DS1151B - leotech
Deviser
DS1151B
Operation Manual
Ver 1.1
The best selection
Part Number:
DK210101151
Warranty
The instrument is warranted for a period of 18 months since
the date of shipment, when used under normal operating
conditions and within the service conditions. The obligation of
DEVISER arising from a warranty claim shall be limited to
repairing or at its option, replacing without charge, any
assembly or component (except batteries).
Deviser shall have no responsibility for any defect or damage
caused by improper use, improper maintenance or for any
product which has been repaired or altered by any one other
than DEVISER or its authorized service center.
Buyer shall prepay shipping charges and insurance for
products returned for warranty service. Deviser or its
authorized service center shall pay costs for return of products
to buyers.
INDEX
1.General Introduction.··················1
1.1
1.2
1.3
Introduction.···
···················1
Equipment.····
···················2
Features.······
···················2
Level Test······················3
Channel Spectrum Scanning·······4
Frequency Spectrum··············4
Title/Level List················4
C/N·················
············5
Data Logging····················5
Auto Test·······················5
Limit Test······················6
Voltmeter·······················6
Printer Output··················6
QAM Measurement·················6
Multiple User Plan··············7
Intelligence Power Management···7
Fast Charge·····················7
2.Product Description.···················8
2.1 Introduction···········
············8
2.2 Getting Acquainted with the Keypad.···9
2.3 Display Description····
············10
3. Using The Instrument·················11
3.1 Introduction.···········
···········11
3.2 Learn User Plan.·······
············11
3.3 Level Measurement.····
············14
3.4 Full Scan·························21
3.5 Frequency Spectrum····
············25
3.6 Tilt/Level List··········
···········34
3.7 Limit Test.············
············37
3.8 Auto Test ························41
3.9 Save and Load·········
············53
3.10 Voltage Meter.,·········
············57
3.11 C/N·····
·························57
3.12 Print.····
·························58
4. Setup.······
·························61
4.1 Instruction.························61
4.2 Information.·······················62
4.3 General.·
·························64
4.4 Measurement.······················70
4.5 Channel Plan·······················75
5. Power Supply·
···
········
·············85
6. Port·······
··························88
Appendix
Specification··························91
1.1 INTRODUCTION
The DS1151B Signal Level Meter is designed for cable
system, and provides you with favorite features for
reduced cost.
It can measure and display the amplitude of the carriers
singly, as a group (up to 12 channels), or as a full
span(at most 150 channels). Especially, the occasional
disturbance can be detected by the peak-hold level
measurement feature in LEVEL and SPECTRUM mode.
The Analogue channel, QAM and FM channel can be
set and tested efficiently, it also supports C/N
measurement, voltmeter function, spectrum, auto test
and data logging. It is easy for DS1151B to obtain a
report of installation data via PC or printer.
The meter is favorite, has many functions and is easy to
use in a wide range of conditions.
1
1.2 EQUIPMENT
*Charger.
*RF connector
*Operator’s manual
*Data cable
.
*Serial printer cable.
*Com Software
*Package
.
DS10030
P.110628J8J F-F
DK210101185
P.3100805L9
P.310075L25
DK401101182
PK100151182
The following options are also available:
*Parallel printer adapter:
0115005
*Noise source:
DS5032
*Reserve signal generator DS5103 DS5112
1.3 FEATURES
The following features are provided by DS1151B
* Level Test
* Channel Spectrum Scanning (up to 150 channels)
* Frequency Spectrum
* Tilt/Level list (up to 12 channels)
* C/N
* Data Logging
* Auto Test
* Limit Test
* Voltmeter Feature
2
* Printer Output by serial or parallel printer
* QAM Measurement
* Remote test
* Multiple User Plan
* Intelligence Power Management
* Fast Charge during 3 hours
* Alphanumeric Inputting
* Fast Setup
LEVEL TEST
The DS1151B takes accurate signal level test on
analogue TV, QAM and single frequency channel.
When turned to a channel, the levels of video and audio,
△V /A also with the peak-hold level and △Peak–Peak
value can be easy displayed. To get more details of a
channel, a single channel SPECTRUM mode is
supported at level function.
Eight base channel plans and up to five user plans are
provided. You can determine the user plan automatically
by a special fast building mode.
The test data of level measurement can be saved and
printed out to a serial or parallel printer.
3
CHANNEL SPECTRUM SCANNING
A full span of audio and video carriers can be displayed
efficiently. You can zoom in or out for 5 levels to
observe the overall flatness and amplitude of your
system. By using marker, you can change the scanned
channels and read out the test data of video and audio
carriers. At most 150 channels are supported in SCAN
mode.
It is easy to save the scanning data or print it out in
report format.
FREQUENCY SPECTRUM
Frequency spectrum is applied ranging form 2.5 MHz to
62.5 MHz, or according to selected user plan. To
determine the problem of disturbance, the peak-hold
level spectrum can be enabled, and two curves are
displayed in one screen. One is for the real test spectrum,
another is for the peak-hold level spectrum. You can
read out the test data and frequency by marker and delta
maker function. Especially, the data of frequency
spectrum could be saved and printed out.
TILT/LEVEL LIST
To fast check the flatness and amplitude, the
4
TILT/LEVEL LIST function is favorite for you. The
level of up to 12 channels can be displayed as spectral
graph or in level table by pressing the TILT button. You
can save or print the test data of TILT/LEVEL LIST
measurement as you like.
C/N
C/N test on line is supported by the meter. To get the
more accurate C/N data, the method of shutting off
carrier is enabled.
DATA LOGGING
Up to 110 files can be saved in one meter depending on
the size of the saved files.
You can select and save the test data of LIMIT TEST,
LEVEL mode, SCAN mode, SPECTRUM mode, TILT
mode or all of them in one file. The file name can be
determined directly to define the saved contents, and the
saved files can be transmitted to PC for data analysis.
AUTO TEST
DS1151B provides a convenient way to automatic test
the cable system according to test commands edited by
user. You can select the test items including LEVEL,
5
SPECTRUM, LIMIT and TILT. Automated tests can be
executed immediately and the test data will be saved
automatically.
LIMIT TEST
You can fast check the cable system by the limit test
function. Each enabled channel will be tested according
to the limits set by user, and marked by pass /fail
indicators. Limit test results are video level, audio level,
△V/A, and pass/fail indicator can be viewed on the
LCD screen.
VOLTMETER
The problem of power supplies or drops can be found
out by a built-in voltmeter of DS1151B with voltages
ranging from 0 to 100V.
PRINTER OUTPUT
You can get a hardcopy of the saved file by a serial or
parallel printer, and the measured data can be printed in
report format. It is useful to document a problem.
QAM MEASUREMENT
The instrument is capable of measuring the actual power
of QAM signal, and showing the shape of the
6
modulation wave. This function takes it enable to detect
the flatness of the QAM signal and found out the
problem of the digital transmission system.
MULTIPLE USER PLAN
You can build up to five user plans in one instrument. It
makes it easy to quickly select the correct user plan for
the cable system at which you are working.
INTELLIGENCE POWER MANAGEMENT
To conserve battery life, the difference power supplying
method is used for difference working condition. The
instrument can work with the built-in battery for 6 hrs
on maximum power supply status. When working in
power saving mode, part circuits of the meter will
power down automatically.
FAST CHARGE
The meter has the capability to full charge the battery
during 3hrs. A specified charger is used, and the
charging time, temperature, voltage and charging
current are detected when charging. The charging
process will be stopped automatically as soon as the full
charge is detected.
7
Product
Description
2.1 INTRODUCTION
DS1151B is a lightweight, easy to use level meter.
Before the first using you must become familiar with its
controls.
RF INPUT PORT
LCD
Soft Key
Digit And Function Key
Power Switch
RS232 Interface
Up And down Key
Charge Socket
8
2.2 GETTING ACQUAINTED
WITH THE KEYPAD
Soft Buttons
There are three soft buttons (F1, F2 and F3) located on
the right top of the instrument. They are used to access
the functions represented by the icons displayed on the
bottom of the display.
Power ON/OFF
Turns the instrument on and off
Arrow Buttons.
The UP and DOWN buttons can be used to move the
marker or change the values.
2nd Button
The keypad is put into a second function after pressing
the 2nd button, and some keys perform as
alphanumerical keys.
C/N.
Enter Carrier-to-Noise measurement Mode.
FILE.
To save or Load a file by pressing the button.
LEVEL Enter the level measure mode
SCAN
Channel spectrum scanning
SPECT Enter frequency spectrum
TILT
Toggles between tilt and level list mode
LIMIT Limit test for all active channels
AUTO
Automated test according to the commands
SET
Enter the fast setup and full setup mode
9
Alphanumeric Input
The keypad will be put into a alphanumeric keypad by
pressing 2nd button. The keys labeled on one numeral
and up to three characters can perform as alphanumeric
keys. To access a numeral or a character you can
repeatedly pressing the key. Once the desired numeral
or character is input and displayed, move the cursor to
the next place by pressing F3 key. The F2 key performs
backspace function. Be sure to confirm the entry by
pressing the F1(ENTER) key.
2.3 DISPLAY DESCRIPTION
Most display contains the following sections or features:
Time
Batt
Plan
Mode
Main
Display
Screen
F1
F2
F3
Figure 2-1
10
Using
The Instrument
3.1 INTRODUCTION
Now you can use the DS1151B according to the
description of chapter 3. Each measurement mode can
be accessed via buttons on the keypad, and it will be
discussed with detailed descriptions on how to perform
the measurement. But before the test operation, It is
important to make the meter to be compatible with your
cable system by learning your user plan at first.
3.2 LEARN USER PLAN
The measurement is performed depending on proper
user plan. It includes three elements as following.
*Channel number
*Channel format (TV, single freq, DIGI)
*Carriers (video and audio)
DS1151B can learn up to five user plans depending on
the cable system and storing it as built-in plans. To build
11
a user plan by operating step-by-step as following.
1. Connect the instrument with the cable system.
2. Operating as following:
SET+ CHANNEL PLAN
The screen displays as figure 3-1
Figure 3-1
Highlight SELECT USER PLAN option, and press
UP or DOWN button to select a user plan (A, B, C, D,
E) which will be built.
3. Select LEARN USER PLAN option, the LCD
displays as figure 3-2.
12
Figure 3-2
4. Press F1 to enter the next menu with the instrument
connected to the cable system. Then select a base
channel plan to build the user plan. Refer to figure
3-3.
Figure 3-3
13
DS1151B is capable of identifying which channels are
on the cable plant. It takes it easy to quickly generate a
customized user plan. After above the new user plan
will be saved automatically. For more details, you can
view the section 4.5.
3.3 LEVEL MEASUREMENT
Press LEVEL button to enter LEVEL mode. The level
test is able for analog TV, digital channel, single
frequency channel and frequency level measurement.
ANALOG CHANNEL LEVEL
The video level, audio level and Δ V/A can be
measured and displayed in data.
Figure 3-4
14
An analogue channel can be defined in FAST SETUP
mode by pressing SET button in LEVEL mode. And the
TV indicator will be shown on the right top of the
screen in channel LEVEL mode.
DIGITAL CHANNEL LEVEL
DS1151B is capable of measuring the average power of
digital channel with QAM or QPSK modulation. A
digital channel with the center and BW can be defined
in FAST SETUP mode by pressing SET button.
Figure 3-5
FREQUENCY CHANNEL LEVEL
You can define a frequency channel in FAST SETUP
mode with specified center and span, the peak level in
the frequency span can be measured and displayed in
large size as figure 3-6.
15
Figure 3-6
The delta frequency and delta level between peak level
point and center frequency point will be shown in small
size on the bottom of the screen.
MEASUREMENT OF PEAK-HOLD LEVEL
You can read out the peak-hold level in data by pressing
F2(PEAK) button. The instrument will start to calculate
the peak-hold value and display it with P indicator.
MEASUREMENT OF PEAK-TO-PEAK VALUE
The peak-to-peak value can be calculated and displayed
by pressing F2(Δ p-p) key with Δ indicator shown.
TUNING BY CHANNEL
When you tuned the meter to a channel, the channel
number, type and label of the tuned channel will be
16
displayed on the top of the screen. Use the UP and
DOWN buttons, you can tune the channel according to
the selected user plan. To tune to a desired channel
directly, you can press the alphanumeric keys directly
after pressing the 2nd key. The standard(alphanumeric)
channel number and EIA(numeric) channel number are
enabled for the meter.
EDITING USER PLAN
The user plan is a critical element to operating the meter.
To make the meter to be more compatible with your
cable system, you can edit the channel of the selected
user plan. Press SET button in channel level mode.
Figure 3-7
The EIA and standard channel number, label, type and
carrier frequency(BW) can be edit, and the
17
measurement of the channel can be disabled or enabled
by selecting the ACTIVE STATUS item. For more
details, refer to the section 4.5.
TUNING BY FREQUENCY
The F1 key toggles between channel and frequency
level mode. The UP and DOWN buttons can be used to
tune the measured frequency according to step size
specified in fast SETUP mode. The step size is ranging
from 10kHz, 100kHz, 1MHz, 10MHz or channel step.
You can use numeric keys to enter the desired frequency
directly.
Figure 3-8
Figure 3-8
The built-in speaker will work depending on the setup.
The volume can be tuned by 4 levels, or mute.The
18
FAST SETUP mode can be accessed by pressing SET
button in FREQUENCY tuned mode.
Figure 3-9
REFERENCE LEVEL AND SCALE
Press F3 to enter the next soft key menu, the LCD
display as following:
Figure 3-10
The scale of analog graph can be adjusted cycled from
1,2,5,10dB. To adjust the reference level, press
F2(REF), the button icon will be displayed as REF. Use
the UP and DOWN keys to adjust the reference level of
analog graph. To exit from the adjusting reference level
mode by pressing F2 button again.
19
SINGLE CHANNEL SPECTRUM
To observe the spectrum of single channel or frequency,
press LEVEL button again, the spectrum will be
displayed with the specified center and span. The
channel number will be shown on the bottom of the
spectrum graph. If the LEVEL button is pressed in
single frequency mode, the channel number will be
displayed as "----", and the center is defined same as the
tested frequency.
To tune the center by moving the marker to the end of
the viewing range. The level and frequency of marked
point can be read out in data immediately with the
moving of the marker. Refer to figure 3-11.
Figure 3-11
To adjust the reference level and scale by pressing
20
F3(▼) to enter the next soft key menu.
Figure 3-12
Press F1(scale) repeatedly to change the scale ranging
from 1,2,5,10dB. Press F2(REF) to enter reference
level adjusting mode with the icon being shown as REF.
Adjust the reference level by pressing UP and DOWN
keys. Press F2(REF) again to exit from the mode.
3.4 FULL SCAN
To quick test the flatness and amplitude of all carries in
the cable system by pressing SCAN button, a spectral
graph of the selected user plan is displayed as following.
Figure 3-13
21
The meter will scan the full span of the cable system
according to the selected user plan.
The measurement data of up to 120 channels will be
displayed in the viewing range. To extend the viewing
range by moving the marker to the end of the viewing
range. You can read out the level and channel number
with the moving of the marker.
The audio level and video level will be displayed by one
bar with difference contrast. The contrast of video level
graph is darker than audio level.
To freeze scanning process by pressing F2(STOP)
button. The graph is retained even if the cable is
disconnected with the meter. To resume scanning, press
F2 (TRIG). To move the vertical marker to the desired
channel, you can use the UP and DOWN keys, or tune
directly by using alphanumeric keys in 2nd keypad.
The measurement data and channel number of the
marked channel will be displayed. To read out the
frequency of the marked channel, press SET button to
enter the FAST SETUP of SCAN mode and set the
MARKER DISPLAY option to be “FREQ”.
22
THE REFERENCE LEVEL AND THE SCALE
The reference level and the scale appear on the top of
the spectral graph. To adjust the two parameters quickly
by pressing F1(AUTO) button, the reference level and
scale will be adjusted automatically to get optimal
scope.
Press F3(▼) to enter the next soft key menu, you can
prefer to set the reference level and scale. The desired
scale (1,2,5,10dB) can be selected by pressing F1(scale)
repeatedly. To tune the reference level by pressing
F2(REF), the icon REF will be shown, and the
reference level can be adjusted by using UP and DOWN
keys. Press F2(REF) again to exit from the reference
level adjusting mode.
ZOOM IN AND OUT
The DS1151B supports 5 levels of magnification. To
zoom in or out by pressing SCAN button repeatedly.
The spectral graph will zoom in on or out from the
marked channel.
LIMIT DISPLAY
Enter the FAST SETUP mode of scanning function by
pressing SET button.
23
Set the LIMIT DISPLAY item to be ON status, the limit
will be displayed in spectral graph as shadow. You can
check that the carrier levels are within limits. Refer to
figure 3-14.
Figure 3-14
EDIT LIMIT
You can set the parameters of limit by selecting LIMIT
EDIT option in FAST SETUP mode. The minimum
video level (40dBµV to 120dBµV), maximum video
level (40 to 120dBμV ), maximum delta video level
(2dB to 30dB), maximum V/A(5dB to 30dB) and
minimum △V/A(0dB to 15dB) can be selected by
yourself. Refer to figure 3-15.
24
Figure 3-15
After all parameters is determined, select SAVE AND
EXIT to save the setup and return to the previous menu.
3.5 FREQUENCY SPECTRUM
The DS1151B is capable to measure the frequency
spectrum with spans ranging from 2.5 to 62.5 MHz. A
special full span can be set, and the DS1151B does as
the spectrum measurement according to the selected
user plan. The amplitude of each video carriers will be
sampled and displayed.
To do the absolute spectrum measurement by pressing
SPECT button with transmission option being set as
“OFF”. The screen displays as figure 3-16.
25
Figure 3-16
RELATIVE MEASUREMENT
The DS1151B enables you to do the relative spectrum
test. The loss insertion for cable, branch line, distributor,
amplifier and filter can be test simply through the
function. The DS1151B utilizes the signal of your CATV
system or noise generator to achieve this.
The TRANSMISSION option needs to be set as “ON”
status to perform the relative spectrum measurement.
Scroll to MEASUREMENT line in SETUP mode. The
screen displays as following by pressing F1(ENTER).
Scroll to TRANSMISSION line and set the option as
“ON” status by pressing UP and DOWN buttons. The
‘ON” indicates that the relative spectrum measurement
is enabled once pressing the SPECT key. Otherwise the
absolute spectrum measurement is enabled.
26
Figure 3-17
Once selecting the “ON”, the screen shows as fig3-18.
Figure 3-18
You have the option to build the new spectrum reference
data by pressing F1(YES). To use the original zero
reference data by pressing F2(NO). If you select to
27
build a new zero reference data by pressing F1(YES).
The DS1151B displays following prompt:
Figure 3-19
To build a new zero reference data according to the
signal of CATV system, you need to connect the cable
with the RF input port of the meter. You may have a
noise generator (such as DS5032). You can connect the
noise generator with the meter and power on the noise
generator.
Press F1(ENTER), the DS1151B scans the amplitude
automatically according to the video carriers of
selected user plan. The span of spectrum measurement
will be set as full span of selected user plan. After
building the new zero reference data, the DS1151B
displays the following prompt:
28
Figure 3-20
Press F1(ENTER), the DS1151B will perform the
relative spectrum measurement according to the new
zero reference data saved in memory.
Figure 3-21
The “REF” indicates △dB to show the delta level
between the reference line and the zero reference data.
29
The relative measurement level of each frequency point
can be read out in data by moving the marker line.
NOTE:
The span can not be changed in relative spectrum
measurement mode.
The zero reference data remains in the memory even
when turn off the meter. You can do a new relative
spectrum measurement anytime by pressing SPECT
button. To resume the absolute spectrum measurement
by putting the TRANSMISSION option into OFF status.
CENTER AND SPAN
To change the center frequency directly by pressing 2nd
button to put the keypad into alphanumeric keypad,
directly input the desired center and press F1(ENTER)
to confirm the entry. The center frequency can be
increased or decreased by moving the marker line to the
end of the viewing range.
You can press F2(SPAN) to change the sweeping
bandwidth. The span cycles through 2.5, 6.25, 12.5, 25,
62.5M and full span. When the full span is selected, the
span is displayed as the label of selected user plan (A,
30
B,C,D or E), the start frequency will be displayed
instead of center frequency, and it can not be changed.
FREEZE SWEEPING
You may want to freeze the sweeping to study the graph,
simply press F1(HOLD), the sweeping process will be
stopped on current sweeping point. To resume the
sweeping process, press F1(TRIG).
REFERENCE LEVEL AND SCALE
To adjust the reference level and scale by pressing
F3(▼) to enter the next soft key menu.
Figure 3-22
You can change the scale by repeatedly pressing the
F1(SCALE) button. The scale cycles through 1dB, 2dB,
5dB and 10dB. To adjust the reference level, you can
press F2(REF) button, and the soft key icon is displayed
as REF .In this mode, the UP and DOWN keys can be
used to tune the reference level, and the graph will be
redrew with the adjusting. To exit from the mode by
pressing F2(REF) again.
You may wish to set the reference level and the scale
31
automatically, simply press F3(▼▲) in figure 3-21.
The new soft key menu displays as figure 3-23. To
automatically adjust the reference level and scale by
pressing F1(AUTO).
Figure 3-23
MARKER AND DELTA MARKER
The amplitude and frequency of each measured point
can be read out in data through the marker line. To tune
the marker by UP and DOWN arrow buttons. You can
also set a delta marker to read out the delta value of two
specified frequency points.
To set a delta marker, switch the soft key menu to figure
3-23 by repeatedly pressing the F3 button. Move the
marker line to the desired frequency point at first, press
F2(△MKR), a delta marker line will be active and be
shown on the point which the marker is placed. Now
you can move the marker line to another desired
frequency point by using UP and DOWN buttons, and
the delta frequency and delta level between the two
frequency points can be shown on the bottom of the
screen.
32
Figure 3-24
PEAK-HOLD LEVEL MEASUREMENT
You can do the peak-hold measurement in SPECTRUM
mode by pressing SPECT key again.
Figure 3-25
The "PEAK" will be displayed in the title line. The
33
peak-hold curve will be displayed with the real test one.
The peak-hold curve will be shown in darker wave, and
the measurement data of peak-hold curve can be read
out
though
the
marker
in
PEAK-HOLD
MEASUREMENT mode.
To make a fresh peak-hold curve by pressing SET
button to enter the FAST SETUP mode. Select the
FRESH P-H DATA option, the original peak-hold curve
will be deleted, and the DS1151B will restart calculating
the new peak-hold data.
3.6 TILT/LEVEL LIST
To fast check the flatness and the gain distribution of the
cable system, press TILT button to enter the
TILE/LEVEL LIST mode. You can fast test up to twelve
specified channels and read out the measurement data in
spectral graph or data.
TILT MODE
Press the TILT button, the screen displays as figure
3-25. All of the tilt channels will be sampled and be
displayed as spectral graph. The level of the lowest and
highest channel is displayed. And the tilt value is
displayed on the bottom of the spectral graph.
34
Figure 3-26
Before do the TILT measurement, at least four tilt
channels must be enabled, Otherwise the following
prompt will pop out.
Press F1(SETUP) to set the tilt channels. The F2(▼)
and F3( ▲ ) buttons can be used to highlight the
channels, and resume the selection by pressing
F1(ENTER). The "√" indicator will be shown to mark
the enabled channel. To disable the channels, highlight
the channel and press F1(ENTER) again, the "√" will
disappear. The channel number of the enabled channels
will be shown in the bottom box according to the
frequency carriers.
35
Figure 3-27
After the setup, press TILT button to start the TILT
measurement, and the TILT setup will be saved
automatically.
The scale can be changed by repeatedly pressing
F3(SCALE) button. To adjust the reference level of
graph, press F2(REF) button to enter the adjusting
reference level mode. Press UP and DOWN buttons to
increase or decrease the reference level. To exit from the
mode by pressing F2(REF) button again.
LEVEL LIST MODE
The TILT button toggles between TILT mode and
LEVEL LIST mode. You can read out the measuring
level of TILT channels in data.
36
Figure 3-28
3.7 LIMIT TEST
The DS1151B is capable of fast checking the cable
system by the specified test limits.
Figure 3-29
37
Press LIMIT button, the amplitude of each channel of
selected user plan is checked one by one. The channel
number, video level and audio level will be shown with
the measuring process.
NOTE:
THE DS1151B CHECKS THE ANALOGUE
CHANNEL ONLY IN THE LIMIT TEST MODE.
TEST RESULTS LIST
After the limit test scanning, the general test results of
cable system will be listed as following:
Figure 3-30
The test items are consist of minimum video level,
maximum video level, maximum delta video level,
minimum △V/A, maximum △V/A and maximum
△ADJ channels.
38
To observe the test results of each channel, press
F1(LIST), the screen displays as following.
Figure 3-31
Highlight the desired channel by F2 and F3 buttons,
and press F1(ENTER) to observe the limit test detail.
Figure 3-32
39
The video level, audio level and △V/A of highlighted
channel are shown on the top of the screen, and the limit
test results of each test item are displayed on the bottom
of the screen.
EDIT LIMIT
You can edit the limit setup in SETUP mode. To enter
the limit editing mode by pressing SET+
MEASUREMENT+ LIMIT SETUP +EDIT LIMIT
The screen displays as following:
Figure 3-33
Select the limit test item by using F2 (▼)and F3
(▲)buttons, the highlighted limit item can be enabled or
disabled by pressing F1(ENTER). The "√" indicates
that the limit test item will be performed in limit test
mode. You can input the desired limit value in 2nd
keypad, or by UP and DOWN buttons.
40
3.8 AUTO TEST
The DS1151B can be set to perform automated level
measurement sequentially. The measurement process
can be programmed and saved in a project. At most
seven projects can be programmed and stored. Once you
running a project, the programmed test item will be
executed sequentially, the measurement data can be
shown on the screen, and after all the measurement data
will be stored into a file automatically.
CONFIGURING AN AUTO TEST
You can enter the auto test menu by pressing AUTO
button. The existing projects are listed on the screen.
The DS1151B can store up to seven projects. You can
highlight the desired project by pressing UP and DOWN
buttons. The highlighted project can be executed by
pressing F1(START).To delete a highlighted project by
pressing F2(DEL).
You may wish to study the measurement process of the
highlighted project, simply press F3(INFO), the
measurement items of the selected project will be listed
as figure3-34. The UP and DOWN buttons are used to
select the item, The test details can be shown by
pressing F1(ENTER). To return to the previous menu
by pressing F3(BACK).
41
Figure 3-34
The screen displays as figure3-35 when moving the
cursor down to a blank row. Your DS1151B will guide
you step by step to build a new project.
Figure 3-35
Be sure to press F1(NEW) to create a new project. You
must name the new project at first. Press F1(NAME) or
42
2nd button to put the keypad into alphanumeric keypad
mode and input the name of the project.
Figure 3-36
After determining the name of the project, press
F2(NEXT) to perform the step 2.
figure 3-37
The auto test includes four types measurement LEVEL
43
TEST, FREQUENCY SPECTRUM, LIMIT TEST and
TILT measurement.
The " √ " indicates the measurement that can be
executed in auto test process. Select the desired
measurement type by UP and DOWN buttons, and
disable or enable it by pressing F1(ENTER) key. To
enter the STEP 3 by press F2 (NEXT)
Figure 3-38
The enabled measurement items will be listed on the
screen. Select and highlight the measurement item by
pressing UP and DOWN buttons and press F1(ENTER)
key to edit the test procedure.
You can select which channel can be tested in LEVEL
item of auto test procedure. Move the cursor by UP and
DOWN buttons and disable or enable the measured
channels by pressing F1(ENTER). The channel marked
by "√" will be tested in LEVEL item of auto test
44
procedure. Be sure to confirm the entry by pressing
F3(BACK) button.
Figure 3-39
Select SPECTRUM in figure 3-38,you can configure the
center frequency and bandwidth of spectrum item.
Figure 3-40
45
Input the center frequency by 2nd keypad. The
bandwidth can be set by F1(ENTER) button. To
confirm the entry by pressing F3(BACK) key.
Select the LIMIT item in figure3-38, the meter will
enter the limit setup menu. Press F1(ENTER) to set the
limit automatically according to the FCC standard when
the SET FCC DEFAULT option being highlighted.
Figure 3-41
Use the F2(▼) and F3(▲) to move the cursor. The
enabled limit test items are marked by "√" indicators.
To disable it by pressing F1(ENTER). The limits can be
increased or decreased by pressing UP and DOWN keys.
To confirm the entry by pressing F3(BACK) button.
To set the measured tilt channels of auto measurement
by selecting TILT item in figure 3-38. The screen
46
displays as following, confirm the entry by pressing
F3(BACK) button.
Figure 3-42
After configuring the auto test, you can start the auto
test procedure by pressing F1(START).
Figure 3-43
47
The auto test results will be stored as a file. You may
wish to save the programmed project only, simply start
the auto test procedure by pressing F1(START), then
exit from the auto test procedure by pressing F1(EXIT),
then the programmed project will be stored with the
input name.
If your DS1151B is capable of automatically measuring
for 24 hours (option), you can press F2(24H) in figure
3-43 to do the setup of auto measurement for 24 hours.
PERFORMING THE AUTO TEST
To execute a stored project by pressing AUTO button.
Refer to 3-44.
Figure 3-44
You can highlight the desired project by pressing UP
and DOWN buttons and start the auto test procedure by
48
pressing F1(START).
The programmed measurement item will be executed
and displayed the test procedure on the screen. When
the auto test procedure is completed, the test results will
be stored automatically, and your DS1151B will display
a list of stored files of auto test procedures that are
available for viewing and printing.
VIEW THE AUTO TEST RESULTS
The auto test results is saved as a file with "*" indicator
displayed beside the file name. The files can be recalled
by pressing FILE button. The existing files will be
listed with the file name, date and time.
Figure 3-45
Highlight the desired file by pressing UP and DOWN
49
buttons, press F3 (LOAD) to view the stored contents.
Figure 3-46
The file name, user plan and measurement items
which have been executed will be listed. The auto test is
executed according to the user plan showed in the
screen. You can select the desired measurement item by
pressing UP and DOWN buttons, and study the
measurement data by pressing F1(LIST) or F2(VIEW)
button. If a proper printer is connected with the
instrument, the measurement data of the desired file can
be printed immediately by pressing F3(PRINT) button.
The F2(VIEW) button can be used to view the
measurement graph of spectrum measurement or limit
test when the correlated item being highlighted. To exit
the viewing mode of figure 3-47 by pressing FILE
50
button on the keypad. The DS1151B will return to
figure 3-46.
Figure 3-47
You can list the test data by pressing F1(LIST) key. The
figure3-48 is the list of LIMIT TEST item.
Figure 3-48
The stored spectrum data can be listed as figure 3-49. To
51
return to figure3-46 by pressing F1(BACK).
Figure 3-49
The DS1151B lists the TILT MEASUREMENT data as
figure 3-50.
Figure 3-50
The LEVEL MEASUREMENT data is listed as figure
52
3-51.
Figure 3-51
3.9 SAVE AND LOAD
DS1151B supplies 32K bytes memory space to store the
measurement data. The measurement data of LEVEL
mode, SCAN mode, TILT, LIMIT TEST, and frequency
SPECTRUM can be saved in one file. The system
distributes difference memory size for file that includes
difference measurement data. Up to 110 files can be
saved in the DS1151B depending on the size of saved
files.
THE FILE DIRECTORY
Press FILE button to enter the menu, you are presented
with a directory of existing files.
53
Figure 3-52
The stored files are labeled by file names, date and time.
The file which is saved in AUTO TEST procedure is
marked with “* ” indicator.
You can highlight the desired file by using UP and
DOWN buttons. To delete the highlighted file, press
F2(DEL) button, and the system will release the
distributed memory space of the deleted file.
STORING A FILE
Press F1(NEW) button to store the current measurement
data into a file. A screen will appear to enable you to
define the file name. The current selected user plan and
current measurement data that can be stored will be
displayed. Refer to figure 3-53.
54
Figure 3-53
The file name can be defined by using alphanumeric
key after pressing 2nd key. Use F2(▼) and F3(▲) to
scroll to the desired measurement data to which you
want to save. The ON/OFF label appears with the
measurement data to flag the storing selection, and the
label can be changed by pressing UP and DOWN
buttons. The measurement data marked by ON can be
saved. To store the file by pressing F1(OK). Before
storing, be sure to enter the file name.
NOTE:
1.Only the test data of the measurement that has
been performed can be displayed and saved.
2.At least one measurement item must be enabled,
otherwise the saving operation can not be
performed.
55
VIEWING A FILE
You are able to recall a stored file and view or print out
it, and adjust the display setting the same as viewing a
real measurement. Use the UP and DOWN buttons to
scroll to a file line to which you wish to view, and press
F3(LOAD) to view the file. See the section3.8 to find
out how to access the existing files.
NOTE:How to save the current status?
If you would like to enter the precious measuring
menu with the expected setup when switching on
again, please PRESS “FILE” KEY TWICE to save
it before power off.
56
3.10 VOLTAGE METER
To perform the voltmeter function by pressing
SET + MEASUREMENT + VOLTAGE
Figure 3-54
In the interface the battery energy and the coaxial cable
voltage can be tested at the same time. Each voltage can
be shown as a column graph. The battery energy
displays as a voltage. When the battery voltage is lower
than 3.2V, a prompt will be displayed, the instrument
will automatically shut down after a while. The voltage
of the coaxial cable is tested with the AC/DC judging
automatically.
3.11 C/N
Connect the cable system with the instrument and press
the C/N key to enter the C/N mode. The carrier level
and the C/N ratio will be tested and displayed on the
57
screen. To improve the C/N measurement accuracy, you
can shut off the carrier at this time, and the instrument
will test the noise level according to the status and get
the new C/N ratio.
Figure 3-55
To test the C/N ratio, the input level of carrier must be
higher than 85dBμV.
3.12 PRINT
The meter provides you the function of printing the data
you have saved and tested in the LEVEL, SCAN,
SPECTRUM and TILT mode. You can review the result
through the printing report. And the system can print
files by several models of printer. The printer you want
to use should be set in the SETUP mode. You can select
one of 4 types of printers, includes CANON BJ-N,
58
LQ1600K, LQ300 and HP-III. Refer to figure 3-56.
Figure 3-56
The more printer supporting software will be supplied
by DEVISER, the user can download the new software
form the website www.deviser.com.cn , and upgrade the
software of the instrument through the communication
port using the data cable and PC.
To connect to the serial printer (such as EPSON
LQ300),you
need
a
serial
printer
cable
(5pin~25pin,P.310075L25).
To connect to the parallel printer, a data cable , parallel
printer adapter (option)and a standard printer cable are
needed. For more connecting details, you can view the
PRINTER SETUP section of chapter 4 and 6.
59
TILT
DS1151B
REPORT
SERIAL NO. 000001 PRINT DATE:2001/01/02
PRINT TIME: 05:02 SAVE DATA: 2001/01/02 SAVE TIME: 04:53
FILE NAME:123
CHANNEL PLAN: A
PLAN NAME: CHINA
SPARE SPACE: 94.5% BATTERY ENERGY: 3.8V TEMP:21.8℃
**************************************************
<TILT>
UNIT: dBμV
CHAN
TYPE
LEVEL
1
TV
5
TV
69.8
79.8
11
TV
89.0
122
TV
83.2
60
4.1 INSTRUCTION
You can easily configure the DS1151B by using the
SETUP. To access the SETUP mode by pressing SET
button.
The DS1151B has two SETUP mode, one is FAST
SETUP mode, another is FULL SETUP mode. The
FAST SETUP mode can be entered by pressing SET
key in the measurement mode. The items of FAST
SETUP mode is correlated with the specified
measurement mode.
To enter the FULL SETUP mode, you may press the
SET button for up to two times, this brings up the FULL
SETUP menu as figure4-1.
Use F2(▼) and F3(▲) to scroll to the option to which
you desired. Press F1(ENTER) to select the display.
61
Figure 4-1
General information of the
DS1151B, includes manufactory, version and
the setup status.
*GENERAL: The setup includes backlight, data
and time, power management, display
contrast and printer selection.
*MEASUREMENT: The setup for measurement
includes level units, limit setup, voltage
measurement, relative spectrum measurement
and some other.
*CHANNEL PLAN: The setup for channel plan
includes user plan selection, learn and edit
user plan and tilt channel setup.
*INFORMATION:
4.2 INFORMATION
Display the information of the instrument. The
62
information page can be page up and down by pressing
F2(▼) and F3(▲) buttons to view the more details.
Figure 4-2
Repeatedly press the F2(▼), the display lists the setup
of the DS1151B in three pages.
Figure 4-3
63
4.3 GENERAL
Use the GENERAL setup to configure the general
parameters of the DS1151B.
Figure 4-4
The active parameters of the highlighted option are
displayed on the bottom of the screen.
BACKLIGHT
The backlight function is designed to satisfy the
difference requirement. You can set the backlight to be
always on or to be shut off anytime. And the backlight
can be set and work in intelligence power saving mode.
It can be set to shut off after specified time for the
inactive keypad, and the backlight will be turned on
once any key being pressed.
64
Figure 4-5
To configure the backlight parameter by pressing UP
and DOWN buttons. You may set the backlight
parameter to be ON, OFF, 3 minutes, 5 minutes, 10
minutes and 30 minutes.
LCD CONTRAST
The LCD contrast is difference in difference
temperature environment. To get the optimal viewing,
you may adjust the LCD contrast by pressing UP and
DOWN buttons.
Figure 4-6
SHUTDOWN TIME
To save the power, the instrument can be set to
shutdown automatically for inactive keypad after 3
minutes, 5 minutes, 10 minutes or 30 minutes. For some
special measurement that will be performed for long
time, the SHUTDOWN TIME option must be disabled
by setting the option to be “ON”.
65
Figure 4-7
If the shutdown function is enabled, after the specified
time with inactive keypad, a prompt will be displayed
ahead of 20 seconds before the shutdown of the
DS1151B. And the measurement process will be
stopped.
Figure 4-8
You can read out the remaining time in seconds. And the
setup of SHUTDOWN is displayed. Press any other
buttons except UP and DOWN on the keypad to recall
the original display. Of course you may change the
SHUTDOWN parameter efficiently by pressing UP and
DOWN keys.
66
TEMPERATURE UNIT
You can read out the internal temperature of the
DS1151B in centigrade or Fahrenheit unit specified by
the TEMPERATURE UNIT option. Use the UP and
DOWN buttons to adjust the setup.
Figure 4-9
DATE &TIME
Highlight the option and press F1(ENTER) to set the
date and time.
Figure 4-10
You can adjust displaying order of the date to any
desired order (Year/Month/Date, Month/Date/Year or
Date/Month/Year) by pressing UP or DOWN button
when the date option is highlighted.
67
Scroll to the option to which you wish to set by pressing
F2(▼)and F3(▲) buttons. To input the desired date and
time by pressing numeric keys in 2nd keypad.
NOTE:
The first two digits of year is fixed as 20, you only
need to input the last two digit of year.
PRINTER SETUP
You can select the printer type in the screen.
Figure 4-11
The enabled printer can print out the stored files in
report format through the communication port. The
DS1151B supports 4 models of printer as following:
68
PRINTER
MODEL
HARDCOPY
PORT
MANUFACTORY
LQ1600K
PARALLEL
EPSON
LQ300
SERIAL
EPSON
BJ-N
PARALLEL
CANON
HP-III
PARALLEL
HP
The new software which can support more printers will
be supplied by Deviser. You can download the software
from the Deviser website (http://www.deviser.com) and
upgrade the meter through the communication port by
using the data cable (P.3100805L9) and the PC.
To connect to the serial printer (such as LQ300), you
need a serial printer cable (Deviser part number
P.310075L25). To connect to the parallel printer, a data
cable, parallel printer adapter (option, part number
0115005) and a standard printer cable are needed. For
more details, you can view the chapter 6.
LCD TEST
The LCD can be tested through the option. You can
press F1(ENTER) repeatedly to make the screen display
69
as all white, all black, vertical stripe and checker board
step by step.
4.4 MEASUREMENT
The measurement parameters can be accessed and set in
the display.
Figure 4-12
TRANSMISSION
The frequency spectrum measurement can be performed
to test the transmission character of the circuit parts or
your system through relative measurement set by
TRANSMISSION option.
Figure 4-13
70
When the option is set as ON, the relative spectrum
measurement can be performed in SPECT mode. The
zero reference data can be built up according to the
prompt displayed in the screen when setting the option
as ON. When the option is set as OFF, the absolute
spectrum measurement can be performed in SPECT
mode. For more details, you can view the section 3.4.
LEVEL UNITS
You can set the DS1151B to test for dBμV, dBmV or
dBm by pressing UP and DOWN buttons.
Figure 4-14
FREQ TEST SETUP
FREQ TEST SETUP menu is same as the FAST SETUP
of FREQ LEVEL mode, refer to figure 4-15.
You can use the UP and DOWN buttons to adjust the
frequency tuning step size ranging from 10KHz,
100KHz, 1MHz, 10MHz, or setting the step size as
channel step of current selected user plan.
The volume of built-in speaker can be increased or
71
decreased by pressing UP and DOWN buttons. To shut
off the speaker when the level is too low in FREQ
LEVEL mode, you can configure the MUTE FOR
NOISE option as YES status.
Figure 4-15
SCAN SETUP
The SCAN SETUP displays as the FAST SETUP of
SCAN measurement mode. It includes LIMIT SETUP
and MARKER DISPLAY options. Refer to figure 4-16.
The limits can be displayed as shadow in SCAN graph
by setting the LIMIT DISPLAY option in YES status.
You can set the marked channel to be labeled as
channel number or frequency in SCAN mode by
configuring the MARK DISPLAY option as “CHN” or
“FREQ”.
72
Figure 4-16
To edit the test limits, you can highlight the EDIT
LIMIT option and press F1(ENTER), the screen
displays as figure 4-17.
Figure 4-17
The test limits can be set according to FCC requirement
73
by selecting SET FCC DEFAULT option and pressing
F1(ENTER).
Item
Limit
63dBμV
Minimum Video level
78 dBμV
Maximum Video level
10 dB
Maximum delta Video level
10dB
Minimum △ V/A
17dB
Maximum △ V/A
3dB
Maximum △ ADJ
FCC standard
To set the test limit as you desired, you can scroll to the
correlated limit item to which you wish to adjust by
pressing F2(▼)and F3( ▲ ) buttons, and increase or
decrease the limit value by using UP or DOWN keys.
After completing the setup of test limits, be sure to
confirm the entry by selecting the SAVE AND EXIT
option and pressing F1(ENTER).
NOTE:
THE LIMIT TEST CAN BE PERFORMED TO
CHECK THE ANALOG CHANNELS ONLY. THE
DS1151B CAN’T DO THE LIMIT CHECK FOR
OTHER TYPES OF CHANNEL.
74
For more details of LIMIT SETUP by viewing EDIT
LIMIT of section 3.4 and EDIT LIMIT of section 3.7.
FRESH P-H DATA
When performing the FREQUENCY SPECTRUM
measurement, the DS1151B calculates the peak-hold
level and displays the peak-hold level waveform in
peak-hold level measuring mode. To fresh the peak-hold
data of spectrum measurement, you can highlight the
option and press F1(ENTER) to confirm the operation.
4.5 CHANNEL PLAN
The DS1151B performs the measurement according to
the channel plan configuration of the user plan.
Figure 4-18
75
It is the most important setup of the DS1151B to
measure the cable system quickly and efficiently. The
screen displays as figure 4-18 when entering the channel
plan setup menu.
SELECT USER PLAN
Up to five user plans can be built and saved in the meter
to support the measurement for difference cable system
plant. The user plans are labeled as A, B, C, D, E. The
current selected user plan can be determined by pressing
UP and DOWN buttons. Refer to figure 4-19.
Figure 4-19
The base channel plan which is used to built the user
plan will be displayed on the top right of figure 4-19.
The label of the current selected user plan is displayed
in the title line on the top of the screen, and it will be
changed as well as the setting.
Figure 4-20
The meter will do the measurement according to the
selection of the user plan.
76
NOTE:
THE METER WILL LOAD THE DEFAULT USER
PLAN WHEN SELECTING THE USER PLAN
THAT HASN’T BEEN BUILT AND SAVED.
CHAN NUM DISPLAY
You can set the channel number to be displayed in
EIA(numeric) mode or STANDARD(alphanumeric)
mode by pressing UP and DOWN buttons with the
CHAN NUM DISPLAY option being highlighted.
Figure 4-21
After the selecting, the DS1151B will show the channel
number as you desired in any measurement mode. The
EIA(numeric) channel number and STANDARD
(alphanumeric) channel number can be entered and
edited in EDIT USER PLAN option.
LEARN USER PLAN
You can build and stored up to five user plans in one
meter. After purchasing the DS1151B you should build
the user plan before the first measuring to make the
meter to be compatible with your cable system.
77
NOTE:
The LEARN USER PLAN option will build the user
plan and overwrite the original current selected user
plan. To built other user plan, you can change the
current selected user plan by SELECT USER PLAN
option, then do the LEARN USER PLAN operation.
Use the following step-by-step procedures to build a
new user plan.
1. Press F1(ENTER) to start the procedure of
learning user plan with the LEARN USER PLAN
option being highlighted. The screen displays as
figure 4-22.
2.
Figure 4-22
Connect the cable with the DS1151B, and press
F1( ENTER) to confirm the operation.
78
NOTE:
The DS1151B will wait for the confirmation for
about ten seconds, then enter the next menu
automatically.
3.
4.
5.
The base channel plans came with the meter are
listed. You can select one from the eight channel
plans to build the user plan.
Figure 4-23
Using F2(▼)and F3( ▲ ) buttons to highlight a
desired base channel plan, and press F1(ENTER)
to proceed the learning procedure.
The DS1151B start to search for active channels
and display the channel number of which channel
79
being checked. The meter will store the new user
plan after completing the procedure, and enter the
LEVEL measurement mode automatically. For
more details you can view the section 3.2.
NOTE:
The channels whose amplitude is higher than
45dBμV will be enabled in user plan. The meter
searches for analog channels only when building
the user plan. If the number of channels whose
amplitude is higher than 45dBμV is less than
three, all channels of the base channel plan will
be enabled to build the user plan.
EDIT USER PLAN
The user plan consist of elements as following:
EIA(numeric) channel number
Standard (alplanumeric) channel number
Channel Label
Channel type (Analogue, DIGI, SINGLE FREQ)
Carrier frequency
Activation status
Each element can be edited manually. An accurate user
80
plan is essential for testing the cable system.
Enter the EDIT CHANNEL PLAN screen by pressing
F1( ENTER) in figure4-18 when the EDIT CHANNEL
PLAN option being highlighted, the screen displays as
following:
Figure 4-24
Highlight the desired channel and press F1(ENTER) to
edit the channel. Refer to figure 4-25.
Highlight the desired item by pressing F2(▼)and F3(▲)
buttons and enter the desired value by UP, DOWN or
alphanumeric keys.
You can enter the label of channel as “B4U” by
alphanumeric keys, and the label will be displayed in
81
channel level measurement graph. Be sure to store the
entry by performing the SAVE AND EXIT item.
Figure 4-25
For more details, you can viewing the EDIT USER
PLAN of section 3.3.
TILT/LEVEL LIST
Up to twelve channels can be tested in TILT/LEVEL
LIST measurement mode. To set the tilt channels you
can press the F1(ENTER) when the TILT/LEVEL LIST
option being highlighted, or press SET button in
TILT/LEVEL LIST measurement mode. Refer
figure4-26.
Highlight the desired channel by pressing F2(▼)and
F3(▲)buttons, and enable or disable the channel by
82
pressing the F1(ENTER) button. The “√” indicator is
shown to mark the enabled channel.
Figure 4-26
The DS1151B will assign the channel and display the
channel number in the bottom box of the screen by the
order according to the frequency.
NOTE:
IF THERE IS TWELVE CHANNELS HAVE BEEN
ENABLED, THE ENABLING OPERATION WILL
BE INACTIVE.
For more details you can view the TILT/LEVEL LIST
section 3.6.
83
LOAD DEFAULT
You may wish to recall the manufactory setup to correct
the error configure caused by error operation. Select the
option and press F1(ENTER), the default setup will be
recalled to overwrite the error setup.
84
The built-in 3.6V 3.5AH Ni-MH battery supplies the
power for the instrument. It can be used continuously
for above 6 hours on condition that it is fully charged.
When the voltage of the battery drops below 3.3V, the
battery indicator in the title line flashes at the top of the
display screen. When the voltage of the battery is lower
than 3.2V, the instrument will shut down automatically,
and you must charge it for about 3 hours.
NOTE:
THE INSTRUMENT IS EQUIPPED WITH Ni-MH
BATTERY, ONLY THE CHARGER(PW320111030)
SPECIFIED FOR THE INSTRUMENT CAN BE
USED TO CHARGE THE BATTERY! THE
INSTRUMENT CAN NOT BE CHARGED
BEYOND 10 ℃ ~35 ℃ ,OTHERWISE THE
BATTERY WILL BE DAMAGED.
85
The intelligence technology is used in power
management. For difference work mode, the difference
power managing method is used to save the power. Part
of the circuits will be shut down in power saving mode.
The working time is above 6 hrs in the highest power
cost mode with full charged built-in battery. When the
voltage of the battery drops below 3.2V, the battery
indicator of title line will flash to indicate the user that
the voltage of the battery is too low and the red light
beside the charge socket will flash. You should charge
the battery immediately. If the voltage of the battery
drops below 3.2V, the instrument will shut off
automatically.
A specified charger is used to charge the battery. To full
charge the battery, the charge time is about 3 hrs. Plug
the DC plug into the charge socket, and power on the
meter, the charging procedure is executed, and the meter
detects the charger type at first, then begin the charging
process. The charging curve and battery energy will be
shown as figure 5-1. If the power adapter is error or the
battery is full charged when begin charging process, the
prompt will be displayed as following.
FULL CHARGED BATTERY
OR ERROR POWER ADAPTER
86
CAUTION
BE SURE THE SCREEN IS DISPLAYED AS
FIGURE 5-1 WHEN CHARGING. OTHERWISE
THE CHARGING PROCESS ISN’T EXECUTED.
Figure 5-1
Figure 5-1
*Battery—B.1100603R5(3.6V,3.6AH Ni-MH battery)
*AC charger—PW320111030
NOTE:
1. USE ONLY THE ADAPTER SUPPLIED WITH
THE INSTRUMENT. ANY OTHER ADAPTER
WILL DAMAGE THE UNIT.
2. THE DS1151B MUST BE TURNED ON TO
PERFORM THE CHARGING OPERATION,
AND THE SCREEN SHOWS AS FIGURE 5-1.
87
The instrument can communicate with a PC or a printer
through the 5 Pin communication port. Refer to figure
6-1.
Figure 6-1
1 ---- TXD
3 ---- NC
2 ---- RXD
.
4 ----+5V
5 ---- GND
Communication software is supplied by manufactory.
Connect the instrument with the PC by data cable and
run the software. You can remotely control the
instrument to do the measurement and the measurement
data will be displayed as graph on the PC monitor.
88
*Data cable—P.3100805L9
(5 pins Din connector to 9 pins D-sub connector)
The windows OS environment is needed to run the
software.
The serial printer such as LQ300 and parallel printer are
supported by the instrument. The measurement data can
be printed out in report format.
NOTE:
ONLY THE EXISTING FILE CAN BE PRINTED.
A serial printer cable is supplied with the instrument.
*Serial printer cable—P.310075L25
(5pins Din connector to 25pins D-sub male
connector)
Connect the Din connector end to the instrument and
D-sub connector to the printer, select the right printer
type, enter the LOAD mode and select the desired file to
begin the printing operation.
To print the report by parallel printer, a parallel printer
adapter and data cable are needed.
*Parallel printer adaptor--0115005
(9pins D-sub connector to 25pins D-sub connector)
89
Connect the instrument with the parallel printer as
following:
Figure 6-2
Select the printer type and saved file to begin the
printing operation.
90
Frequency:
Range:
Accuracy:
Resolution:
5(46)MHz ~ 870MHz
±50×10-6 (20℃±5℃)
10kHz
Channel Type:
Analogue TV: TV
Digital TV:
QAM QPSK
FM channel: SIGL
Level Measurement
Range:
Accuracy:
Resolution:
Input
impedance:
30dBµV~120dBµV
LEVEL±1.5dB (10℃-30℃) (≥35dBµV)
.±3dB (0℃-40℃)
SCAN(±2dB )10℃-30℃
0.1dB
75 Ω (unbalance, BNC or F type
connector)
91
Channel Scan
Number
of channels:
Scanning
speed:
Scale:
Zoom:
150 channels max.
10 channels / s
1,2,5,10 dB/div
1X, 2X, 3X, 4X,5X five levels of
magnification
Frequency spectrum
Bandwidth:
2.5MHz, 6.25MHz, 12.5MHz,
25MHz, 62.5MHz
Scale:
1,2,5,10 dB/div
Carrier noise ratio (C/N):
Input range:
Over 85dBµV (25dBmV)
Range:
20 dB ~50dB max.( depending on
inputting level)
Accuracy:
±2dB
Resolution:
0.1dB
Digital Channel (average) Power
Bandwidth:
0.28 MHz ~9.99MHz
Center frequency:
In the range of the measured
Digital modulation: QAM, QPSK
92
Tilt measurement:
Number of channels: 4~12
Resolution:
0.1dB
Limit line:
Min video
Max video
Max△video
Min△V/A
Max△V/A
Max△ADJ
40 dBµV ~120dBµV
40 dBµV ~120dBµV
2 dB ~30dB
0 dB ~15dB
5 dB ~30dB
0 dB ~20dB
Trunk Voltage measurement:
Input range:
0 V ~100V (AC/DC)
Accuracy:
±1V
Resolution:
0.1V
Others:
Storage:
Communication
Port:
Printer:
Audio Output:
Dimensions:
Weight:
Display:
32K byte
RS 232C
Canon and Epson
Built-in speaker
218mm×95mm×49mm
600g
128×128LCD with backlight
93
Power supply
Battery:
Charger:
Work time:
Charge time:
Accessories:
Charger:
RF connector:
Operation
manual:
Data cable:
Serial printer
cable:
Com Software
Package
3.6V 3.5AH Ni-MH battery
AC 90 V ~240V 50/60Hz
1.8A 7V(max)
Average 6~8 hours
(full charged battery)
Less than 3 hrs
DS10030 Charger 1pc
Type F-F
.
1pc
DK210101185
1pc
P.3100805L9
1pc
P.310075L25
DK401101182
PK100151182
1pc
1pc
94