SMLP4-4 User Guide

SMLP4-4 User Guide
OLS Series Light Sources,
OPM Series Optical Power Meters,
and Related Test Kits
User’s Guide
Limited Warranty
One Year Limited Warranty
All Noyes products are warranted against defective material and workmanship for a period of one year
from the date of shipment to the original customer.
Any product found to be defective within the warranty period will be repaired or replaced by Noyes.
In no case will Noyes liabilities exceed the original purchase price of the product.
Exclusions
The warranty on your equipment shall not apply to defects resulting from the following:
•• Unauthorized repair or modification
•• Misuse, negligence, or accident
CE Information
These instruments have been designed and tested to comply with the relevant sections of
any applicable specifications including full compliance with all essential requirements of
all applicable EU Directives.
Returning Equipment
To return equipment, please contact Noyes to obtain additional information and a Service Request (S.R.)
number. To allow us to serve you more efficiently, please include a brief description specifying the reasons
for the return of the equipment.
AFL Telecommunications
Noyes Test & Inspection
16 Eastgate Park Road
Belmont, NH 03220
Tel:
800-321-5298
603-528-7780
Fax: 603-528-2025
2
© 2010, AFL Telecommunications, all rights reserved. OLTK-00-1000 Rev. 1C, 2010-06-29
Specifications are subject to change without notice.
Table of Contents
Safety Information..................................................................................4
General Information................................................................................5
OPM-Series Optical Power Meters..........................................................6
OPM4 Models............................................................................6
Ports and Display..................................................................6
Dual Function Keys................................................................7
Test Mode: To Set Reference.................................................8
Test Mode: To View Reference...............................................8
Display Readings - Splash and Auto Off Screen......................8
Display Readings in Test Mode...............................................9
OPM5 Models............................................................................10
New Feature.........................................................................10
Ports and Display..................................................................10
Dual Function Keys................................................................11
Display Readings - Splash and Auto Off Screen......................12
Display Readings in Test Mode...............................................13
File Manager: To Select Current File.......................................14
File Manager: To Select Current Fiber.....................................14
Test Mode: To Set Reference.................................................14
Test Mode: To View Reference...............................................15
Recall Mode: To View Stored Records....................................16
To Edit Stored Records..........................................................17
Clear Mode: To Erase Stored File...........................................17
Clear Mode: To Erase Stored Fiber.........................................17
OLS-Series Light Sources .....................................................................18
OLS1-Dual LED Light Source......................................................18
OLS4 Integrated LED and Laser Light Source..............................23
OLS7-3 Triple Wavelength Laser Source......................................25
OLS7-FTTH Triple Wavelength Laser Source...............................27
Measuring Optical Power........................................................................29
Testing Multimode Links.........................................................................30
Testing Single-mode Links......................................................................33
Maintenance Tips..................................................................................36
Cleaning Optical Ports................................................................36
Repair and Calibration................................................................37
Battery Replacement..................................................................37
3
Safety Information
!
WARNING! Use of controls or adjustments other than those specified herein may result in
hazardous radiation exposure.
The OLS1 optical light source is a CLASS I LED PRODUCT.
The OLS2 optical light source is a registered CLASS I LASER PRODUCT.
The OLS4 optical light source is a registered CLASS I LED and LASER PRODUCT.
The OLS7 optical light source is a registered CLASS I LASER PRODUCT.
!
WARNING! To avoid the danger of fire and electrical shock:
• Never use a voltage that is different from that for which the AC adapter is rated.
• Do not plug the unit into a power outlet that is shared by other devices.
• Never modify the power cord or excessively bend, twist, or pull it.
• Do not allow the power cord to become damaged. Do not place heavy objects on the power
cord or expose it to heat.
• Never touch the AC adapter while your hands are wet.
• Should the power cord become seriously damaged (internal wiring exposed or shorted),
contact the manufacturer to request servicing.
!
WARNING! Use only the specified AC adapter. Use of another type of AC adapter can damage
the instrument and create the danger of fire and electrical shock.
!
CAUTION! To avoid serious eye injury, never look directly into the optical outputs of fiber optic
network equipment, test equipment, patch cords, or test jumpers. Always assume that optical
outputs are on.
!
NOTICE! Noyes power meters and light sources contain no user serviceable parts. Except
for changing batteries and cleaning optical ports, these units must be returned to Noyes or
authorized agents for repair and calibration.
IMPORTANT! Proper care in handling should be taken when using any precision optical test
equipment. Scratched or contaminated optical connectors can impact the performance of the
instrument. It is important to keep the dust caps in place when the unit is not being used.
4
General Information
Introduction
The purpose of this User’s Guide is to explain how to use and maintain Noyes test equipment. Please
check our web site at www.AFLtele.com/go/Noyes for updates to this manual, software updates, and
additional application information. If you have any questions about your instruments and recommended
accessories, or if you need technical or sales support, please contact Noyes Customer Service.
Contacting Noyes Customer Service
You may call Noyes Customer Service between 8 a.m. and 5 p.m., United States Eastern Time, as follows:
Phone: 800-321-5298 • 603-528-7780
Fax:
603-528-2025
Email: [email protected]
Unpacking and Inspection
These instruments have been carefully packed in accordance with standard shipping procedures. Examine
the equipment for damage that may have occurred during shipment. If you find any damage, please
contact Noyes.
Recommended Accessories
You will need fiber optic test jumpers to connect instruments to the fiber optic system under test. A test
jumper must have the same core and cladding size as the fiber under test. The connector at one end of
the test jumper must mate with the optical port on each instrument. The connector on the other end must
mate with the fiber optic system under test.
A Connector adapter is required to mate fiber optic test jumpers.
An OPM optical port must be equipped with an adapter cap. A variety of adapter caps, connector
adapters, and test jumpers with a variety of lengths and connector styles are available from AFL - Noyes.
Optical ports and connector end faces must be kept free from dirt or other contaminates to ensure
accurate measurements and operation.
For cleaning connector end faces on OLS light sources, test jumpers, and in fiber frames or adapters, use
optical quality cleaning fluid such as AFL FCC2 connector cleaning fluid and AFL CCT molded cleaning
tips.
For cleaning OPM optical ports and adapter caps, use lint-free optical cleaning wipes such as AFL
FiberWipes and optical quality cleaning fluid such as AFL FCC2 connector cleaning fluid (or IPA - Reagent
Grade Isopropyl Alcohol 99% or better) and a can of filtered compressed air.
Visit our web at www.AFLtele.com/go/Clean for more information.
5
OPM-Series Optical Power Meters
OPM4 Models
2 Adapter cap
1 Optical input adapter cap mount
AFL Telecommunications
Optical Power Meter
0000
WAVE ID
µW
dB/dBm/µW key
dB
dBm
0.00
0000
0.00
Hz
nm
µW
dB
Hz
nm
3 Display
µW
dB
Set
Ref
λ
Wavelength/Backlight
key
Set Reference key
Power key
OPM4
Ports and Display
Ref # Description
6
1
Optical input (adapter cap mount) - accepts Noyes thread-on adapter caps.
2
Adapter cap - the OPM4 optical input must be equipped with an adapter cap.
3
Display - depending on the selected mode, shows measured power dBm or µW or insertion
loss dB, calibrated wavelengths, tone signal Hz, wavelength ID, references, and battery charge
status.
OPM4 Models
Dual Function Keys
YY
Secondary function of a key
Press and hold to activate
XX
Primary function of a key
Press and release to activate
Power key
Provides two functions:
•• Press the key to turn the OPM4 on/off. The unit will turn off automatically five minutes
after the last key press.
•• Press and hold the key to disable/enable the Auto Off feature.
Auto power down will occur after 5 minutes of inactivity.
- to disable the Auto Off feature, press and hold the power key until “HELD P on”
appears.
- to enable the Auto Off feature, press and hold the power key until “HELD AOFF”
appears.
Wavelength/Backlight key
Provides two functions:
•• Press and hold to toggle the Backlight on or off.
•• Press to cycle through the calibrated wavelengths A .
Note: When the OPM4 is receiving a wavelength ID’d light, it automatically switches to
the ID wavelengths. WAVE ID indicator B appears on the display.
8888
-88.88
B
A
nm
8888
-88.88
µW
dB dBm
B
WAVE ID
WAVE ID
8888
-88.88
dB/dBm/µW key
dB
dBm
Provides two functions:
A
µW
dB dBm
nm
A
µW
dB dBm
Dual WAVE mode, or
Tri WAVE mode - first page
Single WAVE mode, or
Tri WAVE mode - second page
W
nm
•• Press to toggle test readings between insertion loss in dB and power in dBm A .
•• Press and hold to view power in µW B .
8888
-88.88
WAVE ID
8888
-88.88
nm
µW
dB dBm
Hz
nm
µW
dB dBm
B
A
B
A
7
OPM4 Models
Set
Ref
Ref/Set key
Provides two functions:
•• To set Reference
•• To display Reference
Test Mode: To Set Reference
1. Set your OPM4 to the desired measurement mode (dB, dBm, µW).
2. To store the currently measured level or multiple levels as the new reference levels, press and hold
the Ref key until the label HELD SET is displayed.
3. Once the new reference is set, the OPM4 switches to the dB measurement mode.
Note: When setting references from multiple wavelength ID’d sources, allow several seconds prior to
setting the references for wavelengths to be ID’d and measured.
Test Mode: To View Reference
1. Press the Ref key to display the stored reference level for the currently selected wavelength or
detected WAVE ID wavelength(s). The reference indicator REF 1 and reference value(s) will be
briefly (for about 2-3 seconds) displayed as follows:
CW mode - the reference value will be displayed in the field 2
Single WAVE mode - the reference value will be displayed in the field 2
Dual WAVE mode - the reference values will be displayed in the fields 2 and 3
Tri WAVE mode - The reference value will be displayed in the fields 2 and 3 on the first page and in
the field 2 on the second page (replaces first page automatically in ~ 2 seconds).
1550
1
-5.48
REF
1310
nm
dBm
WAVE ID
2
1
REF
WAVE ID
-5.15
1490
-5.42
Single WAVE mode, or
Tri WAVE mode - second page
nm
dBm
2
nm
3
dBm
Dual WAVE mode, or
Tri WAVE mode - first page
Display Readings - Splash and Auto Off Screen
8888
A
8888
B
Splash (Version) screen: During power up, the OPM4
software version is shown on the top line A . The OPM type
and detector type are shown on the next line B . The Splash
screen will be automatically replaced by the Test screen after
~ 2 seconds.
Auto Off screen: When the unit is about to power off due to
inactivity, then display will show “OFF” B and count down
the number of seconds until power will be shut off.
8
OPM4 Models
Display Readings in Test Mode
8888 nm
1
2
REF
3
-88.88
WAVE ID
8888
4
µW
dB dBm
5
Hz
6
Single WAVE mode, or
Tri WAVE mode - second page
8888 nm
1
2
3
REF
-88.88
WAVE ID
8888
-88.88
4
µW
dB dBm
5
Hz
nm
6
µW
dB dBm
7
Dual WAVE mode, or
Tri WAVE mode - first page
Ref. # Description
1
Battery status indicator - displays estimated battery life remaining. The 3 bars indicate
the charge level. When the battery is low, the battery indicator will flash.
2
REF indicator - appears to indicate previously stored reference value(s).
On the Ref key press, the reference indicator REF 2 and reference value(s) for the currently
selected wavelength or detected WAVE ID wavelength(s) will be displayed briefly (for about 2-3
seconds).
CW mode - the reference value will be displayed in the field 5 .
Single WAVE mode - The reference value will be displayed in the field 5 .
Dual WAVE mode - The reference values will be displayed in the fields 5 and 7 .
Tri WAVE mode - The reference value will be displayed in the field 5 and 7 on the first
page and in the field 5 on the second page (replaces first page automatically in ~ 2 seconds).
3
WAVE ID indicator - Will be displayed if the OPM4 is receiving a wavelength ID’d light.
4
CW mode - This field displays the currently enabled wavelength.
Single WAVE mode - This field displays the detected WAVE ID wavelength.
Dual WAVE mode - This field displays the first detected WAVE ID wavelength.
Tri WAVE mode - This field displays the first WAVE ID wavelength and the third WAVE ID
wavelength that is shown on a second page (replaces first page automatically in ~ 2 seconds).
5
This field displays measured power dBm or µW or insertion loss dB at the wavelength displayed
in the field 4 .
If power or loss is too high or low for the OPM4 to measure, this field will display HI or LO.
6
CW mode - This field displays modulation frequency if detected.
Dual WAVE mode - This field displays the second wavelength.
7
Dual WAVE mode - This field displays measured power dBm or µW or insertion loss dB at the
wavelength displayed in the field 6 .
If power or loss is too high or low for the OPM4 to measure, this field will show HI or LO.
9
OPM5 Models
3 Adapter cap
Power port 1
2 Optical input
4 MINI-USB port
AFL Telecommunications
Optical Power Meter
0000
WAVE ID
Set Reference key
0.00
0000
0.00
µW
dB/dBm/µW key
Hz
nm
µW
dB
Hz
nm
µW
dB
Set
dB
Ref
dBm
λ
File
Store key
Store
Wavelength/Backlight
key
Arrow/File key
Recall
Clear
Power key
5 Display
Fiber
OPM5
Arrow/Fiber key
Recall key
New Feature
File Management system allows simple organization of power or insertion loss measurements into multiple
files. Using the supplied Test Results Management - TRM software (Windows® compatible) and USB
connection, test records are transferred to a PC for analyzing, generating professional test reports, and
printing. For more information, refer to the TRM software User’s Guide (available on supplied CD and
www.AFLtele.com/go/Software).
Ports and Display
Ref # Description
Power port - this is the interface for the AC power adapter/charger.
1
10
2
Optical input (adapter cap mount) - accepts Noyes thread-on adapter caps.
3
Adapter cap - the OPM5 optical input must be equipped with an adapter cap.
4
Mini-USB type B port - allows connection to a PC for downloading stored test results.
5
Display - depending on the selected mode, shows measured power dBm or µW or insertion loss
dB, calibrated wavelengths, tone signal Hz, wavelength ID, references, file and fiber number,
and battery charge status.
OPM5 Models
Dual Function Keys
YY
Secondary function of a key
Press and hold to activate
XX
Primary function of a key
Press and release to activate
Power key
Provides two functions:
•• Press the key to turn the OPM5 on/off. The unit will turn off automatically five minutes
after the last key press.
•• Press and hold the key to disable/enable the Auto Off feature.
Auto power down will occur after 5 minutes of inactivity.
- to disable the Auto Off feature, press and hold the power key until “HELD P on”
appears.
- to enable the Auto Off feature, press and hold the power key until “HELD AOFF”
appears.
Arrow/File key
File
Provides several functions:
•• Testing mode: press and hold to enter the File Manager - File mode screen.
•• File mode: press to increment the File number.
•• Fiber and Recall mode: press to increment the Fiber number.
Arrow/Fiber key
Fiber
Provides several functions:
•• Testing mode: press and hold to enter the File Manager - Fiber mode screen.
•• Fiber mode: press to decrement the Fiber number.
•• File and Recall mode: press to decrement the File number.
Wavelength/Backlight key
Provides two functions:
•• Press and hold to toggle the Backlight on or off.
•• Press to cycle through the calibrated wavelengths A .
Note: When the OPM5 is receiving a wavelength ID’d light, it automatically switches to
the ID wavelengths. WAVE ID indicator B appears on the display.
8888
-88.88
B
nm
A
8888
-88.88
µW
dB dBm
WAVE ID
B
WAVE ID
8888
-88.88
Single WAVE mode, or
Tri WAVE mode - second page
nm
A
µW
dB dBm
nm
A
µW
dB dBm
Dual WAVE mode, or
Tri WAVE mode - first page
11
OPM5 Models
W
dB
dBm
dB/dBm/µW key
Provides two functions:
•• Press to toggle test readings between insertion loss in dB and power in dBm A .
•• Press and hold to view power in µW B .
8888
-88.88
8888
WAVE ID
-88.88
Set
Ref
B
nm
µW
dB dBm
Hz
nm
A
B
µW
dB dBm
A
Ref/Set key
Provides two functions:
•• Press to display the stored reference level for the currently selected wavelength or
multiple wavelengths.
•• Press and hold until HELD SET is displayed to store the currently measured level or
multiple levels as the new reference levels. Once the new reference is set, the OPM5
switches to the dB measurement mode.
Store
Store key
Provides two functions :
•• Test mode: press this key to save the currently displayed measured data in the next
available memory location (next Fiber) of the current File.
•• File/Fiber/Recall/Clear mode: press this key to return to the Test mode.
Recall
Store
Recall key
Press this key to enter the Recall mode that allows viewing and deleting stored test records.
Press again to exit the Recall mode.
Recall
Clear
Clear
The Clear mode screen may be accessed and used as follows:
•• From the File mode screen to delete a file
•• From the Recall mode screen to delete a fiber
Display Readings - Splash and Auto Off Screen
8888
A
8888
B
Splash (Version) screen: During power up, the OPM5
software version is shown on the top line A . The OPM type
and detector type are shown on the next line B . The Splash
screen will be automatically replaced by the Test screen after
~ 2 seconds.
Auto Off screen: When the unit is about to power off due to
inactivity, then display will show “OFF” B and count down
the number of seconds until power will be shut off.
12
OPM5 Models
Display Readings in Test Mode
8888 nm
1
2
REF
3
-88.88
WAVE ID
8888
4
µW
dB dBm
5
Hz
6
Single WAVE mode, or
Tri WAVE mode - second page
8888 nm
1
2
3
REF
-88.88
WAVE ID
8888
-88.88
4
µW
dB dBm
5
Hz
nm
6
µW
dB dBm
7
Dual WAVE mode, or
Tri WAVE mode - first page
Ref. # Description
1
Battery status indicator - displays estimated battery life remaining. The 3 bars indicate
the charge level. When the battery is low, the battery indicator will flash.
2
REF indicator - appears to indicate previously stored reference value(s).
On the Ref key press, the reference indicator REF 2 and reference value(s) for the currently
selected wavelength or detected WAVE ID wavelength(s) will be displayed briefly (for about 2-3
seconds).
CW mode - the reference value will be displayed in the field 5 .
Single WAVE mode - The reference value will be displayed in the field 5 .
Dual WAVE mode - The reference values will be displayed in the fields 5 and 7 .
Tri WAVE mode - The reference value will be displayed in the field 5 and 7 on the first
page and in the field 5 on the second page (replaces first page automatically in ~ 2 seconds).
3
WAVE ID indicator - Will be displayed if the OPM5 is receiving a wavelength ID’d light.
4
CW mode - This field displays the currently enabled wavelength.
Single WAVE mode - This field displays the detected WAVE ID wavelength.
Dual WAVE mode - This field displays the first detected WAVE ID wavelength.
Tri WAVE mode - This field displays the first WAVE ID wavelength and the third WAVE ID
wavelength that is shown on a second page (replaces first page automatically in ~ 2 seconds).
5
This field displays measured power dBm or µW or insertion loss dB at the wavelength displayed
in the field 4 .
If power or loss is too high or low for the OPM5 to measure, this field will display HI or LO.
6
CW mode - This field displays modulation frequency if detected.
Dual WAVE mode - This field displays the second wavelength.
7
Dual WAVE mode - This field displays measured power dBm or µW or insertion loss dB at the
wavelength displayed in the field 6 .
If power or loss is too high or low for the OPM5 to measure, this field will show HI or LO.
13
OPM5 Models
File Manager: To Select Current File
1. Press and hold the
Arrow/File key to enter the File Manager - File mode screen.
2. A blinking cursor A to the left of FILE indicates that you are in the File mode.
B
FILE
4
A
B - File indicator
3. Use the
/
C
FibS
D
22
E
C - File number
D - Fiber indicator
E - Number of fibers in the selected file
arrow keys to increment/decrement the File number.
•• Press and hold
arrow key to go to the first file.
•• Press and hold
arrow key to go the last file.
4. Press the Store key to make the selected file current and exit the File mode.
5. The indicator FILE blinks for ~ 1.5 seconds before the OPM5 returns to test measuring screen.
File Manager: To Select Current Fiber
1. Press and hold the
Arrow/Fiber key to enter the File Manager - Fiber mode screen.
2. A blinking cursor A to the left of Fiber indicates that you are in the Fiber mode.
FILE
4
A
3. Use the
/
B
C
FIBS
D
22
E
arrow keys to increment/decrement the Fiber number.
•• Press and hold
arrow key to increment by 10 fibers.
•• Press and hold
arrow key to decrement by 10 fibers.
4. Press the Store key to make the selected Fiber number current and exit the Fiber mode.
Test Mode: To Set Reference
1. Set your OPM5 to the desired measurement mode (dB, dBm, µW).
2. To store the currently measured level or multiple levels as the new reference levels, press and hold
the Ref key until the label HELD SET is displayed.
3. Once the new reference is set, the OPM5 switches to the dB measurement mode.
Note: When setting references from multiple wavelength ID’d sources, allow several seconds prior to
setting the references for wavelengths to be ID’d and measured.
14
OPM5 Models
Test Mode: To View Reference
1. Press the Ref key to display the stored reference level for the currently selected wavelength or
detected WAVE ID wavelength(s). The reference indicator REF 1 and reference value(s) will be
briefly (for about 2-3 seconds) displayed as follows:
CW mode - the reference value will be displayed in the field 2
Single WAVE mode - the reference value will be displayed in the field 2
Dual WAVE mode - the reference values will be displayed in the fields 2 and 3
Tri WAVE mode - The reference value will be displayed in the fields 2 and 3 on the first page and in
the field 2 on the second page (replaces first page automatically in ~ 2 seconds).
1550
1
REF
-5.48
1310
nm
2
dBm
1
WAVE ID
REF
WAVE ID
-5.15
1490
-5.42
nm
dBm
2
nm
3
dBm
Dual WAVE mode, or
Tri WAVE mode - first page
Single WAVE mode, or
Tri WAVE mode - second page
Test Mode: To Store Measurements
File Management system allows simple organization of power or insertion loss measurements into multiple
files. Using the supplied Test Results Management - TRM software (Windows® compatible) and USB
connection, test records are transferred to a PC for analyzing, generating professional test reports, and
printing. For more information, refer to the TRM software User’s Guide (available on the supplied CD and
from the AFL - Noyes web at www.AFLtele.com/go/Software).
1. Set your OPM5 to the desired measurement mode (dB, dBm, µW) and wavelengths.
Note: When the OPM5 is receiving wavelength ID, measured data is saved at all currently received
ID’d wavelengths.
2. Test the fiber.
3. Press the Store key to save the currently displayed measured data.
4. The OPM5 will briefly display the following screen:
A
Stor
4
WAVE ID
B
1310 nm
0.22
dB
C
D
A - Blinking label Stor, B - File number, C - Tested wavelength, D - Stored result
Note: For dual/tri-wavelength testing, the stored result for a subsequent wavelength will be shown on
the second/third page that automatically changes in ~ 2 seconds.
15
OPM5 Models
5. The current measurements are stored in the next available memory location (next Fiber) of the current
File.
Recall Mode: To View Stored Records
1. Press and hold the
Arrow/File key to enter File mode and select the desired File to view.
2. Press the Recall key to display the Recall mode screen.
3. Once the Recall key is pressed and before entering the Recall mode screen, the number of currently
open file A and number of fibers B stored in the current file will flash for ~ 1.5 seconds.
FILE
4
A
FIBS
22
B
4. Once in the Recall mode screen, the display changes to show the following data:
FIBR
D
FIBR
4
1310 nm
WAVE ID
0.22
4
A
dB
B
1310
C
0.22
Hz
nm
E
dB
A - Fiber number, B - Wavelength, C - Stored test result,
D - WAVE ID icon if WAVE ID was active when the measurement was taken,
E - Hz icon, if Tone was active when the measurement was taken.
5. Use the /
this location.
arrow keys to increment/decrement the fiber number and view the test result stored in
•• Press and hold the
arrow key to go to the last fiber.
•• Press and hold the
arrow key to go to the first fiber.
•• For multiple-Wave test results, press the wavelength
key to change wavelengths.
6. Press the Recall key to exit.
7. Press the Store key to exit the File mode.
Note: you can go directly to the Recall mode for the current file by pressing the Recall key from the Test
screen.
16
OPM5 Models
To Edit Stored Records
1. If needed, press and hold the Arrow/File key to enter File mode and select the desired File. Press
the Store key to exit the File mode.
2. Press and hold the
arrow key to enter the Fiber mode and select the fiber to edit.
•• Press and hold
arrow key to increment by 10 fibers.
•• Press and hold
arrow key to decrement by 10 fibers.
3. Press the Store key to return to measuring mode.
4. Retest the fiber.
5. Press the Store key to save new measurements in the selected memory location.
Clear Mode: To Erase Stored File
1. Press and hold the
Arrow/File key to enter File mode and select the desired File to erase.
2. Press and hold the Store
Recall
keys to erase the current file.
Clear
3. The message “Stor to clr FILE” appears on the display.
4. Press the Store key to delete all fibers for the current file. While clearing (may take a while for many
fibers stored in a file) the display will show “Clr”.
5. When complete, the display will briefly show “FILE Clr”.
Clear Mode: To Erase Stored Fiber
1. Press the Recall key to enter the Recall mode (recall records stored in the current file).
2. Use the
/
arrow keys to select the desired fiber to erase.
3. Press and hold the Store
Recall
keys to erase the current fiber information (all wavelengths).
Clear
4. The message “Stor to clr Fibr” appears on the screen.
5. Press the Store key to erase the current fiber.
6. When complete, the display will briefly show “Fibr Clr”.
17
OLS-Series Light Sources
OLS1-Dual LED Light Source
2 Output port
Power port 7
External Power 6
indicator
2 Active Output
indicator
9V
850nm 1300nm
AFL Telecommunications
OLS1 DUAL
Optical Light Source
CW
850nm
1300nm
3 850/1300/CW
wavelengths key
Power key 4
NOYES
5 Low Battery
indicator
OLS1-Dual Features
1. Output port - Universal Connector Interface (UCI). Accepts swappable SC connector (ST, FC, and LC
connector styles are available).
2. Active Output indicators (illuminated when the corresponding output port is ON) - The Active
Output indicators light up as follows:
Test Mode
Active Output indicator status
WAVE ID dual Wave
850 and 1300 indicator - ON continuous light
WAVE ID single Wave corresponding indicator - ON continuous light
CW
corresponding indicator - ON blinking light
3. 850/1300/CW key - Wavelengths select key provides several functions.
WAVE ID Mode
•• Press the key to select wavelengths in the following sequence:
18
OLS1-Dual LED Light Source
Dual 850/ 1300 nm
850 nm
1300 nm
Dual 850/ 1300 nm
...
•• Press and hold the key to switch to the CW mode at the currently selected wavelength.
Note: If the OLS1 operates in the Dual WAVE ID mode, it will switch the output port to the next
wavelength in a sequence as follows:
Dual 850/1300 nm WAVE ID
key Press and hold
850 nm CW
CW Mode
•• Press and hold the key to select wavelengths in the following sequence:
850 nm
1300 nm
850 nm
...
•• Press the key to return to the WAVE ID mode at the currently selected wavelength.
4. Power key - Press and hold the key until all indicators light up to turn the OLS1-Dual on or off. The
instrument turns on in its default condition: 850/1300 nm Dual mode sending Wavelength ID.
5. Low Battery indicator - Lights up when a Low Battery condition exists; batteries require
replacement.
6. External Power indicator - Lights up, whether the unit is on or off, when the correct AC power
adapter is connected.
7. Power port  (located on the top panel) - This is the interface for an AC power adapter.
19
OLS2-Dual Laser Source
5 Output port
Power port 2
6 Active Output
indicator
9V
External Power 3
indicator
1310nm 1550nm
AFL Telecommunications
Wave ID
CW
2KHz
1KHz
330HZ
270Hz
OLS2 DUAL
Optical Laser Source
Wavelengths- l key 7
λ
10 Active Mode
Indicators
Mode
9 Mode key
Output
Adjust
8 Output Adjust keys
Power key 1
NOYES
4 Low Battery
indicator
OLS2-Dual features
1. Power key - Press and hold the key until all indicators light up to turn the OLS2-Dual On or Off. The
instrument turns on in its last operating condition it was in prior to being turned off.
2. Power port  (located on the top panel) - This is the interface for an AC power adapter.
3. External Power indicator -  Lights up, whether the unit is on or off, when the correct AC power
adapter is connected.
4. Low Battery indicator - Lights up when a Low Battery condition exists; batteries require
replacement.
5. Output port - Universal Connector Interface (UCI). Accepts swappable SC connector (ST, FC, and LC
connector styles are available).
6. Active Output indicators - The Active Output indicators light up when the corresponding output port
is on.
7. Wavelengths -l key -  Wavelength select key changes the active output wavelength(s) as indicated
below for each operational mode.
20
OLS2-Dual Laser Source
WAVE ID Mode (WAVE ID indicator on)
•• Press the l key to select wavelengths in the following sequence:
Dual 1310/ 1550 nm
1310 nm
1550 nm
Dual 1310/ 1550 nm
...
Note:  When a wavelength is selected, the corresponding Active Output indicator on the front
panel is illuminated. If the Dual WAVE ID mode is selected, both Active Output indicators are
illuminated.
CW Mode or any of the Tone Modes
(WAVE ID indicator off, CW or 2kHz/ 1kHz/ 330Hz/ 270Hz indicator illuminated):
•• Press the l key to select wavelengths in the following sequence:
1310 nm
1550 nm
1310 nm
...
Note:  When a wavelength is selected, the corresponding Active Output indicator on the front
panel is illuminated.
8. Output Adjust keys - The ∆ and ∇ arrow keys allow entering the Output Adjust mode and changing
the power level currently being transmitted at the output port within approximately 3 dB power range
between the factory preset maximum and minimum power levels.
∆ arrow key increases the optical output power up to the allowed maximum level.
∇ arrow key decreases the optical output power down to the allowed minimum level.
Output Adjust Mode Characteristics
a) The Output Adjust mode is valid in CW and single WAVE ID modes only.
Note:
•• If the Output Adjust mode is enabled while the OLS2-Dual is transmitting Dual WAVE ID, the
unit will switch to the 1310nm WAVE ID mode.
•• If the Output Adjust mode is enabled while the OLS2-Dual is transmitting any of the Tone
frequencies, the unit will switch to the CW mode.
b) To enable the Output Adjust mode and begin power change, perform the following steps:
•• Hold either the ∆ or ∇ arrow keys for approximately 4 seconds.
•• An Active Mode indicator of the currently enabled mode (CW or WAVE ID) will turn ON blinking light at approximately ½ second rate.
•• The optical output will begin to change correspondingly to the currently pressed key. ( ∆ key increase output power, ∇ key - decrease output power).
c) Key functions while in the Output Adjust mode are indicated below:
•• Mode key - -toggles between single WAVE ID and CW mode.
•• Wavelength key - toggles optical output between 1310 & 1550 nm.
•• ∆ or ∇ arrow key - increases or decreases the optical output power level.
d) The Output Adjust Mode is automatically disabled after approximately 8 seconds with no key
presses. The OLS2-Dual will return to its previous operating mode it was in prior to entering the
Output Adjust mode.
Output Adjust Mode Operation
To adjust the output power, perform the following steps:
21
OLS2-Dual Laser Source
a) Press and hold either ∆ or ∇ arrow key for ~. 4 seconds to enable the Output Adjust mode.
b) When the OLS2-Dual enters the Output Adjust mode, an Active Mode indicator of the currently
enabled mode (CW or WAVE ID) turns ON - blinking light, the Output Adjust mode is enabled. The
currently selected output can now be adjusted.
c) Press the ∆ or ∇ arrow key to increase or decrease the current power level.
•• Press momentarily to adjust the optical output in a small step.
•• Press and hold to adjust in small steps for a total of six steps at a rate of ~. 2 steps per
second. Continuing to press the same arrow key will cause the unit to increase step size to
and continue at the same repetition rate. Releasing the held key at any time resets the step
size to small and resets the “step size counter” to zero.
Note: End of adjust range indication - Noyes recommends using an Optical Power Meter to
monitor the output when adjusting the OLS2-Dual power level. However, this is not necessary to
set the OLS2-Dual to the maximum or minimum output level. When the end of the adjustment
range has been reached, the OLS2-Dual will turn off the Active Mode Indicator and blink the
Active Output indicator at a rate of approximately 1/4 second. Releasing the arrow key restores
the LED Indicators to their previous states.
9. Mode key - Press this key to select the desired operating mode
When the OLS2-Dual operates in WAVE ID Mode (WAVE ID indicator on)
Press the Mode key to switch the unit from the WAVE ID to CW mode.
•• The WAVE ID indicator will turn off and the CW indicator will turn on.
•• If multiple ID wavelengths were being transmitted, the unit will continue to transmit the shortest
wavelength - 1310nm.
When the OLS2-Dual operates in CW mode (CW indicator on)
Press the Mode key to switch the unit from the CW mode to 2kHZ Tone mode.
•• The CW indicator will turn off and the 2kHz Tone indicator will turn on.
•• The unit will start transmitting the selected wavelength with 2kHz modulation.
When the OLS2-Dual operates in 2kHz Tone mode (2kHz indicator on)
Press the Mode key to switch the unit from 2kHz Tone mode to 1kHz Tone mode.
•• The 2kHz Tone indicator will turn off and the 1kHz Tone indicator will turn on.
•• Press the Mode key again to select the next supported frequency. The unit will cycle through the
available frequencies in the following sequence:
2kHz
1kHz
330 Hz
270 Hz ...
•• The corresponding Tone indicator will turn on.
When the OLS2-Dual operates in 270Hz Tone mode (270 Hz indicator on)
Press the Mode key to switch the unit from the 270Hz Tone mode to WAVE ID mode.
•• The 270Hz Tone indicator will turn off and the WAVE ID indicator will turn on.
•• The unit will start transmitting the wavelength selected on the embedded wavelength ID.
10. Active Mode indicators - The Active Mode indicators light up when the corresponding mode is
enabled.
22
OLS4 Integrated LED and Laser Light Source
Power port 2
6 SM output port
MM output port 5
MM Active Output 7
indicators
7
9V
850nm 1300nm
SM Active Output
indicators
1310nm 1550nm
AFL Telecommunications
External Power 3
indicator
OLS4
Optical Laser Source
with Wave ID
850/1300/CW key 8
MM wavelengths key
Power key 1
CW
CW
850nm
1300nm
1310nm
1550nm
9 1310/1550/CW key SM wavelengths key
Tone
10 Tone key & indicator
NOYES
4 Low Battery indicator
OLS4 Features
1. Power key - Press and hold the key until all indicators light up to turn the OLS4 ON or OFF. The
OLS4 turns on in its default condition: 850/1300 nm Dual mode sending Wavelength ID.
2. Power port  (located on the top panel) - This is the interface for an AC power adapter.
3. External Power indicator - Lights up, whether the unit is on or off, when the correct AC power
adapter is connected.
4. Low Battery indicator - Lights up when a Low Battery condition exists; batteries require
replacement.
5. MM output port - Universal Connector Interface (UCI). Accepts swappable SC connector (ST, FC,
and LC connector styles are available).
6. SM output port - Universal Connector Interface (UCI). Accepts swappable SC connector (ST, FC,
and LC connector styles are available).
23
OLS4 Integrated LED and Laser Light Source
7. Active Output indicators (illuminated when the corresponding output port is ON) - The Active
Output indicators light up as follows:
Test Mode
WAVE ID dual Wave
WAVE ID single Wave
CW or Tone
Active Output indicator status
corresponding indicators - ON continuous light
corresponding indicator - ON continuous light
corresponding indicator - ON blinking light
8. 850/1300/CW key - Multimode wavelengths select key provides several functions as follows:
WAVE ID Mode
•• Press the key to select wavelengths in the following sequence:
•• Dual 850/ 1300 nm
850 nm
1300 nm
Dual 850/ 1300 nm
...
•• Press and hold the key to switch to the CW mode at the currently selected wavelength.
Note: If the OLS4 operates in the Dual WAVE ID mode, it will switch the MM output port to the next
wavelength in a sequence as follows:
Dual 850/1300 nm WAVE ID
key Press and Hold
850 nm CW
CW Mode
•• Press and hold the key to select wavelengths in the following sequence:
850 nm
1300 nm
850 nm
...
•• Press the key to return to the WAVE ID mode at the currently selected wavelength.
9. 1310/1550/CW key - Single-mode wavelengths select key provides several functions as follows:
WAVE ID Mode
•• Press the key to select wavelengths in the following sequence:
Dual 1310/ 1550 nm
1310 nm
1550 nm
Dual 1310/ 1550 nm
...
•• Press and hold the key to switch to the CW mode at the currently selected wavelength.
Note: If the OLS4 operates in the Dual WAVE ID mode, it will switch the SM output port to the
next wavelength in a sequence as follows:
Dual 1310/1550 nm WAVE ID
key Press and Hold
1310 nm CW
CW Mode
•• Press and hold the key to select wavelengths in the following sequence:
1310 nm
1550 nm
1310 nm
...
•• Press the key to return to the WAVE ID mode at the currently selected wavelength.
10. Tone key and Indicator -  
Tone key - Pressing the Tone key enables or disables the 2kHz Tone mode on the SM port at 1550
nm wavelength. The WAVE ID mode is disabled when the unit is transmitting a Tone signal. When the
Tone mode is disabled, the unit returns to the WAVE ID mode at the currently selected wavelength.
Tone indicator - The Tone indicator is ON-continuous light when the OLS4 is transmitting a Tone
signal.
24
OLS7-3 Triple Wavelength Laser Source
5 Output port
Power port 2
6 Active Output
indicators
9V
1310nm
External Power 3
indicator
1550nm 1625nm
OLS7
Optical Light Source
Wavelengths- lkey 7
λ
Wave ID
CW
2KHz
1KHz
330HZ
270Hz
Mode
9 Active Mode
Indicators
8 Mode key
Power key 1
NOYES
4 Low Battery indicator
OLS7-3 Features
1. Power key - Press this key to turn the OLS7-3 on or off. When the unit is turned on, it returns to the
last operating condition it was in prior to being turned off.
2. Power port  (located on the top panel) - This is the interface for an AC power adapter.
3. External Power indicator - Lights up, whether the unit is on or off, when the correct AC power
adapter is connected.
4. Low Battery indicator - Lights up when a Low Battery condition exists; batteries require
replacement.
5. Output port - Universal Connector Interface (UCI). Accepts swappable SC connector (ST, FC, and LC
connector styles are available).
6. Active Output indicators - The Active Output indicators light up when the corresponding output port
is on.
7. Wavelength - l key - Wavelength select key changes the active output wavelength(s) as indicated
below for each operational mode.
WAVE ID Mode (WAVE ID indicator on)
•• Press the l key to select wavelengths in the following sequence:
25
OLS7-3 Triple Wavelength Laser Source
Triple 1310/ 1550/ 1625 nm
1310 nm
1550 nm
1625 nm
Dual 1310/ 1550 nm
Dual 1550/ 1625 nm
Triple 1310/ 1550/ 1625 nm
...
Note:  When a wavelength is selected, the corresponding Active Output indicator on the front
panel is illuminated. If the Triple WAVE ID mode is selected, all three Active Output indicators are
illuminated.
CW Mode or any of the Tone Modes
(WAVE ID indicator off, CW or 2kHz/ 1kHz/ 330Hz/ 270Hz indicator illuminated):
•• Press the l key to select wavelengths in the following sequence:
1310 nm
1550 nm
1625 nm
1310 nm
...
Note:  When a wavelength is selected, the corresponding Active Output indicator on the front
panel is illuminated.
8. Mode key - Press this key to select the desired operating mode
When the OLS7-3 operates in WAVE ID Mode (WAVE ID indicator on)
Press the Mode key to switch the unit from the WAVE ID to CW mode.
•• The WAVE ID indicator will turn off and the CW indicator will turn on.
•• If multiple ID wavelengths were being transmitted, the unit will continue to transmit the shortest
wavelength.
When the OLS7-3 operates in CW mode (CW indicator on)
Press the Mode key to switch the unit from the CW mode to 2kHZ Tone mode.
•• The CW indicator will turn off and the 2kHz Tone indicator will turn on.
•• The unit will start transmitting the selected wavelength with 2kHz modulation.
When the OLS7-3 operates in 2kHz Tone mode (2kHz indicator on)
Press the Mode key to switch the unit from 2kHz Tone mode to 1kHz Tone mode.
•• he 2kHz Tone indicator will turn off and the 1kHz Tone indicator will turn on.
•• Press the Mode key again to select the next supported frequency. The unit will cycle through the
available frequencies in the following sequence:
2kHz
1kHz
330 Hz
270 Hz ...
•• The corresponding Tone indicator will turn on.
When the OLS7-3 operates in 270Hz Tone mode (270 Hz indicator on)
Press the Mode key to switch the unit from the 270Hz Tone mode to WAVE ID mode.
•• The 270Hz Tone indicator will turn off and the WAVE ID indicator will turn on.
•• The unit will start transmitting the wavelength selected on the embedded wavelength ID.
9. Active Mode indicators - The Active Mode indicators light up when the corresponding mode is
enabled.
26
OLS7-FTTH Triple Wavelength Laser Source
5 Output port
Power port 2
6 Active Output
indicators
9V
1310nm
External Power 3
indicator
1490nm 1550nm
OLS7-FTTH
Optical Light Source
Wavelengths- l key 7
λ
Wave ID
CW
2KHz
1KHz
330HZ
270Hz
Mode
9 Active Mode
Indicators
8 Mode key
Power key 1
NOYES
4 Low Battery indicator
OLS7-FTTH Features
1. Power key - Press this key to turn the OLS7-FTTH on or off. When the unit is turned on, it returns to
the last operating condition it was in prior to being turned off.
2. Power port  (located on the top panel) - This is the interface for an AC power adapter.
3. External Power indicator - Lights up, whether the unit is on or off, when the correct AC power
adapter is connected.
4. Low Battery indicator - Lights up when a Low Battery condition exists; batteries require
replacement.
5. Output port - Universal Connector Interface (UCI). Accepts swappable SC connector (ST, FC, and LC
connector styles are available).
6. Active Output indicators - The Active Output indicators light up when the corresponding output port
is on.
7. Wavelength - l key - Wavelength select key changes the active output wavelength(s) as indicated
below for each operational mode.
27
OLS7-FTTH Triple Wavelength Laser Source
WAVE ID Mode (WAVE ID indicator on)
•• Press the l key to select wavelengths in the following sequence:
Dual 1490/ 1550 nm
1310 nm
Triple 1310/1490 / 1550 nm
...
Note:  When a wavelength is selected the corresponding Active Output indicator on the front
panel is illuminated. If the Triple WAVE ID mode is selected all three Active Output indicators are
illuminated.
CW Mode or any of the Tone Modes
(WAVE ID indicator off, CW or 270 Hz / 330 Hz / 1 kHz/ 2 kHz indicator illuminated):
•• Press the l key to select wavelengths in the following sequence:
1310 nm
1490 nm
1550 nm
1310 nm
...
Note:  When a wavelength is selected the corresponding Active Output indicator on the front
panel is illuminated.
8. Mode key - Press this key to select the desired operation mode
When the OLS7-FTTH operates in WAVE ID Mode (WAVE ID indicator on)
Press the Mode key to switch the unit from the WAVE ID to CW mode.
•• The WAVE ID indicator will turn off and the CW indicator will turn on.
•• If multiple ID wavelengths were being transmitted, the unit will continue to transmit the shortest
wavelength.
When the OLS7-FTTH operates in CW mode (CW indicator on)
Press the Mode key to switch the unit from the CW mode to 2kHZ Tone mode.
•• The CW indicator will turn off and the 2kHz Tone indicator will turn on.
•• The unit will start transmitting the selected wavelength with 2kHz modulation.
When the OLS7-FTTH operates in 2kHz Tone mode (2kHz indicator on)
Press the Mode key to switch the unit from 2kHz Tone mode to 1kHz Tone mode.
•• The 2kHz Tone indicator will turn off and the 1kHz Tone indicator will turn on.
•• Press the Mode key again to select the next supported frequency. The unit will cycle through the
available frequencies in the following sequence:
2kHz
1kHz
330 Hz
270 Hz ...
•• The corresponding Tone indicator will turn on.
When the OLS7-FTTH operates in 270Hz Tone mode (270 Hz indicator on)
Press the Mode key to switch the unit from the 270Hz Tone mode to WAVE ID mode.
•• The 270Hz Tone indicator will turn off and the WAVE ID indicator will turn on.
•• The unit will start transmitting the wavelength selected on the embedded wavelength ID.
9. Active Mode indicators - The Active Output indicators light up when the corresponding mode is
enabled.
28
Measuring Optical Power
It is important to keep all optical connections and surfaces free from dirt, oils, or other
contaminants to ensure proper operation. Always clean all test jumpers before conducting the
test procedures outlined in this Guide (see Section 5: “Maintenance” for details).
!
To Measure Optical Power
1. Turn on the OPM optical power meter.
2. Select the appropriate fiber optic test jumper. The fiber type of this jumper must be the same as the
fiber type normally connected to the output being measured.
3. Mount the appropriate adapter cap on the OPM optical input. This adapter cap must match the
connector on the end of the test jumper you will connect to the OPM.
4. Connect one end of the test jumper to the OPM adapter cap and the other end to the optical output to
be measured.
5. Press the λ key to select the calibrated wavelength that matches the nominal wavelength of the
source being measured.
6. If the OPM is presently measuring loss dB, press the dB/dBm key to display power in dBm, or press
and hold to display power in µW.
7. OPM5 power meters only: You may press the Store key to save the displayed measurement in the
next available memory location.
fiber optic equipment
Output
Input
4
3
adapter cap
4
9V
AFL Telecommunications
OPTICAL POWER METER
0000
WAVE ID
0.00
0000
0.00
µW
Set
6
dB
dBm
Ref
7
Store
test jumper
Hz
nm
µW
dB
Hz
nm
2
µW
dB
5
Recall
Clear
1
OPM 5
NOYES
29
Testing Multimode Links
Step I - Set the Reference (One Jumper Method)
1. Turn on the OPM optical power meter and OLS optical light source. Allow the light source to stabilize
(minimum of 2 minutes).
2. If not using the WAVE ID feature, set both instruments to the desired test wavelength.
3. Select the appropriate fiber optic transmit and receive test jumpers. The fiber type of these jumpers
must match the fiber type of the link to be tested.
4. Wrap and secure the transmit jumper five times around the appropriate diameter mandrel.
Clean both ends of the transmit jumper!
5. Connect the transmit jumper to the multimode output port of the OLS.
6. Mount an adapter cap on the OPM that matches the free connector on the transmit jumper.
7. Connect the free end of the transmit jumper to the OPM.
If necessary, press the dB/dBm key to display optical power in dBm.
8. If measured output power is outside of the normal range (specified by manufacturer), clean all fiber
connections or replace the transmit jumper. Repeat steps 4 - 7.
9. To set the reference level at the current wavelength(s): on the OPM, press and hold the Ref/Set key
until the label HELD SET is displayed to store the currently measured level or pair of levels as the new
reference levels.
Once the new reference is set, the OPM switches to the loss dB measurement mode. The OPM
should display 0 dB ± 0.05 dB.
Note: You may press the Ref/Set key to display the stored reference levels.
Note: When setting references from multiple wavelength ID’d sources, allow several seconds prior to
setting the references for wavelengths to be ID’d and measured.
4
Mandrel wrap
Transmit jumper
7
5
6
0 dB
OPM
OLS
Set the Reference
30
Testing Multimode Links
Step II - Verify Test Jumpers
10. Disconnect the transmit jumper from the OPM.
Do not disturb the transmit jumper at the OLS end!
11. If necessary, change the OPM adapter cap to match the connector on the receive jumper that will be
connected to the OPM.
Clean both ends of the receive jumper!
12. Connect the receive jumper to the OPM.
13. Mate the free ends of the transmit and receive jumpers using the appropriate adapter.
14. Verify that the insertion loss of this mated connector pair is under 0.75 dB, the maximum allowed by
the TIA (Noyes recommends 0.4 - 0.5 dB typical), as follows:
•• Observe the displayed power level. This is the mated connector pair insertion loss of the test
jumpers in dB.
•• If the insertion loss is not acceptable, disconnect the transmit and receive jumpers at the adapter,
clean the free ends of both test jumpers and repeat step 13 & 14.
•• If the insertion loss is still not acceptable, replace test jumpers and repeat steps 1-14.
15. If the insertion loss is acceptable, disconnect the transmit and receive jumpers at the adapter.
16. Move the OPM and OLS to opposite ends of the link to be tested.
Adapter
Receive jumper
Transmit jumper
Mandrel wrap
13
12
11
Do NOT disturb
this connection
New adapter cap
(if necessary)
0.4 dB
OLS
OPM
Verify Test Jumpers
31
Testing Multimode Links
Step III - Measure Multimode Link Insertion Loss
17. Connect the free ends of the transmit and receive jumpers to the multimode link under test.
Clean jumper end that connects to patch panel prior to every test!
18. OPM will measure and display the insertion loss of the link under test.
19. Perform one of the following:
If using OPM4: Record link insertion loss at the current test wavelength.
If using OPM5: Press the Store key on the OPM5 to save the displayed measurement in the next
available memory location.
20. Repeat steps 17-19 for all links to be tested at the current wavelength.
Transmit jumper
Mandrel wrap
Patch panel
~
~
Patch panel
Receive jumper
Link under test
2 dB
OLS
OPM
Measure Multimode Link Insertion Loss.
32
Testing Single-mode Links
Step I - Set the Reference (One Jumper Method)
1. Turn on the OPM optical power meter and OLS optical light source. Allow the light source to stabilize
(minimum of 2 minutes).
2. If not using the WAVE ID feature, set both instruments to the desired test wavelength.
3. Select the appropriate fiber optic transmit and receive test jumpers. The fiber type of these jumpers
must match the fiber type of the link to be tested.
4. Make a 30 mm (1.2 in) loop in a transmit jumper and secure it with a piece of tape (TIA testing only).
Make sure the loop remains in place during referencing and in all testing steps.
Clean both ends of the transmit jumper!
5. Connect the transmit jumper to the single-mode output port of the OLS.
6. Mount an adapter cap on the OPM that matches the free connector on the transmit jumper.
7. Connect the free end of the transmit jumper to the OPM.
If necessary, press the dB/dBm key to display optical power in dBm.
8. If measured output power is outside of the normal range (specified by manufacturer), clean all fiber
connections or replace the transmit jumper. Repeat steps 4 - 7.
9. To set the reference level at the current wavelength(s): on the OPM, press and hold the Ref/Set key
until the word HELD SET is displayed to store the currently measured level or pair of levels as the new
reference levels.
Once the new reference is set, the OPM switches to the loss dB measurement mode. The OPM
should display 0 dB ±0.05 dB.
Note: You may press and the Ref/ Set key to display the stored reference level for the currently
selected wavelength or pair of wavelengths.
Note: When setting references from multiple wavelength ID’d sources, allow several seconds prior to
setting the references for wavelengths to be ID’d and measured.
30 mm loop
Transmit jumper
4
7
5
6
0 dB
OLS
OPM
Set the Reference
33
Testing Single-mode Links
Step II - Verify Test Jumpers
10. Disconnect the transmit jumper from the OPM.
Do not disturb the transmit jumper at the OLS end!
11. If necessary, change the OPM adapter cap to match the connector on the receive jumper that will be
connected to the OPM.
Clean both ends of the receive jumper!
12. Connect the receive jumper to the OPM.
13. Mate the free ends of the transmit and receive jumpers using the appropriate adapter.
14. Verify that the insertion loss of this mated connector pair is under 0.75 dB, the maximum allowed by
the TIA (Noyes recommends 0.4 - 0.5 dB typical), as follows:
•• Observe the displayed power level. This is the mated connector pair insertion loss of the test
jumpers in dB.
•• If the insertion loss is not acceptable, disconnect the transmit and receive jumpers at the adapter,
clean the free ends of both test jumpers and repeat step 13 & 14.
•• If the insertion loss is still not acceptable, replace test jumpers and repeat steps 1-14.
15. If the insertion loss is acceptable, disconnect the transmit and receive jumpers at the adapter.
16. Move the OPM and OLS to opposite ends of the link to be tested.
Receive jumper
Transmit jumper
30 mm loop
13
12
11
Do NOT disturb
this connection
New adapter cap
(if necessary)
0.4 dB
OLS
OPM5
Verify Test Jumpers
34
Testing Single-mode Links
Step III - Measure Single-mode Link Insertion Loss
17. Connect the free ends of the transmit and receive jumpers to the single-mode link under test.
Clean jumper end that connects to patch panel prior to every test!
18. OPM will measure and display the insertion loss of the link under test.
19. Perform one of the following:
If using OPM4: Record link insertion loss at the current test wavelength.
If using OPM5: Press the Store key on the OPM5 to save the displayed measurement in the next
available memory location.
20. Repeat steps 17-19 for all links to be tested at the current wavelength.
Transmit jumper
Patch panel
~
~
Patch panel
Receive jumper
30 mm loop
Link under test
2 dB
OLS
OPM5
Measure Multimode Link Insertion Loss.
35
Maintenance Tips
Cleaning Optical Ports
CAUTION! Before conducting the following procedures be sure to have the instruments turned OFF.
Optical ports must be kept free from dirt or other contaminants to ensure accurate measurements and
operation. Always clean all test jumpers before conducting the test procedures outlined in this Guide. It is
important to keep dust caps in place when instruments are not being used.
For cleaning connector end faces on OLS light sources, test jumpers, and in fiber frames or adapters, use
optical quality cleaning fluid such as AFL FCC2 connector cleaning fluid and AFL CCT molded cleaning
tips.
For cleaning OPM optical ports and adapter caps, use lint-free optical cleaning wipes such as AFL
FiberWipes and optical quality cleaning fluid such as AFL FCC2 connector cleaning fluid (or IPA -Reagent
Grade Isopropyl Alcohol 99% or better) and a can of filtered compressed air.
Cleaning OPM Optical Ports
1. Unscrew the adapter cap from the adapter cap mount.
2. Use lint-free optical cleaning wipes such as AFL FiberWipes and optical quality cleaning fluid such as
AFL FCC2 connector cleaning fluid.
Note: if using isopropyl alcohol (IPA), be sure to use 99% pure IPA that has not been contaminated.
•• Dampen a portion of the wipe with the cleaning fluid.
•• Gently wipe the exposed OPM port starting with the wet section of the wipe and pulling it to the
dry section.
Note: Starting with the wet cleaning and finishing in the dry improves cleaning action, reduces static
buildup, and finishes with the end-face dry.
3. Using a can of filtered compressed air (held vertically), blow out any contaminants from the adapter
cap.
4. Replace the adapter cap once the cleaning is complete.
Cleaning OLS Optical Ports without removing adapters
AFL One-Click Cleaner method
1. Remove the protective dust cover from the tip of the One-Click Cleaner.
2. Insert the tip of the One-Click Cleaner into the optical port adapter and gently press the body of the
One-Click Cleaner until an audible “click” is heard.
3. Remove the One-Click Cleaner.
Cleaning OLS Optical Ports with Adapters Removed
1. Rotate the adapter base counterclockwise approximately four times.
2. Pull the adapter directly out away from the adapter mount.
3. Perform one of the following:
AFL FCC2 fluid and CCT stick method
•• Leaning the FCC2 can back (30°), press the button on the FCC2 to fill the well.
•• Dip the CCTP tip into the well of the FCC2 to dampen the tip with optical cleaning fluid.
36
•• Place the damp tip over the ferrule to be cleaned.
•• Rotate the tip clockwise 10 revolutions while applying varying pressure to create a gentle pumping
action where the tip contacts the ferrule.
•• Discard the CCTP stick after using both tips.
If using lint-free optical cleaning pads and isopropyl alcohol
•• Be sure to use 99% IPA that has not been contaminated.
•• Dampen the wipe with the alcohol and gently wipe the exposed ferrule. Then dry the ferrule using
a new optical wipe
4. Using a can of filtered compressed air (held vertically), blow out any contaminants from the adapter.
5. Once completed, replace the adapter making sure that the alignment pin on the adapter aligns with
the mating hole on the base.
Repair and Calibration
Repair of the Noyes test equipment in the field is NOT recommended.
Calibration is recommended every 12 months. Noyes Calibration Department is in compliance with ANSI/
NCSL Z540-1, ISO 10012-1, MIL STD 45662A, ISO Guide 25 and traceability to the National Institute
of Standards and Technology. Call Customer Service to obtain a Service Request (S.R.) number before
sending units in for calibration.
Battery Replacement
To replace a battery:
6. Remove the protective rubber boot from the instrument.
7. Remove the battery compartment cover located on the back of the instrument.
8. Replace the discharged battery.
9. Replace the battery compartment cover and rubber boot.
Mandrels for Multimode Transmit Jumpers
When testing multimode links using an overfilled LED source, always wrap the transmit jumper five (5)
times around the proper diameter mandrel. This is specified by TIA/EIA-568-B and will improve insertion
loss measurement repeatability and accuracy. Do NOT use mandrels on multimode receive jumpers or
single-mode jumpers.
Available mandrels for test
jumpers with 3 mm jackets
Diameter (D) for other
jacket diameters
Fiber type
Part #
(mm)
62.5µm
5400-00-0201 D = 20 mm minus
jacket diameter
50µm
5400-00-0202 D = 25 mm minus
jacket diameter
Kit with both 5400-00-0900
mandrels
Multimode transmit
jumper
D
(5) Wraps
Mandrel
37
Thank you for choosing Noyes Test & Inspection
16 Eastgate Park Road
Belmont, NH 03220
Phone:800-321-5298
603-528-7780
Fax: 603-528-2025
www.AFLtele.com/go/Noyes
N OY E S
ISO
9001
C E RT I F I E D