SMX2064/2060, SM2064/2060
SMX2064/2060,
SM2064/2060
1
7-
/2 Digit Digital Multimeter
Features
Operating Systems
Flexible, full-featured auto-ranging DMM
• Windows 98/NT/2000/XP/2003
7-1/2 digit resolution
Recommended Software
20,000 readings/second (SMX2064, SM2064)
• VB/VC++/BCB/Delphi
1,400 readings/second (SMX2060, SM2060)
• DAQBench
DC & AC voltage & current, 2-wire, 4-wire,
6-wire (SMX2064, SM2064) ohms
Frequency and time measurements
Capacitance & inductance measurements (SMX2064, SM2064)
Voltage & current sourcing (SMX2064, SM2064)
Precise trigger capabilities
330 V isolation barrier
Self-calibrating
Plug-and-Play software with graphical user interface
Fully programmable with Visual Basic, C++, LabVIEW™ etc.
2-Wire Resistance
Introduction
Accuracy ± (% of reading + Ω) [1]
1
The SMX2064, SMX2060, SM2064 and SM2060 are 7- /2 digit high-speed
Digital Multimeters in PXI and PCI, which maintain high accuracy at high
measurement rates. The measurement speed is up to 20,000 readings/second
for SMX2064 and SM2064. The SMX206x and SM206x series provide a
comprehensive set of DMM capabilities, including 2-wire, 4-wire, and 6-wire
guarded resistance measurements; inductance and capacitance; leakage and
temperature; frequency and timing; sourcing of voltage and current; and much
more. The SMX2064, SMX2060, SM2064 and SM2060 Digital Multimeters are
easy to setup and use, have sophisticated analog and digital circuitry to provide
repeatable measurements, and are protected to handle any unexpected situation
your measurement environment may encounter. With high performance and
variable applications, the SMX2064, SMX2060, SM2064 and SM2060 are
suitable for automated production testing, laboratory automation, and
portable/field testing.
DC Functions
DC Voltage
Accuracy ± (% of reading + Volts) [1]
Range Full scale
7-1/2 Digits
Resolution Input
24 hours
Resistance 23˚C±1˚C
90 Days
23˚C±5˚C
240 mV 240.00000 mV
10 nV
>10 GΩ
0.003 + 1 µV
0.004 + 1.5 µV 0.005 + 2 µV
0.0025 + 5 µV
One Year
23˚C±5˚C
2.4 V
2.4000000 V
100 nV
>10 GΩ
0.002 + 3 µV
24 V
24.000000 V
1 µV
10 MΩ
0.004 + 120 µV 0.005 + 130 µV 0.006 + 150 µV
240 V
240.00000 V
10 µV
10 MΩ
0.003 + 250 µV 0.004 + 300 µV 0.005 + 0.5 mV
330 V
330.00000 V
10 µV
10 MΩ
0.005 + 550 µV 0.01 + 700 µV
0.003 + 5 µV
0.015 + 0.8 mV
[1] With Aperture set to ≥ 0.5 sec, and within one hour from Self Calibration (S-Cal)
DC Current
Accuracy ±(% of reading + Amps) [1]
Resolution Max Burden 24 hours
Voltage
23˚C±5˚C
90 Days
23˚C±5˚C
Current 23˚C±1˚C
90 Days
One Year
23˚C±5˚C
23˚C±5˚C
24 Ω[2]
24.000000 Ω
1 µΩ
10 mA
240 Ω
240.00000 Ω
10 µΩ
1 mA
0.0037 + 4.5 mΩ 0.0046 + 5 mΩ
0.007 + 6 mΩ
2.4 kΩ
2.4000000 kΩ
100 µΩ
1 mA
0.0023 + 28 mΩ
0.006 + 33 mΩ
0.0038 + 1.4 mΩ 0.005 + 1.6 mΩ 0.008 + 2 mΩ
0.004 + 32 mΩ
24 kΩ
24.000000 kΩ
1 mΩ
100 µA
0.0025 + 300 mΩ 0.004 + 330 mΩ 0.006 + 350 mΩ
240 kΩ
240.00000 kΩ
10 mΩ
10 µA
0.0055 + 3.2 Ω
0.006 + 4 Ω
0.007 + 5 Ω
2.4 MΩ
2.4000000 MΩ 100 mΩ
1 µA
0.018 + 40 Ω
0.03 + 50 Ω
0.04 + 70 Ω
24 MΩ
24.0000 MΩ
100 Ω
100 nA
0.12 + 400 Ω
0.13 + 500 Ω
0.2 + 600 Ω
1 kΩ
10 nA
0.8 + 20 kΩ
1.0 + 30 kΩ
1.3 + 50 kΩ
240 MΩ[2] 240.000 MΩ
[1] With Aperture set to ≥0.5 Sec, and within one hour from Self Calibration (S-Cal).
[2] Ranges are only with the SMX2064, SM2064.
[3] Test voltages are 2.4 V max with the exception of the 24 Ω and 240 Ω ranges 240 mV.
Range [3] Full scale
240 nA[2] 240.0000 nA 0.1 pA
100 µV
0.07 + 40 pA
0.1 + 45 pA
0.17 + 60 pA
2.4 µA[2]
2.400000 nA 1 pA
100 µV
0.05 + 70 pA
0.08 + 90 pA
0.21 + 150 pA
24 µA[2]
24.00000 µA 10 pA
Resolution Source 24 hours
7-1/2 Digits
Current 23˚C±1˚C
90 Days
One Year
23˚C±5˚C
23˚C±5˚C
24 Ω[2]
24.000000 Ω
1 µΩ
10 mA
240 Ω
240.00000 Ω
10 µΩ
1 mA
0.0037 + 3 mΩ
0.0046 + 4 mΩ
0.007 + 5 mΩ
2.4 kΩ
2.4000000 kΩ
100 µΩ
1 mA
0.0023 + 28 mΩ
0.004 + 32 mΩ
0.006 + 33 mΩ
0.0038 + 0.7 mΩ 0.005 + 0.8 mΩ 0.008 + 1 mΩ
24 kΩ
24.000000 kΩ
1 mΩ
100 µA
0.0025 + 300 mΩ 0.004 + 330 mΩ 0.006 + 350 mΩ
240 kΩ
240.00000 kΩ
10 mΩ
10 µA
0.0055 + 3.2 Ω
0.007 + 4 Ω
0.007 + 5 Ω
2.4 MΩ
2.4000000 MΩ 100 mΩ
1 µA
0.018 + 40 Ω
0.03 + 50 Ω
0.04 + 70 Ω
24 MΩ
24.0000 MΩ
100 nA
0.12 + 400 Ω
0.13 + 500 Ω
0.2 + 600 Ω
100 Ω
[1] With Aperture set to ≥0.5 Sec, and within one hour from Self Calibration (S-Cal).
[2] Ranges are only with the SMX2064, SM2064.
[3] Test voltages are 2.4 V max with the exception of the 24 Ω and 240 Ω ranges 240 mV.
Diode Characterization
Maximum Diode
Available DC current
Typical Current
Typical Voltage Value
Voltage Compliance
Uncertainty
Value
Uncertainty
4V
100 nA, 1 µA, 10 µA, 100 µA and
1%
0.02%
One Year
23˚C±5˚C
1 mA (SMX2064 and SM2064
10 mA & 12.5 mA)
AC Functions
AC Voltage (true RMS)
One Year Accuracy ± (% of reading + Volts), 23˚C±5˚C
100 µV
0.05 + 400 pA
0.08 + 600 pA
0.13 + 0.8 nA
240 µA[2] 240.000 µA
10 nA
2.5 mV
0.052 + 200 nA 0.07 + 300 nA
0.1 + 400 nA
2.4 mA
2.40000 mA
10 nA
25 mV
0.05 + 300 nA
0.07 + 550 nA
24 mA
24.0000 mA
100 nA
250 mV
0.05 + 350 nA
0.065 + 450 nA 0.08 + 550 nA
240 mV 240.00000 mV 100 nV
3.2 + 430 µV 0.4 + 200 µV
0.15 + 120 µV 0.27 + 230 µV 2.0 + 400 µV
240 m
240.000 mA
1 µA
55 mV
0.05 + 50 µA
0.055 + 60 µA
0.065 + 80 µA
2.4 V
2.4000000 V 1 µV
3.2 + 2.5 mV 0.4 + 1.7 mV
0.065 + 1.2 mV 0.35 + 1.5 mV 2.1 + 2 mV
2.4 A
2.40000 A
10 µA
520 mV
0.3 + 60 µA
0.4 + 70 µA
0.45 + 90 µA
24 V
24.000000 V 10 µV
3.3 + 20 mV
0.073 + 13 mV 0.22 + 25 mV
240 V
240.00000 V 100 µV
3.3 + 200 mV 0.4 + 150 mV 0.06 + 130 mV 0.30 + 200 mV 1.6 + 300 mV
330 V
330.00000 V 100 µV
3.3 + 200 mV 0.45 + 250 mV 0.09 + 230 mV 0.32 + 300 mV 1.6 + 400 mV
0.06 + 400 nA
[1] With Aperture set to ≥0.96 sec, and within one hour from zero (Relative control).
[2] Available only with the SMX2064 and SM2064.
3-24
7-1/2 Digits
Accuracy ± (% of reading + Ω) [1]
Specifications subject to change without notice.
For the most current and complete specifications, please refer to the user manual.
Full scale
6-1/2 Digits
Resolution Source 24 hours
4-Wire Resistance
Specifications
Range
Range [3] Full scale
www.adlinktech.com
Range Full scale
Resolution 10Hz - 20Hz 20Hz - 47Hz
47Hz - 10kHz 10kHz - 50kHz 50kHz-100kHz
7 -1/2 Digits
0.4 + 16 mV
1.5 + 40 mV
Digital Multimeter
AC Current (true RMS)
One Year Accuracy ± (% of reading + Amps), 23˚C±10˚C
Range Full scale
Resolution Max Burden
6-1/2 Digits
10Hz - 20Hz[1] 20Hz - 47Hz[1] 47Hz - 1kHz[1] 1kHz -
Voltage (RMS)
2.4 mA 2.400000 mA 1 nA
25 mV
10kHz[1]
2.9 + 4 µA
1.0 + 4 µA
0.12 + 4 µA
0.22 + 4 µA
24 mA 24.00000 mA 10 nA
250 mV
2.8 + 30 µA
1.0 + 30 µA
0.16 + 30 µA
0.4 + 40 µA
240 mA 240.0000 mA 100 nA
55 mV
2.8 + 400 µA
1.0 + 400 µA
0.2 + 220 µA
0.4 + 400 µA
2.4 A
520 mV
2.7 + 5 mA
0.9 + 6 mA
0.35 + 4 mA
0.5 + 5 mA
2.400000 A
1 µA
Other measurement functions of SMX2064 and SM2064 6-wire guarded
1
resistance, extended resistance, AC peak-to-peak voltage, AC crest factor,
Software
Solutions
AC median value, leakage current, RTD temperature, thermocouple temperature
[1] All AC Current ranges have typical measurement capability to 20 kHz.
Time Functions
Frequency and Period
ACV Mode
Input RMS
Input Impedance Frequency Range Period Range Resolution Uncertainty
Voltage range
24 mV - 250 V 1 MΩ with < 300 pF 2 Hz - 300 kHz
2
Source Functions (SMX2064 and SM2064 only)
0.5 s - 3.33 µs 5 1/2 digits ± 0.002% of reading
DC Voltage Source
Output range: -10.000 V to +10.000 V
DAC resolution: 18 bits (closed loop), 12 bits (open loop)
PXI/
CompactPCI
Platforms
3
AC Voltage Source
Output range: 900 mV to 8 V
DAC resolution: 12 bits
Frequency range/resolution: 1 Hz to 200 kHz/2 mHz
Modular
Instrument
4
DC Current Source
PXI/
CompactPCI
Modules
Output range: 1.25 µA to 12.5 mA
ACI Mode
Input Signal
Input Impedance
Frequency
Period Range Resolution Uncertainty
range
Range
0.33 mA - 2.5 A 10 Ω(3 mA & 30 mA)
2 Hz - 500 kHz 0.5 s - 2.0 µs 5 1/2 digits
Trigger Functions
± 0.01% of reading
Trigger input voltage level range: +3 V to +15 V
Minimum trigger input current: 1 mA
0.1 Ω (330 mA & 2.5 A)
Pulse Width
Polarity
Frequency Range Resolution Width Range Typical Uncertainty
Positive or negative pulse widths 2 Hz to 100 kHz
1 µs
2 µs to 1 s
0.01% of reading ±4 µs
Threshold DAC
Trigger point: selectable positive or negative transition of set threshold
6
Captures up to 120 post-trigger readings for apertures > 625 µ Sec.
GPIB
Interface
Captures up to 80 post-trigger readings for apertures > 625 µ Sec.
Delayed Hardware Trigger
Threshold range
Threshold DAC
Highest allowed
Typical one year
Up to 65 m Sec with 1 µ Sec resolution
Range
(DC level)
resolution
input Vp-p
setting uncertainty
Up to 1 s with 2 µs resolution
240 mV
-1.0 V to +1.0 V
0.5 mV
1.900 V
0.2% + 4 mV
2.4 V
-10.0 V to +10.0 V
5.0 mV
19.00 V
0.2% + 40 mV
24 V
-100.0 V to +100.0 V 50 mV
190.0 V
0.2% + 0.4 V
240 V
-400 V to 400 V
850.0 V
0.2% + 4 V
Totalizer
Active edge polarity
Maximum Count Allowed rate
Positive or negative transition 10^9
General Specifications
Overload Protection (voltage inputs): 330 VDC, 250 VAC
Isolation: 330 VDC, 250 VAC from earth ground
Maximum Input (Volt x Hertz):
Condition
1 to 30,000 events per second Uses Threshold DAC
Capacitance and Inductance Specifications (SMX2064 and SM2064 only)
Capacitance
Accuracy ±(% of reading + farads) [1]
Range
Full scale
Reading
Resolution
One Year 23˚C±5˚C
Measurement Time
Measurement
Rate (rps)
8x106 volts x Hz normal mode input
1x106 volts x Hz common mode input
Calibration: Calibrations are performed by Signametrics in
constants are stored in a text file.
Operating Temperature: -10 to 65˚C
Storage Temperature: -40 to 85˚C
0.1 pF
1.5 ± 0.25 pF
52.3 ms
19.1
Power requirements: +5 volts, 300 mA maximum
12 nF
11.999 nF
1 pF
1.2 ± 5 pF
118 ms
8.5
Dimensions (not including connectors):
120 nF
119.99 nF
10 pF
1.0
127 ms
7.9
SMX2064/2060: 160 mm x 100 mm
1.2 µF
1.1999 µF
100 pF
1.0
175 ms
5.7
SM2064/2060: 208 mm x 112 mm
11.999 µF
1 nF
1.0
480 ms
2.1
120 µF
119.99 µF
10 nF
1.0
50.3 ms
19.9
1.2 mF
1.1999 mF
100 nF
1.2
151.5 ms
6.6
12 mF
50.000 mF
1 µF
2
170 ms
5.9
[1] Within one hour of zero, using relative control. Accuracy is specified for values higher
than 5% of the selected range with the exception of the 1,200 pF range.
Inductance
Accuracy ± (% of reading + henrys)
Range
Test Frequency
Full Scale 4 1/2 Digits
Resolution
One Year Accuracy
23˚C±5˚C [1]
24 µH
75k Hz
33.000 µH
1 nH
3.0% + 500 nH
240 µH
50 kHz
330.00 µH
10 nH
2.0% + 3 µH
2.4 mH
4 kHz
3.3000 mH
100 nH
1.5% + 25 µH
24 mH
1.5 kHz
33.000 mH
1 µH
1.5% + 200 µH
240 mH
1 kHz
330.00 mH
10 µH
2.5 + 3 mH
2.4 H
100 Hz
3.3000 H
100 µH
3.0 + 35 mH
[1] Accuracy is specified for values greater than 5% of the selected range.
7
PCI/PCI
®
Express
DAQ Cards
8
PCI/PCI
®
Express
DIO Cards
9
PC/104-Plus
Modules
a computer at 23˚C internal temperature rise. All calibration
1,200 pF 1,199.9 pF
12 µF
Bus
Interface
Analog Threshold Trigger
Selected VAC
500 mV
5
External Hardware Trigger (at DIN-7 connector)
Safety: Designed to IEC 1010-1, Installation Category II
Ordering Information
SMX2064
7-1/2 digits PXI Digital Multimeter with LCR Meter
SMX2060
7-1/2 digits PXI Digital Multimeter
SM2064
7-1/2 digits PCI Digital Multimeter with LCR Meter
SM2060
7-1/2 digits PCI Digital Multimeter
10
ISA DAS/
DIO Cards
11
System
Product
12
Wiring
Termination
Boards
13
Motion, HSL,
Vision, COM
& GEME
14
Remote I/O
Modules
15
Industrial
Computers
www.adlinktech.com 3-25
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertisement