Beamex MC4

Beamex MC4
DOCUMENTING PROCESS CALIBRATOR
39
Document as you go
40
MC4: a compact, easy-to-use documenting process calibrator
The Beamex MC4 is a documenting process calibrator.
Instrument data can be sent from a computer to the MC4 and calibration results can be uploaded from the MC4 to a computer containing the Beamex CMX calibration software.
Being a multifunctional calibrator, the MC4 is suitable for calibrating various process parameters, such as pressure, temperature and electrical signals.
High accuracy is one of the important features of the MC4. A standard feature of the MC4 includes an accredited calibration certificate as proof of its accuracy. The correction coefficients of a PRT probe can be programmed into the MC4 to further improve temperature accuracy. The large graphical display, menubased multilingual user interface and full numerical keyboard make it easy to use.
Main features of MC4
Communication with calibration software
Using the MC4 together with a calibration software provides you with a complete documenting calibration system that produces calibration certificates automatically.
All-in-one functionality
The MC4 is a versatile calibrator with many different functions.
There’s no need to take several different measurement devices to the field – the MC4 does the job.
Accuracy guaranteed
The MC4 is a highly accurate process calibrator. As a proof of this, each calibrator is delivered with a traceable, accredited calibration certificate.
Calibration is quick and easy
The large graphical display, menu-based multilingual user interface and full numerical keyboard make the MC4 quick and easy to use.
BEAMEX MC4
| FIELD EQUIPMENT
41
42
Advanced features of MC4
FEATURE
Calibration mode
PRT sensor coefficients
Error % display
Error display in input or output units
% display
Scaling
SPECIFICATION
The MC4 includes a versatile calibration mode making it easy and effective to create and calibrate process instruments.
The MC4 compensates sensor errors because it includes the possibility to record PRT sensor correction coefficients.
When calibrating a transmitter, its output may be displayed in an error % unit rather than in an engineering unit.
When calibrating a transmitter, the transmitter’s output may be displayed as an error in input or output engineering units.
User setups
Leak testing
Step and ramp
Programmable alarms
Damping
Bar graph
Difference
Deviation
Redundancy
Additional information
Any measurement or generation may be presented in percentages within the user-programmable range.
A versatile, programmable scaling function allows the user to scale any measured or generated unit into any custom unit.
Scaling also includes a rooting transfer function for flow applications as well as custom transfer functions.
The unit has a large number of user-configurable setups that make it easy to save and quickly recall a desired configuration.
The leak test function indicates pressure drops and leak rates during the user-programmable period.
The unit includes a versatile and programmable automatic step and ramp function as well as a manual step function.
An alarm based on a measurement value or rate of change can be programmed into the device.
Programmable damping allows the user to select different filters for measurements.
The bar graph allows the user to display a measurement or generation as an analogue bar, including programmable starting and ending points.
Difference measurement allows the user to measure the difference between two pressure modules.
The deviation function allows the user to display a deviation between a given reference value and the actual measurement.
Redundancy measurement allows the user to measure the same pressure using two pressure modules (internal and external) simultaneously. The unit’s alarm sounds if the readings excessively differ from each other.
The unit also allows the user to view additional information such as min, max, rate, internal temperature, thermocouple’s thermovoltage, RTD sensor’s resistance etc.
FIELD EQUIPMENT | BEAMEX MC4
General specifications
FEATURE
Display
Weight
Dimensions
Keyboard
Battery type
Charging time
Charger supply
Battery operation
Battery operation with optional dry battery cartridge and 4 alkaline AA cells
Operating temperature
Operating temp. while charging batteries
Storage temperature
Humidity
Warm-up time
Max. input voltage
Safety
EMC
SPECIFICATION
60 mm x 60 mm (2.36" x 2.36"), 160 x 160 pixels, back lit LCD
720 … 830 g (1.59 … 1.83 lbs)
215 mm (8.5") x 102 mm (4") x 49 mm (1.9") (d/w/h)
Membrane keyboard
Rechargeable NiMH pack, 4000 mAh, 3.6V DC
5 hours
100 … 240 VAC, 50–60 Hz
13 … 24 hours in measurement mode, back light off.
8 … 12 hours when sourcing an average of 12 mA to loop, with back light on.
4 … 8 hours in measurement mode, back light off.
3 … 4 hours when sourcing an average of 12 mA to loop, with back light on.
−10 … 50 °C (14 … 122°F)
0 … 35 °C (32 … 95°F)
−20 … 60 °C (−4 … 140°F)
0 to 80 % R.H. non-condensing
Specifications valid after a 5-minute warm-up period.
30 V AC, 60 V DC
Directive 73/23/EEC, EN 61010-1
Directive 89/336/EEC, EN 61326
VOLTAGE MEASUREMENT −1 … 60 V DC
RANGE
±0.25 V
±(0.25 … 1 V)
1 … 25 V
25 … 60 V
RESOLUTION
0.001mV
0.01 mV
0.1 mV
1 mV
FEATURE
Temperature coefficient
Input impedance
Supported units
Display update rate
1 YEAR UNCERTAINTY (±)
(1
0.02 % RDG + 5 µV
0.02 % RDG + 5 µV
0.02 % RDG + 0.25 mV
0.02 % RDG + 0.25 mV
SPECIFICATION
< ±0.0015 % RDG / °C outside of 18 … 28 °C
< ±0.0008 % RDG / °F outside of 64.4 … 82.4°F
>1 MΩ
V, mV, µV
3 / second
mA MEASUREMENT ±100 mA
RANGE
±25mA
±(25 … 100 mA)
FEATURE
Temperature coefficient
Input impedance
Supported units
Display update rate
RESOLUTION
0.0001 mA
0.001 mA
1 YEAR UNCERTAINTY (±)
(1
0.02 % RDG + 1.5 µA
0.02 % RDG + 1.5 µA
SPECIFICATION
< ±0.0015 % RDG / °C outside of 18 … 28 °C
< ±0.0008 % RDG / °F outside of 64.4 … 82.4°F
< 7.5 Ω mA, µA
3 / second
LOOP SUPPLY
FEATURE
Maximum output current
Output voltage
Output impedance in HART compatible mode
SPECIFICATION
> 25 mA, short circuit protected
24 V ±10 %
300 Ω ±20 %
1) Uncertainty includes reference standard uncertainty, hysteresis, non-linearity, repeatability and typical long-term stability for the mentioned period (k=2).
BEAMEX MC4
| FIELD EQUIPMENT
43
44
Electrical measurements
FREqUENCY MEASUREMENT 0.0027 … 50 000 Hz
RANGE
0.0027 … 0.5 Hz
0.5 … 5 Hz
5 … 50 Hz
50 … 500 Hz
500 … 5000 Hz
5000 … 50000 Hz
RESOLUTION
0.000001 Hz
0.00001 Hz
0.0001 Hz
0.001 Hz
0.01 Hz
0.1 Hz
1 YEAR UNCERTAINTY (±)
(1
0.01 % RDG
0.01 % RDG
0.01 % RDG
0.01 % RDG
0.01 % RDG
0.01 % RDG
FEATURE
Temperature coefficient
Input impedance
Trigger level
Minimum signal amplitude
Supported units
Gate period
SPECIFICATION
Specification valid from −10 to 50 °C (14 … 122°F)
> 1 MΩ
−1 … 14 V in 1 V steps and open collector inputs
2 Vpp (< 10 kHz), 3 Vpp (10 … 50 kHz)
Hz, kHz, cph, cpm, 1/Hz (s), 1/kHz (ms), 1/MHz (µs)
267 ms + 1 signal period
1) Uncertainty includes reference standard uncertainty, hysteresis, non-linearity, repeatability and typical long-term stability for the mentioned period (k=2).
PULSE COUNTING 0 … 9 999 999 PULSES
FEATURE
Range
Input impedance
Trigger level
Minimum signal amplitude
SPECIFICATION
0 to 9 999 999 pulses
> 1 MΩ
−1 … 14 V in 1 V steps and open collector inputs
2 Vpp (pulse length > 50 µs), 3 Vpp (pulse length 10 … 50 µs)
SwITCH TEST
FEATURE
Potential free contacts
Voltage level detection
SPECIFICATION
Test voltage (trigger level)
Trigger level
Input impedance
3 V, 0.13 mA (1 V) or 24 V, 35 mA (2 V)
−1 … 14 V in 1 V steps
> 1 MΩ
FIELD EQUIPMENT | BEAMEX MC4
Pressure measurements
INTERNAL PRESSURE MODULES (NPM)
INTERNAL MODULE
3)
NPM200mC
NPM2C
NPM20C
NPM160
Barometric option
UNIT RANGE
2)
RESOLUTION 1 YEAR UNCERTAINTY (±)
(1 kPa mbar iwc kPa bar psi
±20
±200
±80
−100 to 200
−1 to 2
−14.5 to 30
0.001
0.01
0.001
0.001
0.00001
0.001
0.035 % FS + 0.05 % RDG
0.015 % FS + 0.035 % RDG kPa bar psi
MPa bar psi
−100 to 2000
−1 to 20
−14.5 to 300
0 … 16
0 … 160
0 … 2400
0.01
0.0001
0.01
0.0001
0.001
0.01
0.015 % FS + 0.035 % RDG
0.015 % FS + 0.035 % RDG
Also enables absolute pressure measurement for the above pressure inputs. When using the barometric option, add 0.1 kPa (0.0146 psi) uncertainty for absolute pressure measurement.
FEATURE
Temperature coefficient
Maximum overpressure
Pressure port
Media compatibility
Supported pressure units
Display update rate
SPECIFICATION
< ±0.001 % RDG / °C outside 15 … 35 °C
< ±0.0006 % RDG /°F outside 59 … 95 °F
2 × Range
G 1/8” female with G 1/8” male (ISO 228/1) 60° internal cone adapter
NPM160: G 1/8” female
Wetted parts: AISI316 stainless steel, Nitrile rubber.
Pa, hPa, kPa, MPa, mbar, bar, lbf/ft2, psi, ozf/in2, gf/cm2, kgf/cm2, kgf/m2, kp/cm2, at, mmH
2 mH
2 cmH
O, iwc, ftH2O, mmHg, cmHg, mHg, inHg, mmHg(0 °C), inHg(0 °C), mmH
2
2
O(4 °C; 60°F; 68°F/20 °C), inH
+ four (4) user-configurable units
2
O(4 °C; 60°F; 68°F/20 °C), ftH
2
O, cmH
2
O,
O(4 °C; 60°F; 68°F/20 °C),
O(4 °C; 60°F; 68°F/20 °C), torr, atm,
2.5 / second
EXTERNAL PRESSURE MODULES (EXT) STANDARD ACCURACY
EXTERNAL MODULE
EXT200mC-s
EXT2C-s
EXT20C-s
EXT160-s
RANGE
2)
±200 mbar
−1 … 2 bar
−1 … 20 bar
0 … 160 bar
±80 iwc
−14.5 … 30 psi
−14.5 … 300 psi
0 … 2400 psi
RESOLUTION
0.01 mbar
0.01 iwc
0.0001 bar
0.001 psi
0.001 bar
0.01 psi
0.01 bar
0.1 psi
EXTERNAL PRESSURE MODULES (EXT) HIGH ACCURACY
MODULE
Barometric
EXT10mD
EXT100m
EXT400mC
EXT1C
EXT2C
EXT6C
EXT20C
EXT60
EXT100
EXT160
EXT250
EXT600
EXT1000
RANGE
2)
800 … 1200 mbar abs
±10 mbar differential
0 … 100 mbar gauge
±400 mbar
±1 bar
−1 … 2 bar
−1 … 6 bar
−1 … 20 bar
0 … 60 bar
0 … 100 bar
0 … 160 bar
0 … 250 bar
0 … 600 bar
0 … 1000 bar
23.6 … 35.4 inHg a
±4 iwc differential
0 … 40 iwc
±160 iwc
−14.5 … 15 psi
−14.5 … 30 psi
−14.5 … 90 psi
−14.5 … 300 psi
0 … 900 psi
0 … 1500 psi
0 … 2400 psi
0 … 3700 psi
0 … 9000 psi
0 … 15000 psi
1 YEAR UNCERTAINTY (±)
(1
0.05 % RDG + 0.05 % FS
0.05 % FS
0.05 % FS
0.05 % FS
1 YEAR UNCERTAINTY (±)
(1
0.5 mbar (0.015 inHg)
0.05 % Span + 0.1 % RDG
0.025 % FS + 0.025 % RDG
0.02 % FS + 0.025 % RDG
0.015 % FS + 0.025 % RDG
0.01 % FS + 0.025 % RDG
0.01 % FS + 0.025 % RDG
0.01 % FS + 0.025 % RDG
0.01 % FS + 0.025 % RDG
0.01 % FS + 0.025 % RDG
0.01 % FS + 0.025 % RDG
0.015 % FS + 0.025 % RDG
0.015 % FS + 0.025 % RDG
0.015 % FS + 0.025 % RDG
1) Uncertainty includes reference standard uncertainty, hysteresis, non-linearity, repeatability and typical long-term stability for the mentioned period (k=2).
2) The internal pressure module’s range may also be displayed in absolute pressure if a barometric module is used.
3) The MC4 calibrator can hold one internal pressure module and the barometric option.
All external pressure modules (EXT) are also compatible with Beamex MC2, MC5, MC5P and MC6 calibrators.
45
BEAMEX MC4
| FIELD EQUIPMENT
mV MEASUREMENT (T/C-TERMINALS) −25 … 150 mV
RANGE
−25 … 150 mV
FEATURE
Temperature coefficient
Input impedance
Supported units
Display update rate
RESOLUTION
0.001 mV
1 YEAR UNCERTAINTY(±)
1)
0.02 % RDG + 4 µV
SPECIFICATION
< ±0.0015 % RDG / °C outside of 18 … 28 °C
< ±0.0008 % RDG / °F outside of 64.4 … 82.4°F
> 10 MΩ
V, mV, µV
3 / second
46
mV GENERATION (T/C-TERMINALS) −25 … 150 mV
RANGE
−25 … 150 mV
FEATURE
Temperature coefficient
Maximum load current
Load effect
Supported units
RESOLUTION
0.001 mV
1 YEAR UNCERTAINTY (±)
(1
0.02 % RDG + 4 µV
SPECIFICATION
< ±0.0015 % RDG / °C outside of 18 … 28 °C
< ±0.0008 % RDG / °F outside of 64.4 … 82.4°F
5 mA
< 5µV/mA
V, mV, µV
VOLTAGE GENERATION −3 … 12 V
RANGE
±0.25 V
−3 … −0.25 V
0.25 … 12 V
RESOLUTION
0.01 mV
0.1 mV
0.1 mV
FEATURE
Temperature coefficient
Maximum load current
Load effect
Supported units
1 YEAR UNCERTAINTY (±) (1
0.02 % RDG + 0.1 mV
0.02 % RDG + 0.1 mV
0.02 % RDG + 0.1 mV
SPECIFICATION
< ±0.0015 % RDG / °C outside of 18 … 28 °C
< ±0.0008 % RDG / °F outside of 64.4 … 82.4°F
5 mA
< 50 µV/mA
V, mV, µV
mA GENERATION (SOURCE/SINk) 0 … 25 mA
RANGE
0 … 25 mA
RESOLUTION
0.0001 mA
FEATURE
Temperature coefficient
Max load impedance (source)
Max loop voltage (sink)
Supported units
1 YEAR UNCERTAINTY (±)
(1
0.02 % RDG + 1.5 µA
SPECIFICATION
< ±0.0015 % RDG / °C outside of 18 … 28 °C
< ±0.0008 % RDG / °F outside of 64.4 … 82.4°F
750 Ω (0 … 20 mA), 600 Ω (20 … 25 mA)
60 V mA, µA
1) Uncertainty includes reference standard uncertainty, hysteresis, non-linearity, repeatability and typical long-term stability for the mentioned period (k=2).
FIELD EQUIPMENT | BEAMEX MC4
RESISTANCE MEASUREMENT 0 … 4000 Ω
RANGE
0 … 250 Ω
250 … 2650 Ω
2650 … 4000 Ω
RESOLUTION
1 mΩ
10 mΩ
100 mΩ
FEATURE
Temperature coefficient
Measurement current
Supported units
Display update rate
1 YEAR UNCERTAINTY (±)
(1
4-wire connection:
0.02 % RDG + 3.5 mΩ
3-wire connection:
0.02 % RDG + 13.5 mΩ
SPECIFICATION
< ±0.0015 % RDG / °C outside of 18 … 28 °C
< ±0.0008 % RDG / °F outside of 64.4 … 82.4°F
Pulsed, bi-directional 1 mA (0 … 500 Ω), 0.2 mA (>500 Ω)
Ω , kΩ
3 / second
RESISTANCE SIMULATION 0 … 4000 Ω
RANGE
0 … 400 Ω
400 … 4000 Ω
RESOLUTION
10 mΩ
100 mΩ
1 YEAR UNCERTAINTY (±)
(1
0.04 % RDG or 30 mΩ (whichever is greater)
0.04 % RDG or 30 mΩ (whichever is greater)
FEATURE
Temperature coefficient
Maximum resistance excitation current
Settling time (pulsed currents)
Supported units
SPECIFICATION
< ±0.0015 % RDG / °C outside of 18 … 28 °C
< ±0.0008 % RDG / °F outside of 64.4 … 82.4°F
5 mA (0 … 650 Ω) Iexc × Rsim < 3.25 V (650 … 4000 Ω)
1 ms
Ω, kΩ
Specification valid with an exitation current >0.2 mA (0 … 400 ohm), >0.1 mA (400 … 4000 ohm).
FREqUENCY GENERATION 0.0005 … 10 000 Hz
RANGE
0.0005 … 0.5 Hz
0.5 … 5 Hz
5 … 50 Hz
50 … 500 Hz
500 … 5000 Hz
5000 … 10000 Hz
RESOLUTION
0.000001 Hz
0.00001 Hz
0.0001 Hz
0.001 Hz
0.01 Hz
0.1 Hz
FEATURE
Temperature coefficient
Maximum load current
Output amplitude positive square wave
Output amplitude symmetric square wave
Duty cycle
Supported units
Jitter
1 YEAR UNCERTAINTY (±)
(1
0.01 % RDG
0.01 % RDG
0.01 % RDG
0.01 % RDG
0.01 % RDG
0.01 % RDG
SPECIFICATION
Specification valid from −10 to 50 °C (14 … 122°F)
5 mA
0 … 12 Vpp ±(0.2 V+5 %)
0 … 6 Vpp ±(0.2 V+5 %)
1 … 99 % (0.0009 … 500 Hz), high / low time: min 25µs, max 1165 s
Hz, kHz, cph, cpm, 1/Hz (s), 1/kHz (ms), 1/MHz (µs)
< 0.28 µs
PULSE GENERATION 0 … 9 999 999 PULSES
FEATURE
Range
Resolution
Maximum load current
Output amplitude positive pulse
Output amplitude symmetric pulse
Pulse frequency
Duty cycle
SPECIFICATION
0 to 9 999 999 pulses
1 pulse
5 mA
0 … 12 Vpp ±(0.2 V+5 %)
0 … 6 Vpp ±(0.2 V+5 %)
0.0005 … 10 000 Hz
1 … 99 % (0.0009 … 500 Hz), high / low time: min 25µs, max 1165 s
1) Uncertainty includes reference standard uncertainty, hysteresis, non-linearity, repeatability and typical long-term stability for the mentioned period. (k=2).
47
BEAMEX MC4
| FIELD EQUIPMENT
48
THERMOCOUPLE MEASUREMENT AND SIMULATION
Thermocouple types available as standard
TYPE RANGE (°C)
B
2)
0 … 1820
R 2)
S
2)
E 2)
J
2)
K
2)
N
2)
T
2)
U 4)
L
4)
C 5)
G 6)
D
5)
−50 … 1768
−50 … 1768
−270 … 1000
−210 … 1200
−270 … 1372
−270 … 1300
−270 … 400
−200 … 600
−200 … 900
0 … 2315
0 … 2315
0 … 2315
RANGE (°C)
0 … 200
200 … 500
500 … 800
800 … 1820
−200 … 0
0 … 600
−200 … 0
0 … 900
0 … 900
900 … 2000
2000 … 2315
0 … 70
70 … 200
200 … 1600
1600 … 2000
2000 … 2315
0 … 1000
1000 … 2000
2000 … 2315
−50 … 0
0 … 50
50 … 1400
1400 … 1768
−50 … 0
0 … 50
50 … 1500
1500 … 1768
−270 … −200
−200 … 0
0 … 600
600 … 1000
−210 … −200
−200 … 0
0 … 1200
−270 … −200
−200 … 0
0 … 1000
1000 … 1372
−270 … −200
−200 … −100
−100 … 0
0 … 750
750 … 1300
−270 … −250
−250 … −200
−200 … 0
0 … 400
FEATURE
Resolution
Temperature coefficient
Input impedance
Supported units
Display update rate
Maximum load current
Load effect
MEASUREMENT
0.01 °C
< ±0.0015 %
of thermovoltage / °C outside of 18 … 28 °C
< ±0.0008 %
of thermovoltage / °F outside of 64.4 … 82.4°F
>10 MΩ
°C, °F, K
3 / second
–
–
1 YEAR UNCERTAINTY (±) 1)
3)
2.0 °C
0.8 °C
0.6 °C
1.0 °C
0.7 °C
0.5 °C
0.6 °C
1.0 °C
0.7 °C
0.6 °C
0.7 °C
3)
0.07 °C + 0.08 % RDG
0.07 °C + 0.015 % RDG
0.026 % RDG
3)
0.08 °C + 0.07 % RDG
0.08 °C + 0.02 % RDG
3)
0.1 °C + 0.1 % RDG
0.1 °C + 0.02 % RDG
0.03 % RDG
3)
0.2 % RDG
0.15 °C + 0.05 % RDG
0.15 °C + 0.01 % RDG
0.03 % RDG
3)
0.7 °C
0.1 °C + 0.1 % RDG
0.1 °C + 0.01 % RDG
0.15 °C + 0.1 % RDG
0.15 °C + 0.01 % RDG
0.13 °C + 0.07 % RDG
0.13 °C + 0.02 % RDG
0.4 °C
0.045 % RDG
1.2 °C
3)
1.0 °C
0.5 °C
0.7 °C
1.0 °C
0.4 °C
0.04 % RDG
1.2 °C
SIMULATION
0.01 °C
< ±0.0015 %
of thermovoltage / °C outside of 18 … 28 °C
< ±0.0008 %
of thermovoltage / °F outside of 64.4 … 82.4°F
–
°C, °F, K
–
5 mA
< 5 µV/mA
FIELD EQUIPMENT | BEAMEX MC4
INTERNAL REFERENCE JUNCTION
RANGE (°C)
−10 … 50 °C
1 YEAR UNCERTAINTY
±0.25 °C
1) Uncertainty includes reference standard uncertainty, hysteresis, non-linearity, repeatability and typical long-term stability for the mentioned period (k=2).
Uncertainty does not include reference junction uncertainty.
2) IEC 584, NIST MN 175, BS 4937, ANSI MC96.1
3) ±0.02 % of thermovoltage + 4 µV
4) DIN 43710
5) ASTM E 988 - 96
6) ASTM E 1751 - 95e1
RTD MEASUREMENT AND SIMULATION
SENSOR TYPE
Pt 50 … 1000
Ni 100
Ni 120
Cu10
FEATURE
Temperature coefficient
RANGE
Measurement current
Maximum resistance excitation current
Supported units
Display update rate
Settling time
(pulsed currents)
−200 … 0 °C
0 … 850 °C
−60 … 180 °C
−80 … 260 °C
−200 … 260 °C
RESOLUTION
0.01 °C
0.01 °C
0.01 °C
0.01 °C
MEASUREMENT
< ±0.0015 %
of resistance / °C outside of 18 … 28 °C
< ±0.0008 %
of resistance / °F outside of 64.4 … 82.4°F
Pulsed, 1 mA (0..500 Ω),
0.2 mA (>500 Ω)
–
°C, °F, K
3 / second
1 ms
MEASUREMENT 1 YEAR
UNCERTAINTY (±)
1)
0.06 °C
0.06 °C + 0.025 % RDG
0.06 °C
0.06 °C
0.2 °C
SIMULATION 1 YEAR
UNCERTAINTY (±)
1) 2)
0.10 °C
0.10 °C + 0.025 % RDG
0.12 °C
0.12 °C
0.8 °C
SIMULATION
< ±0.0015 %
of resistance / °C outside of 18 … 28 °C
< ±0.0008 %
of resistance / °F outside of 64.4 … 82.4°F
–
–
–
5 mA (0 … 650 Ω)
Iexc × Rsim < 3.25 V (650 … 4000 Ω)
°C, °F, K
RTD TYPES AVAILABLE AS STANDARD
Pt50 (385)
Pt100 (385)
Pt200 (385)
Pt400 (385)
Pt500 (385)
Pt1000 (385)
Pt100 (3926)
Pt100 (391)
Pt100 (375)
Pt100 (3923)
Ni100 (618)
Ni120 (672)
Cu10 (427)
1) Uncertainty includes reference standard uncertainty, hysteresis, non-linearity, repeatability and typical long-term stability for the mentioned period (k=2).
2) Specification valid with an excitation current >0.2 mA (0 … 400 Ω), >0.1 mA (400 … 4000 Ω).
MC4 supports Callendar van Dusen correction coefficients for PRT sensors to compensate sensor error.
STANDARD ACCESSORIES
• User guide
• Accredited calibration certificate
• Internal rechargeable NiMH battery pack + battery charger
• Test leads and clips
• USB cable
• Adapter pressure connector – from G1/8" female to G 1/8" male with 60° internal cone (included in models with internal pressure module)
OPTIONAL ACCESSORIES
• Pressure T-hose
• Soft carrying case
• Connection cable for external pressure modules
• Dry battery cartridge
• Calibration hand-pumps
49
BEAMEX MC4
| FIELD EQUIPMENT
SUMMARY
Beamex MC4
DOCUMENTING PROCESS CALIBRATOR
50
The Beamex MC4 is a documenting process calibrator.
Instrument data can be sent from a computer to the MC4, and calibration results can be uploaded from the MC4 to a computer using the Beamex CMX calibration software.
Being a multifunctional calibrator, the MC4 is suitable for calibrating various process parameters, such as pressure, temperature and electrical signals. High accuracy is one of the important features of the MC4. A standard feature of the
MC4 includes an accredited calibration certificate as proof of its accuracy.
Communication with calibration software
Using the MC4 together with calibration software provides you with a complete documenting calibration system that produces calibration certificates automatically.
all-in-one functionality
The MC4 is a versatile calibrator with many different functions. There’s no need to take several different measurement devices to the field –
MC4 does the job.
accuracy guaranteed
The MC4 is a highly accurate process calibrator. As a proof of this, each calibrator is delivered with a traceable, accredited calibration certificate.
Calibration is quick and easy
The large graphical display, menu-based multilingual user interface, and full numerical keyboard make the MC4 quick and easy to use.
Main features
Automated and documented calibrations quickly and easily
Calibration capabilities for pressure, temperature, electrical and frequency signals
Compact size and design
Documenting – communicates with Beamex calibration software
FIELD EQUIPMENT | BEAMEX MC4