1
2
3
Company network with DHCP and name server . . . . . . . . . . . . . . . . . . . . . . 5
Workgroup with DHCP router and local shared folder . . . . . . . . . . . . . . . . . . . 7
Workgroup with switch and local shared folder . . . . . . . . . . . . . . . . . . . . . . . 9
Read this manual carefully before installing and running your system.
Store the manual in the immediate vicinity of the PC running the software, so that it can be consulted
at any time.
This manual is copyright. Information from it may not be reproduced, distributed, or used for competitive purposes, nor made available to third parties. The manufacture of any component with the aid of
this manual without prior written agreement is also prohibited.
3
K-375 Network connection, Version A
Table of contents
4
K-375 Network connection, Version A
1 Company network with DHCP and name server
Company network with DHCP and name server
To work in a company network with DHCP and name server proceed as follows:
1. Go to “Settings  Network”. The following screen appears:
2. For “Obtain IP address via DHCP” select “Yes”.
3. Assign a unique device name for the instrument to be recognized within the network.
Fileserver with network
data share e.g.
everyone on \\filesrv01
Company network
with DHCP and
name server
LA
N
LAN
1
everyone
(\\filesrv01) (W:)
PC with KjelLink
installed (optional)
User e.g. MrWho with password
1234 and with read/write access
authorization to everyone
5
K-375 Network connection, Version A
1 Company network with DHCP and name server
4. Switch off the K-375 before connecting it to the LAN to avoid any network conflicts.
5. Connect the K-375 to your company network by using the LAN socket (cable not provided).
6. Make sure that the displayed settings in “Settings  Network” for the IP address, the subnet mask
and the default gateway match your IT guidelines.
7. Create a network share in the Windows Explorer, e.g. everyone on \\filesrv01.
8. Go to “Settings  Import and export”. The following screen appears:
9. Under “Use USB device” tab Yes to enble to store data additionally on a memory stick.
10.Under “Use network data share” tab Yes.
11.Under “Network host name” enter the name of your fileserver, e.g. \\filesrv01.
12.Under “Network share name” enter the name of your network share, e.g. everyone.
13.Leave the “Domain to access network data share” empty, if you do not have one.
14.Under “User name to access network data share” enter your user name and password for the
created network share. Usually that is your PC account name, e.g. MrWho.
15.Under “Password to access network data share” enter your password to access to the created
network share. Usually that is your PC account password, e.g. 1234.
16.Under “Relative path to network data share folder” tab to browse to your directory or enter the
name.
17.Under “Export results automatically” tab Yes.
18.Tab TEST EXPORT to verify if your settings are correct.
NOTE
Please ask your IT department for assistance.
6
K-375 Network connection, Version A
2 Workgroup with DHCP router and local shared folder
2
Workgroup with DHCP router and local shared folder
To work in a network workgroup with DHCP router and local shared folder proceed as follows:
1. Go to “Settings  Network”. The following screen appears:
2. For “Obtain IP address via DHCP” select “Yes”.
3. Assign a unique device name for the instrument to be recognized within the network.
192.168.0.20
255.255.255.0
Router with
DHCP server
192.168.0.22
255.255.255.0
PC01 with KjelLink
installed (optional)
LAN
LAN
N
LA
192.168.0.23
255.255.255.0
PC02 for storing your data
User e.g. MrWho with password
1234 and with read/write access
authorization to transferK375
transferK375
shared folder
4. Switch off the K-375 before connecting it to the LAN to avoid any network conflicts.
5. Connect the K-375 to your workgroup by using the LAN socket (cable not provided).
6. Make sure that the displayed settings for the IP address, the subnet mask and the default gateway
match your workgroup setup.
7. Use the Windows Explorer to create a local shared folder on PC02 e.g. c:\transferK375.
7
K-375 Network connection, Version A
2 Workgroup with DHCP router and local shared folder
8. Go to “Settings  Import and export”. The following screen appears:
9. Under “Use USB device” tab Yes to enble to store data additionally on a memory stick.
10.Under “Use network data share” tab Yes.
11.Under “Network host name” enter the name of your fileserver, e.g. PC02.
12.Under “Network share name” enter the name of your network share, e.g. transferK375.
13.Leave the “Domain to access network data share” empty.
14.Under “User name to access network data share” enter your user name and password for the
created shared folder. Usually that is your PC account name, e.g. MrWho.
15.Under “Password to access network data share” enter your password to access to the created
network share. Usually that is your PC account password, e.g. 1234.
16.Under “Relative path to network data share folder” tab to browse to your directory or enter the
name.
17.Under “Export results automatically” tab Yes.
18.Tab TEST EXPORT to verify if your settings are correct.
NOTE
Please ask your IT department for assistance.
8
K-375 Network connection, Version A
3 Workgroup with switch and local shared folder
3
Workgroup with switch and local shared folder
To work in a network workgroup with switch and local shared folder proceed as follows:
1. Go to “Settings  Network”. The following screen appears:
2. For “Obtain IP address via DHCP” select “No”.
3. Assign a unique device name for the instrument to be recognized within the network.
4. Enter the IP address and the subnet mask manually.
NOTE
In this setup all IP adresses and subnet masks have to be assigned manually. Plase make sure that no
combination is used more than once.
192.168.0.2
255.255.255.0
PC01 with KjelLink
installed (optional)
Switch
LAN
LAN
N
LA
192.168.0.3
255.255.255.0
PC02 for storing your data
User e.g. MrWho with password
1234 and with read/write access
authorization to transferK375
transferK375
shared folder
5. Switch off the K-375 before connecting it to the LAN to avoid any network conflicts.
6. Connect the K-375 to your workgroup by using the LAN socket (cable not provided).
7. Create a local shared folder on PC02, e.g. c:\transferK375.
9
K-375 Network connection, Version A
3 Workgroup with switch and local shared folder
8. Go to “Settings  Import and export”. The following screen appears:
9. Under “Use USB device” tab Yes to enble to store data additionally on a memory stick.
10.Under “Use network data share” tab Yes.
11.Under “Network host name” enter the name of your fileserver, e.g. PC02.
12.Under “Network share name” enter the name of your network share, e.g. transferK375.
13.Leave the “Domain to access network data share” empty.
14.Under “User name to access network data share” enter your user name and password for the
created shared folder. Usually that is your PC account name, e.g. MrWho.
15.Under “Password to access network share” enter your password to access to the created
network share. Usually that is your PC account password, e.g. 1234.
16.Under “Relative path to network data share folder” tab to browse to your directory or enter the
name.
17.Under “Export results automatically” tab Yes.
18.Tab TEST EXPORT to verify if your settings are correct.
NOTE
Please ask your IT department for assistance.
10
K-375 Network connection, Version A
Quality in your hands
BUCHI Affiliates:
BÜCHI Labortechnik AG
CH – 9230 Flawil 1
T +41 71 394 63 63
F +41 71 394 65 65
buchi@buchi.com
www.buchi.com
BUCHI Italia s.r.l.
IT – 20010 Cornaredo (MI)
T +39 02 824 50 11
F +39 02 57 51 28 55
italia@buchi.com
www.buchi.it
BUCHI Russia/CIS
United Machinery AG
RU – 127787 Moscow
T +7 495 36 36 495
F +7 495 981 05 20
russia@buchi.com
www.buchi.ru
Nihon BUCHI K.K.
JP – Tokyo 110-0008
T +81 3 3821 4777
F +81 3 3821 4555
nihon@buchi.com
www.nihon-buchi.jp
BUCHI Korea Inc
KR – Seoul 153-782
T +82 2 6718 7500
F +82 2 6718 7599
korea@buchi.com
www.buchi.kr
BÜCHI Labortechnik GmbH
DE – 45127 Essen
Freecall 0800 414 0 414
T +49 201 747 490
F +49 201 747 492 0
deutschland@buchi.com
www.buechigmbh.de
BÜCHI Labortechnik GmbH
Branch Office Benelux
NL – 3342 GT
Hendrik-Ido-Ambacht
T +31 78 684 94 29
F +31 78 684 94 30
benelux@buchi.com
www.buchi.be
BUCHI China
CN – 200052 Shanghai
T +86 21 6280 3366
F +86 21 5230 8821
china@buchi.com
www.buchi.com.cn
BUCHI India Private Ltd.
IN – Mumbai 400 055
T +91 22 667 75400
F +91 22 667 18986
india@buchi.com
www.buchi.in
BUCHI Sarl
FR – 94656 Rungis Cedex
T +33 1 56 70 62 50
F +33 1 46 86 00 31
france@buchi.com
www.buchi.fr
BUCHI UK Ltd.
GB – Oldham OL9 9QL
T +44 161 633 1000
F +44 161 633 1007
uk@buchi.com
www.buchi.co.uk
BUCHI (Thailand) Ltd.
TH – Bangkok 10600
T +66 2 862 08 51
F +66 2 862 08 54
thailand@buchi.com
www.buchi.co.th
PT. BUCHI Indonesia
ID – Tangerang 15321
T +62 21 537 62 16
F +62 21 537 62 17
indonesia@buchi.com
www.buchi.co.id
BUCHI Corporation
US – New Castle,
Delaware 19720
Toll Free: +1 877 692 8244
T +1 302 652 3000
F +1 302 652 8777
us-sales@buchi.com
www.mybuchi.com
BUCHI do Brasil
BR – Valinhos SP 13271-570
T +55 19 3849 1201
F +41 71 394 65 65
latinoamerica@buchi.com
www.buchi.com
BUCHI Support Centers:
South East Asia
Latin America
Middle East
BÜCHI NIR-Online
BUCHI (Thailand) Ltd.
TH-Bangkok 10600
T +66 2 862 08 51
F +66 2 862 08 54
bacc@buchi.com
www.buchi.com
BUCHI Latinoamérica Ltda.
BR – Valinhos SP 13271-570
T +55 19 3849 1201
F +41 71 394 65 65
latinoamerica@buchi.com
www.buchi.com
BUCHI Labortechnik AG
UAE – Dubai
T +971 4 313 2860
F +971 4 313 2861
middleeast@buchi.com
www.buchi.com
DE – 69190 Walldorf
T +49 6227 73 26 60
F +49 6227 73 26 70
nir-online@buchi.com
www.nir-online.de
We are represented by more than 100 distribution partners worldwide.
Find your local representative at: www.buchi.com
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising