Beamex MC5-IS

Beamex MC5-IS

INTRINSICALLY SAFE MULTIFUNCTION CALIBRATOR

63

Made for extreme environments

64

Made for extreme environments

MC5-IS: designed for use in potentially explosive environments

The ATEX- and IECEx- certified MC5-IS is designed for use in  potentially explosive environments, such as offshore platforms, oil refineries, chemical and petrochemical plants where inflammable gases may be present. There is probably no other intrinsically safe calibrator that can outperform the MC5-IS in  terms of functionality. It is a documenting, multifunction calibrator that has calibration capabilities for pressure, temperature, electrical and frequency signals. Its modular design  allows configuration based on your specific needs.

Main features of MC5-IS

Safe and robust field calibrator

The ATEX- and IECEx- certified, IP65-rated MC5-IS with impact  protectors and membrane keyboard is robust and made for tough use.

Accuracy guaranteed

The MC5-IS is a highly accurate all-in-one calibrator. It is  delivered with a traceable, accredited calibration certificate.

Communication with calibration software

Using the MC5-IS together with calibration software provides  you with a complete documenting calibration system that produces calibration certificates automatically.

Unmatched functionality

No other intrinsically safe calibrator can outperform the MC5-IS  in terms of functionality.

BEAMEX MC5-IS

| FIELD EQUIPMENT

65

Additional features

Accuracy guaranteed

The MC5-IS is among the most accurate process calibrators  available. As proof of this, each MC5-IS calibrator is delivered  with a traceable, accredited calibration certificate.

66

Made for tough use

The IP65-rated robust casing, along with integrated impact  protectors, makes the MC5-IS an ideal calibrator for use in wet  and dusty environments subject to wide temperature variations. 

Modularity means versatility

The MC5-IS is an extremely versatile calibrator with many  different functions. The modular construction of the MC5-IS  provides flexibility for the user. For instance, the MC5-IS can be  ordered as a pressure or temperature stand-alone calibrator, and  then later expand it into a data-logging, documenting  multifunction calibrator.

Communication with calibration software

Using the MC5-IS together with calibration software provides  you with a complete documenting calibration system that produces calibration certificates automatically. The benefits of the system include automated calibration procedures and paperless calibration management.

Calibrate safely

The MC5-IS is an intrinsically safe, ATEX- certified (EEx ia IIC T4  and ATEX directive II 1 G) multifunction calibrator. It is designed for use in potentially explosive environments, such as offshore platforms, oil refineries, chemical and petrochemical plants where inflammable gases may be present.

Fieldbus instruments must also be calibrated

Fieldbus installations are growing rapidly worldwide. Beamex  is the first company in the world to answer to this demand. We  have introduced the MC5-IS fieldbus calibrator, which provides  capability for calibrating Foundation Fieldbus H1 or Profibus PA  transmitters. It offers the safest possible way to calibrate fieldbus transmitters.

FIELD EQUIPMENT | BEAMEX MC5-IS

Specifications

GENERAL SPECIFICATIONS

GENERAL

Display

Weight

Dimensions

Case protection

Keyboard

Battery type

Battery operation

Charger supply

Operating temperature

Storage temperature

Humidity

Measurement sample rate

Warranty

96 x 72 mm (3.78" x 2.83"), 320 x 240 pixels, LCD

1.7 – 2.3 kg (3.7 – 5.1 lbs)

245 mm (9.6") x 192 mm (7.5") x 74 mm (2.9") (d/w/h)

IP65 (dust and water proof)

Membrane protected individual keys

Rechargeable NiMH, 1200 mAh, 8.4V DC

Average 5 hours

100 … 240 VAC, 50-60 Hz

−10 … 50 °C (14 … 122°F)

−20 to 60 °C (-4 to 140°F)

0 to 80 % R.H. non-condensing

2.5 / second

Standard: 3 years for MC5-IS; 1 year for battery pack. Warranty extension programs are also available.

FEATURES OF MODULES

FEATURE

Internal pressure modules

External pressure modules

Current measurement

Voltage measurement

Low voltage measurement

Frequency measurement

Pulse counting

Switch sensing

RTD measurement / simulation

Resistance measurement / simulation

TC measurement / simulation

Low voltage measurement / generation

Voltage generation

Frequency generation

Pulse generation

Internal TC reference junction compensation

INT = Internal pressure module

EXT = External pressure module

E = Electrical measuring module

ET = Electrical and temperature module

RJ = Thermocouple reference junction module

INT

EXT

E

ET

RJ

67

BEAMEX MC5-IS

| FIELD EQUIPMENT

68

INTERNAL & EXTERNAL PRESSURE MODULES

INTERNAL MODULES

1)

EXTERNAL MODULES RANGE

2)

INT B-IS

INT10mD-IS

INT100m-IS

INT400mC-IS

INT1C-IS

INT2C-IS

INT6C-IS

INT20C-IS

INT60-IS

INT100-IS

INT160-IS

EXT B-IS

EXT10mD-IS

EXT100m-IS

EXT400mC-IS

EXT1C-IS

EXT2C-IS

EXT6C-IS

EXT20C-IS

EXT60-IS

EXT100-IS

EXT160-IS

EXT250-IS

EXT600-IS

EXT1000-IS

−100 to 2000 kPa

−1 to 20 bar

−14.5 to 300 psi

0 to 6000 kPa

0 to 60 bar

0 to 900 psi

0 to 10 MPa

0 to 100 bar

0 to 1500 psi

0 to 16 MPa

0 to 160 bar

0 to 2400 psi

0 to 25 MPa

0 to 250 bar

0 to 3700 psi

0 to 60 MPa

0 to 600 bar

0 to 9000 psi

0 to 100 MPa

0 to 1000 bar

0 to 15000 psi

80 to 120 kPa a

800 to 1200 mbar a

11.6 to 17.4 psi a

±1 kPa diff

±10 mbar diff

±4 iwc diff

0 to 10 kPa

0 to 100 mbar

0 to 40 iwc

±40 kPa

±400 mbar

±160 iwc

±100 kPa

±1 bar

−14.5 to 15 psi

−100 to 200 kPa

−1 to 2 bar

−14.5 to 30 psi

−100 to 600 kPa

−1 to 6 bar

−14.5 to 90 psi

Temperature coefficient ±0.001 % Rdg/ °C outside 15 … 35 °C (59 … 95 °F)

INT10mD-IS / EXT10mD-IS < ±0.002 % Span/ °C outside 15 … 35 °C (59 … 95°F)

RESOLUTION ACCURACY

3)

(±)

0.0001

0.001

0.01

0.001

0.01

0.1

0.001

0.01

0.1

0.001

0.01

0.1

0.01

0.0001

0.001

0.1

0.001

0.01

0.0001

0.001

0.01

0.01

0.1

0.001

0.0001

0.001

0.001

0.0001

0.001

0.001

0.001

0.01

0.001

0.001

0.00001

0.0001

0.001

0.00001

0.0001

0.01

0.0001

0.001

0.03 kPa

0.3 mbar

0.0044 Psi

0.05 % Span

0.015 % FS + 0.0125 % RDG

0.01 % FS + 0.0125 % RDG

0.007 % FS + 0.0125 % RDG

0.005 % FS + 0.01 % RDG

0.005 % FS + 0.01 % RDG

0.005 % FS + 0.01 % RDG

0.005 % FS + 0.0125 % RDG

0.005 % FS + 0.0125 % RDG

0.005 % FS + 0.0125 % RDG

0.007 % FS + 0.0125 % RDG

0.007 % FS + 0.01 % RDG

0.007 % FS + 0.01 % RDG

1 YEAR UNCERTAINTY (±)

4)

0.05 kPa

0.5 mbar

0.0073 psi

0.05 % Span + 0.1 % RDG

0.025 % FS + 0.025 % RDG

0.02 % FS + 0.025 % RDG

0.015 % FS + 0.025 % RDG

0.01 % FS + 0.025 % RDG

0.01 % FS + 0.025 % RDG

0.01 % FS + 0.025 % RDG

0.01 % FS + 0.025 % RDG

0.01 % FS + 0.025 % RDG

0.01 % FS + 0.025 % RDG

0.015 % FS + 0.025 % RDG

0.015 % FS + 0.025 % RDG

0.015 % FS + 0.025 % RDG

1) The MC5-IS calibrators can hold three internal pressure modules.

2) Every internal/external pressure module’s range may also be displayed in absolute pressure if the barometric module (B) is installed.

3) ‘Accuracy’ includes hysteresis, nonlinearity, repeatability and reference standard uncertainty (k=2).

4) ‘1 Year Uncertainty’ includes hysteresis, nonlinearity, repeatability and typical long-term stability for mentioned period (k=2).

All intrinsically safe external pressure modules (EXT-IS) are also compatible with Beamex MC2, MC2-IS, MC4, MC5, MC5P and MC6.

Supports the following pressure units as standard:

Pa, hPa, kPa, MPa, mbar, bar, lbf/ft2, psi, gf/cm2, kgf/cm2, kgf/m2, kp/cm2, at, mmH

2 mmH

2

O(4 °C), inH

2

O(4 °C), ftH

2

O(4 °C), inH

2

O(60°F), mmH

2

O(68°F), inH

2

O(68°F), ftH

2

O, cmH

2

O, mH

2

O, iwc, ftH

O(68°F), torr, atm.

2

O, mmHg, cmHg, mHg, inHg, mmHg(0 °C), inHg(0 °C),

INT B-IS / EXT B-IS; M5 (10/32") female.

INT10mD-IS and EXT10mD-IS; Two M5 (10/32") female threads with a hose nipple included.

INT100m-IS/EXT100m-IS – INT20C-IS/EXT20C-IS; G1/8" (ISO228/1) female. A conical 1/8” BSP male with 60° internal cone adapter included for Beamex hose set.

INT60-IS, INT100-IS, INT160-IS; G1/8" (ISO228/1) female. EXT60-IS, EXT100-IS, EXT160-IS, EXT250-IS, EXT600-IS, EXT1000-IS; G ¼" (ISO228/1) male.

Wetted parts AISI316 stainless steel, Hastelloy, Nitrile rubber.

Maximum overpressure;

B module; 1200 mbar abs. 10mD module; 200 mbar. EXT600; 900 bar. EXT1000; 1000 bar.

For all other modules, the maximum overpressure is twice the nominal range.

HART is a registered trademark of HART Communication Foundation.

FIELD EQUIPMENT | BEAMEX MC5-IS

ELECTRICAL MODULE (E)

FUNCTION RANGE mV measurement

V measurement

(3 mA measurement

Hz measurement

(5

(2

(4

±250 mV

±30 V

±100 mA

0.0028 to 50000 Hz

Pulse counting

(5

0 to 9 999 999 pulses mA Sink 0 to 25 mA

Temperature coefficient < ±0.001 % RDG / °C outside of 15 … 35 °C (59 … 95°F)

RESOLUTION

0.001 mV

0.00001–0.001 V

0.0001–0.001 mA

0.000001− 0.1 Hz

1 pulse

0.0001 mA

1 YEAR UNCERTAINTY (±) (1

0.02 % RDG + 5 µV

0.02 % RDG + 0.25 mV

0.02 % RDG + 1.5 µA

0.01 % RDG

N/A

0.02 % RDG + 1.5 µA

1) Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long-term stability for mentioned period (k=2).

2) Bias current <10 nA

3) Impedance >1 MΩ

4) Impedance < 7.5 Ω

5) MC5-IS; Impedance > 1 MΩ. Frequency measurement minimum amplitude 1 Vpp (< 10 kHz),

3 Vpp (10 … 50 kHz). Pulse counting minimum amplitude 1 Vpp (pulse length > 50 µs),

3 Vpp (pulse length 50 µs … 10 µs).

Trigger level range –1 … +15 V.

TEMPERATURE ELECTRICAL MODULE (ET)

FUNCTION mV generation

(2

V generation

(3 mA sink

Hz generation (4

Pulse generation

(5

Ohm simulation (6

Ohm measurement

(8 mV measurement

(9

RANGE

±250 mV

–2.5 to 10 V

0 to 25 mA

0.00028 to 50 000 Hz

0 to 9 999 999 pulses

1 to 4000 Ω

0 to 4000 Ω

±250 mV

Temperature coefficient < ±0.001 % RDG / °C outside of 15 … 35 °C (59 … 95°F)

RESOLUTION

0.001 mV

0.00001–0.0001 V

0.0001 mA

0.000001 - 0.1 Hz

1 pulse

0.0–0.1 Ω

0.001–0.1 Ω

0.001 mV

1 YEAR UNCERTAINTY (±)

(1

0.02 % RDG + 4 µV

0.02 % RDG + 0.1 mV

0.02 % RDG + 1 µA

0.01 % RDG

N/A

0.04 % RDG or 30 mΩ (7

0.02 % RDG + 3.5 mΩ

0.02 % RDG + 4 µV

1) Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long-term stability for mentioned period (k=2).

2) Load effect < 5 µV/mA. Maximum output current 1 mA.

3) Load effect < 100 µV/mA. Maximum output current 1 mA (0 … 10 V)

4) Amplitude range 0 … 5 Vpp (positive), 0 … 5 V (symmetric). Amplitude setting accuracy up to 5kHz ±(200 mV + 5 % of set value).

Waveforms: square wave (positive / symmetric) and sinewave (above 40 Hz).

5) Pulse generation frequency range 0.1 … 1000 Hz. Amplitude range 0 … 5 Vpp (positive), 0 … 5 V (symmetric).

6) Valid with measurement current 0.2 … 2 mA (1 … 250 Ω), 0.05 < I meas

 • Rsim < 0.5 V (250 … 4000 Ω). Ω/RTD simulation settling time 1 ms.

7) Whichever is greater.

8) Specification valid with 4 wire connection. In 3 wire connection add 10 mΩ.

9) Bias current < 10 nA.

69

BEAMEX MC5-IS

| FIELD EQUIPMENT

70

THERMOCOUPLE MEASUREMENT AND SIMULATION

TYPE

B

(2

R

(2

S

(2

E

(2

J

(2

K (2

N

(2

T

(2

U

(4

L (4

C (5

G

(6

D (5

RANGE (°C)

0 … 1820

−50 … 1768

−50 … 1768

−270 … 1000

−210 … 1200

−270 … 1372

−270 … 1300

−270 … 400

−200 … 600

−200 … 900

0 … 2315

0 … 2315

0 … 2315

RANGE (°C)

−270 … −200

−200 … 0

0 … 1000

1000 … 1372

−270 … −200

−200 … −100

−100 … 0

0 … 750

750 … 1300

−270 … −250

−250 … −200

−200 … 0

0 … 400

−200 … 0

0 … 600

−200 … 0

0 … 900

0 … 900

900 … 2000

2000 … 2315

0 … 70

70 … 200

200 … 1600

1600 … 2000

2000 … 2315

0 … 1000

1000 … 2000

2000 … 2315

0 … 200

200 … 500

500 … 800

800 … 1820

−50 … 0

0 … 150

150 … 1400

1400 … 1768

−50 … 0

0 … 50

50 … 1500

1500 … 1768

−270 … −200

−200 … 0

0 … 600

600 … 1000

−210 … −200

−200 … 0

0 … 1200

1 YEAR UNCERTAINTY (±)

(1

(3

2.0 °C

0.8 °C

0.6 °C

1.0 °C

0.7 °C

0.5 °C

0.6 °C

1.0 °C

0.7 °C

0.6 °C

0.7 °C

(3

0.08 % RDG + 0.07 °C

0.015 % RDG + 0.07 °C

0.026 % RDG

(3

0.07 % RDG + 0.08 °C

0.02 % RDG + 0.08 °C

(3

0.1 % RDG + 0.1 °C

0.02 % RDG + 0.1 °C

0.03 % RDG

(3

0.2 % RDG

0.05 % RDG + 0.15 °C

0.01 % RDG + 0.15 °C

0.03 % RDG

(3

0.7 °C

0.1 % RDG + 0.1 °C

0.01 % RDG + 0.1 °C

0.1 % RDG + 0.15 °C

0.01 % RDG + 0.15 °C

0.07 % RDG + 0.13 °C

0.02 % RDG + 0.13 °C

0.4 °C

0.045 % RDG

1.2 °C

(3

1.0 °C

0.5 °C

0.7 °C

1.0 °C

0.4 °C

0.04 % RDG

1.2 °C

Resolution 0.01 °C.

With internal reference junction (module RJ) add 0.1 °C uncertainty.

Other thermocouple types also available as an option.

1) Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long-term stability for mentioned period (k=2).

2) IEC 584, NIST MN 175, BS 4937, ANSI MC96.1

3) ±(0.02 % of thermovoltage + 4 µV)

4) DIN 43710

5) ASTM E 988 - 96

6) ASTM E 1751 - 95e1

REFERENCE JUNCTION MODULE (RJ)

RANGE (°C)

−10 … 50 °C

1 YEAR UNCERTAINTY (±)

(1

0.1 °C

1) Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long-term stability for mentioned period (k=2).

FIELD EQUIPMENT | BEAMEX MC5-IS

RTD MEASUREMENT AND SIMULATION

FUNCTION

Pt-sensors

RANGE (°C)

−200 to 850 °C

RANGE (°C)

−200 to 0 °C

0 to 850 °C

MEASUREMENT

1 YEAR UNCERTAINTY (±)

(1

0.06 °C

0.025 % RDG + 0.06 °C

SIMULATION

1 YEAR UNCERTAINTY (±)

(1

0.1 °C

0.025 % RDG + 0.1 °C

1) Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long-term stability for mentioned period (k=2).

RTD TYPES AVAILABLE AS STANDARD

Pt50 (385)

Pt100 (385)

Pt200 (385)

Pt400 (385)

Pt500 (385)

Pt1000 (385)

Pt100 (3923)

Pt100 (391)

Pt100 (375)

Pt100 (3926)

Ni100 (618)

Ni120 (672)

Cu10 (427)

To improve uncertainty with PRT (platinum RTD) sensors, the MC5-IS includes a standard possibility that allows you to create customized PRT sensors using the

Callendar van Dusen correction coefficients. The easy-to-use Beamex PRT Tool PC software is used to create the sensor and to send it to the MC5-IS. Up to 100 customized PRT sensors can be stored in MC5-IS at one time.

This function may be also used to create new, non-supported PRT sensors in the MC5-IS. Both measurement and simulation can be done with the customized sensors.

71

STANDARD ACCESSORIES

•  Accredited calibration certificate

•  User guide

•  Computer cable

•  Battery charger / eliminator

•  Internal NiMH battery pack

•  Test leads and clips

•   Appropriate pressure T-hose with internal low pressure modules

•  CD-ROM with user manual, software tools and product information

OPTIONAL ACCESSORIES

•  Soft carrying case

•  Hard carrying case

•  Spare battery pack

BEAMEX MC5-IS

| FIELD EQUIPMENT

SUMMARY

Beamex MC5-IS

INTRINSICALLY SAFE MULTIFUNCTION CALIBRATOR

72

The Beamex MC5-IS is ATEX- and IECEx- certified and designed for use in potentially explosive environments, such as offshore platforms, oil refineries, chemical and petrochemical plants where inflammable gases may be present. The MC5-IS is a documenting, multifunction calibrator that has calibration capabilities for pressure, temperature, electrical and frequency signals. Its modular design allows configuration based on your specific needs.

Safe and robust field calibrator

The ATEX- and IECEx- certified, IP65-rated MC5-IS with impact protectors and membrane keyboard is robust and made for tough use.

accuracy assured

The MC5-IS is a highly accurate all-in-one calibrator. It is delivered with a traceable, accredited calibration certificate.

Communication with calibration software

Using the MC5-IS together with calibration software provides you with a complete documenting calibration system that produces calibration certificates automatically.

Unmatched functionality

No other intrinsically safe calibrator can outperform the MC5-IS in terms of functionality.

Main features

Highly accurate all-in-one calibrator

Designed for use in potentially explosive environments

Calibration capabilities for pressure, temperature, electrical and frequency signals

Certified in accordance with the ATEX directive and IECEx scheme

HART

Foundation Fieldbus H1 and Profibus PA support

FIELD EQUIPMENT | BEAMEX MC5-IS