BST-IT705 5kV HIGH VOLTAGE INSULATION TESTER

BST-IT705 5kV HIGH VOLTAGE INSULATION TESTER
BST-IT705
5kV HIGH VOLTAGE INSULATION
TESTER
INSTRUCTION MANUAL
Index
Page
1. Safety Precautions.................................................. 1-2
2. Overview.................................................................
3. Features..................................................................
2
3
..........................................................
4
5. Connections............................................................
5
6. Instrument layout....................................................
6
4.
7. Measuring procedure.............................................. 7-14
8. Charge....................................................................
15
9. Maintenance & repair.............................................. 16
10. Interface connection and operation....................... 17-24
1. Safety Precautions
Electricity can cause severe injuries even with low voltages or
currents.
Therefore it is extremely important that you read the following
information before using your high voltage insulation tester.
a. This Instrument must only be used and operated by a
competent trained person and in strict accordance with the
instructions. we will not accept liability for any damage or injury
caused by misuse or non compliance with instructions and
safety procedures.
b. This instrument must not be used on live circuits. Ensure
all circuits are de-energised before testing. see paragraph
for details of built-in warning features should your high
voltage insulation tester be connected to a live system.
c. Always inspect your high voltage insulation tester and test
leads before use for any sign of abnormality or damage. If any
abnormal conditions exist (broken test leads, cracked case,
display faulty etc...) do not attempt to take any measurement
or use the tester. Return your high voltage insulation tester to
your nearest distributor for service.
d. Your high voltage insulation tester has been designed with
your safety in mind. However, no design can completely
protect against incorrect use. Electrical circuits can be
dangerous and / or lethal when a lack of caution or poor safety
practice is used.
e. Pay attention to cautions and warnings which will inform you of
potentially dangerous procedures.
f. Your high voltage insulation tester has a live circuit warning
beeper. If it is connected to a live circuit, a rapid pulsating
bleep will be heard. DO NOT proceed to test and immediately
disconnect the instrument from the circuit. In addition your
tester will display the warning message.
-1-
g. Rated environmental conditions :
(1) Indoor use.
(2) Installation Category .
(3) Pollution Degree 2.
(4) Altitude up to 2000 meters.
(5) Relative humidity 80% max.
(6) Ambient temperature 0ºC~40ºC.
h. Observe the international Electrical Symbols listed below :
Meter is protected throughout by double insulation or
reinforced insulation.
Warning ! Risk of electric shock.
Caution ! Refer to this manual before using the meter.
Earth
2. Overview
This is a 5kV high voltage insulation tester which has output
voltages of 500V, 1000V, 2500V, 5000V.
The top line of the display shows the elapsed time at the start
of the test. Digital readout of the total time will remain displayed
even after testing has ceased.
This instrument displays a voltage warning and sounds when AC
or DC is present before injecting the test voltage.
-2-
3. Features
High contrast OLED display
Microprocessor-controlled
Tests insulation resistance up to 10
4 Insulation test voltages:
500V, 1000V, 2500V, 5000V
AC/DC Voltmeter
Short-circuit current up to 5mA
PI (Polarization Index) indication
DAR (Dielectric Absorption Ratio) indication
Auto-ranging on all insulation ranges
Optical USB to RS-232 data transmission
Well isolated from contact
Well protected from surges
2 built-in optical LEDs for data transfer
Visual and audio warning of external voltage presence
( 30Vac or 30Vdc)
Auto-hold function to freeze reading
Overload protection
Adjustable testing duration: 1~30 minutes
Internal memory for data storage
Displays testing duration for insulation measurement
Auto-off function
200 measurement results can be saved in memory and
recalled on display
-3-
4.
Test Voltage
500V, 1000V, 2500V, 5000V
Insulation resistance
1T / 500V
2T / 1000V
5T / 2500V
10T / 5000V
Accuracy
0~100G / 500V
0~200G / 1000V
0~500G / 2500V
0~1000G / 5000V
±(5.0%rdg + 5dgt)
100G~1T / 500V
200G~2T / 1000V
±12%rdg
500G~5T / 2500V
1000G~10T / 5000V
Resolution
: 1M
10G : 0.01G
100G : 0.1G
1T : 1G
10T : 10G
Short circuit current
up to 5mA
PI (Polarization Index)
DAR (Dielectric Absorption Ratio)
Voltmeter
ACV: 30~600V
DCV: 30~600V
Accuracy: ±(2.0%rdg + 3dgt)
Resolution: 1V
Current measurement
0.5nA ~ 0.55mA
(Depending on the insulation resistance)
Power source
Rechargeable battery
Dimensions
Input: 100-240Vac, 0.4A, 50-60Hz
Output: 24Vdc, 0.62A
330(L) × 260(W) × 160(D)mm
Weight
Approx. 3760g
Safety standard
EN 61010-1 CAT IV 600V
EN 61010-2-030
EN 61326-1
Accessories
Instruction manual
Test leads
Data transmission cable CA-232
Compact disk (CD) for PC interface
Alligator clip
Charger
Test report
Adapter
-4-
5. Connections
FIRST MEASUREMENT
MEASURE WITHOUT THE GUARD TO TAKE EVERYTHING INTO
ACCOUNT AND FIND OUT IF NEED CLEANING.
DIRTY
ELECTRICAL EQUIVALENT
INSULATOR CIRCUIT DIRTY INSULATOR
SECOND MEASUREMENT
MEASURE WITH THE
GUARD TO ENSURE
INSULATOR IS CORRECT.
TYPICAL TEST
RESISTANCE DUE TO CONTAMINATION
CAN BE VERY LOW AND LOWER THE
TOTAL RESISTANCE.
CLEANING PERIODICALLY CAN ALSO
REDUCE SYSTEM POWER CONSUMPTION.
-5-
6. Instrument layout
5kV HIGH VOLTAGE
INSULATION TESTER
High contrast OLED display
ENTER / SAVE button
+HV terminal
TEST / STOP button
GUARD terminal
Function rotary switch
-HV terminal
Connection socket for
ESC button
data transmission
SELECT button
Battery-charge socket
"
" button
"
" button
charge indicator
-6-
7. Measuring procedure
Main Functions:
1. 500V voltage insulation resistance test.
2. 1kV voltage insulation resistance test.
3. 2.5kV voltage insulation resistance test.
4. 5kV voltage insulation resistance test.
5. AC/DC voltage measurement.
Minor functions:
(A) Insulation resistance measurement test (main function)
Note:
2. T
r’s safet
4. W
(a)
measurement.
-7-
(b) Measuring procedure:
1. Connect specimen by test cable.
2. Switch
the function rotary switch to turn on power
and select test voltage from 500V, 1kV, 2.5kV or 5kV,
respectively.
3. Be sure that the cable connecting the specimen, if there
is a exterior voltage (above AC30V or DC30V) exerted,
beeper activates in response; and LCD displays
the warning picture as shown in the AC/DC voltage
measurement function:
Now, test cannot be performed. To go on the test it
should remove the exterior voltage.
TEST / STOP button to test.
4. Then, press
TEST / STOP button
(1) Rotate another angle to lock
for testing continuously.
(2) While test is running, beeper activates to remind
operator that test is underway.
(3) After the set test time is due (see Function 2: the
test time setting), test stops and system will
automatically lock down the test value.
(4) To read the test value on the LCD display
-8-
5. Read the test value from LCD display.
(ENTER/SAVE); LCD displays
6. To store the data, press
the picture shown in below:
(B) Measure voltage (V
the function rotary switch to turn on power and
1. Switch
select AC/DC voltage measurement.
2. Read the data measured from LCD display.
-9-
(C) Date/time adjustment (RTC
the function rotary switch to turn on power and
1. Switch
no meter which function selected (Main page)
2. Press
SELECT button to enter the minor function
selection, LCD display the following pictures respectively:
(ENTER/SAVE) button to enter the Date/Time
3. Press
Adjustment .
4. Press
SELECT button to select year, month, day, hour
or minute for adjustment, LCD display the following
pictures respectively:
5. Press
(value-add) "up" or
the correct value is reached.
(value-reduce) "down" till
6. After all adjustments are complete, press
(ENTER/
Note: if Date/Time unit (year, month, day, hour or minute)
(ESC) to skip it and
needs not to be adjusted, press
return back to the minor function selection.
7. Press
(ESC) again to return back to the main page.
-10-
(D) Measurement time setting (Test T
the function rotary switch to turn on power and
1. Switch
no meter which function selected (Main page)
2. Press
SELECT button to enter the minor function
selection.
3. Press
(value-add) "up" or
(value-reduce) "down" to
following pictures respectively:
(ENTER/SAVE), LCD displays the picture
4. Press
shown in below:
5. Press
(value-add) "up" or
(value-reduce) "down"
to set the test Time (form 1 to 30 minutes).
6. After setting is complete, press
(ENTER/SAVE) to
Note: if Test Timer unit needs not to be adjusted, press
(ESC) to skip it and return back to the minor function
selection.
7. Press
(ESC) again to return back to the main page.
-11-
(E)
the function rotary switch to turn on power and
1. Switch
no meter which function selected (Main page)
2. Press
SELECT button to enter the minor function
selection.
3. Press
(value-add) "up" or
(value-reduce) "down" to
following pictures respectively:
4. Press
(ENTER/SAVE), LCD displays the picture shown
in below:
(ENTER/SAVE) to query the subpage of data
5. Press
(include Resistance value, testing voltage, PI, DAR,
Saving Date & Time).
6. Press
(value-add) "up" or
(value-reduce) "down" to
select the required data value. If no data available, LCD
displays the picture shown in below:
7. After the query is over, press
minor function selection.
8. Press
(ESC) return back to the
(ESC) again to return back to the main page.
-12-
c
(F)
the function rotary switch to turn on power and
1. Switch
no meter which function selected (Main page)
SELECT button to enter the minor function
2. Press
selection.
3. Press
(value-add) "up" or
(value-reduce) "down" to
pictures respectively:
4. Press
(ENTER/SAVE) to inquire whether to clear up the
data or not; LCD displays the picture shown in below:
Note: if the erasing unit needs not to be done, press
(ESC) to skip it and return back to the minor function
selection.
(ENTER/SAVE) again to clear up the data ; LCD
5. Press
displays the picture shown in below:
, then return back to the minor function selection.
6. Press
(ESC) to return back to the main page.
-13-
(G) Introduction of other functions
1. Dielectric absorption ratio (DAR):
Ratio of insulation resistance between 1-min and 30-sec
DAR :
1-min insulation resistance
30-sec insulation resistance
2. Polarization index (PI):
Ratio of insulation resistance between 10-min and 1-min
PI :
10-min insulation resistance
1-min insulation resistance
Lower insulation resistance tested takes longer test
time, which would deteriorate the specimen. Thus,
higher DAR or PI (as close to 1) would create better
insulation grade of specimen.
Operation:
During the test run, wait for one minute, DAR will be
displayed automatically; wait for 10 minutes, PI will be
displayed automatically.
3. AUTO OFF:
System will shutdown automatically after 3 minutes without
operation.
-14-
8. Charge
(A) Timing:
After "Low Battery" is displayed on LCD display, perform
battery charge; LCD displays the picture shown in below:
(B) Process:
1. Plug one end of charger into the battery-change socket
and the other end into the ACV power socket.
2. If ACV plug socket is energized, the LED charge
indicator is lit on indicating that charge is underway.
If the ACV plug socket isn’t energized, remove to
another power-energized one making power charge.
3. After the voltage reaches 12.6V, charge process is
complete. (It can be observed on LCD display.)
No measurement can be performed during the charging
process.
-15-
9. Maintenance & repair
(A) To avoid and electric-shock or device damage, do not wet
inner part of the tester.
(B) Avoid the tester from being dropped down that would
damage or disconnect devices apart.
(C) Wipe the tester surface with soft, dry cloth and mild
detergent. Prohibit from using sand paper or solvent.
Note:
1. This tester is HV operated; user should not open the
outer casing. If any damage occurs, take the tester back to
manufacturer for repair.
2. If the tester isn’t under operation for a long period, charge
it at least once per month as to protect the battery from
being deteriorated.
-16-
10. Data Transmission Interface connection and
operation
(A) 5KV Insulation Tester Installation Steps:
1. This 5KV Insulation Tester Installation program will be
installed on the computer automatically.
2. Click the “Next” key to set.
3. If you want to install a different folder, click Browse, and
select another folder.
If it’s not necessary, click the “Next” key.
-17-
4. Click the “Next” key.
personal computer.
-18-
6. It will show the information of Insulation tester has been
successfully installed and then click “Finish” key.
(Note: If your personal computer system is Windows 7, it will
indicate the driver automatically. It’s necessary to install the
driver if your computer system is not windows 7, then the
driver is in the compact disk (CD). The directory is “ E:USB
-19-
(B) Windows Comm Port setting:
1. Plug data transmission cable into your personal
computer USB port.
Panel” Application.
Control Panel
Sound” Application.
Hardware and Sound
-20-
4.
Manager” Application.
Device Manager
5. On the Device Manger Application, click right button of the
6. On the Universal Serial Bus Controllers, Find the USB
Serial Port (COM 3)
Universal Serial Bus controllers
USB Serial Port (COM3)
(Note: USB Serial Port will indicate different
“COM”automatically)
-21-
(C) 5KV Insulation Tester Software Comm Port setting:
1. Connect data transmission cable to the Insulation
Tester.
2. Click the icon of the “5KV Insulation Tester” on your
desktop of your personal computer.
3. On the 5KV Insulation Tester Windows, select the correct
“Comm Port” and click the RS 232 connection button.
Click
Comm Port
-22-
(D) 5KV Insulation Tester Interface Layout:
RS 232 Connection.
Main operation interface.
Memory saving and downloading interface.
Click the "Download Log from Device" button to download
-23-
Click the "Chart Display" button to see the chart,
This Chart Display can also show Resistance and Current
diagram by clicking the dot
Clic
below:
-24-