Huvitz HS-7000 Operator's Manual
Add to my manuals
49 Pages
Huvitz HS-7000 is a slit lamp microscope designed for eye examinations of the anterior eye segments, from the cornea epithelium to the posterior capsule. It aids in diagnosing diseases or trauma affecting the structural properties of the anterior eye segment. Features include a halogen projection bulb, magnification changer, applanation tonometer, and a blue filter.
advertisement
------------------------------------------------------ -----------------------------SLIT LAMP H
IS-7000 1
Operator’s Manual
SLIT LAMP MICROSCOPE
HIS-7000
2 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
IMPORTANT NOTICE
!
WARNING
P ote ntial e le ctro magne tic or o ther inte rfe re nce be twe e n me dic al e quipme nts and o the r de vic e s being o perate d to ge ther in the same e nviro nme ntal may e xpe rt an ad ve rse influe nce o n func tio ning o f the me dic al e quipme nt. No n me dic al e quipme nts no t in c o mplianc e with the re quire me nts of EN 606 01 -1 and EN 6 060 1 -1 -2 sho uld no t be use d to ge ther in the same e nviro nme ntal as the me dic al e quipme nts
This e quipme nt has be e n te ste d and fo und to c o mply with the limits fo r me dic al de vic es in
IEC 6 0 6 01 -
1-
2:200 1 . The se limits are de signe d to pro vide re aso nable pro tec tio n against harmful interfe re nc e in a typic al me dic al installatio n
P o we r Co rd
Fo r use of e quipme nt in rate d vo ltage le ss than 12 5 Vac , minimum 6 A, Type SJT o r S VT,
1 8/3 AW G, 10 A, ma x 3 .0 m lo ng: One e nd with Ho spital Grade Ty pe , NEM A 5 15 P o the r e nd with a pplianc e co uple r. Fo r use of e quipme nt in rate d vo ltage le ss than 2 50 Vac
, minimum 6 A, Type SJT o r SVT, 1 8 /3 AW G, 1 0 A, max 3 .0 m lo ng:
O ne e nd te rminate d with blade attac hme nt plug (HAR) Ty pe , NEM A 6 15 P .
This pro duct may malfunc tio n due to elec tro magne tic wave s c ause d by po rtable pe rso nal te le pho nes, transce ive rs, radio -co ntro lle d to ys, e tc . Be sure to avo id ha vi ng o bje c ts suc h as, whic h affe c t this pro duc t, bro ught ne ar the pro duc t.
The informatio n in this public atio n has bee n c are fully c hec ke d and is be lie ve d to be e ntire ly ac c urate at the time of public atio n. HUVITZ assume s no respo nsibility, ho we ve r, fo r possible e rrors o r o missio ns, o r for any c o nse que nc es re sulting fro m the use o f the o rmatio n c o ntaine d he rein.
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 3
HUVITZ re se rve s the right to make c hange s in its pro ducts or pro duct spec ific atio ns at any time and witho ut prio r notic e , and is no t re quire d to update this doc ume ntatio n to refle ct suc h c hange s.
© 20 09 HUVITZ Co ., Ltd.
6 89 -3 , Ge umjeo ng
do ng, Gunpo
si, Gye o nggi
-do,
4 35 -86 2 , Re public of Kore a
All rights are re se rve d.
Unde r c o pyright laws, this manual ma y no t be c o pie d, in who le o r in part, witho ut the prio r w ritte n co nse nt of HUVITZ Co ., Ltd.
Do c ume nt Numbe r: 9 000 ENGJ2 01
-A.
4 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
Contents
1.
Introduction ....................................................................................................... 6
1.1.
1.2.
Indication for use
...........................................................................................6
Classification............................................................................................ 6
2.
Safety Infor mation.............................................................................................. 6
2.1.
2.2.
2.3.
2.4.
Introduction .............................................................................................. 7
Safety Symbols ........................................................................................ 8
Environment factors................................................................................ 10
Safety Precautions.................................................................................. 12
3.
Features & Operating Condit ion ...................................................................... 14
3.1.
3.2.
Brief Description ..................................................................................... 14
Operating Condition ................................................................................ 14
4.
Standard Components ..................................................................................... 16
5.
Accessories ..................................................................................................... 17
6.
Composit ion .................................................................................................... 18
7.
Installation ....................................................................................................... 24
7.1.
Assembling the Instrument ...................................................................... 24
8.
Trial Operations ............................................................................................... 28
9.
Main Features .................................................................................................. 29
10. Ordinary Maintenance ...................................................................................... 30
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 5
10.1.
Halogen Projection bulb .......................................................................... 30
10.2.
Protection fuses replacement................................................................... 32
10.3.
Cleaning & Disinfection ........................................................................... 33
11.
EMC Infor mation .............................................................................................. 34
11.1.
Guidance and manufacturer’s declaration – electromagnetic emissions...... 34
11.2.
Guidance and manufacturer’s declaration – electromagnetic immunity ....... 35
11.3.
Guidance and manufacturer’s declaration – electromagnetic immunity ....... 37
11.4.
Recommended separation distances between portable and mobile RF communications equipment and the EUT.............................................................. 39
11.5.
Immunity and Compliance Level .............................................................. 40
11.6.
Guidance and manufacturer’s declaration – electromagnetic immunity ....... 41
12. Electric Diagram............................................................................................... 43
13. Technical Specifications .................................................................................. 44
14. Service Information .......................................................................................... 46
15. Particular Information ……………………………………………………………………….48
6 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
1.
Introduction
1.1. Indication for use (intended use)
The SLIT LAMP,
HIS-7000 is intended for use in eye examination of the ant eye segments, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affects
the structural properties of the anterior eye segment.
1.2. Classification
-
-
-
-
-
-
Protection against electric shock: Class I
Applied Parts: Type B applied parts.
Protection against harmful ingress of water: Ordinary, IPX0
Degree of safety in the presence of a flammable anesthetics mixture with air or with oxygen or with nitrous oxide: Not suitable for use in the presence of a flammable anesthetics mixture with air or with oxygen or with nitrous oxide.
Mode of operation: Continuous
Light Hazard protection Classification: Group 2 by ISO15004-2:2007
Stationary: Do not try to move this equipment without help. This microscope is heavy and should maintain stationary. Always get two persons to help lift this unit
!
CAUTION
Federal law restricts this device to sale by or on the of a physician
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 7
2.
Safety Information
2.1. Introduction
Safety is everyone’s responsibility. The safe use of this equipment is largely dependent upon the installer, user, operator, and maintainer. It is imperative that personnel study and become familiar with this entire manual before attempting to install use, clean, service or adjust this equipment and any associated accessories. It is paramount that the instructions contained in this manual are fully understood and followed to enhance safety to the patient and the user/operator. It is for this reason that the following safety notices have been placed appropriately within the text of this manual to highlight safety related information or information requiring special emphasis. All users, operators, and maintainers must be familiar with and pay particular attention to all Warnings and
Cautions incorporated herein.
!
WARNING
“ Warning ” indic ate s the prese nc e o f a hazard that c o uld re sult in se ve re perso nal injury, de ath o r substantial pro perty damage if igno re d
.
NOTE
“
No te ” de sc ribe s informatio n fo r the installatio n, o peratio n, o r mainte nanc e o f whic h is impo rtant but ha zard re late d if ignore d.
!
CAUTION
“Cautio n” indic ate s the prese nce o f a hazard th at c o uld re sult in mino r injury, or pro pe rty damage d if igno re d.
8 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
2.2. Safety Symbols
The International Electro technical Commission (IEC) has established a set of symbols for medical electronic equipment which classify a connection or warn of any potential hazards. The classifications and symbols are shown below.
This applies only to the instrument that has the certification symbol printed explicitly on the product label or sticker.
Save this instruction
I an d O o n po we r switc h re prese nt ON and OFF re spe ctive ly.
Ty pe B Iso late d patie nt c o nnec tio n.
This symbo l ide ntifies a safety no te . Ensure yo u unde rstand the func tio n of this co ntro l before using it. Co ntro l o n is de scribe d in the appro priate User ’s or Se rvic e M anual.
Ide ntifie s the po int whe re the syste m safety gro und is faste ne d to the c hassis. P ro te ctive e arth co nnec te d to c o nduc tive parts o f Class I e quipme nt fo r safe ty purposes.
Ho t surface .
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 9
UL6 06 01 -1
CAN/CS A C2 2 .2
NO.6 01 .1
M EDIC AL EUIP M ENT
W ITH RESP EC T TO ELEC TRIC SHOCK
FIRE AND M E CH ANIC AL H AZ ARDS
ONLY IN ACCO RD ANCE W ITH U L 6 0 601 -
1 , AND C AN/CS A C2 2 .2 NO.601 .1
Disposal of yo ur old applianc e
W he n this crosse d-o ut whe e le d bin symbo l is attac he d to a pro duc t it me ans the pro duc t is co ve re d by the Euro pe an
Direc tive 2 00 2/96/EC.
All e lec tric al and e le ctro nic pro duc ts sho uld be dispo se d of se parately fro m the municipal waste stre am via de signate d c olle ctio n fac ilities appo inte d by the go ve rnme nt or the lo c al autho rities.
The c orre ct disposal of yo ur o ld applianc e will he lp pre ve nt po te ntial ne gative c o nse que nce s for the e nviro nme nt and human he alth.
4 . Fo r mo re de taile d info rmatio n abo ut disposal of yo ur o ld applianc e , ple ase c o ntact yo ur c ity offic e , waste dispo sal servic e or the sho p whe re yo u purc hase d the pro duc t.
Alte rnating Curre nt
10 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
2.3. Environment factors
Avoid the following environments for operation or storage:
W he re the e quipme nt is e xpo se d to wate r vapo r.
Do n ’ t o perate e quipme nt with a we t han d.
W he re the e quipme nt is e xpo se d to direc t sunlight.
W he re the te mperature c hange s e xtre me ly
.
Humidity is fro m 3 0 % to 75 % .
Altitude is fro m 70 0 to 1 060 hpa
W he re it is ne ar the he at e quipme nt.
W he re the humidity is e xtre me ly high o r the re is a ve ntilatio n pro ble m.
W he re the e quipme nt is subje ct to e xce ssive sho c ks or vibratio ns.
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 11
W he re e quipme nt is e xpose d to c he mic al material or e xplo sive gas.
Be c are ful not to be inse rte d dust, e spe cially, me tal.
Do n
’ t disasse mble the pro duc t or o pe n. We are n
’ t respo nsible fo r it fo r no thing.
Be c are ful no t to close the fan lo c ate d o n the lateral or bac k side o f the e quipme nt.
Do n
’ t plug the AC po we r c o rd into the o utle t before the c o nnec tio n be twe e n de vic es of the e quipme nt is co mplete d.
This c an ge ne rate the de fec t.
P ull o ut the po we r c ord with ho lding the plug, no t the c ord.
Avoid places where the ambient temperature falls below 10
℃
or exceeds 40
℃
for normal operation, or below -10
℃
or exceeds 40
℃
(14
℉
-104
℉
) for transportation and storage. Humidity should be maintained between 30 and for normal operation, transportation and storage. The atmospheric pressure in operation, transportation and storage is from 700 to 1060hpa.
12 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
2.4. Safety Precautions
This equipment has been developed and tested according to safety standards as well as national and international standards. This guarantees a very high degree of safety for this device. The legislator expects us to inform the user expressively about the safety aspects in dealing with the device. The correct handling of this equipment is imperative for its safe operation. Therefore, please read carefully all instructions before switching on this device. For more detailed informatio please contact our Customer
Service Department or one of our authorized representatives.
1.
This equipment must not be used (a) in an area that is in danger of explosions and (b) in the presence of flammable, explosive, or vo tile solvent such as alcohol, benzene or similar chemicals.
2.
Do not put or use this device in humid rooms. Humidity should be maintained between 30 and 75% for normal operation. Do not expose the device to water splashes, dripping water, or sprayed water. Do not place containers containing fluids, liquids, or gases on top of any electrical equipment or devices.
3.
The equipment must be operated only by, or under direct supervision of a properly trained and qualified person.
4.
Modifications of this equipment may only be carried out by Huvitz’s service technicians or other authorized persons.
5.
Customer maintenance of this equipment may only be performed as stated in the User’s Manual and Service Manual. Any additional maintenance may only be performed by Huvitz’s service technicians or other authorized persons.
6.
The manufacturer is only responsible for effects on safety, reliability, and performance of this equipment when the following requirements are fulfilled: (1)
The electrical installation in the respective room corresponds to the specifications stated in this manual and (2) This equipment is used, operated,
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 13
And maintained according to this manual and Service Manual.
7.
The manufacturer is not liable for damage caused by unauthorized tampering with the device(s). Such tampering will forfeit any rights to claim under warranty.
8.
This equipment may only be used together with accessories supplied by
Huvitz’s. If the customer makes use of other accessories, use them only if their safe usability under technical safety aspects has been proved and confirmed by Huvitz or the manufacturer of the accessory.
9.
Only persons who have undergone proper training and instructions are authorized to install, use, operate, and maintain this equipment.
10.
Keep the User’s Manual and Service Manual in a place easily accessible at all times for persons operating and maintaining the equipment.
11.
Do not force cable connections. If a cable does not connect easily, be sure that the connector (plug) is appropriate for the receptacle (socket). If you cause any damage to a cable connector(s) or receptacle(s), let the damage(s) be repaired by an authorized service technician.
12.
Please do not pull on any cable. Always hold on to the plug when disconnecting cables.
13.
Before every operation, visually check the equipment for exterior mechanical damage(s) and for proper function.
14.
Do not cover any ventilation grids or slits.
15.
Immediately turn off and unplug any equipment that gives off smoke, sparks, strange noises, or odors.
16.
Power Supply Cord set (international) - (Detachable) Rated 10 A, 250 V. Plug type CEE 7/VII, Connector type IEC 60320/C13, and Cord type H05VV-F3G, min. 0.75 mm2, 3-conductor terminating in molded-on grounding type attachment plug. <HAR> marked on the cord. Maximum 4.0 m long.
14 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
3.
Features & Operating Condition
3.1. Brief Description
-
-
The device works like all slit lamps:
Stereomicroscopic examination of the eye under the slit light.
Microscopic examination of the ocular fundus and posterior vitreou body (using hruby’s lens).
These devices are designed for use by ophthalmologists and optometrists (within the realms of their respective professions) for specific diagnostic procedures (biomicroscopic examination of the eye)
3.2. Operating Condition
l
-
-
As long as the slit lamp is in its original package it can be exposed to the following environmental conditions for up to 15 weeks shipping and storage without risk of damage:
Relative humidity range: from 30% to75%; l
-
-
The device is made to operate under the following environmental conditions:
Relative humidity range: from 30% to 75%;
The device neither generates nor receives electromagnetic interference when used near other equipment. Therefore, no preventive or corrective measures are required
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 15
!
CAUTION
Be c ause pro lo nge d inte nse light e xposure c an damage the re tina, the use of the de vice for o c ular e xaminatio n sho uld no t be unne c e ssarily pro lo nge d, and the brightne ss se tting sho uld no t e xce e d what is ne e de d to pro vide c l visualizatio n o f the target struc ture s. This de vic e sho uld be use d with filte rs that e liminate UV radiatio n
(<4 00 nm) and, whe ne ve r po ssible , filters that e liminate sho rt wa ve le ngth blue light
(<4 20 nm).
The re tinal e xpo sure dose for a pho to c he mic al hazard is a pro duc t o f the radianc e and the e xposure time . If the value o f radianc e we re re duc in half, twic e the time wo uld be ne e de d to re ac h the maximum e xpo sure limit.
W hile no ac ute o ptic al radiatio n hazards ha ve be e n ide ntifie d for slit lamps, it is rec o mme nde d that the inte nsity o f light direc te d into patie nt’s e ye be limite d to the minimum le ve l whic h is nece ssary fo r diagno sis. Infants, aphake s and pe rso ns with dise ase d e ye s will be at gre ater risk. The risk may also incre ase d o f the pe rso n be ing e xamine d has ha d a ny e xpo sure with the same instrume nt or any o the r o phthalmic instrume nt using a visible light so urce during the pre vio us 24 ho urs. This will ap ply partic ularly if the e ye has be e n e xpose d to re tinal pho to graphy
16 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
4.
Standard Components
The instrument is delivered in package. Taking it out the package, make sure that none of the following parts is missing.
l
A table top(N.B. the table top isn’t supplied with the slit lamp for twin table) on which are placed: A SMPS box(15) including the main switch & a warning light (13), Two sliding guides for the base with orthogonal movements(17), A sliding plate for the position device(19), A drawer(16); l
A base with orthogonal movements.
l
A microscope with either dual magnification or a system (31) including eyepieces.
l
An optical group of the slit projector.
l
A head rest module (59).
l
A microscope arm (45).
l
This manual.
l
A set of accessories including; two wheels for the sliding guides(18); a setting rod(43); a protective cover(46); a spare bulb 12V 30W ogen bulb(53); two protection fuses(48); a breath screen(19); a chin-rest paper(47);a hole cover(45); a power cable; two patient’s handles(50)
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 17
5.
Accessories
The following standard accessories are provided with the instrument; l protection cover(46) l a setting rod(43) l spare bulbs
12V 30W Halogen bulb (53) l chin-rest paper(47) l spare fuses
250Vac T2.0AL
l a breath screen(5) l power cable
[ Figure 1. Accessories ]
18 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
6.
Composition
1) Fixation point
2) Control lever for changing enlargement
3) Removable eye-pieces
4) Knob for clamping the breath screen
5) Breath screen
6) Control handle for the slit rotation 90-0-90
7) Connecting plug of the projector lamp
8) Knurled ring for filters insertion
9) Joystick for lateral, longitudinal and vertical movements(x y z)
10) Base with orthogonal movements
11) Shaped tabletop
12) Warning light indicating the ignition of the SMPS is on
13) Main switch & Warning light
14) Luminosity selector
15) SMPS box
16) Drawer
17) Toothed guides
18) Wheel protection crankcases
19) Sliding plate
20) Knob to lock the projector
21) Scale for the projector position
22) Knob to lock the arm of the microscope
23) Hole cover
24) Knurled knob to lock the Lamp housing
25) Control handle for the vertical setting of the chin-rest
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 19
26) Locking pins for chin rest paper
27) Chin rest
28) Knobs to change the slit width
29) Pointer for the eye positioning
30) Forehead Rest
31) Microscope
32) Lamp socket
33) Knob to lock the base of the instrument
34) Knob to lock the microscope
35) Control handle for horizontal tilting
36) Control lever for vertical tilting
37) Microscope position stop
38) Graduated ring 90-0-90 determining the inclination of the slit during rotation
39) Lamp cover
40) fuse slots
41) Connecting plug
42) Fixation point connector
43) Setting rod
44) Microscope separation knob
45) Microscope arm
46) Protection cover
47) Chin rest papers
48) Guide screw
49) An optical group of the slit projector
50) Patient’s handle
51) Head rest fastening screw
52) Wheel
53) Halogen bulb 12V 30W
20 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
54) Bulb fastening knob
55) Mirror
56) Capture connector ( For the image system)
57) Capture switch
58) Luminosity selector connector ( Connect the Luminosity selector to the
SMPS)
59) Head rest module
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 21
[ Figure 2. Composition I ]
22 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
[ Figure 3. Composition Ⅱ]
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 23
[ Figure 4. Composition Ⅲ]
24 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
7.
Installation
NOTE
The applianc e s are always supplie d with e xc e lle nt pac kaging to be able to stand up to all kinds o f transport and standard sto rage c o nditio ns. P le ase co ntac t the supplie rs if any irre gularities are fo und in the applianc e due to transpo rt after unpac king
7.1. Assembling the Instrument
l
Put the table top on a very stable support. Block the wo parts together by tightening the screws
[ Figure 5. Table Assembling ]
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 25
l
Place the head rest module (59) and make sure that its holes fit together with those of the table top. Tighten the screw with the screw driver l
[ Figure 6. Assembling the instrument I]
Place the base with orthogonal movements on the guides situated on the upper part of the table top, checking that the wheels are in line. Then Place the crankcase (59) and make sure that its holes fit together with those of the toothed guides (17). Tighten the screw with the screw driver(+) l
Assemble the microscope arm (45) on the Base with orthogonal move ments
(10).
l
Assemble an optical group of the slit projector (49) on the Base with orthogonal
26 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
movements (10). Then tighten the Guide screw with the screw driver(-) l
Put the microscope(31) in the right position making sure that it is placed well in the stop(37); then tighten it with the knob(34)situate the right side of the microscope.
l
Fasten the breath screen(5) with the knob(4) l
Insert the connecting cable into the socket.
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 27
[ Figure 7. Assembling the instrument Ⅱ]
28 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
8.
Trial Operations
l
Assemble the chin-rest paper (47) on the chin-rest (27).
l
Let the patient sit down comfortable with his chin on the chin-rest and his forehead against the forehead-rest.
l
Lift or lower the chin-rest by moving the handle (25) so that the patient’s eyes are in line with the notches on the chin-rest(29) l
Switch the instruments on, press the switch (13). You will see the warning light on (12).
l
Adjust the luminous intensity by moving the selector(14) l
Frame and focus the eye to be examined by moving the lever(9)
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 29
9.
Main Features
Principal operative characteristics of the instrument: l
Observation without settling or tiring and rapid change of enlargements.
l
Base with cross-wise movement (which can be stopped) commanded by a single joystick for all adjustments(x y z) l
Stereomicroscopic observation of the eye with enlargements 6X-10X-16X-25X-
40X (5 enlargements model).
l l
Microscopy of the back part and of the posterior vitreous body by means of the
Hruby lens or a contact options. (Goldman lens).
l
Observation of the eye and valuation of the position of contact lenses in a fluorescent light.
l
Tonometry with Goldman tonometer.
l
Measurement of the cornea diameter and contact lenses.
l
Survey which reveals the axis of the contact lens, tha to the possible reading of the rotation angle of the slit which includes all the meridians.
30 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
10. Ordinary Maintenance
!
WARNING
The re pairs illustrate d be lo w sho uld all be c arrie d o ut with the po we r c able disco nne cte d fro m the main supply plug. P le ase co ntac t the suppliers for any o ther pro ble ms that c anno t be resolve d by the o pe ratio ns de s ibe d be lo w.
!
CAUTION
P le ase turn o ff the po we r switc h be fo re re plac ing lamps. P le ase place lamps after c oo ling do wn. The he ate d lamp may burn yo ur finge rs.
NOTE
Do n
’ t to uc h any o the r parts o f glass tube with the bare finge rs. Finge rprints, if any, must be wipe d o ff fro m the glass tube with alc o ho l
10.1. Halogen Projection bulb
To replace the Halogen projection bulb proceed as follow.
l
Before doing anything else, disconnect the plug from the main supply socket; l
Remove the screws with the driver(24), Remove the cover(39), l
Remove the Lamp socket(32) from the bulb stems; l
Bulb fastening knob (54) to loosen the bulb stop l
Remove the burnt out bulb; be careful , it could be very hot;
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 31
l
Replace it with a new one(12V30W Halogen) using the pl tic envelope(never touch the bulb whit fingers) l
Fasten the bulb with the knob l
Make sure that the stems of the bulb are completely inserted; insert the plug in the main supply socket
[ Figure 8. Replace Halogen bulb ]
!
WARNING
Do n ’ t To uc h to the Lamp Ho use , During the Ope ratio n.
It’s He ating up.
32 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
10.2. Protection fuses replacement
To replace the fuses proceed as follows: l
The protection fuses are placed on the back of the SMPS, and inserted power socket. Before doing anything else, disconnect the plu from the main supply socket l
Remove the burned-out fuses l
Substitute the fuses and check l
Replace the plug in the main supply socket.
[ Figure 9. Replace the fuse ]
!
CAUTION
Re plac e fuses with the same rating and ty pe (25 0 Vac , T2 .0 AL)
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 33
10.3. Cleaning & Disinfection
l
To keep the device clean, apply clinical alcohol and cidex or cidex plus on a soft cloth and scrub the body and the probe once a month. Do not use lacquer, thinner, ethylene, or the oxidizing substance. l
If the instrument is not in use, dust must be kept away by covering the instrument with the supplied plastic cover. Dust accumulating during the functioning of the instrument must be periodically removed with a very soft cloth or with a small pump.
34 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
11. EMC Information
11.1. Guidance and manufacturer’s declaration – electromagnetic emissions
The EUT is intended for use in the electromagnetic environment specified below.
The customer or the user of the EUT should assure that it is used in such an environment.
Immunity test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Compliance
Group 1
Class B
Electromagnetic environment guidance
The EUT uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment
The EUT is suitable for use in ail establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
Flicker emissions
Class A
Complies
IEC 61000-3-3
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 35
11.2. Guidance and manufacturer’s declaration – electromagnetic immunity
The EUT is intended for use in the electromagnetic environment specified below.
The customer or the user of the EUT should assure that it is used in such an environment.
Immunity test IEC 60601
Test level
Compliance level
Electromagnetic environment -guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±6kV Contact
±8kV air
±6kV Contact
±8kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
±2kV for power supply lines
± lines
1kV input/output for
±1kV differential mode
±2kV common mode
±2kV for power supply lines
± 1kV for input/output lines
±1kV differential mode
±2kV common mode
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions and voltage variations on power supply input lines
IEC 61000-4-11
<5% U?
(>95% dip in
U?) for 0.5cycle
40% U?
(60% dip in
U? ) for 5 cycle
70% U?
(30% dip in U?) for 25 cycle
<5% U?
(>95% dip in U?) for 0.5cycle
40% U?
(60% dip in U? ) for 5 cycle
70% U?
(30% dip in U?) for 25 cycle
<5% U?
(<95% dip in
Mains power quality should be that of a typical commercial or hospital environment. If the user of the EUT image intensifier require continued operation during power mains interruptions, it is recommended that the
36 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
<5% U?
(<95% dip in
U? ) for 5 s
U? ) for 5 s
EUT image intensifier be
Powered from an uninterruptible power supply or a battery.
Power frequency
(50/60Hz) magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should be at levels character ristic of a typical location in a typical commercial or hospital environment.
NOTE U? is the a.c. mains voltage prior to application of the test level.
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 37
11.3. Guidance and manufacturer’s declaration – electromagnetic immunity
The EUT is intended for use in the electromagnetic environment specified below.
The customer or the user of the EUT should assure that it is used in such an environment.
Immunity test
Conducted
RF
IEC
61000-4-6
IEC
60601 test level
3 Vrms
150 kHz to
80MHz
Complianc e level
3 Vrms
150 kHz to
80MHz
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no closer to any part of the EUT, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
Radiated
RF
IEC
61000-4-3
3 V/m
80
MHz to
2.5GHz
3 V/m
80MHz to
2.5GHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer
38 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as deter-mined by an electromagnetic site a survey, should be less than the compliance level in each frequency range. b
Interference may occur in the vicinity of equipment marked with the following symbol :
NOTE 1) At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2) These guidelines may not apply in all situatio Electromagnetic propagation is affected by absorption and reflection from structures, objects and a people.
Field strengths from fixed transmitters, such as base tations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the EUT is used exceeds the applicable RF compliance level above, the EUT should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the EUT.
b
Over the frequency range 150kHz to 80MHz, field strengths should be less than
[V
1
] V/m.
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 39
11.4. Recommended separation distances between portable and mobile RF communications equipment and the EUT
This is intended for use in an electromagnetic environment which radiated RF disturbances are controlled. The customer or the user the EUT can help Prevent electromagnetic interference by maintaining a minimum nce between portable and mobile RF communications equipment (transmitters) the EUT as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter
[W]
Separation distance according to frequency of transmitter [m]
150kHz to 80MHz 80MHz to
800MHz
800MHz to
2.5GHz
0.01
0.1
1
V
1
=3Vrms
0.116
0.368
1.166
E
1
=3V/m
0.1166
0.3687
1.1660
E
1
=3V/m
0.2333
0.7378
2.3333
10
100
3.687
11.660
3.6872
11.6600
7.3785
23.333
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where p is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1) At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
40 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
11.5. Immunity and Compliance Level
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601 Test
Level
3Vrms
150kHz to 80MHz
3Vrms
80MHz to 2.5GHz
Actual Immunity
Level
3Vrms
3V/m
Compliance
Level
3Vrms
3V/m
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 41
11.6. Guidance and manufacturer’s declaration – electromagnetic immunity
The EUT is intended for use in the electromagnetic environment specified below.
The customer or the user of the EUT should assure that it is used in such an electromagnetic environment.
Immunity test
Conducted
RF
IEC
61000-4-6
IEC 60601 test level
3 Vrms
150 kHz to
80MHz
Compliance level
3 Vrms
150 kHz to
80MHz
Electromagnetic environment guidance
The EUT must be used only in a shielded location with a minimum RF shielding effectiveness and, for each cable that enters the shielded location with a minimum RF shielding effectiveness and, for each cable that enters the shielded location
Radiated
RF
IEC
61000-4-3
3 V/m
80 MHz to
2.5GHz
3 V/m
80MHz to
2.5GHz
Field strengths outside the shielded location from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than 3V/m.
a
Interference may occur in the vicinity of equipment marked with the following symbol:
42 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
NOTE 1) These guidelines may not apply in all situatio Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
NOTE 2) It is essential that the actual shielding effectiveness and filter attenuation of the shielded location be verified to assure that they meet the a minimum specification.
Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength outside the shielded location in which the EUT is used exceeds 3V/m, the EUT should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as relocating the EUT or using a shielded location with a higher RF shielding effectiveness and filter attenuation.
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 43
12. Electric Diagram
44 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
13. Technical Specifications
Type
Magnification
Eyepiece
MICROSCOPE - H
IS-7000 (5x)
Total magnification
Real field of view
Interpupillary adjust
Galilean converging binocular
5 positions rotation drum
12.5x (Option 10x)
6x, 10x, 16x, 25x, 40x(12.5x eyepiece ) or 5x, 8x, 12x, 20x, 32x( 10x eyepiece )
38.5, 24.0, 15.0, 9.0, 6.0 mm
55-80 mm
SLIT ILLUMINATION
Slit length
Slit width
Slit projection
Aperture diaphragms
Filters
0.3-12 mm(continuous)
0-12 mm(continuous)
1x
0.3/1/3/5/9/12 mm cobalt blue, red-free, gray and heatabsorption
Slit rotation
Angle of incidence surface working distance
80 mm
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 45
BASE
Vertical Movement
Longitudinal Movement
Lateral Movement
Fine base Movement
Power supply
Power consumption
Instrument Voltage
Halogen Projection bulb
Fixation point LED
Min. 28mm
Min. 78mm.
Min. 98mm
Min. 10mm
100~240 V AC 50/60HZ 2.0A
70VA
12V DC
12V 30W
3.4V 20mA
46 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
14. Service Information
Ho w to co ntac t servic e : If there are any pro ble ms with the e quipme nt, ple ase follo w the ste ps be lo w: l l l
First o f all, refer to the 1 0 . M ainte nanc e and 11 . Tro uble shoo ting se ctio ns ac cording to the pro ble m that yo u are e nc o unte re d. And the n fo llo w the sugge ste d se que nce s.
If the pro ble m pe rsists, ple ase co ntac t the loc al distributo r in yo ur pro vinc e o r co untry at first.
Be fo re c alling to the loc al distributo r, yo u ’ d be tter c hec k the se informatio n suc h as
M o de l and Se rial Numbe rs. To do so , fill up the follo wing table as soo n as yo u purc hase o ur pro duc t. Yo u c an lo o k up the se informatio n at any time . The se rial numbe r is fo und o n the bac k o f this unit. The serial numbe r is unique to this unit.
Yo u sho uld re tain this manual as a pe rmane nt re co rd of yo ur purc hase . P le ase re tain yo ur purc hase rec e ipt as yo ur proo f o f purc hase .
Date o f P urchase:
Deal er
’ s Nam e:
Deal er Address:
Deal er P hone No . :
M odel No .:
Seri al No .:
l
If yo u c an
’ t c o ntac t with yo ur lo c al distributo r, yo u c an direc tly ge t in to uc h with the servic e de partme nt of the HUVITZ using the pho ne numbe r and the a ddre ss writte n in the be lo w table .
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 47
How to Co ntact HUVITZ Co ., Ltd
Wri te us at:
HUVITZ Co ., Ltd.
Hu vitz B/D, 6 8 9
-
3 Ge umjeo ng
do ng
Gunpo
si Gyeo nggi
do , So uth Ko re a
4 35 -86 2
Tel
: +82 -3 1
-
442 -8 868
F ax
: +8 2
-
31 -44 2-8 619
e
-m ail:
HIS-support@ huvitz.com
48 SLIT LAMP
HIS- 7000------------------- ----------------------------------------------------------------
15. Particular Information
15.1. Relative spectral output graph
(1) Halogen Lamp
(2) Fixat ion Lamp
------------------------------------------------------ -----------------------------SLIT LAMP
HIS-7000 49
Note: this graph be shown for the beam after it exist the instrument, when the instrument is operating at maximum light intensity and maximum aperture.
!
CAUTION
“CAUTION – The light emitted from this instrument is potentially hazardous. The longer the duration of exposure, the greater the risk of ocular damage. Exposure to light from this instrument operated at maximum output will exceed the safety guideline after
_80seconds__
(e. g. 80seconds for source halogen lamp, 151057 seconds for fixation lamp,).”
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 6 Introduction
- 6 Classification
- 6 Safety Infor mation
- 7 Introduction
- 8 Safety Symbols
- 10 Environment factors
- 12 Safety Precautions
- 14 Features & Operating Condit ion
- 14 Brief Description
- 14 Operating Condition
- 16 Standard Components
- 17 Accessories
- 18 Composit ion
- 24 Installation
- 24 Assembling the Instrument
- 28 Trial Operations
- 29 Main Features
- 30 10. Ordinary Maintenance
- 30 Halogen Projection bulb
- 32 Protection fuses replacement
- 33 Cleaning & Disinfection
- 34 EMC Infor mation
- 34 Guidance and manufacturer’s declaration – electromagnetic emissions
- 35 Guidance and manufacturer’s declaration – electromagnetic immunity
- 39 communications equipment and the EUT
- 40 Immunity and Compliance Level
- 41 Guidance and manufacturer’s declaration – electromagnetic immunity
- 43 12. Electric Diagram
- 44 13. Technical Specifications
- 46 14. Service Information
- 48 15. Particular Information