CAIRE Companion 5 User manual

Add to my manuals
16 Pages

advertisement

CAIRE Companion 5 User manual | Manualzz

Companion 5

TM

Oxygen Concentrator

User Manual

User Controls & System Status Indicators

Symbol Definition

Read user manual before operation. See user manual for instructions.

Symbol

I/O

On/Off Switch

No Smoking Icon: Do not smoke near unit. Use no oil or grease.

Definition

Warnings / ALERT (Yellow) Indicator

Class II equipment

0

2

Oxygen Output

Type BF Applied Part (degree of protection against electric shock)

This symbol is to remind the equipment owners to return it to a recycling facility at the end of its life, per Waste Electrical and Electronic Equipment

(WEEE) Directive.

Authorized representative in the European

Community

No open or naked flames.

No serviceable parts inside. Do not open cover.

IPX1

Certified for both the U.S. and Canadian markets, to the applicable U.S. and Canadian standards.

This device complies with the requirements of

Directive 93/42/EEC concerning medical devices. It therefore bears the CE marking as shown.

Drip Proof Equipment-IPX1: The Companion 5 provides protection against the harmful effects of the ingress of liquids. (IPX1, per IEC 60529)

Name and address of manufacturer

2 — 15062781 E

Contents

User Controls & System Status Indicators ..............................................................................................2

Quick Start Guide ......................................................................................................................................4

Indications for Use ....................................................................................................................................7

Contraindications for Use .........................................................................................................................7

Safety Guidelines ......................................................................................................................................7

Recommended Operating Environments ................................................................................................9

Operating Instructions ............................................................................................................................10

Alarm Conditions .....................................................................................................................................13

Troubleshooting .......................................................................................................................................14

Cleaning, Care & Routine Maintenance .................................................................................................15

Companion 5 User Manual — 3

Quick Start Guide

1

Unpack Your Companion 5

Companion 5

2

Getting to Know Your Companion 5

Review all Warnings, Cautions and additional device information in the rest of this manual.

Become familiar with the key features of the Companion 5 and the User Control Panel.

FRONT

Power Switch

Outer Case

Oxygen

Outlet Barb

LED Display

Hour Meter

Flow Meter

Humidifier

Bottle Support

Stand (Bottle optional)

BACK

Carrying Handle &

Air Intake Location

Cooling Air

Intake Location

Air Intake

Location

Exhaust Vent

4 — 15062781 E

The Companion 5 Components

Outer Case Durable plastic case that encloses and protects the inner components of the CAIRE Companion 5.

Power Switch ON/OFF power switch used to initiate or stop power supply to the unit. The ON position is indicated by the

“l” symbol, and the OFF position is indicated by the

“O” symbol.

Outlet Barb Oxygen exits the Companion 5 here after it has been filtered and concentrated. Either a single lumen nasal cannula or standard oxygen tubing is attached to deliver to the patient. The maximum length of oxygen tubing attached to the concentrator is 50 ft. (15.2 m).

LED Display Contains a green, red and yellow* LED light. The green light indicates normal operation, and the yellow* and red lights indicate alarms conditions. (See the Alarm

Conditions section of the instruction for more information.)

Hour Meter

Flow Meter

Digital display of the elapsed operation time of the concentrator. Displays to the nearest tenth of an hour and cannot be reset. Displays diagnostic alarm codes when the concentrator experiences an alarm condition.

Used to adjust and display the flow rate of oxygen that is delivered. Flow rates range from 0.5 - 5 liters per minute (LPM) and can be adjusted by turning the knob.

Humidifier

Bottle Support

Stand (Bottle optional)

Location to attach a humidifier bottle. The stand contains an elastic band used to secure the bottle on the stand. A humidifier bottle provides additional moisture to the oxygen flow. The bottle is filled with distilled water, and the moisture is used to prevent drying of the nasal tissues.

*Oxygen Concentrator Status Indicator (OCSI)

models only

3

Attaching Humidifier Bottle & Cannula

If using a humidifier bottle, first fill the bottle with distilled water to the manufacturer’s specified level. Place bottle in designated support stand and secure with the elastic band. Connect the threaded end of the humidifier tube to the bottle, and the other to the Companion 5 outlet barb. Connect oxygen tubing and/or a nasal cannula to the outlet barb on the humidifier bottle.

If not using a humidifier bottle, connect the oxygen tubing and/or a nasal cannula directly to the Companion 5 outlet barb.

Companion 5 User Manual — 5

4

Power On and Warm Up

Plug the AC power cord into an AC outlet.

5

Adjust Flow Setting to Prescribed Level

Use the Flow Control Knob to select the flow prescribed by your physician.

Toggle the Power Switch to the On position.

Power

Switch

WARNING: IT IS VERY IMPORTANT TO SELECT ONLY THE

PRESCRIBED LEVEL OF OXYGEN. DO NOT CHANGE THE

FLOW SELECTION UNLESS YOU HAVE BEEN DIRECTED TO DO

SO BY A LICENSED CLINICIAN. THE OXYGEN CONCENTRATOR MAY BE

USED DURING SLEEP UNDER THE RECOMMENDATION OF A QUALI-

FIED CLINICIAN.

When the Companion 5 is powered on properly, the green power indicator on the LED Display will light up. All LED lights will illuminate upon start-up. After the concentrator completes the warm-up cycle, only the green light will remain on.

Breathe normally through the nasal cannula.

Note: After initially powering ON the device, please allow up to ten (10) minutes for the device to reach its performance specifications.

For OCSI Models: The O2 Light will illuminate for four (4) seconds upon start-up. After this initial start-up, this light is disabled for approximately five (5) minutes while the concentration rises to specifications.

To adjust flow rate:

Turn counter-clockwise to increase flow.

Turn clockwise to decrease flow.

The middle of the ball indicates flow rate. See picture below.

Power

Indicator

LED

6 — 15062781 E

Important!

Safety Instructions are defined as follows:

WARNING: IMPORTANT SAFETY INFORMATION FOR HAZ-

ARDS THAT MIGHT CAUSE SERIOUS INJURY.

CAUTION: Important information for preventing damage to the Companion 5.

Note: Information needing special attention.

Indications for Use

The Companion 5 is intended for the administration of supplemental oxygen. The device is not intended for life support, nor does it provide any patient monitoring capabilities.

WARNING: IT IS VERY IMPORTANT TO SELECT ONLY THE

PRESCRIBED LEVEL OF OXYGEN. DO NOT CHANGE THE

FLOW SELECTION UNLESS YOU HAVE BEEN DIRECTED TO DO

SO BY A LICENSED CLINICIAN. THE OXYGEN CONCENTRATOR MAY BE

USED DURING SLEEP UNDER THE RECOMMENDATION OF A QUALI-

FIED CLINICIAN.

WARNING: FEDERAL (USA) LAW RESTRICTS THIS DEVICE TO SALE OR

RENTAL BY ORDER OF A PHYSICIAN OR OTHER LICENSED HEALTH

CARE PROVIDER.

WARNING: THIS UNIT IS NOT TO BE USED FOR LIFE SUPPORT. GERIAT-

RIC, PEDIATRIC, OR ANY OTHER PATIENT UNABLE TO COMMUNICATE

DISCOMFORT WHILE USING THIS DEVICE MAY REQUIRE ADDITIONAL

MONITORING. PATIENTS WITH HEARING AND/OR SIGHT IMPAIR-

MENT(S) MAY NEED ASSISTANCE WITH MONITORING ALARMS.

Contraindications for Use

WARNING: IN CERTAIN CIRCUMSTANCES, THE USE OF

NON-PRESCRIBED OXYGEN CAN BE HAZARDOUS. THIS

DEVICE SHOULD ONLY BE USED WHEN PRESCRIBED BY A

PHYSICIAN.

WARNING: NOT FOR USE IN THE PRESENCE OF FLAMMABLE ANES-

THETICS.

WARNING: AS WITH ANY ELECTRICALLY POWERED DEVICE, THE

USER MAY EXPERIENCE PERIODS OF NON-OPERATION AS A RESULT

OF ELECTRICAL POWER INTERRUPTION, OR THE NEED TO HAVE THE

COMPANION 5 SERVICED BY A QUALIFIED TECHNICIAN. THE COM-

PANION 5 IS NOT APPROPRIATE FOR ANY PATIENT WHO WOULD

EXPERIENCE ADVERSE HEALTH CONSEQUENCES AS THE RESULT OF

SUCH TEMPORARY INTERRUPTION.

Safety Guidelines

WARNING: NO MODIFICATION OF THIS EQUIPMENT IS

PERMITTED

WARNING: THE MANUFACTURER RECOMMENDS AN ALTER-

NATE SOURCE OF SUPPLEMENTAL OXYGEN IN THE EVENT

OF A POWER OUTAGE, ALARM CONDITION, OR MECHANI-

CAL FAILURE. CONSULT YOUR PHYSICIAN OR EQUIPMENT PROVIDER

FOR THE TYPE OF RESERVE SYSTEM REQUIRED.

WARNING: THIS DEVICE SUPPLIES HIGH-CONCENTRATION OXYGEN

THAT PROMOTES RAPID BURNING. DO NOT ALLOW SMOKING OR

OPEN FLAMES WITHIN THE SAME ROOM OF (1) THIS DEVICE, OR (2)

ANY OXYGEN-CARRYING ACCESSORY. FAILURE TO OBSERVE THIS

WARNING CAN RESULT IN SEVERE FIRE, PROPERTY DAMAGE, AND/OR

CAUSE PHYSICAL INJURY OR DEATH.

WARNING: DO NOT OPERATE UNIT IN A RESTRICTED OR CONFINED

SPACE (I.E., A SMALL CASE) WHERE VENTILATION CAN BE LIMITED.

THIS CAN CAUSE THE OXYGEN CONCENTRATOR TO OVERHEAT AND

IMPAIR PERFORMANCE.

WARNING: THE CONCENTRATOR SHOULD BE LOCATED AS TO AVOID

SMOKE, POLLUTANTS OR FUMES.

WARNING: THE USE OF SOME OXYGEN ADMINISTRATING ACCESSO-

RIES NOT SPECIFIED FOR THIS OXYGEN CONCENTRATOR MAY IMPAIR

ITS PERFORMANCE. RECOMMENDED ACCESSORIES ARE REFERENCED

WITHIN THIS MANUAL.

WARNING: IF THE OXYGEN CONCENTRATOR HAS BEEN DROPPED,

DAMAGED OR EXPOSED TO WATER PLEASE CONTACT YOUR HOME

CARE PROVIDER FOR INSPECTION OR POSSIBLE REPAIR OF THE

DEVICE. DO NOT USE THE OXYGEN CONCENTRATOR IF IT HAS A DAM-

AGED POWER CORD OR PLUG.

WARNING: DO NOT ALLOW EITHER THE AIR INTAKE OR THE AIR

OUTLET VENTS TO BECOME BLOCKED. DO NOT DROP OR INSERT ANY

OBJECTS INTO ANY OPENINGS ON THE DEVICE. THIS CAN CAUSE THE

OXYGEN CONCENTRATOR TO OVERHEAT AND IMPAIR PERFORMANCE.

WARNING: DO NOT OVERFILL THE OPTIONAL HUMIDIFIER. FILL THE

OPTIONAL HUMIDIFIER WITH WATER ONLY TO THE LEVEL SHOWN BY

THE MANUFACTURER OF THE HUMIDIFIER.

WARNING: DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST

READING AND UNDERSTANDING THIS MANUAL. IF YOU ARE UNABLE

TO UNDERSTAND THE WARNINGS AND INSTRUCTIONS, CONTACT

YOUR EQUIPMENT PROVIDER BEFORE ATTEMPTING TO USE THIS

EQUIPMENT; OTHERWISE, INJURY OR DAMAGE MAY RESULT.

WARNING: IF YOU FEEL DISCOMFORT OR ARE EXPERIENCING A MEDI-

CAL EMERGENCY, SEEK MEDICAL ASSISTANCE IMMEDIATELY.

Companion 5 User Manual — 7

WARNING: OPERATING THE OXYGEN CONCENTRATOR

OUTSIDE OF THE OPERATIONAL TEMPERATURE SPECIFI-

CATIONS CAN LIMIT THE CONCENTRATOR’S ABILITY TO

MEET OXYGEN CONCENTRATION SPECIFICATION. REFER TO

THE SPECIFICATION SECTION OF THIS MANUAL FOR TEMPERATURE

LIMITS.

WARNING: USE NO OIL, GREASE, OR PETROLEUM-BASED OR OTHER

FLAMMABLE PRODUCTS WITH THE OXYGEN-CARRYING ACCESSORIES

OR THE OXYGEN CONCENTRATOR. ONLY WATER BASED, OXYGEN

COMPATIBLE LOTIONS OR SALVES SHOULD BE USED. OXYGEN ACCEL-

ERATES THE COMBUSTION OF FLAMMABLE SUBSTANCES.

WARNING: THE OXYGEN CONCENTRATOR SHOULD NOT BE USED AD-

JACENT TO OR STACKED WITH OTHER EQUIPMENT. IF ADJACENT OR

STACKED USE IS UNAVOIDABLE, THE DEVICE SHOULD BE OBSERVED

TO VERIFY NORMAL OPERATION.

CAUTION: In the event of an alarm or you observe the

Oxygen Concentrator is not working properly; consult the Troubleshooting section in this manual. If you cannot resolve the problem, consult your Equipment Provider.

CAUTION: To prevent a void warranty, follow all manufacturers’ instructions.

WARNING: DO NOT ATTEMPT ANY MAINTENANCE OTHER

THAN THE POSSIBLE SOLUTIONS LISTED WITHIN THIS

MANUAL. DO NOT REMOVE COVERS, ONLY YOUR EQUIP-

MENT PROVIDER OR A QUALIFIED SERVICE TECHNICIAN SHOULD

REMOVE THE COVERS OR SERVICE THIS DEVICE.

WARNING: ONLY ACCESSORIES RECOMMENDED BY THE MANUFAC-

TURER. USE OF ANY OTHER MAY BE HAZARDOUS, CAUSE SERIOUS

DAMAGE TO YOUR OXYGEN CONCENTRATOR AND WILL VOID THE

WARRANTY.

WARNING: THE MANUFACTURER RECOMMENDS AN ALTER-

NATE SOURCE OF SUPPLEMENTAL OXYGEN IN THE EVENT

OF A POWER OUTAGE, ALARM CONDITION, OR MECHANI-

CAL FAILURE. CONSULT YOUR PHYSICIAN OR EQUIPMENT

PROVIDER FOR THE TYPE OF RESERVE SYSTEM REQUIRED.

WARNING: SMOKING WHILE USING OXYGEN IS THE NUM-

BER ONE CAUSE OF FIRE INJURIES AND RELATED DEATHS.

YOU MUST FOLLOW THESE SAFETY WARNINGS:

WARNING: DO NOT ALLOW SMOKING, CANDLES, OR OPEN FLAMES

IN THE SAME ROOM WITH THE DEVICE OR THE OXYGEN-CARRYING

ACCESSORIES.

WARNING: SMOKING WHILE WEARING AN OXYGEN CANNULA MAY

CAUSE FACIAL BURNS AND POSSIBLY DEATH.

WARNING: IF YOU SMOKE, THESE 3 STEPS MAY SAVE YOUR LIFE:

TURN OFF THE OXYGEN CONCENTRATOR, TAKE OFF THE CANNULA,

AND LEAVE THE ROOM WHERE THIS DEVICE IS LOCATED.

WARNING: “NO SMOKING – OXYGEN IN USE” SIGNS MUST BE

PROMINENTLY DISPLAYED IN THE HOME, OR WHERE OXYGEN IS IN

USE. PATIENTS AND THEIR CAREGIVERS MUST BE INFORMED ABOUT

THE DANGERS OF SMOKING IN THE PRESENCE OF, OR WHILE USING,

MEDICAL OXYGEN.

WARNING: DO NOT USE YOUR OXYGEN CONCENTRATOR IN THE

PRESENCE OF FLAMMABLE GASES. THIS CAN RESULT IN RAPID

BURNING CAUSING PROPERTY DAMAGE, BODILY INJURIES OR

DEATH.

WARNING: REMOVING THE CANNULA AND PUTTING IT ON CLOTH-

ING, BEDDING, SOFAS, OR OTHER CUSHION MATERIAL WILL CAUSE

A FLASH FIRE WHEN EXPOSED TO A CIGARETTE, HEAT SOURCE, OR

FLAME.

WARNING: DO NOT LEAVE A NASAL CANNULA ON OR UNDER CLOTH-

ING, BED COVERINGS OR CHAIR CUSHIONS. IF THE UNIT IS TURNED

ON BUT NOT IN USE, THE OXYGEN WILL MAKE THE MATERIAL FLAM-

MABLE. SET THE I/0 POWER SWITCH TO THE 0 (OFF) POSITION WHEN

THE OXYGEN CONCENTRATOR IS NOT IN USE.

CAUTION: Always place oxygen supply tubing and power cords in a manner that prevents a trip hazard.

Note: Portable and mobile RF communications equipment can effect medical electrical equipment.

8 — 15062781 E

Recommended Operating Environments

For proper use of the device, the following chart provides important information concerning the recommended operating environments, or operating conditions.

Specifications

Operating

Temperature

Operating

Humidity

Transport/

Storage

Temperature

41º F to 104º F (5° to 40°C) up to 95% at an 82.4° F (28° C) dew point

-40º F to 158º F (-40º C to 70º C)

Humidity: up to 95% Non-Condensing

Electrical

Use no extension cords. Use no electrical outlets controlled by a switch.

Operating

Altitude

-1253–9879 feet (-382–3011 meters)

Placement

DO NOT block the air intake locations (top rear or underneath) or the exhaust vent (back bottom left and right).

Place device a minimum of 12 inches (30 cm) away from walls, draperies, furniture, etc.

Environment Must be smoke, pollutant, and fume free.

Operating

Time

up to 24-hours a day

Proper Placement of the Companion 5

Select a location for the device that avoids the intake of smoke, fumes and pollutants. Correct placement of the device should allow intake of air through the 3 air intake locations at the top rear and underneath the cabinet, and also allow for exhaust air to freely leave the exhaust vent at the bottom left of the device.

Air intakes are located on the upper back and bottom of the concentrator. Locate the unit so that there is at least 12 inches (30 cm) of space between the concentrator and any walls, furniture, curtains, or other obstructions.

Place the device such that the alarms can be heard.

Position the oxygen supply tubing in such a way that it does not kink or occlude.

WARNING: THIS DEVICE SUPPLIES HIGH-CONCENTRATION

OXYGEN THAT PROMOTES RAPID BURNING. DO NOT ALLOW

SMOKING OR OPEN FLAMES WITHIN THE SAME ROOM

OF (1) THIS DEVICE, OR (2) ANY OXYGEN-CARRYING ACCESSORY.

FAILURE TO OBSERVE THIS WARNING CAN RESULT IN SEVERE FIRE,

PROPERTY DAMAGE, AND/OR CAUSE PHYSICAL INJURY OR DEATH.

Note: If the Oxygen Concentrator has been stored for an extended period of time outside its normal operating temperature range, the unit should be allowed to return to normal operating temperature before being turned on. (Refer to the Specifications section in this manual.)

CAUTION: Do not operate unit in a restricted or confined space (i.e., a small case) where ventilation can be limited. This can cause the

Oxygen Concentrator to overheat and impair performance. Do not allow either the air intake or the air outlet vents to become blocked.

This can cause the Oxygen Concentrator to overheat and impair performance.

Companion 5 User Manual — 9

Operating Instructions

Before Operating

This user manual serves as your reference to help you operate and maintain the device. If you have any questions or concerns please call your home care provider.

Step 2: If the Humidifier Bottle WILL NOT be Used

a. Connect a nasal cannula or oxygen tubing directly to the oxygen outlet barb.

Important! DO NOT attempt to operate the Companion 5 without first reading the Safety Guidelines section of this manual.

Please follow all of the operating instructions. Please observe all Warnings on the device and in the Users Manual. In order to reduce the risk of fire, personal injury and serious damage to the

Companion 5, please observe all of the safety precautions.

The CAIRE Companion 5 does not create its own oxygen. It produces highly concentrated oxygen from surrounding room air and delivers it to you. As it naturally exists, room air contains approximately only 21% oxygen by volume.

Room air is drawn into the concentrator by a compressor. The air then travels through a series of filters and a molecular sieve bed that chemically adsorbs nitrogen molecules. As a result, the oxygen-enriched air that exits the concentrator is delivered at

87–95% oxygen by volume.

WARNING: CLEAN THE CABINET, CONTROL PANEL, AND

POWER CORD ONLY WITH A MILD HOUSEHOLD CLEANER

APPLIED WITH A DAMP (NOT WET) CLOTH OR SPONGE,

AND THEN WIPE ALL SURFACES DRY. DO NOT ALLOW ANY LIQUID TO

GET INSIDE THE DEVICE. PAY SPECIAL ATTENTION TO THE OXYGEN

OUTLET FOR THE CANNULA CONNECTION TO MAKE SURE IT RE-

MAINS FREE OF DUST, WATER, AND PARTICLES.

CAUTION: Do not allow either the air intake or the air outlet vents to become blocked. This can cause the Oxygen

Concentrator to overheat and impair performance.

b. Proceed to step 4.

Step 3: If the Humidifier Bottle WILL be Used

a. Unscrew the lid from the humidifier bottle.

b. Fill the bottle using distilled water. Make sure the water level is between the manufacturer’s specified maximum and minimum levels shown on the bottle.

c. Secure the lid of the humidifier bottle. Ensure that there are no leaks.

d. Connect a nasal cannula or standard oxygen tubing to the outlet barb on the humidifier bottle.

Note: Use of a humidifier not specified for use with this concentrator might affect its performance.

Step 1: Positioning Your Companion 5 for Use

Place the Companion 5 in a well-ventilated, well lit area. Be sure the air inlet and exhaust vents are not obstructed. Position the

Companion 5 so that all audible and visual indicators or alarms can be easily seen and heard.

Plug the device into an AC Power outlet.

Air Intake Location

Cooling Air Intake

Location

Air Intake Location

Exhaust Vent

10 — 15062781 E

e. Connect the threaded end of the humidifier tube to the lid of the humidifier bottle.

Step 4: Power On and Warm Up

a. Plug the electrical cord of the CAIRE Companion 5 into an

AC outlet.

WARNING: INSPECT THE ELECTRICAL CORD FOR DAMAGE

BEFORE USE. IF THE CORD IS DAMAGED, DO NOT PLUG IT

INTO AN ELECTRICAL OUTLET OR ATTEMPT TO OPERATE

THE CONCENTRATOR.

f. Connect the other end of the humidifier tube to the outlet barb of the CAIRE Companion 5.

b. Press the power switch in the “ON” (I) position. When the

Companion 5 is powered on properly, the green indicator on the LED Display will light up. All LED lights will illuminate upon start-up. After the concentrator completes the warm-up cycle, only the green light will remain on.

Note: After initially powering ON the device, please allow up to ten (10) minutes for the device to reach its performance specifications.

g. Place the bottle in its designated space on the Humidifier

Bottle Support Stand.

h. Secure the bottle on the stand using the elastic band.

For OCSI Models: The O2 Light will illuminate for four (4) seconds upon start-up. After this initial start-up, this light is disabled for approximately five (5) minutes while the concentration rises to specifications.

i. Proceed to step 4.

Companion 5 User Manual — 11

Step 5: Adjust Flow Control Rate

a. Turn the flow control knob to the oxygen flow rate (LPM) prescribed by your physician.

Step 6: Verify Flow and Breathe Normally

a. Verify that oxygen is indeed flowing through the nasal cannula and that there are no kinks, bends, or blockages in the tubing.

Note: Ensure the cannula is fully inserted and secure. You should hear or feel oxygen flow to the prongs of the nasal cannula.

b. Properly position your nasal cannula and breathe normally through the nasal cannula.

To adjust flow rate:

Turn counter-clockwise to increase flow.

Turn clockwise to decrease flow.

The middle of the ball indicates flow rate. The image below indicates a flow rate of 4.0 LPM

WARNING: IT IS VERY IMPORTANT TO SELECT ONLY THE

PRESCRIBED LEVEL OF OXYGEN. DO NOT CHANGE THE

FLOW SELECTION UNLESS YOU HAVE BEEN DIRECTED TO DO

SO BY A LICENSED CLINICIAN. THE OXYGEN CONCENTRATOR MAY BE

USED DURING SLEEP UNDER THE RECOMMENDATION OF A QUALI-

FIED CLINICIAN.

12 — 15062781 E

Step 7: Turn Off

a. Press the power switch in the “OFF” (O) position when the

CAIRE Companion 5 is no longer in use.

CAUTION: Use only accessories recommended by the manufacturer. Use of any other may be hazardous, cause serious damage to your oxygen concentrator and will void the warranty.

CAUTION: Do not use extension cords with this unit or connect too many plugs into the same electrical outlet. The use of extension cords could adversely affect the performance of the device. Too many plugs into one outlet can result in an overload to the electrical panel causing the breaker/fuse to activate or fire if the breaker or fuse fails to operate.

CAUTION: Use of cables and adapters other than those specified, with the exception of cables and adapters sold by the manufacturer of the medical electrical equipment as replacement parts for internal components, may result in increased emissions of decreased immunity of the Oxygen Concentrator.

WARNING: DO NOT ALLOW EITHER THE AIR INTAKE OR THE

AIR OUTLET VENTS TO BECOME BLOCKED. THIS CAN CAUSE

THE OXYGEN CONCENTRATOR TO OVERHEAT AND IMPAIR

PERFORMANCE.

Note: To Equipment Provider: The following oxygen administration accessories are recommended for use with the Oxygen Concentrator:

• Nasal Cannula with 7 feet (2.1 m) of tubing: CAIRE Part Number

CU002-1

• Humidifier Adaptor Tubing: CAIRE Part number 20843882

• Humidifier Bottle: CAIRE Part Number HU003-1

• For a list of additional recommended accessories, please see the Accessories Catalog (PN ML-LOX0010) available on www.CAIREmedical.com.

• CAIRE offers the OxySafe™ as an optional accessory. This is intended to be used in conjunction with the Companion 5 concentrator. For customers in regions requiring compliance to EN ISO 8359:1996-Ammendment

1:2012, this accessory will meet this need.

The OxySafe™ is a thermal fuse to stop the flow of gas in the event that the downstream cannula or oxygen tubing is ignited and burns to the OxySafe. It is placed in-line with the nasal cannula or oxygen tubing between the patient and the oxygen outlet of the Companion 5.

For proper use of the OxySafe, always refer to the manufacturer’s instructions (included with each OxySafe kit).

CAIRE offers an OxySafe™ kit that includes a cannula headpiece,

OxySafe™, and 7’ tubing: PN 20628667

Alarm Conditions

Your CAIRE Companion 5 uses a combination of two or three

(for models with OCSI) LED lights and an audible alarm to alert you when there is a malfunction with your concentrator.

1) Green LED - Indicates normal operation.

2) Red LED - Indicates system malfunction. See Alarm

Conditions Table below for more details.

3) Yellow LED (OCSI Models Only) - Indicates low oxygen concentration. See Alarm Conditions Table below for more details.

The Alarm Conditions Table on the following page shows the different alarm conditions that can be displayed by the concentrator. It outlines possible causes and the actions you should take if the alarm is experienced. In the case of any alarm, it is recommended that you contact your healthcare provider as soon as possible.

WARNING: DO NOT IGNORE ALARMS

WARNINGS: THE MANUFACTURER RECOMMENDS AN

ALTERNATE SOURCE OF SUPPLEMENTAL OXYGEN IN THE

EVENT OF A POWER OUTAGE, ALARM CONDITION, OR MECHANICAL

FAILURE. CONSULT YOUR PHYSICIAN OR EQUIPMENT PROVIDER FOR

THE TYPE OF RESERVE SYSTEM REQUIRED.

Green

Red

Yellow

Green

Red

Above: Companion 5 with OCSI

Below: Companion 5 without OCSI

Companion 5 User Manual — 13

Alarm Conditions Table

Audible

Alarm

Off

Colored LED

Green Solid

Continuous

Continuous

Beeping

Continuous

Beeping

RED Solid

RED Solid

RED Flashing

Yellow Solid

Yellow Flashing

Cause Your Action

The concentrator is working properly. None

The concentrator has lost power but the power switch is still in the “ON” position.

Verify that the concentrator’s electrical cord is plugged into an outlet and that the outlet has power. Try a different outlet. If the problem continues, connect to a back-up oxygen supply and contact your healthcare provider.

System Malfunction

Product Flow Rate Too High or Too

Low

The Companion 5 has detected low oxygen levels.

The Companion 5 has detected low oxygen levels.

Turn the power switch in the “OFF” position immediately. Disconnect the AC

Power from the wall outlet. Wait 5 minutes. Connect the AC Power back into the wall outlet and turn the concentrator back on again. If the alarm continues, service is required. Connect to a backup oxygen supply and contact your healthcare provider immediately.

1. Ensure that the cannula is not kinked or blocked. If used with a humidifier bottle, ensure that it is filled properly and not creating a blockage.

2. Ensure that the Companion 5 has proper ventilation. It needs to be at least 12 inches (30 cm) from any surface to ensure the vents are not blocked.

3. If the problem persists, switch to an alternate source of oxygen and contact healthcare provider for assistance.

1. Ensure the air intake filter and exhaust locations are not clogged or restricted.

2. Ensure the Companion 5 is in a well-ventilated area. Make sure there are at least 12 inches between the back and sides of the Companion and any obstructions (furniture, curtain, etc.).

3. If the condition persists, switch to an alternate source of oxygen and contact your healthcare provider immediately.

Troubleshooting

It is recommended that you contact your healthcare provider if your concentrator displays any alarms or if you feel that your concentrator is not working properly. You should not attempt to repair the concentrator yourself. Repair should only be performed by a trained service technician. However, there are

User Troubleshooting Table

Problem

some troubleshooting steps that you can take if you experience problems with your CAIRE Companion 5. They are outlined in the following table. You can use this chart by following the checks for your problem in numerical order.

Possible Cause

1. The electrical cord is not plugged in to an outlet.

1. Plug in the electrical cord.

Your Action

The concentrator is turned on, but it is not running.

2. The electrical outlet is not providing power.

3. Internal failure.

1. The flow control setting has been changed.

2. Tubing has been disconnected.

2. Check your household fuse and circuit. Try a different outlet.

3. Connect to a back-up oxygen supply. Contact your healthcare provider immediately.

1. Verify that the flow control knob is set on your appropriate Liter Per

Minute (LPM) setting prescribed by your doctor.

2. Verify that the tubing is tight at all connections (outlet barb, humidifier bottle, water trap, etc). Re-secure any connections if necessary.

You are experiencing low or no oxygen flow.

3. The oxygen tubing or cannula is kinked.

4. There is a leak in your cannula or tubing.

5. Air flow into the concentrator is impeded.

6. Internal failure.

3. Check for kinks or blockages in your tubing. Repair if needed.

4. Inspect all tubing and the cannula for leaks. You can do this by either listening or feeling for escaping air. Replace leaky tubing or cannula.

5. Verify that nothing is blocking the inlets of the concentrator. Be sure it is at least 12 inches (30 cm) away from furniture, walls, or other obstructions.

6. Connect to a back-up oxygen supply. Contact your healthcare provider immediately.

Note: If your Companion 5 is still alarming after using the Alarm Conditions and Troubleshooting tables, please contact Technical Service at

1-800-482-2473 with the alarm code displayed on the LCD Display.

14 — 15062781 E

Cleaning, Care & Routine Maintenance

Routine Maintenance

Servicing of the Companion 5 must be conducted by your equipment provider. CAIRE recommends routine maintenance every 2 years. Your equipment provider will arrange and conduct maintenance on your Companion 5.

User Care and Cleaning of the Device

Cannula Replacement

Replace your supply tubing and cannula on a regular basis as recommended by your home care provider. Your physician or home care provider will provide you with cleaning, disinfection and replacement information.

Note: Always follow the cannula manufacturer’s instructions for proper use. Replace the disposable cannula as recommended by the cannula manufacturer or your Equipment Provider. Additional supplies are available from your Equipment Provider.

Humidifier Bottle (if Applicable)

You should check your humidifier bottle daily to ensure that its water level is between the specified minimum and maximum levels. Refill the bottle as necessary using distilled water. Clean the humidifier bottle in accordance with the manufacturers’ instructions.

Cleaning Procedure:

Turn OFF the Companion 5 and disconnect from AC power before any cleaning or disinfecting activity. DO NOT spray the outer case directly. Use a damp (not wet) cloth or sponge.

Spray the cloth or sponge with a mild detergent solution to clean the cabinet and power cord. To disinfect the Companion 5, use

Lysol

®

Brand II disinfectant (or equivalent). Dilute as directed by the manufacturer of the cleaning product, but do not spray liquid directly on the Companion 5.

Air Intake Filter

The Air Intake Filter should only be accessed or changed by your healthcare provider. The Air Intake Filter should be replaced every 2 years as part of the preventative maintenance schedule.

Replacement schedules and procedures are located in the CAIRE

Companion 5 Technical Manual.

Gross Particle Filter (Optional):

Your equipment provider may choose to install a gross particle

(foam) filter on your Companion 5. If so, this is located inside the device handle at the upper back of the Companion 5.

If this filter is installed, you should inspect and clean it once weekly. To clean the filters, use the following process:

1) Remove the gross particle filter.

2) Wash the filter in warm tap water using a mild soap detergent solution

3) Rinse the filter thoroughly with clear tap water and squeeze out the excess water

4) Allow the filter to air dry

5) Reinsert the filter in the cabinet

WARNING: ELECTRICAL SHOCK HAZARD. TURN OFF THE

UNIT AND DISCONNECT THE POWER CORD FROM THE

ELECTRICAL OUTLET BEFORE YOU CLEAN THE UNIT TO

PREVENT ACCIDENTAL ELECTRICAL SHOCK AND BURN HAZARD. ONLY

YOUR EQUIPMENT PROVIDER OR A QUALIFIED SERVICE TECHNICIAN

SHOULD REMOVE THE COVERS OR SERVICE THE UNIT.

WARNING: ELECTRICAL SHOCK HAZARD. DISCONNECT THE POWER

CORD FROM THE ELECTRICAL OUTLET BEFORE YOU CLEAN THE UNIT

TO PREVENT ACCIDENTAL ELECTRICAL SHOCK AND BURN HAZARD.

ONLY YOUR EQUIPMENT PROVIDER OR A QUALIFIED SERVICE TECHNI-

CIAN SHOULD REMOVE THE COVERS OR SERVICE THE UNIT.

WARNING: CARE SHOULD BE TAKEN TO PREVENT THE OXYGEN CON-

CENTRATOR FROM GETTING WET OR ALLOWING FLUIDS TO ENTER

THE UNIT. THIS CAN CAUSE A MALFUNCTION OR SHUT DOWN, AND

CAUSE AN INCREASED RISK FOR ELECTRICAL SHOCK OR BURNS.

WARNING: CLEAN THE CABINET, CONTROL PANEL, AND POWER

CORD ONLY WITH A MILD HOUSEHOLD CLEANER APPLIED WITH A

DAMP (NOT WET) CLOTH OR SPONGE, AND THEN WIPE ALL SURFACES

DRY. DO NOT ALLOW ANY LIQUID TO GET INSIDE THE DEVICE. PAY

SPECIAL ATTENTION TO THE OXYGEN OUTLET FOR THE CANNULA

CONNECTION TO MAKE SURE IT REMAINS FREE OF DUST, WATER,

AND PARTICLES.

WARNING: DO NOT USE LIQUID DIRECTLY ON THE UNIT. A LIST OF

UNDESIRABLE CHEMICAL AGENTS INCLUDES BUT IS NOT LIMITED TO

THE FOLLOWING: ALCOHOL AND ALCOHOL-BASED PRODUCTS, CON-

CENTRATED CHLORINE-BASED PRODUCTS (ETHYLENE CHLORIDE),

AND OIL- BASED PRODUCTS (PINE-SOL®, LESTOIL®). THESE ARE NOT

TO BE USED TO CLEAN THE PLASTIC HOUSING ON OXYGEN CONCEN-

TRATOR, AS THEY CAN DAMAGE THE UNIT’S PLASTIC.

Note: The Manufacturer does not recommend the sterilization of this equipment.

Note: Do not operate the Oxygen Concentrator without the air intake filter in place.

Companion 5 User Manual — 15

CAIRE Inc.

2200 Airport Industrial Drive, Ste. 500

Ball Ground, GA 30107 USA

1-800-482-2473

Worldwide: www.CAIREmedical.com

Medical Product Services GmbH

Borngasse 20

35619 Braunfels, Germany

Worldwide: www.CAIREmedical.com

CAIRE Inc. reserves the right to discontinue its products, or change the prices, materials, equipment, quality, descriptions, specifications and/or processes to its products at any time without prior notice and with no further obligation or consequence. All rights not expressly stated herein are reserved by us, as applicable.

Companion 5 ® is a trademark of CAIRE Inc., Ball Ground, GA 30107

Lysol ® is a registered trademark of Reckitt Benckiser, UK.

Copyright © 2017 Chart Industries PN 15062781 E

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement