Logic Analysis Support for the Texas Instruments

Logic Analysis Support for the Texas Instruments
User’s Guide
Publication number E8137-97000
October 2000
For Safety information, Warranties, and Regulatory information,
see the pages behind the index.
© Copyright Agilent Technologies 1994-2000
All Rights Reserved
Logic Analysis Support for
the Texas Instruments
TMS320C6000 DSP Family
Logic Analysis Support for the
Texas Instruments TMS320C6000 DSP Family—
At a Glance
Inverse Assembler
The Agilent Technologies E9617A inverse assembler, in conjunction with an
Agilent Technologies 16700-series logic analyzer, allows you to view C6000
assembly instructions that are executing in your target system.
The inverse assembler is identified as “Agilent Technologies E8137A” in the
Setup Assistant.
Source Correlation Tool Set
The Agilent Technologies B4620B Source Correlation Tool Set lets you set up
logic analyzer triggers based on source code (even in SDRAM), and it lets you
view the source code associated with signal values captured by the logic
analyzer.
In This Book
This book documents the following products:
Product Ordered
Supports
Includes
E9617A Option 001
inverse assembler
Texas Instruments
TMS320C6202/6203/
6204/6205/6211 DSP
The E8137A inverse assembler
and B4620B Source Correlation
Tool Set
2
Logic Analysis Support for the TI C6000
Tips To Save You Time
Use the Setup Assistant
Click here to connect the logic analyzer
cables, and automatically load the correct
configuration files. See page 12.
Use the appropriate Run button
Click here to start a measurement.
If your system includes an
emulation probe/module,
click here to run the target
microprocessor.
Logic Analysis Support for the TI C6000
3
Additional Information Sources
Newer editions of this manual may be available. Contact your local Agilent
Technologies representative.
Application notes may be available from your local Agilent Technologies
representative or on the World Wide Web at:
http://www.agilent.com/find/logicanalyzer
The measurement examples include valuable tips for making emulation and
analysis measurements. You can find the measurement examples under the
system help in your Agilent Technologies 16700-series logic analysis system.
4
Logic Analysis Support for the TI C6000
Contents
At a Glance 2
In This Book 2
Tips To Save You Time 3
1
Equipment and Requirements
Setup Assistant
11
12
Equipment and Software Supplied
14
Additional equipment required 15
Additional equipment supported 15
Compatible Logic Analyzers
2
16
Preparing the Target System
17
Preparing for Logic Analysis (and Inverse Assembly)
Design Considerations
18
18
Designing Logic Analyzer Connectors into Your Target System
Using High-Density Connectors 19
Recommended Connector Layout and Signal Routing
3
21
Setting Up the Logic Analysis System
Power-ON/Power-OFF Sequence
19
25
26
To power-ON the Agilent Technologies 16700-series logic analysis
systems 26
5
Contents
To power-OFF
26
Installing Logic Analyzer Modules
Installing Software
28
To install the software from CD-ROM
4
27
29
Probing the Target System
31
Connecting the Logic Analyzer to the Target System
5
Configuring the Logic Analyzer
Configuring 16700-series Logic Analysis Systems
32
33
35
To load configuration files (and the inverse assembler) from hard disk
36
To load configuration files (and the inverse assembler) from floppy disk
37
To list software packages that are installed 38
Logic analyzer configuration files 38
Using the Inverse Assembler
39
To use the Invasm menu
39
Loading the Inverse Assembler 39
Setting Inverse Assembler Preferences
40
To set the inverse assembler preferences 40
To set the Memory Map preferences 41
To set the C6000 DSP Options preferences 44
To set the Decoding Options preferences 46
Loading Symbol Information
6
47
Contents
To view user-defined symbols 47
To load object file symbols 48
To compensate for relocated code 51
Setting Up Labels for Groups of Signals
52
Predefined Label Descriptions 52
To define additional labels 53
Changing the Analysis Mode
54
To change to state analysis 54
To change to timing analysis 55
6
Capturing DSP Execution
Setting Up Logic Analyzer Triggers
To set up logic analyzer triggers
Triggering on Source Code
59
59
61
To avoid capturing library code execution
7
57
61
Displaying Captured DSP Execution
To display the captured state data 64
To display symbols 65
To interpret the inverse assembled data 66
To use the inverse assembler filters 67
To view the source code associated with captured data
To display captured timing analysis mode data 72
63
69
7
Contents
8
General-Purpose ASCII (GPA) Symbol File
Format 73
General-Purpose ASCII (GPA) Symbol File Format
GPA Record Format Summary
SECTIONS 78
FUNCTIONS 79
VARIABLES 80
SOURCE LINES 81
START ADDRESS 82
Comments 82
9
10
76
Specifications and Characteristics
Operating Characteristics
Solving Logic Analyzer Problems
Intermittent data errors 87
Unwanted triggers 87
No activity on activity indicators
No trace list display 88
87
88
89
Target system will not boot up 89
Erratic trace measurements 90
Capacitive loading 90
Solving Inverse Assembler Problems
8
83
84
Troubleshooting the Logic Analyzer
Solving Probing Problems
74
91
85
Contents
No inverse assembly or incorrect inverse assembly
Inverse assembler will not load or run 92
Solving Intermodule Measurement Problems
An event wasn’t captured by one of the modules
Logic Analyzer Messages
Index
93
93
94
“. . . Inverse Assembler Not Found” 94
“No Configuration File Loaded” 94
“Selected File is Incompatible” 95
“Slow or Missing Clock” 95
“Time from Arm Greater Than 41.93 ms”
“Waiting for Trigger” 96
Glossary
91
95
97
103
9
Contents
10
1
Equipment and Requirements
11
Chapter 1: Equipment and Requirements
Setup Assistant
Setup Assistant
The Setup Assistant is an on-line tool for connecting and configuring your
logic analysis system for microprocessor and bus analysis. The Setup Assistant
is available on the 16700-series logic analysis systems. You can use the Setup
Assistant in place of the connection and configuration procedures provided in
this manual.
This menu-driven tool will guide you through the connection procedures for
connecting the target system to an analysis probe, an emulation module, or
other supported equipment. It will also guide you through connecting the logic
analyzer pods to connectors on the target system.
The inverse assembler is identified as “Agilent Technologies E8137A” in the
Setup Assistant. Start the Setup Assistant by clicking
in the system
window.
If you ordered this product with your Agilent Technologies 16700-series logic
analysis system, the logic analysis system has the latest software installed,
including support for this product.
12
Logic Analysis Support for the TI C6000
Chapter 1: Equipment and Requirements
Setup Assistant
Install software
Connect probes
Load configuration
file
Create executable with
symbol information
Use source
correlation?
Yes
Download executable to
target
Load program symbols
into analyzer
No
Installation done. Begin
making measurements.
E8137F01.VSD
Logic Analysis Support for the TI C6000
13
Chapter 1: Equipment and Requirements
Equipment and Software Supplied
Equipment and Software Supplied
The inverse assembler (Agilent Technologies E9617A Option 001) includes:
•
Logic analyzer configuration files and the inverse assembler software on a
CD-ROM (for Agilent Technologies 16700-series logic analysis systems).
•
This User’s Guide.
The inverse assembler is identified as “E8137A” in the Setup Assistant.
14
Logic Analysis Support for the TI C6000
Chapter 1: Equipment and Requirements
Equipment and Software Supplied
Additional equipment required
In addition to the items listed above, the following is required:
•
Connector headers on your target system which supply the necessary
signals to the logic analyzer. See Chapter 2, “Preparing the Target System,”
beginning on page 17 for information on designing the appropriate
connectors into the target system.
•
Agilent Technologies termination adapter cables to attach your target
system to a logic analyzer.
Additional equipment supported
Agilent Technologies B4620B Source Correlation Tool Set. The inverse
assembler may be used with the Agilent Technologies B4620B Source
Correlation Tool Set. The software is already installed on the Agilent
Technologies 16600/16700-series logic analysis system’s disk. All you need is
the entitlement certificate for licensing the source correlation tool set
software. The CD-ROM is included in case you need to re-install the software.
Logic Analysis Support for the TI C6000
15
Chapter 1: Equipment and Requirements
Compatible Logic Analyzers
Compatible Logic Analyzers
A minimum of four logic analyzer pods (68 logic analysis channels) are
required for inverse assembly. For inverse assembly of both instructions and
data, eight pods are required. If optional signals are used, such as PCI analysis,
additional logic analyzer pods are required.
The inverse assembler only works in 16700-series logic analysis systems. The
16500 logic analysis system and 166x/167x portable logic analyzer families are
not supported.
The inverse assembler works with logic analysis system software version
A.02.10 or greater. The latest logic analysis system software version is on the
CD-ROM shipped with this product.
Given these restrictions, the following logic analyzers can be used:
Agilent Technologies
Logic Analyzer
16550A
16554A
16555A
16555D
16556A
16556D
16557D
16710A
16711A
16712A
16715A
16716A
16717A
16718A
16719A
16750A
16751A
16752A
16
Logic Analysis Support for the TI C6000
2
Preparing the Target System
17
Chapter 2: Preparing the Target System
Preparing for Logic Analysis (and Inverse Assembly)
Preparing for Logic Analysis (and Inverse
Assembly)
The inverse assembler requires a minimum of four logic analyzer pods.
Design Considerations
Cache memory
The processor supplies no external information when the cache is enabled.
This poses a problem for logic analysis, since there are no linear address and
data cycles to reconstruct code flow. Therefore, when the cache is on, the
inverse assembler can only track transactions on the visible bus, but cannot
provide mnemonic disassembly.
The processor provides no external information for accesses to the internal
cache. The inverse assembler does not support accesses to the internal cache.
Electrical requirements
Any probed signal line must be able to supply a minimum of 600 mV to the
probe tip and handle a minimum loading of 90 kΩ shunted by 10 pF. The
maximum input voltage for the logic analyzer is +/- 40 volts peak.
For all signals, the logic analyzers require a minimum combined setup/hold
window:
•
For 16715/16/17/18/19/50/51/52A logic analyzers the combined window
must be at least 2.5 ns (1.25 ns using eye finder).
•
For 16710/11/12A logic analyzers, the combined window must be at least
4.0 ns.
•
For 16557 logic analyzers, the combined setup/hold time must be at least
3.0 ns.
•
For 16550/54/55/56 logic analyzers, the combined setup/hold time must be
at least 3.5 ns.
18
Logic Analysis Support for the TI C6000
Chapter 2: Preparing the Target System
Designing Logic Analyzer Connectors into Your Target System
Designing Logic Analyzer Connectors into Your
Target System
The logic analyzer can be connected directly to connectors on your target
system. This section describes what kind of connectors to use, and how to
connect the correct signals to the connectors.
Using High-Density Connectors
High-density MICTOR (Matched Impedance ConnecTOR) connectors are
recommended for connecting the target system to the logic analyzer because
they require less board space and provide higher signal integrity than
medium-density connectors. Each connector carries 32 signals and two
clocks.
•
Each 32-signal high-density header connector requires approximately 1.1”
x 0.4” of printed-circuit board space.
•
The part number for the high-density MICTOR connector is: AMP P/N 2767004-2 or Agilent Technologies 1252-7431.
•
Each MICTOR connector requires one Agilent Technologies E5346A highdensity termination adapter cable to attach to the logic analyzer. This is a
Y-cable where the single end connects to the high-density header
connector, and each of the two opposite ends connects to a logic analyzer
pod.
•
Any probed signal line must be able to supply a minimum of 600 mV to the
probe tip and handle a minimum loading of 90 kOhms shunted by 10 pF.
The maximum input voltage for the logic analyzer is +/- 40 volts peak.
•
If a printed-circuit board already has a header connector attached, but the
signal pinouts do not match the requirement, an adapter (Agilent part
number E5346-60002) can be used to route the signals to the correct pods.
•
A plastic shroud (Agilent part number E5346-44701) is available to secure
the mechanical connection of the high-density cable to the MICTOR
header connector.
Logic Analysis Support for the TI C6000
19
Chapter 2: Preparing the Target System
Designing Logic Analyzer Connectors into Your Target System
See Also
More information on this connector is available in Portable Document Format
(PDF) from the web site:
http://www.tm.agilent.com/
When you reach this web site, search on E5346A.
High-Density Connector Mechanical Specifications
Dimensions of the AMP MICTOR 2-767004-2 surface mount connector are
shown below. The holes for mounting a support shroud are off-center to allow
0.40 in (1.20 mm) centers when using multiple connectors.
The high-density connector pin assignment and recommended circuit board
routing are shown in the following figure.
20
Logic Analysis Support for the TI C6000
Chapter 2: Preparing the Target System
Designing Logic Analyzer Connectors into Your Target System
Five center inline pins on the connector are the signal ground returns and
must be connected to ground.
Recommended Connector Layout and Signal
Routing
The following shows the recommended connector layout and signal routing.
Recommended Configuration Connection Notes
•
'nc' pins MUST be a true no-connect on the target. The signals are used for
other functions unavailable to target probing.
•
Five center inline pins on the connector are the signal ground returns and
must be connected to ground.
•
Any blank pins can be used for user defined signals.
•
'#' or overscore denotes an active low signal.
•
J1-J2: Connectors required for inverse assembly.
Recommended Connector Layout
Minimize trace lengths from the DSP to the connectors. Keep trace lengths
less than 1/5 of the rise time of the signal.
Logic Analysis Support for the TI C6000
21
Chapter 2: Preparing the Target System
Designing Logic Analyzer Connectors into Your Target System
Recommended Signal Routing
MICTOR Connector J1
J1
Pin
Signal
J1
Pin
Signal
1
No Connect
2
No Connect
3
No Connect
4
No Connect
5
ECLKOUT/CLKOUT2 (*)
6
ARE/SDCAS/SSADS
7
CE3
8
A15
9
CE2
10
A14
11
CE1
12
A13
13
CE0
14
A12
15
BE3
16
A11
17
BE2
18
A10
19
BE1
20
A9
21
BE0
22
A8
23
ARDY
24
A7
25
AWE/SDWE/SSWE
26
A6
27
A21 (MSB)
28
A5
29
A20
30
A4
31
A19
32
A3
33
A18
34
A2
35
A17
36
37
A16
38
22
Even Cable
Odd Cable
Logic Analyzer Pod 2
Logic Analyzer Pod 1
Logic Analysis Support for the TI C6000
Chapter 2: Preparing the Target System
Designing Logic Analyzer Connectors into Your Target System
MICTOR Connector J2
J2
Pin
Signal
J2
Pin
Signal
1
No Connect
2
No Connect
3
No Connect
4
No Connect
5
AOE/SDRAS/SSOE
6
CLKOUT1(*)
7
D31 (MSB)
8
D15
9
D30
10
D14
11
D29
12
D13
13
D28
14
D12
15
D27
16
D11
17
D26
18
D10
19
D25
20
D9
21
D24
22
D8
23
D23
24
D7
25
D22
26
D6
27
D21
28
D5
29
D20
30
D4
31
D19
32
D3
33
D18
34
D2
35
D17
36
D1
37
D16
38
D0 (LSB)
Even Cable
Odd Cable
Logic Analyzer Pod 4
Logic Analyzer Pod 3
* Use CLKOUT1/2 for C6202/3/4/5
Logic Analysis Support for the TI C6000
23
Chapter 2: Preparing the Target System
Designing Logic Analyzer Connectors into Your Target System
24
Logic Analysis Support for the TI C6000
3
Setting Up the Logic Analysis System
25
Chapter 3: Setting Up the Logic Analysis System
Power-ON/Power-OFF Sequence
Power-ON/Power-OFF Sequence
Listed below are the sequences for powering on and off a fully connected
system. Simply stated, your target system is always the last to be powered on,
and the first to be powered off.
To power-ON the Agilent Technologies 16700-series
logic analysis systems
Ensure the target system is powered off.
1 Turn on the logic analyzer. The Setup Assistant will guide you through
the process of connecting and configuring the logic analyzer.
2 When the logic analyzer is connected to the target system, and
everything is configured, turn on your target system.
To power-OFF
Turn off power to your system in the following order:
1 Turn off your target system.
2 Turn off your logic analysis system.
26
Logic Analysis Support for the TI C6000
Chapter 3: Setting Up the Logic Analysis System
Installing Logic Analyzer Modules
Installing Logic Analyzer Modules
You should install logic analyzer, oscilloscope, or pattern generator modules in
your logic analysis system before you install an emulation module (if
applicable) and software.
CAUTION:
Electrostatic discharge can damage electronic components. Use appropriate
ESD equipment (grounded wrist straps, etc.) and ESD-safe procedures when
you handle and install modules.
Refer to the Agilent Technologies 16700-series logic analysis system’s
Installation Guide for instructions on installing modules.
Logic Analysis Support for the TI C6000
27
Chapter 3: Setting Up the Logic Analysis System
Installing Software
Installing Software
This section explains how to install the software you will need for your inverse
assembler or emulation solution.
Installing and loading
Installing the software will copy the files to the hard disk of your logic analysis
system. Later, you will need to load some of the files into the appropriate
measurement module.
Logic analysis system or logic analyzer frame
CD-ROM
Install
Load
Hard Disk
Logic analyzer
or emulation
module
What needs to be installed
If you ordered an inverse assembler or emulation solution with your logic
analysis system, the software was installed at the factory.
The following files are installed when you install a processor support package
from the CD-ROM:
•
Logic analysis system configuration files.
•
Inverse assembler (automatically loaded with the configuration files).
•
Personality files for the Setup Assistant.
•
Emulation module firmware (for emulation solutions).
•
Emulation Control Interface (for emulation solutions).
The Agilent Technologies B4620B Source Correlation Tool Set is installed with
the logic analysis system’s operating system.
28
Logic Analysis Support for the TI C6000
Chapter 3: Setting Up the Logic Analysis System
Installing Software
To install the software from CD-ROM
Installing a processor support package from a CD-ROM will take just a few
minutes. If the processor support package requires an update to the Agilent
Technologies 16700 operating system, installation may take approximately 15
minutes.
If the CD-ROM drive is not connected, see the instructions printed on the CDROM package.
1 Turn on the CD-ROM drive first and then turn on the logic analysis
system.
If the CD-ROM and analysis system are already turned on, be sure to save any
acquired data. The installation process may reboot the logic analysis system.
2 Insert the CD-ROM in the drive.
3 Select the System Administration icon.
4 Select the Software Install tab.
5 Select Install....
Change the media type to “CD-ROM” if necessary.
6 Select Apply.
7 From the list of types of packages, double-click “PROC-SUPPORT.”
NOTE:
For touch screen systems, double select the “PROC-SUPPORT” line by quickly
touching it twice.
A list of the processor support packages on the CD-ROM will be displayed.
8 Select the “TIC6000” package.
If you are unsure whether this is the correct package, click Details for
information about the contents of the package.
9 Select Install.
The Continue dialog box will appear.
10 Select Continue.
Logic Analysis Support for the TI C6000
29
Chapter 3: Setting Up the Logic Analysis System
Installing Software
The Software Install dialog will display “Progress: completed successfully”
when the installation is complete.
11 If required, the system will automatically reboot. Otherwise, close the
software installation windows.
The configuration files are stored in /logic/configs/hp/tic6000. The inverse
assemblers are stored in /logic/ia.
See Also
The instructions printed on the CD-ROM package for a summary of the
installation instructions.
The online help for more information on installing, licensing, and removing
software.
30
Logic Analysis Support for the TI C6000
4
Probing the Target System
31
Chapter 4: Probing the Target System
Connecting the Logic Analyzer to the Target System
Connecting the Logic Analyzer to the Target
System
You can also use the Setup Assistant to guide you through the connection
process. See page 12.
CAUTION:
Be sure to power down the target system before connecting or disconnecting
cables. Otherwise, you may damage circuitry in the analyzer or target system.
Find the labels that were shipped with the high-density cables and use them
to help identify the connections. Attach the labels to the cables after
connecting the cables to the logic analyzer.
If you have several logic analyzer cards connected together as a single
“machine,” be sure to use the pods from the master card.
Analyzer Card
Pod 4
Target Connector
J2, Even J2, Odd
ADDR/
ADDR
STAT
Signal
32
Pod 3
Pod 2
Pod 1
J1, Even J1, Odd
DATA
DATA
Logic Analysis Support for the TI C6000
5
Configuring the Logic Analyzer
33
Chapter 5: Configuring the Logic Analyzer
The sections of this chapter describe setting up and using the inverse assembler.
Because your target system is designed uniquely according to your needs, it is
important that you specify the available signals and memory regions to the inverse
assembler.
The information in this chapter is presented in the following sections:
•
Loading the configuration file and the inverse assembler
•
Tables showing configuration file names
•
Using the inverse assembler
•
Setting the inverse assembler preferences
•
Symbols
•
Changing analysis mode
34
Logic Analysis Support for the TI C6000
Chapter 5: Configuring the Logic Analyzer
Configuring 16700-series Logic Analysis Systems
Configuring 16700-series Logic Analysis Systems
You configure the logic analyzer by loading a configuration file. Normally this
is done using the Setup Assistant (see page 12). If you did not use the Setup
Assistant, you can load the configuration and inverse assembler files from the
logic analysis system hard disk.
The configuration file to load is determined by what model of logic analyzer
card you are using.
The information in the configuration file includes:
•
Label names and channel assignments for the logic analyzer
•
Inverse assembler file name
If the cache is turned off, the inverse assembler decodes captured data into
software addresses (SW_ADDR label) and assembly language mnemonics.
Logic Analysis Support for the TI C6000
35
Chapter 5: Configuring the Logic Analyzer
Configuring 16700-series Logic Analysis Systems
To load configuration files (and the inverse assembler) from
hard disk
If you use Setup Assistant, it will load configuration files and the inverse
assembler for you. This is the preferred method. If you did not use Setup
Assistant, you can load the configuration and inverse assembler files from the
logic analysis system hard disk.
1 Select the File Manager icon. Use File Manager to ensure that the
subdirectory /logic/configs/hp/tic6000/ exists.
If the above directory does not exist, you need to install the TIC6000
Processor Support Package. Close File Manager, then use the procedure on
the CD-ROM jacket to install the TIC6000 Processor Support Package before
you continue. See “To install the software from CD-ROM” on page 29 for
details.
2 Using File Manager, select the configuration file you want to load in the
/logic/configs/hp/tic6000/ directory, then select Load. If you have more
than one logic analyzer installed in your logic analysis system, use the
Target field to select the machine you want to load.
The logic analyzer is configured for TIC6000 analysis by loading the
appropriate configuration file. Loading the indicated file also automatically
loads the inverse assembler. The configuration file names are shown in the
table on page 38.
3 Close File Manager.
36
Logic Analysis Support for the TI C6000
Chapter 5: Configuring the Logic Analyzer
Configuring 16700-series Logic Analysis Systems
To load configuration files (and the inverse assembler)
from floppy disk
If you use Setup Assistant, it will load configuration files and the inverse
assembler for you. This is the preferred method. If you did not use Setup
Assistant, you can load the configuration and inverse assembler files from the
logic analysis system hard disk or floppy disk; however, the preferred method
is to install this functionality from the CD-ROM onto the hard disk and load
from the hard disk.
To install a configuration and inverse assembler file from a floppy disk:
1 Insert the floppy disk in the floppy drive on the logic analysis system
mainframe.
2 In the logic analysis System window, select the File Manager icon.
3 In the File Manager window:
•
Set Current Disk to Flexible Disk.
•
Set Target to the analyzer you wish to configure.
•
Select the name of the desired configuration file in the Contents frame.
The Contents frame lists the configuration files and inverse assembler files
available on the floppy disk. These may be either DOS or LIF format files.
Either format can be loaded directly into the appropriate logic analyzers.
Note that the logic analyzers read both DOS and LIF formats. However, only
DOS formatted floppy disks can be used to store configurations and data. LIF
format floppy disks are read-only.
4 Select Load.
The configuration file you choose will set up the logic analyzer and associated
tools. You may see Information, Error, and Warning dialogs that say your
configuration has been loaded, and advise you about making proper
connections.
5 Select the Workspace window icon to see the arrangement of analysis
tools in your configuration.
6 Select the logic analyzer icon in your configuration and choose its Setup
button to see the way your configuration file defined the Config,
Format, and Trigger options.
Logic Analysis Support for the TI C6000
37
Chapter 5: Configuring the Logic Analyzer
Configuring 16700-series Logic Analysis Systems
NOTE:
Under the Format tab, buses are labeled, and bits included in each bus are
identified by an asterisk “*”.
This procedure restores the configuration that was in effect when the
configuration file was saved. Because the file was not saved using your system,
you may receive error messages about loading the enhanced inverse
assembler or about pods that were truncated. Select the Config, Format, and
Trigger tabs and modify the configuration to satisfy your measurement desires.
Then you can save your customized configuration to DOS format using the
File→Save Configuration selection in any of your tool windows, or selecting
the Save tab in the File Manager. For details about how to save configuration
files, open the Help window.
To list software packages that are installed
• In the System Administration Tools window, select List....
Logic analyzer configuration files
The following table shows the configuration files for the supported logic
analyzers.
No-data
Configuration File
16550A
C6000F_1
16554/5/6/7
C6000M_1
16710/11/12A
C6000F_1
16715/16/17/18/19A C6000L_1
16750/51/52A
C6000L_1
Analyzer Model
38
Logic Analysis Support for the TI C6000
Chapter 5: Configuring the Logic Analyzer
Using the Inverse Assembler
Using the Inverse Assembler
This section discusses the general output format of the inverse assembler and
processor-specific information.
Traditional inverse assembly, in which the external processor bus states are
captured and decoded, may be implemented by disabling the target’s cache.
However, this will slow the target significantly, and may induce timing related
problems. The target system’s performance will be much better if the cache-on
trace reconstruction feature is enabled when using the inverse assembler.
To use the Invasm menu
The Invasm menu provides four choices: Load, Preferences, Filter, and
Options. Access the Invasm menu in the listing window.
You must use the Preferences dialog to configure the inverse assembler to
match the memory controller configuration. The other dialogs assist in
analyzing and displaying data. The following sections describe these dialogs.
Loading the Inverse Assembler
The Load dialog lets you load a different inverse assembler and apply it to the
data in the Listing window. In some cases you may have acquired raw data;
you can use the Load dialog to apply an inverse assembler to that data.
Logic Analysis Support for the TI C6000
39
Chapter 5: Configuring the Logic Analyzer
Setting Inverse Assembler Preferences
Setting Inverse Assembler Preferences
The Invasm Preferences dialog lets you give the inverse assembler information
about your target system so that it can properly disassemble signal values
captured by the logic analyzer.
To set the inverse assembler preferences
To open the Preferences dialog:
1 Load a configuration file, if needed.
2 Open the Listing window.
3 Select Preferences... from the Invasm menu at the top of the Listing
window.
4 Look at each of the tabs in the Preferences dialog and adjust the
settings to match your target system. At a minimum, you must set the
memory map preferences.
40
Logic Analysis Support for the TI C6000
Chapter 5: Configuring the Logic Analyzer
Setting Inverse Assembler Preferences
To set the Memory Map preferences
The DSP does not provide all of the signals required to reconstruct physical
addresses. It is therefore necessary to reconstruct the address from various
pieces of information:
•
Memory map configuration (set in the Preferences dialog)
•
The RAS and CAS signals
•
The address signals
It is necessary to configure the memory map in the Preferences dialog before
using the inverse assembler. None of the memory banks are enabled by default.
Logic Analysis Support for the TI C6000
41
Chapter 5: Configuring the Logic Analyzer
Setting Inverse Assembler Preferences
Bank Enable/Disable. Enable the banks for all of the chip selects that are
being used in the system; otherwise, the inverse assembler may not function
correctly. These enables are required because chip selects are active low and
unconnected logic analyzer channels float low; without the enables, the
inverse assembler cannot tell the difference between active and unconnected
chip selects.
Base Address. Specifies the starting address of the memory bank.
This information is used for upper address bit reconstruction, and is required
for proper source correlation and source line triggering.
Memory Type.
Depending upon the type of memory that is accessed, there are different
signals that assert when a valid address is on the bus. For this reason, the
inverse assembler must know what type of memory is being accessed.
Memory Width.
If your target system uses 8-bit or 16-bit memory, change this from the default
of 32 bits.
42
Logic Analysis Support for the TI C6000
Chapter 5: Configuring the Logic Analyzer
Setting Inverse Assembler Preferences
Inst/Data.
All regions of memory are assumed to contain instructions by default. Set up
the Region Map dialog for each memory bank which contains data.
Click Apply when you have finished configuring the memory map.
Logic Analysis Support for the TI C6000
43
Chapter 5: Configuring the Logic Analyzer
Setting Inverse Assembler Preferences
To set the C6000 DSP Options preferences
The Processor Options tab of the Preferences dialog lets you tell the inverse
assembler which mode the DSP is operating in.
CPU Type
The C6202/3/4/5 can support both synchronous and asynchronous memory by
using two clocks, but a logic analyzer card can’t use both clocks at once. If
your target system uses a mix of synchronous and asynchronous memory, set
up a second logic analyzer “machine” to probe the other memory system.
44
Logic Analysis Support for the TI C6000
Chapter 5: Configuring the Logic Analyzer
Setting Inverse Assembler Preferences
Cache Mode
If the cache of the DSP is turned off, you can view disassembled opcodes in
the Listing window.
If the cache is turned on, you can only view bus activity (mostly memory
reads). See the documentation for your DSP for information on disabling
cache.
Endian Mode Selection
The inverse assembler is designed to support both the big endian mode (mostsignificant byte has the smallest address) and the little endian mode (leastsignificant byte has the smallest address) of operation. When operating in
little-endian mode, the processor uses a technique known as “address
munging” to convert internal little endian addresses into external big endian
addresses.
CAS Latency
If your target system uses SDRAM, the inverse assembler needs to know the
CAS latency of the memory (the time delay that passes before the SDRAM
starts to carry out a read command after receiving it).
Number of Column Address Bits
If your target system uses SDRAM, set the number of column bits strobed by
CAS (depends on the size of the SDRAM).
Logic Analysis Support for the TI C6000
45
Chapter 5: Configuring the Logic Analyzer
Setting Inverse Assembler Preferences
To set the Decoding Options preferences
Display Alias Registers. Displays certain registers in terms of their
common names. For example, b15 will be displayed as sp.
Display Simplified Instructions Decoding. Displays certain opcodes
in a simplified form. For example, add 0x00,a10,b4 will be displayed as
mv 10,b4.
Display Decimal Offsets/Constants. Displays decimal instead of
hexadecimal values in the Inverse Assembler column of the listing.
46
Logic Analysis Support for the TI C6000
Chapter 5: Configuring the Logic Analyzer
Loading Symbol Information
Loading Symbol Information
Symbols are more easily recognized than hexadecimal address values in logic
analyzer trace displays, and they are easier to remember when setting up
triggers.
Agilent logic analyzers let you assign user-defined symbol names to particular
label values.
Also, you can download symbols from certain object file formats into Agilent
Technologies logic analyzers.
To view user-defined symbols
User-defined symbols are symbols you create in the logic analyzer by assigning
symbol names to label values. Typically, you assign symbol names to address
label values, but you can define symbols for data, status, or other label values
as well.
User-defined symbols are saved with logic analyzer configurations. The logic
analyzer configuration files included with this inverse assembler do not
contain predefined symbols for logic analyzer labels.
To display any defined symbols:
1 Open the logic analyzer’s Setup window.
2 Select the Symbol tab.
3 Select the User Defined tab.
4 Choose a label name from the Label list.
The logic analyzer will display the symbols associated with the label.
Logic Analysis Support for the TI C6000
47
Chapter 5: Configuring the Logic Analyzer
Loading Symbol Information
To load object file symbols
The most common way to load program symbols into the logic analyzer is from
an object file that is created when the program is compiled. The object file
containing symbolic debug information must be in a format the logic analyzer
understands. To use the Source Correlation Toolset with this inverse
assembler, the object file must be in COFF Version 2.0 format.
If your compiler generates object files in a format that the logic analyzer
doesn’t understand, you can use a General-Purpose ASCII (GPA) symbol file
(see Chapter 8, “General-Purpose ASCII (GPA) Symbol File Format,” on page
73).
To load symbols in the logic analysis system:
1 Open the logic analyzer module’s Setup window.
2 Click the Symbol tab.
3 Click the Object File tab.
Make sure the label is ADDR.
48
Logic Analysis Support for the TI C6000
Chapter 5: Configuring the Logic Analyzer
Loading Symbol Information
From this dialog you can select object files and load their symbol information.
When you load object file symbols into a logic analyzer, a database of symbol/
line number to address assignments is generated from the object file.
Logic Analysis Support for the TI C6000
49
Chapter 5: Configuring the Logic Analyzer
Loading Symbol Information
The Symbol Selector dialog allows you to use a symbol in place of a
hexadecimal value when defining trigger patterns, trigger ranges, and so on.
To select this dialog, select Trigger tab, then Pattern. Set the numeric base
field to Symbols, then select the field to the right of the numeric base field.
50
Logic Analysis Support for the TI C6000
Chapter 5: Configuring the Logic Analyzer
Loading Symbol Information
To compensate for relocated code
When code segments are relocated, or when memory management units
produce fixed code offsets, you can compensate by using the address offset
field in the Symbol Selector dialog.
Entering the appropriate address offset will cause the logic analyzer to
reference the correct symbol information for the relocatable or offset code.
Logic Analysis Support for the TI C6000
51
Chapter 5: Configuring the Logic Analyzer
Setting Up Labels for Groups of Signals
Setting Up Labels for Groups of Signals
Predefined Label Descriptions
The logic analyzer configuration files automatically set up labels for most
C6000 signals. The following tables show some of the predefined labels for the
most commonly used signals.
ADDR Label
ADDR
Signal Name
Bit 0
..
Bit 21
ADDR[0]
..
ADDR[21]
DATA Label
DATA
Signal Name
Bit 0
...
Bit 31
D[31]
...
D[0]
EXTRA Label
ADDR
Signal Name
Bit 0
Bit 1
ADDR[0]
ADDR[1]
STAT Label
STAT
Signal Name
Bit 0
Bit 1
Bit 2
Bit 3
Bit 4
Bit 5
Bit 6
Bit 7
SDWE/AWE
ARDY
BE[0]
BE[1]
BE[2]
BE[3]
CE[0]
CE[1]
52
Logic Analysis Support for the TI C6000
Chapter 5: Configuring the Logic Analyzer
Setting Up Labels for Groups of Signals
STAT
Signal Name
Bit 8
Bit 9
Bit 10
Bit 11
Bit 12
Bit 13
CE[2]
CE[3]
CLKOUT1
SDRAS/AOE
SDCAS/ARE
ECLKOUT
Clock & Qualifier
CLK
Signal Name
J
K
L
M
CLKOUT1
SDRAS/AOE
SDCAS/ARE
ECLKOUT
To define additional labels
1 Open the Setup window.
2 Click the Format tab.
3 Click a label and select Insert before... or Insert after....
4 Click the signals under the appropriate pod, then select which bits to
include in the label.
Logic Analysis Support for the TI C6000
53
Chapter 5: Configuring the Logic Analyzer
Changing the Analysis Mode
Changing the Analysis Mode
The logic analyzer can be set up to operate in the following analysis modes:
•
State.
•
Timing.
Inverse assembly is available in the state analysis mode.
To change to state analysis
In state mode, the logic analyzer uses the CLKIN signal from the target system
to capture data synchronously. This mode allows inverse assembly and is the
default mode set up by the configuration files.
To configure the logic analyzer for state mode:
• Load the appropriate logic analyzer configuration file (see “To load
configuration files (and the inverse assembler) from hard disk” on
page 36).
You can change the master clock setting by opening the logic analyzer’s Setup
window, selecting the Format tab, and clicking the Master Clock button to open
the master clock dialog.
54
Logic Analysis Support for the TI C6000
Chapter 5: Configuring the Logic Analyzer
Changing the Analysis Mode
To change to timing analysis
In timing mode, the logic analyzer samples the processor pins asynchronously,
according to an internal, adjustable sample rate clock. The minimum sample
period for a 250 MHz timing analyzer is 4 ns.
Inverse assembly is not available in the timing analysis mode.
To configure the logic analyzer for timing analysis:
1 Load the appropriate logic analyzer configuration file (see “To load
configuration files (and the inverse assembler) from hard disk” on
page 36).
2 Open the logic analyzer’s Setup window.
3 Select the Sampling tab.
4 Change the type option from State Mode to Timing Mode.
Logic Analysis Support for the TI C6000
55
Chapter 5: Configuring the Logic Analyzer
Changing the Analysis Mode
56
Logic Analysis Support for the TI C6000
6
Capturing DSP Execution
57
Chapter 6: Capturing DSP Execution
The normal steps in using the logic analyzer are:
1. Configure the logic analyzer.
2. Format labels for the logic analyzer channels (that is, mapping logic
analyzer channels to target system signal names).
3. Load symbols from the program’s object file.
4. Set up the trigger, and run the measurement.
5. Display the captured data.
The inverse assembler is loaded, the logic analyzer is configured, and labels
are created (formatted) for the logic analysis channels when configuration
files are loaded (see “To load configuration files (and the inverse assembler)
from hard disk” on page 36).
You can load program object file symbols into the logic analyzer when
configuring it (see “Loading Symbol Information” on page 47).
This chapter describes setting up logic analyzer triggers when using the
inverse assembler and the Agilent Technologies B4620B Source Correlation
Tool Set.
See Chapter 7, “Displaying Captured DSP Execution,” beginning on page 63
for information on displaying captured data.
58
Logic Analysis Support for the TI C6000
Chapter 6: Capturing DSP Execution
Setting Up Logic Analyzer Triggers
Setting Up Logic Analyzer Triggers
To set up logic analyzer triggers
1 Open the logic analyzer’s Setup window.
2 Select the Trigger tab.
3 Define the patterns, ranges, and other resources that will be used in
the logic analysis measurement.
Logic Analysis Support for the TI C6000
59
Chapter 6: Capturing DSP Execution
Setting Up Logic Analyzer Triggers
4 Set up the trigger sequence.
5 Run the measurement.
See Also
The 16700-series logic analysis system’s on-line help for more information on
setting up logic analyzer triggers.
60
Logic Analysis Support for the TI C6000
Chapter 6: Capturing DSP Execution
Triggering on Source Code
Triggering on Source Code
When setting up trigger specifications to capture program execution:
•
Use the COFF 2.0 object file format.
•
Be sure to correctly configure the memory type.
ASYNC and SBSRAM can use straightforward triggering, but SDRAM
requires a multi-stage trigger. Configuring the memory type will
automatically set up the right kind of source line trigger.
•
Remember that triggers will occur on prefetches.
To avoid capturing library code execution
When viewing the source code associated with captured data, the source
correlation tool set can exhibit long response times to requests for the next
source line if the current trace listing corresponds to code from a library that
is not in the source code search path. Logic analyzer storage qualification can
be used to avoid capturing library code routines.
It is not practical to configure the logic analyzer’s storage qualification
capabilities to exclude idle states, because of prefetching.
Logic Analysis Support for the TI C6000
61
Chapter 6: Capturing DSP Execution
Triggering on Source Code
62
Logic Analysis Support for the TI C6000
7
Displaying Captured DSP Execution
63
Chapter 7: Displaying Captured DSP Execution
To display the captured state data
1 Open the Listing display window.
The logic analyzer displays captured state data in the Listing display.
64
Logic Analysis Support for the TI C6000
Chapter 7: Displaying Captured DSP Execution
The inverse assembler is already loaded when state configuration files are
loaded, but it can also be loaded into a Listing display using the Invasm menu.
The name of the inverse assembler file is IC6000E, and it is located in the
/logic/ia directory.
See Also
“To use the inverse assembler filters” on page 67 for information on displaying
or hiding certain types of bus cycles.
The Agilent Technologies 16700-series logic analysis system on-line help for
information on using the Listing display.
To display symbols
• Over a Listing display’s label base, right-click the mouse button, and
select Symbols.
Any symbols that have been defined will be displayed for equivalent captured
values.
See Also
“To load object file symbols” on page 48.
Logic Analysis Support for the TI C6000
65
Chapter 7: Displaying Captured DSP Execution
To interpret the inverse assembled data
Constants, registers, and instructions are displayed using the usual formats.
Here are a few things to be aware of:
Numbers.
Numerical data is displayed in hexadecimal, with a prefix of “0x”, or decimal,
without a prefix.
You can display offsets and constants in the inverse assembly listing in either
hexadecimal or decimal (see “To set the Decoding Options preferences” on
page 46). You can change the numeric base of the other labels in the listing
(such as ADDR) by selecting the label at the top of the display.
Instructions.
If an opcode is invalid or otherwise not recognized, it is shown as "Unknown
Opcode."
Instructions which are executed in parallel are marked with “||”.
66
Logic Analysis Support for the TI C6000
Chapter 7: Displaying Captured DSP Execution
To use the inverse assembler filters
• In the Listing display window, choose the Filter command from the
Invasm menu.
The inverse assembler filtering options allow you to display or hide certain
Logic Analysis Support for the TI C6000
67
Chapter 7: Displaying Captured DSP Execution
types of bus cycles or memory bank accesses.
Because the filter options do not affect the data that is stored by the logic
analyzer (they only affect whether that data is displayed), they let you display
the same data in different ways.
Filtering allows faster analysis in two ways:
•
Unneeded information can be taken out of the display. For example,
suppressing idle/wait states will let you view more instruction cycles
without scrolling the listing.
•
Particular operations can be isolated by suppressing all other operations.
For example, Branch instructions can be shown, with all other states
suppressed, allowing quick analysis of branch instructions.
You can also use color to distinguish between cycle types or memory bank
accesses (when they are displayed). Color can be used for distinguishing
between memory bank accesses or cycle types, but not both at the same time.
You can display or hide the following types of cycles:
See Also
•
Idle/Wait States.
•
Extension Words.
•
Memory Commands.
•
Branch Instructions.
•
Load/Store Instructions.
•
Other Instructions.
•
Data Reads.
•
Data Writes.
The address ranges for memory banks 0-11 are specified in the Preferences
menu (see “To set the inverse assembler preferences” on page 40).
68
Logic Analysis Support for the TI C6000
Chapter 7: Displaying Captured DSP Execution
To view the source code associated with captured data
• In the Listing display window, select Source Viewer from the Source
menu.
• Or, open the Source Viewer window from the logic analyzer’s icon in
the main system window.
The source correlation tool set correlates the logic analyzer’s address label
with a line of high-level source code whose address, symbol name, file name,
and line numbers are described in a symbol file downloaded to the logic
analyzer (see “To load object file symbols” on page 48).
Logic Analysis Support for the TI C6000
69
Chapter 7: Displaying Captured DSP Execution
Inverse Assembler Generated SW_ADDR (Software
Address) Label
In the Agilent Technologies 16700-series logic analysis system, the inverse
assembler generates a “SW_ADDR” label. The SW_ADDR label is displayed as
another column in the Listing tool. This label is also known as the software
address generated by the inverse assembler.
The “Goto this line in listing” commands in the Agilent Technologies 16700series logic analysis system perform a pattern search on the SW_ADDR label in
the Listing display (when an inverse assembler is loaded). Because the inverse
assembler is called for each line that is searched, the search can be slow,
especially with a deep memory logic analyzer.
Also, a single line of source code will generate many assembly instructions.
The “Goto this line in listing” commands will not find a given line of source
code unless the first assembly instruction generated from the source line has
been acquired by the logic analyzer.
For example, if the compiler unrolls a loop in the source code, the trace could
70
Logic Analysis Support for the TI C6000
Chapter 7: Displaying Captured DSP Execution
begin after the first assembly instruction of the loop has been executed. A
“Goto this line in listing” command would not find the source line.
Access to Source Code Files
The source correlation tool set must be able to access the high-level source
code files referenced by the symbol information so that these source files can
be displayed next to and correlated with the logic analyzer’s execution trace
acquisition. This requires you to be aware of a number of issues.
Source File Search Path. Verify that the correct file search paths for the
source code have been entered into the source correlation tool set. The
Agilent Technologies B4620B Source Correlation Tool Set can often read and
access the correct source code file from information contained in the symbol
file if the source code files have not been moved since they were compiled.
Network Access to Source Files. If source code files are being
referenced across a network, the Agilent Technologies logic analyzer
networking must be compatible with the user’s network environment. Agilent
Technologies logic analyzers currently support Ethernet networks running a
TCP/IP protocol and support ftp, telnet, NFS client/server and X-Window
client/server applications. Some PC networks may require extensions to the
normal LAN protocols to support the TCP/IP protocol and/or these
networking applications. Users should contact their LAN system
administrators to help set up the logic analyzer on their network.
Source File Version Control. If the source code files are under a source
code or version control utility, check the file names and paths carefully. These
utilities can change source code file paths and file names. If these names are
changed from the information contained in the symbol file, the source
correlation tool set will not be able to find the proper source code file. These
version control utilities usually provide an “export” command that creates a
set of source code files with unmodified names. The source correlation tool set
can then be given the correct path to these files.
See Also
More information on configuring and using the source correlation tool set can
be found in the on-line help for your logic analysis system.
Logic Analysis Support for the TI C6000
71
Chapter 7: Displaying Captured DSP Execution
To display captured timing analysis mode data
• Open the Waveform display for your logic analyzer.
You can also use the Waveform display in the state analysis mode to display
state timing diagrams.
72
Logic Analysis Support for the TI C6000
8
General-Purpose ASCII (GPA) Symbol
File Format
73
Chapter 8: General-Purpose ASCII (GPA) Symbol File Format
General-Purpose ASCII (GPA) Symbol File Format
General-Purpose ASCII (GPA) Symbol File
Format
General-purpose ASCII (GPA) format files are loaded into a logic analyzer just
like other object files, but they are usually created differently.
If your compiler does not include symbol information in the output, or if you
want to define a symbol not in the object file, you can create an ASCII format
symbol file.
Typically, ASCII format symbol files are created using text processing tools to
convert compiler or linker map file output that has symbolic information into
the proper format.
You can typically get symbol table information from a linker map file to create
a General-Purpose ASCII (GPA) symbol file.
Various kinds of symbols are defined in different records in the GPA file.
Record headers are enclosed in square brackets; for example, [VARIABLES].
For a summary of GPA file records and associated symbol definition syntax,
refer to the “GPA Record Format Summary” that follows.
Each entry in the symbol file must consist of a symbol name followed by an
address or address range.
While symbol names can be very long, the logic analyzer only uses the first 16
characters.
The address or address range corresponding to a given symbol appears as a
hexadecimal number. The address or address range must immediately follow
the symbol name, appear on the same line, and be separated from the symbol
name by one or more blank spaces or tabs. Ensure that address ranges are in
the following format:
beginning address..ending address
Example
main
test
var1
00001000..00001009
00001010..0000101F
00001E22
#this is a variable
This example defines two symbols that correspond to address ranges and
one point symbol that corresponds to a single address.
74
Logic Analysis Support for the TI C6000
Chapter 8: General-Purpose ASCII (GPA) Symbol File Format
General-Purpose ASCII (GPA) Symbol File Format
For more detailed descriptions of GPA file records and associated symbol
definition syntax, refer to these topics that follow:
•
SECTIONS
•
FUNCTIONS
•
VARIABLES
•
SOURCE LINES
•
START ADDRESS
•
Comments
Logic Analysis Support for the TI C6000
75
Chapter 8: General-Purpose ASCII (GPA) Symbol File Format
General-Purpose ASCII (GPA) Symbol File Format
GPA Record Format Summary
[SECTIONS]
section_name start..end attribute
[FUNCTIONS]
func_name start..end
[VARIABLES]
var_name start [size]
var_name start..end
[SOURCE LINES]
File: file_name
line# address
[START ADDRESS]
address
#Comments
If no record header is specified, [VARIABLES] is assumed. Lines without a
preceding header are assumed to be symbol definitions in one of the
VARIABLES formats.
Example
This is an example GPA file that contains several different kinds of
records:
[SECTIONS]
prog
00001000..0000101F
data
40002000..40009FFF
common
FFFF0000..FFFF1000
[FUNCTIONS]
main
00001000..00001009
test
00001010..0000101F
[VARIABLES]
total
40002000
value
40008000
76
4
4
Logic Analysis Support for the TI C6000
Chapter 8: General-Purpose ASCII (GPA) Symbol File Format
General-Purpose ASCII (GPA) Symbol File Format
[SOURCE LINES]
File: main.c
10
00001000
11
00001002
14
0000100A
22
0000101E
File: test.c
5
00001010
7
00001012
11
0000101A
Logic Analysis Support for the TI C6000
77
Chapter 8: General-Purpose ASCII (GPA) Symbol File Format
General-Purpose ASCII (GPA) Symbol File Format
SECTIONS
[SECTIONS]
section_name
start..end
attribute
Use SECTIONS to define symbols for regions of memory, such as sections,
segments, or classes.
section_name A symbol representing the name of the section.
start The first address of the section, in hexadecimal.
end The last address of the section, in hexadecimal.
attribute This is optional, and may be one of the following:
•
NORMAL (default)—The section is a normal, relocatable section, such as
code or data.
•
NONRELOC—The section contains variables or code that cannot be
relocated; this is an absolute segment.
Define sections first
To enable section relocation, section definitions must appear before any other
definitions in the file.
Example
[SECTIONS]
prog
data
display_io
00001000..00001FFF
00002000..00003FFF
00008000..0000801F
NONRELOC
If you use section definitions in a GPA symbol file, any subsequent function or
variable definitions must be within the address ranges of one of the defined
sections. Functions and variables that are not within the range are ignored.
78
Logic Analysis Support for the TI C6000
Chapter 8: General-Purpose ASCII (GPA) Symbol File Format
General-Purpose ASCII (GPA) Symbol File Format
FUNCTIONS
[FUNCTIONS]
func_name start..end
Use FUNCTIONS to define symbols for program functions, procedures, or
subroutines.
func_name A symbol representing the function name.
start The first address of the function, in hexadecimal.
end The last address of the function, in hexadecimal.
Example
[FUNCTIONS]
main
00001000..00001009
test
00001010..0000101F
Logic Analysis Support for the TI C6000
79
Chapter 8: General-Purpose ASCII (GPA) Symbol File Format
General-Purpose ASCII (GPA) Symbol File Format
VARIABLES
[VARIABLES]
var_name
start [size]
var_name
start..end
You can specify symbols for variables either by using the address of the
variable, the address and the size of the variable, or a range of addresses
occupied by the variable. If you specify only the address of a variable, the size
is assumed to be one byte.
var_name A symbol representing the variable name.
start The first address of the variable, in hexadecimal.
end The last address of the variable, in hexadecimal.
size This is optional, and indicates the size of the variable, in bytes, in decimal.
Example
[VARIABLES]
subtotal
total
data_array
status_char
80
40002000
4
40002004
4
40003000..4000302F
40002345
Logic Analysis Support for the TI C6000
Chapter 8: General-Purpose ASCII (GPA) Symbol File Format
General-Purpose ASCII (GPA) Symbol File Format
SOURCE LINES
[SOURCE LINES]
File: file_name
line# address
Use SOURCE LINES to associate addresses with lines in your source files.
file_name The name of a file.
line# The number of a line in the file, in decimal.
address The address of the source line, in hexadecimal.
Example
[SOURCE LINES]
File: main.c
10
00001000
11
00001002
14
0000100A
22
0000101E
Logic Analysis Support for the TI C6000
81
Chapter 8: General-Purpose ASCII (GPA) Symbol File Format
General-Purpose ASCII (GPA) Symbol File Format
START ADDRESS
[START ADDRESS]
address
address The address of the program entry point, in hexadecimal.
Example
[START ADDRESS]
00001000
Comments
#comment text
Use the # character to include comments in a file. Any text following the #
character is ignored. You can put comments on a line alone or on the same line
following a symbol entry.
Example
#This is a comment.
82
Logic Analysis Support for the TI C6000
9
Specifications and Characteristics
83
Chapter 9: Specifications and Characteristics
Operating Characteristics
The following operating characteristics are not specifications, but are
typical operating characteristics for the E8137A inverse assembler.
Operating Characteristics
Microprocessor Compatibility
Texas Instruments TMS320C6202/6203/6204/
6205/6211 DSP. This product does not support
the TMS320C6201.
Agilent Technologies Logic
Analyzers Supported
16550A
16554/55/56/57
16710/11/12A
16715/16/17/18/19A
16750/51/52A
Accessories Required
For state and timing analysis, the E5346A
high-density cables are required.
Probes Required
At least four 16-channel probes are required
for disassembly.
Electrical Characteristics
See the documentation for the logic analyzer card you are using
84
Logic Analysis Support for the TI C6000
10
Troubleshooting the Logic Analyzer
85
Chapter 10: Troubleshooting the Logic Analyzer
If you encounter difficulties while making measurements, use this chapter to
guide you through some possible solutions. Each heading lists a problem you
may encounter, along with some possible solutions.
If you still have difficulty using the analyzer after trying the suggestions in this
chapter, please contact your local Agilent Technologies service center.
CAUTION:
When you are working with the analyzer, be sure to power down both the
analyzer and the target system before disconnecting or connecting cables, and
probes. Otherwise, you may damage circuitry in the logic analyzer or target
system.
86
Logic Analysis Support for the TI C6000
Chapter 10: Troubleshooting the Logic Analyzer
Solving Logic Analyzer Problems
Solving Logic Analyzer Problems
This section lists general problems that you might encounter while using the
analyzer.
Intermittent data errors
This problem is usually caused by poor connections, incorrect signal levels, or
marginal timing.
❏ Remove and reseat all cables and probes, ensuring that there are no
bent pins or poor connections.
❏ Adjust the threshold level of the data pod to match the logic levels in
the system under test.
❏ Use an oscilloscope to check the signal integrity of the data lines.
Clock signals for the state analyzer must meet particular pulse shape and
timing requirements. Data inputs for the analyzer must meet pulse shape and
setup and hold time requirements.
See Also
See “Capacitive loading” on page 90 for information on other sources of
intermittent data errors.
Unwanted triggers
Unwanted triggers can be caused by instructions that were fetched but not
executed.
❏ Set the trigger at an address which is not the target of a conditional
branch to avoid this problem. For example, set the trigger in the middle
of a subroutine, rather than at the first instruction of a subroutine.
The logic analyzer captures prefetches, even if they are not executed. When
you are specifying a trigger condition or a storage qualification that follows an
instruction that may cause branching, an unused prefetch may generate an
Logic Analysis Support for the TI C6000
87
Chapter 10: Troubleshooting the Logic Analyzer
Solving Logic Analyzer Problems
unwanted trigger.
No activity on activity indicators
❏ Check for loose cables or board connections.
❏ Check for bent or damaged pins.
No trace list display
If there is no trace list display, it may be that your trigger specification is not
correct for the data you want to capture, or that the trace memory is only
partially filled.
❏ Check your trigger sequencer specification to ensure that it will
capture the events of interest.
❏ Try stopping the analyzer; if the trace list is partially filled, this should
display the contents of trace memory.
88
Logic Analysis Support for the TI C6000
Chapter 10: Troubleshooting the Logic Analyzer
Solving Probing Problems
Solving Probing Problems
This section lists probing problems that you might encounter when using a
logic analyzer. If the solutions suggested here do not correct the problem, you
may have a damaged logic analyzer. Contact your local Agilent Technologies
Sales Office if you need further assistance.
Target system will not boot up
If the target system will not boot up after connecting the target system, the
processor (if socketed) or the probe cables may not be installed properly, or
they may not be making electrical contact.
❏ Ensure that you are following the correct power-on sequence for the
logic analyzer and target system.
1. Power up the logic analyzer.
2. Power up the target system.
If you power up the target system before you power up the logic analyzer,
interface circuitry may latch up and prevent proper target system operation.
❏ Verify that the logic analyzer cables are in the proper target system
connector headers and are firmly inserted.
Logic Analysis Support for the TI C6000
89
Chapter 10: Troubleshooting the Logic Analyzer
Solving Probing Problems
Erratic trace measurements
There are several general problems that can cause erratic variations in trace
lists and inverse assembly failures.
❏ Do a full reset of the target system before beginning the measurement.
Some designs require a full reset to ensure correct configuration.
❏ Ensure that your target system meets the timing requirements of the
processor with the logic analyzer connected.
See “Capacitive loading” on page 90. While logic analyzer probe loading is
slight, pin protectors, extenders, and adapters may increase it to unacceptable
levels. If the target system design has close timing margins, such loading may
cause incorrect processor functioning and give erratic trace results.
❏ Ensure that you have sufficient cooling for the processor.
Ensure that you have ambient temperature conditions and airflow that meet
or exceed the requirements of the processor manufacturer.
Capacitive loading
Excessive capacitive loading can degrade signals, resulting in incorrect
capture by the logic analyzer, or system lockup in the processor. All logic
analyzer probes add additional capacitive loading, as can custom probe
fixtures you design for your application.
Careful layout of your target system can minimize loading problems and result
in better margins for your design. This is especially important for systems that
are running at frequencies greater than 50 MHz.
❏ Remove as many pin protectors, extenders, and adapters as possible.
90
Logic Analysis Support for the TI C6000
Chapter 10: Troubleshooting the Logic Analyzer
Solving Inverse Assembler Problems
Solving Inverse Assembler Problems
This section lists problems that you might encounter while using the inverse
assembler.
When you obtain incorrect inverse assembly results, it may be unclear
whether the problem is in the logic analyzer or in your target system. If you
follow the suggestions in this section to ensure that you are using the logic
analyzer and inverse assembler correctly, you can proceed with confidence in
debugging your target system.
No inverse assembly or incorrect inverse assembly
This problem may be due to incorrect synchronization, modified
configuration, incorrect connections, or a hardware problem in the target
system. A locked status line can cause incorrect or incomplete inverse
assembly.
❏ Ensure that each logic analyzer pod is connected to the correct
connector.
There is not always a one-to-one correspondence between analyzer pod
numbers and probe cable numbers. Probes must supply address (ADDR), data
(DATA), and status (STAT) information to the analyzer in a predefined order.
The cable connections for each probe are often altered to support that need.
Thus, one probe might require that you connect cable 2 to analyzer pod 2,
while another will require you to connect cable 5 to analyzer pod 2. See
“Connecting the Logic Analyzer to the Target System” on page 32 for
connection information.
❏ Check the activity indicators for status lines locked in a high or low
state.
❏ Verify that the STAT, DATA, and ADDR format labels have not been
modified from their default values.
These labels must remain as they are configured by the configuration file. Do
not change the names of these labels or the bit assignments within the labels.
Some inverse assemblers also require other data labels. See “Configuring the
Logic Analyzer” on page 33 for more information.
Logic Analysis Support for the TI C6000
91
Chapter 10: Troubleshooting the Logic Analyzer
Solving Inverse Assembler Problems
❏ Verify that all processor caches and memory managers have been
disabled.
In most cases, if the processor caches and memory managers remain enabled
you should still get inverse assembly; however, it will consist of cache line fills.
❏ Verify that storage qualification has not excluded storage of all the
needed opcodes and operands.
❏ Check that the inverse assembler preferences (especially the memory
map) have been set correctly.
Inverse assembler will not load or run
You need to ensure that you have the correct system software loaded on your
analyzer.
❏ Ensure that the inverse assembler is on the same disk as the
configuration files you are loading.
Configuration files for the state analyzer contain a pointer to the name of the
corresponding inverse assembler. If you delete the inverse assembler or
rename it, the configuration process will fail to load the disassembler.
See “Configuring the Logic Analyzer” on page 33 for details.
92
Logic Analysis Support for the TI C6000
Chapter 10: Troubleshooting the Logic Analyzer
Solving Intermodule Measurement Problems
Solving Intermodule Measurement Problems
Some problems occur only when you are trying to make a measurement
involving multiple modules.
An event wasn’t captured by one of the modules
If you are trying to capture an event that occurs very shortly after the event
that arms one of the measurement modules, it may be missed due to internal
analyzer delays. For example, suppose you set the oscilloscope to trigger upon
receiving a trigger signal from the logic analyzer because you are trying to
capture a pulse that occurs right after the analyzer’s trigger state. If the pulse
occurs too soon after the analyzer’s trigger state, the oscilloscope will miss the
pulse.
❏ Adjust the skew in the Intermodule menu.
You may be able to specify a skew value that enables the event to be captured.
❏ Change the trigger specification for modules upstream of the one with
the problem.
If you are using a logic analyzer to trigger the oscilloscope, try specifying a
trigger state one state before the one you are using. This may be more difficult
than working with the skew because the prior state may occur more often and
not always be related to the event you are trying to capture with the
oscilloscope.
Logic Analysis Support for the TI C6000
93
Chapter 10: Troubleshooting the Logic Analyzer
Logic Analyzer Messages
Logic Analyzer Messages
This section lists some of the messages that the analyzer displays when it
encounters a problem.
“. . . Inverse Assembler Not Found”
This error occurs if you rename or delete the inverse assembler file that is
attached to the configuration file.
Ensure that the inverse assembler file is not renamed or deleted, and that it is
located in the correct directory:
•
For Agilent 16700-series logic analysis systems it should be in
/logic/ia.
•
For other logic analyzers it should be in the same directory as the
configuration file.
“No Configuration File Loaded”
This is usually caused by trying to load a configuration file for one type of
module/system into a different type of module/system.
❏ Verify that the appropriate module has been selected from the Load
{module} from File {filename} in the disk operation menu. Selecting
Load {All} will cause incorrect operation when loading most
configuration files.
See Also
See “Configuring the Logic Analyzer” on page 33 for a description of how to
load configuration files.
94
Logic Analysis Support for the TI C6000
Chapter 10: Troubleshooting the Logic Analyzer
Logic Analyzer Messages
“Selected File is Incompatible”
This occurs when you try to load a configuration file for the wrong module.
Ensure that you are loading the appropriate configuration file for your logic
analyzer.
“Slow or Missing Clock”
❏ This error message might occur if the logic analyzer cards are not
firmly seated in the Agilent Technologies 16700-series logic analysis
system frame or in the Agilent Technologies 16701A expansion frame.
Ensure that the cards are firmly seated.
❏ This error might occur if the target system is not running properly.
Ensure that the target system is on and operating properly.
❏ If the error message persists, check that the logic analyzer pods are
connected to the proper connectors on the target system. See
“Connecting the Logic Analyzer to the Target System” on page 32 to
determine the proper connections.
“Time from Arm Greater Than 41.93 ms”
The state/timing analyzers have a counter to keep track of the time from when
an analyzer is armed to when it triggers. The width and clock rate of this
counter allow it to count for up to 41.93 ms before it overflows. Once the
counter has overflowed, the system does not have the data it needs to
calculate the time between module triggers. The system must know this time
to be able to display data from multiple modules on a single screen.
Logic Analysis Support for the TI C6000
95
Chapter 10: Troubleshooting the Logic Analyzer
Logic Analyzer Messages
“Waiting for Trigger”
If a trigger pattern is specified, this message indicates that the specified
trigger pattern has not occurred. Verify that the triggering pattern is correctly
set.
❏ When analyzing processors that fetch only from word-aligned
addresses, if the trigger condition is set to look for an opcode fetch at
an address not corresponding to a word boundary, the trigger will
never be found.
96
Logic Analysis Support for the TI C6000
Glossary
Analysis Probe A probing solution
connected to the target
microprocessor. It provides an
interface between the signals of the
target microprocessor and the inputs
of the logic analyzer. Formerly called
a “preprocessor.”
Background Debug Monitor Also
called Debug Mode, In Background,
and In Monitor. The normal processor
execution is suspended and the
processor waits for commands from
the debug port. The debug port
commands include the ability to read
and write memory, read and write
registers, set breakpoints and start
the processor running (exit the
Background Debug Monitor).
Debug Mode See Background
Debug Monitor.
Debug Port A hardware interface
designed into a microprocessor that
allows developers to control
microprocessor execution, set
breakpoints, and access
microprocessor registers or target
system memory using a tool like the
emulation probe.
Elastomeric Probe Adapter A
connector that is fastened on top of a
target microprocessor using a
retainer and knurled nut. The
conductive elastomer on the bottom
of the probe adapter makes contact
with pins of the target
microprocessor and delivers their
signals to connection points on top of
the probe adapter.
Emulation Migration The
hardware and software required to
use an emulation probe with a new
processor family.
Emulation Module An emulation
module is installed within the
mainframe of a logic analysis system.
An E5901A emulation module is used
with a target interface module
(TIM) or an analysis probe. An
E5901B emulation module is used
with an E5900B emulation probe
and does not use a TIM.
Emulation Probe An emulation
probe is a standalone instrument
connected via LAN to the mainframe
of a logic analyzer or to a host
computer. It provides run control
within an emulation and analysis test
setup. Formerly called a "processor
probe" or "software probe."
Emulation Solution A set of tools
for debugging your target system. A
solution includes probing, inverse
assembly, the B4620B Source
Correlation Tool Set, and an
emulation module.
97
Glossary
Emulator An emulation module or
an emulation probe.
general-purpose flexible adapter to
the analysis probe.
Extender A part whose only
function is to provide connections
from one location to another. One or
more extenders might be stacked to
raise a probe above a target
microprocessor to avoid mechanical
contact with other components
installed close to the target
microprocessor. Sometimes called a
"connector board."
High-Density Adapter Cable A
cable assembly that delivers signals
from an analysis probe hardware
interface to the logic analyzer pod
cables. A high-density adapter cable
has a single MICTOR connector that
is installed into the analysis probe,
and two cables that are connected to
corresponding odd and even logic
analyzer pod cables.
Flexible Adapter Two connection
devices coupled with a flexible cable.
Used for connecting probing
hardware on the target
microprocessor to the analysis probe.
High-Density Termination
Adapter Cable Same as a HighDensity Adapter Cable, except it has
a termination in the MICTOR
connector.
Gateway Address An IP address
entered in integer dot notation. The
default gateway address is 0.0.0.0,
which allows all connections on the
local network or subnet. If
connections are to be made across
networks or subnets, this address
must be set to the address of the
gateway machine.
In Background, In Monitor See
Background Debug Monitor.
General-Purpose Flexible
Adapter A cable assembly that
connects the signals from an
elastomeric probe adapter to an
analysis probe. Normally, a male-tomale header or transition board
makes the connections from the
IP address Also called Internet
Protocol address or Internet address.
A 32-bit network address. It is usually
represented as decimal numbers
separated by periods; for example,
192.35.12.6.
98
Inverse Assembler Software that
displays captured bus activity as
assembly language mnemonics. In
addition, inverse assemblers may
show execution history or decode
control busses.
Glossary
Jumper Moveable direct electrical
connection between two points.
JTAG (OnCE) port See debug
port.
Label Labels are used to group and
identify logic analyzer channels. A
label consists of a name and an
associated bit or group of bits.
Link-Level Address The unique
address of the LAN interface. This
value is set at the factory and cannot
be changed. The link-level address of
a particular piece of equipment is
often printed on a label above the
LAN connector. An example of a linklevel address in hexadecimal:
0800090012AB. Also known as an
LLA, Ethernet address, hardware
address, physical address, or MAC
address.
Mainframe Logic Analyzer A logic
analyzer that resides on one or more
board assemblies installed in a 16500,
1660-series, or 16600/700-series
mainframe.
Male-to-male Header A board
assembly that makes point-to-point
connections between the female pins
of a flexible adapter or transition
board and the female pins of an
analysis probe.
MICTOR Connector A high-density
matched impedance connector
manufactured by AMP Corporation.
High-density adapter cables can be
used to connect the logic analyzer to
MICTOR connectors on the target
system.
Monitor, In See Background
Debug Monitor.
Pod A collection of logic analyzer
channels associated with a single
cable and connector.
Preprocessor See Analysis Probe.
Preprocessor Interface See
Analysis Probe.
Probe Adapter See Elastomeric
Probe Adapter.
Processor Probe See Emulation
Probe.
Run Control Probe See
Emulation Probe and Emulation
Module.
Setup Assistant Wizard software
program which guides a user through
the process of connecting and
configuring a logic analyzer to make
measurements on a specific
microprocessor. The setup assistant
icon is located in the main system
99
Glossary
window.
separated by periods; for example,
255.255.255.0.
Shunt Connector. See Jumper.
Solution A set of tools for debugging
your target system. A solution
includes probing, inverse assembly,
the B4620B Source Correlation Tool
Set, and (optionally) an emulation
module.
Stand-Alone Logic Analyzer A
standalone logic analyzer has a
predefined set of hardware
components which provide a specific
set of capabilities. A standalone logic
analyzer differs from a mainframe
logic analyzer in that it does not offer
card slots for installation of additional
capabilities, and its specifications are
not modified based upon selection
from a set of optional hardware
boards that may be installed within
its frame.
State Analysis A mode of logic
analysis in which the logic analyzer is
configured to capture data
synchronously with a clock signal in
the target system.
Subnet Mask A subnet mask blocks
out part of an IP address so the
networking software can determine
whether the destination host is on a
local or remote network. It is usually
represented as decimal numbers
100
Symbol Symbols represent patterns
and ranges of values found on labeled
sets of bits. Two kinds of symbols are
available:
1) Object file symbols — Symbols
from your source code, and symbols
generated by your compiler. Object
file symbols may represent global
variables, functions, labels, and
source line numbers.
2) User-defined symbols — Symbols
you create.
Target Board Adapter A daughter
board inside the E5900B emulation
probe which customizes the
emulation probe for a particular
microprocessor family. The target
board adapter provides an interface
to the ribbon cable which connects to
the debug port on the target system.
Target Control Port An 8-bit, TTL
port on a logic analysis system that
you can use to send signals to your
target system. It does not function
like a pattern generator or emulation
module, but more like a remote
control for the target’s switches.
Target Interface Module A small
circuit board which connects the 50pin cable from an E5901A emulation
module or E5900A emulation probe
Glossary
to signals from the debug port on a
target system. Not used with the
E5900B emulation probe.
TIM See Target Interface Module.
Timing Analysis A mode of logic
analysis in which the logic analyzer is
configured to capture data at a rate
determined by an internal sample
rate clock, asynchronous to signals in
the target system.
Transition Board A board
assembly that obtains signals
connected to one side and rearranges
them in a different order for delivery
at the other side of the board.
Trigger Specification A set of
conditions that must be true before
the instrument triggers. See the
printed or online documentation of
your logic analyzer for details.
1/4-Flexible Adapter An adapter
that obtains one-quarter of the
signals from an elastomeric probe
adapter (one side of a target
microprocessor) and makes them
available for probing.
101
Glossary
102
Index
A
access to source code files, 71
accessories required, 84
active low signal, 21
activity indicators, 88, 91
adapter, 19
adapters, 90
ADDR label, 48, 91
address offset field, 51
address range
memory banks 0-11, 68
addresses
mask, 39
Agilent Technologies 1252-7431,
19
Agilent Technologies 16550A logic
analyzer, 38
Agilent Technologies 16554A logic
analyzer, 38
Agilent Technologies 16710A logic
analyzer, 38
Agilent Technologies 16715/16/17A
logic analyzers, 38
Agilent Technologies B4620B
source correlation tool set, 58
Agilent Technologies E5346-44701
shroud, 19
Agilent Technologies E5346-60002
adapter, 19
Agilent Technologies E5346A highdensity termination adapter
cable, 19
Agilent Technologies E9611A
Option 001 inverse assembler,
14
analysis mode
changing, 54
state, 54, 72
timing, 54
analysis probe
definition, 97
arm, 93, 95
ASCII format (GPA), 74
assembly language mnemonics, 35
asynchronous sampling, 55
B
B4620B source correlation tool set,
2, 28
background debug monitor, 97
bank enable/disable, 42
base address, 42
blank pins, 21
board space, 19
branch instructions, 68
bus cycles, 68
C
cache memory, 18
caches, 92
capacitive loading, 90
captured data, source code
associated with, 69
captured execution, displaying, 63
capturing execution, 57
cards
See logic analyzers
CD-ROM, 14, 16
CD-ROM, installing software from,
29
center inline pins, 21
characteristics, 83
CLKIN signal, 54
clock signals, 87
color, 68
comments, in GPA files, 82
configuration files, 14, 28, 38, 47,
54, 55, 58, 65, 92, 94, 95
configuring the logic analyzer, 35
connection notes
recommended configuration, 21
connector board, 98
connector layout, 21
recommended, 21
cooling, 90
Logic Analysis Support for the TI C6000
counter overflow, 95
creating GPA symbol files, 74
custom probe fixtures, 90
cycle types, 68
D
data cache, 18
DATA label, 91
data reads, 68
data writes, 68
debug mode, 97
debug port, 97
definition, 97
decoding
simplified mnemonic, 46
decoding options, 46
deep memory logic analyzer, 70
design considerations, 18
directories
configuration files, 36
display timing analysis mode data,
72
displaying captured execution, 63
download symbol information, 47
E
E8126A inverse assembler, 2
E9503A emulation solution, 2
E9603A Option 001 inverse
assembler, 2
elastomeric probe adapter
definition, 97
electrical characteristics, 84
Emulation Control Interface, 28
emulation migration
definition, 97
emulation module
definition, 97
emulation probe
definition, 97
emulation solution, 2
definition, 97
103
Index
equipment, 11
erratic trace measurements, 90
Ethernet networks, 71
event wasn’t captured, 93
extender, 98
extenders, 90
extension words, 68
F
filters, inverse assembler, 67
fixed code offsets, 51
flexible adapter, definition, 98
frequencies greater than 50 MHz,
90
ftp, 71
FUNCTIONS in GPA format, 79
G
gateway address, definition, 98
General-Purpose ASCII (GPA)
symbol file, 48
General-Purpose ASCII format, 74
address format, 74
comments, 82
FUNCTIONS, 79
record format summary, 76
record headers, 74
SECTIONS, 78
simple form, 74
SOURCE LINES, 81
START ADDRESS, 82
VARIABLES, 80
general-purpose flexible adapter
definition, 98
ground returns, 21
H
high-density adapter cable
definition, 98
high-density connector
mechanical specifications, 20
pin assignment, 20
104
high-density connectors, 19
high-density termination adapter
definition, 98
high-level source code, 69
I
idle/wait states, 68
incorrect inverse assembly, 91
incorrect signal levels, 87
information sources, 4
inline pins, 21
input voltage, maximum, 18
installing analyzer modules, 27
installing software, 28
intermittent data errors, 87
intermodule measurement
problems, 93
internal analyzer delays, 93
internal data cache, 18
internal instruction cache, 18
internal sample rate clock, 55
invalid opcode, 66
Invasm menu, 65
inverse assembled data, 66
inverse assembler, 18, 28, 35, 39,
65, 70, 92
definition, 98
equipment supplied, 14
file name, 94
filters, 67
loading files, 36, 37
not found, 94
preferences, 40
problems, 91
requirements for, 39
software, 14
will not load, 92
inverse assembly, 39, 54, 92
incorrect, 91
traditional, 39
IP address
definition, 98
J
jumper, definition, 99
L
labels
defining, 53
definition, 99
predefined, 52
LAN protocols, 71
LAN system administrators, 71
layout, 90
library code execution, 61
link-level address
definition, 99
Listing display window, 64
Load menu, 39
load/store instructions, 68
loading configurations, vs.
installing, 36
loading object file symbols, 48
loading symbol information, 47
loading, minimum, 18
locked status line, 91
logic analysis
preparing target for, 18
logic analysis system
setting up, 25
software version, 16
logic analyzer, 86
configuration files, 38, 47
configuring, 33
connecting to target system, 32
deep memory, 70
maximum input voltage, 18
messages, 94
modules, installing, 27
pods, 91, 95
solving problems, 87
storage qualification, 61
trigger setup, 59
triggers, 58
troubleshooting, 85
Logic Analysis Support for the TI C6000
Index
logic analyzers
configuring, 36, 37
supported, 16, 84
M
mainframe logic analyzer
definition, 99
male-to-male header
definition, 99
marginal timing, 87
master clock dialog, 54
maximum input voltage, 18
mechanical specifications
high-density connector, 20
memory bank, 68
accesses, 68
memory banks, 39
memory management units, 51
memory managers, 92
memory map, 41
memory map information, 41
messages
logic analyzer, 94
Mictor (Matched Impedance
ConnecTOR) connectors, 19
MICTOR connector, definition, 99
minimum loading, 18
mnemonics, assembly language, 35
monitor, background debug, 97
N
network access to source files, 71
NFS client/server, 71
no configuration file loaded, 94
no inverse assembly, 91
no-connect, 21
numerical data, 66
O
object file formats, 47
object file symbols, 58
loading, 48
opcode fetch, 96
Options menu, 39
oscilloscope, 87, 93
modules, installing, 27
other instructions, 68
other options, 45, 46
P
pattern generator modules,
installing, 27
patterns, 59
PCI analysis, 16
personality files, 28
pin protectors, 90
plastic shroud, 19
pods, logic analyzer, 91, 95, 99
poor connections, 87
power-on sequence, 89
power-ON/OFF sequence, 26
predefined symbols, 47
preferences, inverse assembler, 40
prefetches, 87
preprocessor
See analysis probe
preprocessor interface
See analysis probe
probe fixtures, custom, 90
probed signal lines, 18
probes required, 84
probing the target system, 31
problems
intermodule measurement, 93
inverse assembler, 91
logic analyzer, 87
probing, 89
processor caches, 92
processor compatibility, 84
processor options, 44
processor support package, 28, 29
pulse shape requirements, 87
Logic Analysis Support for the TI C6000
R
ranges, 59
recommended circuit board
routing, 20
recommended configuration
connection notes, 21
recommended connector layout,
21
recommended signal routing, 21,
22
record format, General-Purpose
ASCII, 76
record headers, 74
references, 4
relocated code, compensating for,
51
requirements, 11
resources, 59
run control tool
See emulation control interface
S
sample period, 55
sample rate clock, internal, 55
search path, source code, 61, 71
Section Format, 74
SECTIONS in GPA format, 78
selected file is incompatible, 95
setup and hold time requirements,
87
Setup Assistant, 12, 26, 28
setup assistant
definition, 99
Setup window, 47
setup/hold window, 18
shroud, 19, 20
signal ground returns, 21
signal integrity, 19, 87
signal routing
recommended, 22
simplified instructions, 46
simplified mnemonics, 46
105
Index
skew, 93
slow or missing clock, 95
software
list of installed, 38
software addresses, 35, 70
software supplied, 14
software version, 16
software, installing, 28
solution, definition, 100
source code, 69
associated with captured data, 69
search path, 61
triggering on, 61
source correlation tool set, 2, 69,
71
source files
network access to, 71
search path, 71
version control, 71
SOURCE LINES in GPA format, 81
specifications, 83
stand-alone logic analyzer
definition, 100
START ADDRESS in GPA format,
82
STAT label, 91
state analysis, 100
definition, 100
state mode, 54, 72
changing to, 54
status lines, locked, 91
storage qualification, 61, 87, 92
subnet mask
definition, 100
supplied equipment and software,
14
support shroud, 20
supported logic analyzers, 16
suppressing all other operations,
68
surface mount connector, 20
SW_ADDR label, 35, 70
106
symbol file, 69
symbol files
creating, 74
symbol information
loading, 47
Symbol Selector dialog, 50
symbols
definition, 100
predefined, 47
Symbols tab, 47
symbols, displaying, 65
synchronization, 91
system administrators, 71
trigger specification
definition, 101
trigger tool, 47
triggering on source code, 61
troubleshooting, 85
T
target board adapter
definition, 100
target control port, 100
target interface module (TIM)
definition, 100
target system, 86, 91
connecting logic analyzer to, 32
preparing, 17
probing, 31
won’t boot up, 89
TCP/IP protocol, 71
telnet, 71
threshold level, 87
time from arm greater than 41.93
ms, 95
timing analysis, 101
definition, 101
timing analysis mode, 54
changing to, 55
data, displaying, 72
timing requirements, 87, 90
transition board
definition, 101
trigger condition, 87
trigger pattern, 96
trigger sequence, 60
trigger sequencer specification, 88
U
undefined opcode, 66
unneeded information, 68
unused prefetch, 87
unwanted triggers, 87
user defined signals, 21
User Defined Symbols tab, 47
user-defined symbols, 47
V
VARIABLES in GPA format, 80
version control, source file, 71
version, software, 16
W
waiting for trigger, 96
Waveform display, 72
web sites
logic analyzers, 4
See Also under debugger names
word-aligned addresses, 96
X
X-Window client/server, 71
Y
Y-cable, 19
Logic Analysis Support for the TI C6000
© Copyright Agilent Technologies
1994-2000
All Rights Reserved.
Reproduction, adaptation, or
translation without prior written
permission is prohibited, except
as allowed under the copyright
laws.
Restricted Rights Legend
Use, duplication, or disclosure by
the U.S. Government is subject to
restrictions set forth in
subparagraph (C) (1) (ii) of the
Rights in Technical Data and
Computer Software Clause in
DFARS 252.227-7013. Agilent
Technologies, 3000 Hanover
Street, Palo Alto, CA 94304 U.S.A.
Rights for non-DOD U.S.
Government Departments and
Agencies are set forth in FAR
52.227-19 (c) (1,2).
Document Warranty
The information contained in this
document is subject to change
without notice.
Agilent Technologies makes
no warranty of any kind with
regard to this material,
including, but not limited to,
the implied warranties of
merchantability or fitness for
a particular purpose.
Agilent Technologies shall not be
liable for errors contained herein
or for damages in connection with
the furnishing, performance, or
use of this material.
Safety
This apparatus has been designed
and tested in accordance with
IEC Publication 1010, Safety
Requirements for Measuring
Apparatus, and has been supplied
in a safe condition. This is a
Safety Class I instrument
(provided with terminal for
protective earthing). Before
applying power, verify that the
correct safety precautions are
taken (see the following
warnings). In addition, note the
external markings on the
instrument that are described
under "Safety Symbols."
Warning
• Before turning on the
instrument, you must connect the
protective earth terminal of the
instrument to the protective
conductor of the (mains) power
cord. The mains plug shall only be
inserted in a socket outlet
provided with a protective earth
contact. You must not negate the
protective action by using an
extension cord (power cable)
without a protective conductor
(grounding). Grounding one
conductor of a two-conductor
outlet is not sufficient protection.
• Only fuses with the required
rated current, voltage, and
specified type (normal blow, time
delay, etc.) should be used. Do
not use repaired fuses or shortcircuited fuseholders. To do so
could cause a shock of fire hazard.
Agilent Technologies
P.O. Box 2197
1900 Garden of the Gods Road
Colorado Springs, CO 80901-2197, U.S.A.
• Service instructions are for
trained service personnel. To
avoid dangerous electric shock,
do not perform any service unless
qualified to do so. Do not attempt
internal service or adjustment
unless another person, capable of
rendering first aid and
resuscitation, is present.
• If you energize this instrument
by an auto transformer (for
voltage reduction), make sure the
common terminal is connected to
the earth terminal of the power
source.
• Whenever it is likely that the
ground protection is impaired,
you must make the instrument
inoperative and secure it against
any unintended operation.
• Do not operate the instrument
in the presence of flammable
gasses or fumes. Operation of any
electrical instrument in such an
environment constitutes a definite
safety hazard.
• Do not install substitute parts or
perform any unauthorized
modification to the instrument.
• Capacitors inside the
instrument may retain a charge
even if the instrument is
disconnected from its source of
supply.
Safety Symbols
!
Instruction manual symbol: the
product is marked with this
symbol when it is necessary for
you to refer to the instruction
manual in order to protect against
damage to the product.
Hazardous voltage symbol.
Earth terminal symbol: Used to
indicate a circuit common
connected to grounded chassis.
WARNING
The Warning sign denotes a
hazard. It calls attention to a
procedure, practice, or the like,
which, if not correctly performed
or adhered to, could result in
personal injury. Do not proceed
beyond a Warning sign until the
indicated conditions are fully
understood and met.
CAUTION
The Caution sign denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like, which, if not correctly
performed or adhered to, could
result in damage to or destruction
of part or all of the product. Do
not proceed beyond a Caution
symbol until the indicated
conditions are fully understood or
met.
Product Warranty
This Agilent Technologies
product has a warranty against
defects in material and
workmanship for a period of one
year from date of shipment.
During the warranty period,
Agilent Technologies Company
will, at its option, either repair or
replace products that prove to be
defective.
For warranty service or repair,
this product must be returned to
a service facility designated by
Agilent Technologies.
For products returned to Agilent
Technologies for warranty
service, the Buyer shall prepay
shipping charges to Agilent
Technologies and Agilent
Technologies shall pay shipping
charges to return the product to
the Buyer. However, the Buyer
shall pay all shipping charges,
duties, and taxes for products
returned to Agilent Technologies
from another country.
Agilent Technologies warrants
that its software and firmware
designated by Agilent
Technologies for use with an
instrument will execute its
programming instructions when
properly installed on that
instrument. Agilent Technologies
does not warrant that the
operation of the instrument
software, or firmware will be
uninterrupted or error free.
Limitation of Warranty
The foregoing warranty shall not
apply to defects resulting from
improper or inadequate
maintenance by the Buyer,
Buyer- supplied software or
interfacing, unauthorized
modification or misuse, operation
outside of the environmental
specifications for the product, or
improper site preparation or
maintenance.
No other warranty is
expressed or implied. Agilent
Technologies specifically
disclaims the implied
warranties of merchantability
or fitness for a particular
purpose.
Exclusive Remedies
The remedies provided herein are
the buyer’s sole and exclusive
remedies. Agilent Technologies
shall not be liable for any direct,
indirect, special, incidental, or
consequential damages, whether
based on contract, tort, or any
other legal theory.
Assistance
Product maintenance agreements
and other customer assistance
agreements are available for
Agilent Technologies products.
For any assistance, contact your
nearest Agilent Technologies
Sales Office.
Certification
Agilent Technologies certifies that
this product met its published
specifications at the time of
shipment from the factory.
Agilent Technologies further
certifies that its calibration
measurements are traceable to
the United States National
Institute of Standards and
Technology, to the extent allowed
by the Institute’s calibration
facility, and to the calibration
facilities of other International
Standards Organization members.
About this edition
This is the Logic Analysis
Support for the Texas
Instruments TMS320C6000
Family User’s Guide.
Publication number
E8137-97000, October 2000
Printed in USA.
Print history is as follows:
Many product updates do not
require manual changes, and
manual corrections may be done
without accompanying product
changes. Therefore, do not expect
a one-to-one correspondence
between product updates and
manual updates.
Was this manual useful for you? yes no