HM8012 - PowerTronics
HM8012
4¾-Digit Programmable Multimeter
HM8012
HZ15 (included)
R 4¾-Digit Display with 50,000 Counts
R Basic Accuracy 0.05 %
R Max. Resolution: 10 µV, 0.01 dBm, 10 nA, 10 mΩ, 0.1 °C
R Offset Function/Relative Value Measurement
WDM8012 Software
(included)
Mainframe HM8001-2
required for Operation
R RS-232 Interface and Software included
Measurement voltage:
4¾-Digit Programmable Multimeter HM8012
All data valid at 23 °C after 30 minutes warm-up.
DC voltage
500 mV, 5 V, 50 V, 500 V, 600 V
Measurement ranges:
10 µV, 100 µV, 1 mV, 10 mV, 100 mV
Resolution:
Accuracy:
5 V, 500 V, 600 V:
±(0.05 % of reading +0.002 % of full scale)
500 mV, 50 V:
±(0.05 % of reading +0.004 % of full scale)
Overload protection:
V/Ω/T°/dB/
to
850 Vp at max. 60 Hz or 600 Vdc
COM and to chassis:
COM against chassis:
250 Vrms at max. 60 Hz or 250 Vdc
Input resistance:
50 V, 500 V, 600 V:
10 MΩ II 90 pF
500 mV, 5 V:
>1 GΩ II 90 pF
10 pA
Input current:
Common mode rejection ratio: ≥100 dB (50/60 Hz ±0.5 %)
Serial mode rejection ratio: ≥60 dB (50/60 Hz ±0.5 %)
dB Mode
Accuracy:
Resolution:
DC current
Measurement ranges:
Resolution:
Accuracy:
0.5…500 mA:
10 A:
Voltage drop:
10 A range:
500 mA range:
other ranges:
AC voltage
Measurement ranges:
Resolution:
Accuracy 0.5…50 V:
40 Hz…5 kHz:
20 Hz…20 kHz:
Accuracy 500 V and 600 V:
40 Hz…1 kHz:
20 Hz…1 kHz:
Overload protection:
V/Ω/T°/dB/
to
COM and to chassis:
COM against chassis:
Input impedance
AC mode:
AC + DC mode:
Bandwidth at -3 dB:
dB mode:
Accuracy
-23.8…59.8 dBm:
Resolution:
CMRR1):
Crest factor:
AC current
Measurement ranges:
Resolution:
Accuracy:
0.5…500 mA:
10 A:
±(0.02 dB + 2 digits) (display >-38.7 dBm)
0.01 dB above 18 % of rating
500 µA, 5 mA, 50 mA, 500 mA, 10 A
10 nA, 100 nA, 1 µA, 10 µA, 1 mA
±(0.2 % of reading + 0.004 % of full scale)
±(0.3 % of reading + 0.004 % of full scale)
0.2 V max.
2.5 V max.
0.7 V max.
500 mV, 5 V, 50 V, 500 V, 600 V
10 µV, 100 µV, 1 mV, 10 mV, 100 mV
±(0.4 % of reading + 0.07 % of full scale)
±(1 % of reading + 0.07 % of full scale)
±(0.4 % of reading + 0.07 % of full scale)
±(1 % of reading + 0.07 % of full scale)
Temperature
2-wire resistance
­measurement
Range:
Resolution:
Measurement current:
Display:
Accuracy:
10 V typical for open inputs, depending
on the value of resistance to be measured.
Negative polarity of measurement voltage
is across common terminal.
with linearization for PT100 sensors as per
standard EN60751
-200…+500 °C
0.1 °C
approx. 1 mA
in °C, °F
±(0.4 °C + 0.0005 x T) from -200…+200 °C
±(0.5 °C + 0.0005 x T) from +200…+500 °C
(T in °C, sensor tolerance not included)
Temperature coefficient: (reference 23 °C)
V = 500 mV, 50 V
30 ppm/°C
600 V range
80 ppm/°C
other ranges
20 ppm/°C
V ~ 600 V range
80 ppm/°C
other ranges
50 ppm/°C
mA all ranges
200 ppm/°C
mA~all ranges
300 ppm/°C
Ω 5 MΩ, 50 MΩ ranges
200 ppm/°C
other ranges
50 ppm/°C
Miscellaneous
Power supply (from mainframe):
+5 V
300 mA
~26 V
140 mA
+5…+40 °C
Operating temperature:
-20…+70 °C
Storage temperature:
5…80 % (non condensing)
Rel. humidity:
Dimensions (W x H x D)
(without 22-pole flat plug): 135 x 68 x 228 mm
approx. 0.5 kg
Weight:
Accessories supplied: Operator’s Manual, Interface cable (HZ14),
PVC test leads (HZ15), Software CD
Recommended accessories:
HZ10S 5 x silicone test lead (measurement connection in black)
HZ10R 5 x silicone test lead (measurement connection in red)
HZ10B 5 x silicone test lead (measurement connection in blue)
HZ812 PT100 Temperature probe
850 Vp at max. 60 Hz or 600 Vdc
250 Vrms at max. 60 Hz or 250 Vdc
1 MΩ II 90 pF
10 MΩ II 90 pF
80 kHz typical
20 Hz…20 kHz
±0.2 dBm
0.01 dB above 9 mV
≥60 dB (50/60 Hz ±0.5 %)
7 max.
500 µA, 5 mA, 50 mA, 500 mA, 10 A
10 nA, 100 nA, 1 µA, 10 µA, 1 mA
±(0.7 % of reading + 0.07 % of f.s.)
40 Hz…5 kHz
±(1 % of reading + 0.07 % of full scale)
AC + DC measurements
As shown for AC + 25 digit
Resistance
500 Ω, 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ, 50 MΩ
Measurement ranges:
10 mΩ, 100 mΩ, 1 Ω, 10 Ω, 100 Ω, 1 kΩ
Resolution:
Accuracy:
500 Ω…500 kΩ:
±(0.05 % of reading + 0.004 % of f.s.+50 mΩ)
5…50MΩ:
±(0.3 % of reading + 0.004 % of full scale)
Input protection max. 300 Vrms
500 Ω…5kΩ range:
1mA
Measurement current:
50 kΩ range:
100µA
500 kΩ range:
10 µA
5…50 MΩ range:
100 nA
HM8012E/121010 · C&E · Subject to change without notice · © HAMEG Instruments GmbH® · DQS-certified in accordance with DIN EN ISO 9001:2008, Reg.-No.: 071040 QM08
HAMEG Instruments GmbH · Industriestr. 6 · D-63533 Mainhausen · Tel +49 (0) 6182 8000 · Fax +49 (0) 6182 800100 · www.hameg.com · [email protected]
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertisement