Operating Instructions CTSTestr™ 50 Pocket Testers - Cole

Operating Instructions CTSTestr™ 50 Pocket Testers - Cole
Operating Instructions
Backlight Settings
Calibration for Conductivity, TDS, or Salinity
1. P
ress MENU/v to enter setup window. Scroll down by pressing MENU/v to
select Settings.
For best results, periodic calibration with an accurate standard is recommended prior
to measurement. Use the calibration standard value that is close to your intended
sample value.The tester will retain one calibration value in each mode (conductivity,
TDS, salinity) when the instrument is powered off. The conductivity value can be
calibrated automatically or manually, while the TDS & salinity values require manual
calibration. The tester will begin in the measurement mode that was used when it was
powered off. See “Measurement Parameter Setting” to change the desired parameter.
2. Press HOLD/ to select Settings. The display shows TDS Factor and Backlight.
3. S
croll down by pressing MENU/v to toggle between TDS Factor and Backlight.
Press HOLD/ to select Backlight.
4. T
he display shows ON and OFF. Scroll down by pressing MENU/v to toggle
between ON and OFF. Backlight ON increases readability in low-light conditions.
CTSTestr™ 50 Pocket Testers
5. Press HOLD/ to select the desired backlight option. The display shows the
selected backlight option with a .
Applications
menu
Key Info
• Aquaculture
• Electroplating rinse tanks
Scroll
• Aquariums
• Printing industry
EXIT
• Aquariums and fish farms
• Swimming pools
• Boiler blow-down
•V
erification of reverse osmosis
system operation
• Drinking water
• Water & wastewater treatment
• Hydroponics
Accept
hold
hold
Measure
Settings
Temp Set
Reset
hold
TDS Factor
Backlight
Backlight
ON
OFF
OFF
Automatic Calibration for Conductivity
1. Remove the cap and press ON/OFF ( ) to power on.
2. Dip the sensor in at least 30 mm of calibration standard.
3. Stir gently and press CAL/ESC key to begin the calibration.
4. The display will show CAL followed by the default value. CAL is indicated on the
display during calibration mode.
Temperature Settings
1. P
ress MENU/v to enter setup window. Scroll down by pressing MENU/v to
select Temp Set. Press HOLD/ to select Temp Set. The display shows Set °C/°F,
Temp Cal and Temp Coeff.
2. Press HOLD/ to select Set °C/°F. Scroll down by pressing MENU/v to toggle
between °C and °F.
5. If the reading is within the calibration range of the automatically recognized
standards; 80 (84 μS/cm), 1410 (1413 μS/cm), or 12.90 (12.88 mS/cm), the icon
is displayed when the automatic calibration standard value has been detected.
6. Press HOLD/ to accept the auto conductivity standard and finish the calibration.
7. Display returns to Measurement window.
CAL
3. Press HOLD/ to select temperature unit. The display shows the selected
temperature setting with a .
COND
CAL
μS/cm
844
1413
Place the tester in
the calibration STD
to start or press
to escape.
cal
esc
μS/cm
844
23.5 °C
cal
esc
menu
Key Info
Scroll
Accept
EXIT
hold
Measure
Settings
Temp Set
Reset
hold
hold
Set °C/°F
Temp Cal
°C
°C
°F
Temp Coeff
Manual Calibration
When the conductivity reading is outside calibration range of the automatic conductivity
standards or when TDS or salinity is used, the tester will require manual adjustment.
Getting Started
Temperature Calibration
The CTSTestr 50 Pocket Tester has been factory calibrated and usually works well
out of the box. However, after extended periods of non-use, it is best to remove the
sensor cap and soak the sensor in warm tap water for 10 minutes or so. Prior to
taking the measurements, periodic callibaration with certified standards is
recommended for best accuracy.
1. P
ress MENU/v to enter setup window. Scroll down by pressing MENU/v to
select Temp Set.
2. Press HOLD/ to select Temp Set. The display shows Set °C/°F, Temp Cal and
Temp Coeff.
Note: The adjustment will decrease only, however the adjustment will eventually cycle
to the highest available value after decreasing by 40% of the initial value.
Measurement Parameter Setting
3. S
croll down by pressing MENU/v to toggle between Set °C/°F, Temp Cal and
Temp Coeff. Press HOLD/ to select Temp Cal.
3. Press HOLD/ to accept and finish the calibration when the desired value is selected.
1. Press ON/OFF ( ) to power on the tester.
2. P
ress MENU/v to enter setup window. Press HOLD/ to select Measure. The
display shows Cond, TDS and Salinity.
4. T
he lower display shows the current measured temperature reading based on the
last set offset and the upper display shows the current measured temperature
reading based on factory default calibration.
3. Scroll down by pressing MENU/v to toggle between Cond, TDS and Salinity.
Press HOLD/ to select Cond.
5. D
ip the tester into a solution of known temperature and allow time for the built-in
temperature sensor to stabilize.
4. The display shows the selected measure with a
6. P
ress MENU/v to adjust the temperature value or press the HOLD/ to confirm the
calibrated value as the new temperature value of the solution.
menu
Key Info
hold
Measure
Settings
Temp Set
Reset
Scroll
Accept
EXIT
.
hold
Note: To exit this program without confirming the calibration, press CAL/ESC.
Cond
Cond
TDS
Salinity
menu
Key Info
Scroll
TDS Factor Setting
Accept
EXIT
1. Press MENU/v to enter setup window. Scroll down by pressing MENU/v to
select Settings.
2. Press HOLD/ to select Settings. The display shows TDS factor and
Backlight.
3. P
ress HOLD/ to select the default TDS factory setting or MENU/v to adjust
the setting.
4. P
ress HOLD/ to confirm the selection of the setting. The display shows the
set TDS factor with a .
hold
hold
Measure
Settings
Temp Set
Reset
Set °C/°F
hold
Temp Cal
23.4 °C
23.4 °C
Temp Cal
Temp Coeff
23.4°C
Temp Coefficient
1. P
ress MENU/v to enter setup window. Scroll down by pressing MENU/v to
select Temp Set.
2. Press HOLD/ to select Temp Set. The display shows Set °C/°F, Temp Cal and
Temp Coeff.
3. S
croll down by pressing MENU/v to toggle between Set °C/°F, Temp Cal and
Temp Coeff.
menu
Scroll
Accept
EXIT
Measure
Settings
Temp Set
Reset
hold
hold
hold
TDS Factor
Backlight
Adjust
TDS Factor
0.71
0.66
Note: To abort calibration, press CAL/ESC to escape.
4. Once the calibration is finished and user has accepted the changes, measurement
window will now show the calibrated reading.
Note: The auto conductivity standards are 84 μS/cm, 1413 μS/cm & 12.88 mS/cm.
CAL
cal
esc
CAL
CAL
Place the tester in
the calibration STD to
start or press
to
escape.
cal
esc
μS/cm
844
1413
μS/cm 844
820
COND
μS/cm
844
24.8 °C
CTS Measurement
1. Press ON/OFF ( ) to power the tester on if needed.
CAUTION: Testing dry samples is not accurate and can lead to sensor
damage or breakage. Soils must be wet and free of particulates that may
scratch the glass sensor. Excessive force into dry samples can cause glass
breakage.
COND
µs/cm
820
24.8 °C
3. Note the CTS value or press HOLD/ to freeze the reading. To release the
reading, press HOLD/ again.
4. Press ON/OFF ( ) for 3 seconds to turn off tester. If key is not pressed for 8.5 minutes,
the tester will automatically shut off to conserve batteries.
5. P
ress HOLD/ to confirm the Temp Coeff value. The new value is automatically
confirmed with a .
menu
Key Info
Scroll
Accept
EXIT
1065O100_MAN_35634-55,-65.indd 1
2. Press MENU/v to manually adjust the value to the desired reading.
2. Dip the electrode about 2 cm to 3 cm into the test solution. Stir and let the reading
stabilize. The timer icon will blink during this time. Once the reading is stabilized, the
timer stops blinking and will appear to indicate the stability of the reading.
4. Press HOLD/ to select Temp Coeff or MENU/v to adjust the Temp Coeff.
Key Info
1. Repeat steps 1 to 4 from “Automatic Calibration for Conductivity”.
hold
hold
Measure
Settings
Temp Set
Reset
Set °C/°F
Temp Cal
Temp Coeff
hold
Adjust
Temp Coeff
2.00
2.00
8/25/2017 7:27:05 AM
User Reset
Sensor Replacement
Reset the CTS calibration to the user’s default settings by using the user reset function.
Temperature user calibration is not affected by the user reset function.
Replace the sensor module at the fraction of the cost of a new tester. When the
tester fails to calibrate or gives fluctuating readings in calibration standards, you
need to change the electrode.
1. P
ress MENU/v to enter setup window. Scroll down by pressing MENU/v to select Reset.
Press HOLD/ to select Reset. The display shows User Reset and Factory Reset.
2. Press HOLD/ to select User Reset.
3. T
he display automatically shows No and Yes. Scroll down by pressing MENU/v
to toggle between No and Yes.
4. Press HOLD/ to confirm either No or Yes. Display shows User Reset option with a .
Key Info
menu
Scroll
Accept
EXIT
hold
hold
Measure
Settings
Temp Set
Reset
User Reset
No
Factory Reset
Yes
Self-Diagnostic Messages
1. W
ith dry hands, grip the ring with sensor facing you. Twist the ring counterclockwise. Save the ring for later use.
2. Pull the old sensor module away from the tester.
3. Align the four tabs on the new module so that they match the four slots on
the tester.
4. Gently push the module onto the slots to sit it in position. Push the smaller
O-ring fully onto the new sensor module. Push the other O-ring over the
module and thread it into place by firmly twisting clockwise.
Note: It is necessary that you recalibrate your tester prior to measurement after
a sensor replacement.
Factory Reset
Reset the CTSTestr to the factory default settings by using the factory reset function.
1. Press MENU/v to enter setup window. Scroll down by pressing the MENU/v
to select Reset. Press HOLD/ to select Reset. The display shows User Reset and
Factory Reset.
2. Scroll down by pressing MENU/v to toggle between the resets. Press HOLD/
to select Factory Reset.
3. The display automatically shows No and Yes. Scroll down by pressing MENU/v
to toggle between No and Yes.
4. Press HOLD/ to confirm either No or Yes. The display shows the Factory Reset
option with a .
Key Info
menu
Scroll
Accept
EXIT
hold
Measure
Settings
Temp Set
Reset
hold
Factory Reset
Yes
This feature helps to freeze the display for a delayed observation.
COND
µs/cm
820
23.5 °C
820
23.5 °C
Press
release
2. Press HOLD/ again to release measurement.
COND
µs/cm
µs/cm
to
820
23.5 °C
Press
Appears when calibration is attempted but the reading is not yet stable. Wait for
the reading to stabilize or manually confirm the calibration by pressing enter.
over range
The reading is above the measuring range of tester.
Specifications
Conductivity
Conductivity range
Resolution
Relative accuracy
Normalization temperature
Temperature co-efficient
Calibration points
TDS
TDS range
Resolution
Relative accuracy
Calibration points
TDS factor
Salinity
Salinity range
Resolution
Relative accuracy
Calibration points
Temperature
Temperature range
Temperature resolution
Temperature accuracy
No
HOLD Function
COND
stable error
hold
User Reset
Replacing the Batteries
1. Press HOLD/ button to freeze measurement.
Batteries are weak and need replacement soon.
COND
hold
µs/cm
820
23.5 °C
to
release
Sensor Maintenance
1. Always keep the sensors electrodes clean. Rinse the electrodes with
deionized water and wipe them dry with clean cloth before storing with its
protective cap. For cup type electrodes, remove the white plastic cup and
insert to thoroughly clean viscous solutions. Never scratch electrodes with
a hard substance.
2. For better performance, soak the electrode in alcohol for 10 to 15 minutes and
rinse with de-ionized water before starting any measurement process. This is
to remove dirt and oil stains on the electrode which may affect the accuracy
of the measurements.
The CTSTestr 50 Pocket Tester uses four
AAA 1.5 V batteries.
1. To remove the battery cover, see Figure 1.
Clear the front catch and then the back
catch, before sliding the cover off.
Fig. 1: Removing battery cover
Fig. 2: Push to unlock
2. T
o remove the battery plate, push the
center tab towards the front of the tester
as shown in Figure 2. Once unlocked,
remove the plate to access the batteries.
3. Invert the tester upside down to remove
the batteries. Each side uses two
AAA batteries. Orient each battery with
positive terminal facing downward.
Fig. 3: Align tabs
Fig. 4: Push down to lock
4. T
o lock the battery plate, align the small
tabs (Figure 3) into the guide ribs on
the housing and then press down.
See Figure 4.
Fig. 5: Battery plate unlocked
Fig. 6: Battery plate locked
Warranty
This instrument is supplied with a warranty against manufacturing defects for
a period of one year from the date of purchase.
CTSTestr 50
0.0 to 200.0 μS, 200 to 2000 μS, 2.00 to 20.00 mS
0.1 μS, 1 μS, 0.01 mS
±1% full scale
25.0ºC (77ºF)
0.0% to 10.0%
Up to 3 points
0.0 to 100.0 ppm, 100 to 1000 ppm, 0.10 to 10.00 ppt (TDS Factor 0.5)
0.1 ppm, 1 ppm, 0.01 ppt
±1% full scale
Up to 3 points
0.40 to 1.00 (selectable)
0.00 to 10.00 ppt
0.10 ppt
±1% full scale
1
0 to 60°C (32.0 to 140.0°F)
0.1°C / 0.1°F
From 0 to 50ºC (±0.5ºC / ±0.9ºF + 1 LSD);
from 50 to 60ºC (±1.0ºC / ±1.8ºF + 1 LSD)
Yes (Automatic Temperature Compensation)
Temperature compensation
General
Display
Graphics, dot matrix 80 x 100 pixel
Auto off
8.5 minutes (from last key press)
Reset
User / Factory
Power requirement
Four AAA 1.5 V batteries
Battey life
>150 hours
Water proofing
IP67
Environmental operating conditions
Ambient operating
5 to 45°C / 41 to 113°F
temperature
Relative humidity
5% to 85% noncondensing
Storage temperature
–20 to 60°C / –4 to 140°F
Storage humidity
5% to 85% noncondensing
Accessories
Ordering Code
35634-55
35634-65
35634-57
35634-07
35634-09
09376-00
17101-45
Product Description
CTSTestr 50P pin-style pocket tester with case, lanyard, and batteries
CTSTestr 50C cup-style pocket tester with case, lanyard, and batteries
Replacement sensor module for CTSTestr 50P pin-style tester
Replacement sensor module for CTSTestr 50C cup-style tester
Replacement sensor cap
Replacement alkaline batteries; AAA, 1.5 V. Pack of 12
NIST-traceable calibration with data for pocket testers
Return of Items
Authorization must be obtained from your distributor before returning items
for any reason. When applying for authorization, please include information
regarding the reason the item(s) are to be returned.
We reserve the right to make improvements in design, construction and
appearance of products without notice. Prices are subject to change without
notice.
www.4oakton.com
1065O100_MAN_35634-55,-65
1065O100_MAN_35634-55,-65.indd 2
August, 2017
8/25/2017 7:27:14 AM