Model 664

User’s Guide

Version 1.1

Model 664 User’s Guide

Copyright © 2008 GSE. All rights reserved.

Published by: GSE

Information in this Technical Reference Manual is subject to change without notice due to correction or enhancement. The information described in this manual is solely the property of

GSE. No part of this manual may be reproduced or transmitted in any form or by any means, electronic or mechanical, including photocopying and recording and sold for any monetary figure without the express written permission of GSE.

GSE

1525 Fairlane Circle

Allen Park, MI 48101

U.S.A. www.gse-inc.com

T

ABLE OF

C

ONTENTS

CHAPTER 1: INTRODUCTION .................................................... 3

U

NPACKING THE

S

CALE

...................................................................... 3

I

NSTALLATION

..................................................................................... 3

E

NCLOSURE

......................................................................................... 4

K

EYPAD

.............................................................................................. 4

S

ETUP

M

ENU

....................................................................................... 5

CHAPTER 2: OPERATION ............................................................... 7

C

OUNTING

M

ETHODS

.......................................................................... 7

Save Or Recall An NDC ................................................................. 7

Quick Sample .................................................................................. 8

M

ANUAL

NDC E

NTRY

...................................................................... 10

C

LEAR A

D

RUG FROM THE

D

ATABASE

............................................... 10

A

DD

N

EW

D

RUG

............................................................................... 11

C

OMPOUNDING

.................................................................................. 12

P

RESCRIPTION

V

ALIDATION

.............................................................. 13

CHAPTER 3: LEGAL FOR TRADE ............................................... 15

NTEP ................................................................................................ 15

Sealing And Audit Trails .............................................................. 15

CHAPTER 4: CALIBRATION ........................................................ 17

O

UT OF

T

OLERANCE

.......................................................................... 17

C

ALIBRATION

C

HECK

........................................................................ 17

F

ULL

C

ALIBRATION

........................................................................... 18

General Notes on Calibration ...................................................... 18

C

ALIBRATION

E

RROR

M

ESSAGES

...................................................... 19

CHAPTER 5: ERROR CODES / MESSAGES ............................... 21

Operational Mode Error Codes ................................................... 21

Setup Mode Error Codes .............................................................. 21

Hardware Problem Error Codes .................................................. 22

i

Calibration Error Codes ............................................................... 22

General Error Codes .................................................................... 23

Setup Error Codes ........................................................................ 24

Communications Error Messages ................................................. 25

Miscellaneous Messages ............................................................... 26

ii

Chapter 1: I

NTRODUCTION

Unpacking the Scale

CAUTION! The weigh module is a fragile precision mechanism. Do not turn the scale upside down allowing the load cell stem (or weighing surface if installed) to touch any surface, and do not put pressure on the load cell stem. Pressure on the load cell stem could damage the load cell and void the warranty. (The stem is the post protruding from the top of your scale that supports the weighing surface.)

Unpack all items carefully. If possible, do not discard the packaging material. Only

GSE packaging material may be used to ship this product. Shipping this product without the GSE packaging material will void the warranty.

Installation

Note

– Do not plug the power cord into the scale until step 6!

1.

Place the scale on an unobstructed surface.

2.

Level the Model 664 by referring to the level bubble located on the top of the scale, under the weighing platter. Adjust the leveling feet on the front corners until the bubble is in the center.

3.

Gently place the stainless steel weigh platter and it’s support structure onto the load cell stem located on the top of the scale.

4.

Unfold the scanner gooseneck and swivel the top holder 180

°

.

5.

Plug the scanner into the connecter marked “Barcode Scanner” located in the back of the Model 664. (Note: Be careful not to use the other similar connector!). Secure the cable to the Model 664 by tightening the screw posts on each side of the connector. Place the scanner in the holder at the top of the scanner stand.

6.

Insert the white tip at the end of the AC power module cable into the hole labeled “Power” on the back of the Model 664, keeping the notched edge facing upward. The black over-molding should engage into the rear panel hole slightly.

7.

Plug the AC power supply into a standard 110 V

AC

wall socket.

3

The display area should illuminate and within a few seconds and the

Scan NDC

screen shall be displayed. Several beeps from the main unit and barcode scanner during power-up are normal.

Enclosure

The Model 664 enclosure is made of strong ABS plastic material. To clean, use a slightly damp cloth.

Keypad

Refer to Figure 1-1 for the keypad layout and Table 1-1 for explanations of the functions of the keys.

CLASS II, d=0.001g, e=0.01g

Temperature Limit:

10°C to 30°C (50°F to 86°F)

MPW=0.03g, MSS= 10 pieces

Counting Feature For Prescription Filling Only

ZERO

CAL

DATA

BASE

LOGOUT

WEIGH

COUNT

QUICK

SAMPLE

ENTER

YES

7

4

1

0

8

5

2

.

9

6

3

CANCEL

NO

Figure 1-1: Model 664 Keypad

Table 1-1: Keypad Function Explanations

Key Function

CAL

DATABASE

Starts the calibration process

Add or clear a drug from memory

QUICK SAMPLE Initiates the pill weight sampling process

WEIGH/COUNT

ENTER/YES

Switches between the weigh mode (used for compounding) and the Scan NDC label screens

1) Enter or 2) Answers YES to a question

CANCEL/NO

0 – 9, .

1) Aborts an operation or 2) Answers NO to a question

Used to enter numeric values

Reference Page

N/A

17

N/A

N/A

7

10

N/A

N/A

N/A

4

Setup Menu

The setup menu provides access to view and change settings such as display contrast, audit trail etc. Refer to Table 1-2 for settings and explanations.

To access the setup menu, press

[ENTER]

when the GSE splash screen appears during power-up to gain access to the setup menu. The setup menu will appear after the Model 664 finishes the boot up process.

Figure 1-2: Setup Menu Screen

To view or change the settings, press the number associated with the desired item.

Then press

[CANCEL/NO]

to make a change and press

[ENTER]

to save the change. Press

[ENTER]

again to exit setup.

Table 1-2: Setup Menu Explanation

Number Setting Description

1

2

4

Contrast

Weigh Increment *

Network

Set the optimum viewing angle of the display

Set the displayed resolution to either 0.01 grams or

0.001 grams. See Setup Note below.

Automatically zeroes weight of scoop on scale when

Count Drug Screen appears.

Enable/disable network capability.

Only appears when

Network

(choice 4) is enabled.

View and change scale IP address, Host IP, Port,

Mask, and Mac ID.

The PSC QS2500 is automatically configured. If other is selected, the PSC QS2500 will not be configured.

Scan a label and verify the data displayed 7 Scanner Test

9 Validation

0 Not used at this time

View the calibration and setup counters. See page 15 for more details.

Compare the prescription label to the stock bottle to verify they match.

*

SET UP NOTE: If the scale is being sealed as “Legal for trade”, change the weigh increment to the desired setting before sealing. Each time the weigh increment is changed, the setup audit trail counter will increase which may require the scale to be re-inspected.

Or, if a physical seal is used, this parameter cannot be changed unless the program switch is set to “YES”. Access to the switch will require breaking of the seal and possibly reinspection.

5

This page is intentionally left blank

6

Chapter 2: O

PERATION

After the Model 664 goes through the power up sequence,

Scan NDC

will be displayed. From this mode it is possible to count or compound. Follow the instructions in this chapter for different methods of counting and compounding.

Counting Methods

S

AVE

O

R

R

ECALL

A

N

NDC

Scan a stock bottle to recall the drug from the database or store a new drug to the database.

1.

Scan the stock bottle. If the NDC is not found in the database, a screen will prompt the user to select from one of the displayed NDC numbers. If the NDC number was found, the Model 664 will proceed to step 4.

2.

Compare the printed NDC number found on the stock bottle label to the ones displayed. Select the correct NDC number by pressing the digit associated with that number. See

Select the Correct NDC Number

on page 8 for more details.

3.

The display will prompt

“Save as new NDC?”

. If the number is correct, place the counting boat on the scale platter and press

[ENTER/YES]

to save it to the database.

4.

If the NDC was saved the count screen will appear. Add 30 pieces and press

[ENTER/YES]

or key in a desired amount, add that amount, and press

[ENTER/YES]

.

7

The NDC and average piece weight are now stored to the database. Now each time the same NDC number is scanned, the average piece weight will be recalled and the

Model 664 will go right to the count screen.

NOTE:

Press

[CANCEL/NO]

at any time to cancel and return to the

Scan NDC

screen.

Select the Correct NDC Number

The barcode on the stock bottle is a 10-digit number. The NDC number needs to be an 11-digit number. There are three fields within the NDC number. See the table below for the required digits per field. A leading zero or zeros will be added to the field that does not match.

First Field

Second Field

Third Field

Number of Digits

5

4

2

Example: The NDC number printed on the stock bottle is 62037-871-30.

6 2 0 3 7 –

8 7 1

– 3 0

5 digits 3 digits 2 digits

In this case the second field would have a leading zero since it only has 3 digits instead of 4. The correct NDC number choice would be:

[2]

62037-0871-30.

Q

UICK

S

AMPLE

The quick sample mode offers a fast way to count without database recall or scanning a label first.

Sample 30 pieces:

1.

If a container is being used to count pieces in, place it on the scale platter now.

8

2.

Press

[QUICK SAMPLE]

to enter the count mode. The scale will zero and prompt

Add 30

. The display may read

Must Sampl

if an average piece weight has not been established or power has been cycled.

3.

Place a sample of 30 pieces on the scale platter or in a container and press

[ENTER]

.

4.

Continue adding pieces to count.

NOTE:

Press

[CANCEL/NO]

at any time to abort the quick count mode.

Sample A Known Amount:

1.

If a container is being used to count pieces in, place it on the scale platter now.

2.

Press

[QUICK SAMPLE]

to enter the count mode. The scale will zero and prompt

Add 30

. The display may read

Must Sampl

if an average piece weight has not been established or power has been cycled.

3.

Place the sample on the scale platter or in the container. Key in the known sample size and press

[ENTER]

. Example:

[9] [0] [ENTER]

4.

Continue adding pieces to count.

The keypad of the

Model 664

is marked with MSS: 10 and MPW: 0.03 g. The

MSS is the minimum sample size required in Legal for Trade applications. The

MPW is the minimum piece weight required in Legal for Trade applications. Each individual piece weight (MPW) must be at least 30 mg and the minimum sample size (MSS) must be equal to or greater than 10 pieces. The total piece weight will be a minimum of 300 mg.

If the above conditions are not met, an error will be displayed. Follow the instructions on the display to continue.

9

Manual NDC Entry

In some cases a stock bottle will not have a barcode to scan. However, the NDC number is printed on the label. Follow the steps below to enter a NDC number and save it to the database.

1.

Locate the NDC number on the stock bottle. In most cases, the stock bottle will be missing a digit in at least one of the fields. If this is the case, key in a zero or zeroes at the beginning of the field and then the number in that field.

See the table below for the required digits per field. See

Select the Correct

NDC Number

on page 8 for more details.

First Field

Second Field

Third Field

Number of Digits

5

4

2

2.

Enter the NDC number for each field.

NOTE: If the required number of digits is entered in a field, the cursor will proceed to the next field and pressing [ENTER] is not necessary.

NOTE: If a mistake was made during entry, pressing

[CANCEL]

will clear the entire field.

3.

Review the number entered on the display. If the number is correct press

[ENTER]

to save the NDC to the database.

Clear a Drug from the Database

1.

Scan the stock bottle from the

Scan NDC

screen.

2.

If the NDC number was found in the database, the count screen will be displayed.

3.

Press the

[DATABASE]

key.

10

5.

Press

[1]

to delete the drug.

4.

The next screen will affirm whether you want to delete the drug. Press

[ENTER/YES]

to delete the selected drug or

[CANCEL/NO]

to cancel.

5.

If

[ENTER/YES]

was pressed,

“Drug Deleted”

will be shown for a few seconds and then the display will return to

Scan NDC.

6.

If

[CANCEL/NO]

was pressed, the display will return to the counting mode.

Use the

[CANCEL]

SELECTIONS

menu. or

[DATABASE]

key to exit out of the

DATABASE

Add New Drug

This will only be necessary if two NDC number are similar or the wrong NDC number was chosen previously.

1.

Scan the stock bottle from the

Scan NDC

screen. If the NDC number was found in the database, the count screen will be displayed and the drug does not need to be added. Otherwise, go to step

2.

Press the

[DATABASE]

key.

3.

Press

[2]

to add new drug.

11

4.

Select the correct NDC number for the choices displayed.

5.

If the NDC was saved the count screen will appear. Add 30 pieces and press

[ENTER/YES]

or key in a desired amount, add that amount, and press

[ENTER/YES]

.

Compounding

The weigh increment for compounding is set to 310 x 0.001 grams as a default. It is possible to change the value to be 310 x 0.01 grams. Refer to page 5 for further instructions on changing the weigh increment. If the scale is set to 310 x 0.001 grams, the last digit will be inverted. This inverted digit is required by NTEP. See page 15 for more details on legal for trade requirements.

If the last digit is fluctuating up and down, this is most likely caused by air currents or vibrations being transmitted to the scale through the countertop.

Either eliminate these external influences or change the weight increment to show only two digits after the decimal using the setup menu, as described on page 5.

If the scale is being sealed as legal for trade, change the weigh increment before sealing. Each time the weigh increment is changed, the setup audit trail counter will increase. If a physical seal is used, the program switch will need to be moved to allow changing the increment size.

1.

To access the compounding mode from the Scan NDC Label screen press

[WEIGH/COUNT]

2.

If desired place an empty container on the scale. Then press zero to re-zero the scale. Once the displayed weight has re-zeroed, you can proceed. This may take a while if there are any vibrations or air currents.

3.

Add the ingredient until the desired weight is achieved.

4.

Press

[WEIGH/COUNT]

to return to the Scan NDC Label screen.

12

Prescription Validation

Protect the patient by validating the medication dispensed is correct. Both prescription label and stock bottle label NDC identifiers must match before a prescription can be counted and filled.

Scan the prescription label from the

“Scan Label”

prompt. Then scan the stock bottle from the

Scan NDC”

prompt.

If the prescription and stock bottle labels match, the scale will permit the prescription to be filled. If the labels do not match, the scale displays a message that the incorrect stock bottle was selected.

13

This page is intentionally left blank

14

Chapter 3: L

EGAL

F

OR

T

RADE

NTEP

The National Type Evaluation Program (NTEP) is a widely accepted weights and measures standard in the United States. Most states abide by some or all of the requirements set forth by NTEP. A complete list of these regulations is available in the “Handbook 44” publication distributed by the National Institute of Standards and

Technology (NIST). For more information on this and other NIST publications, visit their web site at http://www.nist.gov.

The Model 664 NTEP Certificate of Conformance (C.O.C.) is 04-094

S

EALING

A

ND

A

UDIT

T

RAILS

Most legal-for-trade installations will require the unit to be sealed. A sealed unit cannot be accessed for setup or calibration changes without breaking a physical seal or incrementing an event counter, thus providing evidence that the unit has been tampered with. The

Model 664

offers two types of sealing provisions:

Physical seal used in conjunction with an program switch

Two-event audit trail counter

Check with your local weights and measures authority to determine which method(s) are required.

Physical Seal

The most common sealing method is a lead-wire seal. The

Model 664

has two tamper-proof screws installed for sealing the bottom plate to the top enclosure. A lead-wire seal can be applied by passing the lead-wire seal wire through a hole in the two tamperproof screws, thus preventing the screws from being removed without breaking the seal. Refer to Figure 3-1 for an illustration.

Before applying a wire seal, be sure to move the program switch on the back panel to the ‘NO’ position to prevent access to the setup and calibration modes.

15

Figure 3-1: Physical Seal

Audit Trail Parameters

Two separate incrementing, non-resettable audit trail parameters are used to indicate changes to various parameters:

Calibration

Setup

An audit trail counter will increment only once upon exiting the setup mode and saving changes regardless of how many settings were changed. Each audit trail counter will increment to 99999 before beginning again at 00001.

Accessing Audit Trails

The audit trails may be accessed from the weigh mode. An access code is not required to view audit trail parameters.

1.

Cycle power on the

Model 664

unit by removing and reapplying power.

2.

Press the

[ENTER]

key at the GSE Scale Systems splash screen. This will gain access to the setup menu.

3.

Press the corresponding number key to Audit Trail in the setup menu.

4.

The display will show the

Calibration

and

Setup Mode

audit trail parameters for approximately five seconds. The display will return to the setup menu after the five second period is over. Repeat step 3 as many times as needed to view the audit trail parameters.

5.

Press

[ENTER]

to return to “

Scan NDC”

.

16

Chapter 4: C

ALIBRATION

The Model 664 only needs to be calibrated if it is out of tolerance. It is possible to check the calibration tolerance with the

Calibration Check

method by pressing the

[CAL]

key once. Refer to the

Calibration Check

section below for details.

Refer to the

Full Calibration

section if the calibration is out of tolerance. To access the

Full Calibration

mode, press and hold the

[CAL]

key.

Out of Tolerance

The

“OUT OF TOLERANCE”

message will be displayed after a

Calibration Check

is performed and the calibration values are outside the tolerance band. Make sure the scale is level and perform another

Calibration Check

. Refer to the

Calibration Check

section below for details.

If the

“OUT OF TOLERANCE”

message is still displayed, a full calibration will be necessary. Refer to the

Full Calibration

section for details on calibrating the Model

664.

Calibration Check

The calibration check feature allows the calibration accuracy to be checked without going through the full calibration routine and possibly incrementing the calibration audit trail counter.

To Check Calibration:

1.

Make sure the scale is level by referring to the level bubble located on the top of the scale, under the weighing platter. Adjust the leveling feet on the front corners until the bubble is in the center.

5.

Without a load on the scale platter, press the

[CAL]

key.

6.

The unit will automatically perform a zero function. After the zero reference is established the display will prompt

Add 200 g

.

7.

Place a certified calibrated 200 gram weight on the scale platter.

8.

The unit will determine whether the weight is within an acceptable tolerance.

The display will show

“GOOD”

or

“OUT OF TOLERANCE”.

17

If

“GOOD”

is displayed, the calibration is within tolerance and no further action is required.

If

“OUT OF TOLERANCE”

is displayed, a full calibration needs to be performed.

9.

Remove the 200 gram test weight to return to the “

SCAN NDC”

prompt.

Full Calibration

The full calibration only needs to be performed when the Model 664 calibration is out of tolerance.

To Perform a Calibration:

1.

Make sure the scale is level by referring to the level bubble located on the top of the scale, under the weighing platter. Adjust the leveling feet on the front corners until the bubble is in the center.

2.

Remove all weight from the scale platter.

3.

Press and hold the

[CAL]

key for 3 seconds.

4.

The unit will automatically perform a zero function. After the zero reference is established the display will prompt

Add 200 g

.

5.

Place a certified calibrated 200 gram weight on the scale platter.

6.

After calibration is complete the calibration is saved and the display will return to the gross mode.

Please be aware that the calibration audit trail will be incremented every time a full calibration is performed. See page 15 for more details.

If the

Model 664

has a physical seal the full calibration will be locked out. Contact a qualified service technician to calibrate your

Model 664

unit. See page 15 for more details.

G

ENERAL

N

OTES ON

C

ALIBRATION

Pressing

[CANCEL]

at any point will exit the calibration mode.

The digital filter is automatically set to 4 seconds during full calibration.

A motion delay is enforced during zero and span calibration.

New calibration is not permanently saved until the calibration mode is exited.

18

Calibration Error Messages

If calibration problems occur despite an apparently proper configuration, carefully inspect all hardware. Moisture, obstructions, defective or worn components, improper grounding techniques, and an unsuitable environment are factors that could contribute to calibration problems. Refer to the

Full Calibration

section outlined on page18 for proper calibration technique.

If problems persist contact your GSE service representative.

19

This page is intentionally left blank

20

Chapter 5: E

RROR

C

ODES

/ M

ESSAGES

This section describes all error codes generated by the

Model 664

. Most error codes show a two-digit code reference along with a short text message.

Possible causes and remedies are described for each error.

O

PERATIONAL

M

ODE

E

RROR

C

ODES

Information Parameter

Code02

Under

Load!

Description

Input signal less than negative full scale. If this is due to excessive loading, reduce the load. Otherwise check the load cell connections.

Input signal is greater than positive full scale. Use same check as for underload.

CoDE03

Over-

Load!

CoDE04

#>

Dsply

Number to be displayed will not fit within 6 digits. This will not normally occur for the Gross Weight, but may result while displaying the accumulated totals if the amount exceeds 999,999. Either clear the totals or settle for only being able to transmit the totals.

An attempt was made to zero out more than allowed.

CoDE05 Zero>

Max.!

CoDE08 Check

Conn

The signal into the A/D is greater than +/- 2 times the expected full scale signal.

S

ETUP

M

ODE

E

RROR

C

ODES

Information Parameter

CoDE10 Entry

>Max!

CoDE14

Must

Keyin

CoDE16

CHECK

JUMPR

Description

An entry was made which had more characters than allowed.

The choice for the current parameter must be keyed in.

A programming operation was attempted when the program switch is set to “NO”.

21

H

ARDWARE

P

ROBLEM

E

RROR

C

ODES

Information Parameter

CoDE18

CoDE19

CoDE20

Description

The accumulative

total

buffer size for both the TX and RX buffers of all COMM ports exceeds the reserved storage capacity. If entries to the “buffer size” parameters (P207-

P208) exceed this total, this error message will be displayed.

Certain combinations of protocol are not available. The protocol combination selections are in P201, P202 and

P203. This error occurs if an illegal protocol combination is selected.

The A/D calibration data is corrupt. When the information modes are accessed (P61100) this message will be displayed for 1 second and P61113 – P61116 are defaulted to a factor of 1. This will also happen on power-up if the check-sum for the A/D data is corrupted.

Error

reading

data from the FRAM.

CoDE21

CoDE22

CoDE23

CoDE24

CoDE25

CoDE26

CoDE27

CoDE29

Error

writing

data to the FRAM.

Supplementary error message for above errors.

The setup being attempted requires more FRAM than is currently installed.

Upon power-up the indicator has

not

found the proper codes. Therefore, all parameters have been reset to factory default values.

The stored data has a checksum error. Check all parameters or re-load setup.

The indicator cannot use the FRAM for data storage and is attempting to power-up again to cure the problem.

The FRAM is corrupted in the PIN section. Check FRAM for problems. The access code is then defaulted to the manufacturer (GSE) access code. Also refer to Error 11.

C

ALIBRATION

E

RROR

C

ODES

Information Parameter

CoDE30 F.S.>

MAX!

CoDE31

F.S.<

.1mVv

Description

The entered calibration weight, together with the currently applied signal, indicates that the full scale signal will be greater than the allowed maximum of the controller.

The entered calibration weight, together with the currently applied signal, indicates that the full scale signal will be less than the allowed minimum of the controller.

22

Information Parameter

CoDE32

ADD

MORE!

Description

The applied weight during calibration was less than 0.1% of capacity. More weight than this is required.

CoDE33

ReCAL

Req'd

The just completed calibration is insufficient to guarantee accurate results due to either the cal weight being less than 5% of capacity or this was the first calibration.

G

ENERAL

E

RROR

C

ODES

Information Parameter

CoDE42

Check

Setup

Description

A sample operation was performed with a sample size of “0” at parameter P34.

CoDE43

CoDE48 dbNOT

SETUP

E2Typ

Error

An attempt to access an undefined database occurred.

This error could also indicate that the database option has not yet been initialized at P65010.

Both FRAM and standard EERPOM are installed. Memory types cannot be intermixed. Power down and install one or two FRAM chips

or

one or two FRAM chips and power up.

The sample placed on the platform is too small to accurately compute the piece weight. Increase the sample size.

CoDE51

Too

Small

CoDE52

Can't

Count

There is an insufficient quantity on the platform to perform an accurate count.

CoDE53

CoDE54

Accy<

Req'D

Scale

Disbl

The accuracy is less than required. The accuracy requirements specified at P183 has not been achieved.

Increase the sample size.

An attempt to select a

disabled

scale was made. Make sure the intended scale is enabled and properly configured beginning at P108.

CoDE56 dbNOT

INIT

The database option has not initialized at 65010. This should not happen unless the indicator is powered down in the middle of a optional RAM test .

CoDE57 dbOpt

Error

The FRAM database is corrupt (valid signature, invalid checksum).

CoDE58 NewDb

Opt'n

Database format has changed. This is most likely the result of upgrading to a new FLASH with a new database definition.

CoDE60

CoDE61

CoDE62

New

FLASH

Need

Code

Comm#

Error

A new FLASH file has been uploaded via the ReFlash or

BDM flash utilities in order to upgrade the firmware. Press any key to acknowledge.

The setup mode access code has not been entered prior to attempting to change the configuration of a setup parameter. Key in the access code (i.e.

100 [LOG OUT]

23640 [ENTER]

).

An entry error occurred at the “EnterCOMM#” prompt. This message appears during a setup download (P64000,

P64001), database download, database print, etc. if the

23

Information Parameter

CoDE71 TxNot

Exist

Description

COMM port number was omitted or invalid.

The specified custom transmit does not exist. An invalid custom transmit was entered at the “Which Tx#?” prompt.

Make sure the intended custom transmit is configured beginning at P989.

S

ETUP

E

RROR

C

ODES

Information Parameter

CoDE95

CoDE95

CoDE95

SyErr

NvRam

SyErr

HSR00

SyErr

Typ04

Code96

CoDE99

P___

Description

The FRAM size is too small to allocate the database requested.

An error occurred at startup or during operation.

An error occurred at startup.

Erase

Boot!

Can't

Set!

Invld

Mode!

Mode

<100!

Press

Enter

An attempt was made to enter an instrument serial number or board serial number with the flash already programmed.

An attempt to enter a value for a parameter which is not field changeable, such as the serial numbers or the audit trail counter results in this message.

An attempt was made to access a non-existent parameter. Key in a valid parameter and press

[LOG

OUT]

.

An attempt was made to access a weigh mode parameter from within the setup mode. Valid setup mode parameters contain three or more digits.

The

[LOG OUT]

key was pressed at a “pick instance” prompt. You must press

[ENTER]

to select an instance, or

[CANCEL]

to abort the instance entry.

This is not an error. Press

[ENTER]

to acknowledge your entry, or

[CANCEL]

to re-enter.

Okay?#

####

Cksum error

Upon each power-up, the unit tests the integrity of its firmware. If the result is not correct this message is displayed and the indicator is not usable. ReFlash the unit.

24

C

OMMUNICATIONS

E

RROR

M

ESSAGES

Information Parameter

prtyX error ovrnX error frmgX error portX error tx on hold tx abort tx

Con'd

Description

The parity of a received character did not match the parity specified in the setup mode at P202. This could also result if the baud rate (P200) or the number of data bits (P201) are incorrect. The ‘X’ in the error message represents the

COMM port number on which the problem occurred.

An overrun error occurred where additional characters were received while the receive buffer was full. The additional characters will be lost. The ‘X’ in the error message represents the COMM port number on which the problem occurred.

The stop bit of a received character did not occur when it was expected. This could be the result of an incorrect baud rate (P200), incorrect number of data bits (P201), or incorrect parity setting (P202). The ‘X’ in the error message represents the COMM port number on which the problem occurred.

The indicator did not check its receive data register in time, thus missing a character. To prevent the problem, try reducing the baud rate (P200). The ‘X’ in the error message represents the COMM port number on which the problem occurred.

Occurs if a data transmission is held up for two seconds of more due to a deasserted handshake. Refer to the description of parameter P209 for more information.

Occurs if the

[CANCEL]

key is pressed when the tx on hold error message is shown or if P209 is set for abort and the transmit buffer becomes full.

Appear briefly when the handshake is re-asserted after the tx on hold message occurs.

BadTx

Port

Appears briefly after an attempt was made to put a byte in an invalid comm port receive buffer.

Wrong

Comm#

An invalid communication port number was specified.

25

M

ISCELLANEOUS

M

ESSAGES

Information Parameter

Entry

Error

Description

An invalid entry was made. When entering data, make sure the values are within the acceptable limits and of the proper type as required by the entry mode.

Enter

Comm#

Clear

All?

Enter toCLR

Sure?

???

No

Instn

This prompt appears during a setup download (P64000,

P64001), database download, database print, etc. if a

COMM port was not specified.

This prompt will appear when the

[CANCEL]

key is pressed when the cursor is at the end of a custom transmit table or macro table. Press

[ENTER]

to clear all information or any other key to retain the table information.

This prompt is used at P65010. When

[ENTER]

is pressed this message will appear. It is then followed by

“Enter=Dflt”.

This prompt is displayed for verification of resetting or clearing information. This message occurs at parameters such as P65001, P65002, P65010, etc. Press

[ENTER]

to clear or reset all information or any other key to retain the information.

Indicates an instance was specified when entering an operating parameter that does not have multiple instances.

An invalid parameter instance was specified when attempting to key in a parameter value.

Invld

Instn

Insuf

OpRAM

NoOpt

RAM!

There is not enough operational RAM available to perform the requested function.

There is no database present or it has not been initialized at P65010.

OutOf

Memry

The current setup requires more RAM than is currently installed. This was previously a “Code 28 NoRAM

AVAIL” error code.

There is not enough FRAM installed to accommodate the current request.

Fram2

Small

>Max

Rows

Invld

Mode

Number of rows received during upload exceeded

2,147,483,648. Contact GSE if this error occurs.

Parameter selected does not exist or feature is disabled.

26

Model 664

User’s Guide

Version 1.1

Part Number: 39-10-43166