kanomax usa, inc.
KANOMAX USA, INC.
219 US Hwy. 206
PO Box 372
Andover, NJ 07821
Tel: 1-800-247-8887, 1-973-786-6386 (from outside of US)
Fax: 1-973-786-7586
URL: www.kanomax-usa.com
2nd Edition Rev.2
Kanomax Group
Since our inception more than 50 years ago, Kanomax has been the most promising manufacturer of a broad range
of precision measuring instruments for fluid mechanics research, environmental, aerosol research, particles
measurement, and customized system applications.
As a company that prides itself in technology, product quality,
and service, we have been enjoying an unsurpassed reputation in the Japanese industrial and academic fields.
Japanese customer demand for high quality has driven us to produce the finest manufactured products in the
industry.
We are pleased to introduce our quality products to the global market. To further benefit our customers
worldwide, we have launched our global operations aiming to introduce our dependable technology and to provide
our wide range of measuring solutions and services.
Kanomax USA, Inc. (NJ)
Kanomax Holdings, Inc. (NY)
Kanomax Japan, Inc.
Shenyang Kanomax
Instrument Co., Ltd. (China)
Kanomax FEASibility
Kanomax makes your idea feasible and realizes your needs.
Environmental Measurement Instruments
Fluid Measurement Instruments
Having established worldwide brand recognition and product
loyalty, Kanomax Fluid Measurement Instruments have
become the measurement standard in Japan.
Our Fluid
Velocity Systems have wide-ranging industrial applications to
include Elimination of Locomotive Wheel Noise, Research to
Increasing Energy Efficiency, Architectural Research and
Development, Civil Engineering, Chemistry, and Medical
Science Research to name a few.
has
developed
demand for improved monitoring and control of air temperature,
humidity, flow and particulates.
Kanomax produces a number of
Air Measurement and Control Instruments currently utilized by
industry professionals. These devices are vital tools for
maintaining Constant Air Quality and Comfort in critical locations
that include public and private offices, factories, and medical
facilities. Kanomax Anemometers currently enjoy a 67.9% market
share among top 10 Japanese subcontractors.
Aerosol Measurement Instruments
Kanomax
Ever increasing public awareness for air quality is driving the
Particle
System Application
Measurement Kanomax
has
developed
custom
Wind
Tunnel,
Technologies to address a variety of applications from Environmental Test and Performance Test systems for a
Semiconductor Production Facilities requiring rigorous air variety of areas that include Aviation, Environmental
cleanliness to general office work environments.
Our Assessment, and Automotive. As an example, during the
Particle Measurement Technology is generally divided into development of an intercontinental rapid transit scramjet
three
areas,
Environmental
Comfort
Monitoring, engine, Kanomax tested flame stabilization, air inlet shape,
Cleanroom Environment Monitoring, and Aerosol Research and many other effects using Kanomax supersonic wind
Instruments.
tunnels.
Customer Services
Kanomax fully understands service to be an essential part of the total solution provided to our valued customers.
Having already established a worldwide service network, we continuously strive to improve our support services.
NIST (National Institute of Standards and Technology)
Traceable Calibration Services
Our Calibration Laboratory in New Jersey maintains the longest and most
accurate wind tunnel of its kind.
Kanomax provides the highest quality of service
available with a quick turnaround time. Our service specialists are well trained
and will calibrate your instruments to the highest standards. We recommend
that all instruments be calibrated on an annual basis.
Kanomax Global Calibration Facility
Kanomax’s global (Japan and US) calibration facilities are directly traceable to the national standards and ensure the
highest precision measurements for our valued customers. We are fully committed to providing the best calibration
services possible utilizing our global facilities.
Kanomax Anemometer Calibration Facility
Calibration Facility
Temperature
Variable Wind
Tunnel
Low Velocity
Wind Tunnel
High Velocity
Wind Tunnel
High Temperature
Wind Tunnel
Open Jet Wind
Tunnel
Humidity Calibrator
Pressure
Generator
Type
GÖTTINGEN
(closed-circuit)
EIFFEL
(open-circuit)
GÖTTINGEN
(closed-circuit)
GÖTTINGEN
(closed-circuit)
Centrifugal fan
a) Two temperature type
b) shunt type
Pump type
Specifications
59 to 9840fpm
(0.3 to 50m/s)
41 to 176ºF
(5 to 80ºC)
10 to 492fpm
(0.05 to 2.5m/s)
590 to 9840fpm
(3 to 50m/s)
Room temperature
to 752ºF (400ºC)
20 to 9840fpm
(0.1 to 50m/s)
50 to 6000fpm
(0.25 to
30.5m/s)
a) 3 to 100%, 41 to
104ºF (5 to 40ºC)
b) 0 to 100%, 41 to
140ºF (5 to 60ºC)
±0 to 10kPA
Applications
Temperature
compensating
calibration
Low velocity
range air velocity
calibration
High velocity
range air velocity
calibration
High temperature
range air velocity
calibration
Anemometers
and Pitot tubes
calibration
Humidity calibration
Pressure
calibration
Standard Particle Generator
Open Jet Wind Tunnel
Indoor
Environment
Airconditioning
Equipment
Hightemperature
Environment
6815
6813
6812
1570
1560
○ ○ ○
Air-condition and environmental
measurement in buildings and factories
○ ○ ○ ○
○ ○
○
○ ○ ○
Air-conditioning capacity test and
maintenance check
○ ○ ○ ○
○ ○
○
○ ○
Air velocity measurement in cleanroom
○ ○ ○ ○ ○ ○ ○ ○
Performance check for HEPA filter, etc.
○ ○ ○ ○ ○
Stack gas and combustion air measurement
○
for incinerator, electric furnace, boiler, etc.
Hot air measurement for drying air,
sterilization, etc
Product performance check for cooling
efficiency, drying efficiency, etc.
○
○
○
○ ○
○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
○
Air velocity control for blowers
Multipoint
measurement
1550
○ ○ ○ ○ ○ ○ ○ ○ ○
(e.g. inside PCs) for cooling efficiency, etc.
Automotive
6312
6162
6113
A543
A542
A541
A533
A531
Air environmental measurement in
residence and office buildings
Air velocity measurement in tight quarters
Inspection,
Control
A031
Application
A004
Area
A003
Anemometers Selection Guide
Automobile velocity measurement
○ ○ ○ ○
○ ○
Air velocity measurement for engine
compartment and vehicle interior
○ ○ ○ ○
○ ○
Air velocity distribution measurement in
indoor and inside ducts
○ ○ ○ ○
○ ○
Air velocity monitoring in cleanroom
and indoor
○
○ ○ ○
○ ○ ○
○ ○ ○ ○
Solution Technologies for Automotive R&D
R&D competition is increasingly keen in the modern day automobile industry.
Kanomax continually monitors trends
and user needs of collective automobile technologies.
Kanomax designs and produces testing equipment utilizing state of the art technologies. We offer customized
engineering systems and products, applying our many competitive technologies. These include fluid mechanics
research instruments, particle measurement devices, optical measurement systems and control technology for the
automobile, aeronautics, shipbuilding, and architectural industries. Kanomax optimizes system applications to meet
each customer’s individual needs, drawing on our solid product knowledge, years of experience and many successful
engagements.
CONTENTS
Velocity, Air Temperature,
1 Air
Relative Humidity, Differential Pressure
1
2 Thermohygrometer
13
3 IAQ Monitor, Gas Monitor
14
4 Particle / Dust Monitor
18
5 Cleanroom Facility Monitoring
23
6 Aerosol Research Instrumentation
27
7
Fluid Mechanics Research Instrumentation
29
8 R & D / Test Instrument and System
30
Air Velocity, Air Temperature,
Relative Humidity,
Differential Pressure
Anemomaster Model A004
Palm-Size Hot-Wire Anemometer
NIST Traceable
HVAC Testing
IAQ Investigation
Industrial Hygiene
Cleanrooms
Features:
• Probe Compatibility feature allows utilization of a spare probe
• Detachable probe
• Compact, light-weight, and affordable
• Simple operation
• Wide ranges of measurement
• Built-in temperature compensation circuit
Benefits:
• Palm-size and feather-weight main body is hardly noticeable
• Accuracy of +/- 3% is maintained to suit your needs
• Soft keys and large display make it a breeze to conduct measurement
• Temperature compensation maintains accuracy at any temperature in
the range
Specifications
Model
A004
Range
Accuracy
Air Velocity
Response Time
Resolution
20 to 3940 fpm (0.10 to 20.0m/s)
+/-3% of reading or +/-3 fpm (+/-0.015m/s) whichever is greater
Less than 1second at 196 fpm (1m/sec), 90% response
0 to 9.99 m/sec: 0.01 m/sec (minimum)
10.0 to 20.0 m/sec: 0.1 m/sec
Range
Air
Accuracy
Temperature
Response Time
32 - 122ºF (0-50ºC)
+/-0.2ºF (1ºC)
Less than 30 seconds at 196 fpm (1 m/sec), 90% response
Resolution
1ºF (0.1ºC)
4 x 1.5V AA cells Mn battery, alkaline battery, or Ni-Cd battery
Power Supply
(An appropriate charger to be used for the Ni-Cd battery)
Battery Life
Approx. 4 hrs. Continuous at air velocity 196 fpm (1 m/s) with Mn batteries
Operating
Main Unit
41 to 104ºF (5 to 40ºC)
Environment
Probe
32 to 122ºF (0 to 50ºC)
Storage Environment
14-122ºF (-10 to 50ºC)
Probe: approx. 0.24" or 0.40" (6 or 10 mm) in diameter x 7.9" (200 mm) in length
Dimensions
Probe cable: 0.13" (3.3 mm) in diameter x 59" (1500 mm) in length
Main body: 2.4" (60 mm) x 4.7" (120 mm) x 1.2" (30 mm)
Weight
Approx. 0.4lbs (180g) including batteries
Standard Kit
Optional Accessories
Operation Manual, AA Batteries
Spare Probe, Extension Rod (Telescopic 6.5" - 35.8" / 166mm to 909mm)
1
Air Velocity, Air Temperature,
Relative Humidity,
Differential Pressure
Anemomaster Model A031 Series
NIST Traceable
HVAC Testing
IAQ Investigations
Filter Face Velocity Measurements
Cleanroom Studies
Industrial Hygiene
Features:
• Single probe capable of measuring air velocity and temperature simultaneously
• Built-in memory allows storage of measured data
• Highly- visible LCD is capable of displaying air velocity and temperature
simultaneously
• RS232 and Analog (option) outputs are available
Benefits:
• Data can be reviewed on-screen, printed, or downloaded to a computer
• Articulating probe for various applications
• Etched length increments on the telescopic probe make duct traverse measurements easier
• Telescopic probe for hard-to-reach areas and in-duct measurements
Specifications
Air Velocity
Range
Accuracy
Resolution
20-6000fpm (0.10-30.0m/s)
+/-2% of reading or +/-3fpm whichever is greater
(+/-2% of reading or +/-0.015m/s whichever is greater)
1fpm (0.01m/s from 0 to 9.99m/s, 0.1m/s from 10.0 to 30.3m/s)
3
Volumetric Flowrate
Temperature
Differential
Pressure
(Option)
Output
Range
Accuracy
Resolution
Range
Accuracy
Resolution
Digital
Analog
Power Supply
Data Storage
Probe
Accessories
Options
Straight Probe Model
A031/A041
Articulating Probe
Model A034/A044
At 20 fpm: 9031 ft3/min, at 6000 fpm: 2,709,360 ft /min
3
3
(At 0.1m/s: 2,341m /s, At 30 m/s: 702,270 m /s)
-4 to 140.0 ºF (-20 to 60.0ºC)
+/-1ºF (+/-0.3ºC)
0.1ºF (0.1ºC)
-5.00 to + 5.00kPa
+/- (3% of reading + 0.01)kPa
0.01kPa
RS 232C (Baud rate 4800, 9600, 19200 and 38400 bps)
DC 0-3 V (Select from Air Velocity, Air Temperature and Pressure)
6 x 1.5V AA batteries (AC adaptor: AC90-240V/Optional)
Timed and multipoint average calculation.
Total of 800 data for Velocity and Temperature /
Volumetric Flowrate and Temperature / Pressure
Length: 39.4 in. (100 cm)- Telescopic / Diameter of Tip: 0.236 in.
(6.0mm), Diameter of Base: 0.472 in. (12.0 mm)
Length: 39.4 in. (100 cm)- Telescopic / Diameter of Tip: 0.236 in. (6.0mm),
Diameter of Base: 0.472 in. (12.0 mm) Articulating section length: 3.543 in. (90.0 mm)
Carrying case, Operation manual, 6 x AA Batteries, RS-232C cable,
Software (for Windows), AC adaptor
Analog output, Printer, AC adaptor (for Printer)
2
Climomaster Series
Air Velocity, Air Temperature,
Relative Humidity,
Differential Pressure
NIST Traceable
Multi-Function Thermal Anemometer
Ventilation Testing
Laboratory Control
Cleanrooms
IAQ Investigation
Industrial Hygiene
Quality Control
A531
A541
A542
A533
Features:
• Probe Compatibility feature allows utilization of a spare probe
• Detachable probe allows for easy replacement, and for
compatibility with multi-function thermo anemometers
• Automatic Flow Rate Calculation function
• RS232C terminal for data logging
• Stores up to 1500 measurement data
• Differential Pressure available as an option
A543
Benefits:
• If you have more than one unit, detachable probe allows you to share the main unit or the probe.
• Gives automatic flow rate readings by registering duct sizes (up to 25 duct sizes).
• Well designed software allows you to access the data instantly.
• You can store the data or upload it to your PC via RS232C cable and software (both optional)
Specifications
Model
Air Velocity Range
Air Temperature Range
Relative Humidity Range
Differential Pressure (Option)
Accuracy
Output
Digital
Analog
Power Supply
Battery Life
Operating
Main Unit
Environment
Probe
Storage Environment
Weight
Standard Kit
Optional Accessories
A531
A541
A542
A533
A543
20 to 6000fpm (0.10 to 30.0m/s)
10 to 1000fpm (0.05 to 5.00m/s)
32.0 to 140.0ºF (0.0 to 60.0ºC)
2.0 to 98.0%RH
-2.0 to 98.0%RH
--5.00 to +5.00kPa
Velocity: +/-2% of reading or +/-3fpm (+/-0.015m/s) whichever is greater
Air Temperature: +/- 1 ºF (+/-0.5 ºC)
Relative Humidity: +/-2.0%RH from 2 to 80%RH, +/-3%RH from 80 to 98%RH
Differential Pressure: +/- (3% of reading +0.01)kPa
RS-232C (Baud Rate 4800, 9600, 19200 and 39400bps)
DC0-1V (Select from Air Velocity, Air Temperature, Relative Humidity and Pressure)
6 x 1.5V AA Batteries, AC Adapter (Optional): AC100-240V
Approx. 10hrs. Continuous at 984fpm (5m/s), 68ºF (20ºC) with alkaline batteries
41 to 104ºF (5 to 40ºC)
32 to 140ºF (0 to 60ºC)
41-104ºF (5 to 40ºC)
Approx. 0.9lbs (400g)
Carrying Case, Operation Manual, AA Batteries, Probe Cable
Spare Probe, Analog Output, Pressure Sensor, Extension Rod, Printer, Printer Cable,
Communication Cable, Software (for Windows), AC Adapter: AC100-240V 50/60Hz
Selection Guide
Model
A531
A541
A542
A533
A543
Measuring Range
20 - 6,000 fpm
20 - 6,000 fpm
20 - 6,000 fpm
10 - 1,000 fpm
10 - 1,000 fpm
Probe Type
Rod
Rod
Needle
Spherical
Spherical
Directivity
Mono
Mono
Omni
Omni
Omni
3
Velocity & Temp.
○
○
○
○
○
Relative Humidity
○
×
×
○
×
Air Velocity, Air Temperature
Relative Humidity, Differential Pressure
Anemomaster Model 6812, 6813, 6815
Rotating Vane Digital Anemometer
NIST Traceable
HVAC Testing
IAQ Investigations
Industrial Hygiene
Cleanroom Testing
Features:
Display switchable in m/s (MPS) or ft/min (FPM) for
air velocity and cubic ft/min (CFM) or cubic meter/hr (CMH)
for calculated air volume flow.
Very high accuracy due to use of microprocessor
Digital memory for maximum and minimum values
Average measurements over two or sixteen seconds
for air velocity
Display hold for easy reading of measurements
Long battery life
Specifications
Model
Air Velocity
Probe AP275
Probe AP100
Accuracy
Resolution
Air Volume Flow
Temperature
6812
Air Vel. Probe
Optional Temp Probe
Accuracy
Resolution
Range
Relative
Accuracy
Humidity
Resolution
Display
Operating Temperature
6813
40 to 7800 FPM (0.2 to 40 m/s)
60 to 6800 FPM (0.3 to 35 m/s)
±1.0% of Reading ±1 digit
1 FPM or 0.01 m/s
-
0.0 to 9999 CFM(ft3/min) / 0.0 to 9999 m3/hr
-22 to 212 F (-30 to 100C)
-139 to 392 F (-95 to 200C)
±0.3℃ ±0.2% of Reading in ℃
-
Dimensions
Standard Accessories
0.1 F or 0.1℃ (1 F below -99.9 F)
5.0 to 95.0 %RH
±2.0 %RH
0.1 %RH
0.5 inch LCD, 4 digits with LED Backlight
Instrument: 32 to 125F (0 to 50C)
Probe: -4 to 212F (-20 to 100C)
Power Supply
Operating Time (Battery)
Instrument
Probe AP275
Probe AP100
RH/Temp Probe
6815
3 AA Alkaline Batteries
Approximately 150 hours
W 3.2 x D 1.5 x H 6.5 inches (8.2 x 3.8 x 16.5 cm)
2.75 inch diameter
-
1.00 inch diameter
6 inch x 1 inch diameter
1 x Rotating Vane Probe Head, your choice of AP275 or AP100
HTP 201 combination RH / Temp Probe (available for 6815)
Extension Rod w/ Handle Grip, Flexible Rod
3 x AA 1.5V Alkaline Batteries, Operation Manual, Carrying Case
※Please contact us for Standard Kit & Optional Accessories for those products
4
Air Velocity, Air Temperature,
Relative Humidity,
Differential Pressure
Anemomaster Model 6113, 6114
Multi-Function Thermal Anemometer
NIST Traceable
Built-in Printer (Model 6113)
Probe Compatibility
RS-232C Terminal
Features:
• Probe Compatibility feature allows utilization of a spare probe
• Simultaneously measures air velocity, temperature, and pressure
measurements
• Large, easy-to-read LCD
Benefits:
• Quick printing of measurements at site
• Simple operation with buttons for printing and calculation
• Easily keeps record of measurements
• Suitable for measurements inside ducts
Specifications
Air Velocity
Range
Air
Temperature
Range
Pressure
(Option)
Range
Measuring
Functions
32 to 212ºF (0 to 100ºC)
+/-2.0ºF (1.0ºC)
-5.00 to +5.00kPa
Accuracy
+/-(3% of reading +0.01)kPa
Settings
Display hold, time constant setting (1, 5, or 10 sec.), remaining battery life (5 steps),
measuring unit setting (m/s, fpm, Celsius, Fahrenheit, kPa).
Data Storage
Calculation
Instantaneous storage, average (over 60 sec. max.), Storage 100 max.
Maximum and minimum values, averaging, raw data display
Calendar function, air velocity bar graph display.
Digital Output
RS-232C (4800, 9600, 19200, 38400bps) for communication with PC.
Printer Output
Printing calculation results and measurements.
Analog Output
(Option)
DC 0 to 1V (1ch. Selected from air velocity, temp. and pressure)
Power Supply
6 × 1.5 volt C cells (Mn, alkaline, or Ni-Cd)
Battery Life
Operating
Environment
+/-(3% of reading +20fpm (0.1m/s))
Accuracy
Display
Outputs
20 to 9840 fpm (0.10 to 50.0 m/s)
Accuracy
10 hours continuous operation (with alkaline batteries, at 5m/s, 20C, without printer use)
Main Body
41 to 104ºF (5 to 40ºC)
Probe
32 to 212ºF (0 to 100ºC)
Storage Environment
41 to 104ºF (5 to 40ºC)
Weight (including batteries)
Approx. 2.2 lbs. (Approx. 1000 g)
Dimensions
Standard Kit
Optional Accessories
7.9” x 5.9” x 3.9” (200 x 150 x 100mm)
1 x operation manual, 6 x 1.5V Mn C cells, 1 x probe with 2-meter (6.6-feet) cable,
1 x extension rod, 1 x shoulder strap
Spare Probe, Analog Output, Pressure Measurement, Printer Paper, Communication Cable,
Data Management Software (for Windows), AC Adapter
5
Air Velocity, Air Temperature,
Relative Humidity,
Differential Pressure
Anemomaster Model 6162
NIST Traceable
Middle and High Temperature Anemomaster
Probe Model 0203 (Middle temp. Up to 392ºF (200ºC))
Model 0204 (High temp. Up to 932ºF (500ºC))
Features:
• Simultaneous display of air velocity and temperature
• Improved response time by the addition of secondary
temperature compensation circuit
• Easy review of time history by graphic display
• Memory function of maximum 999 separate measurement data
• Built-in RS-232 C serial interface for connection to PC. Analog
output and remote control terminal also equipped
• Probe Compatibility feature allows you to change the probe easily
Specifications
Model
Probe Model 0203 (for middle temp.)
32 – 392F (0 – 200ºC)
Measuring Range
Probe Model 0204 (for high temp.)
32 – 752F (0 – 400ºC)
40 – 9840fpm (0.2 – 50m/s): 32 – 212ºF (0 – 100ºC)
80 –9840fpm (0.4 – 50m/s): 212 – 392ºF (100 – 200ºC )
--
138 – 9840fpm (0.7 – 50m/s): 392 – 572ºF (200 – 300ºC)
--
197 – 9840fpm (1.0 – 50m/s): 572 – 752ºF (300 – 400ºC)
Measuring Accuracy
Air velocity:
+/-3%F.S.
Air temperature:
+/-(1%rdg+1ºC)
Temp. Compensation
Less than 984fpm (5m/s): +/-10%F.S.
Less than 984fpm (5m/s): +/-15%F.S.
Accuracy (Air Velocity)
984fpm(5m/s) to 9,840fpm(50m/s): +/-6%F.S.
984fpm(5m/s) to 9,840fpm(50m/s): +/-10%F.S.
Heat-resistance
Teflon coating (Probe side): 392ºF (200ºC)
of Cable
Vinyl code (Extension cable): 176ºF (80ºC)
Length of Cable
Extension Rod (Option)
Probe Dimensions
Display
Input/output Terminal
Power Supply
Teflon coating 4.9ft (1.5m)
Teflon coating 7.5ft (2.3m)
Vinyl code 16.4ft (5m)
Vinyl code 32.8 (10m)
0.65”(MAX)x31.5” (16.5x800mm)
0.87”(MAX)x81.5” (22x2070mm)
Dimension: φ0.43”x8.2” (11x208mm)
Dimension: φ0.43”x39.4” (11x1000mm)
Digital (simultaneous display of air velocity and temperature)
Remote terminal:
START/STOP key
Analog output terminal:
Output voltage 0 – 1 V, Output impedance 47 Ω
Digital output terminal:
RS-232C (serial interface)
Dry battery drive: US type (1.5V x 6 pcs = 9V), Alkaline battery, Mn battery
AC adapter: 12.5V, 450mA (AC100V +/- 10%, 50/60Hz)
Operating Temp.
41 – 104ºF (5 – 40ºC)
Battery Life
Dimensions
Weight
Approx. 8 hours
8.7” x 3.3” x 5.9”
220 x 85 x 150 mm
Main body: Approx. 4.0lbs (1.8kg), Probe Model 0203: 7.1oz(200g), Model 0204: 17.6oz (500g)
6
4 Channel Anemomaster Model 1570
Air Velocity, Air Temperature,
Relative Humidity,
Differential Pressure
NIST Traceable
Product Efficiency Control
Ventilation Testing
IAQ Investigation
Cleanroom
Features:
• 4CH simultaneous measurement in a compact body.
• Wide probe selection available for various applications.
• Capable of taking measurements at high-time-resolution (0.1 sec)
• RS232C terminal for data logging.
Benefits:
• Each channel display can be easily switched over by the touch of a button
• Probe compatibility saves your downtime and shipping cost
• Windows software gives you simultaneous measurement on air velocity and flow rate
• Measurement data saved in text format for easy processing using other software such as Excel
Main Body Specifications:
Display
Data Transfer
Software
Function
Output
LCD display
Transfers data of all channels
Burst Mode
Channel Mode
Transfers data of specific channel only
Instant Mode
Displays instantaneous values every specified sampling time
Average Mode
Displays average values of a specified sampling time
Digital
Analog
RS 232C (Baud rate 9600bps fixed)
0-5 V
Main Body Operation Temperature
41-104ºF (5-40ºC)
External Dimensions
7.8” x 10.2” x 2.8” (200 x 260 x 70mm)
Weight
Power Supply
Standard Kit
5.7 lbs (Approx. 2.6 kg)
AC 85V – 276V 50Hz/60Hz
RS-232 cable, Power cable, 2xFuse, Data Processing Software
Probe Specifications:
Model
Measuring Range & Accuracy
Temperature Compensation
Response Time
(1 m/s, 90% response)
0962-00 / 0963-00
0964-01 / 0964-02
0965-00/01/03/04/07/08
20-1000fpm: +/-20fpm
(0.1-4.99m/s: +/-0.1m/s)
1000-2000fpm: +/-40fpm
(5.00-9.99m/s: +/-0.2m/s)
2000-5000fpm: +/-100fpm
(10.0-24.9m/s: +/-0.5m/s)
5000-10000fpm:m+/-200fpm
(25.0-50.0m/s: +/-1.0m/s)
20-1000fpm: +/-30fpm
(0.1-4.99m/s: +/-0.15m/s)
1000-2000fpm: +/-60fpm
(5.00-9.99m/s: +/-0.3m/s)
2000-5000fpm: +/-150fpm
(10.0-24.9m/s: +/-0.75m/s)
5000-10000fpm:m+/-300fpm
(25.0-50.0m/s: +/-1.5m/s)
20-1000fpm: +/-30fpm
(0.1-4.99m/s: +/-0.15m/s)
1000-2000fpm: +/-60fpm
(5.00-9.99m/s: +/-0.3m/s)
2000-5000fpm: +/-120fpm
(10.0-25.0m/s: +/-0.6m/s)
41 – 176ºF (5 – 80ºC)
Approx. 1 sec
Approx. 3 sec
7
Approx. 7 sec
Air Velocity, Air Temperature,
Relative Humidity,
Differential Pressure
Multi-Channel Anemomaster Model 1550 & 1560
NIST Traceable
Real-Time Air Quality Monitoring System
1550
1560
Features:
•
Main Unit Features: One unit of Model 1550 has 64 Channels for air velocity.
•
For a larger system, connect up to 5 units in a cascade and add a computer for control.
•
Up to 320 channels of measuring points.
•
Flexibility in system configuration means greater freedom, simplicity, and efficiency in measurement.
Specifications:
Model
1550
1560
LCD (with backlight) Simultaneously displays on a module basis
Display
MAN: Select modules by operating the UP/DOWN switches
AUTO: Automatically selects modules every 2 seconds to display data
Burst Mode: Transfers data of all channels at high speed
Functions
Channel Mode: Transfers data of the specific channel only
Data Selection: Outputs instantaneous values every specified sampling time, and outputs the average
Data transfer: RS232C (standard),
Extended RS232C (optional)
GP-IB (optional)
Interface
Baud Rate: 300, 600, 1200, 2400, 4800, 9600, 19200bps (selectable)
Printer output: Centronics
Dimensions
16.9” x 19.6” x 5.5” (430 x 500 x 140 mm)
8.9” x 12.8” x 5.5” (226 x 325 x 140mm)
Weight
Approx. 22lbs (Approx. 10kg)
Approx. 11lbs (Approx. 5kg)
8
Probes:
Velocity & Temperature
Velocity
Model 0963-00 A-200 (Rod)
V: 20 – 9840 fpm (0.1 – 50 m/s)
Model 0963-21 A-200 Rod
V: 20 – 9840 fpm (0.1 – 50 m/s)
T: 32 – 212ºF (0 – 100ºC)
Model 0962-00 Caramel
V: 20 – 9840 fpm (0.1 – 50 m/s)
Model 0962-21 Caramel
V: 20 – 9840 fpm (0.1 – 50 m/s)
T: 32 – 212ºF (0 – 100ºC)
Model 0965-21 Spherical
V: 20 – 4920 fpm (0.1 – 25 m/s)
T: 32 – 212ºF (0 – 100ºC)
Model 0965-00/01 Spherical
V: 20 – 4,920 fpm (0.1 – 25 m/s)
Model 0963-31 A-200 Rod
V: 20 – 9840 fpm (0.1 – 50 m/s)
T: 32 – 140ºF (0 – 60ºC), H: 5 – 95%RH
L-Shaped (04)
Model 0965-03/04 Miniature with Built-in
Temperature Compensation
V: 20 – 4920 fpm (0.1 – 25 m/s)
I-Shaped (03)
Velocity, Temperature, & Humidity
L-Shaped (08)
Model 0965-07/08 Miniature with Independent
Temperature Compensation
V: 20 – 4920 fpm (0.1 – 25 m/s)
I-Shaped (07)
Model 0965-31 Spherical
V: 20 – 9840 fpm (0.1 – 50 m/s)
T: 32 – 140ºF (0 – 60ºC), H: 5 – 95%RH
Model 0964-01/02 Omni-directional
V: 20 – 4920 fpm (0.1 – 25 m/s)
Modules
Model 1504
V Module 4 ch.
Model 1511
VT Module 2 ch.
Model 1512
TVH Module 1 ch.
Model 1503
Pressure Module 2 ch.
Model 1510
Analog Output Module
Module Features:
Five types of modules add more freedom to system configuration.
• 4-channel air velocity module Model 1504
• 2-channel air velocity and temperature module Model 1511
• 1-channel air velocity, temperature, and humidity module Model 1512
• Static pressure module Model 1503
• Analog output module Model 1510
* Combine these modules and design a multi-channel system freely. More modules can be added whenever
necessary
9
Air Velocity, Air Temperature,
Relative Humidity,
Differential Pressure
Tabmaster
Features:
•
•
•
•
•
•
•
•
•
•
•
•
•
Designed For Maximum Contractor Productivity
30 hour battery life
Largest selection of hoods in the industry
Displays air volume flow, velocity and temperature in metric or imperial
measurements
Self-averaging, measuring sensor complies with ISO 3966
Detachable meter can be used with optional 4” rotating vane kit for face
velocity measurements
Optional 4” vane head and Aircone Hood Kit
Display indicates supply or exhaust flow-Memory, avg/max/min/sum,
communication to Windows® via RS232 serial port
Folding hood frame, no onsite assembly required
4 handle locations available (2 handles included)
Measurements down to 50 cfm without a low flow screen
Durable rubber grill seal outlasts typical foam seals.
True ambient readout. Displays flow in actual or standard CFM with onboard
conversion via user input of temperature and pressure
2' x 2'
16" x 16"
2' x 4'
610 x 634 mm
410 x 434 mm
610 x 1234mm
3' x 3'
1' x 4'
1' x 5'
910 x 934 mm
310 x 1234 mm
310 x 1534 mm
Specifications
Volume Flow
Velocity Range (using 100 mm head)
Display Resolution Volume
50-2400 cfm, 85-4078 m3/hr.,
24-1133 l/s
50-6000 ft/min, 0.25-30 m/sec
1 CFM,
1m3/hr,
1 l/sec
Temperature
1°F,
Weight (Instrument and 2’x2’ hood)
0.1°C
9.7 lbs.,
Memory Locations
4.4kg
99
Output
Battery Type/Life
RS232
4xAA Alkaline 35 hours
Standard
Volume flow kit with 2x2 foot hood, soft case
Options
Fabric Hood (Red, Yellow, Orange, Blue, Gray) Carrying Case, 4" vane kit
10
Air Velocity, Air Temperature,
Relative Humidity,
Differential Pressure
Airflow Transducer Model 6332 & 6332D
Features:
• 10 interchangeable probe options
• Probes quickly and easily attach/detach
• Digital Display Option (model 6332D)
• Selectable Output (Output can be set to either
voltage: 0 to 5V or current: 4 to 20mA)
Model 6332 (without display)
Main Unit Specifications
Measuring Object
Measuring Range
Measuring Accuracy
Temperature
Compensation Range
Selectable Output
Option
Selectable Range
Option
Display Resolution
(Model 6332D only)
Power Supply
Power Consumption
(Reference Value)
Connection Wire Size
Temp Range
Dimensions
Weight
Standard Accessories
Optional Accessories
Clean air and compatible, non-combustible gases
Maximum range 0.1 ~ 50 m/s (Range dependant on probe type*)
± (3% of reading + 0.1) m/s
0.1-4.99m/s
5.00-9.99m/s
10.0-24.9m/s
25.0-50.0m/s
5~40°C
±0.25m/s
±0.50m/s
±1.25m/s
±2.50m/s
40~80°C
±0.35m/s
±0.70m/s
±1.75m/s
±3.50m/s
(Set via onboard Dipswitch). Current output: DC 4~20mA (Max. load resistance: 250Ω)
Voltage output: DC 0~5V
(Rotary Switch selectable).
0~2, 0~5, 0~10, 0~25, or 0~50 m/s
Velocity Range: 0~2, 0~5, 0~10 m/s → Display Resolution: 0.01m/s
Velocity Range: 0~25, 0~50 m/s
→
Display Resolution: 0.1m/s
DC 12~24V
Approx. 2.0W (Under the following conditions - power supply: 12V, air velocity: approx. 10m/s, using a unit
with a display with probe model 0965-03.)
Power consumption rate is subject to change according to conditions such as air velocity, probe type and
use of display.
2
0.5 ~ 1.5 mm
Operating Temp: 41 to 104°F (5 to 40°C), Storage Temp: 14 to 122°F (-10 to 50°C)
Approx. 5” x 3.1” x 1.2” (128 × 78 × 30 mm)
Approx. 11oz (320g)
Operation Manual : 1, Main Unit Case: 1
Probe Cable (33ft (10m), 66ft (20m), 98ft (30m)), Display Unit (Supplied with a case with a display window
and 2 installation screws.), Dedicated AC Adapter (DC12V)
Probe Specifications
MODEL
0962-00, 0963-00
0964-01, 0964-02
0965-00, 0965-01
Measuring Range
0.1~50.0m/s
0.1~25.0m/s
0965-03, 0965-04
0.1~25.0m/s
0965-07, 0965-08
Sensor Type
Uni-directional
Omni-directional (Needle)
Omni-directional (Spherical)
Mini-temperature-compensation-sensor integrated type
Omni-directional (Spherical)
Mini-temperature-compensation-sensor independent type
Omni-directional (Spherical)
11
Airflow Transducer Model 6312
Air Velocity, Air Temperature,
Relative Humidity,
Differential Pressure
Features:
• Keep high quality of products by airflow control
• Maintain cleanliness by airflow control
• For creation of comfortable environment
• Optimal for various air-conditioning (air velocity and
airflow rate) control equipment
• Space saving design
• Traceable certificate available
• Current output range is 4 to 20mA
Main Unit Specifications
20 – 394 fpm (0.1 – 2.0 m/s)
Air Velocity Measuring Range
Accuracy: +/-30fpm (0.15m/s)
Temp. Compensation
64 – 82ºF (18 – 28ºC)
Signal Output
4 – 20mA
Power Supply
DC 24V
Operational Conditions
Clean airflow with normal temperature and humidity
Accessories
Probe Cable: 16 feet (5m)
Probe Specifications
Model
Dimensions
Quality of material
0941
0942
0.2” x 0.4” x 3.0”
φ0.4” x 4.3”
(5.5 x 11 x 75mm)
(φ11 x 108mm)
Molded Resin
SUS
12
Thermohygrometer,
Sounds / Noise
Measurement
Thermohygrometer Model 6841
Indoor Environmental Measurement
Products Quality Control
Features:
• Ultra light and palm-size
• Easy to use
(Select temperature, humidity, or power on/off)
Specifications
Measuring Object
Measuring range
Clean airflow
Temperature
-4 to 113ºF (-20 to 45ºC)
Humidity
Display resolution
5 to 95%RH
Temperature
0.1ºF or 0.1ºC
Humidity
Accuracy
0.1%RH
Temperature
+/-0.9ºF (+/-0.5ºC)
Humidity
+/-3%RH
Responsiveness
30 sec
Measuring Function
Dimensions
Conversion between ºF and ºC
Main Unit
Approx. 2.4” x 4.7” x 1.3” (60 x 120 x 34mm)
Probe
Approx. φ19 x 170mm
Cable
Approx. φ3.3 x 1.5m
Power Supply
Operating Environment
AAA Battery x 4 (Mn, Alkaline, or Ni-Cd)
Main Unit
41 to 104ºF (5 to 40ºC)
Probe
-4 to 113ºF (-20 to 45ºC)
Storage
14 to 122ºF (-10 to 50ºC)
Temperature
Weight (including batteries)
Approx. 6.3oz (180g)
Accessories
AAA Mn batteries x 4
13
IAQ Monitor,
Gas Monitor
Indoor Air Quality Monitor Model 2211
Multi-Parameter Indoor Air Quality Monitor
IAQ Investigation
Environmental Health and General Safety
Thermal Comfort Measurements
Features:
• Simultaneously measure and log data with multiple parameters
such as CO, CO2, Temperature, Relative Humidity, to monitor
indoor air quality conditions
• Calculates Dew Point, Web Bulb Temperature, Absolute Humidity,
Humidity Ratio and % Outside Air
• Built-in memory allows user to store multiple point measurements
Benefits:
• Easy user calibration with built-in calibration function
• Capable to continuously monitor parameters via PC
• Large LCD shows multiple parameters at a glance
Specifications
CO
CO2
Temperature
Range
Accuracy
Accuracy
0 – 5000PPM
+/-3% of reading or +/-50PPM whichever is greater
Range
-4 – 140ºF (-20 – 60ºC)
Accuracy
Range
Humidity
Accuracy
Normal Mode
Calculation Mode
+/- 1ºF (+/-0.5ºC)
2 – 98%RH
2 – 80%RH: +/-2%RH,
Calibration Mode
Digital
Analog (Option)
80 – 98%RH: +/-3%RH
Normal measuring: Relative humidity, CO, CO2, Dew point, Wet bulb temperature,
Absolute humidity, and Humidity Ratio
Store the data, and calculate the maximum, minimum and average value
% OA Mode
Data Output Mode
Outputs
+/-3% of reading or 3PPM whichever is greater
Range
Relative
Functions
0 – 500PPM
Measure ventilation ratio
Perform re-calculation of stored data and output of data to PC and Printer.
Perform the calibration of CO and CO2.
RS-232C (Baud rates:4800,9600,19200,38400bps)
DC0~1V (Select 1 ch among CO, CO2, temperature, and relative humidity)
Power Supply
AA Battery×6 (Alkaline or Ni-Cd), AC adapter:AC100~240V (50/60Hz)(Optional)
Battery Life
Approximately 10 hours (at 77ºF (25ºC) with alkaline batteries, without RS232C)
Operating
Main Body
Environment
Probe
41-104ºF
Storage Environment
Standard Kit
Optional Accessories
(5~40ºC)
- 4 - 140ºF
(-20~60ºC)
- 4- 140ºF
(-20~60ºC)
Mn Battery × 6, Calibration cap, Connection tube
Analog Output, Serial Printer, Communication Cable (For Computer and Printer),
AC Adapter
14
IAQ Monitor,
Gas Monitor
Aeroqual Series 200
Multi-Sensor Handheld Gas Monitor
The Series 200 can be hand held or fixed in position. It
is a simple, easy-to-use, low-cost monitor that displays
the gas concentration. The monitor is compatible with
the full range of Aeroqual gas sensors
Sensor Head
Monitor Base Specification
Measurement units
ppm
Power supply
12 VDC, 800 mA
Standard 9.6V (940mA/hr Ni-Cd)
Long Life 9.6V (2100mA/hr Ni-MH )
Rechargeable battery pack (2 x options)
Permanently fixable
Screw fix
Enclosure rating
IP20 & NEMA 1 equivalent
Size (with sensor head)
7.6 x 4.8 x 2.1 (in); 195 x 122 x 54 (mm)
Weight (with sensor head and battery)
< 16 oz; < 460 g
Functions
Remote sensor capability , Removable / replaceable sensor
head, Low battery indication, Sensor condition status,
Standby mode
Approvals
Part 15 of FCC Rules, EN 50082-1: 1997, EN50081-1: 1992
Sensor Head Specification
Ozone
Measurement
range
Accuracy
T90 response
Low
0.000 - 0.500
ppm
< 0.008 ppm (0 0.100 ppm)
±10% (0.100 –
0.500 ppm)
< 60 seconds
(T90)
Sensor type
Operating
temperature range
Operating relative
humidity range
Approvals
VOC*
High
Leak
0.50 - 20.00 ppm
0.00 - 50.00 ppm
±10% (0.20 2.00 ppm)
±15% (2.00 20.00 ppm)
< 35 seconds
(T90)
±20%,
15 seconds after
Reset
< 10 seconds
Gas-sensitive semiconductor
23°F to 122°F ; -5°C to 50°C (sensor head and base unit)
95% maximum (sensor head and base unit)
---
---
---
* Specific Calibrations to other VOC’s available – contact Kanomax Sales
15
Ammonia
CO
0 to 400 ppm
toluene
< ±10 ppm (0 –
200 ppm);
< ±10% ( > 200
ppm)
0 to 1000 ppm
< 60 s
< 60 s
< 150s
Gas-sensitive
semiconductor
-4°F to 140°F;
-20°C to 60°C
5 to 95%
non-condensating
---
Gas-sensitive
semiconductor
-4°F to 140°F ;
-20°C to 60°C
5 to 95%
non-condensating
---
Gas-sensitive
semiconductor
32°F to 158°F;
0°C to 70°C
5 to 95%
non-condensating
UL 2034, BS 7860
< ±5 ppm (0 – 100
ppm);
< ±10% (100 –
1000 ppm)
0 to 2000 ppm
carbon monoxide
< ±10 ppm in the
range 0 to 400
ppm
IAQ Monitor,
Gas Monitor
Aeroqual Series 300 & 500
Multi-Sensor Handheld Gas Monitor
•
•
The Series 300 can be hand held or fixed in position and
provides a high level of functionality & monitoring capability.
The Series 500 includes the same features and functionality
of the Series 300 with the added feature of onboard and PC
data logging.
Sensor Head
Monitor Base Specification
Measurement units
External signal for alarms & control
External signal functions
Analog output
Power supply
Rechargeable battery pack (optional)
Permanently fixable
Enclosure rating
Data interface with PC (Series 500 only)
On Board Data Logging (Series 500 only)
Size (with sensor head)
Weight (with sensor head and battery)
ppm or mg/m3
Transistor output, 150 mA max
Low Alarm, High Alarm & Control
0-5V
12 VDC, 800 mA
9.6V Ni-MH (5 hours operation)
Screw fix
IP20 & NEMA 1 equivalent
Serial RS232
8,000 data points
7.6 x 4.8 x 2.1 (in); 195 x 122 x 54 (mm)
< 16 oz; < 460 g
Remote sensor capability, Removable / replaceable sensor head,
On-board alarm, Alarm status displayed, Low battery indication, Alarm
mute function, Sensor condition status, Standby mode
Functions
(Series 500 only): Data logging independent of PC, Data logging direct
to PC, On-board real-time clock, Data logging software supplied
Approvals
Part 15 of FCC Rules, EN 50082-1: 1997, EN50081-1: 1992
Sensor Head Specification
Ozone
Measurement
range
Accuracy
T90 response
Low
0.000 - 0.500
ppm
< 0.008 ppm (0 0.100 ppm)
±10% (0.100 –
0.500 ppm)
< 60 seconds
(T90)
Sensor type
Operating
temperature range
Operating relative
humidity range
Approvals
VOC*
High
Leak
0.50 - 20.00 ppm
0.00 - 50.00 ppm
±10% (0.20 2.00 ppm)
±15% (2.00 20.00 ppm)
< 35 seconds
(T90)
±20%,
15 seconds after
Reset
< 10 seconds
Gas-sensitive semiconductor
23°F to 122°F ; -5°C to 50°C (sensor head and base unit)
95% maximum (sensor head and base unit)
---
---
---
* Specific Calibrations to other VOC’s available– contact Kanomax Sales
16
Ammonia
CO
0 to 400 ppm
toluene
< ±10 ppm (0 –
200 ppm);
< ±10% ( > 200
ppm)
0 to 1000 ppm
< 60 s
< 60 s
< 150s
Gas-sensitive
semiconductor
-4°F to 140°F;
-20°C to 60°C
5 to 95%
non-condensating
---
Gas-sensitive
semiconductor
-4°F to 140°F ;
-20°C to 60°C
5 to 95%
non-condensating
---
Gas-sensitive
semiconductor
32°F to 158°F;
0°C to 70°C
5 to 95%
non-condensating
UL 2034, BS 7860
< ±5 ppm (0 – 100
ppm);
< ±10% (100 –
1000 ppm)
0 to 2000 ppm
carbon monoxide
< ±10 ppm in the
range 0 to 400
ppm
IAQ Monitor,
Gas Monitor
Tracer Gas Hardware
As Per ANSI/ASHRAE Standard 110-1995
In-line
Tracer Gas
Diffuser
Flow
(Ejector)
IAQ Investigation
Industrial Hygiene
Quality Control
Meter
Features:
The Kanomax tracer gas diffuser is for use in performing
the Tracer Gas test in accordance with ANSI/ASHRAE
Standard 110-1995. Our diffuser is an improved design
and meets the specifications of Standard drawing
#110-83M. The diffuser is placed in the fume hood and
sulfur hexafloride gas is injected at a supply pressure of
30 psig.The internal critical orifice ensures a flow rate of
Regulator
Dif-Kit
4 liters per minute. Other orifice sizes can be inserted.
Also available is the Kanomax model Dif-Kit.
The Dif-Kit supplies all the required hardware from the cylinder regulator to the diffuser. This includes the
diffuser, tank regulator, in-line flow meter, shut-off valve, pressure gage, and 25 ft. of tubing. The in-line
flow meter provides a secondary verification of flow rate to the critical orifice, as required by the
Standard, so as to alert the user to any clogging or wear of the orifice. The orifice and other components
are serviceable by Kanomax.
Specifications
Critical Orifice
Regulator
Flow Meter
Pressure Gage
Options
4 or 8 Liter Per Minute (4Lpm Standard)
Dual Stage, Specialty Gas
Calibrated 150mm/200psi, Glass Tube
Large Dial, 0 to 60psi
Rebuild Kit (includes new Critical Orifice, washers, Bronze Sinter Filter)
8Lpm Swappable Ejector Base
17
Particle / Dust
Monitor
Piezobalance Dust Monitor Model 3521
Respirable Aerosol Mass Monitor
NIST Traceable
IAQ Investigations
Environmental Measurements
Product Quality Control
Laboratory Research
Features:
• Data logging ability allows user to log 500 samples
• Simple cleaning mechanism for easy maintenance
• RS232C port available as standard
• Back-lit display is easy to read in poor lighting conditions
Benefits:
• While conventional dust meters “count” particulates, the Piezobalance dust
meter “weighs” mass concentration of particulates
• Data can be reviewed on-screen, printed or downloaded to a computer
• Software included for easy data download and processing
• Easy operation requires no special training
Specifications
Measuring Object
Airborne Particulate Matter < φ10 µm
Measuring Range
0.02-10mg/m3
Sampling Flow Rate
1liter/min
Preset: 120 sec or 24 sec
Measuring Time
Accuracy
Cleaning Frequency
Data Logging
Digital Output
(To PC or printer)
Power Supply
Manual: 10sec to 3600sec
+/-10% of rdg +/-1 digit (0 to 5mg / m3)
+/-20% of rdg +/-1 digit (5 to 10mg / m3)
Cleaning is required after every 10-20 measurements.
"CLEANING" will appear on the display when it needs to be cleaned.
Max 500 date and time stamped samples
RS232C (Baud Rates 4800, 9600, and 19200)
AC (85V-240V) or DC Ni-MH battery
2.6” x 5.9” x 7.1”
Dimensions
(65 x 150 x 180mm)
Weight
Approx. 4.4lbs. (Approx. 2kg)
18
Particle / Dust
Monitor
Digital Aerosol Monitor Model 3431
Compact Dust/Aerosol Monitor
NIST Traceable
Indoor Air Quality Investigation
Industrial Hygiene
Health & Safety Applications
Exposure Monitoring
Manufacturing Process Control
Features:
• State-of-the-art components based on advanced
light-scattering technology
• Compact and Light Weight
• Easy operation
Benefits:
• Accuracy is ensured to meet the latest needs of dust measuring in extremely clean indoor environments
• All measurements can be made with a touch of a button on the front panel.
• Allows the user to pre-enter the known concentration conversion value and let the unit automatically convert the
dust count into mass concentration
Specifications
Measuring Method
Light Scattering Method
Measuring Object
Measuring Range
Accuracy
Aerosol particles, φ10µm or smaller
3
0.001 – 3.999mg/m (1 – 9999cpm, 1cpm=0.001mg/m3 for φ0.3µm stearic acid particles)
+/-10% of reading +/-1cpm (or 0.001mg/m3)
Linear Characteristics
Measuring Time
+/- 5% of reading
1 minute / 3 minutes / 10 minutes with built-in timer, and continuous mode
Light Source
Laser diode
Detector
Photo diode
Operating Environment
Display
Indicators
Output
Power Supply
Dimensions
Weight
Accessories
Options
35 – 104ºF (5 – 40ºC) under 85%RH
4-digit LCD, dust count value (cpm), relative mass concentration value
Cleaning, measuring mode, battery status, concentration conversion operation, zero point
confirmation
Analog output, 0 – 10,000cpm = 0 – 4V
AC 100V adapter and dry cell batteries (6 x AA batteries)
6.4” x 2.4” x 39” (162 x 62 x 100mm) not including the nozzles
2.2 Lbs. (1kg) not including batteries
AC adapter, micro-screwdriver, alkaline dry cell batteries (6 x AA) shoulder strap
Carrying case, Soft case, Analog output cable
19
Particle / Dust
Monitor
Handheld Laser Particle Counter MODEL 3887
NIST Traceable
Cleanroom certification
IAQ investigations
Food Industry
Filter Testing
Aerospace
Hospital Surgical Rooms
Paint Spray Booths
Features:
• Displays 3 particle sizes simultaneously (0.3, 0.5, 5.0 um) in cf or m3
• Built-in flow sensor (0.1 CFM +/- 10%)
• User can log up to 8,000 measurements with easy transmission to PC or Printer.
• Fits in your palm
Benefits:
• Batteries can be altered with regular AA alkaline batteries
• Comes with a network capable of up to 8 units using optional Windows software
Specifications
as a standard accessory
Measuring Particle Size
0.3, 0.5, 5.0um
Light Source
Counting Efficiency
Zero Count
Coincidence Loss
Laser Diode
50% @ 0.3 um; 100% for particles > 0.45 um (per JIS B9921: 1997)
≤1 count / 5 minutes (per JIS B9921: 1997)
<5% @ 2,000,000 particles/cf
Flow Rate
Sampling Time
Sampling Frequency
0.1 cfm (2.83l/min)
1 sec to 99 minutes 59 sec (1 sec increments)
1 to 99 times or Continuous
Count Alarm
Mode of Measurements
1 - 70,000,000 counts
Single / Repeat / Continuous / Calculation / Remote / ISO <C4
Display
Error
20 letters, 4 lines
Max. concentration, Laser Power, Flow Rate and Battery
Interface
RS232C or RS485
Baud Rate
9600bps
Buffer Memory
8000 measurements
Power Supply
4 x AA NiMH battery or Alkaline, AC Adapter (100 - 240V)
Operating Hours
Dimensions
Approximately 3 hours (NiMH)
4.3” x 7.7” x 2.7”
Weight
Operating Conditions
Accessories
Options
(108 x 196 x 68 mm)
1.5 lbs (680g)
50 - 95ºF (10 - 35ºC)
AC Adapter, Zero Filter, Software, Communication Cable, 4xAA NiMH, Battery Charger
Printer, Printer Cable, Carrying Case, Tripod
20
Particle / Dust
Monitor
Handheld Laser Particle Counter MODEL 3886
GEO-α
NIST Traceable
Cleanroom certification
IAQ investigations
Food Industry
Filter Testing
Aerospace
Hospital Surgical Rooms
Paint Spray Booths
Features:.
• Measures 5 particle sizes simultaneously (0.3, 0.5, 1.0, 3.0 and 5.0
um) in cf or m3, displaying 2 channels
• Built-in flow sensor (0.1 CFM +/- 10%)
• Temp./RH and Air Velocity probes available as an option
• Capable of measuring up to 4 parameters in one instrument.
Specifications
5 Size Channels
0.3 um; 0.5 um; 1.0 um; 3.0 um; 5.0um
Flow Rate
0.1 cfm (2.83 l/min)
Light Source
Laser Diode
Calibration
Counting Efficiency
Zero Count
Coincidental Loss
Sampling Time
PSL particles in air
50% @ 0.3 um; 100% for particles > 0.45 um (per JIS B9921: 1997)
≤1 count / 5 minutes (per JIS B9921: 1997)
<5% @ 2,000,000 particles/cf
1 sec - 99 min 59 sec, 1 second increment
Sampling Frequency
Count Modes
1 - 99 or continuous
Single, Repeat, Continuous, Calculation and Remote
Display
20 letters, 4 line LCD
Interface
RS232C or RS485
Baud Rate
9600 bps
Buffer Memory
AC Adapter
Power
500 sample records
5VDC at 2.5A, 100 - 240VAC, 50 to 60 Hz
4 x AA NiMH
Rechargeable
Battery
Vacuum Source
Dimensions
3.5 hrs with 1600mAh batteries
Internal pump, flow controlled
4.5" x 2.8" x 8.5"
Weight
Accessories
Options
(115 mm x 70 mm x 211 mm)
2.2 lbs (1 kg)
AC Adapter, Zero Filter
Isokinetic Probe, Air Velocity Probe, Thermal Printer,
Temperature/Relative Humidity Probe, Carrying Case
21
Particle / Dust
Monitor
Airborne Particle Counter Model 3900
Portable Laser Particle Counter
NIST Traceable
Monitor and verify cleanrooms
Pharmaceutical, Food Industry, Semiconductor
Filter testing
Facility certification
Features:
28.3 L/min (1.0 cfm) flow rate
0.3 micron minimum sensitivity
Store up to 5000 measurements
All stainless body
Large touch screen
Multi-function: Measuring particle, airflow, temperature
relative humidity, and differential pressure
Specifications
22
Particle / Dust
Monitor
Cleanroom Facility
Monitoring
Condensation Nucleus Counter
NIST Traceable
MODEL 3855
Cleanroom Monitoring
IC Production
HDD Testing
MEMS
Features:
• Condensation Nucleus Counter for continuous multi-point
monitoring in cleanroom.
• 0.01μm sensitivity.
• Less downtime by less alcohol replacement.
• Easy checking of the current status on the LEDs
Benefits:.
• Ideal for ultrafine particle detection
• Easy maintenance
• No more headaches in cleanroom using Propylene Glycol
Specifications
Type
Principle
Maximum Concentration
Sensitivity
Sensor-Counter Isolated Type
Vapor/Air Mixing Type Condensation Particle Counting
3×106 Particles / cf
0.01µm (50% Counting Efficiency)
Measuring Accuracy
Coincidence Loss
Sampling Flow Rate
Flow Rate
(100 Particles/cc)
0.02µm: 100% ± 10%
Less than 5% at 3,000,000 Particles/cf
0.1CFM
(2.83L/min) ± 10%
4.2 ± 0.4L/min (Requires external vacuum source)
Alcohol
Propylene Glycol
Indicators
Ready, Liquid Level, Flow, Optics, Particle
Interface
RS‐232C and RS‐485
Output
Operating Environment
Dimensions
Output for M/P or Analog Output
Temperature:
9.8” x 9.8” x 11” (250 × 250 × 280mm) ( Excludes alcohol bottle )
Weight
Approx. 22lbs (10Kg)
Power Supply
Accessories
Options
20~30ºC ( 68~86ºF )
Clean Air Environments (e.g. Clean Rooms)
AC100 ~ 240 V
AC Cable, Operation Manual, Alcohol Bottle ( 200cc ), Bottle Holder
Alcohol Bottle ( 1000cc ), Bottle Holder
23
( for 1000cc )
Particle / Dust
Monitor
Cleanroom Facility
Monitoring
Facility Monitoring Laser Particle
Sensor MODEL 3792
NIST Traceable
Cleanroom monitoring
Pharmaceutical, Aerospace/ Defense
Electronics, MEMS, Semiconductor
Food processing
Medical / Hospitals
Features:.
•
•
•
•
•
•
0.2um/0.3um sensitivity
Built-in sonic nozzle for accurate and consistent flow
Built-in LEDs for sensor status at a glance
Compact stainless body
Temperature/Relative Humidity sensors available as option
Status LED shows Particle, LD Status, and Flow
Specifications
Model
Sampling Air Flow
Sensitivity
Condensation Range
Type A
Output
Type B
Display
Status LED
Power Supply
Dimensions
Weight
3792-01
3792-03
2.83 l/min
1 l/min
0.2 μm, 0.3 μm
0.2 μm, 0.3 μm
1,000,000 pieces/cf
3,000,000 pieces/cf
RS-485 Output
Connector Type RS-485 IN
Modular Connector 8-pin (rear panel)
RS-485 OUT Modular Connector 8-pin (rear panel)
Transmission Setting: DIP switch 8-pin (front panel)
Address Setting(0-999): Rotary DIP switch, 10 positions x 3 (front panel)
Multiplexer Output
Connector Type: BNC Connector
4-20mAOutput
Output: Data for 2 particle size ranges and Sensor Status
Connector Type: D-sub Connector 9-pin (rear panel)
Indicative Resolution: 1/1000 of the full scale
Full Scale: 3 levels (100、10,000、1,000,000 pieces / cf)
Full Scale Selection: Sliding DIP switch 8-pin (front panel)
Multiplexer Output
Connector type: BNC Connector
5-Digit LCD
Optional Display unit available (needs to be exchanged with the supporting casing)
Condensation Display(Running Average for pre-set time)
SIZE key gives selection of Particle size range to monitor
PARTICLE: Turns Red when picking the signal of the particle
STATUS: Steady Green when Optics is operating normally.
FLOW:
Turns to Green when Sampling Flow rate is normal
AC 100V-240V
4.6” x 5.3” x 6.7”
(118×135×170 mm)
(Excluding sampling air nozzle on the upper surface and other interface outfit.
Also excludes rubber feet.)
Approx. 6.6lbs (Approx. 3kg)
24
Cleanroom Facility
Monitoring
Facility Monitoring Laser Particle Sensor MODEL 3714/3715
NIST Traceable
Cleanroom Facility Monitoring
•
Pharmaceutical
•
Aerospace/ Defense
•
Electronics
•
Semiconductor
•
Food Processing
•
Medical / Hospitals
Features:
• 0.3um/0.5um for Model 3714 & 0.5 um /5.0 um for
Model 3715 sensitivity
• Built-in sonic nozzle for accurate and consistent flow
• Built-in LEDs for sensor status at a glance
Benefits:
• Compact stainless body
• Temperature/Relative Humidity sensors available as option
• Status LED shows Particle, Status (LD), and Flow (Optional)
Specifications
Model
Particle Sizes
3714
3715
0.3um/0.5um
0.5um/5.0um
Flow Rate
Concentration Range
Flow Control
Light Collection
Light Source
Light Detector
Status LED
Measuring Accuracy
0.1CMF ( 2.83 L/min )
0~1,000,000 particles / cf
Sonic nozzle
( Requires external vacuum source )
Laser-light-scattering, wide-range light collection by side mirror
780nm Laser diode
Photodiode
PARTICLE, STATUS (LD), FLOW
Power Supply
Dimensions
(Optional)
+/-10% of reading
Interface
Connector Type
( LD )
RS-485
RJ45 connector
Supplied from the Distributor or Power Unit
2.8” x 5.0” x 1.6”
(72 × 126 × 40mm)
Weight
Approx. 2.2lbs (Approx. 1kg)
25
Cleanroom Facility
Monitoring
Cleanroom Monitoring System
Features:.
• On-site monitoring of condensation by built-in LCD
• Light and compact body does not require large space
• CPU loaded and RS485 output
• Steel stand and attachment plate for various installation needs
• Quick connect cabling via modular connection
• Simple vacuum tubing/pressure connected
• Compact laser particle sensor (LPS) with
built-in LCD for continuous monitoring of
aerosol level
• Multi-function, use-friendly monitoring software
• Reduced piping & wiring using a single
compact pump for entire system
• Interface unit for environmental sensors such
as temperature, relative humidity, differential
pressure, etc.
Distributor (Model 3770)
Supplies power to a maximum of 8 sensors or interface units.
Input
8 channel RS-485
Interface
Power
Dimensions
Weight
26
Baud: 4800/9600/19200bps
AC 85-132V/170-264V
11.8” x 3.9” x 7.9”
(300 x 100 x 200mm)
Approx. 6.6lbs
(Approx. 3kg)
Aerosol Research
Instrumentation
Hand-held Condensation Particle Counter Model 3800
NIST Traceable
Indoor Air Quality Investigation
Aerosol Research
Filter Tests
Environmental Monitoring for
•
Electronics
•
Food Processing
•
Pharmaceutical
•
Medical / Hospital, etc.
Features:
• 0.015 µm sensitivity
• Concentration range of 0 to 100,000 particles/cm3
• Programmable data-logging capabilities
• Power Supply can be selected from alkaline/Ni-MH battery or AC adapter
• Simple to download the data to your computer via USB
Specifications
Measuring Particle Size
Concentration Range
Counting Efficiency
Zero Count
0.015 to > 1µm
0 to 100,000 particles/cm3
50nm: 100±20% (15nm: More than 50%)
Less than 1 particle/ cm3
Flow Rate
Alcohol
Supply
Aerosol Flow: 100cc/min
Total Flow: 700cc/min
Type
Hours per Fill
100% reagent-grade isopropyl alcohol
Approx. 5 hours at 23ºC(73° F)
Absolute Pressure Sensor
Mode of Measurements
150 to 1150 hPa
Repeat / Program / Counter
Display
Interface
Built-in LCD ( 128 x 64 dots)
USB
Buffer Memory
Power Supply
Type
Operating hours
Environmental Operating
Condition
Dimensions
Weight
Standard Accessories
Options
Max. 10,000 measurements
6 pieces of AA-size Alkaline / Ni-MH battery or AC adapter (Input 100 – 240V)
Approx. 5 hours ( By Alkaline batteries );
Approx. 8 hours ( By Ni-MH batteries )
Ambient temperature range: 10 to 35ºC ( 50 to 95° F)
4.7”(W) x 11”(H) x 5.1”(D)
(120 x 280 x 130mm)
Approx. 3.3lbs (1.5kg) ( without batteries )
AC adapter, Zero filter , Alkaline battery×6, Operation manual,
Software(for Windows), PC Communication Cable, Carrying Case
Ni-MH battery (1.2V-2500mA) x 6, Charger x 2,
Printer, Printer cable, Printer AC adapter
27
Aerosol Research
Instrumentation
Aerosol Particle Mass Analyzer Model APM-3600
This analyzer classifies the mass of a single aerosol based on the balance
between centrifugal force and electrostatic power
Diesel Exhaust Aerosol Research
Nano-Particle Aerosol Research
Atmospheric Aerosol Research
Bio-Chemical Applications
Features:
The APM-3600 classifiers particles for each particle size, based on the aerodynamic movement of the particles.
This analyzer uses our own rotating mechanism and seal technologies, as well as taking into consideration the
static effects and flow of the fluid.
Main Body Specifications
Classifying method
Classifying mass range
Double cylinder rotating
Double cylinder high voltage
Double cylinder dimensions
Sampling Flowrate
Dimensions
Weight
Classification is based on the balance between centrifugal force and static power.
Approx. 0.01~100 Femto-gram(For particle concentration of 1g/cm3 approx.
40nm~500nm, classification accuracy is within ±30% regarding center mass)
~8,000rpm
~3,000V
Inner cylinder diameter: 3.9” (100mm), Outer cylinder diameter: 4.1” (104mm),
Cylinder length: 9.8” (250mm)
Above 1L/min
15.7” x 15.7” x 47.2” (400mm x 400mm x 1200mm)
Approx. 165lbs (75kg)
Control, Display
Control function
Display function
Control Method
Dimensions
Weight
Power supply
Rotation rate and applied voltage
Applied voltage / Rotation rate / Differential pressure (Panel display)
Control by panel, Control by PC (By Manual / Remote Switch)
16.9” x 16.9” x 11.8” (430mm x 430mm x 300mm)
Approx. 55lbs (25kg)
AC100V、50/60Hz
28
Fluid Mechanics
Research
Instrumentation
Smart LDV System
High Quality and Compact LDV System
Features:.
• The best device for precision PIV calibration and
measurement of medium or low velocity
• Easy to use
Specifications
Available velocity
-10m/s~30m/s (Please contact us of the application to over 30 m/s.)
Laser
Focal Distance
Optics
Laser Diode, λ=635nm, 10mW
150mm (Options:170mm, 200mm, 250mm)
Measurement Volume
0.09mm×0.7mm
Measurement Method
Back Scattering/Forward Scattering
(A retro reflector mirror is used with a standard model to obtain high signal quality)
Probe Size
60mmφ×300mm
Shift frequency
Signal
Processor
0.01-10MHz(1-2-5step)
Signal Processing
8-bit FFT(512,256,128 points)
Frequency Range
1kHz~40MHz (8 ranges)
Max. Data Rate
8000 Data/sec
Effective Judgment
Burst spectrum ratio
Computer
IBM PC Compatible
Max. Number of Data
Software
99,000
Realtime Monitor
Burst signal, Burst spectrum, Velocity histogram
Data Processing
Mean velocity, Turbulence intensity, Skewness factor,
Flatness factor, Velocity histogram, Time history
Data Output
CSV Format
29
R & D / Test
Instrument and
System
Amenity Manikin
For Automotive Interior Environmental Measurement
R&D
Design
Engineering
Features:
The life sized Manikin is designed to accurately emulate
the positioning of a driver or passengers torso, arms and
legs while seated in the vehicle passenger compartment.
Placement and positioning is simple and data collection
can be started immediately. The Manikin deploys over
120 sensors that can be custom configured by the end-user
on-sight. There are 4 sensor types to choose from that
include Air Velocity, Air Temperature, Relative Humidity and
Radiant Heat. The system components include the
Manikin with built-in data processor, Sensors, Power and
Data Transfer Unit, PC Interface, Monitoring PC and
Data-Logging Software. Up to 2 Manikins can be linked to
a single monitoring PC.
Manikin Specifications
Height
Approx 5’7” (Approx. 170cm)
Weight
88lbs (40kg) including data processor
Material
Plastic (FRP)
Sensor allocations (example)
Air Velocity
Temperature
Relative Humidity
Radiant Heat
Head
4
12
1
3
Upper Torso
12
33
0
5
Lower Torso
20
33
1
4
Total
36
45
2
12
Sensors Specifications
Range
Air Velocity
Temperature
Accuracy
Range
20 – 984fpm (0.1 - 5 m/s)
20 – 394fpm (0.1-2m/s)
±10fpm (0.05m/s)
394 - 984fpm (2-5m/s)
±20fpm (0.10m/s)
-22 – 212F (-30 - 100ºC)
Accuracy
±4.6F (3ºC)
Range
Humidity
3-95%RH
3-30%RH
Accuracy
Wavelength
Radiant Heat Range
Accuracy
±3%RH
30-75%RH
±2%RH
75-95%RH
±3%RH
0.3-40µm
0-1Kw/m2
±7%
30
34
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF