A&D AD-4212D-32 Instruction manual

Add to my manuals
54 Pages

advertisement

A&D AD-4212D-32 Instruction manual | Manualzz

AD-4212D-33

AD-4212D-32

AD-4212D-302

AD-4212D-301

Production Weighing Unit

INSTRUCTION MANUAL

1WMPD4003129

.

© 2016 A&D Company, Limited. All rights reserved.

No part of this publication may be reproduced, transmitted, transcribed, or translated into any language in any form by any means without the written permission of A&D

Company, Limited.

The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement without notice.

Windows and Excel are registered trademarks of Microsoft Corporation.

CONTENTS

1.

INTRODUCTION............................................................................................................................ 2

1.1.

Features ........................................................................................................................................ 2

1.2.

Compliance ................................................................................................................................... 5

2.

UNPACKING THE BALANCE....................................................................................................... 6

2.1.

Unpacking...................................................................................................................................... 6

2.2.

Installing the Balance .................................................................................................................... 8

3.

PRECAUTIONS ........................................................................................................................... 11

3.1.

Before Use ...................................................................................................................................11

3.2.

When Building Into a System ...................................................................................................... 12

3.3.

During Use .................................................................................................................................. 14

3.4.

After Use...................................................................................................................................... 14

3.5.

Power Supply .............................................................................................................................. 14

4.

BASIC OPERATON (WHEN USING THE AD-1691).................................................................. 15

4.1.

Changing the Weighing Speed Using AD-1691 .......................................................................... 16

4.2.

Calibration ................................................................................................................................... 16

4.3.

Calibration Using the Internal Weight.......................................................................................... 17

4.4.

Calibration Using An External Weight ......................................................................................... 18

5.

SMART RANGE FUNCTION....................................................................................................... 19

6.

UNDERHOOK.............................................................................................................................. 20

7.

FUNCTION TABLE...................................................................................................................... 21

7.1.

How to set(When using the AD-1691) .................................................................................... 21

7.2.

Details Of The Function Table ..................................................................................................... 22

7.3.

Description Of The Class "Environment, Display"....................................................................... 24

7.4.

Description Of The Item "Data Output Mode" ............................................................................. 25

7.5.

Description Of The Item "Data Format"....................................................................................... 26

7.6.

Data Format Examples................................................................................................................ 29

8.

STANDARD INPUT AND OUTPUT INTERFACE ....................................................................... 30

8.1.

RS-232C Interface....................................................................................................................... 30

8.2.

Commands .................................................................................................................................. 32

9.

MAINTENANCE........................................................................................................................... 38

9.1.

Treatment Of The Balance .......................................................................................................... 38

9.2.

Error Codes ................................................................................................................................. 38

9.3.

Checking The Balance Performance And Environment.............................................................. 40

9.4.

Asking For Repair........................................................................................................................ 40

10.

SPECIFICATIONS.................................................................................................................... 41

11.

EXTERNAL DIMENSIONS ...................................................................................................... 42

12.

OPTIONS.................................................................................................................................. 44

13.

TERMS/INDEX ......................................................................................................................... 46

1

This manual describes how the AD-4212D series balance works and how to get the most out of it in terms of performance.

Read this manual thoroughly before using the balance and keep it at hand for future reference.

1.1. Features

or glove boxes.

The weighing unit is compact.

Model

AD-4212D-33

Weighing capacity

32g

Minimum weighing value

0.001mg

AD-4212D-32 32g/5.1g

*2

0.01mg/0.001mg

*2

AD-4212D-302 320g 0.01mg

AD-4212D-301 320g 0.1mg

*1 With FAST selected under good environment

*2 Smart range (Refer to “5. SMART RANGE FUNCTION”.)

Stabilization time

*1

7 seconds

7 seconds

 The AD-4212D series can output the weighing digital data directly. Therefore, the AD-4212D series (the weighing unit) can be connected to a personal computer or a PLC directly.

 The AD-4212D series has a patented shock absorber under the weighing pan and can cope with movement in all directions, protecting the weighing unit from an actuator malfunction.

display of weighing values for the AD-4212D and manual operation of the re-zero and calibration functions. of AND-MEET can be saved in PDF format in USB memory. analog outputs.

 When connected to an optional remote controller, an AD-8923-BCD or AD-8923-CC, the remote controller can receive the weighing data and manually perform re-zero and calibration.

The AD-8923-BCD outputs the data using BCD and the AD-8923-CC outputs the data using CC-Link.

Note: CC-Link is a high-speed field network able to simultaneously handle both control and information data.

 When connected to an optional AD-8526 Ethernet converter, it can convert the AD-4212C

RS-232C data, for output to a local area network (LAN).

2

Example 1 (Connecting to a AD-1691, a personal computer or an AD-8922A)

When connecting to AD-1691 and PC

RS-232C cable for PC/PLC

(Sold separately: AX-KO4472-1000)

Connector for the PC/PLC

AD-1691

AD-4212D

RS-232C cable for weight indicator (Accessory: 2m)

Connector for A&D’s external display

USB converter

(Sold separately:

AX-USB-9P)

USB port

When only connecting the AD-4212D to a PC

Connector for the PC/PLC

AD-4212D

RS-232C cable for the PC/PLC

(Sold separately:

AX-KO4472-1000)

USB converter

(Sold separately:

AX-USB-9P)

Personal computer

USB port

When connecting the AD-4212D to the AD-8922A (External display) and PC

RS-232C cable for PC/PLC

Connector for the PC/PLC

(Sold separately: AX-KO4472-1000)

AD-8922A

(Sold separately)

AD-4212D

RS-232C cable for weight indicator (Accessory: 2m)

Connector for A&D’s external display

USB converter

(Sold separately:

AX-USB-9P)

Personal computer

Personal computer

USB port

3

Example 2 (Connecting to a PLC)

When connecting using the RS-232C

Connector for the PC/PLC

AD-4212D

D-sub 9-pin (female)

* Depending on the

PLC to be connected, a converter for the

D-sub 9-pin plug may be required.

RS-232C cable for the PC/PLC

(Sold separately:

AX-KO4472-1000)

PLC

When connecting using the BCD

AD-4212D

Connector for A&D’s external display

AD-8922A-01

Connection cable*

BCD input

AD-8922A

(Sold separately)

PLC

BCD output board

(Sold separately:

AD-8922A-01)

BCD cable

(Fabricate locally)

* Specify the AD-8922A-01 connection cable when ordering the AD-8922A-01.

AX-KO3705-200 2 m

AX-KO3705-500 5 m

AX-KO3705-1000 10 m

When connecting using the CC-LinK

Connector for A&D’s external display

AD-4212D

RS-232C cable for the PC/PLC

(Sold separately: AX-KO4472-1000)

AD-8923-CC

When connecting to a LAN

To other remote stations

Connector for the PC/PLC

AD-4212D

CC-Link network

RS-232C cable for PC/PLC

(Sold separately:

AX-KO4472-1000)

PLC

LAN port

AD-8526

(Sold separately)

LAN cable

PLC

4

1.2. Compliance

Compliance with FCC Rules

Please note that this equipment generates, uses and can radiate radio frequency energy. This equipment has been tested and has been found to comply with the limits of a Class A computing device pursuant to Subpart J of Part 15 of FCC rules. These rules are designed to provide reasonable protection against interference when the equipment is operated in a commercial environment. If this unit is operated in a residential area, it may cause some interference and under these circumstances the user would be required to take, at his own expense, whatever measures are necessary to eliminate the interference.

(FCC = Federal Communications Commission in the U.S.A.)

Compliance With Directives of CE mark

This device features radio interference suppression, safety regulation and restriction of

Hazardous Substances in compliance with the following Council Directives

Council directive 2004/108/EC EN61326 EMC directive

Council directive 2006/95/EC EN60950 Safety of Information Technology Equipment

Council directive 2011/65/EU EN50581 Restriction of the use of certain Hazardous Substances

The CE mark is an official mandatory European marking.

Please note that any electronic product must comply with local laws and regulations when sold or used anywhere outside Europe..

5

2. UNPACKING THE BALANCE

2.1. Unpacking

 The balance is a precision instrument. Unpack the balance carefully. Keep the packing material to be used for transporting the balance in the future.

 When unpacking, see the illustration below to confirm that everything is contained.

AD-4212D-33

AD-4212D-32

Weighing pan

AD-4212D-302

AD-4212D-301

RS-232C connector

(For PC,PLC)

RS-232C connector

(For A&D's external display)

Screws

(Flat head:

M3 X 8)

Pan support

Fine range ring

Grounding terminal

(See note below.)

Pilot lamp

Pan support boss

Dust plate

Bubble spirit level

Weighing unit

Waterproof cap for cleaning

Leveling foot

Accessories

AC adapter

*

AC adapter

ID label

Rear view

AC adapter jack

Arrow mark face up

AC adapter plug

Sold separately:

AX-KO4472-1000

RS-232C cable for the PC/PLC (10 m)

Small breeze break

RS-232C cable for the weight indicator (2 m)

AX-KO3590-1000

Weighing pan

Pan support

Screws (Flat head: M3 X 8)

Power plug stopper

* Please confirm that the AD adapter type is correct for your local voltage and

receptacle type.

RS-232C cable for the weight indicator

(2 m)

Note

When the AD-4212D series weighing unit is built into a system, be sure to earth ground the weighing unit using the grounding terminal.

6

ATTACHING THE AD-4212D SERIES WEIGHING UNIT DIRECTLY ON THE MOUNTING BASE

When the AD-4212D series weighing unit is built into a system, remove the three leveling feet and use the screw holes to secure the weighing unit to the base.

Attachment Procedure

1. Remove the three leveling feet.

Weighing unit

Caution

Do not remove any other screws from the bottom of the weighing unit. Doing so could damage the weighing sensor.

Leveling foot

2. Secure the weighing unit from above.

(Prepare appropriate screws with a size corresponding to an M5 screw. The screw hole diameter: 6 mm).

In case of securing the weighing unit from the under side, use M6 screws with 1-mm

M5 screw x 3 pitch.

Notes

The screws to secure the weighing unit to the mounting base are not provided.

For the position of the holes, refer to “11.

EXTERNAL DIMENSIONS”.

M6 screw x 3

There are cases when an error in the leveling adjustment occurs due to the installation conditions. In this case, place a shim between the lower surface of the weighing unit and the securing surface, or use the double-nut method using M5 or M6 nuts to perform the height adjustment and secure it. If the adjustment is still difficult to perform, perform calibration before use. Then, the balance will function normally.

7

2.2. Installing the Balance

Install the balance as follows:

1. Refer to “3. PRECAUTIONS” for installing the balance.

2. Refer to "2. UNPACKING THE BALANCE" on page 6, to attach the pan support and the weighing pan on the weighing unit.

3. Insert the RS-232C cable into the jack located on the rear of the weighing unit.

Rear of the weighing unit

Sold separately:

AX-KO4472-1000

D-sub 9-pin

(PC, PLC)

D-sub 9-pin

A&D's external display

(AD-1691, AD8922A)

RS-232C cable

Arrow mark face up

How to disconnect the cable from the weighing unit

Slide the connector sleeve in the direction of the arrow to unlock and gently pull the connector out.

Slide the connector sleeve in the direction of the arrow

Unlocked

8

4. Follow the numerical sequence to assemble the small breeze break.

*

* The cover's hole is not positioned in the center of the cover. Attach the cover on the breeze break so that the hole is positioned just above the weighing pan.

Weighing pan

5. Confirm that the AC adapter type is correct for the local voltage and power receptacle type.

6. Plug the AC adapter plug into the AC adapter jack located on the rear of the weighing unit and plug the AC adapter into the electrical outlet. Warm up the balance for one hour or more with nothing on the weighing pan.

How to use the power plug stopper

1) Remove the screw attached on the main body.

2) Insert the power plug stopper to the slit at the thickest part of the AC adapter plug.

3) Insert the AC adapter plug into the AC adapter input jack and secure the power plug stopper to the main body with a screw.

1

2

3

3

9

7. Set the pan unit and I/O unit to adapt to the peripheral system. Set the following for the I/O unit.

 RS-232C (Refer to“8. STANDARD INPUT AND OUTPUT INTERFACE”)

Set the weighing speed to adapt to the ambient conditions.

Note

Make sure that the personal computer is in operation before connecting to the AD-4212D.

8. After the balance has been installed, calibrate the balance using the internal weight or external calibration weight. For details, refer to “4.2. Calibration”.

9. By removing the provided weighing pan, a locally fabricated weighing pan can be attached to the

AD-4212D series by using the four screw holes (M3 x 8) on the pan support.

MASS OF THE LOCALLY FABRICATED WEIGHING PAN

Design the weighing pan so that the mass falls within the ranges shown in the table below:

Mass of

AD-4212D-33/32

AD-4212D-302/301

Pan support is used (with weighing pan removed)

Pan support boss is used (with the weighing pan and pan support removed)

Pan support is used (with weighing pan removed)

Pan support boss is used (with the weighing pan and pan support removed)

AD-4212D-33/32

When the weighing pan is removed

When the weighing pan and pan support are removed

When the weighing pan is removed

10 to 21

35 to 46

20 to 230

70 to 280

AD-4212D-302/301

When the weighing pan and pan support are removed

32 g

Mass range to be weighed

21 g

32 g

Mass range to be weighed

21 g

320 g

Mass range to be weighed

110 g

320 g

Mass range to be weighed

110 g

10 g

21 g

35 g

Mass of locally fabricated pan

46 g

20 g

Mass of locally fabricated pan

230 g

70 g

Mass of

locally fabricated pan

280 g

Notes

If the balance is to be used in a range other than shown above, contact the local A&D dealer.

To avoid the effects of static electricity or magnetism, use materials other than resin or magnetic materials when designing a weighing pan.

When a locally fabricated weighing pan is used, the zero point may be shifted greatly right after the AC adapter is plugged in. Perform re-zero before weighing to cancel the amount of zero drift as necessary.

10

To get the optimum performance from the balance and acquire accurate weighing data, note the following:

3.1. Before Use

 Install the weighing unit in an environment where the temperature and humidity are not excessive. The best operating temperature is about 20°C / 68°F at about 50% relative humidity.

 Install the weighing unit where it is not exposed to direct sunlight and it is not affected by heaters or air conditioners.

 Install the weighing unit where it is free of dust.

 Install the weighing unit away from equipment which produces magnetic fields.

 Install the weighing unit in a stable place avoiding vibration and shock. Corners of rooms on the first floor are best, as they are less prone to vibration.

 Ensure a stable power source when using the AC adapter.

 Connect the power to the balance using the AC adapter at least one hour before use, but in the case of the AD-4212D-33/32, four hours are required. It is recommended to connecting the power to the balance.

 Calibrate the balance before use or after having moved it to another location.

In addition, calibrate it periodically to maintain the accuracy.

Caution

Do not install the balance where flammable or corrosive gas is present.

11

3.2. When Building Into a System

Errors due to a static charge and measures to take

When the ambient humidity is less than 45% RH, insulators such as plastic or glass are prone to static electricity. When charged material comes close to the balance, a pulling force is generated by static induction between the charged material and the weighing pan. This causes an unstable weight value.

To protect the balance against a discharge generated by charged material when it comes close to the balance, be sure to earth ground the weighing unit and the display unit.

Note that static electricity generated by static induction will not be canceled by earth-grounding.

Use a humidifier or the AD-1683 DC static eliminator,

Charged plastic breeze break

The pulling force causes an unstable weight value

Charged sample

Weighing unit

A pulling force generated between the charged sample and the balance causes an unstable weight value.

Measures to take (Plastic is used in the examples below. They can be applied to glass, too.)

When the sample or devices are plastic

 Use a static eliminator that generates no air blow such as the

AD-1683, DC static eliminator, to remove static electricity.

 Place the sample in a container that is made of a conductive material such as metal and that can be sealed, and then weigh it.

Charged material

Metal case

When the sample is powdery

 When the balance is used in combination with a feeder for batch weighing of powdery samples, samples may be charged by rubbing sample particles against each other.

Use a static eliminator and perform weighing while removing static electricity.

Grounding

When the sample container is made of material that is prone to static electricity such as plastic

 Cover the outside of the container with a metal such as aluminum foil.

Apply an anti-static agent onto the container.

When making a breeze break using plastic

 Apply an anti-static agent onto the breeze break.

Use a conductive acrylic fiber.

When plastic exists in the balance installation site

 Cover the plastic with a grounded conductive material.

 Apply an anti-static agent onto the plastic.

When an operator is static charged

If an operator’s clothes are static charged, especially in winter, it may be a cause for unstable weight values.

12

 Wear an anti-static wrist strap.

Errors due to air flow and measures to take

Where the influence of ambient air flow is great such as: close to an air conditioner or fan, door or passage way. Even very subtle air flow that is hard to be detected may influence the weighing operation.

 Avoid those areas as a weighing site.

 If weighing is to be performed in such an area, use a breeze break or take other appropriate measures.

Influence of the air flow makes the weight value unstable

Where the influence of heat or drafts is great

 Eliminate temperature differences between a sample and the environment.

When a sample is warmer (cooler) than the ambient temperature, the sample will be lighter (heavier) than the true weight.

This error is due to a rising (falling) draft around the sample.

 Do not touch the sample directly with your hand. Use tweezers or other tools.

If you touch the sample, the same type

Draft

20 C

40 C error described above will occur.

 Do not perform weighing where it is exposed to direct sunlight. Weighing errors may occur due to sudden temperature change or drafts.

Errors due to vibration and measures to take

Where the influence of vibration is great, such as:

(1) Soft ground (2) Second or higher floor (3) Near center of a floor far from pillars (4) Seismic isolated structures (5) Near tall buildings.

In the areas listed above, the scale may yield unstable weight values on windy days or after an earthquake. Especially in case of (4) and (5), weight values may be unstable for a long period of time because of long lasting low-frequency vibration after strong winds or an earthquake.

Errors due to other causes and measures to take

Change in temperature or humidity

A sudden change in temperature or humidity can generate a draft and cause the balance to absorb or exude moisture, which leads to unstable displays.

 Avoid sudden change in temperature or humidity.

 Use an air conditioner or humidifier to control the temperature or humidity.

Magnetic material

The balance uses a strong magnet as part of the balance assembly, so use much care when weighing magnetic materials.

 Place a non-magnetic object such as aluminum or brass between the sample and the balance to keep an appropriate distance between them while weighing.

13

3.3. During Use

 To minimize the affect by electrical noises, earth ground the weighing unit and the display unit (option).

 Do not drop things upon the weighing pan, or place a sample on the pan that is beyond the balance weighing capacity. Place a sample in the center of the weighing pan to minimize corner-load errors.

 To prevent possible errors, before each weighing, perform re-zero using the RS-232C command, or calculate the difference between the weight value before and after weighing.

 Take into consideration the affect of air buoyancy on a sample when measuring the absolute value with the minimum weighing value of 0.1 mg.

 While cleaning, attaching the waterproof cap to the unit instead of the weighing pan results in the AD-4212D becoming dust-proof and drip-proof and conforming to IP65. Be sure to clean the weighing pan and keep the conditions around the pan clean after weighing powdery, fluid samples or metallic strips.

 Balance status is indicated by the pilot lamp.

Lamp status

Turned off

Balance status

The power with the balance is not turned on.

Red When turning on the power, when an error has occurred, or during a re-zero or calibration

3.4. After Use

 Avoid mechanical shock to the weighing unit.

Periodically calibrate the balance using the internal weight or a calibration weight.

 Do not disassemble the weighing unit. Contact the local A&D dealer if the balance needs service or repair.

 Do not use organic solvents to clean the weighing unit. Clean the weighing unit with a lint free cloth that is moistened with warm water and a mild detergent.

 Avoid dust and water so that the weighing unit weighs correctly. Protect the internal parts from liquid spills and excessive dust.

3.5. Power Supply

 When the AC adapter is connected, the balance is in the standby mode if the standby indicator is on (refer to “4. BASIC OPERATON (WHEN USING THE AD-1691)”). This is a normal state and does not harm the balance. For accurate weighing, plug in the AC adapter and warm up the balance for the appropriate duration before use.

14

4. BASIC OPERATON (WHEN USING THE AD-1691)

Using the provided AD-1691 weighing environment analyzer as a remote controller enables performing the following functions.

・ Weighing value display

Calibration

・ Changing of the minimum display digit

Changing of the weighing speed (Response)

・ Re-zero

Preparations before use

1. Connect the RS-232C connector for the A & D's external display to the AD1691's arbitrary

RS-232C connector using the cable.

* Factory setting of the AD1691 is set to COM1.

2. Connect the AD-4212D to the power using the AC adapter to turn the power on and put the

AD-4212D to weighing mode.

3. Connect the AD-1691 to the power using the AC adapter and turn the power on.

4. Perform functions as the balance weighing display from the home screen.

5. Touch the Start button to display the weighing value.

* AD-4212D weighing indicator RS-232C has a fixed 2400bps baud rate with parity-even and data bits-7 bits.

Descriptions for the display and buttons

Stabilization indicator

Weighing value

Each button

Button Function

ON:OFF Toggles between the weighing mode and standby mode.

CAL

EXC

Performs the calibration using the internal weight.

Performs the calibration using the calibration weight.

MODE Changes the weighing speed (Response speed).

RANGE / SAMPLE Changes the minimum display digit.

ENTRE(PRINT)

RE-ZERO

Output

ION

Outputs the weighing value. (COM port that is set at the external output port)

Sets the display value to zero.

Outputs the current weighing value.

Has no function.

15

4.1. Changing the Weighing Speed Using AD-1691

The weighing speed can be selected from the following three rates to minimize the influence on weighing that is caused by drafts and/or vibration at the place where the balance is installed.

Indicator Speed Stability

Weighing speed setting

FAST

MID.

SLOW

Fast Sensitive value

Slow Stable value

Operation

1. Touch the MODE button to select a weighing speed.

Either FAST, MID. or SLOW can be selected.

MODE

Touch

MODE

Each touching switches the indicators

4.2. Calibration

Calibration

Calibration using the calibration weight.

Caution

 Do not allow vibration or drafts to affect the balance during calibration.

Caution on using an external calibration weight

 The accuracy of the weight can influence the accuracy of weighing. Select an appropriate weight as listed below.

Model

AD-4212D-30/32

Usable calibration weight

1g, 2g, 5g, 10g, 20g, 30g

AD-4212D-302/301

10g, 20g, 50g, 100g, 200g, 300g

Display

The calibration weight in bold type: factory setting

 This indicator means “the balance is measuring calibration data”. Do not allow vibration or drafts to affect the balance while this indicator is displayed.

16

4.3. Calibration Using the Internal Weight

This function calibrates the balance using the internal weight.

1 Connect the AC adapter and warm up the balance for at least one hour with nothing on the weighing pan.

2 Touch the CAL key to display Cal in .

3 The balance performs calibration using the internal weight.

Do not allow vibration or drafts to affect the balance.

4 The balance will automatically return to the weighing mode after calibration.

CAL

17

4.4.

Calibration Using An External Weight

Calibration procedure

This function calibrates the balance using the calibration weight. (Display example: AD-4212D-301)

Operation

1. Plug in the AC adapter and warm up the balance for 30 minutes or more with nothing on the pan.

2. Touch the EXC button. Cal 0 is displayed.

 If you want to cancel calibration, touch the CAL button. The balance will return to the weighing mode.

 If you want to change the calibration weight value, press the

SAMPLE button. Touch the RE-ZERO button to select the weight value, and touch the ENTER button to store it. Cal 0 1 is displayed.

3. Confirm that there is nothing on the pan and touch the ENTER button. The balance measures the zero point. Do not allow vibration or drafts to affect the balance.

The balance displays the calibration weight value.

EXC

ENTER

4. Place a calibration weight, of the weight value displayed, on the pan and touch the ENTER button. The balance measures the calibration weight. Do not allow vibration or drafts to affect the balance.

5. The balance displays end . Remove the weight from the pan.

6. The balance will automatically return to the weighing mode.

7. Place the calibration weight on the pan and confirm that calibration was performed correctly. If not, check the ambient conditions such as drafts or vibration, and repeat steps 2 through 7.

ENTER

18

The AD-4212D-32 is equipped with two ranges. The precision range has a higher resolution. The standard range has a normal resolution.

The range switches automatically depending on the value displayed. Touching the RE-ZERO button or sending the R command allows weighing in the precision range regardless of the tare value. The range can be fixed to the standard range by changing the minimum weighing value using the

RE-ZERO button or R command.

Weighing pan

Precision range

1 Touch the RE-ZERO button or send the R command.

The balance will start weighing, using the precision range.

Container

Standard range

2 Place a container on the weighing pan.

The weight value exceeds the precision range value and the balance will switch to the standard range.

Precision range

3 Touch the RE-ZERO button or send the R command.

The balance will switch to the precision range.

4 Place a sample in the container.

The weight value is within the precision range value and the balance will perform a weighing, using the precision range.

Sample

Precision range

Precision range/standard range

Model

AD-4212D-32

Range used

Precision range

*

Standard range

Weighing range

5.1 g to 32 g

Available minimum weighing value

0.000001g (0.001mg)

0.00001g (0.01mg)

 The factory setting is the precision range.

* When the weight value exceeds 5.1 g while the minimum weighing value of 0.000001 g

(precision range) is used, the balance will switch to the minimum weighing value of

0.00001 g (standard range). By touching the RE-ZERO button or sending the R command, the balance can weigh from that point up to 5.1 g, using the minimum weighing value of

0.000001 g (precision range).

19

6. UNDERHOOK

The underhook can be used for magnetic materials or density measurement. The built-in underhook is revealed by removing the cap on the bottom of the balance. Use the underhook as shown below.

Caution

Do not apply excessive force to the underhook.

When not in use, attach the cap to prevent dust from getting into the balance.

Do not push the underhook upward

When turning the balance over, parts and weighing matter in the weighing chamber will fall off. (Parts: the weighing pan and fine range ring)

Remove them first.

Underhook

When the cap is removed, a weighing error may occur, because of drafts entering into the internal

Cap

portion of the balance. Arrange the room condition.

How to use

1 Remove the cap on the bottom of the balance.

2 Hang your weighing pan on the underhook.

Place the balance on the solid table.

Sturdy table equipped with a hole.

Weighing pan

Bottom

20

The function table reads or rewrites the parameters that are stored in the balance.

These parameters are maintained in non-volatile memory, even if the AC adapter is removed.

The function table menu consists of two layers. The first layer is the "Class" and the second layer is the "Item".

7.1. How to setWhen using the AD-1691

Using the provided AD-1691 weighing environment analyzer as a remote controller enables changes or confirmation of function settings.

1. Connect the RS-232C connector for the A&D's external display to the AD-1691's arbitrary

RS-232C connector using the cable.

2. Connect the AD-4212D to the power using the AC adapter to turn on the power and have the

AD-4212D enter the weighing mode.

3. Connect the AD-1691 to the power using the AC adapter and turn the power on.

4. Perform the balance function settings at the AD-1691.

5. Touch the Take Data button to start communication between the AD-1691 and AD-4212D.

6. The screen shows the function settings list.

7. Click the arrow mark to change the setting values.

* The changed contents have yet to be confirmed at this step.

8. Touch the Confirm button to start the communication and change the balance function settings.

The device will be restarted when the setting changes were ended.

21

7.2. Details Of The Function Table

Class Item and Parameter Description

0 Fast response, sensitive value FAST

Cond

Condition

2 Slow response, stable value

Can be changed by

SLOW adjustment. ba5fnc

Environment

Display dout

Data output

#4

5t-b

Stability band width

0 Stable range is ±1 digit

The stabilization indicator digit

2 Stable range is ±3 digits fluctuation is within the range. ploc

off OFF

Display lock function

on

Display hold function to measure the volume dispensed from the pipette. trc

Zero tracking

0 OFF

1 Normal

2 Strong

3 Very strong

5pd

0 5 times/second #3

Display refresh rate

1 10 times/second #3

Keeps zero display by tracking zero drift.

Period to refresh the display pnt

Decimal point

0 Point (.)

1 Comma (,)

0 Normal mode

Decimal point format

Outputs data by commands. prt

Data output mode ap-p

Auto print polarity

1

Auto print mode A

Reference = zero

2

Auto print mode B

Reference = last stable value

3

Stream mode /

Interval mode

0 Plus only

1 Minus only

Outputs data when the display is stable and conditions of ap-p, ap-b and the reference value are met.

2 Both

Displayed value > Reference

Displayed value < Reference

Regardless of displayed value ap-b

Auto print difference

0 10 digits

1 100 digits

2 1000 digits

Difference between reference value and displayed value

#2

: Factory settings.

#1 The unit of minimum display is digit.

Example: If 1 mg display is selected touching the RANGE button for the AD4212D-301, 1 mg is one digit.

#2 Usable minimum display of the balance is one digit.

Example: In gram display, one digit is 0.00001 mg for the AD4212D-302 and 0.0001 mg for the

AD4212D-301.

#3 Factory settings of the AD4212D-33 and AD4212D-32 is set to 5 times per second.

Factory settings of the AD4212D-302 and AD4212D-301 is set to 10 times per second.

#4 These items apply only for RS-232C output for the PC or PLC.

22

5if

Serial interface

#4

Class dout

Data output

#4

5if

Serial interface

#4

Item and Parameter

0 Every measurement

1 2 seconds

2 5 seconds int

Interval time

3 10 seconds

4 30 seconds

5 1 minutes

6 2 minutes

7 5 minutes

8 10 minutes

0 No output

1 Time only

5-td

Time/Date output at-f

Auto feed

2 Date only

3 Time and date

5-id

off No output

ID number output on pU5e

off No pause

Data output pause on

off Not used ar-d

off Not used

Zero after output on

5-tp

off Not output

Temperature output on bp5

Baud rate

0 600 bps

1 1200 bps

2 2400 bps

3 4800 bps

4 9600 bps

5 19200 bps btpr

Data bit, parity bit

Crlf

Terminator type

Data format

0 7 bits, even

1 7 bits, odd

2 8 bits, none

0 CR LF

1 CR

0 A&D standard format

1 DP format

2 KF format t-Up

Timeout

3 MT format

4 NU format

5 CSV format

off No limit

on 1 second erCd

AK, Error code

off No output

on Output

: Factory settings.

Description

Interval time in the interval memory mode when using prt 3

Selects whether or not the time or date is added to the weighting data.

Selects whether or not the ID number is output.

Selects the data output interval.

Selects whether or not automatic feed is performed.

Adjusts zero automatically after data is output

Adds temperature to data output

CR: ASCII code 0Dh

LF: ASCII code 0Ah

Refer to "

7.5. Description Of The

Item "Data Format" ".

Selects the wait time to receive a command.

AK: ASCII code 06h

23

Caution The balance may not transmit the data completely at the specified refresh rate, depending on the baud rate or data added to the weighing data such as time, date and ID number.

#4 These items apply only for RS-232C output for the PC or PLC.

7.3. Description Of The Class "Environment, Display"

Condition ( Cond )

Cond 0

This parameter is for sensitive response to the fluctuation of a mass value. Used for powder target mass, weighing a very light sample or when quick response weighing is required.

Cond 2 This parameter is for stable weighing with slow response. Used to prevent a mass value from drifting due to vibration or drafts.

Notes In automatic response adjustment, this parameter is selected automatically.

Stability band width ( 5t-b )

This item controls the width to regard a mass value as a stable value. When the fluctuation per second is less than the parameter, the balance displays the stabilization indicator and outputs.

The parameter influences the "Auto print mode". Currently displayed minimum display means 1 digit.

Example) If the RANGE button was touched to select the 1 mg display at the AD4212D-301, 1 mg means 1 digit.

5t-b 0 This parameter is used for sensitive response of the stabilization indicator. Used for exact weighing.

5t-b 2 This parameter ignores slight fluctuations of a mass value. Used to prevent a mass value from drifting due to vibration or drafts.

Zero tracking ( trc )

This function tracks zero point drift caused by changes in the environment and stabilizes the zero point. When the weighing data is only a few digits, turn the function off for accurate weighing. trc 0 The tracking function is not used. Used for weighing a very light sample. trc 1 The normal tracking function is used. trc 2 The strong tracking function is used. trc 3 The very strong tracking function is used. Used for stable zero display.

Display refresh rate ( 5pd )

The periodic time to refresh the display. This parameter influences "Baud rate", "Data output pause" and the data output rate of "Stream mode".

Decimal point ( pnt )

The decimal point format can be selected.

24

7.4. Description Of The Item "Data Output Mode"

The parameter setting of "Data output mode (prt)" applies to the performance when the data is transmitted using the RS-232C interface

.

Normal mode

The output is performed only when receiving a command such as a Q command.

Required mode

Auto print modes A and B

When the displayed value is stable and the conditions of "Auto print polarity", "Auto print difference" and reference value are met, the balance outputs the weighing data once.

Auto print mode A

Example For weighing each time a sample is placed and removed, with "ar-d" set to "1" (to adjust zero after the data is output). setting prt 1

Auto print mode A (reference = zero) dout ap-p Auto print polarity dout ap-b

Auto print difference dout ar-d on Zero after output

Auto print mode B

Example For weighing while a sample is added. setting prt 2

Auto print mode B (reference = last stable value) dout ap-p Auto print polarity

Stream mode

dout ap-b

Auto print difference

The balance outputs the weighing data continuously regardless of the display condition. When the display refresh rate is set to 5 times / second (5pd 0), the data output rate is also set to the same 5 times / second. The display does not blink in this mode.

Example For monitoring data on a computer. setting mode ba5fnc 5pd Display refresh rate

5if

Caution The balance may not transmit the data completely at the specified refresh rate, depending on the baud rate or time, date and ID number appended to the weighing data.

Interval mode

The weighing data is periodically output.

Example In case of monitoring the weighing value periodically, the time and date can be added to the weighing data depending on settings of 5-td.. setting mode dout int

time

25

7.5. Description Of The Item "Data Format"

* The output data format applies only to RS-232C output for the PC or PLC.

A&D standard format

This format is used when the peripheral equipment can receive the A&D format.

This format consists of fifteen or sixteen characters excluding the terminator.

5if type 0

When numerical characters without decimal point are eight characters for the AD-4212D-30,

AD-4212D-32 and AD-4212D-302, the format becomes sixteen characters.

A header of two characters indicates the balance condition.

The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus sign is applied.

The unit, consisting of three characters, follows the data.

Header Data

Stable header

Unstable header

Overload header

DP (Dump print) format

Stable header of counting mode

Header Data Unit

5if type 1

This format consists of sixteen characters excluding the terminator.

A header of two characters indicates the balance condition. No overload header is used.

The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is not zero or overloaded.

The unit, consisting of three characters, follows the data.

Header Data

Stable header

Unstable header

Stable header of counting mode

KF format

5if type 2

This is the Karl-Fischer moisture meter format and is used when the peripheral equipment can only communicate using this format.

This format consists of fourteen characters excluding the terminator.

This format has no header characters.

The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is not zero or overloaded.

This format outputs the unit only for a stable value.

Stable value

Unstable value

26

MT format

5if type 3

This format is used when the peripheral equipment of other manufacturer is connected. Please note that the connection is not guaranteed.

A header of two characters indicates the balance condition.

The polarity sign is used only for negative data.

The weighing data uses spaces in place of the leading zeros.

The character length of this format changes dependent upon the unit

Header Data

Stable header

Unstable header

Overload header

NU (numerical) format

This format outputs only numerical data.

5if type 4

This format consists of ten characters excluding the terminator.

The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus sign is used.

CSV format

Data Terminator

5if type 5

This format separates the data of A&D standard format and the unit by a comma ( , ).

This format outputs the unit even when the data is overloaded.

When a comma ( , ) is selected for the decimal point, the separators are set to semicolon ( ; ).

When the ID number, time and date are added at "Data output (dout)" of the function table, outputs ID number, date, time and weighing data in this order and separates each item by a comma and treats all the items as one group of data.

LAB-0123, 2004/07/01, 12:34:56, ST,+0000.1278,

ID number Date Time Weighing data

S T , + 0 0 0 . 1 2 7 8 , g CR LF

O L , + 9 9 9 9 9 9 9 E + 1 9 ,

ID number

The number to identify a specific balance.

This format consists of eight characters excluding the terminator.

~~ g

g<CR><LF>

CR LF dout 5-id on

Date

The date output order can be changed in " Time/Date output (5-td)" .

The year is output in a four-digit format.

2 0 1 5 / 0 9 / 0 1

CR LF dout 5-td 2 or 3

27

Time

This format outputs time in 24-hour format.

Temperature

This format outputs temperature. The unit is "C".

Caution Humidity and air pressure are not output.

dout 5-td 1 or 3 dout 5-tp on

28

7.6. Data Format Examples

Stable

Unstable

Overload

Positive error

Overload

Negative error

Space, ASCII 20h

Carriage Return, ASCII 0Dh

Line Feed, ASCII 0Ah

29

8. STANDARD INPUT AND OUTPUT INTERFACE

8.1. RS-232C Interface

The balance is a DCE device. Connect the balance to a personal computer (DTE) using a straight through cable.

Transmission system

Transmission form

Transmission rate

Data format

: EIA RS-232C

: Asynchronous, bi-directional, half duplex

: 10 times/second or 5 times/second (same as data refresh rate)

: Baud rate : 600, 1200, 2400, 4800, 9600, 19200bps

Data bits : 7 or 8 bits

Parity : Even, Odd (Data bits 7 bits)

Stop bit

None

: 1 bit

(Data bits 8 bits)

-5V to -15V

+5V to +15V

Pin assignments

Data bits

Start bit

Stop bit

Parity bit

Pin assignments of the RS-232C communication cable for PC or PLC (D-Sub 9 pin side)

Pin No. Signal name Direction

1 - -

Description

No connection D-Sub 9 pin plug of RS-232C cable for the PC or PLC

3

4

TXD (RXD)

-

Input Receive data

5 4 3 2 1

9 8 7 6

6 DSR Output Data set ready

#4-40UNC

9 -

Pin

names mean one for the DTE side.

Balance side

DCE

+5V

Computer side

DTE

RTS 7

RTS

CTS 8

CTS

DSR 6

DSR

SG 5

SG

Inside of balance

Terminal example of the PL or PLC side

30

Connection to a computer using the RS-232C Interface

Features

The balance is of the DCE type (Data Communication Equipment), which can be connected to a personal computer using the RS-232C interface. Before connection, read the personal computer manual thoroughly.

Using Windows Communication Tools Software (WinCT)

When Windows is used as an operating system in a personal computer, the WinCT software can be used to transmit the weighing data to the personal computer.

The WinCT software has three communication methods: "RsCom", "RsKey" and "Rsweight ".

Refer to the WinCT instruction manual.

The current version of the WinCT can be downloaded from the A&D website.

RsCom

RsCom can transmit commands to control the balance.

RsCom can make bi-directional communication between the balance and a personal computer using the RS-232C interface.

RsCom can display or store the data using a text file format. RsCom can also print the data using a printer connected to the personal computer.

When several ports of a personal computer have balances connected, the personal computer can communicate with each balance simultaneously.

RsCom can share a personal computer with other application software.

RsKey

RsKey can transmit the weighing data output from the balance directly to other application software such as Microsoft Excel.

RsKey can be used with most application software.

RsWeight

Can retrieve the weight data from the balance and display the data as a graph in real time.

Can calculate and display the maximum, minimum, average, standard deviation and coefficient of variation.

Using the WinCT software, the balance can do the following :

Analyzing the weighing data and the statistics input by "RsKey"

The weighing data can be input directly into an Excel worksheet. Then, Excel can analyze the data to obtain total, average, standard deviation, maximum and minimum value, and display them in a graph.

Controlling the balance using commands from a personal computer

By using "RsCom", the personal computer sends commands such as "re-zero" or "send weighing data" to the balance and controls the balance.

Receiving weighing data at a certain interval

The weighing data can be received at a certain interval and data characteristic with elapsed time can be obtained.

Using a personal computer as an external indicator

With the "RsKey" test mode function, a personal computer can be used as an external display for the balance. (To do this, set the balance data output mode to stream mode.)

31

8.2. Commands

Command List

Note A command has a terminator added, that is specified using "5if, Crlf" of the function

table, and is sent to the balance.

Commands to query weighing data

C

Cancels the S or SIR command.

Q

Requests the weighing data immediately.

S

SI

SIR

Requests the weighing data when stabilized.

Requests the weighing data immediately.

Requests the weighing data continuously.

Commands to control the balance

CAL

Performs calibration using the internal weight.

EXC

Performs calibration using the external weight.

OFF

Goes into standby mode. (Weighing value output is not performed)

ON

Goes into weighing mode from standby mode.

P

Switches between weighing mode and standby mode.

PRT

Confirms operations at the calibration etc.

R

Performs re-zero. #1

RNG

Switches the minimum display.

TR

Tares the balance. #1

U

Changes the weighing speed.(Response speed)

Commands to request data

?ID

Outputs the identification number.

?PT

Outputs the tare value. #1

?SN

Outputs the serial number of the balance.

?TN

Outputs the model name of the balance.

?TM

Outputs time.

?DT

Outputs date.

Command to set data

PT:. g

Sets the tare value. #1

TM:::

Sets time.

Example)

12:34:56

TM: 12/34/56

DT://

ID://

Sets date.

Example) October 1, 2015

DT: 15/10/01

Sets the ID number.

#1 Command

Command

The canceled tare value can be read using the command ?PT.

32

Command

An example of this command sets a negative target value and fills with the sample until the display becomes zero. For details, refer to page 37.

Step 1 Place a container.

Step 2 Set the display to zero using the R command.

Step 3 Set a target value using the PT: command.

33

Acknowledge Code And Error Codes

When the "Serial interface function (5if)" parameter is set to "erCd on", the balance outputs

<AK> code or error code to each command as follows:

<AK> (06h) Acknowledge in ASCII code.

When the balance receives a command to request data and can not process it, the balance transmits an error code (EC, Exx).

When the balance receives a command to request data and can process it, the balance outputs the data.

When the balance receives a command to control the balance and can not process it, the balance transmits an error code (EC, Exx).

When the balance receives a command to control the balance and can process it, the balance transmits the acknowledge code.

<AK> (06h) Acknowledge in ASCII code.

Among commands to control the balance, the following transmit the acknowledge code both when the balance receives the command and when the balance has accomplished the command.

If the command can not be processed properly, the balance transmits an error code (EC, Exx).

This error can be released using the CAL command. xx is error code number.

CAL command

(Calibration command using internal weight)

EXC command

ON command

P command

R command

(Display ON command)

(Display ON/OFF command)

(RE-ZERO command)

TR command

(Tare command)

When a communication error has occurred due to external noise, or a parity error has occurred due to transmission error, the balance transmits an error code (EC, E xx

). In this case, send the command again. xx

is error code number.

Settings Related To RS-232C

Concerning the RS-232C, the balance has two functions: "Data output (dout)" and "Serial interface

(5if)". Set each function as necessary.

34

Command Examples

This example uses the "erCd on" of "5if" so that the <AK> (06h) code is output.

ON command (turning on the balance)

Computer Time

ON command

Balance

Standby mode

Command received

All segments

All segments are displayed

Command completed

Zero display

R command (Re-zeroing the display)

Computer Time

R command

Balance

Weighing mode

Command received

In process

Command completed

Zero display

CAL command (Calibration with internal weight)

Computer Time

CAL command

Balance

Zero display

Command received

Command completed

Zero display

35

Error code

This example is of an error using the R command. "erCd on" is used. The balance transmits an error code when the received command can not be achieved.

Computer Time

R command

Balance

Weighing mode (unstable)

CAL command

Command received

In process

When the R command can not be achieved due to unstable weighing:

Error code

When the CAL command is received, the balance returns to the weighing mode.

Command completed.

Weighing mode

Weighing with a tare

This example uses "erCd on" of "5if" so that the <AK> (06h) code is output.

Computer Time

R command

TR command

S command

?PT command

Balance

Sets the display to zero

Command received

In process

Command completed

Zero display

Place a container

Tares

Command received

In process

Command completed

Zero display

(Net value)

(Tare value)

36

Setting a negative target value and filling with a sample until the display becomes zero

This example uses "erCd on" of "5if" so that the <AK> (06h) code is output.

R command

PT: command to set a target weight of 10 g

Place a container

Before command execution

Command received

In process

Command completed

Zero display

Command received

Target weight display

Fill with sample until the display becomes zero

37

9. MAINTENANCE

9.1. Treatment Of The Balance

When cleaning, remove the weighing pan and pan support from the balance and attach the waterproof cover for cleaning the balance.

Do not use organic solvents to clean the balance.

Do not disassemble the balance. Contact the local A&D dealer if the balance needs service or repair.

Use the original packing material for transportation. When packing, fasten the waterproof cover to the balance using the tape by following the instructions at the figure below. balance.

1

Remove the weighing pan and pan support .

9.2. Error Codes

Error code

2

EC,E00

Waterproof cover for cleaning

3

Fasten the waterproof cover to the balance using tape.

Description

Internal error

If displaying this error continuously, contact your local A&D dealer to repair the balance.

EC,E11

EC,E16

EC,E17

EC,E20

Stability error

The balance can not stabilize due to an environmental problem. Check around the pan. Prevent vibration, drafts, temperature changes, static electricity and magnetic fields, from influencing the balance.

Internal weight error

Applying the internal weight does not yield a change in the weight value as specified. Confirm that there is nothing on the pan and perform the weighing operation from the beginning again.

Internal weight error

The internal weight application mechanism does not function properly. Perform the weighing operation from the beginning again.

Calibration weight error

The calibration weight is too heavy. Confirm the calibration weight value. Press the

1 CAL 1 key to return to the weighing mode.

38

code

E

Error code

EC,E21

-E

EC,E00

EC,E01

EC,E02

EC,E03

EC,E04

EC,E06

EC,E07

Description

Calibration weight error

The calibration weight is too light. Check around the pan. Confirm the calibration weight value. Press the 1 CAL 1 key to return to the weighing mode.

Overload error

A sample beyond the balance weighing capacity has been placed on the pan.

Remove the sample from the pan.

Underload error

The weight value is too light. Confirm that the weighing pan and pan support are properly installed. Press the ON:OFF key two times to return to the weighing mode.

If the error still persists, calibrate the balance.

Communications error

A protocol error occurred in communications.

Confirm the format, baud rate and parity.

Undefined command error

An undefined command was received.

Confirm the command.

Not ready

A received command can not be processed.

Example:

The balance received a "Q" command, but not in the weighing mode.

The balance received a "Q" command while processing a RE-ZERO command.

Adjust the delay time to transmit a command.

Timeout error

If the timeout parameter is set to "t-Up on", the balance did not receive the next character of a command within the time limit of one second. Confirm the communication.

Excess characters error

The balance received excessive characters in a command.

Confirm the command.

Format error

A command includes incorrect data.

Example:

The data is numerically incorrect. Confirm the command.

Parameter setting error

The received data exceeds the range that the balance can accept. Confirm the parameter range of the command. contact the local A&D dealer.

39

9.3. Checking The Balance Performance And Environment

The balance is a precision instrument. When the operating environment or the operating method is inadequate, correct weighing can not be performed. Place a sample on the pan and remove it, and repeat this several times. If the balance seems to have a problem with repeatability or to perform improperly, check as described below. If improper performance persists after checking, contact the local A&D dealer for repair.

Checking that the operating environment or weighing method is proper

Operating environment

Is the weighing table solid enough?

Is the balance level?

Is the operating environment free from vibration and drafts?

Is there a strong electrical or magnetic noise source such as a motor near the balance?

Weighing method

Does the weighing pan rim touch anything? Is the weighing pan installed correctly?

Is the RE-ZERO button touched before placing a sample on the weighing pan?

Is the sample placed in the center of the weighing pan?

Has the balance been calibrated using the internal weight (one-touch calibration)?

Has the balance been warmed up for one hour before weighing? (At the AD-4212D-33/32, duration of four hours or more are required.)

Sample and container

Has the sample absorbed or lost moisture due to the ambient conditions such as temperature and humidity?

Has the temperature of the container been allowed to equalize to the ambient temperature?

Is the sample charged with static electricity?

Is the sample of magnetic material such as iron? There are cautions about weighing magnetic materials.

Checking that the balance performs properly

Check the balance performance using an external weight. Be sure to place the weight in the center of the weighing pan.

Check the balance repeatability, linearity and calibrated value using external weights with a known value.

Check whether the calibration is performed correctly. If an error occurs and the calibration can not be performed quickly, the balance may not measure correctly.

9.4. Asking For Repair

If the balance needs service or repair, contact your local A&D dealer.

The balance is a precision instrument. Use much care when handling the balance and observe the following when transporting the balance.

Use the original packing material for transportation.

Remove the weighing pan, pan support, rings and dust plate from the main unit.

40

10. SPECIFICATIONS

Weighing capacity

Maximum display

Minimum display

Repeatability

(Standard

AD-4212D-33 AD-4212D-32 AD-4212D-302 AD-4212D-301

32 g

32.000084 g

0.001 mg



32 g

5.1 g

32.00008 g

5.100009 g



0.01 mg

0.001 mg

0.02 mg

0.004 mg



320 g

320.00084 g

0.01 mg

0.02 mg



320.0084 g

320 g

0.1 mg

0.07 mg



Linearity ±0.02 mg ±0.03 mg ±0.15 mg ±0.25 mg

Stabilization time

(Typical at FAST)

Approx. 7 seconds

Approx. 7 seconds

Approx. 7 seconds

Approx. 3 seconds

Sensitivity drift

Operating environment

Display refresh rate

Environmental sensors

±2 ppm/°C (10 °C to 30 °C)

5 °C to 40 °C, 85 %RH or less (No condensation)

5 times/second or 10 times/second

Resolution: Temperature ±1.5 °C, Humidity ±10 %, Air pressure ±10 hPa

Applicable range: 5 °C to 40 °C gram

RS-232C

Display mode

Communication function

External calibration weight

30 g

20 g

10 g

5 g

2 g

1 g

30 g

20 g

10 g

5 g

2 g

1 g

Size of loading area for

Weighing pan:φ34 mm

300 g

200 g

100 g

50 g

20 g

10 g

300 g

200 g

100 g

50 g

Weighing pan:φ50 mm

External dimensions

Power supply &

AC adapter type

259(W) x 466(D) x 326(H) mm

Power consumption: Approx. 30VA (supplied to the AC adapter )

Confirm that the adapter type is correct for the local voltage and power receptacle type.

Approx. 3.6 kg Mass

The operating environment does not include excessive change of ambient temperature, humidity, vibration, drafts, magnetic fields and static electricity.

The internal weight may change due to corrosion or other damage caused by the operating environment, or due to aging. Check the internal weight periodically and correct the internal weight value if necessary.

 When placing the tare weight within the standard range, if using the smart range, the weighing sample can be weighed within the precision range.

 This is a figure measured by loading and unloading a 1g weight on the device using the automatic machine.

 The repeatability for 10 gram within the weighing capacity.

41

11. EXTERNAL DIMENSIONS

- AD-4212D-33/32

94

82.5 ± 1

Pan support dimensions

26

φ

32.8

4-M3

2- φ 3.4 hole

153

133 ± 0.2

φ

34

3-M6

When securing, use the M5 screws.

14.5

56.5

11 0.5

Unit: mm

42

- AD-4212D-302/301

Pan support dimensions

26

φ

48.8

26 10

4-M3

2- φ 3.6 hole

153

133 ± 0.2

φ

50

3-M6

When securing, use the M5 screws.

connected. Approx.5

94

82.5 ± 1

14.5

56.5

11 0.5

Unit: mm

43

12. OPTIONS

AX-KO3590-1000 (For external display) RS-232C cable 10m

AX-KO4472-1000 (For PCPLC) RS-232C cable 10m

AX-USB-9P: USB converter/Cable set

 USB converter (D-Sub 9 pin - USB)

 RS-232C cable (D-Sub 9 pin - D-Sub 9-pin)

* This RS-232C cable is not used with the AD-4212D series.

AD-8922A: Remote controller

 The AD-8922A can receive the weighing data, and performs re-zero and calibration manually.

 The AD-8922A can output the data using BCD

(AD-8922A-01), comparator (AD-8922A-04), or analog out (AD-8922A-06), by selection of the option installed in the AD-8922A.

AD-8923-BCD/CC: Remote controller

 The AD-8923-BCD or AD-8923-CC can receive the weighing data, and performs re-zero and calibration manually.

 The AD-8923-BCD can output the data using BCD.

The AD-8923-CC can output the data using CC-Link.

AD-8526: Ethernet converter

 The AD-8526 can convert the AD-4212D RS-232C data, for output to a local area network (LAN).

AD-1683: DC static eliminator

 A compact design with efficient static elimination

 No air blowing from a fan allows precision weighing

44

AD-1684A: Electrostatic field meter

 This option measures the amount of the static charge on the sample, tare or peripheral equipment and displays the result.

If those are found to be charged, discharge them using the AD-1683 DC static eliminator.

AD-8121B Printer

 Compact thermal dot-matrix printer

 Statistical function, clock and calendar function, interval print function, graphic print function, dump print mode

 5 x 7 dots, 16 characters per line

 Print paper (AX-PP143, 45 (W) x 50 (L) mm , ø65 mm)

 AC adapter or alkaline battery

45

13. TERMS/INDEX

Terms

Stable value

Environment

Calibration

Output

Zero point

The weight value when the stabilization indicator appears.

Ambient conditions such as vibration, drafts, temperature, static electricity or magnetic fields which affect the weighing operation.

Adjustment of the balance using the calibration weight so that it can weigh accurately.

To output the weighing data using the RS-232C interface.

A weighing reference point or the zero display. Usually refers to the value displayed when nothing is on the weighing pan.

Digit

Tare

Re-zero

Unit of minimum display. Used for the balance, the minimum displayable weighing value.

To cancel the weight of a container which is not included in the weighing data.

To set the display to zero.

GLP Good Laboratory Practice.

Repeatability Variation in measured values obtained when the same weight is placed and removed repetitively. Usually expressed as a standard deviation. e.g. Standard deviation=1 digit: This means that the measured values fall within

1 digit at a probability of about 68%.

Stabilization time Time required after a sample being placed, until the stabilization indicator illuminates and the weighing data is displayed.

Sensitivity drift An affect that a change in temperature causes to the weighing data. Expressed as temperature coefficient. e.g. Temperature coefficient = 2 ppm/

C : If a load is 100 g and the temperature changes by 10

C, the value displayed changes by the following value.

0.0002%/

C x 10C x 100 g = 2 mg

46

Index

Symbols

Space mark .................................................29

AC adapter.............................................. 6, 9, 11, 14

AC adapter type.....................................................41

AD-1683.......................................................... 12, 44

AD-1684A ..............................................................45

AD-1691.............................................................3, 21

AD-8121B ..............................................................45

AD-8526.............................................................4, 44

AD-8922A ..................................................... 3, 4, 44

AD-8923-BCD ...................................................2, 44

AD-8923-CC......................................................4, 44

AK code ................................................34, 35, 36, 37 erCd AK, Error code .....................................23

AK, Error code .......................................................23

Ambient humidity...................................................12

AND-MEET..............................................................2

Anti-static agent .....................................................12 ap-b Auto print difference .......................... 22, 25 ap-p Auto print polarity .............................. 22, 25 ar-d Zero after output ............................... 23, 25 at-f Auto feed .................................................23

Auto feed..................................................................23

Auto print........................................................... 22, 25

Auto print difference ..............................................22

Auto print polarity...................................................22

AX-KO3705-1000....................................................4

AX-KO3705-200......................................................4

AX-KO3705-500......................................................4

AX-KO4472-1000............................................3, 4, 8

AX-USB-9P........................................................3, 44

B ba5fnc Environment & Display.........................22

Baud rate........................................................... 23, 30

BCD............................................................... 2, 4, 44 bp5 Baud rate ...................................................23

Breeze break......................................................6, 12 btpr Data bit, parity bit......................................23

Bubble spirit level.....................................................6

Cable set ................................................................44

Cal in One-touch calibration ...........................17

Calibration .................................................16, 18, 46

One-touch calibration

................................... 17

Calibration weight....................................... 6, 16, 18

CC-Link ......................................................... 2, 4, 44

Charged material...................................................12

Command .............................................32, 35, 36, 37

Communication function .......................................41

Cond Condition .......................................... 22, 24

Conductive acrylic fiber.........................................12

Connect................................................................3, 4

CR Carriage return...........................................29

Crlf Terminator .......................................... 23, 32

Data bits ...................................................................23

Data bits & parity bits.................................................30

Data format..................................................23, 26, 29

Data output...............................................................23

Data output mode .............................................. 22, 25

Data output pause.....................................................23

Date.................................................................. 23, 27

DC static eliminator......................................... 12, 44

Decimal point...............................................22, 24, 27

Digit.........................................................................46

Discharge...............................................................12

Display lock function................................................22

Display mode.........................................................41

Display refresh rate......................................22, 24, 41 dout Data output .................................22, 23, 25

47

Dust plate .................................................................6

E ............................................................................39

-E ............................................................................39

Earth ground ..........................................................12

EC,E00............................................................ 38, 39

EC,E01...................................................................39

EC,E02...................................................................39

EC,E03...................................................................39

EC,E04...................................................................39

EC,E06...................................................................39

EC,E07...................................................................39

EC,E11 ...................................................................38

EC,E16...................................................................38

EC,E17...................................................................38

EC,E20...................................................................38

EC,E21...................................................................39

Electrostatic field meter.........................................45

Environment.............................................................40

Environmental sensors .........................................41 erCd

AK, Error code ......................34, 35, 36, 37

Error code ......................................................... 36, 38

Ethernet converter.............................................2, 44

EXC button ............................................................18

External calibration weight....................................41

External dimensions..............................................41

FAST ..................................................................2, 16

Fine range ring.........................................................6

Grounding terminal..................................................6

ID number ......................................................... 23, 27

Insulators................................................................12 int

Interval time ...............................................23

Interval mode..................................................... 22, 25

Interval time..............................................................23

LAN...........................................................................4

Leveling foot.........................................................6, 7

LF

Line feed .....................................................29

Linearity..................................................................41

Magnetic substance..................................................20

Mass.......................................................................41

Maximum display...................................................41

MID.........................................................................16

Minimum display....................................................41

Minimum weighing value ........................................2

Operating environment .........................................41

Pan support..............................................................6

Pan support boss.....................................................6

Parity................................................................. 23, 30

Personal computer..................................................3

Pilot lamp............................................................6, 14

Plastic .....................................................................12

PLC...........................................................................4 ploc Display lock function ................................22 pnt

Decimal point ...................................... 22, 24

Powdery samples..................................................12

Power plug stopper .............................................6, 9

Power supply .........................................................41

Precision range......................................................19

Printer .....................................................................45 prt

Data output mode ............................... 22, 25 pU5e Data output pause....................................23

R command ...........................................................19

Remote controller ..................................................44

48

Repeatability ................................................... 41, 46

RESPONSE...................................................... 22, 24

RS-232C .................................................3, 4, 30, 31

RS-232C cable ....................................................6, 8

RS-232C connector ................................................6

RsCom Command mode for WinCT...................31

RsKey Data output mode for WinCT .................31

RsWeight................................................................31

Sensitivity drift ................................................. 41, 46

Serial interface.......................................................23

5-id ID number output.....................................23

5if

Serial interface ...........................23, 25, 26

5if Serial interface .......................................23

SLOW ....................................................................16

Smart range .............................................................2

Smart Range Function..........................................19

5pd

Display refresh rate .......................22, 24, 25

Stability band width............................................. 22, 24

Stabilization time ...................................... 2, 41, 46

Stable value..................................................... 16, 46

Standard range......................................................19

Static charge ..........................................................13

Static electricity................................................ 10, 46

5t-b Stability band width...............................22

5t-b Stability detection..................................24

5-td Time & Date output..................................23

Stop bit.....................................................................30

5-tp Temperature output..................................23

Stream mode..................................................... 22, 25

Tare..........................................................................36

Temperature.............................................................28

Temperature output...................................................23

Terminator ......................................................... 23, 32

Time.................................................................. 23, 28 t-Up Timeout....................................................23

Timeout...................................................................23 trc

Zero tracking ....................................... 22, 24 type

Data format ........................................ 23, 26

Underhook................................................................20

USB ..........................................................................3

USB converter ...................................................3, 44

Warm up.............................................................9, 14

Waterproof cap for cleaning....................................6

Weighing capacity ...................................... 2, 14, 41

Weighing pan.....................................................6, 41

Weighing speed.............................................. 10, 16

WinCT Communication Tools............................31

Zero

Zero tracking

......................................... 22, 24

Zero after output ....................................................23

Zero point ........................................................ 18, 46

49

MEMO

50

MEMO

51

MEMO

52

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals