Foxcroft FX-1500 toxic gas detector Instruction manual
Below you will find brief information for toxic gas detector FX-1500. This is a single channel toxic gas detector that can be located up to 2000 feet away from the hazard area. The FX-1500 is designed to give an early warning of developing toxic gas leaks or low oxygen levels by monitoring the air. When it senses toxic gas or low oxygen levels, it sounds its built-in horn and flashes its red blinking LED.
Advertisement
Advertisement
Foxcroft Equipment & Service, Co. Inc.
Model FX-1500
Single Channel Toxic Gas Detector
Instruction Manual
Warning!
Please Read Carefully and Save.
The FX-1500 Single Channel Toxic Gas Detector includes an instruction manual that contains important information about its operation. Purchasers who install this toxic gas detection system for use by others must leave this instruction manual or a copy with the user.
Copyright © 2012 FOXCROFT
Every effort has been made to ensure the accuracy of this document at the time of printing. In accordance with the company’s policy of continuous product improvement, Foxcroft reserves the right to make product changes without notice.
No liability is accepted for any consequential losses, injury or damage resulting from the use of this document or from any omissions or errors herein.
Table of Contents
Introduction
Specifications 8
INSTALLATION
1.1 Unpacking
9
9
1.2 Mechanical Installation 10
1.3 Layout
1.4 Typical Installation
11
12
14
1.6 Internal Wiring 25
3
4
2 CONTROL PANEL 28
30
SENSOR
31
4.2 Frequency of Testing
CALIBRATION
5.1 Calibration
32
32
35
5.3 Full Sensor Calibration 36
5.4 Oxygen Sensors 37
38
6 SENSOR
8 MONTHLY SENSOR TEST LOG
44
45
46
10 SERVICE 48
10.1 System Repair 48
2
IMPORTANT SAFETY INFORMATION
Please read and observe the following:
All functions of this gas detection system must be verified on a regular basis.
HELPFUL IDENTIFIERS
In addition to information on installation and operation, this instruction manual may contain WARNINGS pertaining to user safety, CAUTIONS regarding possible instrument malfunction, and NOTES on important, useful operating guidelines.
WARNING: A warning looks like this. It warns you of the potential for personal
injury
CAUTION: A caution looks like this. It alerts you to the possibility of instrument
malfunction or damage.
NOTE: A note looks like this. It alerts you to important operating information.
3
Introduction
General Information
Instrument Description and Capability
The FX-1500 Single Channel Toxic Gas Detector is an electronic system designed to rapidly detect and warn of hazardous toxic gas concentrations or low oxygen levels. The system consists of a toxic gas or oxygen sensor and a remote alarm unit that can be located up to 2000 feet away from the hazard area.
The FX-1500 is designed to give an early warning of developing toxic gas leaks or low oxygen levels by monitoring the air. When it senses toxic gas or low oxygen levels, it sounds its built-in horn and flashes its red blinking LED. The FX-1500 provides 3 alarm trip levels, with each level indicating specific concentration of toxic gas or percentage of oxygen. These levels are fixed and cannot be adjusted.
Put an FX-1500 in every place necessary to protect personnel. Place it in the proper location according to which type of gas the air is being monitored for.
Dimensions
6.000”
7.2139”
6.000”
7.2139”
Figure 1
4
Figure 2
FX-1500 Introduction
WARNING: If the alarm buzzer sounds a loud pulsating sound and you are
not testing the toxic gas detector, it has sensed a gas leak. The alarm buzzer is warning of a possible serious situation and it requires your immediate attention.
WARNING: All functions of the FX-1500 must be checked and verified on
a regular basis. Even the ability for the gas sensor to sense toxic gas or low oxygen levels must be verified on a regular basis. It is recommended that the gas sensor be tested by means of a certified gas standard at least once every six months.
WARNING: If the FX-1500 is altered in any way, the warranty will be void
(e.g., drilling additional holes in the enclosure to mount conduits, changing operating range of the gas sensor and altering the gas alarm trip points).
WARNING: Never disconnect an AC powered gas detector to stop nuisance
alarms. The source of the problem must be corrected.
5
FX-1500 Introduction
General Limitations
The FX-1500 Single Channel Toxic Gas Detector will not work without AC power. AC powered gas detectors will not work if the power source is cut off for any reason. Some examples are a power failure at the power station, an open fuse or circuit breaker, corrosion of the electrical system that destroys the wires, a failure of an electrical switching device.
If there are any concerns about the limitations of AC power, a battery backup should be installed.
The Green LED Digital Display should be clearly illuminated at all times. This indicates the AC power is turned on.
The FX-1500 will not sense toxic gas or oxygen if the gas which it is monitoring for does not reach the sensing chamber.
The FX-1500 may not sense toxic leaks or low oxygen levels in another room or another floor
of the building (e.g., an outside mounted gas detector may not sense a toxic gas leak inside a building; therefore, place a gas detector wherever it is needed to protect your personnel).
FX-1500 may not be heard or seen. The loudness of the alarm in the FX-1500 Toxic Gas Detector meets (or exceeds) current standards. However, if the FX-1500 is placed in a closed room, it may not be seen or heard by personnel, especially if the door does not have a window to see the visual alarm.
Even normal noise such as traffic, motors running, radios, and air conditioners may prevent personnel from hearing the audible alarm.
FX-1500 Toxic Gas Detectors may not be heard or seen by persons
who are hard of hearing or visually impaired. It is recommended that the remote warning device be installed to alert personnel before entering a room. Several detectors are recommended both outside the building or in any room that contain personnel, where toxic gas or low oxygen levels can develop.
The FX-1500 is not foolproof. Like all other electronic devices, FX-1500 Toxic Gas Detectors have limitations. Detectors cannot be expected to sense a dangerous toxic gas leak or low oxygen levels if the cavity is blocked by debris or the detector is carelessly tampered with by personnel.
The FX-1500 has a limited life. This equipment contains many parts. Just as with any other device any one of these parts could fail at any time. Therefore, you must test your detector on a regular basis.
Be sure to have it repaired or replaced when it fails to test properly. In no case should the detector be used for more than 10 years. All functions of this detector must be checked and verified on a regular basis. Even the ability of the sensor must be verified by means of a certified gas standard at least 2 times a year.
Life and Replacement: Most sensors are designed for a minimum life span of 2 years and are warranted for 1 year from date of shipment. Life expectancy can be as high as 10 years. Oxygen sensors have a life span of 2 years and are warranted for a period of 1 year from the date of shipment.
To verify expiration date, see serial number data tag or the data sheet provided.
6
FX-1500 Introduction
Range and Alarm Trip Point Chart
Standard Range (ppm or %)
Optional Ranges and Alarm Trip Points may be available.
Refer to the Serial Number label inside the remote unit for the type of gas being monitored, or the labeling on the sensor itself.
Gas
Chlorine (Cl
2
)
Sulfur Dioxide (SO
2
)
Carbon Monoxide (CO)
Hydrogen (H
2
)
Hydrogen Sulfide (H
2
Nitric Oxide (NO)
S)
Nitrogen Dioxide (NO
2
)
)
Oxygen (O
2
Ammonia (NH
3
)
Hydrogen Cyanide (HCn)
Ozone (O
3
)
Hydrogen Chloride (HCl)
Chlorine Dioxide (ClO
2
)
Standard
Range
0-10 ppm
0-10 ppm
0-500 ppm
0-200 ppm
0-10 ppm
0-50 ppm
0-10 ppm
0-25%
0-100 ppm
0-10 ppm
0-2 ppm
0-10 ppm
0-1 ppm
Caution Trip
Point
0.3 ppm
0.5 ppm
20 ppm
5.0 ppm
0.5 ppm
10 ppm
0.5 ppm
19.5%
25 ppm
0.5 ppm
0.03 ppm
0.5 ppm
0.05 ppm
Warning Trip
Point
0.5 ppm
1.0 ppm
35 ppm
35 ppm
1.0 ppm
15 ppm
1.0 ppm
18%
35 ppm
1.0 ppm
0.05 ppm
1.0 ppm
0.1 ppm
Danger Trip
Point
1.0 ppm
3.0 ppm
50 ppm
50 ppm
3.0 ppm
25 ppm
3.0 ppm
16%
50 ppm
3.0 ppm
0.1 ppm
3.0 ppm
0.3 ppm
7
Sensor input
Sensor signal type
Warranty
Range and alarm trip Points
Response time
Accuracy (ppm @ 20°C)
Resolution
Operating temperature range
Operating humidity
Drift
Operating pressure
Digital display type
Output signal
Auxiliary relay contacts
Configurable relay
Primary sensitivity
Audible alarm
Visual alarms
Local sensor
Power
Sensor distance
Enclosure
Specifications
One sensor
4-20 mA, 200 ohms
1 year (remote electronics only);
Sensor warranty varies depending on gas type - see individual toxic gas specifications sheets.
See page 7 for specific gas trip levels and ranges.
Less than 8 seconds to full alarm
0.1 percent
0.1ppm
-15°C to 40°C (5°F to 104°F)
15% to 90% (non-condensing)
Less than 2% per month
900 to 1100 mBar
Standard, Red LED 3 digits, 0.56” high
4-20 mA, 600 VDC (optional)
Problem - SPST
Normally open (NO), 1A @ 125 VAC non-latching relay
Relay 1 - SPDT, 10A @ 125 VAC
Relay 2 - SPDT, 10A @ 125 VAC
Direct PPM
Pulsating piezo buzzer (90dB)
Caution - yellow LED
Warning - yellow LED
Danger - red blinking LED
Problem - green blinking LED
Danger alarm - red blinking LED
110 or 220 VAC - 50/60 Hz @ 200 mA
25’ Std. (2000’ max)
Remote unit - NEMA 4X
Sensor unit - NEMA 4X
8
Installation
Unpacking
After unpacking, it is recommended to save the shipping carton and packing materials if the instrument must be stored or re-shipped. Inspect the equipment and packing materials for signs of shipping damage. If there is any evidence of damage, notify the transit carrier immediately.
The shipping container consists of the following:
1 FX-1500 Single Channel Toxic Gas Detector
1 Calibration Nose
1 Calibration Nose O-Ring
2 Calibration Nose Caps
1 Set of Four Mounting Feet with Four Screws
1 Instruction Manual
1 Calibration Manual
Serial Number
Should technical assistance be required, refer to the serial number to identify your system.
9
Installation (Mechanical)
Mechanical Installation
Sensor Location
Proper placement of the sensor is essential to the operation of the FX- 1500 Single Channel Toxic Gas
Detector. Depending on its location, the FX-1500 will either sound a rapid alarm of hazardous leaks or low oxygen levels, or may delay or fail to warn of hazardous conditions.
Correct installation of the sensor depends mostly on what type of gas the detector is monitoring the air for. If the gas is heavier than air, then the sensor should be placed closer to the floor. If the gas is lighter than air, the sensor should be placed further from the floor.
• Never mount the FX-1500 in direct sunlight or direct light.
• Do not install in very dusty or dirty areas. Dust and dirt can build up on the sensing chamber, making it overly sensitive. Dirt can block openings to the sensing chamber and keep the gas detector from sensing toxic gas or low oxygen levels.
• Do not install in areas of wash down. The toxic gas detector should not be washed with water or mounted in areas where water is used for cleaning.
• Do not install near fresh air vents, or very drafty areas. Take into account ventilation air flow patterns. If installing in an outdoor area, the sensor should generally be placed downwind of the source. More than one system may be needed to efficiently protect some outdoor locations.
• Do not install in insect-infested areas. If insects enter a sensing chamber, they may cause a nuisance alarm. They can also block the sensing chamber and prevent toxic gas or oxygen from entering the chamber.
• Outdoor areas may require more than one FX-1500 for protection.
Physical Installation
Installation of the FX-1500 consists of physically mounting the remote alarm unit and sensor, and wiring in AC power, sensor connections, and output alarm contacts.
Remote Alarm Unit
Attach the four mounting feet to the rear of the remote enclosure. Anchor securely to a wall. Be sure the remote alarm unit is mounted in a location that can be seen clearly by all personnel. The remote alarm unit must be mounted outside the hazard area.
Detector Mounting
Physical installation of the gas detector simply requires mounting the detector to a wall.
10
FX-1500 Installation (Mechanical)
Figure 3
Gas Sensor Opening
WARNING: The toxic gas detector cannot efficiently monitor the air if the
sensing chamber is blocked in any way.
Layout
Mounting Holes (4)
Gas Sensor
Enclosure
Remote Alarm
Unit
Figure 4
NOTES: It is recommended that the remote alarm be mounted outside the hazard area.
Always mount the sensor chamber facing down towards the floor.
11
Installation (Typical)
Typical Installation
Physical Installation
This drawing indicates recommended locations for minimum security and a higher degree of security.
The FX-1500 Single Channel Toxic Gas Detector will not sense a toxic gas leak or oxygen if the gas does not reach the sensing chamber.
12
Figure 5
FX-1500
Example: Chlorine gas ONLY
Installation (Typical)
13
Figure 6
Installation (Electrical)
Wiring
The FX-1500 Single Channel Toxic Gas Detector should be wired to its own AC power source which is uninterrupted. A battery backup is recommended where AC power may be interrupted.
The FX-1500 will not work without AC power.
Use 14 gauge stranded wire only (maximum).
All electrical connections are made to a terminal strip located on the power supply card. The power supply card is located in the remote unit. To gain access to the terminal strip open the door and remove the yellow control panel by unscrewing 4 thumb screws. Next remove the 4 short standoffs then remove the upper amplifier card. Be careful not to pull the amplifier card out of the enclosure too far. The cards are attached together by a ribbon cable. Set the amplifier card on top of the enclosure.
The ribbon cable may also be disconnected from the amplifier card. Be careful not to bend the pins on the connector. Four conduit holes are provided. The power supply is wired to the conduit hole on the far back right. The unit is equipped to operate on 110VAC or 220VAC, 50/60 Hz. The other conduit holes are provided for relay outputs. The danger relays are non-configurable. Relay 1 and Relay 2 are configurable. Factory set of these relays is Relay 1, (Caution Trip Alarm) and Relay 2, (Warning Trip
Alarm). Refer to Configuration Chart on page 18.
The sensor wires come attached to the sensor input terminals. The sensor wires may be disconnected from the sensor input terminals only. Do not disconnect the sensor wires from the Sensor Unit Enclosure.
The sensor cable may be shortened or installed in a conduit. It is recommended the sensor cable be installed in a metal conduit where RF noise is present. Do not splice the sensor cable because a continuous length of cable must be installed. The metal conduit must be secured to an earth ground however, do not earth ground the sensor unless instructed by the factory.
Use a 4 conductor, 22 gauge stranded wire only for sensor connections.
14
FX-1500
Installation (Electrical)
TB-9 AC Power input terminal is shown below. Maximum 14 gauge stranded wire only. Do not use solid wire. Always connect to an earth ground.
15
Figure 7
FX-1500 Installation (Electrical)
Remote Reset
A remote reset switch terminal (TB-8) is provided. This permits the Gas Detector to be reset from a remote location. A normally closed switch must be used. The normally closed switch is not provided and is optional. First, remove the jumper from the terminal and then attach the normally closed switch.
Only remove the jumper when connecting a remote normally closed (NC) switch.
Finally, replace the amplifier card and standoffs and secure the control panel. If the ribbon cable was removed, be very careful when reconnecting, be sure pins are lined up properly and polarity is observed.
16
Figure 8
FX-1500 Installation (Electrical)
Alarm relay terminals shown below. Located on bottom circuit board, power supply card.
Figure 9
Danger Alarm Relays
Two (2) latching relays are provided for the Danger Trip Point. These relays have a maximum rating of
10A @ 125VAC. They are non- configurable and must be manually reset either by pressing the reset button or remotely.
Do Not Over Load Relays
10 amp Maximum, @ 125 VAC
Circuit Board Damage will occur if relays are over loaded.
Auxiliary Relays
Two (2) auxiliary relays are provided, Relay 1 and Relay 2. They are both normally open (NO) with a maximum rating of 10A @ 125 VAC. DO NOT OVERLOAD RELAYS. Each relay is user configurable to one of any four combinations. To configure the relays simply move the blue relay 1 jumper and the red relay 2 jumper to the selection required.
The selections are as follows:
1. Danger - 2 points provided (latching relay, this option allows two additional latching relays or 4 total danger relays. The alarm trip points for these relays are factory set. To unlatch these relays, they must be manually reset by either locally pressing the reset button or by a user remote reset button).
17
FX-1500 Installation (Electrical)
2. Danger Auto/Reset - 2 points provided (non-latching relay, this option allows the relay to trip at
the danger alarm point. When the concentration of toxic gas falls below the alarm trip point, the
relay will automatically reset).
3.
Warning - 2 points provided (non-latching relay, this option allows the relay to trip at the
warning point. When the concentration of toxic gas falls below the alarm trip point, the relay will
automatically reset).
4. Caution - 2 points provided (non-latching relay, this option allows the relay to trip at the caution
level. When the concentration of toxic gas falls below the trip level, the relay will automatically
reset).
Refer to page 7 for alarm trip points and gas ranges.
Auxiliary Relay Configuration Chart
Before configuring the relays, the AC power source of the FX-1500 Single Channel Toxic Gas Detector
MUST be turned off. The relay configuration is located on the power supply card just above the relays themselves. To configure simply move the blue relay 1 and or the red relay 2 jumpers to any of the four selections. Be sure the plug is secured properly and the selection is what is required. After configuration is complete, the AC power may be turned back on.
Factory configuration:
Relay 1 (blue jumper) - Caution Alarm
Relay 2 (red jumper) -Warning Alarm
See page 20 for additional information
18
Figure 10
FX-1500 Installation (Electrical)
Below is shown the auxiliary configuration area for relay 1 and relay 2. Relay 1 blue jumper and
Relay 2 red jumper. Select one configuration for each relay.
Factory relay configuration:
Figure 11
WARNING: Do Not Over Load Relays; 10 amp Maximum, @ 125 VAC.
Circuit Board Damage will occur if relays are over loaded.
Important: The Danger Alarm Relays associated with the “Danger Alarm” LED is normally a latching
relay. When the alarm is energized, the relay will hold in until it is manually reset using the “Reset” button on the front panel. However, relay 1 and or relay 2 can be configured to a non- latching relay by selecting the auto/reset option.
Alarm trip points are factory set and calibrated. Warranty is automatically voided if any alarm trip points are adjusted.
WARNING: Never create a toxic gas leak to test and verify the function of a
toxic gas detector. A certified method of testing is recommended.
19
FX-1500
Relay Schematic
Installation (Electrical)
TB2 Problem Relay non-latching, 1 amp rating @125VAC
TB3 Danger Alarm Relay latching, 10 amp rating @ 125VAC
TB4 Danger Alarm Relay latching, 10 amp rating @ 125VAC
TB5 Relay 1 configurable, 10 amp rating @125VAC
TB6 Relay 2 configurable, 10 amp rating @125VAC
Figure 12
20
FX-1500
Battery Backup System
Installation (Electrical)
21
Figure 13
FX-1500
4-20 mA Retransmitter Card (optional)
Installation (Electrical)
Figure 14
Important: Be careful not to connect the ribbon cable backwards. Follow all polarity markings on
cable and printed circuit board. Damage will occur if ribbon cable is connected backwards.
WARNING: Do not connect any external device to the 4-20mA sensor signal
loop. Connecting an external device will prevent the monitor from alarming.
Typical Audible Horn and Visual Lamp Wiring
22
Figure 15
FX-1500
Typical Exhaust Fan Wiring
Installation (Electrical)
Do Not Over Load Relays
10 amp Maximum, @ 125 VAC
Figure 16
23
FX-1500
Typical Telephone Dialer Wiring
Installation (Electrical)
Figure 17
The drawing demonstrates a typical wiring of the FX-1104 or FX-1108 Telephone Dialing System.
For more information refer to the FX- 1104 and FX-1108 manuals.
Important: Configure Relay 2 to Auto/Reset Selection
24
Installation (Internal Wiring)
Sensor Cable
Only use style number 2576 if sensor cable requires lengthening. It is recommended that whenever possible the proper length of cable be factory installed.
Standard 25 feet of cable, up to 2000 feet maximum!
Figure 18
25
FX-1500
Sensor Input Wiring
Installation (Internal Wiring)
Figure 19
WARNING: Do not connect any external device to the 4-20mA sensor
signal loop. Connecting an external device will prevent the monitor from alarming.
26
FX-1500
Audible Alarm
Installation (Internal Wiring)
Figure 20
Never connect an audible alarm to the piezo buzzer output terminal. This connection is only for the factory piezo buzzer provided.
27
Control Panel
Remote Alarm Unit
1.
Digital Readout Direct PPM
Red LED Displays concentration of toxic gas in parts per million (ppm) or % of oxygen. Standard,
3 digits, .56” high.
2.
Caution Alarm
Yellow LED
Please refer to page 7 for caution alarm trip point.
3.
Warning Alarm
Yellow LED
Please refer to page 7 for warning alarm trip point.
4.
Danger Alarm
Red Blinking LED
Please refer to page 7 for danger alarm trip point.
5.
Problem Alarm
Green Blinking LED
Indicates a possible problem with the sensor signal. The cell failure indicating LED is not foolproof.
Like all other electronic devices, it contains many parts. Any of these parts could fail at any time.
Therefore, you must test your detector on a regular basis.
6.
Silence Button
Silences piezo buzzer. (Not Shown)
Piezo Buzzer Audible alarm which sounds at 90 db. when toxic gas concentration reaches danger trip level.
7.
Reset Button
Manually resets the danger alarm.
Sensor Enclosure
8.
Toxic Gas Sensor Chamber
9. Remote Danger
Alarm Red Blinking LED
Provides local sensor danger trip point indication. Indicating light only, no audible alarm provided with sensor enclosure.
10. Electronic Test Button
Tests all electronic functions. Does not check the ability of the sensor to sense toxic gas or oxygen levels.
28
FX-1500
Gas Sensing
Chamber
Electronic Test Button
Remote Danger Alarm
Figure 21
Control Panel
29
Figure 22
Startup
CAUTION: Remove the protective plug nose from the sensor cavity. This plug
is clearly marked with a sticker. “Remove protective plug before operating”.
Store protective nose in a safe place. It will be required for verification and calibration of the sensor.
Operations
1. Apply AC power to the FX-1500 Single Channel Toxic Gas Detector.
2. The Red LED Digital Display will illuminate and may start out with a negative number. The digital display within 5 minutes should settle down and indicate 00.0 ppm.
3. The “problem” LED indicator will blink when you first apply AC power. It will go out in approximately 5 minutes.
4. The Danger Alarm may alarm during startup, press the reset button to reset alarm.
5. Wait 5 minutes for the monitor to warm up.
6. Press the “Test” button located on the sensor enclosure and hold 10 seconds.
7. All alarms should illuminate and the piezo buzzer should sound.
8. Silence the piezo buzzer by pressing the “Silence” button.
9. After 2 minutes press the “Reset” button.
10. Never operate the detector if any or all of the alarm trip points fail to illuminate the LED’s or sound the piezo buzzer
11. Allow 24 hours for full sensor stabilization before placing detector into operation.
12. System is now ready for operation.
30
Sensor Test System
Sensor Test System
While the electronic “TEST” will only test all electronic functions, it is important that the sensor be tested periodically to ensure its gas sensitivity. A certified method is recommended.
Important: It is very important to maintain Weekly and Monthly Test Logs. Regular testing ensures the
proper operation and life of the detector.
Frequency of testing your Toxic Gas Detector
Bump Test
The Toxic Gas Sensor is designed to be used in safety critical applications. To ensure that the sensor and/ or instrument in which it is used, are operating properly, it is a requirement that the function of the device is confirmed by exposure to target gas DAILY. Failure to carry out such tests may jeopardize the safety of people and property.
Calibration Check
The Toxic Gas Sensor is designed to be used in safety critical applications. To ensure that the sensor and/ or instrument in which it is used, are operating properly, it is a requirement that the function of the device is confirmed by exposure to target gas MONTHLY. Failure to carry out such tests may jeopardize the safety of people and property.
Full Calibration
The Toxic Gas Sensor is designed to be used in safety critical applications. To ensure that the sensor and/ or instrument in which it is used, are operating properly, it is a requirement that the function of the device is confirmed by exposure to target gas EVERY SIX MONTHS. Failure to carry out such tests may jeopardize the safety of people and property.
Definitions
a. Bump Test (Function Check) - A qualitative function check where a challenge gas is passed over the sensor(s) at a concentration and exposure time sufficient to activate all alarm indicators to present at least their lower alarm setting. The purpose of this check is to confirm that gas can get to the sensor(s) and that all the alarms present are functional. This is typically dependent on the response time of the sensor(s) or a minimum level of response achieved, such as 80% of gas concentration applied. Note this check is not intended to provide a measure of calibration accuracy.
b. Calibration Check - A quantitative test utilizing a known traceable concentration of test gas to demonstrate that the sensor(s) and alarms respond to the gas within manufacturer’s acceptable limits. This is typically ±10-20% of the test gas concentration applied unless otherwise specified by the manufacturer, internal company policy, or a regulatory agency.
c. Full calibration – The adjustment of the sensor(s) response to match the desired value compared to a known traceable concentration of test gas. This should be done in accordance with the manufacturer’s instructions.
31
Calibration
The FX-1500 Single Channel Toxic Gas Detector has been factory calibrated prior to shipment. If you wish to check its calibration, a low toxic gas standard is required. The best method for this is to use a certified calibration mixture.
It is recommended that the FX-1500 be verified by the use of a low level toxic gas standard at least once every six months.
Important: It is very important to maintain Weekly and Monthly Test Logs. Regular testing ensures the
proper operation and life of the detector.
All functions of this gas detection system must be checked and verified on a regular basis.
WARNING: Do not attempt to calibrate any trimmers in the Remote Alarm
Unit. These trimmers are factory set.
Important: All trimmers required in the calibration procedures are in the Gas Sensor Enclosure.
32
FX-1500
Calibration
Important: To calibrate the gas sensor a certified target calibration kit is required. The Calibration Gas
Kit includes one target gas cylinder, regulator, tubing and certification sheet. A separate calibration kit is required for each type of gas you monitor.
Tools required:
• Jeweler’s Screwdrivers (full calibration only)
• Calibration nose with O-Ring and caps (Provided with all gas sensors)
• Calibration Gas Kit (pictured below)
• Digital Voltmeter (full calibration only, optional)
Tubing
Regulator
Gas Cylinder
Calibration Nose
Figure 23
33
FX-1500 Calibration
WARNING: Toxic gas sensors are supplied precalibrated, and the sensitivity
of the sensor should not drift by more than 2% of full signal per month.
The full calibration procedure is not a requirement for normal operation of the
FX-1500 Single Channel Toxic Gas Detector. Full calibration is only necessary when it has been determined that there is a problem with the sensor, and only after the problem has been repaired, or the sensor has been replaced.
However, full gas calibration is a good safety procedure, and is recommended as such, in this instruction manual, for the gas detector. If your on-site safety procedures require a gas detector test (which they should), we recommend a short gas leak test rather than a full calibration. There are two reasons for this:
1. The gas sensor is designed for ambient air conditions, and can eventually be “poisoned” by repeated overexposure to the test and target gases. This is usually noted as a gradual elevation in the zero reading over time and exposure to the test and target gases.
2. If the calibration is done improperly, or with old test gas (it has a shelf life listed on the cylinder), it could seriously impair the sensitivity and function of the gas detector.
WARNING: Target calibration gas contains toxic gas and should only be used
by personnel who are fully trained in gas calibration procedures. All safety procedures and warnings on the gas cylinder should be read, understood, and observed.
34
FX-1500 Calibration
Figure 24
Short Gas Leak Test
Important: This test is not a calibration; it is only to test the gas sensor response to a gas leak. Short
gas leak tests can only be done with a known target concentration gas.
It is
very important that this test sets off all the alarms and the target concentration gas level is
displayed on the LED display.
Full calibration is required if the test fails to set off alarms and the target concentration gas level is not displayed on the LED display.
1. The Remote alarm unit must be at zero on the LED display.
2. Install the calibration nose in the collar plug; be sure the nose is sealed properly and is isolated from the atmosphere.
3. Connect the target calibration gas sample tubing to the nose.
4. Apply target calibration gas with a known concentration.
5. Wait for a stable reading to be obtained (five minutes is a suitable period).
6. The known target gas concentration level should be indicated on the LED display and all the alarms should be set off including the buzzer.
7. If the target calibration gas level is indicated on the LED display and the alarms are set off including the buzzer, the test is complete. Remove calibration nose, LED display should zero in a few minutes. Reset danger alarm.
8. If the target calibration gas level is not indicated on the LED display and or fails to set off all alarms a full calibration is required.
35
FX-1500 Calibration
Full Sensor Calibration
Each sensor is supplied precalibrated, and the sensitivity of the sensor should not drift by more then
2% of full signal per month. The oxygen sensor has a single trimmer for span adjustment (see page
37) and has a separate calibration routine.
Important: Full sensor calibration is only as accurate as the digital millivolt meter being used (or the
FX-1500 Single Channel Toxic Gas Detector LED display), and the freshness of the target gas being used. So it is important to use an accurate digital meter and a fresh target gas cylinder. The FX-1500 remote alarm unit already has a factory precalibration and calibration should not be required. If you have reason to believe that the display or alarm level calibration of the FX-1500 remote alarm unit is off, we recommend that you contact Foxcroft, to arrange an “RMA”, to return the entire unit with the sensor, for a complete calibration. Foxcroft maintains a serial numbered calibration database for all of our gas detectors. Foxcroft does not recommend electronic calibration of the remote unit in the field.
1. The gas sensor enclosure mounting system allows easy zeroing and aspiration using the calibration nose provided (see page 43). With the plug in place and sealing caps fitted, the sensor is completely isolated from the atmosphere. Removing the cap seals allows the calibration target gas to be connected to one of the inlets and exhausted through the other. A flow of 500 milliliter/ minute is recommended.
2. Ensure the sensor is free from the target gas being measured either by purging the sensor with an inert gas, or blanking the sensor from the atmosphere with the calibration nose supplied.
3. Apply a target calibration gas of a known concentration to the sensor.
4. Wait for a stable reading to be obtained (five minutes is a suitable period of time).
Important: It is very important to maintain weekly and monthly test logs (see pages 45-47). Regular
testing ensures the proper operation and life of the detector.
36
FX-1500 Calibration
Oxygen Sensors
Calibration should be carried out in ambient air, and is done simply by adjusting the span potentiometer until a reading of 20.9% on a scale of 0-25% is displayed on the LED display.
Figure 25
37
FX Easy Calibration
Performance Characteristics (Chlorine only)
Sensor Type Used
Expected Operating Life
Resolution
Temperature Range
Pressure Range
Pressure Coefficient
T
90
Response Time
Relative Humidity Range
Maximum Zero Shift (20°C - 40°C)
Long Term Output Drift
Repeatability
Output Linearity
3CLH
Two years in air
0.1ppm
-20°C to +50°C
Atmospheric +/- 10%
No data
≤ 60 seconds
15-90% non-condensing
-0.2ppm equivalent
< 2% signal loss/year
2% of signal
Linear
N.B. All performance data is based on conditions at 20C, 50%RH and 1013mBAr.
Electrical Properties (Chlorine only)
Minimum Input Span
Maximum Input Span
Linearity @ 25°C
Thermal Drift (Input)
Linearisation
Output
Maximum Output Range
Operating Voltage
Thermal Drift (Output)
Output Accuracy
Protection
Maximum Loop Load
5uA
+/-100uA
0.01% Full Scale
0.02% per °C
Linear Response
4-20mA, 2 wire loop powered
3.8 to 21mA
10 to 30VDC
2uA per °C
+/- 5uA
Reverse Polarity Protected
R
LOAD
= (V supply
-10)*50 e.g. 700Ω at 24V
Physical Characteristics (Chlorine only)
Weight
Position Sensitivity
Storage Life
Recommended Storage Temperature
Linearisation
Warranty Period
58g (including mounting accessory)
None
Six months in CTL Container
0-20°C
Linear Response
12 months from date of despatch
38
FX-1500 FX Easy Calibration
Calibration Notes
Recalibration is only possible if the output of the sensor at full scale is greater than 50% of the original factory calibration. Failure can occur if:
1. Attempts to recalibrate to a range less than 50% of the original factory calibrated range
2. The output of the sensor has fallen by more than 50% (replace sensor)
3. Incorrect span gas used (use correct span gas)
4. Insufficient time for output to settle after exposing sensor to span gas (apply span gas for 2-5 mins before setting 20mA level)
Transmitter Error Condition
Any time the error condition is set the output will be forced to 21mA and the LED will be held on. Carry out the Reset function to reset the error.
Reset to Factory Calibration
1. Remove loop power from transmitter.
2. Hold down push button and connect loop power to transmitter.
3. LED will blink as soon as power is applied (approx. 2Hz).
4. Transmitter output set to 21mA.
5. Releasing button will start an 8 second timout period.
6. After 8 seconds the factory calibration will overwrite user calibration.
7. LED cleares and transmitter goes back into normal mode.
User Calibration Method
(Span gas = 20mA signal)
1. Connect loop power to the transmitter.
2. Apply required zero gas/air to transmitter for 2 minutes.
3. Hold down button until LED flashes at approx. 1Hz. (4mA level now set)
4. Apply required span gas to transmitter for 2 to 5 minutes.
5. Press and release button (20mA level now set)
6. There will be an 8-second timeout period and flashing LED approx. 8Hz.
7. Calibration is complete, LED clears and returns to normal mode.
39
FX-1500 FX Easy Calibration
Non Standard Span Gas Calibration (Span gas ±5% full range)
1. Calculate the mA signal expected for span gas.
Example:
1. Required 4-20mA range is 0-500ppm
2. Available span gas is 480ppm
3. Dynamic range is 16mA
4. Therefore 1mA = 31.25ppm
5. Expected signal at 480ppm = (15.36mA + 4mA) = 19.36mA
6. Each time button is pressed signal increases by 40uA. When the output reaches 20.8mA the next button press will take the output to 19.2mA.
Further button presses will again increase the output by 40uA.
7. The output range available is 19.2mA to 20.8mA.
2. Carry out steps 1-4 of User Calibration Method
3. Press and release button to set 20mA level.
4. Continued momentary pressing of the button within an eight second period will trim the output by
40uA. When the required output is reached, allow the transmitter to timeout.
5. Calibration is complete, LED clears and returns to normal mode.
Non Standard Span Gas Calibration (Span gas not within ±5% of full range; current source required)
1. Measure the current of the sensor with a known gas concentration. This can then be used to calculate the µA/ppm for the specific sensor.
2. Now calculate the expected current when the sensor is exposed to the full scale of target gas.
Example:
1. Required 4-20mA range is 0-500ppm
2. Available span gas is 300ppm
3. Current from sensor when exposed to span gas = 33uA
4. Therefore sensitivity = 0.11 uA/ppm
5. Expected sensor output at 500ppm = 55uA
3. Connect current loop power to the transmitter.
4. Connect a current source to the transmitter. For oxidizing sensors (CO, H
2
S, SO
2 connect the negative to ‘SEN’ and positive to ‘CNT’. For reducing sensors (Cl
2
or NO
2
, NO, ...)
) connect the positive to ‘SEN’ and negative to ‘CNT’. ‘CNT’ and ‘REF’ terminals on the transmitter board should be connected.
5. Set the current source to zero.
6. Hold down button until LED flashes at approx. 1Hz.
(4mA level now set in RAM)
7. Set the current source to the value calculated for the full scale of target gas.
8. Press and release button
(20mA level now set into RAM)
9. There will be an 8-second timeout period and flashing LED approx. 8Hz.
10. Calibration is complete, LED clears and returns to normal mode.
40
FX-1500
Chlorine - Output vs Temperature
FX Easy Calibration
Figure 26
41
Gas Sensor Replacement
WARNING: Even the ability of the gas sensor to sense toxic gas or low oxygen
levels must be verified on a regular basis. It is recommended that the gas sensor be tested by means of a certified gas standard at least every six months.
WARNING: All functions of this FX-1500 Gas Detector must be checked and
verified on a regular basis.
Gas sensor replacement in the FX-1500 Single Channel Toxic Gas Detector is easily done with minimal tools.
Tools required:
• Small Jeweler’s Screwdriver
• General Small Screwdriver
• Needlenose Pliers
1. Disconnect the AC power from the detector.
2. Remove the sensor enclosure cover by removing (4) screws.
3. Remove the front cover of the sensor enclosure.
4. Carefully pull off the orange wire connector.
5. Unscrew the bulk head nut (see page 43 for diagram). Place the tips of a pair of needlenose pliers in the two holes of the sensor bulk head nut. Turn the ring counterclockwise to loosen the ring, and remove the entire sensor from the enclosure.
6. The transmitter PC board is mounted to the sensor with disconnect pins. Hold the PC board by the edges and carefully pull the board off of the sensor. Some very early sensors, pre- ”1994,” had the PC board soldered directly to the sensor. This type is not removable. If you have this type, stop here. Do not attempt to solder the new sensor to the transmitter board. The heat from the soldering process will damage the new sensor. You will need to replace the entire sensor/ transmitter assembly. All new sensors have removable boards, so future sensor replacements will not be hindered by the additional cost of the transmitter board.
9. Remove the shorting wire from the gold pins on the new sensor, and carefully line up the 3 gold pins on the new sensor with the 3 sockets on the bottom of the transmitter board. Press evenly on the board until it fully seats on the gold pins.
10. Installation of the completed sensor/transmitter assembly is the reverse of its removal.
11. After completing the sensor replacement procedure, power-up the gas detector and allow it to stabilize for 20 minutes (some sensor gas types take up to 24 hours to stabilize). The LED display, should show 0.0 ppm +/-0.1 ppm. If the display is off by more than 0.2 ppm, or the cell failure/problem light is blinking, the sensor requires a full sensor calibration.
12. Refer to page 35 and follow the short gas leak test.
42
FX-1500
Cap
Seals
Calibration Nose
Plug O-ring
Bulkhead
Nut
Collar Plug
O-ring
TFE Mesh
Membrane
O-ring
Gas Sensor Replacement
Gas Sensor Transmitter
Figure 27
43
Figure 28
Parts Listing
Electronics
7
8
9
10
5
6
3
4
Item Number
1
2
Description
Gas Sensor Only
Gas Sensor Transmitter Only
Gas Sensor with Transmitter
Amplifier Card
Power Supply Card
Ribbon Cable
Sensor Cable (25’ standard)
Sensor Cable (specify length)
4-20mA Retransmitter Kit
Normally Closed Switch Kit
Part Number
Specify gas type
Specify gas type
Specify gas type
FX-CL21001A
FX-CL21001P
FX-9431323-07
FX-2W979-25
FX-2W979
4-20 MA XMT
FX-2751548
Calibration Gas Kits
Remote Alarm
Figures 29-31
Telephone Dialing Systems
44
Date
Month of
Month of
Month of
Month of
Month of
Month of
Month of
Month of
Monthly Sensor Test Log
Initials Date
Month of
Month of
Month of
Month of
Month of
Month of
Month of
Month of
Initials
45
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Date
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Weekly Electronic Test Log
Initials Initials
46
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Date
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Date
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Weekly Electronic Test Log
Initials Initials
47
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Date
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Week of
Service
System Repair
Customer Service Department
If you need spare parts, assistance in troubleshooting, or repair service, please contact Foxcroft
Customer Service at:
Foxcroft Equipment and Service, Co. Inc.
2101 Creek Road, P.O. Box 39
Glenmoore, PA 19343
Tel: (800) 874-0590
Fax:
(610) 942-2888
(610) 942-2769
Email: [email protected]
www.foxcroft.com
Customer Repair / Returns Policy
All systems returned for repair or replacement must be freight prepaid and include the following information:
1. A clearly written description of the malfunction.
2. Name of person to contact and the phone number where they can be reached.
3. Proper return address for shipping system back. Include preferred shipping method.
4. A purchase order if the system is out of warranty to cover costs of repair.
5. A Return Material Authorization Number (RMA) is required before shipping any products for service. Call telephone number above to receive a RMA number.
NOTE: Returns will only be held at Foxcroft for 90 days. If a decision is not made regarding the
repair, the product will be returned.
48
Product Warranty
Foxcroft Equipment & Service warrants all products obtained hereunder to be free from defects in material and workmanship for a period of one year from the date of shipment. In the event of a product failure or defect requiring warranty repair, the customer must obtain an RMA number by calling 1-800-
874-0590, before returning the product, at the customer’s expense to Foxcroft for repair. Warrantor (Foxcroft Equipment and Service) will repair the unit, without charges for parts, labor and return freight.
Foxcroft Equipment & Service is not responsible for damage to its products through improper installation, maintenance, act of God, use or attempts to operate such products beyond their functional capacity, intentionally or otherwise, or for any unauthorized repair.
Buyer agrees to hold Foxcroft Equipment & Service harmless from all claims for damages arising out of injury or death to any person or damage to any facility, or any other property, or loss of use of any such property, whether such person or property is on or off the installation or activity site for which the equipment or material furnished hereunder is destined and whether such damage, loss destruction or loss of use, injury or death results directly or indirectly from a nuclear incident or for any other cause.
Statements and instructions set forth herein are based upon the best information and practices known to Foxcroft Equipment & Service but it should be assumed that every acceptable safety procedure is contained herein. Of necessity this company cannot guarantee that actions in accordance with such statements and instructions will result in the complete elimination of hazards and it assumes no liability for accidents that may occur.
49
Serial Number Label of FX-1500 Single Channel Toxic Gas Detector:
Foxcroft Equipment and Service, Co. Inc.
2101 Creek Road, P.O. Box 39
Glenmoore, PA 19343
Tel: (800) 874-0590
Fax:
(610) 942-2888
(610) 942-2769
Email: [email protected]
www.foxcroft.com
Advertisement
Key Features
- Remote alarm unit
- 3 alarm trip levels
- Sensor test system
- Calibration
- Gas sensor replacement
- Battery backup system
- 4-20 mA retransmitter card