OMEGA DVG64 Electronic Vacuum Gauge Instruction sheet

Below you will find brief information for Electronic Vacuum Gauge DVG64. This device is used for measuring vacuum in a variety of applications. It reads vacuum in 7 international units, uses standard 9V alkaline battery, and features a cleaning port for quick cleaning and drying of the sensor. The device comes with a built-in hanger for easy storage.

PDF Download
Document
Electronic Vacuum Gauge DVG64 Instruction Sheet | Manualzz

Advertisement

Advertisement

Advertisement

Key features

  • Reads vacuum in 7 international units
  • Uses standard 9V alkaline battery
  • Cleaning port provides quick cleaning and drying of sensor
  • Convenient built-in hanger
  • Auto shut off after 10 minutes when vacuum reading is above 12,000 Microns (12 Torr)
  • Continuous Usage Over 35 hours
  • Updated every Ω second, 1 micron resolution below 200 micron of vacuum
  • Operating Temp. Range 35°F to 125°F ( 2°C to 52°C)
  • Accuracy +/-10% (0 to 1000 microns)

Frequently asked questions

The DVG64 can read vacuum in 7 international units: Microns, PSI, Inches of mercury (InHg), milliBars, Pascals, Torr and milliTorr.

To change the scale, press the Scale button. The scale order is: Microns, PSI, Inches of mercury (InHg), milliBars, Pascals, Torr and milliTorr. The DVG64 will keep the scale settings even if the power is turned OFF.

To clean the sensor, close the Vacuum port with the supplied cap, open the Auxiliary port, pour about 2 teaspoons of ordinary rubbing alcohol into the Auxiliary port, close the Auxiliary port, shake the DVG64 unit for about 10 seconds, open both the Vacuum and the Auxiliary ports, empty the alcohol, and air dry the sensor.

The DVG64 can be used continuously for over 35 hours with a 9 Volt Alkaline battery.
Preparing document for printing…