Features
• Rugged, handheld, lightweight
• Designed for field use
• Certify multimode fiber links per TIA/EIA
standards
• Wave ID - automatic wavelength identification and switching
• Dual Wave ID, single Wave ID, and CW
• Large LCD with Backlight
• Power measurements in dBm or µW;
insertion loss in dB
• Reference power level storage
• Up to 500 test records per wavelength
storage and download
• USB port for download of stored records
• AA alkaline, optional rechargeable NiMH
battery pack or AC adapter
• Low battery indicator
• Free Windows® compatible software to
view, print, and archive test records
• Free 50 µm and 62.5 µm mandrels
• Cost-effective, easy to use
• N.I.S.T traceable
The MLP 5-2 test kit combines the OPM 5-2D optical power meter and OLS 1-Dual LED
light source and is ideally suited for testing multimode fiber optic networks.
The OLS 1-Dual features 850 nm and 1300 nm LED output from a single output port
and is easy to operate with only a power button and a wavelength select button. The LED
output is stabilized to ensure accurate test results per current TIA/EIA requirements. This
light source offers 3 modes of operation: Dual wavelengths sending ID, single wavelength
sending ID, and CW. [Active Output], [Battery], and [External Power] indicators identify the
currently enabled operating mode, battery charge status, and external power presence.
The output port is equipped with a removable SC (FC & ST available) adapter to allow the
output connector to be inspected and cleaned.
The OPM 5-2D offers Wave ID, automatic wavelength identification and switching, when
used with the OLS 1-Dual light source. The OPM 5-2D stores optical references for each
calibrated wavelength. An easy to read Dual Wavelength LCD display with Backlight shows
measured power [dBm or µW] or insertion loss [dB], calibrated wavelengths, tone signal
[Hz], wavelength ID, and the battery charge status. Up to 500 records per wavelength of
power or insertion loss measurements may be stored in internal non-volatile memory.
Using the supplied Windows® compatible software and USB connection, test records may
be transferred to a PC for storage, display, analysis, and printing.
In addition to being powered by two AA alkaline, the OPM 5-2D or OLS 1-Dual can be
powered by optional AC adapter and/or rechargeable NiMH battery pack.
The OPM 5-2D optical input port accepts Noyes thread-on style adapter caps. Adapter
caps are required for operation and must be ordered separately. The OPM 5-2D and
OLS 1-Dual are fully N.I.S.T. traceable.
Ordering Information
Model
Includes
MLP 5-2
OLS 1-Dual optical light source, OPM5-2D optical power meter, AA batteries, protective rubber
boots, adapter cap, USB cable, Windows® compatible software and user's guide, 50 and 62.5 µm
mandrels, user's guide, and carry case.
Test jumpers and connector adapters are required for operation (purchased separately).
Test jumpers with a variety of connector styles and fiber types and adapter caps for most
common connectors may be purchased from AFL Telecommunications.
continued on the next page
Test & Inspection
© 2002, AFL Telecommunications, all rights reserved. MLP5-30-2000 Revision A, 6.28.05
Specifications are subject to change without notice.
1
MLP 5-2 multimode test kit with Wave ID and data storage
OLS 1-Dual specifications
Optical
OLS 1-Dual (single port)
850 ±30 nm
Output wavelength
Spectral width (typ) (FWHM)
1300 +50/-10 nm
40 nm
120 nm
>-20 dBm*
Output power
62.5 µm**
Fiber size
Output connector
SC (FC & ST available)
Emitter classification
Class 1 (IEC 60825 - 1)
Stability
General
± 0.1 dB over 8 hours (after 5 min. warm-up)
Power
2 x AA batteries, optional NiMH or AC adapter
Battery life (2 x AA)
Typical 30 hours
Minimum 20 hours
Operating temperature
-10 to 50°C, 90% RH (non-condensing)
Storage temperature
-30 to 60°C, 90% RH (non-condensing)
Size (H x W x D)
5.5 x 3.2 x 1.5 in (14.0 x 8.1 x 3.8 cm)
0.65 lb (.29 kg)
Weight
* Output power will be approximately 3 dB less if a 50 µm mandrel-wrapped jumper is used instead of
a 62.5 µm mandrel-wrapped jumper.
** May be used to test 50 or 62.5 µm fiber with supplied mandrels.
All specifications at 25°C
OPM 5-2D specifications
Optical
OPM 5-2D
Calibrated wavelengths
850, 1300, 1310,1550 nm
Detector type
Germanium (Ge)
Measurement range
+6 to -60 dBm
Tone detect range
+6 to -50 dBm
+6 to -45 dBm for 850nm
Wavelength ID range ()
+6 to -50 dBm
+6 to -45 for 850nm
Accuracy*
± 0.25 dB
Resolution
0.01 dB
Measurement units
General
dB, dBm, µW
Power
2 x AA batteries, optional NiMH or AC adapter
Battery life (2 x AA)
300 hours with backlight [Off]
20 hours with backlight [On]
Operating temperature
-10 to 50°C, 90% RH (non-condensing)
Storage temperature
-30 to 60°C, 90% RH (non-condensing)
Size (H x W x D)
5.5 x 3.2 x 1.5 in (14.0 x 8.1 x 3.8 cm)
Weight
0.58 lb (0.26 kg)
* Accuracy measured at 25ºC and -10 dBm per N.I.S.T. standards.
All specifications at 25°C
Test & Inspection
© 2002, AFL Telecommunications, all rights reserved. MLP5-30-2000 Revision A, 6.28.05
Specifications are subject to change without notice.
2
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertisement