AquaShock® pH Kit 850057K
AquaShock® pH Kit
850057K
AquaShock® pH Kit 850057K
Copyright ©2014 by Sper Scientific
ALL RIGHTS RESERVED
Printed in the USA
The contents of this manual may not be reproduced or transmitted
in any form or by any means electronic, mechanical, or other
means that do not yet exist or may be developed, including
photocopying, recording, or any information storage and retrieval
system without the express permission from Sper Scientific.
Sper Scientific
8281 E. Evans Rd., Suite #103, Scottsdale, AZ 85260 Tel: (480) 948-4448 Fax: (480) 967-8736
Web: www.sperscientific.com
2
TABLE OF CONTENTS
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
FEATURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
MATERIALS SUPPLIED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
LCD DISPLAY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
POWER SUPPLY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
KEYPAD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
SETUP MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
MEASUREMENT PROCEDURES . . . . . . . . . . . . . . . . . . . . . . 12
OPTIONAL ACCESSORIES . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3
INTRODUCTION
The Sper Scientific AquaShock® pH Meter is the most waterproof,
rugged and reliable pH meter on the market. It is fully waterproof,
including probe connections, when used with the AquaShock®
probe. As with all AquaShock® meters, the pH Meter is fully
shockproof, floats, and features a double injection molded
case with an integral protective soft grip outer layer, high end
antishock, LED backlight display, and is powered by the latest
environmentally friendly, long lasting, rechargeable Lithium Ion
battery. The meter is highly accurate and stable, very intuitive and
includes all of the functions required by most users. Sper Scientific
guarantees you will not find any meters more rugged and reliable
than AquaShock®.
4
FEATURES
• Floats
• Shockproof
• IP67 Waterproof
• Protective soft grip outer layer
• Rechargeable Lithium Ion battery
• Also accepts standard BNC probes
• Multi-line display LCD screen
• Reads pH and mV
• Automatic or manual temperature compensation
• Automatic buffer recognition
• Celsius or Fahrenheit selectable
• 99 data points for analysis
• Calibration point review
• Internal clock and calendar
• Maximum, minimum and average
• Hold function
• Low battery indicator
• Backlight
MATERIALS SUPPLIED
• Meter
• Dual pH/ATC Probe
• pH Buffers, Distilled Water
• 1 Lithium 1450 mAh 7.4 V Battery
• DC Power Adapter
• Instruction Manual
• Hard Carrying Case
5
LCD DISPLAY
1
3
4
5
6
2
7
8
1
2
Name
Ready
3
4
5
Electrode
Durability Icon
Scale Unit
Measuring Results
ATC/MTC
6
Temperature Scale
7
Temperature Value
8
Battery Icon
Description
Displayed when measuring
results are stable
Current electrode durability
Current measurement scale unit
Current measurement value
Automatic/Manual Temperature
Compensation indicator
Current temperature
measurement unit
Current temperature
measurement value
Current battery power level
6
LCD DISPLAY
Probe Life Icon:
More than 90% life remaining
More than 85 – 90% life remaining
More than 80 -85% life remaining
Less than 80% life remaining
Battery Power Icon:
Battery voltage is more than 7.7V shows full.
Battery voltage between 7.2V and 7.7V shows 2 cells.
Battery voltage between 6.6V and 7.2V shows 1 cell.
Battery voltage less than 6.6V shows 0 cells and flashes.
Display showing battery charging from
less than 6.6V to fully charged.
POWER SUPPLY
This meter is powered by one rechargeable Lithium 7.4V (1450
mAh) battery. The DC power adapter can be used to charge the
battery even when the meter is turned off. This meter includes a
charge protection function and will automatically stop charging
when the battery is full. A complete re-charge takes approximately
8 hours.
7
KEYPAD
1
2
3
4
5
6
7
8
Name
1st Function
2nd Function
1
Power/Backlight/
Low Battery
Power on/off
Backlight on/off
2
Setup
Setting parameters
Confirm setting
3
MI↑
Store data
↑
4
MR↓
Recall data
↓
5
Max/Min/Ave
Max/Min/Ave
6
Range/Temp
Switch measuring
scale
Set temperature value
7
Hold/Exit
Hold
Exit
8
CAL
Calibration
8
SETUP MODE
The Setup Mode allows you to customize the following
meter preferences and defaults:
• Temperature Units
• Clock Setting
• Restore Factory Defaults
• View Calibration Data
• View Probe Parameters
1.Press POWER for 1 second to turn the meter on before
performing any setup function.
Temperature Units
1. Press SETUP to enter the Setup Mode.
2. Press SETUP to enter the Set Temperature screen.
3. The unit °C or °F will flash on the LCD.
4. P
ress ↑ or ↓ to select the °C or °F temperature unit.
5. Press SETUP to save and return to Normal Mode.
6. Press EXIT to return to Normal Mode without saving the selection.
Real Time Clock
1. Press SETUP to enter the Setup Mode.
2. Press ↑ or ↓ to select the Set Clock screen.
3. Press SETUP to set the year. The value will flash on the LCD.
4. P
ress ↑ to increase the value by 1. Hold down ↑
to continually increase the value.
Press ↓ to decrease the value by 1. Hold down ↓
to continually decrease the value.
Press SETUP to save and the meter automatically
moves to the month setting.
5. Repeat step 4 to adjust the month, day, hour, and minute.
6. Press EXIT to return to Normal Mode without saving.
Note...
This procedure adjusts the meter’s internal clock. The internal
clock will function when the meter is turned off but only when the
battery is installed.
9
Restore Factory Settings
1. Press SETUP to enter the Setup Mode.
2. Press ↑ or ↓ to select the Restore Factory Settings screen.
3. Press SETUP to confirm selection. “NO” will flash on the LCD.
4. Press ↑ or ↓ to select “YES” or “NO.”
5. When “YES” is displayed, press SETUP to confirm and all
parameters will restore to factory settings (except the clock)
and the meter returns to Normal Mode.
6. Press EXIT to return to Normal Mode without saving.
View Calibration Data
1. Press SETUP to enter the Setup Mode.
2. Press ↑ or ↓ to select the View CAL Data screen.
3. Press SETUP to enter the mode.
4. P
ress ↓ to view the previous calibration point. Press ↑ to
view the next calibration point. The date and time is on
cyclical display.
5. Press EXIT to return to Normal Mode.
View Probe Parameters
1. Press SETUP to enter
the Setup Mode.
2. P
ress ↑ or ↓ to select the View
Probe Parameters screen.
Note...
If the meter is not calibrated, the screen will show:
Press SETUP or EXIT to
return to Normal Mode.
10
CALIBRATION
Calibration should be performed whenever the probe has been
out of use for 48 hours or longer. To prepare for calibration:
1. Immerse the pH probe in standard buffer solution for
30 minutes or longer.
Calibrate With Temperature Sensor/ATC pH Probe
1. Press CAL for 2 seconds to enter Calibration Mode.
2. T
he meter will automatically recognize the calibration
buffer. The value and “Finished” will display on
the LCD when the procedure is complete.
3. Press EXIT to return to Normal Mode.
4. Repeat Steps 1-2 to calibrate other points.
Note...
All other keys will be inoperable while using the calibration function.
Calibrate Without Temperature Sensor/MTC pH Probe
1. Press CAL for 2 seconds to enter Temperature Setup Mode.
The temperature value will flash on the LCD.
2. P
ress ↑ to increase the value. Hold down ↑
to continually increase the value.
Press ↓ to decrease the value. Hold down ↓
to continually decrease the value.
For °C, the range is 0 to 60.0°C; for °F,
the range is 32 to 140.0°F.
Press SETUP and the meter will automatically recognize the
calibration buffer. The value and “Finished” will display on the
LCD when the procedure is complete.
Repeat Step 2 to calibrate other points.
3. Press EXIT to return to Normal Mode.
11
MEASUREMENT PROCEDURES
Turning the Unit On/Off
1. Press POWER to turn the meter on.
2. Press and hold POWER for 2 seconds to turn the meter off.
Backlight
1. With the meter on press POWER to turn the backlight on/off.
Scale Selection
1. Press RANGE to switch between the pH and mV scales.
Setting the MTC Temperature
Do NOT connect the temperature sensor or non-ATC probe
while setting the temperature.
1. From Normal Mode, press RANGE for 2 seconds to enter
Set Temperature Mode. The temperature value will flash
on the LCD.
2. P
ress ↑ to increase the value. Press and hold ↑
to increase the value continuously.
Press ↓ to decrease the value. Press and hold ↓
to decrease the value continuously.
For °C, the range is 0 to 60.0°C; for °F, the range is 32 to 140.0°F.
3. Press SETUP to save and the meter returns to Normal Mode.
4. P
ress EXIT to return to Normal Mode without saving.
Saving to Memory
1. Press MI to save the current value and view the data point
number. There are 99 data points available.
12
Recall Memory
1. Press RANGE to select the scale (pH or mV) for recall.
2. P
ress MR to enter Recall Mode and view the last measured
value. The date and time are on cyclical display.
3. Press ↑ or ↓ to scroll saved data points.
4. Press EXIT to return to Normal Mode.
Clear Memory
1. Press MI and MR simultaneously for 2 seconds to clear all
memory. The meter will return to Normal Mode.
Maximum, Minimum, and Average
1. Press RANGE to select the scale.
2. P
ress MAX/MIN/AVE, the maximum recorded value will
appear on the LCD. The date and time is on cyclical display.
3. P
ress MAX/MIN/AVE, the minimum recorded value will
appear on the LCD. The date and time is on cyclical display.
4. P
ress MAX/MIN/AVE, the average recorded value will appear
on the LCD. The date and time is on cyclical display.
5. Press MAX/MIN/AVE to return to Normal Mode.
Hold
1. Press HOLD to hold the value.
2. Press HOLD to release the hold value.
Note...
When in Hold Mode, all other functions will be disabled
except turning the meter off, backlight function, saving, and
exiting Hold Mode.
13
OPTIONAL ACCESSORIES
840016 840049 840093 850057P 860008 860009 860010 860011 860012 Standard pH Probe (non-ATC)
Spear Tip pH Probe (non-ATC)
Field Tripod
AquaShock® ATC pH Probe
pH 4 Buffer Solution - 3 bottles of 40ml each
pH 7 Buffer Solution - 3 bottles of 40ml each
pH 10 Buffer Solution - 3 bottles of 40ml each
Deionized Water - 3 bottles of 40ml each
pH Buffer Set - 1 each pH 4, 7, 10
14
SPECIFICATIONS
Mode
Range
pH
0 to 14 pH
Resolution
Accuracy
Input
Calibration
0.01 pH
±0.01 pH
BNC
Connector
Up to 5 points
Memory
Up to 99 data points
pH Slope
Range
Temp
Compensation
Operating
Temp
Battery
Battery Life
Temperature
32 to 140°F
(0 to 60°C)
0.1°F (0.1°C)
±0.9°F (±0.5°C)
Temperature
Probe Connector
mV
-1999 to
1999 mV
0.1 mV
±0.3 mV
BNC
Connector
75% to 110%
Automatic (ATC) or manual (MTC) from
32 to 140°F or 0 to 60°C
32 to 122°F (0 to 50°C)
Lithium battery 7.4 V (1450 mA/h)
> 200 hours
Power Button Indicator Light
Power Off
Power On
Low Battery
Charging Battery
Light off
Green light for 5 seconds, then light off
Red light flashes every 5 seconds
Blue light on
15
WARRANTY
Sper Scientific warrants this product against defects in materials
and workmanship for a period of five (5) years from the date
of purchase, and agrees to repair or replace any defective unit
without charge. If your model has since been discontinued, an
equivalent Sper Scientific product will be substituted if available.
This warranty does not cover probes, batteries, battery leakage,
or damage resulting from accident, tampering, misuse, or abuse of
the product. Opening the meter to expose its electronics will break
the waterproof seal and void the warranty.
To obtain warranty service, ship the unit postage prepaid to:
SPER SCIENTIFIC LTD.
8281 East Evans Road, Suite #103
Scottsdale, AZ 85260
The defective unit must be accompanied by a description of the
problem and your return address. Register your product online at
www.sperwarranty.com within 10 days of purchase.
Revised 2/8/2017