Digital Test Gauge
Operating
Instructions
© Ashcroft Inc. 2005 I&M008-10080-12/11 GBE Dwg. 83B234-01 Rev. C
– 2 –
Congratulations on your purchase of the Ashcroft ® digital test gauge with total error band full-scale accuracy and the largest display readout in the industry of .66˝ high. Other industry-leading features include twelve selectable engineering units, seven languages, and password-protected disable and calibration functions. With the range printed on the keypad, the Ashcroft digital gauge meets the ASME B40.7 digital gauge specification.
See a complete listing of product features and specifications on pages 14 & 15.
– 3 –
– 4 –
TABLE OF CONTENTS
Quick Reference Guide
Keypad Functions
• ON/OFF KEY
• BACKLITE KEY
• MIN/MAX KEY
• ZERO/CLEAR KEY
• ENTER KEY
• CONFIG KEY
Configurable Functions (CONFIG Mode)
• Units (Engineering)
• Update Rate
• Auto Off
• Backlite
• Languages
• Damping
• Contrast
• Calibrate (Gauge Calibration)
• Default Password
• Zero Calibration
• Zero Span
• Zero Disable
• Access Keypad
• Disable
Available Ranges
Specifications
Installation & Battery Replacement
• Mounting
• Battery Replacement & Installation
Page
6-7
8
12
13
13
14
15
15
9
9-10
10
10
10
11
11
11
16
17-18
19
– 5 –
QUICK REFE
.66˝ high display, 5 digit
Bar graph
% of full scale
Flashing display when unit pressured below zero
Press to indicate minimum or maximum pressure gauge has measured
Press again to return to
pressure units
While in max or min mode, press to clear minimum or maximum displayed values
UNITS (Pressure)
ENGLISH
PSI
INHG
INH2O with Temperature
(Options: 60°F,
4°C, 20°C)
FTSW
METRIC
BAR
MBAR
KPA
MPA mmHG
CMH2O mmH2O
KG/CM2
BACKLITE
(Off options)
ON/OFF*
10 SEC
30 SEC
1 MIN
5 MIN
– 6 –
UPDATE RATE
(Pressure measurement per second)
10x*
5x
2x
1x
DAMPING
(averages gauge reading
None*
AVG 2
AVG 4
AVG 6
AVG 8
LANGUAGE(s)
English*
French
Spanish
German
Italian
Portuguese
Dutch
ERENCE
Press to turn unit on or off
Press to turn backlite on or off
Flashing display when unit
pressured beyond full-scale
This bar graph indicates battery level; the more segments, the closer the battery is to full charge
Range on keypad; complies with ASME B40.7
AUTO OFF
(Turns unit off after option selected)
Never*
2 minutes
5 minutes
15 minutes
30 minutes
While in unit of measurement mode (eg: psi), press the ZERO CLEAR button to rezero the gauge. This feature functions when displayed pressure is within ±5% of zero value
CALIBRATE
Zero and span adjustments, password protected
CONTRAST
(Customizes display readout)
7 available
Selection 4 is
Default*
DISABLE
Allows for
“lockout” of
CONFIG options
*Indicates Default
– 7 –
KEYPAD FUNCTIONS
ON/OFF
BACK
LITE
MAX
MIN
ZERO
CLR
Turns the gauge on and off. When pressing the
ON/OFF key while in the off position, gauge startup display first indicates the software version
followed by the model number and gauge pressure range. The gauge will then display indicated pressure and be ready for use.
Manually turns backlite on and off. (See
CONFIG
mode for options).
Allows review of minimum and maximum pressure values since unit start-up or last push of the clear button. Press key to:
1) Indicate maximum pressure.
2) Indicate minimum pressure.
3) Exit MAX/MIN mode and return to unit of pressure measurement mode. To clear minimum and maximum values press ZERO/CLR button.
Must be in MAX/MIN mode.
Note:
MIN/MAX data is lost when unit is turned off.
Press this key prior to gauge usage to rezero any initial offset less than ±5% of the rated gauge range. If indicated pressure is greater than 5% of range, the rezero feature becomes inoperable. This prevents accidental tare of a pressurized gauge.
To clear minimum and maximum values, press
ZERO/CLR button (when min/max values are indicated.
ENTER
Used in conjunction with CONFIG key, see next page.
CONFIG
This key allows for customization of the gauge.
Pressing the CONFIG key allows cycling through the main menu items; UNITS, UPDATE, AUTO OFF,
BACKLITE, LANGUAGE, DAMPING, CONTRAST &
CALIBRAT.
– 8 –
KEYPAD FUNCTIONS
UNITS:
12 units of measurement are available, both English and metric, by cycling through the
UNITS key; psi, ˝Hg, ˝H
2
O (with three temperature options, 60°F, 4°C and 20°C), ftSw, Bar, mBar, kPa, mPa, mmHg, cmH
2
O, mmH
2
O, kg/cm 2 .
Step 1:
Press the CONFIG key until the word
UNITS appears.
Step 2:
Press the ENTER key.
Step 3:
Press the CONFIG key once to select
ENGLISH or again to select METRIC.
Step 4:
Press the ENTER key with selection of
ENGLISH or METRIC.
Step 5:
Press CONFIG key to select unit of
measurement.
Step 6:
Press ENTER key to finalize unit of
measurement.
UPDATE:
this option allows for changing the rate at which pressure is updated on the display screen. The default rate measures pressure at the maximum rate of
10
* updates per second or 100 milli-seconds. Optional rates of measurement are measured in updates per second. The options are
10
*
,
5, 2 or 1 update of pressure measurement per second.
Since customer processes vary, update rates should be selected based on the application.
To use the UPDATE option:
Step 1:
Press the CONFIG key until the word
UPDATE appears.
Step 2:
Press the ENTER key.
Step 3:
Press the CONFIG key to select the desired update rate.
Step 4:
Press ENTER key to finalize
UPDATE
rate.
AUTO OFF:
this option sets the amount of time before the gauge will turn itself off after no activity.
Offerings are
Never*,
2, 5, 15, or 30 minutes.
– 9 –
*Indicates default.
KEYPAD FUNCTIONS
To use the AUTO OFF option:
Step 1:
Press the CONFIG key until the word
AUTO OFF appears.
Step 2:
Press the ENTER key.
Step 3:
Press the CONFIG key to select the desired AUTO OFF rate.
Step 4
Press the ENTER key to finalize the AUTO
OFF rate.
BACKLITE:
5 options are available. They include
ON/OFF*,
10 seconds, 30 seconds 1 or 5 minutes. With the ON option pressed, the gauge backlite will remain lit whenever the gauge is in the ON mode or until the backlite button is pushed again.
Options allow the backlite to automatically turn-off after a selected period of time.
Note:
leaving backlite button on will decrease battery life.
To use the BACKLITE option:
Step 1:
Press the CONFIG key until the word
BACKLITE appears.
Step 2:
Press the ENTER key.
Step 3:
Press the CONFIG key to select one of the available time options.
Step 4:
Press the ENTER key to finalize your choice of BACKLITE options.
LANGUAGE:
available in seven different languages, this option allows the user to change the default language in the CONFIG mode. The languages include
English*,
French, Spanish, German,
Italian, Portuguese and Dutch.
Step 1:
Press the CONFIG key until the word
LANGUAGE appears.
Step 2:
Press the Enter key.
Step 3:
Press the CONFIG key to select one of the available LANGUAGE options.
Step 4:
Press the ENTER key to finalize your
LANGUAGE option
– 10 –
*Indicates default.
KEYPAD FUNCTIONS
DAMPING:
with six different options, this mode allows for taking process pressure readings and averaging them. This option is particularly useful when there is pulsation in the process. The options are
NONE*,
AVG 2, 4, 6 or 8.
Step 1:
Press the CONFIG key until the word
DAMPING appears.
Step 2:
Press the ENTER key.
Step 3:
Press the CONFIG key to select one of the available DAMPING options.
Step 4:
Press the ENTER key to finalize your
DAMPING option.
CONTRAST:
this mode allows for BACKLITE contrast level. Seven options are available,
1, 2, 3,
4*,
5, 6 and 7.
Step 1:
Press the CONFIG key until the word
CONTRAST appears.
Step 2:
Press the ENTER key.
Step 3:
Press the CONFIG key to select one of the available CONTRAST options.
Step 4:
Press the ENTER key to finalize your
CONTRAST selection.
Note:
setting high contrast levels will decrease battery life.
CALIBRAT.:
Gauge Calibration:
Both zero and span adjustments are available. This gauge has been configured with a default password of ØØØØØ. This factory password does not allow access to calibration. To access the calibration mode, it is necessary to configure a
user password.
Once the user password is configured, it will become the default password that allows access to gauge calibration
.
– 11 –
*Indicates default.
KEYPAD FUNCTIONS
To access the factory default password:
Step 1:
Press the CONFIG key until the word
CALIBRAT appears.
Step 2:
Press the ENTER key.
Step 3:
The letters/asterisks… PW***** appear.
Step 4:
Press the CONFIG key. An Ø appears in the first position.
Step 5:
Press the ENTER key once.
Step 6:
Press the CONFIG key until Ø appears. Ø will appear in the second position.
Step 7:
Press ENTER.
Step 8:
Use this format until all the asterisks are replaced with Ø.
There now should be a total of five Ø’s on the
keyboard display. The zero in the fifth position should be blinking.
Step 9:
Press the ENTER key. You are now prompted to SET PW (or set password).
Step 10:
Press the ENTER key.
Step 11:
Decide on a five number user password, then follow the procedure above inserting a number in the flashing display until all five numbers are inserted.
Step 12:
A SAVE prompt will then appear. If the selected user password is acceptable, press ENTER. If the selected user password is not acceptable press ZERO
CLEAR to refigure the user password.
After the password is configured, the default factory password will be replaced with the user password. Once configured, the factory password is no longer accessible.
If an incorrect password is entered, the system will display WRONG. Press the CONFIG key to reenter the correct password.
Step 13:
Press ENTER again to begin calibration.
Note:
Calibration feature allows recalibration of zero and span.
– 12 –
*Indicates default.
KEYPAD FUNCTIONS
Zero Calibration:
Step 14:
Press the CONFIG key once and the word
CALIBRA appears. Press ENTER. (This mode allows for 0 and full- scale adjustment of span.) The gauge will now display 0.00. Ensure the gauge is not pressurized,
1
then press ENTER to zero the gauge. Zero calibration is now complete.
Full Scale Calibration
Step 15:
The gauge will now display full-scale range (e.g. 100.00 psi). Pressurize the gauge to 100% of the range (which is equal to the displayed value) utilizing a pressure standard with accuracy four times greater than the unit being calibrated.
2
Press ENTER. Full-scale calibration is now complete.
Notes:
1.
If the digital gauge under test is not pressurized while in span adjustment of full-scale range, an
ERROR message will be displayed when the
ENTER button is pressed. If this occurs, press the ZERO CLEAR button on the keypad to return to the previous screen.
2.
ASME B40.7-1998, section 6.1.1.1 recommends the working standard for the gauge being tested is 4X better than the digital gauge under test.
This means the pressure standard measuring the full-scale pressure being applied to the gauge should have an accuracy four times greater than the unit being spanned.
Zero SP (span):
This feature allows setting
the % of span in which the zero button will operate. Span is limited to prevent accidental tare of process pressures.
Options are
5%,
10% or DISAB
(5% is the factory default and means the unit can be rezeroed between ±5% of span). If DISAB is selected, the zero button is deactivated and no display change will occur when the zero button is pushed.
– 13 –
KEYPAD FUNCTIONS
Step 1:
Press the CONFIG key until the word ZERO
SP appears.
Step 2:
Press ENTER.
Step 3:
Enter user five digit password (PW). This is the same password established to access the CONFIG mode in the menu.
Step 4:
Press the CONFIG key to select the desired option.
Step 5:
Press ENTER to finalize the selection.
Notes:
Selecting the DISAB feature does not disable the
CLEAR button on the keypad for the MAX/MIN feature. If the DISAB feature is selected, pressing the
ZERO button on the keypad will cause the display to read DISAB for two seconds. The gauge will then revert back to the unit of measure of the gauge. The
DISAB feature disables the zero feature of the gauge.
Zero Disable Feature
:
This feature allows disabling the Zero/Clear button on the keypad. It also allows for a zero tolerance of either 5% (default) or 10% of the gauge range.
Step 1:
Press the CONFIG key until the word ZERO
SP appears.
Step 2:
Press ENTER.
Step 3:
A prompt appears to enter PW (enter password). The ZERO SP password is the same password as discussed on page 10 and the heading CALIBRAT:, Gauge
Calibration. Follow the instructions on page 10 to enter a password.
Step 4:
Press the CONFIG key to select the zero tolerance, either 5% or 10% of range, or press the CONFIG key again and the word
DISAB appears. Press ENTER to select the new default setting.
– 14 –
KEYPAD FUNCTIONS
If the user password is lost or stolen, contact
Ashcroft Inc., customer service at (203) 378-
8281 for a new factory password that will allow the user to establish a new user password.
DISABLE:
allows “lockout” of
individual
CONFIG options. The default is ENABLE for all options in the CONFIG mode.
Step 1:
Press the CONFIG key until the word
DISABLE appears.
Step 2:
Press the ENTER key.
Step 3:
Insert the
user
password following the procedure as described in steps 3 through
13 above.
Note:
This is the same user password as in the
CALIBRAT mode.
Step 4:
Press the ENTER key.
Step 5:
The first option in the CONFIG menu will now be displayed (UNITS).
Step 6:
To DISABLE the UNITS option press the
ENTER key until the word DISABLE appears.
Step 7:
Press the CONFIG key. The UNITS option is now DISABLED.
Step 8:
Proceed through the balance of the
CONFIG menu options by pressing the
CONFIG key. Follow steps 6-8.
– 15 –
DIGITAL TEST GAUGE RANGES
600
800
1000
1500
2000
2500
3000
5000
7000
psi psi psi
(Gauge) (Compound) (Absolute)
vac.
–15/+15
5 –15/+30
15
25
50
100
160
200
300
500
10
15
30
60
bar/kb/cm 2 bar
(Gauge) (Compound)
1
1.6
–1/0
–1/1
–1/2
–1/3
–1/30
16
25
40
60
160
6
10
2.5
4
250
400
500
mmH
2
O
(Gauge)
3000
5000
10,000
mPa
(Gauge)
6
10
40
1
1.6
2.5
mBar/cmH
2
O
(Gauge) kPa
(Gauge)
250
300
400
500
600
1000
1600
2000
400
600
2500 1000
4000
5000
6000
10,000
25
40
60
100
160
250
– 16 –
SPECIFICATIONS
Type
2089 (0.05% accuracy), 2086 (0.10% accuracy),
2084 (0.25% accuracy)
Accuracy
0.05%, 0.10%, 0.25% all Full Scale Terminal Point Total
Error Band (TEB) Accuracy Including Hysteresis,
Linearity, Repeatablilty & Temperature (–18/65°C)
(0/150°F)
Dial Size
3"
Case Material
300 Series SS
Case Finish
Electropolished
Case Enclosure Rating
Weatherproof, IP65
Socket Material
316 SS
Socket Size
1 ⁄
4
NPT Male (others on application)
Connection Location
Lower, 3:00, 9:00
Ranges
Vac thru 7000 psi (see engineering units below for other units of measurement)
Operating Temperature
0/150°F
Storage Temperature
–40/180°F
DISPLAY:
Type
LCD
Display Digits
5
Character Height
.66˝
Backlit
Off By Default
Bar Graph
Yes
Battery Life
1000 Hrs.
Agency Approvals
CE, FM, CSA
KEYPAD FUNCTIONS:
On/Off
Manually Turns Unit On & Off
(auto off options in config menu)
Backlit
Manually Turns Backlit On & Off
(auto off options in config menu)
Min/Max
Stores Min & Max Values
Zero/Clear
Zeros Display or Clears Min/Max Values When
Displayed
– 17 –
SPECIFICATIONS
Enter
Selects Items In CONFIG Menu
Config Mode
Allows Scrolling Through CONFIG Menus
Engineering Units
Psi, ˝Hg, ˝H
2
O (with three temperature options, 60°F, 4°C and 20°C), ft.SW, bar, mbar, kPa, mPa, mmHg, cmH
2
O, mmH
2
O, kg/cm 2
Update Rate
4 Options: 10x/sec, 5x/sec, 2x/sec, 1x/sec
Auto Off
6 Options: Never, 2 Min., 5 Min., 15 Min., 30 Min.,
Dampening
6 Options: None, Average 2, 4, 6, 8 x update rate
Languages
7 Languages: English, Spanish, French, Italian, German,
Portuguese, Dutch
Backlite
5 Options: On/Off, 10 Sec., 30 Sec., 5 Min., 15 Min.
Field Recalibration
Zero & Span (password protected)
Contrast
7 Available Options
Disable Config Options
Allows disabling of Config Options (password protected)
Calibration Chart
10 Point Individual NIST Traceable Calibration Chart,
Standard
Standard Accessories
300 Series SS Protective Cover
Nylon Protective Carrying Pouch
WARNING AND ERROR MESSAGES
Display
Flashing
0% or 100%
LOW BAT
ERROR
RES ERROR
Description
Gauge over/underpressured beyond 105% of range
Low battery, replace
Internal error, call customer service (203) 378-8281
Pressure unit conversion exceeds display resolution or gauge pressured beyond resolution
– 18 –
Gauge Installation:
Pipe Mount –
The Ashcroft digital test gauge comes standard with a
1
⁄
4
NPT connection. Good piping practices recommend using teflon tape or a pipe sealant on the gauge threads. Utilize a
7
⁄
16
˝ wrench on the wrench flat of the gauge to tighten the gauge to the process.
NEVER TIGHTEN GAUGE THREADS BY HOLDING THE BODY OF THE GAUGE.
DOING SO MAY DAMAGE THE GAUGE AND MAKE THE GAUGE INOPERABLE.
Panel Mount –
The lower connected Ashcroft digital test gauge is available with an optional flange for panel mounting. Please refer to illustration and dimensions below.
.
1 7
HOLE
CUTOUT
DIAMETER
3.406
(3 13 ⁄
32
± 1 ⁄
32
)
1 .
86 0
4.00
Battery Installation and Replacement:
The gauge comes standard with a quantity of three AAA alkaline batteries
(installed). Use either Duracell MN2400, MX2400 or Energizer E92BP,
X92RP AAA alkaline, non-rechargeable batteries.
Batteries have a life of approximately 1000 hours. Battery life is dependent on gauge usage, backlite settings and power off settings. When the display flashes LOW BAT, batteries should be replaced.
To replace the batteries:
1) Remove the single screw on the back of the gauge case.
2) Hold the keypad in the palm of hand.
3) Carefully remove the three batteries from the holder and replace the batteries. Use only AAA alkaline non-rechargeable batteries.
!
Pipe to which gauge is attached must be properly grounded.
– 19 –
Ashcroft Inc.
250 Main Street
Stratford, CT 06614-5145
Tel: (203) 378-8281
Fax: (203) 385-0602 e-mail: [email protected]
www.ashcroft.com
© Ashcroft Inc. 2005 I&M008-10080-12/11 GBE Dwg. 83B234-01 Rev. C