DX-4015
DX-4015
Multicomponent FTIR Gas Analyzer
General parameters
Gasmet On-site Series includes portable multicomponent gas
analyzers for demanding applications. The Gasmet DX-4015
features a Fourier transform infrared (FTIR) spectrometer, a
temperature controlled sample cell with a built-in pump, and
signal processing electronics. The sample cell can be heated
to 50 °C. Sample cell absorption path length is selected
according to the application. This analyzer offers versatility and
high performance for different applications.
Measuring principle:
Fourier transform infrared, FTIR
Performance:
Simultaneous analysis of up to 50
gas compounds
Response time, T90:
Typically < 120 s, depending on
the gas flow and measurement
time
Operating temperature:
The Gasmet DX-4015 is designed for on site measurements at
low concentrations in ambient air. Typical usage areas include
industrial hygiene and emergency response situations.
Short term 0 - 40°C
long term 5 - 30°C
non-condensing
Storage temperature:
-20 - 60°C, non condensing
Power supply:
100-115 or 230 V / 50 -60 Hz
12 VDC
The Gasmet DX-4015 is factory calibrated using certified
single component calibration gases. There is no need for the
user to do any span calibrations after that. The user can also
easily configure the analyzer for a new set of compounds.
Power consumption:
Average 150 W, maximum 300 W
Spectrometer
Resolution:
-1
-1
8 cm or 4 cm
Scan frequency:
10 scans / s
Detector:
Peltier cooled MCT
Source:
SiC, 1550 K
Beamsplitter:
ZnSe
Window material:
ZnSe
Wave number range:
900 - 4 200 cm
-1
Sample cell
Analysis software (for external PC)
Structure:
Multi-pass, fixed path length 9.8 m
Operating system:
Windows XP
Material:
100 % rhodium coated aluminium
Analysis software:
Calcmet for Windows
Mirrors:
Fixed, protected gold coating
Volume:
0.4 liters
Options
Connectors:
Inlets Swagelok 6 mm
Outlet Swagelok 6 mm
Sample cell:
Multi-pass, fixed path
2.5 m or 5.0 m
Gaskets:
Viton O-rings
Pressure measurement:
Inside sample cell
Temperature:
50 °C, maximum
Valve:
Manual Swagelok
Window material:
Ar coated ZnSe
®
Measuring parameters
Analog signals (ext PC): ADAM 5000/TCP module (for
analog inputs, outputs, relays)
®
®
Teflon coated Viton or Kalrez
Power connection:
12 VDC
Power supply cables:
12 V cables with battery clips or
cigarette lighter connector
Wheeled cart for the analyzer and
laptop computer
Zero point calibration:
24 hours, calibration with nitrogen
(4.0 or higher N2 recommended)
Trolley:
Zero point drift:
< 2 % of measuring range per zero
point calibration interval
Enclosure
Sensitivity drift:
none
Material:
Aluminium
Linearity deviation:
< 2 % of measuring range
Dimensions (mm):
438 * 164 * 445
Temperature drifts:
< 2 % of measuring range per 10
K temperature change
Weight:
14.9 kg
CE label:
According to EMI guideline
89/336/EC
Pressure influence:
1 % change of measuring value
for 1 % sample pressure change.
Ambient pressure changes
measured and compensated
Electrical connectors:
Digital Interface:
9-pole D-connector for RS-232
Analyzer is connected to an
external computer via RS-232C
cable. The external computer
controls Gasmet.
Power connection:
Standard plug CEE-22
Gas inlet and outlet conditions
Gas temperature:
Non-condensing, the sample gas
temperature should be the same
as the sample cell temperature
Flow rate:
120 - 360 liters per hour
Gas filtration:
Filtration of particulates (2 µ)
required
Sample gas pressure:
Ambient
Sample pump:
Internal, for ambient air only
Electronics
A/D converter:
Dynamic range 95 dB
Signal processor:
32-bit floating point DSP
120 MFLOPS speed
Computer:
External, not included
®
Sample cell gaskets:
DX-4000
Multicomponent FTIR Gas Analyzer
Gasmet On-site Series includes portable multicomponent gas
analyzers for demanding applications. The Gasmet DX-4000
incorporates a Fourier transform infrared, FTIR spectrometer,
a temperature controlled sample cell, and signal processing
electronics. The analyzer offers versatility and high
performance for all users.
The Gasmet DX-4000 is designed for short term on site
measurements with wide dynamic ranges. It is an ideal tool to
measure trace concentrations of pollutants in wet, corrosive
gas streams. The sample cell can be heated up to 180 °C.
Sample cell absorption path length is selected according to the
application.
The Gasmet DX-4000 allows simple calibration using only
single component calibration gases. The user can easily
configure the analyzer for a new set of compounds.
General parameters
Measuring principle:
Fourier transform infrared, FTIR
Performance:
Simultaneous analysis of up to 50
gas compounds
Response time, T90:
Typically < 120 s, depending on
the gas flow and measurement
time
Operating temperature:
Short term 0 - 40°C
long term 5 - 30°C
non-condensing
Storage temperature:
-20 - 60°C, non-condensing
Power supply:
100-115 or 230 V / 50 -60 Hz
Power consumption:
Average 150 W, maximum 300 W
Spectrometer
-1
-1
Resolution:
8 cm or 4 cm
Scan frequency:
10 scans / s
Detector:
Peltier cooled MCT
Source:
SiC, 1550 K
Beamsplitter:
ZnSe
Window material:
ZnSe
Wave number range:
900 - 4 200 cm
-1
Sample cell
Analysis software (for external PC)
Structure:
Multi-pass, fixed path length 5.0 m
Operating system:
Windows XP
Material:
100 % rhodium coated aluminium
Analysis software:
Calcmet for Windows
Mirrors:
Fixed, protected gold coating
Volume:
0.4 liters
Options
Connectors:
Inlet Swagelok 6 mm
Outlet Swagelok 8 mm
Sample Cell:
Multi-pass, fixed path length 2.5 m
or 9.8 m
Gaskets:
Viton O-rings
Pressure measurement:
Inside sample cell
Temperature:
180 °C, maximum
Window material:
BaF2
®
Analog signals (ext PC): ADAM 5000/TCP module (for
analog inputs, outputs, relays)
®
®
®
Sample cell gaskets:
Teflon coated Viton or Kalrez
Measuring parameters
Trolley:
Wheeled cart for the analyzer and
laptop computer
Zero point calibration:
24 hours, calibration with nitrogen
(4.0 or higher N2 recommended)
Enclosure
Zero point drift:
< 2 % of measuring range per zero
point calibration interval
Material:
Aluminium
Dimensions (mm):
390 * 445 * 164
Sensitivity drift:
None
Weight:
13.9 kg
Linearity deviation:
< 2 % of measuring range
CE label:
Temperature drifts:
< 2 % of measuring range per 10
K temperature change
According to EMI guideline
89/336/EC
Pressure influence:
1 % change of measuring value
for 1 % sample pressure change.
Ambient pressure changes
measured and compensated
Electrical connectors:
Digital Interface:
9-pole D-connector for RS-232
Analyzer is connected to an
external computer via RS-232C
cable. The external computer
controls Gasmet.
Remote control connection for
Portable sampling unit
Power connection:
Standard plug CEE-22
PSS connection:
Remote connection of PSS
(Portable Sampling System)
Gas inlet and outlet conditions
Gas temperature:
Non-condensing, the sample gas
temperature should be the same
as the sample cell temperature
Flow rate:
120 - 600 liers per hour
Gas filtration:
Filtration of particulates (2 µ)
required
Sample gas pressure:
Ambient
Sample pump:
External, not included
Electronics
A/D converter:
Dynamic range 95 dB
Signal processor:
32-bit floating point DSP
120 MFLOPS speed
Computer:
External, not included
DX-4030
The Portable FTIR Gas Analyzer
The Gasmet DX-4030 is the first truly portable
FTIR gas analyzer in the world. The Gasmet DX–
4030 analyzer combines Fourier transform
infrared (FTIR) spectrometer, rhodium-gold
coated sample cell, built-in sample gas pump and
signal processing electronics in a compact unit.
In battery use, operating time is 2.5 hours with
one charge. AC power supply is also included.
There is a filter for particulates inside the probe
and no sample preparation is needed. The
lightweight analyzer can be used with or without
the durable Teflon coated backpack.
The Gasmet DX-4030 is designed for on-site
measurements of different compounds (both
organic & inorganic) at low concentrations in
ambient air. Typical usage areas include industrial
hygiene and emergency response situations.
Analysis results of up to 25 pre-calibrated
compounds are displayed on screen of a PDA
(Personal Digital Assistant). The communication
between the analyzer module and the PDA is
wireless (with Bluetooth protocol).
In the standard configuration, concentrations of 15
gases of interest can be simultaneously
monitored. As an option, ten additional gases can
be added to the analysis.
Furthermore, with the optional Calcmet 4030
Standard or Professional softwares the analyzer
can be connected to a laptop PC for extended
analysis capability (e.g. identification of unknown
compounds with library spectra).
There are no consumable parts that would need
replacing on regular basis. In addition, due to
FTIR technology, the calibrations remain very
stable. Hence no span calibrations are needed.
Also, cross-interferences (i.e. interference from
other gases) are automatically compensated in
the analysis algorithm during the calculation of the
results.
To sum up, the Gasmet Dx-4030 provides a very
cost-effective, easy-to-use solution for multicomponent gas analysis in ambient air.
General parameters
Electrical connectors:
Measuring principle:
Fourier transform infrared, FTIR
Performance:
Simultaneous analysis of up to 25
gas compounds (PDA), 50 gas
compounds with optional Calcmet
4030 Std or Pro software
Digital Interface:
Bluetooth protocol & RS-232
The analyzer is connected to a
PDA with Bluetooth connection
(RS-232 optional). The PDA
provides the analysis results.
Response time, T90:
Typically < 120 s, depending on
the gas flow and measurement
time
Option: sample spectra transfer to
laptop (PC) for additional
analytical capabilities
Operating temperature:
Ambient temperature
Short term 0 - 40°C
Long term 5 - 30°C
Non-condensing
Gas inlet and outlet conditions
Storage temperature:
-20 - 60°C, non condensing
Gas temperature:
Power supply:
12 VDC or 115 / 230 VAC
Ambient temperature (0 – 50 C),
non-condensing
Flow rate:
120 – 360 liters per hour
Gas filtration:
Filtration of particulates included in
the sample probe
Sample gas pressure:
Ambient
Sample pump:
Flow 2 l/min, for ambient air only
Approximately 2.5 hours operation
time with battery
Spectrometer
Resolution:
8 cm
o
-1
Electronics
Scan frequency:
10 scans / s
Detector:
Peltier cooled MCT
Source:
SiC, 1550 K
Beamsplitter:
ZnSe
Window material:
ZnSe
Wave number range:
900 - 4 200 cm
A/D converter:
Dynamic range 95 dB
Signal processor:
Two 32-bit floating point DSP’s
120 MFLOPS speed
-1
Analysis software (PDA)
Sample cell
Structure:
Multi-pass, fixed path length 9.8 m
Material:
100 % rhodium coated aluminum
Mirrors:
Fixed, protected gold coating
Volume:
0.4 liters
Temperature:
Ambient
Operating system:
Windows Mobile 6.0
Analysis software:
Calcmet Lite
Options
Software:
Measuring parameters
Calcmet 4030 Standard or
Professional for additional analysis
functions (identification of
unknowns etc.). Laptop PC +
Windows XP required. For more
information read Calcmet 4030
Std and Pro Technical data.
Zero point calibration:
24 hours
Zero point drift:
< 2 % of measuring range per zero
point calibration interval
Sensitivity drift:
none
Linearity deviation:
< 2 % of measuring range
Enclosure
Temperature drifts:
< 2 % of measuring range per 10
K temperature change
Material:
Aluminium
Pressure influence:
1 % change of measuring value
for 1 % sample pressure change.
Ambient pressure changes
measured and compensated.
Weight:
11.5 kg
CE label:
According to EMI guideline
89/336/EC
All technical information may change without notice.