ALMEMO 4390-2 | ALMEMO 2890-9 | ALMEMO 5690-2CPU | ALMEMO 5790-2CPU | ALMEMO 8590-9 | ALMEMO 2450-1L | ALMEMO 8690-9A | ALMEMO 2470-2S | ALMEMO 2470-1S | ALMEMO 5790-2M09 | ALMEMO 5690-1CPU | ALMEMO 5690-1M09 | User manual | AHLBORN ALMEMO 2450-1, 2450-1L, 2490-1, 2490-2, 2490-1L, 2490-2L, 2470-1S, 2470-1SRH, 2470-2S, 2470-2, 2590-2A, 2590-4AS, 2690-8A, 2890-9, 710, 8590-9, 8690-9A, 5690-1M09, 5690-2M09, 5790-2M09, 5690-1CPU, 5690-2CPU, 5790-2CPU, 4390-2 Product catalog

ALMEMO 4390-2 | ALMEMO 2890-9 | ALMEMO 5690-2CPU | ALMEMO 5790-2CPU | ALMEMO 8590-9 | ALMEMO 2450-1L | ALMEMO 8690-9A | ALMEMO 2470-2S | ALMEMO 2470-1S | ALMEMO 5790-2M09 | ALMEMO 5690-1CPU | ALMEMO 5690-1M09 | User manual | AHLBORN ALMEMO 2450-1, 2450-1L, 2490-1, 2490-2, 2490-1L, 2490-2L, 2470-1S, 2470-1SRH, 2470-2S, 2470-2, 2590-2A, 2590-4AS, 2690-8A, 2890-9, 710, 8590-9, 8690-9A, 5690-1M09, 5690-2M09, 5790-2M09, 5690-1CPU, 5690-2CPU, 5790-2CPU, 4390-2 Product catalog

Add to My manuals
311 Pages

Below you will find brief product information for measuring instrument ALMEMO 2450-1, measuring instrument ALMEMO 2450-1L, measuring instrument ALMEMO 2490-1, measuring instrument ALMEMO 2490-2, measuring instrument ALMEMO 2490-1L, measuring instrument ALMEMO 2490-2L, measuring instrument ALMEMO 2470-1S, measuring instrument ALMEMO 2470-1SRH, measuring instrument ALMEMO 2470-2S, measuring instrument ALMEMO 2470-2, measuring instrument ALMEMO 2590-2A, measuring instrument ALMEMO 2590-4AS, measuring instrument ALMEMO 2690-8A, measuring instrument ALMEMO 2890-9, measuring instrument ALMEMO 710, measuring instrument ALMEMO 8590-9, measuring instrument ALMEMO 8690-9A, measuring instrument ALMEMO 5690-1M09, measuring instrument ALMEMO 5690-2M09, measuring instrument ALMEMO 5790-2M09, measuring instrument ALMEMO 5690-1CPU, measuring instrument ALMEMO 5690-2CPU, measuring instrument ALMEMO 5790-2CPU, measuring instrument ALMEMO 4390-2, transmitter ALMEMO 2450-1R02, transmitter ALMEMO 2490-1R02, transmitter ALMEMO 2490-2R02, reference measuring instrument ALMEMO 1020-2, reference measuring instrument ALMEMO 1030-2, reference measuring instrument ALMEMO 1036-2 The ALMEMO 2450-1 is a compact measuring instrument with a single measuring input, designed for a wide range of applications. It features a 2-row segment display for clear readings and a user-friendly interface with seven buttons for navigation. The 2450-1 comes with a selection of over 35 measuring ranges for various sensor types, including thermocouples, NTC sensors, humidity sensors, pressure transducers, and more. The device allows for setting zero points and saving maximum/minimum values for enhanced data analysis. The 2450-1L is similar to the 2450-1 but runs exclusively on battery power.

advertisement

AHLBORN measuring instrument ALMEMO 2450-1 Product Catalog | Manualzz

IJ

FT

L

=lo

www.ahlborn.com

Product

Gatalog

Measuring instruments and sensors

-.-,..

1'Hl ry w

00:

"C

25.80

rE- s*rrd mfiúU

0O!8

ALMEMO

®

Measuring Instruments

Reference Measuring Instruments

ALMEMO

®

input connectors

Output modules

Network technology

Software

General accessories

Temperature Sensors

Infrared measuring technology

Air Humidity Sensors

Air Flow Sensors

Pressure Sensors

Sensors for Force, Displacement, Speed and Flow

Sensors for Electrical variables

Sensors for Meteorology and Room air conditions

Sensors for Building physics and Moisture in materials

Sensors for Optical radiation

Sensors for Water analysis

Sensors for Gas concentrations in the air

Calibration certificates

ALMEMO

®

Measuring Instruments

Contents

The ALMEMO

®

system

General technical specifications

Measuring ranges

ALMEMO

®

measuring instruments, overview

Measuring ranges, ALMEMO

®

2450, 2490, 2470, 2590

Compact ALMEMO

®

measuring instrument ALMEMO

®

2450

Basic measuring instrument ALMEMO

®

2490

Professional measuring instrument ALMEMO

®

2470

Professional measuring instrument ALMEMO

®

2590

Precision measuring instrument ALMEMO

®

2690-8A

Precision measuring instrument ALMEMO

®

2890-9

Precision measuring instrument ALMEMO

®

710

Precision measuring instrument for measured data acquisition

ALMEMO

®

8590/8690

Precision measuring instrument for measured data acquisition

ALMEMO

®

5690/5790

Universal ALMEMO

®

transmitter 2450 / 2490 01.50

Precision measuring instrument ALMEMO

®

4390 in fitted panel design 01.52

Reference measuring instrument ALMEMO

®

1030-2

Reference measuring instrument ALMEMO

®

1020-2

Reference measuring instrument ALMEMO

®

1036-2

01.54

01.55

01.58

01.02

01.05

01.06

01.09

01.10

01.12

01.14

01.16

01.19

01.22

01.24

01.26

01.29

01.31

01.01

ALMEMO

®

Measuring Instruments

The ALMEMO

®

system

The ALMEMO

®

system comprises an

ALMEMO

®

measuring instrument and intelligent ALMEMO

®

connectors for the relevant sensor equipment.

An extensive range of measuring instrument variants is thus available - from the single-channel transmitter right through to data acquisition systems with over 1000 measuring points.

The only differences between most of the measuring instruments in the ALMEMO

® series concern their housing (i.e. handheld instruments, desktop instruments, 19-inch systems, fitted panel instruments, transmitters, etc.), the number of measuring inputs (1 to 250), the display, output, and operating controls, and their respective power supplies.

As soon as a sensor or interface cable is connected, the ALMEMO

®

measuring instrument will, thanks to the intelligent

ALMEMO

®

connector system, be completely programmed right through to process scheduling.

These measuring instruments provide a uniform range of functions with many configurable options. All parameters can be accessed via the interface and can, since the media in the connectors are always overwritten, be freely modified as and whenever necessary.

The ALMEMO

®

principle: Only one measuring instrument for all sensors

An extensive range of transducers, sensors, and signals can be connected to any measuring input on virtually any ALMEMO

® measuring instrument - all via the patented ALMEMO

®

plug system Since all the sensor data is saved in the connector, no extra programming is required; as soon as a sensor is connected, the measuring instrument is configured automatically. The sensor data memory (EEPROM) ensures that each sensor can be identified, scaled, and calibrated - all on the basis of its own unique designation. This system of individual sensor designations avoids confusion and makes the measuring setup clear and logical. Sensor errors can be corrected within the plug, turning simple sensors into precision transducers.

Standard signals can be displayed in their original dimensions. For multi-purpose sensors (e.g. temperature and humidity) only one shared plug will usually be required. Programming can be protected by a graduated locking function.

With ALMEMO

®

measuring instruments you will not need new sensors

For your existing sensors we will provide you with a matching adapter that you can fit quickly and easily. You can also program ALMEMO

®

plugs yourself quickly and easily via keypad, terminal, or software. The data medium in the plug can be overwritten as and whenever necessary.

ALMEMO

®

measuring instruments are ideal for all sorts of application

All incorporate the same measuring input circuitry. For applications that are not sector-specific there are more than 60 standard measuring ranges available, e.g. for measuring :

Temperature, humidity, flow velocity, flow rate, heat flow, pressure, rotational speed, frequency, resistance, current, voltage, force, strain factor, displacement, pH value, redox potential, conductivity, O

2

, CO

2

, CO,

, etc. Maximum and minimum values O

3 are saved automatically. Measured values can be averaged over a series of individual measurements, over the output cycle, or over the actual measuring duration; limit values can be monitored in terms of programmable maximum / minimum values.

Measured values can be corrected with regard to zero point and gain and can be scaled by factor, base value, exponent, and units.

01.02

ALMEMO

®

Measuring Instruments

ALMEMO

®

measuring instruments are real individuals

ALMEMO

®

instruments automatically recognize the specifications of a sensor as it is connected. Specific functions will only be activated as and when the appropriate connector, interface cable, or module is detected. With humidity sensors the dew point, mixture ratio, vapor pressure, and enthalpy will be calculated automatically. Measuring operations involving psychrometers, dynamic pressure probes, or probes for solute oxygen may require pressure compensation; for this purpose the prevailing atmospheric pressure can be entered manually or calculated automatically by an integrated pressure transducer.

When measuring dynamic pressure, pH value, atmospheric humidity, solute oxygen, or conductivity it is possible similarly to perform temperature compensation.

When using flow sensors to measure volume flow the appropriate cross-section can be entered. For certain special sensors there are connectors available incorporating an integrated adapter circuitry.

ALMEMO

®

measuring instruments meet even the most stringent requirements

ALMEMO

®

devices incorporate a highresolution 16-bit A/D converter, digital linearization (for Pt100 sensors with the new ITS 90 temperature scale), and digital calibration. Optimal cold junction compensation is ensured by means of precision thermistors incorporated in the socket spring. Measuring inputs, power supply, and interfaces are all electrically isolated from each other.

The ALMEMO

®

data acquisition system adapts to your requirements

The internal measured data memory incorporated in ALMEMO

®

data loggers can be expanded by adding external capacity and can be configured either as linear or ring memory.

This memory can be read out selectively according to time or number. The switchover between measuring points is electrically isolated using semiconductor relays that are totally wear-resistant. Continuous measuring point scanning at 10 or 50 measuring operations per second can thus be performed trouble-free. Measuring point scans can be individually programmed. Measuring cycles and output cycles can be selected independently; measured values, average values, and maximum / minimum values can be selectively output and / or saved to memory. The start / stop of each measuring point scan can be variably controlled (by keypad or interface, by date and time-of-day, by limit values, or by an external signal). All measuring instruments can be addressed via interface and are thus fully network-capable. Up to

100 devices can be networked either via cable or over a wireless link. The output of measured values from all devices in the whole network can be initiated from any one such device. For covering longer distances RS422 drivers and distributors are available. This system minimizes hardware requirements, cabling costs, and possible EMC problems, and can be expanded as and when required.

ALMEMO

®

measuring instruments accept virtually any peripheral equipment while maintaining optimal data transmission

Analog or digital interfaces are not integrated in the measuring instruments themselves but in the connectors and connecting cables. Depending on requirements a wide variety of adapters can be connected, e.g. analog outputs, various interfaces (RS232,

RS422, optic fiber, current loop, Ethernet,

Bluetooth), alarm signaling devices, or trigger inputs. Data can also be transmitted via a standard fixed-line telephone

(analog or ISDN) or a wireless modem at a maximum baud rate of 9600 baud for remote interrogation purposes.

ALMEMO

®

measuring instruments provide evaluation of measured data easily and conveniently

Suitable output formats are provided for printers or spreadsheet software. For the graphical presentation and the evaluation of measured data there are various software packages available.

ALMEMO

®

instruments can be programmed quickly and easily

The software protocol and the commands list are identical for all devices. Only one terminal is enough to program all parameters and to scan the measured data.

There is a free WINDOWS configuration software, AMR-Control, with terminal, available for this purpose.

01.03

ALMEMO

®

Measuring Instruments

Measuring humidity and moisture

ALMEMO

®

atmospheric humidity sensors provide 4 channels that can be programmed optionally for any of the variables - temperature, relative atmospheric humidity, dew point, mixture ratio, partial vapor pressure, or enthalpy. The first 4 variables are provided as standard.

All measuring functions (maximum, minimum, limit values) and all programming functions can be used for all these channels.

With psychrometers the atmospheric pressure function will also be activated, so that any strongly deviating atmospheric pressure (e.g. at high altitudes above mean sea level) can be entered and used for compensation purposes.

Probes for measuring moisture in materials can be set using the base value for a wide variety of materials, e.g. in the material groups - construction materials, wood, paper.

Measuring air flow velocity

When using hot-wire thermoanemometers, rotating vanes, or dynamic pressure transducers universal ALMEMO

®

measuring instruments 2590-2 and above can activate averaging functions, volume flow, cross section area, and diameter. The volume flow is calculated over the cross section area by matrix measuring with averaging over a series of individual values or continuous averaging. Since calculation of flow velocity in Pitot tubes is strongly influenced by air temperature, automatic temperature compensation can be activated. It is also possible to set an attenuation filter with a selectable time constant, thus ensuring that relatively smooth values can be applied to particularly critical measuring points.

Non-contacting temperature measurement

When measuring infra-red temperature the emissivity factor and background temperature must always be considered. As soon as an infra-red probe is connected these two functions are activated and the associated parameters are stored in the plug.

Radiant temperature - WBGT measurement

Wet-bulb globe temperature (WBGT) is used e.g. for evaluating heat stress in the workplace. Using a psychrometer with disengageable motor and a globe thermometer, WBGT is calculated from the dry temperature TD, the natural wet temperature

TW, and the globe temperature TG.

WBGT = 0.1µ TD + 0.7µ TW + 0.2µ TG

A function channel, WBGT, is provided for evaluating this formula.

Measuring heat flow, thermal coefficient, and transmittance (U value)

The calibration value for each heat flux plate is saved as a factor in the plug, so that heat flow measuring operations can be performed without having to reset the calibration each time. It is also possible to use function channels to determine the average heat flow and the average temperature difference and, from the quotient of these two average values, to determine a thermal coefficient. Depending on how the temperature sensors are arranged, the thermal surface transfer coefficient (α), the thermal conductance coefficient (λ) or the thermal transmittance coefficient (U value) can be determined.

Force measurement including adjustment of zero-point and final value

With force transducers the basic load (tare weight) can be adjusted to zero and the final value can be entered as setpoint. From these values the correction factor will then be calculated automatically. For force transducers with an integrated reference resistor there is a connector available that switches this on for adjustment purposes.

Adjustment and temperature compensation for pH probes

Probes for measuring pH are subject to ageing and must therefore be recalibrated at regular intervals. Zero-point and gain can be calibrated at the touch of a button using the standard reference solutions. A big advantage here is that the calibration setting will be saved in the plug, thus ensuring that the probe can also be operated with other instruments. It is even possible to use several probes with their own individual calibration settings.

Temperature compensation can be performed either automatically using a combined temperature / pH probe or manually by entering the temperature of the medium.

Measuring conductivity - with temperature compensation

The conductivity probe measures the temperature of the medium and calculates conductance referred to 25 °C.

01.04

ALMEMO

®

Measuring Instruments

General technical specifications

Inputs

Channel switching between input sockets

Cold junction compensation (CJC)

Nominal temperature

Sensor power supply

Self-calibration

Monitoring functions

4-contact with photo-MOS relays

Potential separation maximum 50 V

Measuring modules with higher potential separation (see chapter „Input modules“)

Offset voltage <5 µV effective in range -30 to +100 °C, Accuracy ±0.2 K (±0.01 K / °C)

22 °C ±2 K

6 to 12 V depending on power supply

Automatic zero-point correction, measuring current calibration

Automatic sensor recognition and sensor breakage detection

Precision class

ALMEMO

Overload at 2.5 mops

®

Input current

series

Measuring rates

Measuring operations per second (mops)

Input range

Measuring current

System accuracy

Temperature drift

C

2450, 2420

2,5 mops

Basic measuring instruments

B

2490, 2590

2,5, 10mops

Professional measuring instruments

A

2470, 2790

2590A

2,5, 10mops

Precision measuring instruments

2890, 4390

5690, 8490

8590, 8690

AA

2690A, 710

2,5, 10, 50, 100mops

Option 400mops* | Option 500mops *

0.26 to +2.6 V

-4 to +5 V

< 2nA

-2 to +5 V

-2 to +5 V

< 20nA

0.1% of measured value ±4 digits

0.01% / K (100 ppm)

Pt100/1000:

0.3mA

0.03% of measured value ±4 digits

0.005% / K (50 ppm) meas. range 2.6 V:

-2 to +3 V in all other meas. ranges

-1.9 to +2.9 V

-2 to +5 V

100pA

Pt100/1000: 0.3mA

meas. range 2.6 V:

-3 to +3 V in all other meas. ranges

-2.3 to +1.3 V

± 12V meas. range 2.6 V:

-2 to +3 V in all other meas. ranges

-1.9 to +2.9 V

± 12V

Meas. range 2.6 V:

500 nA in all other meas. ranges

500 pA

100pA

Pt100: 1mA, Pt1000: 0.1mA

0.03% of measured value ±3 digits

0.003% / K (30 ppm)

0.02% of measured value ±2 digits

0.003% / K (30 ppm)

*Measuring rate 400 mops (Option SA0000Q4)

*Measuring rate 500 mops (Option SA0000Q5):

It is also possible, in addition to the standard conversion rates, to set 400 or 500 mops (measuring operations per second). At the rate of 400 or

500 mops just one selected measuring channel can be saved. This can only be used with sensors with voltage or current ranges or with NTC sensors. Nor is it possible to change channels in the course of a measuring operation.

The resolution, accuracy, and sensitivity to disturbance caused by mains hum or electromagnetic interference are comparable with measuring operations performed at a rate of 50 mops. Care must be taken to ensure that the environment is free from interference and that the sensor lines are kept short.

Data can only be output to a micro SD card. Accessories ZA1904SD Memory connector with micro SD Data is saved in table format (separated by semi-colons) and with a time-stamp resolution of 0.0001 seconds. This format can be processed using the WinControl software (as of version 6.1.1.6).

Measuring instrument

Interface to all ALMEMO

®

plugs / modules

Operating temperature

Storage temperature

Humidity range

Electromagnetic compatibility Safety standards

I2C bus

-10 to +60 °C

-30 to +60 °C

10 to 90 % (non-condensing)

EN 61010-1: 2001, EMC: EN 61326-1: 2013

01.05

ALMEMO

®

Measuring Instruments

Measuring ranges

Sensor type Type

Resistance temperature detectors:

Pt100 / Pt1000 -1 4-wire FP Axxx

Pt100 / Pt1000 -2 4-wire FP Axxx

Pt100 -3 4-wire

Ni100/1000 4-wire

NTC type N

Thermocouples

NiCr-Ni (K)

NiCroSil-NiSil (N)

Fe-CuNi (L)

Fe-CuNi (J)

Cu-CuNi (U)

Cu-CuNi (T)

PtRh10-Pt (S)

PtRh13-Pt (R)

PtRh30-PtRh6 (B)

AuFe-Cr

Electrical and digital signals:

FP Axxx

FN Axxx

FT Axxx

Ampere DC (sampling rate 1 kHz) (Example)

Milliamperes DC

Percent (4 / 20mA DC)

Ohms

Ohms

Frequency

Pulses / measuring cycle

Digital interface

Digital input

Capacitive humidity sensors:

Rel: humidity FH A646

Rel: humidity with TC FH A646-R

Dew-point temperature

Mixture ratio

Partial vapor pressure

Enthalpy

Psychrometer

Wet temperature

Relative humidity

Dew-point temperature

Mixture ratio

Partial vapor pressure

Enthalpy

FN A846

range

5.0 to

5.0 to

-200.0 to +850.0

-200.00 to +400.00

-8.000 to + 65.000

-60.00 to + 240.00

-50.00 to +125.00

-200.0 to +1370.0

-200.0 to +1300.0

-200.0 to +900.0

-200.0 to +1000.0

-200.0 to +600.0

-200.0 to +400.0

0.0 to +1760.0

0.0 to +1760.0

+400.0 to +1800.0

-270.0 to +60.0

Millivolts DC

Millivolts 1 DC

Millivolts 2 DC

Volts DC

-10.0 to +55.0

-26.0 to +26.0

-260.0 to +260.0

-2.6 to +2.6

Volts DC -26 to +26

For measuring bridges Supply 5 V (Example) -26.0 to +26.0

For potentiometers Supply 2.5 V

Volt AC (50 Hz to 2 kHz) (Example)

Volt AC (11 Hz to 250 Hz) (Example)

Ampere AC (11 Hz to 250 Hz) (Example)

Volts DC (sampling rate 1 kHz) (Example)

-2.6 to +2.6

0 to +26

0 to +400

0 to +10.00

0 to +400

0 to +10.00

-32.0 to +32.0

0.0 to 100.0

0.00 to 500.00

0.0 to 5000.0

0 to 15000

0 to 65000

0 to 65000

0.00 to 100.00

98.0

98.0

-25.0 to +100.0

0.0 to 500.0

0.0 to 1013.2

0.0 to 400.0

0.00 to +100.00

0.0 to +100.0

-25.0 to +100.0

0.0 to 500.0

0.0 to 1013.2

Units Resolution

°C

°C

°C

°C

°C

°C

°C

°C

°C

°C

°C

°C

°C

°C

°C mV mV mV

*

V mV

*

V

V

A

V

A

*

%

*

*

Hz

%

%H

%H

°C g/kg mbar kJ/kg

°C

%H

°C g/kg mbar

0.0 to 400.0 kJ/kg

0.01 K

0.001 K

0.1 K

0.01 K

V

1 µV

V

mA

0,01 %

0,1 %

0,1 %

0.1 K

±0,5 %

±0.2 K

0.1 g/kg ±0.5 % of measured value

0.1 mbar ±0.1 mbar ±0.1 % of measured value

0.1 kJ/kg

0.1 %

0.1 kJ/kg

0.1 K

0.1 K

0.1 K

0.1 K

1 µV

1 µV

0.01 mV

1 mV

0.1 V

1 V

0.01 A

1 V

0.01 A

1 Hz

0.01 K

0.1 K

0.1 g/kg

-

0.1 K ±0.05 K ±0.05 % of measured value ZA 9030 FS1/4

±0.05 K

±0.002 K

±0.05 K

±0.05 K

±0.3 K

±0.3 K

±0.3 K

±0.1 K

0.1 mV

0.1 mV

1 µA

0.01 Ω

0.1 Ω

±0.5 % of measured value

ZA 9846 AK

±0.05 K

±1,0 %H

±0.2 K

±0.5% of measured value

ZA 9030 FS2 / 5

ZA 9030 FS7

ZA 9030 FS3 / 6

ZA 9040 FS

0.1 K ±0.05 K ±0.05 % of measured value ZA 9020 FS

0.1 K ±0.05 K ±0.05 % of measured value ZA 9021 FSN

0.1 K ±0.05 K ±0.05 % of measured value ZA 9021 FSL

0.1 K ±0.05 K ±0.05 % of measured value ZA 9021 FSJ

0.1 K ±0.05 K ±0.05 % of measured value ZA 9000 FSU

0.1 K ±0.05 K ±0.05 % of measured value ZA 9021 FST

ZA 9000 FSS

ZA 9000 FSR

ZA 9000 FSB

ZA 9000 FSA

ZA 9000 FS0

ZA 9000 FS1

ZA 9000 FS2

– ZA 9000 FS3

ZA 9602 FS

ZA9650 FS1V

- ZA9025 FS3

ZA 9603 AK3

ZA 9903 AB5

ZA 9904 AB2

ZA 9900 AB5

ZA 9901 AB4

– ZA 9601 FS1

ZA 9601 FS2

– ZA 9003 FS

– ZA 9003 FS2

ZA 9909 AK1U

ZA 9909 AK2U

ZA 9919 AKxx

ZA 9000 ES2

0.1 mbar ±0.1 mbar ±0.1% of measured value

±0.5% of measured value.

* Data may vary depending on device. (see relevant device data sheet)

01.06

ALMEMO

®

Measuring Instruments

Sensor type Type

Flow sensors

Rot. vane, snap-on head FV AD15-Sx (e.g.) 0.50 to 40,00

Rotating vane Macro

Water turbine

FV AD15-MA1

FV AD15-WM1

Dynamic pressure sensor FD A602-S1K

Dynamic pressure sensor FD A602-S6

Hot-wire anemometer FV A935-TH4

Hot-wire anemometer FV A935-TH5

Hot-wire anemometer FV A605-TA1

Hot-wire anemometer FV A605-TA5

Chemical probes

Conductivity

O

O

O

2

2

2

in gases

O

3

in gases

CO probe

CO

2

in gases pH probe

Redox probe

Lux measuring probe

Lux measuring probe

Lux measuring probe

UV measuring probe

FY A641-LF (e.g.)

dissolved saturation FY A640-O2

dissolved, concentr: FY A640-O2

FY 9600-O2

FY 9600-O3

FY A600-CO (e.g.)

0.10 to 20.00

0.00 to 5.00

0.5 to 40.0

1.8 to 90.0

0 to 2.000

0 to 20.00

0.01 to 1.000

0.15 to 5.00

0 to 20.000

0 to 260

0.0 to 40.0

1 to 100

0 to 300

0 to 300

FY A600-CO2 (e.g.) 0.000 to 2.500

FY96PH-Ex

FY96RX-Ex

Optical radiation (Examples)

FL A613-VL

FL A603-VL2

FL A603-VL4

FL A613-UV

0.0 to 14.00

0.0 to 2600.0

0 to 260000

0.05 to 12500

1 to 250000

0 to 87.00

Units

m/s m/s m/s m/s m/s m/s m/s m/s m/s mS

% mg/l

% ppb ppm

% pH mV lux lux lux

W/m

2

Resolution

0.01 m/s

0.01 m/s

0.01 m/s

0.1 m/s

0.1 m/s

0.001 m/s

0.01 m/s

0.001 m/s

0.01 m/s

0.001 mS

1%

0.1 mg/l

1%

20 ppb

1 ppm

-

± 0.1 m/s

± 0.1 m/s

0,01% ±0.2% of measured value

0.01 pH

0.1 mV

1 lux

0.01 lux

1 lux

0.01 W/m

2

UVA measuring probe FL A603-UV24 0.0004 to 100 mW/cm

2

0.1 µW/cm

2

±0.2% of measured value

±0.2 mg/l

ZA 9610 AKY4W

ZA 9610 AKY5W

Radiometric probe

Photosynthesis probe

FL A603-RW4 0.00004 to

FL A603-PS5

Other connectable sensors / transducers (Examples)

10 mW/cm

0.0002 to 100 mmol/m

2

2

0.01 µW/cm s 0.1µmol/m

2

2 s

Heat flow plates FQ Axxx

Moisture content probe FH A696-MF

Differential pressure

Barometer

Alarm value

Thermal coefficient

FD A612-SR

FD A612-SA

ZA 9909-AK2U

M (q) / M (∆T)

-260.0 to +260.0

0 to 50.0

0 to 1000

0.0 to 1050 mbar

Pressure transducer FDA FD A602-xx (e.g.) 0.00 to 10.00

Force transducer FK Axxx (e.g.)

Displacement transducer FW Axxx(e.g.)

Tachometer

Function values

Differential

Maximum value

Minimum value

Average value over time

Average value over measuring point

Summation over measuring points

Total number of pulses ZA 9909-AK2U

Pulses / print cycle

Wet-bulb globe temperature (WBGT) (0.1 TD + 0.7 TW +0.2 TG)

Measured value

FU A919-2

0.0 to 50.00

0.0 to 150.00

8 to 30000

0 to 65000

0 to 65000

0 to 65000

0.0 to 100.00 mV

% mbar bar kN mm rpm

%

0.01 mV

0,1%

0.1 mbar

0.1 mbar

0.01 bar

0.01 kN

0.01 mm

1 rpm

-

ZA 9007 FS

ZA 9909 AK4U

Cold junction temperature

Number of averaged values

Volume flow 0 to 65000 m

3

/h 1 m

3

°C

/h

01.07

ALMEMO

®

Measuring Instruments

Outputs

ALMEMO

®

socket A1

ALMEMO

®

socket A2

Digital interface

Analog output

Networking

Saving data

Analog output

Trigger input

Relay output

Relay output

Baud rates 150, 300, 600, 1200, 2400, 4800, 9600 baud, 57.6, 115.2 kilobaud

Data : 8 bit serial, 1 start bit, 1 stop bit, no parity

ALMEMO

®

data link via USB, RS232, Ethernet wireless link via Bluetooth or RS422

(see chapter „Networking“)

ALMEMO

®

analog cable and analog interface

(see chapter „Output modules“)

ALMEMO

®

network cable or wireless via Bluetooth

(see chapter „Networking“)

ALMEMO

®

memory connector with memory card

(see chapter „General accessories“)

ALMEMO

®

analog cable and analog interface

(see chapter „Output modules“)

ALMEMO

®

trigger cable and trigger interface

(see chapter „Output modules“)

ALMEMO

®

relay cable and relay interface

(see chapter „Output modules“)

ALMEMO

®

relay cable and relay interface

(see chapter „Output modules“e

Mains adapter and DC supply cable

see chapter „General accessories“

Input connector

ALMEMO

®

plug

In the ALMEMO

®

measuring system, depending on the sensor and measuring instrument, up to 4 measuring channels can be accessed at any one measuring input.

The patented ALMEMO

®

plug incorporates 6 screw terminals - 2 for the sensor‘s power supply and 4 for its measuring signal. With Pt100 sensors using 4-conductor circuitry all 4 free connections will be required for the measuring signal.

Only one sensor of this type can be connected therefore per measuring input. Electrical signals only require 2 connections for the measuring signal. One plug can thus acquire two different measuring signals over just one measuring channel. An atmospheric humidity sensor can example usually be combined with a temperature sensor. The associated operands (e.g. dew point, mixture ratio, partial vapor pressure, enthalpy) are programmed in the plug as additional measuring channels. Up to maximum four measuring channels can be output per measuring input.

ALMEMO

®

D6 plugs for digital sensors

• The digital ALMEMO

®

D6 sensor can be connected to any ALMEMO

®

measuring instrument without in any way affecting its measuring accuracy. The A/D converter incorporated in the ALMEMO

®

D6 sensor is exclusively responsible for the measuring accuracy of the whole system.

• The digital ALMEMO

®

D6 sensor is calibrated without involving the ALMEMO

®

measuring instrument

(DAkkS / factory) and can be replaced or exchanged as and whenever necessary.

• The connecting cable for the digital ALMEMO

®

D6 sensor can be extended using pluggable extension cables quickly and easily and without any line losses. (see chapter „General accessories“)

• These digital extension cables provide high transmission reliability; they have no effect on measuring accuracy.

• The digital ALMEMO

®

D6 sensor can be connected via USB directly to a PC or be incorporated via Ethernet in an ALMEMO

®

network. Measured values can be processed directly using the AMR WinControl software package. (see chapter „Software“)

• These digital ALMEMO

®

D6 sensors can be confi gured (e.g. measuring range selection) directly on the PC using

USB adapter cable ZA1919AKUV (see page 04.05).

01.08

ALMEMO

®

Measuring Instruments

ALMEMO

®

measuring instruments, overview

Measuring inputs Expansions Display Graphics display Data logger functionIntegrated memory Interface / outputs Precision class Measuring rate (mops) max. Catalog page

Compact measuring instrument

ALMEMO

®

2450-1

ALMEMO

®

2450-1L

Basic measuring instrument

ALMEMO

ALMEMO

ALMEMO

ALMEMO

Professional measuring instrument

ALMEMO

®

2470-1S/-1SRH 1

ALMEMO

®

2470-2S

ALMEMO

®

2470-2

2

2

ALMEMO

ALMEMO

Precision measuring instrument

ALMEMO

®

2690-8A

ALMEMO

®

®

®

®

®

®

®

2490-1

2490-2

2490-1L

2490-2L

2590-2A

2590-4AS

2890-9

ALMEMO

®

710

1

1

5

9

2

4

1

2

1

2

ALMEMO

ALMEMO

ALMEMO

®

5690-1M09

ALMEMO

®

5690-2M09

ALMEMO

®

5790-2M09

ALMEMO

ALMEMO

ALMEMO

®

®

®

®

®

8590-9

8690-9A

5690-1CPU

5690-2CPU

5790-2CPU

10

9

9

4

4

4

4

4

4

4 4

4

4

A 10 65

4

4 4

4

4

A 10 65

4

4 4

A 10 65

4

4 4 4

A 10 65

4

4 4

4

4

A 10 65

4

9 opt.

9 opt.

9 opt.

4

4

4

4

4

4

4

4

4

4

4

4

4

4

B 10 65

4

4

B 10 65

4

4

4

4

4

4

4

4

C 2,5 35

C 2,5 35

B 10 65

B 10 65

AA 100 66 opt. 4

AA 100 66 opt. 4

AA 100 66 opt. 4

opt. 4 AA 100 66 opt.

opt. 4 AA 100 66 opt.

opt. 4 AA 100 66 opt.

4

4

AA 100 66 opt.

opt. 4 AA 100 66 opt.

4

4

4

4

4

4

4

4

4

01.12

01.12

01.14

01.14

01.14

01.14

01.16

01.16

01.16

01.19

01.19

01.22

01.24

01.26

01.29

01.29

01.32

01.32

01.32

opt.

4

4

4

AA 100 66 opt.

4

01.42

opt.

4 4

4

4

AA 100 66 opt.

4

01.42

opt.

4 4

4

4

AA 100 66 opt.

4

01.42

1

4 4

4

4

AA 100 66

4

01.52

ALMEMO

®

4390-2

Compact device (transmitter)

ALMEMO

®

2450-1R02 1

4 4

C 2,5 35

4

01.50

Basic device (transmitter)

ALMEMO

®

2490-1R02

ALMEMO

®

2490-2R02

Reference measuring instrument

ALMEMO

®

1020-2

ALMEMO

®

1030-2

ALMEMO

®

1036-2

2

2

2

1

2

4

4

4

4

4

4

4

4

4

4

4

4

4

B 10 65

B 10 65

AS 1,25 4

AS 1,25 1

AS 1,25 7

4

4

4

4

4

4

4

4

01.50

01.50

01.54

01.55

01.58

01.09

ALMEMO

®

Measuring Instruments

Measuring ranges, ALMEMO

®

2450, 2490, 2470, 2590A series

Sensor type / Measuring range

ALMEMO

®

series

Precision class

Type

2450

C

2490

B

2470

A

Temperature

Thermocouple sensor

NiCr-Ni Typ K (NiCr)

NiCroSil-NiSil Typ N (NiSi)

Fe-CuNi Typ L/J (FeCo/IrCo)

Cu-CuNi Typ U/T (CuCo/CoCo)

PtRh10-Pt Typ S (Pt10)

PtRh13-Pt Typ R (Pt13)

PtRh30-PtRh6 Typ B (EL18)

AuFe-Cr (AuFe)

Resistance temperature detectors

Pt100/1000 (P104, P204)

Ni100/1000 (N104)

NTC Typ N (NTC)

Heat flow

Atmospheric humidity

Capacitive with NTC

Digital temperature / humidity sensor

Digital temperature / humidity sensor

Psychrometric with NTC

Psychrometric with Pt100 (2 plugs)

Digital psychrometer

Dew point

Digital dewpoint sensor

Dew detector

Moisture in materials

FTA xxx

FPA xxx

FNA xxx

FQA xxx, FQADxx

FHA 646 xxx

FHAD 46x

FHAD 36 Rx

FNA 846

FPA 8363

FNAD46, FNAD463

FH A646 DTC1

FHA 9461

Water detection probe FHA 936 WD

Sensor for measuring moisture in materials FHA 696 MF

Moisture probe for wood FHA 636 MFx, FHA 696 MFS1

Material moisture sensor for granulates FHA 696 GF1

Moisture in the soil

Air flow

FDA 602 TM1

Rotating vanes for air

Differential pressure for Pitot tube

Thermo-anemometer probe

Thermo-electric flow sensor

** Smoothing is possible for 1 measuring channel

Pressure

FVAD 15 Sxxx, FVAD 15 MA1

FDA 602 S1K, FDA 602 S6K

FVAD 35 THxx

FVA 605 TAxx

7

*

Range

7

*

7

*

7

7

7

7

*

*

*

*

* An average value channel is not possible with flow measurement; (no start of continuous or cyclic measuring)

Pressure transducer for liquid and gaseous media FDA 602 Lxx

Temp.-compensated pressure transducer FD 8214

Differential transmitter

Digital pressure sensor

FDA 602 D

FDAD 33, FDAD 35M

Pressure transducer, for wall mounting FD 8612 DPS / APS / DPT

Barometric pressure

Barometric pressure, digital

Plug-in probe for differential pressure

Force

FDA 612 SA

FDAD 12 SA

FDA6 12 SR, FDA 602 SxK

7

7

7

7

7

Range

7

Range

7

7

7

7

7

7

7

7

Push / pull force FKA xxx

* Only temporary zero-setting is possible; (no final value adjustment)

Tachometer

FUA 9192 Tachometer

7

*

7

7

*

7

7

7

7

7

7

Range

Range

Range

Range

Range

7

7

7

7

7

Range

Range

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7 7

7

7

7

7

7 7

Function Function

Function Function

7 7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7 7

Function Function

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

**

**

**

**

7

*

7

7

7

7

7

7

7

7

01.10

2590A

A

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

ALMEMO

®

Measuring Instruments

Measuring ranges, ALMEMO

®

2450, 2490, 2470, 2590A series

Sensor type / Measuring range

ALMEMO

®

series

Precision class

Type

Displacement

Displacement transducer, potentiometric FWA xxx T

Displacement gauge, potentiometric

Flow

FWA xxx TR

* Only temporary zero-setting is possible; (no final value adjustment)

Axial turbine flowmeter for liquids

Flow sensor with temperature

Electrical variables

FVA 915 VTHxxx

FVA 645 GVx

Split-core-type transformer for AC current FEA 6042, FEA 604 MN,

ALMEMO

®

measuring modules for

FEA 6044 N

DC voltage, DC

AC voltage, AC

Meteorology

ZA 9900 ABx, ZA 9901 ABx,

ZA 9903 ABx, ZA 9904 ABx

Meteo Multi (2 plugs)

Wind velocity sensor

Wind direction sensor

Rainfall and precipitation sensor

Rainfall detector

Radiation probe head

Star pyranometer

* for ALMEMO

®

2470-2 - function missing

Indoor climate and air conditioning

Globe thermometer

Optical radiation

Radiation sensor

Radiation sensor

Radiation sensor

Digital color temperature sensor

Water analysis pH One-Bar Measuring Chain

Redox-One-Bar Measuring Chain

Conductivity probe

Oxygen sensor

Gas concentrations in air

FMA 510, FMA 510H

FVA 615-2

FVA 614

FRA 916, FRA 916 H

FRA 616 D

FLA 613 x

FLA 628 S

FPA 805 GTS

FLA 603 x

FLA 613 x

FLA 623 x

FLAD 23 CCTx

FY 96 PH x

FY 96 RXEK

FYA 641 LF xxx

FYA 640 O2

Digital carbon dioxide sensor, hand-held FYAD 00 CO2

Carbon dioxide probe

Carbon monoxide probe

Oxygen probe

Ozone measuring transducer

Gas probes

Infra-red temperature measurement

FYA 600 CO2

FYA 600 CO

FYA 600 O2

FYA 600 O3

FYA 600 Ax

ALMEMO

®

infra-red probe head

Infra-red probe

Hand-held IR device

Digital IR sensor

* Emissivity cannot be modified

FIA 844

MR 7838, MR 7842

MR 781420 SB

FIAD 43

2450

C

7

*

7

*

7

7

7

7

*

7

7

Range

7

7

7

7

Adjustment

Adjustment

Range

Adjustment

7

Range

7

Adjustment

7

7

2490

B

7

*

7

*

7

7

7

7

7

7

7

7

7

7

Function

7

7

7

7 7

Function Function

7

7

7

7

7

7

7

7

7

7

*

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

2470

A

7

*

7

*

7

7

7

7

7

7

7

7

7

7

*

7

7

7

7

7

7

7

7

7

7

7

7

7

7

*

7

7

7

7

7

7

7

7

2590A

A

7

7

7

7

7

7

7

7

Prerequisites missing for perfect functioning

- Range: Measuring range missing or restricted -> Measured value cannot be shown.

- Function: Function missing for showing sensor-specific measured data (e.g. average value / cycle) or for necessary programming

- Adjustment: Measured value adjustment of this sensor is not possible (pressure, force, displacement, O2, pH, conductivity)

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

7

01.11

ALMEMO

®

Measuring Instruments

ALMEMO

®

2450

Compact ALMEMO® measuring instrument

1 measuring input, over 35 measuring ranges

Technical data and functions Serie ALMEMO

®

2450

• Generously dimensioned 2-row segment display including units

• Easy and convenient to operate by means of 7 keys.

• Over 35 measuring ranges for

- Thermocouple and NTC sensors

For the customer’s own sensors ready-to-use ALMEMO

®

connectors are available. (see chapter 07)

- Atmospheric humidity sensor, capacitive, dewpoint sensor,

water detection probe, moisture in wood FHA636MF

(see chapter 13)

- Pressure transducer FDA602L/D, FD8214, FD8612,

Tachometer, turbine fl owmeter (see chapter 10)

Current clamps FEA604, Voltage / current measuring

modules ZA990xAB (see chapter XREF)

- Meteorological radiation probe heads FLA613

(see chapter XREF)

- Carbon dioxide sensor FYAD00CO2, Carbon monoxide

probe and ozone probe (see chapter 15),

• ALMEMO

®

plugs with multi-point adjustment are supported.

• Measuring functions

Measured value, zero-setting, saving of maximum / minimum values, hold function

• Test functions

Segment monitoring, range monitoring, sensor breakage indication, battery voltage check and display.

Technical data, ALMEMO

®

2450 series

Measuring input

Precision class

1 ALMEMO

®

socket

C (see page 01.05)

Measuring rate 2.5 mops

Measuring ranges (see 01.06 / 01.07) NiCr-Ni(K),

Fe-CuNi(J),

NTC

Voltage

NiCroSil-NiSil(N),Fe-CuNi(L),

Cu-CuNi(U), Cu-CuNi(T), PtRh10-Pt(S),

Current

-200 to +950 °C

-20 to +100 °C

-26 to +26 mV, -260 to +260mV, 0 to 2.6V

0 to 26 mA, 4 to 20 mA

Double connectors with 2 x differenti al voltage / differential current

(input D - B) are not possible.

Humidity, capacitive 0 to 100 % RH, (% RH, HcRH, HRH)

Dew point, mixture ratio, partial vapor pressure, enthalpy, rotating vanes, digital process (0 / 100 %), frequency, pulse, rotational speed, digital

Resolution (see page 01.06 / 01.07)

Linearization accuracy (see page 01.06 / 01.07)

Standard equipment

LCD 7 segments

16 segments

Keypad

Measured value 5 characters, 15 mm

Function 4½ characters, 9 mm

Units 2 characters, 9 mm

9 symbols

7 silicone keys

Power supply

Battery set 3 AA alkaline batteries

Current consumption approx. 10 mA without input modules

Housing ABS (max. 70 °C) 127 x 83 x 42 mm (LxWxH)

Operating temperature -10 to +60 °C

Atmospheric humidity (ambient) 10 to 90 % RH (non-condensing)

ALMEMO

®

2450 series, accessories

Rubberized impact protection, gray

DIN rail mounting

ZB2490GS2

ZB2490HS

Magnetic fastening

Instrument case

Order no.

ZB2490MH

ZB2490TK2

DIN rail mounting Rubberized protection Magnetic fastening

01.12

ALMEMO

®

2450-1

ALMEMO

®

Measuring Instruments

ALMEMO

®

2450-1L

Compact measuring instrument with interface. Runs in battery mode or via mains unit

Technical data and functions

• Technical data and functions, as for ALMEMO

®

2450 series

• 2 ALMEMO

®

output sockets, suitable for all interface cables, network cables, trigger / relay cables

• Complete sensor and device programming via interface

• ALMEMO

®

DC socket for mains adapter.

Compact measuring instrument with interface. Runs in battery mode

Technical data and functions

• Technical data and functions, as for ALMEMO

®

2450 series

Technical data

Technical data, as for ALMEMO

®

2450 series

Sensor power supply

Option U

9 V, maximum 0.5 A

9 V, maximum 70 mA

Power supply

Mains adapter

10 to 30 VDC not electr. isolated

ZA1312NA7 230 VAC to 12 VDC, 1 A

Outputs 2 ALMEMO

®

sockets, suitable for all interface cables

with option OA2450I only Internal RS485 interface, electrically isolated, via DC socket

Technical data

Technical data as for ALMEMO

®

2450 series

Sensor power supply 9 V, maximum 0.5 A

Accessories

Mains adapter 12 V, 1 A, with ALMEMO

®

plug

DC adapter cable

10 to 30 VDC, 12 V / 0.25 A, electrically isolated

Connecting cables

Order no.

ZA1312NA7

ZA2690UK

USB data cable, electrically isolated

Ethernet data cable, electrically isolated

V24 data cable, electrically isolated

ZA1919DKU

ZA1945DK

Analog output cable, -1.25 to +2.0 V, 0.1 mV / digit

ZA1601RK

ZA1909DK5

Network technology, Bluetooth modules (see chapter „Networking“)

Option

Order no.

Power supply, electrically isolated, 10 to 30 VDC, 80 mA including ALMEMO

®

plug for DC socket

OA2450U

RS485 interface, internal including ALMEMO

®

DC socket option

OA2450I

Analog outputs (socket P0), electrically isolated, integrated internally (see page 01.05) ALMEMO

®

transmitter

Measuring instrument IP54

(if water-proof plugs are used)

OA2450W

Option

Measuring instrument IP54

(if water-proof plugs are used)

Order no.

OA2450W

Standard delivery

Order no.

Batteries, operating instructions, manufacturer’s test certifi cate

Compact measuring instrument ALMEMO

®

2450-1

MA24501

Standard delivery

Order no.

Batteries, operating instructions, manufacturer’s test certifi cate

Compact measuring instrument ALMEMO

®

2450-1L

MA24501L

DAkkS or works calibration KE90xx, electrical, for measuring instrument (see chapter „Calibration certifi cates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

01.13

ALMEMO

®

Measuring Instruments

ALMEMO

®

2490

ALMEMO

®

basic measuring instrument

Ideal for all sorts of application, quick and easy to operate

1 or 2 measuring inputs, over 65 measuring ranges

Technical data and functions ALMEMO

®

2490 series

• Generously dimensioned 2-row static 7 / 16 segment display including units

• Easy and convenient to operate by means of 7 keys

• Over 65 standard measuring ranges

• Memory suffi cient for 100 measured values, can be called up and viewed in the display

• Good measuring accuracy, measuring rate up to 10 measuring operations per second (mops)

• Support for ALMEMO

®

plugs with multi-point adjustment, special linearization, and special measuring ranges

• Measuring functions

Measured value, zero-setting, sensor adjustment, saving of maximum / minimum values, memory for 100 values, cold junction compensation, and temperature compensation

• Test functions

Segment monitoring, range monitoring, sensor breakage indication, battery voltage check and display

Technical data ALMEMO

®

2490 series

Precision class

Measuring rate

B (see page 01.05)

2.5 / 10 measuring operations per second

Measuring ranges as on page XREF - but

Milliamperes DC -26 to +26 mA

Battery set 3 AA alkaline batteries

Current consumption approx. 20 mA without input modules

Standard equipment

LCD 7 segments

16 segments

Keypad

Housing

Measured value 5 characters, 15 mm

Function 4½ characters, 9 mm

Units 2 characters, 9 mm

9 symbols

7 silicone keys

ABS (maximum 70 °C)

127 x 83 x 42 mm (LxWxH)

ALMEMO

®

2490 series, accessories

DIN rail mounting

Rubberized impact protection, green

ZB2490HS

ZB2490GS1

Magnetic fastening

Instrument case

DIN rail mounting Rubberized protection Magnetic fastening

Order no.

ZB2490MH

ZB2490TK2

01.14

ALMEMO

®

2490-1 / -2

ALMEMO

®

Measuring Instruments

ALMEMO

®

2490-1L / -2L

ALMEMO

®

2490-1 ALMEMO

®

2490-2 ALMEMO

®

2490-1L ALMEMO

®

2490-2L

Basic measuring instrument with interface

Runs in battery mode or via mains unit

Technical data and functions

• Technical data and functions, as for ALMEMO

®

2490 series

• 2 ALMEMO

®

output sockets, suitable for all interface cables, network cables, trigger / relay cables

• Complete sensor and device programming via interface

• ALMEMO

®

DC socket for mains adapter.

Technical data

Technical data, as for ALMEMO

®

2490 series

Measuring input

2490-1

2490-2

1 ALMEMO

®

input socket

2 ALMEMO

®

input sockets, el. isol., with semicond. relays (50V)

Additional channels

Sensor power supply

Option U

4 function channels, device-internal

9 V, maximum 0.5 A

9 V, maximum 70 mA

Power supply

Mains adapter

10 to 30 VDC not electr. isolated

ZA1312NA7

230 VAC to 12 VDC, 1 A

Outputs 2 ALMEMO

®

sockets, suitable for all interface cables

with option OA2490I only RS485 interfac

Accessories

Mains adapter 12 V, 1 A, with ALMEMO

®

plug

DC adapter cable

10 to 30 VDC, 12 V / 0.25 A, electrically isolated

Connecting cables

Order no.

ZA1312NA7

ZA2690UK

USB data cable, electrically isolated

Ethernet data cable, electrically isolated

V24 data cable, electrically isolated.

ZA1919DKU

ZA1945DK

Analog output cable, -1.25 to +2.0 V, 0.1 mV / digit

ZA1601RK

ZA1909DK5

Network technology, Bluetooth modules (see chapter „Networking“)

Option

Order no.

Power supply, electrically isolated, 10 to 30 VDC, 80 mA including ALMEMO

®

plug for DC socket

RS485 interface, internal, including option U

Analog outputs, electrically isolated, integrated internally

(see page 01.50) ALMEMO

®

transmitter

Measuring instrument IP54

(if water-proof plugs are used)

OA2490U

OA2490I

OA2490W

Basic measuring instrument

Runs in battery mode

Technical data and functions

• Technical data and functions, as for ALMEMO

®

2490 series

Technical data

Technical data, as for ALMEMO

®

2490 series

Measuring inputs

2490-1L

2490-2L

Sensor power supply

Outputs

Option

Measuring instrument IP54

(if water-proof plugs are used)

1 ALMEMO

®

input socket

2 ALMEMO

®

input sockets, el. isol., with semicond. relays (50 V)

9 V, maximum 0.5 A

None

Order no.

OA2490W

Standard delivery

Order no.

Batteries, operating instructions, manufacturer’s test certifi cate

Basic measuring instrument ALMEMO

®

2490-1 MA24901

Basic measuring instrument ALMEMO

®

2490-2 MA24902

Standard delivery

Order no.

Batteries, operating instructions, manufacturer’s test certifi cate

Basic measuring instrument ALMEMO

Basic measuring instrument ALMEMO

®

®

2490-1L MA24901L

2490-2L MA24902L

DAkkS or works calibration KE90xx, electrical, for measuring instrument (see chapter „Calibration certifi cates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

01.15

ALMEMO

®

Measuring Instruments

ALMEMO

®

2470

ALMEMO

®

professional measuring instrument with data logger function

Functions for all application areas, 1 or 2 measuring inputs

Also with integrated sensor for temperature, atmospheric humidity, atmospheric pressure

Technical data and functions, ALMEMO

®

2470 series

• new

Segmented color display with bright, white illumination

• Clear and easy-to-understand display of programming and measured values in 5 different colors and alarm display on a red background

• new

In the event of a limit value being overshot / undershot various freely confi gurable alarm messages are available, namely acoustic signal, visual LED signal, alarm display on a red background.

• new

With the 2470-1S /-2S these alarm messages are also confi gurable for long-term recording; in sleep mode the messages remain active and the most recent measured value is displayed continuously.

Good measuring accuracy, measuring rate up to 10 measuring operations per second (mops)

• More than 65 standard measuring ranges

• Support for ALMEMO® plugs with multi-point adjustment, special linearization, and special measuring ranges

• Easy and convenient to operate by means of 7 keys, with confi gurable locking for keys and functions

• Measuring functions : Maximum and minimum values, measured value smoothing, zero-setting, sensor adjustment

• Programming functions : Limit values, sensor correction with base value and factor

• All ALMEMO

®

functions programmable via interface

• Modern, compact housing (IP54 option)

Technical data, ALMEMO

®

2470 series

Precision class

Measuring rate

A (see page 01.05)

2.5 / 10 measuring operations per second

Sensor power supply Battery mode

With mains adapter

Sensor voltage 6 V, 400 mA

9 V, 300 mA and 12 V, 200 mA

12 V, 400 mA

Standard equipment

Display 16 segments

7 segments

Keypad

Measured value 5 characters, 15 mm

Units 2 characters, 9 mm

Function 4½ characters, 9 mm

21 symbols, Illumination 2 RGB LEDs

7 silicone keys

Power supply

Mains adapter

1 ALMEMO

®

DC socket

ZA1312NA7 230 VAC to 12 VDC, 1A, electrically isolated

DC adapter cable, el. isol. ZA2690UK 10 to 30 V, 0.25 A

Current consumption (without input and output modules)

Active without illumination approx. 12 mA

Active with illumination approx. 30 mA

Sleep mode approx. 60 μA

Housing 127 x 83 x 42 mm (LxWxH)

ABS (maximum 70 °C), 290g

ALMEMO

®

2470 series, accessories

Rubberized impact protection, gray

Instrument case

Mains adapter 12 V / 1 A

ZB2490GS2

ZB2490TK2

ZA1312NA7

DC cable 10 to 30 V, 12 V / 0.25 A, electr. isol.

DIN rail mounting

Magnetic fastening

Order no.

ZA2690UK

ZB2490HS

ZB2490MH

Automatic alarm (red background). Display shows incorrect measured value

01.16

Dual display

1. Humidity Measured value overshoots limit value (red).

2. Temperature

1. Measured value is inside limit values (green).

2. Peak value MAX overshoots limit value (red)

Programming of

1. Save-to-memory cycle

2. Sleep mode

ALMEMO

®

2470-1S

ALMEMO

®

Measuring Instruments

ALMEMO

®

2470-1SRH

Professional measuring instrument,

1 measuring input

Data logger with integrated memory

Technical data and functions

• Technical data and functions as for ALMEMO

®

2470 series

• Data logger functions: Internal EEPROM, memory cycle, real-time clock

• Long-term recording in sleep mode with AA batteries

• Operating time up to 1.5 years with memory cycle of 15 minutes and temperature / humidity sensor..

Professional measuring instrument,

1 measuring input, Data logger with integrated memory, Integrated sensor for temperature, atmospheric humidity, atmospheric pressure

Technical data and functions

• Technical data and functions, as for ALMEMO

®

2470 series

• Data logger functions

• Internal EEPROM, memory cycle, real-time clock

• Long-term recording in sleep mode with AA batteries

• Operating time up to 1.5 years with memory cycle of 15 minutes and temperature / humidity sensor.

Technical data

Measuring inputs

Outputs

Memory, internal

Date and time-of-day

Power supply

Connecting cable

1 ALMEMO

®

input socket

ALMEMO

®

DC socket for mains adapter or USB cable with supply

ZA 1919 DKU5

EEPROM suffi cient for 100,000 measured values

Real-time clock, buffered by device battery

3 AA batteries

USB data cable with 5-V power supply

Order no.

ZA1919DKU5

Technical data

Measuring inputs

Outputs

1 ALMEMO

®

input socket

ALMEMO

®

DC socket for mains adapter or USB cable with supply

ZA 1919 DKU5

Memory, internal

Date and time-of-day

Power supply

EEPROM suffi cient for 100,000 measured values

Real-time clock, buffered by device battery

3 AA batteries

Digital atmospheric pressure sensor, integrated in the measuring instrument Measuring range

Accuracy

700 to 1100 mbar

±2.5 mbar (at 0 to 65 °C)

Digital sensor for measuring temperature / atmospheric humidity

FH0D 462 plugged in on the measuring instrument

General description and other technical data (see chapter „Atmospheric humidity“)

Connecting cable

USB data cable with 5-V power supply

Order no.

ZA1919DKU5

Option

Measuring instrument IP54

(if water-proof plugs / sensors are used)

Order no.

OA2470W

Option

Measuring instrument IP54

(if water-proof plugs / sensors are used)

Order no.

OA2470W

Standard delivery

Order no.

Batteries, operating instructions, manufacturer’s test certifi cate

Professional measuring instrument ALMEMO

®

2470-1S

MA24701S

Standard delivery

Order no.

Batteries, digital plug-in sensor for temperature / atmospheric humidity, operating instructions, manufacturer’s test certifi cate

Professional meas. instrument ALMEMO

®

2470-1SRH

MA24701SRH

DAkkS or works calibration KE90xx, electrical, for measuring instrument (see chapter „Calibration certifi cates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

01.17

ALMEMO

®

Measuring Instruments

ALMEMO

®

2470-2

ALMEMO

®

2470-2S

Professional measuring instrument,

2 measuring inputs

Technical data and functions

• Technical data and functions, as for ALMEMO

®

2470 series

• Power supply, 3 AA rechargeable NiMH batteries, with charging via the device itself.

Technical data

Measuring inputs

Additional channels

Outputs

2 ALMEMO

®

input sockets el. isol., with semicond. relays (50 V)

4 channels, device-internal

(e.g. difference)

ALMEMO

®

sockets A1 and A2, suitable for all output modules

(analog, data, trigger, relay cables, etc.)

(see chapter „Networking“)

Individual value memory 99 individual measured values

Power supply 3 AA rechargeable NiMH batteries

Integrated charge circuitry

Professional measuring instrument,

2 measuring inputs,

Data logger with internal memory

Technical data and functions

• Technical data and functions, as for ALMEMO

®

2470 series

• Power supply, 3 AA rechargeable NiMH batteries, with charging via the device itself

• Data logger functions: Internal EEPROM or external memory connector (accessory), memory cycle, real-time clock

• Long-term recording in sleep mode, internal memory, AA rechargeable NiMH batteries. Operating time up to 1 year with memory cycle of 15 minutes and temperature / humidity sensor.

Technical data

Measuring inputs

Additional channels

Outputs

2 ALMEMO

®

input sockets el. isol., with semicond. relays (50 V)

4 channels, device-internal

(e.g. difference)

ALMEMO

®

sockets A1 and A2, suitable for all output modules

(analog, data, trigger, relay cables, etc.)

(see chapter „Networking“)

Memory, internal EEPROM suffi cient for 100,000 measured values

Date and time-of-day Real-time clock, buffered by device battery

Power supply 3 AA rechargeable NiMH batteries

Integrated charge circuitry

Accessories

Memory connector with micro SD card

Order no.

ZA1904SD

Connecting cables

Order no.

USB data cable, electrically isolated

USB data cable with 5-V power supply

V24 data cable, electrically isolated

Ethernet data cable, electrically isolated

ZA1919DKU

ZA1919DKU5

ZA1909DK5

ZA1945DK

Analog output cable, -1.25 to +2.0 V, 0.1 mV / digit

ZA1601RK

Trigger and relay cable (2 relays, 500 mA, 50 V)

ZA1006EKG

Network technology, Bluetooth modules (see chapter „Networking“)

Option

Measuring instrument IP54

(if water-proof plugs / sensors are used)

Order no.

OA2470W

Connecting cables

Order no.

USB data cable, electrically isolated

USB data cable with 5-V power supply

V24 data cable, electrically isolated

Ethernet data cable, electrically isolated

ZA1919DKU

ZA1919DKU5

ZA1909DK5

ZA1945DK

Analog output cable, -1.25 to +2.0 V, 0.1 mV / digit

ZA1601RK

Trigger and relay cable (2 relays, 500 mA, 50 V)

ZA1006EKG

Network technology, Bluetooth modules (see chapter „Networking“)

Option

Measuring instrument IP54

(if water-proof plugs / sensors are used)

Order no.

OA2470W

Standard delivery

Order no.

Rechargeable batteries, operating instructions, manufacturer‘s test certifi cate, carry case, mains unit

Professional measuring instrument ALMEMO

®

2470-2

MA24702KN

Standard delivery

Order no.

Rechargeable batteries, operating instructions, manufacturer‘s test certifi cate, carry case, mains unit

Professional measuring instrument ALMEMO

®

2470-2S

MA24702SKN

DAkkS or works calibration KE90xx, electrical, for measuring instrument (see chapter „Calibration certifi cates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

01.18

ALMEMO

®

2590A

ALMEMO

®

Measuring Instruments

ALMEMO

®

professional measuring instrument with data logger function,

Comprehensive range of functions for all application areas,

Graphics display for showing measured values and programming,

2 or 4 measuring inputs

Technical data and functions, ALMEMO

®

2590A series

• New variant, further developed

• Good measuring accuracy, measuring rate up to 10 measuring operations per second (mops)

• Over 65 standard measuring ranges

• Support for ALMEMO

®

plugs with multi-point adjustment, special linearization, and special measuring ranges

• Graphics display with white illumination, easy and convenient operation by means of 4 soft-keys and cursor block

• Clear and easy-to-understand menu system

• 3 measuring menus (1 menu can be freely confi gured by user from a range of 50 functions), measured values displayed numerically, 1 to 12 measured values can be displayed in two sizes or graphically in bar chart form.

• Intelligent sensor readings with sensor-specifi c functions

Cold junction compensation, temperature compensation, and atmospheric pressure compensation

• Measuring functions

Measured value, zero-setting, setpoint adjustment

• Function menus

Maximum value, minimum value, memory suffi cient for 99 measured values, average value over time / individual values

/ measuring points, smoothing, volume fl ow with center point measuring, two-point adjustment, scaling, data logger with confi guration menus

• Option VN

Volume fl ow determined from matrix measuring as per DIN

EN 12599

• Programming menus for clear and easy-to-understand sensor programming, range, units, designation, right through to special functions, confi guration of device parameters and of output modules

• Choice of languages : German, English, French (other options also available)

• 2 ALMEMO

®

output sockets, suitable for digital interfaces, analog output, trigger input, alarm contacts, memory card

• External memory connector with micro SD can simply be plugged in.

• Sleep mode for long-term recording

Technical data ALMEMO

®

2590A series

Precision class

Measuring rate

Additional channels

Sensor power supply

Outputs

Standard equipment

Display

Keypad

Date and time-of-day

A (see page 01.05)

2.5 / 10 measuring operations per second

4 function channels, device-internal

6 / 9 / 12 V, maximum 0.5 A

2 ALMEMO

®

sockets, suitable for all output modules (analog / data / trigger / relay cables, memory, etc.)

Graphics display, 128 x 64 pixels, 8 rows

Illumination 2 white LEDs

7 silicone keys (of which 4 soft-keys)

Real-time clock, buffered by battery

Power supply

Battery set

Mains adapter

3 AA alkaline batteries

ZA1312NA7

230 VAC to 12 VDC, 1 A electrically isolated

DC adapter cable, electrically isolated ZA2690-UK 10 to 30 V, 0.25 A

Current consumption (without input and output modules)

Active mode

With illumination

Sleep mode approx. 12mA approx. 32 mA approx. 0.05 mA

Housing 127 x 83 x 42 mm (LxWxH)

ABS (maximum 70 °C) 290 g

01.19

ALMEMO

®

Measuring Instruments

Serie ALMEMO

®

2590A

Accessories

Memory connector with micro SD (see page 06.02)

Mains adapter 12 V / 1 A

DC adapter cable, 10 to 30 VDC, 12 V / 0.25 A, electrically isolated

Rubberized impact protection, green

Magnetic fastening

DIN rail mounting

Instrument case

Network technology, Bluetooth modules (see chapter „Networking“)

Connecting cables

USB data cable, electrically isolated

Ethernet data cable, electrically isolated

Analog output cable, -1.25 to +2.0 V, 0.1 mV / digit

V24 data cable, electrically isolated.

Network technology, Bluetooth modules (see chapter „Networking“)

Order no.

ZA1904SD

ZA1312NA7

ZA2690UK

ZB2490GS1

ZB2490MH

ZB2490HS

ZB2490TK2

Order no.

ZA1919DKU

ZA1945DK

ZA1601RK

ZA1909DK5

01.20

ALMEMO

®

2590-2A

ALMEMO

®

Measuring Instruments

ALMEMO

®

2590-4AS

Professional measuring instrument,

2 measuring inputs, Data logger with external memory connector (accessory)

Technical data and functions

• Technical data and functions as for ALMEMO

®

2590A series

Technical data

Technical data as for ALMEMO

®

2590A series

Measuring inputs 2 ALMEMO

®

input sockets, el. isol., with semicond. relays (50V)

Professional measuring instrument,

4 measuring inputs, Data logger with internal memory or external memory connector

Technical data and functions

• Technical data and functions, as for ALMEMO

®

2590A series

• Internal EEPROM suffi cient for 100 000 measured values, confi gurable as linear or ring memory

Technical data

Technical data as for Serie ALMEMO

®

2590A series

Measuring inputs

4

ALMEMO

®

input sockets, el. isol., with semicond. relays (50V)

Memory, internal EEPROM suffi cient for 100,000 measured values

Option

Volume fl ow determined from matrix measuring as per DIN EN 12599

Temperature ranges for 8 refrigerants

Measuring instrument IP54

(if water-proof plugs are used)

Order no.

OA2590VN

SB0000R2

Option

Volume fl ow determined from matrix measuring as per DIN EN 12599

Temperature ranges for 8 refrigerants

Measuring instrument IP54

(if water-proof plugs are used))

Order no.

OA2590VN

SB0000R2

OA2590W OA2590W

Standard delivery

Order no.

Measuring instrument, batteries, operating instructions, manufacturer’s test certifi cate

Professional measuring instrument

ALMEMO

®

2590-2A MA25902A

Standard delivery

Measuring instrument, batteries, operating instructions, manufacturer’s test certifi cate.

Professional measuring instrument

ALMEMO

®

2590-4AS

Order no.

MA25904AS

Measuring instrument, batteries, rubberized impact protection ZB2490GS1, Mains unit ZA1312NA7, USB data cable

ZA1919DKU, Case ZB2490TK2, Operating instructions, manufacturer‘s test certifi cate

Professional measuring instrument ALMEMO

®

2590-4AS

Case set MA25904ASKSU

DAkkS or works calibration KE90xx, electrical, for measuring instrument (see chapter „Calibration certifi cates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

Humidity reading with further humidity variables, e.g. temperature, dew point, mixture ratio

Temperature / humidity display in bar chart form

Flow reading, measured value with automatic temperature compensation and atmospheric pressure compensation

Overview of all sensors connected pH reading, measured value with automatic temperature compensation

Function menus

01.21

ALMEMO

®

Measuring Instruments

ALMEMO

®

2690-8A

ALMEMO

®

precision measuring instrument with data logger function. Comprehensive range of functions for all application areas. Increased measuring accuracy, fast measuring rate.

Generously dimensioned graphics display, bright illumination, 5 measuring inputs. Runs on rechargeable batteries, charging via the device itself

Technical data and functions ALMEMO

®

2690-8A

• Increased measuring accuracy and stability

• Fast measuring rate, up to 50 measuring operations per second

With SD memory card, up to 100 mops, optional for 1 channel up to 500 mops

• 5 measuring inputs, electrically isolated

• Integrated atmospheric pressure sensor, for automatic pressure compensation, inter alia for Pitot tube fl ow measurement and humidity variables

• Over 65 standard measuring ranges

• Support for ALMEMO

®

plugs with multi-point adjustment, special linearization, and special measuring ranges

• Option KL for independent multi-point adjustment or special linearization programmable in 30 points and management of calibration data saved in the sensor connector and the measuring instrument

• Option GT for higher measuring quality thanks to electrical isolation between measuring inputs and device power supply

(device ground)

• Improved cold junction compensation with 2 sensors

• Data logger with internal EEPROM, suffi cient for 200,000 measured values, confi gurable as linear or ring memory

• Memory connector with micro SD (accessory)

• Sleep mode for long-term recording

• Generously dimensioned graphics display, bright illumination, large display of measured values

• Measured values can be displayed graphically in line chart or bar chart form or numerically in various sizes.

• 3 user-defi ned menus can be freely confi gured from a range of

50 functions.

• Easy to operate by means of 4 soft-keys and cursor block, menu-guided with wizards and context-sensitive help windows

• Choice of languages : German, English, French (other options also available)

• 2 ALMEMO

®

output sockets, suitable for digital interfaces, analog output, trigger input, alarm contacts, memory card

• Runs on rechargeable batteries (standard), high-speed charging in the device itself using the mains unit, included in delivery

• Modern housing with rubberized impact protection and folding stand, splash-proof

Technical data

Precision class

Measuring rate

Measuring inputs

Electrical isolation

Option GT

Additional channels

AA (see page 01.05)

2.5 / 10 / 50 / 100 mops

5 ALMEMO

®

input sockets

with semiconductor relay?*?s (50 V) for analog sensors

Additional electrical isolation between measuring inputs and power supply

(device ground)

4 function channels, device-internal

Sensor power supply

Rechargeable battery/ies 6 / 9 / 12 V, maximum 0.5 A

Mains adapter 12 V, maximum 0.5 A

new:

Atmospheric pressure sensor Integrated

Measuring range

Accuracy

700 to 1100 mbar

±2.5 mbar (at 0 to 65 °C)

Outputs

Graphics display

Illumination

2 ALMEMO

®

sockets, suitable for all output modules (analog / data / trigger / relay cables, memory, etc.)

128 x 128 pixels, 16 rows

5 white LEDs, 3 brightness levels

01.22

Keypad

Memory

9 tactile silicone keys (4 soft-keys)

EEPROM suffi cient for 200,000 measured values

Date and time-of-day Real-time clock, buffered with battery

Power supply

Rechargeable battery/ies 3 AA batteries NiMH or alkaline

Mains adapter

DC adapter cable integrated, high-speed charging (2.5 hours)

ZA1312NA7 230 VAC to 12 VDC, 1 A electrically isolated electrically isolated

ZA2690-UK2 10 to 30 V, 1 A

Current consumption (without input and output modules)

Active mode

With illumination

Sleep mode

approx. 17 mA

approx. 25 to 140 mA

approx. 0.05 mA

Housing 209 x 107 x 54 mm (LxWxH)

ABS (maximum +70 °C), 570 g

Protective class IP54

(if water-proof plugs / sensors are used)

ALMEMO

®

2690-8A

ALMEMO

®

Measuring Instruments

Precision measuring instrument, 5 measuring inputs

Data logger with internal memory or external memory connector (accessory)

Accessories

Memory connector with micro SD, including USB card reader (see chapter „General accessories“)

DC adapter cable, 10 to 30 VDC, 12 V / 1 A, electrically isolated

Generously dimensioned carry case, aluminum profi le frame / ABS

Connecting cables

Ethernet data cable, electrically isolated

ZA1945DK

Analog output cable, -1.25 to +2.0 V, 0.1 mV / digit

ZA1601RK

Order no.

ZA1904SD

ZA2690UK2

ZB2590TK2

Order no.

Trigger and alarm cable (2 relays, 0.5 A, 50 V)

ZA1006EKG

Network technology, Bluetooth modules (see chapter „Networking“)

Options

Measuring module electrically isolated

Multi-point adjustment, special linearization, management of calibration data

Temperature ranges for 8 refrigerants

Measuring rate 500 mops (SD card required)

DIN rail mounting

Order no.

OA2690GT

OA2690KL

SB0000R2

SA0000Q5

OA2290HS

Standard delivery

Order no.

3 rechargeable NiMH batteries, rubberized protection, desktop mains unit ZA1312NA7, USB data cable ZA1919DKU,

Case ZB2490TK2, Operating instructions, manufacturer‘s test certifi cate

Precision measuring instrument ALMEMO

®

2690-8A in case set

as above but with RS232 data cable ZA1909DK5

Precision measuring instrument ALMEMO

®

2690-8A in case set

MA26908AKSU

MA26908AKS

DAkkS or works calibration KE90xx, electrical, for measuring instrument (see chapter „Calibration certifi cates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

Operating concept as for precision measuring instruments ALMEMO

®

2690, 2890 und 5690 / 5790

Menu selection Standard display

Multi-channel display Measuring points list

Bar chart Line diagram

Programming rmenu Assistent menu

01.23

ALMEMO

®

Measuring Instruments

ALMEMO

®

2890-9

ALMEMO

®

precision measuring instrument with data logger function. Comprehensive range of functions for all application areas. Increased measuring accuracy, fast measuring rate.

Generously dimensioned graphics display, bright illumination. 9 measuring inputs

Runs on rechargeable batteries, charging via the device itself

Technical data and functions

• Increased measuring accuracy and stability

• Fast measuring rate, up to 50 measuring operations per second

With SD memory card, up to 100 mops, optional for 1 channel up to 400 mops

• 9 measuring inputs, electrically isolated

• Over 65 standard measuring ranges

• Support for ALMEMO

®

plugs with multi-point adjustment, special linearization, and special measuring ranges

• Option KL for independent multi-point adjustment or special linearization programmable in 30 points and management of calibration data saved in the sensor connector and the measuring instrument

• Higher measuring quality thanks to electrical isolation between measuring inputs and device power supply (device ground)

• Improved cold junction compensation with 2 sensors

• Data logger with internal EEPROM, suffi cient for 100,000 measured values, confi gurable as linear or ring memory

• Memory connector with micro SD (accessory)

• Sleep mode for long-term recording

• Generously dimensioned graphics display, bright illumination, large display of measured values

• Measured values can be displayed graphically in line chart or bar chart form or numerically in various sizes.

• 3 user-defi ned menus can be freely confi gured from a range of

50 functions.

• Easy to operate by means of 4 soft-keys and cursor block, menu-guided with wizards and context-sensitive help windows

• Additional thumb-wheel for extra cursor speed

• Choice of languages : German, English, French (other options also available)

• 2 ALMEMO

®

output sockets, suitable for digital interfaces, analog output, trigger input, alarm contacts, memory card

• Runs on rechargeable batteries (as standard), high-speed charging in the device itself using mains unit, included in delivery

Technical data

Precision class

Measuring rate

Measuring inputs

Electrical isolation

for analog sensors

AA (see page 01.05)

2.5 / 10 / 50 / 100 mops

(measuring operations per second)

9 ALMEMO

®

input sockets with semiconductor relays (50 V)

Additional electrical isolation between measuring inputs and power supply

(device ground)

Additional channels 4 function channels, device-internal

Sensor power supply

Rechargeable battery/ies 9 or 12 V, maximum 0.5 A

Mains adapter 12 V, maximum 0.3 A

Outputs

Standard equipment

Display

Graphics display

Illumination

2 ALMEMO

®

sockets, suitable for all output modules (analog / data / trigger / relay cables, memory, etc.)

128 x 128 pixels, 16 rows

5 white LEDs, 3 brightness levels

Keypad

Memory, EEPROM

Date and time-of-day

9 membrane keys (4 soft-keys), thumb-wheel suffi cient for 100,000 measured values

Real-time clock, buffered with battery

Power supply

Rechargeable battery pack 6 rechargeable NiMH batteries, 1600 mA

Mains adapter

DC adapter cable

Integrated high-speed charging (2.5 h)

ZB1112NA7 230 VAC to 12 VDC, 1 A electrically isolated electrically isolated

ZB2590-UK 10 to 30 V, 1 A

Current consumption (without input and output modules)

Active mode

With illumination

Sleep mode approx. 37 mA approx. 45 to 100 mA approx. 0.05 mA

Housing 204 x 109 x 44 mm (LxWxH)

ABS (maximum 70 °C), 550g

01.24

ALMEMO

®

Measuring Instruments

ALMEMO

®

2890-9

Precision measuring instrument, 9 measuring inputs

Data logger with internal memory or external memory connector (accessory)

Accessories

Memory connector with micro SD, including USB card reader (see chapter „General accessories“)

DC adapter cable, 10 to 30 VDC, 12 V / 1 A, electrically isolated

Generously dimensioned carry case, aluminum profi le frame / ABS

Connecting cables

V24 data cable, electrically isolated

Ethernet data cable, electrically isolated

Analog output cable, -1.25 to +2.0 V, 0.1 mV / digit

Trigger and alarm cable (2 relays, 0.5 A, 50 V

)

Network technology, Bluetooth modules (see chapter „Networking“)

Order no.

ZA1904SD

ZB2590UK

ZB2590TK2

Order no.

ZA1909DK5

ZA1945DK

ZA1601RK

ZA1006EKG

Options

Multi-point adjustment, special linearization, management of calibration data

Temperature ranges for 8 refrigerants

Measuring rate 400 mops (SD card required)

Order no.

OA2690KL

SB0000R2

SA0000Q4

Standard delivery

Rechargeable battery pack, desktop mains unit ZB1112NA7,

USB data cable ZA1919DKU

, case ZB2490TK2,

Operating instructions, manufacturer‘s test certifi cate

Precision measuring instrument ALMEMO

®

2890-9

DAkkS or works calibration KE90xx, electrical, for measuring instrument (see chapter „Calibration certifi cates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

Order no.

MA28909

01.25

ALMEMO

®

Measuring Instruments

ALMEMO

®

V7

ALMEMO

®

710

ALMEMO

®

precision measuring instrument, latest V7 generation

With data logger function and touchscreen.

Comprehensive range of functions for all application areas.

Increased measuring accuracy, fast measuring rate.

10 measuring inputs

Data logger from our latest V7 generation

Data logger ALMEMO

®

710 offers outstanding functions - thanks to our latest D7 sensors.

High-quality display - easy and convenient touchscreen

operation

The brightly illuminated, generously dimensioned 5.7-inch color graphics display shows all measured values and functions clearly and precisely. The device is operated easily and conveniently via touchscreen. The menu guidance system, incorporating wizards and help windows, has a clear, straightforward structure.

Measured values, peak values, average values, and limit values can all be displayed in an easy-to-understand way in various forms, namely list, bar chart, or line graph (up to 4 lines).

Users can even confi gure their own customized user menus to display those parameters required by a particular application.

Choice of languages : German, English, French, Czech

One measuring instrument for every use

The measuring instrument is enclosed in a handy, compact housing with rubberized impact protection. This device can be used in a wide variety of ways, in mobile applications or as a desktop unit, on a folding stand or as a stationary unit in a wallmounted housing.

It incorporates a powerful rechargeable lithium battery to ensure a long operating time.

Data logger for all storage applications

For the purpose of saving measured values the device incorporates an 8-MB fl ash memory. This can also be confi gured as a ring memory for monitoring tasks.

To save larger data quantities an external memory is available in the form of a plug-in SD card.

For autonomous long-term monitoring the data logger can also be run in energy-saving sleep mode.

Measuring inputs for 10 ALMEMO

®

sensors, all generations

Data logger ALMEMO

®

710 incorporates 10 measuring inputs.

All new and already existing sensors designed for any measurable variable can be connected and evaluated.

Sensors using analog signals pass via the integrated high-speed, high-resolution A/D converter. Additional electrical isolation between measuring inputs and power supply (device ground) increases measuring quality.

Digital D6 and the latest digital D7 sensors transfer measured values to the measuring instrument directly in digital form.

The measuring instrument supports all ALMEMO® plug connectors and sensor functions. Digital D6 / D7 sensors can be confi gured directly via the touchscreen.

New digital ALMEMO

®

D7 sensors

With these digital ALMEMO

®

D7 sensors the ALMEMO

®

system is enhanced by many new functions.

They operate via an all-digital interface to the ALMEMO

®

710 measuring instrument ensuring high-speed serial transmission of all measured values.

The measuring ranges of ALMEMO

®

D7 plugs are independent of the measuring instrument and can be expanded as and when required for new applications.

Measured values can be displayed with up to 8 digits (depending on range) and the units with up to 6 characters. Sensor designation and information can be up to 20 characters.

Each connected D7 sensor has its own processor. These all work in parallel at their sensor-specifi c sampling rate. D7 sensors thus attain very high measuring speeds in dynamic measuring operations. Scanning times on the ALMEMO

®

710 can be set individually for quick-acting and slow-acting sensors.

The ALMEMO

®

D7 plug can process up to 10 channels for measured values and function values. This includes new applications, especially for multi-purpose sensors (e.g. Meteo sensors) and for linking up to complex third-party devices (e.g. chemical analysers, power analysers).

Other equipment

With 3 ALMEMO

®

output sockets it is possible to connect simultaneously a PC / network, an ALMEMO

®

output interface with relays and analog output, and an SD memory card.

The ALMEMO

®

710 incorporates an atmospheric pressure sensor to ensure automatic pressure compensation for measuring operations involving inter alia air fl ow or humidity variables.

With option KL it is possible - for an ALMEMO

®

sensor (e.g. temperature or pressure sensors) - to program multi-point adjustment or linearization in the ALMEMO

®

plug itself.

This option is possible with all ALMEMO

®

plug versions.

Standard connector (analog or DIGI), ALMEMO

®

D6 and D7

01.26

ALMEMO

®

V7

ALMEMO

®

710

ALMEMO

®

Measuring Instruments

Precision measuring instrument, latest V7 generation, 10 measuring inputs

Data logger with internal memory or external memory connector (accessory)

Technical data

Measuring inputs

Precision class

10 ALMEMO

®

input sockets for

ALMEMO

®

sensors, all generations analog sensors, D6 and D7 sensors

AA (see page 01.05)

Measuring rate for analog sensors, D6 sensors

2.5 / 10 / 50 / 100 mops

(measuring operations per second)

Electrical isolation

for analog sensors

Channels with semiconductor relays (50 V)

Additional electrical isolation between measuring inputs and power supply

(device ground)

Up to 100 measuring channels per device

Sensor power supply 6 / 9 / 12 V, maximum 2 x 400 mA for supply via mains adapter

12 V, maximum 2 x 400 mA

Atmospheric pressure sensor Integrated, meas. range 700 to 1100 mbar

Accuracy ±2.5 mbar (at 0 to 65 °C)

Outputs 3 ALMEMO

®

sockets, suitable for all output modules (data / analog / trigger / relay cables, memory connector, etc.)

Standard equipment

Display

Graphics display

Illumination

Keypad

Memory

Date and time-of-day

5.7-inch

TFT LCD VGA, 640 x 480 pixels white LED, dimmable

Capacitive touchscreen and 3 additional touch keys

8-MB fl ash memory

(400,000 up to 1.5 million meas. values)

Real-time clock (4.7 ppm) buffered with lithium battery

Power supply

Rechargeable battery/ies 2 rechargeable lith. batteries, total 15.6 Ah

Mains adapter

Integrated, high-speed charging (3 hours)

ZA1312NA9

230 VAC to 12 VDC, 2.5 A, electr. isol.

Current consumption (without input and output modules)

Active mode

Sleep mode approx. 300 to 500 mA approx. 0.05 mA

Housing

ALMEMO

®

710

ALMEMO

®

710 WG

222 x 169 x 61 mm (WxDxH) 1200 g

ABS / TPE, 2-shot technology with rubberized impact protection with folding stand with DIN rail fi xture for wall-mounting, connections facing downwards

Accessories

Memory connector with micro SD, including USB card reader (see chapter „General accessories“)

Large carry case, aluminum profi le frame / ABS, inside dimensions 48 x 35 x 6+6 cm (WxDxH)

Order no.

ZA1904SD

ZB2590TK2

Connecting cables

Ethernet data cable, electrically isolated

USB data cable with 5V device supply from PC not electrically isolated

Analog output cable -1.25 to +2.0 V

Trigger and alarm cable (2 relays, 0.5 A, 50 VDC)

Order no.

ZA1945DK

ZA1919DKU5

ZA1601RK

ZA1006EKG

Note on WinControl measuring software

As measuring software WinControl is suitable for current version 7 and above. For version 6 or earlier a WinControl update is required.

Variants and description (see chapter „Software“).

Options

Multi-point adjustment and / or linearization can - with all ALMEMO

®

plug versions - be programmed by users themselves

Sampling rate 500 measuring operations per second (mops) for 1 analog measuring channel with voltage or current ranges or NTC sensors

Order no.

OA710KL

OA710Q5

Standard delivery

USB data cable ZA1919DKU, Mains unit 12 V / 2.5 A ZA1312NA9, Manufacturer’s test certifi cate

Mobile device with folding stand, in case ZB9710TK

Precision measuring instrument ALMEMO

Stationary device with wall-mounting,

Precision measuring instrument ALMEMO

®

710WG

®

710

DAkkS or works calibration KE90xx, electrical, for measuring instrument (see chapter „Calibration certifi cates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

Order no.

MA710

MA710WG

01.27

ALMEMO

®

Measuring Instruments

ALMEMO

®

V7

ALMEMO

®

710 Clear, precise display - easy and convenient touchscreen operation

List of active measuring channels Display of measured values as a bar chart

Display of measured values as a line graph Keypad for programming

Generously dimensioned display of measured values

01.28

Settings for all sensor and device parameters

ALMEMO

®

Measuring Instruments

ALMEMO

®

8590 /8690 series

ALMEMO

®

precision measuring instrument for measured data acquisition, with data logger function. Comprehensive range of functions for all application areas. Increased measuring accuracy, fast measuring rate

9 measuring inputs.

Operates as data logger or PC interface, also with rechargeable batteries.

Technical data and functions, ALMEMO

®

8590 /8690

• Increased measuring accuracy and stability

• Fast measuring rate, up to 50 measuring operations per second

With SD memory card, up to 100 mops, optional for 1 channel up to 400 mops

• 9 measuring inputs, electrically isolated

• Over 65 standard measuring ranges

• Support for ALMEMO

®

plugs with multi-point adjustment, special linearization, and special measuring ranges

• Option KL for independent multi-point adjustment or special linearization programmable in 30 points and management of calibration data saved in the sensor connector and the measuring instrument

• Higher measuring quality thanks to electrical isolation between measuring inputs and device power supply (device ground)

• Improved cold junction compensation with 2 sensors

• Data logger option

• Internal EEPROM suffi cient for 100,000 measured values

(option S) confi gurable as linear or ring memory - or memory connector with micro SD (accessory)

• Sleep mode for long-term recording

• 2 ALMEMO

®

output sockets, suitable for digital interfaces, analog output, trigger input, alarm contacts, memory card

• 5 LEDs for indicating various operating states

• Key for switching on and start / stop measuring

• Complete sensor and device programming by means of AMR-

Control software (included in delivery).

Technical data ALMEMO

®

8590 /8690

Precision class

Measuring rate

Measuring inputs

Electrical isolation

for analog sensors

Additional channels

Outputs

AA (see page 01.05)

2.5 / 10 / 50 / 100 mops

9 ALMEMO

®

input sockets with semiconductor relays (50 V)

Additional electrical isolation between measuring inputs and power supply

(device ground)

4 function channels, device-internal

2 ALMEMO

®

sockets, suitable for all output modules (analog / data / trigger / relay cables, memory, etc.)

Operation 1 key, 5 LEDs, 2 coding switches

Internal memory (option S) Internal EEPROM suffi cient for 100,000 measured values, confi gurable as linear or ring memory

External memory (accessory) ALMEMO

®

memory connector with micro SD card

Date and time-of-day Real-time clock, buffered with lithium battery

Current consumption (without input and output modules)

Active mode

Sleep mode approx. 25 mA approx. 0.05 mA

ALMEMO

®

8590 /8690, accessories

Memory connector with micro SD, including USB card reader (see chapter „General accessories“)

DC adapter cable, 10 to 30 VDC, 12 V / 1 A, electrically isolated

ALMEMO

®

8590 /8690, connecting cable

USB data cable, electrically isolated

V24 data cable, electrically isolated

Ethernet data cable, electrically isolated

Analog output cable, -1.25 to +2.0 V, 0.1 mV / digit

Trigger and alarm cable (2 relays, 0.5 A, 50 V)

Network technology, Bluetooth modules (see chapter „Networking“)

Order no.

ZA1904SD

ZB3090UK2

Order no.

ZA1919DKU

ZA1909DK5

ZA1945DK

ZA1601RK

ZA1006EKG

01.29

ALMEMO

®

Measuring Instruments

ALMEMO

®

8590-9 ALMEMO

®

8690-9A

Precision measuring instrument, 9 measuring inputs

Data logger option with internal memory or external memory connector (accessory)

Precision measuring instrument, 9 measuring inputs

Data logger option with internal memory or external memory connector (accessory)

Runs on rechargeable batteries, charging via the device itself

Technical data and functions

• Technical data and functions, as for ALMEMO

®

8590 / 8690

Technical data

Technical data, as for ALMEMO

®

8590 / 8690

Sensor power supply Mains adapter 12 V, maximum 0.5 A

Power supply

Mains adapter

DC adapter cable

ZB1212NA7 230 VAC to 12 VDC, 1 A, electrically isolated

ZB3090UK2 10 to 30 VDC, 1 A, electrically isolated

Housing 180 x 49 x 137 mm (LxWxH)

Polystyrene (PS) Weight approx. 490 g

Technical data and functions

• Technical data and functions, as for ALMEMO

®

8590 / 8690

• Runs on rechargeable batteries, high-speed charging in the device itself using mains unit, included in delivery

Technical data

Technical data, as for ALMEMO

®

8590 / 8690

Rechargeable battery pack 8 rechargeable NiMH batteries,

9 to 11 V, 1600 mAh

With intelligent high-speed charging

(3.5 hours)

Sensor power supply

Mains adapter 12 V, maximum 0.5 A

Runs on rechargeable batteries 9 to 11.5 V, maximum 0.5 A

Power supply

Mains adapter

DC adapter cable

ZB1212NA9

90 to 260 VAC, 12 VDC, 2.5 A electrically isolated ZB3090-UK2

10 to 30 VDC, 12 VDC, 1 A

Housing 218 x 77 x 145 mm (LxWxH)

Polystyrene (PS) Weight approx. 1.2 kg

Options

Internal data memory suffi cient for 100,000 values

Multi-point adjustment, special linearization, management of calibration data

Temperature ranges for 8 refrigerants (see 10.08)

Measuring rate for 1 measuring channel, 400 mops

(SD card required)

DIN rail mounting

Order no.

OA8590S

OA8590KL

SB0000R2

SA0000Q4

OA2290HS

Options

Internal data memory suffi cient for 100,000 values

Multi-point adjustment, special linearization, management of calibration data

Temperature ranges for 8 refrigerants (see 10.08)

Measuring rate for 1 measuring channel, 400 mops

(SD card required)

DIN rail mounting

Order no.

OA8590S

OA8590KL

SB0000R2

SA0000Q4

OA2290HS

Standard delivery

Order no.

Mains plug assembly ZB1212NA7, operating instructions, manufacturer‘s test certifi cate

Precision measuring instrument ALMEMO

® for measured data acquisition

8590-9

MA85909

Standard delivery

Order no.

Rechargeable batteries, mains plug assembly ZB1212NA9,

Operating instructions, manufacturer‘s test certifi cate

Precision measuring instrument ALMEMO

® for measured data acquisition

8690-9A

MA86909A

DAkkS or works calibration KE90xx, electrical, for measuring instrument (see chapter „Calibration certifi cates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

01.30

ALMEMO

®

Measuring Instruments

Data acquisition systems ALMEMO

®

5690 und 5790

ALMEMO

®

5690-1M09 fully equipped (example)

ALMEMO

®

5690-2 with graphics display

ALMEMO

®

5690-1CPU fully equipped (example)

01.31

ALMEMO

®

Measuring Instruments

ALMEMO

®

5690 data acquisition system

ALMEMO

®

precision measuring instrument for measured data acquisition, with data logger function.

Comprehensive range of functions for all application areas. Increased measuring accuracy, fast measuring rate. Up to 99 / 190 measuring inputs

Operates as data logger or PC interface, also with generously dimensioned graphics display.

Technical data and functions, ALMEMO

®

5690 and 5790 series

• Multi-functional data acquisition systems with up to 99 or 190 measuring inputs (applies to ALMEMO

®

5690-xCPU with option XU or XM)

• Increased measuring accuracy and stability

• Fast measuring rate, up to 50 measuring operations per second

With SD memory card, up to 100 mops, optional for 1 channel up to 400 mops (does not apply to ALMEMO

®

5690-xCPU with option XM)

• Measuring rate increased to over 100 channels / second with several measuring circuit boards (applies to ALMEMO

®

5690-xCPU with option XM)

The measuring circuit boards operate in parallel, thus ensuring short scanning times for a large number of channels.

• Over 65 standard measuring ranges

• Option KL for independent multi-point adjustment or special linearization programmable in 30 points and management of calibration data saved in the sensor connector and the measuring instrument

• Higher measuring quality thanks to electrical isolation between measuring inputs and device power supply (device ground)

• Improved cold junction compensation with 2 sensors per input card

• Operates as data logger (internal EEPROM / RAM or SD memory card, sleep mode for long-term recording) or as interface for PC online operation

• ALMEMO

®

5690-1 (variant without display), ALMEMO

®

5690-2 (variant with display and operating controls)

• 5 LEDs for displaying the operating status of the measuring circuit and the CPU

• 8 rechargeable NiMH batteries with high-speed battery charging (accessory)

• Relay / trigger / analog interface as plug-in board (accessory) for output of alarm and control signals

• ALMEMO

®

output sockets, suitable for digital interfaces, analog output, trigger input, alarm contacts, memory card

• Housing in several variants: Desktop housing TG1, TG3, TG8

Wall-mounted housing WG3, Rack housing BT8

Protected industrial housingIG2.

Technical data, ALMEMO

®

5690 and 5790 series

Precision class

Measuring rate

AA (see page 01.05)

2.5 / 10 / 50 / 100 mops

Electrical isolation with semiconductor relays (50 V)

for analog sensors Additional electrical isolation between measuring inputs and power supply

(device ground)

Date and time-of-day Real-time clock, buffered with lithium battery

Supply current For system boards and sensor supply

Entire system, max. 2.5 A, per board max. 0.5 A

Power supply

Mains adapter ZB1212NA9 90 to 260 VAC, 12 VDC, 2.5 A

DC adapter cable electrically isolated

ZB3090-UK2 10 to 30 VDC, 12 VDC, 1 A

Rechargeable battery pack 8 rechargeable NiMH batteries,

9 to 11 V, 1600 mAh With intelligent high-speed charging (3.5 hours)

Supply current Entire system maximum 1.5 A

ALMEMO

®

5690 and 5790 series, accessories

Rechargeable batteries, 1600 mAh, 1 slot

DC cable, 10 to 30 VDC, 12 VDC, 1.25 A

Relay / trigger / analog board (see chapter „Output modules“) 2 slots

Carry case, aluminum profile frame / ABS, suitable for ALMEMO

®

5690 in desktop housing TGx

Rack case with handle, suitable for ALMEMO

®

5690 in rack housing BT8

Order no.

ES5690AP

ZB3090UK2

ES5690RTA5

ZB5600TK3

ZB5090RC

ALMEMO

®

5690 and 5790 series, connecting cables

Order no.

USB data cable, electrically isolated

Ethernet data cable, electrically isolated

Trigger and relay cable (2 relays, 0.5 A, 50 V)

Analog output cable, -1.25 to +2.0 V, 0.1 mV / digit

ZA1919DKU

ZA1945DK

ZA1006EKG

ZA1601RK

V24 data cable, electrically isolated

ZA1909DK5

Network technology, Bluetooth modules (see chapter „Networking“) Relay trigger analog adapter (see chapter „Output modules“)

01.32

ALMEMO

®

Measuring Instruments

ALMEMO

®

data acquisition systems - a comparison

Function

System type 5690-xM09 5690-xCPU 5690-xCPU 5690-xCPU

with option XU with option XM

Measuring circuit

Measuring inputs

Number of channels

Expansions

Selector switch boards

Master measuring circuit board with 9 measuring inputs up to 99 inputs up to 99 channels up to 9 up to 100 inputs up to 100 channels up to 10

Measuring circuit

CPU board (without measuring inputs) up to 190 inputs up to 250 channels up to 19 up to 190 inputs up to 250 channels

None

Expansions

Active measuring circuit boards

None None None up to 19

Scanning time (approx.)

At conversion rate 10 Hz

At conversion rate 50Hz

ALMEMO

®

plug with special measuring range

/ multi-point calibration, linearization

ALMEMO

®

outputs

For 1 to 99 channels in total

0.1 to 10 seconds

0.02 to 2 seconds

Up to 9 ALMEMO

®

plugs

(master measuring circuit)

Sockets A1 and A2

For 1 to 100 channels in total

0.1 to 10 seconds

0.02 to 2 seconds

Up to 100 ALMEMO

® plugs

For 1 to 190 channels in total

0.1 to 19 seconds

0.02 to 4 seconds

Up to 190 ALMEMO plugs

®

(relay / trigger / analog outputs)

For 100 / 190 channels in total = 10/19 measuring circuit boards with 10 channels each

... 1.1 / 1.1 seconds*

... 0.3 / 0.5 seconds*

*for systems without display

Up to 190 ALMEMO plugs

Sockets A1 to A5 for expanding the periphery, optional socket P0

®

Operating modes

System type 5690-1M09 5690-2M09 5690-1CPU 5690-2CPU

Online operation via PC

Display and operating controls

Data logger

Internal memory yes no yes

Accessory ZA1904SD

Memory connector including micro SD

Micro SD drive, integrated, including micro SD

(as standard)

512-KB EEPROM (option) yes no yes

Accessory ZA1904SD

Memory connector including micro SD

Micro SD drive, integrated, including micro SD

(as standard)

2-MB RAM, battery-buffered (standard) or 2-MB FeRAM, non-volatile (option)

01.33

ALMEMO

®

Measuring Instruments

ALMEMO

®

5690-1M09

Technical data and functions

• Technical data and functions, as for ALMEMO

®

5690 series

• Master measuring circuit, 9 ALMEMO

®

input sockets, electrically isolated, suitable for 9 ALMEMO

®

sensors

• Up to 9 ALMEMO

®

connectors; special ranges / multi-point calibration / linearization possible

(only on master measuring circuit)

Technical data

Technical data, as for ALMEMO

®

5690 series

Measuring inputs 9 ALMEMO

®

input sockets

Expansion up to 99 inputs by means of selector switch boards

Measuring channels Expansion up to maximum

99 measuring channels

Internal memory (option S) Internal EEPROM sufficient for

100,000 measured values, configurable

• Expansion up to 99 inputs by means of various selector switch boards, maximum 99 measuring channels

• Data logger option with internal EEPROM or external ALMEMO

®

memory connector with micro SD card as linear or ring memory

External memory (accessory) ALMEMO

®

memory connector with micro SD card

Outputs

Operation

2 ALMEMO

®

sockets, suitable for all output modules (analog / data / trigger / relay cables, etc.)

Alarm signal transmitter, internal

1 key, 5 LEDs, 2 coding switches

Accessories

Memory connector with micro SD, including USB card reader (see chapter „General accessories“)

Expansions

Selector switch boards U-A10, U-MU, U-TH, U-KS

Relay / trigger / analog board, 2 slots Per system up to 7 boards are supported. (see chapter „Output modules“)

ZA1904SD

Order no.

(see page 01.40)

ES5690RTA5

Optionen

Internal data memory sufficient for 100,000 values

Multi-point adjustment, special linearization, management of calibration data

Temperature ranges for 8 refrigerants (see 10.08)

Measuring rate for 1 measuring channel, 400 mops (SD card required)

Order no.

OA5690S

OA5690KL

SB0000R2

SA0000Q4

Standard delivery

Precision measuring instrument, data acquisition system with master measuring circuit board MM-A9, mains plug assembly

ZB1212NA9, Operating instructions, manufacturer‘s test certificate

DAkkS or works calibration KE90xx, electrical, for measuring instrument (see chapter „Calibration certificates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

01.34

ALMEMO

®

Measuring Instruments

ALMEMO

®

5690-1M09TG3 ALMEMO

®

5690-1M09TG1

Dimensions:

77 x 145 x 218 mm

(WxHxD)

Data acquisition system in desktop housing TG1, 9 inputs,

1 free slot

MA56901M09TG1

Expansion with

1 U-MU board or U-TH or U-KS (10 inputs)

Dimensions:

179 x 158 x 232 mm

(WxHxD)

Data acquisition system in desktop housing TG3, 9 inputs,

6 free slots

MA56901M09TG3

Expansion with

3 U-A10 boards or U-TH (30 inputs) or 6 U-MU boards or U-KS (60 inputs) or 3 RTA5 boards

ALMEMO

®

5690-1M09TG8

ALMEMO

®

5690-1M09BT8

Dimensions:

444 x 158 x 232 mm

(WxHxD)

Data acquisition system in desktop housing TG8, 9 inputs,

19 free slots

MA56901M09TG8

Expansion with

9 U-A10 boards or U-TH or U-MU or U-KS (90 inputs) or 7

RTA5 boards

Dimensions:

483 x 132 x 273 mm (WxHxD)

Data acquisition system in 19-inch rack housing, 9 inputs,

19 free slots

MA56901M09BT8

Expansion with

9 U-A10 boards or U-TH or U-MU or U-KS (90 inputs) or 7 RTA5 boards

Carry case, aluminum profile frame ZB5600TK3 for ALMEMO

®

5690-1/ -2

Rack case with handle ZB5090RC for ALMEMO

®

5690-xxBT8 in 19-inch rack housing

01.35

ALMEMO

®

Measuring Instruments

ALMEMO

®

5690-2M09

Technical data and functions

• Technical data and functions, as for ALMEMO

®

5690 series

• Master measuring circuit, 9 ALMEMO

®

input sockets, electrically isolated, suitable for 9 ALMEMO

®

sensors

• Up to 9 ALMEMO

®

connectors; special ranges / multi-point calibration / linearization possible

(only on master measuring circuit)

• Expansion up to 99 inputs by means of various selector switch boards, maximum 99 measuring channels

• Generously dimensioned graphics display, bright illumination, large display of measured values

• Measured values can be displayed graphically in line chart or bar chart form or numerically in various sizes.

• 3 user-defined menus can be freely configured from a range of

50 functions.

• Easy to operate by means of 4 soft-keys and cursor block, menu-guided with wizards and context-sensitive help windows

• Choice of languages : German, English, French (other options also available)

• Data logger with micro SD (standard)

• Option, internal EEPROM.

Technical data

Technical data, as for ALMEMO

®

5690 series

Measuring inputs

Measuring channels

Memory

9 ALMEMO

®

input sockets

Expansion up to 99 inputs by means of selector switch boards

Expansion up to maximum 99 measuring channels

Micro SD card, integrated drive

Internal memory (option S) Internal EEPROM sufficient for

100,000 measured values, configurable as linear or ring memory

Outputs

Display

Graphics display

Illumination

Operation

2 ALMEMO

®

sockets, suitable for all output modules (analog / data / trigger / relay cables, etc.)

Alarm signal transmitter, internal

128 x 128 pixels, 16 rows

5 white LEDs, 3 brightness levels

9 keys (4 soft-keys and cursor block)

9 status LEDs on front panel

Expansions

Selector switch boards U-A10, U-MU, U-TH, U-KS

Relay / trigger / analog board, 2 slots Per system up to 7 boards are supported. (see chapter „Output modules“)

Order no.

(see page 01.40)

ES5690RTA5

Options

Internal data memory sufficient for 100,000 values

Multi-point adjustment, special linearization, management of calibration data

Temperature ranges for 8 refrigerants (see 10.08)

Measuring rate for 1 measuring channel, 400 mops (SD card required)

Order no.

OA5690S

OA5690KL

SB0000R2

SA0000Q4

Standard delivery

Precision measuring instrument, data acquisition system with graphics display and operating controls, master measuring circuit board MM-A9, micro SD card, USB card reader, mains plug assembly ZB1212NA9, operating instructions, manufacturer‘s test certificate

DAkkS or works calibration KE90xx, electrical, for measuring instrument (see chapter „Calibration certificates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

01.36

ALMEMO

®

Measuring Instruments

ALMEMO

®

5690-2M09WG3 ALMEMO

®

5690-2M09TG3

Dimensions:

179 x 158 x 232 mm

(WxHxD)

Data acquisition system in desktop housing TG3, 9 inputs,

6 free slots

MA56902M09TG3

Expansion with

3 U-A10 boards or U-TH (30 inputs) or 6 U-MU boards or U-KS (60 inputs) or 3 RTA5 boards

Dimensions:

209 x 207 x 153 mm (WxHxD)

(width includes fastening strips)

Data acquisition system in wall-mounted housing WG3,

9 inputs, 1 free slot

MA56902M09WG3

Expansion with

3 U-A10 boards or U-TH (30 inputs) or 6 U-MU boards or U-KS (60 inputs) or 3 RTA5 boards

The boards have their connections facing downwards. To facilitate wall-mounting four holes (5.3 mm) are provided on the protruding strips to the left and right of the housing’s backplate

(which cannot itself be removed).

ALMEMO

®

5690-2M09TG8

ALMEMO

®

5690-2M09BT8

Dimensions:

444 x 158 x 232 mm

(WxHxD)

Data acquisition system in desktop housing TG8, 9 inputs,

19 free slots

MA56902M09TG8

Expansion with

9 U-A10 boards or U-TH or U-MU or U-KS (90 inputs) or 7 RTA5 boards

Dimensions:

483 x 132 x 273 mm

(WxHxD)

Data acquisition system in 19-inch rack housing, 9 inputs,

19 free slots

MA56902M09BT8

Expansion with

9 U-A10 boards or U-TH or U-MU or U-KS (90 inputs) or 7 RTA5 boards

01.37

ALMEMO

®

Measuring Instruments

ALMEMO

®

5790-2M09IG2

Technical data and functions

• Technical data and functions, as for ALMEMO

®

5690 series

• Robust aluminum housing, protective class IP65

• Master measuring circuit, 9 ALMEMO

®

input sockets, electrically isolated, suitable for 9 ALMEMO

®

sensors

• Up to 9 ALMEMO

®

connectors; special ranges / multi-point calibration / linearization possible

(only on master measuring circuit)

• Expansion up to 29 inputs by means of various selector switch boards

• Generously dimensioned graphics display, bright illumination, large display of measured values

• Measured values can be displayed graphically in line chart or bar chart form or numerically in various sizes.

• 3 user-defined menus can be freely configured from a range of

50 functions.

• Easy to operate by means of 4 soft-keys and cursor block, menu-guided with wizards and context-sensitive help windows

• Choice of languages : German, English, French (other options also available)

• Data logger option with internal EEPROM or external AL-

MEMO

®

memory connector with micro SD card

Technical data

Technical data, as for ALMEMO

®

5690 series

Measuring inputs

Measuring channels

9 ALMEMO

®

input sockets

Expansion up to 29 inputs by means of selector switch boards

Expansion up to maximum 99 measuring channels

Internal memory (option S) Internal EEPROM sufficient for

100,000 measured values, configurable as linear or ring memory

External memory (accessory) ALMEMO

®

memory connector with micro SD card

Outputs 2 ALMEMO

®

sockets, suitable for all output modules (analog / data / trigger / relay cables, etc.)

Alarm signal transmitter, internal

Display

Graphics display

Illumination

Operation

128 x 128 pixels, 16 rows

5 white LEDs, 3 brightness levels

9 keys (4 soft-keys and cursor block)

Power supply

Screwed cable glands

Housing

Dimensions

Weight

Protective class

Wall-mounting

9 status LEDs on front panel

Mains unit ZB1212NA6, installed on a fixed basis, 100 to 240 VAC, connected via appliance socket, including safety connecting cable

Plastic, with multiple inserts, slotted

24 drilled holes for cables d= 4 mm

2 drilled holes for cables d= 7 mm for all supply lines (sensor cables, output cables, e.g. data cable, mains supply cable) including dummy plugs for all holes

Aluminum

233 x approx. 350 x 121 mm (WxHxD)

(height includes PGs)

19-inch design Plastic insert, 16 DUs approx. 6 kg

IP65

4 x M4 thread, including 2 aluminum profiles

Accessories

Memory connector with micro SD, including USB card reader (see chapter „General accessories“)

Order no.

ZA1904SD

Expansions

Selector switch boards U-A10, U-MU, U-TH, U-KS

Relay / trigger / analog board, 2 slots, maximum 1 board (see chapter „Output modules“)

Order no.

(see page 01.40)

ES5690RTA5

Options

Internal data memory sufficient for 100,000 values

Multi-point adjustment, special linearization, management of calibration data

Temperature ranges for 8 refrigerants (see 10.08)

Measuring rate for 1 measuring channel, 400 mops (SD card required)

Power supply via rechargeable battery module

Rechargeable battery set (8 NiMH cells, 1600 mAh), 1 slot

Standard delivery

Precision measuring instrument, data acquisition system with graphics display and operating controls, master measuring circuit board MM-A9, mains unit ZB1212NA6 installed on a fixed basis, safety connecting cable, operating instructions, manufacturer‘s test certificate

DAkkS or works calibration KE90xx, electrical, for measuring instrument (see chapter „Calibration certificates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

01.38

Order no.

OA5690S

OA5690KL

SB0000R2

SA0000Q4

OA5790A

ES5690AP

ALMEMO

®

5790-2M09IG2

ALMEMO

®

Measuring Instruments

Data acquisition system in industrial housing, 9 inputs, 2 free slots

Expansion with 1 U-A10 board U-TH or 2 U-MU boards U-KS or 1 RTA5 board

Dimensions:

233 x approx. 350 x 121mm

(WxHxD)

(with PGs)

MA57902M09IG2

01.39

ALMEMO

®

Measuring Instruments

Master measuring circuit board, selector switch boards, and expansions for the ALMEMO

®

5690-1M09 and 5690-2M09 systems

Master measuring circuit board

MM-A9

U-A10 U-MU U-TH U-KS (U/I) AP RTA5

Selector switch boards for ALMEMO

®

5690-1M09 and 5690-2M09

Technical data and functions of selector switch boards

• Selector switch boards for expanding the ALMEMO

®

5690-

1M09 and 5690-2M09 systems by additional inputs

• There are several design variants for different installations / input plugs.

Selector switch boards U-A10

10 inputs for ALMEMO

®

single connectors

For flexible applications with individual sensors and measuring signals.

Selector switch boards U-MU

10 inputs for ALMEMO

®

10 MU connectors

For permanently installing groups of 10, especially temperature sensors.

Technical data

Measuring inputs 10 ALMEMO

®

input sockets, electrically isolated

Measuring ranges All ranges (see page 01.06)

Sensor supply

Footprint

12 V, max. 0.3 A (per system max. 2.5 A)

2 slots

Standard delivery

Selector switch board U-A10

Order no.

ES5690UA10

ALMEMO

®

connector must be ordered separately.

Technical data

Measuring inputs 10 inputs, electrically isolated, socket strip for ALMEMO

®

10-way MU connector

Measuring ranges all thermocouples, Pt100, Ni100, ohms,

2.6 V, 260 mV, 55 mV, 26 mV

Sensor supply None

Footprint 1 slot

Standard delivery

Selector switch board U-MU

ALMEMO

®

10-way MU connector

Order no.

ES5690UMU

ZA5690MU

01.40

ALMEMO

®

Measuring Instruments

Selector switch boards U-TH

10 inputs for miniature thermal connectors

For any individual thermocouple temperature sensors with miniature thermal connector.

Selector switch boards U-KS

10 nputs, electrically isolated, sensor connection via socket block

For permanently installing groups of 10.

Technical data

Measuring inputs 10 miniature thermal sockets, electr. isolated

ALMEMO

®

sensor parameters are saved in the measuring instrument.

Measuring ranges all thermocouples

Sensor supply None

Footprint 2 slots

Technical data

Measuring inputs 10 inputs, electrically isolated, male strip connector for socket block

ALMEMO

®

sensor parameters are saved in the measuring instrument.

Measuring ranges Pt100, Ni100, NTC, ohms, 2.6V, 260mV,

55mV, 26mV

Sensor supply

Footprint

None

1 slot

Standard delivery

Order no.

Selector switch board U-TH

ES5690UTH

Miniature thermal connectors must be ordered separately.

Standard delivery

Selector switch board U-KS

including socket block

Socket block (spare)

Order no.

ES5690UKS

ZB5600KS

Selector switch boards U-KSU

10 inputs, electrically isolated, sensor connection via socket block

For permanently installing groups of 10 with voltages 10 V

Selector switch boards U-KSI

10 inputs, electrically isolated, sensor connection via socket block

For permanently installing groups of 10 with currents 20mA

Technical data

Measuring inputs 10 inputs, electrically isolated, male strip connector for socket block

ALMEMO

®

sensor parameters are saved in the measuring instrument.

Measuring ranges Voltage -26 to +26 V (integrated divider)

Accuracy, divider ±0.1 % of measured value

Sensor supply None

Footprint 1 slot

Technical data

Measuring inputs 10 inputs, electrically isolated, male strip connector for socket block

ALMEMO

®

sensor parameters are saved in the measuring instrument.

Measuring ranges Current -32 to +32 mA (integrated shunt)

Accuracy, shunt

Sensor supply

Footprint

±0.1 % of measured value

None

1 slot

Standard delivery

Selector switch board U-KSU including socket block

Socket block (spare)

Order no.

ES5690UKSU

ZB5600KS

Standard delivery

Selector switch board U-KSI including socket block

Socket block (spare)

Order no.

ES5690UKSI

ZB5600KS

01.41

ALMEMO

®

Measuring Instruments

ALMEMO

®

5690-1CPU

Technical data and functions

• Technical data and functions, as for ALMEMO

®

5690 series

• CPU board with measuring circuit (without measuring inputs) and output sockets

• Up to 100 measuring inputs / 100 measuring channels via selector switch boards

• Option XU - up to 190 measuring inputs / 250 measuring channels via selector switch boards

• Option XM - high-speed measuring operations, up to 190 measuring inputs / 250 measuring channels via active measuring circuit boards

The measuring circuit boards operate in parallel, thus ensuring short scanning times for a large number of channels. The scanning time is determined by the measuring circuit board with the highest number of active measuring channels - or, at conversion rate 50 Hz, also by the processing time of the

CPU.

• Option - 5 ALMEMO

®

output sockets for digital interfaces, analog outputs, trigger, alarm contacts, socket P0 for integrated relay outputs

• Data logger with internal RAM (standard) or FeRAM (option) or external ALMEMO

®

memory connector with micro SD card

Technical data

Technical data, as for ALMEMO

®

5690 series

CPU board Measuring circuit (without measuring inputs), input boards (see page 01.48)

Measuring inputs / measuring channels

Standard

Option XU

Option XM up to 100 inputs / 100 meas. channels via selector switch boards up to 190 inputs / 250 meas. channels via selector switch boards up to 190 inputs / 250 meas. channels via active measuring circuit boards

Memory, internal

Standard

Option SF sufficient for 400,000 values, linear or ring memory

RAM (buffered by battery)

FeRAM (non-volatile)

External memory (accessory) ALMEMO

®

memory connector with micro SD card

Outputs 5 ALMEMO

®

sockets, suitable for all output modules (analog / data / trigger / relay cables, etc.) . Alarm signal transmitter, internal Socket P0 for integrated relay outputs (option)

Or trigger and analog output

(by request)

Operation 1 key, 5 LEDs, 2 coding switches

Accessories

Memory connector with micro SD, including USB card reader (see chapter „General accessories“)

Input boards / expansions

Option XM - selector switch boards and active measuring circuit boards

Relay / trigger / analog board, 2 slots Per system up to 4 boards are supported. (see chapter „Output modules“)

ZA1904SD

Order no.

(see page 01.48)

ES5690RTA5

Options

Up to 190 measuring inputs / 250 measuring channels

For active measuring circuit boards, up to 190 measuring inputs / 250 measuring channels

Data memory, internal FeRAM, non-volatile (instead of battery-buffered RAM)

Multi-point adjustment, special linearization, management of calibration data

Temperature ranges for 8 refrigerants (see 10.08)

Measuring rate for 1 measuring channel, 400 mops (SD card required) This cannot be combined with option XM.

For output socket P0

SH2 2 semiconductor relays (normally open) internal, 0.5 A, 50 V

OH2 2 additional relays (normally closed) for option SH2 (thus 2 changeover relays)

Order no.

OA5690XU

OA5690XM

OA5690SF

OA5690KL

SB0000R2

SA0000Q4

OA5690SH2

OA5690OH2

Standard delivery

Precision measuring instrument, data acquisition system with CPU board

Measuring circuit (without measuring inputs) Input boards must be ordered separately. (see page 01.48)

Mains plug assembly ZB1212NA9, Operating instructions, manufacturer‘s test certificate

DAkkS or works calibration KE90xx, electrical, for measuring instrument (see chapter „Calibration certificates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

01.42

ALMEMO

®

Measuring Instruments

ALMEMO

®

5690-1CPUTG3 ALMEMO

®

5690-1CPUTG1

Dimensions:

77 x 145 x 218 mm

(WxHxD)

Data acquisition system in desktop housing TG1

CPU board, 1 free slot

MA56901CPUTG1

Messeingänge über:

Measuring inputs via 1 MU / TH / KS board (10 inputs)

Dimensions:

179 x 158 x 232 mm

(WxHxD)

Data acquisition system in desktop housing TG3

CPU board, 6 free slots

MA56901CPUTG3

Measuring inputs via three A10 or TH boards (30 inputs) or 6 MU or KS boards (60 inputs) or three RTA5 output boards

ALMEMO

®

5690-1CPUTG8

ALMEMO

®

5690-1CPUBT8

Dimensions:

444 x 158 x 232 mm

(WxHxD)

Data acquisition system in desktop housing TG8

CPU board, 19 free slots

MA56901CPUTG8

Measuring inputs via nine A10 or TH boards (90 inputs) or 19 MU or KS boards (190 inputs) or four RTA5 output boards

Dimensions:

483 x 132 x 273 mm

(WxHxD)

Data acquisition system in 19-inch rack housing

CPU board, 19 free slots

MA56901CPUBT8

Measuring inputs via nine A10 or TH boards (90 inputs) or 19 MU or KS boards (190 inputs) or four RTA5 output boards

Carry case, aluminum profile frame ZB5600TK3 for ALMEMO

®

5690-1/ -2

Rack case with handle ZB5090RC for ALMEMO

®

5690-xxBT8 in 19-inch rack housing

01.43

ALMEMO

®

Measuring Instruments

ALMEMO

®

5690-2CPU

Technical data and functions

• Technical data and functions, as for ALMEMO

®

5690 series

• CPU board with measuring circuit (without measuring inputs) and output sockets

• Up to 100 measuring inputs / 100 measuring channels via selector switch boards

• Option XU - up to 190 measuring inputs / 250 measuring channels via selector switch boards

• Option XM - high-speed measuring operations, up to 190 measuring inputs / 250 measuring channels via active measuring circuit boards

The measuring circuit boards operate in parallel, thus ensuring short scanning times for a large number of channels. The scanning time is determined by the measuring circuit board with the highest number of active measuring channels - or, at conversion rate 50 Hz, also by the processing time of the CPU.

Technical data

Technical data, as for ALMEMO

®

5690 series

CPU board Measuring circuit (without meas. inputs)

Input boards (see page 01.48)

Measuring inputs / measuring channels

Standard

Option XU

Option XM up to 100 inputs / 100 measuring channels via selector switch boards up to 190 inputs / 250 measuring channels via selector switch boards up to 190 inputs / 250 measuring channels

via active measuring circuit boards

Memory, internal

Standard

Option SF sufficient for 400,000 values, linear or ring memory

RAM (buffered by battery)

FeRAM (non-volatile)

• Option - 5 ALMEMO

®

output sockets for digital interfaces, analog outputs, trigger, alarm contacts, socket P0 for integrated relay outputs

• Generously dimensioned graphics display, bright illumination, large display of measured values

• Measured values can be displayed graphically in line chart or bar chart form or numerically in various sizes.

• 3 user-defined menus can be freely configured from a range of

50 functions.

• Easy to operate by means of 4 soft-keys and cursor block, menu-guided with wizards and context-sensitive help windows

• Choice of languages : German, English, French (other options also available)

• Data logger with internal RAM (standard) or FeRAM (option) and with micro SD card (standard).

Memory

Outputs

Display

Graphics display

Illumination

Operation

Micro SD card, integrated drive

5 ALMEMO® sockets, suitable for all output modules (analog / data / trigger / relay cables, etc.)

Alarm signal transmitter, internal

Socket P0 for integrated relay outputs

(option)

Or trigger and analog output (by request)

128 x 128 pixels, 16 rows

5 white LEDs, 3 brightness levels

9 keys (4 soft-keys and cursor block)

9 status LEDs on front panel

Input boards / expansions

Option XM - selector switch boards and active measuring circuit boards

Relay / trigger / analog board, 2 slots Per system up to 4 boards are supported. (see chapter „Output modules“)

Order no.

(see page 01.48)

ES5690RTA5

Options

Up to 190 measuring inputs / 250 measuring channels

For active measuring circuit boards, up to 190 measuring inputs / 250 measuring channels

Data memory, internal FeRAM, non-volatile (instead of battery-buffered RAM)

Multi-point adjustment, special linearization, management of calibration data

Temperature ranges for 8 refrigerants (see 10.08)

Measuring rate for 1 measuring channel, 400 mops (SD card required) This cannot be combined with option XM.

For output socket P0

SH2 2 semiconductor relays (normally open) internal, 0.5 A, 50 V

OH2 2 additional relays (normally closed) for option SH2 (thus 2 changeover relays)

Order no.

OA5690XU

OA5690XM

OA5690SF

OA5690KL

SB0000R2

SA0000Q4

OA5690SH2

OA5690OH2

Standard delivery

Precision measuring instrument, data acquisition system with graphics display and operating controls, CPU board

Measuring circuit (without measuring inputs) Input boards must be ordered separately. (see page 01.48) Micro SD card,

USB card reader, mains plug assembly ZB1212NA9, Operating instructions, manufacturer‘s test certificate.

DAkkS or works calibration KE90xx, electrical, for measuring instrument (see chapter „Calibration certificates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

01.44

ALMEMO

®

Measuring Instruments

ALMEMO

®

5690-2CPUWG3 ALMEMO

®

5690-2CPUTG3

Dimensions:

179 x 158 x 232 mm

(WxHxD)

Dimensions:

209 x 207 x 153 mm

(WxHxD)

(width includes fastening strips)

Data acquisition system in desktop housing TG3

CPU board, 6 free slots

MA56902CPUTG3

Measuring inputs via three A10 or TH boards (30 inputs) or 6 MU or KS boards (60 inputs) or three RTA5 output boards

Data acquisition system in wall-mounted housing WG3

CPU board, 6 free slots

MA56902CPUWG3

Measuring inputs via three A10 or TH boards (30 inputs) or 6 MU or KS boards (60 inputs) or three RTA5 output boards

The boards have their connections facing downwards. To facilitate wall-mounting four holes (5.3 mm) are provided on the protruding strips to the left and right of the housing’s backplate

(which cannot itself be removed).

ALMEMO

®

5690-2CPUTG8

ALMEMO

®

5690-2CPUBT8

Dimensions:

444 x H158 x T232 mm

(WxHxD)

Data acquisition system in desktop housing TG8

CPU board, 19 free slots

MA56902CPUTG8

Measuring inputs via nine A10 or TH boards (90 inputs) or 19 MU or KS boards (190 inputs) or four RTA5 output boards

Dimensions:

483 x 132 x 273 mm

(WxHxD)

Data acquisition system in 19-inch rack housing

CPU board, 19 free slots

MA56902CPUBT8

Measuring inputs via nine A10 or TH boards (90 inputs) or 19 MU or KS boards (190 inputs) or four RTA5 output boards

01.45

ALMEMO

®

Measuring Instruments

ALMEMO

®

5790-2CPUIG2

Technical data and functions

• Technical data and functions, as for ALMEMO

®

5690 series

• Robust aluminum housing, protective class IP65

• CPU board with measuring circuit (without measuring inputs) and output sockets

• Up to 20 measuring inputs / 80 measuring channels via selector switch boards

• Option XM - high-speed measuring operations, up to 20 measuring inputs / 80 measuring channels via active measuring circuit boards

The measuring circuit boards operate in parallel, thus ensuring short scanning times for a large number of channels. The scanning time is determined by the measuring circuit board with the highest number of active measuring channels - or, at conversion rate 50 Hz, also by the processing time of the CPU.

• Option - 5 ALMEMO

®

output sockets for digital interfaces, analog outputs, trigger, alarm contacts, socket P0 for integrated relay outputs

• Generously dimensioned graphics display, bright illumination, large display of measured values

• Measured values can be displayed graphically in line chart or bar chart form or numerically in various sizes.

• 3 user-defined menus can be freely configured from a range of

50 functions.

• Easy to operate by means of 4 soft-keys and cursor block, menu-guided with wizards and context-sensitive help windows

• Choice of languages : German, English, French (other options also available)

• Data logger with internal RAM (standard) or FeRAM (option) or external ALMEMO

®

memory connector with micro SD card

Technical data

Technical data, as for ALMEMO

®

5690 series

Measuring inputs / measuring channels

Standard up to 20 inputs / 80 measuring channels via selector switch boards

Option XM up to 20 inputs / 80 measuring channels via active measuring circuit boards

Memory, internal

Standard

Option SF sufficient for 400,000 values, linear or ring memory

RAM (buffered by battery)

FeRAM (non-volatile)

External memory (accessory) ALMEMO

®

memory connector with micro SD card

Outputs

Display

Graphics display

Illumination

5 ALMEMO

®

sockets, suitable for all output modules (analog / data / trigger / relay cables, etc.)

Alarm signal transmitter, internal

Socket P0 for integrated relay outputs

(option)

Or trigger and analog output

(by request)

128 x 128 pixels, 16 rows

5 white LEDs, 3 brightness levels

Operation

Power supply

Screwed cable glands

Housing

Dimensions

19-inch design

Weight

Protective class

Wall-mounting

Accessories

Memory connector with micro SD, including USB card reader (see chapter „General accessories“)

9 keys (4 soft-keys and cursor block)

9 status LEDs on front panel

Mains unit ZB1212NA6, installed on a fixed basis, 100 to 240 VAC, connected via appliance socket, including safety connecting cable

2 PGs with multiple inserts, slotted

24 drilled holes for cables d= 4 mm

2 drilled holes for cables d= 7 mm for all supply lines (sensor cables, output cables, e.g. data cable, mains supply cable) including dummy plugs for all holes

Aluminum

233 x approx. 350 x 121 mm (WxHxD)

(height includes PGs)

Plastic insert, 16 DUs approx. 6 kg

IP65

4 x M4 thread, including

2 aluminum profiles

ZA1904SD

Input boards

Option XM - selector switch boards and active measuring circuit boards

Order no.

see page 01.48

01.46

ALMEMO

®

5790-2CPUIG2

ALMEMO

®

Measuring Instruments

Data acquisition system in industrial housing, CPU board, 2 free slots

Measuring inputs via one A10 or TH board (10 inputs) or two MU or KS boards (20 inputs)

Dimensions:

233 x approx.350 x 121mm

(WxHxD),

(with PGs)

MA57902CPUIG2

Options

for active measuring circuit boards, up to 20 inputs / 80 channels

Data memory, internal FeRAM, non-volatile (instead of battery-buffered RAM)

Multi-point adjustment, special linearization, management of calibration data

Temperature ranges for 8 refrigerants (see 10.08)

Measuring rate for 1 measuring channel, 400 mops (SD card required) This cannot be combined with option XM.

For output socket P0

SH2 2 semiconductor relays (normally open) internal, 0.5 A, 50 V

OH2 2 additional relays (normally closed) for option SH2 (thus 2 changeover relays)

Power supply via rechargeable battery module

Rechargeable battery set (8 NiMH cells, 1600 mAh), 1 slot

Standard delivery

Precision measuring instrument, data acquisition system with graphics display and operating controls, CPU board

Measuring circuit (without measuring inputs) Input boards must be ordered separately. (see page 01.48)

Integrated mains unit ZB1212NA6, safety connecting cable, Operating instructions, manufacturer‘s test certifi cate

DAkkS or works calibration KE90xx, electrical, for measuring instrument (see chapter „Calibration certifi cates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

Order no.

OA5690XM

OA5690SF

OA5690KL

SB0000R2

SA0000Q4

OA5690SH2

OA5690OH2

OA5790A

ES5690AP

01.47

ALMEMO

®

Measuring Instruments

CPU board, selector switch boards, active measuring circuit boards and expansions for CPU systems ALMEMO

®

5690-1CPU and 5690-2CPU

CPU

U-A10

M-A10

U-MU

M-MU

U-TH

M-TH

U-KS (U/I)

M-KS (U/I)

AP RTA5

Input boards for ALMEMO

®

5690-1CPU and 5690-2CPU

Technical data and functions

• Selector switch boards U-xx for CPU systems without options

XU / XM or with option XU

• Active measuring circuit boards M-xx with own A/D converter for CPU systems with option XM

• There are several design variants for different installations / input plugs.

Input board U-A10 / M-A10

10 inputs for ALMEMO

®

single connectors.

For flexible applications with individual sensors and measuring signals.

Input board U-MU / M-MU

10 inputs for ALMEMO

®

10 MU connectors.

For permanently installing groups of 10, especially temperature sensors.

Technical data

Measuring inputs 10 ALMEMO

®

input sockets, electrically isolated

Measuring ranges All ranges (see page 01.05)

Sensor supply

Footprint

12 V, maximum 0.3 A (per system max. 2.5 A)

2 slots

Standard delivery

Selector switch board U-A10

Active measuring circuit board M-A10

(for CPU system with option XM)

Order no.

ES5690UA10

ES5690MA10

Technical data

Measuring inputs 10 inputs, electrically isolated, socket strip for ALMEMO

®

10-way MU connector

Measuring ranges all thermocouples, Pt100, Ni100, ohms, 2.6 V,

260 mV, 55 mV, 26 mV

Sensor supply

Footprint

None

1 slot

Standard delivery

Selector switch board U-MU

Active measuring circuit board M-MU

(for CPU system with option XM)

ALMEMO

®

10-way MU connector

Order no.

ES5690UMU

ES5690MMU

ZA5690MU

01.48

ALMEMO

®

Measuring Instruments

Input board U-TH / M-TH

10 inputs for miniature thermal connectors.

For any individual thermocouple temperature sensors with miniature thermal connector.

Input board U-KS / M-KS

10 inputs, electrically isolated, sensor connection via socket block.

For permanently installing groups of 10

Technical data

Measuring inputs 10 miniature thermal sockets, electr. isolated

ALMEMO

®

sensor parameters are saved in the measuring instrument.

Measuring ranges all thermocouples

Sensor supply None

Footprint 2 slots

Standard delivery

Order no.

Selector switch board U-TH

Active measuring circuit board M-TH

ES5690UTH

(for CPU system with option XM) ES5690MTH

Miniature thermal connectors must be ordered separately

Technical data

Measuring inputs 10 inputs, electrically isolated, male strip connector for socket block

ALMEMO

®

sensor parameters are saved in the measuring instrument.

Measuring ranges Pt100, Ni100, NTC, ohms, 2.6 V, 260 mV,

55 mV, 26 mV

Sensor supply

Footprint

None

1 slot

Standard delivery

Order no.

Selector switch board U-KS including socket block

ES5690UKS

Active measuring circuit board M-KS including socket block

(for CPU system with option XM)

Socket block (spare)

ES5690MKS

ZB5600KS

Input board U-KSU / M-KSU

10 inputs, electrically isolated, sensor connection via socket block.

For permanently installing groups of 10 with voltages 10 V.

Technical data

Measuring inputs 10 inputs, electrically isolated, male strip connector for socket block

ALMEMO

®

sensor parameters are saved in the measuring instrument.

Measuring ranges Voltage -26 to +26 V (integrated divider)

Accuracy, divider ±0.1 % of measured value

Sensor supply None

Footprint 1 slot

Input board U-KSI / M-KSI

10 inputs, electrically isolated, sensor connection via socket block.

For permanently installing groups of 10 with currents

20 mA.

Technical data

Measuring inputs 10 inputs, electrically isolated, male strip connector for socket block

ALMEMO

®

sensor parameters are saved in the measuring instrument.

Measuring ranges Current -32 to +32 mA (integrated shunt)

Accuracy, shunt

Sensor supply

Footprint

±0.1 % of measured value

None

1 slot

Standard delivery

Selector switch board U-KSU including socket block

Active measuring circuit board M-KSU including socket block

(for CPU system with option XM)

Socket block (spare)

Order no.

ES5690UKSU

ES5690MKSU

ZB5600KS

Standard delivery

Selector switch board U-KSI including socket block

Active measuring circuit board M-KSI including socket block

(for CPU system with option XM)

Socket block (spare)

Order no.

ES5690UKSI

ES5690MKSI

ZB5600KS

01.49

ALMEMO

®

Measuring Instruments

Universal ALMEMO

®

transmitter 2450 / 2490

• 1 or 2 measuring inputs

• Various outputs - digital, analog

• Various power supplies

ALMEMO

®

transmitter - a comparison

Measuring ranges

(see Table, page 01.10 / 01.11)

ALMEMO

®

2450

Compact measuring instrument

Over 35 measuring ranges, inter alia thermocouples, NTC, temperature / humidity, capacitive

C

ALMEMO

®

2490

Basic measuring instrument

Over 65 measuring ranges, inter alia

Pt100, Pt1000, thermocouples, NTC temperature / humidity, capacitive temperature / humidity, psychrometric

B Precision class technical data

(see page 01.05)

Measuring inputs ALMEMO

®

2450-1x 1 measuring input

Other technical data (see ALMEMO

®

2450, page 01.12)

Common technical data

Analog outputs

0.0 to 10.0 V

0.0 / 4.0 to 20.0 mA

Accuracy

Temperature drift

Time constant

10 V or 20 mA (programmable)

16-bit DAC, electrically isolated

0.5 mV / digit, load >100 kilohms

0.1 mA / digit, load <500 ohms

0.1% of meas. v. +0.1% of final v.

10 ppm / K

100 ms

Standard equipment

Housing

Operating temperature

Atmospheric humidity

ALMEMO

®

2490-1x 1 measuring input

ALMEMO

®

2490-2x 2 measuring inputs

(see ALMEMO

®

2490, page 01.14)

LCD screen, keypad

ABS (maximum 70 °C)

127 x 83 x 42 mm (LxWxH)

-10 to +60 °C

10 to 90 % RH (non-condensing)

Programming the analog output (Example)

Analog - start

01.50

Analog - end

ALMEMO

®

Measuring Instruments

Compact measuring instrument

ALMEMO

®

2450-1x

Universal transmitter with display for a wide variety of ALMEMO

®

sensors

Technical data

Measuring input

ALMEMO

®

2450-1x

Measuring ranges

1 ALMEMO

®

socket

(see Table, page 01.10 / 01-11)

Over 35 measuring ranges, inter alia

Thermocouples, NTC, temperature, humidity, capacitive

Other common data (see page 01.50)

Basic measuring instrument

ALMEMO

®

2490-1x / -2x

Universal transmitter with display for all ALMEMO

®

sensors

Technical data

Measuring input

ALMEMO

®

2490-1x

ALMEMO

®

2490-2x

Measuring ranges

1 ALMEMO

®

socket

2 ALMEMO

®

sockets

(see Table, page 01.10 / 01-11)

Over 65 measuring ranges, inter alia

Pt100, Pt1000, thermocouples, NTC

Temperature / humidity, capacitive

Temperature / humidity, psychrometric

Other common data (see page 01.50)

Variants

Digital transmitter

Option

Protective class IP54

(if water-proof plugs are used)

Option U Power supply, electrically isolated

Option I RS485 interface

Order no.

Measuring input for ALMEMO

®

sensors, LCD screen, 7 keys, with interface via 2 ALMEMO

®

output sockets A1, A2, and

1 ALMEMO

®

DC socket for mains adapter including 3 AA alkaline batteries, operating instructions, manufacturer’s test certificate,

Compact measuring instrument ALMEMO

®

2450-1

1 measuring input

MA24501

Analog transmitter, like the digital transmitter described above, plus integrated analog output via socket P0, electrically isolated

(scaling via keypad), including ALMEMO

®

clamp connectors

2 analog outputs (common ground), electrically isolated, 10 V or

20 mA (programmable)

Compact measuring instrument ALMEMO

1 Messeingang

®

2450-1,

MA24501R02

OA2450W

OA2450U

OA2450I

Variants

Digital transmitter

Order no.

Measuring input for ALMEMO

®

sensors, LCD screen, 7 keys, with interface via 2 ALMEMO

®

output sockets A1, A2, and

1 ALMEMO

®

DC socket for mains adapter including 3 AA alkaline batteries operating instructions, manufacturer’s test certificate

Basic measuring instrument ALMEMO

®

2490-1

1 measuring input

2490-2

MA24901

Basic measuring instrument ALMEMO

®

2 measuring inputs

MA24902

Analog transmitter, like the digital transmitter described above, plus integrated analog output via socket P0, electrically isolated

(scaling via keypad), including ALMEMO

®

clamp connectors

2 analog outputs (common ground), electrically isolated, 10 V or

20 mA (programmable)

Basic measuring instrument ALMEMO

®

1 measuring input

Basic measuring instrument ALMEMO

®

2490-2

2 measuring inputs

Option

Protective class IP54

(if water-proof plugs are used)

2490-1

Option U Power supply, electrically isolated

Option I RS485 interface

MA24901R02

MA24902R02

DAkkS or works calibration KE90xx, electrical, for measuring instrument (see chapter „Calibration certificates“)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

OA2490W

OA2490U

OA2490I

Accessories, options

Digital interface (see chapter „Networking“)

(please order separately)

Power supply

230 VAC via desktop mains unit 12 V, 1 A

10 to 30 VDC, maximum 80 mA, electrically isolated, integrated including ALMEMO

®

clamp connector

10 to 30 VDC, maximum 200 mA, electrically isolated, via DC adapter cable, with banana plugs

ZA1312NA7 see option U

ZA2690UK

10 to 30 VDC, not electrically isolated (not suitable for thermocouple measuring) including ALMEMO

®

clamp connector ZA1312FS1

Limit value contact (see chapter „Output modules“)

(Programming via digital interface, see above)

2 normally open contacts, 50 VDC / 500 mA

(can also be programmed as inverted) via ALMEMO

®

relay cable, V6, clamped connection ZA1006EKG

ALMEMO

®

limit value cable with banana plugs

(for electrical socket adapter)

Electrical safety socket adapter, 250 V / 6 A

(for ALMEMO

®

limit value cable)

Installation

ZA1006GK

ZB2280RA

USB interface via ALMEMO

®

USB cable

Ethernet interface via ALMEMO

®

Ethernet cable

RS232 interface via ALMEMO

®

RS232 cable

ZA1919DKU

ZA1945DK

ZA1909DK5

DIN rail

Magnet

ZB2490HS

ZB2490MH

RS485 interface, integrated including ALMEMO

®

clamp connector see option I

01.51

ALMEMO

®

Measuring Instruments

ALMEMO

®

4390-2

ALMEMO

®

precision measuring instrument in fi tted panel design with data logger function.

Comprehensive range of functions for all application areas

Increased measuring accuracy, fast measuring rate, 1 measuring input, 2 limit value relays, integrated. Option with double analog output.

Technical data and functions

• Increased measuring accuracy and stability

• Fast measuring rate, up to 50 measuring operations per second

With SD memory card, up to 100 mops, optional for 1 channel up to 400 mops

• 1 ALMEMO

®

input socket, suitable for all ALMEMO

®

sensors

• or 6-contact clamp connector socket, also for 26 V and 20 mA

• More than 65 standard measuring ranges

• Support for ALMEMO

®

plugs with multi-point adjustment, special linearization, and special measuring ranges

• Higher measuring quality thanks to electrical isolation between measuring inputs and device power supply (device ground)

• Data logger with internal EEPROM, suffi cient for 16,000 measured values, confi gurable as linear or ring memory

• Memory connector with micro SD (accessory)

• As standard 2 limit value relays can also be driven via interface

• Option with double analog output can also be driven via interface

Technical data

Precision class

Measuring rate

AA (see page 01.05)

2.5 / 10 / 50 / 100 mops

Measuring inputs

Accuracy

Channels

1 ALMEMO

®

input socket, suitable for all ALMEMO

®

sensors or 6-contact screw connector with input for 26 V (integrated divider) or 20 mA

(integrated shunt)

Divider / shunt ±0.1 % of measured value

4 channels for double sensors and function channels

Electrical isolation for analog sensors between measuring input and power supply (device ground)

Sensor power supply 12 V / 0.1 A; 9 V / 0.15 A; 6 V / 0.2 A

Outputs

2 limit value relays

2 ALMEMO

®

sockets, suitable for all output modules (analog / data / trigger / relay cables, memory, etc.)

Mechanical changeover, 230 V, 2 A

• 2 ALMEMO

®

output sockets, suitable for digital interfaces, analog output, trigger input, alarm contacts, memory card

• 8-character alphanumeric 14-segment display

• new:

Programming functions displayed in normal text (3 languages)

• new:

5 programming menus

Measuring function, memory, sensor, device, output

• Measuring functions

Measured value, dual display, smoothing, zero-setting, setpoint adjustment, maximum / minimum / average values, temperature compensation, atmospheric pressure compensation

• Sensor programming: Measuring range, measured value correction, scaling, units, limit value monitoring, graduated locking of functions, scaling of analog output

• Device programming: Conversion rate, real-time clock with date, output cycle, baud rate, choice of languages

Option with double analog output 10 V or 20 mA (programmable)

0.0 to 10.0 V

0.0 to 20.0 mA

Accuracy

Temperature drift

Time constant

16-bit DAC, electrically isolated

0.5 mV / digit, load >100 kilohms

0.1 mA / digit, load <500 ohms

0.1 % of fi nal value

10 ppm / K

100 μs

Standard equipment

Display

Keypad

8-character 14-segment LED display

5 membrane keys

Date and time-of-day Real-time clock, buffered with battery

Memory, internal EEPROM suffi cient for 16,000 measured values

Power supply

Mains operation

Option U

90 to 250 VAC, 50 / 60 Hz

10 to 30 V, 0.5 A, electrically isolated

Housing

Panel opening

Standard plastic housing

96 x 48 x 132 mm (WxHxD)

90 x 42.5 mm

Accessories

Memory connector with micro SD, including USB card reader (see chapter „Output modules“)

Order no.

ZA1904SD

Options

Measuring rate 400 mops (SD card required)

Power supply 10 to 30 VDC, electrically isolated

2 analog outputs (common ground), electrically isolated 10 V or 20 mA (programmable)

Temperature ranges for 8 refrigerants

Order no.

SA0000Q4

OA4390U

OA4390R02

SB0000R2

Standard delivery

Operating instructions, manufacturer‘s test certifi cate,

Precision measuring instrument ALMEMO

®

4390-2

DAkkS or works calibration KE90xx, electrical, for measuring instrument (see chapter „Calibration certifi cates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

Order no.

MA43902

01.52

High-precision measuring

The new reference measuring instruments ensure very high levels of resolution, precision, and linearity. They are thus ideally suitable as reference instruments in calibration laboratories and quality assurance procedures. They measure with resolution up to 0.001 K. These devices are offered in a set including sensor. They come in a compact design (with an optional variant with protective class IP54), an illuminated graphics display, and easy and convenient operation by soft-keys and the cursor block. There are two output sockets which can be used for connection to a PC or for networking. There is also a plugon measured value memory available as an option. Delivery includes evaluation software, data cable, temperature sensor,

DAkkS calibration certifi cate, mains unit, and measuring instrument case.

01.53

Reference Measuring Instruments

ALMEMO

®

1030-2

Reference measuring instrument for temperature.

High-precision measuring with

Pt100 sensors

Resolution 0.001 K

Technical features

• Temperature measurement with very high resolution, precision, and linearity, using Pt100 sensors

• Suitable as reference device in calibration laboratories and quality assurance procedures

• Very high accuracy thanks to multi-point adjustment of the

Pt100 temperature sensor

• 2 electrically isolated measuring inputs for Pt100 sensors

• Resolution can be set to 0.001 or 0.01 K.

• Units °C, °F, K

• High-resolution A/D converter, delta-sigma, 24-bit, 1.25 mops

(measuring operations per second)

• Two output sockets for digital interface, ALMEMO

®

memory connector

• Compact, modern, ergonomic design

• Graphics display, illuminated with white light

• Easy and convenient to operate by means of 4 soft-keys and cursor block

• Measured value display 2 measured values and differential

• Measuring functions: Zero-setting, smoothing, maximum / minimum values, individual value memory for 100 values

• Data logger with ALMEMO

®

memory connector (accessory)

• Sensor programming: Smoothing, designation, units, resolution

• Device confi guration: Illumination, contrast, device address, baud rate

• Choice of language: German, English, French

Technical data

Measuring inputs 2 ALMEMO

®

input sockets for Pt100 sensors

Electrical isolation Semiconductor relay (50 V)

A/D converter

Measuring range

Resolution

Delta-sigma, 24-bit, 1.25 mops

Pt100, -200 to +400 °C

0.001 K or 0.01 K

Measuring current 1 mA

Accuracy ±0.010 K ±1 digit in range -50 to +400 °C

Nominal conditions 23 °C ±2 K, 1013 mbar, battery mode

Temperature drift typical 2 ppm / K

Outputs 2 ALMEMO

®

sockets for interface cable and ALMEMO

®

memory connector

Standard equipment

Display

Illumination

Keypad

Graphics display, 128 x 64 pixels, 8 rows

2 white LEDs

7 silicone keys (of which 4 soft-keys)

Date and time-of-day Real-time clock, buffered by device battery

Individual value memory, internal 100 measured values

Power supply

Battery set

Mains adapter

3 AA alkaline batteries

ZA1312NA7 230 VAC to 12 VDC, 1 A, electrically isolated

Current consumption (without input and output modules) approx. 20 mA

With illumination approx. 40 mA

Housing 127 x 83 x 42 mm (LxWxH)

ABS (maximum 70 °C) 290 g

Pt100 temperature sensor FPA923L0250

Measuring element Pt100 as per DIN EN 60751

Class

Measuring tip

Response time T

90

1/10 B (DIN EN 60751) at 0 °C

Operative range -50 to +400 °C

5 seconds

250 mm Nominal length

Shaft Stainless steel, diameter 3 mm

Connecting cable 2 meters, FEP / silicone

ALMEMO

®

plug Resolution 0.001 K

Other sensor designs are available on request.

Accessories

Ethernet data cable

ALMEMO

®

memory connector with micro SD

Order no.

ZA1945DK

ZA1904SD

Rubberized impact protection, gray

DIN rail mounting

Order no.

ZB2490GS2

ZB2490HS

Standard delivery

Order no.

Reference measuring instrument for temperature measurement with accessories, evaluation software, and Pt100 temperature sensor. Complete set including DAkkS calibration certificate:

Reference measuring instrument ALMEMO

®

1030-2 including 3 AA alkaline batteries, Desktop mains unit ZA1312NA7,

USB data cable ZA1919DKU, Instrument case, evaluation software ALMEMO

®

View SW5500AV (see page 06.06) and Pt100 temperature sensor FPA923L0250 with DAkkS calibration certificate

(2 temperature points at 0 and 100 °C, including adjustment)

SP10302D

01.54

ALMEMO

®

1020-2

Reference Measuring Instruments

Reference measuring instrument for temperature

High-precision measuring by means of thermocouples

Types N, S, R, B

Resolution 0.01 K, up to 1800 °C

Technical features

• Temperature measurement with very high levels of resolution, precision, and linearity, using thermocouples Types N, S, R, B

• Suitable as reference device in calibration laboratories and quality assurance procedures

• Very high accuracy thanks to multi-point adjustment of the thermocouple temperature sensor

• Each temperature sensor has its own cold junction stored in the ALMEMO

®

plug or externally. The cold junction temperature in the ALMEMO

®

plug is measured to a very high resolution of 0.001 K by means of an NTC sensor.

• Two electrically isolated measuring inputs for thermocouples, types N, S, R, B

• Resolution 0.01 K

• Units °C, °F, K

• High-resolution A/D converter, delta-sigma, 24-bit, 1.25 mops

(measuring operations per second)

• Two output sockets for digital interface, ALMEMO® memory connector

• Compact, modern, ergonomic design

• Graphics display, illuminated with white light

• Easy and convenient to operate by means of 4 soft-keys and cursor block

• Measured value display : 2 measured values, differential, measuring point list, cold junction temperature

• Measuring functions : Zero-setting, smoothing, maximum / minimum values, individual value memory for 100 values

• Data logger with ALMEMO

®

memory connector (accessory)

• Sensor programming : Smoothing, designation, units

• Device confi guration : Illumination, contrast, device address, baud rate

• Choice of language : German, English, French

Technical data ALMEMO

®

1020-2

Measuring inputs

Electrical isolation

A/D converter

Measuring ranges

NiCrSi-NiSi Type N

PtRh10-Pt Type S

PtRh13-Pt Type R

Resolution

Accuracy

Type N

Type S

2 ALMEMO

®

input sockets for thermocouples

Semiconductor relay (50 V)

Delta-sigma, 24-bit, 1.25 mops

-200 to +1300 °C

-50 to +1768 °C

-50 to +1768 °C

PtRh30-PtRh6 Type B +250 to +1820 °C

Type R

Type B

Nominal conditions

Temperature drift

0.01 K

±0.1 K ± 1 digit in range

-200 to +1300 °C

+50 to +1760 °C

+100 to +1760 °C

+500 to +1800 °C

23 °C ±2 K, 1013 mbar, battery mode typical 10 ppm / K

Cold junction temperature Measuring operations with 0.001 K resolution

Outputs

2 ALMEMO

®

sockets for interface cable and ALMEMO

®

memory connector

Standard equipment

Display

Illumination

Keypad

Graphics display, 128 x 64 pixels, 8 rows

2 white LEDs

7 silicone keys (of which 4 soft-keys)

Date and time-of-day Real-time clock, buffered by battery

Individual value memory, internal 100 measured values

Power supply

Battery set

Mains adapter

Current consumption (without input and output modules)

With illumination

3 AA alkaline batteries

ZA1312NA7 230 VAC to 12 VDC, 1 A, electrically isolated approx. 20 mA approx. 40 mA

Housing 127 x 83 x 42 mm (LxWxH)

ABS (maximum 70 °C), 290g

Accessories

Ethernet data cable

ALMEMO

®

memory connector with micro SD

Rubberized impact protection, gray

DIN rail mounting

Order no.

ZA1945DK

ZA1904SD

ZB2490GS2

ZB2490HS

01.55

Reference Measuring Instruments

Variants

Complete set comprising reference measuring instrument for temperature plus accessories, evaluation software, thermocouple sensor, with DAkkS calibration certificate

Reference measuring instrument ALMEMO

®

1020-2, including 3 AA alkaline batteries, desktop mains unit ZA1312NA7,

USB data cable ZA1919DKU, instrument case, and evaluation software ALMEMO

®

View SW5500AV (see page 06.16)

Set with high-precision sheathed thermocouple sensor type N

Set Order no.

with sheathed thermocouple sensor type N FTAN926L0500P2 with DAkkS calibration certificate at 0 / 100 / 500 / 1000 °C, including adjustment

SP10202ND

Technical data:

Sheathed thermocouple sensor type N FTAN926L0500P2

Measuring element NiCrSi-NiSi, type N, class 1

Measuring tip

Operative range

Connecting cable

ALMEMO

®

plug

Mineral-insulated sheathed line, d = 6 mm, L = 500 mm

-200 to +1150 °C

1.5 meters, thermal line (stranded wire)

FEP / silicone (-50 to +200 °C)

Resolution 0.01 K with integrated cold junction compensation sensor

Set with high-precision thermocouple sensor type S

Set Order no.

with thermocouple sensor type S FTAS917L0700P2 replacement ceramic protective tube, case for sensors

ZB9000TK2 with DAkkS calibration certificate at 500/1000/1200 °C, including adjustment

SP10202S1D

Technical data:

Thermocouple sensor type S FTAS917L0700P2

Measuring element PtRh10-Pt, Type S, Class 1

Measuring tip

Operative range

Thermowire, d = 0.5 mm in ceramic protective tube diameter = 7 mm, length = 700 mm up to +1400 °C

Connection head

Connecting cable

ALMEMO

®

plug ceramic protective tube, screwed

1.5 meters, compensation line

FEP / silicone (-50 to +200 °C)

Resolution 0.01 K with integrated cold junction compensation sensor

Set with precision thermocouple sensor type S, with external cold junction

Set Order no.

with thermocouple sensor type S, with external cold junction

FTAS907L0700P2, replacement ceramic protective tube,

Case for sensors ZB9000TK2 with DAkkS calibration certificate at 500 / 1000 / 1200 °C, including adjustment

SP10202S2D

Technical data:

Thermocouple sensor type S, with external cold junction

FTAS907L0700P2

Measuring element PtRh10-Pt, Type S, Class 1

Measuring tip

Operative range

Connection head

Connecting cable

Cold junction

Connecting cable

ALMEMO

®

plug

Thermowire, d = 0.5 mm in ceramic protective tube diameter = 7 mm, length = 700 mm up to +1600 °C ceramic protective tube, screwed

0.75 meters, insulated, thermo-wires

PtRh10-Pt as far as cold junction

Stainless steel protective tube diameter = 5 mm, length = 250 mm

2 meters, stranded copper wire

Resolution 0.01 K

01.56

Reference Measuring Instruments

Certificates

Calibration certificate for ALMEMO® 1020-2 with precision sheathed thermocouple sensor type N

(Example)

Calibration certificate for ALMEMO® 1020-2 with precision thermocouple sensor type S,

with external cold junction (Example)

Other certificates for measuring instruments and sensors (see chapter „Calibration certificates“)

01.57

Reference Measuring Instruments

ALMEMO

®

1036-2

Reference measuring instrument for humidity

High-precision measuring with Pt100 psychrometer

Resolution Temperature 0.001 K

Relative humidity 0.01 %

Dew point 0.01 K

Technical features

• Humidity measurement with very high resolution, precision, and linearity, using Pt100 psychrometer

• Suitable as reference device in calibration laboratories and quality assurance procedures

• Very high level of accuracy using the Pt100 psychrometer thanks to multi-point adjustment of the two temperature sensors

• Pt100 psychrometer optimized for measuring operations involving high humidity levels performed over long periods

• new:

Automatic atmospheric pressure compensation is provided for pressure-dependent humidity variables by means of a digital atmospheric pressure sensor integrated in the AL-

MEMO

®

device.

• new:

Humidity calculation on the basis of formulae as per Dr.

Sonntag and the enhancement factor as per W. Bögel (correction factor fw(t,p) for real mixed gas systems). This substantially widens the measuring range and improves the accuracy of humidity variable calculations.

• Resolution : Temperature Pt100 0.001 K,

Relative humidity 0.01%, Dew point 0.01 K

• The humidity variables are calculated from the three primary measuring channels (real measurable variables). Dry temperature (°C), humid temperature (°C), atmospheric pressure

(mbar)

• Three humidity variables displayed simultaneously, freely selectable : Relative humidity (%), dew point (°C), mixture

(g/kg),

new:

Absolute humidity (g/m³), vapor pressure (mbar), enthalpy (kJ/kg)

• Two electrically isolated measuring inputs for Pt100 sensors

• High-resolution A/D converter, delta-sigma, 24-bit, 1.25 mops

(measuring operations per second)

• Two output sockets for digital interface, ALMEMO

®

memory connector

• Compact, modern, ergonomic design

• Graphics display, illuminated with white light

• Easy and convenient to operate by means of 4 soft-keys and cursor block

• Measured value display : Sensor display (up to 4 measured values), measuring points list, atmospheric pressure

• Measuring functions : Zero-setting, smoothing, maximum / minimum values, individual value memory for 100 values

• Data logger with ALMEMO

®

memory connector (accessory)

• Sensor programming : Smoothing, designation, measuring range selection, locking

• Device confi guration : Illumination, contrast, device address, baud rate, atmospheric pressure

• Choice of language : German, English, French

• Humidity measurement in temperature range -100 to +200 °C, with precision digital capacitive temperature / humidity sensors FHAD 36 Rx, with ALMEMO

®

D6 connector (Accessories, see chapter „Atmospheric humidity“). Confi guration of

ALMEMO

®

D6 sensors on ALMEMO

®

device itself

Technical data ALMEMO

®

1036-2

Measuring inputs

Electrical isolation

A/D converter

Measuring range

Resolution

Two ALMEMO

®

input sockets for Pt100 psychrometer FPA 836-3P3 or Precision digital capacitive temperature / humidity sensors FHAD 36 Rx

Semiconductor relay (50 V)

Delta-sigma, 24-bit, 1.25 mops

Pt100, -200 to +400 °C

0.001 K

Measuring current

Accuracy

1 mA

±0.010 K ±1 digit in range -50 to +400 °C

Nominal conditions 23 °C ±2 K, 1013 mbar, battery mode

Temperature drift typical 2 ppm / K

Calculated humidity quantities Analytic equation

(not an approximation)

Digital atmospheric pressure sensor (integrated in the device)

Measuring range

Accuracy

700 to 1100 mbar

±2.5 mbar (at 0 to +65 °C)

Outputs Two ALMEMO

®

sockets for interface cable and ALMEMO

®

memory connector

Standard equipment

Display

Keypad

Graphics display, 128 x 64 pixels, 8 rows

Illumination 2 white LEDs

7 silicone keys (of which 4 soft-keys)

Date and time-of-day Real-time clock, buffered by battery

Individual value memory, internal 100 measured values

Power supply

Battery set

Mains adapter

Current consumption (without input and output modules)

With illumination

3 AA alkaline batteries

ZA1312NA7 230 VAC to 12 VDC, 1 A, electrically isolated approx. 20 mA approx. 40 mA

Housing 127 x 83 x 42 mm (LxWxH)

ABS (maximum 70 °C), 290g

01.58

Reference Measuring Instruments

Psychrometer FPA 836-3P3

Technical data Pt100 psychrometer FPA 836-3P3

Operating temperature up to +90 °C (no ice)

Humidity measuring range approx. 10 to 100 % RH

Measuring system

Accuracy psychrometric

< ±1 % RH under nominal conditions

Dimensions 175 x 50 x 75 mm (LxWxH)

Ventilator power supply 12 VDC via mains unit cable, approx. 1.5 meters

(included in delivery)

Nominal conditions

Temperature sensors

Housing

23 °C ±2 K, 1013 mbar, 50 % RH

2 x Pt100, class B, ALMEMO

®

adjusted

Plastic PMMA

Connecting cables

ALMEMO

®

plug

2 cables, each 5 meters,

FEP / silicone

Pt100, resolution 0.001 K

Accessories

Ethernet data cable

ALMEMO

®

memory connector with micro SD

Rubberized impact protection, gray

DIN rail mounting

Spare wicks (2 pieces)

Extension cable for mains units, 3-pin bayonet coupling, length 5 meters

Order no.

ZA1945DK

ZA1904SD

ZB2490GS2

ZB2490HS

ZB98462ED

ZB5090VK05

Variants

Reference measuring instrument for humidity measurement with accessories, evaluation software, and Pt100 psychrometer, Complete set including DAkkS calibration certificate

Order no.

Reference measuring instrument ALMEMO

®

1036-2, with integrated digital atmospheric pressure sensor including 3 AA alkaline batteries, desktop mains unit ZA1312NA7, USB data cable ZA1919DKU, instrument case, and evaluation software ALMEMO View SW5500AV (see page 06.16) and Pt100 psychrometer FPA 836-3P3 including mains unit, water bottle, pair of wicks with DAkkS calibration certificate

Temperature at approx. +25 °C, relative humidity at approx. 30 % / 70 % RH, and atmospheric pressure in range 700 to 1100 mbar (5 points)

SP10362D

01.59

01.60

Input connectors and adapter cables

Content

Make the most of your existing sensor technology

High-precision measuring operations using inexpensive standard sensors - thanks to Multi-point Correction.

Connectors for Thermocouple

ALMEMO

®

Measuring Module for Thermocouples

Connector for Thermocouple

Connector with integrated cold junction sensor

Connector for Pt100 Sensors/Pt1000 Sensors

Connector for Ni100 Sensors/Ni1000 Sensors

Connector for Ntc Sensors

Connector for Resistance

Connector for Potentiometer pickoffs

Connector for Volt DC

Input connector for Measuring Bridges

Measuring Module for DC Voltage

Connector for DC Current

Measuring Module for DC

Adapter Cables for AC

Measuring Module for AC Voltage

Measuring Module for AC

Adapter Cables for Frequency / Pulse / Rotational Speed

Adapter Cable for Digital Input Signals

Universal Adapter Cables with Free Ends

MU Connector for ALMEMO

®

Plug-In Boards

Connector Adapter Cable, Digital Input of Third Party Device to ALMEMO

®

Device

02.02

02.06

02.06

02.07

02.07

02.09

02.09

02.10

02.11

02.03

02.04

02.04

02.05

02.05

02.05

02.06

02.11

02.12

02.12

02.13

02.13

02.14

02.14

02.15

02.01

Input connectors and adapter cables

ALMEMO

®

Input connectors

Make the most of your existing sensor technology

Our patented intelligent connector makes the ALMEMO

®

measuring system extraordinarily fl exible. Thus, instead of our pre-confi gured ALMEMO

®

sensors, you can use your own existing sensors.

We can supply you with ALMEMO

® connectors specially pre-programmed for this purpose with the necessary sensor parameters and the appropriate measuring range. These have six screw terminals and can be easily and conveniently connected.

All devices and connectors offer the following functions :

• Each measuring point can be assigned a specifi c designation.

• The sensor signals can be scaled.

• Measured values can be corrected for zero-point and gain.

The new measuring instruments with

ALMEMO

®

connectors also offer the following additional functions :

• Multi-point calibration data can be saved in the connector.

• User-defi ned linearization with up to

30 points can be programmed in the connector.

• Control points with actual / setpoint value table can be entered quickly and easily via the AMR-Control software.

• Any special measuring ranges programmed in the connector can be processed.

• Calibration schedules can be managed in the connector and are detected automatically.

• The connector’s exact designation can be called up.

The already high level of precision and overall performance quality provided by

ALMEMO

®

measuring technology is thus raised even further.

Give us a description of your measuring tasks !

And we shall provide you with comprehensive advice and fi nd the most cost-effective solution.

Please do not hesitate to ask !

02.02

Input connectors and adapter cables Input connectors and adapter cables

High-precision measuring operations using inexpensive standard sensors - thanks to multi-point correction. Linearization and correction of non-linear sensors

Linearization and correction at over 30 points - performed by the user - without further processing on the PC

Although special-purpose sensors with a non-linear output can usually be connected to existing measuring systems, the lack of linearization in the sensor’s output signal means that the measured value will need subsequent correction to make it at all usable. AHLBORN now offers customers a revolutionary new feature - also available with hand-held devices. An option is now available allowing the user to perform linearization and multi-point correction on

ALMEMO

®

measuring instruments. Not only all the relevant sensor characteristics but also the linearization or multi-point correction data are saved in the patented

ALMEMO

®

connector. The measuring instrument automatically recognizes each sensor that is connected to it and shows the appropriate measured values precisely in its display.

Sensor-specific linearization data can be saved by the user in the connector itself

Thanks to further development of the flexible and intelligent ALMEMO

® connector it is now possible to save

complex tables for linearization or multipoint correction - all in the connector itself. For the user this means that it

is now also possible to connect sensors with a non-linear output. The device displays measured values already in

linearized form; this ensures that the whole process can be monitored right from the outset. A further advantage

is the enormous saving in time when evaluating special measuring operations of this nature. For each sensor the linearization data is saved in the connector; then as soon as the sensor is plugged into the measuring instrument this data is loaded automatically. The linearization table is buffered in the main working memory on the device for the period of the measuring operation in question or for as long as the sensor remains connected. With effect from the

ALMEMO

®

2690-8 the user can use this

“KL” option to program linearization processes of this nature quickly and easily.

Individual linearization processes can be applied in the voltage, current, resistance, or frequency ranges. On request - or for other devices - readily pre-programmed connectors can be obtained from the factory. Various already implemented special-purpose linearizations are also available.

High-precision measuring operations - thanks to multi-point correction

A sensor’s output signal can also be corrected at various specific points.

Inexpensive standard sensors made by third-party manufacturers can be calibrated. Deviations are then saved in the sensor connector as fine corrections.

This can be performed either by users themselves or on request in advance at the factory - for example for temperature calibrations. It is now possible to save not only previous characteristics but also over

30 correction points - all in the connector itself. In the new digital ALMEMO

®

D6 sensors (code „D6“) stored at the factory all matching data in the digital sensor element. The multi-point correction / linearization using the ALMEMO

®

device with KL option is NOT applicable.

Programming via software

In the AMR-CONTROL software package the measuring protocol for a multi-point correction or a linearization table can be transferred to a table of reference points. Over 30 such reference points are possible. During a measuring operation the measured values between these are interpolated on a linear basis. The AMR-

CONTROL software is included with all our instruments free-of-charge.

System requirements

Connector - the new generation (code

“E4”), not for digital ALMEMO

®

D6sensors (additional code „D6“) For evaluation purposes : ALMEMO

®

devices in version V6 (2490, 2470, 2590-2/3S/4S,

2690, 2890, 4390, 8590, 8690, 5690)

For user-defined programming : Option

“KL” with devices 2690-8, 2890-9,8590,

8690 and 5690

02.03

Input connectors and adapter cables

ALMEMO

®

Connector for Thermocouple Types K, N, L, J, T

Variants ( with thermal material)

Model Meas. Range

NiCr-Ni (K) –200.0 to +1370.0°C.

NiCroSil-NiSil (N) –200.0 to +1300.0°C.

Fe-CuNi (L)

Fe-CuNi (J)

Cu-CuNi (T)

–200.0 to +900°C.

–200.0 to +1000°C.

–200.0 to +400°C.

Resolution

0.1 K

0.1 K

0.1 K

0.1 K

0.1 K

Order no.

ZA9020FS

ZA9021FSN

ZA9021FSL

ZA9021FSJ

ZA9021FST

ALMEMO

®

measuring module for thermocouples, types K, J, T, electrically isolated, up to 1000 V Type ZAD 950 AB

• Electrically isolated measurement of thermocouples (in particular bare thermo-wire types) on live parts

• Digital transfer of measured values to the ALMEMO

®

measuring instrument

• Connecting cable, fitted with ALMEMO

®

plug

Technical data

Sensor Thermocouple

Measuring range

ZAD950ABK NiCr-Ni (K) -200 to 1370 °C

ZAD950ABJ Fe-CuNi (J) -200 to 1000 °C

ZAD950ABT Cu-CuNi (T) -200 to 400 °C

Resolution 0.1 K

Linearization accuracy ±0.05 K ±0.05 % of measured value

Precision class C (see page 01.05)

Electrical isolation 1 kV DC/AC permanent, 4 kV for 1s

Sensor connection 4-mm safety sockets and safety plugs

(with screw terminals)

Power supply 6 to 13 VDC via ALMEMO

®

device

Current consumption approx. 30 mA

Connecting cable 1.5 meters with ALMEMO

®

plug

Housing Dimensions (LxWxH) 127x83x38mm,

ABS (acrylonitrile butadiene styrene)

Types:

ALMEMO

®

measuring module for NiCr-Ni (K), including 1.5 meters ALMEMO

®

connecting cable

ALMEMO

®

measuring module for Fe-CuNi (J) including 1.5 meters ALMEMO

®

connecting cable

ALMEMO

®

measuring module for Cu-CuNi (T) including 1.5 meters ALMEMO

®

connecting cable

Please note : thermocouple must be ordered extra; e.g. thermo-wires see Chapter Temperature

DAkkS- or Factory calibration KE90xx, electrically, for digital measuring module, see Chapter Calibration.

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

02.04

Order no.

ZAD950ABK

ZAD950ABJ

ZAD950ABT

Input connectors and adapter cables

ALMEMO

®

Connector for Thermocouple Types U, S, R, B, AuFe-Cr

Types

Model

Cu-CuNi (U)

PtRh10-Pt (S)

PtRh13-Pt (R)

PtRh30-PtRh6 (B)

AuFe-Cr (A)

Meas. Range

–200.0 to +600.0°C

0.0 to +1760.0°C

0.0 to +1760.0°C

+400.0 to +1800.0°C

–270.0 to +60.0°C

Resolution

0.1 K

0.1 K

0.1 K

0.1 K

0.1 K

Order no.

ZA9000FSU

ZA9000FSS

ZA9000FSR

ZA9000FSB

ZA9000FSA

ALMEMO

®

Connector with integrated cold junction sensor for all thermocouples

Types:

Model

NiCr-Ni (K)

NiCroSil-NiSil (N)

Fe-CuNi (L)

Fe-CuNi (J)

Cu-CuNi (T)

Cu-CuNi (U)

PtRh10-Pt (S)

Meas. Range

–200.0 to +1370.0°C.

–200.0 to +1300.0°C.

–200.0 to +900°C.

–200.0 to +1000°C.

–200.0 to +400°C.

–200.0 to +600.0°C

0.0 to +1760.0°C

Resolution

0.1 K

0.1 K

0.1 K

0.1 K

0.1 K

0.1 K

0.1 K

For especially exacting applications demanding the highest possible level of precision or performed under unfavorable conditions (e.g. subject to thermal irradiation)

Programming:

1st channel, NTC, integrated cold junction sensor, resolution

0.01 K

2nd channel, thermocouple, resolution 0.1 K; please specify type !

Order no.

ZA9400FSK

ZA9400FSN

ZA9400FSL

ZA9400FSJ

ZA9400FST

ZA9400FSU

ZA9400FSS

ALMEMO

®

Connector for Pt100 Sensors/Pt1000 Sensors

Types:

Model Meas. Range

Pt100 4-conductor

Pt100 4-conductor

–200.0 to +850.0°C

–200.0 to +400.0°C *

Pt1000 4-conductor –200.0 to +850.0°C *

Pt1000 4-conductor –200.0 to +400.0°C *

Order no.

Resolution

0.1 K

0.01 K

0.1 K

0.01 K

ZA9030FS1

ZA9030FS2

ZA9030FS4

ZA9030FS5

* Data may vary depending on device; (see data sheet per device)

02.05

Input connectors and adapter cables

ALMEMO

®

Connector for Ni100 Sensors/Ni1000 Sensors

Types:

Model

Ni100

Ni1000

Meas. Range

–60.0 to +240.0°C

–60.0 to +240.0°C

Resolution

0.1 K

0.1 K

ALMEMO

®

Connector for Ntc Sensors

Order no.

ZA9030FS3

ZA9030FS6

Types:

Model

Ntc Typ N

2xNtc Typ N

Meas. Range

–50.0 to +125.0°C

–50.0 to +125.0°C

Resolution

0.01 K

0.01 K no electrical isolation

Order no.

ZA9040FS

ZA9040FS2

ALMEMO

®

Connector for Resistance

Types:

Model

Ohm

Ohm kOhm

Meas. Range

0.00 to 500.00

0.0 to 5000.0*

0 to 110.00 kOhm

Resolution

0.01 Ω∗

0.1 Ω∗

0.01 kOhm

Technical Data ZA9003SS4:

Connection 2-wire

Linearization accuracy: ±0,2 % ± 0,02 kOhm

Linearization is saved in the

ALMEMO

®

connector; (this is not available with ALMEMO

®

2450, 8390)

Order no.

ZA9003FS

ZA9003FS2

ZA9003SS4

* Data may vary depending on device; (see data sheet per device)

02.06

Input connectors and adapter cables

ALMEMO

®

Connector for Potentiometer pickoffs

Sensor supply 2,5V potentiometer

Technical Data

Sensor supply:

Temperature coefficient:

2.5 V

< 50 ppm/K

Types:

Model Meas. Range Resolution

2.6 V DC Differenz –2.6 to +2.6* 0.1 mV

* Data may vary depending on device; (see data sheet per device)

ALMEMO

®

Connector for Voltage Millivolt

Order no.

ZA9025FS3

Types:

Model

55 mV DC

26 mV DC

260 mV DC

Meas. Range

–10.0 to +55.0

–26.0 to +26.0

–260.0 to +260.0

ALMEMO

®

Connector for Volt DC

Resolution

1 µV

1 µV

10 µV

Technical Data

Accuracy divider: only 5.5 / 26 V connector,

±0.1% of measured value

Order no.

ZA9000FS0

ZA9000FS1

ZA9000FS2

Types:

Model

2.6 V DC

5.5 V DC (divider 100:1)

26 V DC (divider 100:1)

2 mal 26 V DC (2 x divider)

Meas. Range

–2.6 to +2.6*

-1.0 to 5.5

–26.0 to +26.0

–26.0 to +26.0

Order no.

Resolution

0.1 mV

ZA9000FS3

0.1 mV

1 mV

ZA9602FS4

ZA9602FS

1 mV no electrical isolation

ZA9602FS2

* Data may vary depending on device; (see data sheet per device)

02.07

Input connectors and adapter cables

ALMEMO

®

Connector for DC voltage difference millivolts / volt

for sensors / transmitters, Supply from ALMEMO

®

device

Technical Data

Sensor supply

Accuracy divider:

(for voltage see technical data of ALMEMO

®

device) only 26V connector

±0,1% of measured value

Types:

Model

55 mV DC

26 mV DC

260 mV DC

2.6 V DC

26 V DC (divider 100:1)

Meas. Range Resolution

–10.0 to +55.0

–26.0 to +26.0

–260.0 to +260.0

–2.6 to +2.6*

1 µV

1 µV

10 µV

0.1 mV

–26.0 to +26.0 1 mV

(Connection diagram for connectors with 4 clamps, see below)

* Data may vary depending on device; (see data sheet per device)

Order no.

ZA9000FS0D

ZA9000FS1D

ZA9000FS2D

ZA9000FS3D

ZA9602FS3

ALMEMO

®

Connector for DC Millivolt / Volt Differential

for sensors / transmitters, Supply : 12 V from the ALMEMO

®

device

Sensor

Technical Data

Sensor supply U

Device voltage U

:

G

:

Output current:

Accuracy divider:

Types:

Model

55mV DC

26mV DC

260mV DC

2.6V DC

26V DC

Meas. Range

Order no.

Resolution

–10.0 to +55.0

–26.0 to +26.0

–260.0 to +260.0

–2.6 to +2.6*

-26.0 to +26.0

1 µV

1 µV

10 µV

0.1 mV

1 mV

* Data may vary depending on device; (see data sheet per device).

12.2 ... 12.5V (15V on request)

8 ... 12 V

100mA at U

G

= 9 ... 12V only 26V connector

±0,1% of measured value

ZA9600FS0V12

ZA9600FS1V12

ZA9600FS2V12

ZA9600FS3V12

ZA9602FS3V12

02.08

Input connectors and adapter cables

ALMEMO

®

Connector for measuring bridges, millivolt / volt differential

With zero-symmetrical voltage supply of ±2.5 V stabilized from the ALMEMO

®

device

Bridge circuit

Technical Data

Sensor supply

Voltage U

F

:

Temperature coefficient:

Output current:

Ruhestrom:

new:

Energy saving

5V ± 0.05V

<50ppm/°C max. 100mA approx. 3 mA

So long as the measuring point is not selected, the bridge voltage remains switched OFF.

Types:

Model

55mV DC

26mV DC

260mV DC

2.6V DC

Meas. Range Resolution

–10.0 to +55.0

–26.0 to +26.0

–260.0 to +260.0

–2.6 to +2.6*

1 µV

1 µV

10 µV

0.1 mV

* Data may vary depending on device; (see data sheet per device)

ALMEMO

®

Measuring Module for DC Voltage, with Electrical Isolation, 4kV

Technical Data

see Chapter Electrical variables

Order no.

ZA9105FS0

ZA9105FS1

ZA9105FS2

ZA9105FS3

Types:

Measuring range

±2.000 V

±20.00 V

±200.0 V

±400 V

Resolution

0.001V

0.01V

0.1V

1V

Overload

400 V

500 V

500 V

1000 V

Internal resistance

800 kΩ

1 MΩ

1 MΩ

4 MΩ

DAkkS- or Factory calibration KE90xx, electrically, for digital measuring module, see Chapter Calibration.

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

Order no.

ZA9900AB2

ZA9900AB3

ZA9900AB4

ZA9900AB5

02.09

Input connectors and adapter cables

ALMEMO

®

Connector for DC Current mA

Technical Data

Accuracy shunt: ±0,1% of measured value

Types:

Model

32 mA DC

4/20 mA DC

2 mal 32 mA DC

2 mal 4/20 mA DC

Meas. Range

–32.0 to +32.0*

0 to 100%

–32.0 to +32.0*

0 to 100%

Order no.

Resolution

1 µA

0.01 %

1 µA no electrical isolation

0.01 % no electrical isolation

ZA9601FS1

ZA9601FS2

ZA9601FS3

ZA9601FS4

* Data may vary depending on device; (see data sheet per device)

ALMEMO

®

Connector for DC mA Differential

for sensors / transmitters, Supply from the ALMEMO

®

device

Technical Data

Sensor supply

Accuracy shunt:

(for voltage see technical data of ALMEMO

®

device)

±0,1% of measured value

Types:

Model

32 mA DC

4/20 mA DC

Meas. Range

–32.0 to +32.0*

0 to 100%

Resolution

1 µA

0.01 %

Order no.

ZA9601FS5

ZA9601FS6

* Data may vary depending on device; (see data sheet per device)

ALMEMO

®

for DC mA Differential

for sensors / transmitters, Supply 12V from the ALMEMO

®

device

Technical Data

Sensor supply U

F

Device voltage U

:

G

:

Output current:

2-wire transmitter

(3-wire transmitter can also be connected)

Accuracy shunt:

Types:

Model

32mA DC

4-20mA DC

02.10

Meas. Range Resolution

–32.0 to +32.0* 1 µA

0 to 100% 0.01 %

* Data may vary depending on device; (see data sheet per device)

12,2 ... 12,5V

8 ... 12V

100mA at U

G

= 9 ... 12V

±0,1% of measured value

Order no.

ZA9601FS5V12

ZA9601FS6V12

Input connectors and adapter cables

ALMEMO

®

Measuring Module for DC, with Electrical Isolation, 4kV

Technical Data

see Chapter Electrical variables

Types:

Measuring range

±20.00 mA

±200.0 mA

±2.000 A

±10.00 A

Resolution

0.01mA

0.1mA

0.001A

0.01A

DC via external shunt:

±200.0 mV 0.1mV

Overload Internal resistance

Order no.

0.1 A* 10 Ω

ZA9901AB1

1 A*

10 A*

20 A*

1 Ω

0.1 Ω

0.01 Ω

ZA9901AB2

ZA9901AB3

ZA9901AB4

*Without fuse, overload condition only up to 1 minute maximum

40 V 50 kΩ

DAkkS- or Factory calibration KE90xx, electrically, for digital measuring module, see Chapter Calibration.

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

ZA9900AB1

ALMEMO

®

Adapter Cable for AC Voltage

Technical Data

Frequency range: 50 Hz to 10 kHz

Accuracy: ± 0.2% of final val. ± 0.5% of meas. val.

(40Hz … 2kHz sinusoidal),

Crest factor: 3 (add. error 0.7%), 5 (add. error 2.5%)

!

NEVER connect voltages higher than 50V!

DANGER!

Types:

Meas. Range

5 to 260mV eff

0.05 to 2.6V

eff

0.5 to 26.0V

eff

Resolution

0.1 mV

0.001 V

0.01 V

Order no.

ZA9603AK1

ZA9603AK2

ZA9603AK3

02.11

Input connectors and adapter cables

ALMEMO

®

Measuring Module for AC Voltage, with Electrical Isolation, 4kV

Technical Data

see Chapter Electrical variables

Types:

Meas. range Resolution Peak

130.0mV

eff

0.1mV 0.2V

Overload

400V

Internal resistance

0.5MΩ

1.300V

eff

13.00V

eff

130.0V

eff

400V eff

1mV

10mV

0.1V

1V

2V

20V

200V

1000V

400V

500V

500V

1000V

0.8MΩ

1MΩ

1MΩ

4MΩ

DAkkS- or Factory calibration KE90xx, electrically, for digital measuring module, see Chapter Calibration.

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

ALMEMO

®

Measuring Module for AC, with Electrical Isolation, 4kV

Technical Data

see Chapter Electrical variables

Order no.

ZA9903AB1

ZA9903AB2

ZA9903AB3

ZA9903AB4

ZA9903AB5

Types:

Measuring range

1.000A

eff

10.00A

Resolution

1mA

10mA

Peak

2A

20A

Order no.

Overload

10A*

20A*

Internal resistance

0.10Ω

0.01Ω

DAkkS- or Factory calibration KE90xx, electrically, for digital measuring module, see Chapter Calibration.

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

02.12

ZA9904AB1

ZA9904AB2

Input connectors and adapter cables

ALMEMO

®

Adapter Cable for Frequency / Pulse / Rotational Speed

for sensors, Supply : 5 V or direct from ALMEMO

®

device

Technical Data

Frequency range:

Speed range:

0 to 15000 Hz (Resolution 1 Hz)

0 to 3200.0 Hz (Resolution 0.1 Hz)

8 to 32000 rpm (Resolution: 1 rpm)

Max. pulse count:

Pulse length:

65000

> 50 ms

Input voltage 4 to 40 V, square-wave via optocoupler

Current consumption: 3 mA switch NPNoutput

PNPoutput

Sensor supply 5 V or direct from ALMEMO

®

device

(for voltage see technical data of ALMEMO

®

device)

Option V12

Sensor supply:

Output current:

13.5V ±0.5V

100mA at U

50mA at U

G

20mA at U

G

G

= 12V

= 9V

= 7V (U

G

= device voltage)

Types:

(Cable lengths, 1.5 meters)

Model Meas. Range

Frequency

Frequency

0 to 15000 Hz

0 to 3200,0 Hz

Pulses / Cycle

Speed

0 to 65000 Imp

8 to 32000 UpM

Option sensor supply 12 V

Resolution

1 Hz

0.1 Hz

(can, by inserting wire jumper, be switched to)

1 Imp

1 UpM

Order no.

ZA9909AK1U

ZA9909AK2U

ZA9909AK4U

OA9909V12

ALMEMO

®

Adapter Cable for Digital Input Signals

channel 1 channel 2 channel 3

Types:

(cable length, 1.5m each)

3 digital inputs, (optocoupler), for floating contacts, 5V auxiliary voltage led out

4 digital inputs, electrically isolated (optocoupler) for external voltage, 4 to 30 V

Order no.

ZA9000ES2

ZA9000EK2

02.13

Input connectors and adapter cables

ALMEMO

®

Universal Adapter Cable with Free Ends

Types:

Order no.

The ALMEMO

®

universal connector ZA 9000-FS is also available with connecting cable and free ends, as adapter cable

ZA9000AK. The sensor supply voltage is present on terminal U+; it is supplied by the ALMEMO

®

device (sensor supply voltage 5 V, can be stabilized on request). Connecting cable : 8-wire, 8 x 0.14 mm

2

, black, Length 1.5 m The wiring diagram and color code of the wires are consistent for all ALMEMO

®

sensors and cables, so that any pin configuration can be quickly and easily identified.

ZA9000AK

ALMEMO

®

10-Fold MU Connector for ALMEMO

®

Plug-In Boards with 64-Pin Spring Contact Strip

!

NOT suitable for sensors needing interface circuitry

(e.g. 26 V, AC voltage, mA, humidity sensors, rotating vanes, frequency, pulse, rotational speed) no sensor supply possible)

!

The current MU connector version, ZA5690MU, can only be used in conjunction with the new ALMEMO

®

5690 systems.

The old MU connector version, ZA5590MU, can of course be used in conjunction with the old ALMEMO

®

5590/5990 systems but is subject to certain restrictions with the current 5690 systems (e.g. only 1 measuring channel per input, no multi-point adjustment or connector linearization)

Types:

ALMEMO

®

10-fold connector (64-pin) with EEPROM sensor memory for connecting 10 sensors; on request pre-programmed to your specifications for Data acquisition systems ALMEMO

®

5690 (not for ALMEMO

®

5590 / 5990)

For Data acquisition systems ALMEMO

®

5590 und 5990

Order no.

ZA5690MU

ZA5590MU

02.14

Input connectors and adapter cables

ALMEMO

®

Connector Adapter Cable, Digital Input of Third Party Device to ALMEMO

®

Device Type ZA 1000A KSW / ZAD 919A Kxx

Existing equipment incorporating a digital interface can, thanks to our flexible ALMEMO

®

system, continue being used. For this purpose, we can offer you the following service :

1. We program a device type protocol for you, which matches the output interface of your device.

2. We fit the interface cable for your device with the matching ALMEMO

®

connector.

Description:

• Data acquisition from external devices with digital interface and integration in the data acquisition with ALMEMO

® devices.

• The digital connector of the adapter cable provides an electrically isolated serial interface and includes an interface processor for protocol conversion.

• Value-adding to existing measuring technology at a very interesting price-performance ratio.

Examples:

• Scales and weighing equipment

• Dial gauges and displacement transducers

• Multimeters

• Incremental displacement transducers

• Flue gas analysers

Types:

Order no.

For the purposes of programming the interface, please provide us with a detailed description of the output interface of the third-party device you want to have integrated, or a matching cable, or a connector including the pin configuration, plus the third-party device itself for the purposes of testing and checking.

Interface programming for the device type protocol of the device to be integrated

ALMEMO

®

connector adapter cable

ZA1000AKSW

ZAD919AK

02.15

02.16

ALMEMO

®

Output modules

Content

ALMEMO

®

trigger cable ZA1000ET/ZA1006EK2

ALMEMO

®

trigger / relay cable V6 Typ ZA1006EKG/ETG

ALMEMO

®

relay cable, V6, ZA 1006 GK and electrical socket relay adapter ZB2280RA

ALMEMO

®

analog output cable ZA1601RK

ALMEMO

®

relay trigger adapter, analog ZA8006-RTA3

ALMEMO

®

trigger output interface ES5690-RTA5

03.03

03.03

03.04

03.04

03.05

03.06

03.01

ALMEMO

®

Output modules

ALMEMO

®

Output modules

A modern measuring instrument must be able to communicate with its environment, i.e. transfer its measured data to peripheral equipment, execute commands from a computer, trigger alarm signals, and respond to switching pulses.

To cover all possibilities while also keeping the hardware needed to a minimum all necessary interfaces have been integrated in our ALMEMO

®

output connector. This concept allows the user - with one and the same ALMEMO

®

measuring instrument

- to choose freely from a wide variety of output interfaces to best suit the particular task in hand .

For the purposes of connecting the modules virtually all ALMEMO

®

devices are equipped with two output sockets

A1 and A2; these also allow the devices to participate in digital networking. The output modules, just like the sensors, are detected automatically; no extra programming is required.

Please note that many ALMEMO

® output modules can only be operated in conjunction with ALMEMO

®

devices version 6 and above (not 2390, 8390).

Labeled V6 (device fi rmware update may be needed).

Describing all the many options provided by the ALMEMO

® log.

system with output modules would be beyond the scope of this cata-

Please ask for our ALMEMO

® modules.

Manual. It will provide you with valuable tips and a detailed description of our ALMEMO

®

output

We shall of course be pleased to offer you competent advice and support to help you solve your particular measuring tasks.

Or you can arrange a date for a demonstration. Our experts will be pleased to visit you - to introduce and explain the numerous application options that the ALMEMO

®

system offers.

03.02

ALMEMO

®

trigger cable ZA 1000 ET / ZA 1006 EK2

ALMEMO

®

Output modules

Technical Data

Trigger input

ZA1000ET

ZA1006EK2

Trigger variants can be programmed with key

For external zero-potential contact

(not electrically isolated) and for external voltage 4 to 30 VDC (optocoupler), trigger variants can be programmed

Current consumption approx. 3 mA

Cable length

Connection

1.5 meters

(see variants)

Variants

ALMEMO

®

trigger cable, V5 / V6, with 1 key

ALMEMO

®

trigger cable, V5 / V6, with 1 trigger input for external voltage, with 2 banana plugs

ALMEMO

®

trigger cable, V6, with 2 trigger inputs for external contacts or voltages, with clamp connector

Order no.

ZA1000ET

ZA1000EK

ZA1006EK2

ALMEMO

®

trigger / relay cable V6 ZA 1006 EKG / ETG

Technical Data:

Trigger input For external zero-potential contact

(not electrically isolated) or for external voltage 4 to 30 VDC (optocoupler)

New

Trigger variants - can be programmed

(V6 only)

Relay

New

Can also be programmed as inverted

Normally open contact

(semiconductor relay)

(V6 only) Load capacity:

50 VDC, 0.5 A, 1 ohm

Current consumption approx. 3 mA

Cable length 1.5 meters

Connection Clamp connector

Variants

ALMEMO

®

trigger / relay cable, V6, with 2 trigger inputs (programmable trigger variant) for external voltages and 2 normally open contacts

ALMEMO

®

trigger / relay cable, V6, with 2 trigger inputs (programmable trigger variant) for external zero-potential contacts and 2 normally open contacts for ALMEMO

®

ALMEMO

®

devices, version V5

trigger / relay cable, V5, with 1 trigger input (Start / Stop only) for 1 external zero-potential contact or for voltage and 2 normally open contacts

Order no.

ZA1006EKG

ZA1006ETG

ZA1000EGK

03.03

ALMEMO

®

Output modules

ALMEMO

®

relay cable, V6, ZA 1006 GK and electrical socket relay adapter, ZB 2280 RA

Technical Data

Relay cable, V6, type ZA 1006 GK

Relay Normally open (semiconductor relay)

New

Can also be programmed as inverted

(V6 only)

Load capacity 50 VDC, 0.5 A, 1 ohm

Current consumption approx. 3 mA

Cable length 1.5 meters

Connection Banana plugr

Variants

Order no.

ALMEMO

®

relay cable, V6, with 1 normally open contact for ALMEMO

ALMEMO

®

®

devices, version V5

relay cable, V5, with 1 normally open contact

ZA1006GK

ZA1000GK

Technical Data

Relay adapter ZB2280RA

Control input

Output

Switching status for optocoupler output or switching contact R <10 kW

Electrical safety socket, mechanical relay, load capacity 250 V, 6 A

OFF idle; ON alarm

Variants

Order no.

Relay adapter for switching mains supplied devices combined with relay cable ZA1006GK/ZA1000GK

ZB2280RA

ALMEMO

®

analog output cable ZA 1601 RK

• Measured values can be recorded using a chart recorder or a similar output device.

• A signal converter is integrated in the connector.

• The device signal is converted into a voltage corresponding to the linearized measured value.

• To obtain a high response speed a conversion rate of 10 mops can be set in the ALMEMO

®

device.

• The output signal can be scaled as required..

Technical Data:

Output voltage -1.250 to 2 000 V, not electr. isolated

Gain

Load

0.1 mV / digit

>100 kW

Accuracy ±0.1% ± 6 digits

Temperature drift 1 digit / K

Time constant 100 ms

Current consumption approx. 3 mA

Cable length 1.5 meters

Variants

Analog output cable -1.250 to 2.000 V (0.1 mV / digit) not electrically isolated

03.04

Order no.

ZA1601RK

ALMEMO

®

Output modules

ALMEMO

®

relay trigger adapter, analog ZA 8006 RTA3 for connecting to ALMEMO

®

devices

menu selection all peripherals

• Universal trigger output interface for connecting to output sockets on ALMEMO

®

devices - from version V6 up (not 2390,

8390). device fi rmware update may be needed.

• Up to 10 peripheral elements (relays, trigger inputs, analog outputs) each with individually confi gurable function

• Relay functions, total alarm, assignment to particular limit values, or addressing via interface

• Integrated alarm signaling device can be assigned to all relay functions.

• Inverse relay addressing for alarm in the event of power failure

• Programmable messages to be issued when relays are activated

• Comprehensive trigger features with the aid of command macros, addressing via 2 keys or electrical signals

• Either 2 or 4 analog outputs (10 V or 20 mA) can be assigned to any measuring channels, scalable sub-areas, or alternatively addressing via interface.

• New:

Analog output type 10 V or 20 mA (programmable)

• All programming and peripheral states shown on illuminated graphics display relais messages

• Keypad for selecting menu and port trigger inputs analog outputs

• Watchdog function in the event of a failure of ALMEMO

® device or computer addressing

• Connection of peripherals via ALMEMO

®

clamp connectors, cable with anti-kink protective sleeve and strain relief

• Power supply via the ALMEMO output option a mains adapter may also be required.

• Modern, compact housing - also suitable for DIN top-hat rail mounting

®

device; in case of the analog

On request: ALMEMO

®

output interface ZA8006RTA4 for connection to the PC (directly or via network).

Technical Data

Trigger inputs

Relay

Analog outputs

0.0 to 10.0 V

0.0 to 20.0mA

Accuracy

Temperature drift

Time constant

Power supply

Optocoupler, 4 to 30 V, Ri >3 kohms

Semiconductor relay 50 V, 0.5 A, 1 ohms

10 V or 20 mA (programmable)

16-bit DAC, electrically isolated

0.5m V / digit, Load > 100 kohms

0.1 mA / digit, Load <500 ohms

0.1% of meas. val. +0.1% of final val.

10 ppm / K

100 µs via ALMEMO

®

device

Basic version 2 trigger inputs and 4 normally open relays

Options 2 additional relays (normally open) OA8006SH2

Per normally open pair 2 additional normally closed relays

(with normally open relays 2 changeover relays)

OA8006OH2

2 analog outputs (common ground), electrically isolated

10 V or 20 mA (programmable) OA8006R02 or mains adapter

(with 9V supply)

ZA1312NA7

(recommended for analog output option)

Current consumption approx. 10 mA, Lighting approx. 15 mA

2 analog outputs approx. 30 mA + 1.6 I

Out

Display Graphics 128 x 64 (55 x 30 mm)

Lighting 2 white LEDs

Keypad

Housing

7 silicone keys (4 soft-keys)

127 x 83 x 42 mm (LxWxH)

ABS (maximum 70°C), 290 g

Possible combinations

1x OA8006SH2 (+2 relays) or 1x OA8006SH2 (+2 relays) + 1x OA8006R02

(+2 analog outputs) or 2 x OA800R0H2 (+4 analog outputs)

Accessories

Mains unit, 12 V, 1 A

DIN tophat rail mounting

ZA1312NA7

ZB2490HS

Variants

ALMEMO

®

relay trigger adapter with 2 trigger inputs, 4 normally open relays, DC socket, graphics display, and keypad, including 1.5-meter ALMEMO

®

connecting cable and 3 ALMEMO

®

clamp connectors

Order no.

ZA8006RTA3

03.05

ALMEMO

®

Output modules

ALMEMO

®

trigger output interface, ES 5690 RTA5, for ALMEMO

®

data acquisition systems

Technical Data:

Trigger inputs Optocoupler 4 to 30 V, Ri > 3 kohms

Relays Semiconductor relays 50 V, 0.5 A, 1 ohm

Analog outputs

0.0...10.0 V

0.0...20.0 mA

10 V or 20 mA (programmed)

16 bit DAC. electrically isolated

0.5 mV/Digit. Load > 100 kohms

0.1 mA/Digit. Load < 500 ohms

Accuracy 0.1% of meas. val. +0.1% of final val.

Temperature drift 10 ppm/K

Time constant 100 µs

Power supply

Current consumption Standard: approx. 10 to 20 mA via ALMEMO

®

measuring system

2 analog outputs: approx. 15 mA + 1.8·IOut

Module 19“ 8-DU (2 slots)

• Universal trigger output interface for ALMEMO

®

5690 data acquisition systems

• System (master measuring circuit or CPU module) addressed via an internal SPI bus

• Up to 10 peripheral elements (relays, trigger inputs, analog outputs) each with individually configurable function

• Relay functions, total alarm, assignment to particular limit values, or addressing via interface

• Inverse relay addressing for alarm in the event of power failure

• Relay states shown via LEDs

• Watchdog function in the event of a failure of ALMEMO

® device or computer addressing

• Comprehensive trigger features with the aid of command macros, addressing via electrical signals

• Either 2 or 4 analog outputs (10 V or 20 mA programmable) can be assigned to any measuring channels, scalable subareas, or alternatively addressing via interface.

• On request : 10 analog outputs per plug-in module (without trigger inputs, without relays)

• Connection of peripherals via ALMEMO

• Power supply via ALMEMO

®

system.

®

clamp connectors, cable with anti-kink protective sleeve and strain relief

ALMEMO

®

clamp connector

Basic version 2 trigger inputs and 4 normally open relays

Options 2 additional relays (normally open)

OA8006SH2

Per normally open pair 2 additional normally closed relays

(with normally open relays 2 changeover relays)

OA8006OH2

2 analog outputs (common ground), electrically isolated

10 V or 20 mA (programmable) OA8006R02

Possible combinations

2x OA8006SH2 (+4 relays) or 1x OA8006SH2 (+2 relays) + 1x OA8006R02

(+2 analog outputs) or 2 x OA8006R02 (+4 analog outputs)

Variants

ALMEMO

®

relay trigger module - with 2 trigger inputs,

4 normally open relays, and 3 ALMEMO

®

clamp connectors

03.06

Order no.

ES5690RTA5

ALMEMO

®

Network technology

Content

ALMEMO

®

networking technology

ALMEMO

®

PC connection using USB data cable ZA 1919 DKU

ALMEMO

®

PC connection using Ethernet data cable ZA1945-DK

Data cable for digital ALMEMO

®

D6 probe

ALMEMO

®

Network Interface Cables

ALMEMO

®

Network Interface Cables with Fiber Optics

Wireless data links using ALMEMO

®

Bluetooth modules

Wireless PC link with Bluetooth

Bluetooth USB CPU module ZA 1719 BCU

Wireless PC link with Bluetooth

Bluetooth device CPU ZA2719BC

Wireless device connection with Bluetooth

Bluetooth measuring instrument ALMEMO

®

2790 with integrated Bluetooth slave

Wireless sensor connection with Bluetooth

Sensor connection with Bluetooth sensor measuring instrument

RS422 network distributor ZA 5099 NVL connection via optic fiber

Ethernet network driver ZA 5045 AK

RS422-network distributor ZA 5099 NVB device connection via screw terminals

Data connection via GPRS

GPRS mobile communications modem ZA 1709 GPRS

GPRS connection 1+1

GPRS connection 1+3

GPRS connection 3+1

GPRS connection 3+3

04.07

04.08

04.09

04.10

04.11

04.12

04.12

04.02

04.05

04.05

04.05

04.06

04.06

04.13

04.14

04.15

04.16

04.17

04.18

04.18

04.19

04.19

04.01

ALMEMO

®

Network technology

ALMEMO

®

networking technology

The ALMEMO

®

system provides optimal support for networked, decentralized measured data acquisition. Measured data can be acquired locally on site using short sensor signal lines and small modular measuring instruments and can then be evaluated all together on a central computer. This not only minimizes wiring requirements but also goes a long way to solving EMC problems (especially if optic fi ber cables are used).

Via the cascadable interface provided by ALMEMO

®

devices it is possible, thanks to our ALMEMO

®

networking technology, to manage up to 100

ALMEMO

®

measuring instruments from just one computer. User-friendly software packages (see Chapter 05) are available for automatically scanning measuring points within the network, for evaluating the measured values, and for graphically representing results in line chart or bar chart form. This permits measuring setups in which devices can be used with such high operational reliability and with such great fl exibility that even the most demanding measuring tasks can be solved.

For example:

• Data connection from the PC to

ALMEMO

®

devices via USB, Ethernet,

RS232, RS422, Bluetooth, GPRS mobile communications, modem.

• Can be combined in a wide variety of ways via the output sockets A1 and

A2 on the ALMEMO

®

measuring instrument

• Various networking arrangements can be implemented.

• Measuring instruments can be installed in separate rooms; considerable distances can be bridged.

• ALMEMO

®

devices / networks can be connected to the PC via an existing

Ethernet network.

• New

PC and devices can be connected over a wireless link using Bluetooth modules.

• Measured data can be acquired and also read out from the measured value memory on an ALMEMO

®

data logger

- all online - using the WinControl software package

USB

(up to 5 meters )

PC connection via USB

(over a wireless Bluetooth link, see page 04.03)

Inexpensive for relatively short distances (up to

5 m) several connections in parallel (star-configured network) for mobile use, e.g. notebook

Necessary component

ZA 1919 DKU see page 04.05

04.02

Ethernet

ALMEMO

®

Network technology

RS232 up to 15 meters with optic fiber up to 50 meters

ALMEMO

®

network cable (up to 50 meters)

PC connection via Ethernet

(over a wireless Bluetooth link, see page 04.03)

Measured data acquisition, on a decentralized basis, using existing LAN cabling (bus networking), relatively long distances, via Internet worldwide.

Necessary component(s)

ZA 1945 DK see page 04.05

PC connection via RS232

(over a wireless Bluetooth link, see page 04.03)

Single connection via COM interface (also USB with converter), up to 15 meters, and with optic fiber up to 50 meters

Necessary component

ZA 1909 DK5 see page 04.05

Connection between ALMEMO

®

measuring instruments over ALMEMO

®

network cable

(over a wireless Bluetooth link, see page 04.03)

Inexpensive linear network solution, flexible, plug-and-play, easy to expand.

Necessary component ZA 1999 NK5 see page 04.06

USB

Bluetooth

(up to 300 meters)

Wireless Bluetooth link PC - USB

Inexpensive USB for mobile applications expandable for up to 7 ALMEMO

®

measuring instruments in parallel (star-configured network).

Necessary components

ZA1719BPVU,

ZA1719BT1XS or Bluetooth meas. instrument

MA2790BT1XS see page 04.08

Bluetooth

(up to 300 meters)

Wireless PC link with Bluetooth

Highly flexible irrespective of location expandable for up to 7 ALMEMO

®

measuring instruments in parallel (star-configured network) display and configuration of (multiple) connections via Bluetooth device CPU.

Necessary components

ZA2719BPVU or ZA2719BPVN

ZA1719BT1XS or Bluetooth meas. instrument

MA2790BT1XS see page 04.09

04.03

ALMEMO

®

Network technology

Wireless Bluetooth link between ALMEMO

® measuring instruments

For mobile networking highly flexible network topology (linear / star-configured network)all connections expandable for up to 7 ALMEMO

® measuring instruments in parallel.

Necessary components

ZA1719BNV,

ZA1719BT1XS or Bluetooth meas. instrument

MA2790BT1XS see page 04.10

Bluetooth

(up to 300 meters)

Bluetooth

(up to 300 meters)

Wireless sensor connection via Bluetooth (AL-

MEMO

®

wireless sensor)

Single connection from a measuring ALMEMO

® device (wireless sensor) to a receiving ALMEMO

® device with display and saving of measured values

(also without PC). Any number of sensor connections in parallel.

Necessary components

ZA1719BT1XFV or ZA2790BT1XFV

(with Bluetooth measuring instrument) see page 04.12

Connection between ALMEMO

®

measuring instruments over ALMEMO

®

RS422 network

(over a wireless Bluetooth link, see page 04.03)

Fixed installation, measured data acquisition on a decentralized basis, linear / star-configured network, relatively long distances, good resistance to radio interference affecting transmission.

Necessary components

ZA 5099 NTL or ZA 5045 AK

ZA 5099 NVL see pages 04.13 to 04.15

RS422 (up to max. 1km)

Modem Modem

Telephone network

GPRS-Modem

Internet Mobile

internet

PC connection via fixed-line telephone network

Fixed installation any distance, worldwide.

Necessary components on request

PC connection via GPRS mobile modem

Mobile operation over any distance.

Necessary components:

ZA 1709 GPRS see page 04.16 to 04.19

04.04

ALMEMO

®

Network technology

ALMEMO

®

PC connection using USB data cable ZA 1919 DKU

RS232 data cable, type ZA 1909 DK5, USB adapter cable ZB 1909 USB

• ALMEMO

®

-USB data cable for data connection between an

ALMEMO

®

device and a PC with a USB interface

• ALMEMO

®

RS232 data cable with a DSUB socket for data connection between an ALMEMO® device and a PC with a

COM interface

• ALMEMO

®

optic fiber cable (RS232 or with adapter to USB) for absolute electrical isolation and extensive protection against lightning.

Types:

USB data cable, electrically isolated, maximum 115.2 kbaud, cable length 1.5 meters, including CD with Windows driver

As above but cable length 5 meters

RS232 data cable electrically isolated, max. 115.2 kbaud,

Current consumption : approx. 1 mA, Cable length : 1.5 m

As above, but cable lengths 5m / 10m / 15m

RS232 data cable with optic fiber, max. 115.2 kbaud, Cable length 1,5 m

Longer optic fiber (up to 50 m) for interiors, Duplex plastic 2.2 x 4.3mm, per meter

Converter, USB to RS232, 9-pin DSUB for ALMEMO

®

data cable ZA1909DKx, including WINDOWS driver

Order no.

ZA1919DKU

ZA1919DKU-05

ZA1909DK5

ZA1909DK5-05 /-10 /-15

ZA1909DKL

LL2243L

ZB1909USB

ALMEMO

®

PC connection using Ethernet data cable ZA 1945 DK

Technical data

Ethernet:

ALMEMO

®

Socket RJ45 (10/100 base-T)

Automatic switchover 10 / 100 MHz

ALMEMO

®

connector for socket A1

Baud rate standard 9600 bd, max. 115.2 kbd

(can be changed via Device-Installer and browser)

• For connecting almost any ALMEMO

®

measuring instrument to an Ethernet PC network.

• Linking up to the Internet now possible.

• Terminal operation using our AMR-Control software, available free-of-charge.

• Device-Installer configuration software also included on the

AMR CD.

• Measured data acquisition via several Ethernet modules using our Win-Control software. (Version SW5600WC2 and above, see chapter Software).

Power supply 12 V DC via measuring instrument

(suitable mains supply unit recommended)

Current consumption <60 mA (10 MHz), <90 mA (100

MHz)

Accessories

Patch cable RJ45, plug / plug, 2 meters

Type

Ethernet data cable, RJ45 socket on ALMEMO

®

connector, cable length 1.5 meters

Order no.

ZB1904PK2

Order no.

ZA1945DK

Data cable for digital ALMEMO

®

D6 sensors

Types

ALMEMO

®

USB adapter cable length 1.5 meters for connecting an ALMEMO

®

D6 sensor directly to the USB port on a PC (power supply via USB)

ALMEMO

®

Ethernet adapter cable total length 3 meters for connecting an ALMEMO

®

D6 sensor directly to an Ethernet PC network, including mains unit

Order no.

ZA1919AKUV

ZA5045AKFBV

04.05

ALMEMO

®

Network technology

ALMEMO

®

Network Interface Cables ZA 1999 NK5

!

The device network will be blocked if the measuring instrument fails to operate.

No further peripheral devices can be connected

(analog output, alarm relay etc.)

Uses:

• Especially suitable for short distances and mobile measuring setups.

• Up to 100 ALMEMO

®

measuring instruments can be networked.

Advantages:

• Devices can be quickly and easily interconnected and networked.

• Low power consumption (approx. 1 mA) without additional power supply.

• You can easily assemble the network cable yourself, up to

50m in length, using just two single network connectors

ZA1999FS5 (a couple) and one four-wire cable.

Types

Network cable for cascading several devices for baud rates up to 57.6 kbaud current loop, electrically isolated, 1.5 m long

As above, but cable lengths 5m / 10m / 15m

2 Network connectors (a couple) with screw terminals for local self-assembly

Order no.

ZA1999NK5

ZA1999NK5 -05/ -10 / -15/ -xx

ZA1999FS5

ALMEMO

®

Network Interface Cables with Fiber Optics ZA 1999 NKL

!

The device network will be blocked if the measuring instrument fails to operate.

No further peripheral devices can be connected

(analog output, alarm relay etc.)

Uses:

• Especially suitable for safe and reliable data transmission in industrial environments with high levels of interference.

• Up to 10 ALMEMO

®

measuring instruments can be networked (at 9600 baud, double this number, if the transmission rate is halved).

Advantages:

• Devices can be quickly and easily interconnected and networked.

• No EMC problems, highest possible immunity to interference, absolute electrical isolation of the instruments - even under high voltages.

• No additional voltage supply is required.

• You can easily assemble the network cable with plastic optic fiber yourself, up to 50m in length, using just two single network connectors ZA1999FSL, without needing any special tools.

Types

Network cable with optic fiber for cascading several devices 1.5 m long for baud rates up to 57.6 kbaud

As above, but cable lengths 5m / 10m / 15m

Longer optic fiber cable for interiors, Duplex plastic 2.2 x 4.3 mm

Network connector with optic fiber converter for local self assembly

Order no.

ZA1999NKL

ZA1999NKL -05/ -10 / -15/ -xx

LL2243L

(please specify length L)

ZA1999FSL

04.06

ALMEMO

®

Network technology

Wireless data links using ALMEMO

®

Bluetooth modules

Various types of connection are possible

Wireless PC connection see page 04.08/04.09

Wireless connection from a PC with ALMEMO® Bluetooth

CPU to up to 7 ALMEMO

®

measuring instruments each with

Bluetooth slave

Wireless device connection see page 04.10

Wireless connection from an ALMEMO

®

measuring instrument with Bluetooth CPU to up to 7 ALMEMO® measuring instruments each with Bluetooth slave

Wireless sensor connection see page 04.12

Wireless sensor connection from a measuring ALMEMO

®

device with Bluetooth slave to the measuring input on a receiving

ALMEMO

®

device with Bluetooth sensor module.

Up to 4 measuring channels can be transmitted per connection..

Common technical data

Class 1 with active antenna

Protocol

Operating range

ALMEMO

®

data rate

Module housing

SPP (sequence packet protocol)

(128-bit encryption)

300 meters (free field)*

1200 baud up to 115.2 kbaud

(LxWxH) 61 x 30 x 12 mm

Polystyrene (-10 to +70 °C)

Cable length for plug-in module with option OA1719BK

Length = 1 meter

*

Inside a building the operating range of the wireless link will

be substantially lower.

Advantages of ALMEMO

®

connections using Bluetooth compared with other wireless technologies

• Bluetooth wireless technology is industrial standard in compliance with IEEE 802.15.1; it ensures high transmission reliability.

• The frequency hopping procedure used ensures robustness against interference. The Bluetooth partners move continually to and from among the 79 wireless channels available.

• Any number of Bluetooth connections can operate in parallel with complete reliability.

• The multi-digit PIN code ensures that all Bluetooth participants are identified reliably and unequivocally.

• These links - once configured - will, as soon as the device is switched ON, be automatically setup - and, in the event of interruption, be automatically restored.

• One Bluetooth CPU supports up to 7 parallel connections to

Bluetooth slaves.

• These powerful new Bluetooth class 1 wireless modules incorporate an integrated active antenna ensuring an especially wide operating range (up to 300 meters free field); there is no need for an extra antenna.

Common technical features

• Bluetooth links are supplied already paired, i.e. simply plug in and start measuring.

• In the event of interruption to the Bluetooth connection the

USB / COM interface in the PC remains available for the software being used. For continuous monitoring purposes this ensures very high transmission reliability.

Advisory note : The Bluetooth links integrated in some laptops / PCs cannot be used for these purposes because in the event of interruption the operating system deactivates the

COM interface and this must then be reactivated manually each time.

• Any ALMEMO

®

measuring instrument with a Bluetooth slave module connected can be used.

• Using the Bluetooth CPU on the PC or a plug-in Bluetooth

CPU module on the ALMEMO

®

measuring instrument up to

7 measuring instruments with Bluetooth slave modules can participate in a star-configured network. Compared with paired single connections star-configured networking saves on additional master modules.

• To extend the operating range or raise the number of parallel connections further CPUs can be cascaded as repeaters or routers (increasing the switchover times for device scanning in the WinControl software).

• The plug-in module variant with a 1-meter cable can, in order to optimize the wireless link, be positioned away from the measuring instrument between the ALMEMO® connector and the module (option OA1719BK) and specially aligned (using

Velcro fastener).

• All (multiple) connections can be configured end-to-end quickly and easily either with the AMR-Control software or on the Bluetooth device CPU via the display and keypad.

• To search through and select from all the available Bluetooth slave partners the user simply enters the appropriate PIN codes. The Bluetooth device CPU can also be configured fully automatically by simply plugging in the slave module;

(pairing is performed automatically in an exchange of PIN codes and hardware addresses).

04.07

ALMEMO

®

Network technology

Wireless PC link with Bluetooth

Bluetooth USB CPU module ZA 1719 BCU

Wireless connection from a PC with ALMEMO

®

Bluetooth CPU to up to 7 ALMEMO

®

measuring instruments with Bluetooth slave

Technical data

Common technical data see page 04.07

Cable

Voltage supply

ZA1719BCU

ZA1719BT1XS

ZA1719BCU Length = 1.5 meters via USB interface on the PC via ALMEMO

®

measuring instrument, approx. 35 mA (9 V)

ZA 1719 BCU ZA 1719 BT1XS

• Connection of the CPU module to the USB interface on a PC

• Connection of the plug-in slave module to socket A1 on an

ALMEMO

®

device

Option for plug-in module ZA1719BT1XS

Cable between ALMEMO

®

connector and module Length = 1 meter

Order no.

OA1719BK

Variants

Order no.

Paired wireless PC connection (USB) for 1 ALMEMO

®

measuring instrument (configured and ready-to-operate)

Bluetooth CPU module with USB (ZA1719BCU) and plug-in Bluetooth slave module (ZA1719BT1XS)

Paired connection with Bluetooth measuring instrument ALMEMO

®

2790 see page 04.11

ZA1719BPVU

Extension for multiple connections

Plug-in Bluetooth slave module for 1 ALMEMO

®

device

Bluetooth measuring instrument ALMEMO

®

2790 see page 04.11

ZA1719BT1XS

04.08

ALMEMO

®

Network technology

Wireless PC link with Bluetooth

Bluetooth device CPU ZA2719BC

Wireless connection from a PC with ALMEMO

®

Bluetooth CPU to up to 7 ALMEMO

®

measuring instruments with Bluetooth slave.

Technical features of the device CPU

Common technical features see page 04.07

• Modern, compact housing - also suitable for DIN top-hat rail mounting

• Graphic display - shows status of connections - can be illuminated

• All (multiple) connections can be configured end-to-end using the display and keypad.

• Use as repeater

This extends the operating range or raises the number of parallel connections. An ALMEMO

®

Bluetooth slave module is connected to socket A1 on the CPU. Power is supplied via a mains unit.

ZA 2719 BC

ZA 1719 BT1XS

• Connection of the device CPU to the USB interface on a PC

Or, alternatively, an ALMEMO

®

data cable can be used

(Ethernet, RS232, or RS422).

• Connection of the plug-in slave module to socket A1 on an

ALMEMO

®

device

Display

Technical data

Common technical data see page 04.07

ALMEMO

®

Bluetooth device CPU ZA 2719 BC

Display Graphics display 128x64 (55x30mm)

Illumination

Keypad

Housing

2 white LEDs

7 silicone keys (of which 4 softkeys)

(LxWxH) 127 x 83 x 42 mm

ABS (-10 to +70 °C) 290 g

Voltage supply

ZA2719BC with USB data cable ZA1919DKU5 via USB interface on the PC or with connector mains unit 12V 1A

ZA1312NA7 or battery set (3 AA cells), approx. 40 mA (5 V)

ZA1719BT1XS via ALMEMO

®

measuring instrument with illumination approx. 70 mA (5 V) approx. 35 mA

(9 V)

Selection menu main menu

Accessories for device CPU ZA2719BC:

Fixture for DIN rail mounting

Rubberized impact protection

ZB2490HS

ZB2490GS2 device configuration connecting menu

Option for plug-in module ZA1719BT1XS:

Cable between ALMEMO

®

connector and module

Length = 1 meter OA1719BK

Variants

Paired wireless PC connection (USB) for 1 ALMEMO

®

Order no.

measuring instrument (configured and ready-to-operate) comprising :

Bluetooth device CPU (ZA2719BC) including USB cable ZA1919DKU5 and plug-in Bluetooth slave module (ZA1719BT1XS)

ZA2719BPVU

Paired wireless PC connection for 1 ALMEMO® measuring instrument (configured and ready-to-operate) comprising :

Bluetooth device CPU (ZA2719BC) including connector mains unit ZA1312NA7 (without data cable) and plug-in Bluetooth slave module (ZA1719BT1XS)

ALMEMO

®

RS232 data cable

ZA2719BPVN

ZA1909DK5

ZA1945DK

ALMEMO

®

Ethernet data cable

Paired connection with Bluetooth measuring instrument ALMEMO

®

2790 see page 04.11

Extension for multiple connections

Plug-in Bluetooth slave module for 1 ALMEMO

®

device

ZA1719BT1XS

Bluetooth measuring instrument ALMEMO

®

2790 see page 04.11

04.09

ALMEMO

®

Network technology

Wireless device connection with Bluetooth

Wireless connection from an ALMEMO

®

measuring instrument with Bluetooth CPU to up to 7 ALMEMO

®

measuring instruments with Bluetooth slave.

Technical data

Common technical data see page 04.07

Voltage supply

ZA1719BC

ZA1719BT1XS via ALMEMO

®

measuring instrument, approx. 20 mA (9 V) via ALMEMO

®

measuring instrument, approx. 35 mA (9 V)

ZA 1719 BC ZA 1719 BT1XS

• Connection of the plug-in CPU module to socket A2 on an

ALMEMO

®

device

• Connection of the plug-in slave module to socket A1 on a second ALMEMO

®

device

Option for plug-in module ZA1719BT1XS:

Cable between ALMEMO

®

connector and module Length = 1 meter

Variants

Paired wireless device connection (configured and ready-to-operate) between 2 ALMEMO

®

measuring instruments comprising:

Plug-in Bluetooth CPU module (ZA1719BC) and plug-in Bluetooth slave module (ZA1719BT1XS)

Paired connection with Bluetooth measuring instrument ALMEMO

®

2790 see page 04.11

Extension for multiple connections:

Plug-in Bluetooth slave module for 1 ALMEMO

®

device

Bluetooth measuring instrument ALMEMO

®

2790 see page 04.11

04.10

Order no.

OA1719BK

Order no.

ZA1719BNV

ZA1719BT1XS

ALMEMO

®

Network technology

RS422 network distributor ZA 5099 NVL

RS232 / RS422 network driver ZA 5099 NTL, Device / PC connection via optic fiber

Uses:

• Standard solution for stationary measuring setups in industrial environments.

• Suitable for relatively long distances, up to 1 km.

• Up to 100 ALMEMO

®

measuring instruments can be networked.

Advantages:

• Absolute electrical isolation of connected instruments - even under high voltages.

• Common mode interference on the transmission line is largely suppressed.

• Trouble-free implementation of branches and stub lines, directly inter-connectable, also as RS485 bus master.

• Easy to install - using a surface-mount housing, fastening brackets, and a screw terminal connector.

• Further peripheral devices can be connected to the AL-

MEMO

®

device, (analog output, alarm relays, etc.).

!

Technical Data:

Connection :

ZA5099NVL: 3 x RS422, 4-wire, via terminal connector

1 x optic fiber cable, 1.5 m long via

ALMEMO

®

connector to ALMEMO

®

device

ZA5099NTL: 2 x RS422, 4-wire, via terminal connector

1 x RS232 optic fiber cable, 1.5 m long via 9-pin sub-D to the PC

Wiring arrangements: RS422, 4-wire plus voltage supply, 2-wire data line,

(2 x 2 wires, duplicated)stranded in pairs

Max. line length: between two RS422 distributors 1 km optic fiber cable to the ALMEMO

® device or PC, 50 m

Power supply: 10 to 12 V DC, via terminal connector

Current consumption: approx. 10 to 18 mA

Dimensions: L 71,5/90 x W 61,5/95 x H 30 mm

The distributor is supplied via the RS422 network or via its own mains power unit. The network remains functional - even when the ALMEMO® device is switched off or disconnected.

ALMEMO device

Power supply

Signal from controller

Signal from meas. instr.

Signal from all devices green yellow red red

Network Distr. Fib. Opt.

fiber optics

Network Driver Fib. Optics

Transmitter

Receiver fiber optics mains adapter

DSUB

9-pin

Receiver

Transmitter

Types

RS422 network distributor, ALMEMO, device connection via optic fiber (length = 1.5 m),

Power supply via the mains supply unit

RS232 / RS422 network driver ZA5099NTL, computer connection via optic fiber (length = 1.5 m)

Power supply via the mains supply unit

Mains supply unit, 12 V DC / 2.5 A

Cable housing for ZA5099NVx (1 set = 3 pieces)

Data line 4 x 2 wires, stranded in pairs, per meter (power supply, 2 x 2 wires, duplicated)

Order no.

ZA 5099 NVL

ZA 5099 NTL

ZB1012NA9

ZB5099KG

LD0042

04.13

ALMEMO

®

Network technology

Ethernet network driver ZA 5045 AK

Technical data

Ethernet:

RS422

Baud rate

Power supply

Socket RJ45 (10/100 base-T) automatic switchover 10 / 100 MHz

6-pin screw terminal connector, 4-wire TX+,

TX-, RX+, RX- and supply +12 V, -12 V; line length between driver and distributor, maximum 1 kilometer maximum 115.2 kbaud

9 to 12 V DC, <60 mA (10 MHz),

<90 mA (100 MHz)

• Connection of all ALMEMO

®

networks to an Ethernet PC network.

• Linking up to the Internet now possible.

• Terminal operation using our AMR-Control software, available free-of-charge.

• Configuration software XPort/Device-Installer is also included on the AMR CD.

• Measured data acquisition using our WinControl software

(see Chapter Software).

• Extension between driver and network distributor up to

1 kilometer now possible.

• Can also be used as RS485 bus driver.

• The driver in conjunction with network distributor ZA5099-

NVL replaces previous Ethernet network distributor ZA5099-

NVE.

Types

Ethernet network driver, RJ45 to RS422, 4-wire

Mains adapter, 12 V DC, 2.5 A, with free ends, also for supplying other network distributors via the bus

Patch cable RJ45, plug / plug, 2 meters

Optic fiber network distributor RS422 to ALMEMO

®

optic fiber and 2 x RS422

Data line 4 x 2 wires, stranded in pairs, per meter (power supply, 2 x 2 wires, duplicated)

04.14

Order no.

ZA5045AK

ZB1012NA9

ZB1904PK2

ZA5099NVL

LD0042

ALMEMO

®

Network technology

RS422 network distributor ZA 5099 NVB

RS232 / RS422 network driver ZA 5099 AS, device connection via screw terminals

Uses:

• Especially suitable for relatively long distances, up 1 km, and for stationary measuring setups.

• Up to 100 ALMEMO

®

measuring instruments can be networked.

Advantages:

• Common mode interference on the transmission line is largely suppressed.

• Trouble-free implementation of branches and stub lines, directly inter-connectable, also as RS485 bus master.

• Easy to install - using a surface-mount housing, fastening brackets, and a screw terminal connector.

• Further peripheral devices can be connected to the

ALMEMO

®

device, (analog output, alarm relays, etc.).

!

Technical Data:

Connection :

ZA5099NVB : 3 x RS422, 4-wire, via terminal connector

1 x cable, 1.5 m, via ALMEMO connector to the ALMEMO device

ZA5099AS 1 x RS422, 4-wire, via terminal connector

1 x RS232, via 9-pin sub-D, to the PC

Wiring arrangements : RS422, 4-wire data line, stranded in pairs

Max. line length : between two RS422 distributors, 1 km

Power supply :

ZA5099NVB : via ALMEMO device (standard)

ZA5099AS No external supply necessary

Current consumption : approx. 25 to 35 mA

Dimensions :

ZA5099NVB L 71,5/90 x W 61,5/95 x H 30 mm

ZA5099AS L 50 x W 33 x H 16 mm

The power for the distributor is, as standard, supplied via the ALMEMO

®

device. The network is only functional when the ALMEMO

®

device is switched on. Alternatively, the power for the distributor can be supplied via the RS422 network or via its own mains power unit.

Power supply

Signal from controller

Signal from meas. instr.

Signal from all devices green yellow red red

Network Driver

mains adapter

Distributor

RS422/485 screw terminal

4-pin

ALMEMO device

Types

RS422 network distributor, ALMEMO device connection via cable (length = 1.5 m),

Supply via ALMEMO device or via network (selectable by jumpers)

RS232 / RS422 network driver, can be connected directly to the computer

Mains supply unit, 12 V DC / 2.5 A

Cable housing for ZA5099NVx (1 set = 3 pieces)

Data line 4 x 2 wires, stranded in pairs, per meter (power supply, 2 x 2 wires, duplicated)

Order no.

ZA5099NVB

ZA 5099 AS

ZB1012NA9

ZB5099KG

LD0042

04.15

ALMEMO

®

Network technology

Mobile Internet and terms such as UMTS (universal mobile telecommunications system) and GPRS (general packet radio service) are on everyone‘s lips. Our solutions access ALMEMO

®

measuring technology over a mobile Internet connection (GPRS). It makes no difference whether our measuring technology is being used on a mobile basis or is installed in the remote locations. Operation may involve measuring instruments all over the world but it will be as though they were set up right next to your computer.

Mobile communication via GSM

:

The meas. instrument is accessed via the telephone network.

Mobile Internet via GPRS

:

The measuring instrument is accessed via the Internet.

:

Connection setup is controlled by schedule and the measuring instrument memory is read out automatically.

:

Connection setup is controlled by schedule and the measuring instrument memory is read out automatically.

-

Given the costs structure communication with the measuring instrument will be limited basically to reading out from the measuring instrument memory at fairly infrequent intervals.

:

The measuring instrument is connected with the computer online. The measuring instrument on site can save measured values and simultaneously these can be read out at regular, frequent intervals.

:

No additional computer hardware is required.

-

An additional modem is required at the computer end.

-

Connection is set up via a conventional telephone line and for a limited period of time.

:

The measuring instrument connects to your network automatically and is then available continuously.

-

It is not possible to scan multiple devices simultaneously because the number of telephone lines / modems is limited.

-

Charges are calculated according to connection duration.

:

:

Measured data can be acquired simultaneously from an unlimited number of devices.

Connection charges are calculated on a real utilization basis, i.e. according to the volume of data transmitted.

04.16

ALMEMO

®

Network technology

GPRS mobile communications modem ZA 1709 GPRS

• Remote interrogation and remote control of ALMEMO

® devices

• Ideal for measuring operations at remote sites

• Automatic memory readout or inexpensive 24-hour online measuring - thanks to a charges structure according to actual data usage.

Technical data

Frequency range Quad band 850 / 900 / 1800 / 1900 MHz

Output power

Connections

Power supply

2 W for EGSM 850 / 900

1 W for GSM 1800 / 1900

RS-232

(9600 baud, 9-contact. sub-D socket)

FME antenna connection (male)

Power supply, SIM card reader

8 to 30 V, via mains unit, included in delivery

Current consumption 30 mA at 12 V (basic consumption) maximum 190 mA at 12 V (sending)

Operating temp.

Dimensions

-30 to +65 °C (mains unit 0 to +40 °C)

65 x 74 x 33 mm

Weight

Mains unit approx. 110 g

Input voltage 110 to 240 VAC

Output voltage 10.5 to 13.5 VDC

Operating temperature 0 to +40 °C

!

Advisory note

For technical reasons a special data tariff and a VPN access are required; these can be arranged via „akrobit software GmbH“. Akrobit software GmbH offers various tariffs for VPN and mobile communications; depending on the tariff chosen, the GPRS modem can be used within Germany, within Europe, or worldwide.

A VPN client software must also be installed on the computer used for evaluation. The VPN client software is included in delivery free-of-charge.

For automatic memory readout the software AMR Win-

Control is required together with additional module

„Automatic ALMEMO

®

memory readout“

SW5600WCZM9.

Accessories

Additional protocol „Automatic memory readout“ for WinControl (SW5600WC1/2/3/4)

Order no.

SW5600WCZM9

Variants Order no.

GPRS mobile communications modem for connecting to AL-

MEMO

®

devices, including data cable ZA1909DK5, adapter

ZA1709AS, mains unit, documentation, antenna with magnetic base Cable approx. 2.5 meters.

ZA1709GPRS

Other variants are available on reques:

GPRS modem for texting SMS, with digital inputs, alarmdriven by the ALMEMO

®

device.

GPRS connections and cost accounting - examples

Advisory note

These cases are provided as examples only; the number of VPN1 accesses is for illustration purposes and can be modified as required. However, at least two accesses are always required (1x PC + 1x modem).

The software AMR WinControl can, depending on requirements, normally be used. The modem option is not always necessary; however, if several modems / devices need to be addressed simultaneously, at least one WC2 (standard) will be required. For device-internal data recording (especially with a memory card) we strongly recommend the optional software module for automatic memory readout (SW5600WCZM9).

The costs incurred as per the mobile communications tariff Vodafone and by Telekom Deutschland depend on actual data usage. All data tariffs permitting use of an alternative APN

2

are supported. Prepaid cards are not supported. A suitable tariff can be arranged by akrobit software GmbH. Customers preferring to make their own arrangements must have set up the mobile communications contract before ordering the modem. Rental solutions for modem, VPN, and mobile communications accesses, and provision of a test access by akrobit software GmbH are all available on demand.

04.17

ALMEMO

®

Network technology

GPRS connection 1+1

Installation of the VPN software on one computer for the purpose of addressing one modem with one or more connected devices

Required

GPRS modem

Data cable ZA1909-DK5

Modem adapter

Costs (net)

see price list included included

Note

Preconfigured with RS-232 connection

RS-232

VPN access

Automatic memory readout

(option) approx. EUR 15 / month 1x mobile + 1x PC (minimum contractual term 12 months)

Mobile communications card approx. EUR 14 / month m2m 5 MB included (minimum contractual term 24 months)

3

AMR WinControl software see price list SW5600WC1 (Light version for 1 device, 20 meas. channels) see price list SW5600WCZM9

GPRS connection 1+3

Installation of the VPN software on several computers for the purpose of addressing one modem with one or more connected devices. Each such computer is allocated a separate access with its own unique IP address; however, only one such computer can establish a connection to the modem at any one time.

Required

GPRS modem

Data cable ZA1909-DK5

Modem adapter

VPN access

Costs (net)

see price list included

Note

Preconfigured with RS-232 connection

RS-232 included approx. EUR 30 / month 1x mobile + 3x PC (minimum contractual term 12 months)

Mobile communications card approx. EUR 14 / month m2m 5 MB included (minimum contractual term 24 months)

3

3x AMR WinControl software see price list 3x SW5600WC1

(Light version for 1 device, 20 meas. channels)

3x automatic memory readout

(option) see price list 3x SW5600WCZM9

04.18

GPRS connection 3+1

Installation of the VPN software on one computer for the purpose of simultaneously addressing several modems each with one or more connected devices. Each such modem is allocated a separate access with its own unique IP address; all connected devices can be interrogated simultaneously (requires at least SW5600WC2).

ALMEMO

®

Network technology

Required

3x GPRS modem

3x data cable ZA1909-DK5

3x modem adapter

Costs (net)

see price list included included

Note

Preconfigured with RS-232 connection

RS-232

VPN access

3x mobile communications card

AMR WinControl software approx. EUR 30 / month 3x mobile + 1x PC (minimum contractual term 12 months) approx. EUR 42 / month m2m 5 MB included (minimum contractual term 24 months) see price list

Automatic memory readout (option) see price list

SW5600WC2 (standard version)

SW5600WCZM9

GPRS connection 3+3

Installation of the VPN software on several computers for the purpose of addressing several modems each with one or more connected devices.

Each such computer and each such modem is allocated a separate access with its own unique IP address. Each computer can establish connections to several modems; however, one modem can only be connected to one computer at any one time.

Required

3x GPRS modem

3x data cable ZA1909-DK5

3x modem adapter

VPN access

Costs (net)

see price list included included

Note

Preconfigured with RS-232 connection

RS-232 approx. EUR 45 / month 3x mobile + 3x PC (minimum contractual term 12 months)

3x mobile communications card

3x AMR WinControl software approx. EUR 42 / month m2m 5 MB included

(minimum contractual term 24 months)

3 see price list

3x automatic memory readout (option) see price list

3x SW5600WC2 (standard version)

3x SW5600WCZM9

1)

VPN (virtual private network) is a non-public network that uses the infrastructure of another - usually public - network (e.g. the Internet).

2)

APN (access point name) is the name of a connection point in a GPRS network that permits access to an external packet data network

(e.g. the Internet).

3)

The prices quoted are examples only; real prices will fluctuate depending on terms currently offered by providers.

04.19

04.20

Content

State-of-the-art measuring instruments must be able to interconnect with their environment.

AMR-Control

(included in the delivery)

AMR WinControl the software for all ALMEMO

®

instruments

RMT WinControl software for evaluating, monitoring, networking

WinControl Client OCX and Simple ASCII Server

ALMEMO

®

View

05.02

05.02

05.06

05.16

05.17

05.18

Software

05.01

Software

State-of-the-art measuring instruments must be able to interconnect with their environment.

Special ALMEMO

®

software programs give you complete control of the whole measuring setup and ensure convenient device handling.

Once measured values have been acquired by the ALMEMO

®

measuring instrument, this data can be transmitted to a computer via modem, data line, optic fiber, or radio link.

“AMR-Control”, the WINDOWS configuration software, is included freeof-charge with all ALMEMO

®

devices.

This software package lets you program all the device parameters and scan all measured data via a single computer.

The “WinControl” package has been specially developed for data acquisition and measured data processing with

ALMEMO

®

equipment.

Acquired measured values can be displayed, arithmetically processed, stored, printed out, and exported to other software applications for further processing. ALMEMO

®

measuring instruments can thus be addressed in an already established corporate network.

A demo version of AMR WinControl can be downloaded free-of-charge from www.

ahlborn.com

AMR-Control : Full Control over the Instrument Setup and Convenient Device Handling

The software AMR-Control is supplied with each ALMEMO

®

data logger and allows for the complete programming of the sensors, for the configuration of the measuring instrument and for the readout of the data memory via serial interface.

The only item required is an ALMEMO

® data cable. The integrated terminal even allows to obtain online measurements from the PC.

As a result, you can keep a constant overview and can completely control your measuring task!

The latest program version is available for download from www.ahlborn.com.

05.02

AMR-Control, initial screen

Devices list

Device Programming

System Configuration

(programming of devices and connectors) testing / saving

Programming user menus

(on ALMEMO

®

2690 and 2890)

Software

Reading out from the measured value memory

05.03

Software

List of connectors / measuring points

Programming measuring points / programming connectors

Saving / loading connector / measuring point programs

Creating / saving multipoint calibration, special linearization

see Chapter

Input connectors

05.04

Measured values list with zero-setting / adjusting/ deleting

Software

Output modules list

Terminal for online measuring operations and for direct programming

05.05

Software

AMR WinControl the software for all ALMEMO

®

measuring instruments

Software Description:

• The AMR WinControl software package has been specially developed for data acquisition and measured data processing with ALMEMO

®

equipment.

• This software makes it very easy and convenient for the user to program and operate ALMEMO

®

devices.

• The acquired measured values can be displayed, arithmetically processed, stored, printed and for further data processing (also

ONLINE) can be exported to other software packages.

• It is possible to derive alarm conditions from the acquired or calculated variables and perform control measures.

• The Windows desktop and the context-sensitive online help ensure a quick familiarisation and a safe handling of the software.

• At our site (www.akrobit.de) you can find all the latest information regarding software versions and updates and also download the most recent trial version of the software.

Software Versions:

For 20 measuring points and one instrument

For any number of measuring points and instruments

For any number of measuring points and instruments, all options included

(except Data server, Web server and additional modules)

For any number of measuring points and devices, all options included (except add-on modules), with an integrated data server (simultaneous access by several RMT WinControl clients).

Update of the latest software version for older versions

Update of the latest software version for newer versions

Order no.

SW5600WC1

SW5600WC2

SW5600WC3

SW5600WC4

SW5600WCU3

SW5600WCU4

Options:

Network capability (for addressing several ALMEMO

®

devices)

Automatic generation of measured data files (daily files / weekly files)

Modem support

relays, starting other applications) Alarm function (alarm record, output to ALMEMO

®

Data server see 05.13

Web server see 05.14

New

Extended evaluation functions see page 05.11

Additional modules

Thermal comfort and air-conditioning calculations (as per DIN 1946, EN ISO 7730); (see 05.12)

Password protection (see 05.12)

Test bench manager (prerequisite : WC3 / WC4 or WC1 / WC2 + WCO2) (see 05.15)

Thermal transmittance (U) wizard (see 06.11)

Thermal quantity wizard (see 05.12 and chapter Building physics)

OPC export (see 05.12)

Additional protocol (selectable, requires WC3 / WC4) (see system integration, page 05.13)

Security package (requires WC3 / WC4) (see 05.15) including watchdog card

The memory is read out automatically (see 05.13 connecting options)

Complete packages (see 05.13 - 05.14)

Long-term / continuous monitoring

PIMEX - combined measured value recording, video recording, and evaluation functions

Hardware copy protection (see 05.15)

USB dongle

USB network dongle

Minimum system requirements :

Components :

Computer

Operating system

Memory

Free hard-disk capacity

Interfaces

Minimum configuration

IBM-compatible PC

Windows XP, 2003, Vista, 2008, 7, 8

(32 and 64 bit)

256 MB

25 MB

USB

Recommended configuration

IBM-compatible PC

Windows 7

1024 MB

100 MB

Order no.

SW5600WCO1

SW5600WCO2

SW5600WCO3

SW5600WCO5

SW5600WCO8

SW5600WCO9

SW5600WCO10

SW 5600 WCZM1

SW 5600 WCZM2

SW 5600 WCZM3

SW 5600 WCZM4

SW 5600 WCZM5

SW 5600 WCZM6

SW 5600WCZM7

SW 5600 WCZM8

SW 5600 WCZM9

SW 5600 WCV

SW 5600 WCP

SW 5600 HL

SW5600NHL

COM (RS232), USB,network card

Modem or ISDN

05.06

Function overviewy

Measured values - scanning

Number of measuring points supported

Number of connections supported

Support for ALMEMO

®

network

Connection types

Serial (COM), TCP/IP

Modem, GSM, and wireless modem support

Schedule-controlled connection setup

Measured values - display

Display of measured values (numeric, bar chart, wind rose, round gauges)

Line graph (YT), XY graph

Save / load presentation characteristics as format type

Table, overview

Zoom functions

Project icons

Work surfaces

Measured values - saving

Saving to hard disk - manual

Saving to hard disk - automatic

Automatic generation of daily, weekly, monthly files

Automatic saving on an event-controlled basis

Automatically saved files - sent by e-mail

Automatically saved files - backed up automatically

Fail-safe (only devices with failsafe mode and internal memory)

Measured values - analysis

Two measuring cursors with statistics function

Displaying local maximum and minimum values in a line graph

Loading comparative characteristics in a line graph

Measured values - processing

Global arithmetic channels

Local arithmetic channels for files already saved

Calculations based on external table values

Linking /splitting of files

Grouping measured value files in a particular folder (wild card search)

Grouping measured value files over a particular period of time

Exporting measured values

Clipboard

File formats

(MS-Excel XLS / XLSX, TXT / CSV, FAMOS, QS-STAT, DIAdem, binary)

Dynamic data exchange (DDE, OLE)

ONLINE data transmission to MS-Excel

ODBC export (e.g. SQL databases)

Measured values - import

ASCII (list, columns, table formats)

ALMEMO

®

View files

ODBC import (e.g. SQL databases)

Programming of measuring points and devices

Programming the characteristics of measuring points and devices

Automated scaling of third-party sensors

Measuring points programming - save to file / load from file

Editing the programmed file (similar to Excel tables)

Software

WC1 WC2 WC3 WC4 WCV

20

1 unlimited. unlimited. unlimited.

unlimited.

unlimited. unlimited. unlimited.

unlimited.

   

 

05.07

Software

Data reduction

Averaging function (ONLINE and OFFLINE)

Smoothing (over time / over number of values, ONLINE and OFFLINE)

Data logger functions

Programming the data logger (including averaging functions)

Read out from device memory (all / selective measured values)

Display of memory occupancy status

Alarm functions

Alarm value display in measuring points list and in all measured value displays

Alarm report with confirmation and comments text

Events list (audit trail)

Start a program in the event of a particular fault

Send e-mail / SMS in the event of an alarm

Switch ALMEMO

®

output relays (specific to measuring point).

Control commands dependent on measured values (KwikScript)

Advance warning alarm

Alarm log printout

Schedules for alarm processing

Automatic checking of system configuration

Password protection

Protection against unauthorized access

Protection against operator error by assigning individual access rights

Traceability of activities by means of an events list

Alarm confirmation with user identification

Control and regulation

Two-point controller with ALMEMO

®

output relay?*?s

Proportional controller with ALMEMO

®

analog output modules

PID controller with ALMEMO

®

analog output modules and arithmetic channels

User-definable command buttons

Keys and buttons in project icons and as a toolbar

Setting constants

Starting / stopping a measuring operation

Switching relays

Setting analog output values

Configuration management

Save / load program configuration

Printout

Diagrams, meas. value tables, meas. point list, file overview including comments

Network server functions

Displaying measured values and diagrams on Intranet or Internet

Embedding diagrams and project icons on your own Internet pages

Accessing the integrated web server via any browser

Accessing measured data and history data via TCP/IP (open text protocol)

Forwarding measured data to RMT WinControl

Availability of already acquired measured data even after program restart

Alarm confirmation via web server

After initial installation AMR WinControl will run in demo mode - comprising the full functionality of the professional version

(WC3) - for a trial period of 30 days, after which time it will have to be registered. All the functions incorporated in the professional version can be tried without restriction and without engagement. If further functions (additional modules) are needed for test purposes, these can also be enabled on a temporary basis. Users can thus try the software for the duration of the trial period with the full range of functions normally needed and then place an order after the system has been running to their complete satisfaction.

Registration does not need reinstallation.

05.08

Main Window/General View

• The main window is the platform for all operations with AMR WinControl. All actions run within this window and can be minimised to a symbol, within the window or together with the window, and run in the background.

• The measuring data can be presented as follows: Numeric presentation of measured values, bar diagram, wind rose, round instruments, line diagram, XY diagram, table, file overview.

• Windows can be distributed over various work surfaces between which it is possible to switch by means of tabs.

• The program can be operated by means of menu commands. Only those commands, which can be executed in the corresponding situation, will be available. For a faster operation context-sensitive menus, keyboard commands and symbols in the tool bar are available.

• Comprehensive help information is available via the function descriptions in the status line, notes in the tool bar and a context-sensitive help system.

List of Measuring Points, devices and connections

• As soon as the program is started and the serial interface is assigned, all sensors that are programmed and connected to the measuring instrument(s) will be recognized automatically and displayed in the list of measuring points.

• Apart from sensor specific data regarding the measuring range, comment, limit values, correction values the list also contains symbols for limit value exceeding, sensor breakage and online storage.

• Device-specific information, e.g. device type, memory occupancy, and settings for operating the data logger are also displayed.

• Measuring instruments can be connected via various interfaces (COM, TCP, modem) simultaneously; mixed-mode operation over various connections is possible.

Information regarding the current status of connections is displayed here.

ONLINE and OFFLINE Arithmetical Operations

• The arithmetical functions of the program allow to calculate physical variables from the measured data.

• The required variables can be defined via a formula editor and can be set as an arithmetic channel (virtual measuring point).

• Data acquired through these arithmetic channels can be, ONLINE and OFFLINE, further processed and presented.

• Depending on its definition an arithmetic channel can be globally available in the whole program or only locally within a data record (line or XY diagram, table).

• Even data records that have already been stored can be extended by arithmetic channels, as required.

Measuring Cursors/Statistic Function

• In the „Line Diagram“ view, the acquired data can be analysed using two measuring cursors (ONLINE and OFFLINE).

• The movement of the cursors can be performed in any area within the line diagram.

• Corresponding to the position of the measuring cursors, the measured values of all displayed lines located below the cursors will be displayed in a table.

• Through the integrated statistic function the difference of the values under the cursors, the minimum and maximum value and the average value of the area defined by the cursor positions will be calculated and also provided within the table.

• It is possible to perform a printout of the diagram and of the table displayed in the window, together or individually, or to copy them to the clipboard.

Software

05.09

Software

Data Logger Functions

• The settings required for data logger operation can be programmed within a dialogue.

• In the dialogue „Program Data Logger“ the current settings of the device, e.g. memory occupancy, start and end time, measuring cycle and print cycle will be displayed.

• The read-out of the device memory can be, individually or together, performed for all data loggers within the measuring network.

• An optional setting allows to defi ne that not all measured values but only a selection of the stored data should be read out of the device memories. The selection criteria available include „Average Values Only“, „Alarm Values Only“, „Meas. Values with

Number Only“ and „Only Values Within A Time Frame“.

The memory is read out automatically

• This module greatly facilitates the task of reading out from the device memory of an autonomous data logger.

• Saving data to the data logger is interrupted, its memory is read out, and, if this is successful, the memory content is deleted. The time-of-day is synchronized and saving data to the data logger is resumed.

• Reading out from memory can be completely automated in the form of schedules.

• All steps and possible errors are documented in the events list..

Monitoring Functions

• An alarm can be triggered by a component failure or a limit value infringement.

• Alarm processing can be activated individually for each measuring point.

• Alarms are reported visually and / or acoustically.

• The cause and the duration of events responsible for triggering the alarm are documented in an events list.

• Alarm reports can be confi rmed either individually or all together.

• If the cause of an alarm persists uncorrected an alarm reminder is issued to ensure that the alarm is not forgotten after it has been confi rmed.

• A line graph with settable history can be generated for the variable triggering an alarm.

• In the event of an alarm being triggered e-mails can be sent, ALMEMO

® relays can be switched, and programs or scripts can be executed.

• Alarm reports can be forwarded via the network.

output

• In the event of a limit value being infringed program control commands can be executed (KwikScript).

Control and regulation

• Two-point controllers, proportional controllers, and time-based controls are available.

• It is also possible, using arithmetic channels, to defi ne PID controllers.

• Setpoint curves und process sequences can be specifi ed by means of fi les with coordinates pairs.

• Values can be specifi ed and process sequences can be modifi ed - all via command buttons in project icons or the toolbar.

Automatic saving-to-memory

• Measured data can be saved to memory - manually or on a time-controlled or eventdriven basis.

• Not only daily / weekly / monthly fi les can be specifi ed but also fi les with any random periods of time.

• Data is saved to memory automatically in the background - irrespective of any opened diagrams, tables, or displays.

• Measured value fi les can be exported automatically on completion of a save-tomemory cycle and be sent by e-mail (as an option with the events list).

05.10

Extended evaluation functions

• Measured value files can be incorporated in new or already existing line charts in the form of comparative characteristics.

• Folders containing a large number of measured value files can be conveniently grouped using various patterns based on file names and filters according to time and measuring point.

• Local maximum and minimum values can be shown in a line chart as any measured value curve required. The search radius between maximum and minimum can be freely set.

Data Export:

• The data files can be, ONLINE and/or at any later point in time, stored in the following formats Excel (XLS / XLSX), ASCII (TXT / CSV), WK1, FAMOS,

QS-STAT, DIAdem, binary.

• With ODBC measured data can be exported in SQL databases (structured query language). This supports all data sources for which an ODBC driver is installed and set up on the system.

• The line and XY diagrams and the tables can be copied to the clipboard and, for example, be inserted into a protocol text.

• Via dynamic data exchange (DDE) it is possible to transfer measured values ON-

LINE to other applications, for example MS-EXCEL.

• Furthermore, line diagrams can be embedded into text documents (e.g. a MS Word document) via the OLE function.

Project Illustrations

• Project illustrations allow for visualising the setups of measurements and processes by using individually designed graphics and/or photographs (bitmaps).

• The presentation of the acquired data is provided in data fields that can be freely positioned; size and colours can be freely selected.

• Text fields can be filled with legend information and descriptions and can be freely positioned.

• Command buttons (keys and switches) can be freely positioned in the project icon and allow changes to values for performing calculations or controlling processes

(switching of relays or valves, etc.).

• The design of the command buttons can be changed in any way in the form of icons; the measurement setup can thus be visualized in a completely integrated way.

• Any number of project illustrations can be opened at the same time, allowing, for example, to have a presentation of the total view and detailed views of a project.

Thermal transmittance (U) wizard

• The thermal transmittance (U) wizard is available for ONLINE and OFFLINE calculations; it guides the user through all the required steps.

• The user can choose from a selection of calculation methods - for the experimental thermal transmittance value, for the thermal transmittance value according to DIN

4108, and for the official calculated value.

• Determination of the currently calculated value and the sliding average value.

• The calculation methods will be described and the allocation of the corresponding measuring variables will be provided.

• After completing all steps a line diagram will be created, which will then be filled with the measuring data and the calculated variables.

• The cursor function can be used to open the statistic table, which provides further evaluation options (see above).

Software

05.11

Software

PPD / PMV wizard (comfort index measurement)

• Calculation of thermal comfort as per DIN 1946 Part 2 and ISO 7730

• User guidance by means of a wizard and easy-to-understand evaluation

• Output in the form of “predicted mean vote” (PMV) and “predicted percent of dissatisfied“ (PPD)

• Online and offline calculation of PMV and PPD in real time or on the basis of measured values already existing

• Graphical representation of measured data and calculated values in a format suitable for export (e.g. ASCII, MS Excel, DiaDEM, etc.)

• Calculation parameters can be saved as a model for subsequent calculations.

• Additional PMV / PPD functions are available for use in arithmetic channels.

Thermal quantity wizard

• The thermal quantity is calculated automatically from the volume flow and the temperature difference.

• You can enter settings easily and conveniently using the wizard.

• Data tables for water are included in delivery; users can define their own extensions for other media themselves.

• The thermal quantity can be calculated in real time or on the basis of existing measured value files.

δQ = c

v • m

• dT

Password protection

• Integrated user management system protects AMR WinControl against unauthorized access. This policy reduces security risks to a minimum.

• Each change of user is logged in the events list for subsequent evaluation if this becomes necessary.

• Rights of access can be defined individually per user.

• Alarm confirmations can be assigned unequivocally to a particular user.

OPC export

ODBC

• „Openness, Productivity, and Collaboration”

• OPC is an established industrial standard for access procedures on a multi-vendor basis irrespective of manufacturer.

• AMR WinControl operates as an OPC client; it writes current measured values to the global variables provided by an OPC server.

• Data can be transferred in parallel to several OPC servers.

• Data from AMR WinControl can, with the aid of OPC, be visualized online in Lab-

View™.

• Open database connectivity

• ODBC is a standardized database interface used by SQL as its database language.

• Recorded measured values can thus be transferred to a database.

• Current measured values can be interrogated from a database per measuring cycle.

• A suitable ODBC driver for the database must be installed and set up on the system.

05.12

Connecting Options

• AMR WinControl can handle single measuring instruments as well as a network of measuring instruments of the ALMEMO

®

series.

• The connection to the measuring instrument(s) can be established via serial interface, USB, Bluetooth, or (GSM) modem.

• In a similar way, the measuring instruments can be addressed via a computer network (TCP/IP address) and VPN.

• Connections can be set up on a time-controlled basis. Reading out from the memory on ALMEMO

®

devices can be automated. The memory can on request be cleared and saving to memory can be resumed automatically. Any problems encountered are noted in the events list.

System Integration

• AMR WinControl also provides optional support for protocols used by devices from other manufacturers for measured value scanning in parallel for any number of connections.

• “SimpleASCII” is an open text protocol that can be used for the simple and straightforward integration of various measuring instruments in AMR WinControl.

• With the “OPC-Import-Protocol” data from an OPC server (e.g. Labview) can be read into AMR WinControl and processed by it.

• Data from the climate chambers (Feutron und Weiss Umwelttechnik) can be acquired and recorded in much the same way as e.g. gas analysis data from Emerson devices “XStream” and “NGA” or the “MRU Nova H8” device .

• Using the highly flexible Modbus protocol means that many other devices that support this protocol can be addressed.

• For the purposes of measuring and recording electrical variables (current, voltage, output, power factor, energy, etc) various protocols for “Simeas-T” and “Yokogawa” devices are available.

• Communication with a PLC can also be established via a further serial interface or

TCP/IP connection.

• This provides an open design for implementing automated test processes.

Measured value server

• With the measured value server up to 200 users simultaneously can access current measured values and the measured values history via a TCP network (Intranet /

Internet).

• Interface to any data acquisition and process control system

• Online transmission of measured data to other operating systems (e.g. LINUX,

WINDOWS CE, UNIX, etc.)

• Data distribution according to any specified criteria

• Customized solutions can be implemented using straightforward ASCII commands issued via the TCP protocol; all these commands are fully documented.

• Open “read-only” interface for any user-defined connection software

• “REMOTE WinControl” and “WinControl Client OCX” provide powerful standard solutions for the measured value clients.

Software

05.13

Software

Web server

• AMR WinControl provides a full range of web server functions for publishing web pages (HTML) in the Intranet / Internet. It also incorporates additional functions that can be used to output the contents of AMR WinControl windows directly onto web pages.

• Current measured values and measured value histories can be displayed in a variety of ways (line diagrams, XY diagrams, project illustrations) in the Intranet / Internet.

• Visualization of processes and systems

• Visual remote monitoring

• Confirmation of alarms via the browser (only with alarm function and password protection)

• Linking presentation and real-time data on web pages

• The way in which measured values are displayed does not depend on the operating system; only a browser is needed (MS Internet Explorer, Firefox, Chrome, Opera, etc.).

• Diagrams and measured values can also be displayed on smartphones and tablet PCs.

• Security provided by SSL / TLS and user authentication

• Very easy to use : Images generated from the contents of a window can be transmitted as soon as the program starts - without needing any further settings. For particularly demanding tasks the HTML pages must first be adapted and connected to the web server.

• The wide variety of image formats and special parameters make for transparency and loss-free scaling and permit automatic updating. Powerful real-time compression algorithms minimize the volume of data to be transmitted.

• All the layout facilities available in HTML, DHTML, and CSS can be exploited; combining with JavaScript is also possible.

• Graphics, text, and measured value displays can be combined and merged completely seamlessly.

• The web designer is free to specify, more or less independently of AMR WinControl, how the measured value displays are to appear.

• The user receives current measured data without being exposed to any sort of security risk - because there is no need for Java or special plug-ins.

SW5600WCV Package for long-term / continuous monitoring

This package, based on the AMR WinControl “professional” version, contains all the options and modules needed to implement long-term and continuous monitoring of critical measurable variables.

• Integrated user management with individually settable access rights and password protection

• Tamper-proof event list with sort and filter functions

• Trend monitoring - pre-alarm for signaling trend developments

• Signaling of alarms and events with user-specific confirmation and comments

• Alarm confirmation per web server (authentication and SSL / TLS available)

• Schedules : Automatic switching ON / OFF of alarm treatment for each measuring point, e.g. alarm treatment on working days between 06 and 18 o’clock only.

• Temporarily withdrawing certain measuring points from alarm treatment, e.g. for defrosting a cold room

• In the event of alarm an MS-Excel log can be printed out automatically. Users can modify the log provided or create their own.

• Failsafe : Automatic reading out of the device memory after loss of connection to the device

• Requirements: ALMEMO

®

device with failsafe mode and internal memory

• System configuration

• Integrity check on all measuring points and measuring instruments after program start

• Processing of measured and calculated variables in control and regulation functions

• Automatic printout and / or e-mail with daily files and event lists

• Including security package.

Security package

• Data security : Automatic backup of all automatically recorded data (daily and weekly files, measured values recorded on an event-controlled basis, event lists, etc.)

• Fail-safe : In the event of failure a watchdog is triggered for PC restart and / or signaling via relay.

• Including watchdog card

05.14

SW5600WCP: PIMEX

• Simultaneous acquisition of measured values from ALMEMO

®

devices together with video data from a digital source

• The measured data and video signal are synchronized and displayed together.

• The modes available are preview, record, and playback.

• It is also possible as an option to generate presentation videos from the acquired data.

• Possible applications : Documentation / visualization of the process environment

(e.g. for safety in the workplace, quality management, etc.)

Test bench manager

• Several autosave managers can be operated and organized via a convenient, easy-touse graphical user interface.

• Measured data can thus be saved simultaneously to different files.

• Autosave managers can be started and stopped independently of one another and according to various criteria (time-driven or event-driven).

• Different measuring locations (operating in parallel) can thus be treated separately.

• Measured value files can be indicated as write-protected already during recording.

• Including 10 autosave managers (optionally more available)

Copy protection

• AMR WinControl incorporates a copy protection system which requires a PCdependent code to enable it. To receive this code the user must first register the software by telephone, fax, or e-mail. Per licence purchased the software may be installed and operated on one computer.

• It is also possible as an option to request a hardware copy protection mechanism, a dongle; with this the software can be installed on any number of computers but will only run on that PC into which the dongle is currently plugged.

• A network dongle may contain more than one licence; with this it is possible - without the inconvenience of moving the dongle - to run the software simultaneously on as many computers in a company network as there are licences encoded in the dongle.

Software

05.15

Software

RMT WinControl software for evaluating, monitoring, networking

Program description

• Access to measured values on one or more AMR WinControl data servers in a local network or via the Internet

• Access to one measuring system by any number of users simultaneously

• Open and evaluate AMR files

• Same range of functions as AMR WinControl except for device access

• At our site (www.akrobit.de) you can find all the latest information regarding software versions and updates and also download the most recent trial version of the software.

RMT WinControl can perform the following:

• Monitoring of measured data from WinControl data servers at various locations

• Evaluation of acquired measured data / files independent of the recording computer

• Safe and secure access to the data acquisition system by “read-only” protocol

• Additional alarm handling and recording independent of the recording computer

• Since the measured value history is scanned, the evaluating computer therefore does not need to run continuously.

Software versions

Basic version (like SW5600WC2 except for device access and maximum 1 connection)

Professional version (like SW5600WC3 except for device access and maximum 1 connection)

Web server (like SW5600WC4 except for device access and any number of connections)

Update to the latest software version

Options

Automatic generation of measured data files (daily files / weekly files)

Modem support

Alarm function (event list, alarm e-mail / SMS, switching of ALMEMO

®

output relays)

Data server see page 05.13

Web server see page 05.14

Extended evaluation functions see page extended evaluation 05.11

Additional modules

Thermal comfort calculations as per DIN 1946, EN ISO 7730 see page 05.12, 12.14

Password protection see page 05.12

Test bench manager (prerequisite : WCR3 / WCR4 or WCR2 + WCRO2) see page 05.15

Thermal transmittance (U) wizard see page 05.11, 13.03

Thermal quantity wizard see page 05.12

OPC export see page 05.12

Hardware copy protection see page 05.15

Hardlock USB dongle

Hardlock USB network dongle

Minimum system requirements

Component

Computer:

Operating system

RAM

Free hard-disk capacity

Interfaces

05.16

Minimum configuration

IBM-compatible PC

Windows XP, 2003, Vista, 2008, 7, 8

(32 and 64 bit)

256 MB

25 MB

Network cardTCP/IP protocol

Recommended configuration

IBM-compatible PC

Windows 7

1024 MB

100 MB

Network card TCP/IP protocol

Internet or VPN connection

Order no.

SW5600WCR2

SW5600WCR3

SW5600WCR4

SW5600WCRU

Order no.

SW5600WCRO2

SW5600WCRO3

SW5600WCRO5

SW5600WCRO8

SW5600WCRO9

SW5600WCRO10

SW5600WCRZM1

SW5600WCRZM2

SW5600WCRZM3

SW5600WCRZM4

SW5600WCRZM5

SW5600WCRZM6

SW5600HL

SW5600NHL

Software

WinControl client OCX and SimpleASCII server

WinControl Client OCX

• Access to measured values on a WinControl data server in a local network or via

Internet

• MS ActiveX

®

universal components for integrating in your own applications

• Client licence for data server included

• Including documentation and simple application example for MS Excel

• This requires an AMR WinControl WC4 or option WCO8.

• Measured values from a WinControl data server can be transferred to your own applications by the WinControl client OCX.

• Current values and the measured value history can be scanned.

• Using OCX shortens development times appreciably because it relieves the developer of tasks involving communication with the data server.

• It can be incorporated in any application supporting OLE (Object Linking and Embedding) (e.g. MS Excel, Matlab, MS Access, MS SQL Server, etc.).

• Any programming language can be used for this purpose (C++, C#, Visual Basic

(VB, VBA, VBS), Delphi, etc.).

• Since multiple objects can be used simultaneously, data from various data servers can be acquired and recorded. OCX needs to be installed on the system only once.

Simple ASCII server

• Server component for sending data to AMR WinControl using the SimpleASCII protocol via TCP/IP

• MS ActiveX

®

universal components for integrating in your own applications

• including SimpleASCII protocol licence .

• Including documentation and simple application example for MS Excel.

• Using the SimpleASCII server measured values or data can be transferred to AMR

WinControl from another source (application or measuring instrument).

• Using this component shortens development times appreciably because it relieves the developer of tasks involving the programming of a TCP/IP server; (in programming languages (e.g. VBA, VBS) this is not possible without additional components).

• It can be incorporated in any application supporting OLE (Object Linking and Embedding) (e.g. MS Excel, Matlab, MS Access, MS SQL Server, etc.).

• Any programming language can be used for this purpose (C++, C#, Visual Basic

(VB, VBA, VBS), Delphi, etc.).

• With ActiveX-Control you can e.g. develop your own driver for incorporating an additional measuring instrument in AMR WinControl.

Software version Order no.

Client licence with OCX (client licence for the AMR WinControl server and OCX developer’s licence)

SW5600COCX

SimpleASCII server (SimpleASCII protocol licence for AMR WinControl with ActiveX-Control)

SW5600WCZM7

Minimum system requirements

The configuration actually needed depends on the software in which ActiveX-Control is integrated.

05.17

Software

ALMEMO

®

View

ALMEMO

®

View is a software package that can be used to evaluate and display measured data on one ALMEMO measurement channels.

®

device with up to four

ALMEMO

®

View runs under MS-Windows and can be used to drive an ALMEMO

®

device with up to four measuring points.

As soon as the connection between the computer and the measuring instrument has been established the program detects and lists these measuring points automatically.

The measured values are then read at a sampling rate selected by the user.

Datenlogger

The measured value memory on an ALMEMO grammed with ALMEMO

®

View

®

data logger (maximum four measuring points) can be read out, displayed as a line chart or table, and saved to a file. The parameters needed to operate the measuring instrument can be set via a dialog and pro-

Display of measured values

The recorded data can be displayed in numeric form, in a table, and as a line chart. It is possible to display just one measuring point or several measuring points at the same time in different modes.

Saving measured values

Measured values can be archived in line chart or table form.

Printing out

ALMEMO

®

View can also be used directly to print out diagrams, tables, or a list of all measuring points with their associated correction values, e.g. for the purposes of technical documentation. The results can be viewed in advance before printing out in the print preview. The program supports all printers that can normally be installed under MS-Windows.

Documentation

To compile protocols using some other software the line charts, tables, and lists in ALMEMO

®

View can be copied via the MS-

Windows clip-board to other application programs.

Software versionen

Basic ALMEMO

®

View software for maximum four measuring channels

(recommended for 1 measuring instrument with maximum 4 inputs, connection via COM-Port)

Order no.

SW5500AV

System requirements:

ALMEMO

®

View runs under MS-Windows from Windows 98 and higher.

05.18

General accessories

Content

ALMEMO

®

memory connector with micro-SD ZA 1904 SD 06.02

Extension cable for all other measuring probes (except thermocouples) 06.03

Extension Cable for NiCr-Ni Sensors 06.04

GPS mouse for determining position

Accessories for measuring instruments

Batteries, Rechargeable Batteries, battery supply

Mains Adapter, Power Supply Cables

Carry cases, Rack case

06.05

06.05

06.06

06.07

06.08

06.01

General accessories

ALMEMO

®

memory connector with micro-SD ZA 1904 SD

• for ALMEMO

®

data loggers, as of version 6

• Large memory

• High data security

• Measured values can be saved to a text file.

• The memory card in the data logger can be replaced quickly and easily on site.

• Files can be transferred to a PC quickly and easily via a card reader

Technical data

Measuring instruments for ALMEMO® 2590-2/-3S/-4S, 2690,

2890, 4390, 5690, 5790, 8490, 8590

Memory connector on device output socket A2

ALMEMO

®

memory connector

Integrate drive for micro-SD card

Memory card MicroSD industry standard

(Industrial Grade SSD SLC Technology) with high performance, reliability and durability, possible up to 2 GB, standard FAT16 format

Measured values

Ring memory

File format

Reading device

Measuring software

Variants

ALMEMO

®

memory connector with micro-SD memory card (512 MB) including USB card reader

Micro-SD memory card (512 MB as replacement)

With 128 MB approx. 8 million measured values no

ASCII text file, measured values in table format, separated by semi-colons

USB card reader for removable storage media

WinControl (as of version 6), see Chapter Software

Order no.

ZA1904SD

ZB1904SD

Micro-SD memory card (as replacement)

Micro-SD memory card, including USB card reader

06.02

General accessories

Extension cable for all other measuring probes (except thermocouples)

Extensions, 5 meters and longer

Longer sensor lines

If distances exceeding this really are necessary, then - instead of extension cables - longer sensor lines should be used. For this purpose the sensor connector must be detached, the sensor cable extended in the conventional way, and the connector then refitted to the end.

Extensions up to 4 meters

Passive extension cables ZA9060VK

Passive 8-pin extension cables with ALMEMO

®

connectors are available in lengths of 1, 2, and 4 meters (ZA 9060 VK1/2/4); these are suitable for all sensors (except thermocouples).

The cable length between sensor connector and measuring instrument must not exceed four (4) meters; if this maximum length is exceeded, communications with the connector EEPROM may be adversely affected.

Intelligent extension cable ZA 9090 VKC with RS485 from 5 meters and above

If the sensor cable cannot be extended as such , it is still possible to use new intelligent extension cable(s) ZA9060-VKC with microcontroller for up to 100 meters. Two microcontrollers transmit EEPROM data from the sensor connector and measured data from digital sensors (DIGI) in both directions via RS485 and make this interference-resistant data available for the measuring instrument. Sensors can thus be freely interchanged as and when necessary (e.g. calibrated sensors with correction values,.multipoint calibration or special linearization (ZAxxxxSS)).

new:

also for digital ALMEMO

®

D6 probes

!

The total length of all passive extension cables connected to an ALMEMO

®

measuring instrument must not exceed four (4) meters. If the total length exceeds this, the device’s internal data bus may, depending on environmental conditions, be subject to interference.

!

These intelligent extension cables cannot be used for thermocouples or for sensors with a frequency / pulse output (e.g. turbines / rotating vanes FVA915, frequency

/ pulse / rotational speed ZA 9909 AKx / FUA 9192, DC measuring modules ZA 99xx AB).

When using extension cable(s) operation in sleep mode is not possible

Types :

Extension cable for all other sensors

1 meter long (passive)

2 meters long (passive)

4 meters long (passive)

New

5 meters long (intelligent, with microcontroller, not sensor-specific)

10 meters long (intelligent, with microcontroller, not sensor-specific)

20 meters long (intelligent, with microcontroller, not sensor-specific)

30 meters long (intelligent, with microcontroller, not sensor-specific)

50 meters long (intelligent, with microcontroller, not sensor-specific)

100 meters long (intelligent, with microcontroller, not sensor-specific)

Order no.

ZA9060VK1

ZA9060VK2

ZA9060VK4

ZA9090VKC5

ZA9090VKC10

ZA9090VKC20

ZA9090VKC30

ZA9090VKC50

ZA9090VKC100

06.03

General accessories

Extension Cable for NiCr-Ni Sensors

Extensions, 5 meters and longer

Longer sensor lines

If distances exceeding this really are necessary, then - instead of extension cables - longer sensor lines should be used.

For this purpose the sensor connector must be detached, the sensor cable extended in the conventional way with a compensation line, and the connector then refitted to the end.

Extensions up to 4 meters

Passive extension cables ZA9020VK (NiCr-Ni)

Passive 8-pin extension cables with ALMEMO® connectors are available in lengths 1, 2, and 4 meters; for NiCr-Ni thermocouples special extension cables with an integrated compensation line are available (ZA 9020 VK1/2/4).

The cable length between sensor connector and measuring instrument must not exceed four (4) meters; if this maximum length is exceeded, communications with the connector EEPROM may be adversely affected.

Active extension cables ZA9020VKP (NiCr-Ni)

If for some reason it is not possible to extend the sensor cable itself, an active extension cable ZA9020VKP (NiCr-Ni) can be used.

This cable incorporates an ALMEMO

®

connector with an integrated EEPROM for data storage; this connector is a copy of the sensor connector. A compensation line is used for the extension.

The terminals in the connectors are also made from thermo material.

If correction values or other sensor-specific settings have been programmed in the sensor connector (e.g. comments, average values, etc.), these must also be programmed (being a copy) in the

ALMEMO

®

connector on the extension cable.

The active extension cable is then sensor-specific.

!

The total length of all passive extension cables connected to an ALMEMO

®

measuring instrument must not exceed four (4) meters. If the total length exceeds this, the device’s internal data bus may, depending on environmental conditions, be subject to interference.

Types :

Extension cable, from NiCr-Ni compensation line

1 meter long (passive)

2 meters long (passive)

4 meters long (passive)

10 meters long (active, with EEPROM, sensor-specific)

06.04

Order no.

ZA9020VK1

ZA9020VK2

ZA9020VK4

ZA9020VKP10

General accessories

GPS mouse for determining position

Variants

Fix for current location via GPS

Order no.

Measuring channels for latitude and longitude

GPS mouse with cable and ALMEMO

®

connectorr

ZAD919GPS

Accessories for measuring instruments ALMEMO

®

2450, 2490, 2590 and output interface ZA 8006 RTA

Rubber safety holster, green

Rubber safety holster, gray including carry strap

Order no.

ZB2490GS1

ZB2490GS2

Top hat rail mounting

Order no.

1 battery compartment cap with top hat rail holder fitted, including top hat

ZB2490HS

Magnetic fastening

2 pot magnets, including 2 screws

(for battery compartment cap)

Order no.

ZB2490MH

06.05

General accessories

Batteries and Rechargeable Batteries

Types:

9V battery (spare)

Order no.

ZB2000B9

Charger, int. in connector incl. 9V bat.

ZB2000LS

9V rechargeable battery

AA battery, 1.5 V

ZB2000A9

ZB2000B1

AA NiMH rechargeable battery, 1.2 V, 1600 mA, coded for charging in ALMEMO

(e.g. ALMEMO

®

2690-8)

®

unit

ZB2000A1NM

Rechargeable batteries

Types Order no.

Rechargeable battery, 12 V, 1600 mAh, NiMH with intelligent high-speed charging housed in case 174 x 29 x 137 mm (LxWXH) (without plug connections) voltage output via 3-pin socket

ZB5690AP

Connector mains unit, 90 to 260 VAC for charging the battery

ZB1212NA9

Connecting cable from battery to ALMEMO

®

device length = 1.5 meters, with ALMEMO

®

plug for ALMEMO

®

2450, 2490, 2470, 2590-2/-3S/-4S, 2690

ZA1012AKA

With 3-pin bayonet coupling for ALMEMO

®

5690, 8590, 8690

With hollow connector for ALMEMO

®

2890, 6290

ZB5090EKA

ZB2290EKA

06.06

Mains Adapter

ZA 1312 NA7

ZB 1212 NA9

General accessories

Variants Order no.

Transformer power supply / desktop version 230 VAC

12 VDC, 1 A DIN hollow connector for ALMEMO

®

2890-9, 6290-7B2

ZB1112NA7

12 VDC, 1 A ALMEMO

® devices ALMEMO

®

connector e.g. for hand-held

2450, 2490, 2590, 2690 ZA1312NA7

12 VDC, 1 A 3-pin bayonet coupling e.g. for ALMEMO

®

8590-9

12 VDC, 1 A With free ends

ZB1212NA7

ZB1012NA7

Switching power supply / connector variant 90 to 260

VAC

12 VDC, 2.5 A 3-pin bayonet coupling e.g. for ALMEMO

®

5690, 8690

ZB1212NA9

12 VDC, 2.5 A With free ends

ZB1012NA9

Accessories

Conversion connector for mains-powered devices

Euro-plug to US standard (flat-pin)

ZB1000UA

DC Power Supply Cables

Supply cables for DC voltages

• Usage for car and electric fence batteries.

• For instruments that need to be supplied from the car battery.

Variants Order no.

10 to 30 V DC, electrically isolated, with DIN hollow connector for ALMEMO

®

2890-9, 6290-7B2

Output : 12V DC / 1 A (max.)

ZB2590UK

10 to 30 V DC, electrically isolated, with ALMEMO

®

connector for ALMEMO

®

2450, 2490, 2590, 2690-8

Output: 12 V DC / 250 mA (max.)

ZA2690UK

Output: 12 V DC / 1 A (max.)

ZA2690UK2

10 to 30VDC, electr. isol., with bayonet coupling for

ALMEMO

®

8590

Output: 12VDC/250mA (max.)

ZB3090UK

10 to 30VDC, electr. isol., with bayonet coupling, for

ALMEMO

®

5690-9, 8690 output: 12V DC / 1.25A (max.)

ZB3090UK2

Adapter cable with universal car connector

ZB1000AKU

New

ALMEMO

®

Programming 1 A

power supply plug, 9 to 12 VDC, not electr. isolated, with clamp connector for ALMEMO socket on hand-held devices ALMEMO

2690

Programming 0.2 A

®

®

DC

2450, 2490, 2590,

ZA1312FS1

ZA1312FS8

06.07

General accessories

Instrument Cases

ZB 2590 TK2 ZB 5600 TK3

Types Order no.

Carry cases (approx. dimensions in cm)

Carry case, large, aluminum profile frame / ABS (acrylonitrile butadiene styrene) - e.g. for ALMEMO

®

710, 2690,

2890 data logger, Inside dimensions 48 x 35 (WxD) x 6 (H) +

6 cm (removable insert)

ZB2590TK2

Carry case, universal, high, aluminum profile frame / ABS, e.g. for ALMEMO® 5690 measuring systems

Inside dimensions 48 x 25 (WxD) x 16 (H) + 10 cm (removable insert)

ZB5600TK3

Instrument case for all ALMEMO® handheld devices, inside dimensions (WxDxH) 42 x 30 x 9 (divided into compartments, see photograph)

ZB2490TK2

ZB 2490 TK2

Rack case (approx. dimensions in cm)

Rack case with carrying handle, for ALMEMO

®

MA5690xxBT8 measuring systems, in 19-inch sub-rack, 84

DU, height 5 HU Outside dimensions (WxDxH) 54 x 50 x

27, with integrated lockable rack draw, inside dimensions

(WxDxH) 40 x 37 x 7 (for cables, accessories, or laptop)

ZB5090RC

ZB 5090 RC

ALMEMO

®

input connector also for existing sensors (see Chapter Input Connectors)

ALMEMO

®

output modules (analog, relay, trigger) (see Chapter Output Connectors)

ALMEMO

®

data connection, network technology, Bluetooth modules

Wireless and modem transmission (see Chapter Network Technology).

Software for the presentation and evaluation of measuring data, including many notes, is described in Chapter Software.

The software ´AMR-Control´ for measurement setup and convenient device handling, as well as the manual, are included with the delivery of all ALMEMO digital outputs.

®

devices with

06.08

Temperature

Content

The Right Temperature Sensor For Any Measuring Task

Types and Fields of Application

Sheathed sensors

Accessories for Sheathed sensors

Low-cost universal sensors

NiCr-Ni sensors in various designs

Thermowires and compensation lines

Connectors for thermocouples

Pt100 cable sensor

Pt100 sensors and NTC sensors in various designs

Plug-in laboratory sensor

Pt100 glass thermometer with immersion depths as per ASTM

Insertable sensor

Infrared measuring technology

Emissivity values for different materials

Digital infra-red sensor for measuring surface temperature

Compact infra-red probe head AMiR FIA 844

Infra-red transmitter for measuring surface temperature

Infrared Measuring Heads in Two-Wire Design AMiR 7838

07.21

07.22

07.26

07.27

07.28

07.30

07.33

07.34

07.36

07.38

07.40

07.02

07.04

07.06

07.09

07.10

07.11

07.18

07.20

07.01

Temperature

The Right Temperature Sensor For Any Measuring Task

Selecting the right type of temperature sensor depends on your measuring task. For example, thermocouples, resistor-based sensors (Pt100 and Ntc) and pyrometers (infrared sensors) are available.

Rule of Thumb:

• Thermocouples are very fast and provide a large measuring range.

• Resistor-based sensors are more accurate but slower.

• Ntc sensors are very fast, accurate, but they have a limited measuring range.

• Infrared sensors do not contact the device under test and they have very small time constants, but they depend on the emission grade.

• The larger the measuring range, the more universal the possible range of applications.

Selection Criteria:

Select the temperature sensor that suits your measuring task according to the criteria below:

• Meas. range

• Accuracy

• Response time

• Stability

• Type of construction

07.02

Temperature

Thermocouples

Thermocouples consist of two spot-welded wires of different metals or alloys. The thermoelectric effect at the contact surface is used to measure temperatures. A relatively small thermoelectric voltage is caused, which depends on the temperature difference between the measuring point and the connecting terminals.

Accuracy, Operating Temperatures:

The basic values for the thermoelectric voltages and for the permissible tolerances of thermocouples are specified in standard DIN/

IEC 584. Our thermocouple sensors are available in two tolerance classes as per DIN/IEC 584-2. Following limit values apply

(highest value in each case) : type K / N

Class 1:

Class 2:

±1.5 °C or ±0.004 x l t l

±2.5 °C or ±0.0075 x l t l

(-40...1000°C)

(-40...1200°C)

Our thermocouple sensors generally comply with Class 2 as per DIN/IEC 584-2. The specified Tmax values refer to the tip of the sensor. The specified T

90

times refer to measuring operations in a moving liquid. The sensor handles and cables are usually resistant to temperatures up to +80 °C. Heat-resistant cables are also available on request.

Various types of thermocouples are available; these can be distinguished in terms of their temperature range, sensitivity, and in particular their compatibility with the test substance. The most popular thermocouple is the NiCr-Ni (type K).

new

Connecting cable with thermal line (stranded wire)

There is no adverse temperature effect at the juncture from measuring element to cable.

With immediate effect, the sensor connecting cables for many sensor types will use a new thermal line (stranded wire, thermal line class 2) instead of the conventional compensation line. The transition from measuring element (sensor tip) to connecting cable (in the cable sleeve or in the handle) thus remains, even over a wide temperature span (up to 200 °C), unaffected by temperature error; the usual measuring errors caused by temperature differences at the juncture when using a conventional compensation line can thus with the new thermal line be avoided.

For just a few sensor types and extension cables a compensation line will continue to be used as previously. The compensation lines generally comply with Class 2 as per DIN 43722. For type K the operating temperature range of the compensation line is 0 to

150 °C.

Resistor-Based Sensors (Pt100 Sensors)

When measuring the temperature the increase in resistance at increasing temperatures is utilised at the Pt100 sensors. The measuring resistor is fed with a constant current and the voltage drop at the resistor is measured as a function of the temperature. Due to the small resistance variation (0.3 to 0.4W/°C) the 4-conductor circuit should always be used to exclude any influences from the lead wires.

Accuracy, Operating Temperatures:

Pt100 sensors are, as standard, used with Class B (DIN/IEC 751) measuring resistors (surcharge for DIN Class A or 1/5 DIN Class

B accuracy). The specified Tmax values relate to the tip of the sensor. The specified T on request.

90

times are related to measurements in a moving liquid. The sensor handles and cables are usually resistant to temperatures up to +80 °C. Heat-resistant cables are available

Measuring ranges, resolution

PT100 probes FP Axxx are by default assigned measuring range PT100-1 (resolution 0.1 K). Measuring range PT100-2 (resolution

0.01K) can be programmed as alternative on the 1st channel or in addition on the 2nd channel.

New

Measuring range PT100-3 (resolution 0.001K) in range 0 to 65 °C (for V6 devices, with effect from 2690-8, 2890-9,

85/8690-9, 5690-1/2)

Measurement Accuracies of Resistor-Based Temperature Sensors

Designation Range Maximum Deviation

Test resistances

Pt 100 Ω at –200°C at –100°C at –50°C at 0°C at +100°C at +200°C at + 300°C at + 400°C higher accuracies for an additional charges

DIN Class B

±1.3 K

±0.8 K

±0.3 K

±0.8 K

±1.3 K

±1.8 K

±2.3 K

DIN Class A

±0.25 K*

±0.15 K

±0.35 K

±0.55 K

±0.75 K

1/5 DIN Class B

±0.06 K

±0.16 K

±0.26 K

±0.36 K

Order no. OPG2 Order no. OPG5**

* Range -50 °C only for sheathed sensors with 2mm diameter and bigger

** On request, depending on the sensor design

07.03

Temperature

Thermistors (NTC Sensors)

NTC sensors (thermistors) have a significantly higher resistance than Pt100 sensors. When measuring temperatures their negative temperature coefficient is utilised, i.e. the resistance is decreasing with increasing temperatures.

Accuracy, Operating Temperatures:

The accuracy data of the normalised NTC sensors are based on manufacturer specifications. The specified T tip of the sensor. The specified T stant to temperatures up to 90°C.

90 max

values relate to the

times are related to measurements in a moving liquid. The sensor handles and cables are resi-

Accuracies

Designation

NTC element

(10K at 25°C)

Range

–20 to 0°C

0 to 70°C

70 to 125°C

Maximum Deviation

±0.4 K

±0.1 K

±0.6 K

Types and Fields of Application

The construction variants of temperature sensors are as many and diverse as the measuring tasks.

T max

is the maximum operating temperature of the sensor tip.

T

90

is the time required by the sensor to reach 90% of the step response after a jump in temperature . The specified T to measuring operations in a moving liquid.

90

times refer

The temperature sensors listed are also available, on request, with other lengths and diameters

Surface sensors with flat measuring tip

For measurements on good heat conductors, on even and plain surfaces.

Surface sensor with spring-type thermocouple band

For quick measurements, also on non-plain surfaces.

Immersion probes

For measurements in liquids, as well as powdery substances, air and gases.

Sensors with heat-resistant measuring tip

For measurements at extremely high temperatures.

Sensor with penetrating tip

For measurements in plastic and pasty substances.

Sword probe

For measurements in paper, cardboard and textile stacks.

Transducer with free sensor

For measurements in air and gases

!

If you do not find a suitable sensor in this catalogue, we can manufacture it according to your specifications

(technical drawing or detailed specification) and supply you with a customised sensor!

07.04

Temperature

Temperature Measurement à la ALMEMO

®

All ALMEMO

®

sensors can be adjusted, i.e. the correction values of the sensor can be stored in the connector. This considerably increases the accuracy of measurement.

As a result of the DAkkS and factory-set calibrations performed by us, the corrective factors are automatically determined, stored in the connector plug and locked. Maximum accuracy can then be achieved.

Ordering Information

ALMEMO

®

sensors are available in different designs. The type designation can be identified by:

„P“

„N“

„T“

= temperature sensor with Pt100W test resistance

= temperature sensor with NTC element

= temperature sensor with NiCr-Ni element

All temperature sensors with an ALMEMO

®

flat connector can be identified by the „A“ in the order no.

Naturally, they are also available for the measuring instruments of our THERM series. In this case they will have a circular connector.

When ordering please replace the letter „ A“ by the number „9“.

Example: FT A1201 (with ALMEMO

®

connector) >> FT 91201 (with circular connector for THERM devices)

Describe your measuring task to us!

We can provide you with comprehensive advice and find the most cost-effective solution for you.

Please do not hesitate to ask !

Use Your Existing Sensor Technology!

The patented idea of the intelligent connector makes the ALMEMO

®

system an extraordinarily flexible measuring system.

Instead of our pre-configured ALMEMO

®

sensors you can also use your own, existing sensors.

• We can supply you with pre-programmed ALMEMO

®

connectors that contain the corresponding sensor parameters and matching measuring ranges. They have six screw terminals and can be easily connected.

• You can correct the errors of the sensors, which means that even simple sensors become precision transducers

• Listing all the combinations and application options would be beyond the scope of this catalogue. Special programming, range extensions and linearisations for other sensor technology are always available for ALMEMO

®

devices.

• The pricing for this results from the efforts and the number of devices required.

ALMEMO

® sensor connector with 6 terminal screws and

EEPROM.

07.05

Temperature

Sheathed sensors

• These reasonably priced sensors are for universal use (-200 to +1100 °C) and suitable for immersion measurements in liquids, air, and gases. The sheathed line, depending on diameter, can be bent - within certain limits.

• Different connection variants :

With cable and ALMEMO

®

connector Order no. FxAxx, with cable and free ends, Order no. Fx0xx.

Connector options :

With THERM circular connector : Option T9020RS,

with miniature Thermo flat connector : Option OT9020FS.

Thermocouple sheathed sensors FTAxx and FTANxx

Measuring element:

Sensor tip, sheathed line : diameter, length, operating temperature; see table; material Inconel 2.4816

Cable sleeve :

Standard cable :

FTAxx; NiCr-Ni thermocouple, type K, DIN class 1 (see 07.03)

FTANxx; NiCrSi-NiSi thermocouple, type N, DIN class 1 (see 07.03)

Here the sensor tip and sheathed line are of the same diameter.

These types are therefore also suitable for mounting with clamped screw connections.

Brass, hexagonal, L = 65 mm, circumdiameter = 9 mm, operating temp. –40 to +160 °C

New 1.5 meter FEP / silicone thermal line (stranded wire)* Operating temp. -50 to +200°C

Cable options :

There is no adverse temperature effect at the juncture from measuring element to cable.

Compensation line, PVC / PVC, insulated, operating temperature –20 to +105 °C

The compensation line is also available, on request, with FEP / FEP, insulated.

ALMEMO

®

connector FTAxx NiCr-Ni ZA9020FS with resolution 0.1 K

FTANxx NiCrSi-NiSi ZA9021FSN with resolution 0.1 K

Pt100 sheathed sensors FPAxx

Measuring element :

Options :

Sensor tip :

Sheathed line :

Cable sleeve :

Pt100 4L, DIN class B (see 07.03)

DIN class A, 1/5 DIN class B (see 07.03) diameter, length, operating temperature; see table; material stainless steel diameter, length; see table; material stainless steel

On certain types the sensor tip and sheathed line are of different diameter; (i.e. the sensor tip is thicker). These types are therefore not suitable for mounting with clamped screw connections. Types suitable for clamped screw connections are available on request.

Brass, hexagonal, L = 65 mm, circumdiameter = 9 mm, operating temp. –40 to +160 °C

Standard cable :

Cable options :

1.5 meters line, FEP / silicone, insulated, operating temperature –50 to +200 °C

Line, PVC / PVC, insulated, operating temperature –20 to +105 °C

The line is also available, on request, with FEP / FEP, insulated.

ALMEMO

®

connector Pt100, ZA9030FS1, with resolution 0.1 K

Option : Pt100 ZA9030FS2 with resolution 0.01 K (standard with 1/5 DIN class B)

NTC sheathed sensors FNAxx

Measuring element :

Sensor tip :

Sheathed line :

Cable sleeve :

Standard cable :

Cable options :

NTC type N (see 07.04) diameter, length, operating temperature; see table; material stainless steel diameter, length; see table; material stainless steel

On certain types the sensor tip and sheathed line are of different diameter; (i.e. the sensor tip is thicker). These types are therefore not suitable for mounting with clamped screw connections. Types suitable for clamped screw connections are available on request.

Brass, hexagonal, L = 65 mm, circumdiameter = 9 mm, operating temp. –40 to +160 °C

1.5 meters line, PVC / PVC, insulated, operating temperature –20 to +105 °C

Line, FEP / silicone, insulated, operating temperature –50 to +200 °C

The line is also available, on request, with FEP / FEP, insulated.

ALMEMO

®

connector NTC, ZA9040FS, with resolution 0.01 K.

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.06

Temperature

Sheathed sensors

Sensor with :

Sensor tip, dimensions d1, sheathed line, dimensions d2, overall length (including sensor tip) L,

Cable sleeve, dimensions length = 65 mm, circumdiameter = 9 mm, Cable

Thermocouple sheathed sensors NiCr-Ni, type K

Typical Application: universal, in range -40 ° C to 900 ° C

Diameter d1=d2

0.5 mm

0.5 mm

0.5 mm

0.5 mm

0.5 mm

1.5 mm

1.5 mm

1.5 mm

1.5 mm

3.0 mm

3.0 mm

3.0 mm

3.0 mm

Operating temperature

Sensor tip

-200...900°C

-200...900°C

-200...900°C

-200...900°C

-200...900°C

-200...1100°C

-200...1100°C

-200...1100°C

-200...1100°C

-200...1100°C

-200...1100°C

-200...1100°C

-200...1100°C

Length L

50 mm

100 mm

250 mm

500 mm

1000 mm

100 mm

250 mm

500 mm

1000 mm

100 mm

250 mm

500 mm

1000 mm

Order no

FTA05L0050

FTA05L0100

FTA05L0250

FTA05L0500

FTA05L1000

FTA15L0100

FTA15L0250

FTA15L0500

FTA15L1000

FTA30L0100

FTA30L0250

FTA30L0500

FTA30L1000

Connection cable

FEP/silicone

Thermal line

(stranded wire)

PVC/PVC

Compensation line

Operative range

-50...200°C

-20...105°C

Length

1.5 m

5 m

1.5 m

5 m

Order no

default

OTK01L0050

OTK02L0015

OTK02L0050

Thermocouple sheathed sensors NiCrSi-NiSi, type N

Typical application: in the range -200 ° C to 1150 ° C, long-term stability at high temperatures

Diameter d1=d2

1.5 mm

1.5 mm

1.5 mm

3.0 mm

3.0 mm

3.0 mm

6.0 mm

6.0 mm

6.0 mm

Connection cable

FEP/silicone

Thermal line

(stranded wire)

Operating temperature

Sensor tip

-200...1150°C

-200...1150°C

-200...1150°C

-200...1150°C

-200...1150°C

-200...1150°C

-200...1150°C

-200...1150°C

-200...1150°C

Operative range

-50...200°C

Length L

500 mm

750 mm

1000 mm

500 mm

750 mm

1000 mm

500 mm

750 mm

1000 mm

Length

1.5 m

Order no

FTAN15L0500

FTAN15L0750

FTAN15L1000

FTAN30L0500

FTAN30L0750

FTAN30L1000

FTAN60L0500

FTAN60L0750

FTAN60L1000

Order no

default

5 m OTNK01L0050

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.07

Temperature

Resistor-based sensors Pt100 4L

Typical Application: universal, in range -40°C to 500°C

Diameter d1 Sensor tip

1.5 mm

1.5 mm

1.5 mm

2.2 mm*

2.2 mm*

2.2 mm*

3.2 mm*

3.2 mm*

3.2 mm*

Diameter d2, Sheathed line

1.5 mm**

1.5 mm**

1.5 mm**

2.0 mm

2.0 mm

2.0 mm

2.8 mm

2.8 mm

2.8 mm

Operating temp.

Sensor tip

-40...500°C

-40...500°C

-40...500°C

-40...500°C

-40...500°C

-40...500°C

-40...500°C

-40...500°C

-40...500°C

Length L

100 mm

250 mm

500 mm

100 mm

250 mm

500 mm

100 mm

250 mm

500 mm

Order no.

FPA15L0100

FPA15L0250

FPA15L0500

FPA22L0100

FPA22L0250

FPA22L0500

FPA32L0100

FPA32L0250

FPA32L0500

Options

PT100 measuring resistor

Accuracy

Class B

Class A

Class 1/5 DIN Class B

Ceramic measuring resistor

operating range -200 ... 600 ° C

Connection cable Operative range

FEP/silicone -50...200°C

PVC/PVC -20...105°C

Order no.

default

OPG2

OPG5

OPM1

Length

1.5 m

5 m

1.5 m

5 m

Resistor-based sensors NTC

Typical Application: universal, in range 0°C to typ. 70°C

Diameter d1 Sensor tip

2.0 mm

2.0 mm

2.0 mm

3.2 mm*

3.2 mm*

3.2 mm*

Diameter d2, Sheathed line

2.0 mm

2.0 mm

2.0 mm

2.8 mm

2.8 mm

2.8 mm

Operating temp.

Sensor tip

-20...100°C

-20...100°C

-20...100°C

-20...100°C

-20...100°C

-20...100°C

Length L

100 mm

250 mm

500 mm

100 mm

250 mm

500 mm

Order no.

default

OPK01L0050

OPK02L0015

OPK02L0050

Order no.

FNA20L0100

FNA20L0250

FNA20L0500

FNA32L0100

FNA32L0250

FNA32L0500

Connection cable Operative range

PVC/PVC -20...105°C

Length

1.5 m

5 m

Order no.

default

OPK02L0050

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.08

Temperature

Handle for sensors with hexagonal cable sleeve

Option Handle including fitting

Order no. OFH1

Clamp srew connection ZT943xKV

Operative range

For sheath elements

Option:

Notched steel ring

(once fitted, cannot be removed),

T max

= 800 °C

For ZT9431KV

Order no. OT9431ST

For ZT9432KV

Order no. OT9432ST

Variants (with PTFE clamping ring) for types

FTA15Lxxxx, FPA16Lxxxx for types

FTA30Lxxxx, FPA30Lxxxx

and FNA30Lxxxx

Order no.

ZT9431KV

ZT9432KV

Technical data

Operating temperature

Thread up to maximum 250 °C with option up to 800 °C

M8x1, 14 AF

Heat-conducting paste ZB9000WP

For surface measurement, operative range -30 to +200 °C, heat-conducting paste, tube, 12 ml

Order no. ZB9000WP

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.09

Temperature

NiCr-Ni-sensor FTA 15 P

For immersion measurement

Pt100-sensor FPA 32 P

For immersion measurement

NTC-sensor FNA 305

For Indoor air measurements

Meas. element: NiCr-Ni Class 1 *

Measuring tip: Operative range -200...+1100 °C

200x1.5 mm, sheathed line, Inconel

T

90

: *

Cable:

1.5 s approx. 1.4 m FEP/silicone with spray-coated ALMEMO

®

connector

L = 200 mm

Sensor with handle

(No variants available)

Order no. FTA15P

Order no. FTA15PH

Meas. element: Pt100, Class B *

Measuring tip: Operative range -40...+500 °C

200 x 2.8/3.2 mm, sheathed line

T

90

: *

Cable:

10 s approx. 1.4 m PVC with spray-coated ALMEMO

®

connector

L = 200 mm

Sensor with handle

(No variants available)

Order no. FPA32P

Order no. FPA32PH

Meas. element NTC*

Measuring tip Operative range -10 to +60 °C

T

90

(non-condensing)

Protective tube in stainless steel

Diameter = 3.0mm, length = 50 mm mounted directly on ALMEMO

®

connector

8 s

L = 50 mm

Order no. FNA305

(No variants available)

* For general technical data, see page 07.03

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.10

Temperature

NiCr-Ni sensor with handle FTA 120x

Meas. element: NiCr-Ni class 1 *

Measuring tip: Operative range -200...+400 °C

T

90

: *

Handle: *

Cable:

3 s

138 mm

1.5 m PVC

L = 30 mm

Order no. FTA1201

L = 150 mm

Order no. FTA1202

For surface measurement and immersion measurement

NiCr-Ni sensor with handle FTA 122 LxxxxH

For surface measurement and immersion measurement

T

Meas. element: NiCr-Ni class 1 *

Measuring tip: Operative range -200...+400 °C

90

: *

Silver rivet, level, not electr. isolated

3 s

Handle: * 127 mm

Cable:

new

1.5 m FEP/silicone thermal line**

L = 50 mm

Order no. FTA122L0050H

L = 100 mm

Order no. FTA122L0100H

L = 200 mm

Order no. FTA122L0200H

NiCr-Ni sensor with handle FTA 121 LxxxxH

For surface measurement and immersion measurement

Meas. element: NiCr-Ni class 1 *

Measuring tip: Operative range -200...+400 °C

T

90

: *

Silver rivet, level, angled, not electrically isolated

3 s

127 mm

Cable:

new

1.5 m FEP/silicone thermal line**

L = approx. 50 mm

L = approx. 200 mm

Order no. FTA121L0050H

Order no. FTA121L0200H

NiCr-Ni sensor with handle FTA 150 LxxxxH

T

Meas. element: NiCr-Ni class 1 *

Measuring tip: Operative range -200...+800 °C

90

: *

Handle: *

neu:

(for brief periods1000°C)

Stainless-steel rivet, level, electrically isolated

3 s

127 mm

For surface measurement and immersion measurement

L = 350 mm

Order no. FTA150L0350H

L = 700 mm

Order no. FTA150L0700H

* For general technical data, see page 07.03.

L = 1250 mm

Order no. FTA150L1250H

** There is no adverse temperature effect at the juncture from measuring element to cable. see page 07.03

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.11

Temperature

NiCr-Ni sensor FTA 109 P

For surface measurement

Meas. element: NiCr-Ni class 2 *

Measuring tip: Operative range -50...+500 °C

Thermal ribbon, not electr. isolated

Measuring head approx. 15 mm diameter

T

90

: *

Cable:

1 s approx. 1.5 m PVC

L = approx. 180 mm

Sensor with handle

(No variants available)

Order no. FTA109P

Order no. FTA109PH

NiCr-Ni sensor FTA 104 P

For surface measurement

Meas. element: NiCr-Ni class 2 *

Measuring tip: Operative range -50...+500 °C

Thermal ribbon, not electr. isolated

Measuring head approx. 15 mm diameter

T

90

: *

Cable:

1 s approx. 1.5 m PVC

L = approx. 180 mm, with 90° angle, approx. 50 mm

Order no. FTA104P

Sensor with handle

(No variants available)

Order no. FTA104PH

NiCr-Ni sensor with handle FTA 153 LxxxxH

For surface measurement

Meas. element: NiCr-Ni class 2 *

Measuring tip: Operative range -200...+250 °C

T

90

: *

Handle: *

Cable:

Thermal ribbon, crossed, not electrically isolated

1.5 s

127 mm

1.5 m PVC

L = 100 mm

Order no. FTA153L0100H

L = appr. 180 mm angled 45°, 160/50mm

Order no. FTA1533L0180H

NiCr-Ni sensor with handle FTA 1535 LxxxxH

For surface measurement

T

Meas. element: NiCr-Ni class 2 *

Measuring tip: Operative range -200...+250 °C

90

: *

Handle: *

Cable:

Thermal ribbon, not electr. isolated

2 s

127 mm

1.5 m PVC

L = 100 mm

Order no. FTA1535L0100H

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.12

Temperature

NiCr-Ni sensor with handle FTA 420 LxxxxH

For surface measurement on level surfaces

Meas. element: NiCr-Ni Class 1 *

Measuring tip: Operative range -50...+500 °C

T

90

: *

Handle: *

Cable:

Silver disc, spring-loaded, not electrically isolated

2 s

127 mm

1.5 m PVC

L = 150 mm

Order no. FTA420L0150H

NiCr-Ni sensor FTA 025 P

Meas. element: NiCr-Ni Class 2 *

Measuring tip: Operative range -50...+300 °C

T

90

: *

Cable:

Thermal ribbon, not electr. isolated

Fastened by magnet

1.5 s approx. 2 m PVC

Magnet sensor

(No variants available)

Order no. FTA025P

Magnet sensor for surface measurement

Magnet sensor with Velcro fastener e.g. for pipework

Klettband:

Operating range: approx. 400 mm, for pipe diameter appr. 10 to 75 mm

-10 ... +110 °C mounted on sensor tip

Magnet sensor, including Velcro fastener

Order no. FTA025PKB

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.13

Temperature

NiCr-Ni sensor FTA 131

T

Meas. element: NiCr-Ni Class 2 *

Measuring tip: Operative range -50...+100 °C

90

: *

Cable:

Silver rivet, level, spring-loaded, not electrically isolated

Fastened by magnet

3 s

3 m FEP/silicone

Magnet sensor

Order no. FTA131

Magnet sensor For surface measurement

NiCr-Ni sensor FTA 026 P

Meas. element: NiCr-Ni Class 1 *

Measuring tip: Operative range -50...+300 °C

T

90

: *

Cable:

Thermal ribbon, not electrically isolated

1.5 s approx. 0.9 m line, fabric insulation

Ribbon sensor

Order no. FTA026P

(No variants available)

For surface measurement

NiCr-Ni sensor FTA 8068

T

Meas. element: NiCr-Ni Class 2 *

Measuring tip: Operative range -50...+120 °C

90

: *

Thermal ribbon, not electr. isolated

Fastened by pipe clamp

(spring-loaded)

3 s

Pipe diameter: 12...25 mm

Cable: 1.2 m PVC

Pipe clamp sensor

Order no. FTA8068

For surface measurement on pipes

NiCr-Ni sensor FTA 8069

For surface measurement on pipes

Meas. element: NiCr-Ni Class 2 *

Measuring tip: Operative range -50...+120 °C

T

90

Thermal ribbon, not electr. isolated

Fastened by pipe clamp

: * 3 s

Pipe diameter: 12...30 mm

Cable: 1.2 m PVC

Pipe clamp sensor

Order no. FTA8069

* For general technical data, see page 07.03.

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.14

Temperature

NiCr-Ni film thermocouple FTA 683

For surface measurement

T

Meas. element: NiCr-Ni Class 2*

Measuring tip: Operative range -100 to +200°C

90

: *

Folie, Insulation Kresol

2 s

new

With permanently connected FEP / silicone thermal line

(stranded wire)**

-50 to +200°C, 2 meters, with ALMEMO

®

connector

Order no. FTA683

Measuring element without cable, free ends

(for your own sensors)

Order no. FT0683

NiCr-Ni sensor FTA 390 x

For surface measurement

T

Meas. element: NiCr-Ni Class 2 *

Measuring tip: Thermowire, welded,

90

: *

Wire: not electrically isolated

3 s

1.5 m

Insulation, glass fiber,

Operative range -25...+400 °C

Order no. FTA3900

Insulation FEP,

Operative range -200...+205 °C

Order no. FTA39010

* For general technical data, see page 07.03.

** There is no adverse temperature effect at the juncture from measuring element to cable. see page 07.03

Digital infra-red sensor for measuring surface temperature FIAD43

Operative range: -40...600 °C,

Miniature probe head, with cable and ALMEMO

®

D6 plug and 1 mounting nut

Cable length = 1 m

Cable length = 3 m

For technical data, see page 07.34

Order no. FIAD4332

Order no. FIAD4332L3

DAkkS or factory calibration KI9xxx temperature for digital sensor (see chapter Calibration certificates)

Compact infra-red probe head FIA844

Operative range: -20...500 °C,

Probe head, with cable and ALMEMO

®

plug and 2 mounting nuts

Cable length = 1 m

Cable length = 3 m

For technical data, see page 07.36

Order no. FIA844

Order no. FIA844L3

Factory calibration KI9xxx temperature for sensor (see chapter Calibration certificates)

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.15

Temperature

NiCr-Ni sensor with handle FTA 05 L0050H

T

Meas. element: NiCr-Ni Class 1 *

Measuring tip: Operative range -200...+500 °C

90

: *

Sheathed line, Inconel

0.8 s

Handle: * 127 mm

Cable:

new

1.5 m FEP/silicone thermal line**

L = 50 mm

Order no. FTA05L0050H

For immersion measurement

NiCr-Ni sensor with handle FTA 125 LxxxxH

For immersion measurement

T

Meas. element: NiCr-Ni Class 1 *

Measuring tip: Operative range -200...+800 °C

90

: *

Handle: *

Cable:

Sheathed line, Inconel

1.5 s

127 mm

new

1.5 m FEP/silicone thermal line**

L = 300 mm

Order no. FTA125L0300H

NiCr-Ni sensor with handle FTA 126 LxxxxH

For immersion measurement

T

Meas. element: NiCr-Ni Class 1 *

Measuring tip: Operative range -200...+900 °C

90

: *

Sheathed line, Inconel

2.5 s

Handle: * 127 mm

Cable:

new

1.5 m FEP/silicone thermal line**

L = 600 mm

Order no. FTA126L0600H

NiCr-Ni sensor with handle FTA 1261 LxxxxH

T

Meas. element: NiCr-Ni Class 1 *

Measuring tip: Operative range -200...+500 °C

90

: *

Sheathed line, Inconel

3 s

Handle: * 127 mm

Cable:

neu:

1.5 m FEP/silicone thermal line**

For immersion measurement in plastic and pasty substances, e.g. bitumen

L = 150 mm

Order no. FTA1261L0150H

L = 300 mm

Order no. FTA1261L0300H

* For general technical data, see page 07.03.

** There is no adverse temperature effect at the juncture from measuring element to cable. see page 07.03

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.16

Temperature

NiCr-Ni sensor with handle FTA 123 LxxxxH

For immersion measurement in plastic and pasty substances

T

Meas. element: NiCr-Ni Class 1 *

Measuring tip: Operative range -200...+300 °C

90

: *

Penetrating tip

3 s

Handle: * 127 mm

Cable:

new

1.5 m FEP/silicone thermal line**

L = 50 mm

Order no. FTA123L0050H

L = 100 mm

Order no. FTA123L0100H

NiCr-Ni sensor with handle FTA 1231 LxxxxH

For immersion measurement in plastic and pasty substances

Meas. element: NiCr-Ni Class 1 *

Measuring tip: Operative range -200...+400 °C

T

90

: *

Handle: *

Penetrating tip, cone stainless steel 1.4541

6 s

127 mm

Cable:

new

1.5 m FEP/silicone thermal line**

L = 250 mm

Order no. FTA1231L0250H

* For general technical data, see page 07.03.

** There is no adverse temperature effect at the juncture from measuring element to cable. see page 07.03

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.17

Temperature

NiCr-Ni thermowire T 190-0

Thermowire: NiCr-Ni, class 2

*

Insulation : Glass fiber (wires and sheath)

Operating temp.: -25°C to +400°C

Wire diameter: 0.5 mm

External diameter: approx. 1.3 x 2.1 mm

NiCr-Ni thermowire per meter with glass fiber covering

Order no. LT01900

NiCr-Ni thermowire sensor, welded tip, with

ALMEMO

®

connector 1.5m long

Order no. FTA3900

ALMEMO

®

connector 5m long

Order no. FTA3900L05

NiCr-Ni thermowire T 190-1

Thermowire: NiCr-Ni, Class 2

*

Insulation : Glass fiber (wires and sheath)

Operating temp.: -25°C to +400°C

Wire diameter: 0.2 mm

External diameter: approx. 0.6 x 1.0 mm

NiCr-Ni thermowire per meter with glass fiber covering

Order no. LT01901

NiCr-Ni thermowire sensor, welded tip, with

ALMEMO

®

connector 1.5 m long

Order no. FTA3901

ALMEMO

®

connector 5m long

Order no. FTA3901L05

NiCr-Ni thermowire T 190-2

Thermowire: NiCr-Ni, Class 2

*

Insulation : PVC (wires and sheath)

Operating temp.: -10°C to +105°C

Wire diameter: 0.5 mm

External diameter: approx. 2.2 x 3.4 mm

NiCr-Ni thermowire per meter with PVC insulation

Order no. LT01902

NiCr-Ni thermowire sensor, welded tip, with

ALMEMO

®

connector 1.5 m long

Order no. FTA3902

ALMEMO

®

connector 5 m long

Order no. FTA3902L05

NiCr-Ni thermowire T 190-3

Thermowire: NiCr-Ni, Class 2

*

Insulation : Silicone (wires and sheath)

Operating temp.: -45°C to +200°C

Wire diameter: 0.5 mm

External diameter: approx. 4 mm

NiCr-Ni thermowire per meter with silicone insulation

Order no. LT01903

NiCr-Ni thermowire sensor, welded tip, with

ALMEMO

®

connector 1.5 m long

Order no. FTA3903

ALMEMO

®

connector 5 m long

Order no. FTA3903L05

* For general technical data, see page 07.03

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.18

Temperature

NiCr-Ni thermowire T 190-10

Thermowire: NiCr-Ni, class 2

*

Insulation : FEP (Wires and sheath)

Operating temp.: -200°C to +205°C

Wire diameter: 0.5 mm

External diameter: approx. 1.5 x 2.5 mm

NiCr-Ni thermowire per meter with FEP insulation

Order no. LT019010

NiCr-Ni thermowire sensor, welded tip, with

ALMEMO

®

connector 1.5m long

Order no. FTA39010

ALMEMO

®

connector 5m long

Order no. FTA39010L05

NiCr-Ni thermowire T 190-11

Thermowire: NiCr-Ni, class 2

*

Insulation :

Wire diameter: 0.2 mm

External diameter:

FEP (Wires and sheath) approx. 1.3 x 2.0 mm

NiCr-Ni thermowire per meter with FEP insulation

Order no. LT019011

NiCr-Ni thermowire sensor, welded tip, with

ALMEMO

®

connector 1.5m long

Order no. FTA39011

ALMEMO

®

connector 5m long

Order no.

FTA39011L05

NiCr-Ni thermowire T 190-7

Nur für trockene, nicht agressive Umgebung!

Thermowire: NiCr-Ni, Class 2

*

Insulation : Ceramic fiber (Wires and sheath)

Operating temp.: -40°C to +1200°C

Wire diameter: 0.8 mm

External diameter: approx. 3 x 4 mm

NiCr-Ni thermowire per meter with ceramic fiber insulation

Order no. LT01907

NiCr-Ni thermowire sensor, welded tip, with

ALMEMO

®

connector 1.5m long

Order no. FTA3907

ALMEMO

®

connector 5m long

Order no. FTA3907L05

NiCr-Ni compensation line T 191-1

compensation line:

Insulation :

NiCr-Ni

PVC (Wires and sheath)

Operating temp.: -10°C to +105°C

Wire diameter: 0.5 mm

External diameter: approx. 3.6 mm

NiCr-Ni bunched conductor with PVC insulation, for each meter

Order no. LT01911

Other types are available on request.

LT01912 Insulation Silicone/silicone/glass filament, up to 200°C

LT01913 Insulation PVC / screening film / PVC, up to 105°C

NiCr-Ni thermal line (Litze) T 191-6

Thermal line (stranded wire)**: NiCr-Ni*

Insulation: Wires : FEP, sheath : silicone

Operating temp.: -50...+200°C

Wire diameter: 0.7 mm

External diameter: approx. 3.8 mm

NiCr-Ni thermal line (stranded wire) with FEP / silicone insulation, per meter

Order no. LT01916

* For general technical data, see page 07.03.

** There is no adverse temperature effect at the juncture from measuring element to cable. see page 07.03

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.19

Temperature

ALMEMO

®

connector for thermocouples

(see Chapter Input connectors)

For Types K, N, L, J, T

(no thermo-electric transition / with thermal material)

NiCr-Ni (K)

NiCroSil-NiSil (N)

Fe-CuNi (L)

Fe-CuNi (J)

Cu-CuNi (T)

Order no. ZA9020FS

Order no. ZA9021FSN

Order no. ZA9021FSL

Order no. ZA9021FSJ

Order no. ZA9021FST

For Types U, S, R, B, AuFe-Cr

Cu-CuNi (U)

PtRh10-Pt (S)

PtRh13-Pt (R)

PtRh30-PtRh6 (B)

AuFe-Cr (A)

Order no. ZA9000FSU

Order no. ZA9000FSS

Order no. ZA9000FSR

Order no. ZA9000FSB

Order no. ZA9000FSA

ALMEMO

®

adapter plug with miniature flat socket

For Types K, J, T, S

NiCr-Ni (K)

Fe-CuNi (J)

Cu-CuNi (T)

PtRh-Pt (S)

Order no. ZKA029RA

Order no. ZJA029RA

Order no. ZTA029RA

Order no. ZSA029RA

Miniature flat connectors for thermocouples types K, J, T, S, E

Examples for NiCr-Ni (K):

NiCr-Ni flat socket

NiCr-Ni flat connector

Order no. ZK9029FB

Order no. ZK9029FS

Locking plate (10 pieces)

NiCr-Ni single built-in socket

Order no. ZB9026VP

Order no. ZK9029FE

1-row panel with NiCr-Ni socket

Order no. ZK9029FB1

6-row panel with NiCr-Ni socket

Order no. ZK9029FB6

• Connectors with thermo contacts for avoiding voltage corruption at thermocouple junctions.

• For ambient temperatures –183 to +200 °C.

• Locking plate for complete coupling.

Order numbers for the above examples are compiled from the following coding elements : Z9029F.

The coding elements can be taken from the table below.

Ordering:

Type

NiCr-Ni (K)

Fe-CuNi (J)

Cu-CuNi (T)

NiCr-CuNi (E)

PtRh-Pt (S)

Color (IEC 584)

Variant  green Male connector = S black brown lilac

Female connector = B

Panel

1-er (1-rhg)

6-er (1-rhg)

12-er (1-rhg)

24-er (2-rhg) orange

Panel dimensions

38 x 38 x 2.5 mm

113 x 38 x 2.5 mm

203 x 38 x 2.5 mm

203 x 76 x 2.5 mm mounting depth: 25.4 mm

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.20

Temperature

Pt100 cable sensor

Inexpensive resistance-based temperature sensors, for universal use. For immersion measurements in air and gases. Rigid protective tube made from stainless steel

A wide variety of cable variants.

Operating temperature (depending on variant) -40 to +400°C.

Technical features

Measuring element : Pt100 4L, DIN class B, For technical data see page 07.03.

Protective tube: Diameter, length see Variants, stainless steel 1.4301

Junction of protective tube / connecting cable: Direct, hard-crimped for dry uses

Cables: Length = 1.5 meters, Other lengths are available as options. Cable diameter is never larger than the diameter of the protective tube.

Operating temperature: see variants, Always for whole sensor (i.e. sensor tip and cable)

ALMEMO

®

connector: Pt100 ZA9030FS2 with resolution 0.01 K .

Variants

With FEP / FEP cable (black),

Operative range -40...+250°C:

Diameter

3.0 mm

3.0 mm

4.0 mm

4.0 mm

Length

50 mm

100 mm

50 mm

100 mm

Order no.

FPA30K03L0050

FPA30K03L0100

FPA40K03L0050

FPA40K03L0100

With FEP / silicone cable (red),

Operative range -40...+200°C:

Diameter Length

5.0 mm

5.0 mm

50 mm

100 mm

Order no.

FPA50K01L0050

FPA50K01L0100

6.0 mm

6.0 mm

50 mm

100 mm

FPA60K01L0050

FPA60K01L0100

A longer cable is available as an option

Total length 5 m

Total length 10 m

OPK01L0050

OPK01L0100

A longer cable is available as an option

Total length 5 m

Total length 10 m

OPK03L0050

OPK03L0100

Other designs are available on request:

Pt100 cable sensors FPA30K20L0020 vapor-tight (protective class IP69K), inter alia for temperature measuring in autoclaves, sterilizing units, high-temperature steam applications, vacuum applications, freeze drying units, -30. to +150 °C, protective tube in stainless steel with PFA cable.

Cable with glass-fiber / glass-fiber / VA wire shielding,

Operative range -40...+400°C:

Diameter Length

5.0 mm

5.0 mm

50 mm

100 mm

Order no.

FPA50K06L0050

FPA50K06L0100

6.0 mm

6.0 mm

50 mm

100 mm

FPA60K06L0050

FPA60K06L0100

A longer cable is available as an option

Total length 5 m

Total length 10 m

OPK06L0050

OPK06L0100

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.21

Temperature

Pt100 sensor with handle FPA 106 LxxxxH

T

Meas. element: Pt100, class B *

Measuring tip: Operative range -40...+500 °C

90

: *

Handle: *

Cable:

Sheath element, stainless steel

8 s

127 mm

1.5 m FEP/silicone

L = 100 mm

Order no. FPA106L0100H

For immersion measurement

Pt100 sensor with handle FPA 123 LxxxxH

For immersion measurement in plastic and pasty substances

T

Meas. element: Pt100, Class B *

Measuring tip: Operative range -40...+500 °C

90

: *

Handle: *

Cable:

Penetrating tip

8 s

127 mm

1.5 m FEP/silicone

L = 100 mm

Order no. FPA123L0100H

Pt100 sensor with handle FPA 124 LxxxxH

T

Meas. element: Pt100, Class B *

Measuring tip: Operative range -40...+300 °C

90

: *

Handle: *

Cable:

Silver rivet, level

10 s

127 mm

1.5 m FEP/silicone

L = 100 mm

Order no. FPA124L0100H

For surface measurement and immersion measurement

* For general technical data, see page 07.03

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.22

NTC sensor with handle FNA 106 LxxxxH

Temperature

T

Meas. element: NTC *

Measuring tip: Operative range -20...+100 °C

90

: *

Handle: *

Cable:

Sheath element, stainless steel

8 s

127 mm

1.5 m PVC

L = 100 mm

Order no. FNA106L0100H

For immersion measurement

NTC sensor with handle FNA 123 LxxxxH

For immersion measurement in plastic and pasty substances

T

Meas. element: NTC *

Measuring tip: Operative range -20...+100 °C

90

: *

Handle: *

Cable:

Penetrating tip

8 s

127 mm

1.5 m PVC

L = 100 mm

Order no. FNA123L0100H

NTC sensor with handle FNA 124 LxxxxH

T

Meas. element: NTC *

Measuring tip: Operative range -20...+100 °C

90

: *

Handle: *

Cable:

Silver rivet, level

10 s

127 mm

1.5 m PVC

L = 100 mm

Order no. FNA124L0100H

For surface measurement and immersion measurement

NTC sensor FNA 305

For room air measurement

Meas. element: NTC*

Measuring tip: Operative range -10...+60°C

T

90

:

(non-condensing), Protective tube in stainless steel diameter = 3.0mm, length = 50mm mounted directly on ALMEMO

®

connector

8 s

L = 50 mm

Order no. FNA305

(No variants available)

* For general technical data, see page 07.03

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.23

Temperature

Pt100 sensor FPA 611 x

For surface measurement

T

Meas. element: Pt100, class B *

Measuring tip: Operative range see below

Copper, level

new

Improved thermal transfer thanks to innovative

90

: * sensor element and new contact technology

Cable:

20 s see below

Surface sensor

-10...+90°C, Cable PVC, 2 m

Order no. FPA611

-10...+110°C, Cable, PFA, 3m for more demanding mechanical stress ALMEMO

®

connector, resolution 0.01 K

Order no. FPA611S01

Accessories

Fixture for fastening with cable ties

Best-Nr. ZB9611RM

Pt100 film sensor FPA 686

Meas. element: Pt100, class B*, gewickelt

Messfläche: Operative range -50...+200 °C,

T

90

*:

Cable:

temperature-resistant foil,

15 x 40 mm, approx. 0.5 mm thick

2 s

Stranded wire PFA, 4-wire twisted

Length 2 m

Length 10 m

Order no. FPA686

Order no. FPA686L10

For surface measurement

Pt100 ceramic chip sensor element FP 0802

Meas. element: Pt100, Class B *

Measuring tip: Operative range -40...+400 °C

Ceramic chip sensor

Connection wires: 10 mm, bare

Ceramic chip sensor

Order no. FP0802

Unprotected sensor element for constructing your own sensors

* For general technical data, see page 07.03

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.24

NTC sensor FNA 611

Temperature

T

Meas. element: NTC *

Measuring tip: Operative range -10...+90 °C

90

: *

Cable:

Copper, level

20 s

2 m PVC

Surface sensor

Order no. FNA611

For surface measurement

NTC sensor FN 0001 K

Unprotected sensor element with cable

Accessories

Fixture for fastening with cable ties

Best-Nr. ZB9611RM

Meas. element: NTC*

Measuring tip: Sensor element, unprotected

Operative range: -20...+100°C

Connection wires: appr. 180 mm, fluoropolymer insulation

Connecting cable: 2 meters, PVC, thin stranded pick-up wire, Operative range -10 to +90 °C

Cable juncture, in shrink-fit

NTC sensor with cable, free ends

Option:

Single connectors for 1 sensor

Order no. FN0001K

ALMEMO

®

connector including assembly

Order no. OT9040AS

Double connector for 2 sensors Order no. OT9040AS2

NTC sensor element FN 0001

Unprotected sensor element for constructing your own sensors

Meas. element: NTC *

Measuring tip: Operative range -20...+100 °C

Sensor

Connection wires 180 mm, fluoropolymer insulation

Sensor

Order no. FN0001

* For general technical data, see page 07.03

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.25

Temperature

Pt100 Plug-in laboratory sensor FPA 416

Meas. element: Pt100, class B *

Measuring tip: Operative range -40...+150 °C

T

90

:*

Cable:

15 s

Silicone/FEP 3m

ALMEMO

®

connector: resolution 0.01 °C

Plug-in laboratory sensor

Order no. FPA416

Measuring element PT100, 4-conductor class B, integrated in the socket of a 6 mm laboratory connector made of brass (nickel-plated).

Pt100 Plug-in laboratory sensor FPA 414

Meas. element: Pt100, Class B *

Measuring tip: Operative range -40...+150 °C

T

90

:* 15 s

Cable: Silicone/FEP 3m

ALMEMO

®

connector: resolution 0.01 °C

Plug-in laboratory sensor

Order no. FPA414

Measuring element PT100, 4-conductor class B, integrated in the socket of a 4 mm laboratory connector made of brass (goldplated).

Plug-in laboratory sensor, examples of use

Measuring object with hole for inserted PT100 plug-in laboratory sensor.

* For general technical data, see page 07.03

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.26

Temperature

Pt100 glass thermometer with immersion depths as per ASTM

Operative range:

For immersion measurement in liquid media at low immersion depths.

Variants Order no.

Pt100 glass thermometer with immersion depths as per

ASTM specifications, with ALMEMO

®

connector (including

2-meter FEP / silicone cable)

FPA910

Technical data

Meas. element: Pt100, class A

Measuring tip

Shaft

Operative range -50 to +310 °C

Glass, tapered

Diameter = 3 mm, length = 15 mm

Glass, Diameter = 6 mm

NL= 250 mm (total nominal length)

Labeling codes for immersion depths : identification rings on the shaft as per

ASTM specifications (American Society for Testing and Materials)

T

90

2.5 seconds

Cable junction sleeve Stainless steel, 8 x 40 mm

Cable exit secured with shrink-fit sleeve

Cable 2 meters, FEP / silicone

ALMEMO

®

connector Resolution 0.01 K

Also available on request

Resolution 0.001 K, in range -8 to +65 °C

On devices with effect from ALMEMO

®

2690

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.27

Temperature

Insertable sensor NiCr-Ni with round mounting plug T 820-6

Operative range:

Measuring tip, spring-loaded, for surface and immersion measurement.

Technical data

Measuring element NiCr-Ni class 2*

Measuring tip

T

90

*

Insert length

Fixture

Connection

Operative range -40 to +400 °C

Silver rivet, level, spring-loaded not electrically isolated

3 s

60 mm (see layout drawing)

Plastic, Ø 20 mm, resistant up to +120 °C

Round mounting plug

Accessories:

ALMEMO

®

connecting cable,

2 meters Order no. ZA9020BK2

Types

Insertable sensor NiCr-Ni with round mounting plug

Order no.

FT98206

Screw-fit sensor NiCr-Ni, Pt100, NTC, with fitted cable Fx 0710 L27M10

Operative range:

For immersion measurement

Technical data

Meas. element:

Sensor materials

Operative range

Thread

Insert length

Cable see under variants

Stainless steel see under variants

M10

27 mm (see layout drawing)

3 meters, free ends see under variants

Options:

ALMEMO

®

connector, including assembly, for NiCr-Ni sensors

Order no. OT9020AS

For Pt100 sensors

Order no. OT9030AS

For NTC sensors

Order no. OT9040AS

Variants

Screw-fit sensor, with cable, free ends

Order no.

NiCr-Ni class 2*, -100 to +400 °C Thermal line

Glass filament / glass filament / VA wire shielding

Option Cable length 5 meters

NTC*, -20 to +100 °C Cable, PVC,

Cable juncture, in shrink-fit

FT0710L27M10

OTK06L0050

Pt100 class B* -40 to +200 °C Cable FEP / silicone

Cable juncture, in shrink-fit

Option Cable length 5 meters

FP0710L27M10

OPK01L0050

FN0710L27M10

Option Cable length 5 meters

OPK02L0050

* For general technical data, see page 07.03

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.28

Temperature

Einbausensor Pt100 mit Anschlußkopf FP 0463

Operative range:

For immersion measurements, pressure-sealed up to 15 bar.

Technical data

Meas. element:

Sensor tube

Operative range:

Thread

Insert length

Terminal head

Pt100, class B*

Stainless steel

-40...+350°C

1/2”, with copper ring seal, pressure-sealed up to 15 bar

70 mm (see layout drawing)

Clamp connector

Variants Order no.

(on request with cable and ALMEMO

®

connector)

Insertable sensor with terminal head

Pt100, Class B*

FP0463

Insertable sensor PtRh-Pt (S) with terminal head FT 0425

Operative range:

For immersion measurements, up to 1400 or 1600 °C.

Technical data

Measuring element Thermowire PtRh-Pt (S) see under variants

Measuring tip

Operative range

Insert length

Protective tube

Cable

Ceramic tube see under variants see under variants

500 mm

Ceramic, replaceable,

7 x 1 mm

2-meter compensation line silicone insulation, free ends

Accessories

Ceramic protective tube for

T04251 Order no. ZB9425SR1

Ceramic protective tube for

FT04252 Order no. ZB9425SR2

Options

ALMEMO

®

connector with assembly Order no. OT9020AS

Variants

(including 2-meter compensation line)

Order no.

PtRh-Pt(S), T ceramic 610 max

= 1400°C, element-Ø = 0.35 mm,

FT04251

PtRh-Pt(S), T ceramic 710 max

= 1600°C, element-Ø = 0.5 mm,

FT04252

* For general technical data, see page 07.03

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.29

Infrared measuring technology

Why Infrared Measurements?

Infrared measuring instruments provide large advantages with regard to measuring tasks that cannot be solved with conventional contact thermometers. Examples:

• Measurements of very high temperatures not allowing the use of thermocouples.

• Measurements at surfaces with low thermal conduction and bodies with low thermal capacity.

• Measurements at moving, inaccessible or live parts with a high rate of response

(<1s).

• Measurements at objects, which must not be infl uenced by contact measurements.

What is Infrared Radiation?

Every substance with a temperature above absolute zero emits an infrared radiation

(spectral range of wavelengths from 0.7 to

1000µm) that corresponds to its temperature. This range is located below the longer red wavelength range and is not visible to the human eye. For measurements the most interesting range is located between

0.7 and 20µm.

The infrared radiation emitted by the test object follows the known optical rules and, therefore, can be deviated, bundled with lenses or refl ected from catoptric elements.

The emissivity of a test object indicates how much infrared energy has been absorbed or released by radiation. The value can be between 0 and 1.0. The fact that the emissivity depends on the wavelength is relevant for measurements. With increasing object temperature the radiation maximum shifts to the short wave range.

Therefore, IR thermometers are equipped with fi lters, which allow only one particular wavelength to pass through for the measurement. The spectral range for specifi c materials must be considered for the application.

How Infrared Thermometers Operate

The optical system of an infrared thermometer captures the energy emitted from a circular measuring spot and focuses it onto a detector. A material with a high transmission factor is used for the lenses. The energy captured by the detector is electronically amplifi ed and converted into an electrical signal. The optical resolution results from the ratio of the measuring distance to the size of the measuring spot.

The measuring spot must always be smaller than the test object or the measuring point of interest. The higher the optical resolution the smaller the measuring spots can be measured at further distances.

07.30

Infrared measuring technology

What is Intermittent Photometry?

Using intermittent photometry eliminates the thermal drift and immunes devices against thermal shock. The stability resulting from this, combined with noise-optimised signal processing, leads to an excellent temperature resolution and allows the measurement of smallest test objects and fast response times.

Special Infrared Pyrometers

Ratio Pyrometers determine the temperature from the ratio of the energy radiated in each of two wavelength ranges. This method allows for exact measuring results, even in case of a limited view to the test object due to vapour, steam, dust, dirty windows or lenses (up to 95% reduction of meas. signal). Furthermore, test objects, which are smaller than the measuring spot

(e.g. measurement at wires), or low or varying emissivities at fast moving objects, do not affect the measuring result.

Line Scanners measure the object temperature along a line. Fixed installed line scanners provide coloured heat flow charts from a product passing under the measuring head (e.g. conveyors, rotary furnaces), but can also be moved to pass above objects (e.g. heat flow chart of a house wall). The infrared scanner measuring head AMiR 7880 scans up to 256 dots over an angle of 90°. 20 lines can be scanned within one second. One measuring tape can be divided into 3 sectors, side by side or overlapping.

What You Should Consider For Infrared Measurements

What to do in case of dust, vapour and aerosols at the measuring point?

If the atmosphere at the measuring point is contaminated with dust, vapour and aerosols, the radiation energy impinging on the sensor can be influenced by contaminated lenses. This can be avoided by using an air blow attachment that keeps the lens clean.

What to do in case of high ambient temperatures?

If the ambient temperature exceeds the temperature specified for the measuring head of the IR sensor, the measuring head must be protected by mounting an air or water cooling system along with an air blow attachment (to avoid water condensing on the lens). Furthermore, cables and cable routings with high temperature stability must be used.

What to do in case of heat sources located next to the measuring object?

If heat sources are located next to the test object, these can transmit or reflect additional energy. Such ambience radiations occur, for example, at measurements in industrial furnaces where the wall temperature is often higher than the temperature of the test object. Many infrared instruments allow for a compensation of the ambient temperature.

What to do in case of measurements in a vacuum?

In case of vacuum furnaces and similar applications it is necessary to mount the measuring head outside of the vacuum area and to perform the measurement through a window. When selecting the measuring window the transmission values of the window must match the spectral sensitivity of the sensor. Quartz glass or quartz are typically used for high temperatures.

In case of low temperatures within the 8 to

14µm band the use of a special material, which is translucent for IR, is necessary, e.g. germanium, amtir, zinc selenide or sapphire. When selecting the window the temperature requirements, window thickness and pressure difference, as well as the possibility of keeping the window on both sides clean, must be considered. It might be advisable to consider an additional antireflective coating an the window on the window to increase the transmission capacity. Furthermore, it must be considered that not all window materials are translucent in the visible range.

Why is the emissivity so important?

In case of ideal radiators the reflected and transmitted energy equals zero and the emitted energy corresponds 100% to the characteristic temperature. However, many bodies emit less radiation at the same temperature (non-selective radiator). The ratio of real radiation value and that of the ideal radiator is defined as the emissivity ε. For example, a mirror has an emissivity of 0.1 while a so-called ´black body´ has an emissivity of 1.0. Many nonmetals such as wood, rubber, stone, and organic materials have only low reflecting surfaces and, as a result, high emissivities between 0.8 and 0.95. However, metals, especially if they have glossy surfaces, can have ε = 0.1. Therefore, IR thermometers provide an option for setting the emissivity. The emissivity should be known as exact as possible. If a too high emissivity has been set, the indicated temperature is lower than the actual temperature, given that the temperature of the test object is higher than the ambient temperature. For example, if 0.95 has been set, while the emissivity is actually only 0.9, a temperature that is lower than the actual temperature will be indicated.

How can the emissivity be determined?

Several methods can be used to determine the emissivity. As a first starting point, the following emissivity table can be consulted. The table data only represents average values, as the emissivity of a material is influenced by various factors. These include: temperature, angle of measurement, surface geometry (plane, concave, convex), thickness, surface quality (polished, rough, oxidised, sand-blasted), spectral range of the measurement and transmission capacity (e.g. in case of thin plastic foils)

07.31

Infrared measuring technology

Application Examples for Infrared Thermometers

Temperature

Range

Spectral

Sensitivity

Application

Examples

appr. 0 … 800°C 8 to 14 µm

3 to 5 µm

7 to 15 µm

7 to 18 µm

All non-metals, wood, paper, textiles, floor coverings, asphalt, lime floor, edibles, pharmaceuticals, as well as use with print, coating, laminating, drying/hardening, wave soldering and reflow soldering, for indoor installations, fire control, dust tips etc. appr. 10 … 360°C nominal 7.9 µm appr. 260 … 1650°C nominal 5.0/5.2 µm appr. 200 … 1200°C 3.9 µm appr. 30 … 340°C nominal 3.43 µm appr. 400 … 3000°C 2 to 2.7 µm appr. 200 … 1800°C 1.6 µm appr. 500 … 3000°C 1 µm

Fabrication and processing of polyester foil, fluoroplastics, fluoropolymer, acrylate, nylon (polyamide), acetylene cellulose, polyamides, polyurethanes,

PVC, polycarbonates.

Surface measurement on glass for heating up, forming, sealing, laminating, bending.

Metal finishing, furnaces, melting furnaces, blast furnaces, measurements on thick glass. Measurements slightly influenced by

CO

2

atmosphere (combustion gases).

Fabrication and processing of polyethylene, polypropylene, polystyrene and other foils.

Processing of ferrous and nonferrous metals, induction furnaces, glass production, melting furnaces, lab research.

Heat treatment of steel, bending, hardening, warming up.

Steel production, molten baths, for highest precision with shaping, casting and processing of metals, as well as the processing of glass, ceramics, semiconductors and chemicals.

Compact Glossary of Important Terms

Atmospheric Windows: The wavelength ranges within the infrared spectrum, in which the atmospheric radiation energy is transmitted and the atmospheric absorption is minimal, approximately 3 … 5µm and 8 … 14µm.

Focal Point, Focal Distance: Measuring distance where the maximum optical resolution is reached.

Far Field: Measured distance, which is significantly larger than the focal length of a device, in most cases is larger than ten times the focal length.

Field of View: The test object area, which is measured by the infrared thermometer; the diameter of the measuring spot is proportioned to the distance from the test object; often also specified as an angular variable at the focal point. Also see optical resolution.

Non-Selective Radiator: Radiating body with an emissivity that, for all wavelengths, bears the same constant ratio to the emissivity of a full radiator at the same temperature, which is opaque to radiation of infrared energy.

Background Temperature: From the view of the measuring instrument the ambient temperature or the temperature behind the test object.

Measuring Spot: Diameter of the test object area, which is subject to a temperature measurement; the measuring spot is defined by the circular area, which typically allows to capture 90% of the infrared energy radiating from the test object to the optical receiving aperture of the measuring instrument.

Optical Resolution: Also called the distance ratio: The ´measuring distance/measuring spot size´ ratio (distance ratio

E:M) of an IR measuring spot. The measuring distance is typically defined as the distance from the focal point and the measuring spot size as the diameter of the IR measuring spot measured at the focal point (typically the 90% energy measuring spot diameter). The optical resolution can be also defined for the far field, by using the values for the measuring distance and measuring spot size within the far field.

Degree of Reflection:

Full Radiator:

Spectral Sensitivity:

Ratio of the radiation energy reflected from a surface to the incident radiation of the same surface; for a perfect mirror the value is approximately 1, for a full radiator the reflection is zero.

Also: black body; ideal radiator. Body, which absorbs the whole impinging radiation energy of all wavelengths and which does not reflect nor transmit any radiation. The surface of a full radiator has a uniform emissivity of 1.

Wavelength range for which an infrared thermometer is sensitive.

07.32

Infrared measuring technology

oxidised electrolytic black cold-rolled heavy plate polished sheet metal melt steel oxidised stainless polished oxidised polished oxidised polished

(non-oxidised)

(non alkaline) pane molten mass non-oxidised oxidised oxidised etched polished polished etched oxidised oxidised non-oxidised rusty molten oxidised non-oxidised molten dull alloy oxidised sand-blasted electropolished polished etched oxidised polished high polished oxidised oxidised non-oxidised non-oxidised graphite

(any colour)

(translucent, over 0.5mm)

(cloth)

Emissivities of Various Materials Depending on the Spectral Range

Monel (Ni–Cu)

Nickel

Platinum

Mercury

Silver

Steel

Titanium

Tungsten

Zinc

Tin

Nonmetals

Asbestos

Asphalt

Basalt

Concrete

Ice

Soil

Paint

Gypsum

Glass

Rubber

Wood, natural

Limestone

Metals

Aluminium

Alloy A3003,

Lead

Chromium

Iron

Iron, cast

Iron, wrought

Gold

Haynes

Inconel

Copper

Magnesium

Brass

Molybdenum

Carborundum

Ceramics

Pebble stones

Carbon

Paper

Plastic

Fabric

Sand

Snow

Argil

Water

0.5

0.25

1 µm

0.9

0.65

0.3

0.8–0.9

0.2–0.4

0.04

0.8–0.9

0.35

0.35

0.8–0.9

0.35

0.5–0.75

0.35–0.4

0.6

0.4

0.8–0.95

0.8–0.9

0.35

0.9

0.3

0.5–0.9

0.4–0.9

0.3–0.4

0.2–0.5

0.05

0.05–0.2

0.2–0.8

0.3–0.8

0.8–0.95

0.6

0.5–0.9

0.25–0.35

1 µm

0.1–0.2

0.4

0.2–0.8

0.1–0.2

0.35

0.65

0.4

0.4–0.8

0.35

0.35

0.7–0.9

0.35

0.95

0.8–0.95

0.8–0.9

0.8–0.9

0.8–0.95

0.05

0.1–0.3

2.2 µm

0.8

0.9

0.2

0.4–0.9

0.2–0.6

0.4–0.7

0.1–0.2

0.95

0.05–0.15

0.02

0.6–0.7

0.2

0.25–0.4

0.8–0.9

0.2–0.9

0.2–0.5

0.6–0.8

0.1–0.3

0.15

0.3–0.4

0.95

0.01–0.1

0.6–0.9

0.6–0.9

0.3–0.6

0.25

0.03

0.05–0.2

0.7–0.9

0.05–0.2

0.01–0.05

0.4

0.6

0.4–0.9

0.1–0.3

2.2 µm

0.02–0.2

0.2–0.4

0.4

0.2–0.6

0.02–0.1

0.05–0.2

0.5

0.3–0.7

0.05–0.3

0.7–0.9

0.1–0.3

0.6–0.9

0.4–0.6

0.7–0.95

0.3

0.03

0.05

5.1 µm

0.9

0.95

0.7

0.9

0.4–0.97

0.98

0.9

0.9

0.9–0.95

0.4–0.98

0.1–0.5

0.3–0.6

0.1–0.15

0.9

0.05–0.15

0.02

0.8–0.9

0.5–0.7

0.1

0.1–0.2

0.7–0.9

0.15–0.8

0.1–0.3

0.5–0.7

0.05–0.25

0.1

0.2–0.3

0.9

0.01–0.1

0.3–0.8

0.6–0.9

0.3–0.6

0.15

0.03

0.05–0.15

0.5–0.8

0.03–0.15

0.01–0.05

0.3

0.5

0.3–0.7

0.1–0.15

5.1 µm

0.02–0.2

0.2–0.4

0.4

0.1–0.4

0.02–0.1

0.05–0.2

0.4

0.2–0.7

0.03–0.3

0.6–0.9

0.05–0.25

0.5–0.8

0.65–0.95

0.25

0.9

0.85–0.95

0.95

0.8–0.9

0.7–0.9

0.95

0.95

0.95

0.9

0.85–0.95

0.9

0.95

0.95

0.8–0.9

0.7–0.8

0.95

0.95

0.95

0.9

0.9

0.95

0.93

0.02

0.05

8–14 µm

0.95

0.95

0.7

0.95

0.98

0.9–0.98

0.9–0.95

0.8–0.95

0.85

0.95

0.9–0.95

0.98

0.1–0.14

0.2–0.5

0.05–0.15

0.9

0.05–0.15

0.02

0.7–0.9

0.4–0.6

0.1

0.7–0.9

0.1–0.8

0.05–0.2

0.5–0.6

0.03–0.1

0.1

0.2–0.3

0.9

0.01–0.1

0.3–0.8

0.7–0.95

0.3–0.6

0.15

0.03

0.05–0.1

0.4–0.8

0.02–0.1

0.01–0.05

0.3

0.5

0.2–0.6

0.1

8–14 µm

0.02–0.1

0.2–0.4

0.3

0.1–0.3

0.02–0.1

0.05–0.1

0.4

0.2–0.6

0.02–0.2

0.5–0.9

0.05–0.2

0.5–0.7

0.6–0.95

0.2

07.33

Infrared measuring technology

ALMEMO

®

D6

Digital infra-red sensor for measuring surface temperature FIAD43

Miniature probe head, integrated electronics, ALMEMO

®

D6 plug

Measuring Field

• Digital infra-red probe head with integrated signal processor

• All sensor characteristics and adjustment data are stored in the probe head itself.

• Digital transmission ensures that measured values are not affected by the sensor cable being moved, bent, or twisted.

• Surface temperature is measured over a wide range up to 600°C.

• Robust stainless steel housing, protection class IP65

• The probe head, thanks to its small dimensions, can be installed in cramped and restricted conditions.

• The probe head is threaded for quick and easy installation.

• The sensor cable in polyurethane (PUR) is suitable for industrial use and is resistant to oily, acidic, basic environments.

• The sensor can be connected directly via the cable’s

ALMEMO

®

D6 plug to any ALMEMO

®

device.

• One measuring channel is preprogrammed on leaving our factory - surface temperature (°C).

• Emissivity 0.95 are preprogrammed (on leaving our factory).

• This can be programmed from 0.1 to 1.0 at the current

ALMEMO

®

V6 devices via the device or via interface

(some only via interface).

• Transmittance 1.0 is preprogrammed (on leaving our factory).

Transmittance can be modified directly on the PC using USB adapter cable ZA1919AKUV. (see “General accessories for

ALMEMO

®

D6 sensors” page 04.05).

General features and accessories, ALMEMO

®

D6 sensors

see page 01.08

Options fitted at our factory

Air blower attachment OR7843LB

Deflecting mirror with integrated air blower attachment

OR7843US1

Air-cooled housing and T adapter including air hose, insulation, and air blower attachment

Length of air hose 0.8 meters

Length of air hose 2.8 meters

OR7843KL1

OR7843KL2

Deflecting mirror for air-cooled housing

Standard delivery

Infra-red probe head with cable and ALMEMO

®

D6 plug and 1 mounting nut

Cable length = 1 meter

Cable length = 3 meters

DAkkS or factory calibration KI9xxx temperature for digital sensor (see chapter Calibration certificates).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.34

OR7843US

Order no.

FIAD4332

FIAD4332L3

ALMEMO

®

D6

Infrared measuring technology

Technical data

Digital infra-red probe head (including A/D converter)

Temperature measuring range -40 to +600 °C

Spectral sensitivity 8 to 14 µm

Optical resolution (90 % energy) 10:1 with focal point lens attachment 1 mm at distance of 10 mm

Transmittance can be programmed to 0.75. (see below)

Accuracy

Reproducibility

Nominal conditions

Temperature coefficient

Temperature resolution

±1 % of meas. value or ±1 K (whichever value is higher) ±2 K for meas. values <20 °C

±0.5 % of measured value or ±0.5 K (whichever value is higher)

23 °C ±5 K, emissivity 1.0

±0.05 K / K or ±0.05 % of measured value / K (whichever value is higher)

0.1 K

Response time

Emissivity

Transmittance

Protection class

130 ms (90 %)

0.95 (preprogrammed on leaving our factory) This can be programmed from 0.1 to 1.0 at the current ALMEMO

®

V6 devices via the device (some only via interface).

1.0 (preprogrammed on leaving our factory) This can be programmed from 0.1 to 1.0 directly on the PC using USB adapter cable ZA1919AKUV. (please place a special order)

(see “General accessories for ALMEMO

®

D6 sensors”)

IP65 (NEMA 4) (National Electric Manufacturers Association)

Ambient temperature

Storage temperature

-10 to +120 °C with air-cooled housing -10 to +200 °C

-20 to +120 °C

Relative atmospheric humidity 10 to 95 % non-condensing

Housing Stainless steel

Dimensions

Weight

Connecting cable(s)

ALMEMO

®

D6 plug

Probe head Length 28 mm x Ø 14 mm Thread M12 x 1

Probe head 50 grams with 1-meter cable permanently fitted Polyurethane (PUR) with ALMEMO

®

D6 plug

For available lengths see variants.

Refresh time

Supply voltage

0.25 seconds for all channels

6 to 13 VDC

Current consumption 4 mA

Accessories

Measuring field with focal point lens attachment

Focal point lens attachment (cannot be used together with air blower attachment or air-cooled housing)

Transmittance 0.75 ZR7843CFL

Protective window (cannot be used together with air blower attachment or air-cooled housing)

Transmittance 0.75 ZR7843PW

Mounting bracket, rigid ZR7842H Mounting bracket, adjustable ZR7842JH

07.35

Infrared measuring technology

Compact infra-red probe head AMiR FIA 844 suitable for all ALMEMO

®

devices

• Compact inexpensive infra-red probe head for measuring surface temperature

• Other measuring ranges -20 to +500 °C

• High optical resolution Measuring spot 11.5 mm at distance 150 mm, in distant field 9:1

• Sturdy stainless steel housing

Protection IP65

• Quick and easy to install thanks to screw-fit housing

• Integrated electronics, cable permanently fitted

• Can be connected directly to the ALMEMO

®

device using an

ALMEMO

®

connector.

Accessories

Mounting bracket, rigid

Mounting bracket, adjustable

Air blower attachment Thread M18x1

Variants (including 2 mounting nuts):

ALMEMO

®

infra-red probe head Measuring range -20 to +500 °C with permanently fitted cable and ALMEMO

®

connector, Cable length = 1 meter

Same as above Cable length = 3 meters

Factory calibration KI9xxx temperature for sensor (see chapter Calibration certificates)

07.36

Order no.

ZR7844FB

ZR7844JB

ZR7844APM

FIA844

FIA844L3

Infrared measuring technology

Technical data

Temperature range

Spectral sensitivity

-20 to +500 °C

8 to 14 µm

Optical resolution (90 % energy) 13:1 (11.5 mm at 150 mm distance), distant field 9:1

Accuracy

Reproducibility

Nominal conditions

Temperature resolution

Response time

Emissivity

Voltage supply

Protection

Ambient temperature

Storage temperature

Relative atmospheric humidity

Housing

Dimensions

Connecting cable

Weight

±1.5 % of measured value or ±2 K (whichever value is higher)

±3.5 K for measured values <0 °C

±0.5 % of measured value or ±1 K (whichever value is higher)

23 °C ±5 K, Emissivity 0.95

0.1 K

150 ms (95 %)

0.95, fixed setting via ALMEMO® connector (12 VDC)

IP65

0 to +70 °C

-20 to +85 °C

10 to 95 % non-condensing

Stainless steel

Length 94 mm Thread M18x1 permanently fitted, 1 or 3 meters, -30 to +105 °C including ALMEMO

®

connector, programmed approx. 160 g (1-meter cable)

Mounting bracket

Order no. ZR7844FB

Mounting bracket, adjustable

Order no. ZR7844JB

Air blower attachment

Thread M18x1

Order no. ZR7844APM

07.37

Infrared measuring technology

Infra-red transmitter for measuring surface temperature AMiR 7843

Miniature probe head, transmitter box with display / operating controls, with analog output

• Surface temperature is measured over a wide range up to 600

/ 1000 °C.

• The probe head, thanks to its small dimensions, can be installed in cramped and restricted conditions.

• Robust stainless-steel housing, protective class IP65

• The probe head is threaded for quick and easy installation.

• The sensor cable is suitable for industrial use and is resistant to oily, acidic, and alkaline environments.

• Transmitter box with display and operating controls

• Analog output 10 V / 20 mA, freely selectable and scalable.

!

Infra-red sensor suitable for direct connection to

ALMEMO

®

measuring instruments see Digital sensor

FIAD43x with ALMEMO

®

D6 plug (see page 01.08)

Accessories MR7843 series

Mounting bracket, rigid

Mounting bracket, adjustable

Protective window (cannot be used together with air blower attachment or air-cooled housing)

ZR7842H

ZR7842JH

ZR7843PW

Order no.

Focal point lens attachment (cannot be used together with air blower attachment or air-cooled housing) ZR7843CFL

10:1 optics Measuring spot diameter 1 mm at distance of 10 mm

22:1 optics Measuring spot diameter 0.5 mm at distance of 10 mm.

Accessories for MR7843-12 / -32 / -42

Air blower attachment

Air-cooled housing and T branch, including 0.8-meter air hose, insulation, and air blower attachment

Same as above but with 2.8-meter air hose

ZR7842LB

ZR7842KL1

ZR7842KL2

Order no.

90° defl ecting mirror

(only for air-cooled housing and air blower attachment) ZR7842US

90° defl ecting mirror with integrated air blower attachment

ZR7842US1

Options for MR7843-12 / -32 / -42

Factory test certifi cate

(only with delivery of new devices) OR7843KZ1

Order no.

DAkkSDKD or factory calibration KI9xxx, temperature, for sensors (see chapter „Calibration certifi cates“). DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

Standard delivery

Probe head (including mounting nut) with cable, PUR, mounted on transmitter box

Temperature range Optical resolution Ambient temperature, probe head

-40 to 600°C

-40 to 600°C

2:1

10:1

-10 to 120°C

-10 to 120°C

0 to 1000°C 22:1 -10 to 120°C

Order no.

Probe head cable, 1 m

MR784312

MR784332

MR784342

* Available on request longer probe head cable, 8 / 15 / 30 meters

Order no.

Probe head cable, 3 m*

MR784312L03

MR784332L03

MR784342L03

Options for MR7843-33 / -43

Air blower attachment, only fi tted at our factory

90° defl ecting mirror

(only with air blower attachment OR7843LB1)

OR7843LB1

OR7843KZ1

Order no.

Factory test certifi cate (only with delivery of new devices) OR7843KZ1

DAkkS or factory calibration KI9xxx, temperature, for sensors (see chapter „Calibration certifi cates“). DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

Standard delivery

Probe head (including mounting nut) with cable, fluoropolymer, mounted on transmitter box

Temperature range Optical resolution Ambient temperature, probe head

-40 to 600°C 10:1 -10 to 180°C

0 to 1000°C 22:1 -10 to 180°C

Order no.

Probe head cable, 1 m

MR784333

MR784343

* Available on request longer probe head cable 8 / 15 / 30 meters

Order no.

Probe head cable, 3 m*

MR784333L03

MR784343L03

07.38

Infrared measuring technology

Technical data

Probe head

Temperature measuring range

Spectral sensitivity 8 to 14 µm depending on type -40 to +600 °C or 0 to +1000 °C

Optical resolution (90 % energy)

Response time (90%) 130 ms

Accuracy

Reproducibility depending on type 2:1 / 10:1 / 22:1, typical (21:1 guaranteed)

±1 % of measured value or ±1 K (whichever value is higher) ±2 K for measured values <20 °C

±0.5 % of measured value or ±0.5 K (whichever value is higher)

Nominal conditions at ambient temperature +23 °C ±5 K, Emissivity factor 1.0 and calibration geometry

Temperature coefficient ±0.05 K / K or ±0.05 % of measured value / K (whichever value is higher)

Ambient temperature

Protective class depending on type -10 to +120 °C (with air cooling up to +200 °C) or -10 to +180 °C

IP65 (NEMA-4) / IEC 60529

Relative humidity

Housing

Dimensions

Probe head cable

10 to 95 % non-condensing

Stainless steel

L = 28 mm, Ø = 14 mm, Thread M12x1 depending on type polyurethane (PUR) or fluoropolymer

Weight 50 g (with 1-meter cable)

Transmitter box

Output (selectable) 0 to 5 V / 0 to 10 V; 0 to 20 mA / 4 to 20 mA (Temperature range can be programmed in each case.)

Thermocouple, type J, K, R, S Not electrically isolated from supply voltage

Temperature resolution ±0.1 K for temperature range < 500 °C

Accuracy ±1 K for output mA / V ±1.5 K for output, thermocouple

Temperature coefficient ±0.02 K / K for output mA / V, ±0.05 K / K for output, thermocouple

Emissivity

Transmittance

0.100 to 1.100

0.100 to 1.000

Signal processing

Alarm output

Power supply

Ambient temperature

Saving of maximum / minimum / average value retention period up to 998 seconds zero-potential contact (semiconductor relays) 48 V / 300 mA

8 to 32 VDC, maximum 6 W

-10 to +65 °C

Protective class

Relative humidity

Housing

Dimensions

Weight

IP65 (NEMA-4) / IEC 60529

10 to 95 % non-condensing

Zinc die casting

80 x 60 x 31.5 mm (LxWxH)

370 g

Dimensions

Measuring field

( 90% energy )

07.39

Infrared measuring technology

Infrared Measuring Heads in Two-Wire Design AMiR 7838

• Compact, robust and precise infrared measuring heads.

• Wide range of versions for applications in intelligent process control and monitoring systems, as well as in production and test lab.

• Low cost standard version with fixed set temperature and output current range and emissivity can be manually set at the measuring head.

• The standard version without programming functions is ideally suitable for connecting to ALMEMO

®

devices.

• Measuring heads also available as addressable and remotely programmable versions.

Accessories

ALMEMO

®

connecting cable, 2 meters, ALMEMO

®

connector, programmed for the probe head’s temperature range,

Sensor supply via ALMEMO

®

device (use of the device mains unit is recommended)

(cable not suitable for ALMEMO

®

4490-2, available here on request) for programmable measuring heads MR7838xP

Protective window, snap-on, according to above lens detail

Remote control set incl. HART adapter and software

Industrial mains adapter 110/220V – 24VDC

Order no.

ZA7838AK

ZR7838SF

OR7838SH

ZR7838NT

Options

Other focus point optics (also see page 07.44 / 07.45)

Water/air cooling housing including air blow attachment, factory mounted

Inherent safety (Ex in IIC T4), only available with programmable meas. heads without cooling jacket

Factory test certificate, based on DAkkS/NIST certified sensors (only with delivery of new devices)

OR7838KL

OR7838IS4

OR7800KZ1

Types

(incl. rigid mounting angle and fastening screw)

For universal applications, standard optics OR7838OS1 (Fresnel Lens)

Meas. range –18 to 500°C, spectral range 8 to 14 µm, response time 165ms, optical resolution 15:1

Order no.

MR783810(P)

For universal applications, standard optics OR7838OS3 (Amtir Lens)

Meas. range –18 to 500°C, spectral range 8 to 14 µm, response time 165ms, optical resolution 33:1

For high temperature measurements in metal finishing and in rotary tubular kilns, standard optics OR7838OS3 (Sapphire Lens)

Meas. range 200 to 1000°C, spectral range 3.9 µm, response time 165ms, optical resolution 33:1

For maximum temperature measurements in metal finishing, standard optics OR7838OS6 (Float Glass Lens)

Meas. range 500 to 2000°C, spectral range 2.2 µm, response time 100ms, optical resolution 60:1

MR783811(P)

MR783821(P)

MR783851(P)

For high temperature measurements in glass production and at heating up and hardening, standard optics OR7838OS3 (Calcium Fluoride Lens)

Meas. range 250 to 1650°C, spectral range 5.0 µm, response time 165ms, optical resolution 33:1

MR783831(P)

For low temperature measurements in the production of plastic foils and normal foils, standard optics OR7838OS3 (Calcium Fluoride Lens)

Meas. range 10 to 360°C, spectral range 7.9 µm, response time 165ms, optical resolution 33:1

MR783841(P)

(P) Measuring heads remotely programmable

DAkkS- oder Factory calibration KI9xxx temperature for sensor (see chapter Calibration certificates).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

07.40

Infrared measuring technology

Device Functions only AMiR 7838-xxP (programmable AMiR Heads)

Programming: through PC via HART

®

adapter (OR7838SH)

Emissivity: 0.10 to 1.00 programmable

Data functions: max, min, average value hold, compensation of ambience radiation

Limit value programming: 1 limit value incl. hysteresis, also usable for monitoring the temperature of the measuring head

ALMEMO® application: To acquire and save measured values using those measuring head variants which cannot be addressed and remotely programmed we recommend our ALMEMO

For other ALMEMO

®

®

4390-2 panel meters.

devices please see Chapter 01.Mesuring instruments

Technical Data

Accuracy: ±1% of meas. value or ±1.4°C, the higher value of either is always valid

Reproducibility:

Response time:

±0.5% of meas. value or ±0.7°C, the higher value of either is always valid

165ms, at 7838 - 51(P) 100ms

Nominal temperature: +23°C, ±5°C

Temperature resolution: AMiR 7838 -10, -11: 0.125°C, AMiR 7838 -21, -31, -41, -51: 1°C

Relative humidity:

Power supply:

Output signal:

Emissivity:

10 to 95%, non-condensing, at 30°C max.

12–24VDC, for AMiR 7838xxP: 24VDC

4 … 20mA linear, two-wire technology

0.10 to 1.00 manually adjustable at measuring head (only noprogrammable heads)

Operating temperature: without cooling: 0 to 70°C, with air cooling: 0 to 120°C with water cooling: 0 to 175°C, with protective housing: 0 to 315°C

Protection system:

Shock:

Vibration:

IP 65, (IEC 529)

IEC 68-2-27 (MIL STD 810D), 50G, each axis, 11ms

IEC 68-2-6 (MIL STD 810D), 3G, each axis, 11 to 200Hz

Dimensions:

Weight: without water cooling housing: 187mm long, Ø 42mm with water cooling housing: 187mm long, Ø 60mm without water cooling housing: 330 g with water cooling housing: 595 g

07.41

Infrared measuring technology

Digital Signal Processing and Configuration

HART

®

protocol: The Hart

®

protocol (´Highway Accessible Remote Transducer Protocol´) is one of the most popular intelligent field bus protocols. It is more often used in industry than any other protocol and is supported by a large number of products and software of other manufacturers. The Hart

®

signal combines the standard output of 4 to 20mA with a simultaneously running digital remote data transmission. As a result, the measuring heads can, additionally, digitally communicate through the 2-conductor current loop (4 to 20mA) with the measuring computer.

Single installation:

Parallel working:

The most frequently used installation method is the single current loop. Analog displays and controls, recorders or measuring equipment within the current loop will not be influenced by digital signals in the current loop.

Up to 15 measuring heads can be switched in parallel and the measured values can be digitally further processed. For evaluation a powerful software with a menu-driven and user-friendly interface is available. It allows a graphical display of the ONLINE data including storing the measured values as an ASCII file for data export to other applications.

Configuration examples: Single installation Parallel working.

Measuring Field Diagrams: AMiR 7838-10(P)

meter [mm]

Distance Sensor - Test Object [mm]

*Focus Point E : M = 15 : 1 Distant Field E : M = 14 : 1

Standard Optics OS1 meter [mm]

Distance Sensor - Test Object [mm]

*Focus Point E : M = 7 : 1 Distant Field E : M = 4 : 1

Focal Point Optics OS2

Order no. OR7838OS2

07.42

Infrared measuring technology

Measuring Field Diagrams: AMiR 7838-11(P)/-21(P)/-31(P)/-41(P)

meter [mm]

Distance Sensor - Test Object [mm]

*Focus Point E : M = 33 : 1 Distant Field E : M = 30 : 1

Standard Optics OS3 meter [mm]

Distance Sensor - Test Object [mm]

*Focus Point E : M = 30 : 1 Distant Field E : M = 5 : 1

Focal Point Optics OS4

Order no. OR7838OS4

!

The devices AMiR 7838-31(P) and AMiR 7838-41(P) are only available with standard optics OS3.

meter [mm]

Distance Sensor - Test Object [mm]

*Focus Point E : M = 32 : 1 Distant Field E : M = 10 : 1

Focal Point Optics OS5

Order no. OR7838OS5

Measuring Field Diagrams: AMiR 7838-51(P)

meter [mm]

Distance Sensor - Test Object [mm]

*Focus Point E : M = 60 : 1 Distant Field E : M = 42 : 1

Standard Optics OS6 meter [mm]

Distance Sensor - Test Object [mm]

*Focus Point E : M = 60 : 1 Distant Field E : M = 7 : 1

Focal Point Optics OS7

Order no. OR7838OS7

meter [mm]

Distance Sensor - Test Object [mm]

*Focus Point E : M = 60 : 1 Distant Field E : M = 14 : 1

Focal Point Optics OS8

Order no. OR7838OS8

07.43

Infrared measuring technology

Accessories for All Measuring Heads AMiR 7838, 7845, 7850

Without Use of the Thermojacket Protective Housing

Tube Adapter adjustable tube adapter

Measuring Head

Angle Bracket

Sighting Aid

Measuring Head with Water or Air Cooled Housing adjustable angle bracket

90° Deviation Mirror

Air blow Attachment

Fastening

Screw

Accessories

Rigid mounting angle (spare)

Adjustable mounting angle

Fastening screw (spare)

Sighting aid, screw-on

90° deviation mirror

Air blow attachment

Tube adapter onto 11/2” NPT

07.44

Order no.

ZR7838H

ZR7838JH

ZR7838BM

ZR7838VS

ZR7838US

ZR7838LB

ZR7838RA

Infrared measuring technology

Accessories for All Measuring Heads AMiR 7838, 7845, 7850

With Use of the Thermojacket Protective Housing

ThermoJacket

Sealing Cap

Connecting Flange with Stopper

End Adapter

Smart Version

Mounting Flange with Viewing Tube

Intermediate Ring

ThermoJacket Housing

Front Adapter adjustable mounting stand adjustable mounting flange adjustable tube adapter ceramic viewing tube special steel viewing tube

Accessories

Thermojacket protective housing (3.26kg)

Adjustable mounting stand

Adjustable mounting flange

Mounting flange for anti-reflective tube

30cm anti-reflective tube, special steel

30cm anti-reflective tube, ceramics

Adjustable tube adapter

Connecting flange with stopper and Amtir window (from 3.9 to 14 mm)

Connecting flange with stopper and quartz window (from 1 to 2.2 mm)

Water quantity regulator

Air quantity/pressure regulator

Order no.

ZR7838SH

ZR7838MF

ZR7838JM

ZR7838FR

ZR7838RE

ZR7838RK

ZR7838JR

ZR7838SA

ZR7838SQ

ZR7838WR

ZR7838LR

07.45

07.46

Air humidity

Content

Air Humidity

Digital sensor for temperature, humidity, and atm. pressure FHAD46-x

Digital sensor for temperature, humidity, and atm. pressure FHAD46-4AG in protective all-weather housing

Digital sensor for temperature, humidity, and atm. pressure FHAD 46-4x

Digital sensor for temperature, humidity, and atm. pressure FHAD46-2

Digital sensor for temperature, humidity, and atm. pressure FHAD 46-0

High-precision sensor for temperature, humidity, atm. pressure FHAD 36 Rx

High-precision sensor for temperature, humidity, atm. pressure FHAD 36 RS

High-precision sensor for temperature, humidity, atm. pressure FHAD 36 RIC

High-precision sensor for temperature, humidity, atm. pressure FHAD 36 RHK

Capacitive humidity sensor FHA 646 R, miniature sensor

Digital sensor for measuring temperature and humidity FHAD 46-7 compact screw fit sensor

ALMEMO

®

dewpoint sensor FHA 646 DTC1, dewpoint transmitter MT 8716 DTC1

Digital psychrometers, FNAD 46 series

Psychrometer FPA 8363

Transmitter in wall-mounted housing MA 8646

Digital temperature / humidity transmitter MH8D46

08.06

08.06

08.07

08.08

08.09

08.10

08.11

08.12

08.02

08.04

08.04

08.05

08.13

08.14

08.16

08.17

08.18

08.01

Air humidity

The Right Humidity Sensor for Any Measuring Task

• For humidity measurements various methods are used that differ from each other mainly with regard to their accuracy and their suitability for long term measurements and the substance used for the measurement:

• Capacitive Air Humidity Measurement,

• Psychrometric Air Humidity Measurement,

• Hygrometric Air Humidity Measurement,

• Dielectric Measurement of Moisture in

Materials,

• Measurement of the Moisture in Materials According to the Principle of

Conductivity,

• Dew Point Determination with CCC

Dew Point Probes,

• Dew Point Determination with Dew

Point Mirrors.

Capacitive Air Humidity Measurement

Capacitive sensors contain a glass substrate with a moisture sensitive polymer layer between two metal layers. By absorption of water, corresponding to the relative humidity, the dielectric constant and, as a result, the capacity of the thin-fi lm capacitor are changing. The measuring signal is directly proportional to the relative humidity and does not depend on the atmospheric pressure.

Advantage:

• maintenance-free measurement over longer periods,

• can withstand temperatures below 0°C

• atm. pressure-independent, works when pressure is applied

• fl exible use of the sensor

Disadvantage:

• limited long term stability

• sensitive to dewing and certain aggressive substances

Psychrometric Air Humidity Measurement

Psychrometers are precision devices containing a dry and a moistened temperature sensor. As a result of the evaporation the humidity sensor cools down, with a wind velocity of a minimum of 2m/s being required for the cool down process. The humidity values are calculated from the temperature difference (psychrometric difference). The calculation formulae for AL-

MEMO

®

devices correspond to those used by the German Weather Authority related to 1013mbar. Differences regarding to the atmospheric pressure can be corrected to achieve precise measurements.

Advantage:

• no ageing of the sensor -

• exception: contamination of the wick

• high accuracy

• high quality regarding the measuring technology

• usable without problems up to 100% r.H. in all substances

Disadvantage:

• long term measurement limited by the required water reserve and wick maintenance

• diffi cult to use with temperatures below

0°C and with low humidities

• depending on the atmospheric pressure

Hygrometric Air Humidity Measurement

Hygrometric sensors are equipped with a measuring strip, which lengthens or tightens depending on the humidity. The measuring strip consists of many single fi bers (measuring harp), which are made from organic or synthetic material.

Advantage:

• simple and low cost measuring technology, also usable for contaminated environments

• easy to clean

Disadvantage:

• limited accuracy

• limited measuring range

• slow measurement

Dielectric Measurement of Moisture in Materials

The measurement of the moisture in materials is performed indirectly via the determination of the dielectric constant. This is performed by using a capacity measurement via a high-frequency electrical fi eld,

08.02

which penetrates the material without disturbances.

Advantage:

• simple and fast measuring technology

• non-destructive contact measurement

• long term use is possible

Disadvantage:

• limited accuracy

Air humidity Air humidity

Measurement of the Moisture in Materials according to the Principle of Conductivity

The measurement of the moisture in materials is performed indirectly via the determination of the electrical resistance, which depends on the moisture content of the material.

Advantage:

• simple and fast measuring technology

Disadvantage:

• limited accuracy

• probe insertions

• only for short term control measurements

• measured values depend on various material parametersMaterialparametern

Dew Point Determination with CCC Dew Point Probes

The dew point sensor is equipped with an integrated sensor chip (CCC dew point principle according to Heinze), which is mounted on a cooling element. The sensor unit is also connected to a control circuit that regulates the operating current of the cooling element so that a defined condensate is established. The resulting dew point temperature will be directly measured within the sensor and can be output in a format, which allows for an evaluation.

Advantage:

• high accuracy, reliability and reproducibility

• wide measuring range

Disadvantage:

• high-sophisticated measuring method

• not suitable for quick control measurements

• cannot be used at temperatures below

0°C

Dew Point Determination with Dew Point Mirrors

An optically monitored mirror is mounted on a cascaded Peltier element. The sensor unit is also connected to a control circuit that regulates the operating current of the cooling element so that a defined condensate is established. The dew point temperature will be directly measured within the sensor and can be output in a format, which allows for an evaluation.

Advantage:

• high accuracy, reliability and reproducibility

• independent from atmospheric pressure

• wide measuring range

• suitable for temperatures below 0°C

Disadvantage:

• high sophisticated measuring method

• high current consumption

• risk of contamination

Small Glossary for Humidity/Moisture Measurement Variables

Absolute Humidity

Enthalpy

The absolute humidity indicates the weight of the water vapour contained in one m

3

of a mixture of air and water vapour.

The enthalpy indicates how much heat is stored within the humid air. This value is important for calculating the cooling and heating performance, e.g. when checking heat exchangers.

Mixture Ratio

Relative Humidity

The absolute humidity related to 1kg dry air.

The relative humidity indicates the percentage of air, which is saturated with water vapour, i.e. how much percent of the maximum possible amount of water vapour is currently contained in the air.

Owing to the dependence on temperature the relat. humidity can only ever be indicated for one specific temperature.

Saturation Vap. Pressure Air can only ever contain a certain maximum amount of water vapour. This is called the saturation vapour pressure, specified as g water vapour per kg of humid air. The saturation vapour pressure strongly depends on the air temperature. At low temperatures it will be low and at high temperatures it will be high. Therefore, warm air can accept large amounts of vapour pressure and cold air

Dew Point only small amounts.

The dew point is the temperature where the relative humidity equals 100%. If the dew point is not reached the water vapour will start condensing.

Water Vap. Partial Press. The total pressure in the room determined by the water vapour.

08.03

Air humidity

Digital sensor for temperature, humidity, and atmospheric pressure FHAD46-x

ALMEMO

®

D6

Example:

ALMEMO

®

D6-sensor

FHAD 46-41

Digital sensor for temperature, humidity, and atmospheric pressure FHAD46-x, with ALMEMO

®

D6 plug with integrated atmospheric pressure sensor for automatic pressure compensation

Common technical features FHAD 46x

• Digital capacitive humidity sensor with integrated signal processor

• All sensor characteristics and adjustment data are saved in the humidity sensor element itself.

• Humidity sensor element, plug-in : Spare elements are inexpensive; a replacement can be fitted on site quickly and easily by virtually anyone; it will be fully accurate straight away needing no special adjustment.

• new:

A digital atmospheric pressure sensor integrated in the

ALMEMO

®

D6 plug itself provides automatic pressure compensation for all pressure-dependent humidity variables.

• All relevant ambient parameters are measured with just one sensor.

• new:

Humidity calculation on the basis of formulae as per

Dr. Sonntag and the enhancement factor as per W. Bögel (correction factor fw(t,p) for real mixed gas systems)

This substantially widens the measuring range and improves the accuracy of humidity variable calculations.

• new:

Humidity variable : Absolute humidity in g/m³

• The humidity variables are calculated from the three primary measuring channels (real measurable variables). temperature, relative humidity, atmospheric pressure

• Freely selectable measurable variables

Four measuring channels are programmed (at our factory). temperature (°C, T, t), relative humidity (%H, RH, Uw), dewpoint (°C, DT, td), atmospheric pressure (mbar, AP, p)

Other humidity variables can also be selected. mixture (g/kg, MH, r), absolute humidity (g/m³, AH, dv), vapor pressure (mbar, VP, e), enthalpy (kJ/kg, En, h)

This device can be configured on a PC using USB adapter cable ZA1919AKUV. (see page 04.05).

Common technical data FHAD 46x

Digital temperature / humidity sensor (including A/D converter)

Operative range depending on sensor type

Humidity

Measuring range

Sensor

Accuracy

0 to 100 % RH

CMOSens

®

technology

±1.8 % RH in range 10 to 90 % RH at nominal temperature

Hysteresis

Nominal temperature typical ±1 % RH

Sensor operating pressure Atmospheric pressure

Response time T

63

+25 °C typical 8 seconds at +25 °C, 1 m/s

(without filter)

Accuracy

Reproducibility

Response time T

63

±0.3 K at +25 °C

±0.4 K at +10 to +40 °C

±1.3 K at -20 to +80 °C typical ±0.1 K typical 20 seconds (without filter)

ALMEMO

®

connecting cable

PVC; Length (see variants) with ALMEMO

®

D6 plug

Digital atm. pressure sensor (integrated in ALMEMO

Measuring range

Accuracy

700 to 1100 mbar

®

±2.5 mbar (at 0 to +65 °C)

D6 plug)

Temperature

Sensor CMOSens

®

technology

ALMEMO

®

D6 plug

Refresh rate

Supply voltage

Current consumption

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

2 seconds for all four channels

6 to 13 VDC

12 mA

DAkkS or factory calibration KH9xxx temperature, humidity for digital sensor (see chapter „Calibration certificates“).

Digital sensor for temperature, humidity, and atm. pressure FHAD46-4AG in protective all-weather housing cable length up to 100 meters with ALMEMO

®

D6 plug

Technical data and variants

(see chapter „Meteorology“)

08.04

ALMEMO

®

D6

Digital sensor for temperature, humidity, and atm. pressure FHAD 46-4x

Version in stainless steel, with filter cap with ALMEMO

®

D6 plug

Air humidity

General description and common technical data FHAD 46 x

Technical features

• Four measuring channels are programmed (at our factory).

- temperature (°C, T, t),

- relative humidity (%H, RH, Uw),

- dewpoint (°C, DT, td),

- atmospheric pressure (mbar, AP, p)

Technical data

Operative range

Mechanical design

Sensor tube

-20...+80 °C / 5...98 % RH

Stainless steel, diameter 12 mm

Length (see variants)

Filter cap

Screw-fit cable gland

Metal-mesh filter, SK7

Splash-protected

Variants

including manufacturer’s test certificate

Digital sensor for temperature, humidity, and atmospheric pressure, filter cap, stainless steel tube, with fitted cable and ALMEMO

®

D6 plug.

Sensor length 160 mm, Connecting cable, length 2 meters

Sensor length 160 mm, Connecting cable, length 5 meters

Sensor length 160 mm, Connecting cable, length 10 meters

Sensor length 270 mm, Connecting cable, length 2 meters

Sensor length 270 mm, Connecting cable, length 5 meters

Sensor length 270 mm, Connecting cable, length 10 meters

Sensor length 530 mm, Connecting cable, length 2 meters

Sensor length 530 mm, Connecting cable, length 5 meters

Sensor length 530 mm, Connecting cable, length 10 meters

Replacement sensor element, digital, adjusted, plug-in

Order no.

FHAD4641

FHAD4641L05

FHAD4641L10

FHAD4642

FHAD4642L05

FHAD4642L10

FHAD4643

FHAD4643L05

FHAD4643L10

FH0D46

Protective caps

Dimensions : length approx. 33 mm, diameter 12 mm

SK7

SK6

SK8

Designation

Metal-mesh filter in

PC-housing

PTFE-Sinterfilter

Stainless steel sinter filter

Pore size

100 µm

50 µm

10 µm

Accessories

Brackets for wall mounting, distance from wall approx. 40 mm

ZB9600W

SK7

max. temp.*

120°C

180°C

180°C

SK6

Typical Application

Universal, for medium, contamination, also high humidity

High chemical resistance

For severe mechanical stress, heavy contamination, strong air flow

Movable brass screw connection with plastic sealing ring

ZB9600KV20

Connecting flange for screw connection, hole circle 38 mm Ø

ZB9600F20

SK8

Order no.

ZB9600SK7

ZB9600SK6

ZB9600SK8

* Observe application range

08.05

Air humidity

Digital sensor for temperature, humidity, and atmospheric pressure FHAD46-2

Version in plastic, with slotted sensor cap with ALMEMO

®

D6 plug

ALMEMO

®

D6

FHAD462

Sensor element incorporated in slotted sensor cap compact design, short response time

FHAD462L00

FHAD462 Option with plug-in extension tube

• Four measuring channels are programmed (at our factory).

Temperature (°C, T, t), Relative humidity (%H, RH, Uw)

Dewpoint (°C, DT, td)

Atmospheric pressure (mbar, AP, p).

Technical data

Operative range

Mechanical design

Sensor cap

Plug connection

-20 to +60 °C / 5 to 98 % RH

Ø 8 mm, length 36 mm

Ø approx. 9 mm, IP40

Variants

including manufacturer’s test certificate

Digital sensor for temperature, atmospheric humidity, and atmospheric pressure, with sensor element in slotted sensor cap, plug connector, including ALMEMO

®

connecting cable with coupling and ALMEMO

®

D6 plug.

Connecting cable, length 2 meters

Connecting cable, length 5 meters

Connecting cable, length 10 meters

FHAD462

FHAD462L05

FHAD462L10

Other designs are available on request

Sensor with terminal box FHD 462 KL for wall mounting

Terminal box with plug-in digital temperature / humidity sensor, cable lengths up to 100 meters

Extension tube Ø 8 mm, length 97 mm

General description and common technical data see FHAD 46x

Cable stub approx. : 80 mm

(incl. sensor element)

Spare sensor element for FHAD462, digital, enclosed in slotted sensor cover, adjusted

Extension tube, Ø 8 mm, length 97 mm, plug-in, for FHAD462

Order no.

FHAD462L00

FH0D462

ZB0D462VR

Digital sensor for temperature, humidity, and atm. pressure FHAD 46-0

Uncovered sensor element with ALMEMO

®

D6 plug

FHAD460

Uncovered sensor element most compact design, short response time

• Four measuring channels are programmed (at our factory).

Temperature (°C, T, t), Relative humidity (%H, RH, Uw)

Technical data

Operative range

Mechanical design

-20 to +80 °C / 5 to 98 % RH

Dewpoint (°C, DT, td),

Atmospheric pressure (mbar, AP, p).

Sensor element (dimensions over all) approx. 6 x 14 x 3 mm

Plug connection Width approx. 7 mm

Variants

including manufacturer’s test certificate

Digital sensor for temperature, humidity, and atmospheric pressure, with uncovered sensor element, plug connector, including

ALMEMO

®

connecting cable with coupling and ALMEMO

®

D6 plug.

Connecting cable, length 2 meters

Connecting cable, length 5 meters

FHAD460

FHAD460L05

08.06

Connecting cable, length 10 meters

Replacement sensor element, digital, adjusted, plug-in

Order no.

FHAD460L10

FH0D46

ALMEMO

®

D6

Air humidity

High-precision sensor for temperature, humidity, atmospheric pressure FHAD 36 Rx

Wide operating temperature range Automatic atmospheric pressure compensation

Digital sensor with ALMEMO

®

D6 plug

ALMEMO

®

connecting cable with sensor

(example FHAD 36 RS)

General features,

ALMEMO

®

D6 sensors

see page 01.08

Common technical features FHAD 36 Rx

• Digital capacitive humidity sensor with integrated signal processor, designed to meet the highest accuracy requirements in humidity measurement

• Unique correction and adjustment process

All sensor characteristics and adjustment data are saved in the humidity sensor itself.

• new:

A digital atmospheric pressure sensor integrated in the ALMEMO

®

D6 plug itself provides automatic pressure compensation for all pressure-dependent humidity variables.

• new:

Humidity calculation on the basis of formulae as per

Dr. Sonntag and the enhancement factor as per W. Bögel

(correction factor fw(t,p) for real mixed gas systems)

This substantially widens the measuring range and improves the accuracy of humidity variable calculations.

• new:

Humidity variable, Absolute humidity in g/m³

• All relevant ambient parameters are measured with just one sensor.

• The humidity variables are calculated from the three primary measuring channels (real measurable variables). temperature, relative humidity, atmospheric pressure

• Freely selectable measurable variables

• Four measuring channels are programmed (at our factory). temperature (°C, T, t), relative humidity (%H, RH, Uw), dewpoint (°C, DT, td), atmospheric pressure (mbar, AP, p)

Other humidity variables can also be selected: mixture (g/kg, MH, r), absolute humidity (g/m³, AH, dv), vapor pressure (mbar, VP, e), enthalpy (kJ/kg, En, h)

• This device can be configured directly on a PC using USB adapter cable ZA 1919 AKUV. (see chapter „Networking“).

Common technical data FHAD 36 Rx

Digital temperature / humidity sensor (including A/D converter)

Operative range depending on sensor type

Humidity

Sensor

Measuring range

Adjusted

Accuracy

Reproducibility

Response time T

63 capacitive

0 to 100 % RH at +23 °C and 10%, 35%, 80% RH

±1.3 % RH (at +23°C ±5 K)

0.3 % RH

<15 seconds at typical 1 m/s

(without filter)

Temperature

Sensor

Measuring range

Accuracy at +23 °C ±5 K ±0.2 K

Reproducibility

Pt100 class A

-100 to +200 °C *

Please observe operative range !

(depending on sensor type)

0.05 °C

Sensor connector on the sensor / sensor cable

Plug connector (Materials : anticorodal aluminum, anodized) IP65

Operative range of the electronics

in the connecting cable (coupling) -40 to +90 °C

in the grip (of hand-held sensors) -40 to +85 °C

ALMEMO

®

connecting cable

Coupling (length = 100 mm) with cable, length = 2 or 5 meters

(Materials : TPU, -40 to +90 °C) with ALMEMO

®

D6 plug

Digital atm. pressure sensor (integrated in ALMEMO

®

D6 plug)

Measuring range

Accuracy

700 to 1100 mbar

±2.5 mbar (at 0 to +65 °C)

ALMEMO

®

D6 plug

Refresh rate

Supply voltage

Current consumption

1 second for all four channels

6 to 13 VDC

12 mA

* Persistent use in the high-temperature range (>170 °C) may incur a loss in accuracy and / or damage to the measuring cell.

DAkkS or factory calibration KH9xxx temperature, humidity for digital sensor (see chapter „Calibration certificates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

08.07

Air humidity

ALMEMO

®

D6

High-precision sensor for temperature, humidity, atmospheric pressure FHAD 36 RS

Automatic atmospheric pressure compensation. Digital sensor with ALMEMO

®

D6 plug

General description and common technical data

FHAD 36 Rx (see page 08.07)

Technical data

Operative range

Sensor materials

-50 to +100 °C

Polycarbonate

Accessorie

Brackets for wall mounting (see page 08.05)

Order no.

ZB9600W

Filter Polyethylene

Variants

Including factory test certificate and polyethylene filter

High-precision digital temperature / humidity sensor, with plug connector, including ALMEMO

®

connecting cable with coupling and ALMEMO

®

D6 plug, and integrated digital atmospheric pressure sensor

Connecting cable, length 2 meters

Same as above Connecting cable, length 5 meters

Order no.

FHAD36RS

FHAD36RSL05

Filters

Variants

Polycarbonate filter cartridge with a filter insert made from polyethylene for standard applications good response time and good protection against fine particulates

Polycarbonate filter cartridge with a filter insert made from stainless-steel wire fabric quickest response time not suitable for environments that are bioactive or contaminated with fine particulates (risk of congestion)

Polycarbonate filter cartridge with a filter insert made from PTFE (polytetrafluoroethylene) good protection against fine particulates and salt (maritime environment) slower response time

POM (polyoxymethylene) filter cartridge with a filter insert made from PTFE water-proof very good protection against fine particulates slow response time

08.08

Order no.

ZB9636PE

ZB9636WM

ZB9636TF

ZB9636FD2

ALMEMO

®

D6

Air humidity

High-precision sensor for temperature, humidity, atmospheric pressure FHAD 36 RIC

Industrial-standard design for high temperatures up to +200 °C*

Automatic atmospheric pressure compensation. Digital sensor with ALMEMO

®

D6 plug

General description and common technical data

FHAD 36 Rx (see page 08.07)

Sensor plug, high-temperature cable, sensor

Technical data

Operative range

Sensor materials

-100 to +200 °C *

Sensor length 100 mm

(Other lengths 250 / 400 / 550 / 700 mm are available on request.)

PPS (polyphenylene sulfide)

Accessories

Assembly screw fittings for 15 mm sensor Brass, nickel-plated

Thread M20x1.5 Viton

®

seal, up to +200 °C

ZB9636KV

Filter cartridge

Filter

Response time T

63

Brass, nickel-plated

Stainless-steel wire fabric filter

<10 seconds at typical 1 m/s, without filter

* Persistent use in the high-temperature range (>170 °C) may incur

a loss in accuracy and / or damage to the measuring cell.

Mounting flange Steel, nickel-plated Diameter 80 mm

Order no.

ZB9636F

Variants

Including factory test certificate and stainless-steel wire fabric filter

Order no.

High-precision digital temperature / humidity sensor, industry-standard, with high-temperature sensor cable and plug connector, including ALMEMO

®

connecting cable with coupling and ALMEMO

®

D6 plug

Integrated digital atmospheric pressure sensor

Sensor cable, length = 2 meters, Connecting cable, length 2 meters

Same as above Sensor cable, length = 5 meters, Connecting cable, length 2 meters

Same as above Sensor cable, length = 2 meters, Connecting cable, length 5 meters

Same as above Sensor cable, length = 5 meters, Connecting cable, length 5 meters

FHAD36RIC102

FHAD36RIC105

FHAD36RIC102L05

FHAD36RIC105L05

Filter

for sensors with filter cartridge for FHAD 36 RIC and FHAD 36 RHK

Variants

(up to 200°C)

Stainless-steel wire fabric filter quickest response time not suitable for environments that are bioactive or contaminated with fine particulates (risk of congestion)

Stainless-steel sinter filter best protection in environments heavily contaminated with particulates good response time for low humidities (not to be used for high humidities)

PTFE filter good protection against fine particulates and salt (maritime environment) slower response time

Order no.

ZB9636M15

ZB9636S15

ZB9636T15

Other designs are available on request

Industry-standard humidity sensor FHAD 36 RIM in stainless steel Diameter 15 mm, -100 to +200 °C*

Screw-fit humidity sensor FHAD 36 RIE, up to 100 bar, stainless steel Thread G 1/2-inch, -100 to +200 °C*

* Persistent use in the high-temperature range (>170 °C) may incur a loss in accuracy and / or damage to the measuring cell.

08.09

Air humidity

ALMEMO

®

D6

High-precision sensor for temperature, humidity, atmospheric pressure FHAD 36 RHK

Hand-held sensor for temperatures up to +200 °C*

Automatic atmospheric pressure compensation, Digital sensor with ALMEMO

®

D6 plug

For on-site test measurements, not for stationary installation

Technical data

Sensor plug, cable, sensor with handle

Operative range -100 to +150 / +200 °C* (see variants)

Operative range of the electronics in the grip -40 to +85 °C

Sensor materials

Grip

Shaft PPS (polyphenylene sulfide)

POM (polyoxymethylene)

General description and common technical data FHAD 36 Rx

(see page 08.07)

Filter cartridge

Filter

Response time T

63

Brass, nickel-plated

Stainless-steel wire fabric filter

<10 seconds at typical 1 m/s, without filter

* Persistent use in the high-temperature range (>170 °C) may incur

a loss in accuracy and / or damage to the measuring cell.

Variants

Including factory test certificate and stainless-steel wire fabric filter

Order no.

High-precision digital temperature / humidity sensor

Handle with 2-meter sensor cable and plug connector, including ALMEMO

®

connecting cable, length 0.3 meters, with coupling and ALMEMO

®

D6 plug Integrated digital atmospheric pressure sensor

Operative range up to +150 °C Sensor length 250 mm

Operative range up to +200 °C Sensor length 400 mm

FHAD36RHK25

FHAD36RHK40

Other designs are available on request

Miniature cable humidity sensor

Diameter 4 mm , -40 to +85 °C

Humidity probe with pointed tip, Diameter 5 / 10 mm for taking meas. in loose bulk materials, -40 to +85 °C

Humidity probe with flat blade 18 x 4 mm for taking meas. in paper or textile stacks, -40 to +85 °C

08.10

Air humidity

Capacitive humidity sensor FHA 646 R, miniature sensor

• Compact sensor, extremely small dimensions

• Wide operating temperature range

• Particularly suitable for measuring operations between PCBs,

Technical data

Operative range -30 to +100 °C, 5 to 98 % RH

Humidity measuring circuit

Measuring range

Sensor

Accuracy

Reproducibility

0 to 100 % RH capacitive

±2 % RH in the range <90 % RH at nominal temperature

<1% RH at nominal temperature

Nominal temperature +25 ±3 °C

Response time T63 approx. 10 seconds at 1 m/s inside cases, in walls, ceilings, and insulation layers used in the construction industry, and for the protection of listed historic monuments

Temperature measuring circuit

Sensor

Accuracy

Reproducibility

NTC type N

-20 to 0 ±0.4 °C, 0 to +70 ±0.1 °C

+70 to +100 ±0.6 °C

0.1 °C

Mechanical design

Sensor tube

Protective cap

Cable nickel-plated, 50 mm long, 5 mm Ø

None

High-temperature cable (up to +100 °C),

2 meters long, with ALMEMO

(no other lengths available)

®

plug

!

The sensor can only be operated by plugging DIRECTLY onto an ALMEMO

®

device.

(NOT with extension cables ZA9060VKx or ZA9090VKCx).

Or, alternatively, the following sensor types can be used. FHAD36RS up to +100 °C (see page 08.08)

FHAD462 or FHAD460 Compact design (see page 08.06)

Accessories

PTFE filter, inside diameter 5 mm suitable for protection against dust, not water-proof

Clamped screw connection with thread adapter for telescopic extension / extension set (maximum 80 °C)

Telescopic extension Ø 15 to 24 mm, 330 / 1010 mm

Extension set Ø 15 mm, 4 x 255 mm

Order no.

ZB9646SKR

ZV9915KV

ZV9915TV

ZV9915VR3

Variants

Miniature sensor for temperature / humidity, with fitted high-temperature cable, length 2 meters, with ALMEMO

®

plug

Order no.

FHA646R

DAkkS or factory calibration KH9xxx temperature, humidity for measuring chain (sensor + device) (see chapter „Calibration certificates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

08.11

Air humidity

Digital sensor for measuring temperature and humidity FHAD 46-7,

ALMEMO

®

D6

Pressure-sealed variant up to 16 bar, with ALMEMO

®

D6 plug

• Compact sensor made from stainless steel

• Screw thread, for pressure pipes

• Option - adapter for compressed air pipes

• Digital capacitive humidity sensor with integrated signal processor

• All sensor characteristics and adjustment data are saved in the humidity sensor element itself. Humidity sensor element, plug-in

• Spare elements are inexpensive; a replacement can be fitted on site quickly and easily by virtually anyone; it will be fully accurate straight away needing no special adjustment.

• The humidity variables are calculated from the two primary

Technical data

Operative range

Temperature

Sensor

Accuracy

Reproducibility

-20 to +80 °C, 5 to 98 % RH

Digital temperature / humidity sensor (including A/D converter)

Humidity

Measuring range

Sensor

Accuracy

0 to 100 % RH

CMOSens

®

technology

±1.8 % RH in range 10 to 90 % RH at nominal temperature

Hysteresis typical ±1 % RH

Nominal temperature +25 °C

Sensor operating pressure up to 16 bar

CMOSens

®

technology

±0.3 K at +25 °C

±0.4 K at +10 to +40 °C

±1.3 K at -20 to +80 °C typical ±0.1 K measuring channels (real measurable variables): temperature, relative humidity

• Three measuring channels are programmed: temperature (°C, T, t), relative humidity (%H, RH, Uw), dewpoint (°C, DT, td)

One further humidity variable can also be selected: mixture (g/kg, MH, r), absolute humidity (g/m³, AH, dv), vapor pressure (mbar, VP, e), enthalpy (kJ/kg, En, h)

The system pressure needed for automatic pressure compensation of pressure-dependent humidity variables and the channel configuration are entered directly on the PC using USB adapter cable ZA1919AKUV. (see page 04.05).

ALMEMO

®

connecting cable

PVC Length (see variants) with ALMEMO

®

D6 plug

ALMEMO

®

D6 plug

Refresh time

Supply voltage

2 seconds for all four channels

6 to 13 VDC

Current consumption 12 mA

Mechanical design

Sensor

Filter cap

Process connection

Stainless steel, diameter 12 mm

Overall length approx. 77 mm

PTFE sinter filter SK6

Male thread G 1/2-inch

Fitted length 48 mm, Width across flats 27

Screw-fit cable gland Splash-protected

Accessories

Adapter for compressed air pipes

PTFE sinter filter (spare ) (see page 08.08)

Stainless-steel sinter filter (see page 08.08)

Adapter for compressed air pipes

Order no.

ZB96467AP

ZB9600SK6

ZB9600SK8

Variants

Digitaler sensor for temperature and humidity, filter cap PTFE, pressure-sealed variant, with fitted cable and ALMEMO

®

D6 plug, manufacturer’s test certificate

Connecting cable, length 2 meters

Connecting cable, length 5 meters

Connecting cable, length 10 meters

Replacement sensor element, digital, adjusted, plug-in

DAkkS or factory calibration KH9xxx, temperature, humidity, for digital sensor (see chapter „Calibration certificates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

08.12

Order no.

FHAD467

FHAD467L05

FHAD467L10

FH0D46

Air humidity

ALMEMO

®

dewpoint sensor FHA 646 DTC1, dewpoint transmitter MT 8716 DTC1

• Especially suitable for monitoring pressurized systems

• Digital transfer of measured values to the ALMEMO

®

display device (avoids risk of inaccuracy on connecting lines or display section itself)

• High-level accuracy sustained down to -80 °C

• Quick response time

• Displayed variables temperature, relative humidity, dewpoint

• Process connection for high pressures (option, up to 350 bar).

Technical data

Measuring range

Measuring accuracy

-80 to +20°C dewpoint temperature (DT)

± 0.5 °C from -10 to +20 °C DT typical ±2 °C DT at -40 °C DT

Measuring channels (FHA646DTC1 only)

temperature

Relative humidity

Dewpoint

-20.0 to +70.0 °C

0 to 98.0 % RH

-80.0 to +20.0 °C (DT)

Operating temperature -20 to +70 °C

Process connection Screw thread G 1/2-inch, stainless steel

Protective cap

Pressure range

Storage temperature

Sintered stainless steel filter

-1 to +50 bar standard

-40 to +80 °C

Accessories

Order no.

Screw-on measuring chamber for connecting a dewpoint transmitter to compressed air pipes via a ball valve up to maximum 16 bar including perforated protective cap

ZB9646DTCK

Advantage high-speed measuring without waiting for installation.

Option

Dewpoint sensor for process pressure up to 350 bar

OA9646DTCP

FHA 646 DTC1

Output

Power supply

Connection

MT 8716 DTC1

Output

Power supply

Connection

Housing

Material

Protective class

ALMEMO

®

digital via ALMEMO

®

plug, approx. 5 mA

Cable, 1.5 meters, with ALMEMO

®

plug

4 to 20 mA / -80 to +20 °C (DT), 2 wires

10 to 30 VDC, load <500 ohms

Transmitter connector

Polycarbonate

IP65

Variants

including factory calibration certificate

ALMEMO

®

dewpoint sensor with connecting cable, 1.5 meters long, and ALMEMO

®

plug

Dewpoint transmitter with current output, including connector

Factory calibration KH93xx, dewpoint, for digital sensor (see chapter „Calibration certificates“)

Order no.

FHA646DTC1

MT8716DTC1

08.13

Air humidity

ALMEMO

®

D6

Digital psychrometers, FNAD 46 and FNAD 46-3 with ALMEMO

®

D6 plug with integrated atmospheric pressure sensor, for automatic pressure compensation

General features,

ALMEMO

®

D6 sensors

see page 01.08

• new:

A digital atmospheric pressure sensor integrated in the

ALMEMO

®

D6 plug itself provides automatic pressure compensation for all pressure-dependent humidity variables.

• new:

Humidity calculation on the basis of formulae as per Dr.

Sonntag and the enhancement factor as per W. Bögel (correction factor fw(t,p) for real mixed gas systems)

This substantially widens the measuring range and improves the accuracy of humidity variable calculations.

• new:

Humidity variable Absolute humidity in g/m³

• High-precision NTC sensors for dry temperature and wet temperature

• Temperatures are measured using a 24-bit A/D converter incorporated in the ALMEMO

®

D6 plug.

• The humidity variables are calculated from the three primary measuring channels (real measurable variables):

Dry temperature, wet temperature, atmospheric pressure

• Freely selectable measurable variables

Four measuring channels are programmed (at our factory): dry temperature (°C, TT, t), wet temperature (°C, HT, tw), relative humidity (%H, RH, Uw), atmospheric pressure

(mbar, AP, p)

• Other humidity variables can also be selected: dewpoint (°C, DT, td), mixture (g/kg, MH, r), absolute humidity (g/m³, AH, dv), vapor pressure (mbar, VP, e), enthalpy (kJ/kg, En, h)

This device can be configured directly on a PC using USB adapter cable ZA 1919 AKUV. (see chapter „Networking“ page 05.05).

Technical data, FNAD 46 and FNAD 46-3

Digital atmospheric pressure sensor

(integrated in ALMEMO

®

D6 plug)

Measuring range 700 to 1100 mbar

Accuracy ±2.5 mbar (at 0 to +65 °C)

A/D converter incorporated in ALMEMO® D6 plug

Inputs 2 NTC sensors

(clamped connection in plug)

Resolution 0.01 K

Linearization

Accuracy

Nominal temperature

Temperature drift: error-free computing method according to Galway Steinhart

(no approximations)

±0.05 K

23 °C ±2 K

0,004 %/K (40 ppm)

Calculated humidity variables Analytic equation

(not an approximation)

Refresh rate 0.4 seconds for all four channels

08.14

ALMEMO

®

D6

Hand-held digital psychrometer FNAD 46

Air humidity

Stationary digital psychrometer FNAD 46-3

For test measurements

General description and common technical data

FNAD 46 (see page 08.14)

Technical data

Operating temperature 0 to +60 °C (no ice)

Humidity measuring range 10 to 100% RH

Measuring system

Accuracy

Nominal conditions psychrometric

±1 % RH under nominal conditions

+25 °C ±3 K, 1013 mbar, 50 % RH

Temperature sensors

Accuracy

Ventilator power supply

Housing

Dimensions

Weight

Sensor connector

2 x NTC type N

±0,1 K at 0 to 60 °C via ALMEMO

Plastic

®

D6 plug

Ø 50 mm, length 245 mm approx. 300 g

Built-in plug

ALMEMO

®

connecting cable coupling, 1.5 meters, PVC cable with ALMEMO

®

D6 plug

Supply voltage 9 to 13 VDC

Current consumption 20 mA

Version optimized for long-term measuring operations

Automatic humidification of the wick after filling the water tank.

General description and common technical data

FNAD 46-3 (see page 08.14)

Technical data

Operating temperature 0 to +90 °C (no ice)

Humidity measuring range 10 to 100% RH

Measuring system

Accuracy

Nominal conditions psychrometric

±1 % RH under nominal conditions

+25 °C ±3 K, 1013 mbar, 50 % RH

Temperature sensors

Accuracy

Ventilator power supply

Housing

2 x NTC type N

±0,1 K at 0 to 70 °C,

±0,4 K at 70 to 90 °C

12 VDC via mains unit, cable approx.

1.5 meters (included in delivery)

Plastic PMMA

Dimensions

Weight

175 x 50 x 75 mm (LxWxH) approx. 890 g

ALMEMO

®

connecting cable Cable, FEP / silicone, 5 meters with ALMEMO

®

D6 plug

Supply voltage 6 to 13 VDC

Current consumption 4 mA

Accessories

Extension pipe, 200 mm long

Plastic suction hose, 300 mm long

Spare wicks (2 pieces)

Order no.

ZB9846VR

ZB9846PS

ZB9846ED

Accessories

Extension cable for mains units,

3-pin bayonet coupling, length 5 meters

Spare wicks (2 pieces)

Order no.

ZB5090VK05

ZB98462ED

Variants

Order no.

Hand-held digital psychrometer with NTC sensor

Hand-held psychrometer, connecting cable with ALMEMO

®

D6 plug, integrated digital atmospheric pressure sensor, water bottle, two wicks

FNAD46

Variants

Order no.

Digital psychrometer with NTC sensor

Psychrometer, fitted cable, with ALMEMO

®

D6 plug, integrated digital atmospheric pressure sensor, mains unit, water bottle, two wicks, carry case

FNAD463

DAkkS or factory calibration KH91xx, temperature, humidity, for digital sensor (see chapter „Calibration certificates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

08.15

Air humidity

Psychrometer FPA 8363

• Optimized version for long-term measuring operations

• Especially suitable for high temperatures

• Operative range 0 to 90 °C, 10 to 100% RH

• Possible variables dry temperature, relative humidity, dewpoint, mixture ratio, wet temperature, partial vapor pressure.

Technical data

Humidity

Measuring range

Measuring system

Accuracy

Nominal conditions

Temperature

Sensors

Accuracy

10 to 100% RH psychrometric

±1 % RH under nominal conditions

+25 ±3 °C, 1013 mbar, 50% RH

2 x Pt100

IEC 751, class B ALMEMO

®

adjusted

Electrical supply

Operating voltage

Current consumption

Mechanical design

Housing

Dimensions

Weight

Cable

12 VDC via mains plug

(cable approx. 2 meters) approx. 40 mA

Plastic PMMA

175 x 50 x 75 mm (LxWxH) approx. 890 g

FEP / silicone, 5 meters with ALMEMO

®

plug

2 cables, 2 plugs

Accessories

new:

ALMEMO

®

plug-in pressure probe for measuring barometric pressure 700 to 1100 mbar without pressure connection sleeve

(version with pressure connection sleeve) (see page 10.10)

Technical data (see page 10.10) including programming for automatic atmospheric pressure compensation (comment *P)

FDAD12SA

OA9000PK

Spare wicks (2 pieces)

Extension cable for mains units,

3-pin bayonet coupling, length 5 meters

Order no.

ZB98462ED

ZB5090VK05

Variants

(including mains plug, water bottle, two wicks) Psychrometer with 2 x Pt100 sensors, including connecting cable (two ALMEMO

®

plugs)

Order no.

FPA8363

DAkkS or factory calibration KH91xx, temperature, humidity, for sensor or measuring chain (sensor + device) (see chapter „Calibration certificates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

08.16

Air humidity

Transmitter in wall-mounted housing MA 8646 for capacitive ALMEMO

®

humidity sensor FHA 646

Transmitter MA8646-0 with plug-in sensor FHA6466

• Twin analog transmitters for capacitive ALMEMO

®

humidity sensors (not for dewpoint sensor FHA646DTC1 or digital

ALMEMO

®

D6 sensors)

• Humidity sensor, plug-in, can be exchanged as and when necessary.

Technical data

Operative range

Humidity measuring circuit

Measuring range

Sensor

Accuracy

Reproducibility

Nominal temperature capacitive

±2 % RH in the range <90 % RH at nominal temperature

1 % at nominal temperature

+23 ±3 °C

Transmitter, accuracy ±0.5 % RH

Temperature measuring circuit

Measuring range -50 to +125 °C

Sensor

Accuracy

Reproducibility

Transmitter, accuracy

Outputs

Resolution

Temperature drift

(see humidity sensor)

0 to 100 % relative humidity

(%RH, HRH, HcRH)

NTC type N

0 to +70 °C ±0.1 K

-20 to 0 °C ±0.4 K

+70 to +100 °C ±0.6 K

0.1 K

±0.1 K

2 x 0 to 10 V (load >100 kilohms)

12 bit (4000 digits)

±0.02 % / K

• Analog output range can be scaled on the sensor connector.

• For stationary measuring operations, housing suitable for wall-mounting

• Versions available for different supply voltages.

Nominal temperature

Option R3

Output range

Power supply

Option U

Option U0

Current consumption

+23 ±3 °C

2 x 0/4 to 20 mA (load <500 ohms)

Standard 0 to 100 % RH, -30 to +70 °C

Set to customer-specific needs before leaving our factory or programmed by the user in the sensor connector using

ALMEMO

®

device

Mains 230 V, + 10 to 15 %, 50 to 60 Hz

(Option U5 : 110 V)

10 to 30 VDC, electrically isolated

13 to 28 VDC, not electrically isolated approx. 30 mA (no load)

Connections Screw terminals

Cable bushing to the wall or through grommets at end

Housing

Protective class

Ambient conditions

Operating temperature

Storage temperature

Ambient humidity

Wall-mounted housing, plastic

123 x 68 x 49 mm

IP40

-10 to +60 °C

-30 to +70 °C

10 to 90 % RH non-condensing

Option

Analog output, 2 x 0 to 20 mA

Analog output, 2 x 4 to 20 mA

Other analog output range

PLEASE SPECIFY WHEN ORDERING !

Programming on the humidity sensor connector

OA8646R3

OA8646R4

OA9000PR

Supply voltage 13 to 28 VDC not electr. isolated

Supply voltage 0 to 30 VDC electr. isolated

Supply voltage 110 VAC, 50 - 60 Hz

Order no.

OA8646U0

OA8646U

OA8646U5

Humidity sensor

including manufacturer’s test certificate

Plug-in sensor, –20 to +60 °C

FHA6466

Stainless steel tube, with 1.5-meter cable, –20 to +80 °C

FHA646E1C

Miniature sensor, with 2-meter cable, -30 to +100 °C

FHA646R

Order no.

Advisory note

Dewpoint sensor FHA646DTC1 and digital ALMEMO

®

D6 sensors cannot be connected.

Variants

including manufacturer’s test certificate

Temperature / humidity transmitter in wall-mounted housing. Outputs 2 x 0 to 10 V

(equivalent to 0 to 100 % RH and -30 to +70 °C). Supply voltage 230 VAC including wall unit, without sensor

Order no.

MA86460

DAkkS or factory calibration KH9xxx, temperature, humidity, for measuring chain (sensor + transmitter) (see chapter „Calibration certificates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

08.17

Air humidity

Digital temperature / humidity transmitter MH8D46 with double analog output V or mA

Transmitter with open housing

• Digital sensor element

All key sensor characteristics, settings, and adjustment data are saved in the sensor element itself.

• Plug-in sensor element

Spare elements are inexpensive; a replacement can be fitted on site quickly and easily by virtually anyone; it will be fully accurate straight away needing no special adjustment.

• Digital transfer of measured values from the sensor element to the transmitter

• Factory or DAkkS calibration is performed on the sensor element alone.

Fully accurate - irrespective of connecting cable and transmitter

• Four climate variables can be measured:

Double analog output for temperature and one humidity variable relative humidity / dewpoint / mixture ratio

• Limit value relays available on request

• The transmitters can be configured via the internal display and the keypad.

• The analog output type (10 V or 20 mA) can be selected (via the keypad); the analog output range can be programmed.

• Display of measured value, channel, units, humidity range, analog start, analog end, and analog type

• The sensor tube can be connected either directly by plugging onto the transmitter itself or via a connecting cable.

• Suitable for conduit mounting or wall mounting

Technical data

Operative range Sensor -20 to +80 °C, 5 to 98 % RH

Electronics -10 to +60 °C, IP65

Humidity sensor

Measuring range

Sensor

0 to 100 % RH

CMOSens

®

technology

Fixed measuring period / output period approx. 3 seconds

Accuracy ±1.8 % RH in range 10 to 90 % RH at nominal temperature

Hysteresis typical ±1 % RH

Nominal temperature +25 °C

Sensor operating pressure Atmospheric pressure

Response time T

63 typical 8 seconds at +25 °C, 1 m/s

(without filter)

Temperature sensor

Sensor

Fixed measuring period / output period approx. 3 seconds

Accuracy

Reproducibility

Response time T

CMOSens

®

technology

±0.3 K at +25 °C

±0.4 K at +10 to +40 °C

±1.3 K at -20 to +80 °C typical ±0.1 K typical 20 seconds (without filter)

Outputs

Double analog output Digital-to-analog converter (DAC) electr. isol.

63

0 to 10 V, load >100 kilohms

0 to 20 mA, load <500 ohms

08.18

Output type

Resolution

Accuracy

Temperature drift

Time constant

Connection

Standard equipment

Display, internal

Operation, internal

0 to 10 V, 0 to 20 / 4 to 20 mA, selectable

16 bit

0.1 % of final value

10 ppm / K

100 μs

Cable, via screwless clamp connector, with cable bushing

Cable diameter 2 to 5 mm

Limit value relays available on request

2-row LCD 7 segments

4 1/2 and 5 characters 2 digits

16 segments

3 keys

Power supply

DC voltage 9 to 30 VDC

Current consumption 30 mA + 1.2·IOut

Connection Cable, via screwless clamp connector, with cable bushing

Cable diameter 2 to 5 mm

Mechanical design

Sensor tube

Protective cap

Stainless steel, diameter 12 mm

SK7, metal-mesh filter

Housing

Dimensions

Protective class

Die-cast aluminum, closed cover

100 x 100 x 60 mm (LxWxH)

IP65 (with sensor tube or connecting cable plugged in)

Display of measured values and programming

(housing open)

Air humidity

Measured value display, channel M0, temperature

Measured value display, channel

M1, humidity variable, e.g. relative humidity

Selecting the humidity variable, e.g. relative humidity, % RH

Selecting the analog output type, e.g. 4 to 20 mA

Programming the analog start Programming the analog end

Accessories

Order no.

Angle bracket for wall mounting

Rubber gasket (mat) for mounting the housing

ZB8D00W

directly on a conduit wall (immersion depth = sensor length + approx.

42 mm plug length)

ZB8D00GD

Movable brass screw with plastic sealing ring

(see page 08.05)

Connecting flange for screw connection, pitch circle diameter 38 mm (see page 08.05)

ZB9600KV20

ZB9600F20

Protective caps (see page 08.05)

Mains plug, 230 VAC, 12 VDC, 2.5 A

ZB1012NA9

Connecting cable between sensor tube and transmitter

Length = 2 meters

Same as above Length = 5 meters including protective cap SK7

Sensor length = 125 mm

ZH9D46VK02

ZH9D46VK05

Same as above Length = 10 meters

ZH9D46VK10

Spare sensor, complete Sensor element inside sensor tube

Same as above Sensor length = 265 mm

Same as above Sensor length = 525 mm

Replacement sensor element, digital, adjusted, plug-in

FH9D461K1

FH9D461K2

FH9D461K3

FH0D46

Variants

including manufacturer’s test certificate

Digital transmitter for temperature and humidity

with double analog output, 10 V or 20 mA (selectable via keypad), internal display, 3 keys, aluminum housing, IP65, with plug-in digital sensor, sensor length = 125 mm

Same as above Sensor length = 265 mm

Same as above Sensor length = 525 mm

DAkkS or factory calibration KH9xxx, temperature, humidity, for digital sensor (see chapter „Calibration certificates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

Order no.

MH8D461K1

MH8D461K2

MH8D461K3

08.19

08.20

Content

The right flow sensor for each and every measuring task

Correction factors ensure precise air speed measurement.

Air Speed for selected Pitot tubes

Digital vane anemometer FVAD 15 for air

Differential pressure and Pitot tube measurement

Measuring connector

Pitot Tubes for Differential Pressure Sensors FDA602

Digital thermoanemometer FVAD 35 THx

Thermoelectric flow sensor

09.02

09.03

09.03

09.04

09.06

09.07

09.08

09.09

Air flow

09.01

Air fl ow

The Right Flow Sensor For Any Measuring Task

For measuring the flow velocity, typically, three methods are used, which are particularly different from each other with regard to their measuring range and the operating temperature:

• Pitot tubes

• Rotating vanes

• Thermoanemometer probes

Pitot Tubes

The air velocity is determined by the dynamic pressure and the static pressure.

Pitot tubes are robust and are available in special steel or nickel-plated brass. They connect to ALMEMO® devices by silicone hoses and a differential pressure module.

Rotating Vanes

The flow velocity is determined through a frequency measurement. Our rotating vanes are sensitive transducers with diamond bearings that are very precisely adjusted. This ensures high accuracy.

Thermoanemometers

Thermistors and hot wire anemometers are highly sensitive sensors. The measuring element is continuously heated up. A control circuit keeps the temperature of the element, which has cooled down by the air flow, on a constant value. The control current is proportional to the flow velocity.

Advantage:

suitable for high flow velocities and harsh operating conditions, high ambient temperatures possible, easy to clean

Advantage:

high accuracy at medium flow velocities and medium ambient temperatures, insensitive to turbulent flows

Advantage:

even very small air speeds can be measured (e.g. draught measurements), direction-independent measurements are also possible

Disadvantage:

strongly directional, low flow velocities are not measurable, temperature-dependent, limited accuracy, sensitive to turbulent flows

Disadvantage:

sensitive sensor technology, sensitive to mechanical stress, directional

Disadvantage:

sensitive sensor technology, sensitive to mechanical stress and contamination, sensitive to turbulent flows, high current consumption, limited ambient temperature.

09.02

Air flow Air flow

Correction Factors for Exact Measurements of the Air Speed

Air Temperature

–30°C

–20°C

–10°C

0°C

10°C

20°C

30°C

40°C

50°C

60°C

70°C

80°C

90°C

100°C

150°C

200°C

250°C

300°C

400°C

500°C

600°C

700°C

1.069

1.085

1.102

1.118

1.135

1.151

1.167

1.242

1.314

1.381

1.446

1.567

940 mbar 960 mbar

0.942

0.961

0.980

0.998

0.932

0.951

0.970

0.988

1.016

1.035

1.051

1.005

1.024

1.040

1.68

1.784

1.884

1.663

1.766

1.865

1.057

1.074

1.09

1.106

1.123

1.139

1.154

1.229

1.300

1.367

1.431

1.55

The true air velocity depends on the air temperature and the barometric air pressure. Therefore, the measured value must be corrected according to the above table

1.047

1.063

1.079

1.095

1.111

1.127

1.142

1.216

1.287

1.353

1.416

1.534

980 mbar 1000 mbar 1020 mbar 1040 mbar

0.922

0.941

0.960

0.978

0.913

0.932

0.950

0.968

0.904

0.923

0.941

0.958

0.895

0.914

0.931

0.949

0.995

1.013

1.029

0.985

1.003

1.019

0.975

0.993

1.009

0.966

0.983

0.999

1.646

1.748

1.846

1.036

1.052

1.068

1.084

1.100

1.116

1.131

1.204

1.274

1.339

1.402

1.519

1.629

1.73

1.827

1.026

1.042

1.057

1.073

1.089

1.105

1.120

1.192

1.261

1.326

1.388

1.504

1.613

1.713

1.809

1.597

1.696

1.791

1.016

1.031

1.047

1.063

1.078

1.094

1.109

1.180

1.249

1.313

1.375

1.489

to obtain exact measurements of the air speed.

Example:

Measured air velocity 50m/s, air temperature 80°C, atmospheric pressure 960mbar.

The measured value must be multiplied with the correction value 1.123. The air velocity is, therefore, 56.1m/s.

Air Speed For Selected Dynamic Pressures

(Prandtl Pitot Tube, T = 22°C)

Dynamic Pressure [Pa]

1

2

3

4

5

10

20

30

40

50

100

Dyn. Press. [mm h.o.water]

0.1

0.2

0.3

0.41

0.51

1.02

2.04

3.06

4.08

5.1

10.2

Air Speed [m/s]

1.29

1.83

2.24

2.59

2.89

4.09

5.78

7.08

8.18

9.14

12.93

09.03

Air flow

ALMEMO

®

D6

Digital vane anemometer FVAD 15 for air, with ALMEMO

®

D6 plug

Technical data and functions, FVAD 15 series

• Measuring air flow velocity

• The vane anemometer is in practice unaffected by environmental variables such as pressure, temperature, density, or humidity.

• The design is compact - especially suitable for mobile measuring operations - heating, ventilating, air-conditioning.

• The probe head has an aero-dynamically optimized shape and protected bearings.

• On those variants with a snap-on head the probe head can be exchanged quickly and easily, e.g. for servicing.

• ALMEMO

®

D6 plug with high-resolution frequency measurement

• One measuring channel is programmed (at our factory).

Flow velocity (m/s, v).

General features, ALMEMO

®

D6 sensors

see page 01.08

Technical data FVAD15 series

Operative range

Maximum resolution

Nominal temperature

Connecting cables

ALMEMO

®

adapter cable

ALMEMO

®

D6 plug

Frequency measurement

Refresh rate

Averaging period

Supply voltage

Current consumption

-20 to +140 °C

0.01 m/s

+22 °C ±2 K

Fitted cable, 1.5 meters, with LEMO

®

plug

LEMO

®

coupling cable, 0.2 meters with ALMEMO

®

D6 plug resolution 0.01 Hz

0.5 seconds for all channels

2 seconds

6 to 13 VDC

4.5 mA

Accessories

Extension set Ø 15 mm, 4 x 255 mm

Telescopic extension Ø 15 to 24 mm, 330 / 1010 mm

DAkkS or factory calibration KV90xx air flow for digital sensor (see chapter „Calibration certificates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

Order no.

ZV9915VR3

ZV9915TV

Digital vane anemometer FVAD 15 S120/S140 with snap-on head, mini

Accessories

Spare snap-on head, mini, 20 m/s

Spare snap-on head, mini, 40 m/s

Order no.

ZV9915S120

ZV9915S140

Technical data

Accuracy

Probe head

Insert opening

Sensor shaft

Sensor length

Standard delivery

Digital vane anemometer with snap-on head, fitted cable, adapter cable with ALMEMO

®

D6 plug

Measuring range 0.4 to 20 m/s

Measuring range 0.5 to 40 m/s

±1 % of final value ±1.5 % of measured value

Ø 22 mm, length 28 mm

Replaceable snap-on head from 35 mm

Ø 15 mm

175 mm including probe head

Order no.

FVAD15S120

FVAD15S140

09.04

ALMEMO

®

D6

Air flow

Digital vane anemometer FVAD 15 S220/S240 with snap-on head, micro

Accessories

Spare snap-on head, micro, 20 m/s

Spare snap-on head, micro, 40 m/s

Order no.

ZV9915S220

ZV9915S240

Technical data

Accuracy

Probe head

Insert opening

Sensor shaft

Sensor length

±1 % of final value ±3 % of measured value

Ø 11 mm, length 15 mm

Replaceable snap-on head from 16 mm

Ø 15 mm

165 mm including probe head

Standard delivery

Digital vane anemometer with snap-on head fitted cable, adapter cable with ALMEMO

®

D6 plug

Measuring range 0.6 to 20 m/s

Measuring range 0.7 to 40 m/s

Order no.

FVAD15S220

FVAD15S240

Digital vane anemometer FVAD 15 SMA1 with snap-on head, macro

Technical data

Accuracy

Probe head

Insert opening

Sensor shaft

Sensor length

Accessories

Spare snap-on head, macro, 20 m/s

Carry-case

Order no.

ZV9915SMA1

ZB9605TK

Standard delivery

Digital vane anemometer with snap-on head fitted cable, adapter cable with ALMEMO

®

D6 plug

Measuring range 0.2 to 20 m/s

±1 % of final value ±1.5 % of measured value

Ø 85 mm, length 80 mm

Replaceable snap-on head from 119 mm

Ø 15 mm

235 mm including probe head

Order no.

FVAD15SMA1

Digital vane anemometer FVAD 15 MA1 with brass probe head, macro attachment for measuring air quantity

Air quantity attachment ZV 9915 LM

Technical data

Accuracy

Probe head

Insert opening

Sensor shaft

±0.5 % of final value ±1.5 % of measured value

Ø 80 mm, length 70 mm fitted brass probe head from 108 mm

Ø 15 mm

225 mm including probe head Sensor length

Accessories

Carry-case for rotating vane

Air quantity attachment (plug-in)

Ø 200 mm (up to approx. 275 m³/h)

Order no.

ZB9605TK

ZV9915LM

Standard delivery

Digital vane anemometer with fitted brass probe head fitted cable adapter cable with ALMEMO

®

D6 plug

Measuring range 0.2 to 20 m/s

Order no.

FVAD15MA1

09.05

Air flow

Differential pressure and Pitot tube measurement Measuring connector FDA 602 S1K / S6K

• Pressure measuring connector in compact design for flow measurement with Pitot tubes

• Fitting for connecting hose between Pitot tube and pressure measuring connector

• Pressure measuring connector can be plugged directly onto the measuring instrument.

Measuring connector FDA602S1K / S6K

Technical data

Overload capacity Maximum three times final value

Max. common mode pressure 700 mbar

Accuracy (zero-pt adjusted) ±0.5% of final value in range 0 to positive final value

Nominal temperature

Temperature drift

Compensated temp. range

25 °C

< ±1.5 % of final value

0 to +70 °C

Operating range

Dimensions

Hose terminals

Sensor material

-10 to +60 °C, 10 to 90% RH, non-condensing

74 x 20 x 8.8 mm

Ø 5 mm, 12 mm long aluminum, nylon, silicone, silica gel, brass

!

Advisory note when used in conjunction with

ALMEMO

®

2890, 5690, 5790, 8590, 8690:

The new ALMEMO

®

pressure measuring connector is very slightly higher (8.8 mm). As a result adjacent input sockets on the ALMEMO

®

device may be partly covered. However, the 1st input socket can always be used without restriction. Or, alternatively, the ALMEMO

® pressure measuring connector can be plugged in at any input socket using connecting cable ZA9060AK1.

!

On ALMEMO

®

devices to obtain precise measured results in m/s the wind tunnel temperature can be entered in the -50 to +700 °C range for compensation purposes.

Accessories

Order no.

new

ALMEMO

®

pressure measuring connector for barometric pressure 700 to 1100 mbar, without pressure terminal sleeve

Technical data see page 11.12 including programming for automatic atmospheric pressure compensation (comment *P)

(variant with pressure terminal sleeve, see page 10.10)

FDAD12SA

OA9000PK

Connecting cable, 0.2 meters

Extension cable, 2 meters

1 set of silicone hoses black / colorless, 2 meters

Silicone hose, black, per meter

Silicone hose, colorless, per meter

ZA9060AK1

ZA9060VK2

ZB2295S

ZB2295SSL

ZB2295SFL

Variants

(including manufacturer’s test certificate)

(including one set of silicone hoses, 2 meters)

Measuring ranges ±1250 Pa, Differential pressure (1 to 40 m/s), Measured variables: m/s, Pa,

Measuring connector, independent of position

Measuring ranges ±6800 Pa Differential pressure (2 to 90 m/s) Measured variables m/s, Pa,

Measuring connector, independent of position

Order no.

FDA602S1K

FDA602S6K

DAkkS or factory calibration KV90xx, air flow, and KD90xx, pressure, for sensor or measuring chain (sensor + device) (see chapter „Calibration certificates“). DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

09.06

Air flow

Pitot Tubes for Differential Pressure Sensors FDA602

• Prandtl Pitot tubes with hemispheric head.

• For measuring the dynamic pressure, the tip of the Pitot tube has an opening of 0.3d.

• For measuring the static pressure, a total of 12 holes with 0.1d

Ø have been arranged at a distance of 3d.

!

Mit ALMEMO

®

devices that have an option for entering factors can also be used to perform wind velocity measurements with cylindrical probes, according to VDEH.

The cylindrical Pitot tubes have a probe-related coefficient of 1.7. By entering a factor of 0.767 in the range m/s this coefficient will be considered during the measurement.

Option

Movable screw connection for brass Pitot tubes with shaft diameter x (6; 8; 10; 20mm) for steel Pitot tubes with shaft diameter x (6; 8; 10; 20mm)

Order no.

ZB9912KMx

ZB9912KVx

Types and Technical Data:

Head Diameter (d) Shaft Diameter (D) Length

3 mm

3 mm

6 mm

6 mm

300 mm

300 mm

10 mm

10 mm

10 mm

10 mm

10 mm

10 mm

20 mm

20 mm

5 mm

5 mm

5 mm

5 mm

8 mm

8 mm

8 mm

8 mm

10 mm

10 mm

10 mm

10 mm

20 mm

20 mm

20 mm

20 mm

8 mm

8 mm

8 mm

8 mm

8 mm

8 mm

8 mm

8 mm

400 mm

400 mm

600 mm

600 mm

400 mm

400 mm

800 mm

800 mm

800 mm

800 mm

1000 mm

1000 mm

1500 mm

1500 mm

2000 mm

2000 mm

Tmax Permiss. Dust Material

150°C none Nickel-plated brass

300°C none Chrome-nickel steel

350°C none

500°C none

350°C none

500°C none

350°C low

500°C low

350°C low

600°C low

350°C some

600°C some

350°C some

600°C some

350°C some

600°C some

350°C more

600°C more

Nickel-plated brass

Chrome-nickel steel

Nickel-plated brass

Chrome-nickel steel

Nickel-plated brass

Chrome-nickel steel

Nickel-plated brass

Chrome-nickel steel

Nickel-plated brass

Chrome-nickel steel

Nickel-plated brass

Chrome-nickel steel

Nickel-plated brass

Chrome-nickel steel

Nickel-plated brass

Chrome-nickel steel

Order no.

FD991233MS

FD991233VA

FD991254MS

FD991254VA

FD991256MS

FD991256VA

FD991284MS

FD991284VA

FD991288MS

FD991288VA

FD991296MS

FD991296VA*

FD991297MS

FD991297VA*

FD991298MS

FD991298VA*

FD991299MS

FD991299VA*

*) all VA Pitot tubes can be operated up to 700°C for a short period

09.07

Air flow

ALMEMO

®

D6

Digital thermoanemometer FVAD 35 THx with ALMEMO

®

D6 plug with integrated atmospheric pressure sensor, for automatic pressure compensation

FVAD 35 TH4 / TH5

• new:

Automatic atmospheric pressure compensation is provided for pressure-dependent flow velocity by means of a digital atmospheric pressure sensor integrated in the ALMEMO® D6 plug itself.

• Digital thermoanemometer with A/D converter in the grip or integrated in the cable

• The probe tube has a small diameter, only 6 mm.

• All relevant measurable variables can be measured using just one sensor.

• Three measuring channels are programmed (at our factory):

Temperature (°C, t), Flow velocity (m/s, v), Atmospheric pressure (mbar, AP, p)

General features and accessories, ALMEMO

®

see page 01.08

D6 sensors:

FVAD 35 TH4Kx / TH5Kx

Technical data

Digital thermoanemometer (Sensor including A/D converter)

Flow

Measuring range

FVAD 35 TH4 / TH4Kx 0.08 to 2 m/s

0.2 to 20 m/s FVAD 35 TH5 / TH5Kx

Resolution

FVAD 35 TH4 / TH4Kx

FVAD 35 TH5 / TH5Kx

Response time

Accuracy

FVAD 35 TH4 / TH4Kx

FVAD 35 TH5 / TH5Kx

0.001 m/s

0.01 m/s

<1.5 seconds

Nominal conditions

± (0.04 m/s +1% of meas. val.)

± (0.2 m/s +2% of meas. val.)

22 °C ±2 K, 45 % RH ±10 % RH

1013 mbar

Temperature compensation 0 to +50 °C

Influence of temperature

FVAD 35 TH4 / TH4Kx

FVAD 35 TH5 / TH5Kx

Incidental flow

Angle dependence

Pressure range

Pressure compensation

±0.5 % of measured value /°C at 0.3 to 2 m/s

±0.3% of measured value /°C at 0.3 to 20 m/s bidirectional

<3% of measured value with deviation <15°

Ambient pressure automatic in range 700 to 1100mbar

Accessories

(for FVAD 35 THxK1 / K2 only)

DAkkS or factory calibration KV90xx air flow for digital sensor (see chapter „Calibration certificates“). DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

Temperature

Measuring range

Resolution

Accuracy

Response time T

90

-20 to +70 °C

0.1 °C

±0.7 °C at 0 to 50 °C and >0.5 m/s typical 10 seconds

Digital atmospheric pressure sensor

(integrated in ALMEMO

®

D6 plug)

Measuring range

Accuracy

ALMEMO

®

D6 plug

Refresh rate

Supply voltage

700 to 1100 mbar

±2.5 mbar (at 0 to +65 °C)

0.5 seconds for all 3 channels

6 to 13 VDC

40 mA Current consumption

Dimensions

Probe diameter

Flow aperture

6 mm approx. 10 x 3 mm

FVAD 35 TH4 / TH5

Probe with grip, probe lengths 210 mm

(plus grip) ALMEMO

®

cable 1.5 meters

FVAD 35 TH4Kx / TH5Kx

Probe with detached electronics unit integrated in the

cable, Probe lengths THxK1, 80 mm / THxK2, 300 mm

Probe cable 5 meters to the electronics

ALMEMO

®

cable 1.5 m

Order no.

Clamped screw connection with thread adapter for telescopic extension

/ extension set (maximum 80 °C)

Telescope extension Ø 15 to 24 mm 330 / 1010 mm

Extension set Ø 15 mm 4 x 255 mm

ZV9915KV

ZV9915TV

ZV9915VR3

Variants

(including works certificate)

Order no.

Digital thermoanemometer, fitted cable with ALMEMO

®

D6 plug and integrated digital atmospheric pressure sensor

Sensor 2 m/s, length = 210 mm, (with grip)

Sensor 2 m/s, length = 80 mm, (detached electronics unit)

Sensor 2 m/s, length = 300 mm, (detached electronics unit)

Sensor 20 m/s, length = 210 mm, (with grip)

Sensor 20 m/s, length = 80 mm, (detached electronics unit)

Sensor 20 m/s, length = 300 mm, (detached electronics unit)

FVAD35TH4

FVAD35TH4K1

FVAD35TH4K2

FVAD35TH5

FVAD35TH5K1

FVAD35TH5K2

Other designs are available on request

High-temperature thermoanemometer MT8635THx

Operative range -40 to +120 °C, up to 40 m/s

Probe with detached electronics unit integrated in the cable

09.08

Air flow

Thermoelectric Flow Sensor FV A605 TA

• Probe tube with heated miniature thermistor for flow measurement and precision NTC resistance for automatic compensation.

• Evaluation electronics are located in a separate sensor transmitter module.

• High accuracy as a result of integrated temperature compensation and individual calibration in wind tunnel, with laser

Doppler anemometer as reference system.

• Response time only 2s for smoothing the measured value indicated, optionally without smoothing with 100ms response time.

• Suitable for measuring small flow velocities in gaseous substances, particularly for control systems and monitoring.

• Typical applications include comfort index measurements,

HEVAC applications, environmental technology, clean room technology and process measuring and control technology.

Technical Data

Electronics Box with Sensor

Measuring range:

FV A605 TA1(O)

FV A605 TA5(O)

Resolution:

FV A605 TA1(O)

FV A605 TA5(O)

0.01 to 1m/s

0.15 to 5m/s

0.001m/s

0.01m/s

Accuracy:

FV A605 TA1(O)

FV A605 TA5(O)

Nominal conditions:

±1.0% of final value and

±1.5% of meas. value

±0.5% of final value and

±1.5% of meas. value

22°C, 960hPa

Automatic temperature compensation: effective in range 0 to 40°C

Temperature influence: ±0.5% of fin. value/°C

Sensor

Head size:

Shaft:

Operative range:

Angle of attack:

FV A605 TA1/TA5

FV A605 TA1O/TA5O

Inlet opening:

FV A605 TAx:

FV A605 TAxO:

Ø 8mm

Ø 15mm

0 to 40°C

±30°

±180°

9mm protecting cage 110mm

Sensor length:

FV A605 TAx:

FV A605 TAxO

Sensor cable length:

300mm

310mm

1.5m

Storage temperature: –30 to +90°C

General Technical Specifications

Measurement medium: dry air or inert gases

Response time:

FVA605TAxD

FVA605TAxU smoothened, 1 t = 2s not smoothened, 1 t = 100ms

Power supply:

Current consumption:

Output signal:

Housing:

Dimensions:

Protection system:

Weight:

Operating temperature:

Storage temperature:

Air humidity:

Adjusting reference: through ALMEMO

(approx. 7… 12V) approx. 70mA

®

0 … 1V, linearised,

device load resistance min. 10kohms

100 x 60 x 35mm (L x W x H)

IP 40 (aluminium housing) approx. 250g

0 to 40°C

–30 to 90°C

0 … 90% r.H., non-condensing laser Doppler wind tunnel, adjustment at 22°C/approx. 960hPa,

(certificate according to SN EN 45001)

Types

(incl. clamping holder and ALMEMO

®

connecting cable 1.5m long)

Unidirectional (sensitive in one direction) with protected measuring tip

Measuring range up to 1m/s, smoothened

Measuring range up to 5m/s, smoothened

Measuring range up to 1m/s, not smoothened

Measuring range up to 5m/s, not smoothened

Order no.

FVA605TA1D

FVA605TA5D

FVA605TA1U

FVA605TA5U

Omnidirectional (direction-independent, symmetrical ball tip)with protecting cage (Ø110mm) including carry-case

Measuring range up to 1m/s, smoothened

Measuring range up to 5m/s, smoothened

Measuring range up to 1m/s, not smoothened

Measuring range up to 5m/s, not smoothened

FVA605TA1OD

FVA605TA5OD

FVA605TA1OU

FVA605TA5OU

DAkkS or factory calibration KV90xx, air flow, for sensor or measuring chain (sensor + device) (see chapter „Calibration certificates“).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

09.09

09.10

Pressure, force, displacement, speed, flow

Content

The right pressure sensor for each and every measuring task

Pressure Transducer FDA602L

High-precision pressure sensor FDAD33/35M

10.02

10.03

10.04

10.06

Temperature-Compensated Pressure Sensors FD8214

Pressure transducer for measuring the temperature of refrigerants, absolute pressure FDA602LxAK

Differential pressure transmitter FDA602D

Digital atmospheric pressure sensor FDAD 12 SA, for barometric pressure

10.08

10.09

Pressure measuring connector for barometric pressure FDA 612 SA 10.10

Pressure measuring connector for differential pressure FDA 612 Sxx 10.11

Pressure Sensors for Wall Mounting FD8612DPS/APS

Differential pressure transmitter for smallest pressure with automatic zero-point correction

10.10

10.12

10.13

Technical Features of Force Transducers

The Right Displacement Sensor For Any Measuring Task

Axial turbine flowmeters

Optical Rotational Speed Meters

Tension and Compression Sensor K25

Compression Sensor K22 and K1613

ALMEMO

®

input connector for measuring bridges

10.15

10.15

10.16

10.16

10.17

10.18

10.19

Displacement Sensor, Potentiometric FWA xxx T

Displacement Tracer, Potentiometric FWA xxx TR

Rotational Speed Sensor FUA 9192

Flow sensors for liquids FVA 645 GVx

Axial turbine flowmeter for liquids FVA 915 VTH

Axial turbine flowmeter for liquids FVA 915 VTH25

10.20

10.21

10.22

10.23

10.24

10.25

10.01

Pressure

The Right Pressure Sensor For Any Measuring Task

Different methods are usually used for manufacturing pressure sensors that have been adapted to the corresponding application.

• Thick-Film Sensors

• Thin-Film Sensors

• Piezo-Resistive Sensors

Pressure transducers are principally available with 4 pressure calibrations:

• Relative pressure: Pressure related to the environmental pressure

• Absolute pressure: Pressure related to vacuum (0bar)

• Overpressure: Pressure related to atm. pressure at manufacturing (approx.

1bar)

• Differential press.: Pressure related to a second, variable pressure

Thick-Film Sensors

The expansion-sensitive elements are applied to a special steel membrane by screen printing technology.

Advantage:

Compact design, particularly suitable for use in simple monitoring and control circuits.

Disadvantage:

Limited operating temperature range, measured values are subject to a longterm variation

Thin-Film Sensors

In a demanding manufacturing process, the wire strain gauges are directly formed on a passivated special steel membrane by a chemical vapour deposition process.

Advantage:

Very compact and homogeneous design, high long-term stability and dynamic load capacity, particularly suitable for operation in harsh industrial environments in the range of medium and high relative pressures.

Disadvantage:

Very expensive manufacturing process.

Piezo-Resistive Sensors

A silicone membrane with ‚diffused in‘ expansion-sensitive resistors is used as the pressure-sensitive element. Due to its compatibility with many substances silicone would limit the use of the sensor.

Therefore, a pressure transmission system, consisting of a filling liquid and a special steel membrane has been integrated. The pressure measuring cell is temperaturecompensated and is manufactured in demanding vacuum processes.

Advantage:

High accuracy within a wide temperature range, particularly suitable for use in high sophisticated measurement and control tasks, especially for measurement of absolute pressure and low to medium relative pressure.

Disadvantage:

Generally, an expensive manufacturing process, however, cost-efficient when produced in large quantities.

Two mechanical designs are available in the ALMEMO

®

sensor range:

• Pressure sensors for hose connection:

The measuring cell is housed in a compact plastic housing with two connecting fittings. The pressure sensors are available for wall mounting or as pressure modules that can be directly plugged into measuring instruments, with measuring ranges for relative or differential pressure measurement in gases, and also for atmospheric pressure measurements.

• Built-In Pressure Transducers:

The measuring cell is suspended in an oil-filled, all-welded special steel enclosure. All parts that come into contact with a substance are made from special steel. Therefore, these transducers are also suitable for use in chemically aggressive substances in various industrial applications.

Temperature Measurement with Pressure Sensors for Refrigerants

Option SB0000R

All ALMEMO

®

Version V5/V6 devices, including ALMEMO

®

data loggers and data acquisition systems, can be used for continuous temperature measurement (resolution 0.1K) with absolute pressure sensors (resolution 0.001 bar compulsory !).

Both, pressure and temperature can be selected or continuously indicated and recorded. (cf. page 10.08)

10.02

Pressure

Pressure Transducer FDA 602 L

Technical Data:

Overload

Output signal

Two times final value

0.2 to 2.2 V

Accuracy class ±0.5 % of final value

(linearity + hysteresis + reproducibility)

Total error range

0 to +50 °C

-10 to +80 °C

±1.0 % of final value

±1.5 % of final value

(linearity + hysteresis + reproducibility + temperature

coefficients + zero-point + range tolerance)

Response time (0 to 99 %) <5 ms

Nominal conditions 22°C ±2 K, 10 to 90 % RH, non-condensing

Pressure

• Compact pressure sensors for industrial applications in liquid and gaseous substances.

• Piezo-resistive, flexibly suspended silicone measuring cell in an oil-filled, all-welded special steel enclosure.

• The stable mechanical construction provides a reliable protection for the measuring cell against the test substance and immunes it against pressure peaks and vibrations.

• Available with three calibrations.

Relative pressure: Pressure related to the environmental press.

Absolute pressure: Pressure related to vacuum (0 bar)

Overpressure: Pressure related to atm. pressure at manufacturing (approx. 1bar).

Power supply

Operating temperature

6.5 to 15 VDC, consumption <4 mA via ALMEMO

®

connector

-40 to +100 °C

Pressure terminal male thread G1/4” membrane not flush with front

Material in contact with medium Stainless steel

DIN 1.4404/1.1135

External seal Viton

Weight

Protective class approx. 50 g

IP 65

Quick-release coupling nominal width 5 internal thread G1/4” nominal width 7,2 internal thread G1/4”

Accessories

PTFE sealing tape, -200 to +260 °C, width 10 mm, thickness 0.1 mm, roll of 12 meters

Quick-release coupling, nominal width 5, up to 35 bar Connection internal thread G1/4“, brass

Quick-release coupling, nominal width 7.2, up to 35 bar connection internal thread G1/4“, brass

Order no.

ZB9000TB

ZB9602N5

ZB9602N7

Types: including ALMEMO

®

cable 1.5m long

Measuring ranges relative pressure:

up to 2.5 bar up to 5 bar up to 10 bar

FDA602L3R

FDA602L4R

FDA602L5R

Measuring ranges absolute pressure:

up to 2.5 bar up to 5 bar up to 10 bar

FDA602L3A

FDA602L4A

FDA602L5A

Measuring ranges overpressure:

up to 25 bar up to 50 bar up to 100 bar up to 500 bar

FDA602L2U

FDA602L3U

FDA602L4U

FDA602L6U

Pressure transducer for measuring the temperature of refrigerants see page 10.08.

DAkkS or factory calibration KD9xxx pressure for sensor or measuring chain (sensor + device) (see chapter Calibration certificates).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

10.03

Pressure

ALMEMO

®

D6

High-precision pressure sensor FDAD33/35M

Very accurate over a wide temperature range, digital sensor with ALMEMO

®

D6 plug

• Stable piezo-resistive transducer with integrated A/D converter and signal processor

• Temperature-dependence and non-linearity are eliminated by means of mathematical compensation; this ensures a high level of accuracy.

• Digital output of measured value

• The current value is measured at the sensor’s high sampling rate.

• To acquire transitory pressure fluctuations and pressure peaks the maximum value, minimum value, and average value are calculated from the current values in the ALMEMO® D6 plug and output in three function channels.

• One measuring channel is programmed (at our factory) :

Pressure (bar, p)

Up to three function channels can also be activated

(via LMEMO

®

device V6) :

Maximum value, minimum value, average value. This device can be completely configured directly on a PC via USB adapter cable ZA 1919 AKUV. (see “General accessories for

ALMEMO

®

D6 sensors” page 04.05).

General features and accessories, ALMEMO

®

see page 01.08

D6 sensors:

FDAD 33 FDAD 35M

Technical data

Digital pressure sensor (including A/D converter)

Pressure range

Relative pressure

Overpressure

Absolute pressure

1 to 1000 bar see under variants

Zero-point at ambient atmospheric pressure, current

Zero-point at ambient atmosph. pressure, production

Zero-point, vacuum

Pressure connection

FDAD33

FDAD35M

Outside thread G ¼“

Diaphragm, internal

Diaphragm, flush with front

Outside thread G 1/2“

In pressure range 700/1000 bar

Outside thread G 3/4“

Storage / operating temperature -40 to +120 °C

Accuracy

Error margin* at -10 to +40 °C

Error margin* at -10 to +80 °C

*Linearity, hysteresis, reproducibility, temperature coefficients,

zero-point

0.05 % of final value

0.1 % of final value

Sampling rate, internal

Material in contact with medium

200 Hz

Stainless steel, AISI 316L,

Viton

Protection

Dimensions

Sensor connector

IP65 see dimensional drawings

Built-in plug

ALMEMO

®

connecting cable Coupling, 2-meter PVC cable,

ALMEMO

®

D6 plug

ALMEMO

®

D6 plug

Refresh time

Setting time

0.005 seconds for all channels

0.6 seconds

Delay after sleep mode

Supply voltage

Current consumption

1 second

6 to 13 VDC approx. 11 mA

10.04

ALMEMO

®

D6

Options

Connecting cable Total length = 5 m

Connecting cable Total length = 10 m

Greater lengths up to 100 meters on request..

Pressure

Order no.

OD0D33L05

OD0D33L10

Variants

Digital pressure sensor, plug connection, 2-meter connecting cable with ALMEMO

®

D6 plug, factory test certificate

Pressure range

Relative pressure

0 to 1 bar

0 to 3 bar

0 to 10 bar

0 to 30 bar

Resolution

0.0001 bar

0.0001 bar

0.001 bar

0.001 bar

Overload

2 bar

5 bar

20 bar

60 bar

Order no.

Diaphragm, internal

FDAD3301R

FDAD3302R

FDAD3303R

FDAD3304R

Order no.

Diaphragm, flush with front

FDAD35M01R

FDAD35M02R

FDAD35M03R

FDAD35M04R

Special ranges -1 ... 1 / 3 / 10 bar on request

Overpressure

0 to 100 bar

0 to 300 bar

0 to 700 bar

0 to 1000 bar

0.01 bar

0.01 bar

0.1 bar

0.1 bar

Absolute pressure

0,8 to 1,2 bar

0 to 1 bar

0 to 3 bar

0 to 10 bar

0 to 30 bar

0.0001 bar

0.0001 bar

0.0001 bar

0.001 bar

0.001 bar

200 bar

400 bar

1000 bar

1000 bar

2 bar

2 bar

5 bar

20 bar

60 bar

FDAD3305U

FDAD3306U

FDAD3307U

FDAD3308U

FDAD3300A

FDAD3301A

FDAD3302A

FDAD3303A

FDAD3304A

FDAD35M05U

FDAD35M06U

FDAD35M07U

FDAD35M08U

FDAD35M00A

FDAD35M01A

FDAD35M02A

FDAD35M03A

FDAD35M04A

DAkkS or factory calibration KD9xxx pressure for digital sensor (see chapter Calibration certificates).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

10.05

Pressure

Pressure Sensors FD 8214

• Compact pressure sensors for liquid and gaseous substances.

• Piezo-resistive measuring cell with temperature compensation.

• Pressure membrane and enclosure made from special steel.

• Available with three calibrations.

Relative pressure:

Pressure related to the environmental pressure.

Absolute pressure:

Pressure related to vacuum (0bar).

Overpressure:

Pressure related to atm. pressure at manufacturing

(approx. 1bar).

!

As the pressure is transmitted to the pressure membrane through a small hole in the thread part, the liquids should not be prone to crystallise and gases should not be heavily contaminated with dust.

Options

Linearity 0.1% (for ranges 1 bar to 600 bar)

Substance temperature –25 to +100°C

Substance temperature –25 to +150°C

(version with cooling fins)

Process connection, small flange

(for FD8214xxA absolute pressure)

KF16

Accessories

Coupler socket with 2m cable and ALMEMO

®

connector

Order no.

OR8214G1

OR8214T1

OR8214T2

OR8214KF16

KF25

Food compliant version with vegetable oil ASEOL Food

Throttle against excess pressure

Output 0 to 10V

Output 0 to 20mA

Output 4 to 20mA

Order no.

OR8214KF25

OR8214ML

OR8214DS

OR8214V

OR8214A

OR8214R4

Order no.

ZA8214AK

Coupler socket 6-pin Straight version

Coupler socket 6-pin Angled version

Order no.

ZB9030RB

ZB9030RBW

Types

Order no.

FD 8214:

Standard version with G1/4” internal thread

Other threads available on request

FD 8214 M:

Membrane (welded with end of thread) flush with front, external thread G1/2”, can be sterilised (important for food and pharmaceutical industry)

Other threads available on request

G1/4”internal thread G1/2”external thread

Measuring ranges relative pressure:

0 to 100 mbar

0 to 160 mbar

0 to 250 mbar

FD821401R

FD821402R

FD821403R

FD8214M01R

FD8214M02R

FD8214M03R

0 to 400 mbar

0 to 600 mbar

0 to 800 mbar

0 to 1 bar

0 to 1.6 bar

0 to 2.5 bar

0 to 4 bar

0 to 6 bar

0 to 10 bar

FD821404R

FD821405R

FD821406R

FD821407R

FD821408R

FD821409R

FD821410R

FD821411R

FD821412R

FD8214M04R

FD8214M05R

FD8214M06R

FD8214M07R

FD8214M08R

FD8214M09R

FD8214M10R

FD8214M11R

FD8214M12R

Types

Order no.

G1/4”internal thread G1/2”external thread

Measuring ranges absolute pressure:

Option: Process connection. small flange (see under Options)

0 to 1 bar

0 to 1.6 bar

0 to 2.5 bar

0 to 4 bar

0 to 6 bar

0 to 10 bar

FD821407A

FD821408A

FD821409A

FD821410A

FD821411A

FD821412A

Measuring ranges overpressure:

0 to 10 bar

0 to 16 bar

0 to 25 bar

FD821412U

FD821413U

FD821414U

0 to 40 bar

0 to 60 bar

0 to 100 bar

0 to 160 bar

0 to 250 bar

0 to 400 bar

0 to 600 bar

0 to 1000 bar

FD821415U

FD821416U

FD821417U

FD821418U

FD821419U

FD821420U

FD821421U

FD821422U

other measuring ranges on request

FD8214M07A

FD8214M08A

FD8214M09A

FD8214M10A

FD8214M11A

FD8214M12A

FD8214M12U

FD8214M13U

FD8214M14U

FD8214M15U

FD8214M16U

FD8214M17U

FD8214M18U

FD8214M19U

FD8214M20U

FD8214M21U

FD8214M22U

DAkkS or factory calibration KD9xxx pressure for sensor or measuring chain (sensor + device) (see chapter Calibration certificates).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

10.06

Pressure

Technical Data

Measuring cell:

Overload

Output signal, power supply :

Response time:

Linearity:

Media temperature:

Temperature drift: piezo-resistive

Ranges 600 bar, i.e. 1.5 times the final value (minimum 3 bar, maximum 850 bar)

Ranges >600 bar, 1500 bar

Standard 0 to 2 volts, feed 6.5 to 13 volts (from ALMEMO® device), current <4 mA

Option : 0 to 10 volts, feed 15 to 30 volts, load >10 kilohms, current <4 mA

Option : 0 to 20 mA, feed 9 to 33 volts, (>18 volts at load 500 ohms), current <25 mA

Option : 4 to 20 mA, 2 conductors, feed 9 to 33 volts, (>18 volts at load 500 ohms), current <25 mA

<1.5 ms / 10 to 90 % nominal pressure

Standard ±0.25 % of final value

Option : ±0.1 % of final value for ranges 1 bar and up to 600 bar

0 to +80°C, temperature comp.: 0 to +70°C option:

–25 to +100°C, temperature comp.: –25 to +85°C

–25 to +150°C, temperature comp.: –25 to +85°C

Zero-point <±0.04 % of final value / °C for ranges >0.5 bar span <±0.02 % of final value / °C for all ranges

Nominal temperature:

Material:

22°C ±2 K, 10 to 90% rH non-condensing housing, pressure connector, membrane: special steel 1.4435

Operat. environment/Sealing: IP 67

Dimensions: see drawing

Connecting threads:

Electrical connection

Weight:

Type 8214: internal thread G1/4“, wrench SW 27

Option for absolute pressure: small flange KF16 or KF21

Type 8214 M: external thread G1/2“, wrench SW 27

Other threads are available on request

Flush-mounting connector, binder coupling 723, 5-pin approx.. 180 g

Type FD 8214 standard version with internal thread G1/4“

L = 45 mm (L = 72 mm with option of medium temperature up to 150 °C with cooling ribs)

Type FD8214M membrane flush with front (welded with end of thread), internal thread G1/2“ can be easily sterilized

L = 45mm

(L = 72 mm with option of medium temperature up to 150 °C with cooling ribs)

Accessories

Order no.

PTFE sealing tape, -200 to +260 °C, width 10 mm, thickness 0.1 mm, roll of 12 meters ZB9000TB

Quick-release coupling, nominal width 5, up to 35 bar

Connection G1/4“ external thread, brass ZB8214N5

Order no.

Quick-release coupling, nominal width 7.2, up to 35 bar

Connection 1/4“ external thread, brass ZB8214N7

Quick-release coupling nominal width 5 external thread G1/4”

Quick-release coupling nominal width 7.2 external thread

G1/4”

10.07

Pressure

Pressure transducer for measuring the temperature of refrigerants FDA 602 LxAK

• Compact pressure sensors for industrial applications in liquid and gaseous substances.

• Piezo-resistive, flexibly suspended silicone measuring cell in an oil-filled, all-welded special steel enclosure.

• The stable mechanical construction provides a reliable protection for the measuring cell against the test substance and immunes it against pressure peaks and vibrations.

• Absolute pressure: pressure related to vacuum (0 bar).

Technical Data:

Overload

Output signal

Two times final value

0.2 to 2.2 V

Accuracy class ±0.5 % of final value

(linearity + hysteresis + reproducibility)

Total error range

0 to +50 °C

-10 to +80 °C

±1.0 % of final value

±1.5 % of final value

(linearity + hysteresis + reproducibility + temperature

coefficients + zero-point + range tolerance)

Response time (0 to 99 %) <5 ms

Nominal conditions 22°C ±2 K, 10 to 90 % RH, non-condensing

Power supply

Operating temperature

Pressure terminal

6.5 to 15 VDC, consumption <4 mA via ALMEMO

®

-40 to +100 °C

connector male thread G1/4” membrane not flush with front

Material in contact with medium Stainless steel

DIN 1.4404/1.1135

External seal, Viton

Weight

Protective class approx. 50 g

IP 65

Calculation of the refrigerant temperature with device special version SB0000R2

The ALMEMO

®

Version V6 devices, (2590-2/-3S/-4S, 2690, 2890, 8590, 8690, 5690) can be used a for continuous temperature measurement (resolution 0.1K) with absolute pressure sensors (resolution 0.001 bar compulsory !). Both, pressure and temperature can be selected or continuously indicated and recorded.

Technical data for ALMEMO

®

option SB0000R2:

Refigerant: R22

Pressure Range: 0 to 36 bar

Temperature Range: –90°C to +79°C *

Operation point dew-point

R23 R134a R404a

0 to 49 bar 0 to 40,5 bar 0 to 32 bar

–100°C to +26°C * –75°C to +101°C * –60°C to +65°C * dew-point dew-point dew-point

R404a

0 to 32 bar

–60°C to +65°C * boiling point

Refigerant: R407C

Pressure Range: 0 to 46 bar

Temperature Range: –50°C to +86°C *

Operating point dew-point

R407C

0 to 46 bar

–50°C to +86°C * boiling point

R410A

0 to 49 bar

–70°C to +70°C * dew-point

R417A

0 to 27 bar

–50°C to +70°C * dew-point

R507

0 bis 37 bar

–70°C to +70°C * dew-point

*) The final temperature is obtained from the data of the refrigerant.

For pressure transducer with smaller pressure ranges, the specified final temperature changes. (Linearizations for other refrigerants on request)

Special design refrigerant temperature for ALMEMO

®

devices V6

(Please order when buying new devices or send it to upgrade existing device)

Order no. SB0000R2

Types

including ALMEMO

®

connecting cable, 1.5 m, and programming of a refrigerant measuring channel

Measuring ranges Absolute pressure (resolution 0.001 bar) up to 10bar up to 30bar up to 50bar

Order no.

FDA602L5AK

FDA602L6AK

FDA602L7AK

DAkkS or factory calibration KD9xxx pressure for sensor or measuring chain (sensor + device) (see chapter Calibration certificates).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

10.08

Pressure

Differential pressure transmitter FDA 602 D

Low-pressure version

• This measures the differential pressure in liquid and gaseous media indirectly using two absolute pressure sensors.

• This makes it less expensive but more robust with respect to asymmetrical overload.

• The differential pressure range should be at least 5% of the standard pressure range.

• Each side of the sensor incorporates two pressure connections.

The transmitters can thus be used easily and conveniently in pressure pipes.

• It incorporates a high-speed, high-precision microprocessor.

• All reproducible errors affecting the pressure sensors, i.e. involving non-linearity and temperature dependency, can be completely eliminated by means of mathematical error compensation.

Medium-pressure version

Technical Data:

Standard pressure range (maximum measurable pressure per pressure connection), overload, differential pressure range.

See versions listed below.

Storage / operating temperature -40 to +100 °C

Compensated standard range -10 to +80 °C

Error margin ≤0.05% of final value, typical

≤0.1% of final value, max. with respect to standard pressure range

(linearity + hysteresis + reproducibility + temperature error)

Pressure connections G1/4“ thread, female

(2 per side)

Material in contact with medium Stainless steel, 316L,

DIN 1.4435

Power supply

Output

Electrical connection

6 to 15 VDC via

ALMEMO

®

connector

0 to 2 V

Binder plug, including

ALMEMO

®

connecting cable,

2 meters

CE conformance

Protective class

EN61000-6-1 to 4 with shielded cable

IP 65

Weight

Low-pressure version 475 grams

Medium-pressure version 750 grams

Types

Differential pressure transmitter, including ALMEMO

®

cable, 2 meters

Standard pressure range

Absolute pressure

Low-pressure version

0 to 3 bar

0 to 10 bar

0 to 25 bar

Medium-pressure version

0

0

to 100 bar

to 300 bar

Overload

10 bar

20 bar

40 bar

200 bar

450 bar

Differential pressure range

Please indicate final value

0 to 0.2 to 3 bar

0 to 0.5 to 10 bar

0 to 1.25 to 25 bar

0 to 5 to 100 bar

0 to 15 to 300 bar

Order no.

FDA602D01

FDA602D02

FDA602D03

FDA602D10

FDA602D11

DAkkS or factory calibration KD9xxx pressure for sensor or measuring chain (sensor + device) (see chapter Calibration certificates).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

10.09

Pressure

Digital atmospheric pressure sensor FDAD 12 SA, for barometric pressure

Integrated in ALMEMO

®

D6 plug

General features and accessories, ALMEMO

®

D6 sensors

see page 01.08

Special features

• Digital atmospheric pressure sensor with temperature compensation

• Very accurate over a wide temperature range

• The value measured for atmospheric pressure can also be used to compensate other sensors on the ALMEMO

®

device

(programming comment *P).

• Compact design, without pressure connection sleeve

• Can be connected directly to the measuring instrument.

• One measuring channel is programmed (at our factory).

• Atmospheric pressure (mbar, AP, p)

Technical Data

Digital atm. pressure sensor (integrated in ALMEMO

®

D6 plug)

Measuring range

Accuracy

Operating range

Dimensions

700 to 1100 mbar

±2.5 mbar (at 0 to +65 °C)

-10 to +60 °C

10 to 90 % RH non-condensing

62 x 20 x 7.6 mm

ALMEMO

®

D6 plug

Refresh rate

Supply voltage

Current consumption

1 second for all channels

6 to 13 VDC

4 mA

Variants

(including manufacturer’s test certificate)

Digital atmospheric pressure sensor for barometric pressure, integrated in ALMEMO

®

D6 plug

DAkkS or factory calibration KD92xx atmospheric pressure for digital sensor (see chapter Calibration certificates).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

Order no.

FDAD12SA

Pressure measuring connector for barometric pressure FDA 612 SA

• Compact design - can be plugged directly onto measuring instrument.

• Piezo-resistive pressure sensor - ensures high measuring accuracy.

Technical Data:

Measuring range

Overload capacity

Accuracy

Nominal temperature

Temperature drift

Hose terminals

Accessories

Connecting cable, 0.2 meters

Extension cable, 2 meters

700 to 1050 mbar

(total range 0 to 1050 mbar)

Maximum 1.5 times final value

±0.5 % of final value

25 °C

<±1 % final value at 0 to +70 °C

Ø 5 mm, 12 mm long

Order no.

ZA9060AK1

ZA9060VK2

Sensor material

Operating range

Dimensions

Extension cable, 4 meters aluminum, nylon, silicone, silica gel, brass

-10 to +60 °C, 10 to 90% RH, non-condensing

90 x 20 x 7,6 mm

Order no.

ZA9060VK4

Variants

(including manufacturer’s test certificate)

Pressure measuring connector for barometric pressure with pressure terminal sleeve

Order no.

FDA612SA

DAkkS or factory calibration KD9xxx pressure for sensor or measuring chain (sensor + device) (see chapter Calibration certificates).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

10.10

Pressure

Pressure measuring connector for differential pressure FDA 612 SR, FDA 602 S2K

• New compact design - can be plugged directly onto measuring instrument.

• Piezo-resistive pressure sensor - ensures high meas. accuracy.

!

Advisory note when used in conjunction with

ALMEMO

®

2890, 5690, 5790, 8590, 8690:

The new ALMEMO

®

pressure measuring connector is very slightly higher (8.8 mm). As a result adjacent input sockets on the ALMEMO

®

device may be partly covered. However, the 1st input socket can always be used without restriction. Or, alternatively, the ALMEMO

® pressure measuring connector can be plugged in at any input socket using connecting cable ZA9060AK1.

Technical Data

Overload capacity

FDA612SR

FDA602S2K max. 1.5 times final value maximum 250 mbar

Accuracy (zero-pt adjusted) ±0.5% of final value in range

0 to positive final value

Common mode pressure

Nominal temperature

FDA602S2K max. 700 mbar

25 °C

Temperature drift

FDA612SR < ±1.5 % of final value

compensated temperature range 0 to +70 °C

FDA602S2K < ±2 % of final value

compensated temperature range -25 to +85 °C

Accessories

Connecting cable, 0.2 meters

Extension cable, 2 meters

Operating range -10 to +60 °C, 10 to 90% RH, non-condensing

Dimensions New 74 x 20 x 8.8 mm

Hose terminals Ø 5 mm, 12 mm long

Sensor material aluminum, nylon, silicone, silica gel, brass

Order no.

ZA9060AK1

ZA9060VK2

Extension cable, 4 meters

Order no.

ZA9060VK4

Variants

(including manufacturer’s test certificate)

(including one set of silicone hoses, 2 meters) Pressure measuring connector for differential pressure

Range ±1000 mbar

Range ±250 Pa

(independent of position)

Range ±1250 Pa or ±6800 Pa see page 09.06

Order no.

FDA612SR

FDA602S2K

DAkkS or factory calibration KD9xxx pressure for sensor or measuring chain (sensor + device) (see chapter Calibration certificates).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

10.11

Pressure

Pressure Sensors for Wall Mounting FD 8612 DPS / APS

• Suitable for use in the laboratory, as well as for use in harsh industrial environments, e.g. HEVAC applications, clean room technology, medical technology, filter technology and finishing pass technology.

• The robust mechanics guarantees long term stability, linearity and good reproducibility.

• Temperature drift reduced to a minimum by specific compensation of the sensors.

• Operation is almost maintenance-free, as a result of the freefrom-wear inductive measuring system.

• As standard, the integrated electronics provide a pressure proportional voltage signal from 0 to 2V as output.

Technical Data:

Linearity:

Hysteresis:

±1% of final value, option: ±0.2% or ±0.5%

±0.1% of final value

Nominal temperature:

Overload capacity:

23°C up to 400 mb: 5-fold, from 500 mb: 2-fold

Max. common mode pressure: 1 bar

(at differential measurement)

Power supply: 6 ... 12 VDC, option: 230V 50/60Hz

Power consumption:

Output:

Connection: approx. 3.5mA

0 to 2V, option:

0 to 10V/0(4) to 20mA electrical: screw terminals, screwed cable gland PG 7, pressure: 6.5mm hose connection

Optionen

Linearity 0.2%

(DPS from final value / APS from range) with DPS only in ranges ≥ 2.5 mbar with APS only in range ≤ 100 mbar

Linearity 0.5%

(DPS from final value / APS from range) with DPS only in ranges ≥ 1 mbar with APS only in range ≤ 200 mbar

Order no.

OD8612L2

OD8612L5

Rise time:

Temperature drift:

Zero point

range

Operative range:

Storage temperature:

Housing:

Protection system:

Weight:

Sensor capacity:

Volume increase:

Power supply : 230 V

Output 0 to 10 V

(voltage supply 19 to 31 V DC)

Output 0 to 20 mA

(voltage supply 19 to 31 V DC)

Output 4 to 20 mA

(voltage supply 19 to 31 V DC)

T

90

approx. 0.02s

0.03% of final value / K,

0.03% of final value / K

+10 to +50°C , air humidity 10 to 90% non-condensing

–10 to +70°C material ABS

120 x 80 x 55mm (L x H x D)

Safety class: 0

IP 54 approx. 300g approx. 3ml approx. 0.2ml at nom. press.k

Order no.

OD8612N

OD8612R2

OD8612R3

OD8612R4

Accessories

Order no.

Connecting cable 2m long mounted with connector for connection to ALMEMO

®

devices ZA8612AK2

1 set silicone hoses 2m long black/colourless ZB2295S

Silicone hose black per m

Silicone hose colourless per m

Order no.

ZB2295SSL

ZB2295SFL

Types

Order no.

Measuring ranges relative and differential pressure:

Pressure transducer type DPS 0 to 2.5 mbar ... 1000 mbar

Please specify measuring range

Range 1 mbar (100 Pa), additional charge

Range 0.5 mbar (50 Pa), additional charge

FD8612DPS

OD8612P10

OD8612P05

Order no.

Measuring ranges absolute pressure:

Pressure transducer type APS 0 to 1000 mbar, 900 to 1100 mbar, 800 to 1200 mbar

Please specify measuring range

FD8612APS

DAkkS or factory calibration KD9xxx pressure for sensor or measuring chain (sensor + device) (see chapter Calibration certificates).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

10.12

Pressure

Differential pressure transmitter for smallest pressure with automatic zero-point correction, FD 8612 DPT25R8AZ

For air and non-aggressive gases

• Adjustable differential pressure measuring transducer for the purposes of monitoring the differential pressure in air and in other non-combustible and non-aggressive gases

• Possible uses include : Monitoring of air filters, of forcedair fans and blowers, of industrial air-cooling circuits, of air flows in ventilation conduits, prevention of overheating in air heaters, regulation of airflow valves and fire protection valves, protection against frost in heat exchangers.

Technical Data:

Measuring element

Measuring range

Measuring accuracy

Long-term stability

Reaction time

Maximum pressure

Bursting pressure

Medium

Operating temperature

Piezoelectronic measuring cell

(can be selected via jumper)

-100 to +100 Pa

0 to +100 Pa

0 to +250 Pa

0 to +500 Pa

0 to +1000 Pa

0 to +1500 Pa

0 to +2000 Pa

0 to +2500 Pa

±1,5 % of the measuring range selected

± 6 Pa for measuring ranges

250Pa, 100Pa, ± 100Pa

0.1 % per year (typical)

0.8 or 4.0 seconds (can be selected)

25 kPa

50 kPa

Air and non-aggressive gases

-5 to +50 °C

Storage temperature

Ambient humidity

-20 to +70 °C

0 to 95 % RH, non-condensing

Housing, housing cover, connecting muff, conduit muff : ABS (acrylonitrile butadiene styrene)

Protection

Dimensions

Weight

Pressure connection

IP54

(LxWxH) 90 x 71.5 x 36 mm

150 g

2 hose muffs

Diameter = 5 / 6.3 mm

Electrical connections

Cable entry

Supply voltage

Output signal

(can be selected)

Screw terminals, maximum 1.5 mm

2

M16

24 VAC or 24 VDC, ±10 %

Power <1 W

0 to 10 V

Load 1 kohm minimum

4 to 20 mA, 3 conductors

Load 500 ohms maximum.

Accessories

ALMEMO

®

connecting cable for FD 8612 DPT, differential pressure, 2 cables connected in the transmitter housing

1. ALMEMO

®

connecting cable, PVC, length = 2 meters, with ALMEMO

®

connector

2. Power supply via mains unit ZB1024NA1, 230 VAC / 24 VDC

Order no.

ZA8612DPTAK

Variants

Order no.

Differential pressure transmitter type DPT, for air and non-aggressive gases, with automatic zero-point correction

8 measuring ranges (can be selected via jumper) including standard accessories:

2 fastening screws, 2 plastic conduit muffs, 2-meter plastic hose

FD8612DPT25R8AZ

DAkkS / DKD or factory calibration KD9xxx pressure for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

DAkkS or factory calibration KD9xxx pressure for sensor or measuring chain (sensor + device) (see chapter Calibration certificates).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

10.13

10.14

Force, Displacement, Flow, Speed

Technical Features of Force Transducers

The technical features of the force transducers are substantially fixed by VDI/

VDE guideline 2637. The most important terms are described below:

Measuring range:

The load range, for which the guaranteed error limits will not be exceeded.

Nominal load:

The nominal load is the upper limit of the measuring range. Depending on the sensor, the nominal load can be a tension or compression load.

Working load:

The working load is the load that can be applied to the sensor, as well as the nominal load, without affecting the specified characteristics. The working load range should only be used in exceptional cases.

Load limit:

The load limit is the maximum permissible load that can be applied to the measuring cell without expecting a destruction of the measuring system. At this load the specific error limits are no longer applicable.

Breaking load:

The breaking load is the load where a permanent change or destruction occurs.

Maximum dynamic load:

Rated force related oscillation amplitude of a sinusoidally changing force in direction of the measuring axis of the sensor.

At a load of 107 cycles the sensor, when being repeatedly used up to the rated force, is not subject to significant changes regarding the metrology characteristics.

Drift error:

The drift error is the maximum permissible change of the output signal of the sensor over the specified time at constant load and stable environmental conditions.

ALMEMO

®

Force Measurement

ALMEMO

®

force transducers allow to adjust the constant load (tare) to zero and to enter the final value as nominal value.

The correction value will be automatically calculated from this by the measuring instrument. An ALMEMO

®

connector that switches on this resistor for the adjustment is available for force transducers with integrated reference resistor.

The Right Displacement Sensor For Any Measuring Task

Different methods can be used depending on the limiting and environmental conditions involved with the measuring task:

Linear inductive displacement transducers and tracers: absolutely accurate, high resolution, robust, acceleration resistant, cost-efficient, noise resistant, good long term stability, environmentally stable (contamination, humidity/moisture), point-shaped, almost contactless measurement, easy mounting and handling

Non-contacting displacement measuring systems based on eddy current: very accurate, very fast, high resolution, environmentally stable (contamination, moisture/humidity), noise resistant regarding EMI, temperature stable, long term stability, for devices under test made of all types of electrically conducting materials, nonmagnetic and ferromagnetic, compact sensor designs, extensive application temperature range

Non-contacting inductive displacement measuring systems: accurate, temperature stable, fast, cost-efficient, particularly for ferromagnetic test objects

Long-travel sensors based on eddy current: large measuring paths, robust and compact, no mechanical wear, easy handling, compression-proof

Non-contacting inductive optical displacement measuring systems: point-shaped measurement, accurate, fast, large base distance, material independent

Cable line displacement sensors: very accurate, large measuring paths, easy mounting, cost-efficient

Non-contacting capacitive displacement measuring systems: extraordinary accurate, very temperature stable, fast, high resolution, very good long term stability, material independent for metal objects under test, also suitable for insulating materials, easy to handle, extensive operating temperature range

Conductive plastic potentiometer: high resolution, good linearity, cost-efficient, good temperature and humidity coefficients, extensive operating temperature range

ALMEMO

®

Displacement Measurement

Our Potentiometric displacement sensors have been pre-aligned in the factory by storing the correction values in the

ALMEMO

®

connector before delivery.

The precise adjustment can be locally performed by the user with final measures after the installation

10.15

Force, Displacement, Flow, Speed

Turbine Flowmeters

The sensor contains a vane or paddle that starts rotating when a flow is present. Unlike the optical method, this method also allows for measurements in cloudy and non-transparent liquids. The rotational speed is proportional to the corresponding quantity of flow. The electrical output signal can be generated by two different methods:

• Inductive Proximity Switch:

The rotor blades are provided with special steel caps, therefore, the rotor blades approaching the transducer cause a change of the inductance and the generation of a pulse type output signal.

• Hall Sensor:

The rotor is provided with permanent magnets that affect a Hall sensor, which is located on the transducer. The transducer electronics transforms the

Hall signal into a pulse type electronical output signal.

For measuring the volume flow rate or for dosing tasks, the ALMEMO

®

sensor range includes turbine flowmeters for different measuring ranges and operating conditions:

• Radial turbine flowmeters for large flow quantities.

• Axial turbine flowmeters with rotating vane for small flow quantities

Optical Rotational Speed Meters

The optical reflection method has become the most accepted method for the measurement of revolutions of shafts, wheels, fans etc.

With single unit retroreflective photoelectric sensors the transmitters and receivers form one single unit. The light sent by the transmitter is, by an opposite located object, reflected to the receiver.

The sensor performs a switch when the reflected amount of light exceeds a specific, adjustable limit value at the receiver.

This quantity of light depends on the size and the reflection properties of the object.

Special reflective tapes are used to increase the sensing range and to improve the signal-to-noise ratio.

ALMEMO® rotational speed sensors can be used in two measurement setups:

• Retroreflective photoelectric sensor

(DIN EN 60947: Type D)

Detects only opaque objects.

The sensing range depends on the reflectivity of the object, i.e. on the surface quality and colour.

Sensitive with regard to contamination and against changes of the reflective properties of the object

These influences can (within limits) be compensated by means of a sensitivity adjustment control

Only small mounting efforts are required as the sensor is a single unit device and a rough alignment is sufficient in most cases.

• Retroreflective light barrier (DIN EN

60947: Type R)

Retroreflectors allow for long sensing ranges and an improved signal-to-noise ratio. Low susceptance to interferences, therefore, highly suitable for use under harsh conditions, e.g. outdoor applications or dirty environments

10.16

Force, Displacement, Flow, Speed

Tension and Compression Sensor K25

Force

• Wire strain gauges in four-conductor full-bridge circuit.

• Control resistance for final adjustment of the measuring range.

• All measuring ranges that are specified in Newton can also be supplied in kg ranges

!

All ALMEMO

®

devices provide easy push-button adjustment of no-load and final value.

Technical Data:

Max. load limit:

Maximum dynamic load:

Reference temperature:

150% of final value

70% of final value

23°C

Cable:

Accuracy for tension:

Accuracy for tension and compression:

Nominal measuring path:

3m long, with axial ALMEMO

®

connector

<±0.1% of fin. val.

<±0.2% of fin. val.

<0.15mm

Operative range: –10 to +70°C

Options for all Force Transducers

Indication of measured values with ALMEMO

®

devices in kg

Order no.

OK9000K

Drift error at permanent load: <0.07% per 30min

Permissible lateral forces:

Protection system:

±60% of fin. val.

up to 1kN: IP 65, from 2kN: IP 67

Material:

Dimensions in mm up to 1kN: aluminium

2 to 50kN: stainless steel up to 10kN:

A=50, B=75, C=20, D=M12

20kN, 50kN:

A=65, B=85, C=40, D=M24 x2

Indication of measured values with ALMEMO

®

devices in N and kg

Order no.

OK9000NK

Accessories

Knuckle eyes with external thread M 12 (2 pcs)

(dimensions in mm: D = M 12, E = 16,

F = 32, G = 12, L = 54)

Order no.

ZB902512

Knuckle eyes with external thread M 24 x 2 (2 pcs)

(dimensions in mm: D = M 24 x 2, E = 26,

F = 62, G = 25, L = 94)

Order no.

ZB902524

Types

(including test certificate)

Measuring range 0.02kN 0.05kN, 0.1kN, 0.2kN, 0.5kN, 1kN, 2kN, 5kN or 10kN please specify

Measuring range 20kN

Measuring range 50kN

Order no.

FKA0251

FKA0252

FKA0255

Factory calibration KK9xxx force (traction / thrust) for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

Other designs are available on request

Tension and compression sensor FKA 012 with male thread terminal up to 1000 kN

Tension and compression sensor

FKA 1563 low height, with male thread terminal up to 2 kN

10.17

Force

Compression Sensor K 22

• Wire strain gauges in four-conductor full-bridge circuit.

• Control resistance for final adjustment of the measuring range.

• All measuring ranges that are specified in Newton can also be supplied in kg ranges.

!

All ALMEMO

®

devices provide easy push-button adjustment of no-load and final value.

Technical Data:

Max. load limit:

Maximum dynamic load:

Reference temperature:

Cable:

Accuracy:

150% of final value

70% of final value

23°C radial, 3m long with ALMEMO

®

connector

<±0.5% of final value

Nominal measuring path: <0.2mm

Operative range: –10 to +50°C

Drift error at permanent load: 0.1% per 30min

Protection system:

Material:

IP 65 stainless steel

Type

(including test certificate)

Measuring range 100 N, 200N, 500N, 1000N or 2000N please specify

Order no.

Order no. FKA022

Factory calibration KK9xxx force (tension or compression) for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

Compression Sensor K 1613

• Wire strain gauges in 4-conductor full-bridge circuit.

• Control resistance for final adjustment of the measuring range.

• All measuring ranges that are specified in Newton can also be supplied in kg rangesr.

!

All ALMEMO

®

devices provide easy push-button adjustment of no-load and final value.

Technical Data:

Max. load limit:

Maximum dynamic load:

Reference temperature:

Cable:

Accuracy:

150% of final value

70% of final value

23°C radial, 3m long with ALMEMO

®

connector

<±0.5% of final value

Nominal measuring path:

Operative range:

<0.2mm

–10 to +50°C

Drift error at permanent load: 0.1% per 30min

Protection system:

Material:

IP 65 stainless steel

Type

(including test certificate)

Measuring range 0.5kN, 1kN, 2kN, 5kN, 10kN or 20kN (50 kN on request) please specify

Order no.

FKA613

Factory calibration KK9xxx force (tension or compression) for sensor or measuring chain (sensor + device) (see chapter Calibration certificates)

10.18

Compression Sensor

Other designs are available on request

Torque sensor

Other designs are available on request

Force

Static torque sensor e.g. with square terminal 2 to 5000 Nm

Compression sensor FKA 2528 inexpensive Protective class IP60

0.2 to 10 kN

Compression sensor FKA 013 other measuring ranges from 10 N up to 100 kN

Rotating torque sensor (slip ring) e.g. with square terminal 1 to 5000 Nm

ALMEMO

®

input connector for measuring bridges, millivolt / volt differential

With zero-symmetrical voltage supply of ±2.5 V stabilized from the ALMEMO

®

device

Bridge circuit

Technical Data:

Sensor supply:

Voltage UF: 5V ± 0.05V

Temperature coefficient: <50ppm/°C

Output current:

Quiescent current max. 100mA approx. 3 mA

New

Energy saving So long as the measuring point is not selected, the bridge voltage remains switched OFF.

Types

Model

55mV DC

26mV DC

260mV DC

2.6V DC

Meas. Range

–10.0 to +55.0

–26.0 to +26.0

–260.0 to +260.0

–2.6 to +2.6*

Resolution

1 µV

1 µV

10 µV

0.1 mV

* Data may vary depending on device; (see data sheet per device)

Order no.

ZA9105FS0

ZA9105FS1

ZA9105FS2

ZA9105FS3

10.19

Displacement

Displacement Sensor, Potentiometric FWA xxx T

• Displacement transducers are suitable for direct, accurate measurement of displacements in automatic control and metrology.

• The pickup of the displacement is performed by using a pull rod with a universal joint. This allows for an actuation that is free from backlash and transverse forces, even in case of parallel and angular displacements of transducer and measuring direction.

• Elastomer-damped, independently resilient multi-finger noble metal sliding contact for reliable contact, even at high adjustment speed, shock or vibration.

• Long life span of 100 x 106 strokes, extraordinary linearity up to ±0.075%, pull rod running on two exact bearings, very high adjustment speed of up to 10m/s, shock and vibration resistant.

!

Pre-adjusted in the factory by storing the correction values in the ALMEMO

®

connector.

The precise adjustment can be locally performed by the user with final measures after the installation.

Technical Data:

Independent linearity: T25: ±0.2%; T50: ±0.15%

T75: ±0.1%; T100: ±0.075%

T150: ±0.075%

Housing length (meas. A+1mm): T25: 63mm; T50: 88mm

T75: 113mm; T100: 138mm

T150: 188mm

Mech. stroke (meas. B ±1.5mm): T25: 30mm; T50: 55mm

T75: 80mm; T100: 105mm

T150: 155mm

Total weight (with 2m cable): T25: 140g; T50: 160g

T75: 170g; T100: 190g T150: 220g

Weight of the pull rod incl. coupling and sliding contact block: T25: 35g; T50: 43g

T75: 52g; T100: 58g

T150: 74g

Movability, ball-shaped coupling ±1mm parallel displacement,

±2.5° angular displacement

Operating force (horizontal): ≤ 0.30N

Reproducibility:

Insulation resistance:

Dielectric strength:

Max. permissible torque:

0.002mm

≥ 10MW, (500VDC, 1 bar, 2s)

≤ 1mA, (50Hz, 2s, 1 bar, 500VAC)

140Ncm

Temperature range:

Temperature coefficient:

Vibrations:

Shock:

Life span:

Protection system:

–30 to +100°C typ. 5ppm/°C

5 to 2000Hz/Amax

= 0.75mm/amax = 20g

50g/11ms

> 100 x 106 strokes

IP 40

Option

Plug connection (instead of fixed connected cable), including 3m cable with screwed round socket and ALMEMO

®

connector

Order no.

OWA071AK

Types

Order no.

Working length/resolution, incl. ALMEMO

®

cable 2m long

25 mm / 0,001 mm

50 mm / 0,01 mm

75 mm / 0,01 mm

FWA025T

FWA050T

FWA075T

100 mm / 0,01 mm

150 mm / 0,01 mm up to 3000mm working length

Order no.

FWA100T

FWA150T

on request included with delivery 2 tensioning clamps Z3-31 including 4 cap screws M4x10, 1 ball-shaped coupling

Other designs are available on request

Displacement transducers FWA xxx TEX with pivot joint Protective class IP54, 10 to 300 mm

10.20

Displacement transducers FWA xxx TX2

Protective class IP67 with pivot joint, 25 to 300 mm

Displacement

Displacement Tracer, Potentiometric FWA xxx TR

• Resistor and collector paths made from conducting plastic.

• Suitable for direct measurements of displacement without a form-locking connection, position detection at stationary measuring objects, tolerance measurements and for continuous contour measurement.

• The pull rod, which is supported on both sides, allows for accepting transverse forces that, for example, occur during a continuous scan of curves or spline parts.

• Rear limit stop is used to provide a simple mechanical coupling of automatic retraction systems, such as pneumatic cylinders or electromagnets.

• Long life span of 100 x 106 strokes, extraordinary linearity up to ±0.075%, tracer pin running on two exact bearings, DIN compliant standard measuring inserts can be used, shock and vibration resistant.

!

Pre-adjusted in the factory by storing the correction values in the ALMEMO

®

connector.

The precise adjustment can be locally performed by the user with final measures after the installation.

Technical Data:

Independent linearity: TR25: ±0.2%; TR50: ±0.15%

TR75: ±0.1%;

TR100: ±0.075%

Housing length (meas. A+1mm): TR25: 63mm;

TR50: 94.4mm;

TR75: 134.4mm;

TR100: 166mm

Mech. stroke (meas. B ±1.5mm): TR25: 30mm; TR50: 55mm

TR75: 80mm;

TR100: 105mm

Total weight (with 2m cable): TR25: 120g; TR50: 150g

TR75: 180g; TR100: 200g

Weight of the pull rod incl. coupling and sliding contact block: TR25: 25g; TR50: 36g

TR75: 48g; TR100: 57g

Max. operating frequency: (for most critical application ´probe tip upright´) TR25: 18Hz; TR50: 14

TR75: 11Hz; TR100: 10Hz

Operating force (horizontal): ≤ 5 N

Reproducibility: 0.002mm

Insulation resistance:

Dielectric strength:

Max. permissible torque:

≥ 10MW (500VDC, 1 bar, 2s)

≤ 1mA (50Hz, 2s, 1 bar, 500VAC)

140Ncm

Temperature range:

Temperature coefficient:

Vibrations:

Shock:

Life span:

Protection system:

–30 to +100°C typ. 5ppm/°C

5 to 2000Hz/Amax

= 0.75mm/amax = 20g

50g/11ms

> 100 x 106 strokes

IP 40

Option

Plug connection (instead of fixed connected cable), including 3m cable with screwed round socket and ALMEMO

®

connector

Types

Order no.

Working length/resolution, incl. ALMEMO

®

cable 2m long

25 mm / 0,001 mm

50 mm / 0,01 mm

75 mm / 0,01 mm

FWA025TR

FWA050TR

FWA075TR

Order no.

OWA071AK

100 mm / 0,01 mm

Order no.

FWA100TR

included with delivery

2 tensioning clamps Z3-31 including 4 cap screws M4x10,

1 probe tip with hard-metal ball

10.21

Speed

Rotational Speed Sensor FUA 9192

• Optical probe for measurements of rotational speed, designed as retroreflective photoelectric sensor for photoelectric detection of rotational speeds or events.

• For evaluation of the pulses, the tachometer probe is equipped with a specific frequency meter module that calculates the number of revolutions per minute from the time period between two pulses. A stable read-out is achieved by averaging over a minimum of 500 ms.

• Easy application:

A reflective adhesive tape is attached to the moving part and the probe is aligned with it. For function control purposes a yellow signal lamp at the rear side of the probe will be on when the reflective adhesive tape is recognised.

• To increase the operation reliability the sensitivity can be adjusted through a potentiometer.

Technical Data:

Measuring range:

Bright-up pulse time:

Resolution:

Accuracy:

Detection range:

Sensitivity:

Detectable object:

8 to 30000rpm (maximum)

> 1ms

1rpm up to 15000rpm:

± 0.02% of m.v. ± 1 digit up to 30000rpm:

± 0.05% of m.v. ± 1 digit

20 to 200mm

(depending on the reflector) adjustable with potentiometers opaque or reflector

Distance hysteresis: ≤ 10%

Indication of switching status: LED yellow

Type of light:

Limit for foreign light: red light 660nm sun light: ≤ 20000lux halogen light: ≤ 5000lux

Ambient/storage temperature: –25/–40°C to +55/+70°C

Protection system: IP 67 (accord. to EN 60529)

Optics: 2-lens system PC

Permissible shock load: b ≤ 30g, T ≤ 1ms

Permissible vibrational load: f ≤ 55Hz, a ≤ 1mm

No-load current:

Supply voltage:

≤ 20mA

> 8.5VDC via instrument, mains adapter recommended

Connection:

Material:

Dimensions:

Weight:

Meets standards:

Device connector M12x1 including socket M12x1, angled, with 1.5 meters cable and ALMEMO

®

connector housing: brass, nickel plated, lens opening: PMMA diameter: M12 x 1mm, length: 55mm

15g

EN 60 947-5-2

Accessories

Extension cable, 1 meter long

Extension cable, 2 meters long

Types

For rotational speeds up to 30000rpm max., incl. 5 reflective adhesive tapes

Connecting cable 1.5m long with ALMEMO

®

connector

DAkkS / DKD or factory calibration KU90xx rotational speed for digital sensor (see chapter Calibration certificates).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

10.22

Order no.

ZA9060VK1

ZA9060VK2

Order no.

FUA9192

Flow

Flow sensors for liquids FVA 645 GVx

Variant in stainless steel without any moving parts, with integrated temperature measuring

• Measuring section in robust, industry-quality stainless steel

• Without any moving parts, no wear and tear

• Integrated temperature measuring

• Low pressure loss

• Wide temperature range

• High-speed reaction time

• Using with water and water-glycol mixture

• For heat output measurement in heating systems and cooling plant

Technical Data:

Flow

Measuring principle Pressure pulsation

Kármán vortex street

Measuring range see variants

Accuracy using water as medium at 0 to +100°C

±1.5 % of final value

FVA645GV12QT/40QT: by water-glycol ( 42 % )

30 to +100°C (Viscosity < 4 mm²/s)

±5 % of final value

Resolution see variants

Reaction time (63 %) < 1 s ( < 3 s for FVA645GV12QT)

Temperature

Measuring range

Accuracy

Resolution

0 to +100 °C

±1 K at +25 to +80 °C

±2 K at 0 to +100 °C

0.5 K

Reaction time (63 %) <1 second under flow conditions

50% of final value

Process connection

Pressure

Pressure loss

2x male thread see variants

10 bar (bursting pressure >16 bar)

0.1 bar, typical under flow conditions,

50 % of final value

Suitable conditions

Media Water, water-glycol ( max. 42 % glycol)

FVA645GV12QT/40QT Viscosity < 4 mm²/s,

FVA645GV100QT/200QT Viscosity < 2 mm²/s)

Temp. of medium 0 to +100 °C

Ambient temperature -25 to +60 °C

Ambient humidity up to 95 % RH, non-condensing

Electrical connections

Output signal 2x 0.5 to 3.5 V

Power supply

Connection

5 VDC (±5 %), <10 mA via ALMEMO

®

connector

Sensor with 2.9-meter connecting cable and ALMEMO

®

connector

Fitting length see variants

Materials (in contact with media)

Corrosion-resistant coating EPDM, PPS, PPA 40-GF

Pipe piece Stainless steel 1.4408;

(inside pipe PPA 40-GF)

Variants

Sensor for flow rate and temperature over a measured section, including ALMEMO

®

connecting cable, 2.9 meters

Measuring range

1 to 12 l/min

2 to 40 l/min

5 to 100 l/min

10 to 200 l/min

Resolution

0.06 l/min

0.2 l/min

0.5 l/min

1.0 l/min

Process connection

G 3/4“ male thread

G 3/4“ male thread

G 1“ male thread

G 1 1/4“ male thread

Factory calibration KV91xx flow for sensor (see chapter Calibration certificates)

Fitting length

ca. 110 mm ca. 110 mm ca. 129 mm ca. 137.5 mm

Order no.

FVA645GV12QT

FVA645GV40QT

FVA645GV100QT

FVA645GV200QT

10.23

Flow

Axial turbine fl owmeter for liquids FVA 915 VTH

• For measuring the volume fl ow rate or for dosing tasks with small fl ow rates.

• Extraordinary compact design.

• Wide, usable measuring range.

• Various options for operation:

Cooling water fl ow, medical technology, plastics industry, solar systems, baker‘s equipment, machine tools, catering equipment, photographic laboratory equipment, dispensers, dosing equipment, cooling equipment, heating applications, calorimetry.

swiviel nut

G3/4

Flat gasket

18,5 x 24 x 2

Technical Data:

Nominal diameter

Measuring range

Measuring accuracy

Reproducibility :

±1% of finale value

± 0,2 %

Signal output from 0.3 l/min maximum size of particles in medium 0.5 mm maximum temperature of medium

Nominal pressure PN10

85°C

Process connection

Pressure loss in bar

Protection system

Output signal

Pulse rate / K factor

Resolution

Signal form

Measuring transducer

Supply voltage

DN 15

2 to 40 l / min continuous load max. 20 l/min

G ¾“ external thread and union nuts

Δp = 0.00145 x Q

2

(Q in l/min) approx. 0.6 bar at 20 l / min approx. 2.3 bar at 40 l / min

IP 54

940 pulses / liter

1.1 ml / pulse rectangular signal, NPN, open collector

Hall sensor

4,5 ... 24 V DC

Electrical connection

Materials

pipe section

FV A915 VTH M

FV A915 VTH K

Flat gasket

Turbine cage

Rotating vane

Rotor complements

Axle / bearing

(from ALMEMO

®

device)

4-pin connector M12x1 including PVC line (Tmax =70 °C) with ALMEMO

®

connector brass CuZn36Pb2As plastic PPONoryl GFN3

NBR

PEI ULTEM

PEI ULTEM hard ferrite magnets axle Arcap AP1D with hard metal pins in saphire bearings

Bearing support

Sensor

O-ring

Knurled swivel nut *

Arcap AP1D

PPO Noryl GFN3

NBR

PA GF 30

* not coming into contact with the medium

Types

incl. connecting cable, 6m long with ALMEMO

®

connector turbine body made of brass

Turbine body made of plastic

Factory calibration KV91xx flow for digital sensor (see chapter Calibration certificates)

Order no.

FVA915VTHM

FVA915VTHK

Other designs are available on request

Axial turbine flowmeters FVA 915 VTWx for water-glycol mixture up to

150 °C, 25 bar, 2 to 30 l/min Figure - similar to above

Axial turbine flowmeters FVA915VTPx for water up to 150 °C, 300 bar,

2 to 40 l/min Figure - similar to above

Magnetic-inductive flowmeters FVA 915 VMZx without moving parts, for small flow rates from 0.1 l / min and high flow rates up to 250 l / min

10.24

FVA 915 VMZ

Flow

Axial turbine flowmeter for liquids FVA 915 VTH25

• For measuring the volume flow rate or for dosing tasks with large flow rates.

• Compact design.

• Wide useful operating range.

• Wide variety of applications :

• Cooling water flow, medical technology, plastics industry, solar systems, baker’s equipment, machine tools, catering equipment, photographic laboratory equipment, dispensers, dosing equipment, cooling equipment, heating applications, calorimetry.

Technical Data

Nominal diameter

Measuring range

Continuous load

DN 25

4 to 160 l/min max. 80 l/min

Measuring accuracy

Reproducibility :

±5% of measured value up to 5l/min ±7% of measured value

±0.5%

Signal output from < 1 l/min maximum size of particles in medium 0.63 mm maximum temperature of medium

Nominal pressure PN10

Process connection

FVA915VTH25M

Pressure loss

Protection system

Output signal

Pulse rate / K factor 65 pulses / liter

85°C

G 1¼“ external thread including adapter for R 1“

(absolutely necessary) approx. 0.1 bar at 80 l / min approx. 0.45 bar at 160 l / min

IP 54

Resolution

Signal form

15 ml / pulse

NPN, open collector

Measuring transducer

Supply voltage

Hall sensor

4,5 ... 24 V DC

(from ALMEMO

®

device)

Electrical connection 4-pin connector M12x1 including PVC line (Tmax =70 °C) with ALMEMO

®

connector

Materials

Pipe section FV A915 VTH25M brass, CW602N

Turbine cage

Rotation vane

Rotor complements

Axle / bearing

Sensor socket

O-ring

PPO Noryl GFN 1630V

PPO Noryl GFN 1520V

Hard Ferrite Magnets stainless steel 1.4539 / saphire, PA

PPO Noryl GFN 1630V

EPDM

Type

incl. connecting cable, 6 m long, with ALMEMO

®

connector turbine body made of brass

Factory calibration KV91xx flow for digital sensor (see chapter Calibration certificates)

Order no.

FVA915VTH25M

Other designs are available on request

Axial turbine flowmeters FVA 915 VTH40 6.7 to 417 l/min, DN40

Figure - similar to above

Turbine flowmeters FVA 915 VTRx

Stainless steel, up to 120 °C, up to 250 bar for different flow rates from 1.8 l/min to 1133 l/min

10.25

10.26

Electrical variables

Content

How split-core type transformers work

Split-core type transformer for AC currents

FEA 6049, FEA 604 MN, FEA 6044 N

Measuring module for DC voltages and DC currents ZA9900AB / ZA9901AB

True / effective measuring module for AC voltages and AC currents ZA9903AB / ZA9904AB

ALMEMO

®

input connectors and adapter cables for all sizes see Chapter Input connectors

11.03

11.05

11.06

11.01

Electrical variables

How Split-Core Type Transformers Work

Current transformers are used to acquire high alternating currents without contact and without interrupting the circuit.

In principle, they consist of 2 separate transformator windings (B winding with N

1

windings, B

2

1

= primary

= secondary winding with N

2

windings) on one common iron core (closed magnetic circuit).

If an alternating current I the winding B

1 in the winding B

1

flows through

, a current I

1

2

/N

2

is induced

, which depends on the winding ratio N

2

. In comparison with stationary-installed panel transformers, split-core type transformers must be able to embrace a conductor within a magnetic circuit that is split open.

In practice, the primary winding B

1

consists of only one winding that carries the current to be measured. The transformation ratio of a current transformer is:

I

1

x N

1

= I

2

x N

2

11.02

Electrical variables

Electrical variables

Split-Core Type Transformer for AC Currents FEA 6049

• Perfectly suitable for use in maintenance and monitoring of electrical systems without interrupting their current supply.

• Application oriented design, particularly suitable for measurement in dense wiring.

• Ideal for non-contact control measurements with ALMEMO

® hand-held devices, e.g. for fault currents or at devices with low current consumption.

Technical Data

Measuring range:

Accuracy of meas. at 50/60Hz:

1A to 150A AC

40 to 150A: ± 4%

15 to 40A: ± 3% ± 0.2A

5 to 15A: ± 6% ± 0.2A

1 to 5A: ± 10% ± 0.2A

Encompassing capacity: cable Ø 10mm

Transformation ratio: 100mVDC/1A AC

Output signal:

Nominal conditions

Electrical safety

15VDC

23°C ±3K, 1013 mbar, 20 to 75% RH

EN 61010-2-032 (issue 2/2003)

Admissible voltage

Operating frequency

Operating conditions

Dimensions

300 V category IV or 600 V category III

48 to 500 Hz

-10 to +50°C, 10 to 85% RH

130 x 37 x 25 mm

Weight

Storage temperature approx. 180 grams

-40 to +80°C

Connecting cable Cable, 1.5 meters, with safety laboratory connectors, including safety coupling and 1.5-meter ALMEMO

®

connecting cable with banana plugs

Types

(including manufacturer’s test certificate)

Single-range split-core type transformer with integrated rectifying for small AC currents incl. ALMEMO

® connecting cable (±26VDC)

Order no.

FEA6049

DAkkS or factory calibration KE90xx electrical for sensor (see chapter Calibration certificates).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

Split-Core Type Transformer for AC Currents FEA 604 MN

• Perfectly suitable for use in maintenance and monitoring of electrical systems without interrupting their current supply.

• Asymmetric shape of the jaw of tongs, particularly suitable for encompassing cables and rails.

• With polarity indicator for power measurements.

• Ideal for non-contact control measurements with ALMEMO

® handheld devices, e.g. at low power systems.

Technical Data

Measuring range: 0.5A to 200A AC

(the higher value corresponds to 120% of the max. nominal value)

Accuracy of meas. at 50Hz: ± 3% of meas. val. ±0.5A

Encompassing capacity: cable Ø 20mm rail 20 x 5mm

Transformation ratio:

Output signal:

100mVDC/1A AC

20VDC

Operating frequency:

Safety standards:

40Hz to 10kHz

IEC 1010-1

Overvoltage protection: category III

Dimensions:

Weight:

135 x 50 x 30mm approx. 180g

Nominal conditions: 25°C ±3°C/1013mbar

Operating temperature: –10 to +55°C

Relative humidity:

Storage temperature:

0% to 90% at 40°C max.

–40 to +70°C

Connecting cable: Connecting cable Integrated banana sockets, including 1.5-meter ALMEMO connecting cable with banana plugs

Types

(including manufacturer’s test certificate)

Single-range split-core type transformer with integrated rectifying for small AC currents incl. ALMEMO

® connecting cable (±26VDC)

DAkkS or factory calibration KE90xx electrical for sensor (see chapter Calibration certificates).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

Order no.

FEA604MN

11.03

Electrical variables

Split-Core Type Transformer for AC Currents FEA 6044 N

• Perfectly suitable for use in maintenance and monitoring of electrical systems without interrupting their current supply.

• Asymmetric shape of the jaw of tongs, particularly suitable for encompassing cables and rails.

• With polarity indicator for power measurements.

• Ideal for non-contact control measurements with ALMEMO

® handheld devices, e.g. at low power systems.

Technical Data

Measuring range: 2A to 500A AC

(the higher value corresponds to 120% of the max. nominal value)

Accuracy of meas. at 50Hz: ± 3% of meas. val. ±0.5A

Encompassing capacity: cable Ø 30mm rail 30 x 63mm

Transformation ratio: 1mVDC/1A AC

Output signal:

Operating frequency:

0.5VDC

40Hz to 1kHz

Safety standards: IEC 348, IEC 1010-2-032

Overvoltage protection: no

Dimensions: 215 x 66 x 34mm

Types

(including manufacturer’s test certificate)

Weight: approx. 420g

Nominal conditions: 25°C ±3°C/1013mbar

Operating temperature: –10 to +55°C

Relative humidity:

Storage temperature:

0% to 90% at 40°C max.

–40 to +70°C

Connecting cable: Cable, 1.5 meters, with safety laboratory connectors, including safety coupling and 1.5-meter ALMEMO

®

connecting cable with banana plugs

Single-range split-core type transformer with integrated rectifying for small and medium AC currents incl. ALMEMO

®

connecting cable (±2.6VDC)

DAkkS or factory calibration KE90xx electrical for sensor (see chapter Calibration certificates).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

Order no.

FEA6044N

11.04

Electrical variables

ALMEMO

®

Measuring Modules for DC Voltage and DC Current ZA 9900 AB / ZA 9901 AB

• Acquisition of the momentary, maximum, minimum and average value, plus transferring data of each measuring point scan to the ALMEMO

®

device.

• DC voltage or DC current signal are scanned with 1kHz.

• Pure digital data transmission to the measuring instrument.

• Connector sockets electrically isolated and overvoltage-protected.

Technical Data

Accuracy:

Sampling rate:

Resolution:

0.1% of fin. val. ±2 digits

1kHz

12bit, ±2048 digits

Meas. period/transient time: 0.1s

Meas. cycle, maximum: 14h

Electrical isolation: 1kV permanent, 4kV for 1s

Housing:

Sockets:

Operating voltage:

Current consumption: polystyrene, dimensions L100 x W54 x H31mm touchproof, Ø 4mm

6 … 14V through ALMEMO

®

device

< 40mA (connector and module)

Types

(incl. touchproof connecting cable)

DC Voltage:

Measuring range

±2.000 V*

±20.00 V

±200.0 V

±400 V

Resolution

0.001V

0.01V

0.1V

1V

Overload

±400 V

±500 V

±500 V

±1000 V

Internal resistance

800 kΩ

1 MΩ

1 MΩ

4 MΩ

Order no.

ZA9900AB2

ZA9900AB3

ZA9900AB4

ZA9900AB5

DC Current:

Measuring range

±20.00 mA

±200.0 mA

±2.000 A

±10.00 A

Resolution

0.01mA

0.1mA

0.001A

Overload

±0.1 A*

±1 A*

±10 A*

Internal resistance

10 Ω

1 Ω

0.1 Ω

0.01A ±20 A* 0.01 Ω

*Without fuse. overload condition only up to 1 minute maximum

ZA9901AB1

ZA9901AB2

ZA9901AB3

ZA9901AB4

DC via external shunt:

±200.0 mV 0.1mV ±40 V 50 kΩ

DAkkS or factory calibration KE90xx electrical for digital measuring module (see chapter Calibration certificates).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

ZA9900AB1

11.05

Electrical variables

True/Effective Measuring Modules for AC Voltages and AC Current ZA 9903 AB / ZA 9904 AB

• Independent, full digital acquisition of the true/effective values of an AC variable.

• Measuring signals with any course of a curve are digitised with 1kHz.

• Pure digital data transmission to the measuring instrument.

• Acquisition of the frequency through a second measuring channel.

• Connector sockets electrically isolated and overvoltageprotected.

Technical Data

TRMS

Accuracy:

Sampling rate:

0.1% of fin. val. ± 2 digits

1kHz

Resolution:

Frequency range:

12 bit, ± 2048 digits for Uss

20.0 … 250Hz

Meas. period/transient time: 0.5s

Frequency

Accuracy:

Sampling rate:

Resolution:

Sensitivity:

± 0.1Hz

1kHZ

0.1Hz

10% of final value

Frequency range: 20.0 … 250Hz

Meas. period/transient time: 0.5s

Electrical isolation:

Housing:

Sockets:

Operating voltage:

Current consumption:

1kV permanent, 4kV for 1s polystyrene, dim. L 100 x W 54 x H 31mm touchproof, Ø 4mm

6 … 14V through ALMEMO

®

device

< 40mA

(connector and module)

Types

(incl. touchproof connecting cable)

AC Voltage

Meas. range Resolution Peak Overload Internal resistance

Order no.

130.0mV

eff

1.300V

eff

13.00V

130.0V

400V eff eff eff

1)

0.1mV

1mV

10mV

0.1V

1V

±0.2V

±2V

±20V

±200V

±1000V

±400V

±400V

±500V

±500V

±1000V

0.5MΩ

0.8MΩ

1MΩ

1MΩ

4MΩ

ZA9903AB1

ZA9903AB2

ZA9903AB3

ZA9903AB4

ZA9903AB5

1)

When using the measuring module for the purposes of current measurement with an external shunt. the shunt must be looped into the neutral conductor (not into the phase).

AC Current

Meas. range Resolution Peak Overload

1.000A

eff

1mA ±2A ±10A

2)

10.00A

2) eff

10mA ±20A ±20A

Without fuse, overload condition only up to 1 minute maximum

2)

Internal resistance

0.10Ω

0.01Ω

DAkkS or factory calibration KE90xx electrical for digital measuring module (see chapter Calibration certificates).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

11.06

ZA9904AB1

ZA9904AB2

Meteorology

Content

Small glossary of meteorological terms

Meteo-Multisensor FMA-510

Mobile weather station

Wind direction sensor FVA 614

Wind velocity sensor FVA 615

Rainfall sensor FRA 916

Precipitation detector

Global radiation probe head FLA 613 GS

Radiation probe head FLA 613 VLM

UVA radiation probe head FLA 613 UVA

UVB radiation probe head FLA 613 UVB

Star pyranometer FLA 628S

Digital sensor for temperature, humidity, atmospheric pressure in protective all-weather housing

Comfort index measurement in the workplace

WBGT measurement for evaluating heat-exposed workplaces

12.02

12.03

12.04

12.06

12.07

12.07

12.08

12.09

12.09

12.10

12.10

12.11

12.12

12.14

12.15

12.01

Meteorology

Compact Glossary of Meteorological Terms

Response value The wind velocity at which the cup or the wind vane starts to move.

Barometer General term for the device measuring the atmospheric pressure.

Barometric pressure Pascal [Pa] = Newton per square meter [N/m

2

]; 1hPa=1mbar; 1 bar=10

5

Pa

Beaufort

Damping ratio

Classification for certain wind velocity ranges: bft m/s bft m/s bft m/s bft m/s bft m/s bft m/s

0 0 - 0.2 1 0.3- 1.5 2 1.6- 3.3 3 3.4- 5.4 4 5.5- 7.9 5 8.0-10.7

6 10.8-13.8 7 13.9-17.1 8 17.2-20.7 9 20.8-24.4 10 24.5-28.4 11 28.5-32.6

12 32.7-36.9 13 37.0-41.4 14 41.5-46.1 15 46.2-50.9 16 51.0-56.0 17 56.1-61.2

Measure for the damping of wind vanes. It is the ratio of successive damped deflection amplitudes

(e.g. 3rd to 1st amplitude) in one direction.

Distance constant

Gray code

Altitude formula

Detection limit

Is the distance that has been passed by the wind and which is reached when, after a sudden change of the wind velocity, the velocity has reached 63% of the final value.

One step digital code used for the wind direction.

Mathematical reduction of the barometric air pressure to a reference altitude, at minimum to sea level

(QFF). Example: with each altitude increase of 8m the pressure decreases by approximately 1hPa.

The lowest value of the wind velocity and wind direction where a stable measured value is established.

Normal pressure

QFE

QFF

QNH

The barometric normal pressure (1013.25hPa) that, according to DIN ISO 2533, serves as base value for the ´high pressure´ and ´low pressure´ data.

The atmospheric pressure that has been reduced to the elevation of an airport runway.

Designation used in aviation for the barometric air pressure that has been reduced to sea level (0m).

Also serves as a common base for the barometric air pressure comparison of different weather stations with different elevations of the stations and it is the base for the presentation of the isobars in weather maps.

Designation commonly used in aviation for the barometric air pressure, which has to be entered into an altimeter as an initial value so the altimeter can indicate the altitude above sea level.

Altitude of station The local elevation regarding the installation of the measuring station incl. the barometer above sea level.

Variation The range in which the wind direction has been changing within the preceding 10 minutes (acc. to ICAO).

Wind velocity

Wind direction

Usual practical units: 1m/s = 3.6km/h = 1.9455knots

Specification of which direction the wind comes from. The specification is based on a clockwise setup starting from North to East (90°), South (180°) and West (270°) to North (360°).

Wind travel Is the distance travelled by the wind during a certain period.

12.02

Meteorology

Meteorology

Meteo-Multisensor FMA 510

Meteo multisensor is a compact and light-weight multi-sensor system for measuring all important meteorological variables.

The system can be freely configured to measure temperature, relative humidity, atmospheric pressure, wind velocity, wind direction, and rainfall.

• Eight essential weather parameters all combined in one device.

• Stable and accurate measured results.

• No moving parts.

• Low power consumption.

• Compact and light-weight.

• Quick and easy to set up.

• Low maintenance requirements.

!

This sensor is connected to two input sockets.

To the output of the sensor values a cycle must have started (in the measuring instrument or in the software). The functions of this sensor supported by the devices V6

2590, 2690, 2890, 8590-9 8690-9, 5690-1 2 and devices

V5 (only with the function pressure/measuring cycle).

Operation with the device in SLEEP mode is not possible!

Technical Data

Wind direction

Azimuth

Accuracy

0 to 360 °,resolution: 1°, with average value

±3°

Wind velocity

Range

Accuracy

0,5 to 60 m/s, resolution: 0,1 m/s, with max. value and average value

0 to 35 m/s ± 0,3 m/s or ± 3%, whichever is the largest

36 to 60 m/s ± 5%,

Barometric Pressure

Range

Accuracy

600 to 1100 mbar, resolution: 0,1 mbar

±0,5 mbar at 0 to 30 °C

±1 mbar at -52 to +60 °C

Air temperature

Range

Accuracy

Relative humidity

Range

Accuracy

-52 to 60 °C, resolution: 0,1 K

± 0,3 K at 20 °C (sensor element)

0 to 100 % r.H., resolution: 0,1% r.H.

± 3% r.H. at 0 to 90 % r.H.,

± 5% r.H. at 90 to 100 %

Rainfall - quantity

Surface area measured: 60 cm², resolution: 0,01 mm with sum value

Accuracy* ±5% of daily total, depending on weather conditions

Accessorie

Mounting adapter (mobile weather station see 12.04)

Rainfall-intensity

Range

Dimensions

Height

Diameter

Weight

Cable

Powersupply

0 to 200 mm/h, resolution: 0,01 mm/h with maximum value

240 mm

120 mm

620 g

Sensor cable, fixed, 12 m long with

2 ALMEMO

® digital input cable, 0.3 m

6 to 12V, 22mA from the ALMEMO

Heating (only FMA510H) 12 V DC max. 1.1A

or 24 V DC/AC max. 0.6A

Mounting

direct mounted on cross arm or tube with external diameter Ø 30mm and internal diameter >= Ø 24mm with adapter ZB9510MA27 mounted on tube with external diameter Ø 27 or Ø 30 mm

®

device

* Due to the of the phenomenon, deviations caused by spatial variations may exist in precipitation readings, especially in short time scale. The accuracy specification does not include possible wind induced error.

Order no.

ZB9510MA27

Types

(incl. factory test certificate)

Order no.

Meteo-Multisensor FMA510, sensor cable, fixed, 12 m long with 2 ALMEMO

® digital input cable, 0.3 m

Meteo-Multisensor FMA510, sensor cable, fixed, 12 m long with 2 ALMEMO

® digital input cable, 0.3 m with heating incl. cable, fixed, 12 m long (mains adapter not included)

Factory calibration KH92xx temperature, humidity, atmospheric pressure for digital sensor (see chapter Calibration certificates)

FMA510

FMA510H

12.03

Meteorology

Mobile weather station

Universal mobile weather station for measuring a wide array of meteorological data, e.g. wind direction, wind velocity , relative atmospheric humidity, temperature, atmospheric pressure, rainfall quantity and intensity, and global radiation Quick and easy to install, robust design, and various power supply options

(rechargeable battery, solar cell, car adapter)

Applications :

• Vehicle test tracks

• Racing tracks

• Sporting events

• Site evaluation for wind power plants

• Mobile helicopter landing fields

• Tracing industrial emissions

• Disaster control (tracing clouds of poisonous gas, observing local weather developments)

• Agricultural trials

Mobile weather station with data logger ALMEMO

®

2690-8A

Components

• ALMEMO

®

2690-8A data logger (New resolution, integrated atmospheric pressure sensor and NiMH rechargeable battery pack) including connector mains unit 90 to 260 VAC .

• Weather-proof housing with lockable transparent door, Data logger mounted on DIN rail, Continuous power supply for data logger and Meteo sensor via external supply voltage

Supply 230 VAC : Integrated socket with connecting cable led out, approx. 1.7 meters, for 230 V, with safety plug

Power supply 10 to 30 VDC Two integrated banana sockets, wired to clamp terminal inside housing (cable to external mains unit / rechargeable battery - to be provided by customer) Short-term bridging in the event of power supply failure by means of internal rechargeable battery in ALMEMO

®

2690-8A (New variant).

• For supply 10 to 30 V : ALMEMO

®

supply cable ZA2690UK, electrically insulated, for external rechargeable battery / battery 9 to 12 V, ALMEMO

®

supply cable ZA1012AK, not electrically insulated.

Types Order no.

Meteo sensor for measuring wind direction, wind velocity , relative humidity, temperature, atmospheric pressure, rainfall quantity and intensity, plus 12 meters cable, with 2 ALMEMO

®

plug-in connectors

Probe head for measuring global radiation, 0 to 1200 W/m

2

, with 1.5 meters cable

Longer cable, total length 5 meters

Mobile tripod stand, extendable up to 3.5 meters, with mountable adapter for Meteo sensor FMA510, including set of guys and anchoring fixtures (comprising 3 spring-snap hooks, guy lines (4 meters),

FMA510

FLA613GS

OA9613K05

and ground pegs) Dimensions (non-extended) approx. 1.6 x 0.15 meters - weight approx. 11 kg

Holder for 1 radiation probe head FLA613GS / VLM / UVA / UVB - length approx. 0.5 meters

Carry case (with space for 1 tripod stand including accessories and up to 2 probe head holders)

Data logger set ALMEMO

®

2690-8A (New variant) including connector mains unit and USB data cable

ZB9510ST

ZB9510MH

ZB9510TT

MA26908AKSU

ALMEMO

®

memory connector, with micro SD card including USB card reader

ALMEMO

®

supply cable, 10 to 30 VDC, output 12 VDC 1 A, electrically insulated

ZA1904SD

ZA2690UK

ALMEMO

®

supply cable, 9 to 12 VDC, not electrically insulated

ZA1012AK

Weather-proof housing with lockable transparent door, cable bushings and mast fixture, supply cable led out, approx. 1.7 m, for 230 V, with safety plug, including ALMEMO

®

2690-8 data logger installed on DIN rail (must be ordered separately)

Housing material ABS (acrylonitrile butadiene styrene), 300 x 250 x170 mm (excluding mast fixture), weight

(including measuring instrument) approx. 3.5 kg

Carry case, universal, spacious, robust Exterior dimensions (WxHxD) approx. 51 x 35 x 30 cm

ZB9015AGA

ZB5600TK3

12.04

Meteorology

Mobile weather station with ALMEMO

®

8590-9 measuring module

Components

• Data logger ALMEMO

®

8590-9 including connector mains unit 90 to 260 VAC.

• Weather-proof housing with lockable opaque door, Data logger mounted on DIN rail, Continuous power supply for data logger and Meteo sensor via external supply voltage

Supply 230 VAC : Integrated socket with connecting cable led out, approx. 1.7 meters, for 230 V, with safety plug

Power supply 10 to 30 VDC : 2 integrated banana sockets, wired to clamp connector inside housing (cable to external mains unit / rechargeable battery - to be provided by customer).

• For supply 10 to 30 V : ALMEMO

®

supply cable ZB3090UK, electrically insulated, for external rechargeable battery / battery 9 to 12 V, ALMEMO

®

supply cable ZB5090EK, not electrically insulated.

• Weather-proof housing, with solar power supply, available on request.

Types Order no.

Meteo sensor for measuring wind direction, wind velocity , relative humidity, temperature, atmospheric pressure, rainfall quantity and intensity, plus 12 meters cable, with 2 ALMEMO

®

plug-in connectors

Probe head for measuring global radiation, 0 to 1200 W/m

2

, with 1.5 meters cable

Longer cable, total length 5 meters

Mobile tripod stand, extendable up to 3.5 meters, with mountable adapter for Meteo sensor FMA510, including set of guys and anchoring fixtures (comprising 3 spring-snap hooks, guy lines (4 meters),

Order no. FMA510

FLA613GS

OA9613K05

and ground pegs) Dimensions (non-extended) approx. 1.6 x 0.15 meters - weight approx. 11 kg

Holder for 1 radiation probe head FLA613GS / VLM / UVA / UVB - length approx. 0.5 meters

Carry case (with space for 1 tripod stand including accessories and up to 2 probe head holders)

ALMEMO

®

8590-9 measuring instrument, including connector mains unit 90 to 260 VAC

ZB9510ST

ZB9510MH

ZB9510TT

MA85909

ALMEMO

®

memory connector, with micro SD card including USB card reader

ALMEMO

®

supply cable, 10 to 30 VDC, output 12 VDC 0.2 A, electrically insulated

ZA1904SD

ZB3090UK

ALMEMO

®

supply cable, 9 to 12 VDC, not electrically insulated

ZB5090EK

Weather-proof housing with lockable opaque door, cable bushings and mast fixture, supply cable led out, approx. 1.7 meters, for 230 V, with safety plug, including ALMEMO

®

8590-9 data logger installed on DIN rail (must be ordered separately)

Housing material ABS (acrylonitrile butadiene styrene), 300 x 250 x170 mm (excluding mast fixture), weight (including measuring instrument) approx. 3.5 kg

Carry case, universal, spacious, robust Exterior dimensions (WxHxD) approx. 51 x 35 x 30 cm

ZB9015AGB

ZB5600TK3

Mobile weather station

12.05

Meteorology

Wind Direction Sensor FVA 614

• Wind direction sensor for measuring the horizontal wind direction.

• Wind vane made from robust plastic, electronics in weatherresistant aluminum housing, rotating mechanism on friction bearings.

• A special labyrinth reliably protects without friction and guards against water penetrating into the housing.

• Electronically controlled heating for operation in winter conditions to prevent bearings and external rotating parts from freezing.

Technical Data

Measuring range:

Accuracy:

0 to 360°

±5°

Resolution: 11.25° (5 bit Gray code)

Measuring principle: optoelectronically (slotted disk)

Sensor power supply: 9–30VDC through ALMEMO

®

device

Heating: 24VAC/DC max. 20W

Operative range:

Cable:

-30 to +70 °C, with heating

12m long, LiYCY 6 x 0.25mm

2

Connection:

Installation:

Weight

Adapter cable with ALMEMO

®

connector including supply cable for heating

(length 1.5 m, free ends)

A mains supply unit must be provided by the user on site.

e.g. pole tube with holding thread

PG21 / drilling 29mm Ø

1100 g

Type

Wind vane including ALMEMO

®

connector (0–2V) with 12m cable

Order no.

FVA614

Accessories for wind direction and wind velocity sensors

Cross-arm for separate wind direction and wind velocity sensors inclusive assembly utilitis for mast Ø 48 to 102 mm

Lightning protection rod

Order no.

ZB9015TC

ZB9015BS

12.06

Meteorology

Wind Velocity Sensor FVA 615 2

Technical Data

Measuring range:

Accuracy:

Resolution:

0.5 to 50m/s

±0.5m/s ±3% of meas. value

0.1m/s

Measuring principle: optoelectronically (slotted disk)

Sensor power supply: 9–30VDC through ALMEMO

®

device

Heating:

Operative range:

Cable:

24VAC/DC max. 20W

-30 to +70 °C, with heating

12m long, LiYCY 6 x 0.25mm2

• Wind velocity sensor for measuring the horizontal wind velocity.

• Cup-type made from robust plastic, electronics in weatherresistant aluminum housing, rotating mechanism on friction bearings.

• A special labyrinth reliably protects without friction and guards against water penetrating into the housing.

• Electronically controlled heating for operation in winter conditions to prevent bearings and external rotating parts from freezing.

Connection:

Installation:

Weight

Adapter cable with ALMEMO

®

connector including supply cable for heating

(length 1.5 m, free ends)

A mains supply unit must be provided by the user on site.

e.g. pole tube with holding thread

PG21 / drilling 29mm Ø

750 g

Type

Cup-type anemometer including ALMEMO

®

connector (0–2V) with 12m cable

Order no.

FVA6152

12.07

Meteorology

Precipitation detector, FRA 616 D and FR 8616 D

• The sensor reacts to precipitation in the form of either rain or snow within just a few seconds.

• It detects even very slight precipitation.

• The precipitation detector reacts by switching a relay. It does not provide a continuous measuring signal; it operates with a step function :

If it detects precipitation, display in ALMEMO

®

measuring instrument : 1.0000, if it does not detect precipitation, display in ALMEMO

® measuring instrument : 0.0000.

• The precipitation detector is designed for use for example in automatic ventilation or shading systems, or in automatically controlled greenhouses, etc.

Technical Data

Voltage connection 230 V AC ±10% 6 VA (50/60 Hz)

Power draw

Electronics

Preheating

Total heating

3 VA

1 VA

3 VA

Admissible ambient temperature

Storage temperature

-30 to +60 °C

-30 to +70 °C

Relative humidity 0 to 100 %

Relay drop-out delay 5 minutes ± 15%

Test voltage

Terminals L or N  Electronics 1.5 kV

Electronics  Relay contacts 1.5 kV

Electromagnetic compatibility

EN50081-1; EN50082-2;

EN61010-1

Relay output

Duty classification

Housing

Material

Protection system

Mounting system

Weight

Connection

FR8616D

FRA616D

250 V AC, max. 4 A,

300 VA inductive approx. 1 million operations polycarbonate, gray

IP65

Tubular steel pole, diameter approx. 25 to 50 mm approx 0.8 kg

(incl. mounting materials) with connecting terminals with ALMEMO

®

connector and

12-meter connection cable

Types

Precipitation sensor including mounting materials

Precipitation sensor including mounting materials, ALMEMO

®

connector, and 12-meter cable

12.08

Order no.

FR8616D

FRA616D

Meteorology

Global Radiation Probe Head FLA 613 GS

• Measuring head in anodized aluminium housing with a plastic dome that is transparent to UV light.

• Rain and splash-proof system, additionally with desiccant to prevent dome from inside condensation.

• Particularly suitable for outdoor measurements, e.g. in medical and biological research, weather information and forecast systems, climatology, agriculture and for general public information.

Technical Data

Measuring range:

Spectral sensitivity:

Maximum spectral sensitivity:

Signal output:

0 to approx. 1200W/m

2

400nm to 1100nm

780nm

0V to 2V

Power supply:

Mounting:

Cable passage:

Housing:

Diffusor:

Dome:

+5V to +15V

2 screws M4, in base plate downwards anodized aluminium

PTFE

PMMA

Option

Longer cable Total length = 5 meters

Cos correction:

Linearity: error f2 < 3%

< 1%

Absolute error: < 10%

Residual voltage: (E = 0) < 10mV

Nominal temperature: 22°C ±2°C

Operating temperature: –20°C to +60°C

Dimensions:

Weight: housing: 55 mm high dome 40 mm high diameter: 80 mm approx. 300 g

Order no.

OA9613K05

Type

(including test protocol)

Order no.

Weather-proof measuring head for measuring the global radiation, incl. ALMEMO

®

connector with 1.5m cable

FLA613GS

Factory calibration KL90xx radiation for sensor (see chapter Calibration certificates)

Illuminance measuring head FLA 613 VLM

• Measuring head in anodized aluminum housing, with UVtransparent plastic dome.

• Rain-proof, splash-protected system, with desiccant to prevent condensation forming on the inside of the dome.

• Especially suitable for measuring operations outdoors, e.g. in medical, biological, and climate research, in weather information forecast systems, in agriculture, and for the purposes of general information for the public.

• The spectral sensitivity of the receiver corresponds approximately to that of the human eye.

Technical Data

Measuring range : 0 to 170 klux (approx. 250 W/m

2

)

Spectral sensitivity : 360 to 760 nm

Max. spectral sensitivity : 550 nm

Signal output

Power supply :

0 to 2 V

+5 to +15 V

Mounting :

Cable passage :

Housing :

Diffusor :

Dome :

2 screws, M4, in base plate downwards anodized aluminum

PTFE

PMMA

Cos correction : error f2 <3%

Linearity :

Absolute error :

<1%

< 10 %

Residual voltage (E = 0) : <10 mV

Nominal temperature : 22 ± 2 °C

Operating temperature : -20 to +60 °C

Dimensions :

Weight :

Housing : 55 mm high

Dome : 40 mm high

Diameter : 80 mm approx. 300 g

Type

(including test protocol)

Weather-resistant measuring head for measuring the illuminance including cable, 1.5 m, and ALMEMO

®

connector

Factory calibration KL90xx radiation for sensor (see chapter Calibration certificates)

Order no.

FLA613VLM

12.09

Meteorology

UVA Radiation Probe Head FLA 613 UVA

• Measuring head in anodized aluminium housing with a plastic dome that is transparent to UV light.

• Rain and splash-proof system, additionally with desiccant to prevent dome from inside condensation.

• Particularly suitable for outdoor measurements, e.g. in medical and biological research, weather information and forecast systems, climatology, agriculture and for general public information.

Technical Data

Measuring range:

Spectral sensitivity:

Maximum spectral sensitivity:

Signal output:

0 to approx. 3mW/cm

2

310 to 400nm

335nm

0V to 2V

Power supply:

Mounting:

Cable passage:

Housing:

Diffusor:

Dome:

+5V to +15V

2 screws M4, in base plate downwards anodized aluminium

PTFE

PMMA (transparent to UV)

Cos correction:

Linearity: error f2 < 3%

< 1%

Absolute error: < 10%

Residual voltage: (E = 0) < 10mV

Nominal temperature: 22°C ±2°C

Operating temperature: –20°C to +60°C

Dimensions:

Weight: housing: 55 mm high dome 40 mm high diameter: 80 mm approx. 300 g

Type

(including test protocol)

Weather-proof measuring head for measuring the UVA radiation including cable, 1.5 m, and ALMEMO

®

connector

Factory calibration KL90xx radiation for sensor (see chapter Calibration certificates)

Order no.

FLA613UVA

UVB RadiationProbe Head FLA 613 UVB

• Measuring head in anodized aluminium housing with a plastic dome that is transparent to UV light.

• Rain and splash-proof system, additionally with desiccant to prevent dome from inside condensation.

• Particularly suitable for outdoor measurements, e.g. in medical and biological research, weather information and forecast systems, climatology, agriculture and for general public information.

Technical Data

Measuring range:

Spectral sensitivity:

0 to approx. 50mW/cm

2

265 to 315nm

Maximum spectral sensitivity:

Signal output:

Power supply:

0V to 2V

297nm

+5V to +15V

Mounting:

Cable passage:

2 screws M4, in base plate downwards

Housing:

Diffusor:

Dome: anodized aluminium

PTFE

PMMA (transparent to UV)

Cos correction:

Linearity: error f2 < 3%

< 1%

Absolute error: < 10%

Residual voltage: (E = 0) < 10mV

Nominal temperature: 22°C ±2°C

Operating temperature: –20°C to +60°C

Dimensions:

Weight: housing: 55 mm high dome 40 mm high diameter: 80 mm approx. 300 g

Type

(including test protocol)

Weather-proof measuring head for measuring the UVB radiation including cable, 1.5 m, and ALMEMO

®

connector

Factory calibration KL90xx radiation for sensor (see chapter Calibration certificates)

12.10

Order no.

FLA613UVB

Meteorology

Star Pyranometer FLA 628 S

• Star pyranometer, according to Dirmhirn, for measuring the global radiation, the sky radiation and the short-wave radiation.

• Independent from ambient temperature through differential temperature measurement.

• Cut precision glass cupola for shielding from external environmental effects.

• Levelling by 3 setting screws and an integrated bubble

Technical Data

Measuring range:

Resolution:

Spectral range:

Output:

Impedance:

0 to 1500W/m

2

0.1W/m

2

0.3 to 3µm approx. 15mV/Wm

-2 approx. 35ohms

Operative range:

Accuracy:

–40 to +60°C cosine effect + azimuth effect + tempera- ture influence

Cosine effect: <3% of measured value

(0 to 80° inclination)

Inclination azimuth effect: < 3% of meas. val.

Temperature influence: < 1% of meas. val. (–20 to +40°C)

Nominal temperature:

Linearity:

Stability:

Settling time:

Dimensions:

Weight:

Accessories

Shadow belt with stand

22°C ±2°C

<0.5% (0.5 to 1330W/m

2

)

<1% of the meas. range per year

25s (t

95

)

160mm Ø, 75mm high, hole circle: 134mm Ø, holes: 8mm Ø

1 kg

Order no.

ZB9628SB

Type

(including test protocol)

Star pyranometer including 3m cable with ALMEMO

®

connector and programmed calibration value

Factory calibration KL90xx radiation for sensor (see chapter Calibration certificates)

Order no.

FLA628S

Other variants are available on request

Probe for measuring global radiation FLA 613 T1B11,

3-mode sensor : It measures UVA, VIS, IRA radiation.

Spectral sensitivity from 315 to 1100 nm

Probe for measuring global radiation FLA 613 GS-SDEK,

This measures the global, direct, and diffused solar radiation

(integrated shadow bar).

Spectral sensitivity from 380 to 1100 nm

12.11

Meteorology

ALMEMO

®

D6

Digital sensor for temperature, humidity, atmospheric pressure FHAD 46-4AG in protective all-weather housing with ALMEMO

®

D6 plug

On request

new:

Temperature sensor Pt100 in protective all-weather housing FPA930AG

• new:

All relevant ambient parameters are measured with one sensor.

• Suitable for mounting on a wall or a mast

• new

Sensor cable up to 100 meters long, clamped in terminal box

• Digital capacitive humidity sensor with integrated signal processor

• All sensor characteristics and adjustment data are stored in the humidity sensor element itself.

• Humidity sensor element, plug-in

• Spare elements are inexpensive; a replacement can be fitted on site quickly and easily by virtually anyone; it will be fully accurate and need no special adjustment.

• new

Automatic atmospheric pressure compensation is provided for pressure-dependent humidity variables by means of a digital atmospheric pressure sensor integrated in the terminal box.

• new

Humidity calculation on the basis of formulae as per Dr.

Sonntag and the enhancement factor as per W. Bögel (correction factor fw(t,p) for real mixed gas systems) This substantially widens the measuring range and improves the accuracy of humidity variable calculations.

• new

Humidity variable, Absolute humidity in g/m³

• The humidity variables are calculated from the three primary measuring channels (real measurable variables) : Temperature, relative humidity. and atmospheric pressure.

• Four measuring channels are programmed (at our factory):

Temperature (°C, T,t), Relative humidity (%H, RH, Uw),

Dew point (°C, DT, td), Atmospheric pressure (mbar, AP, p),

Other humidity variables can also be selected:

Mixture (g/kg, MH, r), Absolute humidity (g/m³, AH, dv),

Vapor pressure (mbar, VP, e), Enthalpy (kJ/kg, En, h).

This device can be configured on a PC using USB adapter cable ZA 1919 AKUV.

(see “General accessories for ALMEMO

®

D6 sensors” page

04.05).

Technical Data

Operative range

-30 to +60 °C, 5 to 98 % RH

Digital temperature / humidity sensor (including A/D converter)

Humidity

Measuring range

Sensor

Accuracy

0 to 100 % RH

CMOSens

®

technology

±1.8 % RH in range 10 to 90 % RH at nominal temperature

Hysteresis typical ±1 % RH

Nominal temperature 25 °C

Sensor operating pressure Atmospheric pressure

Temperature

Sensor

Accuracy

Reproducibility

CMOSens

®

technology

±0.3 K at +25 °C

±0.4 K at +10 to +40 °C

±1.3 K at -20 to +80 °C typical ±0.1 K

Digital atmospheric pressure sensor (integrated in the terminal box)

Measuring range

Accuracy

700 to 1100 mbar

±2.5 mbar (at 0 to +65 °C)

ALMEMO

®

connecting cable

PVC, for available lengths see variants

with ALMEMO

®

D6 plug

ALMEMO

®

D6 plug

Refresh time

Supply voltage

2 second for all four channels

6 to 13 VDC

Current consumption 12 mA

Mechanical design

Sensor tube

Filter cap

Plastic, diameter 12 mm

Metal-mesh filter, SK7

All-weather protection Ø 105 mm, height approx. 110 mm

Terminal box 51 x 53 x 36 mm

Screw-fit cable gland Splash-protected

12.12

ALMEMO

®

D6

Accessories

ALMEMO

®

transmitter 2450-1 with double analog output 10 V or 20 mA

(For other data, options, accessories, see page 01.50)

Meteorology

Order no.

MA24501R02

Standard delivery

Digital sensor for temperature, humidity, atmospheric pressure in protective all-weather housing with connecting cable and ALMEMO

®

D6 plug, manufacturer’s test certificate, 2 fixtures for mounting on a mast

Connecting cable

Order no.

Length = 5 meters

Length = 10 meters

Length = 20 meters

Length = 40 meters

Length = 100 meters

Replacement sensor element, digital, adjusted, plug-in

FHAD464AGL05

FHAD464AGL10

FHAD464AGL20

FHAD464AGL40

FHAD464AGL100

FH0D46

DAkkS or factory calibration KH9xxx, temperature, humidity, and KD92xx, atmospheric pressure, for digital sensor

(see chapter Calibration certificates).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

12.13

Room air conditions

Comfort Index Measurement

Technical features

• Thermal comfort and air-conditioning calculations using

WinControl software with add-on module for comfort index measurement as per DIN ISO 7730 and DIN EN 13779 (formerly DIN 1946)

• Independent measuring sequence in real-time mode

• Various display and output options Real-time mode, memory access to offline measuring operations

• Graphical presentation of measured data and calculated data in a format with data export options

• Comprehensive, clear, meaningful evaluation.

Operative range

It is possible with this measuring setup to measure all the physical parameters needed for assessing and evaluating thermal comfort simultaneously on three levels. It reliably evaluates the performance of heating and ventilating systems. The data acquired from the series of measuring operations for operative temperature (globe temperature), room temperature, and room air flow and humidity, and the necessary input parameters (e.g. clothing factor, activity level, mechanical output) is used together to calculate the PMV (predicted mean vote) and PPD (predicted percent dissatisfied) values (as per DIN ISO 7730) and the degree of turbulence ( as per DIN EN 13779, formerly DIN 1946 Part 2); these values are calculated either online or offline using the AMR

WinControl software in conjunction with the add-on module for comfort index measurement.

The software

The averaging number is preset at 200 measuring points but this is variable and can be modified. The PMV and PPD values and the degree of turbulence can be displayed and documented in y/t or x/y diagrams either each one separately or together with other measurable variables. A software wizard is available to guide the user step-by-step through the various settings. If measuring is started online, the first value is indicated after completion of the first 200 measuring operations (as per DIN ISO 7730). These values continue to be calculated, updated, and displayed, and - optionally - also saved and / or exported. (see Chapter 05)

Types

(sensor set for one level)

Globe thermometer

Humidity / temperature sensor

Order no.

FPA805GTS

FHAD4641

Thermo-anemometer, up to 1 m/s, without smoothing, response time 100 ms, including carry case

FVA605TA1OU

Stand for measuring operations at heights of 0.1 to 1.7 meters, including 1 set of instrument holders for 1 level

(traverse including traverse holder and sensor fastening), including carry case

Set of instrument holders for extra levels (as above) optional for assessing air quality Digital carbon dioxide sensor to 10.000 ppm, with handle

ZB1001PPD

ZB1001MH

FYAD00CO2B10

Device selection

ALMEMO

®

2690-8A (new variant) hand-held data logger, 5 inputs, including mains unit and data cable, USB can be used for 1 measuring level (see page 01.22)

MA26908AKSU

ALMEMO

®

2890-9 hand-held data logger, 9 inputs, including mains unit, can be used for 3 measuring levels (see page 01.24)

ALMEMO

®

data cable, USB, electrically insulated

MA28909

ZA1919DKU

PC link via Ethernet, RS232, or wireless with Bluetooth see Chapter 04, ALMEMO

®

networking technology.

Software:

WinControl for 20 measuring points / 1 device including additional module for comfort index measurement

Accessories:

Carry case, universal, spacious, robust, for globe thermometer, humidity sensor, and data logger

Exterior dimensions (WxHxD) approx. 51 x 35 x 30 cm

SW5600WC1

SW5600WCZM1

ZB5600TK3

DAkkS or factory calibration temperature, humidity, air flow, carbon dioxide for sensor (see chapter Calibration certificates).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

12.14

WBGT Measurement

Room air conditions

Application Range

The wet bulb globe temperature (WBGT) is the decisive parameter for evaluating the work stress at heat-exposed working places and the operation and cool-off times involved. Temperature, radiation and relative humidity are determined by measuring the dry temperature, the natural humid temperature of a psychrometer and the globe temperature of a globe thermometer. These are all combined as WBGT.

Note:

For WBGT measurements the use of a psychrometer with a disengageable ventilator is compulsory

Technical Data

Accuracy:

Sensor:

Globe thermometer:

Class B (DIN/IEC 751)

Pt100 4-conductor, arranged in the center matt black copper globe with suspension

Diameter:

Operating temperature:

Cable length approx. 150mm

–50 to 200°C

3 m

Types

Globe thermometer (Pt100 4L)

Psychrometer with disengageable ventilator

Order no.

FPA805GTS

FNA846WB

DAkkS or factory calibration KT90xx temperature for sensor or measuring chain (sensor + device) (see chapter Calibration certificates).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

NTC-sensor FNA 305

For Indoor air measurements

T

Meas. element NTC

Measuring tip Operative range -10 to +60 °C

90

(non-condensing)

Protective tube in stainless steel

Diameter = 3.0mm, length = 50 mm mounted directly on ALMEMO

8 s

®

connector

L = 50 mm

Order no. FNA305

(No variants available)

12.15

12.16

Building physics, Moisture in materials

Content

Measuring thermal transmittance (U) and heat flow

Heat Flow PLates

Digital heat flow plate

Digital sensors for humidity, temperature, dew point

13.02

13.04

13.05

13.06

Moisture in materials

Moisture Sensor FHA 696 MF

13.07

13.07

Wood moisture probe Hand-held probe for mobile test measurements 13.07

Moisture content sensor - for wood for stationary measuring operations FHA696MFS1 13.08

Moisture content sensor - for wood for stationary measuring operations FHA636MFS1 13.09

Sensor for measuring the moisture content in granulated materials

Dew Point Detector FHA 9461

Water Detection Probe FHA 936 WD

13.10

13.11

13.11

Tensiometer FDA 602 TM1 measurement of soil moisture 13.12

13.01

Building physics, Moisture in materials

Measuring thermal transmittance (U) and heat flow

The heat transfer characteristics of any structural element depend on the thermal conductivity of the materials used, on the thickness of its various component layers, on its structural geometry (e.g. flat or cylindrically curved walls, etc.), and on the ambient conditions at the structure’s surfaces inside and outside.

Presentation of the temperature behavior

The thermal transmittance coefficient (U value) of a structural element describes the quantity of heat that passes through it from one side to the other (no matter how many layers) per second and per square meter surface at a constant difference in ambient temperature inside / outside of 1K. This thermal transmittance coefficient (U) thus also includes the surface heat transfer coefficients, i.e. the thermal energy transferred at the boundary surfaces, interior air - structure - exterior air. The thermal transmittance coefficient (U) is measured in in (W/m²K) and is internationally defined in standard ISO 6946.

A structure’s thermal transmittance coefficient (U) is the reciprocal of its total thermal resistance coefficient (R); R is the sum of the thermal transmission resistances between the structure’s various contiguous layers and also the surface heat transfer resistances between the structure and the ambient media on either side (e.g. air).

Total thermal resistance (R) = thermal transmission resistances through the material + surface heat transfer resistances, inside and out

The thermal transmittance coefficient (U value) is an important rating in civil engineering and the construction industry where it is used to define a building’s transmission heat loss through its various structural elements. Transmission heat loss is the term used to describe the energy-saving qualities of a building’s shell (i.e. the thermal insulation of its roof, outside walls, windows, and floors). In Germany each residential structure is assigned a permissible maximum U value (depending on its external surface area and its internal volume); this is based on the most recently amended version of the Energieeinsparverordnung (EnEV) (German energy-saving legislation)

13.02

Building physics, Moisture in materials Building physics, Moisture in materials

ALMEMO

®

Measuring system for Measuring thermal transmittance (U) and heat flow

The thermal transmittance coefficient (U value) is an important rating in civil engineering and the construction industry where it is used to define a building’s transmission heat loss through its various structural elements. It is now possible, with the ALMEMO

®

measuring system, to measure and record all the physical parameters for the component parts of existing buildings (e.g. walls, etc.) in order to calculate their U value and other relevant thermal energy coefficients.

Measuring principle:

The measuring principle involved in quantifying heat loss at partition elements, e.g. walls, heating systems, etc., is based on the method which uses a heat flow plate (sensor) fitted on the surface of the structural element and thus incorporated directly in the heat flow. Using the known thermal characteristics of the heat flow plate and the thermo-electrically measured temperature gradient inside the heat flow plate the

ALMEMO

®

measuring system can thus measure the heat flow density q in W/m².

The ALMEMO

®

measuring system can also be used to measure the surface temperatures on either side the structural element and the respective air temperatures immediately inside and outside; based on these results it is then possible to calculate all the relevant thermal coefficients.

The temperatures and heat flow density data on which these calculations are based are acquired cyclically as average values.

Any influence that the structure’s own thermal capacity may have on these calculations (e.g. time shifts between temperature and heat flow, affecting calculation of the U value) will, given a sufficiently long measuring period, become negligible and the calculated average value will certainly be very close to the structure’s actual U value.

Operative range:

To ensure a stable and meaningful U value calculation it is possible to stipulate that measuring operations only be performed subject to certain specified conditions.

• The temperature difference between interior and exterior ambient air must be sufficiently large (typically 20 K, e.g. inside temperature 20°C and outside temperature 0°C).

• Any fluctuations in these temperatures (e.g. day / night) must throughout the measuring period be as small as possible.

• The measured values must be acquired and recorded on-site over a sufficiently long period (e.g. one whole day or even several days) and the parameters must be calculated on the basis of average values

Ordering information Order no.

ALMEMO

®

measuring system - with 2 temperature sensors and 1 heat flow plate - for determining the U value - with straightforward calculation in the ALMEMO

®

measuring instrument:

ALMEMO

®

data logger 2590-4AS, 4 inputs MA25904AS

Mains unit ZA1312NA7

ALMEMO

®

data cable, RS232 interface, electrically isolated

Outside air temperature Thermo-wire sensor, with glass-fiber insulation, 5 meters long

Inside air temperature Thermo-wire sensor, with glass-fiber insulation, 1.5 meters long

Programming for inside sensor Differential channel and average value

Heat flow plate, including installation materials see page 13.04 / 13.05 e.g. type 118, approx. 120 x 120 mm, cable 2 meters

Programming for Heat flow plate, Average value and U-value channel

ZA1909DK5

FTA3900L05

FTA3900

OA9000PRUT

FQA018C

OA9000PRUQ

ALMEMO

®

measuring system - with 4 temperature sensors and 1 heat flow plate - for determining the U value - using WinControl software (possible both online and offline) :

ALMEMO

®

data logger 2690-8A, 5 inputs, including mains unit and data cable, RS232 interface

Outside air temperature Thermo-wire sensor, with glass-fiber insulation, 5 meters long

MA26908AKS

FTA3900L05

Outside surface temperature Thermo-wire sensor, with glass-fiber insulation, 5 meters long

Inside air temperature Thermo-wire sensor, with glass-fiber insulation, 1.5 meters long

FTA3900L05

FTA3900

FTA3900 Inside surface temperature Thermo-wire sensor, with glass-fiber insulation, 1.5 meters long

Heat flow plate, including installation materials see page 13.04 e.g. type 118, approx. 120 x 120 mm, cable 2 meters

WinControl software for 20 measuring points, 1 device

Additional module U-value wizard

Hardlock USB dongle

FQA018C

SW5600WC1

SW5600WCZM4

SW5600HL

Accessories

Carry case, large ZB2590TK2

13.03

Heat flow

Heat Flow Plates FQAx

• For determining the heat flow density up to max. 150°C.

• Application-oriented designs, consisting of a meander of opposing thermocouples that are embedded in a substrate.

• In case of thick substrates no lateral circulation of the heat flow because of sufficient meander shell zone.

• Software for k value measurement, see chapter Software

Technical Data:

Type Dimensions

(mm)

117

118

119

100 x 30 x 1.5

120 x 120 x 1.5

250 x 250 x 1.5

120

117SI

118SI

150-1

150-2

33 Ø x 1.5

100 x 30 x 3

120 x 120 x 3

180 x 100 x 0.6

500 x 500 x 0.6

Meander Size Substrate

(mm)

80 x 20

90 x 90

180 x 180 epoxy resin epoxy resin epoxy resin

20 Ø

80 x 20

90 x 90

170 x 90

490 x 490 epoxy resin silikone silikone

PTFE

PTFE

!

Each heat flow plate has been assigned a calibration value, which corresponds to the heat flow density in

W/m

2

when the plate provides an output of 1mV. The calibration value will be stored as factory-setting in the

ALMEMO

®

connector so that ALMEMO

®

devices will immediately indicate the current heat flow density in

W/m

2

.

Temperature

Stability

-40 ... 80°C

-40 ... 80°C

-40 ... 80°C

-40 ... 80°C

-40 ... 80°C

-40 ... 80°C

150°C

150°C

Calibr. Val. appr. Accuracy

(W/m

2

≈ mV) of Calibr. Value

< 50

< 15

< 8

5% at 23°C

5% at 23°C

5% at 23°C

< 150

< 50

< 15

< 80

< 10

6% at 23°C

5% at 23°C

5% at 23°C

5% at 25°C

5% at 25°C

Accessories

Adhesive tape for room temperature

Self-adhesive film 24 x 100cm for room temperature

117

118

119

120

117 SI

118 SI

150-1

150-2

Types incl. connecting cable, 2 m, with ALMEMO

®

connector and manufacturer’s test certificate

Model

Application

for even surfaces, e.g. casement sections for universal applications, e.g. solar-electric systems and insulating plates especially for constructional industry, brickwork insulating plates, old buildings small heat flow plate, e.g. for medicine, veterinary medicine, small components etc. flexible heat flow plate, suitable for even surfaces, e.g. casement sections flexible heat flow plate, suitable for even surfaces, e.g. solar-electric systems and insulating plates flexible heat flow plate, particularly suitable for high temperatures e.g. for brickwork, insulated boilers and pipes particularly suitable for high temperatures, especially for the construction industry, masoned walls and insulating plates

13.04

Order no.

ZQ9017KB

ZQ9017KF

Order no.

FQA017C

FQA018C

FQA019C

FQA020C

FQA017CSI

FQA018CSI

FQA0801H

FQA0802H

ALMEMO

®

D6

Heat flow

Digital heat flow plate FQADx, with integrated temperature sensor for automatically correcting the heat flow plate’s temperature coefficient, with ALMEMO

®

D6 plug

• new:

This automatically corrects the heat flow plate’s temperature coefficient using a miniature NTC sensor integrated in the heat flow plate for the purpose of measuring the plate’s mean temperature.

• It measures heat flows and temperatures using a A/D converter incorporated in the ALMEMO

®

D6 plug.

• Two measuring channels are programmed (at our factory).

• Plate’s mean temperature (°C, t)

Heat flow, temperature-compensated (W/m², fq) model 117, 118, 119

Technical Data

Heat flow sensor (see table on page 13.04)

Accuracy of calibration value at nominal

temperature 5 %

Nominal temperature 23 °C

Temperature coefficient -0.12 % / K (epoxide plate) or -0.17 % / K (silicone plates)

Temperature sensor

Sensor element

Accuracy

Miniature NTC type N

±0.5 K at 0 to +80 °C

A/D converter incorporated in ALMEMO

®

D6 plug

Input 1 NTC sensor

(clamp connector in plug)

Resolution

Linearization

Accuracy

0.01 K error-free computing method according to Galway Steinhart (no approximations)

±0.05 K

Nominal temperature 23 °C ±2 K

Temperature drift: 0.004 %/K (40 ppm)

Voltage mV Input 2

Measuring range

Precision class

Refresh rate

Supply voltage

(clamp connector in plug)

0 to 26 mV, 0 to 260 mV

AA see page 01.05

0.4 seconds for both channels

6 to 13 VDC

Current consumption 4 mA

Accessories

see page 13.03

General features and accessories, ALMEMO

®

D6 sensors see page 01.08

Order no.

Variants

including manufacturer’s test certificate

Order no.

Heat flow plate with integrated temperature sensor cable permanently fitted, PVC, length 2 meters with ALMEMO

®

D6 plug.

Type 117 Substrate Epoxy resin, Dimensions 100 x 30 x 1.5 mm

Type 118

Type 119

Substrate Epoxy resin, Dimensions 120 x 120 x 1.5 mm

Substrate Epoxy resin, Dimensions 250 x 250 x 1.5 mm

Type 117SI Substrate Silicone, Dimensions 100 x 30 x 3 mm

Type 118SI Substrate Silicone, Dimensions 120 x 120 x 3 mm

FQAD17T

FQAD18T

FQAD19T

FQAD17TSI

FQAD18TSI

13.05

Air humidity

Digital sensors for humidity, temperature, dew point FHAD46x for measuring the equilibrium moisture content in building materials

ALMEMO

®

D6

Measuring the equilibrium moisture content

A material’s equilibrium moisture content is that level of relative humidity prevailing in the ambient atmosphere at which the material neither gains nor loses moisture.

All construction materials may - to a greater or lesser degree - attract water vapor from or emit water vapor to the ambient air. They are hygroscopic; i.e. they attempt to establish an equilibrium in terms of moisture content with respect to the ambient air. The construction material and the ambient air, depending on their respective temperatures, establish an interactive balance between the adsorption of and the emission of water vapor from / to one another. Each material thus has, depending on temperature and on atmospheric humidity, a certain moisture content level (measured in water as a percentage of overall weight).

In the state of equilibrium the relationship between the water content and the equilibrium humidity of a material can be displayed graphically as a curve, the so called moisture sorption isotherm. The sorption isotherm for the material in question indicates per atmospheric humidity value the corresponding water content value at a given constant temperature. If the composition or quality of the material changes then its sorption behavior - and thus its sorption isotherm - also changes. Given the great complexity of sorption processes these isotherms cannot be determined by calculation; they have to be recorded experimentally.

Digital sensors for humidity, temperature, air pressure FHAD46-0, uncovered sensor element, with ALMEMO

®

D6 plug.

FHAD460

Uncovered sensor element: Smallest design, short response time

Description andtechnical data see page 08.06

Digital sensor for temperature, atmospheric humidity, and atmospheric pressure FHAD46-2

Version in plastic, with slotted sensor cap with ALMEMO

®

D6 plug

FHAD462

Sensor element enclosed in slotted sensor cover, compact design, short response time

FHAD462 Option with plug-in extension tube

Description andtechnical data see page 08.06

FHAD462L00

DAkkS or factory calibration KH9xxx, temperature, humidity, and KD92xx, atmospheric pressure, for digital sensor

(see chapter Calibration certificates).

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025.

13.06

Moisture in materials

Moisture Sensor FHA 696 MF

Technical Data

Measuring method:

Resolution: capacitive

0.1%

Measuring range (moisture): 0 to 50% moisture, referenced to mass

Measuring range (material):

mineral construction materials

woods

paper and cardboard

0 to 20%, moisture

0 to 50%, moisture

0 to 20% moisture

Housing:

Terminal block: plastic handle with integrated electronics

40mm Ø, 130mm long aluminium/plastic 20 x 25 x 70mm

• Moisture sensor for determination of the moisture content in mineral construction materials, wood and cardboard.

• Indirect measurement of the moisture through the determination of the dielectric constant.

• Capacity measurement through a high frequency electromagnetic field, which penetrates the material in a non-destructive way.

Measuring comb:

Weight:

Nominal temperature:

Operative range:

Storage temperature:

Signal output:

Power supply:

Current consumption stainless spring steel 0.5mm, 70 x 35mm

260g

15 to 25°C

0 to +60°C

–20 to +80°C

0 to 2V

+8 to +12V approx. 7 mA

Accessories

Test block for min. construct. materials

Test block for wood, paper, cardboard

Type

Moisture sensor

Order no.

ZB9696PE05

ZB9696PE30

Order no.

FHA696MF

Wood moisture probe FHA 636 MF Hand-held probe for mobile test measurements

• Moisture sensor for determination of the moisture content in wood.

• Indirect moisture measurement according to the principle of conductivity.

• Determination of the moisture content in the material through the dependence of the electrical resistance on the moisture.

Technical Data

Measuring method:

Measuring range:

Housing:

Measuring tips:

Weight: principle of conductivity

7 to 30 % moisture, referenced to mass plastic handle

40mm Ø, 130mm long stainless steel, uninsulated

3mm Ø, 50mm long

260g

Reproducibility: ± 1%

Nominal temperature: 23°C ±2°C

Operating temperature: 0 to +60°C

Storage temperature:

Signal output:

Power supply:

Current consumption

–20 to +80°C

0 to 2V

7.5 to +12V max. 10 mA

Accessories

PTFE-insulated measuring tip - helps avoid measuring errors in the event of surface moisture, 1 piece

(2 pieces are needed per probe)

Type

Wood moisture probe

Order no.

ZB9636MFST

Order no.

FHA636MF

13.07

Moisture in materials

Moisture content sensor - for wood, for stationary measuring operations FHA696MFS1

Capacitive sensor for applying onto the wood‘s surface

• Moisture content sensor for comparative measurement of moisture in wood materials

• The capacitive sensor with the measuring electronics is completely integrated in the damp-proof sensor housing. Plug-in

ALMEMO

®

connecting cable

• This device is designed for stationary installation and longterm monitoring e.g. of wooden parts of buildings, roof structures (with laminated beams).

• It is also suitable for data logger operation in energy-saving sleep mode (intermittent mode).

• The sensor housing is quick and easy to install on the wooden surface in question.

• The material‘s moisture content is measured indirectly by determining its dielectric constant, which is moisture-dependent

(but not temperature-dependent).

• Its capacity is measured via a high-frequency electrical field which penetrates the wood without destroying it.

• The ALMEMO

®

device acquires the material‘s moisture content based on the linearization curve stored in the ALMEMO® plug.

• This measuring operation can be performed using any current

ALMEMO

®

device (version 6 and above).

Technical Data

Measuring method capacitive

Measuring range 0 to 50 % moisture percentage in wood with respect to total mass

(at 23 °C)

Resolution

Reproducibility

0.1 % moisture content

±1 % moisture content

Nominal temperature 23 °C ±2 K

Suitable conditions 0 to +80 °C

Air humidity 0 to 90 % RH

(no dew formation, no ice)

Storage temperature -20 to +80 °C

Housing

Signal connection

Plastic 51 x 53 x 36 mm (LxWxH)

Built-in plug

Protection Housing and plug connection IP64

ALMEMO

®

connecting cable Coupling, PVC cable, 5 meters

ALMEMO

®

plug Linearization for wood, stored in the

ALMEMO

®

plug (for ALMEMO® devices version 6 and above)

Supply voltage via ALMEMO

®

plug (5 V)

Current consumption approx. 7 mA

Accessories

Test block for wood, for testing purposes

Order no.

ZB9696PE08

Variants Order no.

Moisture content sensor for wood, sensor integrated in the sensor housing, with built-in plug, connecting cable 5 meters,

ALMEMO

®

plug for current ALMEMO

®

devices, version 6 and above

FHA696MFS1

13.08

Moisture in materials

Moisture content sensor - for wood, for stationary measuring operations FHA636MFS1

Conductivity measurement with measuring tips that can be screwed into the wood

Sensor with integrated temperature sensor for automatic temperature compensation

• Moisture content sensor for comparative measurement of moisture in wood materials

• Two hanger bolts are screwed into the wood surface and connected via measuring lines to the measuring electronics in the damp-proof sensor housing.

• The sensor housing with the integrated temperature sensor is also fixed in position on the wood surface.

• Plug-in ALMEMO

®

connecting cable

• This device is designed for stationary installation and longterm monitoring e.g. of wooden parts of buildings, roof structures (with laminated beams).

• Data logger operation in sleep mode (intermittent mode) is required in order to protect the wood from salinization or drying out.

• The material‘s moisture content is measured indirectly by determining its electrical conductivity, which is moisturedependent.

• It is also temperature-dependent. However, the displayed moisture value is automatically temperature-compensated by means of an integrated temperature sensor.

• The ALMEMO

®

device acquires the material‘s moisture content based on the linearization curve stored in the ALMEMO

® plug.

• This measuring operation can be performed using any current

ALMEMO

®

device (version 6 and above).

Technical Data

Measuring method Electrical conductivity

Measuring range 5 to 50 % moisture percentage in wood with respect to total mass

(at 23 °C)

Resolution

Reproducibility

0.2 % moisture content

±1 % moisture content

Nominal temperature 23 °C ±2 K

Temperature sensor NTC, integrated in sensor housing

Temperature compensation in range 0 to +80 °C

Suitable conditions 0 to +80 °C

Air humidity 0 to 90 % RH

(no dew formation, no ice)

Storage temperature -20 to +80 °C

Housing Plastic 51 x 53 x 36 mm (LxWxH)

Measuring connection 2 built-in sockets, 4 mm, with transverse hole

Measuring lines

Measuring tips

Clearance

Signal connection

2 lines, PTFE-insulated, length = 0.5 meters with circular cable lugs 4 mm

2 stainless-steel M4 hanger bolts

Total length = 60 mm including 4 stainless-steel nuts,

4 stainless-steel lock washers

2.5 cm at right angles to the grain

Built-in plug

Protection Housing, including connectors IP63

ALMEMO

®

connecting cable Coupling, PVC cable, 5 meters

ALMEMO

®

plug

Supply voltage

Linearization for wood, stored in the

ALMEMO

®

plug (for ALMEMO

® devices version 6 and above) via ALMEMO

Current consumption approx. 5 mA

®

plug (5 V)

Variants Order no.

Moisture content sensor for wood, with measuring tips, measuring line, sensor housing, connecting cable, 5 meters

ALMEMO

®

plug, for current ALMEMO

®

devices, version 6 and above

FHA636MFS1

13.09

Moisture in materials

Sensor for measuring the moisture in materials FHA 696 GF1

For determining the moisture content in granulated materials such as wood chips, wood pellets, and sawdust

• The sensor operates on the principle of an open plate capacitor. The moisture contained in a material can be measured in terms of that material’s dielectric constants.

• Moisture content can be determined in a matter of seconds - in wood chips or wood pellets, and sawdust, in grain and cereals, and other granulated materials.

• The characteristics of the materials to be measured can be specified on a highly customized basis; a wide variety of granulates, e.g. various cereal types, can thus be measured

Technical Data

Measuring principle capacitive

Measuring range

Resolution

0 to 99.9 % water content as a weight percentage H

0.1%

2

O

Measuring radius / penetration depth approx. 10 cm around the sensor

Temp. range of material +5 to +40 °C

Operating temp. range +5 to +40 °C

Storage temp. range -20 to +70 °C

Signal output ALMEMO

®

(voltage)

Power supply 5 V from ALMEMO instrument

Current consumption approx. 5 mA

®

measuring

Dimensions

Sensor head

Extensions

End piece

Cable terminal

Cable

Ø = 22 mm, length = 200 mm

Rounded tip

3 pieces, screw-on

Ø = 18 mm, length = 300 mm

Plastic

Ø = 22 mm, length = 30 mm

Mountable male connector on sensor head

PVC, length = 2 meters with ALMEMO

®

connector

The cable is led through the extension tubes and end piecet.

Option

Determining characteristics for special customer-specific materials

Order no.

1. We need a sample of approx. 10 liters of your granulate (e.g. wood, cereal, plastic). This sample should be sealed in an air-tight package, e.g. shrink-wrapped in plastic film.

2. We use various dried samples to determine the characteristics of your particular material.

3. We then program these characteristics in the ALMEMO

®

connector for the moisture content probe..

Pro rata processing costs per material sample, net (service) Order no. OA9696GFK

Advisory note:

If the material cannot absorb water (not hygroscopic), it will not be possible to measure its moisture content.

In this case the processing fee we charge will be reduced.

Variants

Sensor for measuring moisture in granulated wood chips and pellets comprising :

Order no.

Sensor head, 3 screw-on extensions, end piece, connecting cable 2 meters, with ALMEMO

®

connector programmed for wood chips (also programmable for wood pellets; if required, please indicate) including carry case

FHA696GF1

Test block for FHA696GF for wood chips and wood pellets

ZB9696PE22

13.10

Dew Point Detector, Water Detection Probe

Dew Point Detector FHA 9461

• Dew detector for determination of dew conditions.

• Consisting of one temperature sensor and an integrated sensor chip with CCC dew point sensor.

• Particularly suitable in building physics for control measurements and stationary installation.

• The dew point detector does not provide a measuring signal but a step function: dewed (100%) / no dew (0%).

Technical Data

Principle of measurement: CCC sensor

Operative range: 0°C to +70°C

(no ice formation, no saliferous atmosphere)

Settling time:

Temperature sensor: final value after 2 to 60 seconds

NTC type N (10k at 25°C), accuracy: ±0.1°C

(within operative range)

Signal output: scaled voltage approx. 0 to 1V

Current consumption: approx. 3mA

Heat flow plate: aluminium, 40 x 40mm

Storage temperature: –10°C bis +70°C

Types Order no.

Sensor and electronics integrated in ALMEMO

®

connector, mounted on heat conducting plate made of aluminium

FHA9461

Water Detection Probe FHA 936 WD

Technical Data

Measuring method:

Meas. values:

Housing:

Electrodes:

Electrode types: detection of water

<10% no water

>10% water plastic handle

40mm Ø, 130mm long stainless steel uninsulated with rounded tip:

200mm long, 3mm Ø uninsulated with sharp-edged tip:

50mm long, 3mm Ø spring steel strap:

200mm long, 6mm wide, 0.5mm high

• Water detection probe for instant detection of uncombined water.

• Particularly suitable for construction applications, especially in locations that are difficult to check visually, e.g. at sealing joints, under cement floors etc.

• Indirect moisture measurement according to the principle of conductivity.

• Probe with two collets for easy electrode replacements.

• Electrodes in three different designs for matching any required application.

Weight: 260g

Nominal temperature: 23°C ±2°C

Operating temperature: 0 to +60°C

Storage temperature: –20 to +80°C

Signal output: ALMEMO

®

(approx. 0 to 2V)

Power supply: 7.5 to 15V

Current consumption max. 10 mA

Type

Water detection probe

Order no.

FHA936WD

13.11

Moisture in the soil

Tensiometer FDA 602 TM2

• Measurement of soil moisture through the identifi cation of suction pressure. The suction pressure is the force with which water is being held in the soil or is available for absorption.

This is the force that must be produced by the plant roots in order for water to be absorbed.

• The porous, clay tip of the tensiometer transfers water from within to the drier outer surroundings by means of capillarity, thereby, creating a sub-pressure within the sealed tensiometer tube. This sub-pressure is a measure of the moisture level and can be determined as a value or used directly to activate an electrical switch. The customary unit of measurement is hPa.

• However, a tensiometer also functions in dry air as long as evaporation can take place over the porous, clay chamber.

Therefore, moisture levels can be measured even in coarsegrained or very loose substrate.

• Suction pressure measurements are largely independent of the salt concentration of the substrate or soil.

Insertion Tensiometer L2

Types

Insertion Tensiometer LV

Order no.

ZB9602TML2

Ceramic cell Cylindrical, with tip, Ø 20 x 65 mm

Overall length approx. 340 mm

Insertion depth typical 250 mm

ZB9602TMLV

Typical Suction Pressure at Peat Substrates

30 - 40 hPa very moist

50 – 120 hPa moist

150 – 200 hPa dried

>200 hPa dry

Ceramic cell Cylindrical, with tip, Ø 15 x 40 mm

Overall length approx. 210 mm

Insertion depth typical 120 mm

Typical Suction Pressure at Open fields

(intermediate grade soil)

< 50 hPa saturated

100 – 150 hPa wet to moist

>200 hPa start drying

200 – 500 hPa Irrigation

Insertion Tensiometer LKV2 ZB9602TMKV2

Ceramic cell Cylindrical, with tip, Ø 15 x 40 mm

Overall length approx. 160 mm

Insertion depth typical 70 mm

Moisture tension meter, electronics

Surface Tensiometer FO ZB9602TMFO

Technical Data

Measurement: Measurement of soil moisture through the identification of suction pressure.

Measure range: 0 to -1000 hPa relative (negative pressure)

Output 0,5 to 4,5 V

Power supply

Cable

5 V via ALMEMO

®

connector

Sensor with cable, length = 5m, with ALMEMO

®

connector

Sensor completely porous for measuring in thin layers of substrate.

Dimensions: 65 mm, Ø 70 mm

Sink deep: approx. 30 - 60 mm

Surface Tensiometer FV ZB9602TMFV

Type

Tensiometer electronics for screwing onto the tensiometer with cable and ALMEMO

®

connector

Order no.

FDA602TM2

13.12

Standard model for use on capillary matting, for moist to moderately moist cultivation or for general measurement on moist surfaces.

Dimensions: 65 mm, Ø 70 mm

Optical radiation

Content

What is ´Optical Radiation´?

Definition of Photometric and Radiometric Meas. Variables

Comparison of Photometric and

Radiometric Measurable Variables

Spectral Valuation Function

Determination of Photometric Characteristic Factors

Radiation probe FLA 623 x

Probe for measuring illuminance FLA 623 VL

Probe for UVA radiation FLA 623 UVA

Probe for UVB radiation FLA 623 UVB

Probe for UVC radiation FLA 623 UVC

Probe for global radiation FLA 623 GS

Probe for infra-red radiation FLA 623 IR

Probe for quantum radiation FLA 623 PS

Lux probe head for illuminance FLA 613 VLK

UVA probe head spherical characteristic FLA 613U VAK

Lux probe head for illuminance FLA 603 VLx

Digital sensor for color temperature and illuminance

Luminance Probe Head FLA 603 LDM2

Light Flux Probe Head FLA 603 LSM4

Probe Heads for Outdoor Operation see Chapter Meteorology

14.02

14.02

14.03

14.03

14.04

14.05

14.05

14.06

14.06

14.06

14.07

14.07

14.07

14.08

14.08

14.09

14.10

14.12

14.12

14.01

Optical radiation

What is ´Optical Radiation´?

Optical radiation covers the wave length range from 100nm to 1mm of the electromagnetic radiation spectrum.

It must be considered that, with regard to the range limits, they do not preset a sharp separation, which is compulsory for all applications.

The detection of optical radiation can, for example, be measured by means of radiometric, photometric, photobiological or plant-physiological measurable variables.

Defi nition of Photometric and Radiometric Measurable Variables

Photometry

Limited to the range of the optical spectrum (light) that is visible to the human eye.

Photometric measurable variables include:

Light fl ux, illuminance, luminance and luminous intensity. The main characteristics of photometry is the evaluation of the brightness perception by the spectral luminosity function of the eye for photopic vision or, in rare cases, for scotopic vision

(DIN 5031). Radiation detectors for photometric measuring tasks must, therefore, provide one of these spectral response characteristics.

Light Flux

The luminous power of a light source

(lamp, LED etc.). As lamps do not generally emit a completely parallel luminous beam, the light fl ux measurement is performed by using measurement geometries, which detect the light fl ux independent from its geometric distribution. In most cases Ulbricht globe photometers or goniometers will be used.

Luminous Intensity

The part of a light fl ux, which radiates in one specifi c direction. The luminous intensity is an important variable for calculating the effi ciency and quality of lighting equipment. The measurement is performed by detectors with a defi ned fi eld of view and placed at distances that allow to consider the light source as a point light source.

Luminance

The brightness sensation provided by an illuminated or luminous surface to the eye. In many cases the luminance data will provide signifi cantly better information regarding the quality of a light than the illuminance. For measuring the luminance, measuring heads with a defi ned measuring fi eld angle are used.

Illuminance

The light fl ux of one or several light sources striking a certain surface horizontally or vertically. In case of a non-parallel incidence (which is the typical case in practical photometry) a cosine diffusor must be used as measurement geometries.

Radiometry

Metrological evaluation of optical radiation using the radiometric variables „Radiation Capacity“, „Radiant Intensity“,

„Radiancy“ and „Intensity of Irradiation“.

The main characteristic of radiometry is the wavelength-independent examination of the intensity of radiation. This is the signifi cant difference between radiometry and actively weighted measurable variab-

14.02

Optical radiation Optical radiation

les, such as variables used in photometry, photobiology, plant physiology etc.

Radiation Capacity

The overall power provided by radiation.

Radiant Intensity

The quotient from the radiation capacity emitted by the light source into a certain direction and the solid angle being covered. The radiant intensity is used for the measurement of the geometric distribution of the radiation capacity.

Radiancy

The quotient from the radiation capacity passing through (striking) a plane in a certain direction and the product of the passed solid angle and the projection of the plane to a plane surface, which is perpendicular to the examined direction. The radiancy is used for the evaluation of aperture radiators. Steradian or telescopic adapters can be used as measurement geometries.

Intensity of Irradiation

The quotient of the radiation capacity striking a plane and the illuminated plane.

For measuring the intensity of irradiation the spacial examination of the incident radiation is very important; therefore, a cosine-corrected field view function has been preset.

Comparison of Photometric and Radiometric Variables

Every photometric variable corresponds to a radiometric variable and involves the same interrelationships between them.

The variables can be distinguished by their index v (visual) and index e (energetic).

Spectral Valuation Function

The relative spectral sensitivity of the human eye is specified with different functions for the light-adapted eye (photopic vision) or for the dark-adapted eye (scotopic vision). Due to the individual differences this data can only be considered for average values but is sufficient for most technical purposes. The detailed data of the spectral sensitivity curve are given in table format in the DIN 5031 standard.

The two different spectral action functions result from the different „sensor types“ of the eye.

The relative luminous efficiency for photopic vision (rods, > 10cd/m

2

) is described with the function V(l), which is the function used in most cases. The spectral luminous efficiency for the scotopic vision (cones, < 0.001cd/m

2

) is described with the function V’(l) and can, with regard to the practical use, only be rarely found.

14.03

Optical radiation

Determination of Photometric Characteristic Factors

The metrological evaluation of the properties of materials regarding their reflection, transmission and absorption, as well as the stray light of objectives, is based on internationally accepted recommendations. These mainly include the CIE 130-

1998 „Practical methods for the measurements of reflectance and transmittance“,

DIN 5036 Part 3 „Radiometric and photometric characteristics of materials“, DIN

67507 „Light transmission factor of glazing“, DIN 58186 „Stray light determination of optically image-forming systems“.

Why Measure Optical Radiation?

A large part of the human sense impression is of an optical nature. Light is the only visible part of the electromagnetic spectrum. The human eye perceives different wave lengths of the light as colours.

The spectral response of the eye, with regard to different colours, depends on the wave length. Furthermore, the human system is also influenced by ultraviolet radiation in a short-wave range and the infrared radiation in a long-wave range of the electromagnetic spectrum.

Illumination:

People are used to daylight illumination.

This can be approximately 5000 lux on a dull winter day, while on a sunny summer day approximately 100000 lux are reached. In contrast, only between 100 and

1000 lux are reached with artificial illumination. However, sufficient light is an essential factor for the well-being of people. Symptoms of tiredness, caused by insufficient light, do not generally occur at the eye but affect the whole body.

The standard DIN 5035/2, therefore, contains illumination standard values for health protection at work places.

These are legally bound in the guideline

ASR 7/3 and it is imperative that this is observed.

The following nominal illuminations are valid for inside:

Offices:

Factories:

Hotels:

Shops:

Hospitals:

Schools: office rooms work places for writing and drawing

300 lux

750 lux visual works within the production process 1000 lux recreation rooms, reception, counter (cash) 200 lux front side of show windows patients´ rooms, emergencies lecture rooms, gymnasiums

1500–2500 lux

100–150 lux

500 lux

300 Lux

Global Radiation:

The global radiation is a measuring variable that is especially important for environmental research. It represents the entire diffuse and direct sun radiation that strikes the surface of the earth. The spectral range covers wavelengths from the short-wave range, at 300nm (UV-B), to the long-wave range, at 5000nm (IR).

UVA Radiation:

The long-wave UV radiation (more than

313nm) reaches the surface of the earth almost unfiltered and tans the human skin and strengthens the immune system. In solariums the biological effect of the UVA spectrum is used, combined with other spectral ranges, to trigger the direct pigmentation (melanin colouring). Damages to the connective tissue and premature skin ageing are promoted by too much radiation.

UVB Radiation:

The short-wave UV range (less than

313nm) can cause irreversible damages.

All spectral characteristic functions that can have unfavourable effects on the human skin are summarised in the CIE recommendation. This recommendation is described in DIN 5050 and regarded as a guideline. A popular measure for the

‚sunburn sensitivity‘ is, for example, the

UV index ‚UVI‘ provided by the German

Weather Service. The measuring results provide, directly or in comparison with other spectral ranges, information that is of medical or biological relevance.

14.04

Optical radiation

Radiation probe FLA 623 x

Optical radiation

• Probes for various spectral ranges:

• Illuminance (V lambda), UVA, UVB, UVC, global radiation,

IR, quantum (photosynthesis)

• Sturdy aluminum housing

• ALMEMO

®

connecting cable, plug-in

• For indoor applications

Common technical data

Diffuser

Cosine correction

PTFE

Error f2 <3 %

Linearity

Absolute error

<1 %

<10 % (<5 % for FLA623VL)

V lambda adapter <3 % (for FLA623VL only)

Nominal temperature 22 °C ±2 K

Operating temperature -20 to +60 °C

Signal output 0 to +2 V

Duty cycle

Power supply

<1 second via ALMEMO

®

connector

(5 to 15 VDC)

Electrical connection Mountable male connector, lateral

Connecting cable PVC cable, plug-in, with

ALMEMO

®

connector

Housing

Fixture

Dimensions

Weight

Aluminum, black anodized

2 screws M2 in base plate

Diameter 33 mm, height approx. 29 mm approx. 50 g (without cable)

Probe for measuring illuminance FLA 623 VL

• This measures the V lambda radiation (visible light, equivalent to sensitivity of the human eye).

• For evaluating lighting conditions, e.g. in the workplace

• The sensor complies with device class B as per DIN 5032.

Technical data:

Measuring range V lambda 0 to approx. 170 klx

Measuring channels

Spectral sensitivity

1st channel up to approx. 20,000 lx

2nd channel up to approx. 170.00 klx

380 to 720 nm, max. at 555 nm

Common technical data and image see page 14.05

Variants

(including factory test certificate)

Illuminance probe with ALMEMO

®

connecting cable, length = 2 meters

Options

ALMEMO

®

connecting cable, length = 5 meters

ALMEMO

®

connecting cable, length = 10 meters

Order no.

FLA623VL

OA9623L05

OA9623L10

Factory calibration KL90xx radiation for sensor (see chapter Calibration certificates)

14.05

Optical radiation

Probe for UVA radiation FLA 623 UVA

• This measures long-wave UV radiation (bronzing effect on human skin).

• Its spectral sensitivity is weighted towards global solar radiation.

Technical data:

Measuring range 0 to approx. 50 W/m

2

Spectral sensitivity 310 to 400 nm, maximum at 335 nm

Common technical data and image see page 14.05

Variants

(including factory test certificate)

UVA probe with ALMEMO

®

connecting cable, length = 2 meters

Options:

ALMEMO

®

connecting cable, length = 5 meters

ALMEMO

®

connecting cable, length = 10 meters

Probe for UVB radiation FLA 623 UVB

Order no.

FLA623UVA

OA9623L05

OA9623L10

• This measures short-wave UVB radiation.

• Its spectral sensitivity is weighted towards global solar radiation likely to cause erythema (sunburn) as per CIE recommendation (Commission Internationale de l‘Eclairage). The UV index can be calculated.

Technical data:

Measuring range

Spectral sensitivity

0 to approx. 5 W/m

2

265 to 315 nm, maximum at 297 nm

Common technical data and image see page 14.05

Variants

(including factory test certificate)

UVB probe with ALMEMO

®

connecting cable, length = 2 meters

Options

ALMEMO

®

connecting cable, length = 5 meters

ALMEMO

®

connecting cable, length = 10 meters

Probe for UVC radiation FLA 623 UVC

Order no.

FLA623UVB

OA9623L05

OA9623L10

• This measures UVC radiation, e.g. Hg line at 256 nm.

• This probe can be used inter alia in water disinfection units.

Technical data:

Measuring range

Spectral sensitivity

0 to approx. 1990 mW/m

2

220 to 280 nm, maximum at 265 nm

Common technical data and image see page 14.05

Variants

(including factory test certificate)

UVC probe with ALMEMO

®

connecting cable, length = 2 meters

Options:

ALMEMO

®

connecting cable, length = 5 meters

ALMEMO

®

connecting cable, length = 10 meters

Factory calibration KL90xx radiation for sensor (see chapter Calibration certificates)

14.06

Order no.

FLA623UVC

OA9623L05

OA9623L10

Optical radiation

Probe for global radiation FLA 623 GS

• This measures the solar spectrum in the visible range and in the short-wave IR range.

• Global radiation comprises both direct and diffused solar radiation.

Technical data:

Measuring range

Spectral sensitivity

0 to approx. 1300 W/m

2

400 to 1100 nm, maximum at 780 nm

Common technical data and image see page 14.05

Variants

(including factory test certificate)

Global radiation probe with ALMEMO

®

connecting cable, length = 2 meters

Options:

ALMEMO

®

connecting cable, length = 5 meters

ALMEMO

®

connecting cable, length = 10 meters

Probe for infra-red radiation FLA 623 IR

Order no.

FLA623GS

OA9623L05

OA9623L10

• This measures the solar spectrum in the short-wave IR range

(excluding the visible range).

• Global radiation comprises both direct and diffused solar radiation.

Technical data:

Measuring range

Spectral sensitivity

0 to approx. 400 W/m

2

800 to 1100 nm, maximum at 950 nm

Common technical data and image see page 14.05

Variants

(including factory test certificate)

IR probe with ALMEMO

®

connecting cable, length = 2 meters

Options:

ALMEMO

®

connecting cable, length = 5 meters

ALMEMO

®

connecting cable, length = 10 meters

Order no.

FLA623IR

OA9623L05

OA9623L10

Probe for quantum radiation FLA 623 PS

• This measures the visible light absorbed by the chlorophyll in plants during photosynthesis.

• It determines the level of quantum radiation in the spectral range specified.

• It is used to assess the conditions in which plants develop in open field and greenhouse cultivation.

Technical data:

Measuring range

Spectral sensitivity

0 to approx. 3000 mmol/m²s

380 to 720 nm, maximum at 420 and 700 nm

Common technical data and image see page 14.05

Variants

(including factory test certificate)

Quantum probe with ALMEMO

®

connecting cable, length = 2 meters

Options:

ALMEMO

®

connecting cable, length = 5 meters

ALMEMO

®

connecting cable, length = 10 meters

Order no.

FLA623PS

OA9623L05

OA9623L10

Factory calibration KL90xx radiation for sensor (see chapter Calibration certificates)

14.07

Optical radiation

Illuminance measuring head FLA 613 VLK

• Measuring independent of direction - thanks to the probe head’s spherical characteristics

• Weather-proof aluminum housing, with plastic globe

• Suitable for universal use, inter alia for measuring in photostability tests according to various international standards and

ICH guidelines (International Conference on Harmonization)

• Spectral range of the probe head corresponds to the sensitivity of the human eye (V-lambda radiation).

Technical data:

Measuring range

Spectral sensitivity

Maximum spectral sensitivity

Signal output

0 to 50 klux

360 to 760 nm

555 nm

0 to 2 V

Duty cycle

Power supply

Fastening

Cable passage

<1 second via ALMEMO

®

connector

+5 to +15 V

2 screws, M4, in base plate at side

Housing

Diffuser anodized aluminum

Plastic

Ball Plastic

Directional characteristic see diagram

Linearity

Absolute error

<1%

<10%

Nominal temperature 22 ± 2 °C

Operating temperature –20 to +60 °C

Dimensions

Weight

Ball diameter : 40 mm

Overall height : 76 mm approx. 100 grams

Type

(including test protocol)

Lux probe head for measuring luminous intensity, with spherical characteristic, including 1.5-meter cable and ALMEMO

®

connector

UVA probe head FLA 613 UVAK

Order no.

FLA613VLK

• Measuring independent of direction - thanks to the probe head’s spherical characteristics

• Weather-proof aluminum housing, with plastic globe

• Suitable for universal use, inter alia for measuring in photostability tests according to various international standards and

ICH guidelines (International Conference on Harmonization)

• Measuring head for measuring the UVA

Technical data:

Measuring range

Spectral sensitivity

0 to approx. 50 W/m

2

310 to 400 nm

Maximum spectral sensitivity 355 nm

Signal output

Duty cycle

Power supply

Fastening

0 to 2 V

<1 second via ALMEMO

®

connector

+5 to +15 V

2 screws M4, in base plate

Cable passage

Housing

Diffuser

Ball at side anodized aluminum

PMMA

(polymethyl methacrylate, acrylic)

PMMA (transparent to UV)

Directional characteristic see diagram

Linearity < 1%

Absolute error

Nominal temperature

< 10%

22 ± 2 °C

Operating temperature –20 to +60 °C

Dimensions

Weight

Ball diameter : 40 mm

Overall height: 76 mm approx. 100 grams

Type

(including test protocol)

UVA probe head, with spherical characteristic, including 1.5-meter cable and ALMEMO

®

connector

Factory calibration KL90xx radiation for sensor (see chapter Calibration certificates)

14.08

Order no.

FLA613UVAK

Optical radiation

Illuminance measuring head FLA 603 VLx

• High quality probe head for illuminance of light in lighting engineering or in sunlight and any place where DIN standards recommend the use of a class B luxmeter.

• Spectral adaptation approximated to the photometric valuation function V(λ) for photopic vision, class B, better than 5%.

• Different measuring channels with different sensitivity.

Technical data:

Measuring range:

Smallest resolution:

Sensitivity:

Spectral adaptation:

Max. cos deviation:

Cos diffusor:

Nominal temperature:

FLA603VL2: 0.05 lx to appr. 9600 lx

FLA603VL4: 1 lx to appr. 190 klx

FLA603VL2: 0.01 lx

FLA603VL4: 1 lx approx. 20pA/lx approxim. to photometric valuat. function V(l) for photopic vision, class B, better than 5% class B, < 3% diameter 7mm

24°C ±2K

Operat./storage temperature: 0 to 60°C/–10 to +80°C

Humidity range: 10 to 90% (non-condensing)

Dimensions: Ø 37mm, height 20 mm

Variants

Illuminance measuring head, DIN quality class B with ALMEMO

®

connecting cable 1.5m long, incl. factory calibration certificate with calibration in lx for indoor lighting (3 measuring channels) for ambient light (2 measuring channels)

Order no.

FLA603VL2

FLA603VL4

Factory calibration KL90xx radiation for sensor (see chapter Calibration certificates)

14.09

Optical radiation

ALMEMO

®

D6

Digital sensor for color temperature and illuminance FLAD23CCT with ALMEMO

®

D6 plug

• Color temperature and illuminance are determined as a means to plot and evaluate lighting systems.

• Compact sensor, particularly suitable for mobile applications

• Continuous measuring and updating of measured values

• Digital color temperature sensor with „TrueColorSensorchip“ and integrated signal processor

The TrueColorSensorchip (3 sensors on 1 chip) detects - separately - each of the three colors - red, green, blue (RGB). The respective sensitivities of these 3 color sensors are adapted to the standard spectral curves as per CIE and DIN. (see Figure) On the basis of these RGB values the computer calculates the color point within the RGB range in terms of coordinates

X and Y and determines the correlated color temperature

(CCT) in Kelvin.

• The display shows simultaneously both this color data and the illuminance in lux (lx) or kilolux (klx).

• Freely selectable measurable variables

Two measuring channels are programmed (at our factory):

Color temperature (CCT, K), Illuminance (Ev, lx)

Other measurable variable can also be selected:

Illuminance (Ev, klx), X-value, Y-value

This device can be configured on a PC using USB adapter cable ZA1919AKUV. (see “General accessories for AL-

MEMO

®

D6 sensors” page 04.05).

Technical data:

Spectral sensitivity 380 to 720 nm

Sensor system TrueColor, 3 sensors on 1 chip

Measuring ranges

Correlated color temperature (CCT) 54 to 30,000 K

Accuracy

(at 120 lx to 170 klx)

< 10% in range 1600 to 17000 K

Coordinates resolution (dx, dy)

< 0.005

Illuminance (V-lambda) 0 to 65,000 lx (factory setting)

Accuracy or 0.00 to 170.00 klx

< 10% in range 120 lx to 170 klx

Cosine correction 8 mm diffuser plate

Cosine error

Measuring duration

< 3%

< 3 seconds

Nominal conditions 23 °C ± 3 K, 0 to 90 % RH

(non-condensing)

Operating temperature -10 to +40 °C

Dimensions Diameter 25 mm, length 134 mm

ALMEMO

®

connecting cable Fixed cable, 1.5 meters, with ALMEMO

®

D6 plug

ALMEMO

®

D6 plug

Refresh rate

Setting time

1.5 seconds for all channels

3 seconds

(In order to run the data logger in sleep mode a wakeup delay of

3 seconds must be programmed.)

Supply voltage

Current consumption

6 to 13 VDC approx.. 4 mA

Variants

Digital sensor for color temperature and illuminance, fitted cable, 1.5 meters with ALMEMO

®

D6 plug

14.10

Color temperature sensor with ALMEMO

®

2590-2 (example)

Order no.

FLAD23CCT

ALMEMO

®

D6

Optical radiation

Accessories

Ulbricht integrating sphere

• Ulbricht integrating sphere, for measuring total radiation from any light source

• Especially suitable for measuring operations on site for light sources that have already been installed. This minimizes interference from extraneous light in the environment.

• Dimensions

Measuring aperture 13.5 mm

Sphere diameter

Housing diameter

40 mm

44.5 mm, length 44 mm

Accessories

An Ulbricht integrating sphere can be attached to color temperature sensor FLAD23CCT

Order no.

ZB9623KU

14.11

Optical radiation

Luminance Probe Head FLA 603 LDM2

• Luminance measuring head, equipped with achromatically corrected, low stray light optics and high quality V(l) detector according to DIN class B.

• The external sighting device allows, at a working distance of 1m, to exactly locate the measuring point, therefore, it is particularly suitable for evaluating the luminance for service and constancy tests.

• Three measuring channels with different sensitivity.

• Typical applications:

Luminescent surfaces such as colour monitors, alphanumerical displays, sign plates and light panels, and reflecting surfaces, such as walls and equipment at work places, projecting screens, traffic and sign plates, guided paths and roadway lines.

Technical data:

Measuring range:

Smallest resolution:

Field of view:

Sensitivity:

Spectral adaptation:

Field of view diameter :

0.04 cd/m

2

to appr. 6400 cd/m

2

10 mcd/m

2

1° approx. 30 pA/(cd/m2) approxim. to photometric valuat. function V(l) for photopic vision, class B, better than 6% approx. 30 mm at a distance of 0.5 m approx. 40 mm at a distance of 1 m approx. 120 mm at a distance of 5 m

Nominal temperature: 24°C ±2K

Operat./storage temperature: 0 to 60°C/–10 to +80°C

Humidity range:

Measuring surface:

10 to 90% (non-condensing)

21mm x 21mm at 1m operating distance

Meets standards:

Dimensions:

IEC 61223-2-5, DIN 5032-T.7

diameter 30mm, length 150 mm

Variants

Luminance probe head with 1° field of view and external sighting device, DIN quality class B, with ALMEMO

®

connecting cable 1.5m long, incl. factory calibration certificate calibration in cd/m

2

Order no.

FLA603LDM2

Light Flux Probe Head FLA 603 LSM4

• High quality measuring head, DIN class B for light flux measurement with Ulbricht globe photometer.

• Perfect coating of the globe with BaSO4 for diffuse reflectivity and spectrally neutral reflection quality.

• Suitable for cold light sources, and lamps with high colour temperature and almost monochromatic radiation (as in

LEDs).

• Examples for applications:

Endoscopes, fiber optic bunches, light emitting diodes.

Technical data:

Measuring range:

Smallest resolution:

Sensitivity:

Acceptance angle:

Accuracy:

Nominal temperature:

0.0002 lm to appr. 38 lm

0.001 lm

20nA/lm up to 90 °

DIN quality class B

24°C ±2K

Humidity range: 10 to 90 % non-condensing

Operating temperature: max. 100°C inside globe

Inner diameter of globe: 50mm

Test opening: 12,7 mm

Type

Light flux probe head with ALMEMO

®

connecting cable 2m long and factory calibration certificate

Factory calibration KL90xx radiation for sensor (see chapter Calibration certificates)

14.12

Order no.

FLA603LSM4

Water analysis

Content

pH value, redox potential, conductivity

The Electrical Conductivity

15.02

15.03

Solute oxygen in liquids 15.03

pH one-bar measuring chains FY 96 PHEK, FY 96 PHER, FY 96 PHEN 15.04

pH insertion electrode FY 96 PHEE

Redox one-bar measuring chains FY 96 RXEN

Accessories for pH and redox probes

ALMEMO

®

connecting cable for pH and redox probes

Conductivity probes FYA 641 LFP1 / LFL1

Conductivity probes FYA 641 LFP2 / LFL2, FYA 641 LFP3

15.05

15.05

15.05

15.06

15.07

15.08

Oxygen sensor for O2 measurement in liquids FYA 640 O2 15.09

15.01

Water analysis

The pH Value

The pH value is a logarithmic measure for the concentration of the H ions in a hydrous solution and indicates, by a numerical value, whether the solution has an acid, neutral or alkaline reaction.

The pH scale ranges from pH0 to pH14, pH7 is neutral.

The further the pH value deviates from 7, the more aggressive the sample is. The acidic or alkaline effect will increase by the factor 10 per pH unit.

The illustration on the left shows some examples for pH values of typical substances

The Redox Potential

The level of the Redox potential (measured in mV) indicates the strength of an oxidising or reducing reaction of a measuring solution. A negative voltage value means that the solution has reducing properties compared to a standard hydrogen electrode. A positive value indicates that the solution has an oxidising effect.

As the extermination of microorganisms

(disinfection) is directly related to the strength of the oxidation (e.g. of chlorine) the Redox potential is successfully being used for monitoring disinfection processes, e.g. in swimming baths. However, redox measurements are also performed for controlling the denitrification of waste waters (redox break point determination) at the detoxification in galvanic plants and for monitoring multiple chemical processes (e.g. cyanide oxidation or chromate reduction).

ALMEMO

®

pH and Redox Measurement

By using reference solutions the calibration of pH and redox probes can be started with the push of a button. As the adjustment is stored in the ALMEMO

®

connector, the probe can also be used with other devices. If ALMEMO

®

devices with several input sockets are used, it is even possible to connect more probes with individual adjustments. The calculation of the pH value is based on the electrode steepness at 25°C. If the temperature of the measuring medium largely deviates from the reference temperature, it is possible for all

ALMEMO

®

devices to perform a temperature compensation.

15.02

Water analysis Water analysis

The Electrical Conductivity

The conductivity (unit S/m = Siemens/ meter) is a measure for the ion concentration in a measuring solution.

It is proportional to the salt, acid or base content in the measuring solution. Highpurity waters have a conductivity of approx. 0.05µS/cm (at 25°C), natural waters approx. 100 to 1000µS/m, some bases

(e.g. potassium hydroxide solutions) up to slightly more than 1000mS/cm.

The diagram shows further examples of hydrous solutions relevant for measurements.

In practice, the conductivity measurement is used for monitoring plants, for producing of high-purity waters or for determining the salinity of sea water.

Solute Oxygen

Oxygen is not only a component of the air but it is also contained dissolved in water and, practically, in every liquid. For example, water contains approximately

9mg/l oxygen in saturated compound at a temperature of 20°C and an atmospheric pressure of 1019mbar.

Every liquid accepts as much oxygen until the oxygen partial vapour pressure in the liquid is in a balance with the ´contacting´ air or gas phase. The saturation state

(air-saturated water) is reached when the partial pressure of the physically dissolved oxygen in the liquid equals the partial pressure of the oxygen in the air.

The current oxygen concentration increases with atmospheric pressures and with decreasing temperatures. Relevant for metrology are processes, such as the oxygen consumption involved with microbiological decomposition processes or an oxygen production, e.g. due to the growth of algae.

The oxygen concentration is very important for animals and organisms living in water and for the biological treatment of municipal and industrial waste water. Additionally, corrosion processes in lines and keeping the quality of beverages depend on the solute oxygen in the liquid.

This is only possible with ALMEMO

®

Devices

Through the complete electrical isolation of the measuring inputs it is possible to use only one single ALMEMO

®

device to simultaneously measure various chemical variables, and use several probes in one sampling vessel without having any mutual influences of the probes! Through pre-programmed ALMEMO

®

connectors it is possible to connect any environmental sensor technology.

ALMEMO

®

system with data logger and comprehensive sensor equipment Order no.

For exploring abandoned polluted areas and their environments or for performing groundwater quality tests

ALMEMO

®

data logger including sensor equipment and accessories

• ALMEMO

• pH electrode 1 to 12 pH including connecting cable and buffer solutions pH 4/7/10

®

2690-8 with 5 measuring inputs, including PC data cable

• Temperature sensor -70 to +400 °C

MA26908AKSU

FPA30L0250 + OFS0008

FY96PHEK + ZA9610AKY4W

+ ZB98PHPL4 + ZB98PHPL7

+ ZB98PHPL10 + ZB98PHNL

• Redox electrode including connecting cable and buffer solution 220 mV and KCl solution FY96RXEK + ZA9610AKY5W

+ ZB98RXPL2

• Conductivity probe 0.01 to 20.00 mS/cm including reference solution 2.77 mS/cm

FYA641LFP1 + ZB96LFRL

• Probe for measuring solute oxygen 0 to 40 mg/l or 0 to 260 % saturation including filling solution

FYA640O2

• Adjustment set for the oxygen probe, saturation and zero point adjustment

ZB9640AS

15.03

Water analysis

pH One-Bar Measuring Chain FY96PHEK

Applications:

manual measurements e.g. swimming pools, drinking water ...

Technical Data

pH range::

Operating range

Operating pressure:

Conductivity:

Diaphragm type:

1 … 12

0 ... 13pH / 0 … 60°C unpressurised

> 150 µS / cm glass fiber

Reference:

Shaft length:

Shaft diameter:

Electrode head:

Type

pH-one-bar measuring chain pH 1 … 12, 0 … 60°C for unpressurised operating

pH One-Bar Measuring Chain FY96PHER

Technical Data

pH range:

Operating range max. pressure:

Conductivity:

Diaphragm type:

Reference:

1 … 12

0 ... 13pH / 0 … 80°C

6 bar

> 50 µS / cm

PTFE ring diaphragm

Ag mit AgCl stock

(3mol KCl / polymer)

Type

pH-one-bar measuring chain pH 1 ... 12; 0 … 80°C

Ag / AgCl (3mol KCl / gel)

125 ±3mm

12mm (polycarbon) plug head SN6

Order no.

FY96PHEK

Applications:

Waste water, drinking water, industrial water, chemical industry, paper industry, food industry ...

(not media contained for chlorine and fluride, for not frequent temperature fluctuations).

Shaft diameter: screw connection

Shaft length:

Electrode head:

12mm (glass) thread PG13.5

120 ±3mm plug head SN6

Order no.

FY96PHER pH One-Bar Measuring Chain FY96PHEN

Applications:

manual measurements in the laboratory.

Technical Data

pH range:

Operating range

Operating pressure:

Conductivity:

Diaphragm type:

Reference:

0 … 12

0 ... 13pH / 0 … 80°C unpressurised

> 150 mS / cm, ceramik diaphragm

Ag / AgCl stock

(3mol KCl / liquid)

Shaft length:

Shaft diameter:

Electrode head:

Type

pH-one-bar measuring chain pH 0 … 12, 0 … 80°C for unpressurised operating

15.04

KCl-elektrolyt refillable

160 ±3mm

12mm (material: glass) plug head SN6

Order no.

FY96PHEN

Water analysis

pH Insertion Electrode FY96PHEE

Applications:

pH-measurings in semi-solid or pasty media, e.g. foods like meat, cheese ...

Technical Data

pH range:

Operating range

Operating pressure:

Diaphragm type:

Reference:

1 … 12

0 ... 13pH / 0 … 60°C unpressurised

3 ceramic diaphragms

Ag / AgCl

(3mol KCl / liquid)

Shaft length:

Penetrating tip

Electrode head:

Type

pH-insertion electrode pH 1 … 12, 0 … 60°C for unpressurised operating

KCl-elektrolyt refillable

120 ±3mm (glass) approx. 45 mm, Ø 6 to 8 mm plug head SN6

Order no.

FY96PHEE

Redox-One-Bar Measuring Chain FY96RXEK

Applications:

manual measurements e.g. swimming pools, drinking water ....

Technical Data

Operating temperature

Operating pressure:

Conductivity:

Diaphragm type:

0 … 60°C unpressurised

> 150 µS / cm glass fiber

Metal electrode :

Shaft length:

Shaft diameter:

Electrode head: platinum

125 ±3mm

12 mm (material: plastic) plug head SN6

Type

Redox-one-bar measuring chain 0 … 60°C for unpressurised operating

Order no.

FY96RXEK

Accessories for pH-One-Bar Meas. Chains and Redox-One-Bar Meas. Chain pH-One-Bar Measuring Chains Order no.

ALMEMO

1.2 m

5 m

®

transducer cable* for pH probes,

ZA9610AKY4W

ZA9610AKY4WL05

ALMEMO

1.2 m

5 m

®

transducer cable* for pH and redox probes,

ZA9610AKY6W

ZA9610AKY6WL05

Buffer solution pH 4.0 50 ml

Buffer solution pH 7.0 50 ml

Buffer solution pH 10.0 50 ml

KCl solution, 3-molar, 50ml for refilling and storage

ZB98PHPL4

ZB98PHPL7

ZB98PHPL10

ZB98PHNL

Order no.

Redox-One-Bar Measuring Chain Order no.

ALMEMO

1.2 m

5 m

®

transducer cable* for redox probes,

ZA9610AKY5W

ZA9610AKY5WL05

ALMEMO

®

1.2 m

transducer cable* for pH and redox probes,

5 m

Redox buffer solution 220 mV

KCl solution, 3-molar for refilling and storage, 50ml

ZA9610AKY6W

ZA9610AKY6WL05

ZB98RXPL2

ZB98PHNL

* Cable with spray-coated ALMEMO®connector

15.05

Water analysis

ALMEMO

®

connecting cable for pH and redox probes

Transducer cable with various electrodes

Applications:

Transducer cables are available for all popular electrodes with a coaxial connector. To avoid the measuring signal being corrupted by the measuring instrument itself an extremely high-impedance amplifier is integrated in the ALMEMO

®

connector on the connecting cable . It is also possible, by means of impedance conversion and differential measurement, to measure several electrodes with different potentials, -free from interference and using only one ALMEMO

®

device.

Technical Data

Transducer High-impedance measuring amplifier (>500 Gohm), integrated in the ALMEMO® connector

Electrode terminal For plug-on head S7/SN6 or SMEK (see variants)

Type

ALMEMO

®

connecting cable with transducer (ALMEMO

®

connector, spray-coated)

For probes with plug-on head S7/SN6 (coaxial connector, screw-fit):

Programming for pH probe

Cable length 1.2 meters

Cable length 5 meters

Programming for redox probes

Cable length 1.2 meters

Cable length 5 meters

Programming for pH or redox probe (1 probe connectable at a time)

Cable length 1.2 meters

Cable length 5 meters

Order no.

ZA9610AKY4W

ZA9610AKY4WL05

ZA9610AKY5W

ZA9610AKY5WL05

ZA9610AKY6W

ZA9610AKY6WL05

Type

ALMEMO

®

connecting cable with transducer

For probes with SMEK plug-on head

Cable length 2 meters

Programming for pH probe with integrated temperature sensor NTC (30 kohm at 25 °C), linearization saved in ALMEMO

®

connector (only for current V6 ALMEMO

®

devices)

Programming for pH probe

Programming for redox probe

Order no.

ZA9640AKY8

ZA9610AKY8

ZA9610AKY9

NTC temperature sensor for automatic temperature compensation when measuring pH

Connector programming designation *T for ALMEMO

®

2490 and 2590-2/-3S/-4S and (with effect from 07/2006) for

ALMEMO

®

2690/ 2890/ 5690/ 8590/ 8690

Type

Stainless-steel sheathed sensor (see page 07.06) Diameter 3.0 mm, length 250 mm, Hexagonal cable sleeve with 1.5 meters PVC cable and ALMEMO

®

connector

Safety hose made from PTFE (for aggressive media) Hermetically sealed on one side, inside diameter 3.0 mm, outside diameter 4.0 mm, length 700 mm

Order no.

FNA30L0250T

ZT9000TS7

15.06

Water analysis

Conductivity Probe FYA641LFP1 / LFL1

Applications:

Concentrated waste water, aggressive waters, general aqueous and partly aqueous solutions, beer, emulsions, electroplating, waters, concentrated acidic and alkaline solutions, corrosive acids and alkaline solutions, lacquers and paints, substances containing protein, soaps, detergents, suspensions, titrations in organic substances, environmental analysis.

Technical Data

Measuring range:

Temperature sensor:

Temperature compensation:

Compensation coefficient:

Cell constant:

Electrode material:

Accuracy:

0.01 to 5mS/cm:

5 to 20mS/cm:

Nominal temperature:

Operating temperature:

0.01 to 20mS/cm

LFL1 up to 10mS/cm

NTC, type N (10k at 25°C)

0 to +70°C, automatic

1.9 linear approx. 1cm

-1 special coal

± 1% of meas. val. ± 0.05mS

± 2% of meas. val. ± 0.05mS

25°C ± 3°C

–5 to 70°C

Accessories

Reference solution 2.77mS/cm at 25°C 0.02mol KCl, 250ml

Minimum insertion depth:

Shaft material:

Shaft length/shaft diameter:

Fitting length / thread

Maximum pressure

Cable length:

Power supply:

Current consumption:

30mm

PVC - C

LFP1: 130mm/20mm

LFL1: 130mm/10mm only LFP1 145 mm / G¾”

LFP1: 16 bar at 25 °C

LFL1: not suitable for use

under pressure

1.5m

8 to 12V through meas. instr.

approx ca. 3 mA

Type

(including manufacturer’s test certificate)

Active conductivity probe with automatic temperature compensation, Built-in probe, G 3/4“ thread, suitable for use under pressure up to 20mS/cm

Laboratory probe, not suitable for use under pressure up to 10mS/cm

Factory calibration KY90xx conductivity for measuring chain (sensor + device) (see chapter Calibration certificates)

Order no.

ZB96LFRL

Order no.

FYA641LFP1

FYA641LFL1

15.07

Water analysis

Conductivity Probe FYA641LFP2 / LFL2

Applications:

Low-salt waste water, general aqueous and partly aqueous solutions, fish tanks, emulsions, desalting/ion exchanger, beverages, waters, cold/boiler feed water, lacquers and paints, milk, samples with low ionic strength, substances containing protein, purest water, soaps, detergents, suspensions, drinking water, environmental analysis.

Technical Data

Measuring range: 1 to 200µS/cm

Temperature sensor: NTC, type N (10k at 25°C)

Temperature compensation: 0 to +70°C, automatic

Compensation coefficient:

Cell constant:

1.9 linear approx. 1cm

-1

Electrode material:

Accuracy:

Nominal temperature:

Operating temperature:

Minimum insertion depth: special coal

± 2% of meas. val. ± 0.5µS

25°C ± 3°C

–5 to 70°C

30mm

Zubehör

Reference solution 147µS/cm at 25°C 0.001mol KCl, 250ml

Shaft material: PVC - C

Shaft length/Shaft diameter: LFP2: 130mm/20mm

LFL2: 130mm/10mm

Fitting length / thread only LFP2 145 mm / G¾”

Maximum pressure

Cable length:

LFP2: 16 bar at 25 °C

LFL2: not suitable for use

under pressure

1.5m

Power supply:

Current consumption:

8 to 12V through meas. instr.

approx. 3 mA

Order no.

ZB96LFRL2

Type

(including manufacturer’s test certificate)

Active conductivity probe 0 … 200µS/cm with automatic temperature compensation,

Built-in probe, G 3/4“ thread, suitable for use under pressure

Laboratory probe, not suitable for use under pressure

Factory calibration KY90xx conductivity for measuring chain (sensor + device) (see chapter Calibration certificates)

Order no.

FYA641LFP2

FYA641LFL2

Conductivity Probe FYA641LFP3

Technical Data

Measuring range:

Temperature sensor:

Cell constant:

Electrode:

Accuracy:

Nominal temperature:

Operating temperature:

Minimum insertion depth:

0 to 200 mS/cm

NTC, type N (10k at 25°C) approx. 1cm

-1

4 electrodes, special coal

1 mS/cm ± 1.5% of meas. val.

25°C ± 3°C

0 to 70°C

30mm

Accessories

Reference solution 111.8mS/cm at 25°C 1mol KCl, 250ml

Type

(including manufacturer’s test certificate)

Conductivity probe 0 … 200mS/cm without temp. compensation

15.08

Applications:

Concentrated waste water, aggressive waters, general aqueous and partly aqueous solutions, beer, emulsions, electroplating, waters, concentrated acid and alkaline solutions, corrosive acids and alkaline solutions, lacquers and paints, substances containing protein, soaps, detergents, suspensions, titrations in organic substances, environmental analysis.

Shaft material:

Shaft length:

Shaft diameter:

Fitting length / thread

Maximum pressure

Cable length:

Power supply:

Current consumption:

PVC - C

145mm

20mm

130 mm / G¾”

16 bar at 25 °C

1.5m

8 to 12V through meas. instr.

approx. 15 mA

Order no.

ZB96LFRL3

Order no.

FYA641LFP3

Factory calibration KY90xx conductivity for measuring chain (sensor + device) (see chapter Calibration certificates)

Water analysis

Oxygen Sensor FYA640O2

Applications:

Determination of the conditions of life for fish and microorganisms in waters and fish tanks, biological treatment of municipal and industrial waste water, storage of organic liquids, examinations of drinking water, control of corrosion processes in heating system lines, examination of qualitykeeping of beverages.

Technical Data

Measuring ranges:

Temperature range:

O2 saturation:

O2 concentration:

Measuring principle:

Working electrode:

Reference electrode:

Membrane:

–5.0 … 50°C

0 … 260% saturation

0.0 … 40mg/l (5 … 40°C)

Clark

Pt cathode

Ag/AgCl counter electrode

PTFE

Response time (t

90

%): approx. 10–15s

Zero current at 0% saturation: < 5nA

Meas. current at 100% saturation: approx. 700nA

Accuracy, oxygen measurement: < ± 1% of measured value

Velocity in blower stream: approx. 10cm/s

Storage temperature:

Insertion depth:

–10 … 50°C

40mm

Filling volume (electrolyte): 0.6ml

Temperature sensor: NTC type N (10k at 25°C)

Accuracy of temp. measurement

(at nominal conditions): –20 … 0°C: ±0.4°C,

0 … 70°C: ±0.1°C

Nominal conditions: 25°C ±3°C/1013mbar

Shaft material:

Membrane cap:

Shaft length/shaft diameter:

Connecting cable:

PVC, black replaceable (spare)

145mm/12mm

1.5m long with spray-coated

ALMEMO

®

connector

Polarisation voltage: 650mV

Service life

(with one electrolyte filling): several months

Total service life (durability): several years

Accessories

Adjustment set consisting of:

25g sodium sulphite in 20ml PE bottle for preparation of the null solution, vessel for adjustment of the saturation level

25g sodium sulphite in 20ml PE bottle

20ml filling solution in PE bottle for O2 probe

Spare membrane cap with protection (2 pieces)

Order no.

ZB 9640 AS

ZB 9640 NS

ZB 9640 NL

ZB9640EM

Type

Oxygen sensor for O

2

measurements in liquids incl. connecting cable 1.5m long with spray-coated ALMEMO

®

connector

Order no.

FYA640O2

15.09

15.10

Gas concentrations in the air

Content

The importance of measuring the quality of room air

Digital carbon dioxide sensor FYAD 00 CO2B10

Carbon dioxide probe FYA 600 CO2

Carbon monoxide probe FYA 600 CO

Oxygen probe FYA 600 O2

Ozone sensor, measuring transducer FYA 600 O3

Gas probe for various gases FYA 600 A

16.02

16.04

16.05

16.06

16.06

16.07

16.08

16.01

Gas concentrations in the air

Why is the Measurement of Room Air Quality So Important?

An unsatisfactory room air quality of indoor rooms (e.g. in offices) can easily cause tiredness, poor powers of concentration and even diseases to people. Indicator for the room air quality is the concentration of specific gases in air. The most important ones include:

• Carbon dioxide (CO

• Oxygen (O

2

2

)

• Carbon monoxide (CO)

• Ozone (O

3

)

)

CO

2

-Concentration

An important criterion for the evaluation of the room air quality is the CO

2 tration. A CO

2

concen-

concentration, which is too high due to insufficient ventilation, is experienced as stale or stagnant air. The illustration above shows the range of CO concentrations that are relevant to a hu-

2 man.

CO-Concentration

CO is produced when carbon is only partially combusted (fuel). CO is very dangerous for humans because it is at the same time highly toxic - but invisible and odorless. Reasons for the production of

CO in various combustion processes:

• defi ciency of air

• too high excess of air

• too early cooling down of fl ame

Effects of CO in the ambient air on the human body

CO concentration Inhalation period and consequences

30 ppm 0.0003% Maximum concentration in the workplace per 8-hour shift (German MAK value)

200 ppm 0.02% Slight headache within 2 to 3 hours

400 ppm

800 ppm

0.04%

0.08%

Headache within 1 to 2 hours, first in the forehead and temples, then spreading to the whole head

Dizziness, nausea, and twitching limbs within 45 minutes, unconsciousness within 2 hours

1600 ppm

3200 ppm

6400 ppm

12800 ppm

0.16%

0.32%

0.64%

1.28%

Headache, dizziness, nausea within 20 minutes, death within 2 hours

Headache, dizziness, nausea within 5 to 10 minutes, death within 30 minutes

Headache and dizziness within 1 to 2 minutes, death within 10 to 15 minutes

Death within 1 to 3 minutes

Applications

• measurement, control, and warning system in garages,

• monitoring of room air quality with respect to maximum permissible workplace concentration (MAK value)

• monitoring of outside air or of protected air systems in domestic and large public shelters.

16.02

Gas concentrations in the air Gas concentrations in the air

O

2

-Concentration

The inhaled air consists of vital oxygen at a ratio of 1:5. Oxygen is required for all oxidation processes; for combustion processes, as well as for silent oxidations.

Examples include the rusting of iron, oxidations, which occur in living processes, or the decomposition of organic material. Additionally, all combustion processes that release energy require this gas, for example, heating systems or aircraft engines. However, oxygen is also bound with any type of noxious fires such as forest and heath fires. Due to the permanent cycle of assimilation and photosynthesis in green plants when they are subject to sunshine, oxygen is continuously re-formed from carbon dioxide. The balance between oxygen consumption and oxygen production is disturbed by the continuously increasing combustion of fossil combustibles.

Therefore, many areas require control measurements of the oxygen content in the air, e.g. in air condition systems, air purifiers, oxygen rectifiers, greenhouses and oxygen incubators, as well as for exhaust emission tests, e.g. in the automotive industry.

O

3

-Concentration

The ozone contained in the earth’s atmosphere forms at altitudes of approximately 30km. It provides a protective shield around the earth and filters out approximately 50% of the solar

UV radiation, particularly the short-wave range, which is dangerous for living organisms. However, ozone is toxic and an extremely aggressive trace gas that can cause major burns in human mucous membranes when breathed in high concentrations. Therefore, control measurements for the ozone content in air must be performed in many areas, e.g. leakage tests in industry, protection of health and safety standards at work, mobile-based air quality measurements or for providing environmental data on advertising displays etc.

Calculation Formulae

The following formulae are used for converting the O3 measured value from ppb to µg/m³, depending on the current atm. pressure and the temperature.

Example:

20°C and 1013 hPa = factor 2

Ozone (µg/m³) = 2 x Ozone (ppb)

This is the nominal value for conversion from ppb to µg/m³.

0,57 x Atm. Press. [hPa]

Ozone(g/m³) = x Ozone (ppb)

Temperature [K]

16.03

Gas concentrations in the air

ALMEMO

®

D6

Digital carbon dioxide sensor FYAD 00 CO2B10 with grip, integrated atmospheric pressure sensor for automatic atmospheric pressure compensation, and ALMEMO

®

D6 plug

General features and accessories, ALMEMO

®

D6 sensors:

see page 01.08

• Digital CO2 sensor with integrated signal processor

• All sensor characteristics and adjustment data are stored in the

CO

2

sensor itself.

• The unique automatic calibration procedure (without fresh air intake) automatically compensates any natural ageing effects.

• The sensor is very well protected against the effects of pollution by means of replaceable PTFE filter caps. Long-term stability is outstanding.

• new:

Automatic atmospheric pressure compensation is provided for pressure-dependent CO2 concentrations by means of a digital atmospheric pressure sensor integrated in the grip.

• The relevant ambient parameter, atmospheric pressure, is measured using the same sensor.

• new:

Long-term measuring operations can be performed with an ALMEMO

®

data logger in sleep mode; this applies only to current device types with sleep delay (180 seconds).

• 2 primary measuring channels (real measurable variables)

CO

2

concentration and atmospheric pressure

• Freely selectable measurable variables Two measuring channels are programmed (at our factory).

CO

2

concentration, average value (ppm), Atmospheric pressure (mbar, AP, p).

Alternatively a further variable can be selected.

CO

2

concentration, current value (ppm)

This device can be configured on a PC using USB adapter cable ZA 1919 AKUV. (see “General accessories for AL-

MEMO

®

D6 sensors” page 04.05).

Technical Data

Digital carbon dioxide (CO2) sensor (including A/D converter)

Measuring principle

non-dispersive infrared (NDIR) technology

Sensor

Measuring range

Accuracy

Nominal conditions

2-beam infrared measuring cell

0 to 10,000 ppm

±(100 ppm +5 % of meas. value)

+25 °C, 1013 mbar

Temperature dependence typical 2 ppm CO2 / K in range 0 to +50 °C

Response time

Operative range

Measuring interval

Filter cap

<195 seconds

-40 to +60 °C / 0 to 95 % RH

(non-condensing)

Moving average 165 seconds

(= 11 current values of 15 sec.)

PTFE

Diameter 18 mm

Length appr. 41 mm

Sensor connector

Grip

Dimensions:

Plug connection

with socket, integrated electronics

Diameter 20 mm

Total length including the sensor

245 mm

ALMEMO

®

connecting cable fitted cable, 2 meters

With ALMEMO

®

D6 plug

Digital atmospheric pressure sensor (integrated in grip)

Measuring range

Accuracy

ALMEMO

®

D6 plug

700 to 1100 mbar

±2.5 mbar (at 0 to +65 °C)

Refresh rate

Supply voltage

Current consumption

1 second for all four channels

6 to 13 VDC

25 mA

Type

(including factory test certificate)

Digital CO

2

sensor with grip, fitted cable with ALMEMO

®

D6 plug, and integrated digital atmospheric pressure sensor

Factory calibration KY96xx carbon dioxide concentration for digital sensor (see chapter Calibration certificates)

16.04

Order no.

FYAD00CO2B10

Carbon Dioxide Probe FYA600CO2

Gas concentrations in the air

• Since the gas is supplied by means of free convection, this is especially suitable for climatology measurements.

• Various measuring ranges up to 25%.

Technical Data

Gas:

Measuring principle:

Measuring ranges:

CO

2

IR optics nominal (% CO

2

):

0 … 2.5%, 0 … 10%, 0 … 25%

Accuracy:

Reproducibility:

Resolution:

Output:

±2% of final value

±1% of final value

(depending on measuring range)

<200ppm at 2.5%

0 … 2V on ALMEMO

Current output: referred to GND max. burden (load resist.): 400W

®

connector

Linearization in ALMEMO

®

device

Power supply:

Current consumpt.

6.5 to 12VDC from the ALMEMO

®

device

Operation with mains supply unit recommended !

eff. 50mA/ max. 70mA

Settling time t90: < 60s

Temperature coefficient: typical –0.4% signal/K

Temperature range:

Relative humidity:

5 to +40°C

0 to 95%, noncondensing

Dimensions:

Weight:

Connecting cable:

W 96mm x H 36mm x D 64mm

241g

1.5m long, ALMEMO

®

connector

!

Operation with the device in SLEEP mode is not possible!

When operating more than one CO

ALMEMO

®

device, these CO

2

2

probe on a single

probes will need their own external power supply ! On request we can offer a wide variety of power supply options to suit your particular measuring setup.

Type

Carbon dioxide sensor including connecting cable 1.5m long for CO

2

(Please specify measuring range !)

measurements in air

Order no.

FYA600CO2

Factory calibration KY96xx carbon dioxide concentration for measuring chain (sensor + device) (see chapter Calibration certificates)

16.05

Gas concentrations in the air

Carbon Monoxide Probe FYA600CO

• Applications:

For measurement, control and warnings in garages, for monitoring the air quality with respect to the maximum allowable concentration at work places (MAC value, e.g. in laboratories and engine test benches)

!

Operation with the device in SLEEP mode is not possible!

Technical Data

Gas:

Measuring principle:

Measuring range:

CO electrochemical reaction see types

Zero point error: < 10 ppm CO

Gauge reading balance: < 3 ppm CO

Error of meas. value:

Zero point drift:

±3% of full scale value

< 2% (1 year)

Reproducibility:

Linearity:

Settling time t

90

:

< 2% (1 year)

< 2% of full scale value

< 60s

Transverse sensitivity:

Output:

Supply voltage:

Ambient temperature:

Air humidity:

< 2% by integrated filter

4 ... 20 mA on ALMEMO

®

connector from the ALMEMO

®

measuring instrument

–10 to +40°C, sensor temperature compensated in range

0 to 90% non-condensing

Life span of the meas. cell: approx. 2 years typical

Dimensions of meas. head: Ø 80mm, height 80mm

Weight:

Connecting cable:

600g

1.5m, with ALMEMO

®

connector

Ausführung

(incl. factory test certificate)

Order no.

Carbon monoxide sensor including connecting cable 1.5m long for CO measurements in air range: 0 ... 150 ppm

FYA600COB1

range: 0 ... 300 ppm range: 0 ... 5000 ppm range: 0 ... 5 Vol.%

FYA600COB2

FYA600COB3

FYA600COB4

Oxygen Probe FYA600O2

• Examples from the range of applications:

Measurements in air conditioning systems, air purifiers, oxygen rectifiers, greenhouses and oxygen incubators.

• Approved by PTB and approved for exhaust emission measurements in the automotive industry.

!

A correction value can be stored in the ALMEMO

®

connector plug to compensate for the natural ageing of the probes, so optimum output characteristics can be ensured for the whole operating life.

Technical Data

Gas:

Measuring principle:

Measuring range:

Accuracy :

Resolution :

Response time:

Signal drift:

Offset voltage at 20°C:

O

2 electrochemical cell

1 … 100% O

2

, linear

1% O

2

0.01% O

2

< 40s

< 2% signal/month

(typ. < 5% over operating life)

< 20mV

Operating life:

Nominal conditions:

2 years, if operated in 20.9% O

2

20°C, 50% rH, 1013mbar

Temperature range: –20 to +50°C

Temperature compensation: effective in range –10 to +40°C

Pressure range:

Relative humidity: atm. pressure ±10%

0 to 99% non-condensing

Connecting cable:

Dimensions: adapter cable 1.5m long

H 43 mm x Ø 29,3 mm

Types Order no.

Oxygen sensor including connecting cable 1.5m long for O

2

measurements in air

FYA600O2

16.06

For Reordering:

Oxygen sensor

ALMEMO

®

connecting cable

FY9600O2

ZA9600AKO2

Ozone Measuring Transducer FYA600O3

Gas concentrations in the air

• Suitable for many measuring tasks where ozone measurements for control purposes were too expensive to date, e.g. for leakage tests in industry, for protection of health and safety standards at work, for mobile air quality measurements etc.

• Each ozone sensor is supplied with a manufacturer’s test certificate.

• As a result of the high long-term stability, only small maintenance costs.

Technical Data

Gas:

Measuring principle:

Measuring range:

Detection limit

Accuracy:

Long term accuracy:

Exposure period :

Meas. interval:

Pump flow rate:

Signal output:

O

3

(ozone) electrochemical

three-electrode sensor

0 … 300 ppb

20 ppb typically 5% of final value under nominal conditions (for intermittent operation) after 12 months under nominal conditions typically 5% of final value (for intermittent operation) until specification is reached, at least 2 hours (at 200 ppb); for a prolonged period the device was in an ozone-free environment pump on: 5min pump off: 10min

500ml/min

0 … 2V, load resistance > 100kW

Power supply:

Current consumption:

Overload capacity:

Expected useful life :

Nominal conditions:

Operating range :

Storage temperature:

Dimensions:

Connecting cable:

6 to 14V, stable pump on : 50 mA, typical pump off : 25 mA, typical pump blocked : 180 mA, typical

1 ppm

Sensor, typically 24 months (at 20 °C) pump, typically 6000 hours

20°C, 30% r.H., 1013 mbar, no contaminations of the contact surfaces

-20 to +40 °C / 30 to 80 % RH

0 to 20°C, at 30 to 80% RH non-condensing

L 180mm x W 125mm x H 90mm

1.5m long with ALMEMO

®

connector programmed in ppb

Type

(including manufacturer’s test certificate)

Ozone sensor including connecting cable 1.5m long for O

3

measurements in air

Option:

Pump in continuous operation (fixed factory setting)

Maintenance set : new electro-chemical measuring cell, pump replacement, readjustment, including calibration certificate

Order no.

FYA600O3

OY9600O3D

ZB9600O3S

16.07

Gas concentrations in the air

Gas probe for various gases FYA600A

• Range:

Measurement of gas concentration in air

• multiple ranges / Modelvariants

Technical Data

Gas:

Measuring principle:

Measuring range:

Error of meas. value:

Zero point drift:

Reproducibility:

Linearity:

Settling time t

90

:

Transverse sensitivity: see model variants electrochemical reaction see model variants

±3% of full scale value

< 2% (1 year)

< 2% (1 year)

< 2% of full scale value

< 60s

< 2% by integrated filter

!

Operation with the device in SLEEP mode is not possible!

Output:

Supply voltage:

Ambient temperature:

Air humidity:

4 ... 20 mA on ALMEMO from the ALMEMO instrument

®

®

connector

measuring

–10 to +40°C, sensor temperature compensated in range

0 to 90% non-condensing

Life span of the meas. cell: approx. 2 years typical

Dimensions of meas. head: Ø 80mm, height 80mm

Weight:

Connecting cable:

600g

1.5m, with ALMEMO

®

connector

Model variants

(including factory test certificate)

Gas probe, including connecting cable, 1.5 meters, for measuring gas in air

Ammonia NH

3

Range: 0 ... 250 ppm

Nitrogen dioxide NO

2

Range: 0 ... 30 ppm

Nitrogen oxide NO

Range: 0 ... 50 ppm

Chlorine gas Cl

2

Range: 0 ... 50 ppm

Sulfur dioxide SO

2

Range: 0 ... 20 ppm

Range: 0 ... 50 ppm

Range: 0 ... 250 ppm

Hydrogen sulfide H

2

S

Range: 0 ... 50 ppm

Range: 0 ... 250 ppm

Ethylene oxide C

2

H

4

O

Range: 0 ... 20 ppm

Range: 0 ... 50 ppm

Range: 0 ... 100 ppm

16.08

Order no.

FYA600ANH3

FYA600ANO2

FYA600ANO

FYA600ACL2

FYA600ASO2B1

FYA600ASO2B2

FYA600ASO2B3

FYA600AH2SB2

FYA600AH2SB3

FYA600AC2H4OB1

FYA600AC2H4OB2

FYA600AC2H4OB4

Calibration certificates

Content

What You Should Know About Calibration 17.02

Test instrument

Simulator Typ KA 7531

Adjustment Set for ALMEMO

®

Devices

Calibration certificates - temperature and pressure - sensor deviation reduced to zero

Calibration certificates for:

Temperature

Infrared Temperature Measurement

Relative Air Humidity for Capacitive Humidity Sensors

Relative Air Humidity for Psychrometers

Relative Air Humidity at Temp 95°C

Relative Air Humidity, temperature, and atm. pressure for FMA510

Dew Point

Pressure

Air Flow

Flow measurement for liquids

Conductivity

Gas Concentration

Measurable Variables for Optical Radiation

Optical Speed Sensors

Force

Electrical Calibration of Measuring and Indicating Devices

17.04

17.06

17.06

17.11

17.11

17.12

17.12

17.12

17.12

17.13

17.13

17.07

17.08

17.08

17.09

17.09

17.09

17.09

17.10

17.01

Calibration certifi cates

What You Should Know About Calibration

With the introduction of quality management standards all over the world, the requirements for measuring and test devices have become significantly more demanding. For example, the certification according to DIN EN ISO 9000+ requires an active quality management involving regular calibrations. With consideration of the specific environment this ensures a high reliability regarding the measuring results and the traceability of the measured values to the national standard.

´Calibration´ is Not Identical with ´Calibration by the Bureau of Standards´

´Calibration´ includes those activities involved with determining, under given conditions, the relationship between the output values of an instrument or a measuring equipment and the corresponding values of a variable, which are determined by reference materials.

The Result of a Calibration

1. The result of a calibration allows the evaluation of errors of dimension of the measuring instrument, measuring equipment or the setup of measuring instruments or the allocation of values to any scaled graduation marks.

2. The result of a calibration can be fixed in a document, which is often called a

´calibration report´ or a ´calibration certificate´.

3. In many cases, the result of a calibration is specified as correction or ´calibration factor´ or as ´calibration curve´.

´Calibration by the Bureau of Standards´

The expression ´Calibration by the

Bureau of Standards´ is limited to the legal metrology and denotes governmental verifications, according to the calibration regulations. This type of calibration must only be performed by the

Bureau of Standards responsible for the devices that are appropriate for verification.

17.02

Calibration certificates Calibration certificates

DAkkS Calibration

• The calibration must only be performed within the range of those measurable variables, measuring ranges and measuring incertainties, which are specified in the accreditation document. The customer receives a DAkkS calibration certificate specifying the measured values, the corresponding measuring incertainty, the designation of the calibration method, the environmental conditions and, as required, information on special measurement conditions.

The calibrated object will be identified by a label (red).

The label contains the following data:

• Number of calibration

• Number of the calibration laboratory

• Date of calibration (month/year) and refers to the calibration certificate.

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025 (inter alia for the monitoring of production processes and the quality assurance applied to products).

!

DAkkS-calibrations outside the range of accreditation services provided by the DAkkS calibration laboratory at

Ahlborn Mess- und Regelungstechnik GmbH are performed by DAkkS laboratories run by our various partners.

Factory Calibration

The calibration is performed according to comparison measurements with factory standards. Factory standards are (as far as possible) PTB or DAkkS calibrated measuring instruments, sensors or measuring systems. The customer receives a factory calibration certificate specifying the measured values, the corresponding measuring incertainty, the designation of the calibration method, the environmental conditions and, as required, information on special measurement conditions.

The calibrated object will be identified by a label.

!

Factory calibrations outside the range of accreditation services provided by the calibration laboratory at

Ahlborn Mess- und Regelungstechnik GmbH are performed by laboratories run by our various partners.

How Often To Calibrate?

The time interval between calibrations highly depends on the specific application and is influenced by the following parameters:

• Permissible measuring tolerances

• Results of previous calibrations

• Environmental conditions

• Customer-specific requirements and definitions

• Application frequency

17.03

Test instrument

Simulator KA 7531

Simulator for Pt100, thermocouples, mV, V, mA, Hz

Option PC interface

Technical features

• Universal manual simulator for simulating temperature sensors and process variables when testing measuring instruments, regulators, and other equipment

• Pt100 simulation with 5 fi xed resistors in 4-conductor technology

Voltage and thermocouples simulation with 15-bit D/A converter

Current simulation with 15-bit D/A converter

Frequency and pulse generator with quartz-crystal oscillator

Continuity check with settable threshold

• All signals are available at the same time.

• Signals can be set either manually or automatically, in step or ramp form.

• All signals and all the programming can be shown on the illu-

Technical data

Signal Pt100

Accuracy

Temperature drift

Signal voltage

-10 to +60.000 mV

-3 to 10.000 V

Accuracy

Temperature drift

Time constant

Thermocouples

Accuracy:

CJ - temperature:

Signal current

0 to 20.0 mA

Accuracy

Temperature drift

Time constant

5 discrete resistance values in

4-conductor technology

0 / 50 / 100 / 200 / 300 °C

±0.1°C

0.01°C / K

15-bit DAC electr. isolated load > 1 MΩ load > 100 kΩ

± 0.05% ± 0.05% of fi nal value

20 ppm / K

100 µs type K, N, T, J (ITS90) resolution: 0.1K

type S, R, B (ITS90) resolution: 1K

± 0.05% ± 0.05% fi nal value

-30..100°C

15-bit DAC electr. isolated load < 500 Ω

± 0.05% ± 0.05% of fi nal value

20 ppm / K

100 μs minated graphics display.

• Connection of peripherals via ALMEMO

®

clamp connectors, cable with anti-kink protective sleeve and strain relief

• Power supply via battery, mains unit, USB cable ZA 1919-

DKUV or connection to RS422 network distributor with connector ZA5099-FSV

• Modern, compact housing - also suitable for DIN top-hat rail mounting

• Option of PC-controlled operation via all ALMEMO

®

data cables.

signal frequency

Pulse width

Accuracy

Pulse range

Period

Pulse

Accuracy

Continuity

Threshold

Power supply:

Battery:

Current consumption

Voltage and Current output: with illumination:

Display

Illumination

Keypad

Housing

Operating range:

Operating temperature:

(Storage temperature:

Ambient humidity:

1..4000Hz, 0.01..10.00kHz,

0.1..40.0kHz, 1..100kHz

1 to 99 % corresponds to the resolution

2µs...99.999 ms, 2ms...99.999 s

1µs...99.998 ms, 1ms...99.998 s

0.01 % current approx. 1 mA

0 to 1000 mV

10..12V DC

3 Mignon Alcaline

(Battery): approx. 30 mA approx. 80mA + 4 x IOUT, approx. 40mA additional graphics 128 x 64 (55 x 30 mm)

2 white LEDs

7 silicone keys (4 soft-keys)

(LxWxH) 127 x 83 x 42 mm

ABS (-10 to +70 °C), 290 g

-10 ... +50 °C

-20 ... +60 °C)

10 ... 90 % rH (noncondensing)

17.04

Test instrument

Displays (examples)

Main menu Device configuration

Pt100 resistance values NiCr-Ni step generator

Continuity tester

Frequency pulse generator

0 to 10 V step generator 0 to 20 mA ramp generator

Accessories

ALMEMO

®

clamp connector (for Pt100 or universal use)

ALMEMO

®

connecting cable with 2 banana plugs and 2 test probes

Mains adapter 12 V / 1 A

USB data cable, electrically isolated

As above but with 9 V supply, not electr. isol.

V24 data cable, electrically isolated

Fixture for top-hat rail mounting

Rubber guard, gray

Order no.

ZA1000TS

ZA1000TK

ZA1312NA7

ZA1919DKU

ZA1919DKUV

ZA1909DK5

ZB2490HS

ZB2490GS2

Options

Order no.

Factory calibration certificate for Simulator KA7531:

Electrical Calibration compared reference standards that are traceable to national standards.

Calibration in 6 ranges: Pt100 (5 points), and (3 points each) voltage 10 V, voltage 50 mV, current 20 mA, thermocouple type K, frequency Hz,

Pakkage Offer

KE9006W

addressable PC interface

OA7531I

Included as standard

Simulator, 5 sockets for Pt100, thermocouples or -4 to 10 V, 0 to 20 mA, frequency, continuity tester,

Graphics display and keypad, sockets DC, A1, batteries, including manufacturer’s test certificate,

1 ALMEMO

®

clamp connector (for Pt100) and ALMEMO

®

and ALMEMO

®

connecting cable with 2 banana plugs and 2 test probes

Order no.

KA75311

17.05

Calibration certificates

Adjustment Set for ALMEMO

®

Devices

Type

Adjustment Set for ALMEMO

®

Devices

Input connector with 1.5 m cable and 4 banana plug (for connection to the calibrator of the customer) including ALMEMO

®

Adjustment instructions and software AMR-Control (CD)

Order no.

ZA9090AKA

Calibration certificates - temperature and pressure - sensor deviation reduced to zero

The patented ALMEMO

®

measuring technology from Ahlborn has proved itself many times over in a wide variety of fields; its use in research and development is just one example. Quality assurance, resources monitoring, and calibration systems also constitute a very broad field for ALMEMO

®

measuring technology.

Wherever measured values must be exactly traceable to a known standard and measuring equipment needs to be thoroughly checked at regular calibration intervals ALMEMO

®

is the answer - both as the reference device in the calibration laboratory and as a calibratable device for the customer; it is precisely in this context that the quality of AHL-

BORN measuring instruments is so decisive.

The measuring chain is corrected by means of a new multi-point adjustment function.

When ALMEMO

®

measuring equipment is calibrated the sensor deviation is determined at each and every calibration point and then saved as correction value for that calibration point in the patented

ALMEMO

®

connector. The measured values for the sensor obtained using this multi-point adjustment function are entered in the calibration certificate.

Sensor deviations are thus reduced to zero.

Measuring operations within the calibration interval can thus be performed with much less uncertainty. The measured value displayed on the ALMEMO

®

measuring instrument has already been corrected and can be used directly. Correction of the displayed measured value on the basis of the sensor deviation in the calibration certificate, as would otherwise be the case, is thus no longer necessary.

17.06

Calibration certificates

Temperature

Caibration certificate for temperature measuring chains consisting of a contact temperature sensor and an instrument (also individual sensors). This calibration can only be performed as an immersion measurement with sensors or sensor + measuring instrument (measuring chain).

Temperature range: Pt100 thermocouples

-196°, –90° to +660°C

-196°, –80° bis + 1600°C

DAkkS Calibration Certificate Order no.

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025 (inter alia for the monitoring of production processes and the quality assurance applied to products).

Calibration is performed by a DAkkS authorized office which compares measured values against reference values based on national standards. .

Package offer (basic rate + 3 points 0°C, 50°C, 100°C)

Measuring points, freely selectable

Basic fee per certificate

Measuring point fee per sensor, one measuring point

Measuring point fee per sensor, one measuring point -196 °C

KT9021D

KT9001D

KT9011D

KT9012D

Factory Calibration Certificate

Calibration compares measured values against reference values based on national standards.

Package offer (basic rate + 3 points 0°C, 50°C, 100°C)

Package offer only for temperature/humidity sensor of type FH XXX6-XX, measuring range temperature

(basic rate + 3 points 10°C, 25°C, 40°C)

KT9021W

KT9041W

Measuring points, freely selectable

Basic fee per certificate, temperature range up to +1300 °C

Basic fee per certificate, temperature range up to +1600 °C

Measuring point fee per sensor, one measuring point

Measuring point fee per sensor, one measuring point -196 °C

KT9001W

KT9002W

KT9011W

KT9012W

If ALMEMO

®

series temperature sensors are used, the sensor deviations that have been determined during the calibration will be stored in the sensor connector as correction values for zero point and slope (gain) before the last measurement is performed.

The correction is performed with DAkkS and factory calibrations.

Calibration certificate - temperature - sensor deviation reduced to zero

(see page 17.06)

Multi-point adjustment for ALMEMO

(preferably using Pt100 and NTC sensors)

®

measuring chains Order no.

for the measurable variable temperature, for calibration packages and for single points (at least 3 temperature points, temperature point 0 °C obligatory), additional charge per sensor for factory / DAkkS calibration (German calibration service)

KA9001DW

Calibration and adjustment of the ALMEMO

®

measuring chain are performed for the whole of the sensor’s measuring range at the points in the calibration package.

Calibration and adjustment of the ALMEMO

®

measuring chain are performed (preferably using Pt100 and NTC sensors) at at least 3 temperature points (temperature point 0 °C obligatory). Outside the calibrated range (i.e. below the lowest and above the highest calibration points) linear interpolation is performed up to the limits of the device’s measuring range (e.g. Pt100 0.01

K from -200 to +400 °C).

When ALMEMO

®

measuring equipment is calibrated the sensor deviation is determined at each and every calibration point and then saved as correction value for that calibration point in the patented ALMEMO® connector. The measured values for the sensor obtained using this multi-point adjustment function are entered in the calibration certificate. Sensor deviations are thus reduced to zero.

With thermocouples, as is generally the case, the indicated (zero-based) values in the calibration certificate are only valid if the device is in a stationary, thermally steady-state condition.

Only for device types ALMEMO

®

2450 (not -L), 2490 (not -L), 2590-2/-3S/-4S, 2690, 2890, 4390, 8590, 8690, 5690 and 5790

These device types as of serial number H0802xxxx incorporate this function as standard; for device types of serial number

H0801 and below a device firmware update is possible (noted at incoming inspection as part of the calibration service).

OA0006U

Advisory note :

On temperature sensors with special linearization or special measuring ranges saved to the ALMEMO

®

connector (e.g.

ALMEMO

®

connector ZA9040SS3 NTC 0.001K or ALMEMO

®

connectors with KTY84, YSI400, or customized NTC) multipoint adjustment is not possible..

17.07

Calibration certificates

Infrared Temperature Measurement

Calibration certificate for temperature measuring chains consisting of an IR temperature sensor

and an instrument (also individual sensors).

DAkkS-Calibration Certificate Order no.

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025 (inter alia for the monitoring of production processes and the quality assurance applied to products).

Calibration is performed by a DAkkS authorized office which compares measured values against reference values based on national standards.

For IR transmitters MR7838, MR7842, MR78434, Hand-held IR devices MR7811, MR7814, ALMEMO

®

IR sensor

FIAD43

Package offer: 3 temperature points, 25, 100, 200 °C

KI9201D

Calibration in the range –20°C to +550°C in 3 individually selectable measuring points

Calibration in the range +550°C to +1600°C in 3 individually selectable measuring points

1 additional measuring point, freely selectable, in the range -20 to +1600 °C

Factory Calibration Certificate

Calibration compares measured values against reference values based on national standards.

KI9168D

KI9178D

KI9168DP

For IR transmitters MR7838, MR7842, MR7843, Hand-held IR devices MR7811, MR7814, ALMEMO

®

IR sensors

FIA844, FIAD43

Package offer 3 temperature points, 25, 100, 200 °C

KI9201W

Calibration in the range –20°C to +550°C in 3 individually selectable measuring points

Calibration in the range +550°C to +1600°C in 3 individually selectable measuring points

1 additional measuring point, freely selectable, in the range -20 to +1600 °C

(but not between 550 and 600 °C)

KI9168W

KI9178W

KI9168WP

Relative Air Humidity for Capacitive Humidity Sensors

Calibration certificate for humidity measuring chains consisting of a capacitive humidity sensor and measuring instrument (also individual sensors).

DAkkS Calibration Certificate Order no.

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025 (inter alia for the monitoring of production processes and the quality assurance applied to products).

Calibration is performed by a DAkkS authorized office which compares measured values against reference values based on national standards.

Package offer

(Basic rate + 3 humid. points 20%/53%/75% r.H. + 1 temperature point at approx. 25 °C )

Package offer

(basic rate + 3 humidity points in the range 20% to 85% r.H.) supplement to KH9056D

(1 further humidity point in the range 20% to 85% r.H.)

KH9046D

KH9056D

KH9056DP

Order no.

Factory Calibration Certificate

Calibration compares measured values against reference values based on national standards.

Calibration is performed in a humidity generator at an ambient temperature of approx. 25° C.

Package offer

(Basic rate + 3 humidity points 11% / 53% / 75% r.H. + 1 temperature point at approx. 25°C)

Package offer

(Basic rate + 3 humidity points 11% / 53% / 75% r.H.)

Package offer

(Basic rate + 3 humidity points in the range 20% to 85% r.H.)

Supplement to KH9056W Additional humidity point in the range 20 to 85% RH

For calibration at other temperatures, see below !

17.08

KH9046 W

KH9036W

KH9056W

KH9056WP

Calibration certificates

Relative Air Humidity for Psychrometers

Calibration certificate for humidity measuring chains consisting of psychrometer and measuring instrument.

DAkkS Calibration Certificate Order no.

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025 (inter alia for the monitoring of production processes and the quality assurance applied to products).

Calibration is performed by a DAkkS authorized office which compares measured values against reference values based on national standards.

Package offer (basic rate + 2 humidity points 30% / 75% RH + 1 temperature point at approx. 25 °C)

KH9146D

Factory Calibration Certificate

Calibration compares measured values against reference values based on national standards.

Calibration is performed in a humidity generator at an ambient temperature of approx. 25 °C.

Package offer (basic rate + 2 humidity points 30% / 75% RH + 1 temperature point at approx. 25 °C)

KH9146W

Relative air humidity at temperatures up to +95 °C

DAkkS calibration certificate for temperatures up to +95° C Order no.

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025 (inter alia for the monitoring of production processes and the quality assurance applied to products).

For capacitive humidity sensors and psychrometers. Calibration is performed by a DAkkS authorized office which compares measured values against reference values based on national standards. Calibration is performed in a humidity generator / climate chamber by a DAkkS authorized office.

Measuring points, freely selectable

Basic rate

Points rate per sensor for 1 climate point

Temperature in the range +10 to +95 °C and humidity in the range 10% to 95% RH

KH9166D

KH9166DP

Factory calibration certificate for temperatures up to +95 °C

For capacitive humidity sensors and psychrometers. Calibration compares measured values against reference values based on national standards. Calibration is performed in a humidity generator / climate chamber.

Measuring points, freely selectable

Basic rate

Points rate per sensor for 1 climate point

Temperature in the range +10 to +95 °C and humidity in the range 10% to 95% RH

KH9166W

KH9166WP

Relative air humidity, temperature, and atm. pressure for FMA510

Factory Calibration Certificate

Calibration compares measured values against reference values based on national standards.

Calibration in a climate chamber at approx. 25 °C

Order no.

Package offer 1 temperature point at approx. 25 °C 1 humidity point at approx. 50 % RH, 1 absolute pressure measuring point for atmospheric pressure at time of calibration

KH9246W

Dew point

Calibration certificate - for dewpoint sensor only FHA646DTC1 / MT8716DTC1.

Factory Calibration Certificate

Factory calibration certificate

Calibration is performed based on measurement comparison at an ambient temperature of approx. 25 °C. basic rate + 1 dew point in the range -60 to +20 °C dew point

Supplement for KH9316W

1 additional dew point in the range -60 to +20 °C dew point

Order no.

KH9316W

KH9316WP

17.09

Calibration certificates

Pressure

Calibration according to DIN 16005/16086.

This calibration can be performed in 5 or 10 measuring points with pressure transducers or transducer + measuring instrument

(measuring chain): to 100bar, medium: gas to 700bar, medium: oil

DAkkS Calibration Certificate Order no.

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025 (inter alia for the monitoring of production processes and the quality assurance applied to products).

Calibration is performed by a DAkkS authorized office which compares measured values against reference values based on national standards.

Positive overpressure in the range 0 to 700 bar, 10 points

Positive and negative overpressure for pressure sensors in the range -1 to 700 bar, 10 points

Absolute pressure in the range from 0.03bar to 700bar, 10 points

KD9012D

KD9014D

KD9112D

Factory Calibration Certificate

Calibration compares measured values against reference values based on national standards.

Positive overpressure in the range 0 to 700 bar, 10 points

Positive overpressure in the range 0 to 700 bar, 5 points

Positive and negative overpressure for pressure sensors in the range -1 to 700 bar, 10 points

Absolute pressure in the range from 0.03bar to 700bar, 10 points

Absolute pressure in the range from 0.03bar to 700bar, 5 points

KD9012W

KD9013W

KD9014W

KD9113W

KD9112W

Calibration certificate - pressure - sensor deviation reduced to zero

(see page 17.06)

Multi-point adjustment for ALMEMO

®

measuring chains

For the measurable variable pressure, for calibration packages, additional charge per sensor for factory / DAkkS calibration

Order no.

KA9001DW

When ALMEMO

®

measuring equipment is calibrated the sensor deviation is determined at each and every calibration point and then saved as correction value for that calibration point in the patented ALMEMO

®

connector. The measured values for the sensor obtained using this multi-point adjustment function are entered in the calibration certificate. Sensor deviations are thus reduced to zero.

For analog pressure sensors, e.g. FDA602Lx, FD8214x.

Not for digital pressure sensors FDAD33x, FDAD35x, FDAD12SA or barometric pressure sensors integrated in the

ALMEMO

®

device or in the ALMEMO

®

D6 plug

Only for device types ALMEMO

®

2450 (not -L), 2490 (not -L), 2590-2/-3S/-4S, 2690, 2890, 4390, 8590, 8690, 5690, 5790

These device types as of serial number H0802xxxx incorporate this function as standard; for device types of serial number H0801 and below a device firmware update is possible (noted at incoming inspection as part of the calibration service).

OA0006U

Absolute pressure for digital atmospheric pressure sensor FDAD12SA

Calibration certificate for barometric pressure sensors integrated in the ALMEMO

®

device or in the ALMEMO

®

D6 plug

DAkkS Calibration Certificate Order no.

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025 (inter alia for the monitoring of production processes and the quality assurance applied to products).

Calibration is performed by a DAkkS authorized office which compares measured values against reference values based on national standards.

Absolute pressure 5 points in the range 700 to 1100 mbar

Absolute pressure 10 points in the range 700 to 1100 mbar

KD9213D

KD9214D

Factory Calibration Certificate

Calibration compares measured values against reference values based on national standards.

Absolute pressure 5 points in the range 700 to 1100 mbar

Absolute pressure 10 points in the range 700 to 1100 mbar

KD9213W

KD9214W

17.10

Air Flow

Calibration certificate for rotating vanes, Pitot tubes and thermoanemometers.

Calibration certificates

DAkkS Calibration Certificate Order no.

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025 (inter alia for the monitoring of production processes and the quality assurance applied to products).

Calibration is performed in a wind tunnel based on measurement comparison against a laser Doppler anemometer by a DAkkS authorized office.

Package offer (basic rate + 5 points 0,5 m/s / 1 m/s / 2 m/s / 5 m/s / 10 m/s )

Package offer (basic rate + 5 points 2 m/s / 5 m/s / 10 m/s / 15 m/s / 20 m/s )

Package offer (basic rate + 5 points 0,2 m/s / 0,3 m/s / 0,5 m/s / 0,8 m/s / 1 m/s )

KV9025D

KV9035D

KV9065D

Factory Calibration Certificate

The calibration can be performed with the sensor and the meas. instrument (meas. chain). Calibration in a wind tunnel.

Reference standards: Wind tunnel and reference rotating vanes (calibrated acc. to the laser-Doppler method).

Package offer (basic rate + 3 points 0.5m/s / 5m/s / 10m/s)

Package offer (basic rate + 3 points 5m/s / 10m/s / 19m/s) FVA645TH3: 15m/s

Package offer (basic rate + 3 points 7m/s / 20m/s / 30m/s)

Package offer (basic rate + 3 points 0.5m/s / 1m/s / 1,75m/s)

Package offer (basic rate + 3 points 0.5m/s / 0.8m/s / 1m/s)

Measuring points, freely selectable

Basic rate

Per measuring point and sensor Meas. range 0.5m/s to 40m/s.

KV9025W

KV9035W

KV9045W

KV9055W

KV9065W

KV9005W

KV9015W

Flow measurement in liquids

Calibration certificate for turbine flow meters or flow sensors

Factory Calibration Certificate

Calibration compares measured values against reference values based on national standards.

Calibration of the volume flow rate in l/min (up to maximum 200 l/min) in the test chamber

Measuring medium: Water

Calibration at 3 measuring points 1 point each at start / middle / end of sensor range

Package offer

Supplement to KV9115W 1 additional measuring point in the sensor’s measuring range

Order no.

KV9115W

KV9115WP

17.11

Calibration certificates

Conductivity

Calibration certificate for conductivity measuring chains.

Factory Calibration Certificate Order no.

Calibration in KCl reference solutions. Reference standard: Precision conductivity measurement instrument, which has been calibrated with KCl solutions of the National Institute of Standards and Technology (NIST).

Package offer for conductivity probe FYA641LF /LFP1

(Basic rate + 3 points 0.5mS / 2.77mS / 10mS)

(Basic rate + 2 points 2.77mS / 12.8mS)

Package offer for conductivity probe FYA641LF2 /LFP2

(Basic rate + 3 points 5µS / 147µS / 190µS)

KY9041W

KY9044W

KY9042W

Package offer for conductivity probe FYA641LF3 /LFP3

(Basic rate + 3 points 5mS / 50mS / 111,8mS)

KY9043W

Gas Concentration

Calibration certificate for CO

2

Factory Calibration Certificate Order no.

Calibration is performed based on measurement comparison against a reference gas specified by the manufacturer.

Package offer for CO

2

probe FYA600CO2

(approx. 10 measuring points)

Package offer for CO

2

probe FYAD00CO2B10 (3 measuring points at approx. 1000 / 4000 / 7000 ppm)

KY9620W

KY9626W

Measurable Variables for Optical Radiation

Calibration certificate for broad-band light detectors

Factory Calibration Certificate

single point calibration of absolute size

(not for probes FLA613GS / UVA / UVB / VLM / VLK /UVAK, FLA623x)

Calibration of absolute variable in 3 points

(only for probes FLA613GS / UVA / UVB / VLM / VLK / UVAK, FLA623x)

KL9033W

KL9034W

Optical Speed Sensors

Calibration certificate for contactless tachometers.

DAkkS Calibration Certificate Order no.

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025 (inter alia for the monitoring of production processes and the quality assurance applied to products).

Calibration is performed by a DAkkS authorized office which compares measured values against reference values based on national standards.

Calibration of the optical transducer at 8 measuring points

(not applicable for the tachometer probe FUA919-MF)

KU9029D

Factory Calibration Certificate

Calibration compares measured values against reference values based on national standards.

Calibration of the optical transducer at 8 measuring points

(not applicable for the tachometer probe FUA919-MF)

KU9029W

17.12

Calibration certificates

Force

Calibration for tension and compression sensors

Factory calibration certificate

Calibration is performed based on the measurement comparison method for Ahlborn force transducers;

(other makes available on request).

4 series of measuring operations upwards and 2 series downwards

3 steps (0%, 20%, 60%, 100% of final value) Tension or compression (indicate direction), up to 1 kN

3 steps (0%, 20%, 60%, 100% of final value) Tension or compression (indicate direction), up to 10 kN

3 steps (0%, 20%, 60%, 100% of final value) Tension or compression (indicate direction), up to 100 kN

3 steps (0%, 20%, 60%, 100% of final value) Tension or compression (indicate direction), up to 1000 kN

Order no.

KK9021W

KK9031W

KK9041W

KK9051W

Electrical Calibration of Measuring and Indicating Devices

Calibration certificate for all devices of the THERM and ALMEMO

®

series.

DAkkS Calibration Certificate Order no.

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025 (inter alia for the monitoring of production processes and the quality assurance applied to products).

Calibration is performed by a DAkkS authorized office which compares measured values against reference values based on national standards.

The calibration is performed at approx. 10 to 20 measuring points.

Calibration for one measuring range

Each further measuring range

KE9010D

KE9020D

Calibration of a measuring chain using ALMEMO

®

adapter cable ZA9603AKx, AC voltage or ALMEMO

®

measuring module ZA990xABx, AC / DC voltage, up to 400 V (50 Hz), or AC / DC current, up to 10 A (50 Hz) Package offer, approx. 10 points

Factory Calibration Certificate

Calibration compares measured values against reference values based on national standards.

The calibration is performed at approx. 10 to 20 measuring points.

KE9030D

Calibration for one measuring range

Each further measuring range

Calibration of a measuring chain using ALMEMO

®

adapter cable ZA9603AKx, AC voltage or ALMEMO

®

measuring module ZA990xABx, AC / DC voltage, up to 400 V (50 Hz), or AC / DC current, up to 10 A (50 Hz) Package offer, approx. 10 points

KE9010W

KE9020W

KE9030W

Calibration certificate for all ALMEMO

®

measuring instruments with interface

DAkkS calibration certificate Order no.

DAkkS calibration meets all the requirements regarding test resources laid down in DIN EN ISO/IEC 17025 (inter alia for the monitoring of production processes and the quality assurance applied to products).

Calibration is performed by a DAkkS authorized office which compares measured values against reference values based on national standards.

Full calibration of ALMEMO device in 9 measuring ranges

2.6 V (volt), 55 mV (mV), 26 mV (mV1), 260 mV (mV2), NiCr-Ni (NiCr), Pt100 0.1 K (P104), Pt100 0.01 K (P204), NTC type N (NTC), relative humidity, capacitive (% RH)

Package offer

KE9005D

Factory calibration certificate

Calibration compares measured values against reference values based on national standards.

Full calibration of ALMEMO device in 9 measuring ranges

2.6 V (volt), 55 mV (mV), 26 mV (mV1), 260 mV (mV2), NiCr-Ni (NiCr), Pt100 0.1 K (P104), Pt100 0.01 K (P204), NTC type N (NTC), relative humidity, capacitive (% RH)

Package offer

KE9005W

17.13

17.14

advertisement

Key Features

  • 2-row segment display
  • easy to use interface
  • over 35 measuring ranges
  • multi-point adjustment
  • saving of maximum / minimum values
  • sensor breakage indication
  • battery voltage check and display

Frequently Answers and Questions

What is the measuring rate of the ALMEMO 2450-1?
The measuring rate of the ALMEMO 2450-1 is 2.5 mops (measuring operations per second).
What is the operating temperature range of the ALMEMO 2450-1?
The operating temperature range is -10 to +60 °C.
What type of sensors can I use with the ALMEMO 2450-1?
The ALMEMO 2450-1 supports a variety of sensors, including thermocouples, NTC sensors, humidity sensors, pressure transducers, and more.