True-rms Remote Display Digital Multimeter
Users Manual
September 2009
© 2009 Fluke Corporation. All rights reserved. Specifications are subject to change without notice.
All product names are trademarks of their respective companies.
LIMITED WARRANTY AND LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke’s behalf.
To obtain service during the warranty period, contact your nearest Fluke authorized service center to obtain return authorization information, then send the product to that Service Center with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR
PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR
CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.
11/99
Table of Contents
Title Page
i
233
Users Manual
ii
Contents (continued)
iii
233
Users Manual
iv
List of Tables
Table Title Page
1.
v
233
Users Manual
vi
List of Figures
Page
1.
9.
vii
233
Users Manual
viii
Introduction
The Fluke 233 (hereafter the Meter) is a compact and easy to operate tool for electrical and electronic circuit measurements.
XW
Warning
Read "Safety Information" before you use the
Meter.
1
233
Users Manual
Safety Information
The Meter complies with:
•
ISA-82.02.01
•
CAN/CSA C22.2 No. 61010-1-04
•
ANSI/UL
•
EN
•
EN
•
EN
•
ETSI EN 300 328 V1.7.1:2006
•
ETSI EN 300 489 V1.8.1:2008
•
FCC Part 15 Subpart C Sections 15.207,
15.209, 15.249 FCCID: T68-F233
•
RSS-210 IC: 6627A-F233
•
Measurement Category III, 1000V, Pollution
Degree 2
•
Measurement Category IV, 600V, Pollution
Degree 2
In this manual, a Warning identifies the conditions and procedures that cause a dangerous situation to the user.
A Caution identifies the conditions and procedures that could cause damage to the Meter, equipment under test damage, or permanent data loss.
The symbols used on the Meter and in this manual are shown in Table 1.
2
Warnings and Cautions
XW
Warning
To prevent possible electrical shock or personal injury, follow these guidelines:
•
Use this Meter only as specified in this manual or the protection can be compromised.
•
Do not use the Meter if it is damaged.
Before you use the Meter, examine the case. Look for cracks or missing plastic.
Carefully look at the insulation around the terminals.
•
Make sure the battery door is closed and locked before you operate the Meter.
•
Replace the batteries when the battery
indicator () appears.
•
Remove the test leads from the Meter before the battery door on the Meter base is opened.
•
Examine the test leads for damaged insulation or exposed metal. Measure the test leads for continuity. Replace damaged test leads before you use the
Meter.
•
Do not apply more than the rated voltage, shown on the Meter, between the terminals or between a terminal and earth ground.
•
Do not operate the Meter with the battery door removed or the case open.
•
Be careful around voltages >30 V ac rms,
42 V ac peak, or 60 V dc. These voltages pose a shock hazard.
•
Use only the replacement fuse specified by the manual.
•
Use the correct terminals, function, and range for measurements.
•
Do not work alone.
True-rms Remote Display Digital Multimeter
Safety Information
•
For current measurements, connect the
Meter to the circuit after you remove circuit power. Always put the Meter in series with the circuit.
•
Connect the common test lead before the live test lead and remove the live test lead before the common test lead.
•
Do not use the Meter if it operates incorrectly. Protection can be compromised. If you are unsure, have the Meter examined.
•
Do not use the Meter around explosive gas, vapor or in damp or wet environments.
3
233
Users Manual
•
Use only specified 1.5-V AA batteries
(three in the Meter base and two in the display), correctly installed, for Meter power.
•
Comply with local and national safety requirements when in hazardous locations.
•
Only use test leads that have the same voltage, category, and amperage ratings as the Meter and that are approved by a safety Agency.
•
Measure a known voltage first to make sure that the Meter operates correctly. If you are unsure, have the Meter examined.
•
Use protective equipment, as directed by local or national authorities when in hazardous work areas.
•
Measure the test leads for continuity before use. Do not use if the resistance is high or noisy.
•
Use only specified replacement parts in the Meter.
•
Keep fingers behind the finger guards on the probes.
4
W
Caution
To prevent damage to the Meter or to the equipment under test, follow these guidelines:
•
Disconnect circuit power and discharge all high-voltage capacitors before you do diode tests or measure resistance, continuity, or capacitance.
•
Use the correct terminals, function, and range for all measurements.
•
Before a current measurement, do the fuse test.
Radio Frequency Data
Note
Changes or modifications to the wireless 2.4
GHz radio not expressly approved by Fluke
Corporation could void the user’s authority to operate the equipment.
This device complies with Part 15 of the FCC Rules.
Operation is subject to the two conditions that follow: (1) this device can not cause interference, and (2) this device must accept any interference, including interference that can cause undesired operation of the device.
Class B digital device: A digital device that is marketed for operation in a residential environment notwithstanding use in commercial, business and industrial environments.
Examples of such devices include, but are not limited to, personal computers, calculators, and equivalent electronic devices that are marketed for operation by the general public.
The Meter was tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the
FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, can cause harmful interference to radio communications.
However, there is no guarantee that interference will not
True-rms Remote Display Digital Multimeter
Safety Information
occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the measures that follow:
•
Reorient or relocate the receiving antenna.
•
Increase the separation between the equipment and receiver.
•
Consult the dealer or an experienced radio/TV technician for help.
The term “IC:” before the radio certification number only signifies the device meets Industry’s Canada technical specifications.
5
233
Users Manual
Hazardous Voltage
When the Meter senses a voltage
≥
30 V or a voltage overload (), the symbol is shown on the display and the red high-voltage LED on the Meter base illuminates to tell you a hazardous voltage is at the Meter input. For frequency measurements >1 kHz, the symbol and highvoltage LED is unspecified.
Test Lead Alert
XW
Warning
To prevent personal injury or damage to the
Meter do not make a measurement with a test lead in an incorrect terminal.
To make sure that you have the test leads in the correct terminals, briefly shows in the display and a beep sounds when you move the function switch to or from an
A (Amps) position.
6
True-rms Remote Display Digital Multimeter
Test Lead Alert
B
AC (Alternating Current)
F
DC (Direct Current)
X
Hazardous voltage
Table 1. Electrical Symbols
J
Earth ground
I
Fuse
P
Conforms to European Union directives.
W
Risk of Danger. Important information.
See Manual.
Conforms to relevant Canadian Standards
Association directives.
Battery. Low battery when shown.
T
Double insulated
R
Continuity test or continuity beeper tone.
E
Capacitance
CAT
III
~
®
IEC Measurement Category III
CAT III equipment has protection against transients in equipment in fixedequipment installations, such as distribution panels, feeders and short branch circuits, and lighting systems in large buildings.
Do not discard this product as unsorted municipal waste. Go to the Fluke website for recycling data.
Examined and licensed by TÜV Product
Services.
)
CAT
IV
O
;
IEC Measurement Category IV
CAT IV equipment has protection against transients from the primary supply level, such as an electricity meter or an overhead or underground utility service.
Diode
Conforms to relevant Australian standards.
7
233
Users Manual
Features
See Tables 3 through 4 for a list of Meter features with a short feature description.
Table 2. Display
1
2 3
4
5
14
13
12
11 meter
10
Auto RangeManual Range
remote
7
9
8
8
No. Symbol
1
2
3
Indication
MIN MAX AVG mode on.
MAX MIN AVG Maximum, minimum, or average measurement shown.
Display hold on. Display freezes the measurement.
6 gcc101.eps
True-rms Remote Display Digital Multimeter
Features
No. Symbol
Table 2. Display (cont.)
Indication
4
5
°
C,
°
F
A
Radio connection indicator. degrees Celsius, degrees Fahrenheit
6
V, mV
μ
F, nF
DC AC
Ω
, M
Ω
, k
Ω amperes (amps) volts, millivolts microfarad, nanofarad
Direct current or alternating current.
Hz, kHz
ohm, megohm, kilohm hertz, kilohertz
7 remote Battery low warning for the display module.
8 Manual Range Manual range set.
9 Auto Range Auto range set.
10 meter Battery low warning for Meter base.
11
12
Diode test mode.
Continuity
13
14
Input is a negative value.
X
Hazardous voltage. Measured input voltage
≥
30 V, or voltage overload condition (OL)
9
233
Users Manual
Table 3. Inputs
10
1 3
2
No. Terminal
1
2
3
Description
A
Input for 0 A to 10.00 A current measurements.
COM
Common terminal for all measurements.
Input for voltage, continuity, resistance, diode, capacitance, temperature, and frequency measurements. gcc110.eps
True-rms Remote Display Digital Multimeter
Features
Table 4. Function Switch Positions
Switch
Position
Description
Hz (button)
AC voltage from 0.06 to 1000 V.
Frequency from 5 Hz to 50 kHz.
DC voltage from 0.001 V to 1000 V.
AC voltage from 6.0 to 600.0 mV, dc-coupled. DC voltage from 0.1 to 600.0 mV. e
Ohms from 0.1
Ω
to 40 M
Ω
.
Continuity beeper turns on at
<
20
Ω
and turns off at
>
250
Ω
.
Farads from 1 nF to 9999
μ
F.
Diode Test. OL shows in the display when the input voltage is >2.0 V.
Hz (button)
Temperature.
AC current from 0.1 A to 10 A (>10 to 20 A, 30 seconds on, 10 minutes off).
>
10.00 A display flashes.
>
20 A, OL is shown. DC-coupled.
Frequency from 45 Hz to 5 kHz.
DC current from 0.001 A to 10 A (>10 to 20 A, 30 seconds on, 10 minutes off).
>
10.00 A display flashes.
>
20 A, OL is shown.
Note: All ac functions are true-rms. AC voltage is ac-coupled. AC mV and ac amps are dc-coupled.
11
233
Users Manual
Error Messages
Table 5 contains possible error messages and the steps to clear the error.
Table 5. Error Messages
Error Messages
Display-module batteries must be replaced before the Meter will operate.
Meter-base batteries must be replaced before the Meter will operate.
Calibration necessary. Meter calibration is necessary before the Meter will operate.
Internal error. The Meter must be repaired before it will operate.
Loss of radio connection with the Meter base.
12
Battery Saver™(Sleep Mode)
The Meter powers-down (Sleep mode) if there is no function change, range change, or button push for 20 minutes. The lowest power drain occurs when the display module is docked with the Meter base. With the display module removed from the Meter base, the power drain is more because the radios are turned on.
To wake up the Meter, push a button or turn the function switch.
To disable the Sleep mode, hold down the button while turning the Meter on. The Sleep mode is always disabled in the MIN MAX AVG mode.
MIN MAX AVG Record Mode
The MIN MAX AVG record mode records the minimum and maximum input values, and calculates an average of all measurements. Each new high or low measurement causes the Meter to beep.
•
Set the Meter to the measurement function and range.
•
Push to enter MIN MAX AVG mode.
•
and MAX are shown and the highest measurement detected since is shown.
•
Push to step through the low (MIN), average
(AVG), and present measurements.
True-rms Remote Display Digital Multimeter
Battery Saver™(Sleep Mode)
•
To pause MIN MAX AVG record mode, push .
is shown. A pause does not erase recorded
MIN MAX AVG measurements.
•
To continue the MIN MAX AVG record mode, push
again.
•
To exit and erase recorded measurements, push
for at least one second or turn the function switch.
13
233
Users Manual
Display Hold
XW
Warning
To prevent electrical shock, when Display
HOLD is on, disable Display HOLD to measure the voltage that is possibly different than the Display HOLD measurement.
Display HOLD freezes the display.
1. Push to activate Display HOLD. ( is shown.)
2. To exit and start normal operation, push or turn the function switch.
Manual and Autoranging
The Meter has Manual and Autorange modes.
•
In the Autorange mode, the Meter sets the range to one with the best resolution for the input signal.
•
In the Manual Range mode, you override Autorange and set the range yourself.
When you turn the Meter on, it is set to Autorange and
Auto Range shows in the display.
1. To set the Meter to the Manual Range mode, push
. Manual Range shows in the display.
2. In the Manual Range mode, push to increment the range. After the highest range, the range of the
Meter is set to the lowest range.
Note
You cannot manually change the range in the
MIN MAX AVG or Display HOLD modes.
If you push while in MIN MAX AVG or
Display HOLD, the Meter beeps twice, indicating an invalid operation and the range does not change.
3. To exit Manual Range, push for at least one second or turn the function switch. The Meter is set to
Autorange and Auto Range shows in the display.
14
Backlight
Push
to toggle the backlight on and off. The backlight automatically extinguishes after 40 seconds. To disable backlight auto-off, hold down while turning the Meter on.
Power-Up Options
Hold a button down while the function switch is moved from the OFF position to set a power-up option. The power-up options cancel when the function switch is moved to OFF or the Meter goes into Sleep mode. See
Table 6 for all power-up options.
Table 6. Power-Up Options
Illuminates all display segments.
Disables the beeper. is shown when on.
Disables automatic power-down ("Sleep mode"). is shown when on.
Disables backlight auto-off. is shown when on.
True-rms Remote Display Digital Multimeter
Backlight
15
233
Users Manual
How to Make Measurements
The sections that follow tell how to make measurements with the Meter.
To connect the test leads to the circuit or device, connect the common (COM) test lead first. To remove the test leads, remove the common test lead last.
AC and DC Voltage Measurements
The voltage ranges are 600.0 mV, 6.000 V, 60.00 V,
600.0 V, and 1000 V. To set the 600.0 mV dc or ac range, turn the function switch to . Millivolts ac is set first.
Push to switch to millivolts dc.
Refer to Figure 1 to measure ac or dc voltage.
For voltage measurements, the Meter puts approximately
10 M
Ω
(10,000,000
Ω
) in parallel with the circuit. This load can cause measurement errors in high-impedance circuits.
In most cases, the error is negligible (0.1 % or less) if the circuit impedance is 10 k
Ω
(10,000
Ω
) or less.
AC Voltage
S w itch Box
DC Voltage
V
+
gcc102.eps
Figure 1. AC and DC Voltage Measurements
16
Resistance Measurements
W
Caution
To prevent possible damage to the Meter or to the equipment under test, disconnect power and discharge all high-voltage capacitors before resistance measurements.
The Meter sends a small current through the circuit for resistance measurements. Because this current flows through all possible paths between the probes, the resistance measured is the total resistance of all paths between the probes.
The resistance ranges are 600.0
Ω
, 6.000 k
Ω
, 60.00 k
Ω
,
600.0 k
Ω
, 6.000 M
Ω
, and 40.00 M
Ω
.
Set the Meter as shown in Figure 2 to measure resistance.
Below are some hints for resistance measurements:
•
The measured value of a resistor in a circuit is frequently different from the specified resistor value.
•
The test leads can add 0.1
Ω
to 0.2
Ω
of error to resistance measurements. To measure test lead resistance, touch the probe points together and read the resistance.
•
The resistance function uses sufficient voltage to forward-bias silicon diode or transistor junctions, and cause current to flow. If you think current flows through the junction, push to apply a lower
True-rms Remote Display Digital Multimeter
How to Make Measurements
current in the next higher range. If the value is higher, use the higher value. Refer to the Input Parameters table in the specifications section for typical shortcircuit currents.
17
233
Users Manual
In-Circuit Resistance Measurements
Circ u it Po w er
OFF
Isolating a Potentiometer
1
3
2
Disconnect
1 2
3
Isolating a Resistor
Figure 2. Resistance Measurements
Disconnect gcc106.eps
18
Temperature Measurements
The Meter measures the temperature of a type-K thermocouple (included). Choose between degrees
Celsius (
°
C) or degrees Fahrenheit (
°
F) by pushing
C.
W
Caution
To prevent possible damage to the Meter or other equipment, use a thermocouple rated for the temperatures to be measured. The
Meter is rated for -40.0
°
C to +400.0
°
C and
-40.0
°
F to 752
°
F, but the included type-K thermocouple is rated to 260
°
C.
The temperature ranges are -40.0
°
C to +400
°
C and
-40.0
°
F to 752
°
F. All other temperatures show on the display. When there is no thermocouple connected, the display shows .
To measure temperature:
1. Connect a type-K thermocouple to the COM and terminals of the Meter.
2. Turn the function switch to .
True-rms Remote Display Digital Multimeter
How to Make Measurements
Continuity Tests
W
Caution
To prevent possible damage to the Meter or the equipment under test, disconnect power and discharge all high-voltage capacitors before a continuity test.
The continuity test uses a beeper that sounds when a closed circuit is sensed. The beeper lets you do continuity tests without the necessity to look at the display.
To do a continuity test, set up the Meter as shown in
Figure 3.
19
233
Users Manual
For in-circ u it tests, t u rn circ u it po w er off.
Acti v ates contin u ity b eeper
O N
(closed)
OFF
(open)
Figure 3. Continuity Tests
gcc103.eps
20
Diode Tests
W
Caution
To prevent possible damage to the Meter or to the equipment under test, disconnect power and discharge all high-voltage capacitors before a diode test.
Do a diode test on diodes, transistors, silicon controlled rectifiers (SCRs), and other semiconductor devices. This function sends a current through the semiconductor junction and then measures the voltage drop across the junction. A good silicon junction drops between 0.5 V and
0.8 V.
To do a diode test on a diode out of a circuit, set up the
Meter as shown in Figure 4. For forward-bias measurements on a semiconductor component, put the red test lead on the positive terminal of the component and put the black test lead on the negative terminal of the component.
In a circuit, a good diode has a forward-bias measurement of 0.5 V to 0.8 V. A reverse-bias measurement includes the resistance of other pathways between the probes.
A short beep sounds if the diode is good (<0.85 V). A continuous beep sounds if the measurement is
≤
0.100 V or short circuit. The display shows “” if the diode is open.
True-rms Remote Display Digital Multimeter
How to Make Measurements
21
233
Users Manual
Typical
Reading
For w ard Bias
+
Single Beep
Bad Diode Open
Figure 4. Diode Test
Re v erse Bias
+
Shorted or gcc109.eps
22
Capacitance Measurements
W
Caution
To prevent possible damage to the Meter or to the equipment under test, disconnect power and discharge all high-voltage capacitors before capacitance measurements. Use the dc voltage function to make sure that the capacitor is discharged.
Capacitance ranges are 1000 nF, 10.00
μ
F, 100.0
μ
F, and 9999
μ
F.
To measure capacitance, set up the Meter as shown in
Figure 5.
True-rms Remote Display Digital Multimeter
How to Make Measurements
Select
Capacitance
+
+
+
+
+
+
+
+
+
Figure 5. Capacitance Measurements
gcc104.eps
23
233
Users Manual
AC and DC Current Measurements
XW
Warning
To prevent possible electrical shock or personal injury, do not make an in-circuit current measurement where the open-circuit potential to earth is >1000 V. Meter damage or injury can occur if the fuse blows during such a measurement.
W
Caution
To prevent possible damage to the Meter or to the equipment under test:
•
Do a fuse test before current measurements.
•
Use the correct terminals, function, and range for all measurements.
•
Do not put the probes across (in parallel with) a circuit or component when the test leads are connected to the current terminals.
To measure current, you must break the test circuit, then put the Meter in series with the circuit.
The current ranges are 6.000 A, and 10.00 A. AC current is shown as an rms value.
To measure current (see Figure 6):
1. Remove power from the circuit. Discharge all highvoltage capacitors.
2. Put the black test lead into the COM terminal. Put the red test lead into the A terminal.
3. Set the function switch to for ac current or for dc current.
24
1
3
True-rms Remote Display Digital Multimeter
How to Make Measurements
Total Current to Circuit
Circ u it Po w er:
OFF
to connect meter.
ON
for meas u rement.
OFF
to disconnect meter.
5
Current Through One Component
2
Figure 6. Current Measurements
5 gcc107.eps
25
233
Users Manual
Frequency Measurements
A frequency measurement is a count of the number of times an ac voltage or current signal crosses a threshold point each second.
To make a frequency measurement:
1. Set the function switch to for voltage or for current.
2. Connect the Meter to the signal source.
3.
Push
.
The Meter autoranges to one of four frequency ranges:
99.99 Hz, 999.9 Hz, 9.999 kHz, and 50 kHz.
Below are some hints for frequency measurements:
•
If a measurement shows as 0 Hz or is not stable, the input signal can be below or near the trigger level. A lower range increases the sensitivity of the Meter and can usually repair these problems.
•
An input signal with distortion can cause a frequency measurement to be higher than usual. The distortion can cause multiple triggerings of the frequency counter. A higher voltage range decreases the input sensitivity and can correct this problem. In general, the lowest frequency is the correct one.
Remote Operation
The Meter uses low-power 802.15.4 wireless technology to allow the display module to operate in a different location than the Meter base. Although there is control of some Meter functions (Hold, MIN MAX AVG, Range, and
Backlight), complete remote control of the Meter is not available through the display module.
The wireless radio does not interfere with meter measurements. Usually, the radio is off when the display module is docked on the Meter base. It is possible for the radio to be on when the display module is docked and the function switch is set to OFF. To make sure that the radio is off, remove the batteries from the Meter base and display module.
The display module is synchronized with a Meter base when it is docked on the Meter base and turned on.
Different display modules can be synchronized with a
Meter base but, only one display module can be synchronized to a Meter base at the same time.
26
Remove the Display Module
To remove the display module (see Figure 7):
1. Push in on the latches on the sides of the display module.
2. Pull the display module off of the top end of the
Meter base.
The Meter base and display module can be a maximum of 10 Meters (30 feet) from each other before the radio connection is broken. This distance can change if obstacles are between the Meter base and the display module. There is a radio connection between the display module and Meter base when shows in the display.
When the display module and Meter base lose the radio connection, the display shows dashes and blinks.
Possible causes for this loss are the distance is too far for the environment or the batteries in the Meter base are dead. To reconnect, decrease the distance between the display module and Meter base.
True-rms Remote Display Digital Multimeter
Remote Operation
1
2
Figure 7. Display Module Separation
gcc114.eps
27
233
Users Manual
If the radios in the Meter base and display module do not connect, flashes in the display. Dock the display module with the Meter base and turn the Meter off and then on. When the Meter is turned on, the red highvoltage LED on the Meter base flashes. If not, replace the
Meter base batteries. For maximum battery life, dock the display module to the Meter base when the Meter is off.
The display module has a built-in magnet to attach to metal surfaces.
Dock the Display Module with the Meter base
To dock the display with the Meter base as shown in
Figure 8:
1. Set the display on the top 10 millimeters of the Meter base with the battery compartment of the display in the channel in the top of the Meter base.
2. Push the display nearer the Meter base until the display latches catch.
1
2 gcc115.eps
Figure 8. Dock Display Module with Meter Base
28
Maintenance
XW
Warning
To prevent a possible electrical shock or personal injury, have an approved technician repair the Meter.
General Maintenance
Clean the case with a damp cloth and weak detergent. Do not use a solvent or cleaners with abrasives.
Dirt or moisture in the terminals can cause incorrect measurements. To clean the terminals:
1. Turn the Meter off and remove all test leads.
2. Shake out dirt that can possibly be in the terminals.
3. Soak a clean swab with weak detergent and water.
Move the swab around in each terminal. Dry each terminal with canned air to push the water and detergent out of the terminals.
XW
Warning
To prevent electrical shock or personal injury, remove the test leads and all input signals before you replace the batteries or fuses. To prevent damage or injury, install
ONLY specified replacement parts shown in
Table 7.
True-rms Remote Display Digital Multimeter
Maintenance
Battery Replacement
XW
Warning
To prevent incorrect measurements, possible electrical shock, or personal injury, replace
the battery when the battery indicator ()
appears. If the display shows , the
Meter will not function until the display module batteries are replaced. If the display
shows , the Meter will not function
until the Meter base batteries are replaced.
There are two low-battery indicators in the display: one for the Meter base batteries and one for the display module batteries. Replace the batteries when the lowbattery indicator shows.
29
233
Users Manual
1
2
4
3
Figure 9. Meter Base Battery Replacement
30 gcc112.eps
To replace the batteries in the Meter base:
1. Turn the Meter off and remove all test leads.
2. Lift the tilt stand up as shown in Figure 9.
3. Turn the battery-door latch with a standard screwdriver until the unlocked symbol () aligns with the arrow.
4. Lift off the battery door.
5. Remove the three AA batteries and replace them with new ones. Use the correct battery orientation.
6. Install the battery door.
Turn the battery-door latch until the locked symbol () aligns with the arrow.When the Meter does not power on, the Meter base batteries or display-module batteries can be dead. To find which of the batteries to replace:
1. Dock the display module with the Meter base.
2. Turn the function switch to off and then on.
If the red high-voltage LED on the Meter base flashes, the
Meter base batteries are good. Replace the displaymodule batteries and turn the Meter on.
True-rms Remote Display Digital Multimeter
Maintenance
31
233
Users Manual
To replace the batteries in the display module:
1. Remove the display module from the Meter base.
See the “Remove the Display” section.
2. Remove the battery door of the display module as shown in Figure 10.
3. Remove the two AA batteries and replace them with new ones. Use the correct battery orientation.
4. Replace the battery door on the display module.
Dock the display module with the Meter base and turn the
Meter on.
2
3 gcc111.eps
Figure 10. Display-Module Battery Removal
32
Fuse Test
To do a fuse test:
1. Set the function switch to .
2. Connect a test lead to the jack as shown in
Figure 11.
3. Touch the other end of the test lead to the A jack.
A good fuse will show a resistance of 0.5
Ω
or less.
Replace the fuse if the resistance is higher or is shown.
True-rms Remote Display Digital Multimeter
Maintenance
Good fuse: 0.0 Ω to 0.5 Ω
Replace fuse: OL
Figure 11. Fuse Test
gcc105.eps
33
233
Users Manual
Fuse Replacement
To replace the fuse:
1. Remove the test leads from the Meter.
2. Remove the display module from the Meter base.
See the “Remove the Display Module” section.
3. As shown in Figure 12, remove four screws from the case bottom.
4. Pull the case bottom from the case top.
5. Remove the fuse from its holder and replace it with an 11 A, 1000 V, FAST fuse with a minimum interrupt rating of 17,000 A. Use only Fluke PN 803293.
To re-assemble the Meter, do the steps above in the opposite sequence.
3
4
5
Figure 12. Fuse Replacement
gcc113.eps
34
True-rms Remote Display Digital Multimeter
Service and Parts
Service and Parts
If the Meter fails, replace the battery and do a fuse test.
Read this manual to make sure the Meter is applied correctly.
Replacement parts and accessories are shown in Table 7 and Figure 13.
To get parts and accessories, refer to “How to Contact
Fluke”.
Table 7. Replacement Parts
Battery, AA 1.5 V
W
Fuse, 11 A, 1000 V, FAST
Battery Door – Display Module
Battery Door – Meter Base
233 Display
Alligator Clip, Black
Alligator Clip, Red
Test Lead Set
Integrated DMM Temperature Probe
233 Users Manual CD
233 Getting Started Manual
W
To ensure safety, use exact replacement only.
[1] Contact your local Fluke service center for display replacement.
1
1
5
1
376756
803293
3383770
3383762
1
1
1
AC72
1
1 TL75
1 80BK-A
3465353
3465366
35
233
Users Manual
Battery
AA 1.5 V
Fuse,
11A, 1000 V, Fast
233 Display
Battery Door
Display Module
Battery Door
Meter Base
233
Users Manual CD
233
Getting Started Manual
Figure 13. Replacement Parts
AC72
Alligator Clips
TL75
Test Lead Set
80BK-A Integrated DMM
Temperature Probe gcc116.eps
36
True-rms Remote Display Digital Multimeter
Service and Parts
Table 8. Accessories
Item Description
TPAK
TL223
TL220
AC285
ToolPak Magnetic Hanger
SureGrip™ Electrical Test Lead Set
Industrial Test Lead Set
SureGrip™ Alligator Clips
AC87 i400s
Heavy Duty Bus Bar Clip Set
AC Current Clamp (requires PM9081 adapter)
PM9081 Dual Banana Plug (male) to Female BNC Adapter
Fluke accessories are available from an approved Fluke distributor.
37
233
Users Manual
General Specifications
Maximum voltage between any
terminal and earth ground ........................................... 1000 V rms
W
Fuse for A inputs ..................................................... 11 A, 1000 V 17000A interrupt rating Fuse
Display ........................................................................... 6000 counts, updates 4/sec (Frequency: 9,999 counts, Capacitance: 1,000 counts)
Altitude
Operating .................................................................... 2,000 meters
Storage ....................................................................... 12,000 meters
Temperature
Operating .................................................................... -10
°
C to +50
°
C
Storage ....................................................................... -40
°
C to +60
°
C
Temperature coefficient ............................................... 0.1 X (specified accuracy) /
°
C (< 18
°
C or > 28
°
C)
Electromagnetic Compatibility (EN 61326-1:2006) .... In an RF field of 3 V/m, accuracy = specified accuracy except in temperature: specified accuracy
±
5
°
C (9
°
F)
Wireless Frequency ...................................................... 2.4 GHz ISM Band 10 meter range
Relative Humidity.......................................................... Maximum noncondensing
90 % at 35
°
C
75 % at 40
°
C
45 % at 50
°
C
0 % to 70 % for 40 M
Ω
range
Battery Type
Meter base.................................................................. Three AA Alkaline batteries, NEDA 15A IEC LR6
Display module ........................................................... Two AA Alkaline batteries, NEDA 15A IEC LR6
Battery Life .................................................................... 400 hrs typical (Alkaline)
Shock ............................................................................. 1 Meter drop 6 sides per IEC 61010
Size (H x W x L) ............................................................. 5.3 cm x 9.3 cm x 19.3 cm
38
True-rms Remote Display Digital Multimeter
Detailed Specifications
Weight ............................................................................604 g (1.3 lbs)
Safety Compliance ........................................................Complies with ANSI/ISA S82.01-2004, CSA 22.2 No. 61010-1-04 to 1000 V
Measurement Category III and 600 V Measurement Category IV.
Certifications .................................................................CSA, TÜV (EN61010), P, ; (N10140),VDE, GOST
Detailed Specifications
For all detailed specifications:
Accuracy is specified for 1 yr after calibration, at operating temperatures of 18
°
C to 28
°
C, with relative humidity at 0 % to 90 %. Accuracy specifications take the form of
±
([ % of Reading ] + [ Number of least significant digits ]).
AC Voltage
AC conversions are ac-coupled and valid from 1 % to 100 % of range.
Range
[1]
Resolution
45 – 500 Hz
600.0 mV
6.000 V
60.00 V
600.0 V
1000 V
0.1 mV
0.001 V
0.01 V
0.1 V
1 V
±
(1.0 % + 3)
[1] Crest factor of
≤
3 at 4000 counts, decreasing linearly to 1.5 at full scale.
Accuracy
500 Hz – 1 kHz
±
(2.0 % + 3)
39
233
Users Manual
DC Voltage, Conductance, and Resistance
Function Range Resolution Accuracy mV dc
V dc
600.0 mV
6.000 V
60.00 V
600.0 V
0.1 mV
0.001 V
0.01 V
0.1 V
±
(0.25 % + 2)
1000 V 1 V
Ω
600.0
Ω
0.1
6.000 k
Ω
0.001
60.00 k
Ω
0.01
600.0 k
Ω
0.1
6.000 M
Ω
0.001
40.00 M
Ω
0.01
±
(0.9 % + 2)
±
(0.9 % + 1)
±
(1.5 % + 2)
Continuity
The beeper is guaranteed on <20
Ω
, and guaranteed off >250
Ω
, and detects opens or shorts of 500
μ s or longer.
Temperature
-40
-40
°
°
C to +400
F to +752
°
°
C
F
0.1
°
C
0.1
°
F
[1] Temperature uncertainty (accuracy) does not include error of the thermocouple probe.
±
(1.0 % + 10)
±
(1.0 % + 18)
40
AC Current
Function Range Resolution
A ac
[1,2,3]
6.000 A
10.00 A
0.001 A
0.01 A
[1] All ranges are specified from 5 % of range to 100 % of range.
[2] Crest factor of
≤
3 at 4000 counts, decreasing linearly to 1.5 at full scale.
[3] AC current >10 A is unspecified. 20 A continuous overload for 30 seconds maximum.
DC Current
Function Range
A dc
[1]
6.000 A
10.00 A
Resolution
0.001 A
0.01 A
[1] DC current >10 A is unspecified. 20 A continuous overload for 30 seconds maximum.
True-rms Remote Display Digital Multimeter
Detailed Specifications
Accuracy
(45 – 500 Hz)
±
(1.5 % + 3)
Accuracy
±
(1.0 % + 3)
41
233
Users Manual
Capacitance
1000 nF 1 nF
10.00
μ
F 0.01
100.0
μ
F 0.1
9999
μ
F 1
Diode
2.000 V 0.001 V
±
(1.9 % + 2)
[1]
±
(0.9 % + 2)
42
True-rms Remote Display Digital Multimeter
Detailed Specifications
Frequency
AC coupled, 5 Hz to 50 kHz, for V ac; dc coupled, 45 Hz to 5 kHz for A ac switch position.
99.99 Hz
999.9 Hz
9.999 kHz
50.00 kHz
MIN MAX Recording
Nominal Response
100 ms to 80 %
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
±
(0.1 % + 2)
Accuracy
Specified accuracy
±
12 counts for changes >200 ms in duration (
±
40 counts in ac)
43
233
Users Manual
Ω
R
E
G
Input Characteristics
Function
L
K
Overload
Protection
Input
Impedance
(nominal)
Common Mode
Rejection Ratio
(1 k
Ω
unbalance)
> 100 dB at dc, 50 Hz
1100 V rms >10 M
Ω
<100 pF or 60 Hz
1100 V rms >5 M
Ω
< 100 pF > 60 dB, dc to 60 Hz
Open Circuit
Test Voltage
Full Scale Voltage
To 6 M
Ω
40
1100 V rms
1100 V rms
1100 V rms
1100 V rms
<2.7 V dc
<2.7 V dc
<2.7 V dc
<2.7 V dc
<0.7 V dc <0.9 V dc
<300 mV dc
<700 mV dc
Up to 2.000 V dc
Normal Mode Rejection
> 60 dB at 50 Hz or 60 Hz
Typical Short Circuit Current
<350
μ
A
<350
μ
A
<350
μ
A
1.2 mA
44
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 1 233 Users Manual
- 2 Warranty
- 3 Table of Contents
- 7 List of Tables
- 9 List of Figures
- 11 Introduction
- 11 How to Contact Fluke
- 12 Safety Information
- 12 Warnings and Cautions
- 15 Radio Frequency Data
- 16 Hazardous Voltage
- 16 Test Lead Alert
- 18 Features
- 22 Error Messages
- 23 Battery Saver™(Sleep Mode)
- 23 MIN MAX AVG Record Mode
- 24 Display Hold
- 24 Manual and Autoranging
- 25 Backlight
- 25 Power-Up Options
- 26 How to Make Measurements
- 26 AC and DC Voltage Measurements
- 27 Resistance Measurements
- 29 Temperature Measurements
- 29 Continuity Tests
- 31 Diode Tests
- 33 Capacitance Measurements
- 34 AC and DC Current Measurements
- 36 Frequency Measurements
- 36 Remote Operation
- 37 Remove the Display Module
- 38 Dock the Display Module with the Meter base
- 39 Maintenance
- 39 General Maintenance
- 39 Battery Replacement
- 43 Fuse Test
- 44 Fuse Replacement
- 45 Service and Parts
- 48 General Specifications
- 49 Detailed Specifications
- 49 AC Voltage
- 50 DC Voltage, Conductance, and Resistance
- 50 Continuity
- 50 Temperature
- 51 AC Current
- 51 DC Current
- 52 Capacitance
- 52 Diode
- 53 Frequency
- 53 MIN MAX Recording
- 54 Input Characteristics