB CONNECTION CABLE 2 m length
9151-04 GP-IB CONNECTION CABLE 4 m length
26
Environmental Measuring Instruments
Environmental Measuring Instruments Index
Temperature measurement
Rotation, Illumination
3441/3442
3446-01, 3447-01 3412-50
-100 °C to 1300 °C
Choose from Basic or
Waterproof models
-100˚C to 1000˚C 1 ch (3446) -50 °C to 999 °C
-100˚C to 300˚C 2 ch (3447) With analog output
with built-in memory
. . . . . . . . . . . . . . . p.30
3403 Rotation
3423 Illumination
30 to 100,000 r/min
20 to 200,000 lx, digital
3404 Rotation
30 to 100,000 r/min,
Max. /Min. /Total /Period
. . . . . . . . . . . . . . . p.31 . . . . . . . . . . . . . . . p.31
Non-contact temperature measurement
Environmental Measuring Instruments
3415-01
-50 °C to 500 °C
Two-beam laser marker
Narrow field
measurement
3416-01
-50 °C to 500 °C
LED spot marker
Spot measurement
3418
3443
3444
3445
-50 °C to 500 °C
Without laser marker
Narrow field
measurement
-50.0 °C to 500.0 °C
Two-beam laser marker
Narrow field measurement
Data memory, Memory dump to
printer, RS-232C interface
. . . . . . . . . . . . . . . p.30
-50.0 °C to 500.0 °C
Two-beam laser marker
Narrow field measurement
MAX. MIN. indication, Analog
output, RS-232C interface
. . . . . . . . . . . . . . . p.30
-50.0 °C to 500.0 °C
Two-beam laser marker
Spot measurement
MAX. MIN. indication, Analog
output, RS-232C interface
. . . . . . . . . . . . . . . p.30
Data Loggers (Temperature/Humidity/Instrumentation/DC-Voltage/AC-Current/AC-Voltage/Leak-Current)
2300 Series
3641-20,3632-20,3633-20 3634-20
Remote Measurement System
Various measurement modules
Internal memory
LAN or SS Air Module
Temperature/Humidity
-40~85˚C 0~100% rh
(using the 9680 sensor)
-40.0 °C to 180.0 °C
(external sensor)
. . . . . . . . . . . . . . . p.29
. . . . . . . . . . . . . . . p.28
Instrumentation
0 to 20.00 mA DC
3635-24,-25,-26
DC Voltage
-24: ±500.0 mV DC
-25: ±5.000 V DC
-26: ±50.00 V DC
. . . . . . . . . . . . . . . p.29
. . . . . . . . . . . . . . . p.29
Data Loggers (Pulse/Illumination/DC-Voltage/Communication Base)
27
(via infrared radiation energ y)
3639-20
Pulse Totalizer
9,999counts/interval
(1ch)
3640-20
Illumination
2000/20000/200000 lx
(1ch)
. . . . . . . . . . . . . . . p.29
. . . . . . . . . . . . . . . p.29
3645-20
Multi-range
Voltage Logger
with preheat
signal function
. . . . . . . . . . . . . . . p.29
3911-20,3912-20
Communication Base
to analyze and process
on a personal computer
. . . . . . . . . . . . . . . p.29
3636-20
AC Current (2ch)
0 to 50.00/500.0 A AC
3637-20
AC Voltage (1ch )
0 to 600.0 V AC
. . . . . . . . . . . . . . . p.29
3638-20
AC Leak Current
(2ch) with clamp-on leak
sensor
. . . . . . . . . . . . . . . p.29
Temp Logger
3650
Temperature -40°C to 85°C
Data memory of 2048 data
Button battery size
2300 Series
REMOTE MEASUREMENT SYSTEM
Easily Construct a Centralized Data Management
System for Monitoring Multiple Locations
LAN Module that paves the way for a low cost remote measuring
system by utilizing existing data network
●Various measurement modules for temperature and humidity,
instrumentation, and pulse
●Power measurement module for multiple circuits
●Large internal memory to avoid data loss due to
communication problems
●Communication module with built-in real-time clock tracks the
data of each measurement module to the second
●SS Air Module for constructing wireless systems
Environmental Measuring Instruments
2343-20
2353-20
2301-20 HUMIDITY MODULE
Temperature 1ch and humidity 1ch.
Use with optional sensor 9764
Temperature: -40.0 to 85.0°C
Humidity: 0.0 to 100%RH
2302-20 Pt MODULE
Temperature 2ch (Pt100)
2 types of platinum resistance thermo sensors available
2303-20 TC MODULE
2ch temperature measurement using thermocouples
(K, E, J, T). 4 TC types available
2304-21 PULSE MODULE
For 2ch pulse input (voltage, contact), maximum 16M pulses/
interval
Input pulse: 4kHz max. (voltage/electronic contact signal)
25Hz max. (mechanical contact signal)
2305-20 INSTRUMENTATION
MODULE
Voltage / current 2ch measurement, for 4-20mA, 1-5V
instrumentation signals
2331-20 POWER METER
MODULE
For single circuit power measurement.
Single-phase 2-wire to 3-phase 4-wire
Voltage: AC 100/200V
Current: AC 5A (with 9695-02 CLAMP ON
SENSOR), AC 50A (with 9695-02), AC 100A
(with 9695-03 or 9661-01), AC 500A (with 9661-01)
2332-20 POWER
METER MODULE
For multiple circuit power measurement: from 6 circuits of
single-phase 2-wire to 3 circuits of 3-phase 3-wire installations
Voltage: AC200V (100V accepted at 200V range)
Current: AC 5A (with 9695-02 CLAMP ON
SENSOR), AC 50A (with 9695-02), AC 100A (with
9695-03 or 9661-01), AC 500A (with 9661-01)
*3-phase 4-wire circuit cannot be measured
2341-20 INPUT MODULE
For recording the status of contact signals
Input 8 ch, Input internal bus isolated
Easily capture on/off status with LED
2342-20 OUTPUT MODULE
Acts as receiver of higher order external control device,
outputs control signals and monitors data of measurement
modules
Output 8ch (Open corrector output), Output internal bus isolated
2343-20 RS LINK MODULE
For communicating with RS-232C interface equipped instruments
Interface: RS-232C, Transfer speed: 57.6kbps(max.)
Please inquire regarding compatible instruments.
(Existing compatible HIOKI instruments include
Models 3331 and 3332.)
2351 AIR MODULE
2.4GHz band SS radio system (RS-232C equipped)
Transfer speed: 51.9kbps(fixed), RS-232C: 57.6kbps(max.)
2351-20 : For EU
2351-21 : For USA
2352-20 WIRE MODULE
For small-scale measurement systems or built-in use
Interface: RS-232C, Transfer speed: 57.6kbps(max.)
2353-20 LAN MODULE
For data logging via LAN
Interface: 10BASE-T
2361-20 AC POWER MODULE
Power supply for the communication modules and
measurement modules (max. 10 modules)
Input: AC 100 to 240V, Output: DC5V/2.4A
2362-20 DC POWER MODULE Power supply for the communication modules and
measurement modules (max. 10 modules)
Input: DC 19 to 36V, Output: DC5V / 2.4A
2391 MODULE BASE
For connecting modules (3 slots reserved for power supply
and communication module)
2391-01: For relay and master station,
2391-02: 5 measurement module slots,
2391-03: 10 measurement module slots
*Model 2331 uses 2 slots
2392 MODULE BASE
For connecting MODULES (also connects with additional
MODULE BASEs for increased measurement capabilities)
2392-01: 1 slot; includes power and internal bus
connection terminal
2392-02: 2 slots; connect with 2391-01 for additional
measurement module slots
* POWER MODULE not compatible; must use with
Model 2392-01 to access POWER MODULE
28
3630
Data Loggers
Data Loggers for All Types of Measurements
For HACCP-related Temperature and Humidity Recording
HUMIDITY LOGGER
TEMPERATURE LOGGERs
For Recording Voltage
VOLTAGE LOGGERs
3641-20
3632-20
3633-20
3635-24, -25, -26
3645-20
3637-20
Can alternately record temperature
and humidity on two channels
for temperature and humidity
measurement
-40.0 ˚C to 85.0 ˚C
0.0 %rh to 100.0 %rh
Waterproof with built-in sensor for
temperature measurement
-20.0 ˚C to 70.0 ˚C
External sensor for temperature
measurement
-40.0 ˚C to 180.0 ˚C
For measurement DC voltage
-24: ± 500.0 mV DC
-25: ± 5.000 V DC
-26: ± 50.00 V DC
With preheat function
For measuring DC voltage
Range: ± 50.00 mV to
± 50.00 V DC
For measuring AC voltage
Range: 600.0 V AC
For Recording Precipitation or Illumination
For Recording Load Current and Monitoring Leak Current
ILLUMINATION LOGGER INSTRUMENTATION LOGGER CLAMP LOGGER
PULSE LOGGER
LEAK LOGGER
3912-20: OPTION
3639-20
3640-20
3634-20
3636-20
3638-20
3911-20: OPTION
For cumulative pulse measurement
for precipitation gauges,
flow gauges, etc.
For illumination measurement
Range: 2,000 lux to
200,000 lux
For measuring typical
instrumentation signals
Range: 20.00 mA DC
For measuring alternating current on
two channels
Range: 50.00/500.0 A AC
(Clamp sensors sold separately)
For measuring alternating current on
two channels
Range: 100.0/1000 mA AC
(Clamp sensors sold separately)
Environmental Measuring Instruments
Accessories
9680
HUMIDITY SENSOR
9632
CONNECTION CABLE
9639
CONNECTION CORD
9629
CONNECTION CABLE
9662
LUX SENSOR
Cord length: 1 m
Cord length: 1 m
Cord length: 3 m
Cord length: 5 m
Cord length: 2 m
Options for 3634-20
Options for 3636-20
Options for 3638-20
9650
CLAMP ON SENSOR
9651
CLAMP ON SENSOR
9657
CLAMP ON SENSOR
9658
CLAMP ON SENSOR
AC 100 A f.s./Up to φ 15 mm
Cord length: 3 m
AC 500 A f.s./Up to φ 46 mm
Cord length: 3 m
AC 1.0 A f.s. / Up to φ 40 mm
Cord length: 3 m
AC 1.0 A f.s. / Up to φ 12 × 30 mm
Cord length: 3 m
9631-01/9631-11/9631-21
TEMPERATURE SENSOR
(Molded plastic type)
9631-02
TEMPERATURE SENSOR
(Needle type)
9631-03
TEMPERATURE SENSOR
(Sheathed type)
9631-05
TEMPERATURE SENSOR
(Molded plastic type)
9631-04/9631-14/9631-24
TEMPERATURE SENSOR
(Lug type)
-40.0 ˚C to 180.0 ˚C
Cord length:9631-01:1 m,
9631-11: 5 m, 9631-21:10 m
-40.0 ˚C to 120.0 ˚C
Cord length:1 m
-40.0 ˚C to 120.0 ˚C
Cord length:1 m
-40.0 ˚C to 180.0 ˚C
Cord length:30 mm
-30.0 ˚C to 180.0 ˚C
Cord length:9631-04:1 m,
9631-14: 5 m, 9631-24:10 m
9633
CONNECTION CABLE
9634
CONNECTION CABLE
Cord length: 1 m
Cord length: 1 m
Options for 3641-20 / 3633-20
9680/9680-01/9680-02
HUMIDITY SENSOR (for 3641-20)
-40.0 ˚C to 85.0 ˚C
0.0 %rh to 100.0 %rh
Cord length:9680:1 m (provided),
9680-01: 5 m, 9680-02:10 m
3911-20 3912-20
COMMUNICATION BASE
Analyze and Process Data on a Personal Computer
The 3911-20, 3912-20 COMMUNICATION BASE are
used to transfer data to a personal computer.
3911-20
3912-20
3911-20
3911-20, 3912-20
Recording Capacity
Communication
method
Power supply
Dimensions
and mass
Accessories
SPECIFICATIONS
Max. 16,000 data points × 16 ch, 32,000 data points × 8 ch,
RS-232C (3911-20 to a PC)
USB 1.1 (3912-20 to a PC)
4 × 1.5 V, LR03 (AAA) alkaline dry cell batteries
3911-20: 69W×92H×36D mm, 150g (including batteries)
3912-20: 69W×128H×36D mm, 180g (including batteries)
3911-20: LR03 (AAA) alkaline dry cell batteries (4),
Communication software (1)
3912-20: USB cable (1), LR03 (AAA) alkaline dry cell batteries (4),
Communication software (1)
Options for 3911-20
3912-20
29
Communication software included
with the 3911-20, 3912-20
Compatible OS: WIndows
95/98/NT4.0/Me/2000/XP
9637 RS-232C CABLE
9638 RS-232C CABLE
9-pin to 9-pin crossed
cable/1.8 m
9-pin to 25-pin
crossed cable/1.8 m
3441 3442
TEMPERATURE HiTESTER
Supports temperature management demands of
various applications
●Compact and weighing only 160g
●More than 200 hours of continuous operation on a single battery
●An assortment of 9 optional temperature sensors
●3442 : Waterproof construction
3441,3442 : SPECIFICATIONS
Material type
K type thermocouple (Chromel /Alumel)
Measurement range -100°C to 1300 °C (-148°F to 2372°F)
The actual measurement range is restricted by the temperature probe.
0.1°C*1 or 1°C*2
Resolution
±0.1%rdg. ±0.8°C (1.5 °F)*1 or ±0.2% rdg. ±1°C (1.8°F)*2
Unit Accuracy
(in addition to accuracy of temperature sensor)
*1 during measurement from -100 to 199.9°C (-148°F to 392°F)
*2 during measurement from 200 to 1300°C (392°F to 2372°F)
Display
LCD
Sampling rate
2 times/second
Contact compensation Auto compensation
Functions
Max/Min temperature recording and display,
display data hold, sensor discontinuity display,
Over-range display, °C/°F display switching (3441-02, 3442-03),
auto power save, low battery warning
Place of use
Indoor use to altitude of 2000 m
Power supply
R6P (AA)×4, or LR6 (AA)×4
Operating time
200 hours or better of continuous use (with manganese battery)
Dimensions, mass
74W×155H×24D mm, 160 g
Accessories
Strap band(1), R6P(AA) Batteries(4)
OPTIONS
3442
(Waterproof construction)
3441
3443 3444 3445 TEMPERATURE HiTESTER
Non-contact measurement, quick and easy
temperature management
3443 : SPECIFICATIONS
Measurement range -50.0°C to 500.0°C, 0.1°C resolution
Measurement field diameter ø24mm at a distance of 1 m
Accuracy
●Ideal for daily temperature checks -- 3443 with integrated memory
●For temperature monitoring -- 3444, 3445 with real-time output
●Dedicated software (option) -- For date analysis and management
●Use in tough environments -- dust/splash-proof design (IP54)
Environmental Measuring Instruments
9180 SHEATH TYPE TEMPERATURE PROBE (up to 750 °C)
9181 SURFACE TYPE TEMPERATURE PROBE (up to 400 °C)
9182 SHEATH TYPE TEMPERATURE PROBE (up to 750 °C)
9183 SHEATH TYPE TEMPERATURE PROBE (up to 750 °C)
9386 CARRYING CASE
9476 SURFACE TYPE TEMPERATURE PROBE (up to 500 °C)
(waterproof structure models)
9472 SHEATH TYPE TEMPERATURE PROBE (up to 300 °C)
9473 SHEATH TYPE TEMPERATURE PROBE (up to 800 °C)
9474 SHEATH TYPE TEMPERATURE PROBE (up to 300 °C)
9475 SHEATH TYPE TEMPERATURE PROBE (up to 500 °C)
(refer to P.57)
±1 % rdg. (at 200.1 to 500.0 °C), ±2°C (at 0.0 to 200.0 °C),
±10 % rdg. ±2°C (at -50.0 to -0.1 °C)
1.6 seconds (95% response)
Response time
Date memory function 130 points of data, memory dump to printer
Analog output function None
Interface
RS-232C output (requires 3909 INTERFACE PACK)
Other functions
Auto power save, low battery warning, auto-hold
Power supply
6F22(006P)× l, or AC adapter
Operating time
Continuous use of 20 hours (light on) and 50 hours (light off)
Dimensions, mass
47W × 200H × 48D mm, 280 g
Accessories
Carrying case(1), hand strap(1), 6F22(1), screwdriver(1)
3444, 3445 : SPECIFICATIONS
Measurement range -50.0°C to 500.0°C, 0.1°C, 0.1 or 1°C resolution switchable
Measurement field diameter 3444 : ø24mm at a distance of 1 m
3445 : ø2.5mm at a distance of 7 cm
±1 % rdg. (at 200.1 to 500.0 °C), ±2°C (at 0.0 to 200.0 °C),
Accuracy
±10 % rdg. ±2°C (at -50.0 to -0.1 °C)
1.6 sec (95% at 0.1°C resolution), 0.7 sec (95% at 1°C resolution)
Response time
3443
3444
3445
OPTIONS
3909 INTERFACE PACK
*BLACK BODY SPRAY
(ε= 0.95, 180ml, 2-coat coverage about 1.0m2,, good to 550 ˚C)
*BLACK BODY TYPE
(ε= 0.95, 50mm×10m, 1roll, good to 180 ˚C)
Date memory function None
Analog output function Possible (requires 3909 INTERFACE PACK)
Interface
RS-232C output (requires 3909 INTERFACE PACK)
Other functions
Auto power save, low battery warning
Power supply
6F22(006P)× l, or AC adapter
Operating time
Continuous use of 20 hours (light on) and 50 hours (light off)
Dimensions, mass
47W × 200H × 48D mm, 280 g
Accessories
Carrying case(1), hand strap(1), 6F22(1), screwdriver(1)
*Note: Used for accurate measurement of the temperature of objects which have
low thermal emissivity (ε) such as polished metal, and for calibrating thermal
emissivity.
30
3423
LUX HiTESTER
Digital illumination meter, maximum scale of
199,900 lx
●Easy-to-operate, hand-held unit
●From low light up to a maximum intensity of 199,900lx
●For illumination equipment, lighting work, and
facility management
3423 : SPECIFICATIONS
Measurement range 20 to 200,000 lx full-scale, 5 ranges
Accuracy
±4 % rdg. ±1 dgt. (environment temperature: 23 ±5°C)
Display
1999 full digits, LCD with EL backlight
(Note : in the 20,000 lx range, the maximum is 19990 /10 digits steps,
and in the 200,000 lx range, the maximum is 199900 /100 digits steps)
5 sec. or less (auto range), 2 sec. or less (manual range)
Response time
Silicon photodiode
Receptor element
Sensor separate: Permits remote measurement with
Other functions
the sensor separated from the main unit.(using the 9436)
Analog output: 200 mV DC at full scale rate
R6P (AA) × 2 (Continuous use of 25 hours)
Power supply
or AC adapter (6 V, 300 mA)
74W × 170H × 30D mm, 310 g (including the dry cells)
Dimensions, mass
9376 CARRYING CASE(1), Sensor cap(1), R6P(2)
Accessories
OPTIONS
9436 CONNECTING CABLE (with case, 2 m length)
9094 OUTPUT CORD (1.5 m length)
Environmental Measuring Instruments
3403 3404
TACHO HiTESTER
Precise rotation speed meter
●Minimum 0.01 r/min resolution and high ±1 digit precision make the
tachometer effective for precise measurements.
●By using the optional 9213 contact adapter, the 3403/3404 become
contact-type tachometers, while the period ring can be used in
applications like measurement of conveyer line speed.
●LED and buzzer signals confirm that the light pulses reflected from
the tape on the rotating body are being picked up.
●MIN/MAX mode, TOTAL mode, PERIOD mode, FAST/SLOW sampling
mode, and Analog output Multi-function performance for the 3404
only.
3403: SPECIFICATIONS
Measurement ranges r/min: rotation per minute, r/s: rotation per second
r/min: (30.00 to 199.99)range to (20000 to 99990)range, 4 ranges,
r/s: (0.5000 to 1.9999)range to (200.0 to 1600.0)range, 4 ranges
Sampling period
0.5 second to 2 seconds
Detection distance
Distance of 50 mm to 200 mm
Analog output
None
Power supply
R6P (AA) × 4 (continuous use of 17 hours) or AC Adapter
(6 V, 300 mA)
Dimensions, mass
62W × 180H × 38D mm, 260 g
Accessories
9211 REFLECTIVE TAPE(1Sheet), Carrying case(1)
3404: SPECIFICATIONS
Measurement ranges r/min: (30.00 to 199.99)range to (20000 to 99990)range, 4 ranges
(at slow sampling mode)
r/s: (0.5000 to 1.9999)range to (200.0 to 1600.0)range, 4 ranges
(at slow sampling mode)
Total: Total rotation counts from 0 to 599999
Period: Measure rotation pulses from 600µs to 2 sec.
Slow: 0.5 sec. to 2 sec., Fast: 0.1 sec. to 0.5 sec.
Sampling period
Distance of 50 mm to 200 mm
Detection distance
1V DC at full-scale range output
Analog output
Power supply
R6P (AA) × 4 (continuous use of 16 hours) or AC Adapter (6 V, 300 mA)
62W × 180H × 38D mm, 260 g
Dimensions, mass
9211 REFLECTIVE TAPE (1Sheet), 9094 OUTPUT CORD (1),
Accessories
Carrying case (1)
OPTIONS
9211 REFLECTIVE TAPE
(30 pieces /sheet, 10 sheets /1 set, 12 mm×12 mm /1 piece size)
9213 CONTACT ADAPTER SET
(includes 9032×1, 9033×2, 9212×1)
9212 PERIPHERAL RING
9033 RUBBER CONTACT TIP
9032 METAL CONTACT TIP
3404
31
3403
Clamp Sensors
Clamp Sensors Index
Wide-band frequency, high-precision, ideal for observing waveforms (for AC / DC)
3272
3269
3273-50/3276
3274
3275
3272/3269
9274
9276
3270
DC to 50 MHz /
100 MHz
15 A / 30 A maximum
0.1 V / A output
ø 5 mm core jaw dia.
. . . . . . . . . . . . . . p.33
DC to 10 MHz
150 A maximum
0.01 V / A output
ø 20 mm core jaw dia.
. . . . . . . . . . . . . . p.33
DC to 2 MHz
500 A maximum
0.01 V / A output
ø 20 mm core jaw dia.
. . . . . . . . . . . . . . p.33
Power supply for 3273-50,
3274, 3275, 3276
Single sensor (3272)
Four sensors (3269)
. . . . . . . . . . . . . . p.33
DC to 10 MHz
20A rated
2V / 20A output
ø5mm core jaw dia.
DC to 1 MHz
150A rated
1.5V / 150A output
ø20mm core jaw dia.
Power supply &
amplifier for 9274,9276
1V / range output Single
sensor connectable
High-precision sensors to view waveforms or to use with power meters (for AC/DC, or AC only)
9278
9279
9270
9272
9555
DC to 100 kHz
20 A rated
2 V / 20 A output
ø 20 mm core jaw dia.
DC to 100 kHz
200 A rated
2 V / 200 A output
ø 20 mm core jaw dia.
DC to 20 kHz
500 A rated
2 V / 500 A output
ø 40 mm core jaw dia.
5Hz to 50kHz, 20A
rated
2V / 20A output
ø 20 mm core jaw dia.
10 Hz to 10 kHz
20 or 200 A rated
2 V / 20 or 200 A output
ø 46 mm core jaw dia.
. . . . . . . . . . . . . . p.34
. . . . . . . . . . . . . . p.34 . . . . . . . . . . . . . . p.34 200A rated
2V / 200A output
Power supply for 9270,
9271, 9272, 9277, 9278,
9279
Single sensor
connectable
. . . . . . . . . . . . . . p.34
9271
For power lines (50/60 Hz use)
9010, 9010-10
40Hz to 1kHz
10A to 500A range
200mV / range output
ø46mm core jaw dia.
9018, 9018-10
40Hz to 3kHz
10A to 500A range
200mV / range output
ø46 mm core jaw dia.
. . . . . . . . . . . . . . p.34
. . . . . . . . . . . . . . p.34
9132, 9132-10
40Hz to 1kHz
20A to 1000A range
200mV / range output
ø55 mm core jaw dia.
9005-01
50Hz or 60Hz
3A to 300A range
300mV / range output
ø46 mm core jaw dia.
Clamp Sensors
9277
Conversion Adapter
9290-10
CT-101A
9657-10/9675
AC current up to 1500 A,
secondary current 1/10 of
primary, ø 55 mm dia. or
88 mm width
Superior phase angle
characteristics
AC current up to 15 A,
secondary current 1/1
or 10 times of primary,
ø 25 mm dia.
9657-10
LEAK CLAMP ON SENSOR
AC 25mV/A
Up to ø 40mm
9675
LEAK CLAMP ON SENSOR
AC100mV/A
Up to ø 30mm
Clamp Sensors for 3169/3196 Power meters Clamp Sensors for 8205-10/8206-10, 3636-20
9669
9660
9661
9667
45Hz to 5kHz (±1%)
AC current up to 100A
AC 1mV / A output
. . . . . . . . . . . . . . p.37
45Hz to 5kHz (±1%)
AC current up to 500A
AC 1mV / A output
10Hz to 20kHz (±3dB) 40Hz to 1kHz (2.0%)
AC current up to
AC current up to 1000 A
5000/500A
AC 0.5mV / A output
AC 0.1mV / A, AC
1mV / A output
9694
45Hz to 5kHz (±1%)
AC current up to 5A
AC 10mV / A output
. . . . . . . . . . . . . . p.37
9650
9651
9668
40Hz to 1kHz (±8%)
AC current up to 100A
Secondary current
100mA AC
40Hz to 1kHz (±3%)
AC current up to 500A
Secondary current
500mA AC
40Hz to 1kHz (±3%)
AC current up to
1000A
Secondary current
1000mA AC
. . . . . . . . . . . . . . p.37 . . . . . . . . . . . . . . p.37 . . . . . . . . . . . . . . p.37 . . . . . . . . . . . . . . p.13 . . . . . . . . . . . . . . p.13 . . . . . . . . . . . . . . p.13
32
3273-50 3274 3275 3276
3269/3272
CLAMP ON PROBE
POWER SUPPLY
Wide-range current probe allows direct input to oscilloscope
●3273-50/3276: Wide Band from DC to 50/100 MHz, For Large Current
Measurements (30 A rms)
●3274/3275: Wide Band from DC to 10/2 MHz, For Large Current
Measurements (150/500 A rms)
3275
3274
3276: CLAMP ON PROBE
3273-50
Wide-band model from DC to 100 MHz
SPECIFICATIONS
Frequency bandwidth
Rise time
Continuous maximum
input range
Maximum peak current
Clamp Sensors
Output voltage rate
Amplitude accuracy
Noise
Sensitivity temperature
characteristics
Maximum rated
Power supply voltage
Ambient conditions for usage
External magnetic field
resistance
Maximum voltage in
measurement circuit
Measurement conductor
Dimensions and mass
Cable length
Supplied accessories
3273-50
DC to 50MHz (-3dB)
7 ns or less
30A rms
3276
DC to 100 MHz (-3dB)
3.5 ns or less
30A rms
3274
DC to 10MHz (-3dB)
35 ns or less
150A rms
3275
DC to 2MHz (-3dB)
175 ns or less
500A rms
Non-continuous 50A peak
Non-continuous 50 Apeak
Non-continuous 700A peak
0.1V/A
±1.0% rdg. ±1 mV (0 to 30 A)
(DC, 45 to 66Hz)
2.5mA rms or less (measured with
20MHz bandwidth equipment)
Within ±2%
(from 0 to 40°C)
5.6VA
±12V ±0.5V
0 to 40°C, max. 80% rh (no condensation)
Max. 20mA (equivalent)
(DC and 60Hz, Magnetic field of 400A/m)
300V, CAT-I
(insulated conductor)
Diameter max. 5mm
Sensor: approx. 175W×18H×40D mm; 230g
Termination unit: approx. 27W×55H×18D mm
Sensor cable: approx. 1.5 m (BNC connector)
Power cable: approx. 1 m
Soft case 1
0.1V/A
±1.0% rdg. ±1 mV (0 to 30 A)
(DC, 45 to 66 Hz)
2.5mA rms or less (measured with
20MHz bandwidth equipment)
Within ±2%
(from 0 to 40°C)
5.3 VA
±12 V ±0.5 V
0 to 40°C, 80% rh or less (no condensation)
Max. 5 mA (equivalent)
(DC and 60 Hz, Magnetic field of 400 A/m)
300 V, CAT-I
(insulated conductor)
Diameter max. 5 mm
Sensor: approx. 175W×18H×40D mm; 240g
Termination unit: approx. 27W×55H×18D mm
Sensor cable: approx. 1.5 m (BNC connector)
Power cable: approx. 1 m
Hard case 1
Non-continuous 300A peak
500A peak at pulse width of ≤ 30µs
0.01V/A
±1.0% rdg. ±1mV (0 to 150A / DC, 45 to 66Hz)
±2.0% rdg. (150A to 300A peak / DC, 45 to 66Hz)
25mA rms or less (measured with
20MHz bandwidth equipment)
Within ±2%
(At 55Hz/150A input, 0 to 40°C)
5.5VA (Input within the maximum input range.)
±12V ±1V
0 to 40°C, max. 80% rh (no condensation)
Max. 150mA (equivalent)
(DC and 60Hz, Magnetic field of 400A/m)
600 V CAT-II, 300 V CAT-III
(insulated conductor)
Diameter max. 20 mm
Sensor: approx. 176W×69H×27D mm; 500g
Termination unit: approx. 27W×55H×18D mm
Sensor cable: approx. 2 m (BNC connector)
Power cable: approx. 1 m
Hard case 1
0.01V/A
±1.0% rdg. ±5mV (0 to 500A / DC, 45 to 66Hz)
±2.0% rdg. (500A to 700A peak / DC, 45 to 66Hz)
25mA rms or less (measured with
20MHz bandwidth equipment)
Within ±2%
(At 50Hz/500A input, 0 to 40°C)
7.2VA (Input within the maximum input range.)
±12V ±0.5V
0 to 40°C, max. 80% rh (no condensation)
Max. 800mA (equivalent)
(DC and 60Hz, Magnetic field of 400A/m)
600 V CAT-II, 300 V CAT-III
(insulated conductor)
Diameter max. 20 mm
Sensor: approx. 176W×69H×27D mm; 520 g
Termination unit: approx. 27W×55H×18D mm
Sensor cable: approx. 2 m (BNC connector)
Power cable: approx. 1 m
Hard case 1
● Optional accessories
3269/3272 POWER SUPPLY
Please specify voltage when ordering for use with 120 V, 220 V, or 240 V.
Use the 3269/3272 power supply for general
measurements or when power is not available
from the oscilloscope.
3269
(Four sensors)
33
3272
(Single sensor)
3273-50, 3274, 3275, 3276 CLAMP ON PROBE
3269/3272 SPECIFICATIONS
Suitable sensor model
3273-50, 3274, 3275, 3276 CLAMP ON PROBE
Number of power supply 1 (3272), 4 (3269)
connectors
(connector type: LEMO inc./ FFA..0S.304.CNAC42Z)
Output voltage
±12 V ±0.5 V
Ambient conditions for
0 to 40°C, max. 80%rh (no condensation)
usage
Power requirements
Please specify voltage when ordering for use with 120V,
220V, or 240V.
Maximum rated power
20VA max. (3272)
consumption
170VA max. (3269)
Dimensions and mass
Approx. 73W × 110H × 186D mm; 1.1kg (3272)
Approx. 80W × 119H × 200D mm; 1.2kg (3269)
Power cord×1, spare fuse×1
Supplied accessories
9277 9278 9279
9555
UNIVERSAL CLAMP ON CT
SENSOR UNIT
High-precision sensors to view waveforms
or to use with power meters (for AC/DC)
●9277/9278/9279: Clamp on sensors
●Wide frequency ranges including DC
●Use together with the 9555 SENSOR UNIT for current
waveform monitoring (with a waveform recorder or
oscillograph)
9279
(Non-CE mark product)
9278
9277
9277, 9278
SPECIFICATIONS
Rated current
Frequency band width
Accuracy
(DC or 45 to 66 Hz)
Output rate
(with the 9555)
Max. circuit voltage
Core jaw dia.
Power supply
Dimensions, mass
Accessories
9277
20 A AC/DC (continuous 50 A)
DC to 100 kHz (±5 % f.s.)
±0.5 % rdg. ±0.05 % f.s., phase ±0.2°
9278
200 A AC/DC (continuous 350 A)
DC to 100 kHz (±5 % f.s.)
±0.5 % rdg. ±0.05 % f.s., phase ±0.2°
9279(Non-CE mark product)
500 A AC/DC (continuous 650 A)
DC to 20 kHz (±5 % f.s.)
±0.5 % rdg. ±0.05 % f.s., phase ±0.2°
2 V/rated current range
(waveform output, with the 9555)
600 V peak (insulated wire)
φ20 mm
9555 SENSOR UNIT (required)
176W×63H×34D mm,
430 g, cord length: 3 m
9375 CARRYING CASE (1)
2 V/rated current range
(waveform output, with the 9555)
600 V peak (insulated wire)
φ20 mm
9555 SENSOR UNIT (required)
176W×63H×34D mm,
430 g , cord length: 3 m
9375 CARRYING CASE (1)
2 V/rated current range
(waveform output, with the 9555)
600 V peak (insulated wire)
φ40 mm
9555 SENSOR UNIT (required)
220W×103H×43.5D mm,
860 g , cord length: 3 m
9375 CARRYING CASE (1)
● Optional accessories
power supply for the
9270/9271/9272/9277/9278/9279, single
sensor connectable
9010 9010-10 9018 9018-10
Easy-to-use, current to voltage transformer
●Intended primarily for application as an input sensor for a
recorder
●A secondary output provides an AC voltage waveform, and can
be connected to an input impedance over 1MΩ
9018
9010
9010-10
Clamp Sensors
9555 SPECIFICATIONS
Suitable models
9270/9271/9272/9277/9278/9279, 1 unit max.
85 to 250 V AC, 47 to 440 Hz (universal type)
Power supply
50W × 100H × 180D mm,
Dimensions, mass
700 g
Power cord (1), Fuse (1), 9177 CONNECTION CORD (1),
Accessories
Setting parts (1)
9555
POWER SUPPLY
CLAMP ON PROBE
9010, 9010-10 : SPECIFICATIONS
500 A AC (10 A to 500 A full scale, 6 ranges selectable)
Rated current
Frequency band width 40 to 1 kHz (±6 % max.)
Accuracy (at 50 or 60 Hz) ±3 % f.s.
0.2 V AC at full scale range (waveform output)
Output rate
Insulated BANANA (Model 9010),
Output terminal
Insulated BNC (9010-10)
Max. circuit voltage 600 V AC rms (insulated wire)
φ46 mm, or 50 mm×20 mm busbar
Core jaw dia.
None
Power supply
74W×184H×37D mm,
Dimensions, mass
420 g , cord length: 3m
9018, 9018-10 : SPECIFICATIONS
500 A AC (10 A to 500 A full scale, 6 ranges selectable)
Rated current
Frequency band width 40 to 3 kHz (±1 % max.), phase ±2.5°max.
Accuracy (at 50 or 60 Hz) ±1.5 % rdg. ±0.1 % f.s.
0.2 V AC at full scale range (waveform output)
Output rate
Insulated BANANA terminal(9018), Insulated BNC(9018-10)
Output terminal
Max. circuit voltage 600 V AC rms (insulated wire)
φ46 mm, or 50 mm×20 mm busbar
Core jaw dia.
None
Power supply
74W×184H×37D mm,
Dimensions, mass
410 g , cord length: 3 m
9018-10
34
Power Measuring Instruments
Power Measuring Instruments Index
For high level performance
3193
3194
3196
9624/9624-10
DC, or Single-phase to 3-phase
4-wire Wide-band up to 1 MHz
6 ch-Direct/ Clamp input
Analysis station for Motor
Evaluation Power, Harmonics,
Rotation Speed, Torque,
Converter efficiency
Power quality analyzer
DC, or Single-phase to 3-phase
4-wire Clamp input
PQA- HiVIEW
Software application for 3196
. . . . . . . . . . . . . . . . . . . . . . p.39
. . . . . . . . . . . . . . . . . . . . . . p.39
. . . . . . . . . . . . . . . . . . . . . . p.37
. . . . . . . . . . . . . . . . . . . . . . p.38
For use on production lines
Power Measuring Instruments
35
Wide bandwidth, multi-purpose models
3186
3331
3332
3187
3167
Single-phase V,A,W,VA
Direct input only
Single-phase to 3-phase 3-wire
V,A,W,VA,var,integ., PF Phase
angle, Hz Direct input only
Single-phase 2-wire V,A,W,VA,
var,integ., PF Phase angle, Hz
Direct input only
DC, or Single-phase V,A,W,VA,
var,integ., PF Phase angle, Hz
Direct/ Shunt input
DC, or Single-phase V,A,W,VA,
var,integ., PF Phase angle, Hz
Clamp input only
. . . . . . . . . . . . . . . . . . . . . . p.39
For managing power lines
3169 -20/-21
9625
3286 -20
Single-phase to 3-phase 4-wire
V,A,W,VA,var,integ., PF, Hz
Clamp input only
POWER MEASUREMENT
SUPPORT SOFTWARE
for 3169-20/21
Clamp-On Power Meter
V,A,W,VA,var, PF,Hz, Hamonics
(V, A)
. . . . . . . . . . . . . . . . . . . . . . p.36
. . . . . . . . . . . . . . . . . . . . . . p.36
. . . . . . . . . . . . . . . . . . . . . . p.44
3169-20 3169-21
CLAMP ON POWER HiTESTER
Offering a new approach to energy-related
measurement such as energy conservation,
ISO14001 testing, equipment diagnosis, and
harmonics measurement.
CAT III 600V
STANDARD
●Measure up to two 3-phase, 3-wire systems
●Measure up to four single-phase, 2-wire systems
●5 A to 5000 A range, PC card data storage
●Power recording for individual waveforms
●Simultaneous recording of demand values and harmonics
9661×2, 9669×2 (option)
Single-phase 2-wire, single-phase 3-wire, three-phase 3-wire,
and three-phase 4-wire systems (50/60 Hz)
Measurement item
Voltage, Current, Active power, Reactive power,
Apparent power, Power factor, Iintegrated value, Frequency,
Harmonics
Measurement range
Voltage: 150 V to 600 V, 3 ranges
Current (When using 9660): 5 A to 100 A, 4 ranges
Current (When using 9661): 5 A to 500 A, 5 ranges
Current (When using 9669): 100 A to 1 kA, 3 ranges
Current (When using 9667): 500 A / 5 kA, 2 ranges
Current (When using 9694): 0.5 A to 5 A, 3 ranges
Power: 75 W to 9 MW, 108 combination patterns
±0.2% rdg. ±0.1% f.s. + Clamp accuracy (for active power)
Basic accuracy
Clamp sensor accuracy 9660 (rated for 100 A): ±0.3% rdg. ±0.02% f.s.
9661 (rated for 500 A): ±0.3% rdg. ±0.01% f.s.
9669 (rated for 1000 A): ±1.0% rdg. ±0.01% f.s.
9667 (rated for 5000 A): ±2.0% rdg. ±1.5 mV
9694 (rated for 5 A): ±0.3% rdg. ±0.02% f.s.
Frequency characteristic Fundamental waveforms up to the 50th order ±3% f.s.
+ measurement accuracy
Other functions
PC card, RS-232C, D/A output (3169-21 only, 4 channels),
External I/O
Power supply voltage rating 100 to 240 V AC, 50/60 Hz
Dimensions, mass
210W×160H×60D mm ±5 mm, 1.2 kg ±100 g
(3169-20, 3169-21)
Accessories
9438-03 Voltage cord set (1), Power cord (1),
Input cord label (1), Operating manuals (2), CD-R (1),
9441 Connection cable (1) (for the 3169-21 only)
OPTIONS
● Current measurement
(The 3169-20/-21 cannot be used alone. Measurement requires one or more optional
clamp-on sensors.)
9660 CLAMP ON SENSOR rated current 100 A AC
9661 CLAMP ON SENSOR rated current 500 A AC
9669 CLAMP ON SENSOR rated current 1000 A AC
9667 FLEXIBLE CLAMP ON SENSOR rated current 5000 A AC
9694 CLAMP ON SENSOR rated current 5 A AC
9290-10 CLAMP ON ADAPTER rated current 1500 A AC, output 150 A (10:1 ratio)
● Voltage measurement
9438-03 VOLTAGE CORD (Supplied as standard with 3169-20/-21)
● PC communication
9625 POWER MEASUREMENT SUPPORT SOFTWARE
9612 RS-232C CABLE for connection to PC
9626 PC CARD 32M
9627 PC CARD 64M
9726 PC CARD 128M
9727 PC CARD 256M
9728 PC CARD 512M
Power Measuring Instruments
SPECIFICATIONS
Measurement lines
● Other options
9720 CARRYING CASE
9440 CONNECTION CABLE for external I/O, 2 m length
9441 CONNECTION CABLE (3169-21 standard), for D/A output, 2 m length
● Printer
9442 PRINTER
9443-02 AC ADAPTER for the 9442 PRINTER, EU type
9443-03 AC ADAPTER for the 9442 PRINTER, USA type
9721 RS-232C CABLE for connection to the 9442, 1.5 m length
1196 RECORDING PAPER 112 mm width×25 m, roll type, 10 rolls per set
9442 PRINTER
Print method : Thermal serial dot printing
Paper width : 112 mm
Print speed : 52.5cps
Power supply : 9443-02/03 AC adapter, or supplied nickel-metal hydride battery
(approx. 3000 lines of printing when fully charged and used
with the 9443-02/03)
Dimensions, mass : Approx.160W ✕ 66.5H ✕ 17D mm, approx.580g
When purchasing the 9442 printer, make sure you also purchase the 9721 RS-232C
cable and 9443-02/03 AC adapter so that you can connect it to the 3169-20/21.
9625
POWER MEASUREMENT
SUPPORT SOFTWARE
36
3196
POWER QUALITY ANALYZER
Monitor and record the quality of power to
analyze the cause of trouble when it occurs !
CAT III 600V
Most complete instrument for power
quality troubleshooting
●Measure Dips, Swells, Interruptions,
Flicker, Transients
●Harmonic to the 50th order
●High frequency transient over voltage
●Detection and waveform display
●4 current and 4
voltage channels
STANDARD
STANDARD
9661×4 (option)
SPECIFICATIONS
Measurement lines
Voltage range
Power Measuring Instruments
Current range
Measurement
Method
Measurement
Function
Internal memory
Interface
Power supply
Dimensions, mass
Accessories
37
Single-phase two-wires, Single-phase three-wires,
Three-phase three-wires, Three-phase four-wires
ch1, ch2, ch3:150/300/600V
ch4: 60/150/300/600V(AC), 60/600V(DC)
9660: 50/100A, 9661:50/500A, 9667:500/5000A, 9669:1000A
Transient overvoltage: 2MHz/s
Arithmetic operation: 256points/cycle
Harmonic/Inter-harmonic:
2048points/10cycles (for 50Hz)
2048points/12cycles (for 60Hz)
1.Transient overvoltage
2.Voltage swell, Voltage dip, Voltage interruption
3.Frequency, Voltage, Current, Voltage/Current peak,
Active/Reactive /Apparent power, Power factor, DPF
4.Voltage unbalance ratio, Current unbalance ratio
5.Harmonic voltage/current/power,
Inter harmonic voltage/current,
Harmonic voltage current phase angle,
Total harmonic/inter harmonic distortion
6.IEC flicker[Pst, Plt], K factor
/ ∆V10(Japan)
13MB
PC card (Flash ATA card / up to 528 MB)
RS-232C, LAN (10BASE-T), HTTP server function
9458 AC adapter or 9459 battery pack
298W×215H×67D mm, 2.0 kg
Voltage cord (1set), 9458 AC adapter (1), 9459 Battery pack (1),
strap (1)
9660 CLAMP ON SENSOR
Curent up to 100A
9694 CLAMP ON SENSOR
Current up to 5A AC
9667 FLEXIBLE CLAMP-ON SENSOR
Current up to 5000A AC Diameter up to
254mm Rogovski-type current sensor
9661 CLAMP ON SENSOR
Current up to 500A
9669 CLAMP ON SENSOR
Current up to 1000A
OPTIONS
(The 3196 cannot be used alone. Measurement requires one or more optional sensors.)
● Current measurement
9660 CLAMP ON SENSOR rated current 100A AC
9661 CLAMP ON SENSOR rated current 500A AC
9669 CLAMP ON SENSOR rated current 1000A AC
9667 FLEXIBLE CLAMP ON SENSOR rated current 5000A AC
9694 CLAMP ON SENSOR rated current 5A AC
9290-10 CLAMP ON ADAPTER rated current 1500A AC, output 150A (10:1 ratio)
9445-02 AC ADAPTER (for the 9667, for America, Japan)
9445-03 AC ADAPTER (for the 9667, for Europe)
● Voltage measurement
9438-02 VOLTAGE CORD (standard accessory)
9264-01 WIRING ADAPTER (3P3W)
9264-02 WIRING ADAPTER (3P4W)
● PC communication
9624 PQA HiVIEW (PC application software)
9624-10 PQA HiVIEW PRO (PC application software for advanced data processing)
9642 LAN CABLE (5m, with straight and crossover connectors)
9626 PC CARD 32 M
9627 PC CARD 64 M
9726 PC CARD 128 M
9727 PC CARD 256 M
9728 PC CARD 512 M
● Other options
9458 AC ADAPTER (included)
9459 BATTERY PACK (included)
9339 CARRYING CASE (soft)
9340 CARRYING CASE (hard)
● Printer
9670 PRINTER (with one roll of recording paper)
9671 AC ADAPTER (for 9670)
9672 BATTERY PACK (for 9670)
9673 BATTERY CHARGER (for 9672)
9237 RECORDING PAPER (80 mm×25 m, 4 rolls, for 9670)
9638 RS-232C CABLE (1.5 m, for printer connection)
9624 PQA-HiVIEW
Model 9624
PQA-HiVIEW
software application
is a program for
analyzing binary
codes stored on a
PC card by Model
3196.
9624
PQA-HiVIEW
Measurement data is saved
in binary format
PC card
Data processing is quick
and easy using the 9624
PQA-HiVIEW software
Choose from 2 Easy-to-Use Application Software
Packages for Further Data Analysis
● Viewer function
Use this function to display screens similar to those used for the
3196.
Select from the TIME PLOT screen (voltage fluctuation, RMS
fluctuation, harmonic fluctuation, inter-harmonic fluctuation),
event list screen, event data screen (waveforms, vectors, DMM,
harmonics, event details), ∆V10 screen(Japanese standard), or
settings screen. In the TIME PLOT screen, and use the two cursors
(A and B) to calculate waveforms within a specified interval.
● Demand/integral power consumption function
Calculate demand and integral power consumption from TIME PLOT
data for effective power.
■ TIME PLOT screen
Modem
LAN
Data stored on a PC card can be
transferred using remote operations
via a LAN or modem
• Use it to create reports
• Use data converted to
CSV format
● Binary CSV format conversion function
Convert binary data into CSV format for event waveforms
within the specified range in the TIME PLOT screen or event
waveforms selected in the event waveform screen. Files saved
in CSV format can be used with spreadsheet software on your
PC.
● Print function
Use this function in each screen to output reports to a printer
connected to your PC.
This screen enables you to select four different types of data, including RMS
fluctuation, voltage fluctuation, harmonic fluctuation, and inter-harmonic
fluctuation data, and display the data in graphs corresponding to the TIME
PLOT screen of the 3196.
■ Spot analysis using the cursor
Conduct spot analysis of time series data using
the A and B cursors.
■ Event list screen
This screen displays an event list
corresponding to the event list on the 3196.
Click an event!
Calculations for the details within the
specified interval are displayed.
■ Event data screen
1. Displays detailed data for the
event that you selected in the
event list.
9624-10
2. Displays nine different screens that correspond to the
VIEW screen on the 3196, such as the waveform,
vector, harmonics, and DMM screens.
Power Measuring Instruments
View power, voltage, and current data at
a single glance in the DMM screen!
PQA-HiVIEW PRO
Advanced functions added to the standard Model 9624
● ITIC curve display function
Make ITIC (CBEMA) curve analyses (limit curve) based on the
power quality control standards of the U.S.A.
●EN50160 display functions
(applicable standard is EN50160:1999)
Effectively evaluate and analyze the quality of power according to
EU standards.
● Downloading from LAN
Data (BINARY/TEXT/BMP) recorded on a PC card or the internal
memory of the 3196 can be downloaded via LAN to a personal
computer. (*This can be done without use of the freeware
Down96. Measurement on the 3196 must be halted during
download.)
● Report generation function
Choose from 3 types of report generation settings to take care
of all the troublesome reporting operations, and either send the
data to a printer or save as a Rich Text file. (Automatic: Output
basic items. Individual setting: Select any item for output. Detailed
setting: Specify a time-series graph in details for output.)
● Positive phase, negative phase, and zero phase function
Recalculate event data captured by 3P4W circuits, and display
each component of the voltage/current of the positive phase,
negative phase, and zero phase.
38
3332
SPECIFICATIONS
Measurement lines
Measurement items
POWER HiTESTER
Single-phase/two-wire
Voltage, Current, Current peak, Active power, Apparent power,
Reactive power, Power factor, Phase angle, Frequency, Power
integration, Current integration
Voltage: 15.000V to 600.00V, auto or 6 ranges
Measurement ranges
Current: 1.0000mA to 50.000A, auto or 15 ranges
Power: 15.000mW to 30.000kW, auto or 90 ranges
Frequency: 4Hz to 100kHz, auto or 2 ranges
0 to±999999MAh/ MWh, (integration time up to 10000 hours)
Integration range
Wave peak measurement Current (displays maximum absolute value)
Basic accuracy
±0.1% rdg. ±0.1% f.s. (active power, at 45 to 66Hz)
Frequency characteristics 1Hz to 100kHz (1Hz to 100kHz at 10A or less, 10Hz to 10kHz
at 20A to 30A, 50/60Hz at 30A to 50A )
Analog level: Voltage, current, active power, 5V DC f.s.
Signal output
Waveform monitor: Voltage, current, 1V rms f.s.
Comparator function for 2 items
Other function
Scaling function (PT /CT/SC), Displays a moving average
function, GP-IB/RS-232C interface
5 times /second
Sampling rate
100 to 240 V AC, 50/60 Hz
Power supply
210W×100H×261D mm, 2.5 kg
Dimensions, mass
Accessories
Power cord (1), Connector for EXT I/O (1)
Measure very low effective power, for stand-by
mode of home use equipment
●Ultra high-sensitive measurement, for use to measure
the effective power of equipment in stand-by mode:
Current 1.0000 mA full-scale, 0.1 µA resolution
●Wide measurement range, up to 50.000 A direct input
OPTIONS
9151-02 GP-IB CONNECTION CABLE (2 m)
9151-04 GP-IB CONNECTION CABLE (4 m)
9442 PRINTER
9443-02 AC ADAPTER (for the 9442, EU), 9443-03 (USA)
9444 CONNECTION CABLE (for the 9442)
1196 RECORDING PAPER
(25 m, 10 rolls /1 set, for the 9442)
STANDARD
3193
STANDARD
SPECIFICATIONS
Measurement lines
POWER HiTESTER
Single-phase/two-wires to three-phase/four-wires
(Through the use of various input units)
(When using 9600, 9601, 9602, optional) Voltage, current, voltage
/current peak, effective/reactive/apparent power, power factor, phase,
frequency, current/power integration, load rate, efficiency (When
using 9603, optional) Voltage, torque, RPM, frequency, motor output
(When using 9605, optional) Harmonic, waveform, voltage fluctuation
/ flicker measurement function
Measurement ranges
Voltage: 6.0000V to 1.0000kV (depends on use of the input unit)
Current: 200.00mA to 500.00A (depends on use of the input unit)
Power: Depends on combination of voltage and current ranges
Integration range
0 to ±999999 TAh/ TWh, (integration time up to 10000 hours)
Basic accuracy
±0.1% rdg. ±0.1% f.s. (voltage, current, power, at 45 to 66Hz)
used with 9600 to 9602 Input unit Note: When used together with the 9270 to 9272, and 9277 to 9279, resulting
accuracy is the sum of that indicated above and the accuracy of these clamp sensors.
Frequency characteristics using with 9600: DC, 0.5Hz to 1MHz using with 9601: 5Hz to 100kHz
using with 9602: DC, 0.5 Hz to 200 kHz (depends on the clamp accuracy)
Signal output
Analog level: Voltage, current, active power, 5V DC f.s.
Waveform monitor: Voltage, current, 1V rms f.s.
D/A output: Outputs 8 arbitrarily selected items, DC±5V f.s.
Other functions
6.4 inch TFT color LCD, RMS/MEAN rectification, FDD,
GP-IB/RS-232C interface, scaling, averaging
Power supply
100/120/200/230 V AC (switched automatically), 50/60 Hz
Dimensions, mass
430W×150H×370D mm, 15 kg (with all options)
Accessories
Power cord (1), Connector for EXT I/O (1)
Measurement items
Wide spectrum power meter for comprehensive
device assessment
STANDARD
STANDARD
Power Measuring Instruments
OPTIONS
(The 3193 cannot be used alone. Measurement requires one or more input units.)
9600 AC/DC DIRECT INPUT UNIT
9604 PRINTER UNIT
9601 AC DIRECT INPUT UNIT
9605 HARMONIC / FLICKER MEASUREMENTS UNIT
*9602 AC/DC CLAMP INPUT UNIT
*9270 series CLAMP ON SENSOR
9603 EXTERNAL SIGNAL INPUT UNIT *9290-10 CLAMP ON ADAPTER
*Note: 9600 to 9605 : factory installation only
*Certain Clamp-on sensors are not CE-mark
compliant. Please refer to p.32
3194
MOTOR/HARMONIC HiTESTER
Analysis Station Extends Reach
of Motor Evaluation!
[Strain gauge-type torque
sensor is directly connected
to 9603-01]
3ø3W
Comprehensive measurement of power,
rotation speed, torque, converter efficiency,
and harmonics, all with a single unit
3194 Performs Comprehensive Evaluation of
3-phase Inverter Motors
Using the 9603-01 EXTERNAL SIGNAL INPUT
UNIT, a torque sensor (strain gauge) is directly
connected to chA. By inputting the output of a
tachometer (analog signal or pulse signal) to
chB, a system for measuring torque, rpm and
motor power can be obtained.
39
Inverter
Motor
Torque
Sensor
Load
r /min
Torque Output
1, 2, 3 ch
4, 5, 6 ch
chB
chA
With direct or clamp input unit,
capable of measuring from micro
motors up to large-size motors
1, 2, 3 ch
Primary V/A/W/PF,
Harmonic analysis
4, 5, 6 ch
Secondary V/A/W/PF and motor output
measurement
Efficiency (primary to secondary, secondary
to motor output and primary to motor output)
Secondary harmonic analysis
Supports measurements of everything from
micro motors used for household appliances
and OA equipment up to industrial large-size
motors. Also supports various applications
such as harmonic measurement of
equipment power sources and power quality
measurement.
Clamp Testers
Clamp Testers Index
Current Meters
(for AC only, basic type)
3127-10
3280-10
AC current, up to 300A,
ø 33 mm dia.
AC current, up to 1000A, AC current, up to 1000A, AC current, up to 600A,
ø 33 mm dia., light 100g ø 33 mm dia., light 100g ø33mm dia., Multi-function
. . . . . . . . . . . . . . . p.43
and slim 16 mm
and slim 16 mm
True RMS
3282
3128-10
3280-20
AC current, up to 1500A,
ø 55 mm dia.
. . . . . . . . . . . . . . . p.45 . . . . . . . . . . . . . . . p.42
Current Meters
. . . . . . . . . . . . . . . p.42
3281
3283
3286-20
Leak current, high-sensitivity
10mA range 10µA resolution
Load current up to 200A,
ø40mm dia.
Clamp-On Power Meter
V,A,W,VA,var, PF,Hz,
Harmonics (V, A only)
ø55mm dia. or 80mm
busbar
. . . . . . . . . . . . . . . p.44
AC current, up to 1000A,
ø46mm dia., Multi-function . . . . . . . . . . . . . . . p.43
. . . . . . . . . . . . . . . p.43
(for AC/DC, two-way type)
3287
9692
9693
DC and AC current, up
to 200 A, ø 33 mm dia.,
Multi-function
. . . . . . . . . . . . . . . p.43
AC/DC current, up to 100 DC and AC current, up to 3290(3290-10)+9691
A, ø 35 mm dia., True
2000A
Measure up to 100 A
RMS rectifier
ø35 mm
Choice of three sensors
. . . . . . . . . . . . . . . p.42
3290/3290-10
3290(3290-10)+9692
Measure up to 200 A
ø33 mm
3290(3290-10)+9693
Measure up to 2000 A
ø55 mm
3285
3288
. . . . . . . . . . . . . . . p.45
. . . . . . . . . . . . . . . p.45
DC and AC current, up
to 2000 A, ø 55 mm dia.,
Multi-function
. . . . . . . . . . . . . . . p.43
AC/DC current, up to
1000 A, ø 35 mm dia.,
Average rectifier
. . . . . . . . . . . . . . . p.42
. . . . . . . . . . . . . . . p.45
Conversion Adapter
9290-10
CT-101A
AC current up to 1500 A,
secondary current 1/10 of
primary, ø 55 mm dia. or
88 mm width
Superior phase angle
characteristics
AC current up to 15 A,
secondary current 1/1 or
10 times of primary,
ø 25 mm dia.
9691
. . . . . . . . . . . . . . . p.45
(for AC only, clamp-on type)
Clamp Testers
3284
40
CLAMP ON HiTESTERs
Table of functions
AC Current ranges Other current ranges AC Voltage ranges Other voltage ranges
3280-10
3280-20
True RMS
3281
True RMS
3282
True RMS
3283
True RMS
3284
True RMS
3285
True RMS
3286-20
True RMS
3287
True RMS
3288
3290
True RMS
3290-10
True RMS
3127-10
3128-10
42.00 to 1000A AC,
3 ranges
3280-10: Average
rectifier effective value
3280-20: Effective
value rectifier
3281: 30.00 to 600A
AC, 3 ranges
3282: 30.00 to 1000A
AC, 3 ranges
Effective value rectifier
Clamp Testers
3282
None
AC, AC+DC(RMS or
Peak value)
3284: 20.00 or 200.0A
AC, 2 ranges
3285: 200.0 or 2000A
AC, 2 ranges
Effective value rectifier
DC (Average or Peak
value)
3284: 20.00 or 200.0A
DC, 2 ranges
3285: 200.0 or 2000A
DC, 2 ranges
AC, AC+DC
(RMS or Peak value)
30.00 to 600V AC,
3ranges
Effective value rectifier
20.00 to 1000 A AC,
3 ranges
Effective value rectifier
3287: 10.00 or 100.0A
AC, 2 ranges
Effective value rectifier
3288: 100.0 or 1000A
AC, 2 ranges
Average rectifier
effective value
3127: 6 to 300A AC,
5 ranges
3128:15 to 1500A AC,
5 ranges
Average rectifier
effective value
True RMS
3284
True RMS
3285
150.0/300.0/600 V AC,
3 ranges
Effective value rectifier
DC mode
3287: 10.00 or 100.0 A
DC, 2 ranges
3288: 100.0 or 1000 A
DC, 2 ranges
3287: 4.200 to 600V
AC, 4 ranges
Effective value rectifier
3288: 4.200 to 600V
AC, 4 ranges
Average rectifier
effective value
None
Sampling rate
True RMS
3287
True RMS
3288
3290
True RMS
3290-10
True RMS
3127-10
3128-10
41
Wave peak value at
AC Voltage
up to 750 or 1000V
peak, 2 ranges
Distortion check:
1 to 5 Crest factor
Resistance:1k or 10kΩ range
Temperature*: -50 to 150°C
Frequency: 30.0 to 1000 Hz
Mode: Slow/Peak/C.F./RMS
Record mode/Auto-off/
Conduction
DC (Average or
Peak value)
30.00 to 600V
DC, 3 ranges
DC mode
420.0 mV to 600 V
DC, 5 ranges
None
150 to 750V AC,
3 ranges
Average rectifier
effective value
DC Voltage range:
75 V DC, 1 range
None
AC current:
±1% rdg. ±5 dgt.
AC voltage:
±1% rdg. ±3 dgt.
Peak: ±3% rdg. ±5 dgt.
Frequency:
±0.3% rdg. ±1 dgt.
DC, or AC 1V / f.s.
(200A range:2V / f.s.)
Level output with
REC mode
Waveform output with
MON mode
10m to 10A range:
±1.0 % rdg. ±5 dgt.
200A range:
±1.5 % rdg. ±5 dgt.
Frequency:
±0.3 % rdg. ±1 dgt.
Frequency:1 to 1000Hz
Mode: Max./Min./AVE.
/RMS
Record mode/Auto-off
DC, or AC 1V / f.s.
Level output with REC
mode
Waveform output with
MON mode
AC current:
±1.3% rdg. ±3 dgt.
AC voltage:
±1.0% rdg. ±3 dgt.
Frequency:
±0.3% rdg. ±1 dgt.
9442 PRINTER
(Option)
Resistance: 420.0 to
42.00MΩ, 6 ranges
Accuracy: ±2.0% rdg.
±4 dgt. (at 420 to 420kΩ
range) Continuity: 420.0Ω
(Buzzer sounds at approx.
50Ω or less)
None
DC,or AC Current : 2V/f.s.
Level output with
REC mode
Waveform output with
MON mode
Integ./Frequency : 1V/f.s.
Frequency :
10 to1000 Hz
None
None
AC current:
±1.5 % rdg. ±5 dgt.
AC voltage:
±2.3 % rdg. ±8 dgt.
DC voltage:
±1.3 % rdg. ±4 dgt.
Continuity:
±2.0 % rdg. ±6 dgt.
Frequency:
30.0 to 1000 Hz
Filter function:
180Hz±30Hz/-3dB
Power(Single-phase or
3 phase): 3kW to
600kW(Single-phase)
6kW to 1200kW(3-phase)
Power factor, Phase angle:
Frequency: 30.0 to 1000Hz
Voltage/current harmonic
levels
None
Accuracy
(at 50 or 60Hz)
Resistance:1k or 100kΩ
range
Temperature*: -50 to
200°C
*9021-01TEMPERATURE
PROBE required,
(sold separately)
Crest factor
(RMS)
Effect of external
magnetic fields
Max. circuit
voltage
Core jaw dia.
None (3280-10)
2.5 (3280-20)
No provision
600V AC rms
φ33 mm
None
Frequency
characteristics
AC current / voltage
AC voltage: 50 to 500Hz
AC current: 50 or 60Hz
(3280-10)
AC current: 40 to 1kHz
(3280-20)
40 to 1000 Hz
40 to 2kHz
3284: DC, 10 to 2kHz
3285: DC, 10 to 1kHz
AC current:
±1.3 % rdg. ±3 dgt.
AC voltage:
±1.0 % rdg. ±3 dgt.
AC current: 45 to 1kHz
Power:
AC voltage: 30 to 1kHz
±2.3% rdg. ±5 dgt.(1φ)
±3.0% rdg. ±10 dgt.(3φ)
(Accuracy guaranteed only
for 50/60Hz cosφ=1)
AC current:
±1.5 % rdg. ±5 dgt.
AC voltage:
±2.3 % rdg. ±8 dgt.
DC voltage:
±1.3 % rdg. ±4 dgt.
DC current:
±1.5 % rdg. ±5 dgt.
Continuity:
±2.0 % rdg. ±6 dgt.
3287 AC current:
DC, 10 to 1kHz
3288 AC current:
DC, 10 to 500Hz
AC voltage:
30 to 500Hz
AC/DC/AC+DC
Current :±1.3 % rdg.
+ 3 dgt. (Typical)
Frequency : ±0.3 % rdg.
+ 1 dgt. (Typical)
DC to 500Hz (9691)
DC to 1kHz(9692, 9693)
±2.3 % rdg. + 8 dgt.
AC current: ±3% f.s.
AC/DC voltage:±3% f.s.
50 or 60 Hz
Power supply
Dimensions/mass
Included
accessories
9208 TEST LEAD (1)
9398 CARRYING
CASE (1)
Digital /LCD,
maximum
4199 dgt.
2.5 times /sec or
1 time /3 sec
Digital /3000 dgt.
Bar graph /35 seg.
3281: 1.5A equivalent
max. at 400 A/m
2 or 4 times /sec
2.5 (1.7 at 600A,
(Slow: 1 time /3 sec) 1000A, 600 V range) 3282: 0.2A equivalent
max. at 400 A/m
600V AC rms
(insulated wire)
3281: φ33 mm
3282: φ46 mm
6F22(006P) ×1
3281: 62W×218H×
39D mm /350 g
3282: 62W×230H×
39D mm /400 g
9207-10 TEST LEAD(1)
9399 CARRYING
CASE(1)
Hand strap(1)
Digital /2000 dgt.
Bar graph /35 seg.
corresponds to 5mA,
2 or 4 times /sec
2.5
max. 7.5 mA
(Slow: 1 time /3 sec) (1.5 at 200A range) equivalent max.
at 400A/m
300 V AC rms
(insulated wire)
φ40 mm
6F22(006P) ×1
or AC adapter
62W×225H×39D
mm/400 g
9399 CARRYING
CASE(1)
Hand strap(1)
Current / 2500 dgt.
Voltage / 3750 dgt.
Bar graph /35 seg.
2 or 4 times /sec
(Slow: 1 time /3 sec)
2.5
1.5
(3284: 200A range),
1.42
(3285: 2000A range)
3284: 0.5A equivalent
max. at 400 A/m
3285: 2.0A equivalent
max. at 400 A/m
600V AC rms
(insulated wire)
3284: φ33 mm
3285: φ55 mm
6F22(006P) ×1
or AC adapter
3284: 62W×230H×
39D mm /460 g
3285: 62W×260H×
39D mm /540 g
9207-10 TEST LEAD(1)
9399 CARRYING
CASE(1 for 3284)
9345 CARRYING
CASE(1 for 3285)
Hand strap(1)
Digital /LCD,
maximum 6000 dgt.
Normal: 1 time /sec
(Slow: 1 time /3 sec)
2.5 (1.7 at 1000 A,
600 V range)
1.00 A equivalent
max. at 400 A/m
600 V AC rms
(insulated wire)
φ55 mm or
80mm busbar
6LR61/6LF22
(006P)×1
100W×287H×39D
mm /650 g
9635 VOLTAGE
CORD(1)
9245 CARRYING
CASE(1)
Hand strap(1)
Digital /LCD,
maximum 4199 dgt.
2.5 times /sec
3287
2.5 or less (150A,
1000V maximum)
3288 None
No provision
600 V AC rms
(insulated wire)
3287: φ35 mm
3288: φ35 mm
3287: 57W×180H×
CR2032 (3VDC) ×1 16D mm /170 g
3288: 57W×180H×
16D mm /150 g
Fast : 4 times / sec,
(3290-10
FAST : 10 times/sec )
Normal : 2 times / sec,
Slow : 1 time / 3sec
2.5 or less
9691 : 0.5 A equivalent
max.at 400 A/m
9692 : 0.7 A equivalent
max.at 400 A/m
9693 : 2.0 A equivalent
max.at 400 A/m
600 V AC rms
(insulated wire)
9691 : φ35mm
9692 : φ33mm
9693 : φ55mm
Type 3 alkaline dry
cell (LR6) ×4
or AC adapter
3290 : 155W×98H×
47D mm / 545g
9691 : 53W×129H×
18D mm / 230g
9692 : 62W×167H×
35D mm / 410g
9693 : 62W×196H×
35D mm / 500g
strap
None
None
600V AC rms
(insulated wire)
3127: φ33 mm
3128: φ55 mm,
or 80 mm busbar
R6P(AA)×1
3127: 78W×190H×
34D mm /340 g
3128: 99W×237H×
34D mm /570 g
9067 TEST LEAD (1)
9351 CARRYING
CASE for 3127(1)
9148 CARRYING
CASE for 3128(1)
True RMS
3286-20
DC Voltage range:
420.0mV to 600V DC,
5 ranges
Resistance: 420.0 to
42.00 MΩ, 6 ranges
Accuracy: ±2.0 % rdg.
±4 dgt. (at 420 to
420 kΩ range)
Continuity: 420.0Ω
(Buzzer sounds at
approx. 50Ω or less)
None
None
3290+9691 : AC 20 to 100A
3290+9691 : DC 20 to 100A
3290+9692 : AC 20 to 200A
3290+9693 : AC 200 to 2000A 3290+9692 : DC 20 to 200A
AC+DC, AC RMS, AC MEAN 3290+9693 : DC 200 to 2000A
True RMS
3283
Wave peak value at AC
Current
300.0 or 600V AC,
3281: 75.0 to 1000A
2 ranges
peak, 3 ranges
Effective value rectifier
3282: 75.0 to 1700A
peak, 3 ranges
None
3281
True RMS
None
10.00mA to 200 A AC,
5 ranges
Effective value rectifier
Display
3280-10
3280-20
True RMS
4.200 to 600V AC,
4 ranges
3280-10 : Average
rectifier effective value
3280-20 : Effective
value rectifier
Analog output
Printer output
Other functions
Digital / LCD
maximum 3000 dgt.
Bar graph / 20 seg.
3290-10
maximum 9999 dgt.
Indicator type
No provision
CR2032 (3 VDC) ×1 57W×175H×
16D mm /100 g
9208 TEST LEAD (1)
9398 CARRYING
CASE (1)
3280-10
CLAMP ON HiTESTER
Easy operation !
●1000 A rms, clamp aperture: 33 mm dia.
●Light 100 g, and Slim 16 mm
●Independent-opening double-lever design
●Slim body allows easy clamping even for narrow conductors
●No metal (iron core) exposure, ensuring enhanced safety
9209 (option)
CAT III 600 V
(Current range)
CAT II 600 V
(Voltage range)
OPTIONS
9209 TEST LEADS HOLDER
*CT-101A LINE SPLITTER
*Note: Non-CE mark product
3280-20
CLAMP ON HiTESTER
CAT III 600 V
(Current range)
CAT II 600 V
(Voltage range)
True RMS !
●1000 A rms, clamp aperture: 33 mm dia.
●Light 100 g, and Slim 16 mm
●Independent-opening double-lever design
●Slim body allows easy clamping even for narrow conductors
●No metal (iron core) exposure, ensuring enhanced safety
OPTIONS
9209 TEST LEADS HOLDER
*CT-101A LINE SPLITTER
*Note: Non-CE mark product
CLAMP ON AC/DC HiTESTER
3288
Compact & easy, one-touch maintenance on
all types of AC/DC equipment
3287
●The 3287 can handle even cogenerator
/ inverter energy-saving equipment
(10/100A)
●Use the 3288 for high current
measurements such as UPS emergency
batteries and train motors (100/1000A)
●A slim core of only 10 mm (0.39") for
easy clamping even in crowded wiring
9209 (option)
Clamp Testers
3287 3288
OPTIONS
9209 TEST LEADS HOLDER
*CT-101A LINE SPLITTER
*Note: Non-CE mark product
3287
CAT III 600 V
(AC/DC A)
CAT II 600 V
(AC/DC V)
42
3281 3282
DIGITAL CLAMP ON HiTESTER
True RMS is shown in the distorted waveform
3282
3281: 600 A AC, ø 33 mm 3282: 1000 A AC, ø 46 mm
●AC Current/Voltage, Frequencies, Resistance, Conduction check,
Temperature (probe sold separately)
●Wave peak value, Waveform distortion check
●Auto-power off to prevent the power from being left on
●Non-fuse type protects up to 600 V AC
3281
CAT IV 600 V
OPTIONS
*9462 THERMISTER TEMPERATURE PROBE
*CT-101A LINE SPLITTER
*Note: Non-CE mark product
3283
CLAMP ON LEAK HiTESTER
Easily monitor leak current fluctuations
OPTIONS
●High-sensitivity with a full scale of 10mA (resolution:10µA)
●High-accuracy at ±1%
●True RMS measurement
●Analyzer functions, for filtering and output signals
●Wide bandwidth, 5Hz to 15kHz (Monitor output)
9445-02 AC ADAPTER (for USA)
9445-03 AC ADAPTER (for EU)
9290-10 CLAMP ON ADAPTER
*CT-101A LINE SPLITTER
(cannot be used for leakage current, for use on load current only)
*9094 OUTPUT CORD
*Note: Non-CE mark product
CAT III 300 V
Clamp Testers
3284 3285
3285
CLAMP ON AC/DC HiTESTER
Analysis for DC to distorted waves
●3284: 200 Arms, clamp aperture: 33 mm dia.
●3285: 2000 Arms, clamp aperture: 55 mm dia.
●Inrush current crest value
●RMS value of full-wave rectified waveforms
●Waveform and harmonic analysis
CAT III 600 V
3284
OPTIONS
9445-02 AC ADAPTER (for USA)
9445-03 AC ADAPTER (for EU)
9290-10 CLAMP ON ADAPTER
*Note: Non-CE mark product
43
*CT-101A LINE SPLITTER (cannot be used for DC,
AC+DC current, for use on AC current only)
*9094 OUTPUT CORD
3286-20
CLAMP ON POWER HiTESTER
All powerful ! Easy operation !
True-RMS Clamp-on Power Meter !
●Use as a single-phase power meter or power factor
meter (3kW to 600kW range)
●Simple checking of three-phase lines (6kW to
1200kW range)
●Check power supply fluctuations
●1000 A, 1000 Hz, peak and harmonic measurement
●True RMS (effective value) display method
●Optional printer (9442 PRINTER)
CAT III 600 V
9636-01 RS-232C PACKAGE
SPECIFICATIONS
Power factor/ Phase angle/ Reactivity measurement
Phase discrimination by phase detection (zero crossing)
0.000 (lead) to 1.000 to 0.000 (lag)
90.0°(lead) to 0.0°to 90.0°(lag)
0.000 (lead) to 1.000 to 0.000 (lag)
Measurement items
Measurable conductor diameter
Display
Rectification method
Display update rate
Analog response time
Voltage/ Current/ Power measurement
Range Table
150.0 V
AC
Voltage
300.0 V
600 V
Single-phase
*3-phase(balanced load)
Single-phase
*3-phase(balanced load)
Single-phase
*3-phase(balanced load)
AC Current
20.00 A
200.0 A
1000 A
3.000 kW
30.00 kW
150.0 kW
6.000 kW
60.00 kW
300.0 kW
6.000 kW
60.00 kW
300.0 kW
6.000/12.00 kW 60.00/120.0 kW
600.0 kW
12.00 kW
120.0 kW
600.0 kW
24.00 kW
240.0 kW
600.0/1200 kW
*3-phase power is calculated and displayed on the basis of a balanced, 50/60 Hz, sine wave input.
For apparent power and reactive power, the unit of watts in the above table is replaced by VA and var respectively.
0.000 (lead) to 1.000 to 0.000 (lag); 1φ only
1000 Arms cont.
600 Vrms (insulated conductor)
Voltage: 10 V to 600 V, Current: 10 A to 1000A,
Power: 80 V to 600 V and 10 A to 1000 A
Voltage: 0.6 Vrms, Current: 0.6 Arms
Min. Display value
Display indication range 5 or less are zero-suppressed, and the upper limit is to 125% of
the range setting (to 100% for the 1000 A range)
(RMS value)
2.5 or less (1000 A and 600 V range is 1.7 or less)
Circuit dynamic
Effective value P.F.
Max. allowable current
Max. usable circuit voltage
Effective input range
Wave peak measurement
Effective in the voltage and current functions
Measurement range
150 (375 peak) / 300 (750 peak) / 600 (1020 peak) V
20 (50 peak) / 200 (500 peak) / 1000 (1700 peak) A
Effective Input Range
Effective value of sine wave is within effective input permisible
in the range and within circuit dynamic
Clamp Testers
Frequency measurement
Effective in the voltage and current functions
Measurement range
30.0 Hz to 100 Hz (at 100.0Hz range)
100 Hz to 1000Hz (at 1000Hz range)
Min. input level
Voltage 10 Vrms-sine wave, Current 1 Arms-sine wave
Effective in the voltage and current functions
Harmonic measurement
Measurement items
Level of each order, percentage of each order and total harmonic
distortion (THD-F and THD-R)
Measurement range
Fundamental frequency 45 Hz to 65 Hz
Window width
1 cycle (45 Hz to 65 Hz), Data points: 256 points
Window type
Rectangular
Orders analyzed
Up to 20th
Other functions
Phase detection
Record
Battery capacity
Data hold
Auto power off
Data output
Normal/ reverse/ missing (at 3-phase balanced load)
MAX. value and MIN. value (Effective in the voltage, current and
effective / apparent power functions)
Displayed in % when the unit is powered on
Holds display
Approx. 10 minutes, buzzer sounds just before power is turned off,
can be extended or cancelled
RS-232C interface by optical insulating coupler
OPTIONS
9636-01 RS-232C PACKAGE
9442 PRINTER
9636 RS-232C CABLE (for 9442 printer)
9443-02 AC ADAPTER (for 9442 printer, EU)
9443-03 AC ADAPTER (for 9442 printer, America)
1196 RECORDING PAPER (for 9442 ,10 rolls)
44
3290/3290-10 CLAMP ON AC/DC HiTESTER
9691 9692 9693
CLAMP ON AC/DC SENSOR
9691 Specifications
φ35mm
●Frequency range:DC to 10kHz(-3dB)
●Effect of conductor position:
less than+/-1.0%
●Effect of extemal magnetic
fields:equivalent to 0.5A or less
(in a 400A/m extermal magnetic field)
●Extermal dimensions:approximately
53W×129H×18D mm
●Weight:approximately 230g
100A
All the Functions You Need for
Measurement at DC or 1Hz and Up
●Choice of three sensors (Example combinations)
3290+9691 : Measure up to 100A
3290+9692 : Measure up to 200A
3290+9693 : Measure up to 2000A
●Choice of measurement methods
DC (for battery measurement)
AC+DC RMS (for full-/half-wave rectification measurement)
AC RMS (for current distortion measurement)
PEAK (for peak value measurement) of inrush current, etc.)
●Choice of output (Simultaneous output)
Effective value output, frequency output,
waveform output
●Choice of response times (Switchable
among three response times)
●LPF function (filters out unnecessary
harmonics : fc=550Hz)
●3290-10 Functions
Current integral measurement
(obtain polarity-specific integrated DC values)
Operating time/duty measurement
φ33mm
9692 Specifications
200A
●Frequency range:DC to 20kHz(-3dB)
●Effect of conductor position:
less than+/-0.5%
●Effect of extemal magnetic
fields:equivalent to 0.7A or less
(in a 400A/m extermal magnetic field)
●Extermal dimensions:approximately
62W×167H×35Dmm
●Weight:approximately 410g
φ55mm
2000A
9693 Specifications
●Frequency range:DC to 15kHz(-3dB)
●Effect of conductor position:
less than+/-0.7%
●Effect of extemal magnetic
fields:equivalent to 2A or less
(in a 400A/m extermal magnetic field)
●Extermal dimensions:approximately
62W×196H×35Dmm
●Weight:approximately 500g
3290 & Sensor Common Specification
●Accuracy at 50 or 60 Hz : AC/DC/AC+DC Current (±1.3% rdg.+ 3 dgt.)
●Frequency range : DC to 1kHz (±2.3% rdg.+ 8 dgt.)
●Cable length: 2m(6.56ft)
●Maxmum circuit voltage: 600V
Model
Measurable
conductor
diameter
9691
φ35mm
9692
φ33mm
9693
φ55mm
3290-10
9692
3290
Clamp Testers
9691
3127-10 3128-10
Maximum display
Measurement
range on 3290 Normal measurement Peak measurement
20.00A
25.00A
50.00A
100Arms
150Apeak
200.0A
105.0A
150.0A
20.00A
25.00A
50.0A
200Arms
300Apeak
200.0A
210.0A
300.0A
200.0A
250.0A
500A
2000Arms
2840Apeak
2000A
2100A
3000A
Rated input
CLAMP ON HiTESTER
One meter drop-proof "Tested Tough!"
●With the range of 300 A (3127-10), 1500 A (3128-10)
●Temperature measurement
●Ohmmeter circuit tested to 250 V AC over voltage-OK
●Clamp-on measurement for busbars and thick conductors (3128-10)
OPTIONS
*9021-01 THERMISTER TEMPERATURE PROBE
9290-10 CLAMP ON ADAPTER ( for large AC current )
*CT-101A LINE SPLITTER
*Note: Non-CE mark product
45
Field Measuring Instruments
Field Measuring Instruments Index
Analog Multi Meters
LAN Cable Tester OPTICAL POWER
METER
3030-10
3008
Basic type analog
tester,
Average rectifier
Use for industrial power Affordable LAN cable Optical loss
lines
testing, Wire-Map, Cable measurement tool
length, Direction-Check 3662/63-20 LASER
Average rectifier
LIGHT SOURCE
. . . . . . . . . . . . . p.50 . . . . . . . . . . . . . . . p.50
3660
Noise Search Tester Voltage Detector
3661-20,3662/63-20 3144-20
Digital Multi Meters (basic functions)
3120
Noise Search Tester
Voltage Detector
Frequency range 500 Hz to 30 MHz AC 70 to 600 V(50/60 Hz)
0 dBV and -20 dBV f.s., 2 ranges
. . . . . . . . . . . . . p.50
. . . . . . . . . . . . . p.49
Phase Detector
3246
3244
3245
3256-50, -51
3126-01
Pencil type DMM,
Average rectifier
Card size DMM with
emphasis on safety,
Average rectifier
A card size DMM with Built tough for use with
industrial power lines
solar charged battery,
DMM, Average rectifier
Average rectifier
Terminal shutter interlock
mechanism DMM,
Average rectifier
Phase detector, Rotary Phase Detector,
disk style, 110 to 480V Non-contact types,
AC 70 to 600 V(50/60 Hz)
. . . . . . . . . . . . . p.48
. . . . . . . . . . . . . p.48
. . . . . . . . . . . . . p.49
. . . . . . . . . . . . . p.49
. . . . . . . . . . . . . . . p.53
3255-50
3129
Digital Multi Meters (multi-functional and high precision)
3801-50
3802-50
3803
3804, 3805
Terminal shutter interlock
mechanism DMM,
True RMS rectifier
Multi-function type, 51000
count display,
RS-232C/USB communication,
True RMS rectifier
. . . . . . . . . . . . . . . p.48
Low-cost type, 51000 count
display,
RS-232C/USB communication,
True RMS rectifier
Low-cost type, 4000 count
display,
RS-232C communication,
Average rectifier
. . . . . . . . . . . . . . . p.48
Multi-function type, 9999 High speed DMMs
count display (V range), 199999 count display
RS-232C communication,
True RMS rectifier (3805)
. . . . . . . . . . . . . . . p.22
. . . . . . . . . . . . . . . p.49
3237,3238,3239
Insulation Testers
Earth Testers
3118-11 to -12
3451-11 to -15
3452-11 to -13
3453
3454 -10/-11/-51
3455
3151
Testing voltage 250V
to 1000V, Two ranges,
Indicator type
Testing voltage 125V
to 1000V, Single
range, Compact and
lightweight, Luminous
scale indicator type
. . . . . . . . . . . . . . p.52
Testing voltage 25V
to 1000V, Three
ranges, Compact and
lightweight, Luminous
scale indicator type
. . . . . . . . . . . . . . p.52
Testing voltage 125V to 1000V,
Four ranges in one body,
Compact and lightweight, Digital
display, Comparator and memory
function
. . . . . . . . . . . . . . p.51
Innovative and
low-cost Insulation
resistance tester with
continuity function
Testing voltage 250V
to 5000V,
Five ranges
10 MΩ to 5 TΩ
Seven ranges
. . . . . . . . . . . . . . p.51
Grounding resistance
meter, Two-wire or
three-wire measurement
method, Tough and
durable design
. . . . . . . . . . . . . . p.53
. . . . . . . . . . . . . . p.52
Field Measuring Instruments
3257-50, -51
46
DIGITAL HiTESTERs
Table of functions
DC Voltage
ranges
Resistance
ranges
DC Current
ranges
AC Current
ranges
Frequency
characteristics
at AC Current
Frequency
ranges
Continuity
3244
420mV to 500V,
5 ranges
Best accuracy:
±0.7% rdg. ±4 dg
4.2V to 500V, 4 ranges
Accuracy: ±2.3 % rdg.
±8 dgt.
Average rectifier
effective value
50 to 500 Hz
420Ω to 42 MΩ,
6 ranges
Best accuracy:
±2.0% rdg. ±4 dgt.
None
None
None
None
50Ω ±30Ω
3246
420.0mV to 600V,
5 ranges
Best accuracy:
±1.3 % rdg. ±4 dgt.
4.2V to 600V,
4 ranges
Accuracy:
±2.3% rdg. ±8 dgt.
Average rectifier
effective value
50 to 500 Hz
420.0Ω to 42.00MΩ,
6 ranges
Best accuracy:
±2.0 % rdg. ±4 dgt.
None
None
None
None
(50Ω±30Ω)
3256-50
3256-51
420.0mV to 1000V,
5 ranges
Best accuracy:
±0.5% rdg. ±2 dgt.
420.0mV to 1000V,
5 ranges
Best accuracy:
±1.2% rdg. ±3 dgt.
Average rectifier
effective value
420.0Ω to 42.00MΩ,
6 ranges
Best accuracy:
±0.7% rdg. ±2 dgt.
42.00µA to 10.00A,
6 ranges
Accuracy:
±1.5 % rdg. ±4 dgt.
42.00µA to 10.00 A,
6 ranges
Best accuracy:
±2.5 % rdg. ±5 dgt.
Average rectifier
effective value
0.50Hz to500.0kHz,
5 ranges
input level: 800mV
to 1000V rms
Accuracy:
±0.02% rdg. ±1 dgt.
50Ω ±30Ω or less
51mV to 1000V,
7 ranges
Best accuracy
3801-50: 0.025% rdg.
± 5dgt.
3802-50: 0.03% rdg.
± 10dgt.
51mV to 1000V,
7 ranges
Best accuracy
3801-50: 0.4% rdg.
± 25dgt.
3802-50: 0.6% rdg.
± 25dgt.
510µA to 10A
6 ranges
Best accuracy
3801-50: 0.05% rdg.
± 25 dgt.
3802-50: 0.1% rdg.
± 25 dgt.
510µA to 10A
6 ranges
Best accuracy
3801-50: 0.7% rdg.
± 20 dgt.
3802-50: 0.9% rdg.
± 25dgt.
3801-50:
20 to 100kHz
3802-50:
30 to 20kHz
99.999Hz to 999.99kHz
Best accuracy
0.02% rdg. ± 3 dgt.
10Ω or less
(at 510Ω range)
3803
400.0mV to 1000V,
5 ranges
Best accuracy:
±0.6% rdg. ±2 dgt.
400.0mV to 1000V,
5 ranges
Best accuracy:
±2.0% rdg. ±2 dgt.
Average rectifier
effective value
40 to 500Hz
400.0Ω to 40.00MΩ,
6 ranges
Best accuracy:
±0.6% rdg. ±3 dgt.
400.0µA to 10.00A,
5 ranges
Best accuracy:
±1.5 % rdg. ±2 dgt.
400.0µA to 10.00A,
5 ranges
Best accuracy:
±2.0 % rdg. ±2 dgt.
Average rectifier
effective value
40 to 500Hz
None
34.5Ω or less
(at 400Ω range)
3804
999.9mV to 999.9V,
4 ranges
Best accuracy:
±0.3% rdg. ±2 dgt.
999.9mV to 999.9V,
4 ranges
Best accuracy:
±1.2 % rdg. ±5 dgt.
Average rectifier
effective value
40 to 500 Hz
400.0Ω to 40.00MΩ,
6 ranges
Best accuracy:
±0.6 % rdg. ±3 dgt.
400.0µA to 10.00A,
6 ranges
Best accuracy:
±0.2 %rdg. ±3 dgt.
400.0µA to 10.00A,
6 ranges
Best accuracy:
±1.2 % rdg. ±5 dgt.
Average rectifier
effective value
50 to 2 kHz
None
10Ω or less
(at 400Ω range)
999.9mV to 999.9V,
4 ranges
Best accuracy:
±0.1% rdg. ±2 dgt.
999.9mV to 999.9V,
4 ranges
Best accuracy:
±1.1% rdg. ±5dgt.
Effective value rectifier
40 to 2 kHz
400.0Ω to 40.00MΩ,
6 ranges
Best accuracy:
±0.5 % rdg. ±3 dgt.
400.0µA to 10.00A,
6 ranges
Best accuracy:
±0.1%rdg. ±3dgt.
400.0µA to 10.00A,
6 ranges
Best accuracy:
±0.1 % rdg. ±5 dgt.
Effective value rectifier
50 to 2kHz
1Hz to 50kHz /
0.7V to 300V rms
Best accuracy:
±0.05% rdg. ±4 dgt.
10Ω or less
(at 400Ω range)
3255-50
420mV to 1000 V,
5 ranges,
Best accuracy:
±0.5 % rdg. ±4 dgt.
420mV to 1000 V,
5 ranges,
Best accuracy:
±1.2% rdg. ±4 dgt.
Average rectifier
effective value
None
(45Ω ±35Ω or less)
3030-10
0.3V (16.7kΩ/V),
3/12/30/120/300/
600V(20kΩ/V)
Accuracy:
±2.5% f.s.
12V±4% f.s. (9kΩ/V)
30/120/300/600V
±2.5% f.s.
Average rectifier
effective value
None
None
(3257)
3801-50
True RMS
3802-50
True RMS
3805
True RMS
50 to 500 Hz
510Ω to 510MΩ ,
7(6) ranges (3802-50)
3801-50: 20 to 100kHz Best accuracy
3802-50: 30 to 30kHz 3801-50: 0.05% rdg.
± 5 dgt.
3802-50: 0.08% rdg.
± 5 dgt.
50 to 500 Hz
None
420Ω to 42 MΩ,
6 ranges,
Best accuracy:
±0.7 % rdg. ±4 dgt.
0 to 3kΩ
(center scale 30Ω)
R×1, R×10, R×100,
R×1k
±3.0% of scale length
60µA/30m/300mA
(300mV internal
voltage drop)
Accuracy: ±3% f.s.
None
None
Field Measuring Instruments
Dimensions/
mass
Included
accessories
CR2032×1 batteries
(Continuous use
150 hours)
55W×109H×
9.5D mm/60g
Hard case(1)
CR2032×1 batteries
(Continuous use
150 hours)
30W×182H×
26.5D mm
80 g
R03(AAA)×
2 dry batteries
(Continuous use
100 hours)
76W×167H×
33D mm/260 g
9170 TEST LEAD(1)
Fuse(2)
9378 CARRYING
CASE(1) (3256-50)
Holster (3256-51)
6LR61×1 (9.0V)
(Continuous
use 50 hours)
90W×192H×
37D mm/940 g
3851-10
TEST LEAD(1),
Holster(1)
6F22(006P)×1
76W×167H×
33D mm
400 g
3851-10 TEST LEAD(1)
Holster(1)
3 times/sec
13 times/sec
(Bar graph)
6F22(006P)×1
76W×167H×
33D mm/400 g
3851-10 TEST LEAD(1)
Holster(1)
3 times/sec
(exclusive Hz range)
1 time/sec
(Hz range)
13 times/sec
(Bar graph)
6F22(006P)×1
76W×167H×
33D mm/400 g
3851-10 TEST LEAD(1)
Holster(1)
R03(AAA)×2
dry batteries
(Continuous use
200 hours)
70 mmW×
145 mmH×
31 mmD/200 g
9207-10
TEST LEAD (1),
9371 CARRYING
CASE (1)
95W×141H×
39D mm
280g
9207 TEST LEAD(1)
fuse(1)
9390 CARRYING
CASE(1)
Auto power
save
Range
switching
Display/Safety
3244
None
None
(Cancel impossible)
Auto
Digital/LCD,
maximum 4199 dgt
Safety: IEC1010-1,
Pollution degree 2,
over voltage category II
None
2.5 times /sec
3246
judgement only
(3.4V open
terminal voltage)
None
(Cancel possible)
Auto or Manual
Digital/LCD,
with Back light
max. 4199 dgt.
None
2.5 times/sec
3256-50
3256-51
(3.4 V open
terminal voltage)
Voltage detect
function
(Cancel possible)
Hold-auto function
Relative function
Auto or Manual
LCD, maximum
4200 dgt. (all mode)
maximum 19999 dgt.
(Frequency)
3801-50 only:
AC+DC measurement,
Pulse output
Common functions:
Capacitance, Data hold,
(3.1 V open
terminal voltage) dBm measurement, (Cancel possible)
Duty ratio/Pulse width,
Temperature,
1ms peak hold
Relative, Max/Min/Ave,
RS-232C, USB
Auto or Manual
LCD, max.
51000 dgt.
with Back light
3803
(3 V open
terminal voltage)
(Cancel possible)
Auto or Manual
LCD, max.
4000 dgt.
(Maximum
41 dots)
2.5 times/sec
13 times/sec
(Bar graph)
3804
Capacitance,
Data hold, Refresh
(3.3 V open
hold, Max/Min/Ave/ (Cancel possible)
terminal voltage) Relative/4-20mA%
display, RS-232C
Auto or Manual
LCD, max.
4000 dgt.
max. 9999 dgt.
(at V range)
(Maximum
41 dots)
3805
(3.3 V open
terminal voltage)
Capacitance,
Data hold, Refresh
hold, Max/Min/
Ave/Relative/
4-20mA% display,
RS-232C,
Temperature
(Cancel possible)
Auto or Manual
LCD, max. 4000 dgt.
max. 9999 dgt.
(at V range)
(Maximum
41 dots)
3255-50
judgement only
(3.4 V open
terminal voltage)
CLAMP (ACA)
function
(Clamp-on probe :
Option)
10A to 1000A
7ranges
(Cancel possible)
Auto or Manual
LCD, maximum
4199 dgt.
None
None
Manual
Indicator type
None
True RMS
3802-50
True RMS
True RMS
3030-10
RS-232C,
Data hold
Battery check:
0.9 to 1.8V, load
resistance 10Ω
Temperature:
(3V open terminal
-20 to 150,
voltage)
(9021-01 Thermister
Temperature Probe
is necessary, sold
separately)
Bar graph
display
None
Other functions
3801-50
50 to 500Hz
None
None
Diode check
(3257)
47
Frequency
characteristics at
AC Voltage
AC Voltage
ranges
(Maximum
40 segments)
Sampling rate
2.5 times/sec
(all mode)
5 times/sec
(Frequency)
25 times/sec
(Bar graph)
Maximum 21 dots 3 times/sec
1 times/sec
2.5 times/sec
None
Power supply
R6P(AA)×2
batteries
3246
PENCIL HiTESTER
Pencil-type DMM with Penlight
In addition to being compact, this pencil-type tester comes with
auto-range and data hold functions for incredibly easy
measurement of electrical and electronic circuitry.
●Full-auto ranging, 4199 count display
●Penlight brightly illuminates test points
●Overload protection to 600 V
(Ω and continuity functions)
Penlight brightly illuminates test points
OPTIONS
*9081 10A SHUNT
*Note: Non-CE mark product
CAT III 600 V
3244
CARD HiTESTER
Card size DMM with emphasis on safety
●Only 9.5 mm thick and 60 g in weight
●4199 count display
●Test leads fit neatly inside the case.
●Automatic power saving function saves your
batteries even when you forget to turn off the power.
CAT II 500 V
3801-50
OPTIONS
*9081 10A SHUNT
*Note: Non-CE mark product
CAT IV 600 V
CAT III 1000 V
DIGITAL HiTESTER
OPTION
OPTION
High-precision, high-resolution, and
multi-functional handy DMMs
OPTIONS
3856-01 RS-232C PACKAGE
3856-02 USB PACKAGE
*9014 HIGH-VOLTAGE PROBE
*9180 to *9183 TEMPERATURE PROBE
9472 to 9476 TEMPERATURE PROBE
9617 CLIP ON BASE
(for capacitance measurement)
9618 CLIP TYPE LEAD
(for capacitance measurement)
3803
Field Measuring Instruments
●Display two different parameters simultaneously
●Optional RS-232C or USB package for transferring data
captured by the 3801-50 to a PC
●Measures the AC components in DC voltage or DC current
●1ms peak hold mode makes it possible to capture the peak
value of a waveform
*Note: Non-CE mark product
DIGITAL HiTESTER
Compact and basic DMM
●4000 count display full scale
●Average rectified RMS indication type
●Optional RS-232C package for transferring data
captured by the 3803 to a PC
OPTION
CAT III 600 V
CAT II 1000 V
OPTIONS
3853 CARRYING CASE
3854 RS-232C PACKAGE
*9014 HIGH-VOLTAGE PROBE
*Note: Non-CE mark product
48
3256-50/-51
3257-50/-51
DIGITAL HiTESTER
3256-50/51
CAT II 1000 V
CAT III 600 V
3257-50/51
CAT III 1000 V
CAT IV 600 V
Terminal shutter interlock mechanism
●Terminal shutter interlock
mechanism exposes only the
correct terminals for connection
in the currently selected function
●Wide range, maximum reading
4200 digit
●High-speed response, bar graph
display
●Conforms with IEC1010
●Hold-auto function automatically
displays voltage or current value
and resistance value
Fail-safe!
Shutter mechanism prevents incorrect test lead connection
Single
operation
3257-50
(True RMS)
Simple
operation
with rotary
switch
Voltage ranges
10A range
Only V and COM terminals open
Only A and COM terminals open
* The above photograph shows a special model with a transparent cover.
Check for live lines safely and easily
In the AC V range, the 3256-50 can be used to
check whether power lines are live. When the
sensitivity level is set to 4 and the test head is
placed near a live power line, the built-in buzzer
sounds and a display indicator lights.
Sensitivity threshold: 100 V AC or higher
3256-51
Model
3256-50 (Standard type)
3256-51 (3256-50 with holster)
3257-50 (Standard type)
3257-51 (3257-50 with holster)
OPTIONS
*9014 HIGH-VOLTAGE PROBE
3853 CARRYING CASE (for 3256-51)
*Note: Non-CE mark product
3255-50
DIGITAL HiTESTER
Tough for use on industrial power lines
●Built-in current limiter and fuse capable of withstanding 1000 V to
prevent short-circuit accidents
●Wide range, maximum reading 4199 digit
●Two-terminal configuration eliminates the need for probe
reconnections
●Industrial grade test leads for enhanced safety
Field Measuring Instruments
3120
CAT II 1000 V
CAT III 600 V
OPTIONS
9207-10 TEST LEAD
(Supplied as standard with the 3255-50)
9371 CARRYING CASE
(Supplied as standard with the 3255-50)
VOLTAGE DETECTOR
Twin Light Audible Voltage Detector
●Green for Battery Check
●Red for Voltage Detection
CAT IV 600 V
SPECIFICATIONS
Measurement Function
Voltage Range
Indication
Battery Check
Power Source
Continuous Use
Dimensions•Mass
49
Voltage Detection
AC 70 to 600 V,50/60 Hz(when touching insulated wiring equivalent to IV2mm2)
Red LED and continuous beeping sound
Green LED
"AAA" manganese(R03)or alkaline(LR03)battery×2
Approx.200 hours(when switched to "ON" and in standy using LR03 batteries)
149H×ø18.5mm(excl.protrusions),38g(incl.batteries)
3144-20
NOISE SEARCH TESTER
Identify noise in communication and power lines
3144-20 SPECIFICATIONS
Input unit configuration 9741dedicated input terminal,BNC input terminal(9741 takes priority)
Frequency range band 500Hz to 30MHz,separated into 7ranges (-3 dB)
500Hz to 3kHz(1 kHz range)
7.5kHz to 22.5kHz(15kHz range)
35kHz to 105kHz(70kHz range)
125kHz to 375kHz(250kHz range)
0.5MHz to 1.5MHz(1MHz range)
1.5MHz to 4.5kHz(3MHz range)
10MHz to 30MHz(20MHz range)
(BNC input 50Ω termination)
×1 range 0 dBV(1V)f.s Measurement range(0dBV to-30dBV)
Measurement range
×10 range-20 dBV(0.1V)f.s. Measurement range(-20dBV to-50dBV)
RMS value conversion
Detection method
500Hz to 1 MHz or less±1.5dBV
Detection accuracy
1MHz to 30 MHz±2.0dBV
Display of measurement voltage level of each frequency range in levels on LCD(2.5dBV/SEG equivalent)
Monitoring function
Measurement data and time saved to internal memory according to specified recording interval
Logging function
1/2/5/10/20/30 seconds 1/2/5/10/20/30/60 minutes
Recording interval
Wave monitoring (Output of input signal coming from 9741 or BNC input)
Output function
Audible range monitoring (Output of detection signal envelope Earphone)
USB Ver.1.1
Interface
AA-size alkaline batteries(LR6)×6,DC9 V 500mA
Power
(9445-02 or 9445-03 AC adapter,rated power voltage AC 100V to 240V,50/60Hz,maximum
rated current 250 mA)
Dimensions and mass 98(W)×179(H)×46(D)mm.430g(excluding batteries)
9741 Clamp-on Voltage Sensor,carrying case,PC applincation software CD-R,AA-size alkaline
Accessories
batteries(LR6)×6,USB cable,strap,9445-02 AC adapter(UL),9445-03 AC adapter(CEE),earphone
3661-20
Sensor configuration
Frequency range
Conductor dia
Maximum rated voltage to
earth
Dimensios and mass
Electrostatic coupling non-contact voltage sensor
600Hz to 30MHz(-3dB )
ø20mm
AC200V
62(W)×158(H)×40(D)mm,260g
OPTICAL POWER METER
3662-20 3663-20
LASER LIGHT SOURCE
3661-20 SPECIFICATIONS
Measurement functions Optical power measurement (dBm)
Measure absolute value of input optical power
Optical loss measurement (dB)
Automatically compare measured power with previously input reference
value to calculate and display loss
Calibration wavelength 850 nm, 1310 nm, 1550 nm
-60 dBm to +9 dBm (auto range)
Range
±0.22 dB (±5 %) at -10dBm, CW, 23°C ±5°C
Accuracy
0.01 dBm (optical power), 0.01 dB (optical loss)
Resolution
+10 dBm
Rated max.
FC, SC (using optional connector adapter)
Connector
Single mode, multi mode
Fiber type
(core dia. 62.5µm max. NA: 0.275 max.)
InGaAs (dia. 1 mm)
Light receiver
Approx. 3 times/s (approx. 350 ms)
Display update rate
Max. 1000 data per wavelength
Memory
USB (Ver. 1.1)
Interface
Dedicated PC application software allows transfer of measurement data
from the 3661-20 memory to a computer
Auto power save (after about 10 minutes of inactivity; defeatable)
Functions
Settings backup (settings are automatically stored at power-off)
Battery check (symbol appears when voltage drops below approx. 4 V)
LR6(AA) alkaline battery×4, 0.5 VA
Power supply
Approx. 40 hours (continuous use)
Operating time
Dimensions and mass Approx. 85 W×192 H (including 36 mm cover)×35 D mm,
Approx. 300g (without batteries)
3662-20 (1550mm)
3663-20 (1310mm)
3661-20
3662-20, 3663-20
Light-emitting element
Output connector
Fiber type
Output mode
Output wavelength
Spectrum width
Output level
Output level stability
SPECIFICATIONS
Laser diode
FC, SC (using optional connector adapter)
Single mode
Continuous wave (CW) or modulated light (270 Hz, 1 kHz, 2 kHz)
1310±20 nm (3663-20) 1550±20 nm (3662-20)
5 nm max.
-6 ±2 dBm
Within ±0.1 dB (temperature constant, 5 minutes)
Within 1.0 dB p-p (ambient temperature 0 to 40°C, 8 hours)
Battery check (indicator flashes when battery voltage drops)
Functions
LR6(AA) alkaline battery×2, 0.6 VA
Power supply
Approx. 20 hours (3662-20, continuous CW output)
Operating time
Approx. 36 hours (3663-20, continuous CW output)
Dimensions and mass Approx. 76 W×159 H (including 36 mm cover)×35 D mm,
Approx. 180g (without batteries)
LAN CABLE HiTESTER
SPECIFICATIONS
●Simple operation
Simple checks can be made by
pressing three operation keys.
●Large LCD shows the wiring status
Can simultaneously display the results
of the WireMap check for incorrect
connections, wire breaks and short
circuits, along with the results of
the cable length check that indicates
the location of the wire break or
short circuit, giving you an accurate
understanding of the wiring status.
Measurable cable
Measurable connector
Measurement item
< WireMap >
<Cable length>>
<Direction>>
Display
Ancillary functions
Applicable standards
Allowable input
Ambient temperature of use
Storage temperature
Power supply
Maximum rated power
Operating time
Dimensions and mass
UTP cable: Characteristic impedance 100Ω, CAT 3, 4, 5, 5e
RJ-45 connector
Can check wiring status using the 9336 WIREMAP TERMINATOR
Error detection: Open, short, other wiring errors (Split pairs cannot be detected.)
Measured length: 2 to 300m
when results are displayed in feet: 6.6 to 984 ft
Measurement accuracy: ± (15% rdg. + 1 m)
when results are displayed in feet: ± (15% rdg. + 3.3 ft)
Display resolution: 0.1 m
when results are displayed in feet: 0.3 ft
Can identify five cables using the optional 9337 DIRECTION TERMINATOR
LCD
Power saving mode: Power saving mode after measurement
operation, recovery from power saving mode when measurement
operation was initiated by pressing the TEST key
Battery check: Battery indicator appears when batteries are nearly exhausted.
Safety: EN61010-1:1993+A2:1995
EMC : EN61326-1:1997+A1:1998
3.5 V peak (between pins of the RJ-45 connector on the 3660)
0°C to 40°C, 80%rh or less, no condensation
-10°C to 50°C, 80%rh or less, no condensation
LR6(AA) alkaline battery×6
1.6 VA
Approx. 250 hours (when measuring once per minute)
Approx. 98 W×160 H×38 D mm, Approx. 290g
Field Measuring Instruments
3660
9741 SPECIFICATIONS
50
3455
HIGH VOLTAGE INSULATION HiTESTER
Maximum 5kV Test Voltage - Up to 5TΩ of Insulated Resistance Testing
Safely evaluate the insulation characteristics of high voltage transformers, motors and cables
●Wide voltage range (250V to 5kV) for maximum 5TΩ of insulation resistance measurements
●Automatically calculate and display the PI (Polarization Index) and DAR (Dielectric
Absorption Ratio) for all types of insulation evaluations
●Temperature compensation to accurately respond to variations in insulation material
●Internal memory stores 100 blocks of manually recorded data and 10 sets of log data
●USB interface, compact rugged case, and safe design
SPECIFICATIONS
0.00 MΩ to 250 GΩ,
250 V range
Accuracy : ±5 % rdg. ±5 dgt. (0 to 2.50 GΩ)
±20 % rdg. ±5 dgt. (2.50 to 250 GΩ)
0.00 MΩ to 500 GΩ ,
500 V range
Accuracy : ±5 % rdg. ±5 dgt. (0 to 5.00 GΩ)
±20 % rdg. ±5 dgt. (5.00 to 500 GΩ)
0.00 MΩ to 1.00 TΩ ,
1 kV range
Accuracy : ±5 % rdg. ±5 dgt. (0 to 10.0 GΩ)
±20 % rdg. ±5 dgt. (10.0 to 500 GΩ)
±30 % rdg. ±50 dgt. (500 G to 1.00 TΩ)
0.00 MΩ to 2.50 TΩ,
2.5 kV range
Accuracy : ±5 % rdg. ±5 dgt. (0 to 25.0 GΩ)
±20 % rdg. ±5 dgt. (25.0 to 500 GΩ)
±30 % rdg. ±50 dgt. (500 G to 2.50 TΩ)
0.00 MΩ to 5.00 TΩ ,
5 kV range
Accuracy : ±5 % rdg. ±5 dgt. (0 to 50.0 GΩ)
±20 % rdg. ±5 dgt. (50.0 to 500 GΩ)
±30 % rdg. ±50 dgt. (500 G to 5.00 TΩ)
Insulation resistance mode: Data memory(100 data), measurement value hold, average, bar graph display, timer etc.
Function
Leak current: (1.00nA to 1.20mA), Temperature: (-10°C to 70°C)
Voltage: (DC±50V to 1kV AC 50V to 750V)
All measurement mode: live wire warning, battery indicators, auto power save
USB ver 2.0 (full speed)
Interface
Display
LCD with backlight
Power supply
LR6(AA) alkaline batteries × 6, 9459 BATTERY PACK
9753 AC ADAPTER
260 mmW × 250.6 mmH × 119.5 mmD, 2.8 kg
Dimensions, mass
Accessories
OPTIONS
9750-01 to 03 TEST LEAD (red, black, blue 3m)
9751-01 to 03 ALLIGATOR CLIPs (red, black, blue)
9631-01 TEMPERATURE SENSOR (1m)
9631-05 TEMPERATURE SENSOR (6cm)
9750-11 to 13 TEST LEAD (red, black, blue 10m)
9459 BATTERY PACK
9753 AC ADAPTER
9750-01 to 03 TEST LEAD (red, black, blue 3m),
ALLIGATOR CLIPs (red, black, blue),
LR6(AA) Alkaline batteries × 6, USB CABLE(1)
3453
DIGITAL MΩ HiTESTER
For efficient insulation measurement!
Field Measuring Instruments
●One body with four ranges: 125 V/40 MΩ, 250 V/2000 MΩ, 500 V/2000 MΩ,
and 1000 V/4000 MΩ
●Accurate digital display
●Insulation measurement through sight and sound
●Memorizes on the spot (Memorizes up to 20 data points)
●Recognizes variations of resistance
●Ability to measure AC voltage and low resistance (continuity)
SPECIFICATIONS
Common SPECIFICATIONS
Testing voltage
125 V DC
250 V DC
500 V DC
1000 V DC
4.000 MΩ or 40.00 MΩ 4.000 /40.00 /400.0 /2000 MΩ, (+ 4000 MΩ range at 1000 V mode)
Measurement range
First
±2 % rdg. ±3 dgt.
±2 % rdg. ±3 dgt.
±2 % rdg. ±3 dgt.
±2 % rdg. ±3 dgt.
effective measurement range at 0.100 to 10.00 MΩ at 0.200 to 20.00 MΩ at 0.200 to 50.00 MΩ at 0.200 to 999 MΩ
Discharge function : effective
Display : Digital/4000 dgt. LCD, Bar graph/42 seg.
with backlight
Functions : Insulation resistance mode: comparator, memory
Second
±5 % rdg.
±5 % rdg.
effective measurement range at 10.01 to 40.00 MΩ at 20.01 to 2000 MΩ
Voltage with no load
Min. resistance measurement value
(Resistance value to maintain rated voltage)
Shorting measurement
current
±5 % rdg.
at 50.1 to 2000 MΩ
±2 % rdg. ±6 dgt.
at 0 to 0.199 MΩ
±5 % rdg.
at 1000 to 4000 MΩ
Not more than 1.2 times rated testing voltage
0.125 MΩ
0.250 MΩ
1.2 mA max.
0.500 MΩ
2.000 MΩ
0.6 mA max.
Response time
Infinitude to center, infinitude to zero-MΩ within 5 second (within accuracy range)
Low resistance (conductivity) ±2 % rdg. ±8 % dgt. at 0 to 400.0Ω (aural warning below: 30Ω), Open terminal voltage: 4 V max.
±3 % rdg. ±8 dgt. at 0 to 600 V, 50 to 60 Hz,
AC voltage range and
Input resistance: 170 kΩ
accuracy
51
(20 data), measurement value hold, auto discharge, bar graph display
(measurement switch ON: insulation resistance; measure switch OFF:
voltage across measurement terminals), auto display of measurement
value 1 minute after measurement start, All measurement mode: live
wire warning, battery indicators, auto power save
Sampling rate : 2 times/second
Power supply : R6P(AA) ×4 or LR6(AA)×4
Dimensions : 155W×98H×80D mm, 500g
Accessories : 9294 TEST PROBE(1)
display cover and suspension band(1)
OPTIONS
*9289 TEST PROBE
*9288 BREAKER PIN
*Note: Non-CE mark product
3454-10/-11/-51
DIGITAL MΩ HiTESTER
3454-11/-51
Revolutionary insulation resistance tester with
continuity function all in one low price
SPECIFICATIONS
Model
Insulation testing voltage
Measurement range
Insulation Accuracy
resistance
Voltage with no load
Short circuit current
Response time
Low resistance (continuity)
AC voltage
Test voltage: 250V / 500V / 1000V DC
(3454-51: Non-CE mark product)
3454-10
3454-11/51
50 V DC / 125 V DC / 250 V DC / 500 V DC
250 V DC / 500 V DC / 1000 V DC
4.000/40.00/400.0/200.0 MΩ,400.0/2000 (250V/500V range only) MΩ
4.000/40.00/400.0/500.0 MΩ, 4000 (1000V range only) MΩ
1st effective range: ±3 %rdg. ±4 dgt. / 2nd effective range: ±5 %rdg. ±5 dgt.
Not more than 1.2 5 rated measurement voltage
1.2 mA max.
∞ to center, ∞ to 0 MΩ within 5 s (within accuracy range)
Measurement range: 40.00 / 400.0 / 4.000 k / 40.00 k / 400.0 k / 4.000 MΩ Short circuit current: 200 mA
Accuracy: ±3 %rdg. ±6dgt. (±5 %rdg. ±6dgt. at 400 kΩ range or higher) Open terminal voltage: 4 to 6 V Response time: 5 s max.
Display indication range: 0 to 750 V Accuracy: ±3 %rdg. ±6 dgt.(up to 600V), Frequency range: 50 / 60 Hz, Input impedance: 100kΩ
OPTIONS
9294 TEST PROBE
*9288 BREAKER PIN
*9289 TEST PROBE
*Alligator Clips for Model 9289
9257 CONNECTOR CORD
*Note: Non-CE mark product
Accessories : 9294 TEST PROBE (for 3454-11/3454-10), 9289 TEST PROBE. Alligator Clips (for 3454-51), Strap band (1)
●Other functions: Insulation and low resistance mode - comparator, measurement value hold; Insulation resistance mode - auto discharge;nsulation and AC voltage mode - live wire warning (when voltage
of 70V ±10V exists across measurement terminals); Low resistance mode - zero adjust; All measurement modes - battery indicators, auto power save
●Display update rate: 2 times / second
●Power source: R6P manganese battery×4 or LR6 alkaline battery×4
●Dimensions and Mass: Approx. 175 W×148 H×56 D mm; Approx. 530g (with display cover closed)
3451-11 to 3451-15
MΩ HiTESTER
Compact and lightweight for perfect portability
Common SPECIFICATIONS
SPECIFICATIONS
Model
3451-11
3451-12
3451-13
Testing voltage
250 V DC
500 V DC
125 V DC
Rated resistance
50 MΩ
100 MΩ
20 MΩ
±5 % of scale
±5 % of scale
±5 % of scale
First effective
measurement range and indication at 0.02 MΩ indication at 0.05 MΩ indication at 0.1 MΩ
to 10 MΩ
to 20 MΩ
to 50 MΩ
tolerances
±10 % of scale
±10 % of scale
±10 % of scale
Second effective
measurement range and indication at more than indication at more than indication at more than
10 MΩ to 20 MΩ 20 MΩ to 50 MΩ 50 MΩ to 100 MΩ
tolerances
Excepting 1st and 2nd
effective measurement
range include 0 MΩ &
infinity indication
Shorting measurement
current
Scale center indication
AC voltage scale and
tolerances
3451-14
3451-15
500 V DC
1000 MΩ
1000 V DC
2000 MΩ
±5 % of scale
indication at 1 MΩ
to 500 MΩ
±10 % of scale
indication at more than
500 MΩ to 1000 MΩ
±5 % of scale
indication at 2 MΩ
to 1000 MΩ
±10 % of scale
indication at more than
1000 MΩ to 2000 MΩ
0.7 % of scale length
0.7 % of scale length
0.7 % of scale length
0.7 % of scale length
0.7 % of scale length
1.2 mA max.
1.2 mA max.
1.2 mA max.
0.6 mA max.
0.6 mA max.
0.5 MΩ
±5 % f.s. at 0 to 250 V
(50/60 Hz)
1 MΩ
±5 % f.s. at 0 to 300 V
(50/60 Hz)
2 MΩ
±5 % f.s. at 0 to 500 V
(50/60 Hz)
20 MΩ
±5 % f.s. at 0 to 500 V
(50/60 Hz)
50 MΩ
±5 % f.s. at 0 to 500 V
(50/60 Hz)
Discharge function : effective
Power supply : R6P(AA)×4
Dimensions, mass :
152W×95H×47D mm, 420 g
Accessories :
9292 TEST PROBE(1)
9384 CARRYING CASE(1)
Safety rating :
Conformance to IEC1010, pollution
degree 2, installation category III
OPTIONS
*9288 BREAKER PIN
9293 PIN-TYPE EARTH PROBE
*Note: Non-CE mark product
MΩ HiTESTER
Compact analog three-range insulation resistance meter
Common SPECIFICATIONS
SPECIFICATIONS
Model
Testing voltage
Rated resistance
First effective
measurement range and
tolerances
3452-11
3452-12
3452-13
at 0.01 MΩ
to 5 MΩ
at 0.01 MΩ
to 5 MΩ
at 0.02 MΩ
to 10 MΩ
at 0.02 MΩ
to 10 MΩ
at 0.05 MΩ
to 20 MΩ
at 0.1 MΩ
to 50 MΩ
±5 % of scale indication
±10 % of scale indication
Second effective
measurement range and at 5 M to 10 MΩ, at 5 M to 10 MΩ, at 10 M to 20 MΩ, at 10 M to 20 MΩ, at 20 M to 50 MΩ, at 50 M to 100 MΩ,
tolerances
0.005M to 0.01MΩ 0.005M to 0.01MΩ 0.01M to 0.02MΩ 0.01M to 0.02MΩ 0.02M to 0.05MΩ 0.05M to 0.1MΩ
0MΩ & infinity indication
0.7 % of 100 V scale length
0.7 % of 500 V scale length
Open-circuit terminal voltage
Not more than 1.2 times rated testing voltage
Not more than 1.2 times rated testing voltage
Rated measurement current
1 m to 1.2 mA
1 m to 1.2 mA
Shorting measurement current
Scale center indication
AC voltage scale & tolerances
Discharge function : effective
Power supply : R6P(AA)×4
Dimensions, mass :
152W×95H×47D mm, 420g
Accessories :
at 0.05 MΩ
at 0.1 MΩ
at 2 MΩ
9292 TEST PROBE(1)
to 20 MΩ
to 50 MΩ to 1000 MΩ
9384 CARRYING CASE(1)
±10 % of scale indication
Safety rating :
at 20 M to 50 MΩ, at 50 M to 100 MΩ, at 1000 M to 2000 MΩ,
Conformance to IEC1010,
0.02M to 0.05MΩ 0.05M to 0.1MΩ
1 M to 2 MΩ
pollution degree 2,
installation category III
0.7 % of 1000 V scale length
25 V DC 50 V DC 100 V DC 125 V DC 250 V DC 500 V DC 250 V DC 500 V DC 1000 V DC
10 MΩ
10 MΩ
20 MΩ
20 MΩ
50 MΩ
100 MΩ
50 MΩ
100 MΩ 2000 MΩ
±5 % of scale indication
±5 % of scale indication
±5 % of scale indication
1.2 mA max.
1.2 mA max.
0.2 MΩ
0.2 MΩ
0.5 MΩ
±5 % f.s. at 0 to 150 V (50/60 Hz)
0.5 MΩ
1 MΩ
2 MΩ
±5 % f.s. at 0 to 500 V (50/60 Hz)
Not more than 1.2 times rated testing voltage
1 m to 1.2 mA
0.5 m to 0.6 mA
1.2 mA max.
0.6 mA max.
1 MΩ
2 MΩ
50 MΩ
% f.s. at 0 to 500 V (50/60 Hz)
±5
Field Measuring Instruments
3452-11 to 3452-13
OPTIONS
*9288 BREAKER PIN
9293 PIN-TYPE EARTH PROBE
*Note: Non-CE mark product
52
3151
SPECIFICATIONS
Grounding resistance, Grounding voltage
Measurement item
Measurement ranges 10Ω (0 to 11.5Ω) to 1000Ω (0 to 1150Ω), 3 ranges
30V (0 to 30 VAC), 1 range
*Using the two-wire measurement method; applied to 100Ω /1000Ω range only.
Operating method
AC phase difference
Open terminal voltage 50V AC max.
Measurement current 15mA AC max. *3mA AC max. using two-wire method.
Measurement frequency 575Hz or 600 Hz selectable
Basic accuracy
Grounding resistance: ±2.5% f.s.
Grounding voltage: ±3% f.s.
Power supply
R6P (AA), 6 pieces (at least 500 operations)
or LR6 (AA), 6 pieces (at least 1400 operations)
*Operating time: 30 second measurement, 30 second off
Dimensions, mass
164W×119H×88D mm, 800g (main unit only)
Accessories
9214 AUXILIARY EARTHING ROD(2),
9215 MEASURING CABLE
(one earth: black 5m, yellow 10m, red 20m, 9216 CABLE WINDER: 3),
9393 CARRYING CASE(1)
EARTH HiTESTER
Stable measurement for earth resistance
●Measurement range for grounding resistance increased to 115 % of normal
range
●Elastomer rotary knob fits the hand perfectly.
●Select the "simple" two-wire measurement method, using a low ground
conductor such as the ground side of a commercial power supply, or the
conventional three-wire measurement method
●Select a measurement frequency to reduce the influence of
harmonics of the power supply frequency on the ground
current
OPTIONS
9050 EARTH NET (set of two)
*Use in location where there is no driven-in ground and
where water seepage is present
3129
CAT III 600 V
PHASE DETECTOR
Non-Metallic Contact for Optimal Safety
SPECIFICATIONS
Measurement Function
Voltage Detection Method
Voltage Range
Clamp Diameter
Display
Battery Check Function
Auto Power Off
Power Supply
Dimensions, Mass
Accessories
Phase detection(positive, negative), live wire check(R-S/S-T only)
Electrostatic induction method
70V to 600V AC(50/60Hz)(sine wave, continuous input)
17mm max.of insulated wiring
Phase Detection
Positive:4 LEDs lit in clockwise order and 3 short beeps
Negative:4 LEDs lit in counterclockwise order and one continuous beep
Live Wire Check R-S and S-T lamps will light if voltage between wires are within voltage range
ON lamp blinks to indicate battery low status when instrument is turned on
Auto shut off if no activity is detected after power is turned ON for 15 minutes
Two "AA" size batteries; rated voltage:DC3.0V; maximum rated power:300mVA; continuous use:approx.70 hours(standby)
70W×75H×30D mm, 200g; Cord length:0.7m
Carrying case×1, strap×1, R6P manganese battery×2, spiral tube×1, user manual
14
W
HS-1
45
P
C
F
EXTERNAL
MULTIPLIER
HB-1 SPECIFICATIONS
HB-1(500V), HB-1(750V)
Model & rated voltage
JIS-Class 0.5 (±0.5% at rated voltage)
Accuracy (50/60 Hz)
1mA
Rating
See figure, 75g
Dimensions, mass
None
Accessories
91
74
41
2-φ3.5
45
Used with a 1mA full scale meter
100.5
57
35
HB-1
112
15
M12
HS-2
5
Options & Peripherals
M5
HS-2 SPECIFICATIONS
Model & rated current HS-2(500A), HS-2(750A), HS-2(1000A)
Accuracy (50/60 Hz)
JIS-Class0.5 (±0.5% at rated current)
Rating
50mV
Dimensions, mass
500A type:115(P),155(C),45(W),20(F)mm,740g
750A type:135(P),175(C),60(W),30(F)mm
1000A type:135(P),175(C),60(W),30(F)mm
Accessories
None
Note: The total resistance of the connection cord must be 0.1Ω or less.
φ60
Secondary
terminal
M5
15
K
Secondary
terminal
M5×0.8
k
20
K
K
L
K
φA
L
F
G
H
75
φA
D
φE
2.0
F
D
38
d
20
35
40
140
HS-1 SPECIFICATIONS
Model & rated current HS-1(30A), HS-1(50A), HS-1(75A), HS-1(100A),
HS-1(150A), HS-1(200A), HS-1(300A)
JIS-Class 0.5 (±0.5% at rated current)
Accuracy (50/60 Hz)
50 mV
Rating
30A type:20(E), 6(F),M4(d), M5(d) mm, 110g
Dimensions, mass
50A type:20(E), 8(F),M4(d), M8(d) mm, 150g
75A type:20(E), 8(F),M4(d), M8(d) mm, 155g
100A type:20(E),15(F),M5(d), M8(d) mm, 250g
150A type:20(E),15(F),M5(d), M8(d) mm, 250g
200A type:25(E),15(F),M5(d),M10(d) mm, 320g
300A type:25(E),15(F),M5(d),M10(d) mm, 330g
None
Accessories
2
E
φ4.5
85
120
53
CURRENT TRANSFORMER
Used with a 50mV full scale meter
26
Field Measuring Instruments
Options & Peripherals
CT-2MR CT-5MRN
SHUNTS
HS-1 HS-2 EXTERNAL
70
80
100~200A:φA=φ20
250~400A:φA=φ30
5
20
35
CT-2MR
B
C
2-M6
8
45
52
CT-5MRN
CT-2MR:SPECIFICATIONS
Model & rated current CT-2MR(100A), CT-2MR(120A), CT-2MR(150A),
CT-2MR(200A), CT-2MR(250A), CT-2MR(300A)
JIS-Class1.0 (±1% of rated value)
Accuracy (50/60 Hz)
2VA
Rated load
5A (all models)
Secondary current
Conductor voltage rating 1150VAC
See figure
Dimensions
None
Accessories
CT-5MRN:SPECIFICATIONS
Model & rated current CT-5MRN(100A), CT-5MRN(120A), CT-5MRN(150A),
CT-5MRN(200A), CT-5MRN(250A), CT-5MRN(300A),
CT-5MRN(500A), CT-5MRN(600A)
Accuracy (50/60 Hz)
JIS-Class1.0 (±1% of rated value)
Rated load
5VA
Secondary current
5A (all models)
Conductor voltage rating 1150VAC
Dimensions
100-200A:23(φA),70(B),85(C),65(D),60(φE),45(F),75(G),83(H) mm
250-300A:32(φA),70(B),85(C),76(D),70(φE),50(F),85(G),91(H) mm
500-600A:50(φA),80(B),100(C),89.5(D),85(φE),58(F),100.5(G),105(H)mm
Accessories
None
Wiring: Pass the wire through center of the C.T. When measuring under 100 A, use the
equation below to find the number of times that the wire is to be passed through. Number
of turns = (Primary current of C.T.)÷(Maximum value measured) However, in order to
make the number of turns a full number, select the primary current or full scale.
Advancing power saving and automation
Standard SPECIFICATIONS
2103:±2.5% class, 2104: ±1.5% class
Meter class
Deflecting range
Passing type, full scale
Setting pointer
Lance shape, upper limit and lower limit pointer
Setting accuracy
±1.5% of scale length
Minimum setting width
Within 3% of scale length
Relay power delay circuit Approx. 2 second
Relay output response
Approx. 0.5 second
Output contact capacity
5A (under condition of 250V AC, 30V DC,
resistance load)
Power supply
100/ 200V AC ±10%
●Electronic design assures high accuracy and reliability
●Ultra sensitive 1 mA, 10 mV DC movement
●Replaces relays in plug-in systems
●Includes a display lamp to illuminate movement at a glance
●Relay action delays circuit closure upon power on
●Both power circuitry and relay built-in
OPTIONS
(Special specifications)
●±1.5% class: for Model 2103
●Extended scale: double or triple extended scale
●Segmented scale: magnified scale for up to 40 % of the maximum scale value
●Double deflection meter: for example, zero-centered scale
●Relay response time:time constant 0.05 second fixed (DC) and variable types also
available
●Setting accuracy: Version with ±1.0 % type
●Delay time: Version with variable delay time after power on. 0.1 to 10 seconds: (for
instruments input DC), 2 to 12 seconds: (for instruments input AC)
●Output signal: Version with 1 V DC /f.s. output terminal
*not isolated from input circuit ground.
●Power supply: Version with 110, 120, 220, 230, 240 V AC ±10 %
2103
2104
■ Dimensions
■ Standard Scale Graduations
Full-Scale
Value
Graduations
1,10,100
50
1.5,15,150
30
2,20,200
40
2.5,25,250
50
3,30,300
30
5,50,500
50
6,60,600
30
7.5,75,750
37.5
Graduation Illustration
0
2
4
0
5
0
0
■ Standard Full-Scale Values
DC Ammeter
Std. Full-Scale
Value
DC Voltmeter
Meter
Std. Full-Scale
Sensitivity
Value
Spec.
10
50mV
Full-Scale:
4-20mA
50mV
10 mV
15
30
50 *2
100
150
300
500
1V
1.5
3
5
10
15
30
50
100
150
300
100kΩ/V
100kΩ/V
100kΩ/V
100kΩ/V
100kΩ/V
100kΩ/V
100kΩ/V
10kΩ/V
10kΩ/V
10kΩ/V
10kΩ/V
10kΩ/V
10kΩ/V
10kΩ/V
10kΩ/V
10kΩ/V
10kΩ/V
10kΩ/V
10kΩ/V
Full-Scale:
1-5V
10kΩ/V
1
1
Std. Full-Scale
Value
200 µA
500
1 mA*3
2
5
10
20
50
100
200
500
1A
2
3
5 *4
Meter
Std. Full-Scale
Sensitivity
Value
Spec.
50mV
50 mV
100
150
300
500
1V
1.5
3
5
10
15
30
50
100
150
300
Meter
Sensitivity
Spec.
10kΩ/V
10kΩ/V
10kΩ/V
10kΩ/V
1kΩ/V
1kΩ/V
1kΩ/V
1kΩ/V
1kΩ/V
1kΩ/V
1kΩ/V
1kΩ/V
1kΩ/V
1kΩ/V
1kΩ/V
1kΩ/V
●When the full-scale value is larger than 5A, an external
CT is used with the 5A instrument denoted by.*4
●When the full-scale value is larger than 300V, an
external multiplier is used with the 1-mA instrument
denoted by.*3
20
2
2
3
2
0
15
15
15
2
20
25
3
4
4
4
10
5
6
6
7.5
■ Contact operation
a
HL type
ON
c contact used
OFF
ON
L setting
L type
H setting
ON
OFF
L setting
H type
OFF
ON
H setting
Options & Peripherals
1 µA
10
20
50
100
200
500
1 mA*1
2
5
10
20
50
100
200
500
1 A
2
5
10
20
Meter
Sensitivity
Spec.
Rectifying AC Voltmeter
10
5
0
Rectifying AC Ammeter
8
10
5
0
0
6
●When the full-scale value is larger than 20A, an
external shunt device is used with the 50-mV
instrument denoted by.*2
●When the full-scale value is larger than 300V, an
external multiplier is used with the 1-mA instrument
denoted by.*1
54
Note: * marked products are discontinued models.
Test Leads and Probes ...Included as accessories with main unit, or sold separately (optional products)
Options & Peripherals
55
3851-10
TEST LEAD
. . for 3251*, 3252*, 3253*
. . . . 3801, 3802, 3803
. . . . 3804, 3805, 7015
9014
HIGH VOLTAGE PROBE
. . for 3081*, 3205*, 3207*
. . . . 3208*, 3209*, 3212*,
. . . . . . . 3230*, 3250s
9017
HIGH VOLTAGE PROBE
. . for 3030*, 3012*, 3015*
. . . . . . 3021*, 3030-10
9021-01
THRMISTER TEMPERATURE
PROBE
. . for 3000*, 3007*, 3015*
. . . 3021*, 3030-10, 3100
9060
TEST LEAD
. . . . . for 3008, 3125*
9060-01
TEST LEAD
. . for 3108-01*, 3109-01*
. . . . . . . . . . . . . . 3008
9067
TEST LEAD
. . for 3127-10, 3128-10
. . . 3000*, 3021*, 3100*
. . . 3102*, 3104*, 3261*
. . . 3262*, 3264*, 3265*
9094
OUTPUT CORD
. . . . for 3412-50, 3283
. . . 3284, 3285, 3225*
. . . . . . . . 3404, 3422*
9140
4-TERMINAL PROBE
. . . . . for 3503, 3511-50
. . 3520*, 3521*, 3522-50
. . . . . . . . . . . . 3532-50
9143
PINCHER PROBE
. . . . . for 3503, 3511-50
. . 3520*, 3521*, 3522-50
. . . . . . . . . . . . 3532-50
9153
TEST LEADS WITH FUSE
. . for 3021*, 3030, 3127
. . . . . . . . . . . . . . 3128
9165
CONNECTION CORD
. for 3601*, 7075, 8850*
. . 8852*, 8852-01*, 8855
. . . LCR/Z HiTESTER
9166
CONNECTION CORD
. . . for 3601*, 7075, 8850 *
. . . . . 8852*, 8852-01*,
. . . LCR/Z HiTESTER
9168
INPUT CORD
. . . . . for 7010*, 7011
9170
TEST LEAD
. . for 3156, 3155, 3200*
. . . 3210*, 3215*, 3216*
. . . 3222*, 3223*, 3230*
. . . 3231*, 3233*, 3234*
. . . 3236*, 3256*, 3257*
. . . . . . . . . . . . . . 7011
9177
INPUT CORD
. . for 8815*, 8816*, 8830*
. . . . . . . 8831*, 8832*
. . . . . . . . . . . . . 8833*
. . . . . . . . . . . . . 8821*
. . . . . . . . . . . . . . 9555
9178
VOLTAGE CORD
. . . . . . . . . . . for 3165
9179
VOLTAGE CORD
. . . for 3167, 3192*, 3195*
9185
TEST LEAD
. . . . . . . . . . for 3255*
9186
INPUT CORD
. . for 8852*, 8852-01*
9190
VOLTAGE APPLY PROBE
. . . . . . . . for 3155-01*
9195
ENCLOSURE
PROBE
. . . . . . . . for 3155-01*
9196
APPLY UNIT
. . . . . . . . for 3155-01*
9197
CONNECTION CORD
. . for 8806*, 8807, 8808
. . . . 8826, 8835, 8841
. . . . . . . . . 8842, 8855
9198
CONNECTION CORD
. . for 8806*, 8807, 8808
. . . . 8826, 8835, 8841
. . . . . . . . . 8842, 8855
9199
CONVERSION ADAPTOR
. . for 8806*, 8807, 8808
. . . . 8826, 8835, 8841
. . . . . . . . . 8842, 8855
9207
TEST LEAD
. . . . . . . . . for 3030-10
9207-10
TEST LEAD
. . . . . . . . . . . for 3281,
3282, 3284, 3285
3256-50/-51, 3257-50/-51
9208
TEST LEADS
. . for 3280*, 3280-01*
. . . . . 3280-10, 3280-11
. . 3280-20, 3287, 3288
9209
TEST LEADS HOLDER
. . for 3280*, 3280-01*
. . . . . 3280-10, 3280-11
. . 3280-20, 3287, 3288
9215
MEASURING CABLE
. . . . . . . . . . . for 3151
9217
CONNECTION CORD
. . for 8806*, 8807, 8808
. . . . 8826, 8835, 8841
. . . . . . . . . 8842, 8855
9219
CONNECTION CABLE
. . for 9695-02, 9695-03
9257
CONNECTION CORD
. . for 8205-10, 8206-10
. . . . . . 8220*, 3454-11
. . . . . . . 3454-10, 3453
. . . . . . 3118-11, 3118-12
9261
TEST FIXTURE
. . . . . for 3503, 3511-50
. . 3520*, 3521*, 3522-50
. . 3530*, 3531*, 3532-50
9262
TEST FIXTURE
. . . . . for 3503, 3511-50
. . 3520*, 3521*, 3522-50
. . 3530*, 3531*, 3532-50
9263
THERMISTER
TEMPERATURE PROBE
. . . 3503, 3511-50, 3520*
. . 3521*, 3522-50, 3530*
. . . . . . 3531*, 3532-50
9264-01
WIRING ADAPTER
. . . . . . . . . . . for 3196
9264-02
WIRING ADAPTER
. . . . . . . . . . . for 3196
9265
MEASUREMENT CABLE
. . . . . . . . . . . for 3143
9268
DC BIAS VOLTAGE UNIT
. . . . . for 3503, 3511-50
. . . . . 3522-50, 3532-50
9269
DC BIAS CURRENT UNIT
. . . . . for 3503, 3511-50
. . . . . 3522-50, 3532-50
Note: * marked products are discontinued models.
9289
TEST PROBE
. . . for 3118-11, 3118-12
. . 3154, 3453, 3454-10
. . . . . . . . . . . . 3454-11
9292
TEST PROBE
. . . for 3451-11, 3451-12
. . . . . 3451-13, 3451-14
. . . . . 3451-15, 3452-11
. . . . . . . . . . . . 3452-12
9293
PIN TYPE EARTH PROBE
. . . for 3451-11, 3451-12
. . . . . 3451-13, 3451-14
. . . . . 3451-15, 3452-11
. . . . . . . . . . . . 3452-12
9294
TEST PROBE
. . . for 3117s*, 3118-11
. . 3118-12, 3154, 3453
9296
CURRENT PROBE
. . . . for 3157, 3157-01
9297
CURRENT APPLY PROBE
. . . . for 3157, 3157-01
9299
SWITCHED PROBE
. . . . . . . . . . . for 3154
9300
CONNECTION CABLE
. . . . . . . . . . . for 3541
9326
CONNECTION CORD
. . . for 8205*, 8205-10
9437
CONNECTION CABLE
. . . . . for 8845*/8846*
9438
VOLTAGE CORD
. . . . . . . . . . for 3166*
9438-02
VOLTAGE CORD
. . . . . . . . . . . for 3196
9438-03
VOLTAGE CORD
. . . . . . . . . . . for 3169
9452
CLIP TYPE LEAD
. . for 3227*, 3239, 3540
. . . . 3541, 3555, 3560
9453
FOUR TERMINAL LEAD
. . for 3227*, 3239, 3540
. . . . 3541, 3555, 3560
9454
ZERO ADJUSTMENT BOARD
. . for 3155-01*, 3227*,
. . . . . . . . . 3239, 3540,
3541, 3555, 3560
9455
PIN TYPE LEAD
. . for 3227*, 3239, 3540
. . . . 3541, 3555, 3560
9460
CLIP TYPE LEAD WITH
TEMPERATURE SENSOR
. for 3540, 3550, 3551
9461
PIN TYPE LEAD
. . for 3155-01*, 3227*
. . . . 3239, 3540, 3541
. . . . . . . . . 3555, 3560
9465
PIN TYPE LEAD
. for 3239, 3540, 3541
. . . . 3550, 3551, 3555
. . . . . . . . . . . . . . 3560
9466
REMOTE CONTROL
SWITCH
. . . . . . for 3551, 3560
9467
LARGE CLIP TYPE LEAD
. . for 3227*, 3239, 3540
. . . . 3541, 3550, 3551
. . . . . . . . . . . . . . 3560
9574
INPUT CORD
. . . . . . . . . . . . . . . . for
8840*/45*/46*,8853*,8851*
9615
H.V.TEST LEAD
. for 3153, 3158, 3159
. . . . . . . . . 3173, 3930
9617
CLIP ON BASE
. for 3501, 3801, 3802
. . . . . . . . . 3804, 3805
9618
CLIP-TYPE LEAD
. for 3501, 3801, 3802
. . . . . . . . . 3804, 3805
9635
VOLTAGE CORD
. . . for 3286*, 3286-20
9635-01
VOLTAGE CORD
. . . . . . . . . for 3286-20
9639
CONNECTION CABLE
. . . . . . for 3637, 3645
9641
CONNECTION CABLE
. for 8420-51, 8421-51, 8422-51
9665
10:1PROBE
. . . . . . . . . . . for 8855
9666
100:1PROBE
. . . . . . . . . . . for 8855
9677
SMD TEST FIXTURE
. . . . for 3503, 3511-50,
3522-50, 3532-50, 3535
9699
SMD TEST FIXTURE
. . . . for 3503, 3511-50,
3522-50, 3532-50, 3535
ALLIGATOR CLIPs
. . . . . . (used with Test
Lead, insert), for 9170,
or similar devices
Options & Peripherals
9287-10
CLIP TYPE LEAD
. . for 3155-01*, 3226*
. . . 3227*, 3239, 3540
. . . . 3541, 3555, 3560
9288
BREAKER PIN
. . for 3118-11, 3118-12,
. . . . . 3451-11, 3451-12,
. . . . . 3451-13, 3451-14,
. . . . . 3451-15, 3452-11,
. . . . . 3452-12, 3452-13,
. . . . . . . 3453, 3454-10,
. . . . . . . . . . . . 3454-11
56
Note: * marked products are discontinued models.
RECORDING PAPERs ...Sold separately (optional products)
Model
3155-01 LEAK CURRENT HiTESTER*
3165 CLAMP ON POWER HiTESTER*
3193 POWER HiTESTER
3194 MOTOR/HARMONIC HiTESTER
3196 POWER QUALITY ANALYZER
3541 RESISTANCE HiTESTER
Recording
Paper
9233
9223
9232
9232
9237
9237
8205-10 MICRO HiCORDER
9235
9236-01
8206-10 MICRO HiCORDER
9235
9236-01
8420-01 MEMORY HiLOGGER*
8420-51 MEMORY HiLOGGER
8421-01 MEMORY HiLOGGER*
8421-51 MEMORY HiLOGGER
8422-01 MEMORY HiLOGGER*
8422-51 MEMORY HiLOGGER
8715-01 POWER HiCORDER
8807-01 MEMORY HiCORDER
9234
9234
9234
9234
9234
9234
9234
9234
58 mm × 10 m, 10 rolls
58 mm × 10 m, 10 rolls
74 mm × 10 m, 10 rolls
74 mm × 10 m, 10 rolls
80 mm × 25 m, 4 rolls
80 mm × 25 m, 4 rolls
60 mm × 15 m, 10 rolls
60 mm × 15 m,
climate-resistant, 10 rolls
60 mm × 15 m, 10 rolls
60 mm × 15 m,
climate-resistant, 10 rolls
112 mm × 18 m, 10 rolls
112 mm × 18 m, 10 rolls
112 mm × 18 m, 10 rolls
112 mm × 18 m, 10 rolls
112 mm × 18 m, 10 rolls
112 mm × 18 m, 10 rolls
112 mm × 18 m, 10 rolls
112 mm × 18 m, 10 rolls
Recording
Paper
9234
9234
9234
Model
Paper Size
8807-51 MEMORY HiCORDER
8808-01 MEMORY HiCORDER
8808-51 MEMORY HiCORDER
8826 MEMORY HiCORDER
9229
9229-01
8835-01 MEMORY HiCORDER
8841 MEMORY HiCORDER
8842 MEMORY HiCORDER
8852, 8852-01 MEMORY HiCORDER*
8855 MEMORY HiCORDER
8860, 8861 MEMORY HiCORDER
8992 PRINTER UNIT
8994 PRINTER UNIT
9203 DIGITAL PRINTER
9442 PRINTER
9604 PRINTER UNIT
9670 PRINTER
9221
9231
9231
9221
9231
9231
9234
9231
9233
1196
9232
9237
Paper Size
112 mm × 18 m, 10 rolls
112 mm × 18 m, 10 rolls
112 mm × 18 m, 10 rolls
264 mm × 30 m, 6 rolls
264 mm × 30 m,
perforated, 6 rolls
110 mm × 30 m, 10 rolls
216 mm × 30 m, 6 rolls
216 mm × 30 m, 6 rolls
110 mm × 30 m, 10 rolls
216 mm × 30 m, 6 rolls
216 mm × 30 m, 6 rolls
112 mm × 18 m, 10 rolls
216 mm × 30 m, 6 rolls
58 mm × 10 m, 10 rolls
112 mm × 25 m, 10 rolls
74 mm × 10 m, 10 rolls
80 mm × 25 m, 4 rolls
Temperature probes or sensors ...Included as accessories with main unit, or sold separately (optional products)
Model
Type/Note
9021-01
-50 to 200˚C
THERMISTER TEMPERATURE PROBE
Sheath type, up to 750°C
non-waterproof
Surface type, up to 400°C
non-waterproof
Sheath type, up to 750°C
non-waterproof
Sheath type, up to 750°C
non-waterproof
reference contact compensation
9180 TEMPERATURE PROBE
9181 TEMPERATURE PROBE
9182 TEMPERATURE PROBE
9183 TEMPERATURE PROBE
9184 RJ SENSOR
-25 to 80˚C
9188 TEMPERATURE PROBE
included with the 3540
9451 TEMPERATURE PROBE
9462 THERMISTER TEMPERATURE PROBE -50 to 150˚C
9463 TEMPERATURE HUMIDITY SENSOR fixed type
9464 TEMPERATURE HUMIDITY SENSOR extension type, (2.7m)
included with the 3227
9472 TEMPERATURE PROBE
9472-50 TEMPERATURE PROBE
9473 TEMPERATURE PROBE
9473-50 TEMPERATURE PROBE
9474 TEMPERATURE PROBE
9475 TEMPERATURE PROBE
Sheath type, up to 300°C
waterproof structure
Sheath type, up to 300°C
waterproof structure
Sheath type, up to 800°C
waterproof structure
Sheath type, up to 800°C
waterproof structure
Sheath type, up to 300°C
waterproof structure
Sheath type, up to 500°C
waterproof structure
Compatible Instrument
Model
Type/Note
3030-10, 3127-10, 3128-10
9476 TEMPERATURE PROBE
3412-50, 3441, 3442
9476-50 TEMPERATURE PROBE
3412-50, 3441, 3442
9478 TEMPERATURE PROBE
3412-50, 3441, 3442
9479 TEMPERATURE PROBE
3412-50, 3441, 3442
7011
3227
3540
3281, 3282
3625
3625
3441, 3442
3446-01
3441, 3442
3446-01
3441, 3442
3441, 3442
9680 HUMIDITY SENSOR
9680-01 HUMIDITY SENSOR
9680-02 HUMIDITY SENSOR
9631-01
TEMPERATURE SENSOR
9631-11
TEMPERATURE SENSOR
9631-21
TEMPERATURE SENSOR
9631-02
TEMPERATURE SENSOR
9631-03
TEMPERATURE SENSOR
9631-04
TEMPERATURE SENSOR
9631-14
TEMPERATURE SENSOR
9631-24
TEMPERATURE SENSOR
9631-05
TEMPERATURE SENSOR
3446-01
3447-01(Pt-100)
3447-01(Pt-100)
10 m length
1 m length
3632 / 3633 / 3641-20
5 m length
3632 / 3633 / 3641-20
10 m length
3632 / 3633 / 3641-20
1 m length
3632 / 3633 / 3641-20
1 m length
3632 / 3633 / 3641-20
1 m length
3632 / 3633 / 3641-20
5 m length
3632 / 3633 / 3641-20
10 m length
3632 / 3633 / 3641-20
30 mm length
3632 / 3633 / 3641-20
5 m length
9478 (Pt-100)
φ5.0
φ2.3
φ2.3 (9472-50)/φ4.8(9473-50)
9472, 9473
Options & Peripherals
150 (9472-50)
300 (9473-50)
100
100
1000
9476-50 (K)
9479 (Pt-100)
φ2.3
150
Hand switch
1000
φ5.0
φ20
9474, 9475
25
90
100
1000
100
157
9476
9080
9080-01/02
57
3441, 3442
3641-20
3641-20
3641-20
1 m length
9472-50, 9473-50 (K)
9021-01
Compatible Instrument
Surface type, up to 500°C
non-waterproof
Surface type, up to 500°C
non-waterproof
Sheath type, up to 300°C
Waterproof structure
Sheath type, up to 300°C
Waterproof structure
9631-01
9631-11
9631-21
9631-02
9631-03
9631-04
9631-14
9631-24
9631-05
1000
Note: * marked products are discontinued models.
All Accessories
1195 RECORDING PAPER . . . . . for 3192 (10rolls)*
9151-02 GP-IB CONNECTOR CABLE
9221 RECORDING PAPER . . . . . . . . . . . . . . . . . . .
1196 RECORDING PAPER
. . . . . . . . . . . . . . . . . . . . for 3511, 3330-02*, 3332 (2m)
. . . . for 8801*, 8802*, 8835-01 (110mm × 30m 10rolls)
. . . . . . . . . . . . . for 3154, 3332-02, 3511, 3196 (10rolls)
9151-04 GP-IB CONNECTOR CABLE
9222 RECORDING PAPER
3851-10 TEST LEAD
. . . . . . . . . . . . . . . . . . . . for 3511, 3330-02*, 3332 (4m)
. . . . . . . . . . . . . . . . . . . . for 3224-02*, 3225-02*, 3235*
. . . . . . . . . . . . . . . for 3801, 3802, 3803, 3804, 3805
9153 TEST LEADS WITH FUSE
. . . . . . . . . . . . . . . . . . . . . . 9200 (38mm × 8.5m 5rolls)
3852 RS-232C PACKAGE . . . . . . . . . for 3801, 3802
. . . . . . . . . . . . . . . . . . for 3030, 3127*, 3128*, 3127-11
9223 RECORDING PAPER
3853 CARRYING CASE . . . . . . for 3803, 3804, 3805
9165 CONNECTION CORD
. . . . . . . . . . . . for 3165*, 3191 (80mm × 30m 5rolls)*
3854 RS-232C PACKAGE . . . . for 3803, 3804, 3805
. . . . . . . . . for 3511, 3601*, 7075, 7075-01(BNC-BNC)
9224 RECORDING PAPER
3909 INTERFACE PACK. . . . . . for 3443, 3444, 3445
9166 CONNECTION CORD
. . . . . . . . . . . . . . . . . for 8820 (216mm × 50m 6rolls)*
3920-01 DATA READER for 3650. . . . . . . . for 3650
. . . . . . . . . for 3511, 3601*, 7075, 7075-01 (BNC-Clip)
9226 RECORDING PAPER
9005-01 CLAMP ON PROBE
9168 INPUT CORD . . . . . . . . . . . . . . for 7010*, 7011
. . . . . . . . . . . . . . . . . for 8601 (24mm × 15m 10rolls)*
. . . . . . . for 3000*, 09*, 11*, 15*, 20*, 21*, 30, 3215*
9170 TEST LEAD
9227 RECORDING PAPER
9010 CLAMP ON PROBE . . . . . . . . for 8200*, 3255
. . . for 3200*, 10*, 16*, 22*, 23*, 33*, 34*, 36*, 7011
. . . . . . . . . . . . . . . . . . . for 3234 (38mm × 3m 5rolls)*
9010-10 CLAMP ON PROBE . . . . . . for 8714*, 8715
9177 INPUT CORD . . for 8904*, 8906*, 8932*, 9555
9228 RECORDING PAPER
9014 HIGH VOLTAGE PROBE
9178 VOLTAGE CORD . . . . . . . . . . . . . . . . for 3165*
. . . . . . . . . . . . . . . . . for 8850 (114mm × 30m 10rolls)*
. . . . for 3081*, 3205*, 3207*, 3208*, 3209*, 3212*, 3230*, 3250S
9179 VOLTAGE CORD . . . . . . . . . . . . . . . . for 3195*
9229 RECORDING PAPER . . . . . . . . . . . . . . . . . . .
9017 HIGH VOLTAGE PROBE
9180 SHEATH TYPE TEMPERATURE PROBE
. . . . . . . . . . . . . . . . . for 8825 (264mm × 30m 6rolls)*
. . . . . . . . . . . . . for 3000*, 3012*, 3015*, 3021*, 3030
. . . . . . . . . . . . . . . . . . . . . for 3412-50, 3441/42 (200°C)
9229-01 RECORDING PAPER(PERFORATED)
9018 CLAMP ON PROBE
9181 SURFACE TEMPERATURE PROBE
. . . . . . . . . . . . . . . . . . for 8826 (264mm × 30m 6rolls)
. . . . . . . . . . . . . . . . . for 8205, 8206, 8805, 8807, 8808
. . . . . . . . . . . . . . . . . . . . . for 3412-50, 3441/42 (400°C)
9231 RECORDING PAPER
9018-10 CLAMP ON PROBE . . . . . for 8714*, 8715
9182 SHEATH TYPE TEMPERATURE PROBE
. . . . . . . for 8840*, 8841, 8842 (216mm × 30m 6rolls)
9021-01 THERMISTER TEMPERATURE PROBE
. . . . . . . . . . . . . . . . . . . . for 3412-50, 3441/42 (1000°C)
9232 RECORDING PAPER
. . for 3000*, 3007*, 3015*, 3021*, 3030, -12, 3100*, 3127-11
9183 SHEATH TYPE TEMPERATURE PROBE
. . . . . . . . . . . . . . . . . . . . for 8804*, 8205, 8206, 8806*
9032 METAL CONTACT TIP . . . . 3402*, 3403, 3404
. . . . . . . . . . . . . . . . . . . . . for 3412-50, 3441/42 (800°C)
. . . . . . . . . . . . . . . . . . . . . . 3193 (74mm × 10m 10rolls)
9033 RUBBER CONTACT TIP
9184 TEMPERATURE PROBE
9233 RECORDING PAPER . . . . . . for 9203, 9203-01
. . . . . . . . . . . . . . . . . . . . . . . . . . for 3402*, 3403, 3404
. . . . . . . . . . . . . . . . . . . . for 7010*, 7011 (-25°C~80°C)
9234 RECORDING PAPER . . . for 8807, 8808, 8420
9035 AC ADAPTER
9185 TEST LEAD . . . . . . . . . . . . . . . . for 3154, 3255
9235 RECORDING PAPER . . . . . . . . for 8205, 8206
. for 3108*, 31*, 32*, 61*, 62*, 3220*, 05*, 09*, 3402~04, 22 (6V)*
9186 INPUT CORD . . . . . . . . . . . . . . for 8851(10:1)*
9236-01 RECORDING PAPER . . . . . . for 8205, 8206
9036 AC ADAPTER
9188 TEMPERATURE PROBE. . . . . . . . . . for 3227*
9237 RECORDING PAPER . . . . . . . . . . . . . for 9670
. . . . . . . . . for 3110*, 3204*, 9005, 9006*, 3411 (9V)*
9190 VOLTAGE APPLY PROBE. . . . . . . . . for 3155*
9245 CARRYING CASE . . . . . . . . . . . . . . for 3286-20
9039 AC ADAPTER. . . . . . . . . . . . . . . for 3501 (12V )
9195 ENCLOSURE PROBE. . . . . . . . . . . . for 3155*
9246 CARRYING CASE . . . . . . . . . . . for 3664, 9742
9196 APPLY UNIT . . . . . . . . . . . . . . . . . . . . for 3155*
9257 CONNECTION CORD. . for 8205, 8206, 8220*
9197 CONNECTION CORD
9261 TEST FIXTURE. . . . . . . . . . . . . . . . . . . . for LCR
9060-01 TEST LEAD . . . . . . . . . . . . for 3108*, 3109*
. . . . . for 8806*, -01*, 8807, 8808, 8826, 8835-01, 8841, 8842
9262 TEST FIXTURE
9067 TEST LEAD
9198 CONNECTION CORD
. . . . . . . . . . . . . . for 3502*, 3511, 3520*, 3531*, 3532
. . . . . . . . . . . . . . . . for 3261*, 3262*, 3127-10, 3128-10
. . . . . for 8806*, -01*, 8807, 8808, 8826, 8835-01, 8841, 8842
9263 THERMISTER TEMPERATURE PROBE
9070 AC ADAPTER. . . . . . . . . . . . . . . for 3118, 3119*
9199 CONVERSION ADAPTER
. . . . . . . . . . . . . . . . . . . . . . . . . . . for 3511, 3522, 3532
9073 RECORDING PAPER
. . . . . for 8806*, -01*, 8807, 8808, 8826, 8835-01, 8841, 8842
9264-01 WIRING ADAPTER . . . . . . for 3196, (3P3Ø)
. . . . . . . . . . . . . . . . . for 8201*, 8202*, 8204 (10rolls)*
9203 DIGITAL PRINTER
9264-02 WIRING ADAPTER . . . . . . for 3196, (3P4Ø)
9074 RECORDING PAPER
. . . . . . . . . . . . . . . . . . . . . for 3227*, 3540, 3550, 3560
9265 MEASUREMENT CABLE . . . . . . . . . . for 3143
. . . . . . . . . . . . . . . . . . . . . . . . for 8203 (1ch) (10rolls)*
9207 TEST LEAD
9267 SAFETY TEST DATA MANAGEMENT SOFTWARE
9081 EXTERNAL SHUNT . . . . . . . . . for 3245 (10A)*
. . . . . . . . . . . for 3030-10, 3030-12, 3282-01, 3284, 3285
. . . . . . . . . . . . . . . . . . . . for 3153, 3156/57/58/59, 3332
9083 CARRYING CASE
9207-10 TEST LEAD
9268 DC BIAS VOLTAGE UNIT
. . . . . . . . . . . . . . . . . . for 3108*, 3109*, 3131*, 3132*
. . . . . . . . . . for 3281, 82, 84, 85, 3256-51/51, 3257-50/51
. . . . . . . . . . . . . . . . . . . . . . . . . . . for 3511, 3522, 3532
9084 CARRYING CASE
9208 TEST LEADS
9269 DC BIAS CURRENT UNIT
. . . . . . . . . . . . . . . . . . for 3180*, 3222*, 3223*, 3235*
. . . . . . . . . . . . . . . . . . for 3287, 3280, -01, -10, -11, 3288
. . . . . . . . . . . . . . . . . . . . . . . . . . . for 3511, 3522, 3532
9088 CARRYING CASE . . . . . . . . . for 3030*, 3021*
9209 TEST LEADS HOLDER
9270 CLAMP ON SENSOR
9094 OUTPUT CORD
. . . . . . . . . . . . . . . . . . for 3287, 3280, -01, -10, -11, 3288
. . . . . . . . . . . . . . for 3191*, 3165*, 3192*, 3167 (20A)
. . . . for 3154, 3225*, 3403, 04, 3412-50, 3422*, 3283, 84, 85
9211 REFLECTIVE TAPE . . . for 3402*, 3403*, 3404
9271 CLAMP ON SENSOR
9132 CLAMP ON PROBE . . . . . . . . . . for AC1000A
9212 PERIPHERAL RING. . . for 3402*, 3403*, 3404
. . . . . . . . . . . . . for 3191*, 3165*, 3192*, 3167 (200A)
9132-10 CLAMP ON PROBE . . . . . for 8714*, 8715
9213 CONTACT ADAPTER . . . . . . . . for 3403, 3404
9272 CLAMP ON SENSOR
9135 RECORDING PAPER . . . . . . for 3215 (5rolls)*
9214 AUXILIARY EARTHING ROD . . . . . . . for 3151
. . . . . . . . . . . for 3191*, 3165*, 3192*, 3167 (20/200A)
9140 4-TERMINAL PROBE
9215 MEASURING CABLE . . . . . . . . . . . . . for 3151
9274 CLAMP ON AC/DC SENSOR
. . . . . . . . . . . . . . . . . . . . for 3511, 3520*, 3521*, 3532
9216 CABLE WINDER . . . . . . . . . . . . . . . . . for 3151
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for AC/DC20A
9143 PINCHER PROBE
9217 CONNECTION CORD
9276 CLAMP ON AC/DC SENSOR
. . . . . . . . . . . . . . for 3511, 3520*, 3521*, 3530*, 3532
. . . . . . . . . . . for 8806*, -01*, 8807, 08, 8826, 8841, 42
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for AC/DC150A
9144 CARRYING CASE . . . . . . . . . for 3030*, 3021*
9219 CONNECTION CABLE
9277 UNIVERSAL CLAMP ON CT
9148 CARRYING CASE
. . . . . . . . . . . . . . . . . . for 3169, 3196, 9695-02, 9695-03
. . . . . . . . . . . . . . . . . . . . for 3192*, 3167 (AC/DC20A)
Options & Peripherals
9050 EARTH NETS . . . . . . . . for 3124*, 3150*, 3151
9060 TEST LEAD . . . . . . . . . . . . . . . for 3008, 3125*
. . . . . . . . . . . . . . . . . . . for 3128, 3261*, 3262*, 3263*
58
Note: * marked products are discontinued models.
Options & Peripherals
59
9278 UNIVERSAL CLAMP ON CT
9364 CARRYING CASE . . . . . . . . . . . . . . for 3118-12
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3551, 3560
. . . . . . . . . . . . . . . . . . . for 3192*, 3167 (AC/DC200A)
9365 CARRYING CASE . . . . . . . . . . . . . . . for 3165*
9455 PIN TYPE LEAD
9279 UNIVERSAL CLAMP ON CT
9371 CARRYING CASE . . . . . . . . . . . . . . . . for 3255
. . . . . . . . . . . . . . . . . . . . for 3226*, 3227*, 3540, 3560
. . . . . . . . . . . . . . . . . . . for 3192*, 3167 (AC/DC 500A)
9372 CARRYING CASE . . . . . . . . . for 8804*, 8806*
9458 AC ADAPTER. . . . . . . . . . . . . . . . . . . . for 3196
9287-10 CLIP TYPE LEAD . . . . . . . . for 3226*, 3540
9375 CARRYING CASE . . . . . . . . . . . for 9277~9279
9459 BATTERY PACK. . . . . . . . . . . . . . . . . . for 3196
9288 BREAKER PIN . . . . . . . . . . . . . . . . . . . for 3452
9376 CARRYING CASE . . . . . . . . . . . . . . . for 3423*
9460 CLIP TYPE LEAD WITH TEMPERATURE SENSOR
9289 TEST PROBE . . . . . . . . . . . . . . . . . . for 3454-51
9377 CARRYING CASE . . . . . . . . . . . . . . . . for 3551
. . . . . . . . . . . . . . . . . . . . . . . . . for 3550, 3551, 3551-03
9290-10 CLAMP ON ADAPTER
9378 CARRYING CASE . . . . . . . . . . . . . . . . for 3256
9461 PIN TYPE LEAD. . . . . . . for 3155*, 3555, 3540
. . . . . . . . . . . . . . . . . . . . . . . . . . . for 1000A CT 10 : 1
9380 CARRYING CASE . . . . . . . . . . . . . . . . for 7011
9462 THERMISTER TEMPERATURE PROBE
9291 CLAMP ON SENSOR . . . . . . . . . . . . for 3166*
9382 CARRYING CASE . . . . . . . . . . . for 3550, 3555
. . . . . . . . . . . . . . . . . . . . . . . . for 3281*, 3282, 3282-01
9292 TEST PROBE . . . . . . . . . . . . . . . for 3451, 3452
9384 CARRYING CASE . . . . . . . . . . . . . . . . for 3452
9463 HUMIDITY SENSOR . . . . . . . . . . . . . . for 3625
9293 PIN TYPE EARTH PROBE . . . . for 3451, 3452
9386 CARRYING CASE
9464 HUMIDITY SENSOR . . . . . . . . . . . . . . for 3625
9294 TEST PROBE . . . . . . . . . . . . . . . for 3154, 3453
. . . . . . . . . . . . . . . . . . . . . for 3441, -01, -02, 3442-02, -03
9465 PIN TYPE LEAD
9296 CURRENT PROBE . . . . . . . . for 3157, 3157-01
9386-01 CARRYING CASE
. . . . . . . . . . . . . . . . . for 3550, 3551, 3552, 3555, 3540
9297 CURRENT APPLY PROBE
. . . . . . . . . . . . . . . . . . . . . . for 3441, 3442, 3446, 3447
9466 REMOTE CONTROL SWITCH
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3157, 3157-01
9388 CARRYING CASE . . for 8835*, 8835-01, 3155*
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3551, 9465
9299 SWITCHED PROBE . . . . . . . . . . . . . . for 3154
9390 CARRYING CASE . . . . . . . for 3030-10, 3030-12
9467 LARGE CLIP TYPE LEAD
9300 CONNECTION CABLE . . . . . . . . . . . . for 3541
9391 CARRYING CASE . . . . . . for 8807, 8808, 8420
. . . . . . . . . . . . . . . . . . . . . . . . . for 3550, 3226*, 3227*
9303 PT
9392 CARRYING CASE . . . . . . . . . . . . . . . . for 3625
9472 SHEATH TYPE TEMPERATURE PROBE
. . . . . . . for 8815*, 25*, 30*, 32*, 51* (PT 40:1,20:1 )
9393 CARRYING CASE . . . . . . . . . . . . . . . . for 3151
. . . . . . . . . . . . . . . for 3441, -01, -02, 3442-02, -03, 7015*
9305 TRIGGER CORD
9397-01 CARRYING CASE. . . . for 8855, 8841, 8720
9472-50 SHEATH TYPE TEMPERATURE PROBE
. . . . . . . . . . . for 8801*, 8802*, 8803*, 8820*, 8835-01
9398 CARRYING CASE
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3446 (K)
9315 LOGIC PROBE. . . . . . . . . . . . . . . . . . for 8852*
. . . . . . . . . . . . . . . . . for 3280, -01*, -10, -11, 3287, 3288
9473 SHEATH TYPE TEMPERATURE PROBE
9318 CONVERSION CABLE . . . . . . . . . . . . for 8940
9399 CARRYING CASE
. . . . . . . . . . . . . . . for 3441, -01, -02, 3442-02, -03, 7015*
9319 CONVERSION CABLE . . . . . . . . . . . . for 8940
. . . . . . . . . . . . . . . . . . . . for 3281, 3282, 3284, 3282-01
9473-50 SHEATH TYPE TEMPERATURE PROBE
9320 LOGIC PROBE. . . . . . . . . . . . . . . for 8800series
9400 CARRYING CASE . . . . . . . . . . . . . . . . for 3290
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3446 (K)
9320-01 LOGIC PROBE . . . . . . . . . . . for 8807, 8808
9418-10 AC ADAPTER
9474 SHEATH TYPE TEMPERATURE PROBE
9321 LOGIC PROBE. . . . . . . . . . . . . . . for 8800series
. . . . . . . . . . . . . . . . . . . . . . . . . . . for 3167, 3551, 7011
. . . . . . . . . . . . for 3441, -01, -02, 3442-02, -03, 7015* (K)
9321-01 LOGIC PROBE . . . . . . . . . . . for 8807, 8808
9418-15 AC ADAPTER
9475 SHEATH TYPE TEMPERATURE PROBE
9322 DIFFERENTIAL PROBE . . . . . . . for 8800series
. . . . . . . . . . . . . . . . for 8420/21/22, 8714*/15, 8807/08
. . . . . . . . . . . . for 3441, -01, -02, 3442-02, -03, 7015* (K)
9323 CONVERSION CABLE . . . . . . . for 9320, 9321
9420 BATTERY PACK. . . . . . . . . . . . for 7011, 8804*
9476 SURFACE TYPE TEMPERATURE PROBE
9324 POWER CORD . . . . . . . . . . . . . . . . . . for 9322
9425 CONNECTION CABLE
. . . . . . . . . . . . for 3441, -01, -02, 3442-02, -03, 7015* (K)
9325 POWER CORD . . . . . . . . . . . . . for 8940, 9322
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 9203, 9203-01
9476-50 SURFACE TYPE TEMPERATURE PROBE
9326 CONNECTION CORD. . . . . . . . for 8205, 8206
9433 DC POWER ADAPTER
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3446 (K)
9327 LOGIC PROBE. . . . . . . . . . . . . . . for 8800series
. . . . . . . . . . . . . . . . . . . . . . . . . . . for 8841, 8842, 8720
9478 SHEATH TYPE TEMPERATURE PROBE
9328 POWER CORD . . . . . . . . . . . . . . . . . . for 8855
9436 CONNECTION CABLE . . . . . . . . . . . for 3423*
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3447 (PT-100)
9329 TERMINAL UNIT
9437 CONNECTION CABLE . . . . . . for 8927(8845)*
9479 SHEATH TYPE TEMPERATURE PROBE
. . . . . . . . . . . . . . . . . . . . . for 8420-51, 8421-51, 8422-51
9438 VOLTAGE CORD . . . . . . . . . . . . . . . . for 3166*
. . . . . . . . . . . . . . . . . . . . for 3447 (PT-100 with switch)
9330-01 WAVE PROCESSOR . . . . . . . . . . . for 8835
9438-02 VOLTAGE CORD. . . . . . . . . . . . . . . for 3196
9496 NETWORK A . . . . . . . . . . . . . . . . . . . for 3155*
9331-01 WAVE PROCESSOR . . . . . . . . . . for 8806*
9438-03 VOLTAGE CORD. . . . . . . . . . . . . . . for 3169
9497 NETWORK B . . . . . . . . . . . . . . . . . . . for 3155*
9332 WAVE COMMUNICATOR . . . . . . . . . . for 8808
9439 DC POWER ADAPTER . . . . . . . . . . . . for 8835
9498 NETWORK C . . . . . . . . . . . . . . . . . . . for 3155*
9333 LAN COMMUNICATOR . . . . . . . for 8800series
9440 CONNECTION CABLE . . . . . . . . . . . for 3166*
9499 NETWORK D . . . . . . . . . . . . . . . . . . . for 3155*
9334 LOGGER COMMUNICATOR . . . . . . . for 8420
9441 CONNECTION CABLE . . . . for 3166*, 3169-01
9517 GP-IB INTERFACE . . . . . . . . . for 3186, 3227*
9335 WAVE PROCESSOR. . . . . . . . . . for 8800series
9442 PRINTER . . . . . . . . . . . . . . . . . . . . for 3443, 44
9518-01 GP-IB INTERFACE
9336 WIREMAP TERMINATOR . . . . . . . . . . for 3660
9443-02 AC ADAPTER
. . . . . . . . . . . . . . . . . . . . . . . . . . for 3511, 3531*, 3532
9337 DIRECTION TERMINATOR. . . . . . . . . for 3660
. . . . . . . . . . . . . for 3166*, 3330-02*, 3332, 3511, 3154
9518-02 GP-IB INTERFASE
9338 CARRYING CASE . . . . . . . . . . . . . . . . for 3143
9443-03 AC ADAPTER
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3157, 3157-01
9339 CARRYING CASE . . . . . . . for 3196 (soft case)
. . . . . . . . . . . . . for 3166*, 3330-02*, 3332, 3511, 3154
9540-01 FUNCTION UP DISK
9340 CARRYING CASE . . . . . . . for 3196 (hard case)
9444 CONNECTION CABLE
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 8835-01
9344 CARRYING CASE . . . . . . . . . . . for 8205, 8206
. . . . . . . . . . . for 3154, 3166*, 3196, 3332, 9442, 3511
9549 FUNCTION UP DISK(POWER MONITOR)
9345 CARRYING CASE . . . . . . . . . . . . . . . . for 3285
9445-02 AC ADAPTER . . . . . . . . . for USA, CANADA
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 8855
9347 CARRYING CASE . . . . . . . . . . . . . . . for 8220*
9445-03 AC ADAPTER . . . . . . . . . . . . . . . . . . . for EU
9555 SENSOR UNIT . . . . . . . . . for 9270, 9271, 9272
9349 CARRYING CASE . . . . . . . . . . . . . . . . for 8842
9446 CONNECTION CABLE
9557 RS-232C CARD
9350 CARRYING CASE . . . . . . . . . . . . . . . . for 3660
. . . . . . . . . . . . . . . . . . . . . . . . . . for 3522, 3532, 3330*
. . . . . . . . . for 8826, 8835*, 8835-01, 8841, 8842, 8720
9351 CARRYING CASE . . for 3127*, 3127-11, 3100*
9447 BATTERY PACK. . . . . . . . for 8807, 8808, 8420
9558 GP-IB CARD
9355 CARRYING CASE . . . . for 3265*, 3267*, 3286
9451 TEMPERATURE PROBE. . . . . . . . . . . for 3540
. . . . . . . . . for 8826, 8835*, 8835-01, 8841, 8842, 8720
9359 CARRYING CASE . . . . . . . . . . . . . . . for 3117*
9452 CLIP TYPE LEAD. . . . . . . . . . . . for 3560, 3540
9574 INPUT CORD . . . . . . . . . . . . . . for 8944(8851)*
9360 CARRYING CASE . . . . for 3165* (9270~9272)
9453 FOUR TERMINAL LEAD . . . . . . for 3560, 3540
9588 GP-IB INTERFACE
9363 CARRYING CASE . . . . . . . . . . . . . . for 3118-11
9454 ZERO ADJUSTMENT BOARD
. . . . . . . . . . . . for 3227*, 3167, 3187, 3330*, 3330-02*
Note: * marked products are discontinued models.
9631-21 TEMPERATURE SENSOR (9631-01,10m)
9680 HUMIDITY SENSOR . . . . . . . . . . for 3641 (1m)
9593-01 RS-232C INTERFACE
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3641 (10 m)
9680-01 HUMIDITY SENSOR. . . . . . . . for 3641 (5m)
. . . . . . . . . . . . . . . . . . . . . . . . . . for 3522, 3531*, 3532
9631-24 TEMPERATURE SENSOR (9631-04,10m)
9680-02 HUMIDITY SENSOR. . . . . . . for 3641 (10m)
9593-02 RS-232C INTERFACE . . . for 3157, 3157-01
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3641 (10 m)
9681 HUMIDITY SENSOR . . . . . . . . . . . . . for 8420s
9598 MO UNIT. . . . . . . . . . . . . . . . . . . . . . . . for 8826
9632 CONNECTION CABLE
9685 CARRYING CASE . . . . . . . . . . . . . . . . for 3246
9599 MEMORY BOARD. . . . . . . for 8826 (48Mword)
. . . . . . . . . . . for 3634, 3635-04~06, 3635-24~26, 3636
9686 CARRYING CASE . . . . . . . . . . . . . . . . for 3156
9600 AC/DC DIRECT INPUT UNIT . . . . . . . for 3193
9633 CONNECTION CABLE . . . . . . . . for 3634,3635
9691 CLAMP ON AC/DC SENSOR
9601 AC DIRECT INPUT UNIT . . . . . . . . . . for 3193
9634 CONNECTION CABLE . . . for 3630, 3635-01,02
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3290 AC100A
9602 AC/DC CLAMP INPUT UNIT . . . . . . . for 3193
9635 VOLTAGE CORD . . . . . . . . . . . . . . . . . for 3286
9692 CLAMP ON AC/DC SENSOR
9603 EXTERNAL SIGNAL INPUT UNIT . . . for 3193
9635-01 VOLTAGE CORD. . . . . . . . . . . . . . . for 3268
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3290 AC200A
9603-01 EXTERNAL SIGNAL INPUT UNIT
9636 RS-232C CABLE. . . . . . . . . . . . . . . . . for 3286
9693 CLAMP ON AC/DC SENSOR
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3194
9636-01 RS-232C PACKAGE . . . . . . . . . . . for 3286
. . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3290 AC2000A
9604 PRINTER UNIT. . . . . . . . . . . . . . . . . . . for 3193
9637 RS-232C CABLE (9pin-9pin/1.8m)
9694 CLAMP ON SENSOR
9605 HARMONIC/FLICKER MEASURMENTS UNIT
. . . . . . . . . . . . . . . . . . . . . . . . for 3154, 3630S, 3911-20
. . . . . . . . . . . . . . . . . . . . . . . . . . for 3169, 3196 AC5A
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3193
9638 RS-232C CABLE (9pin-25pin/1.8m)
9699 SMD TEST FIXTURE . . . . . . . . . . . . . . for 3511
9605-01 HARMONIC MEASURMENTS UNIT
. . . . . . . . . . . . . . . . . . . . . . . . for 3154, 3630S, 3911-20
9700-10 HEAD AMP UNIT . . . . . . . . . . . . . . for 3535
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3194
9639 CONNECTION CABLE . . . . . . . . . . . . for 3637
9713-01 CAN CABLE . . . . . . . . . . . . . . . . . . for 8910
9607 MO UNIT . . . . . . . . . . . . . . . . . . for 8841, 8842
9641 CONNECTION CABLE . . . . . . . . . . . . for 8420
9713-02 CAN CABLE . . . . . . . . . . . . . . . . . . for 8910
9608 MEMORY BOARD(24M-WORD)
9642 LAN CABLE . . . . . . . . . . . . . . . . . . . . . for 8420
9714-01 LOGIC CABLE . . . . . . . . . . . . . . . . for 8910
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 8841, 8842
9643 CHARGE STAND. . . . . . . . . . . . . . . . . for 8420
9714-02 LOGIC CABLE . . . . . . . . . . . . . . . . for 8910
9612 RS-232C CABLE
9644 SCSI INTERFACE . . . . . . . . . . . . . . . . for 8855
9720 CARRYING CASE . . . . . . . . . . . . . . . . for 3169
. . . . . . . . . . for DIN 9pin-Dsub 9pin 8807, 8808, 8420
9645 MEMORY BOARD. . . . . . . . . . . . . . . . for 8855
9721 RS-232C CABLE. . . . . . . . . . . . . . . . . for 3169
9613 REMOTE CONTROL BOX(SINGLE)
9645-01 MEMORY BOARD . . . . . . . . . . . . . for 8855
9726 PC CARD 128M
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3158
9646 MO UNIT . . . . . . . . . . . . . . . . . . . . . . . for 8855
. . . . . . . . . . . . . . . . . . . . . . . . . . for 8800S, 3169, 3196
9614 REMOTE CONTROL BOX(DUAL)
9648 CARRYING CASE
9727 PC CARD 256M
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3158
. . . . . . . . . . . for 8807, 8808, 8420, 8421, 8714*, 8715
. . . . . . . . . . . . . . . . . . . . . . . . . . for 8800S, 3169, 3196
9615 H.V.TEST LEAD . . . . . . . . . . . . . . . . . . for 3158
9649 PROTECTIVE CASE
9728 PC CARD 512M
9615-01 H.V.TEST LEAD
. . . . . . . . . . . . . . . . . . . . . . for 8807, 8808, 8420, 8421
. . . . . . . . . . . . . . . . . . . . . . . . . . for 8800S, 3169, 3196
. . . . . . . . . . . . . . . . . . . . . for 3931 Red (High Voltage)
9650 CLAMP ON SENSOR
9730 CARRYING CASE
9615-03 H.V. TEST LEAD
. . . . . . . . . . . . . . . . . . . . for 100A 8205, 8206, 3636-20
. . . . . . . . . . . . . . . . . . . . . . . . . . . for 3661, 3662, 3663
. . . . . . . . . . . . . . . . . . . . . . . . for 3931 Black (Return)
9651 CLAMP ON SENSOR
9731 FC CONNECTOR ADAPTER . . . . . . . for 3661
9616 WARNING LAMP. . . . . . . . . . . . . . . . . for 3158
. . . . . . . . . . . . . . . . . . for 500A 8205*, 8206*, 3636-20
9732 SC CONNECTOR ADAPTER . . . . . . . for 3661
9617 CLIP ON BASE
9652 FIXED STAND
9733 FC CONNECTOR ADAPTER
. . . . . . . . . . . . . . . . . for 3501, 3801, 3802, 3804, 3805
. . . . . . . . . . . . . . . . . . . . . . for 8807, 8808, 8420, 8421
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3662, 3663
9618 CLIP-TYPE LEAD
9652-01 FIXED STAND
9734 SC CONNECTOR ADAPTER
. . . . . . . . . . . . . . . . . for 3501, 3801, 3802, 3804, 3805
. . . . . . . . . . . for 8420-50, 8421-50, 8422-50, 8807, 8808
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3662, 3663
9623 POWER ANALYZER . . . . . . . . . . . . . . for 3193
9653 HUMIDITY SENSOR . . . . . . . . . for 8420, 8421
9735 FC-FC OPTICAL FIBER CABLE
9624 PQA-HiVIEW . . . . . . . . . . . . . . . . . . . . for 3196
9657 CLAMP ON LEAK SENSOR. . . . . . . . for 3638
. . . . . . . . . . . . . . . . . . . . . . . . . . . for 3661, 3662, 3663
9624-10 PQA-HiVIEW PRO . . . . . . . . . . . . . for 3196
9657-10 CLAMP ON LEAK SENSOR
9736 SC-SC OPTICAL FIBER CABLE
9625 POWER MEASUREMENT SUPPORT SOFTWARE
. . . . . . . . . . . . . . . . . . . . . . . . . . . for 8808series (BNC)
. . . . . . . . . . . . . . . . . . . . . . . . . . . for 3661, 3662, 3663
. . . . . . . . . . . . . . . . . . . . . . . . . . for 3166*, 3168, 3169
9658 CLAMP ON LEAK SENSOR. . . . . . . . for 3638
9737 SC-FC OPTICAL FIBER CABLE
9626 PC CARD 32M . . . . . . . for 8800S, 3169, 3196
9660 CLAMP ON SENSOR . . . . . . for 3196 AC100A
. . . . . . . . . . . . . . . . . . . . . . . . . . . for 3661, 3662, 3663
9627 PC CARD 64M . . . . . . . for 8800S, 3169, 3196
9661 CLAMP ON SENSOR . . . . . . for 3196 AC500A
9738 OPTICAL CONNECTOR CLEANER
9628 LAN CABLE . . . . . . . . . . . . . . . . . . . . . for 3660
9662 LUX SENSOR . . . . . . . . . . . . . . . . . . . for 3640
. . . . . . . . . . . . . . . . . . . . . . . . . . . for 3661, 3662, 3663
9629 CONNECTION CABLE . . . . . . . . . . . . for 3639
9663 HD UNIT . . . . . . . . . . . . . . . . . . . . . . . . for 8855
9739 SPARE CLEANER . . . . . . for 3661, 3662, 3663
9631-01 TEMPERATURE SENSOR
9665 10:1PROBE . . . . . . . . . . . . . . . . . . . . . for 8855
CT-101A LINE SPLITTER. . . . . . . . . . for 3127-11 etc.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3641 (1m)
9666 100:1PROBE . . . . . . . . . . . . . . . . . . . . for 8855
220H PAPER WINDER . . . . . . . . . . . TEPTOM-220H
9631-02 TEMPERATURE SENSOR
9667 FLEXIBLE CLAMP ON SENSOR
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3641 (1m)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3196 AC5000A
9631-03 TEMPERATURE SENSOR
9668 CLAMP ON SENSOR
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3641 (1m)
. . . . . . . . . . . . . . . . . . . . for 8205-10/8206-10 AC1000A
9631-04 TEMPERATURE SENSOR
9669 CLAMP ON SENSOR
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3641 (1m)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3196 AC1000A
9631-05 TEMPERATURE SENSOR
9670 PRINTER . . . . . . . . . . . . . . . . for 3196, 3446/47
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3641 (30 mm)
9671 AC ADAPTER . . . . . . . . . . . . . . . . . . . for 9670
9631-11 TEMPERATURE SENSOR (9631-01,5m)
9674 RS-232C PACKAGE . . . . . . . . . . . . for 3446/47
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3641 (5 m)
9677 SMD TEST FIXTURE . . . . . . . . . . . . . . for 3511
9631-14 TEMPERATURE SENSOR (9631-04,5m)
9678 CONNECTION CABLE . . . . . . . . . . . . for 3535
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 3641 (5 m)
9679 CONNECTION CABLE . . . . . . . . . . . . for 3503
ALLIGATOR CLIPs
. . . . . 9170 or similar devices (used with Test Lead, inserted)
Options & Peripherals
9589 PRINTER INTERFACE. . . . . . . . . . . . for 3227*
60
HIOKI - Offering Top Quality Products and Services
HIOKI E.E. CORPORATION TRACEABILITY CHART
National
Standards
Comparative standards
M
a
k
e
r
N
I
S
Intermediate standards
A
AC/DC VOLTAGE
CURRENT STANDARD
J
E
M
I
C
T
management system that reaches to the national standards
✻AC/DC current
0 to 100A
are indispensable.
✻High voltage
1mΩ to 30kV
CURRNET
TRANSFORMER
RESISTANCE
CALIBRATOR
HIGH VOLTAGE
DIVIDER
STANDARD
RESISTOR
POTENTIAL
TRANSFORMER
S
✻AC/DC voltage
0 to 1000 V
MULTIMETER
A
I
Accuracy can be regarded as the heart of a measuring
accountability in the form of a coherent and comprehensive
MULTIMETER
THERMAL TRANSFER
STANDARD
STANDARD SHUNT
Q
Calibrated
parameter
instrument. To maintain accuracy, traceability and
STANDARD
CALIBRATOR
T
J
Applied standards
STANDARD
RESISTOR
ELECTROMETER/
SOURCE
STANDARD VARIABLE
RESISTOR
HIGH RESISTANCE
STANDARDS
WHEATSTONE
BRIDGE
✻DC resistance
1mΩ to 1MΩ
Traceability allows us to manage and maintain instrument
accuracy characteristics that are tied to recognized national
and international standards. How they are managed and
maintained are dependent on the measurement facilities
✻High resistance
106 to 10Ω10
that offer accuracy testing at the various levels, skilled
technicians, as well as a strong link between national
AC GENERATOR
✻AC power
T
E
L
E
C
C
R
T
STANDARD POWER
METER
PHASE METER
PHASE METER
STANDARD SELFINDUCTOR
STANDARD SELFINDUCTOR
✻ Inductance
The HIOKI Traceability System as indicated by the chart on
STANDARD
CAPACITOR
STANDARD
CAPACITOR
✻Capacitance
the left is strictly managed by accounting for each individual
STANDARD
CAPACITOR
✻AC resistance
instrument - from reference instruments to field equipment
LCR METER
NIST =
National Institute of
Standards and Technology
ILLUMINANCE METER
CALIBRATOR
STANDARD LAMP
JQA =
Japan Quality Assurance
Organizaton
UNIVERSAL
COUNTER
JEMIC =
Japan Electric Meters
Inspection Corporation
CRL =
Communications Research
Laboratory
TELEC =
Telecom Engineering
Center
SYNTHESIZER
UNIVERSAL
COUNTER
FUNCTION
GENERATOR
AIST =
National Institute of
Advanced Industrial
Science and Technology
STANDARD DIGITAL
THERMOMETER
measuring instruments, and basic measuring instruments.
✻Phase
✻Illuminance
✻Frequency
we will strive to meet new demands by not only providing
the appropriate measuring instruments for our users,
DIGITAL
THERMOMETER
STANDARD GLASS
THERMOMETER
- and their constant accuracy. As scientific techniques and
manufacturing technology continue to expand and develop,
✻Time
but also enhancing the accuracy in our test instruments
✻Temperature
TEMPERATURE
CALIBRATOR
and maintaining our Traceability System so that they are
constantly on par with global standards.
DEW POINT
MONITOR
✻Humidity
GAGE BLOCK
MULTIMETER
CALIPER CHECKER
SLIDE CALIPERS
✻Length
Note:Only the primary standards are indicated above. For details, please refer each product's TRACEABILITY CHART.
Please also note that the naming of the standards indicated in this chart may differ from the naming used in
each product's TRACEABILITY CHART.
HIOKI's Calibration System
In order to provide dependable and quality-assured products, HIOKI
has acquired the international standard ISO/IEC17025 certification for
calibration, which allows us to meet a wide variety of calibration needs.
By regularly calibrating HIOKI instruments using reference calibrating
equipment traceable to national standards while complying with the
reference equipment organizational chart, customers are guaranteed
complete accuracy. After purchase, it is highly recommended that
customers regularly re-calibrate their HIOKI instruments to maintain
their accuracy. Depending on your needs, calibration and adjustment
can be conducted at HIOKI in one of 4 ways as illustrated on the right.
61
standards, manufacturer reference equipment, field
STANDARD WATT
CONVERTER
About our Company
Established in 1935, HIOKI E. E. CORPORATION has grown to be a
leading developer and manufacturer of advanced test and measurement
technologies for use both in the field and leading edge facilities around
the world. Our goal is simple: contribute to the advancement of society,
while making sure the natural environment is not compromised. As
a reliable producer and member of society, we pledge to continue to
actively contribute to the cultural and educational development of the
local community through activities such as greening efforts, scholarship
programs and sponsoring children's sports teams. With the support of our
customers and worldwide network, we are confident that our values and
beliefs, and products and services, will be brought forth through the 21st
century and beyond.
Internet website
Corporate History
1935
1945
1946
1952
1965
1975
1983
1990
1991
1992
1993
1997
1998
2001
2003
2005
HIOKI starts manufacturing electrical measuring instruments in Tokyo
Move to Nagano Prefecture due to war
Tester No.1 put to market
HIOKI E.E. CORPORATION established
Designated as the manufacturer of MULTITESTER (MIL Standard) for the
U.S. Far East Air Forces
Mass production of VU instruments for recording level adjustments to tape
recorders
Independent development and sale of instruments with internal magnetic
taut bands
Multiple awards received for innovative clamp-style instruments
Move to HIOKI Forest Hills
Registered on the over-the-counter market
Awarded the Afforestation Center Presidential Award for positively
promoting afforestation
ISO9001 certified
ISO14001 certified
HIOKI USA CORPORATION established
HIOKI Shanghai Representative Office established
Listed on the Second Section of the Tokyo Stock Exchange
Listed on the First Section of the Tokyo Stock Exchange
"Solution Fair" - 70th Anniversary Celebration
www.hioki.co.jp
62
ing
t
a
7
r
0
b
e
Y
l
ear
e
C
s
of
Te
c
hn
loo
g
i
c
a
l
ecnellecxE
HEAD OFFICE:
HIOKI USA CORPORATION:
81 Koizumi, Ueda, Nagano
386-1192, Japan
TEL+81-268-28-0562
FAX+81-268-28-0568
E-mail:os-com@hioki.co.jp
6 Corporate Drive
Cranbury, NJ 08512 USA
TEL+1-609-409-9109
FAX+1-609-409-9108
E-mail: hioki@hiokiusa.com
1989
2000
Shanghai Representative Office:
1704,Shanghai Times Square Office
93 Huaihai Zhong Road, Shanghai, 200021, P.R.China
TEL +86-21-6391-0090/0092
FAX +86-21-6391-0360
E-mail: info@hioki.cn
DISTRIBUTED BY
All information correct as of April 10, 2005. All specifications are subject to change without notice.
This brochure has been printed on recycled paper
with special ink blended with recycled soy oil.
0001E 35-54E-25K Printed in Japan