SECULIFE HITAM and SECULIFE HITMD

SECULIFE  HITAM and SECULIFE  HITMD
SECULIFEHITAM and SECULIFEHITMD
TRMS Medical-Multimeter
3-349-453-03
8/4.17
Digital TRMS handheld multimeter with 23 functions
•
•
Voltage measurement
Auto-ranging current measurement from 100 A
(resolution: 10 nA) to 10 A (16 A)
• Capacitance and resistance measurement, diode and
continuity testing
• Measuring categories: 600 V CAT III and 300 V CAT IV
• 1 kHz low-pass filter
• TRMS AC and AC+DC, 20 kHz bandwidth
• Data memory for more than 15,000 measured values
• Extremely rugged, dust and water-proof variant with IP65
• SECULIFE HITAM: Housing, protective rubber holster and set of measuring
cables each with antimicrobial properties
• SECULIFE HITMD: Multimeter with hygiene expertise
Interface
• Bidirectional infrared interface for communication with the PC
• IR-USB adapter available as option
Power Supply
German
• Battery operation
Accreditation Body
• Mains operation via optional broad range variable power pack DAkkS CalibrationD-K-15080-01-01
Certificate as Standard Feature
IP65
Applications
The multimeter has been designed especially for use in the field of
medical technology, amongst other purposes for servicing,
repairing and testing medical devices.
Three Connector Jacks with Automatic Blocking Sockets (ABS) *
All current ranges are implemented via a single connector jack
which prevents any possibility of operator error.
Beyond this, the automatic blocking sockets prevent incorrect connection of the measurement cables, as well as selection of the wrong
measured quantity. Danger to the user, the instrument and the device
under test resulting from operator error is thus ruled out.
Features
* Patented (patent no. EP 1801 598 and US 7,439,725)
SECULIFE HITAM
The multimeters of the SECULIFE HIT AM series have been endowed
with antimicrobial properties. This is to curb the growth of germs,
counteract microbial colonization or kill microorganisms.
SECULIFE HITMD
SECULIFE HITMD is protected against the penetration of fluids and
resistant to disinfectants. A hygiene expertise has been issued by
the Department for Medical Microbiology and Hygiene at the
Philipps University Marburg to confirm the suitability for use in
hygienically sensitive areas.
On the basis of the specifications set forth by the Deutsche
Gesellschaft für Hygiene und Mikrobiologie (DGHM) (German
Association for Hygiene and Microbiology) and the Vereinigung für
Angewandte Hygiene (VAH) (Association for Applied Hygiene) the
possibilities for safe disinfection of the SECULIFE HITMD have been
tested. The tests were performed with regard to test microbes in
analogy and in conformity with the specifications for disinfection
methods set forth by the Deutsche Gesellschaft für Hygiene und
Mikrobiologie (DGHM) within the framework of a stress test in a
practice-related context.
GMC-I Messtechnik GmbH
Overload Protection
The instrument is safeguarded for up to 1000 V in all measuring
functions by overload protection. Voltages of greater than 1000 V
and current of greater than 10 or 16 A are indicated acoustically.
Dangerous contact voltages are indicated when the 1 kHz lowpass filter is activated.
The FUSE display appears in order to indicate that the fuse for the
current measuring input has blown.
RMS Value with Distorted Waveshape
The utilized measuring method allows for waveshape
independent RMS measurement (TRMS AC and AC+DC) for
voltage and current up to 20 kHz.
Activatable Filter for V AC Measurement
A 1 kHz low-pass filter can be activated if required, for example
when measuring at electronic frequency converters or switchedmode power supplies.
High Voltage sensor
The input signal is examined for contact danger regardless of the
selected input function or filtering.
SECULIFEHITAM and SECULIFEHITMD
TRMS Medical-Multimeter
Measuring 5 V Square-Wave Signals
This function makes it possible to test circuits and transmission
cables by measuring the frequency and the keying ratio of pulses
with amplitudes of 2 to 5 V and frequencies of 100 Hz to 1 MHz.
Analog Scale for Quick Trend Display – Bar Graph or Pointer
The analog scale (with additional negative range for zerofrequency quantities) allows for faster recognition of measured
value fluctuation than is possible with a digital display.
The instrument can be switched back and forth between bar
graph and pointer display.
Automatic or Manual Measuring Range Selection
Measured quantities are selected by means of a rotary switch and
a function key. The measuring range is automatically matched to
the measured values. The measuring range can also be selected
and fixed manually with a key.
DAkkS Calibration Certificate
The multimeters are furnished with an internationally valid DAkkS
calibration certificate (recognized by EA and ILAC). After the
specified calibration interval has elapsed (recommended interval:
1 to 3 years), the multimeters can be inexpensively recalibrated in
our own DAkkS calibration laboratory.
Functions
Voltage VAC TRMS (Ri  9 M )
100 mV / 1 V / 10 V / 100 V / 1000 V
Frequency Hz @ VAC
100 Hz / 1 kHz / 10 kHz / 100 kHz
Voltage Lo 1) VAC TRMS (Ri = 1 M)
100 mV / 1 V / 10 V / 100 V / 1000 V
Frequency Hz @ Lo 1) VAC
100 Hz / 1 kHz / 10 kHz / 100 kHz
1kHz @ Lo VAC or @ Hz
Low-pass filter
Voltage VDC (Ri 9 M)
100 mV / 1 V / 10 V / 100 V / 1000 V
Voltage VAC+DC TRMS (Ri 9M)
100 mV / 1 V / 10 V / 100 V / 1000 V
Bandwidth @ VAC+DC or VAC
20 kHz
Fast Acoustic Continuity Test
Testing for short circuiting and interruption is possible with the
selector switch in the
position. The threshold value for
acoustic signaling can be set to 1, 10, 20, 30, 40 or 90 .
Frequency MHz @ 5 V AC
100 Hz...1 MHz
Continuity test
0 ... 100 @ ICONST = 1 mA
Automatic Storage of Measured Values *
The DATA function automatically saves the digitally displayed
measured value after settling in. Acoustic signaling is also used to
indicate whether the new measured value deviates from the initial
reference value by less or more than 0.1% of the measuring
range.
Diode measurement
0 ... 5,1 V @ ICONST = 1 mA
Duty cycle %
2,0 % ... 98 %
Resistance 
100 / 1 k/ 10 k/ 100 k/
1 M/ 10 M/ 40 M
Temperature measurement °C / °F @ TC
Thermocouple Type K
Temperature measurement °C / °F @ RTD
Pt100 / Pt1000
Capacitance measurement F
10 nF / 100 nF / 1 F / 10 F /
100 F / 1000 F
Current ADC
100 A / 1 mA / 10 mA / 100 mA /
1 A / 10 A (16 A)
Storage of Min-Max Values
Comparable to the slave-pointer function of an analog instrument,
the device saves the highest and lowest measured values after
the MIN/MAX function has been activated or reset. These extreme
values can be queried at the display. The values are acquired
using an especially fast sampling rate (40 measurements per
second).
Current AAC+DC TRMS
100 A / 1 mA / 10 mA / 100 mA /
1A / 10 A (16 A)
Current AAC TRMS
100 A / 1 mA / 10 mA / 100 mA /
1 A / 10 A (16 A)
Bandwidth @ AAC+DC or AAC
10 kHz
Frequency Hz @ AAC
100 Hz / 1 kHz / 10 kHz / 30 kHz
Data logger function 2) (memory)
4 Mbit = 500 kByte = 15400 measured values
Battery Charging Status – Power Saving Circuit
The battery charge level is indicated by means of a 4-segment
symbol.
The device is switched off automatically if the measured value
remains unchanged for a period of between 10 and 59 minutes
(adjustable), and if none of the controls are activated during this time.
Automatic shutdown can be deactivated by switching the
instrument to continuous operation.
IR-Interface
38400 Bd
* Patented
Protective Cover for Harsh Conditions
The instrument is protected against damage in the event of
impacts or dropping by means of a soft rubber cover with tilt
stand and test probe holder. The rubber material also assures
that the instrument does not wander if it is set up on a vibrating
surface.
Infrared Data Interface
The device can be remote configured, and momentary and stored
measurement data can be read out via the bidirectional infrared
interface. The USB X-TRA interface adapter and METRAwin 10
software are required to this end (see accessories). Interface
protocol and device driver software for LabVIEW® (National
Instruments™) are available upon request.
The infrared interface can be switched off in the standby mode.
2
3
Power pack connector socket
Protection
IP65
Measurement category
600 V CAT III and 300 V CAT IV
DAkkS calibration certificate
3
Protective rubber cover
3
1) Alternating voltage measurement with specially reduced input impedance
2) Sampling rate adjustable from 0.1 seconds to 9 hours
Applicable Regulations and Standards
IEC/DIN EN 61010 -1
VDE 0411-1
Safety requirements for electrical equipment for
measurement, control and laboratory use
DIN EN 61326-1
VDE 0843-20-1
Electrical equipment for measurement, control and laboratory
use – EMC requirements – Part 1: General requirements
DIN EN 60529
DIN VDE 0470-1
Test instruments and test procedures
– degrees of protection provided by enclosures (IP code)
Voluntary Manufacturer’s Guarantee
24 months for materials and workmanship
1 to 3 years for calibration (depending upon application)
GMC-I Messtechnik GmbH
SECULIFEHITAM and SECULIFEHITMD
TRMS Medical-Multimeter
Characteristic Values
Meas.Function
V
Measuring Range
Resolution at Upper
Range Limit
11999
100 mV
1
V
10
V
100
V
1000
V
10
100
1
10
100
~/
1199
V
V
mV
mV
mV
Intrinsic Uncertainty under Reference Conditions
(... %rdg. + ... d)
(... %rdg. + ... d)
(... %rdg. + ... d)
Input Impedance
9
9
9
9
9
M
M
M
M
M
~ 10)
9 M // < 50 pF
9 M // < 50 pF
9 M // < 50 pF
9 M // < 50 pF
9 M // < 50 pF
0.09 + 5 mit ZERO
0.05 + 3
0.05 + 3
0.05 + 3
0.09 + 3
A

100
1
10
100
1
10
40
100
A 10 nA
mA 100 nA
mA
1 A
mA 10 A
A 100 A
A
1 mA

k
k
k
M
M
M

5.1 V 3)
F
Hz (V)
Hz (A)
Hz (V)
Hz (A)
10
100
1
10
100
1000
10 m
100 m
1 
10 
100 
1 k
10 k
—
—
1 mV
nF
nF
F
F
F
F
100.00 Hz
1.0000 kHz
10.000 kHz
100.00 kHz
30.00 kHz
MHz
100 Hz ... 1 MHz
%
2.0 ... 98 %
5.0 ... 95 %
10... 90 %
0.1 
10
100
1
10
100
1
0.01
0.1
1
10
pF
pF
nF
nF
nF
F
12 mV
120 mV
16 mV
160 mV
40 mV
600 mV
Open-circuit voltage
< 1.4
V
< 1.4
V
< 1.4
V
< 1.4
V
< 1.4
V
< 1.4
V
< 1.4
V
Approx. 8
V
12 mV
120 mV
16 mV
160 mV
40 mV
600 mV
Meas. curr. @ range limit
Approx. 8
Approx. 1 mA const.
V
Approx. 300 A
Approx. 250 A
Approx. 100 A
Approx. 12 A
Approx. 1.2 A
Approx. 125 nA
Approx. 20 nA
Approx. 1 mA const.
Discharge resist.
10 M
1 M
100 k
12 k
3 k
3 k
U0 max
0.7
0.7
0.7
0.7
0.7
0.7
fmin 5)
Hz
Hz
Hz
Hz
10 Hz
0.01
...100 Hz
—
—
—
V
V
V
V
V
V
~
10)
0.5 + 5
1.5 + 10 (> 200 d)
0.5 + 3
1.5 + 10 (> 200 d)
0.5 + 3
1.5 + 10 (> 200 d)
0.5 + 3
1.5 + 10 (> 200 d)
0.9 + 10
1.5 + 10 (> 200 d)
0.9 + 10
1.5 + 10 (> 200 d)
(... %rdg. + ... d)
0.2 + 5 with active ZERO function
0.2 + 5
0.2 + 5
0.2 + 5
0.2 + 5
0.5 + 10
2.0 + 10
3+5
1 + 30 (> 300 d) 1)
1 + 30 (> 300 d)
1 + 30 (> 300 d)
1 + 30 (> 300 d)
1 + 30 (> 300 d)
1.5 + 30 (> 200 d)
1.5 + 30 (> 200 d)
1.5 + 30 (> 200 d)
1.5 + 30 (> 200 d)
1.5 + 30 (> 200 d)
1.5 + 30 (> 200 d)
0.2 A
continous
10 A:  5 min 11)
16 A:  30 s 11)
1000 V
DC
AC
RMS
sine
Max. 10 s
1000 V
DC
AC
RMS
sine
Max. 10 s
Hz (V) 6):
10
Hz
10
Hz
1 ... 100
Hz
0.05 + 3
1
1
1
Hz
Hz
Hz
0.1 R
> 2 V ... 5 V
0.1 R per kHz
> 2 V ... 5 V
0.1 R per kHz
> 2 V ... 5 V
(... %rdg. + ... d)
– 200.0 ...
+850.0 C
– 150.0 ...
Pt1000 +850.0 C 0.1 C
K
– 250.0 ...
(NiCr-Ni) +1372.0 C
continous
0.5 + 3
Hz
100 Hz ... 1 kHz
... 10 kHz
...100 kHz
Time
6)
0.05 + 3 8)
0.01 %
0.01 %
0.01 %
Value
1000 V
DC
AC
RMS
sine
10)
(... %rdg. + ... d)
1 + 6 4) with active ZERO function
1 + 6 4)
1 + 6 4)
1 + 6 4)
5 + 6 4)
5 + 6 4)
(... %rdg. + ... d)
1
Pt100
C/F
10)
1)
1 + 30 (> 300 d)
0.5 + 9 (> 200 d)
0.5 + 9 (> 200 d)
0.5 + 9 (> 200 d)
0.5 + 9 (> 200 d)
Voltage drop, approx. at upper range limit
100
1
10
100
1
10
Overload Capacity 2)
1000 V
Max. 10 s
Hz (A): 7)
> 2 V ... 5 V
0.3 + 15 9)
0.3 + 15 9)
1% + 5 K 9)
1000 V
Max. 10 s
1000 V
DC/AC
RMS
sine
Max. 10 s
1)
Values of less than 200 digits are suppressed in the mV range.
15 (20) ... 45 ... 65 Hz ... 20 (1) kHz sinusoidal. See influence error on page 4.
At 0 ... + 40 C
Displays up to max. 5.1 V, “OL” in excess of 5.1 V.
Applies to measurements at film capacitors
Lowest measurable frequency for sinusoidal measuring signals symmetrical to the zero point
Overload capacity of the voltage measurement input:
power limiting: frequency x voltage max. 3 x 106 V x Hz for U > 100 V
7) Overload capacity of the current measurement input:
See current measuring ranges for maximum current values.
8)
Input sensitivity, sinusoidal signal, 10% to 100% of the measuring range
9) Plus sensor deviation
10)
Residual value deviates within 1 ... 30 d from the zero point due to TRMS converter
when probe tips are short-circuited
11) Off-time > 30 min and T  40 °C
A
2)
3)
4)
5)
6)
Internal Clock
Time format
DD.MM.YYYY hh:mm:ss
Resolution
0.1 s
Accuracy
1 min. per month
Temperature Influence 50 ppm/K
Key: R = measuring range, d= digit(s), rdg. = measured value (reading)
GMC-I Messtechnik GmbH
3
SECULIFEHITAM and SECULIFEHITMD
TRMS Medical-Multimeter
Influencing Quantities and Influence Error
Influencing
Quantity
Sphere of
Influence
Measured Quantity /
Measuring Range 1)
Influence Error
(...% rdg. + ... d) / 10 K
V
0.2 + 10
V
0.4 + 10
100  ... 1 M
0.5 + 10
> 1 M
–10 C ... +21 C
and
+25 C... +50 C
Temperature
1)
0.5 + 10
mA/A
0.8 + 10
10 nF ... 100 F
1+5
Hz
0.2 + 10
C/F (Pt100/Pt1000)
0.5 + 10
C/F thermocouple K
0.2 + 10
Meas. Qty. / Meas.
Range
> 15 Hz ... 45
VAC
Frequency
1.0000 V
...
100.00 V
1000.0 V 2)
AAC
100.00 A
...
10.0000 A
Intrinsic Uncertainty 3)
 ... % rdg. + ... d)
Sphere of Influence
Crest factor CF
Response Time (after manual range selection)
Measured Quantity /
Measuring Range
V
AV
,V
,A
Response Time
Digital Display
Measured Quantity
waveshape
1.5 s
From 0 to 80%
of upper range limit value
2s
10/40 M
5s
Continuity
< 50 ms
C (Pt 100)
Max. 3 s
Hz
3 + 30
1 kHz
2 + 30
> 1 kHz ... 10 kHz
3 + 30
10 nF ... 100 F
Max. 2 s
2+9
1 000 F
Max. 7 s
1+9
>10 Hz
1.5 s
> 65 Hz ...
> 15 Hz ... 45
> 65 Hz ...
Hz
1 kHz
> 1 kHz ...20
kHz
3+9
> 15 Hz ... 45
Hz
2+9
> 65 Hz ...
1 kHz
2+9
> 1 kHz ... 10 kHz
3 + 30
> 15 Hz ... 45
Hz
3 + 10
> 65 Hz ... 10 kHz
Power limiting: frequency x voltage max. 3 x 106 V x Hz for U > 100 V
The accuracy specification for frequency response is valid within a display value
range of 10% to 100% of the measuring range for both measuring modes with the
TRMS converter in the AC and (AC+DC) ranges.
Influencing
Quantity
+23 C 2 K
40 ... 75%
45 ... 65 Hz
Sine
3 V 0.1 V
Ambient temperature
Relative humidity
Measured qty. frequency
Measured qty. waveshape
Battery voltage
100  ... 1 M
100.00 mV
5)
1 + 10
mA/A
With zero balancing
Influencing
Qty.
2)
3)
Reference Conditions
Sphere of
Influence
Measured Quantity/
Measuring Range
1 ... 3
V
> 3 ... 5
Influence Error 5)
 1 % rdg.
,A
 3 % rdg.
From  to 50%
of upper range limit value
1.5 s
From 0 to 50%
of upper range limit value
Data Interface
Type
Data transmission
Protocol
Baud rate
Functions
Optical via infrared light through the housing
Serial, bidirectional (not IrDa compatible)
Device specific
38,400 baud
– Select/query measuring functions
and parameters
– Query momentary measurement data
– Read out stored measurement data
The USB X-TRA plug-in interface adapter (see accessories) is
used for adaptation to the PC’s USB port.
Except for sinusoidal waveshape
Influencing
Quantity
Sphere of
Influence
Measured Quantity
Influence Error
V A, , F, Hz, C
1 x intrinsic uncertainty
ditto
Included in intrinsic uncertainty
75%
Relative
humidity
3 days
Internal Measured Value Storage
Memory capacity
4 MBit / 540 kB for approx. 15,400
measured values with date and time stamp
instrument off
Battery voltage
Influencing
Quantity
1.8 to 3.6 V
Sphere of Influence
Measured Quantity /
Measuring Range
Interference quantity max. 1000 V
Common Mode
Interference
Voltage
Series Mode
Interference
Voltage
V
1V
Interference quantity max. 1000 V
50 Hz ... 60 Hz, sine
, 10 V
Damping
Power Supply
Battery
> 120 dB
> 80 dB
100 V
> 70 dB
1000 V
> 60 dB
Interference quantity: V , respective
nominal value of the measuring range,
max. 1000 V , 50 Hz ... 60 Hz, sine
V
> 50 dB
Interference quantity max. 1000 V
V
> 110 dB
Service life
Battery test
Power OFF function
4
2 ea. 1.5 V mignon cell (2 ea. size AA),
alkaline manganese per IEC LR6
(2 ea. 1.2 V NiMH rechargeable battery
also possible)
with alkaline manganese: approx. 200 hours
Battery capacity display with battery
symbol in 4 segments:
.
Querying of momentary battery voltage via
menu function.
Multimeter is switched off automatically:
– If battery voltage drops to below prox. 1.8 V
– If none of the keys or the rotary switch
are activated for an adjustable duration
of 10 to 59 minutes, and the multimeter
is not in the continuous operation mode
GMC-I Messtechnik GmbH
SECULIFEHITAM and SECULIFEHITMD
TRMS Medical-Multimeter
Power pack socket
If the NA X-TRA power pack has been
plugged into the instrument, the batteries
are disconnected automatically.
Rechargeable batteries can only be
recharged externally.
Display
Electrical Safety
Per IEC 61010-1:2010/DIN EN 61010-1:2011/VDE 0411-1:2011
Safety class
II
Measuring category III
IV
Operating voltage
600 V
300 V
Fouling factor
2
Test voltage
6.7 kV~
LCD panel (65 mm x 36 mm) with analog and digital display
including unit of measure, type of current and various special
functions
Electromagnetic Compatibility (EMC)
Background illumination
Background illumination is switched off approximately 1 minute
after it has been activated.
Interference emission EN 610326-1: 2006, class B
Interference immunity EN 610326-1: 2006
EN 610326-2-1: 2006
Analog
Display
Ambient Conditions
Scaling
Polarity display
Overflow display
Measuring rate
Digital
Display / char. height
Number of places
Overflow display
Polarity display
Measuring rate
Refresh rate
LCD scale with bar graph or pointer, depending on the selected parameter setting
With 4 division lines each, 1 bar/pointer corresponds to 500 digits at the digital display
With automatic switching
With the symbol
40 measurements per second and display
refresh
7-segment characters / 15 mm
4½ place 11,999 steps
“OL” is displayed for 12,000 digits
“–” (minus sign) is displayed
if plus pole is connected to “”
10 and 40 measurements per second with
the Min-Max function except for the
capacitance, frequency and keying ratio
measuring functions
2 times per sec., every 500 ms
Accuracy range
Operating temp. range
Storage temp. range
Relative humidity
Elevation
Deployment
Mechanical Design
Housing
Dimensions
Weight
Protection
IP XY
Protection against penetraIP XY
(1st digit X)
tion of solid particles
(2nd digit Y)
6
Dust-proof
5
Intermittent signal at above 1000 V
Intermittent signal at above 10 A
continuous signal at above 16 A
Fuse
FF (UR) 10 A/1000 V AC/DC;
10 mm x 38 mm,
Switching capacity: 30 kA at 1000 V AC/
DC, protects the current measurement
input in the 100 A through 10 A ranges
2
1
–
1
1
1
GMC-I Messtechnik GmbH
Protection against
penetration by water
Jet-water
Included
1
Fuse
Impact resistant plastic (ABS)
200 x 87 x 45 mm
(without protective rubber cover)
Approx. 0.35 kg with batteries
Housing: IP 65
Table excerpt regarding significance of the IP code
Acoustic Signals
For voltage
For current
0 C ... +40 C
10 C ... +50 C
25 C ... +70 C (without batteries)
Max.75%, no condensation allowed
To 2000 m
Indoors, except within specified ambient
conditions
multimeter
SECULIFE HITAM: 1 pair of safety measurement cables antimicrobial
(1.5 m) with 4 mm test probes, 600 V CAT III, 300 V CAT IV
(KS17-2AMB)
SECULIFE HITMD: 1 pair of safety measurement cables (1.5 m) with
4 mm test probes, 600 V CAT III, 300 V CAT IV (KS17-2)
batteries, 1.5 V, type AA
condensed operating instructions, English/German
Detailed operating instructions for download from our website
at www.seculife.eu
DAkkS calibration certificate
protective rubber cover
HC20 hard case
5
SECULIFEHITAM and SECULIFEHITMD
TRMS Medical-Multimeter
Accessories for Operation at a PC
Interface Adapter for USB Connection
The USB X-TRA bidirectional interface adapter includes the
following functions:
• Configure the SECULIFE HIT AM from a PC.
• Transmit live measurement data to the PC.
• Read out data from memory at the SECULIFE HIT AM.
The adapter does not require a separate power supply.
Its baud rate is 38,400 baud.
A CD ROM is included which contains current drivers for
Windows operating systems.
METRAwin10/METRAHit  Software
METRAwin10/METRAHit  PC software is a multilingual, measurement data logging program for recording, visualizing and
documenting measured values from SECULIFE HIT AM multimeters.
Communication between the PC and the measuring instrument(s)
is established via available interfaces and memory adapters. Telephone modems can be interconnected as well.
Depending upon device type, one or several of the following operating modes are possible:
• Device Configuration
Remote configuration and querying of device-specific functions
and parameters, for example measuring function, measuring
range and memory parameters. Frequently used device settings
can be saved to configuration files for easy recall.
• Online Recording of Measurement Data
Read-in, display and recording of momentarily measured data
from the interconnected device.
– Number of
measuring channels up to10
– Start recording
manual, triggered by measured value, time
triggered
– Recording mode
> time controlled
with sampling interval of 0.05 s* ... 1 s ...
60 min
> manually controlled
> measured value controlled in event
of exceeded limit/delta value
– Recording duration max. 10 million intervals
* Depending upon device type, measuring function, number of measuring channels and
communication (e.g. via modem), sample intervals of less than 1 s cannot be used.
•
Reading Out and Visualizing Stored Data
If supported by the device: read-in and display of offline data
recorded to device memory.
For purposes of analysis, data recorded online or read in from the
device’s memory can be displayed in various formats:
Y(t)-recorder display
XY-recorder display
for up to 6 channels
for up to 4 channels
Multimeter-display
for up to 4 channels
Tabular display
for up to 10 channels
System Requirements
METRAwin 10 (as from version 6.0) can be run on IBM compatible PCs with Microsoft Windows® VISTA, 7, 8 and 10.
6
GMC-I Messtechnik GmbH
SECULIFEHITAM and SECULIFEHITMD
TRMS Medical-Multimeter
Order Information
Designation
Transport Accessories
Type
Article Number
4½-place (12,000 digits) TRMS multimeter with direct, alternating and pulsating voltage
measurement (TRMS values), frequency measurement, resistance measurement,
continuity test, diode measurement and temperature measurement with type K thermocouples, precision temperature measurement with Pt100 or Pt1000 platinum resistance
thermometers, frequency and keying ratio measurement, broad range capacitance measurement,
LCD with 15 mm characters, analog bar graph and background illumination Measuring
categories: 300 V/CAT IV, 600 V/CAT III, with power pack socket and IR interface, 4 MB
data memory
Housing, protective rubber holster and
set of measuring cables with antimicrobial properties
SECULIFE HITAM
M687A
„Hospital“ TRMS multimeter for use in
hygienic sensitive areas,resistance
against desificationliquids
M687B
SECULIFE HITMD
HitBag Cordura Belt Pouch
For multimeters (with/without
protective rubber cover)
HC20 Hard Case
For multimeter (with/without
protective rubber cover)
and accessories
Accessories for operation at a PC
IR-USB bidirectional interface adapter
USB X-TRA
Z216C
METRAwin10 software
METRAwin10
GTZ3240000R0001
Accessories for temp. measurement with resistance thermometer
Pt100 temperature sensor for
surface and immersion measurement,
-40 to +600 C
Z3409
Pt1000 temperature sensor for
measurement in gases and liquids,
-50 to +220 C
TF220
Z102A
Pt100 oven sensor, -50 to +550 C
TF550
GTZ3408000R0001
FF (UR) 10 A /
1000 V AC/DC
Z109L
NA X-TRA
Z218G
GTZ3409000R0001
F836 Ever-Ready Case
For multimeter and accessories
Replacement fuse
Fuses (pack of 10)
Power pack
F829 Carrying Pouch
For multimeters (with/without
protective rubber cover)
and accessories
Designation
Type
Imitation leather carrying pouch
F829
Article Number
GTZ3301000R0003
Cordura belt pouch
HitBag
Z115A
Soft belt pouch large for one Multimeter. Made of rugged and water repellent
Cordura, three separate cases for leads,
clips, manual, CD, etc.
HitBag L
Z115B
Imitation leather ever-ready case with
cable compartment
F836
GTZ3302000R0001
Hard case for one Multimeter
and accessories
HC20
Z113A
Hard case for two Multimeter
and accessories
HC30
Z113A
For additional information regarding accessories please refer to:
•
•
GMC-I Messtechnik GmbH
our Measuring Instruments and Testers catalog.
our website www.gossenmetrawatt.com
7
SECULIFEHITAM and SECULIFEHITMD
TRMS Medical-Multimeter
Current Measuring Accessories
All current sensors and transformers are equipped with a connector cable (1.2 to 1.5 m long) with 4 mm safety banana plugs
Type
Designation
Measuring Range
Meas.
Transformation Frequency
Max.
Range
Category Wire Dia. Factor
DC/AC Current Sensors with Voltage Output
CP30
DC/AC clip-on current sensor, 5 mA to 30 A
300 V /
25 mm 100 mV/A
DC...20 kHz
with battery mode (30 h)
(DC / AC pk)
CAT III
(-3 dB)
300 V /
25 mm 10 mV/A;
DC...20 kHz
CP330
DC/AC clip-on current sensor, Range: 0.5 ... 30 A
(-3 dB)
with 2 measuring ranges,
Range: 5 ... 300 A
CAT III
1 mV/A
battery mode (50 h)
(DC / AC rms)
CP1100 DC/AC clip-on current sensor, Range: 0.5 ... 100 A
300 V /
32 mm 10 mV/A;
DC...20 kHz
(-1 dB)
with 2 measuring ranges,
Range: 5 ... 1000 A
CAT III
1 mV/A
battery mode (50 h)
(DC / AC rms)
300 V /
32 mm 10 mV/A,
DC ... 20 kHz
CP1800 DC/AC current clamp sensor, Range: 0.5 ... 125 A
with 2 measuring ranges,
Range: 5 ... 1250 A
CAT III
1 mV/A
(-1 dB)
battery mode (50 h)
(DC / AC rms)
AC Current Sensors with Voltage Output
WZ12B
AC clip-on current sensor
10 mA~ ... 100 A~
300 V /
15 mm 100 mV/A
45 ... 65
... 500 Hz
CAT III
WZ12C
AC clip-on current sensor,
1 mA~ ... 15 A~,
300 V /
15 mm 1 mV/mA,
45 ... 65
with 2 measuring ranges
1 ... 150 A~
CAT III
1 mV/A
... 400 Hz
WZ11B
AC clip-on current sensor,
0.5 ... 20 A~,
600 V /
20 mm 100 mV/A,
30...48 ... 65
... 500 Hz
with 2 measuring ranges
5 ... 200 A~
CAT III
10 mV/A
Z3512A
AC clip-on current sensor,
1 mA ... 1/10/100/
600 V /
52 mm 1 V/A, 100 mV/A, 10...48 ... 65
with 4 measuring ranges
1000 A~
CAT III
10 mV/A, 1 mV/A ... 3 kHz
METRAF- Flexible AC current sensor
0,5 ... 30 A,
1000 V
176 mm 100 mV/A,
10 Hz ... 20 kHz
LEX3000 with 3 measuring ranges,
0,5 ... 300 A,
CAT III
10 mV/A,
5 ... 3000 A
battery mode (2000 h)
600 V CAT IV
1 mV/A
METRAF- Flexible AC miniature current 1 ... 3 A,
1000 V
50 mm 1 V/A,
20 Hz ... 100 kHz
LEX300M sensor with 3 measuring
1 ... 30 A,
CAT III
100 mV/A,
ranges, battery mode (150 h) 5 ... 300 A
600 V CAT IV
10 mV/A
AC Current Transformer with Current Output
WZ12A
AC clip-on current transformer 15 ... 180 A~
300 V /
15 mm 1 mA/A
45 ... 65
CAT III
... 400 Hz
AC clip-on current transformer 30 mA ... 150 A~
300 V /
15 mm 1 mA/A
45 ... 65
WZ12D
CAT III
... 500 Hz
AC clip-on current transformer 1 ... 200 A~
600 V /
20 mm 1 mA/A
48 ... 65
WZ11A
CAT III
... 400 Hz
Z3511
AC clip-on current transformer 4 ... 500 A~
600 V /
30 x 63 1 mA/A
48 ... 65
CAT III
mm
... 1 kHz
Z3512
AC clip-on current transformer 0.5 ... 1000 A~
600 V /
52 mm 1 mA/A
30...48 ... 65
CAT III
... 5 kHz
Z3514
AC clip-on current transformer 1 ... 2000 A ~
600 V /
64 x 150 1 mA/A
30...48 ... 65
CAT III
mm
... 5 kHz
Shunt Resistors for Multimeters without Current Measuring Function
0 ... 300 mA
300 V /
—
1 mV/mA
DC ...10 kHz
NW300mA Plug-in shunt resistor,
CAT III
encapsulated 1 
NW3A
Plug-in shunt resistor,
0 ... 3 A
300 V /
—
100 mV/A
DC ...10 kHz
encapsulated 0,1 
CAT III
Edited in Germany • Subject to change without notice • PDF version available on the Internet
GMC-I Messtechnik GmbH
Südwestpark 15
90449 Nürnberg • Germany
Phone +49 911 8602-111
Fax
+49 911 8602-777
E-Mail info@seculife.eu
www.seculife.eu
Intrinsic Uncertainty Article
±(% rdg. + ...)
Number
1 % +2 mA
Z201B
1 % + 50 mA
1 % + 100 mA
Z202B
1 % + 100 mA
1 % + 500 mA
Z203B
1% + 100 mA
1% + 500 mA
Z204A
1.5% +0.1 mA
Z219B
3% + 0.15 mA,
2% + 0.1 A
1 ... 3%
Z219C
0.5 ... 3%,
0.2 ... 1%
1% + 0.1 A
1% + 0.1 A
1% + 1 A
1% + 0.2 A
1% + 0.2 A
1% + 1 A
Z225A
Z208B
Z207E
Z207M
3%
Z219A
2.5% +0.1 mA
Z219D
1 ... 3%
Z208A
3% +0.4 A
GTZ3511
000R0001
GTZ3512
000R0001
GTZ3514
000R0001
0.5% ... 0.7%
0.5% +0.1 A
0.5%
Z205C
0.5%
Z205B
Was this manual useful for you? yes no