N2100B PXIT DCA Module

N2100B PXIT DCA Module
N2100B
PXIT Digital
Communications
Analyzer Module
Version 1.02
User’s Guide
Agilent Technologies
ii
Notices
Technology Licenses
© Agilent Technologies, Inc. 2009
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance
with the terms of such license.
No part of this manual may be reproduced in
any form or by any means (including electronic storage and retrieval or translation into
a foreign language) without prior agreement
and written consent from Agilent Technologies, Inc. as governed by United States and
international copyright lays.
Manual Part Number
N2100-90002
Edition
LZW compression/decompression: Licensed
under U.S. Patent No. 4,558,302 and foreign
counterparts. The purchase or use of LZW
graphics capability in a licensed product does
not authorize or permit an end user to use any
other product or perform any other method or
activity involving use of LZW unless the end
user is separately licensed in writing by Unisys.
Portions of this software are licensed under
one or more Open Source licenses, including
version 2 of the General Public License (GPL)
and version 2 of the Library or Lesser General
Public License (LGPL). The source code of the
GPL and LGPL licensed content is governed
under the terms of the respective licenses and
may be found on the CD shipped with the
instrument.
August 2010
Printed in Germany
Agilent Technologies,
Deutschland GmbH
Herrenberger Str. 130
71034 Boeblingen, Germany
Warranty
The material contained in this document is
provided “as is,” and is subject to being
changed, without notice, in future editions.
Further, to the maximum extent permitted by
applicable law, Agilent disclaims all warranties, either express or implied, with regard to
this manual and any information contained
herein, including but not limited to the implied
warranties of merchantability and fitness for a
particular purpose. Agilent shall not be liable
for errors or for incidental or consequential
damages in connection with the furnishing,
use, or performance of this document or of
any information contained herein. Should Agilent and the user have a separate written
agreement with warranty terms covering the
material in this document that conflict with
these terms, the warranty terms in the separate agreement shall control.
Restricted Rights Legend
If software is for use in the performance of a
U.S. Government prime contract or subcontract, Software is delivered and licensed as
“Commercial computer software” as defined
in DFAR 252.227-7014 (June 1995), or as a
“commercial item” as defined in FAR 2.101(a)
or as “Restricted computer software” as
defined in FAR 52.227-19 (June 1987) or any
equivalent agency regulation or contract
clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD
Departments and Agencies of the U.S. Government will receive no greater than
Restricted Rights as defined in FAR 52.22719(c)(1-2) (June 1987). U.S. Government users
will receive no greater than Limited Rights as
defined in FAR 52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as
applicable in any technical data.
Safety Notices
CAUTION
Caution denotes a hazard. It calls attention to
a procedure which, if not correctly performed
or adhered to, could result in damage to or
iii
destruction of the product. Do not proceed
beyond a caution sign until the indicated conditions are fully understood and met.
WARNING
Warning denotes a hazard. It calls attention
to a procedure which, if not correctly performed or adhered to, could result in injury or
loss of life. Do not proceed beyond a warning
sign until the indicated conditions are fully
understood and met.
Trademark Acknowledgements
Microsoft is a U.S. registered trademark of
Microsoft Corporation.
Windows and MS Windows are U.S. registered trademarks of Microsoft Corporation.
Instrument Markings
The CE mark is a registered trademark
of the European Community.
ISM1-A This text denotes the instrument is an
Industrial Scientific and Medical Group 1
Class A product.
ICES/NMB-001 This is a marking to indicate
product compliance with the Canadian Interference-Causing Equipment Standard.
This product complies with the
WEEE Directive (2002/96/EC)
marking requirements. The affixed
label indicates that you must not
discard this electrical/electronic
product in domestic household
waste. Product Category: With reference to
the equipment types in the WEEE Directive
Annex I, this product is classed as a “Monitoring and Control instrumentation” product. Do
not dispose in domestic household. To return
unwanted products, contact your local Agilent
office, or refer to www.agilent.com for more
information.
The C-Tick mark is a registered trademark of the Australian spectrum
Management Agency.
This symbol indicates the Environmental Protection Use Period
(EPUP) for the product’s toxic
substances for the China RoHS
iv
Contents
1 General Information
Introduction 1–2
Electrostatic Discharge Information 1–5
Connector Care 1–7
Returning the N2100B to Agilent 1–15
2 Installation
Introduction 2–2
Upgrading the Instrument Firmware 2–9
3 Configuring the DCA and Acquiring Data
Introduction 3–2
Module Configuration Settings 3–4
Acquiring Data 3–11
Calibrating the N2100B 3–13
4 Measurements
Introduction 4–2
Oscilloscope Measurements Mode 4–5
Non Return to Zero Measurements Mode 4–17
Eye Mask Test Measurements Mode 4–39
Multiple Measurements Mode 4–46
Pattern Sequence D. J. Measurements Mode 4–49
5 Displaying and Saving Results
Introduction 5–2
Program Settings 5–3
Copying and Saving Displayed Test Results 5–13
6 Pattern Acquisition
Pattern Acquisition 6–2
7 Mask & Mask Margin Definitions
Introduction 7–2
How the Mask and Mask Margin are defined 7–3
Agilent Mask Margin Rules 7–6
Tektronix Mask Margin Rules 7–8
Contents-1
Contents
8 Using the Simulator
Introduction 8–2
Configuring the Simulator 8–5
Controlling with User- written Applications 8–7
9 Programming
Introduction 9–2
DLL API Reference 9–7
ActiveX API Methods 9–46
ActiveX API Properties 9–81
10 Specifications
Specifications 10–2
Contents-2
1
Introduction 1- 2
Electrostatic Discharge Information 1- 5
Connector Care 1- 7
Returning the N2100B to Agilent 1- 15
General Information
General Information
Introduction
Introduction
The N2100B Digital Communications Analyzer (DCA) is a PXI based
instrument that automatically performs accurate eye- diagram analysis
to characterize the quality of sources (transmitters) from 1.063 Gb/s to
8.5 Gb/s for production ATE applications. The DCA has five classes of
operation:
•
•
•
•
•
Oscilloscope
NRZ
Eye mask
Multiple measurements
Pattern acquisition and digital filters
Refer to Chapter 4, “Measurements” for more information.
The N2100B implements a coherent vector under- sampling technique
which combines the benefits and measurement capabilities of a real
time scope with the bandwidth of a sampling scope.
Main Features
•
•
•
•
•
Optical input with a 750 to 1650 nm optical wavelength range.
Includes up to four selectable filters in the optical input path.
Electrical input with 1Vpp maximum level.
Input bit rate from 155.52 Mb/s to 11.318 Gb/s.
Internal clock recovery for signals from 155 Mb/s to 2.7 Gb/s; external
clock operation with clock range of 10 MHz to 11.318 GHz.
• Pattern acquisition with software digital 4th- order Bessel- Thompson
filters and no pattern trigger required.
• Optical performance and accuracy guaranteed at the calibrated
wavelength of 850 nm (with offsets applied for 1310 and 1550 nm).
The DCA is a PXI based instrument. Either an embedded PC or a
stand alone PC connected via a remote bridge (such as the NI MXI- 4
product) can be used to control it via the PXI bus. It is also possible
to connect to the instrument via a TCP/IP connection. The module is
supplied with a Windows Control Panel application, an ActiveX inter-
1-2
General Information
General Safety Considerations
face usable from environments such as Visual C++, Visual Basic, and
LabWindows, a programmer's API and libraries callable from user- written applications. Software examples illustrating the use of the module
interfaces are included in the supplied software.
General Safety Considerations
This product has been designed and tested in accordance with IEC
Publication 1010, Safety Requirements for Electronic Measuring Apparatus, and has been supplied in a safe condition. The instruction documentation contains information and warnings which must be followed
by the user to ensure safe operation and to maintain the product in a
safe condition.
1-3
General Information
General Safety Considerations
WA R N I N G
Install the plug-in module according to the enclosure protection provided and placing
filler panels in empty slots. This instrument does not protect against the ingress of
water. This instrument protects against finger access to hazardous parts within the
enclosure.
WA R N I N G
Laser Safety Notice. The N2100B is used to measure optical signals. When
connecting and disconnecting optical cables or equipment, all optical sources MUST
be disabled. Failure to take proper safety precautions may result in eye damage. All
unused optical ports MUST be covered when not in use to prevent light leakage or
contamination.
WA R N I N G
If this product is not used as specified, the protection provided by the equipment
could be impaired. This product must be used in a normal condition (in which all
means for protection are intact) only.
WA R N I N G
No operator serviceable parts inside. Refer servicing to qualified service personnel.
To prevent electrical shock do not remove covers.
WA R N I N G
Use a dry cloth or one slightly dampened with water to clean the external case parts.
Do not attempt to clean internally.
CAUTION
This product is designed for use in Installation Category II and Pollution Degree 2 per IEC 1010
and 664 respectively.
CAUTION
The N2100B is shipped in materials which prevent damage from static. The module should only
be removed from the packaging in an anti-static area ensuring that correct anti-static precautions
are taken.
WA R N I N G
This product is NOT tested for use in medical or clinical applications.
WA R N I N G
Use appropriate caution when using Agilent products for testing lasers.
1-4
General Information
Electrostatic Discharge Information
Electrostatic Discharge Information
CAUTION
Electrical channel input circuits and the trigger input circuit can be damaged by electrostatic
discharge (ESD). Therefore, avoid applying static discharges to the front-panel input connectors.
Prior to connecting any coaxial cable to the connectors, momentarily short the center and outer
conductors of the cable together. Avoid touching the front-panel input connectors without first
touching the frame of the instrument. Be sure that the instrument is properly earth-grounded to
prevent buildup of static charge. Wear a wrist-strap or heel-strap.
Electrostatic discharge (ESD) can damage or destroy electronic components. All work on electronic assemblies should be performed at a
static- safe work station. The following figure shows an example of a
static- safe work station using two types of ESD protection:
• Conductive table- mat and wrist- strap combination.
• Conductive floor- mat and heel- strap combination.
Figure 2. Static-safe Work Station
1-5
General Information
Electrostatic Discharge Information
Both types of techniques, when used together, provide a significant
level of ESD protection. Of the two, only the table- mat and wrist- strap
combination provides adequate ESD protection when used alone. To
ensure user safety, the static- safe accessories must provide at least 1
MΩ of isolation from ground.
WA R N I N G
These techniques for a static-safe work station should not be used when working on
circuitry with a voltage potential greater than 500 volts.
1-6
General Information
Connector Care
Connector Care
Today, advances in measurement capabilities make connectors and
connection techniques more important than ever. Damage to the connectors on calibration and verification devices, test ports, cables, and
other devices can degrade measurement accuracy and damage instruments. Replacing a damaged connector can cost thousands of dollars,
not to mention lost time! This expense can be avoided by observing the
simple precautions presented in this section.
Electrical Connectors
Advances in measurement capabilities make connectors and connection
techniques more important than ever. Observing simple precautions
can ensure accurate and reliable measurements.
•
•
•
•
•
Handling and storage
Keep connectors clean.
Extend sleeve or connector nut.
Use plastic endcaps during storage.
Do not touch mating plane surfaces.
Do not set connectors contact- end down.
Visual inspection
• Inspect all connectors carefully before every connection.
• Look for metal particles, scratches, and dents.
• Do not use damaged connectors.
•
•
•
•
Cleaning
Clean with compressed air first.
Clean the connector threads.
Do not use abrasives.
Do not get liquid onto the plastic support beads.
1-7
General Information
Optical Connectors
•
•
•
•
•
•
•
•
Making connections
Use connector savers.
Align connectors carefully.
Make preliminary connection lightly.
To tighten, turn connector nut only.
Do not apply bending force to connection.
Do not over tighten preliminary connection.
Do not twist or screw in connectors.
Use a torque wrench, and do not tighten past the "break" point of the
torque wrench.
Optical Connectors
Taking care of fiber- optic connectors is critical to making quality measurements. Because fiber- optic connectors are susceptible to damage
that is not immediately obvious to the naked eye, poor measurements
result without the user being aware. Microscopic examination and
return loss measurements are the best way to ensure good measurements. Good cleaning practices can help ensure that optimum connector performance is maintained. With glass- to- glass interfaces, any
degradation of a ferrule or the end of the fiber, any stray particles, or
finger oil can have a significant effect on connector performance.
Where many repeat connections are required, use of a connector saver
or patch cable is recommended. Figure 3 on page 1- 9 shows the basic
components of a typical connectors.
1-8
General Information
Optical Connectors
Figure 3. Basic components of a connector.
Figure 4 shows the end of a clean fiber- optic cable. The dark circle in
the center of the micrograph is the fiber’s 125 μm core and cladding
which carries the light. The surrounding area is either composed of a
ceramic material or soft nickel- silver ferrule. Figure 5 shows a dirty
fiber end from neglect or perhaps improper cleaning. Material is
smeared and ground into the end of the fiber causing light scattering
and poor reflection. Not only is the precision polish lost, but this
action can grind off the glass face and destroy the connector.
Figure 6 shows physical damage to the glass fiber end caused by either
repeated connections made without removing loose particles or using
improper cleaning tools. When severe, the damage of one connector
end can be transferred to another good connector endface that comes
in contact with the damaged one. Periodic checks of fiber ends, and
replacing connecting cables after many connections is a wise practice.
The cure for these problems is disciplined connector care as described
in the following list and in “Cleaning Non- lensed Connectors” on page
1- 12.
Use the following guidelines to achieve the best possible performance
when making measurements on a fiber- optic system:
• Never use metal or sharp objects to clean a connector and never scrape
the connector.
1-9
General Information
Optical Connectors
• Avoid matching gel and oils. The use of matching gel or oil will invalidate the N2100B system calibration. Contact Agilent Customer assistance if you have questions on this issue.
Figure 4. Clean, problem-free fiber end and ferrule.
Figure 5. Dirty fiber end and ferrule from poor cleaning.
Figure 6. Damage from improper cleaning.
1-10
General Information
Optical Connectors
While these often work well on first insertion, they are great dirt magnets. The oil or gel grabs and holds grit that is then ground into the
end of the fiber. Also, some early gels were designed for use with the
FC, non- contacting connectors, using small glass spheres. When used
with contacting connectors, these glass balls can scratch and pit the
fiber. If an index matching gel or oil must be used, apply it to a
freshly cleaned connector, make the measurement, and then immediately clean it off. Never use a gel for longer- term connections and
never use it to improve a damaged connector. The gel can mask the
extent of damage and continued use of a damaged fiber can transfer
damage to the instrument.
• When inserting a fiber- optic cable into a connector, gently insert it in
as straight a line as possible. Tipping and inserting at an angle can
scrape material off the inside of the connector or even break the inside
sleeve of connectors made with ceramic material.
• When inserting a fiber- optic connector into a connector, make sure that
the fiber end does not touch the outside of the mating connector or
adapter.
• Avoid over- tightening connections.
Unlike common electrical connections, tighter is not better. The purpose of the connector is to bring two fiber ends together. Once they
touch, tightening only causes a greater force to be applied to the delicate fibers. With connectors that have a convex fiber end, the end can
be pushed off- axis resulting in misalignment and excessive return loss.
Many measurements are actually improved by backing off the connector pressure. Also, if a piece of grit does happen to get by the cleaning
procedure, the tighter connection is more likely to damage the glass.
Tighten the connectors just until the two fibers touch.
• Keep connectors covered when not in use.
• Use fusion splices on the more permanent critical nodes. Choose the
best connector possible. Replace connecting cables regularly. Frequently
measure the return loss of the connector to check for degradation, and
clean every connector, every time.
All connectors should be treated like the high- quality lens of a good
camera. The weak link in instrument and system reliability is often the
inappropriate use and care of the connector. Because current connectors are so easy to use, there tends to be reduced vigilance in connector care and cleaning. It takes only one missed cleaning for a piece of
grit to permanently damage the glass and ruin the connector.
1-11
General Information
Optical Connectors
Measuring insertion loss and return loss
Consistent measurements with your lightwave equipment are a good
indication that you have good connections. Since return loss and insertion loss are key factors in determining optical connector performance
they can be used to determine connector degradation. A smooth, polished fiber end should produce a good return- loss measurement. The
quality of the polish establishes the difference between the “PC” (physical contact) and the “Super PC” connectors. Most connectors today
are physical contact which make glass- to- glass connections, therefore
it is critical that the area around the glass core be clean and free of
scratches. Although the major area of a connector, excluding the glass,
may show scratches and wear, if the glass has maintained its polished
smoothness, the connector can still provide a good low level return
loss connection.
If you test your cables and accessories for insertion loss and return
loss upon receipt, and retain the measured data for comparison, you
will be able to tell in the future if any degradation has occurred. Typical values are less than 0.5 dB of loss, and sometimes as little as 0.1
dB of loss with high performance connectors. Return loss is a measure
of reflection: the less reflection the better (the larger the return loss,
the smaller the reflection). The best physically contacting connectors
have return losses better than 50 dB, although 30 to 40 dB is more
common.
Visually inspecting fiber ends
Visual inspection of fiber ends can be helpful. Contamination or
imperfections on the cable end face can be detected as well as cracks
or chips in the fiber itself.
WA R N I N G
Always remove both ends of fiber-optic cables from any instrument, system, or
device before visually inspecting the fiber ends. Disable all optical sources before
disconnecting fiber-optic cables. Failure to do so may result in permanent injury to
your eyes.
Use a microscope (100X to 200X magnification) to inspect the entire
end face for contamination, raised metal, or dents in the metal as well
as any other imperfections. Inspect the fiber for cracks and chips. Visible imperfections not touching the fiber core may not affect performance (unless the imperfections keep the fibers from contacting).
Cleaning Non- lensed Connectors
The procedures in this section provide the proper steps for cleaning
fiber- optic cables and Agilent Technologies universal adapters. The initial cleaning, using the alcohol as a solvent, gently removes any grit
1-12
General Information
Optical Connectors
and oil. If a caked- on layer of material is still present, (this can happen if the beryllium- copper sides of the ferrule retainer get scraped
and deposited on the end of the fiber during insertion of the cable), a
second cleaning should be performed. It is not uncommon for a cable
or connector to require more than one cleaning.
CAUTION
Agilent Technologies strongly recommends that index matching compounds not be applied to their
instruments and accessories. Some compounds, such as gels, may be difficult to remove and can
contain damaging particulates. If you think the use of such compounds is necessary, refer to the
compound manufacturer for information on application and cleaning procedures.
CAUTION
Do not use any type of foam swab to clean optical fiber ends. Foam swabs can leave filmy deposits
on fiber ends that can degrade performance.
1 Apply pure isopropyl alcohol to a clean lint- free cotton swab or lens
paper.
Cotton swabs can be used as long as no cotton fibers remain on the
fiber end after cleaning.
2 Clean the ferrules and other parts of the connector while avoiding the
end of the fiber.
3 Apply isopropyl alcohol to a new clean lint- free cotton swab or lens
paper.
4 Clean the fiber end with the swab or lens paper.
Do not scrub during this initial cleaning because grit can be caught in
the swab and become a gouging element.
5 Immediately dry the fiber end with a clean, dry, lint- free cotton swab
or lens paper.
6 Blow across the connector end face from a distance of 6 to 8 inches
using filtered, dry, compressed air. Aim the compressed air at a shallow
angle to the fiber end face.
Nitrogen gas or compressed air dust remover can also be used.
CAUTION
Do not shake, tip, or invert compressed air canisters, because this releases particles in the can into
the air. Refer to instructions provided on the compressed air canister.
7 As soon as the connector is dry, connect it or cover it for later use.
1-13
General Information
Optical Connectors
If the performance, after the initial cleaning, seems poor try cleaning
the connector again. Often a second cleaning will restore proper performance. The second cleaning should be more arduous with a scrubbing action.
1-14
General Information
Returning the N2100B to Agilent
Returning the N2100B to Agilent
The instructions in this section show you how to properly package the
instrument for return to an Agilent Technologies service office. If the
instrument is still under warranty or is covered by an Agilent maintenance contract, it will be repaired under the terms of the warranty
or contract. If the instrument is no longer under warranty or is not
covered by an Agilent maintenance plan, Agilent will notify you of
the cost of the repair after examining the unit.
When an instrument is returned to an Agilent service office for servicing, it must be adequately packaged and have a complete description of the failure symptoms attached.
When describing the failure, please be as specific as possible about the
nature of the problem. Include copies of any instrument failure settings, data related to instrument failure, and error messages along with
the instrument being returned.
Please notify the service office before returning your instrument for
service. Any special arrangements for the instrument can be discussed
at this time. This will help the Agilent service office repair and return
your instrument as quickly as possible.
Call Center
For technical assistance, contact your local Agilent Call Center. In the
Americas, call 1 (800) 829- 4444. In other regions, visit http://www.agilent.com/find/assist. Before returning an instrument for service, you
must first contact your local Agilent Call Center.
Preparing the product for shipping
1 Write a complete reason for returning the product and attach it to the
instrument. Include any specific performance details related to the
problem.
2 Pack the product. Use original packaging or comparable. Original
materials are available through any Agilent office. Or, follow these
recommendations:
• Insert the product in an anti- static bag.
• Use a double- walled, corrugated cardboard carton of 159 kg (350 lb)
1-15
General Information
Returning the N2100B to Agilent
test strength. The carton must allow approximately 7 cm (3 inches)
on all sides of the kit for packing material and be strong enough to
accommodate the weight of the kit.
• Surround the kit with approximately 7 cm (3 inches) of packing material, to protect the kit and prevent it from moving in the carton. If
packing foam is not available, the best alternative is S.D- 240 Air
Cap™ from Sealed Air Corporation (Commerce, California 90001). Air
Cap looks like a plastic sheet filled with air bubbles. Use the pink
(antistatic) Air Cap™ to reduce static electricity. Wrapping the kit
several times in this material will protect the kit and prevent it from
moving in the carton.
3 Seal the carton with strong nylon adhesive tape.
4 Mark the carton “FRAGILE, HANDLE WITH CARE”.
5 Retain copies of all shipping papers.
1-16
2
Introduction 2- 2
Step 1. Inspect the Shipment 2- 2
Step 2. Install the Instrument Driver Software
Step 3. Install the N2100B 2- 4
Quick Confidence Check 2- 5
Powering Off the Instrument 2- 8
Upgrading the Instrument Firmware 2- 9
Installation
2- 3
Installation
Introduction
Introduction
The PXI chassis can be controlled using either an embedded PXI controller or an external PC using a PCI - cPCI/PXI remote bridge (such
as the NI MXI- 4 product).
You can control the DCA with Agilent Vee only through ActiveX.
NOTE
If an external PC is used, the PC must meet the following specifications:
•
•
•
•
NOTE
Windows 2000 or XP operating system
1 GB RAM
Pentium processor, 133 MHz or greater
NI- VISA
When using an external PC as the controller, you must follow the installation steps in the sequence
described for the PC BIOS to locate the instruments in the PXI chassis.
Step 1. Inspect the Shipment
1 Inspect the shipping container and kit for damage. Keep the shipping
container and cushioning material until you have inspected the contents
of the shipment for completeness and have checked the kit mechanically
and electrically.
2 Locate the shipping list. Verify that you have received all of the items
listed.
To contact Agilent Technologies for technical assistance, contact your
local Agilent Call Center. In the Americas, call 1 (800) 829- 4444. In
other regions, visit http://www.agilent.com/find/assist. Before returning
an instrument for service, you must first contact your local Agilent
Call Center.
3 Verify that the following environmental conditions exist before
proceeding with the installation procedure.
2-2
Installation
Step 2. Install the Instrument Driver Software
WA R N I N G
Ensure that the PXI chassis is connected to the specified power source using the
correct power cord (noting country of use).
WA R N I N G
Ensure that the PXI chassis provides adequate earth grounding.
WA R N I N G
Ensure that the air supply to the chassis is working correctly. The N2100B requires
an optimal air flow within the chassis. It is recommended to regularly change filters
on PXI Chassis.
Step 2. Install the Instrument Driver Software
NOTE
This procedure assumes that NI-VISA has already been installed.
1 Log onto the PC with administrator privileges, so that you can install
the software.
2 Go to the Agilent website: www.agilent.com/find/pxit
3 Click on the Technical Support link and then the Drivers link.
4 Download the latest version of the following driver:
Agilent N2100B DCA Driver
5 Once the download has completed, run the N2100B installation file.
a During the installation, you will enter the user name and organization.
b Select the all users option to ensure the software is available to all
users of the PC. Click Next.
6 When installation process is finished, click Finish. The N2100B control
software is now installed.
7 To ensure that the PC BIOS will be able to locate the instruments in
the PXI chassis, follow this step according to the controller that you are
using:
• If you are using an external PC and remote bridge, turn the PXI chassis and PC power off.
• If you are using an embedded controller, turn the PXI chassis and
PC power off and remove all N2100B modules from the chassis.
2-3
Installation
Step 3. Install the N2100B
NOTE
You can control the DCA with Agilent Vee only through ActiveX.
NOTE
As part of the installation, the N2100B User’s Guide is made available on the Windows Start
menu.
Step 3. Install the N2100B
1 With the PC and chassis powered off, install the N2100B module in an
available slot in a PXI chassis.
2 Power on the PXI chassis and wait for the power up sequence to
complete.
3 Turn on the PC.
If the software is correctly installed, you will see an indication that
new hardware is detected and that the system is attempting to locate
the associated software driver. When this process is complete, a notification appears indicating that the hardware is ready for use.
If needed, you can use Windows Device Manager to determine if the
instruments have been correctly identified by the BIOS. There should
be an NI- VISA PXI Devices entry with your N2100B instrument.
NOTE
The N2100B User’s Guide is provided as a PDF file. To locate and view the user’s guide, go to the
Windows Start menu and select All Programs > Agilent > N2100B > Documents >
User’s Guide.
2-4
Installation
Quick Confidence Check
Quick Confidence Check
1 Apply a PRBS signal to the optical or electrical input connector and
then configure the instrument accordingly.
NOTE
For the N2100B, both optical and electrical signals can be applied at the same time. For the
N2100A, a signal can be applied to only one input.
2 Allow the DCA to warm up for at least 15 minutes.
3 Click the Windows Start menu.
4 Click All Programs > Agilent > N2100B.
5 Click N2100B Control Panel.
6 Click I/O Config to open the Set Module I/O Address dialog
box, to then configure the communications between the host
controller and the instrument.
7 Select PXI bus.
8 Click Scan for Instruments and select the appropriate DCA instrument ID
String.
9 Click OK. (When multiple DCAs are present in the chassis, use NI
Measurement and Automation Explorer (NI MAX) to determine which
DCA is in which chassis slot before selecting one.)
10 On the Control Panel, click Connect. A message panel appears
briefly as the connection is established and then disappears.
11 Select the Module Config button.
2-5
Installation
Quick Confidence Check
12 Select the Line Rate, the reference clock source, and the relevant Input
Path and Filter option.
13 Click Start Acquisition.
A pop- up window appears when data acquisition is performed, notifying you that recalibration is advisable. Click Calibrate Now.
Figure 7. Calibration Warning Dialog Box
2-6
Installation
Quick Confidence Check
The Calibration dialog box appears and the calibration starts automatically. When the calibration is complete, the Calibration dialog box disappears.
NOTE
There must be a PRBS input signal to run the calibration. If the input signal is not PRBS, or if there
is no input signal, click OK and ignore the warning that appears.
You should run the Delay Calibration when the following situations
occur:
• The first time an input path/filter is selected.
• When the internal module temperature has changed.
• When the line rate changes. (Recommended only.)
14 The N2100B is functioning correctly when you can see an eye diagram
as shown in the following figure.
If the input signal timing cannot be locked, the following message is
displayed.
2-7
Installation
Powering Off the Instrument
.
Powering Off the Instrument
1 Click Disconnect.
2 Close the Control Panel.
3 Power down the PC.
4 Switch off the power to the PXI chassis.
2-8
Installation
Upgrading the Instrument Firmware
Upgrading the Instrument Firmware
To update the firmware and execute the FPGA Loader program, perform the following steps:
1 Close any open Control Panels.
2 On the Windows Start menu, click All Programs > Agilent > N2100B >
N2100B FW-FPGA Loader.
3 Press the appropriate buttons to connect to a given module, and note
the current module Serial Number, API, and Firmware Versions.
4 Select the new firmware and FPGA files located in the installation
directory (typically C:\Program Files\Agilent\N2100B\fw), and press
start.
A progress bar appears as the module is updated.
5 Once complete, shut down the host PC and cycle the power on the PXI
chassis.
6 After rebooting the PC, connect to the module and obtain Module
Configuration, to confirm proper software upload.
2-9
Installation
Upgrading the Instrument Firmware
2-10
3
Introduction 3- 2
Module Configuration Settings
Acquiring Data 3- 11
Calibrating the N2100B 3- 13
3- 4
Configuring the DCA and Acquiring Data
Configuring the DCA and Acquiring Data
Introduction
Introduction
This chapter describes how to start and configure the DCA using the
Control Panel. For information on controlling the five measurement
modes, refer to Chapter 4, “Measurements”.
All aspects of the N2100B can be managed through the Control Panel.
Use the top row of buttons to configure the module, I/O, and connect
to the module. Use the list box in the top left corner of the main form
to select the mode of operation for the N2100B.
Row of
configuration
buttons
Data
acquisition
Operating
mode
Clipboard
copy
Measurements
Release button for
detatchable
results win-
Markers
Measurements
results area
Figure 8. Control Panel
3-2
Configuring the DCA and Acquiring Data
To start the Control Panel
To start the Control Panel
NOTE
A single host computer or controller can control multiple DCAs up to the capability of the PXI
backplane. An individual DCA cannot be controlled by more than one application at a time.
1 Click the Windows Start menu.
2 Click All Programs > Agilent > N2100B.
3 Click N2100B Control Panel.
4 Click I/O Config to open the Set Module I/O Address dialog
box, to then configure the communications between the host
controller and the instrument.
5 Select PXI bus.
• If you want to set up a TCP/IP interface, refer to “Setting the IP
address for connecting with TCP/IP” on page 3- 4.
6 Click Scan for Instruments and select the appropriate DCA instrument ID
String.
7 Click OK. (When multiple DCAs are present in the chassis, use NI
Measurement and Automation Explorer (NI MAX) to determine which
DCA is in which chassis slot before selecting one.)
8 On the Control Panel, click Connect. A message panel appears
briefly as the connection is established and then disappears.
3-3
Configuring the DCA and Acquiring Data
Module Configuration Settings
Module Configuration Settings
The Configuration Settings dialog box allows you to set up the operating parameters. The current DCA configuration appears in this panel
when it first opens and any time the Refresh Button is pressed.
Changes made on this screen can be applied to the DCA by pressing
the Apply button (when highlighted), or automatically by pressing the
Start Acquisition button. Factory defaults and ten (10) user selectable
preset configurations are also available. Configuration and presets are
persistent and are specific to a particular DCA module.
Setting the IP address for connecting with TCP/IP
1 In the N2100B Control Panel, click Module Config to open the
Configuration Settings dialog box.
2 Click Set IP Addr, located in the bottom- right corner of the dialog box.
3-4
Configuring the DCA and Acquiring Data
Global Parameters
The Set Module TCP/IP Address dialog box appears.
On a local network, the IP address must be within the range of active
addresses for the network. (The DCA does not support DHCP automatic address assignment.) If you are connecting the DCA directly to a
computer, a crossover network cable is required and the computer
must have a static IP address. The first three fields of both the computer address and the DCA address must be the same.
3 Click OK and reboot the PXI system to activate the new IP address
setting.
4 When the system has restarted, open the Control Panel and click Show
Version to check the module version information and the current IP
address setting.
Global Parameters
Line Rate (Mb/s)
Selects the desired Line Rate from the drop down list or 'Custom.'
3-5
Configuring the DCA and Acquiring Data
Global Parameters
Custom Line Rate
Specifies the desired line rate if 'Line Rate' is set to 'Custom.'
Optical Wavelength
Specifies the wavelength of the signal under test. This affects the calibration coefficient used to calculate optical power.
Optical Attenuation
This field allows you to enter the value for an external optical attenuator (if any) in the input test signal path. You can enter it as a ratio
or a value in dB, with no attenuation as the default. All DCA measurements will compensate for this attenuator and report effective values
at its input.
Electrical Attenuation
This is the same function as the Optical Attenuation field described
above, but for the electrical input path. You can set these two values
independently.
Reference Clock
Sets the Reference Clock to one of the following modes:
• Internal. The internal reference clock. The recommended modes of
operation are either the clock recovery or external clock option. The
instrument specifications are valid only with those two modes.
• External. A user- provided clock signal connected to the External
Clock input. The frequency of this clock must be between 10 MHz
and 11.318 GHz and must match the value of the External Reference
Clock Frequency parameter.
• Recovered. The clock derived from the received optical or electrical
data. The recovered clock with a sub- rate pattern of K28.7 is usable
with the API, but not the Control Panel interface. See
FG_REFCLK_RECOVERED_RATIO in Table 8 on page 9- 43.
External Reference
The N2100B can operate with a non- divided down clock signal or an
“at rate” clock signal. It can also operate with a clock divided down
signal. This parameter allows you to enter the reference as a particular
frequency (freq). For example, freq 10 (MHz). You can also enter the
value as a division of the line rate (rate/). For example, rate / 128.
ER Correction Factor
Extinction Ratio correction factor, in [%], is a constant positive or negative number that will be substracted from the measured ER. The
returned ER results will be affected only if the correction is activated
by checking the ‘Active’ box.
3-6
Configuring the DCA and Acquiring Data
DCA Parameters
By using the extinction ratio correction factor, the user can both
decrease the uncertainty in the measurement and improve repeatability
from system to system. Based on the causes of the ER inaccuracies the
correction factor is compensating for (DC effects, AC effects or both), a
correction factor may be applicable to a specific measurement system
or a family of measurement systems.
The formula employed for ER correction:
Final ER [%] = Measured ER [%] – ER correction factor.
NOTE
Extinction ratio can be defined as a linear ratio, in decibels, or as a percentage:
ER linear = ‘1’ power level / ’0’ power level
ER (dB) = 10 log10(ER linear)
ER (%) = 100 X ‘0’ power level/’1’ power level ‘0’
DCA Parameters
Sample Blocks
Specifies the number of sample blocks that will be used to create the
'eye.' Each sample block consists of 1,024 sample points.
Input Path and Filter
Sets either the electrical input path or the optical input path and the
filter.
Pattern Acquisition
Controls the Pattern Acquisition with the following mode choices:
• Off. The standard DCA eye capture is applied.
• Enabled (No filter). The DCA performs a pattern capture but does
not apply any filtering.
• Any specific filter. The DCA performs a pattern capture and then applies the selected digital filter to this pattern.
Pattern
Specifies the length of the repeating pattern sequence (PRBS 27- 1,
PRBS 29- 1, PRBS 211- 1, K28.5, K28.7), and Set Length for patterns of
arbitrary length. When Set Length is selected, the field to the right is
enabled so you can enter the pattern length. The length must be
between 4 and 2047.
3-7
Configuring the DCA and Acquiring Data
DCA Parameters
NRZ Thresholds
Selects threshold levels used for the Rise & Fall times calculation. The
choices are 10% to 90%, 20% to 80%, and 30% to 70% of either the total
Waveform Amplitude (Full Period), or the NRZ- Windowed high and low
values.
NRZ Window Size
Sets the percentage of the bit period (unit interval) used for NRZ measurements. The NRZ window is centered between the eye crossings.
OMA Window Size
Sets the percentage of the bit period (unit interval) used for OMA
measurements. The OMA window is centered on the eye crossing.
3-8
Configuring the DCA and Acquiring Data
Additional Controls
Additional Controls
Refresh
Cancels unapplied changes on the screen and restores the current DCA
configuration.
Apply
Sends the new configuration settings to the DCA, when you have made
changes to parameters. Pressing Start Acquisition also applies any
changes to parameters.
Show Version
Displays information about the Module Hardware and Software.
Set Defaults
Loads DCA Factory Defaults. These are applied with Apply or Start
Acquisition buttons.
Set IP Addr
Sets the Module IP address. The first three fields for this address must
be the same as the address for the computer that is trying to connect
to the instrument.
3-9
Configuring the DCA and Acquiring Data
Additional Controls
Presets
Clicking Save, followed by one of the buttons [1 - 10] stores the current configuration parameters into one of 10 presets. Preset configurations are DCA specific, and persist until deleted. Clicking buttons [1 10] restores the previously saved configuration associated with that
button. To delete a preset configuration, click Save (while holding the
Ctrl key) followed by the preset button [1- 10].
3-10
Configuring the DCA and Acquiring Data
Acquiring Data
Acquiring Data
After configuring the module and the Control Panel,
click Start Acquisition to begin acquiring data. In the
Control Panel, you can select Repetitive to have continuous acquisitions, as shown in the following figure.
Figure 9. Repetitive Selected
Select Repetitive and Count to specifiy a number of acquisitions to
acquire. In the following figure, the count is set to 100. A progress bar
indicates how many acquisitions have happened and how many are
still to be taken.
progress bar
Figure 10. Count Selected with Progress Bar
Use the General tab of the Program Settings dialog box
to configure the repetitive results text.
• Select Overwrite during a repetitive acquisition sequence so each
subsequent measurement overwrites the measurement on the display.
• Select Add to create a list of measurement results on the display
where each new result is added to the list.
• Select Add + Time to create a time stamped list of measurement results. As each
measurement result is added to form a list on the screen, a start time, an end time,
3-11
Configuring the DCA and Acquiring Data
Acquiring Data
and an elapsed time of the most recent data collection is included. A time stamp
is also included in the results list every 30 seconds.
Viewing and moving the measurement results window
The measurement results area is located at the bottom of the Control
Panel. However, you can click the release button to detach the results
window from the Control Panel for more convenient viewing.
Release button for
detatchable
results win-
Measurements
results area
1 Click the release button to detach the measurement results window.
2 Drag the measurement results window to a more convenient viewing
position.
3 Drag the bottom border of the window to enlarge the area for viewing
multiple measurement results without scrolling through the data.
3-12
Configuring the DCA and Acquiring Data
Calibrating the N2100B
Calibrating the N2100B
Clicking Module Cal displays the Calibration dialog box
that provides access to the following features:
• Calibration of the delay lines used in vector sampling
• Dark current calibration
• System log
Figure 11. Calibration Dialog Box
Delay Lines Calibration
The Delay Calibration calibrates the delay timers that position the
additional samples associated with the vector sampling technique. A
PRBS signal must be applied to the N2100B for the calibration to work
correctly. During the calibration, wait until the message “Calibration
Complete” appears before continuing. You should run the Delay Calibration when the following situations occur:
• The first time an input path/filter is selected.
• When the internal module temperature has changed.
• When the Line rate changes. (Recommended only.)
Temperature Monitor and The DCA firmware monitors the internal module temperature and
Delay Calibration
notifies the user or the client application when the temperature has
Notification
changed enough that a delay recalibration is required.
3-13
Configuring the DCA and Acquiring Data
Calibrating the N2100B
In the Control Panel application, a pop- up window appears when data
acquisition is performed, notifying you that recalibration is advisable.
Calibration may be performed by clicking Calibrate Now, or at a later
time using the Calibration dialog box described above.
Figure 12. Calibration Warning Dialog Box
Data acquisition functions that are in the ActiveX control and the
module API return a status value, FG_DELAYCAL_REQD, indicating
that recalibration is advisable.
Dark Current Calibration
Calibration for optical signals depends on the optical detector's dark
current. That is, the value measured with no light applied. The DCA
can compensate for temperature- or time- related dark current drift.
Use the following procedure to perform a dark current calibration:
1 Remove any optical signal from the optical input and cover the
connector.
2 Click Module Cal on the Control Panel.
3 Click Measure Dark Current.
The DCA measures and stores the dark current measurement. This is
relevant to extinction ratio and optical power measurements. To
increase the accuracy of these values, a dark current calibration
should be performed regularly.
3-14
Configuring the DCA and Acquiring Data
Calibrating the N2100B
System Log
The System Log button allows you to view the log of system errors and
events. The information collected depends on the Logging Level selection:
•
•
•
•
•
Logging Level None. Logs nothing.
Logging Level Fatal. Logs fatal errors only.
Logging Level Error. Logs all errors.
Warning. Logs all errors and warnings.
Logging Level Info. Logs errors, warnings, and other events.
3-15
Configuring the DCA and Acquiring Data
Calibrating the N2100B
• Logging Level Debug. Logs errors, warnings, and internal information used for
firmware debugging.
• Application Log. Associated with application-level firmware functions.
• System Log. Associated with system-level driver functions.
• Read Log. Displays the appropriate log file using Notepad. The file is deleted unless the 'Keep File' checkbox is selected.
• Clear Log. Clears the appropriate log. The 'Allow Clear' checkbox must be selected for this to be available.
NOTE
The logging setting should be set to Error, Fatal or None in normal use. Otherwise, the
performance speed of the DCA will be affected.
3-16
4
Introduction 4- 2
Oscilloscope Measurements Mode 4- 5
Non Return to Zero Measurements Mode 4- 17
Eye Mask Test Measurements Mode 4- 39
Multiple Measurements Mode 4- 46
Pattern Sequence D. J. Measurements Mode 4- 49
Measurements
Measurements
Introduction
Introduction
This section describes how to control the various modes of operation
for performing measurements:
•
•
•
•
•
“Oscilloscope Measurements Mode” on page 4- 5.
“Non Return to Zero Measurements Mode” on page 4- 17.
“Eye Mask Test Measurements Mode” on page 4- 39.
“Multiple Measurements Mode” on page 4- 46.
“Pattern Sequence D. J. Measurements Mode” on page 4- 49.
For information on all other aspects of using the Control Panel, refer
to “Configuring the DCA and Acquiring Data” on page 3- 1.
Eye Diagram Display
The N2100B automatically performs accurate eye- diagram analysis to
characterize the quality of transmitters from 1.063 Gb/s to 8.5 Gb/s.
This chapter lists all possible measurements for each mode including:
• Definition (optical and electrical), including formula for parameter calculation
• Examples with screen shots (OC48 (STM4) signals were used)
• Display configuration
• Procedures for measurements using the API
Eye diagrams are a key figure of merit for most computer and communications system standards, including Gigabit Ethernet, Sonet, Infiniband, Rapid IO, PCI Xpress and others. The eye diagram provides a
clear visual representation whether the interconnection, by itself,
would meet the test specification for a given standard.
The Eye Diagram provides a longer term view of the signal, taking into
account the relative time position of successive pulses. It provides a
more thorough analysis of the cumulative effects of jitter.
4-2
Measurements
Introduction
Eye diagrams are able to collect and display multiple waveforms. For a
large number of waveforms (or blocks), the timing and pulse width
variations are displayed as a widening of the eye- diagram traces,
which must remain within the template to meet specification.
Figure 13. Eye Diagram
4-3
Measurements
Introduction
Figure 14. Optical Power
4-4
Measurements
Oscilloscope Measurements Mode
Oscilloscope Measurements Mode
Oscilloscope Measurement mode offers the ability to select the display
time scale and a selection of measurements related to the analog characteristics of the entire signal.
The following table summarizes the measurements the DCA can perform in Oscilloscope Mode. The DCA software installation includes an
HTML document with full details of the ActiveX and API functions
listed in the examples that follow in this section.
Table 1. Oscilloscope Mode Measurements
Measurement
Description
“Top” on page 4-8.
The average of all sample values identified as high-to-high transitions.
“Base” on page 4-9.
The average of all sample values identified as low-to-low transitions.
“Amplitude” on page 4-9.
The difference between top level and base level of a displayed pulse
waveform.
“Peak to Peak” on page
4-11.
The difference between max and min of a displayed pulse waveform.
“RMS” on page 4-11.
The RMS value of the waveform.
“Min” on page 4-13.
The value of the sample point with the lowest amplitude.
“Max” on page 4-14.
The value of the sample point with the highest amplitude.
“Positive Overshoot” on
page 4-14.
The amount by which the Max value exceeds the Top value, expressed
as a percentage of the Amplitude.
“Negative Overshoot” on
page 4-15.
The amount by which the Min value falls below the Base value,
expressed as a percentage of the Amplitude.
“Average Optical Power”
on page 4-16.
The average power read by the PIN diode (optical signals only).
4-5
Measurements
Oscilloscope Measurements Mode
Oscilloscope
Meas. Mode
Measurements
View eye or
pattern
Time and
amplitude
controls
Markers and
histograms
Figure 15. Oscilloscope Measurement Mode Control Panel
X-Axis Control
• UI always displays a single UI.
• Set displays the X axis with user- selectable time/division and offset.
• Scale zooms in and out.
• Offset (UI) adjusts the starting X value of the display.
• HiRes changes the offset by .05 UI or 0.005 UI.
• Zoom sets the closest pair of offset and scale settings to zoom within
the area selected by the markers.
4-6
Measurements
Oscilloscope Measurements Mode
The offset and scale values are displayed on the bottom left side of the
Eye- Diagram Display .
Y-Axis Control
• Auto selects best fit scaling of eye diagram on screen.
• Scale zooms in and out.
• Offset creates an offset the eye within the display.
• Find auto- finds the trace on the screen. The Y- Axis scale and offset are
adjusted accordingly.
• Zoom sets the closet pair of offset and scale settings to zoom within the
area selected by the markers.
The amplitude values are displayed along the Y axis, with the units on
top.
Markers
When you switch on the Markers control, a pair of movable markers in
both X- axis and Y- axis are displayed. The data at the position of the
markers are displayed. The histogram function displays a histogram of
the samples within the region bounded by the markers, selectable over
X or Y values.
You can reposition the markers using the mouse. Select the marker
with the left mouse button and move it by dragging the mouse. If you
hold down the shift key while moving the mouse, the two markers will
lock so that the distance between them is preserved.
The markers can also be used to control the scale of the display. The
normal view (UI) presents the view of two UI. Switching to the Set
option in the X Axis control allows you to zoom in and out on the eye
diagram. In Y Axis (Auto) mode, the signal amplitude is scaled to best
fit within the display screen.
4-7
Measurements
Oscilloscope Measurements Mode
Top
Top is the average of all sample values identified as high- to- high transitions.
Control Panel
Click Top on the left side of the display.
API Call
FGDCAOscilloscopeAPI
ActiveX
MeasOscTopLevel
4-8
Measurements
Oscilloscope Measurements Mode
Base
Base is the average of all sample values identified as low- to- low transitions.
Control Panel
Click Base on the left side of the display.
API Call
FGDCAOscilloscopeAPI
ActiveX
MeasOscBaseLevel
Amplitude
Amplitude is the vertical difference between the Top and Base of the
signal.
Top - Base = Amplitude
4-9
Measurements
Oscilloscope Measurements Mode
Control Panel
Click Amplitude on the left side of the display.
API Call
FGDCAOscilloscopeAPI
ActiveX
MeasOscAmplitude
4-10
Measurements
Oscilloscope Measurements Mode
Peak to Peak
Peak- to- peak (pk- pk) is the difference between the maximum positive
and the maximum negative amplitudes of a waveform.
Control Panel
Click Peak to Peak on the left side of the display.
API Call
FGDCAOscilloscopeAPI
ActiveX
MeasOscPeakToPeak
RMS
Root- mean- square (rms) refers to the most common mathematical
method of defining the effective voltage or current of an AC wave.
To determine rms value, four mathematical operations are carried out
on the function representing the AC waveform:
1 Take the average of all samples.
2 Take the sum of (sample[i] - average)2.
3 Divide the sum by (# of samples).
4 Take the square root of the result.
4-11
Measurements
Oscilloscope Measurements Mode
N
value rms =
1-⎞
2
⎛ --⎝ N⎠ ∑ ( sample[i] – average )
1
Control Panel
Click RMS on the left side of the display.
API Call
FGDCAOscilloscopeAPI
ActiveX
MeasOscRMS
4-12
Measurements
Oscilloscope Measurements Mode
Min
Min is the value of the sample point with the lowest amplitude.
Control Panel
Click Min on the left side of the display.
API Call
FGDCAOscilloscopeAPI
ActiveX
MeasOscMinLevel
4-13
Measurements
Oscilloscope Measurements Mode
Max
Max is the value of the sample point with the highest amplitude.
Control Panel
Click Max on the left side of the display.
API Call
FGDCAOscilloscopeAPI
ActiveX
MeasOscMaxLevel
Positive Overshoot
Positive Overshoot is the difference between the maximum sample
value and the signal's top level, expressed as a fraction of the signal
amplitude.
( Max – Top )
Positive Overshoot = ------------------------------- × 100
( Top – Base )
4-14
Measurements
Oscilloscope Measurements Mode
Control Panel
Click Pos. Overshoot on the left side of the display.
API Call
FGDCAOscilloscopeAPI
ActiveX
MeasOscPosOvershoot
Negative Overshoot
Negative Overshoot is the difference between the minimum sample
value and the signal's base level, expressed as a fraction of the signal
amplitude.
( Base – Min )
Negative Overshoot = -------------------------------- × 100
( Top – Base )
4-15
Measurements
Oscilloscope Measurements Mode
.
Control Panel
Click Neg. Overshoot on the left side of the display.
API Call
FGDCAOscilloscopeAPI
ActiveX
MeasOscNegOvershoot
Average Optical Power
Applicable to optical signals only. The Average Optical Power is the
true average component of an optical signal, expressed in microwatts
or dBm (decibels relative to a power level of one milliwatt). This measurement results from the use of a hardware average- power monitor
circuit rather than from the calculation of digitized waveform data.
Control Panel
Click Avg. Power on the left side of the display.
API Call
FGDCAAvgPowerAPI
ActiveX
MeasAvgPowerUW
MeasAvgPowerDBM
4-16
Measurements
Non Return to Zero Measurements Mode
Non Return to Zero Measurements Mode
NRZ (Non- return- to- zero) Measurement Mode offers a selection of signal and timing measurements typically associated with NRZ signals.
Re- acquisition of data is not required.
The following table summarizes the measurements the instrument is
capable of performing in NRZ:
Table 2. DCA NRZ Measurements (1 of 2)
Measurement
Description
“Extinction Ratio” on page
4-19.
For optical signals only. A measure of the ratio of optical power at the signal's '1' level to the power at
the signal's '0' level.
“Eye Height” on page 4-21.
Eye Height is the vertical opening of the eye measured as the difference between the -3 sigma value
of the high samples within the NRZ window and the +3 sigma value of the low samples within the
NRZ window. The measured value can be expressed as a ratio, percentage, or in dB.
“Eye Amplitude” on page
4-22.
Difference between the average of the high samples within the NRZ window and the average of the
low samples within the NRZ window.
“Eye Width” on page 4-23.
Time between the latest possible occurrences of one eye crossing to the earliest possible occurrence
of the next crossing.
“OMA (Optical Modulation
Amplitude)” on page 4-24.
Difference between the high and low levels measured in a window centered on the eye crossing.
“Low” on page 4-26.
The average of all sample values within the NRZ window identified as low-to-low transitions.
“High” on page 4-27.
The average of all sample values within the NRZ window identified as high-to-high transitions.
“Signal-to-Noise Ratio” on
page 4-28.
Ratio of the signal difference between high level and low level relative to the noise present at both
levels.
“Bit Rate” on page 4-29.
The rate of successive data transitions or potential data transitions of the input signal. Returns a value,
depending on the reference clock selected:
Internal clock returns the line rate set by the user.
External clock returns the frequency at the reference clock input.
Recovered clock returns the measured line rate.
4-17
Measurements
Non Return to Zero Measurements Mode
Table 2. DCA NRZ Measurements (2 of 2)
Measurement
Description
“Rise Time” on page 4-29.
The mean transition time of the data on the upward slope of an eye diagram between two defined
thresholds (for example, 20% to 80%) relative to signal amplitude (the full-UI Top and Base levels), or
relative to the NRZ-window High and Low levels.
“Fall Time” on page 4-31.
The mean transition time of the data on the downward slope of an eye diagram between two defined
thresholds (for example, 90% and 10%) relative to the full-UI Top and Base levels, or the NRZ-window High and Low levels.
“Crossing Percentage” on
page 4-32.
The amplitude of the crossing level relative to the low and high levels.
“Duty Cycle Distortion” on
page 4-32.
The time separation between the rising edge and falling edge at the 50% level of the eye diagram.
“Edge Jitter” on page 4-34.
A measure of the variation in time position of signal transitions, measured at the crossing.
“Single-Edge Jitter” on page
4-35.
A measure of the variation in rising or falling transitions only in time position of signal transitions. A
peak to peak value as well as an rms value is returned.
“Deterministic Jitter” on
page 4-36.
The time difference between transitions that immediately follow another transition (LHL and HLH) and
transitions that immediately follow an interval with no transition (LLH and HHL).
“Crossing Hist.” on page
4-37.
Histogram showing the distribution of samples at the crossing point.
“ISI Histogram” on page
4-38.
Histogram showing the distribution of the four different transition types separately.
4-18
Measurements
Non Return to Zero Measurements Mode
Figure 16. NRZ Measurements Mode Control Panel
Extinction Ratio
Extinction ratio is the ratio of the 'On' optical level (High) to the 'Off'
optical level (Low) measured within the NRZ window. The result is
expressed as Ratio, dB, and Percentage. These are calculated using the
following formulas:
ER ratio = High
-----------Low
High
ERdB = 10 log ⎛ ------------⎞
⎝ Low ⎠
4-19
Measurements
Non Return to Zero Measurements Mode
Low
ER % = ------------ × 100
High
CorrectedER % = ER % – ERCF
Where ERCF = Extinction ratio correction factor.
Control Panel
Click Extinction Ratio on the left side of the display.
API Call
FGNRZExtinctionRatioAPI
ActiveX
MeasNRZExtinctionRatio
MeasNRZExtinctionRatioDB
4-20
Measurements
Non Return to Zero Measurements Mode
Extinction Ratio with histogram
This measurement displays the same results as the Extinction Ratio
measurement; additionally it displays the distribution of sample points
within the NRZ window using a histogram.
Control Panel
Click Ext. Ratio/Hist. on the left side of the display.
API Call
N/A
ActiveX
MeasNRZExtinctionRatioHist
MeasNRZExtinctionRatioDBHist
Eye Height
Eye Height is the vertical opening of the eye and measured as the difference between the - 3 sigma value of the high samples within the
NRZ window and the +3 sigma value of the low samples within the
NRZ window.
Eye Height = ( P top – 3σ top ) – ( P base + 3σ base )
4-21
Measurements
Non Return to Zero Measurements Mode
Control Panel
Click Eye Height on the left side of the display.
API Call
FGNRZEyeHeightAPI
ActiveX
MeasNRZEyeHeight
Eye Amplitude
Eye amplitude is the difference between the logic 1 level and the logic
0 level histogram mean values of an eye diagram measured within the
NRZ window.
amplitude = high – low
4-22
Measurements
Non Return to Zero Measurements Mode
Control Panel
Click Eye Amplitude on the left side of the display.
API Call
FGNRZEyeAmplitudeAPI
ActiveX
MeasNRZEyeAmplitude
Eye Width
Eye Width is the horizontal opening of the eye.
Eye Width = ( T cross2 – 2σ cross2 ) – ( T cross1 + 2σ cross1 )
4-23
Measurements
Non Return to Zero Measurements Mode
.
Control Panel
Click Eye Width on the left side of the display.
API Call
FGNRZEyeWidthAPI
ActiveX
MeasNRZEyeWidth
OMA (Optical Modulation Amplitude)
Optical modulation amplitude is defined as the difference in power
between the high and low levels as measured within the OMA window,
centered on the crossing.
OMA = P 1 – P 0
4-24
Measurements
Non Return to Zero Measurements Mode
Control Panel
Click OMA on the left side of the display.
API Call
FGNRZOMAAPI
ActiveX
MeasNRZOMA
4-25
Measurements
Non Return to Zero Measurements Mode
Low
Low measures within the NRZ window the average of LL samples during the interval defined by the window.
Control Panel
Click Low on the left side of the display.
API Call
FGNRZLowAPI
ActiveX
MeasNRZLowLevel
4-26
Measurements
Non Return to Zero Measurements Mode
High
High measures within the NRZ window the average of HH samples
during the interval defined by the window.
Control Panel
Click High on the left side of the display.
API Call
FGNRZHighAPI
ActiveX
MeasNRZHighLevel
4-27
Measurements
Non Return to Zero Measurements Mode
Signal-to-Noise Ratio
Signal- to- Noise is the ratio of the signal difference between high level
and low level relative to the noise present at both levels.
Control Panel
Click Signal to Noise on the left side of the display.
API Call
FGNRZSNRAPI
ActiveX
MeasNRZSNR
4-28
Measurements
Non Return to Zero Measurements Mode
Bit Rate
The value returned depends on the reference clock selected.
• Recovered: the measured frequency of the recovered clock.
• External: the measured frequency of the applied reference clock. The
Control Panel also displays the implied bit rate.
• Internal: the bit rate as set..
Control Panel
Click Bit Rate on the left side of the display.
API Call
FGNRZBitRateAPI
ActiveX
GetBitRate
Rise Time
The Rise Time measurement is made on the rising edge between the
thresholds selected by the NRZ Threshold configuration setting.
Choices are 10% to 90%, 20% to 80%, and 30% to 70% of either the
total Waveform Amplitude (Full Period Top and Base), or the NRZWindowed High and Low Levels.
4-29
Measurements
Non Return to Zero Measurements Mode
Rise Time = X high threshold – X low threshold
Control Panel
Click Rise Time on the left side of the display.
API Call
FGNRZRiseTimeAPI
ActiveX
MeasNRZRiseTime
4-30
Measurements
Non Return to Zero Measurements Mode
Fall Time
The Fall Time measurement is made on the falling edge between the
thresholds selected by the NRZ Threshold configuration setting. Choices
are 10% to 90%, 20% to 80%, and 30% to 70% of either the total Waveform
Amplitude (Full Period), or the NRZ- Windowed High and Low Levels.
Fall Time = X low threshold – X high threshold
Control Panel
Click Fall Time on the left side of the display.
API Call
FGNRZFallTimeAPI
ActiveX
MeasNRZFallTime
4-31
Measurements
Non Return to Zero Measurements Mode
Crossing Percentage
Crossing Percentage is the signal level at the point where the rising
and falling edges cross expressed as a percentage of the signal amplitude.
Control Panel
Click Crossing Pct. on the left side of the display.
API Call
FGNRZCrossingPercentAPI
ActiveX
MeasNRZCrossingPercent
Duty Cycle Distortion
Duty Cycle Distortion is the time separation between the rising edge
and falling edge at the mid- level of the eye diagram. The 50% threshold
is calculated either on the total waveform amplitude (full period), or
the NRZ- windowed high- low levels.
4-32
Measurements
Non Return to Zero Measurements Mode
Control Panel
Click Duty Cycle Distortion on the left side of the display.
API Call
FGNRZDutyCyDistortionAPI
ActiveX
MeasNRZDutyCycleDistortion
4-33
Measurements
Non Return to Zero Measurements Mode
Edge Jitter
Edge Jitter is a measure of the difference in the time domain of the
earliest transitioning sample and the latest. This is measured at the
crossing point.
Control Panel
Click Edge Jitter on the left side of the display.
API Call
FGNRZJitterAPI
ActiveX
MeasNRZJitterPeakToPeak
MeasNRZJitterRMS
4-34
Measurements
Non Return to Zero Measurements Mode
Single-Edge Jitter
Single- Edge rising Jitter is a measure of the difference in the time
domain of the earliest transitioning rising point and the latest. This is
measured at the crossing point. The single edge falling jitter is also
available.
Control Panel
Click Single Edge Jitter on the left side of the display.
API Call
FGNRZFallJitterAPI, FGNRZRiseJitterAPI
ActiveX
MeasNRZSingleEdgeJitterPP
MeasNRZSingleEdgeJitterRMS
4-35
Measurements
Non Return to Zero Measurements Mode
Deterministic Jitter
Deterministic Jitter is a measure of data dependent jitter.
Control Panel
Click Deterministic Jitter on the left side of the display.
API Call
FGNRZDeterministicJitterAPI
ActiveX
MeasNRZDeterministicJitterRise
MeasNRZDeterministicJitterFall
4-36
Measurements
Non Return to Zero Measurements Mode
Crossing Hist.
Crossing Histogram shows the distribution of samples at the crossing
point.
Control Panel
Click Crossing Hist. on the left side of the display.
API Call
N/A
ActiveX
N/A
4-37
Measurements
Non Return to Zero Measurements Mode
ISI Histogram
The ISI Histogram shows the distribution at the crossing points for
LLH, HLH, HHL and LHL patterns separately.
Control Panel
Click ISI Histogram on the left side of the display.
API Call
N/A
ActiveX
N/A
4-38
Measurements
Eye Mask Test Measurements Mode
Eye Mask Test Measurements Mode
The N2100B allows the Host Application to select an eye mask and
also to perform a comparison of the acquired NRZ Eye against the
selected eye mask.
Use Mask Test mode to determine whether the signal meets defined
mask requirements and calculate how much margin there is before
failure.
• “Mask Selection - Industry Standard and User Defined” on page 4- 41
• “Pass/Fail” on page 4- 43
• “Find Max Margin” on page 4- 45
4-39
Measurements
Eye Mask Test Measurements Mode
Click Select Mask to select a Mask.
Select mask
Mask test pass?
Find mask
margin for
failed points
User selected
mask margin
0% margin
mask
Calculated
mask margin
Failed points
Figure 17. Mask Tests Mode Control Panel
Failing Points Button
If you click Failing Points, the software reports any points that fall in
the mask, or in the mask with margin applied, if a margin is selected.
This reporting will be recalculated each time the Start Acquisition button is clicked or continuously in Repetitive Acquisition Mode. In the
Trace Display, these reported points are highlighted in yellow. In the
Results Display Area, the appropriate text is displayed: Test PASSED, or
Test FAILED with the number of points that failed subtotaled by individual mask region. For example:
4-40
Measurements
Eye Mask Test Measurements Mode
Test FAILED: 7 points of 16384 [45% margin]
Center 7, Upper 0, Lower 0
Find Max Margin Button The Find Max Margin button allows you to find the maximum mask
margin that results in no more than a specified number of failing
points. The figure below shows an example where the user wants to
determine the maxmimum mask margin available for the highest number of failing points, but no more than 25. The answer in this example
is +51% for 22 points. Setting the margin below this level would result
in less failing points total. Setting the margin above this level would
result in fewer failing points.
Figure 18. Find Max Mask Margin
The Find Max Margin calculates results without the need to reacquire
data. This allows you to converge on a desired answer by trying different threshold values and the clicking the Find Max Margin for each
one. When no data are present the error message NO DATA is displayed.
Mask Selection - Industry Standard and User Defined
Clicking Select Mask invokes the Select Mask dialog box shown below.
This allows you to select from a set of industry standard masks, or
create a custom mask for your specific requirements. The predefined
masks are listed in Table 3 on page 4- 43.
4-41
Measurements
Eye Mask Test Measurements Mode
Figure 19. Selecting a Mask - Industry Standard or User Defined
The Copy Another Mask button allows you to easily create a custom
mask by using an existing mask as a starting template. The existing
mask can be either a standard mask or one that you previously
defined. The new Mask is saved under a different mask number (#)
after clicking OK.
4-42
Measurements
Eye Mask Test Measurements Mode
.
Table 3. DCA Predefined Mask Definitions
MASK
x1
x2
x3
x4
stm-1
0.15
0.35
0.65
stm-4
0.25
0.4
0.6
0.4
0.6
stm-16
x5
x6
y0
y1
y2
y3
y4
y5
y6
0.85
0.5
0.2
0.8
Full UI
0.75
0.5
0.2
0.8
Full UI
0.5
0.25
0.75
Full UI
Window
Gb Ethernet
0.22
0.375
0.625
0.758
0.5
0.2
0.8
-0.2
1.3
20%
Fibre channel (1x)
0.215
0.4
0.6
0.785
0.5
0.2
0.8
-0.2
1.3
20%
Fibre channel (2x,4x)
0.22
0.4
0.6
0.78
0.5
0.2
0.8
-0.2
1.3
20%
Fibre channel (8x)
0.250
0.450
0.550
0.750
0.500
0.320
0.680
-0.400
1.400
Fibre channel (10x)
0.300
0.400
0.600
0.700
0,500
0.250
0.750
-0.400
1.400
10 Gb
Ethernet
0.250
0.450
0.550
0.750
0.500
0.250
0.750
-0.400
1.400
0.400
0.600
0.350
0.650
20%
20%
0.280
0.720
20%
Pass/Fail
A mask test fails if any points fall inside the mask regions. When a
positive mask margin is applied, the mask regions are enlarged so that
points falling within the original mask or the margin fail. When a neg-
4-43
Measurements
Eye Mask Test Measurements Mode
ative margin is applied, the mask regions are reduced so that only
points that fall within the reduced mask (original mask minus margin)
fail.
Control Panel
The Eye Mask Test is done automatically after acquisition in this mode.
Click on the Failing Points button for this display.
API Call
FGDCAMaskTestAPI
ActiveX
GetMaskTestFailPoints
4-44
Measurements
Eye Mask Test Measurements Mode
Find Max Margin
This measurement finds the maximum mask margin value that results
in no more than the specified number of sample points that fail the
mask test.
Control Panel
After an Acquisition, enter the desired number of failing points, and
click on Find Max Margin button to perform the test. This sequence can
be repeated on the same set of test data to determine the margin for
several failing point thresholds.
API Call
N/A
ActiveX
FindThresholdMaskMargin
4-45
Measurements
Multiple Measurements Mode
Multiple Measurements Mode
The Multiple Measurements Mode allows you to gather and view all
possible measurements on the eye diagram from one capture. After you
select the measurements that you want to view, the Control Panel displays the results concurrently.
Figure 20. Multiple Measurements Mode Control Panel
4-46
Measurements
Multiple Measurements Mode
Enabling Statistics on the The Multiple Measurements tab includes access to the statistics funcMultiple Measurements tion. When Calc Stats is selected the following values are collected over
Tab
the measurements made during the current repetitive acquisition operation, for each of the parameters selected.
•
•
•
•
Minimum
Maximum
Average
Standard deviation
Clicking Meas / Stats switches the text window between the measured
values and the statistics.
When the statistics results are displayed, clicking the Copy Text button
copies the data displayed. Fields are separated by a single tab character.
Viewing and moving the measurement results window
1 Click the release button to detach the measurement results window.
Release button for
detatchable
results winMeasurements
results area
2 Drag the measurement results window to a more convenient viewing
position.
3 Drag the bottom border of the window to enlarge the area for viewing
4-47
Measurements
Multiple Measurements Mode
multiple measurement results without scrolling through the data.
4-48
Measurements
Pattern Sequence D. J. Measurements Mode
Pattern Sequence D. J. Measurements Mode
The N2100B is capable of performing a pattern capture without the
need for a pattern trigger.
A Pattern Sequence D.J. mode offers deterministic (pattern- dependent) jitter measurements on selected multibit sequences in the
acquired data pattern. between specified bit sequences. This mode only
applies to data acquired with pattern acquisition. For information on
pattern acquisition, refer to “Pattern Acquisition” on page 6- 1. This
section includes procedures for the following measurements:
• “To acquire a pattern” on page 4- 52.
• “To apply software filters with pattern acquisition” on page 4- 53.
• “D.J. based on Pattern Acquisition” on page 4- 53.
Basic deterministic jitter measurement measures the average jitter
between rising transitions 001 (transition following a steady low), and
101 (transition following a transition), and between falling transitions
110 and 010.
Pattern sequence deterministic jitter measurement measures the average jitter between two specified transition sequences up to 16 bits in
length.
4-49
Measurements
Pattern Sequence D. J. Measurements Mode
The DCA Pattern Sequence D.J. mode includes the
following measurement capabilities:
• Standard deterministic measures the time between the earliest- occurring and latest- occurring rising transition, and the time between the
earliest- occuring and latest- occurring falling
transition.
• Custom deterministic jitter that is the deterministic jitter between two specified bit pattern sequences, or between one specified bit pattern
sequence and all other bit pattern sequences.
The transition labeled, "Fast" refers to the transitions normally called HLH and LHL. The transition labeled, "Slow" refers to the transitions
normally called LLH and HHL. These controls
allow you to redefine these transitions with arbitrary bit sequences for
the purposes of this measurement. For example, a sequence consisting
of a low- to- high transition preceded by exactly four low states would
be expressed as 100001.
For the "Slow" transition, the Length selector refers to the number of
low states preceding a rising transition, or the number of high states
preceding a falling transition. The sequence 1001 corresponds to length
2, and the sequence 100001 corresponds to length 4.
Results of these measurements are reported as follows:
D.J., 1001 to 101 Rise 7.3 ps, Fall 4.3 ps
The eye display changes so that the only transition samples shown are
those that match the specified patterns. If display colors are enabled,
these are shown in the corresponding transition colors.
If you select the All choice, the measurement is made between the
specified "Fast" pattern and all patterns of the form 10…01 with
lengths between 2 and 11. When no transitions are found for a specified length, no result is reported for that length. The following example
is a typical output from this measurement:
Reference pattern 101
Length 2
Length 3
Length 4
Length 5
Length 6
4-50
Rise
Rise
Rise
Rise
Rise
7.3 ps, Fall 4.3 ps
4.5 ps, Fall 4.7 ps
1.3 ps, Fall 4.2 ps
- 0.7 ps, Fall - 0.2 ps
- 1.9 ps, Fall 5.0 ps
Measurements
Pattern Sequence D. J. Measurements Mode
Length 7
Length 8
Rise - 5.4 ps, Fall 9.5 ps
Rise - 7.0 ps, Fall 5.6 ps
4-51
Measurements
Pattern Sequence D. J. Measurements Mode
To acquire a pattern
Control Panel
To display the following pattern sequence, perform the following steps:
1 In the Module Config subpanel, set the Pattern Acquisition option to
Enabled (No filter) or the the desired filter selection.
2 Set either the pattern length or the pattern type.
3 To view the acquired pattern, set the instrument to the Oscilloscope
Meas. mode and ensure that the Pattern option is selected.
API Call
N/A
ActiveX
PatternAcqFilterSel
PatternAcqLength
LoadPatternAcqFilterFile
LoadPatternAcqFilterFileByValue
MeasPatternAcqDetJitterRise
MeasPatternAcqDetJitterFall
MeasPatternAcqStdDetJitterRise
MeasPatternAcqStdDetJitterFall
4-52
Measurements
Pattern Sequence D. J. Measurements Mode
To apply software filters with pattern acquisition
Control Panel
To apply the software filters, perform the following steps:
1 In the Module Config sub panel, set the Pattern Acquisition Option to
the desired filter.
2 Set either the pattern length or the pattern type.
API Call
ActiveX
N/A
PatternAcqFilter
Sel PatternAcqLength
LoadPatternAcqFilterFile
LoadPatternAcqFilterFileByValue
D.J. based on Pattern Acquisition
The Pattern Acquisition capability provides a Standard and a Custom
Deterministic Jitter measurement.
The Standard D.J. measurement displays the maximum time deviations
found among all rising transitions, and among all falling transitions,
without the requirement of selecting a sequence. The result is one
number each for rising and falling transitions.
The Custom deterministic jitter, as derived by two particular bit
sequences, can be extracted using the pattern sequence capture option.
You are allowed to define two sequences to compare. You are also
4-53
Measurements
Pattern Sequence D. J. Measurements Mode
allowed to define one base sequence which can then be compared
against all other sequences. The difference in ps of the crossing point
between the two sequences under test is returned.
Control Panel
Select the Pattern Sequence DJ option in the main panel.
API Call
N/A
ActiveX
MeasPatternAcqStdDetJitterRise
MeasPatternAcqStdDetJitterFall
MeasPatternAcqDetJitterRise
MeasPatternAcqDetJitterFall
4-54
5
Introduction 5- 2
Program Settings 5- 3
Copying and Saving Displayed Test Results
Displaying and Saving Results
5- 13
Displaying and Saving Results
Introduction
Introduction
This chapter describes how to display the measurement results on the
Control Panel and how to copy or save those displayed results.
• “Program Settings” on page 5- 3.
• “Copying and Saving Displayed Test Results” on page 5- 13.
5-2
Displaying and Saving Results
Program Settings
Program Settings
Use the Program Settings button on the Control Panel to change how
data are displayed in the Control Panel. The dialog box has two tabs
(General and Density Display), with Display Colors and Mask Display
settings common to both.
Figure 21. Program Settings dialog box
5-3
Displaying and Saving Results
Settings on the General tab
Display Colors
Controls the presentation of the sampled data through the following
choices:
• Display Off. Prevents the window from being displayed, saving the
time required to format and present the measured data.
• No Colors. All samples are displayed using the same color.
• Transitions. Samples are displayed using four colors to represent the
different transition types: LL, LH, HL, and HH.
• Transitions with history. Samples are displayed using eight colors to
represent transition types: LLL, LLH, LHH, LHL, HLL, HHL, HLH, and
HHH. For this selection, the actual sample is taken on the transition
between the latter two bits.
• Sample Density. The sample is displayed based on whichever option
has been selected within the Sample Density tab.
Mask Display
Controls the mask presentation through the following choices:
• No Colors. Sets the display for the mask test to the standard blue
pixel display. In this mode, sample points failng the mask test are
highlighted.
• Selection. Sets the Mask display to the selection in the display colors
option.
Settings on the General tab
Display Traces
Controls the trace presentation through the following choices:
• All. All transitions are displayed.
• Sel'd. Only transitions selected using the LLL- HHH buttons are displayed.
• LLL- HHH. Allows you to switch individual traces On or OFF. The
warning, 'Some traces not displayed' appears above the Trace Display
Area when not all of the transitions are selected.
• Clr All. Sets all the LLL- HHH buttons to the OFF state.
• Set All. Sets all the LLL- HHH buttons to the ON state.
• Trans. Sets all the LLL- HHH buttons that represent a LH or HL tran-
5-4
Displaying and Saving Results
Settings on the General tab
sition to ON.
Y-Scale
Controls the presentation of the Y- Axis through the following choices:
• Relative. Only displays the used part of the Y- axis.
• Fixed. Always includes the zero value on the Y- axis.
Units
Controls the unit of measurement through the following choices:
• Signal: mV (electrical) or μW (optical). For a DCA with the optical
input selected, the option is μW. For a DCA with an electrical input
selected, the option is mV. Both of these are calibrated options.
• Raw: mV. In this mode, the internal (non- calibrated) mV value of the
instrument is returned.
Display Background
Controls the eye diagram background lighting through the following
choices:
• Dark. The eye diagram has a dark background.
• Light. The eye diagram has a light background.
• Lt (Copy). The eye diagram is displayed with a dark background, but
it is copied to the Windows clipboard with a light background for
pasting into a document.
Pixel Size
Small/Large. Selects the pixel size of the sample points that are drawn:
"small" gives a more precise display; "large" gives a brighter display.
Repetitive Results Text
Controls the presentation of repeated measurement results through the
following choices:
• Overwrite. During a repetitive acquisition sequence, each measurement is overwritten by the subsequent measurement in the display
window.
• Add. Each measurement overwrites the previous measurement in the
display window.
• Add + Time. The list of measurements is time stamped. As each measurement result is added to form a list in the display window along
with a start time, an end time, and an elapsed time of the most recent
data collection. A time stamp is also included in the results every 30
5-5
Displaying and Saving Results
Settings on the Density Display Tab
seconds.
Crossing Histogram
Window
Size (%). Defines the size of the sample window used to display the
crossing histogram. Defined as a percentage of amplitude.
Offset (%). Allows the window to be offset from the center by a percentage of the amplitude (+ or - ).
Show Y-Axis Scale
Sets whether the Y- axis scale is displayed.
Update display all at once Controls the elimination of possible display flicker through the following choices:
• Points are drawn individually as each point position is calculated.
• Points are stored in an internal buffer and displayed all at once at
the end of each acquisition/calculation cycle.
Settings on the Density Display Tab
The Density Display tab allows you to adjust the sample density display options.
Sample Density Color
Scheme
Controls the color of the sample density through the following choices:
• Colors 1. Blue and violet represent the highest sample densities.
• Colors 2. Orange and yellow represent the highest sample densities.
• Intensity. Different intensities of blue and green represent different
densities.
• Single Threshold. A two- color display with adjustable density threshold.
Display
Sets how you will view the acquired data through the following
choices:
• Signal. View the Eye Diagram Display.
• Density Distribution. View a Histogram of the samples.
5-6
Displaying and Saving Results
Settings on the Density Display Tab
Resolution
Selects the quantization grid that assigns sample densities.
• Move the slider to the left to select a coarser grid.
• Move the slider to the right to select a finer grid.
Color Intensity Display
Sets the intensity of displayed pixels through the following controls:
• Contrast adjusts the intensity of the low- density pixels.
• Weighting controls the assignment of display colors to bins. Moving
the slider to the right makes high- density samples relatively more
prominent.
5-7
Displaying and Saving Results
Settings on the Density Display Tab
Threshold Display
Sets the display of pixels associated with the threshold through the
following controls:
• Contrast adjusts the intensity of pixels that indicate densities below
the threshold.
• Threshold sets the display threshold value. An up/down button control allows a fine adjustment of this value.
Sample Density Display
Information
The data sample density display is a software tool that enables the eye
diagram to display colors based upon the population of sample values.
The following set of diagrams shows the variety of displays:
• The left diagram shows the typical single- color eye diagram of samples acquired by the DCA.
• The middle diagram shows multi- color pixels based on the types of
transition to which the samples belong.
• The left diagram shows colored pixels based on the number of samples that occur at the location of that pixel.
Figure 22. Variety of Displays for Sample Density Information
To easily identify a data point that occurs many times in the sample
data, you can set the pixel color to be different from a data point that
occurs just once. With multiple pixel colors available in the DCA, you
can control the display of sample density through the following
sequence:
1 Define multiple thresholds for the number of times a given data point
occurs.
2 Indicate each frequency- of- occurrence group in its own color.
5-8
Displaying and Saving Results
Settings on the Density Display Tab
The DCA restricts the data samples to a fixed quantization grid with
X- and Y- axes, where each sample's "density," and resulting color is
based on the number of samples that fall into the same grid square.
The grid remains the same even when the display window size
changes, retaining the same appearance.
NOTE
Sample densitydisplay affects only how the data samples are displayed. This display does not
affect measurements made on the acquired data or the export of acquired data.
Choosing the quantization grid
The quantization grid size is selectable, giving the you some control
over the way samples are grouped:
• Choosing a coarse grid results in the highest frequency-of-occurrence values but
minimal resolution in the X- and Y-axes.
• Choosing a fine grid gives the best resolution in the X- and Y-axes but reduces the
likelihood that samples will fall into the same grid square.
This concept is best shown with the following histograms of sample
density.
Figure 23. Histograms of Quantization Grid Choices
• The histogram on the left shows the finest grid, resulting in the fewest “coincident” samples. In this distribution, half of the samples have unique values, another
fourth occur just twice.
• The histogram on the right shows the coarsest grid. This distribution has a substantially greater fraction of samples that occur multiple times.
Assigning Colors to Sample Densities
The DCA offers you eight display colors and a choice in assigning colors to sample bins:
• You can assign colors so that each color covers the same number of bins, resulting
in every sample being displayed in the higher frequency of occurrence colors. (See
5-9
Displaying and Saving Results
Settings on the Density Display Tab
the left histogram and eye diagram in the following figure.)
• You can assign each bin a separate color until you run out of colors, resulting in
the highest sample densities being displayed very prominently. An intermediate
setting gives results that are probably the most appealing. (See the right histogram
and eye diagram in the following figure.)
Figure 24. Results of Assigning Colors to Sample Densities
Choosing a color scheme
The DCA offers you three color schemes to distinguish between sample
densities: two that use sets of colors, and one that uses intensity levels. In the previous figure, the right eye diagram uses a color- based
scheme where orange and yellow are assigned to the highest sample
densities. In the following figure, the left eye diagram uses a colorbased scheme where blue and violet represent the highest sample densities. The right eye diagram uses an intensity- based scheme, displaying only green and blue.
5-10
Displaying and Saving Results
Settings on the Density Display Tab
Figure 25. Examples of Color-Based and Intensity-Based Schemes
Adjusting the density threshold display
One further display variant uses two colors and a single, adjustable
density threshold.
Figure 26. Example of Adjustable Density Threshold
Adjusting the contrast of pixel intensity
A Contrast adjustment controls the pixel intensity, allowing you to display the high- frequency- of- occurrence samples most prominently. This
adjustment has the greatest effect on the low- density colors. The following examples show the effect of the contrast adjustment.
5-11
Displaying and Saving Results
Settings on the Density Display Tab
Figure 27. Example of Contrast Adjustment
5-12
Displaying and Saving Results
Copying and Saving Displayed Test Results
Copying and Saving Displayed Test Results
The DCA Control Panel provides a convenient mechanism to capture
displayed measurement results, including acquisition data, statistics,
visual traces, and test timestamp information for documentation purposes. Three Control Panel buttons are used for this purpose:
• Data
• Copy Graph
• Copy Text
Data Button
Use this feature to view or copy the actual samples in a text format.
Different data sorting, filtering, and display options are available. The
two examples shown below are sampled point data and pattern data in
hexadecimal format.
Figure 28. Examples of Two Data Samples
5-13
Displaying and Saving Results
Copying and Saving Displayed Test Results
Sorting Data
• To sort the samples based on the order in which they were collected,
select Point.
• To sort the samples based on the X value (their time position within
the eye), select X Val.
• To sort the samples based on what type of transition they belong to,
select Trans. Within transitions, samples are sorted based on X value.
• To put the mask fail points at the top of the list, select Fail. Within
the pass and fail group, the samples are sorted based on X value.
Copying and Saving Data
• To copy the data to the clipboard, click Copy.
• To save the data to a text (*.txt) file, click the Save as drop- down
arrow and select Text, then click Save.
• To save the data to a rich text format (*.rtf) file, click the Save as
drop- down arrow and select Rich text, then click Save.
• To save the data to a tab separated format (*.txt) file that can easily
be brought into a spreadsheet application, click the Save as dropdown arrow and select Tab sep’d, then click Save.
Displaying Transitions
• To display all transitions, click the transition selection button, All.
• To only display transitions that you select using the transition buttons (LLL, LLH, LHH, HHH ....), click Sel'd. These transition buttons
allow individual transitions to be switched on or off.
Displaying Qualified Samples
• To display all qualifying samples in the data results, select All Y val.
• To only display samples that fall in the middle 40% of amplitude
readings, select Mid 40%.
• To display the individual data samples, select Samples.
• To display the reconstructed pattern sequence in hexadecimal format
(pattern acquisition only), select Pattern Hex.
5-14
Displaying and Saving Results
Copying and Saving Displayed Test Results
• To display the reconstructed pattern sequence in binary format (pattern acquisition only), select Pattern Bin.
NOTE
When data are aquired with a mask test and displayed in binary, pattern sequence bits containing
samples that fail the mask test are underlined in the display.
Copy Graph Button
The Copy Graph button copies the graph and all the text data from the
most recent acquisition run as displayed on the screen. The image is
available as an object on the PC clipboard so you may paste it into a
word processing program for future use.
Copy Text Button
The Copy Text button copies all the information from the results display window in text format. This information can be pasted into any
program such as MS Word or Notepad as in this example:
Custom Annotations
Meas.
Average
Minimum
Maximum
Std. deviation
Amplitude
420.637
407.250
429.750
12.222
Peak to peak
514.087
491.250
528.563
15.871
RMS
197.393
190.520
201.963
6.251
Average
384.790
384.553
384.977
0.178
OMA
419.203
404.144
429.215
13.435
You can annotate the graph image caption. Hold down the keyboard’s
CTRL key while clicking Copy Graph or Copy Text. This displays the
dialog box shown in Figure 29 on page 5- 16. The buttons and text
entry field provide a way for you to append additional text to the
visual or textual data captured, as shown in Figure 30 on page 5- 16.
5-15
Displaying and Saving Results
Copying and Saving Displayed Test Results
Figure 29. Annotate Graph and Image Caption Dialog Box
Figure 30. Annotated Graph Example
5-16
Displaying and Saving Results
Copying and Saving Displayed Test Results
The Copy as Excel- compatible array is available only when the following conditions exist:
• Copy Text option is selected when the Multiple Measurements tab is selected.
• Repetitive Results Text is set to 'Add' or 'Add+Time' option in Prog Settings panel.
This option reformats the list of results acquired during a repetitive
measurements run into a tab- separated array of data. This data can
then be directly copied into an Excel workbook or other spreadsheet.
The data are formatted as one column per measured parameter and
one row per acquisition. The first row of the table has parameter
labels. In 'Add+Time' mode, an additional column is populated with the
timestamps of the measurement block at 30 second intervals.
Amplitude
Base
Top
Rise Time
Fall Time
Time
46.1
8.9
55.0
134.6
141.8
18:03:20
46.1
8.9
55.0
136.7
151.1
46.1
8.9
55.0
128.6
142.0
46.1
8.9
55.0
110.1
150.2
50.7
6.6
57.3
159.7
163.5
46.1
8.9
55.0
137.1
148.7
46.1
8.9
55.0
148.9
151.0
18:03:50
5-17
Displaying and Saving Results
Copying and Saving Displayed Test Results
5-18
6
Pattern Acquisition
6- 2
Pattern Acquisition
Pattern Acquisition
Pattern Acquisition
Pattern Acquisition
The DCA normally samples its input signal over an interval of one bit
time (unit interval, or UI) and reconstructs an eye diagram that is an
aggregate of all the signal's states and transitions. In pattern sequence
acquisition, the interval over which the DCA collects samples is equal
to as many bit times as the full pattern sequence length. The reconstructed result is a representation of the entire pattern sequence.
Although the displayed result is an aggregate of all the sequence periods over which the samples were taken, it is coherent with the
sequence. Therefore, any individual transition in the reconstructed
sequence consists only of samples from the corresponding transition of
the input signal.
The reconstruction of the full pattern sequence allows you to do the
following:
• Measure pattern- dependent (deterministic) jitter.
• Determine the effect on the shape and timing of a transition of the data
values that precede it.
• Use digital filtering techniques to realize desired filter characteristics.
Filter functions can be applied that are specific for the signal line rate
and desired filter bandwidth; these can also take into account the
DCA's inherent frequency response so that the overall response is as
close to optimum as possible. When the pattern acquisition feature is
enabled, any specified filtering is applied to the reconstructed
sequence and performed measurements.
Guidelines for using pattern acquisition
• The input pattern must be absolutely repetitive to allow a reconstruction
from samples taken over many periods. The application of this technique is typically limited to unframed signals containing pure data.
• The sequence must be no longer than some upper limit. The sequence
can range from a few bits in length up to PRBS 211 –1.
6-2
Pattern Acquisition
Pattern Acquisition
• The signal reconstruction is an averaging process so measurements of
random (non- periodic) jitter made on the reconstructed signal do not
accurately reflect the jitter present on the input signal.
• The reconstruction process does not preserve individual data samples.
It is possible (due to averaging) that an individual sample that may fail
a mask test will not appear in the reconstructed output.
• Implementation of the filter with computational techniques requires significant processing. When this capability is enabled, instrument performance will be slower than when it is disabled.
Operation
Configuring the pattern sequence acquisition involves specifying the
following characterisitcs:
• Pattern sequence length
• Filter to be applied, if any
Control Panel Application The Pattern Acquisition control has selections for Off, Enabled (No filter), and a list of software filters (if any) that are available for the
module.
The Pattern control specifies the length of the repeating pattern
sequence (PRBS 27- 1, PRBS 29- 1, PRBS 211- 1, K28.5, K28.7), and Set
Length for patterns of arbitrary length. When Set Length is selected,
the field to the right is enabled so you can enter the pattern length.
The length must be between 4 and 2047.
NOTE
The recovered clock with a sub-rate pattern of K28.7 is usable with the API, but not with the
Control Panel interface.
Acquisition
With Pattern Acquisition enabled, any data acquisition operation
acquires data over the sequence length, reconstructs the sequence, and
applies the selected filter. The reconstructed eye diagram is displayed
and any measurements invoked are made on the reconstructed eye.
6-3
Pattern Acquisition
Pattern Acquisition
NOTE
If the Pattern Length setting does not match the actual sequence length, or if the input signal is
not absolutely repetitive, the sequence reconstruction will be incoherent and the eye display will
appear distorted.
Displaying Acquired Data The acquired data pattern can be displayed and processed through the
in Text Format
Data Samples dialog box described in “Displaying and Saving Results”
on page 5- 1. Multiple display modes, sorting fields, and file manipulation functions are available.
Graphical Sequence
Display
The Oscilloscope Measurements tab includes a selector for Eye or Pattern display. (The Pattern selection is enabled only when pattern
sequence acquisition is active.)
If you select Pattern, the display changes to show the entire reconstructed sequence. You can use the Time Axis controls to scroll
through the sequence and display it with suitable horizontal resolution.
Figure 31. Pattern Display in Oscilloscope Mode
Saving Pattern Sequence To ensure the simulator correctly interprets the data in a saved file of
Data for Use with the
data collected with the pattern sequence acquisition, follow these
Simulator
steps:
1 Before saving the data, set the Measurements selector to Oscilloscope
Measurements.
2 Set the Display selector to Pattern.
6-4
Pattern Acquisition
Software Filters with Pattern Acquisition
Measurements (General) With pattern sequence acquisition enabled, measurements are made on
the reconstructed sequence but otherwise function normally.
Software Filters with Pattern Acquisition
The Pattern Acquisition feature offers the ability to apply software filters. A software filter is a digital filter that matches a specific desired
frequency response and rise time when convolved digitally with the
measured response of the DCA. Each filter is specified by a file containing an array of numerical data. The digital filter is calculated from
the deconvolution of the measured known impulse response of the
DCA and the impulse response of the desired filter.
NOTE
Software filters are only available for optical input. The filters are module specific and cannot be
shared between separate modules.
Guidelines for using digital filters
• Agilent digital filters are designed for use when the DCA is configured
with the unfiltered optical path and in the pattern acquisition mode.
The Input Path and Filter Selection must be set to the unfiltered path.
If the DCA is not set to the unfiltered path, the filters are calculated
for use with the highest- bandwidth optical filtered path so that path
should be selected.
• The filter file can only be applied to a signal whose line rate is the
same as that of the filter. Applying a filter file to a different line rate
will result in erroneous data
• Incorrectly applying digital filters can lead to unpredictable and erroneous results. When using Agilent supplied digital filters, ensure the
configuration for capturing the data is set correctly.
NOTE
Because of the inherent averaging of the data when using the pattern acquisition mode,
measurements of random jitter are not valid. It is recommended that you use hardware filters
when performing jitter measurements.
Filter Files
Each DCA has filter files constructed to meet the target rates and
bandwidths as the DCA's physical input filters. These files are stored
on the module, and are automatically copied to the host for use by the
6-5
Pattern Acquisition
Software Filters with Pattern Acquisition
filtering software. Available filter selections appear in the Pattern
Acquisition control on the Configuration Settings page, following the
<Enabled (No Filter)> choice.
Filter List
Filter selections are listed by bandwidth. If an optional 'optstring' field
is present, that string appears in parentheses after the bandwidth
specifier.
Filter File Location
Filter files are stored in the DCA Filter Files subdirectory of the
N2100B installation directory. If the N2100B software is installed to
the default location, this is C:\Program Files\Agilent\N2100B\DCA Filter Files.
Filter File Format
File names have the format: [email protected], where:
• xxxxxxx is the serial number of the DCA to which the filter file applies.
• string is a numeric or mostly- numeric character string related to the filter bandwidth. For example:
• 4250 suggests 4250 Mb/s.
• 2488p32 suggests 2488.32 Mb/s, where p represents the decimal point.
• optstring is a character string used to distinguish two files that might
otherwise have the same name. (It might be used to distinguish between
filters of different length or order.)
6-6
7
Introduction 7- 2
How the Mask and Mask Margin are defined
Agilent Mask Margin Rules 7- 6
Tektronix Mask Margin Rules 7- 8
7- 3
Mask & Mask Margin Definitions
Mask & Mask Margin Definitions
Introduction
Introduction
This chapter describes the structure of masks used for measurements
that determine if acquired sample points pass or fail a defined test.
• “How the Mask and Mask Margin are defined” on page 7- 3.
• “Agilent Mask Margin Rules” on page 7- 6.
• “Tektronix Mask Margin Rules” on page 7- 8.
7-2
Mask & Mask Margin Definitions
How the Mask and Mask Margin are defined
How the Mask and Mask Margin are defined
A mask defines an area in the reconstructed image of the eye where
no sample points should exist. It always consists of an area in the
center of the eye and can optionally have areas above and below the
eye.
Figure 32. Mask Definition
All y- axis values are defined relative to the measured base level (y =
0) with the measured top level being y = 1. The Y level can be determined based on either of the following:
• Top and bottom of the entire eye
• High and low based on a certain percentage around the middle of the
7-3
Mask & Mask Margin Definitions
How the Mask and Mask Margin are defined
eye
All x- axis values are defined relative to the left side crossing point (x
= 0) with the right side crossing point being x = 1.
The center mask can have 4, 6, or 10 sides depending on the checkbox
settings in the setup form below. This form also allows you to setup
the position of all the vertices of the mask.
Figure 33. Select Mask Dialog Box
• Selecting none of the checkboxes sets the mask to center only and the
shape to rectangular based on x2, x3, y1, y2.
• Checking Mask uses x1, x4 adds two additional points, allowing you to
define the center mask as a hexagon.
• Checking Mask used x5, x6, y5, y6 adds four additional points allowing you
to define the center mask to a decagon.
• Checking Mask uses y3, y4 adds the top and bottom mask definitions.
A validity check is performed on the coordinates before values are
accepted. One or more pop- ups appear if there is an inconsistency in
specifying the mask parameters. You must acknowledge this test before
proceeding with such values.
7-4
Mask & Mask Margin Definitions
How the Mask and Mask Margin are defined
The Margin Compatibility option allows you to choose whether a mask
that applies the margin is based on the Agilent rule or the Tektronix
rule.
Figure 34. Margin Compatibility Option
Refer to “Agilent Mask Margin Rules” on page 7- 6 and “Tektronix Mask
Margin Rules” on page 7- 8.
7-5
Mask & Mask Margin Definitions
Agilent Mask Margin Rules
Agilent Mask Margin Rules
Positive Margin
When a positive margin is applied, vertices of the inner mask are
moved by the requested proportion towards the left and right crossing
points and towards the top and base levels. However, the parameters
x2 and x4 remain constant. The top mask region (y4) moves down
toward the top level, and the bottom mask region (y3) moves up
towards the base level in proportion to the margin.
Figure 35. Agilent Mask - Positive Margin
Negative Margin
When a negative mask margin is applied, all points in the center mask
are moved in towards the center by the requested proportion.
7-6
Mask & Mask Margin Definitions
Agilent Mask Margin Rules
The Top and Bottom masks are moved away from the top and base
levels by the requested proportion.
Figure 36. Agilent Mask - Negative Margin
7-7
Mask & Mask Margin Definitions
Tektronix Mask Margin Rules
Tektronix Mask Margin Rules
Available Margin in the X In the Tektronix mask margin conventions, the available margin in the
Dimension
X (time) dimension (ΔX100) is the smaller of the following values:
• Distance from the left eye crossing to the leftmost mask vertex
• Distance from the rightmost mask vertex and the right eye crossing
Figure 37. Tektronix Rectangle Mask Margin Illustration
7-8
Mask & Mask Margin Definitions
Tektronix Mask Margin Rules
The leftmost and rightmost vertices of the center mask are moved outward by the specified percentage (M) of this available margin or
inwards if the margin is negative. In Figure 37, this is represented as
ΔX100 = Min( ΔXL, ΔXR).
Available Margin in the Y For the center mask, the available margin in the Y (amplitude) dimenDimension
sion, ΔY100 is the smaller of the following values:
• Distance from the signal base level (YB) to the lower mask edge
• Distance from the upper mask edge to the signal top level (YT)
ΔY100= Min( ΔYT, ΔYB)
You can define the YT and YB values relative to the full- UI values or a
specified window (p%) about the center of the eye.
The upper and lower edges of the center mask are moved outwards by
the specified percentage of this available margin, or inwards if the
margin is negative.
The dotted lines represent the mask margin while the full straight
mask lines represent the original mask.
Margin for Upper and
Lower Mask Regions
For the upper (lower) masks, the available margin in the Y dimention
is the distance from the mask level to the top (base) level. These
masks are moved by the specified percentage of this available margin,
toward the center mask if the margin is positive or away if the margin
is negative.
Tektronix Margin
Definition
The Tektronix margin definition requires that the upper and lower
mask regions be present and that the center region be defined to be
between the base and top levels, the upper and lower regions outside
of these levels.
7-9
Mask & Mask Margin Definitions
Tektronix Mask Margin Rules
Margin for 6- and 10Point Masks
For the 6- or 10- point mask case, all intermediate points are moved in
proportion to their distance from the mask center to the corresponding outermost vertices. This rule applies independently in the horizontal and, for the 10 point shape, in the vertical axes. This concept is
illustrated in Figure 38.
Figure 38. Tektronix Margin Rules for 6- and 10-Point Masks
7-10
8
Introduction 8- 2
Configuring the Simulator 8- 5
Controlling with User- written Applications
Using the Simulator
8- 7
Using the Simulator
Introduction
Introduction
When the N2100B software is installed in the DCA, a simulator is also
installed. The simulator reads previously captured data from a file on
the system and presents it to the control panel as if it were from an
instrument. The simulator is also a very powerful development tool as
it enables you to create your own code to control the instrument without having any hardware in place.
This chapter describes how to do the following tasks with the simulator:
• “To run the simulator” on page 8- 2.
• “Configuring the Simulator” on page 8- 5.
• “Controlling with User- written Applications” on page 8- 7.
To run the simulator
1 Start the simulator from the Windows Start menu by clicking Start > All
Programs > Agilent > N2100B > N2100B Simulator.
The simulator and the control application communicate using TCP/IP,
even when both run on the same computer. If you are running a firewall program, the simulator will probably be prevented from communicating with the control panel application. Either disable the firewall or
configure it to allow these programs to communicate as they wish.
8-2
Using the Simulator
To run the simulator
2 Click Load DCA File and select a data file.
You could also use a file of data written by the DCA Control Panel
application's Data/Save to File function. When no files of actual data
are available, you can use one of the sample data files in the DCA data
files directory, (Program Files/Agilent/N2100B/dca data files).
3 Start the DCA Control Panel application.
4 Click I/O Config and select the simulator option. The Port is 1500 by
default. It is possible to have up to four simulators running on ports
1500- 1503.
8-3
Using the Simulator
To run the simulator
5 Click Connect. The connected indicator on the simulator GUI will
activate, as shown below.
6 Click Start Acquisition.
The Activity Indicator flashes on the Simulator panel, as the control application reads data from the simulator and displays the data as an eye
pattern. Use the Module Config controls to change the simulator settings,
and use the various measurement and display controls to interact with
the simulator.
NOTE
The 2nd DCA File button allows you to load a second data file. When a second file is loaded
and the checkbox is checked, successive acquisitions alternate the data from the two files.
8-4
Using the Simulator
Configuring the Simulator
Configuring the Simulator
By clicking Settings on the Simulator, the following panel appears. This
panel allows you to view the settings that a user program has applied
to the simulated DCA, or verify consistency with the Module Configuration information on the DCA Control Panel.
Clicking Input Config on the panel opens the Module Input Configuration dialog box, allowing you to match the DCA being simulated. Setting parameters such as serial number, input configuration, signal
routing, and filter frequencies allows you to load filter and data files
specific to the given DCA.
8-5
Using the Simulator
Configuring the Simulator
NOTE
Measurement results from the simulator may not exactly match those from the DCA module. The
simulator is primarily intended as a qualitative simulation of the DCA.
8-6
Using the Simulator
Controlling with User-written Applications
Controlling with User-written Applications
Since the simulator presents the same TCP/IP (network) interface as
the DCA module, it is accessible by any application that can control
the DCA. To control the simulator through another application:
1 Start the simulator and load the desired simulator data file before the
application attempts to connect to it.
2 Within the application, select TCP/IP as the communication link type
and localhost as the module IP address.
The simulator must run on the same computer as the application that
attempts to connect to it.
Running More Than One To run more than one instance of the simulator simultaneously, use
Simulator
the TCP/IP Port selector on the simulator's Settings panel to select a
different port for each simulator instance. Connect each client application to the intended simulator by specifying that port when connecting. Four ports are currently available.
Simulator Data File
Format
The DCA simulator reads data files written by the control panel application's Data panel. To create suitable data files by another means, the
files must follow this format:
• Data are in fixed- format ASCII, with one data point (sample) per line.
• Columns are formatted as listed below.
• Intervening positions are filled with space characters.
0:
0.5000000
9.30
HLL
1:
0.0527776
9.56
HLL
2:
0.6055553
9.50
LLL
8-7
Using the Simulator
Controlling with User-written Applications
Table 4. Simulator Data File Format
Column
Field
1
The sequence number. The simulator does not use this information.
8
A colon that follows the sequence number. It must be present and properly positioned.
12
The sample X value, a floating-point number between 0 and 1.
22
The sample Y value, a floating-point number.
32
The transition type, one of the following strings.
LLL
Sustained low state following a sustained low state.
LLH
Low-to-high transition following a sustained low state.
LHL
High-to-low transition following a low-to-high transition.
LHH
Sustained high state following a low-to-high transition.
HLL
Sustained low state following a high-to-low transition.
HLH
Low-to-high transition following a high-to-low transition.
HHL
High-to-low transition following a sustained high state.
HHH
Sustained high state following a sustained high state.
8-8
9
Introduction 9- 2
DLL API Reference 9- 7
Module Information 9- 7
Module Connection 9- 11
Configuration 9- 12
Data Acquisition 9- 18
Measurement 9- 22
Calibration 9- 30
Status 9- 80
Logging 9- 31
Returned Status, Warning, and Error Values
Data Structures 9- 37
Enum Values 9- 43
Module Information 9- 46
Module Connection 9- 49
Mask Configuration 9- 51
DCA Sample Data Acquisition 9- 54
Measurement 9- 59
Pattern Sequence Acquisition 9- 70
Calibration 9- 75
Control Display 9- 76
DCA Configuration 9- 81
ActiveX API Properties 9- 81
DCA Configuration 9- 81
Pattern Sequence Acquisition 9- 84
Programming
9- 34
Programming
Introduction
Introduction
In addition to using the Control Panel, the DCA can be controlled
automatically through the following two methods:
• Application Programming Interface (API)
• ActiveX Control
The API for N2100B control is provided in a Windows based DLL and
an ActiveX control. The API allows the host application to create and
delete communication channels to instruments, configure the N2100B,
perform data acquisitions, measurements, and eye mask testing. Refer
to “Module Configuration Settings” on page 3- 4 and Chapter 4, “Measurements” for details of the configurable parameters.
9-2
Programming
Introduction
Figure 39. Host Library Overview
The N2100B Host Library consists of the following software components. A program that uses the DCA API calls must include the files
fgapidefs.h and fghostapi.h. These are found in the installed software,
in the include directory.
fghostapi.dll This file is the Windows based DLL that implements the Host Library
API.
fghostapi.lib This file contains all the DLL exported functions that are linked
against the Host Application.
fghostapi.h This header file contains the Host library function prototypes and definitions.
9-3
Programming
Introduction
fgtypes.h This header contains data type definitions common between the
embedded firmware and the host library.
fgstatus.h This header file contains all the status codes used by the N2100B.
fgapidefs.h This header file contains all the data structures that make up the API
between the Host Application, Host Library, and embedded firmware.
The interface between the OEM application and the N2100B is packet
based.
A folder of API examples is included with the downloaded software
and drivers.
Establishing
Communications
The host system communicates with the N2100B module over the PXI
Bus or through a TCP/IP interface. The host controls the N2100B module in the following ways:
• Sets operating modes and parameter values
• Initiates data acquisition operations
• Reads module status and measurement result values
Before any communications can take place, the host system must create a connection to the N2100B. Refer to “Module Connection” on page
9- 11 for more details. Connections must be closed at the end of the
session.
Active X
ActiveX allows developers to create code and applications from a multitude of languages. This also allows you to build a defined interface
code, making it accessible to other applications.
The DCA software interface is implemented as an ActiveX control
which is a software component with a standard interface that allows it
to be included (hosted) by a variety of programming and test environments such as LabView, LabWindows/CVI, Visual C++, Visual Basic, and
Visual C#. Its properties and methods make the DCA functionality
available to the controlling program, and its optional display presents
the same eye diagrams and measurement results as the GUI control
application.
Rules for using the DCA ActiveX control differ depending on the environment that hosts the control. Visual Basic provides direct access to
the control. LabWindows/CVI automatically generates a wrapper with
9-4
Programming
Introduction
callable functions. Visual C++ generates a wrapper class when the
ActiveX control is added to a project. Refer to the documentation for
your language/environment for the details of using ActiveX controls.
Use the following steps to add the DCA to a test application and
acquire data:
1 Add an instance of the activeX control to your project, using the
procedures appropriate to your programming environment.
9-5
Programming
Introduction
2 Call one of the module connection functions OpenTcp, OpenTcpByVal,
OpenPxi, or OpenPxiByVal to establish communications with the DCA
module.
3 Set the DisplayMode property to select whether the eye display window
is visible, and if so, what information it displays.
4 Call the various configuration functions to set module parameters such
as line rate, reference clock select, and number of sample data points.
5 Call one of the data acquisition functions AcquireDCA or
AcquireDCAWithMask to acquire data.
6 Call the various measurement functions to make measurements on the
acquired data and retrieve the results. Or, call one of the data
acquisition functions GetDCASampleData or GetDCASampleDataVariant
to retrieve the data sample values.
7 Update the module parameter settings and/or make additional
measurements as your application requires.
8 At the end, call the Close function to release the communication link to
the module.
9-6
Programming
DLL API Reference
DLL API Reference
A program that uses the DCA API calls must include the files
fgapidefs.h and fghostapi.h. You can find these files in the installed
software’s /include folder.
•
•
•
•
•
•
•
•
•
•
•
“Module Information” on page 9- 7.
“Module Connection” on page 9- 11.
“Configuration” on page 9- 12.
“Data Acquisition” on page 9- 18.
“Measurement” on page 9- 22.
“Calibration” on page 9- 30.
“Status” on page 9- 80.
“Logging” on page 9- 31.
“Returned Status, Warning, and Error Values” on page 9- 34.
“Data Structures” on page 9- 37.
“Enum Values” on page 9- 43.
Module Information
•
•
•
•
•
•
“FGGetBoardInfoAPI” on page 9-7.
“FGGetBoardNameAPI” on page 9-8.
“FGGetSerialNumberAPI” on page 9-8.
“FGGetBoardRevAPI” on page 9-9.
“FGGetInputConfigAPI” on page 9-9.
“FGGetFilterConfigAPI” on page 9-11.
FGGetBoardInfoAPI
Returns the module type string, serial number string, and revision
level.
Call FGGetBoardInfoAPI(FG_HANDLE hModule, char *pszSerialNumber, char
*pszBoardName, ULONG *pnBoardRev)
9-7
Programming
Module Information
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a character string buffer that receives the module serial
number.
• A pointer to a character string buffer that receives the module type
string.
• A pointer to an unsigned long integer that receives the module version
number.
The type string buffer must be at least FG_MAXBOARDNAME_SIZE bytes long and the serial
number buffer must be at least FG_MAXSERIALNUM_SIZE. These values are defined in
fgapidefs.h.
NOTE
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGGetBoardNameAPI
Returns the module type string.
Call FGGetBoardNameAPI(FG_HANDLE hModule, char *pszBoardName)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a character string buffer that receives the module type
string.
NOTE
This buffer must be at least FG_MAXBOARDNAME_SIZE bytes long, as defined in fgapidefs.h.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGGetSerialNumberAPI
Returns the module serial number string.
Call FGGetBoardSerialNumberAPI(FG_HANDLE hModule, char *pszSerialNumber)
9-8
Programming
Module Information
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a character string buffer that receives the module serial
number string.
This buffer must be at least FG_MAXSERIALNUM_SIZE bytes long, as defined in fgapidefs.h.
NOTE
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGGetBoardRevAPI
Returns the module revision level as an integer.
Call FGGetBoardRevAPI(FG_HANDLE hModule, unsigned long *pnBoardRev)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to an unsigned long variable that receives the returned value.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGGetInputConfigAPI
Returns an integer value that specifies the available inputs and the
signal routing configuration.
Call FGGetBoardInputConfigAPI(FG_HANDLE hModule, unsigned long *pnInputConfig)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to an unsigned long variable that receives the returned value.
Returns A status value of type FG_STATUS reflecting the result of the operation.
9-9
Programming
Module Information
Table 5. Returned Values
Available Input
Selections
Valuea
Module Input Configuration
1
5
Optical input to internal filter bank, electrical input direct
2
2
Optical input direct to high-speed path, electrical input direct
3
2
Optical input to high-speed path with filter, electrical input direct
4
5
Optical input to external high-performance filter bank, electrical input direct
5
1
Electrical only, direct
6
1
Electrical only, to high-speed path with filter
7
4
Electrical only, to internal filter bank
8
4
Electrical only, to external high-performance filter bank
9
4
Optical only, to internal filter bank
10
1
Optical only, to high-speed path
11
1
Optical only, to high-speed path with filter
12
4
Optical only, to external high-performance filter bank
13
3
Electrical with differential input
14
5
Optical input to external high-performance filter bank through high-speed path, electrical input
direct
15
4
Optical only, to external high-performance filter bank through high-speed path
16
5
N2100B with four optical filter selections, includes electrical input..
17
4
N2100B with three optical filter selections.
18
3
N2100B with two optical filter selections.
19
2
N2100B with a single optical filter selection.
20
2
N2100B with no filter assembly, electrical and optical input selections.
a. Values 1 through 15 apply to the N2100A. Values 16 through 20 apply the the N2100B.
9-10
Programming
Module Connection
FGGetFilterConfigAPI
Returns the frequencies of the installed input filters.
Call FGGetBoardFilterConfigAPI(FG_HANDLE hModule, FG_FILTERCONFIG
*pFilterConfig)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a structure of type FG_FILTERCONFIG that receives the
returned values.
Returns A status value of type FG_STATUS reflecting the result of the operation.
Module Connection
fg_connect
Sets up a connection to the module.
Call FG_HANDLE fg_connect(FG_COMM_PARAMS * pParams)
Parameter A pointer to a structure of type FG_COMM_PARAMS that specifies the
communications link (TCP/IP or PXI bus) to be used and the address
on that link of the target DCA module.
Returns A handle to the module connection. This handle must be passed as a
parameter to subsequent calls to the module API functions. If the
operation fails, the call returns NULL (0).
fg_disconnect
Closes a currently open module connection.
Call FG_STATUS fg_disconnect(FG_HANDLE hModule)
Parameters The handle to the connection to be closed, as returned by the call to
fg_connect.
9-11
Programming
Configuration
Returns A status value of type FG_STATUS reflecting the result of the operation.
Configuration
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
“FGCfgGetAPI” on page 9-12.
“FGCfgSetAllAPI” on page 9-12.
“FGCfgSetCustomIntfRateAPI” on page 9-13.
“FGCfgSetWaveLength” on page 9-13.
“FGCfgSetOptAttenRatioAPI” on page 9-13.
“FGCfgSetElectAttenRatioAPI” on page 9-14.
“FGCfgSetRefClkSltAPI” on page 9-15.
“FGCfgSetRefClkAPI” on page 9-15.
“FGCfgSetRecoveredClockRatioAPI” on page 9-16.
“FGCfgSetDCAAcqSizeAPI” on page 9-16.
“FGCfgSetInputSelectAPI” on page 9-16.
“FGCfgSetHWFilterAPI” on page 9-17.
“FGCfgSetNRZThresholdAPI” on page 9-17.
“FGCfgSetNRZWinSizeAPI” on page 9-17.
“FGCfgSetNRZOMAWinSizeAPI” on page 9-18.
FGCfgGetAPI
Reads the module’s current configuration parameter values.
Call FG_STATUS FGCfgGetAPI(FG_HANDLE hModule, FG_CONFIGPKT * pData)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a structure of type FG_CONFIGPKT that receives the configuration settings.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGCfgSetAllAPI
Sets all DCA configuration parameters.
Call FG_STATUS FGCfgGetAPI(FG_HANDLE hModule, FG_CONFIGPKT * pData)
9-12
Programming
Configuration
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a structure of type FG_CONFIGPKT that specifies the new
configuration settings.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGCfgSetCustomIntfRateAPI
Sets the line data rate.
Call FG_STATUS FGCfgSeCustomIntfRateAPI(FG_HANDLE hModule, double dRate)
Parameters
• The module connection handle returned by fg_connect.
• A value of type double that specifies the rate in MHz.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGCfgSetWaveLength
Selects the optical input wavelength.
Call FG_STATUS FGCfgSetWaveLength(FG_HANDLE hModule, FG_WAVELENGTH_TYPE
nWavelength)
Parameters
• The module connection handle returned by fg_connect.
• An enum value of type FG_WAVELENGTH_TYPE that specifies the
wavelength.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGCfgSetOptAttenRatioAPI
Sets the value of any external optical input attenuation.
Call FG_STATUS FGCfgSetOptAttenRatioAPI(FG_HANDLE hModule, float fAttenRatio)
9-13
Programming
Configuration
Informs the DCA of an external attenuator in the optical input path.
The DCA measurements use this value so the measured values reflect
the signal at the input of this attenuator.
Parameters
• The module connection handle returned by fg_connect.
• A value of type float that specifies the ratio by which the attenuator
reduces the optical power.
Attenuation specified in dB may be converted to power ratio according
to the following equation:
Ratio = 10dB/10
Power ratio may be converted to dB according to the following equation:
dB = 10 x log10(ratio)
For example, a 3 dB optical attenuator reduces the power by a factor
of 2:1; it would be entered as an attenuation ratio of 2.0. If no attenuator is used, this value should be set to 1.0.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGCfgSetElectAttenRatioAPI
Sets the value of any external electrical input attenuation.
Call FG_STATUS FGCfgSetElectAttenRatioAPI(FG_HANDLE hModule, float fAttenRatio)
Informs the DCA of an external attenuator in the electrical input path.
The DCA measurements use this value so the measured values reflect
the signal at the input of this attenuator.
Parameters
• The module connection handle returned by fg_connect.
• A value of type float that specifies the ratio by which the attenuator
reduces the input voltage.
For example, a 6 dB electrical attenuator reduces the voltage by a factor of 2:1; it would be entered as an attenuation ratio of 2.0.
9-14
Programming
Configuration
Attenuation specified in dB may be converted to voltage ratio according to the following equation:
Ratio = 10dB/20
Voltage ratio may be converted to dB according to the following equation:
dB = 20 x log10(ratio)
For example, a 6 dB optical attenuator reduces the power by a factor
of 2:1; it would be entered as an attenuation ratio of 2.0. If no attenuator is used, this value should be set to 1.0.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGCfgSetRefClkSltAPI
Specifies the timing source (reference clock) for the DCA’s data acquisition.
Call FG_STATUS FGCfgSetRefClkSltAPI(FG_HANDLE hModule, FG_REFCLKSLT
nRefSelect)
Parameters
• The module connection handle returned by fg_connect.
• An enum value of type FG_REFCLKSLT that specifies the timing source.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGCfgSetRefClkAPI
Specifies the clock frequency to be used when the reference clock
source selection is FG_REFCLK_EXTERNAL or
FG_REFCLK_EXTERNAL_AUTO.
Call FG_STATUS FGCfgSetRefClkAPI(FG_HANDLE hModule, double dFreq)
Parameters
• The module connection handle returned by fg_connect.
• A value of type double that specifes the frequency in MHz.
9-15
Programming
Configuration
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGCfgSetRecoveredClockRatioAPI
Specifies the ratio of the line rate to the recovered clock rate when
the reference clock source selection is
FG_REFCLK_RECOVERED_RATIO.
Call FG_STATUS FGCfgSetRecoveredClockRatioAPI(FG_HANDLE hModule, double dFreq)
Parameters
• The module connection handle returned by fg_connect.
• A value of type float that specifes the ratio.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGCfgSetDCAAcqSizeAPI
Sets the number of samples to be taken during data acquisition.
Call FG_STATUS FGCfgSetDCAAcqSizeAPI(FG_HANDLE hModule, unsigned long nSamples)
Parameters
• The module connection handle returned by fg_connect.
• A value of type unsigned long that specifies the number of samples.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGCfgSetInputSelectAPI
Selects the optical or electrical signal input. For the N2100A, the unselected input must have no signal applied.
Call FG_STATUS FGCfgSetInputSelectAPI(FG_HANDLE hModule, unsigned long
nInputSelect)
Parameters
• The module connection handle returned by fg_connect.
• A value of type unsigned long that specifies the input. Parameter values
9-16
Programming
Configuration
are as follows:
1: Electrical input
2: Optical input
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGCfgSetHWFilterAPI
Selects the hardware input filter.
Call FG_STATUS FGCfgSetHWFilterAPI(FG_HANDLE hModule, FG_FILTER_TYPE
nFilterType)
Parameters
• The module connection handle returned by fg_connect.
• An enum value of type FG_FILTER_TYPE that specifies the filter.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGCfgSetNRZThresholdAPI
Selects the low and high amplitude threshold values used for NRZ
measurements.
Call FG_STATUS FGCfgSetNRZThresholdAPI(FG_HANDLE hModule, FG_HILOW_VLTREF
nThresholdSel)
Parameters
• The module connection handle returned by fg_connect.
• An enum value of type FG_HILOW_VLTREF that specifies the thresholds.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGCfgSetNRZWinSizeAPI
Sets the start and end of the time window, in percent of 1 UI, over
which NRZ measurements are made.
9-17
Programming
Data Acquisition
Call FG_STATUS FGCfgSetNRZWinSizeAPI(FG_HANDLE hModule, FG_WINSIZE *
pWindow)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a structure of type FG_WINSIZE that specifies the window.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGCfgSetNRZOMAWinSizeAPI
Sets the start and end of the time window, in percent of 1 UI, over
which NRZ OMA measurements are made. Because the OMA window
spans the eye crossing, the start time of the window must be greater
than the end time.
Call FG_STATUS FGCfgSetNRZOMAWinSizeAPI(FG_HANDLE hModule, FG_WINSIZE *
pWindow)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a structure of type FG_WINSIZE that specifies the window.
Returns A status value of type FG_STATUS reflecting the result of the operation.
Data Acquisition
•
•
•
•
•
•
“FGDCAAcquireAPI” on page 9-18.
“FGDCAAcquireWithMaskAPI” on page 9-19.
“FGDCAGetNextAPI” on page 9-19.
“FGDCAAcquireToMemoryAPI” on page 9-20
“FGDCAAcquireFromMemoryAPI” on page 9-20
“FGDCAAcquireWithMaskFromMemoryAPI” on page 9-21
FGDCAAcquireAPI
Acquires DCA data using the current configuration settings, and
optionally returns a structure containing the first block of sample values.
9-18
Programming
Data Acquisition
Call FG_STATUS FGDCAAcquireAPI(FG_HANDLE hModule, FG_EYE_PLOTDATA *
PlotData)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a structure of type FG_EYE_PLOTDATA in which to return
the sample data. If the pointer is NULL, no sample data are returned.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGDCAAcquireWithMaskAPI
Acquires DCA data using the current configuration settings, and
optionally returns a structure containing the first block of sample values.
Call FG_STATUS FGDCAAcquireWithMaskAPI(FG_HANDLE hModule, FG_MASK_DATA
*pMask, UINT16 nMaskMarginType, UINT16 nMaskMargin, FG_EYE_PLOTDATA *
PlotData)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a structure of type FG_MASK_DATA that specifies the
mask.
• An integer value that specifies the mask margin convention:
0: Mask margins are compatible with Agilent 86100 mask margins.
1: Mask margins are compatible with Tektronix CSA8000 mask margins.
• An integer value that specifies the mask margin in percent.
• A pointer to a structure of type FG_EYE_PLOTDATA in which to return
the sample data. If the pointer is NULL, no sample data are returned.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGDCAGetNextAPI
Returns the next block of data from the most recent FGDCAAcquireAPI or FGDCAAcquireWithMaskAPI call.
9-19
Programming
Data Acquisition
Call FG_STATUS FGDCAGetNextAPI(FG_HANDLE hModule, FG_EYE_PLOTDATA *
PlotData)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a structure of type FG_EYE_PLOTDATA in which to return
the sample data.
Returns A status value of type FG_STATUS reflecting the result of the operation. If no more sample data are available, the return value is
FG_NOMORE.
FGDCAAcquireToMemoryAPI
Acquire DCA sample data and save the samples in a buffer for later
processing.
Call FG_STATUS FGDCAAcquireToMemoryAPI(FG_HANDLE hModule, int nMemory)
Function Acquire DCA data using the current configuration settings, and save
the data in a memory buffer for later processing.
Parameters Retrieve DCA data previously saved in a memory buffer, and optionally
return a structure containing the first block of sample values.
• The module connection handle returned by fg_connect.
• An integer between 1 and 4 that specifies the memory buffer from which
to retrieve the data.
• A pointer to a structure of type FG_EYE_PLOTDATA in which to return
the sample data. If the pointer is NULL, no sample data are returned.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGDCAAcquireFromMemoryAPI
Retrieve sample data from a memory buffer and return the first data
block.
Call FG_STATUS FGDCAAcquireFromMemoryAPI(FG_HANDLE hModule, int nMemory,
FG_EYE_PLOTDATA * PlotData)
Function
9-20
Programming
Data Acquisition
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a structure of type FG_EYE_PLOTDATA in which to return
the sample data.
Returns A status value of type FG_STATUS reflecting the result of the operation. If no more sample data are available, the return value is
FG_NOMORE.
tion.
FGDCAAcquireWithMaskFromMemoryAPI
Retrieve sample data from a memory buffer, perform a mask test, and
return the first data block.
Call FG_STATUS
FGDCAAcquireWithMaskFromMemoryAPI(FG_HANDLE hModule, FG_MASK_DATA
*pMask, UINT16 nMaskMarginType, UINT16 nMaskMargin, int nMemory,
FG_EYE_PLOTDATA * PlotData)
Function Retrieve DCA data previously saved in a memory buffer, check the
samples against a specified mask, and optionally return a structure
containing the first block of sample values.
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a structure of type FG_MASK_DATA that specifies the
mask.
• An integer value that specifies the mask margin convention:
0: Mask margins are compatible with Agilent 86100 mask margins
1: Mask margins are compatible with Tektronix CSA8000 mask margins
• An integer value that specifies the mask margin in percent.
• An integer between 1 and 4 that specifies the memory buffer from which
to retrieve the data.
• A pointer to a structure of type FG_EYE_PLOTDATA in which to return
the sample data. If the pointer is NULL, no sample data are returned.
Returns A status value of type FG_STATUS reflecting the result of the operation..
9-21
Programming
Measurement
Measurement
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
“FGDCAAvgPowerAPI” on page 9-22.
“FGDCAOscilloscopeAPI” on page 9-23.
“FGDCAMaskTestAPI” on page 9-23.
“FGDCAMaskTestAPI” on page 9-23.
“FGNRZExtinctionRatioAPI” on page 9-24.
“FGNRZEyeHeightAPI” on page 9-24.
“FGNRZEyeAmplitudeAPI” on page 9-24.
“FGNRZEyeWidthAPI” on page 9-25.
“FGNRZOMAAPI” on page 9-25.
“FGNRZLowAPI” on page 9-25.
“FGNRZHighAPI” on page 9-26.
“FGNRZSNRAPI” on page 9-26.
“FGNRZBitRateAPI” on page 9-26.
“FGNRZRiseTimeAPI” on page 9-27.
“FGNRZFallTimeAPI” on page 9-27.
“FGNRZCrossingPercentAPI” on page 9-28.
“FGNRZDutyCyDistortionAPI” on page 9-28.
“FGNRZJitterAPI” on page 9-28.
“FGNRZRiseJitterAPI” on page 9-29.
“FGNRZFallJitterAPI” on page 9-29.
“FGNRZDeterministicJitterAPI” on page 9-29.
FGDCAAvgPowerAPI
Returns the average optical power at the optical input.
Call FG_STATUS FGDCAAvgPowerAPI(FG_HANDLE hModule, FG_RSP_DCA_AVGPWR *
pData)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a structure of type FG_RSP_DCA_AVGPWR to receive the
results.
Returns A status value of type FG_STATUS reflecting the result of the operation.
9-22
Programming
Measurement
FGDCAOscilloscopeAPI
Measures the basic oscilloscope measurement values and statistics on
the acquired data.
Call FG_STATUS FGDCAOscilloscopeAPI(FG_HANDLE hModule,
FG_RSP_DCA_OSCILLOSCOPE * pData)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a structure of type FG_RSP_DCA_OSCILLOSCOPE to receive the results.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGDCAMaskTestAPI
Applies the specified mask and margin to the acquired samples and
returns the number of sample points that fail.
Call FG_STATUS FGDCAMaskTestAPI(FG_HANDLE hModule, FG_MASK_DATA * pMask,
UINT16 nMaskMarginType, UINT16 nMaskMargin, FG_RSP_DCA_MASK * pNumFailed)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a structure of type FG_MASK_DATA that specifies the
mask.
• An integer value that specifies the mask margin convention:
0: Mask margins are compatible with Agilent 86100 mask margins.
1: Mask margins are compatible with Tektronix CSA8000 mask margins.
• An integer value that specifies the mask margin in percent.
• A pointer to a structure of type FG_RSP_DCA_MASKTEST that receives
the number of failing points.
Returns A status value of type FG_STATUS reflecting the result of the operation.
9-23
Programming
Measurement
FGNRZExtinctionRatioAPI
Measures extinction ratio on the acquired data.
Call FG_STATUS FGNRZExtinctionRatioAPI(FG_HANDLE hModule,
FG_RSP_DCA_EXTINCTION * pData)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a structure of type FG_RSP_DCA_EXTINCTION to receive
the results.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGNRZEyeHeightAPI
Measures eye height on the acquired data.
Call FG_STATUS FGNRZEyeHeightAPI(FG_HANDLE hModule,
FG_RSP_DCA_EYEHEIGHT * pData)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a structure of type FG_RSP_DCA_EYEHEIGHT to receive
the results.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGNRZEyeAmplitudeAPI
Measures eye amplitude on the acquired data.
Call FG_STATUS FGNRZEyeAmplitudeAPI(FG_HANDLE hModule,
FG_RSP_DCA_EYEAMPLITUDE * pData)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a structure of type FG_RSP_DCA_EYEAMPLITUDE to receive the results.
9-24
Programming
Measurement
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGNRZEyeWidthAPI
Measures eye width on the acquired data.
Call FG_STATUS FGNRZEyeWidthAPI(FG_HANDLE hModule, FG_RSP_DCA_EYEWIDTH
* pData)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a structure of type FG_RSP_DCA_EYEWIDTH to receive the
results.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGNRZOMAAPI
Measures optical modulation amplitude on the acquired data.
Call FG_STATUS FGNRZOMAAPI (FG_HANDLE handle, FG_RSP_DCA_OMA * pData)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a structure of type FG_RSP_DCA_OMA to receive the results.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGNRZLowAPI
Measures the low level in the NRZ window on the acquired data.
Call FG_STATUS FGNRZLowAPI(FG_HANDLE hModule, FLOAT * pData)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a variable of type float to receive the results.
9-25
Programming
Measurement
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGNRZHighAPI
Measures the average high level within the NRZ window on the
acquired data.
Call FG_STATUS FGNRZHighAPI(FG_HANDLE hModule, FLOAT * pData)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a variable of type float to receive the results.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGNRZSNRAPI
Measures signal- to- noise ratio on the acquired data.
Call FG_STATUS FGNRZSNRAPI(FG_HANDLE hModule, FLOAT * pData)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a variable of type float to receive the results.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGNRZBitRateAPI
Returns the signal bit rate. or the reference clock frequency. The value
returned depends on the reference clock selection. For internal clock,
this call returns the line rate set by the user. For external clock, this
call returns the frequency at the reference clock input. For recovered
clock, this call returns the measured line rate. See “FGCfgSetRefClkSltAPI” on page 15.
Call FG_STATUS FGNRZBitRateAPI(FG_HANDLE hModule, FLOAT * pData)
9-26
Programming
Measurement
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a variable of type float to receive the results.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGNRZRiseTimeAPI
Measures NRZ signal rise time between the programmed NRZ thresholds on the acquired data.
Call FG_STATUS FGNRZRiseTimeAPI(FG_HANDLE hModule, FG_RSP_DCA_RISEFALL *
pData)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a variable of type FG_RSP_DCA_RISEFALL to receive the
results.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGNRZFallTimeAPI
Measures NRZ signal fall time between the programmed NRZ thresholds on the acquired data.
Call FG_STATUS FGNRZFallTimeAPI(FG_HANDLE hModule, FG_RSP_DCA_RISEFALL *
pData)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a variable of type FG_RSP_DCA_RISEFALL to receive the
results.
Returns A status value of type FG_STATUS reflecting the result of the operation.
9-27
Programming
Measurement
FGNRZCrossingPercentAPI
Measures the signal level at the eye crossing on the acquired data,
expressed as a percentage of signal amplitude.
Call FG_STATUS FGNRZCrossingPercentAPI(FG_HANDLE hModule, FLOAT * pData)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a variable of type float to receive the results.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGNRZDutyCyDistortionAPI
Measures the duty cycle distortion in percent on the acquired data,
that is, the asymmetry between the time spent in the low state and the
time spent in the high state.
Call FG_STATUS FGNRZDutyCyDistortionAPI(FG_HANDLE hModule, FG_RSP_DCA_DCD
* pData)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a variable of type FG_RSP_DCA_DCD to receive the results.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGNRZJitterAPI
Measures the signal edge jitter at the eye crossing level on the
acquired data.
Call FG_STATUS FGNRZJitterAPI(FG_HANDLE hModule, FG_RSP_DCA_JITTER * pData)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a variable of type FG_RSP_DCA_JITTER to receive the results.
9-28
Programming
Measurement
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGNRZRiseJitterAPI
Measures the rising edge jitter at the eye crossing level on the acquired
data.
Call FG_STATUS FGNRZRiseJitterAPI(FG_HANDLE hModule, FG_RSP_DCA_JITTER *
pData)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a variable of type FG_RSP_DCA_JITTER to receive the results.
Returns
A status value of type FG_STATUS reflecting the result of the operation.
FGNRZFallJitterAPI
Measures the falling edge jitter at the eye crossing level on the
acquired data.
Call FG_STATUS FGNRZRiseJitterAPI(FG_HANDLE hModule, FG_RSP_DCA_JITTER *
pData)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a variable of type FG_RSP_DCA_JITTER to receive the results.
Returns
A status value of type FG_STATUS reflecting the result of the operation.
FGNRZDeterministicJitterAPI
Measures deterministic (data- dependent) edge jitter by measuring the
average time position difference between LLH and HLH transitions,
and between HHL and LHL transitions, at the eye crossing level.
Call FG_STATUS FGNRZDeterministicJitterAPI(FG_HANDLE hModule,
FG_RSP_DCA_DETJITTER *pJitter);
9-29
Programming
Calibration
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a variable of type FG_RSP_DCA_DETJITTER to receive the
results.
Returns A status value of type FG_STATUS reflecting the result of the operation.
Calibration
•
•
•
•
“FGCalDarkCalAPI” on page 9-30.
“FGCalibrateDCAPI” on page 9-30.
“FGCalCancelAPI” on page 9-30.
“FGCalGetCalStatus” on page 9-31.
FGCalDarkCalAPI
Measures and saves the optical dark current level.
Call FG_STATUS FGCalDarkCalAPI(FG_HANDLE hModule)
Parameter The module connection handle returned by fg_connect.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGCalibrateDCAPI
Starts delay calibration for the currently selected input and filter.
Call FG_STATUS FGCalibrateDCAPI(FG_HANDLE hModule)
Parameter The module connection handle returned by fg_connect.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGCalCancelAPI
Stops any delay calibration operation in progress.
Call FG_STATUS FGCalCancelAPI(FG_HANDLE hModule)
9-30
Programming
Logging
Parameter The module connection handle returned by fg_connect.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGCalGetCalStatus
Returns the state of the current delay calibration operation.
Call FG_STATUS FGCalGetCalStatus(FG_HANDLE hModule, unsigned long *pProgressValue)
Parameter
• The module connection handle returned by fg_connect.
• A pointer to an unsigned long integer that receives a value indicating
the progress of the delay calibration operation.
This value starts at 0 and increments as calibration progresses; the
exact final value depends on the delay calibration values found.
Returns A status value of type FG_STATUS reflecting the result of the operation. In particular, the function returns FG_BUSY if calibration is running, or FG_SUCCESS if it has finished.
Logging
• “FGSyslogSyncTimestampAPI” on page 9-31.
• “FGSyslogSetLevelAPI” on page 9-32.
• “FGSyslogGetLevelAPI” on page 9-32.
• “FGSyslogReadAPI” on page 9-32.
• “FGSyslogClearAPI” on page 9-33.
The following logging functions are not available through the ActiveX
control.
FGSyslogSyncTimestampAPI
Sets the module timestamp equal to the host computer’s current time.
Call FGSyslogSyncTimestampAPI(FG_HANDLE hModule)
Parameter The module connection handle returned by fg_connect.
9-31
Programming
Logging
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGSyslogSetLevelAPI
Sets the log level that specifies the types of information to be logged.
Call FGSyslogSetLevelAPI(FG_HANDLE hModule, FG_SYSLOG_LEVEL level)
Parameters
• The module connection handle returned by fg_connect.
• An enum value of type FG_SYSLOG_LEVEL that specifies the log level.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGSyslogGetLevelAPI
Reads the log level. Set the log level that specifies the types of information to be logged.
Call FGSyslogGetLevelAPI(FG_HANDLE hModule, FG_SYSLOG_LEVEL * pLevel)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to an enum value of type FG_SYSLOG_LEVEL that receives
the current log level.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGSyslogReadAPI
Creates a file and copies the specified logged data to that file.
Call FGSyslogReadAPI(FG_HANDLE hModule, char *filename, FG_SYSLOG_TYPE
LogSelect)
Parameters
• The module connection handle returned by fg_connect.
• A pointer to a buffer that receives the system- generated file name.
• An enum value of type FG_SYSLOG_TYPE that specifies whether the
9-32
Programming
Logging
operation is to read the application log or the system (kernel) log.
Returns A status value of type FG_STATUS reflecting the result of the operation.
FGSyslogClearAPI
Clears the specified log.
Call FGSyslogClearAPI(FG_HANDLE hModule, FG_SYSLOG_TYPE LogSelect)
Parameters
• The module connection handle returned by fg_connect.
• An enum value of type FG_SYSLOG_TYPE that specifies whether the
operation is to read the application log or the system (kernel) log.
Returns A status value of type FG_STATUS reflecting the result of the operation.
9-33
Programming
Returned Status, Warning, and Error Values
Returned Status, Warning, and Error Values
This table lists some of the more common status, warning, and error
values (FG_STATUS) returned by module functions. A full list of status
values appears in the API header file fgapidefs.h.
Table 6. Status, Warning, and Error Values (1 of 3)
FG_STATUS Values
Description
FG_SUCCESS
The operation was successful.
FG_BUSY
The calibration operation has not yet finished.
FG_WARN_GENERIC
Generic warning state.
FG_NOT_SUPPORTED
The specified operation is not supported by the module.
FG_CALCANCELLED
The calibration operation was cancelled by the user.
FG_NOMORE
There are no more data sample values available.
FG_DELAYCAL_REQD
Calibration of the delay lines is recommended.
FG_ERR_GENERIC
Generic error message.
FG_INVALID_COMMINTF
The specified module communication interface is not valid.
FG_MSG_SEND
An error occurred sending packet data to the module.
FG_MSG_RCV
An error occurred receiving the packet response from the module.
FG_INVALID_PKTTYPE
The packet type sent to the module was not valid.
FG_PKTID_UNEXPECTED
The response received from the module had a packet ID different from that of the command
packet that prompted the response.
FG_BADCMD
The command sent to the module was invalid.
FG_INVALID_LOG_CMD
The status/error logging command sent to the module was invalid.
FG_NOTSUPPORTED
The specified operation is not supported by the module.
FG_TIMEOUT
A timeout occurred while performing the specified operation.
FG_INVALIDCONFIG
The specified module configuration is not valid.
9-34
Programming
Returned Status, Warning, and Error Values
Table 6. Status, Warning, and Error Values (2 of 3)
FG_STATUS Values
Description
FG_INVALID_CFG_ITEM
The configuration operation specifies an invalid parameter.
FG_INVALID_WAVELENGTH
The optical wavelength specified is invalid.
FG_INVALID_REFCLKSLT
The reference clock source specified is invalid.
FG_INVALID_REFCLK
The external reference clock frequency specified is invalid.
FG_INVALID_FILTER
The filter selection specified is invalid.
FG_INVALID_INTF
The interface type and line rate specified is invalid.
FG_INVALID_HILOREF
The voltage threshold selection specified is invalid.
FG_INVALID_WINSIZE
The window size specified is invalid.
FG_INVALID_TIAWINDOW
The TIA window specified is invalid.
FG_INVALID_CLOCKEDGE
The trigger edge specified is invalid.
FG_INVALID_MASKMARGIN
The specified mask margin is out of range.
FG_INVALID_CUSTOMRATE
The specified line rate is invalid.
FG_INVALID_INPUTSELECT
The specified input selection is invalid.
FG_ACQUIRE_FAILED
The module was unable to acquire sample data.
FG_NOSAMPLES
There is no acquired data on which to perform the measurement.
FG_NO_BUFFERS
The module was unable to allocate a data buffer.
FG_CONFIG_FAILED
An operation to update the module configuration failed.
FG_NVRREAD_FAILED
A failure occurred reading from the non-volatile memory (NVR).
FG_NVRWRITE_FAILED
A failure occurred writing to the NVR.
FG_CALFAILED
The calibration operation failed.
FG_BOARDINFO_INIT_FAILED
A failure occurred reading the module type and revision data.
FG_SETLOGLVLFAILED
The operation to set the status/error logging level failed.
FG_INTERNAL_ERROR
An unspecified internal error occurred.
FG_TIMESETFAILED
An attempt to set the module’s time stamp failed.
9-35
Programming
Returned Status, Warning, and Error Values
Table 6. Status, Warning, and Error Values (3 of 3)
FG_STATUS Values
Description
FG_FILEOPEN
A file open operation failed or no file is currently open.
FG_NOSUCHFILE
The specified file was not found.
FG_NOSUCHDIR
The specified directory was not found.
9-36
Programming
Data Structures
Data Structures
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
“FG_COMM_PARAMS structure” on page 9-37.
“FG_TCPIP_PARAMS structure” on page 9-37.
“FG_BUS_PARAMS structure” on page 9-38.
“FG_FILTERCONFIG structure” on page 9-38.
“FG_CONFIGPKT structure” on page 9-38.
“FG_WINSIZE structure” on page 9-38.
“FG_XY_DATA structure” on page 9-39.
“FG_EYE_PLOTDATA structure” on page 9-39.
“FG_MASK_DATA structure” on page 9-40.
“FG_RSP_DCA_AVGPWR structure” on page 9-40.
“FG_RSP_DCA_OSCILLOSCOPE structure” on page 9-40.
“FG_RSP_DCA_MASKTEST structure” on page 9-41.
“FG_RSP_DCA_VAMP structure” on page 9-41.
“FG_STATISTICS structure” on page 9-41.
“FG_RSP_DCA_EXTINCTION structure” on page 9-41.
“FG_RSP_DCA_EYEHEIGHT structure” on page 9-41.
“FG_RSP_DCA_EYEAMPLITUDE structure” on page 9-41.
“FG_RSP_DCA_EYEWIDTH structure” on page 9-42.
“FG_RSP_DCA_OMA structure” on page 9-42.
“FG_RSP_DCA_RISEFALL structure” on page 9-42.
“FG_RSP_DCA_DCD structure” on page 9-42.
“FG_RSP_DCA_JITTER structure” on page 9-42.
“FG_RSP_DCA_DETJITTER structure” on page 9-42.
FG_COMM_PARAMS structure
typedef struct _FG_COMM_PARAMS {
LONG pktid;
FG_COMM_TYPES type;
unsigned long rcvtimeout;
void *pReserved;
union {
FG_TCPIP_PARAMS tcpip;
FG_BUS_PARAMS bus;
} comm_type;
} FG_COMM_PARAMS;
An integer to identify the packet; sequential calls
typically use sequential numbers.
Specifies a PXI bus connection or a TCP/IP connection.
Timeout value (TCP/IP only).
For system use.
Parameters specific to a TCP/IP connection.
Parameters specific to a PXI bus connection.
FG_TCPIP_PARAMS structure
typedef struct _FG_TCPIP_PARAMS {
int fd;
char ip_address[FG_IPMAXSIZE];
File descriptor, for system use.
The IP address, a string of the form “192.100.40.1”
The constant FG_IPMAXSIZE is defined in the file
fghostapi.h.
9-37
Programming
Data Structures
struct sockaddr_in saddr;
USHORT port;
} FG_TCPIP_PARAMS;
Socket address, for system use.
The port number for the connection.
FG_BUS_PARAMS structure
typedef struct _FG_BUS_PARAMS {
int fd;
char szInstrumentString[256];
UINT32 local_dma_mem_addr;
UINT32 local_dma_fifo_addr[8];
} FG_BUS_PARAMS;
File descriptor, for system use.
The PXI target module address string, eg.
“PXI2::17::INSTR”.
PXI bus address values, for system use.
FG_FILTERCONFIG structure
typedef struct _FG_FILTERCONFIG {
float Filter[6];
} FG_FILTERCONFIG;
Entries 0 through 4 reflect the frequencies of up to
five installed filters. The number of input options
and filters available for use depends on the module’s
input configuration. Entry 5 is unused.
FG_CONFIGPKT structure
typedef struct _FG_CONFIGPKT {
FG_INTF_TYPE
intftype;
FG_REFCLKSLT
refclkslt;
ULONG PatternAcqEnable;
double refclk;
double CustomIntfLineRate;
ULONG PatternAcqType;
ULONG PatternAcqLength;
FG_HILOW_VLTREF hi_low_volt_ref;
FG_TIA_WINDOW TIAWindow;
ULONG DCASampleSize;
FG_WINSIZE NRZWinWidth;
FG_WINSIZE NRZOMAWinWidth;
FG_HILOW_VLTREF NRZThreshold;
FG_WINSIZE RZWinWidth;
FG_HILOW_VLTREF RZThreshold;
FLOATRecoveredClockRatio;
FLOAT ERCF;
FG_HILOW_VLTREF OscThreshold;
FG_FILTER_TYPE HWFilterType;
ULONG ERCFActive;
ULONG WanderCompEnable;
FLOAT OptAttenRatio;
FLOAT ElectAttenRatio;
FG_WAVELENGTH_TYPE WaveLength;
FG_INPUTSELECT InputSelect;
} FG_CONFIGPKT;
(not used)
Reference clock source selection.
Enable pattern sequence acquisition (must be set to 0
when controlling DCA through API).
External reference clock frequency.
Line rate.
Control parameter for pattern sequence acqusition (not
used when controlling DCA through API).
Control parameter for pattern sequence acqusition (not
used when controlling DCA through API).
(Not used).
(Not used).
Acquisition sample count.
NRZ time window boundaries.
OMA time window boundaries.
NRZ amplitude threshold selection.
(Not used)
(Not used)
RecoveredClockRatio
ER Correction factor
(Not used)
DCA input filter selection.
ER Correction factor is activated
Time base wander compensation enable.
Ratio by which optical measurements are multiplied to
compensate for an external optical attenuator.
Ratio by which electrical measurements are multiplied
to compensate for an external electrical attenuator.
Optical wavelength selection.
Optical or electrical input selection.
FG_WINSIZE structure
typedef struct _FG_WINSIZE {
9-38
Programming
Data Structures
ULONG
start;
The position, in percent (0.01 UI), of the start of
the window.
The position, in percent, of the end of the window.
ULONG end;
} FG_WINSIZE;
FG_XY_DATA structure
typedef struct _FG_XY_DATA {
Used as an element of FG_EYE_PLOTDATA, below
FLOAT XCoordinate;
The X coordinate of the sample.
FLOAT YCoordinate;
The Y coordinate of the sample.
} FG_XY_DATA;
FG_EYE_PLOTDATA structure
typedef struct _FG_EYE_PLOTDATA {
USHORT nSamples;
The number of samples contained in this block.
FG_XY_DATA Sample[FG_EYE_MAXSEG_SIZE];Array of sample X, Y coordinate values.
unsigned char transition[FG_EYE_MAXSEG_SIZE];Refer to Table 7 for values.
} FG_EYE_PLOTDATA;
Table 7. Transition Field Values for FG_EYE_PLOTDATA
Bits
Description
Value
Transition (bits 2..0)
LLL
Steady-state low following a steady-state low.
0
LLH
Low-to-high transition following a steady-state low.
1
HHL
High-to-low transition following a steady-state high.
2
HHH
Steady-state high following a steady-state high.
3
HLL
Steady-state low following a high-to-low transition.
4
HLH
Low-to-high transition following a high-to-low transition.
5
LHL
High-to-low transition following a low-to-high transition.
6
LHH
Steady-state high following a low-to-high transition.
7
Invalid (bit 3)
Sample is not valid (see MeasPatternAcqDetJitterFall and MeasPatternAcqDetJitterRise).
8
Fail bits (bits 7..6)
Fail bits are applicable only for the AcquireWithMask acquisition functions.
Fail
A ‘1’ in this bit indicates a point that failed the mask test (as modified by the mask margin,
if a margin is used).
128
(0x80)
Fail, no margin
A ‘1’ in this bit indicates a point that would fail the mask without a margin. This bit is not
used when the margin is 0.
64 (0x40)
9-39
Programming
Data Structures
Table 7. Transition Field Values for FG_EYE_PLOTDATA
Bits
Description
Value
Fail-region bits (bits 5..4)
Fail center
This point failed the mask’s center region.
16 (0x10)
Fail upper
This point failed the mask’s upper (overshoot) region.
32 (0x20)
Fail lower
This point failed the mask’s lower (undershoot) region.
48 (0x30)
FG_MASK_DATA structure
X values have the range 0 to 1 UI; Y values are expressed as a fraction
of the signal amplitude, where 0.0 is the base level and 1.0 is the top
level.
typedef struct _FG_MASK_DATA {
FLOAT X1;
FLOAT X2;
FLOAT X3;
FLOAT X4;
FLOAT X5;
FLOAT X6;
FLOAT Y0;
FLOAT Y1;
FLOAT Y2;
FLOAT Y3;
FLOAT Y4;
FLOAT Y5;
FLOAT Y6;
UINT16 nMaskShape;
UINT16 bHasy3y4;
UINT16 nWindow;
} FG_MASK_DATA;
Left extent of mask.
Left edge of rectangular region.
Right edge of rectangular region.
Right extent of mask.
Left intermediate X value.
Right intermediate X value.
Center mask Y center line.
Center mask lower limit.
Center mask upper limit.
Lower overshoot mask limit.
Upper overshoot mask limit.
Lower intermediate Y value.
Upper intermediate Y value.
0: Rectangular mask. 1: 6-point mask with triangular
end regions. 2: 10-point mask with end regions and
intermediate points.
0: Overshoot masks are not present. 1: Overshoot masks
are present.
0: Reference levels are based on full interval. > 0:
Reference levels are based on specified percentage of
unit interval, centered in eye.
FG_RSP_DCA_AVGPWR structure
typedef struct _FG_RSP_DCA_AVGPWR {
FLOAT dB;
FLOAT watts;
FLOAT ratio;
}
FG_RSP_DCA_AVGPWR;
Power in dBm.
Power in watts.
The ratio of the average power to 0 dBm, equivalent to
the average power in mW.
FG_RSP_DCA_OSCILLOSCOPE structure
This structure is a wrapper for a FG_RSP_DCA_VAMP structure.
typedef struct _FG_RSP_DCA_OSCILLOSCOPE {
FG_RSP_DCA_VAMP amplitude;
9-40
Programming
Data Structures
}
FG_RSP_DCA_OSCILLOSCOPE;
FG_RSP_DCA_MASKTEST structure
typedef struct _FG_RSP_DCA_MASKTEST {
UINT32 nFailTotal;
Total count of failing points.
UINT32 nFailCenter;
Count of points that fail the center region.
UINT32 nFailUpper;
Count of points that fail the upper region.
UINT32 nFailLower;
Count of points that fail the lower region.
} FG_RSP_DCA_MASKTEST;
FG_RSP_DCA_VAMP structure
This structure contains a FG_STATISTICS structure and a number of
individual values.
typedef struct _FG_RSP_DCA_VAMP {
FG_STATISTICS stats;
FLOAT posovershoot;
FLOAT negovershoot;
FLOAT vbase;
FLOAT vtop;
FLOAT vamplitude;
} FG_RSP_DCA_VAMP;
Positive overshoot, in percent of amplitude
Negative overshoot, in percent of amplitude
Signal base level, in uW (optical) or mV (electrical)
Signal top level
Signal amplitude
FG_STATISTICS structure
typedef struct FG_STATISTICS {
float mean;
float stddev;
float maximum;
float minimum;
float pkpk;
float max_pos_deviation;
float max_neg_deviation;
} FG_STATISTICS;
Average value
Standard deviation or RMS value
Maximum value
Minimum value
Peak-to-peak amplitude
Maximum positive deviation
Maximum negative deviation
FG_RSP_DCA_EXTINCTION structure
typedef struct _FG_RSP_DCA_EXTINCTION
float dB;
float percent;
float ratio;
} FG_RSP_DCA_EXTINCTION;
{
Extinction ratio in dB
Extinction ratio in percent
Extinction ratio
FG_RSP_DCA_EYEHEIGHT structure
typedef struct _FG_RSP_DCA_EYEHEIGHT
float dB;
float amplitude;
float min;
float max;
} FG_RSP_DCA_EYEHEIGHT;
{
(Not used)
Eye height
Lower extent of eye height range
Upper extent of eye height range
FG_RSP_DCA_EYEAMPLITUDE structure
typedef struct _FG_RSP_DCA_EYEAMPLITUDE {
float amplitude;
Eye amplitude
float min;
Lower extent of eye amplitude range
float max;
Upper extent of eye amplitude range
9-41
Programming
Data Structures
}
FG_RSP_DCA_EYEAMPLITUDE;
FG_RSP_DCA_EYEWIDTH structure
typedef struct _FG_RSP_DCA_EYEWIDTH {
float ratio;
Eye width in unit intervals (UI)
float time;
Eye width in picoseconds
float start;
Left extent of eye, in unit intervals (UI)
float end;
Right extent of eye
float ymid;
Signal amplitude at eye crossing
} FG_RSP_DCA_EYEWIDTH;
FG_RSP_DCA_OMA structure
typedef struct _FG_RSP_DCA_OMA {
float time;
float start;
float end;
} FG_RSP_DCA_OMA;
Transition time in picoseconds
Start position of transition, in unit intervals (UI)
End position of transition
FG_RSP_DCA_RISEFALL structure
typedef struct _FG_RSP_DCA_RISEFALL {
float time;
Transition time in picoseconds
float start;
Start position of transition, in unit intervals (UI)
float end;
End position of transition
float thresholdL;
Low threshold used for measurement
float thresholdH;
High threshold used for measurement
} FG_RSP_DCA_RISEFALL;
FG_RSP_DCA_DCD structure
typedef struct _FG_RSP_DCA_DCD {
float time;
float percent;
float risemid;
float fallmid;
} FG_RSP_DCA_DCD;
Duty cycle distortion in picoseconds
Duty cycle distortion, in percent
Rising edge midpoint, in unit intervals (UI)
Falling edge midpoint, in UI
FG_RSP_DCA_JITTER structure
typedef struct _FG_RSP_DCA_JITTER {
float pkpk;
float rms;
} FG_RSP_DCA_JITTER;
Peak-to-peak jitter amplitude in unit intervals (UI)
RMS jitter amplitude in UI
FG_RSP_DCA_DETJITTER structure
typedef struct _FG_RSP_DCA_DETJITTER {
float fRiseJitter;
Time difference, in picoseconds, between the average
time position of a LLH transition and the average time
position of a HLH transition.
float fFallJitter;
Time difference, in picoseconds, between the average
time position of a HHL transition and the average time
position of a LHL transition.
float fRiseLLH;
Average time position, in unit intervals (UI), of a
LLH transition.
float fRiseHLH;
Average time position, in UI, of a HLH transition.
float fFallHHL;
Average time position, in UI, of a HHL transition.
float fFallLHL;
Average time position, in UI, of a LHL transition.
} FG_RSP_DCA_DETJITTER;
9-42
Programming
Enum Values
Enum Values
Table 8. Enum Values (1 of 3)
Integer
Value
Enum Value
Description
Module communications link FG_COMM_TYPES enum values
0
FG_COMM_TCPIP
TCP/IP
1
FG_COMM_BUS
PXI bus
Input FG_INPUTSELECT enum values
1
FG_INPUTSELECT_ELECTRICAL
Electrical input
2
FG_INPUTSELECT_OPTICAL
Optical input
Line rate FG_WAVELENGTH_TYPE enum values
1
FG_WAVELENGTH_850
850 nm
2
FG_WAVELENGTH_1310
1310 nm
3
FG_WAVELENGTH_1550
1550 nm
Reference clock source FG_REFCLK_SLT enum values
1
FG_REFCLK_STRATUM3
The internal reference clock.
2
FG_REFCLK_EXTERNAL
A user-provided clock signal connected to the External Clock input. The
frequency of this clock must be between 10 MHz and 8.5 GHz and must
match the value of the External Reference Clock Frequency parameter.
3
FG_REFCLK_RECOVERED
The clock derived from the received optical or electrical data.
4
FG_REFCLK_STRATUM3_AUTO
The internal reference clock. After acquisition, the data are analyzed and
corrected, if possible, for small deviations from the stated line rate.
5
FG_REFCLK_EXTERNAL_AUTO
A user-provided clock signal, as in FG_REFCLK_EXTERNAL. After
acquisition, the data are analyzed and corrected, if possible, for small
deviations from the stated line rate or reference frequency.
9-43
Programming
Enum Values
Table 8. Enum Values (2 of 3)
Integer
Value
Enum Value
Description
6
FG_REFCLK_RECOVERED_RATIO
The clock derived from the received optical or electrical signal. This option
allows specifying the ratio of the line rate to the recovered clock signal, for
example, when the N2100B’s wideband clock recovery detects a K28.7
pattern at one-fifth the actual rate. See the Note on page 9-45.
Hardware filter (API) and input path/filter (ActiveX) selection values FG_FILTER_TYPE enum values
1
FG_HWFILTER_1
First selection, as returned by the FGGetInputConfigAPI and
FGGetFilterConfigAPI API calls and the GetModuleInformation ActiveX
call.
2
FG_HWFILTER_2
Second selection, if present.
3
FG_HWFILTER_3
Third selection, if present.
4
FG_HWFILTER_4
Fourth selection, if present.
5
FG_HWFILTER_HS
Fifth selection, if present.
Thresholds FG_HILOW_VLTREF enum values
1
FG_VLTREF_9010
10% to 90% of amplitude over full UI period.
2
FG_VLTREF_8020
20% to 80% of amplitude over full UI period.
3
FG_VLTREF_7030
30% to 70% of amplitude over full UI period.
4
FG_VLTREF_9010_WIN
10% to 90% of amplitude within NRZ window.
5
FG_VLTREF_8020_WIN
20% to 80% of amplitude within NRZ window.
6
FG_VLTREF_7030_WIN
30% to 70% of amplitude within NRZ window.
Information logged FG_SYSLOG_LEVEL enum values
5
FG_SYSLOG_NONE
Logging is disabled.
4
FG_SYSLOG_FATAL
Fatal errors only.
3
FG_SYSLOG_ERROR
All errors.
2
FG_SYSLOG_WARNING
All errors and warnings.
1
FG_SYSLOG_INFO
All errors and warning, as well as other significant information.
9-44
Programming
Enum Values
Table 8. Enum Values (3 of 3)
Integer
Value
Enum Value
Description
0
FG_SYSLOG_DEBUG
All of the above, plus internal information chosen for debugging purposes.
Log selected FG_SYSLOG_TYPE enum values
1
FG_SYSLOG_APP
Application (user-level) log.
2
FG_SYSLOG_SYSTEM
System (kernel) log.
NOTE
The N2100B clock recovery does not function well with signals that can be interpreted as lower
than the actual bit rate (for example, the K28.7 pattern is a repeating pattern of five ones followed
by five zeroes; this looks the same as a pattern of alternating ones and zeroes at one-fifth of the
actual bit rate). The N2100A does not have this restriction. To deal with this restriction, you must
select enum 6 or FG_REFCLK_RECOVERED_RATIO as the reference clock source (RefClockSel).
This sets the clock recovery mode to expect a signal that will be detected as a lower rate than it
should. You then enter a multiplier using RecoveredClockRatio; the detected clock frequency is
multiplied by this number to obtain the actual line rate. For example, for a K28.7 signal, the
multiplier is 5.
9-45
Programming
ActiveX API Methods
ActiveX API Methods
•
•
•
•
•
•
•
•
“Module Information” on page 9- 46.
“Module Connection” on page 9- 49.
“Mask Configuration” on page 9- 51.
“DCA Sample Data Acquisition” on page 9- 54.
“Measurement” on page 9- 59.
“Pattern Sequence Acquisition” on page 9- 70.
“Calibration” on page 9- 75.
“Control Display” on page 9- 76.
Module Information
GetModuleInformation
Call BSTR GetModuleInformation(long nIndex)
Returns the specified module information (module type, serial number...) as a string.
Parameters An integer value that specifies the information to be returned. Refer to
the table for the list of valid values and the information returned.
Returns A string containing the specified module information. Note that all values, including numeric values, are returned as character strings. Conversion to numeric values or other types is the responsibility of the
user.
Table 9. GetModuleInformation Index Values And Information Returned
nIndex
value
0
Information returned
Description
Module type string
Alphanumeric string, for example, N2100B.
9-46
Programming
Module Information
Table 9. GetModuleInformation Index Values And Information Returned
nIndex
value
Information returned
Description
1
Module serial number
String of digits, for example, 04101234.
2
Module revision number
String of one or more digits.
3
FPGA revision number
String of the form a.bb or aa.bb, where aa and bb are the major and minor FPGA
revision numbers, respectively.
4
API revision number
String of digits, for example, 20.
5
Calibration API revision
String of digits, for example, 20.
6
Maintenance API revision
String of digits, for example, 20.
7
Driver version
String of the form a.bbb or aa.bbb, where aa and bbb are the major and minor
driver revision numbers, respectively.
8
Firmware version
String of the form a.bbb or aa.bbb, where aa and bbb are the major and minor
firmware revision numbers, respectively.
9
Firmware build date
String of the form Nov 18 2004.
10
ActiveX control build date
String of the form Nov 18 2004.
11
Input configuration and filter
specifiers
See Input configuration string, below.
12
Input / filter #1 descriptor
See Input / Filter descriptor, below.
through
through
16
Input / Filter #5 descriptor
17
Pattern acquisition filter files
resident on the module
A comma-separated list of file names for the pattern acquisition filter files that
reside in the DCA’s flash memory.
18
All pattern acquisition filter files
available for use
A comma-separated list of file names for the pattern acquisition filter files in the
local DCA Data Files directory that match the DCA’s serial number.
19
ActiveX control version
Alphanumeric string, for example, 2.4.1.0.
• Input configuration string in the following form:
1,5,O:4250.0000,O:2488.3200,O:2125.0000,O:1062.5000,E:0.0000
9-47
Programming
Module Information
Where each of the fields is separated by a comma:
• A number that identifies the module’s input configuration, as returned by FGGetInputConfigAPI.
• The number of input selections available. Refer to Table 10.
• One or more Input / Filter descriptors (see below), matching the
number of available input selections. The values in Table 10 reflect
the module’s signal routing configuration.
• Input / Filter Descriptor string in the following form:
O:4250.0000
Where the content is:
The letter O or E, specifies Optical or Electrical input, respectively, or
the strings ED, E+, or E- , specify the differential, positive, or negative
inputs of the electrical differential DCA.
A colon.
A string of digits representing a floating- point number that specifies
the frequency of the filter for that input selection. A numeric value of
0 indicates no filter.
Table 10. Returned Values
Available Input
Selections
Valuea
Module Input Configuration
1
5
Optical input to internal filter bank, electrical input direct.
2
2
Optical input direct to high-speed path, electrical input direct.
3
2
Optical input to high-speed path with filter, electrical input direct.
4
5
Optical input to external high-performance filter bank, electrical input direct.
5
1
Electrical only, direct.
6
1
Electrical only, to high-speed path with filter.
7
4
Electrical only, to internal filter bank.
8
4
Electrical only, to external high-performance filter bank.
9
4
Optical only, to internal filter bank.
9-48
Programming
Module Connection
Table 10. Returned Values
Available Input
Selections
Valuea
Module Input Configuration
10
1
Optical only, to high-speed path.
11
1
Optical only, to high-speed path with filter.
12
4
Optical only, to external high-performance filter bank.
13
3
Electrical with differential input.
14
5
Optical input to external high-performance filter bank through high-speed path, electrical input
direct.
15
4
Optical only, to external high-performance filter bank through high-speed path.
16
5
N2100B with four optical filter selections, includes electrical input..
17
4
N2100B with three optical filter selections.
18
3
N2100B with two optical filter selections.
19
2
N2100B with a single optical filter selection.
20
2
N2100B with no filter assembly, electrical and optical input selections.
a. Values 1 through 15 apply to the N2100A. Values 16 through 20 apply to the N2100B.
Module Connection
•
•
•
•
•
“OpenPxi” on page 9-49.
“OpenPxiByVal” on page 9-50.
“OpenTcp” on page 9-50.
“OpenTcpByVal” on page 9-50.
“Close” on page 9-51.
OpenPxi
Establishes a connection between the DCA ActiveX control and the
module via the PXI, specified by the parameter.
Call FG_STATUS OpenPxi(BSTR * pInstrumentString)
9-49
Programming
Module Connection
Parameter An appropriately formatted character string, of the form
PXI2::17::INSTR, that specifies the target DCA module.
Returns A status value of type FG_STATUS reflecting the result of the operation.
OpenPxiByVal
Sets up a connection to a DCA module through the PXI bus, specified
by the parameter. This is an alternative form of OpenPxi, for applications that have difficulty passing the required BSTR pointer to OpenPxi.
Call FG_STATUS OpenPxiByVal(_TCHAR * pszInstrString)
Parameter An appropriately formatted character string, of the form
PXI2::17::INSTR, that specifies the target DCA module.
Returns A status value of type FG_STATUS reflecting the result of the operation.
OpenTcp
Establishes a connection between the DCA ActiveX control and the
module via TCP/IP (network), specified by the parameter.
Call FG_STATUS OpenTcp(BSTR * pIpAddr)
Parameter An appropriately formatted character string, of the form 192.168.22.40,
that specifies the network IP address of the target DCA module.
The connection port may be specified explicitly by including it in the
address string, separated from the IP address by a colon, for example,
192.168.22.40:1500.
Returns A status value of type FG_STATUS reflecting the result of the operation.
OpenTcpByVal
Set up a connection to a DCA module through TCP/IP (network). This
is an alternative form of OpenTcp, for applications that have difficulty
passing the required BSTR pointer to OpenTcp.
Call FG_STATUS OpenTcpByVal(_TCHAR * pszInstrString)
9-50
Programming
Mask Configuration
Parameter An appropriately formatted character string, of the form 192.168.22.40,
that specifies the network IP address of the target DCA module.
The connection port may be specified explicitly by including it in the
address string, separated from the IP address by a colon, e.g.,
192.168.22.40:1500.
Returns A status value of type FG_STATUS reflecting the result of the operation.
Close
Disconnects the connection between the ActiveX control and the DCA
ActiveX control and the DCA module.
Call FG_STATUS Close( )
Parameters None
Returns A status value of type FG_STATUS reflecting the result of the operation.
Mask Configuration
•
•
•
•
“SetDCAMask” on page 9-51.
“SetDCAMaskVariant” on page 9-53.
“SetDCAMaskWindowed” on page 9-53.
“SetDCAMaskWindowedVariant” on page 9-53.
SetDCAMask
Call void SetDCAMask(FLOAT * fMaskArray)
Sets the mask definition to be used for mask tests. (See also “SetDCAMaskVariant” on page 9- 53.)
Parameters An array of floating- point values for the mask definition.
9-51
Programming
Mask Configuration
Parameter values The mask definition is an array of 13 (for SetDCAMask) or 14 (for SetDCAMaskWindowed) values of type FLOAT.
Mask shape
Rectangular: Set X1 and X4 both 0
6- point: Set X5 and X6 both 0
10- point: Use all X, Y values for center mask
Overshoot masks
Not present: Set Y3, Y4 both 0
Present: Set Y3, Y4 to desired limit values
Table 11. Mask Definition (1 of 2)
Offset
9-52
Function
0
X1: Left extent of mask
1
X2: Left edge of rectangular region
2
X3: Right edge of rectangular region
3
X4: Right extent of mask
4
X5: Left intermediate X value
5
X6: Right intermediate X value
6
Y0: Center mask Y center line
7
Y1: Center mask lower limit
8
Y2: Center mask upper limit
9
Y3: Lower overshoot mask limit
10
Y4: Upper overshoot mask limit
11
Y5: Lower intermediate Y value
12
Y6: Upper intermediate Y value
Programming
Mask Configuration
Table 11. Mask Definition (2 of 2)
Offset
13
Function
For SetDCAMaskWindowed and SetDCAMaskWindowedVariant, this parameter
specifies the window width in percent over which the low and high Y reference levels are
measured. For SetDCAMask and SetDCAMaskVariant, this parameter is not used and
need not be present.
SetDCAMaskVariant
Sets the mask parameters used for mask tests. This is an alternative
form of SetDCAMask, for applications that have difficulty directly passing array parameters.
Call void SetDCAMaskVariant(VARIANT * vMaskArray)
Sets the mask definition to be used for mask tests.
Parameters A VARIANT containing an array of floating- point values for the mask
definition.
Parameter values Refer to “Mask Definition” on page 9- 52.
SetDCAMaskWindowed
Sets the mask parameters used for mask tests (Y levels based on windowed low and high measurements).
Call void SetDCAMask(FLOAT * fMaskArray)
Sets the mask definition to be used for mask tests.
(See also “SetDCAMaskVariant” on page 9- 53.)
Parameters An array of floating- point values for the mask definition.
Parameter values Refer to “Mask Definition” on page 9- 52.
SetDCAMaskWindowedVariant
Sets the mask parameters used for mask tests (Y levels based on windowed low and high measurements).
This is an alternative form of SetDCAMaskWindowed, for applications
that have difficulty directly passing array parameters.
9-53
Programming
DCA Sample Data Acquisition
Call void SetDCAMask(VARIANT * vMaskArray)
Sets the mask definition to be used for mask tests.
Parameters A VARIANT containing an array of floating- point values for the mask
definition.
Parameter values Refer to “Mask Definition” on page 9- 52.
DCA Sample Data Acquisition
•
•
•
•
•
•
•
•
“AcquireDCA” on page 9-54.
“AcquireDCAToMemory” on page 9-54
“AcquireDCAFromMemory” on page 9-55
“AcquireDCAWithMaskFromMemory” on page 9-55
“AcquireDCAWithMask” on page 9-56
“GetDCASampleCount” on page 9-56.
“GetDCASampleData” on page 9-56.
“GetDCASampleDataVariant” on page 9-58.
AcquireDCA
Acquires DCA sample data.
Call long Acquire( )
Acquires DCA data using the current configuration settings.
Parameters None
Returns A status value of type FG_STATUS reflecting the result of the operation.
AcquireDCAToMemory
Acquire DCA sample data to a buffer in DCA memory.
Call long AcquireDCAToMemory(int nMemory)
Function Acquire DCA data using the current configuration settings and save
them in a memory buffer for later processing.
9-54
Programming
DCA Sample Data Acquisition
Parameters An integer between 1 and 4 that specifies the memory buffer in which
to save the data.
Returns A status value of type FG_STATUS reflecting the result of the operation.
AcquireDCAFromMemory
Retrieve DCA sample data from a memory buffer.
Call long AcquireDCAFromMemory(int nMemory)
Function Retrieve DCA sample data acquired with AcquireDCAToMemory
Parameters An integer between 1 and 4 that specifies the memory buffer from
which to retrieve the data.
Returns A status value of type FG_STATUS reflecting the result of the operation.
AcquireDCAWithMaskFromMemory
Retrieve DCA sample data from a memory buffer and perform a mask
test.
Call ong AcquireDCAWithMaskFromMemory(int nMemory, int nMaskMarginType, int
nMaskMargin)
Function Retrieve DCA sample data acquired with AcquireDCAToMemory and
perform a mask test on it using the current mask.
Parameters
• An integer between 1 and 4 that specifies the memory buffer from which
to retrieve the data.
• An integer value that specifies the mask margin convention:
0: Mask margins are compatible with Agilent 86100 mask margins
1: Mask margins are compatible with Tektronix CSA8000 mask margins
• An integer value that specifies the mask margin in percent.
Returns A status value of type FG_STATUS reflecting the result of the operation.
9-55
Programming
DCA Sample Data Acquisition
AcquireDCAWithMask
Acquires DCA sample data and performs a mask test.
Call long AcquireWithMask(int nMaskMarginType, int nMaskMargin)
Acquires DCA data using the current configuration settings and performs a mask test on it using the current mask.
Parameters
• An integer value that specifies the mask margin convention:
0: Mask margins are compatible with Agilent 86100 mask margins.
1: Mask margins are compatible with Tektronix CSA8000 mask margins.
• An integer value that specifies the mask margin in percent.
Returns A status value of type FG_STATUS reflecting the result of the operation.
GetDCASampleCount
Returns the number of samples acquired. This call is applicable only
when the ActiveX control window is displayed. Otherwise, the sample
data are never transferred to the ActiveX control and is therefore not
available.
Call long GetDCASampleCount( )
Returns the number of valid samples acquired.
Parameters None
GetDCASampleData
Return the sample values acquired. This call is not applicable when
the DisplayMode property is 0; in that case the sample data are never
transferred to the ActiveX control and is therefore not available.
Call long GetDCASampleData(long nCount, float * fXArray, float * fYArray, short *
nTransArray)
Returns the X / Y values and transition types of the acquired samples.
See also “GetDCASampleDataVariant” on page 9- 58.
9-56
Programming
DCA Sample Data Acquisition
Parameters
• An integer value that specifies the number of sample points to return.
The number of points returned will be this number or the total number
acquired, whichever is less.
• A pointer to an array of type float that receives the sample X coordinates.
• A pointer to an array of type float that receives the sample Y coordinates.
• A pointer to an array of type unsigned short that receives the sample
transition- type codes. See Table 12 for transition type values for these
codes.
Returns A status value of type FG_STATUS reflecting the result of the operation.
Table 12. Transition Field Values (1 of 2)
Bits
Description
Value
Transition (bits 2..0)
LLL
Steady-state low following a steady-state low.
0
LLH
Low-to-high transition following a steady-state low.
1
HHL
High-to-low transition following a steady-state high.
2
HHH
Steady-state high following a steady-state high.
3
HLL
Steady-state low following a high-to-low transition.
4
HLH
Low-to-high transition following a high-to-low transition.
5
LHL
High-to-low transition following a low-to-high transition.
6
LHH
Steady-state high following a low-to-high transition.
7
Invalid (bit 3)
Sample is not valid (see MeasPatternAcqDetJitterFall and MeasPatternAcqDetJitterRise).
8
Fail bits (bits 7..6)
Fail bits are applicable only for the AcquireWithMask acquisition functions.
Fail
A ‘1’ in this bit indicates a point that failed the mask test (as modified by the mask margin,
if a margin is used).
128
(0x80)
9-57
Programming
DCA Sample Data Acquisition
Table 12. Transition Field Values (2 of 2)
Bits
Description
Value
Fail, no margin
A ‘1’ in this bit indicates a point that would fail the mask without a margin. This bit is not
used when the margin is 0.
64 (0x40)
Fail center
This point failed the mask’s center region.
16 (0x10)
Fail upper
This point failed the mask’s upper (overshoot) region.
32 (0x20)
Fail lower
This point failed the mask’s lower (undershoot) region.
48 (0x30)
Fail-region bits (bits 5..4)
GetDCASampleDataVariant
Returns the sample values acquired. This call is not applicable when
the DisplayMode property is 0; in that case the sample data are never
transferred to the ActiveX control and is thus not available.
This is an alternative form of GetDCASampleData, for applications that
have difficulty directly passing array parameters.
Call long GetDCASampleData(long nCount, VARIANT * vXArray, VARIANT * vYArray,
VARIANT * vTransArray)
Returns the X / Y values and transition types of the acquired samples.
Parameters
• An integer value that specifies the number of sample points to return.
The number of points returned will be this number or the total number
acquired, whichever is less.
• A pointer to a variant containing an array of type float that receives
the sample X coordinates.
• A pointer to a variant containing an array of type float that receives
the sample Y coordinates.
• A pointer to a variant containing an array of type unsigned short that
receives the sample transition- type codes. See Table 12 for transition
type values for these codes.
Returns A status value of type FG_STATUS reflecting the result of the operation.
9-58
Programming
Measurement
Measurement
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
“MeasAvgPowerUW” on page 9-60.
“MeasAvgPowerDBM” on page 9-60.
“GetMaskTestFailPoints” on page 9-60.
“GetMaskTestFailPointsAll” on page 9-60.
“GetMaskTestFailPointsAllVariant” on page 9-61.
“FindThresholdMaskMargin” on page 9-61.
“ExpandDCAMaskMargin” on page 9-62.
“ExpandDCAMaskMarginVariant” on page 9-62.
“GetBitRate” on page 9-63.
“MeasOscAmplitude” on page 9-63.
“MeasOscRMS” on page 9-63.
“MeasOscPeakToPeak” on page 9-64.
“MeasOscMinLevel” on page 9-64.
“MeasOscMaxLevel” on page 9-64.
“MeasOscMaxLevel” on page 9-64.
“MeasOscBaseLevel” on page 9-64.
“MeasOscTopLevel” on page 9-64.
“MeasOscPosOvershoot” on page 9-65.
“MeasOscNegOvershoot” on page 9-65.
“MeasNRZExtinctionRatio” on page 9-65.
“MeasNRZExtinctionRatioDB” on page 9-65.
“MeasNRZExtinctionRatioHist” on page 9-66.
“MeasNRZExtinctionRatioDBHist” on page 9-66.
“MeasNRZEyeHeight” on page 9-66.
“MeasNRZEyeAmplitude” on page 9-66.
“MeasNRZEyeWidth” on page 9-67.
“MeasNRZOMA” on page 9-67.
“MeasNRZLowLevel” on page 9-67.
“MeasNRZHighLevel” on page 9-67.
“MeasNRZSNR” on page 9-67.
“MeasNRZCrossingPercent” on page 9-68.
“MeasNRZRiseTime” on page 9-68.
“MeasNRZFallTime” on page 9-68.
“MeasNRZDutyCycleDistortion” on page 9-68.
“MeasNRZJitterPeakToPeak” on page 9-69.
“MeasNRZJitterRMS” on page 9-69.
“MeasNRZSingleEdgeJitterPP” on page 9-69.
“MeasNRZSingleEdgeJitterRMS” on page 9-69.
“MeasNRZDeterministicJitterRise” on page 9-70.
9-59
Programming
Measurement
• “MeasNRZDeterministicJitterFall” on page 9-70.
MeasAvgPowerUW
Measures average optical power in uW.
Call float MeasAvgPowerUW( )
Measures the average optical power.
Parameters NA
Returns A float value representing the optical power in microwatts.
MeasAvgPowerDBM
Measures average optical power in dBm.
Call float MeasAvgPowerDBM( )
Measures the average optical power.
Parameters NA
Returns A float value representing the optical power in dBm.
GetMaskTestFailPoints
Gets the count of failing points from a mask test.
Call long GetMaskTestFailPoints( )
Gets the results of the most recent AcquireDCAWithMask operation.
Parameters NA
Returns A long value representing the count of samples that failed the mask
test.
GetMaskTestFailPointsAll
Gets the count of failing points and separate counts for each mask
region.
Call long GetMaskTestFailPointsAll(long *pResult )
Gets the results of the most recent AcquireDCAWithMask operation.
9-60
Programming
Measurement
Parameters A pointer to an array of four long values that receive the mask test
results, as follows:
[0] The total count of failing sample points.
[1] The count of points that fail the center mask region.
[2] The count of points that fail the upper mask region.
[3] The count of points that fail the lower mask region.
Returns A status value of type FG_STATUS reflecting the result of the operation.
GetMaskTestFailPointsAllVariant
Gets the count of failing points and separate counts for each mask
region. This is an alternative form of GetMaskTestFailPointsAll, for use
with applications that have difficulty passing an array as a parameter.
Call VARIANT GetMaskTestFailPointsAllVariant( )
Gets the results of the most recent AcquireDCAWithMask operation.
Parameters NA
Returns A VARIANT, containing an array of four long values that receive the
mask test results as follows:
[0] The total count of failing sample points.
[1] The count of points that fail the center mask region.
[2] The count of points that fail the upper mask region.
[3] The count of points that fail the lower mask region.
FindThresholdMaskMargin
Determines the mask margin that yields a specified number of failing
points.
Call long FindThresholdMaskMargin(long nFailingPoints)
Determines the maximum mask margin value that results in no more
than the specified number of points that fail the mask test.
Parameters A long value that specifies the maximum number of failing points
allowed.
9-61
Programming
Measurement
Returns A long value representing the mask margin.
ExpandDCAMaskMargin
Returns the coordinates of the mask vertices with a mask margin
applied.
Call long ExpandDCAMaskMargin(float *fArray, long nMarginType, long nMargin)
Applies the specified margin to the current mask and returns the vertices of the mask that results. Use SetDCAMask to specify the current
mask.
Parameters
• An array of floating- point values, in the format used by SetDCAMask,
that receives the result. Refer to “Mask Definition” on page 9- 52 for the
format of this array.
• An integer value that specifies the mask margin convention:
0: Mask margins are compatible with Agilent 86100 mask margins.
1: Mask margins are compatible with Tektronix CSA8000 mask margins.
• An integer value that specifies the mask margin in percent.
Returns A status value of type FG_STATUS reflecting the result of the operation.
ExpandDCAMaskMarginVariant
Returns the coordinates of the mask vertices with a mask margin
applied.
This is an alternative form of ExpandDCAMaskMargin, for applications
that have difficulty directly passing array parameters.
Call long ExpandDCAMaskMargin(VARIANT * pvArray, long nMarginType, long nMargin)
Applies the specified margin to the current mask and returns the vertices of the mask that results. Use SetDCAMask to specify the current
mask.
Parameters
• A pointer to a VARIANT. On return from this function, the VARIANT
9-62
Programming
Measurement
contains an array of floating- point values, in the format used by SetDCAMask, that receives the result. Refer to “Mask Definition” on page
9- 52 for the format of this array.
• An integer value that specifies the mask margin convention:
0: Mask margins are compatible with Agilent 86100 mask margins.
1: Mask margins are compatible with Tektronix CSA8000 mask margins.
• An integer value that specifies the mask margin in percent.
Returns A status value of type FG_STATUS reflecting the result of the operation.
GetBitRate
Returns the current bit rate or equivalent bit rate, depending on the
reference clock selected:
• Internal clock returns the line rate set by the user.
• External clock returns the frequency at the reference clock input.
• Recovered clock returns the measured line rate.
Call float GetBitRate( )
Parameters NA
Returns A float value representing the line rate in Mb/s.
MeasOscAmplitude
Measures base- to- top amplitude.
Call float MeasOscAmplitude( )
Parameters NA
Returns A float value representing the base- to- top amplitude.
MeasOscRMS
Measures AC RMS amplitude.
Call float MeasOscRMS( )
Parameters NA
9-63
Programming
Measurement
Returns A float value representing the RMS amplitude.
MeasOscPeakToPeak
Measures peak- to- peak amplitude.
Call float MeasOscPeakToPeak( )
Parameters NA
Returns A float value representing the peak- to- peak amplitude.
MeasOscMinLevel
Measures minimum level.
Call float MeasOscMinLevel( )
Parameters NA
Returns A float value representing the minimum level.
MeasOscMaxLevel
Measures maximum level.
Call float MeasOscMaxLevel( )
Parameters NA
Returns A float value representing the maximum level.
MeasOscBaseLevel
Measures base level.
Call float MeasOscBaseLevel( )
Parameters NA
Returns A float value representing the base level.
MeasOscTopLevel
Measures top level.
9-64
Programming
Measurement
Call float MeasOscTopLevel( )
Parameters NA
Returns A float value representing the top level.
MeasOscPosOvershoot
Measures the positive overshoot, in percent, of the acquired signal.
Positive overshoot is defined as 100 * (max – top) / amplitude.
Call float MeasOscPosOvershoot
Parameters NA
Returns A float value representing the positive overshoot.
MeasOscNegOvershoot
Measures the negative overshoot, in percent, of the acquired signal.
Negative overshoot is defined as 100 * (base – min) / amplitude.
Call float MeasOscNegOvershoot
Parameters NA
Returns A float value representing the negative overshoot.
MeasNRZExtinctionRatio
Measures the signal’s extinction ratio within the NRZ window.
Call float MeasNRZExtinctionRatio
Parameters NA
Returns A float value representing the extinction ratio, expressed as a ratio.
MeasNRZExtinctionRatioDB
Measures the signal’s extinction ratio within the NRZ window in dB.
Call float MeasNRZExtinctionRatioDB
Parameters NA
9-65
Programming
Measurement
Returns A float value representing the extinction ratio, expressed in dB.
MeasNRZExtinctionRatioHist
Measures the signal’s extinction ratio within the NRZ window and display a histogram of signal amplitude within the window.
Call float MeasNRZExtinctionRatioHist
Parameters NA
Returns A float value representing the extinction ratio, expressed as a ratio.
MeasNRZExtinctionRatioDBHist
Measures the signal’s extinction ratio within the NRZ window and display a histogram of signal amplitude within the window.
Call long Acquire( )
Parameters NA
Returns A float value representing the extinction ratio, expressed in dB.
MeasNRZEyeHeight
Measures eye height.
Call float MeasNRZEyeHeight
Parameters NA
Returns A float value representing the eye height.
MeasNRZEyeAmplitude
Measures eye amplitude.
Call float MeasNRZEyeAmplitude
Parameters NA
Returns A float value representing the eye amplitude.
9-66
Programming
Measurement
MeasNRZEyeWidth
Measures eye width.
Call float MeasNRZEyeWidth
Parameters NA
Returns A float value representing the eye width.
MeasNRZOMA
Measures the OMA, which is the signal amplitude at the crossing, measured within the NRZ OMA window.
Call float MeasNRZOMA
Parameters NA
Returns A float value representing the optical modulation amplitude.
MeasNRZLowLevel
Measures the average low level within the NRZ window.
Call float MeasNRZLowLevel
Parameters NA
Returns A float value representing the low level.
MeasNRZHighLevel
Measures the average high level within the NRZ window.
Call float MeasNRZHighLevel
Parameters NA
Returns A float value representing the high level.
MeasNRZSNR
Measures the signal- to- noise ratio (SNR) within the NRZ window.
Call float MeasNRZSNR
9-67
Programming
Measurement
Parameters NA
Returns A float value representing the signal- to- noise ratio.
MeasNRZCrossingPercent
Measures the level at the eye crossing.
Call float MeasNRZCrossingPercent( )
Measures the signal level at the eye crossing, expressed as a percentage of the signal amplitude.
Parameters NA
Returns A float value representing the eye crossing level in percent.
MeasNRZRiseTime
Measures the signal rise time between the selected NRZ thresholds.
Call float MeasNRZRiseTime
Parameters NA
Returns A float value representing the rise time in picoseconds.
MeasNRZFallTime
Measures the signal fall time between the selected NRZ thresholds.
Call float MeasNRZFallTime
Parameters NA
Returns A float value representing the fall time in picoseconds.
MeasNRZDutyCycleDistortion
Measures the duty cycle distortion, that is, the asymmetry between the
time the signal is in the high state and the time the signal is in the
low state.
Call float MeasNRZDutyCycleDistortion( )
Parameters NA
9-68
Programming
Measurement
Returns A float value representing the duty cycle distortion in percent.
MeasNRZJitterPeakToPeak
Measures the peak- to- peak edge jitter at the eye crossing.
Call float MeasNRZJitterPeakToPeak( )
Parameters NA
Returns A float value representing the peak- to- peak jitter in picoseconds.
MeasNRZJitterRMS
Measures the RMS edge jitter at the eye crossing.
Call float MeasNRZJitterRMS( )
Parameters NA
Returns A float value representing the RMS jitter in picoseconds.
MeasNRZSingleEdgeJitterPP
Measures the peak- to- peak edge jitter for the selected edge at the eye
crossing.
Call float MeasNRZJitterPeakToPeak(int nEdgeSelect )
Parameters An integer value that selects the edge to be measured:
0:
falling edge
1:
rising edge
Returns A float value representing the peak- to- peak jitter in picoseconds.
MeasNRZSingleEdgeJitterRMS
Measures the RMS edge jitter for the selected edge at the eye crossing.
Call float MeasNRZJitterRMS(int nEdgeSelect )
Parameters An integer value that selects the edge to be measured:
0:
falling edge
1:
rising edge
9-69
Programming
Pattern Sequence Acquisition
Returns A float value representing the RMS jitter in picoseconds.
MeasNRZDeterministicJitterRise
Measures the time difference between the average time position of a
LLH transition and the average time position of a HLH transition.
Call float MeasNRZDeterministicJitterRise( )
Parameters NA
Returns A float value representing the time difference in picoseconds.
MeasNRZDeterministicJitterFall
Measures the time difference between the average time position of a
HHL transition and the average time position of a LHL transition.
Call float MeasNRZDeterministicJitterFall( )
Parameters NA
Returns A float value representing the time difference in picoseconds.
Pattern Sequence Acquisition
•
•
•
•
•
•
•
“LoadPatternAcqFilterFile” on page 9-70.
“LoadPatternAcqFilterFileByValue” on page 9-71.
“MeasPatternAcqDetJitterRise” on page 9-71.
“MeasPatternAcqDetJitterFall” on page 9-72.
“MeasPatternAcqStdDetJitterRise” on page 9-73.
“MeasPatternAcqStdDetJitterFall” on page 9-73.
“MeasPatternSpecificRiseTime” on page 9-73.
LoadPatternAcqFilterFile
Loads a filter specification file for use with pattern sequence acquisition.
Call FG_STATUS LoadPatternAcqFilterFile(BSTR * FileString)
Parameter A character string that is part of the file name and which specifies the
particular file to load.
9-70
Programming
Pattern Sequence Acquisition
Filter file names have the format DCAnnnnnnBTsssss.txt, where
nnnnnn is the module’s serial number and sssss is a user- specified
string that gives the file a unique name. The string parameter specifies
this second field; the remainder of the file name is filled in automatically.
Returns A status value of type FG_STATUS reflecting the result of the operation.
LoadPatternAcqFilterFileByValue
Loads the specified filter file for use with pattern sequence acquisition.
This is an alternative form of LoadPatternAcqFilterFile, for applications that have difficulty passing the required BSTR pointer to LoadPatternAcqFilterFile.
Call FG_STATUS LoadPatternAcqFilterFileByValue(char * FileString)
Parameter A character string that is part of the file name and which specifies the
particular file to load.
Filter file names have the format DCAnnnnnnBTsssss.txt, where
nnnnnn is the module’s serial number and sssss is a user- specified
string that gives the file a unique name. The string parameter specifies
this second field; the remainder of the file name is filled in automatically.
Returns A status value of type FG_STATUS reflecting the result of the operation.
MeasPatternAcqDetJitterRise
Measures the rising- edge deterministic (data- dependent) jitter between
the specified bit sequences.
Call float MeasPatternAcqDetJitterRise(long nPatternFast, long nLengthFast, long nPatternSlow,
long nLengthSlow)
Parameters 1:
The bit sequence to be used for the “fast,” or “high- frequency”
transition.
2:
The length (number of bits) of the “fast” sequence.
3:
The bit sequence to be used for the “slow,” or “low- frequency”
transition.
9-71
Programming
Pattern Sequence Acquisition
4:
The length (number of bits) of the “slow” sequence.
Transitions may be up to 16 bits in length. The bit sequences must be
right- aligned in the data word, for example, the bit sequence 101 is
represented as a pattern value of 5 with length 3. Bits are matched
left to right with the acquired data sequence.
Returns A value of type float representing the deterministic jitter in picoseconds.
NOTE
The functions MeasPatternAcqDetJitterRise and MeasPatternAcqDetJitterFall select specific
transitions by marking other transitions as invalid. After these measurements, sample data
returned by GetDCASampleData or GetDCASampleDataVariant will contain samples with type
code 0x08 (invalid) for such transitions.
MeasPatternAcqDetJitterFall
Measures the falling- edge deterministic (data- dependent) jitter between
the specified bit sequences. The functions MeasPatternAcqDetJitterRise
and MeasPatternAcqDetJitterFall.
Select specific transitions by marking other transitions as invalid. After
these measurements, sample data returned by GetDCASampleData or
GetDCASampleDataVariant will contain samples with type code 0x08
(invalid) for such transitions.
Call float MeasPatternAcqDetJitterFall(long nPatternFast, long nLengthFast, long nPatternSlow,
long nLengthSlow)
Parameters 1:
The bit sequence to be used for the “fast,” or “high- frequency”
transition.
2:
The length (number of bits) of the “fast” sequence.
3:
The bit sequence to be used for the “slow,” or “low- frequency”
transition.
4:
The length (number of bits) of the “slow” sequence.
Transitions may be up to 16 bits in length. The bit sequences must be
right- aligned in the data word. For example, the bit sequence 101 is
represented as a pattern value of 5 with length 3. Bits are matched
left to right with the acquired data sequence.
Returns A value of type float representing the deterministic jitter in picoseconds.
9-72
Programming
Pattern Sequence Acquisition
MeasPatternAcqStdDetJitterRise
Measures the maximum rising- edge deterministic (data- dependent) jitter between any bit sequences in the acquired pattern. This function
selects specific transitions by marking other transitions as invalid.
After this measurement, sample data returned by "“GetDCASampleData” on page 9- 56," “GetDCASampleDataVariant” on page 9- 58 will
contain samples with type code 0x08 (invalid) for such transitions.
Call float MeasPatternAcqStdDetJitterRise
Parameters None
Returns A value of type float representing the deterministic jitter in picoseconds.
MeasPatternAcqStdDetJitterFall
Measures the maximum falling- edge deterministic (data- dependent) jitter between any bit sequences in the acquired pattern. This function
selects specific transitions by marking other transitions as invalid.
After this measurement, sample data returned by "“GetDCASampleData” on page 9- 56," “GetDCASampleDataVariant” on page 9- 58 will
contain samples with type code 0x08 (invalid) for such transitions.
Call float MeasPatternAcqStdDetJitterFall
Parameters None
Returns A value of type float representing the deterministic jitter in picoseconds.
MeasPatternSpecificRiseTime
Measures the rise time of transitions described by the length and pattern parameters.
Call float MeasPatternSpecificRiseTime(long nLength, long nPattern).
Parameter 1:
2:
The length (number of bits) of the sequence
The bit state sequence, as a right- justified integer
9-73
Programming
Pattern Sequence Acquisition
Transitions may be up to 16 bits in length. The bit sequences must be
right- aligned in the data word, for example, the bit sequence 101 is
represented as a pattern value of 5 with length 3. Bits are matched
left to right with the acquired data sequence.
For a valid rise time measurement, he transition specified must end
with the states 01. Invalid transition specifiers will return a value of 0
and will set the last error flag.
Returns A value of type float representing the rise time in picoseconds.
MeasPatternSpecificFallTime
Measures the rise time of transitions described by the length and pattern parameters.
Call float MeasPatternSpecificFallTime(long nLength, long nPattern).
Parameter 1:
2:
The length (number of bits) of the sequence
The bit state sequence, as a right- justified integer
Transitions may be up to 16 bits in length. The bit sequences must be
right- aligned in the data word, for example, the bit sequence 101 is
represented as a pattern value of 5 with length 3. Bits are matched
left to right with the acquired data sequence.
For a valid rise time measurement, he transition specified must end
with the states 01. Invalid transition specifiers will return a value of 0
and will set the last error flag.
Returns A value of type float representing the rise time in picoseconds.
GetPatternAcqData
Gets the acquired pattern data sequence as a string of binary or hex
digits.
Call BSTR GetPatternAcqData(int nFormatSel)
9-74
Programming
Calibration
NOTE
The data sequence is aligned so that the longest string of consecutive identical bits (1s or 0s)
comes first. If the longest string of 1s and the longest string of 0s are the same length, the string
of 0s comes first. If there is no unique longest substring, the sequence begins with one of the
longest substrings, chosen randomly; in this case, successive acquisitions may not begin with the
same substring.
Parameter An integer value that specifies the data format, as follows:
0:
Binary data
1:
Hex data
Returns A string containing the data as digits in the selected data format.
Calibration
• “DarkCal” on page 9-75.
• “DelayCalStart” on page 9-75.
• “DelayCalGetState” on page 9-75.
DarkCal
Measures and saves the optical dark current level.
Call void DarkCal( )
Parameters NA
DelayCalStart
Start or stop delay calibration for the currently selected input and filter.
Call void DelayCalStart(int nFunc)
Parameters An integer value that specifies the Parameter values:
0:
Stop delay calibration
1:
Start delay calibration
DelayCalGetState
Return the state of the delay calibration operation in progress.
9-75
Programming
Control Display
Call long DelayCalGetState( )
Parameters NA
Returns 0:
1:
Delay calibration is not running (has finished).
Delay calibration is running.
Control Display
•
•
•
•
•
•
•
•
•
“ClearDisplay” on page 9-76.
“CopyGraphToClipboard” on page 9-76.
“CopyGraphToFile” on page 9-77.
“DisplayMode” on page 9-77.
“EyeDisplayColorSel” on page 9-78.
“EyeDisplayTraceSel” on page 9-78.
“EyeDisplayPixelSizeSel” on page 9-79.
“EyeDisplayDeferredUpdate” on page 9-79.
“EyeDisplayCursorX0 EyeDisplayCursorX1 EyeDisplayCursorY0
EyeDisplayCursorY1” on page 9-79.
• “JpegQuality” on page 9-79.
ClearDisplay
Clear the eye window display.
Call void ClearDisplay( )
Parameters NA
CopyGraphToClipboard
Place a bitmap copy of the eye display window image on the Windows
clipboard.
Call long CopyGraphToClipboard(BSTR szCaption )
Parameters A character string specifying a caption for the image. If the parameter
value is NULL or the caption string is empty, no caption is included.
Returns 0:
The operation succeeded.
- 1: The operation failed because the eye window is not displayed.
9-76
Programming
Control Display
- 2: The operation failed because of insufficient memory or system
resources.
- 3: The operation failed because the Windows clipboard was not
accessible.
CopyGraphToFile
Copy the image on the display window to an image file.
Call long CopyGraphToFile(int nFileType, BSTR szFilePath, BSTR szCaption )
Parameters
• An integer value that specifies the file format:
0:
Bitmap (.bmp) file format
1:
JPEG (.jpg) file format
• A character string specifying the file name and path.
• A character string specifying a caption for the image. If the parameter
value is NULL or the caption string is empty, no caption is included.
NOTE
Portions of this software make use of the CxImage image-processing library,
www.codeproject.com/bitmap/cximage.asp.
Returns 0:
The operation succeeded.
- 1: The operation failed because the eye window is not displayed.
- 2: The operation failed because of insufficient memory or system
resources.
- 4: The operation failed because the file format specifier is invalid.
- 5: An internal error occurred.
- 6: An internal error occurred.
- 7: An error occurred saving the file.
DisplayMode
Selects whether the ActiveX control displays the eye window or is
invisible, and, if visible, selects the information displayed.
Parameters An integer value that specifies the display mode.
9-77
Programming
Control Display
0: Window is not visible; acquired sample data are not transferred to
the host.
1:
Display the acquired data as an eye diagram.
2:
Display the eye with cursors indicating the measurement results.
3:
Display the eye, cursors, and the measured values as text.
4: Window is not visible; acquired sample data are transferred to the
host and may be read using the GetDCASampleData and GetDCASampleDataVariant methods.
EyeDisplayColorSel
Selects display color mode.
Parameters An integer value that specifies the color mode.
0. Disable trace coloring.
1. Display traces in four different colors based on only the current
transition (LL, LH, HL, HH).
2. Display traces in eight different colors based on the current and
previous transitions (LLL, LLH, LHL, LHH, HLL, HLH, HHL, HHH).
EyeDisplayTraceSel
Enables the selective display of different transition types.
Parameters An integer value whose bits enable display of the different transition
types.
1 LLL
2 LLH
4 HHL
8 HHH
16 HLL
32 HLH
64 LHL
128 LHH
Traces are displayed whose corresponding bits are set. Note that a
parameter value of 0 enables the display of all traces.
9-78
Programming
Control Display
EyeDisplayPixelSizeSel
Selects pixel size used for drawing the eye display.
Parameters An integer value that specifies the pixel size.
0
Small pixels (1 x 1)
1
Large pixels (2 x 2)
EyeDisplayDeferredUpdate
Select whether the display is updated incrementally, as points are read
from the DCA module, or only after all points have been read.
Parameters An integer value that specifies the update mode.
Zero (0) updates incrementally.
One (1) updates in a single action, after all points have been read.
EyeDisplayCursorX0
EyeDisplayCursorX1
EyeDisplayCursorY0
EyeDisplayCursorY1
Measurement cursor position values.
Read- only properties that return the positions of the cursors that show
measurement results. The values returned are floats; the units are unit
intervals (UI) for the X cursors, uW or mV for the Y cursors. The values returned pertain to the most recently performed measurement; for
cursors that are not displayed, the value 0 is returned.
Parameters NA
JpegQuality
Sets the image quality and file compression for JPEG image files created by the method, “CopyGraphToFile” on page 9- 77.
Parameters An integer value in the range 0 to 100. Greater values result in higherquality images and larger image files.
9-79
Programming
Status
NOTE
Portions of this software make use of the CxImage image-processing library,
www.codeproject.com/bitmap/cximage.asp.
Status
GetLastError
Returns the result of the last operation performed.
Call long GetLastError( )
Parameters NA
Returns A status value of type FG_STATUS reflecting the result of the operation.
9-80
Programming
ActiveX API Properties
ActiveX API Properties
• “DCA Configuration” on page 9- 81.
• “Pattern Sequence Acquisition” on page 9- 84.
DCA Configuration
•
•
•
•
•
•
•
•
•
•
•
•
“LineRate” on page 9-81.
“Wavelength” on page 9-81.
“OptAttenRatio” on page 9-81.
“ElectAttenRatio” on page 9-82.
“RefClockSel” on page 9-82.
“RefClockFreq” on page 9-83.
“DCASampleBlockCount” on page 9-83.
“DCADisplayBlockCount” on page 9-83.
“DCAInputPathSelect” on page 9-83.
“NRZThresholdSel” on page 9-84.
“NRZWindowSize” on page 9-84.
“NRZOMAWindowSize” on page 9-84.
LineRate
Specifies the line data rate.
Type Value of type double that specifies the rate in MHz.
Wavelength
Configures the DCA for the specified optical input wavelength.
Type Enum value of type FG_WAVELENGTH_TYPE that specifies the wavelength.
OptAttenRatio
Informs the DCA of an external attenuator in the optical input path.
DCA measurements will use this value so that the measured values
reflect the signal at the input of this attenuator.
9-81
Programming
DCA Configuration
Type Value of type float that specifies the ratio by which the attenuator
reduces the optical power. For example, a 3 dB optical attenuator
reduces the power by a factor of 2:1; it would be entered as an attenuation ratio of 2.0
Attenuation specified in dB may be converted to power ratio according
to the following equation:
Ratio = 10dB/10
The power ratio may be converted to dB according to the following
equation:
dB = 10 x log10(ratio)
For example, a 3 dB optical attenuator reduces the power by a factor
of 2:1; it would be entered as an attenuation ratio of 2.0.
If no attenuator is used, this value should be set to 1.0.
ElectAttenRatio
Specifies an external attenuator in the electrical input path. DCA measurements will use this value so that the measured values reflect the
signal at the input of this attenuator.
Type Value of type float that specifies the ratio by which the attenuator
reduces the input voltage. For example, a 6 dB electrical attenuator
reduces the voltage by a factor of 2:1; it would be entered as an attenuation ratio of 2.0.
Attenuation specified in dB may be converted to voltage ratio according to the following equation:
Ratio = 10dB/20
The power ratio may be converted to dB according to the following
equation:
dB = 20 x log10(ratio)
For example, a 6 dB electrical attenuator reduces the voltage by a factor of 2:1; it would be entered as an attenuation ratio of 2.0.
If no attenuator is used, this value should be set to 1.0.
RefClockSel
Specifies the timing source (reference clock) for the DCA’s data acquisition.
9-82
Programming
DCA Configuration
Type Enum value of type FG_REFCLKSLT that specifies the timing source.
RefClockFreq
Specifies the clock frequency to be used when the reference clock
source selection is FG_REFCLK_EXTERNAL.
Type Value of type double that specifies the frequency in MHz.
RecoveredClockRatio
Specifies the ratio of the line rate to the recovered clock rate when
the reference clock source selection is
FG_REFCLK_RECOVERED_RATIO. See a note on page 9- 45 for further
details.
Call FG_STATUS FGCfgSetRecoveredClockRatioAPI(FG_HANDLE hModule, double dFreq)
Parameters
• The module connection handle returned by fg_connect.
• A value of type float that specifes the ratio.
Returns A status value of type FG_STATUS reflecting the result of the operation.
DCASampleBlockCount
Specifies the number of 1024- sample blocks to be taken when the DCA
acquires data.
Type Integer value that specifies the number of 1024- sample blocks.
DCADisplayBlockCount
Specifies the number of sample blocks to be displayed, and available
for transfer to the user’s buffer. A value of 0 specifies “all.”
Type Integer value
DCAInputPathSelect
Selects the optical or electrical input and the input filter, if applicable.
For proper DCA operation, the input not selected must have no signal
applied.
9-83
Programming
Pattern Sequence Acquisition
Parameters Enum value of type FG_FILTER_TYPE that selects the input and filter.
The available selections are returned by the queries, “GetModuleInformation” on page 9- 46 for input configuration and available filters.
NRZThresholdSel
Selects the low and high amplitude threshold values used for NRZ
measurements.
Parameters A value of type FG_HILOW_VLTREF that specifies the thresholds.
NRZWindowSize
Sets the start and end of the time window, in percent of 1 UI, over
which NRZ measurements are made. The window is positioned in the
center of the eye.
Parameters An integer value between 2 and 98 that specifies the window size.
NRZOMAWindowSize
Sets the start and end of the time window, in percent of 1 UI, over
which NRZ OMA measurements are made. The window is centered on
the eye crossing.
Parameters An integer value between 2 and 98 that specifies the window size.
Pattern Sequence Acquisition
•
•
•
•
“PatternAcqLength” on page 9-84.
“PatternAcqFilterSel” on page 9-85.
“PatternAcqDisplayBits” on page 9-85.
“PatternAcqFileLoaded (read-only)” on page 9-85.
PatternAcqLength
Enables pattern sequence acquisition and specifies the sequence
length.
Parameters An integer value.
0. Disable pattern sequence acquisition.
4 ≤ N ≤ 2047:
9-84
Acquire data with the specified sequence length.
Programming
Pattern Sequence Acquisition
PatternAcqFilterSel
Enables or disables the software filtering.
Parameters An integer value.
0: Unfiltered
1: Bessel- Thompson filter
PatternAcqDisplayBits
Selects the number of bits of the reconstructed pattern sequence data
to be displayed.
Parameters An integer value.
0: Display all bits of the reconstructed pattern sequence.
> 0: Display the specified number of bits.
PatternAcqFileLoaded (read-only)
Returns the status of the most recent filter file load operation.
Parameters An integer value.
0: The specified file was not found or no file has been loaded.
1: The specified file was successfully loaded.
9-85
Programming
Pattern Sequence Acquisition
9-86
10
Specifications
10- 2
Specifications
Specifications
Specifications
Specifications
The distinction between specifications and characteristics is described
as follows:
• Specifications describe warranted performance. All specifications apply
after the instrument is turned on for one (1) hour and while self- calibration is valid. Many performance parameters are enhanced through
frequent, simple user calibrations.
• Characteristics provide useful information by giving functional, but
nonwarranted, performance parameters. Characteristics are printed in
italics.
Table 13. General Specifications
Sample rate
160 Ms/s or 155 Ms/s
Points per sample block
1024
Maximum number of sample blocks
1024
Pattern acquisition maximum length
2047 bits
Fixed number of points per bit in pattern acquisition mode
128
Table 14. Display Specifications: Optical Vertical Scale Factor Per Division
Minimum vertical scale factor
1 μW
Maximum vertical scale factor
500 μW
Number of vertical divisions
8
10-2
Specifications
Specifications
Table 15. Display Specifications: Electrical Vertical Scale Factor Per Division
Minimum vertical scale factor
1 mV/div
Maximum vertical scale factor
100 mV/div
Maximum number of divisions
8
Vertical scale settings
1, 2, 5, 10, 20, 50, 100, 200,
and 450
Table 16. Display Specifications: Horizontal Specifications
Minimum X-scale setting
1 ps/div
Maximum X-scale setting
20 ns/div
Number of divisions (fixed)
10
Table 17. Electrical Specifications
Number of channels
1 single ended
BW of electrical input
9.5 GHz (characteristic)
Transition time (10% to 90%, calculated from TR = 0.35/BW)
37 ps (characteristic)
Connection type
SMA, AC Coupled
AC input voltage range
1 V pp (Max)
Impedance
50 ohm (nominal)
Absolute maximum input voltage without damage
2 V pp
Line rate coverage
155 Mb/s to 11.318 Gb/sa
a. Maximum line rate based on an input signal with 100 mVpp minimum signal.
10-3
Specifications
Specifications
Table 18. Clock Specifications
Clock recoverya
155 Mb/s to 2.7 Gb/s
(characteristic)
Clock input frequency range
10 MHz to 11.318 GHz
(characteristic)
Clock input impedance
50 ohm (nominal)
Maximum clock input amplitude without damage
2 Vpp
Clock input voltage range
0.5 to 1 V pp
a. - The CDR works with the specified PRBS up to 231–1.
- The CDR expects a DC balanced pattern.
- The CDR expects a transition density of one transition for every second bit.
Table 19. Optical Specifications (1 of 2)
Number of optical channels
1
Unfiltered Optical BW
7.5 GHz (characteristic)
Unfiltered option only available if
selected.
Optical connector
FC/PC
Fiber input
62/125 μm
Optical input return loss
850 nm: –14 dB (characteristic)
1310 nm: – 24 dB (characteristic)
Maximum non-destruct average
–3 dBm at 1310 nm
Maximum non-destruct peak
+7 dBm at 1310 nm
Average power monitor range
–30 dBm to –2 dBm at 850 nm
Wavelength range
750 to 1650 nm
Number of optical filters
4
10-4
Specifications
Specifications
Table 19. Optical Specifications (2 of 2)
Available filter choices
155 Mb/s, 622 Mb/s, 1.062 Gb/s,
1.250 Gb/s, 2.125 Gb/s, 2.488 Gb/s,
2.5 Gb/s, 3.125 Gb/s, 4.25 Gb/s,
5.0 Gb/s, 6.25 Gb/s, 8.5 Gb/s,
9.956 Gb/s, 10.3725 Gb/s,
10.5 Gb/s and unfiltered
Optical channel RMS noise at 850 nm
6.5 μW (filtered)a (characteristic)
Optical channel RMS noise at 1310/1550 nm
4.5 μW (filtered)a (characteristic)
Average optical power meter accuracy (single-mode)
5% + uncertainty of reference source
+ connector uncertainty + 200 nWb
(characteristic)
Average optical power meter accuracy (multi-mode)
10% + uncertainty of reference
source + connector uncertainty +
200 nW (characteristic)b
Line rate coverage
155 Mb/s to 11.318 Gb/sc
Calibrated wavelengths
850 nm
1310 nm, 1550 nm (characteristic)
a. The measurement was made with no input signal.
b. For -2 dBm to -25 dBm (at 850 nm).
c. Maximum line rate based on an input signal with 200 μW minimum OMA.
Table 20. Environmental Specifications
Temperature range
10oC to 40oC
Power consumption
40 W
Dimensions
Width: Four-slot PXI module
Height: 3U high PXI module
10-5
Specifications
Specifications
10-6
Index
A
Acquiring Data, 3–11
Active- X, 9–2
ActiveX, 9–4
air flow, 2–3
Amplitude, 4–9
Annotations, 5–15
API, 9–2
Average Optical Power, 4–16
Extinction Ratio with histogram,
4–21
Eye Amplitude, 4–22
Eye Diagram, 4–3
Eye Height, 4–21
Eye Mask Test Mode Measurements,
4–39
Eye Width, 4–23
eye- diagram analysis, 4–2
B
F
Base, 4–9
Bit Rate, 4–29
C
Calc Stats, 4–47
cleaning
fiber-optic connections, 1–13
non-lensed connectors, 1–13
Color Intensity Display, 5–7
Configuration Settings, 3–4
Connect, 2–5, 3–3
Control Panel, 3–2
Copy Another Mask, 4–42
Count, 3–11
Crossing Hist., 4–37
Crossing Histogram Window, 5–6
Crossing Percentage, 4–32
custom user masks, 4–42
D
DCA Parameters, 3–7
Deterministic Jitter, 4–36
dimensions, 10–5
Disconnect, 2–8
Display Background, 5–5
Display Colors, 5–3
Display Traces, 5–4
DLL, 9–3
Duty Cycle Distortion, 4–32
E
Edge Jitter, 4–34
electrostatic discharge, 1–5
Environmental Specifications, 10–5
Extinction Ratio, 4–18
Failing Points, 4–40
Fall Time, 4–31
Filter File Format, 6–6
Filter Files, 6–5
Find Max Margin, 4–41
Find Max. Margin, 4–45
Firmware, 2–9
floor mat, 1–5
Mask Margin, 7–3
Mask Margins Rules, Agilent, 7–6
Mask Margins Rules, Tektronix, 7–8
Mask Test mode, 4–39
Max, 4–14
Min, 4–13
Module Config, 2–5
Multiple Measurements Mode, 4–46
N
Negative Overshoot, 4–15
NRZ Measurement Mode, 4–17
NRZ Thresholds, 3–8
NRZ Window Size, 3–8
O
Global Parameters, 3–5
OMA, 4–24
OMA Window Size, 3–8
Optical modulation amplitude, 4–24
Oscilloscope Measurement mode, 4–6
Oscilloscope Mode Measurements,
4–5
H
P
heel strap, 1–5
High, 4–27
ID String, 2–5, 3–3
industry standard masks, 4–41
Inspect, 2–2
instrument
returning for service, 1–15
ISI Histogram, 4–38
Pass/Fail, 4–43
Pattern Acquisition, 3–7, 4–53, 6–2
Pattern Sequence D. J. Mode Measurements, 4–49
Pattern Sequence D.J. mode, 4–49,
4–50
PC, 2–3
Peak to Peak, 4–11
Positive Overshoot, 4–14
Presets, 3–10
Program Settings, 5–3
L
Q
Load DCA File, 8–3
Low, 4–26
quick confidence check, 2–5
G
I
M
Margin Compatibility option, 7–5
Markers, 4–7
Mask Definitions, 4–43
Mask Display, 5–3
R
remote bridge, 2–3
Repetitive, 3–11
Resolution, 5–7
returning to Agilent, 1–15
Rise Time, 4–29
Index-1
Index
RMS, 4–11
S
Sample Density Color Scheme, 5–6
Sample Density Display Information,
5–8
Scan for Instruments, 2–5, 3–3
Select Mask, 4–40, 4–41
service, 1–15
Set Defaults, 3–9
Set IP Addr, 3–9
shipping
procedure, 1–15
Show Version, 3–9
Show Y axis scale, 5–6
Signal to Noise Ratio, 4–28
Simulator data file format, 8–7
Single Edge Jitter, 4–35
software digital filters, 6–5
Specifications, 10–2, 10–4, 10–5
Start Acquisition, 2–6, 3–11, 4–40
static- safe accessories, 1–6
Statistics, 4–47
T
table mat, 1–5
U
Units, 5–5
User- written Applications, 8–7
W
wrist strap, 1–5
X
X Axis Control, 4–6
Y
Y Axis Control, 4–7
Y- Scale, 5–5
Index-2