Advanced and basic life

Advanced and basic life
Advanced and basic life-support
Philips M3535A HeartStart MRx Monitor/Defibrillator
No compromises. That’s how we created the HeartStart MRx, a combination transport monitor, cardiograph,
manual defibrillator, and AED. HeartStart MRx unites our industry-leading patient monitoring capabilities
with our patented biphasic defibrillation therapy in one comprehensive, easy-to-use, rugged, and lightweight
device, built to meet a variety of user needs.
With carefully arranged controls, connections,
and menus, HeartStart MRx efficiently provides
comprehensive diagnostic measurements without
compromising critical, potentially lifesaving therapy.
Monitoring and resuscitation functions are clearly
delineated, with monitoring located on one side of the
device and defibrillation, pacing, and cardioversion on
the other.
The HeartStart MRx is equipped for 3- and 5-Lead ECG
monitoring with arrhythmia detection, and optional 12Lead ECG, pulse oximetry, noninvasive blood pressure,
invasive pressures, temperature and end-tidal CO2
monitoring. Patient status and care data—including
numeric vital signs and up to four waves, as well as text
prompts, alarms, an event timer, event markers, and
battery status indicators—are logically organized on the
device’s large, backlit, color display. Vital Signs Trending
Reports and Event Summaries are also available.
To help caregivers perform high-quality CPR, the
Q-CPR® option is available. It offers real-time
measurement and corrective feedback on the rate,
depth, and duration of compressions, as well as the
frequency of ventilations. It also provides notification of
lack of CPR activity. Now with the CPR meter, feedback
appears on a graphical display right in the line of site of
the caregiver performing compressions.
2
By using the IntelliVue Networking Option, the
HeartStart MRx can connect to and communicate
with the Philips IntelliVue Clinical Network via a wired
network connection or wirelessly (where available).
When on the network, the HeartStart MRx performs
similarly to a bedside monitor, sending waveforms
and patient parameters along with alarms and INOP
messages to the Information Center while still providing
defibrillation and pacing capabilities at the bedside.
The HeartStart MRx also provides a suite of Data
Transmission options. Combining various options allows
you the flexibility of multiple transmissions avenues.
It is our mission to produce the finest quality medical
equipment, allowing you to deliver effective care and
ultimately save lives. We strive to meet the evolving
needs of the healthcare community with outstanding
technology and innovation, while always focusing on
what’s most important – quality patient care.
Whether it's the long-life batteries, automated
ready-for-use testing, industry-leading algorithms or
a data management solution, HeartStart MRx has the
capabilities you need and the performance you demand –
no compromises.
Features/Options/Upgrades
Standard Features
• ST/AR Basic algorithm for arrhythmia detection
• ECG monitoring through monitoring electrodes and
defibrillation pads
• Synchronized cardioversion
• Adjustable ECG size and autogain
• Manual and AED operation
• SMART Biphasic waveform for defibrillation therapy
• Large 4-wave color display
• Strip chart printer
• Individual, adjustable volume of QRS beeper, voice
prompts, and alerts
• Event summary
• Vital Signs Trending Report
• Configuration mode
• Service mode
• Operational checks
• Automated self-tests with “ready-for-use” indicator
• Lithium ion battery with fuel gauge
• ECG out capability (analog)
Standard Accessories
• AC power module
• Lithium ion battery with fuel gauge
• Hands-free multifunction electrode cable
• 3-Lead ECG cable
• Disposable monitoring electrodes
• Printer paper
• Defibrillator test load
• Documentation CD containing Instructions for Use,
User training workbook, and Application notes
• Quick reference cards
Package
M3535A
HeartStart MRx Monitor Defibrillator
Optional Features
• SpO2 with Fourier Artifact Suppression Technology
(FAST)
• Noninvasive Blood Pressure
• Microstream™ EtCO2
• Noninvasive Pacing
• Invasive Pressures (two channels)
• Temperature
• Audio Recording
• 12-Lead ECG with Philips DXL algorithm
• 12-Lead ECG and Event Summary Transmission
• 75mm Printer
• Q-CPR CPR measurement and feedback
• Q-CPR Data Capture
• IntelliVue Networking
• ACI-TIPI and TPI analysis
• Periodic Clinical Data Transmission
• Batch LAN Data Transfer (via LAN cable)
Some options, upgrades and accessories are not available in all countries. Contact your local Philips Sales Representative
for specific information.
3
Option Ordering Information
W23
Two-Year Bench Warranty (Europe
Only *)
A01
SpO2
W24
A02
SpO2 and NBP
Five-Year Bench Warranty with Loaner
(U.S. & Canada only)
A03
SpO2 , NBP, and EtCO2
WA1
A04
EtCO2
Five-Year Biomed Warranty (U.S., Canada
& Australia only)
A05
SpO2 , NBP, EtCO2 and Temperature
WA2
A06
SpO2 , NBP, EtCO2 , Invasive Pressures and
Temperature
Three-Year Biomed Warranty (U.S.,
Canada & Australia only)
W08
A07
SpO2 , NBP, Invasive Pressures and
Temperature
Three-Year Bench Warranty with Loaner
(Australia & New Zealand only)
A11
EtCO2 and SpO2
B01
External Pacing
B02
12-Lead ECG Acquisition
B04
75mm Printer
B05
Asian 75mm printer
B06
12-Lead ECG Transmission – Bluetooth
wireless technology
B08
Q-CPR
B09
Q-CPR Data Capture
B10
Event Summary – Bluetooth
B11
*
May not be available in all countries.
**
Includes support for IntelliVue iX.
Upgrades
861325
Event Summary – Bluetooth
861326
12-lead transmission, Rosetta-Lt
Interface (available in the U.S. only)
861356
IntelliVue Network Enabled
861357
IntelliVue Network Enabled w/radio
(Wired and Wireless – Available in the
U.S. only)
12-Lead Transmission, Rosetta-Lt™
Interface (Available in the U.S. only)
861359
Invasive Pressures
B12
Batch LAN Data Transfer
861360
Temperature
B14
Audio Recording (all modes)
861442
ACI-TIPI and TPI
B15
IntelliVue Network Enabled (Wired)**
861443
Periodic Clinical Data Transmission
B16
IntelliVue Network Enabled w/radio
(Wired and Wireless – Available in the
U.S. only)**
861444
CPR meter
861447
Batch LAN Data Transfer
989803153411
Internal Bluetooth Card
B17
ACI-TIPI and TPI
M3530A
SpO2
B18
Periodic Clinical Data Transmission
M3531A
NBP
Water Resistant External Paddles
M3532A
EtCO2
C03
Data Card
M3533A
Pacing
C04
Carrying Case
M3534A
C05
Additional Battery
C07
Barrel style Pad Cable (Replacement for
Standard Pad Cable)
12-Lead ECG
Option B02 – Acquisition
Option B04 – 75mm Printer
M3801A
12-Lead Transmission (Bluetooth)
C15
5-Lead ECG Cable
M3802A
LP1
Instructions for Use
12-Lead Transmission (RS232 and
Bluetooth)
LP2
User Training Video (English only)
M3806A
Device Software
LP3
User Training DVD (English only)
M3808A
Therapy PCA
SM1
Service Manual (English only)
M4760A
Handle and Cap Plate (for Pads)
SM3
Service Training Video (DVD,
English only)
M4765A
Option BO2 – B-Level Hardware
Upgrade
M4770A
Q-CPR CPR Measurement and Feedback
M4771A
Q-CPR Data Capture Upgrade
M4772A
Audio Recording Upgrade
M5527A
External Paddles with Paddle Tray
Option C01 – Standard Paddles
Option C02 – Water Resistant Paddles
C01
Option Ordering Information
W01
One-Year On-Site Warranty
®
Some options, upgrades and accessories are not available in all countries. Contact your local Philips Sales Representative
for specific information.
4
Supplies/Accessories
External Paddles
Q-CPR Accessories
M3542A
Standard External Paddles
989803162401
CPR meter
M3543A
Water Resistant External Paddles
989803163291
CPR meter Adhesive Pads
M4745A
Sterilizable External Paddles
989803158661
Pads/CPR meter Cable
M4761A
Compression Sensor
M4762A
Sensor Adhesive Pads (Package of 10)
M4763A
Compression Sensor Pads/CPR cable
Internal Paddles
M1741A
7.5cm Switchless
M1742A
6.0cm Switchless
M1743A
4.5cm Switchless
ECG Monitoring Electrodes
M1744A
2.8cm Switchless
M2202A
M4741A
7.5cm Switched
High-Tack Foam; 5 electrodes/pack
(60 packs/case)
M4742A
6.0cm Switched
M4612A
M4743A
4.5cm Switched
Solid Gel Electrodes, 5 electrodes/pack
(60 packs/case)
M4744A
2.8cm Switched
M4613A
Solid Gel Electrodes, 30 electrodes/pack
(10 packs/case)
M4740A
Internal Paddles Adapter Cable (use with
all M174xA series internal paddles)
ECG Cables
Multifunction Electrode Pads
12-Lead Cable Set
White barrel connector
M3525A
2.7 meter 10-lead ECG Trunk Cable,
12-pin Connector (for 3-Lead, 5-Lead and
12-Lead use)
989803147691
1.3 meter 10-lead ECG Trunk Cable,
12-pin Connector (for 3-Lead, 5-Lead and
12-Lead use)
M3526A
3-Lead ECG Set and Plug with Snap
(AAMI)
M3527A
Add 7-Lead ECG Set for 12-Lead use
(AAMI)
M3501A
Defib Adult, AAMI
M3502A
Defib Adult, IEC
M3503A
Defib Pediatric, IEC
M3504A
Defib Pediatric, AAMI
Gray plug connector
M3713A
Adult Plus
M3716A
Adult Radiolucent
M3717A
Pediatric Plus
M3528A
3-Lead ECG Set and Plug with Snap (IEC)
M3718A
Adult Radiotransparent/Reduced Skin
M3529A
Add 7-Lead ECG Set for 12-Lead use (IEC)
M3719A
Pediatric Radiotransparent/Reduced Skin
M5530A
Combiner Plug for 3-wire Lead Set for use
with M3526A/M3528A
M1663A
10-Lead ECG Patient Trunk Cable, 12-pin
ECG Input Connector (for 5-Lead and
12-Lead use)
M1949A
10-Lead ECG Patient Trunk Cable, 12-pin
ECG Input Connector (for 5-Lead and
12-Lead use)
M1968A
10-electrode Cable Set, Extremities,
Grabber (use with M1976A) (AAMI)
M1976A
10-electrode Cable Set, Chest, Grabber
(use with M1968A) (AAMI)
M1971A
10-electrode Cable Set, Extremities,
Grabber (use with M1978A) (IEC)
M1978A
10-electrode Cable Set, Chest, Grabber
(use with M1971A) (IEC)
Pads Cable
M3507A
Defib Hands-free, barrel style
7ft (2.2m)
M3508A
Defib Hands-free, plug style 7ft (2.2m)
05-10200
Pads Adapter (use with M3507A)
989803158661
Defibrillator Pads Hands-Free Cable,
HeartStart pads, CPR meter cable and
connector
5
ECG Cables
989803158061
989803158071
SpO2 Sensors/Cables
5-Lead ECG Lead Set; Limb Leads; Snaps;
Shielded Electrode (AAMI)
5-Lead ECG Lead Set; Chest Leads; Snaps;
Shielded Electrode (AAMI)
989803158081
5-Lead ECG Lead Set; Limb Leads; Snaps;
Shielded Electrode (IEC)
989803158091
5-Lead ECG Lead Set; Chest Leads; Snaps;
Shielded Electrode (IEC)
3-Lead Cable Set
M1191A
Reusable SpO2 Sensor – Adult Finger (2m)
M1191AL
Reusable SpO2 Sensor – Adult Finger (3m)
M1191B
Reusable SpO2 Sensor – Adult Finger (2m)
M1191BL
Reusable SpO2 Sensor – Adult Finger (3m)
M1191T
Reusable Adult Finger Sensor (Nellcor ®
9-pin D-sub connector)
M1192A
Reusable SpO2 Sensor – Pediatric/Small
Adult Finger
M1192T
Reusable Pediatric Finger Sensor (Nellcor ®
9-pin D-sub connector)
M1194A
Reusable SpO2 Sensor – Adult Ear Clip
M1195A
Reusable SpO2 Sensor – Infant
M1196A
Reusable Clip Adult Sensor
M1196T
Reusable Clip Adult Sensor (Nellcor 9-pin
D-sub connector)
M1669A
3-Lead Trunk Cable
M1500A
3-Lead ECG Trunk Cable (AAMI)
M1605A
3-Lead ECG Snaps (AAMI)
M1510A
3-Lead ECG Trunk Cable (IEC)
M1615A
3-Lead ECG Snaps (IEC)
M1671A
3-Lead ICU Grabber (AAMI)
M1672A
3-Lead ICU Grabber (IEC)
M1941A
SpO2 Extension Cable, 2m (6.5ft)
M1673A
3-Lead ICU Snaps (AAMI)
M1943A
1m Nellcor adapter
M1674A
3-Lead ICU Snaps (IEC)
M1131A
Disposable SpO2 Sensor – Adult/Pediatric
M1675A
3-Lead OR Grabber (AAMI)
M1678A
3-Lead OR Grabber (IEC)
989803170171
3-Lead OR Trunk Cable (AAMI/IEC)
5-Lead Cable Set
M1668A
5-Lead Trunk Cable
M1520A
5-Lead ECG Trunk Cable (AAMI)
M1625A
5-Lead ECG Snaps (AAMI)
M1530A
5-Lead ECG Trunk Cable (IEC)
M1635A
5-Lead ECG Snaps (IEC)
M1968A
5-Lead ICU Grabber (AAMI)
M1971A
5-Lead ICU Grabber (IEC)
M1644A
5-Lead ICU Snaps (AAMI)
M1645A
5-Lead ICU Snaps (IEC)
M1973A
5-Lead OR Grabber (AAMI)
M1974A
5-Lead OR Grabber (IEC)
M1976A
5-Lead Chest ICU Grabber (AAMI)
M1978A
5-Lead Chest ICU Grabber (IEC)
M1979A
5-Lead Chest OR Grabber (AAMI)
M1984A
5-Lead Chest OR Grabber (IEC)
M1602A
5-Lead Chest ICU Snaps (AAMI)
M1604A
5-Lead Chest ICU Snaps (IEC)
989803170181
5-Lead OR Trunk Cable (AAMI/IEC)
NBP
Interconnect Cables
M1598B
Adult Pressure 5ft. (1.5m)
M1599B
Adult Pressure 10ft. (3m)
Reusable Blood Pressure Cuffs
40400A
Reusable NBP Cuff Kit, 3 sizes (pediatric,
adult, large adult)
40400B
Reusable NBP Cuff Kit, 5 sizes (infant,
pediatric, adult, large adult, thigh)
40401A
Traditional Reusable NBP Cuff – Infant
40401B
Traditional Reusable NBP Cuff – Pediatric
40401C
Traditional Reusable NBP Cuff – Adult
40401D
Traditional Reusable NBP Cuff –
Large Adult
40401E
Traditional Reusable NBP Cuff – Thigh
M4552B
Easy Care Reusable NBP Cuff – Infant
M4553B
Easy Care Reusable NBP Cuff – Pediatric
M4554B
Easy Care NBP Cuff – Small Adult
M4555B
Easy Care Reusable NBP Cuff – Adult
M4557B
Easy Care Reusable NBP Cuff –
Large Adult
M4559B
Easy Care Reusable NBP Cuff – Thigh
M1572A
Multi-Patient Comfort Cuffs – Pediatric
M1573A
Multi-Patient Comfort Cuffs – Small Adult
M1783A
Sync Cable 8ft (2.4m)
M1574A
Multi-Patient Comfort Cuffs – Adult
M5526A
Sync Cable 25ft (7.6m)
M1575A
Multi-Patient Comfort Cuffs – Large Adult
Sync Cables
Some options, upgrades and accessories are not available in all countries. Contact your local Philips Sales Representative
for specific information.
6
NBP
Temperature
Disposable Blood Pressure Cuffs
Disposable Temperature Probes
M4572B
Soft Single-Patient Disposable Cuff –
Infant
21090A
Esophageal/rectal
M4573B
Soft Single-Patient Disposable Cuff –
Pediatric
21091A
Skin surface
21093A
Esophageal stethoscope
21094A
Esophageal stethoscope
21095A
Esophageal stethoscope
Foley Catheter
M4574B
Soft Single-Patient Disposable Cuff –
Small Adult
M4575B
Soft Single-Patient Disposable Cuff – Adult
21096A
M4576B
Soft Single-Patient Disposable Cuff –
Adult X-Long
21097A
Foley Catheter
Soft Single-Patient Disposable Cuff –
Large Adult
M1837A
Esophageal/rectal
M2255A
Foley Catheter
M4577B
M4578B
Soft Single-Patient Disposable Cuff –
Large Adult X-Long
M4579B
Soft Single-Patient Disposable Cuff – Thigh
Reusable Temperature Probes
21075A
Esophageal/rectal – adult
21076A
Esophageal/rectal – pediatric
21078A
Skin surface
Reusable Temperature Probe Extension Cables
EtCO2
Intubated Circuits
M1920A
FilterLine Set – Adult/Pediatric
(25 sets/case)
M1921A
Filter H Set – Humidified Adult/Pediatric
(25 sets/case)
M1923A
Filter H Set – Humidified Infant/Neonatal
(yellow, 25 sets/case)
Non-Intubated Dual Purpose Circuits (CO2 + O2)
M2520A
Smart CapnoLine – Pediatric
M2522A
Smart CapnoLine – Adult
Non-Intubated Single Purpose Circuits (CO2)
M2524A
Smart CapnoLine – Pediatric
M2526A
Smart CapnoLine – Adult
21082A
3.0m 2-pin plug extension for mini
phone plug
21082B
1.5m 2-pin plug extension for mini
phone plug
Power
M3538A
Lithium Ion Battery with fuel gauge
M3539A
AC Power Module
989803135301
2-Bay Battery Support System for Lithium
Ion Batteries
989803135331
4-Bay Battery Support System for Lithium
Ion Batteries
989803135341
4-Bay Battery Support System for Sealed
Lead Acid and Lithium Ion Batteries
Invasive Pressures
Networking
CPJ840J6
Reusable Pressure Transducer
989803148551
CPJ84022
Sterile disposable pressure dome for use
with CPJ840J6
Instrument Telemetry 1.4GHz radio and
A/C module (available in the U.S. only)
M4842-61300
CPJ84046
Transducer holder for CPJ840J6
1.4GHZ replacement antenna (available in
the U.S. only)
M1567A
Single channel disposable blood pressure
kit (Available in Europe and Asia only)
M1568A
M1634A
Dual Line blood pressure kit for
measuring CVP, ABP and other pressure
measurements (available in Europe and
Asia only)
Reusable adapter cable (available in
Europe and Asia only)
Disposable Transducers*
TransPac ®
IV ICU Medical, Inc.
TruWave ® PX212 Edwards Lifescience
Paper
40457C
50 mm Chemical Thermal, Gray Grid
(10 rolls)
40457D
50 mm Chemical Thermal, Gray Grid
(80 rolls)
989803138171
75 mm Chemical Thermal, Red Grid
(10 rolls)
989803138181
75 mm Chemical Thermal, Red Grid
(80 rolls)
DTX Plus™ DT-4812Becton, Dickinson and Co.
Available for purchase/service from their respective
*
manufacturers.
7
Specifications
Defibrillator
Color Handle
M5521A
Green and color label set
M5522A
Blue and color label set
M5523A
Yellow and color label set
M5524A
Rose and color label set
M5525A
Gray and color label set
Waveform:
Biphasic Truncated Exponential. Waveform
parameters adjusted as a function of patient
impedance
Shock Delivery:
Via multifunction defib electrode pads, or
paddles
Delivered Energy Accuracy:
Nominal Delivered Energy vs. Load Impedance
Miscellaneous
M1781A
Test Load for use with M3507A Pad Cable
M3725A
Test Load for use with M3508A Pad Cable
M3541A
Carrying Case (includes 3 accessory
pouches and shoulder strap)
989803146981
Data Card and Tray
M3537A
Bedrail Hook mount
M3549A
Wide Bedrail Hook mount
M4737A
Display cover
Selected Load Impedance (ohms) ± 2%
Energy
25
50
75
100 125 150
175
1J
1.2
1.3
1.2
1.1
1.0
0.9
0.8
±2 J
2J
1.8
2.0
2.0
1.9
1.7
1.6
1.5
±2 J
3J
2.8
3.0
3.0
3.1
3.0
2.9
2.7
±2 J
4J
3.7
4.0
4.0
4.1
4.2
4.2
4.0
±2 J
5J
4.6
5.0
5.1
5.1
5.2
5.2
5.0
±2 J
6J
5.5
6.0
6.1
6.2
6.3
6.3
6.1
±2 J
7J
6.4
7.0
7.1
7.2
7.3
7.3
7.1
±2 J
8J
7.4
8.0
8.1
8.2
8.4
8.3
8.1
±2 J
9J
8.3
9.0
9.1
9.3
9.4
9.4
9.1
±2 J
10 J
9.2
10
10
10
10
10
10
±2 J
15 J
14
15
15
15
16
16
15
±15%
20 J
18
20
20
21
21
21
20
±15%
30 J
28
30
30
31
31
31
30
±15%
50 J
46
50
51
51
52
52
50
±15%
70 J
64
70
71
72
73
73
71
±15%
100 J
92
100
101
103
104
104
101
±15%
120 J
110
120
121
123
125
125
121
±15%
150 J
138
150
152
154
157
156
151
±15%
170 J
156
170
172
175
177
177
172
±15%
200 J
184
200
202
206
209
209
202
±15%
Charge Time:
Accuracy
Less than 5 seconds to 200 joules with a new,
fully charged Lithium Ion battery pack at 25ºC.
Less than 15 seconds when operating without
a battery, using the M3539A AC power module
alone at 90% rated mains voltage.
Patient Impedance Range
Minimum:
15 ohm (internal defibrillation); 25 ohm
(external defibrillation)
Maximum:
180 ohm
Note: Actual functional range may exceed the above values.
8
General
ECG and Arrhythmia Monitoring
Dimensions
with pads:
12.4in (W) x 8.3in (D) x 11.7in (H)
(31.5cm x 21.0cm x 29.5cm)
Dimensions
with paddles:
13.4in (W) x 8.3in (D) x 13.6in (H)
(34.0cm x 21.0cm x 34.5cm)
Weight:
13.2lbs. (5.99kg) including pads, pads
cable, full roll of paper, and battery.
Incremental weight of external standard
paddles and paddle tray is 2.5lbs. (1.1kg).
Additional battery weighs less than 1.8lbs.
(0.82kg).
Manual Mode
Manual
Output Energy
(Selected):
1 – 10, 15, 20, 30, 50, 70, 100, 120, 150,
170, 200 J; maximum energy limited to 50
J for internal defibrillation
Controls:
On/Off Therapy Knob, Charge, Shock,
Sync, Print, Mark Event, ECG Lead Select,
Alarm Pause, Event Review, Disarm
Energy Selection Front panel Therapy Knob
Control:
Charge Control: Front panel button; button on external
paddles
Shock Control:
Front panel button; buttons on external
or switched internal paddles
Synchronized
Control:
Front panel SYNC button
Indicators:
Text prompts, audio alerts, QRS beeper,
battery status, Ready For Use, external
power, Sync mode
Armed
Indicators:
Charging tone, charged tone, flashing
Shock button, and energy level indicated
on display
Inputs:
Up to four (4) ECG waves may be
viewed on display and up to 2 waves
printed simultaneously. Lead I, II, or III is
obtained through the 3-lead ECG cable
and separate monitoring electrodes.
With a 5-lead cable, leads aVR, aVL,
aVF, and V can also be obtained. Pads
ECG is obtained through 2 multifunction
electrode pads.
Lead Fault:
Leads Off message and dashed line appear
on the display if an electrode or lead wire
becomes disconnected. Lead Off indicator
in wave sector.
Pad Fault:
Dashed line appears on the display if a pad
becomes disconnected.
Heart Rate
Display:
Digital readout on display from 15 to 300
bpm, with an accuracy of ±10%
Heart Rate/
Arrhythmia
Alarms:
HR, Asystole, VFIB/VTACH, VTACH,
Extreme Tachy, Extreme Brady, PVC rate,
Pacer not capture, Pacer not pacing
ECG Cable
Length:
9ft (2.7m)
Common Mode
Rejection:
Greater than 90 dB measured per AAMI
standard for cardiac monitors (EC 13)
ECG Size:
2.5, 5, 10, 20, 40 mm/mV, autogain
Frequency Response Bandwidth
AC Line Filter:
60Hz or 50Hz
3-Lead, 5-Lead,
and Pads:
Pads ECG for Display: Monitor (0.15 –
40Hz) or EMS (1 – 30Hz); Pads ECG for
Printer: Monitor (0.15 – 40Hz) or EMS (1
– 30Hz); Leads ECG for Display:
Monitor (0.15 – 40 Hz) or EMS (1 – 30Hz);
Leads ECG for Printer: Diagnostic (0.05 –
150Hz) or Monitor (0.15 – 40Hz) or EMS
(1 – 30Hz)
12-Lead:
ECG for Display: (0.0 5– 150Hz), (0.05
– 40Hz), (0.15 – 40Hz); ECG for Report:
(0.05 – 150Hz), (0.05 – 40Hz), (0.15 –
40Hz), (0.05 – 150Hz)
AED Mode
AED Energy
Profile:
150 joules nominal into a 50 ohm test load
Text and Voice
Prompts:
Extensive text/audible messages guide
user through configured protocol
AED Controls:
On/Off, Shock
Indicators:
Monitor display messages and prompts,
voice prompts, battery status, Ready For
Use, external power
Armed
Indicators:
Charging tone, charged tone, Flashing
shock button, energy level indicated on
display, and voice prompts
ECG Analysis:
Evaluates patient ECG and signal quality
to determine if a shock is appropriate
and evaluates connection impedance for
proper defibrillation pad contact
Shockable
Rhythms:
Ventricular fibrillation and certain
ventricular tachycardias, including
ventricular flutter and polymorphic
ventricular tachycardia
Shock Advisory
Algorithm
Sensitivity and
Specificity:
Meets AAMI DF-39
Patient Isolation (defibrillation proof):
ECG:
Type CF
SpO2:
Type CF
EtCO2:
Type CF
NBP:
Type CF
Invasive
Pressures:
Type CF
Temperature:
Type CF
External Defib:
Type BF
Internal Defib:
Type CF
CPR meter:
Type BF
9
Thermal Array Printer
Display
Size:
8.4in diagonal (128mm x 171mm)
Type:
TFT Color LCD
Resolution:
640 x 480 pixels (VGA)
Sweep Speed:
25mm/s nominal (stationary trace;
sweeping erase bar) for ECG, Invasive
Pressures and SpO2; 6.25mm/s for CO2
Wave Viewing
Time:
5 seconds (ECG)
Battery
Type:
Rechargeable, Lithium Ion; minimum
6.45Ah, 14.4V, 92WH
Dimensions:
6.5in (H) x 3.8in (W) x 1.6in (D)
(165mm x 95mm x 42mm)
Weight:
Less than 1.8lb (0.82kg)
Charge Time:
Approximately 3 hours to 100%.
Approximately 2 hours to 80%, indicated
by battery fuel gauge. Charging the
battery at temperatures above 45ºC may
degrade battery life.
Capacity:
At least 5 hours of monitoring with ECG,
SpO2 , CO2 , temperature and two invasive
pressures monitored continuously,
NBP measured every 15 minutes, and
20 200-joule discharges (on a new, fully
charged battery at room temperature,
25º C). At least 3.5 hours while pacing
at 180 ppm at 160 mA and monitoring as
described above
Battery
Indicators:
Fuel gauge on battery, capacity indicator
on display; flashing RFU indicator, chirp,
and LOW BATTERY message appears on
display for low battery condition*
Storage:
Storing the battery for extended periods
at temperatures above 40ºC will reduce
battery capacity and degrade battery life.
*Low battery condition triggered with at least 10 minutes of
monitoring time and 6 maximum energy discharges remaining
(with a new battery at room temperature, 25º C)
10
Continuous
ECG Strip:
The Print key starts and stops the strip.
The printer can be configured to run real
time or with a 10-second delay. The strip
prints the primary ECG lead with event
annotations and measurements.
Auto Printing:
The printer can be configured to
automatically print on Marked Events,
Charge, Shock, and Alarm. When an
alarm condition occurs, the unit prints the
primary ECG wave and the alarming wave,
if configured.
Reports:
The following reports can be printed:
Event Summary, Vital Signs Trending, 12Lead, Operational Check, Configuration,
Status Log, and Device Information
Speed:
25 or 50 mm/s with an accuracy of ± 5%
Amplitude
Accuracy:
± 5% or ± 40 uV, whichever is greater
Paper Size:
50mm (W) by 30m (100ft) (L)
75mm (W) by 30m (100ft) (L)
Non-Invasive Pacing
Waveform:
Monophasic Truncated Exponential
Current Pulse
Amplitude:
10 – 175mA (5mA increments); accuracy
10% or 5mA, whichever is greater
Pulse Width:
40 ms with ± 10% accuracy
Rate:
30 – 180 ppm (10 ppm increments);
accuracy ± 1.5%
Modes:
Demand or Fixed Rate
Refractory
Period:
340msec (30 – 80 ppm); 240msec
(90 – 180 ppm)
SpO2
Non-Invasive Blood Pressure
Range:
SpO2: 0 – 100%
Pressure Range:
Pulse rate:
30 – 300 bpm
Systolic:
40 – 260mmHg
Maximum Power
Output:
≤ 15mW
Diastolic:
20 – 200mmHg
Wavelength
Range:
500 – 1,000nm
Initial Pressure:
160mmHg (adult); 120mmHg (pediatric)
280mmHg
Resolution:
1%
Maximum
Pressure:
Display Update
Period:
1 sec. typical numeric update rate
Overpressure
Safety Limits:
Maximum of 300mmHg
Cuff Inflation
Time:
75 second maximum (pediatric or adult)
Pressure
Transducer
Accuracy:
±3 mmHg
Accuracy with sensor
M1191A
1 standard deviation 70 – 100%, ± 2.0%
M1191AL
1 standard deviation 70 – 100%, ± 2.0%
M1191B
1 standard deviation 70 – 100%, ± 2.0%
M1191BL
1 standard deviation 70 – 100%, ± 2.0%
M1191T
1 standard deviation 70 – 100%, ± 2.0%
M1192A
1 standard deviation 70 – 100%, ± 2.0%
M1192T
1 standard deviation 70 – 100%, ± 2.0%
M1194A
1 standard deviation 70 – 100%, ± 3.0%
M1195A
1 standard deviation 70 – 100%, ± 3.0%
M1196A
1 standard deviation 70 – 100%, ± 3.0%
M1196T
1 standard deviation 70 – 100%, ± 3.0%
M1131A
1 standard deviation 70 – 100%, ± 3.0%
Pulse Rate
Accuracy:
2% or 1 bpm (whichever is greater)
Pulse Alarm
Range:
Low Limit:
30 – 195 (adults); 30–235 (pediatric)
High Limit:
35 – 200 (adult); 35–240 (pediatric)
SpO2 Alarm
Range:
Low Limit:
50 – 99% (adult/pediatric)
High Limit:
51 – 100% (adult/pediatric)
SpO2 and Pulse
High/Low
Alarm Signal
Generation
Delay:
10 seconds
Alarm Range:
Systolic high
limit:
35 – 270 (adult), 35 – 180 (pediatric)
Systolic low
limit:
30 – 265 (adult), 30 – 175 (pediatric)
Diastolic high
limit:
15 – 245 (adult), 15 – 150 (pediatric)
Diastolic low
limit:
10 – 240 (adult), 10 – 145 (pediatric)
Mean high limit:
25 – 255 (adult), 25 – 160 (pediatric)
Mean low limit:
20 – 250 (adult), 20 – 155 (pediatric)
Calibration
Schedule:
Yearly or every 10,000 cycles
Auto Mode
Repetition
Time:
1, 2.5, 5, 10, 15, 30, 60, or 120 minutes
Measurement
Time:
Auto/manual mode: 30 seconds (average)
@ HR > 60 bpm, 170 seconds (maximum)
Interconnect
Tube Length:
M1598B Connect tubing 5ft (1.5m)
M1599B Connect tubing 10ft (3m)
Note: The above referenced sensors were validated for use with
the HeartStart MRx using the Philips picoSAT II SpO2 module
with Fourier Artifact Suppression Technology (FAST). This
module is not available as a stand-alone device.
11
End-Tidal Carbon Dioxide
Invasive Pressures
Range:
0 – 99mmHg at sea level
1mmHg (0.1kPa)
Transducer
Sensitivity:
5uV/V mmHg (37.5uV/V/kPa)
Resolution:
Accuracy:
For values between 0 – 38mmHg:
±2 mmHg. For values between 39 –
99mmHg: ±5% of reading + 0.08% for
every 1mmHg (above 40mmHg). For
breath rates above 80 and EtCO2 values >
18mmHg, accuracy is 4mmHg or ± 12% of
reading, whichever is greater.
Sensitivity
Adjustment
Range:
±10%
Drift of
Measurement
Accuracy:
Over any 24-hour period, the accuracy
claims above are maintained.
Alarm Range:
Low Limit:
10 – 94mmHg (adult/pediatric)
High Limit:
20 – 95 mmHg (adult/pediatric)
Calibration
Schedule:
Yearly or every 4,000 hours
Sample Size:
50ml per min.
Drift of
Measurement
Accuracy:
Over a 24-hour period, accuracy claims
above are maintained
Airway Respiration Rate
Range:
0–150 rpm
Resolution:
1 rpm
Accuracy:
0 – 40 rpm ±1 rpm
41 – 70 rpm ±2 rpm
71 – 100rpm ± 3 rpm
101 – 150rpm ± 5 rpm
Alarm Range:
Low Limit: 0 – 99 rpm (Adult/Pediatric)
High Limit: 10 – 100 rpm (Adult/Pediatric)
Apnea Alarm Time: 10 – 40 seconds, in
increments of 5
Calibration Gas for CO2 Measurement System
12
Ingredients:
5% Carbon Dioxide, 21% Oxygen,
74% Nitrogen
Cylinder Size:
BD
Method of
Preparation:
Gravimetric
Blend Tolerance:
0.03%
Accuracy:
0.03% absolute
Moisture:
10 PPM Maximum
Expiration
Period:
2 years
Pressure:
144 PSIG, Volume: 10 L
Transducer
195 – 2200 ohms
Load Resistance:
Transducer
Output
Resistance:
0 – 3000 ohms
Frequency
Response:
0 – 12Hz or 0 – 40Hz
Zero
Adjustment
Range:
±200mmHg (±26.7kPa)
Zero
Adjustment
Accuracy:
±1.0mmHg (±0.1kPa)
Zero Setting
Drift:
<0.1mmHg/ºC (0.013kPa/ºC)
Gain accuracy
(excluding
transducers):
±1% of reading or 1mmHg (0.1kPa)
whichever is greater
Gain Drift:
Less than 0.05% / ºC
Overall
Accuracy
(included listed
transducers):
±4% of reading or 4mmHg (0.5kPa)
whichever is greater
Measurement
Range:
-40 to 361mmHg (-5.3 to 48.1kPa)
Measurement
Resolution:
1mmHg (0.1kPa)
Noise:
<1mmHg (0.1kPa)
Transducer/
Dome Volume
Displacement:
Refer to the specific device's
specifications.
Additional Noise from EMI if operating under conditions
according to EMC standard EN60601-1-2 (Radiated Immunity
3 V/m or Conducted Immunity 3 VRMS): ≤3 mmHg
Pulse Rate
Range:
25 – 350 bpm
Pulse Rate
Accuracy:
1% of full range
Pulse Rate
Resolution:
1 bpm
Q-CPR
Temperature
Measurement
Range:
0º – 45ºC (32º – 113ºF)
Measurement
Resolution:
0.1ºC (0.2ºF)
Measurement
Accuracy
(excluding any
adapter cable):
+0.1ºC from 25ºC to 45ºC; +0.3ºC from
0ºC to 24.9ºC
Settling Time
Constant:
Measurements
Compressions:
Depth, rate, release (complete or
incomplete), and duty cycle.
Ventilations:
Volume, rate, and inflation time.
Feedback Type
Verbal:
Prioritized, corrective, verbal feedback for
all measurements.
<10 seconds
Numerical:
Measurement values for compression rate,
ventilation rate, and no flow time.
Averaging Time:
1 second
Graphical:
Minimum
measurement
time:
See the probe's Instructions for Use to
obtain minimum measurement times for
accurate readings. The HeartStart MRx
does not add any clinically significant time
to obtain accurate readings.
Compression wave with correct depth
target zone. Lung icon for ventilation
volume.
User Interface:
Integrated into Code (ALS resuscitation)
and AED (BLS resuscitation) views
CPR Meter
12-Lead ECG
Inputs:
ECG Bandwidth
Filters:
With a 10-Lead cable, leads I, II, III, aVR,
aVL, aVF, V/C1-V/C6 can be obtained. All
12-Lead ECG waves can be viewed on the
display simultaneously. All 12 leads can be
printed on the strip chart printer in 3x4
format.
0.15 – 40 Hz
0.05 – 40 Hz
0.05 – 150 Hz
Cellular transmission via a device with Bluetooth wireless
technology or a cell phone with an RS-232 connection.
12-Lead ECGs are transmitted through an ISP and 12-Lead
Transfer Station to printer, fax and/or TraceMaster ECG
Management System with version A.02.01 or greater.
Destinations:
Once a 12-lead reaches the 12-Lead
Transfer Station, it can be displayed,
printed, faxed, emailed, or forwarded to
another 12-Lead Transfer Station. It can
also be forwarded to the TraceMaster
ECG Management System or other
ECG management systems (via the
DatamedFT).
Patient Data Storage
Internal Event Summary:
The internal Event Summary stores up to 12 hours of 2
continuous ECG waves, 1 CO2 wave and 2 invasive pressure
waves, events and trending per event summary. There is a
maximum capacity of 55 Event Summaries or 240 megabytes
(62 megabytes if a 64 megabyte card is installed) of patient
data, whichever comes first.
Data Card Event Summary:
The Data Card has a maximum capacity of 60 Event
Summaries or 240 megabytes (62 megabytes if a 64 megabyte
card is installed) of patient data, whichever comes first.
Dimensions:
154mm x 64mm x 28mm with a .91m
integrated cable.
Weight:
8.3oz (235g)
Input Voltage:
4.0 – 6.0V at 170mA. The CPR meter
is electrically and galvanically isolated
from the defibrillator power and
communication sources.
Storage
Temperature:
-20ºC to 60ºC (-4ºF to 140ºF)
Operating
Temperature:
0ºC to 50ºC (32ºF to 122ºF).
Storage Relative
Humidity:
0 – 75%
Operating
Relative
Humidity:
0 – 95%
Solids/Water
Resistance:
IP55. Meets ISO/IEC 60529.
EMC:
Meets IEC 60601-1-2 and RTCA/
DO-160E.
Patient Adhesive Pads
Dimensions:
39mm x 90 mm
Temperature:
Storage:
-20°C to 60°C (-4°F to 140°F)
Operating:
0°C to 50°C (32°F to 22°F)
Relative
Humidity:
Storage: 0 – 75%
Operating: 0 – 95%
Material:
Foam pad with biocompatible adhesive on
both sides
Shelf life:
2 years when applied to the CPR meter or
4 years in an unopened package.
13
Bluetooth Wireless Technology Card
Environmental
Bluetooth
Class I:
100 meters (approximately 300 feet)
maximum transmission range. Dependent
upon transmission range of lowest class
Bluetooth device. Most Bluetooth devices
are Class II, which transmit at maximum
ranges of up to 10 meters (33 feet).
Temperature:
0º C to 45ºC operating, -20º to 70ºC
storage
Humidity:
Up to 95% relative humidity
Atmospheric
Pressure Range:
Operating and Storage – 1014hPa to
572hPa (0 to 15,000ft; 0 to 4,500m)
Bluetooth
Stacks:
Tested with Toshiba™ 4.20.11, IVT™
2.1.2.0 (Product)/05.04.11.20060301
(stack), Widcomm™ 4.0.1.2400.
Shock:
Operating: Half-sine waveform, duration
< 3ms, acceleration > 145g, 1 time on all
six face
Bluetooth
Version:
1.1 or greater
Non Operating:
Trapezoidal waveform, acceleration ≥ 30
g, velocity change=742cm/s ±10% on all
six faces
Vibration:
Operating: Random vibration, 0.30 grms, 5
– 500Hz for ≥10 minutes/axis PSD=0.0002
g2/Hz from 5 to 350Hz, -6 dB/octave slope
from 350 to 500Hz
Storage: Random vibration, 2.41 grms,
5 – 500 Hz for ≥10 minutes/axis PSD=0.02
g2/Hz from 5 – 100 Hz, -6 dB/octave slope
from 100 – 137 Hz, 0 dB/octave slope
from 137 – 350 Hz, -6 dB/octave slope
from 350 – 500 Hz
Water/Solids
Resistance:
IP24 if used with water-resistant paddles
(M3543A) or pads only. IP20 if used with
other paddles.
EMC:
Complies with the requirements of
standard EN 60601-1-2:2001
Safety:
Meets the UL 2601-1, CSA C22.2 No.
601-1, EN 60601-1 and 60601-2-4
standards.
Other
Considerations:
Device not suitable for use in the presence
of concentrated oxygen or a flammable
anesthetic mixture with air, oxygen, or
nitrous oxide
Mode of
Operation:
Continuous
AC Power
Module:
Input: 100-240 VAC, 50-60 Hz, 1-0.46 A
(Class 1)
Output: 18V, 5A, 90W
Battery:
Minimum 14.4 V Rechargeable, Lithium Ion
Bluetooth devices used with the MRx must support the
Bluetooth Dialup Networking Profile (DUN) or the File
Transfer Profile (FTP). DUN devices must also have a data
transfer plan that supports packet data transmission. Event
summaries can only be transmitted via Bluetooth File
Transfer Profile (not DUN).
Networking
Connection:
Connects via Local Area Network (LAN)
to the Philips IntelliVue Clinical Network
(System L or later)
Connection
types:
Wired or wireless. Wireless uses a radio
module attached to the AC power supply
and is available in the U.S. only.
Telemetry
channels:
The maximum number of telemetry
channels may be limited in some
geographies
Device meets FCC 47 CFR Part 15 and Part 95 for WMTS
transmission.
14
15
Philips Healthcare is part of
Royal Philips Electronics
How to reach us
www.philips.com/healthcare
healthcare@philips.com
The Bluetooth word mark and logos are registered trademarks owned by the Bluetooth
SIG, Inc., and any use of such marks by Philips Medical Systems is under license.
Koninklijke Philips Electronics, N.V. is an Associated Member of the Bluetooth SIG.
Please visit www.philips.com/heartstart
© 2013 Koninklijke Philips Electronics N.V.
All rights are reserved.
Philips Healthcare reserves the right to make changes in specifications and/or to discontinue any product at any time without notice or
obligation and will not be liable for any consequences resulting from the use of this publication.
Printed in The Netherlands.
4522 962 92881 * FEB 2013
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising